mx400 PDF
mx400 PDF
mx400 PDF
Patient Monitori ng
1 Table of Contents
1 Introduction 13
Safety Information 14
Security Information 15
Introducing the Monitor 16
Devices for Acquiring Measurements 18
Operating and Navigating 27
Operating Modes 36
Understanding Screens 37
Connecting Additional Displays to the Monitor 38
Using the XDS Remote Display 39
Using the Visitor Screen 39
Understanding Profiles 39
Understanding Settings 41
Changing Wave Speeds 42
Freezing Waves 42
Using Labels 44
Entering Measurements Manually 46
Changing Monitor Settings 47
Checking Your Monitor Revision 48
Getting Started 48
Disconnecting from Power 49
Networked Monitoring 50
Using the Integrated PC 50
Using Your Monitor with a Monitor in Companion Mode 52
2 What's New? 55
3 Alarms 59
3
Alarm Recordings 73
4
Updating ST Baseline Snippets 168
Recording ST Segments 168
About the ST Measurement Points 168
ST Alarms 171
STE Alarms 171
Viewing ST Maps 172
About QT/QTc Interval Monitoring 175
QT Alarms 178
Switching QT Monitoring On and Off 179
5
Starting and Stopping Measurements 207
Enabling Automatic Mode and Setting Repetition Time 208
Enabling Sequence Mode and Setting Up The Sequence 208
Choosing the NBP Alarm Source 209
Switching Pulse from NBP On/Off 209
Assisting Venous Puncture 210
Calibrating NBP 210
6
Measuring CO2 using M3014A or X2 242
Measuring Mainstream CO2 using M3016A 246
Measuring Microstream CO2 using M3015A/B 248
Setting up all CO2 Measurements 250
Understanding the IPI Numeric 252
7
Changing the BIS Smoothing Rate 290
Switching BIS and Individual Numerics On and Off 290
Changing the Scale of the EEG Wave 290
Switching BIS Filters On or Off 290
BIS Safety Information 291
23 Trends 311
24 Calculations 325
8
26 Event Surveillance 333
27 ProtocolWatch 349
28 Recording 357
9
Language Conflict with External Device Drivers 390
10
38 Maintenance and Troubleshooting 415
39 Accessories 419
40 Specifications 443
11
M4605A Battery Specifications 475
Measurement Specifications 475
Safety and Performance Tests 495
Index 523
12
1
1 Introduction
These Instructions for Use are for clinical professionals using the IntelliVue MX400/MX450, MX500/
MX550, and MX600/MX700/MX800 patient monitor.
This basic operation section gives you an overview of the monitor and its functions. It tells you how to
perform tasks that are common to all measurements (such as entering data, switching a measurement
on and off, setting up and adjusting wave speeds, working with profiles). The alarms section gives an
overview of alarms. The remaining sections tell you how to perform individual measurements, and
how to care for and maintain the equipment.
Familiarize yourself with all instructions including warnings and cautions before starting to monitor
patients. Read and keep the Instructions for Use that come with any accessories, as these contain
important information about care and cleaning that is not repeated here.
This guide describes all features and options. Your monitor may not have all of them; they are not all
available in all geographies. Your monitor is highly configurable. What you see on the screen, how the
menus appear and so forth, depends on the way it has been tailored for your hospital and may not be
exactly as shown here.
In this guide:
• A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to
observe a warning may result in death or serious injury to the user or patient.
• A caution alerts you to where special care is necessary for the safe and effective use of the
product. Failure to observe a caution may result in minor or moderate personal injury or damage
to the product or other property, and possibly in a remote risk of more serious injury.
MX400/ Whenever a monitor's identifier appears to the left of a heading or paragraph, it means that the
MX450 information applies to that monitor only. Where the information applies to all models, no distinction is
made.
For installation, repair, testing and troubleshooting instructions, refer to the Service Guide for your
monitor model.
Rx only: U.S. Federal Law restricts this device to sale by or on the order of a physician.
13
1 Introduction
Safety Information
The following warnings apply to the monitors in general. Warnings that apply to specific
measurements or procedures can be found in the corresponding chapters.
WARNING
Grounding: To avoid the risk of electric shock, the monitor must be grounded during operation. If a
three-wire receptacle is not available, consult the hospital electrician. Never use a three-wire to two-
wire adapter.
Electrical shock hazard: Do not open the monitor or measurement device. Contact with exposed
electrical components may cause electrical shock. Refer servicing to qualified service personnel.
Leakage currents: If multiple instruments are connected to a patient, the sum of the leakage currents
may exceed the limits given in IEC/EN 60601-1, IEC 60601-1-1, UL 60601-1. Consult your service
personnel.
Radio frequency interference: The monitor generates, uses and radiates radio-frequency energy, and
if it is not installed and used in accordance with its accompanying documentation, may cause
interference to radio communications.
Use Environment
WARNING
Explosion Hazard: Do not use in the presence of flammable anesthetics or gases, such as a
flammable anesthetic mixture with air, oxygen or nitrous oxide. Use of the devices in such an
environment may present an explosion hazard.
Positioning Equipment: The monitor should not be used next to or stacked with other equipment.
If you must stack the monitor, check that normal operation is possible in the necessary configuration
before you start monitoring patients.
Environmental Specifications: The performance specifications for the monitors, measurements and
accessories apply only for use within the temperature, humidity and altitude ranges specified in
“Environmental Specifications” on page 458.
Liquid Ingress: If you spill liquid on the equipment, battery, or accessories, or they are accidentally
immersed in liquid, contact your service personnel or Philips service engineer. Do not operate the
equipment before it has been tested and approved for further use.
Prohibited Environments: The monitors are not intended for use in an MRI environment or in an
oxygen-enriched environment (for example, hyperbaric chambers).
14
1 Introduction
Alarms
WARNING
• Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of alarm
volume to a low level or off during patient monitoring may result in patient danger. Remember
that the most reliable method of patient monitoring combines close personal surveillance with
correct operation of monitoring equipment.
• Be aware that the monitors in your care area may each have different alarm settings, to suit
different patients. Always check that the alarm settings are appropriate for your patient before you
start monitoring.
Accessories
WARNING
Philips' approval: Use only Philips-approved accessories. Using other accessories may compromise
device functionality and system performance and cause a potential hazard.
Reuse: Never reuse disposable transducers, sensors, accessories and so forth that are intended for
single use, or single patient use only. Reuse may compromise device functionality and system
performance and cause a potential hazard.
Electromagnetic compatibility: Using accessories other than those specified may result in increased
electromagnetic emission or decreased electromagnetic immunity of the monitoring equipment.
Damage: Do not use a damaged sensor or one with exposed electrical contacts.
Cables and tubing: Always position cables and tubing carefully to avoid entanglement or potential
strangulation.
MR Imaging: During MR imaging, remove all transducers, sensors and cables from the patient.
Induced currents could cause burns.
Security Information
Protecting Personal Information
Protecting personal health information is a primary component of a security strategy. Each facility
using the monitors must provide the protective means necessary to safeguard personal information
consistent with country laws and regulations, and consistent with the facility’s policies for managing
this information. Protection can only be realized if you implement a comprehensive, multi-layered
strategy (including policies, processes, and technologies) to protect information and systems from
external and internal threats.
As per its intended use, the patient monitor operates in the patient vicinity and contains personal and
sensitive patient data. It also includes controls to allow you to adapt the monitor to the patient's care
model. To ensure the patient's safety and protect their personal health information you need a security
concept that includes:
15
1 Introduction
• Physical security access measures - access to the monitor must be limited to authorized users.
It is essential that you consider physical security measures to ensure that unauthorized users
cannot gain access.
• Operational security measures - for example, ensuring that patients are discharged after
monitoring in order to remove their data from the monitor.
• Procedural security measures - for example, assigning only staff with a specific role the right to
use the monitors.
In addition, any security concept must consider the requirements of local country laws and regulations.
Always consider data security aspects of the network topology and configuration when connecting
patient monitors to shared networks. Your medical facility is responsible for the security of the
network, where sensitive patient data from the monitor may be transferred.
Note: Log files generated by the monitors and measurement modules are used for system
troubleshooting and do not contain patient data.
16
1 Introduction
and the MX400/MX450, MX500/MX550, and MX700/MX800 use the touch screen as primary input
device. All monitors have a remote control for convenient access to the five main keys and numeric
data input.
The monitor stores data in trend, event, and calculation databases. You can see tabular trends (vital
signs) and document them on a printer. You can view measurement trend graphs, with up to three
measurements combined in each graph, to help you identify changes in the patient's physiological
condition. You can view fast-changing measurement trends with beat to beat resolution and see up to
four high resolution trend segments. Event surveillance enhances documentation and review of
physiologically significant events by automatically detecting and storing up to 50 user-defined clinical
events over a 24 hour period.
MX600/700/ With the optional Integrated PC, you have computer functionality directly in the monitor. You can use
800 standard applications (e.g. Web browsers), connect to the hospital network or intranet, and run a
second independent display with content from the patient monitor.
An IntelliVue X2 can be connected to your monitor, where it acts as a multi-measurement module,
acquiring measurements for the host monitor. When the X2 is disconnected from the original host
monitor, it continues to monitor the patient as a fully independent, battery powered patient monitor,
eliminating the need for a separate transport monitor. On connection to a new host monitor, the X2
resumes its role as multi-measurement module, ensuring fully continuous monitoring.
17
1 Introduction
MX600/700
MX800
18
1 Introduction
monitor, and switch on automatically when you turn on the monitor. A green power-on LED indicates
when they are drawing power from the monitor. A permanently illuminated, or flashing, red LED
indicates a problem with the unit that requires the attention of qualified service personnel.
All symbols used on the front panels are explained in “Symbols” on page 445.
WARNING
When connecting devices for acquiring measurements, always position cables and tubing carefully to
avoid entanglement or potential strangulation.
1 X1 Multi-Measurement Module
2 Multi-Measurement Module
mount
3 Flexible Module Rack FMS-8
4 Power on LED
5 Interruption indicator
19
1 Introduction
The maximum number of specific module types that can be used simultaneously in an FMS-4 is: four
pressure modules, four temperature modules, four VueLink or IntelliBridge modules (any
combination).
Connect the FMS to the monitor via the measurement link cable (MSL). Use the MSL connector on
the left-hand side (if you have the appropriate option) to connect an additional MMS. Use the
connector on the back to connect to the monitor.
Measurement Modules
MX500/ You can use up to three plug-in modules in the optional module slots. Available modules are:
MX550 • Invasive blood pressure (M1006B)
• Temperature (M1029A)
• Oxygen saturation of arterial blood (SpO2) (M1020B)
• Cardiac output (M1012A), and Continuous cardiac output with M1012A Option #C10
• Intravascular Oxygen Saturation - ScvO2 or SvO2 (M1011A)
• Spirometry (M1014A)
• EEG (M1027A)
• NMT (865383)
• IntelliBridge EC10 (865115)
• Recorder (M1116B/C)
MX600/ You can use up to eight measurement modules with the Flexible Module Rack (M8048A). Available
MX700/ modules are:
MX800 • Invasive blood pressure (M1006B)
• Temperature (M1029A)
• Oxygen saturation of arterial blood (SpO2) (M1020B)
• Cardiac output (M1012A), and Continuous cardiac output with M1012A Option #C10
• Transcutaneous gas (M1018A)
• Mixed venous oxygen saturation - SvO2 (M1021A)
• Intravascular Oxygen Saturation - ScvO2 or SvO2 (M1011A)
• EEG (M1027A)
• Bispectral Index - BIS (M1034A)
• Spirometry (M1014A)
• NMT (865383)
• VueLink device interface (M1032A)
• IntelliBridge EC10 (865115)
• Recorder (M1116B/C)
MX500/550/ You can plug in and unplug modules during monitoring. Insert the module until the lever on the
600/700/800 module clicks into place. Remove a module by pressing the lever upwards and pulling the module out.
A measurement automatically switches on when you plug the module in, and switches off when you
20
1 Introduction
unplug it. Reconnecting a module to the same monitor restores its label and measurement settings,
such as alarms limits. If you connect it to a different monitor, the module remembers only its label.
The connector socket on the front of each module is the same color as the corresponding connector
plug on the transducer or patient cable.
Press the Setup key on the module's front to display the measurement's setup menu on the monitor
screen. When the setup menu is open, a light appears above the key. Some modules have a second key.
On the pressure module, for example, it initiates a zeroing procedure.
1 Module name
2 Setup key LED
3 Setup key to enter setup menu of measurement modules or
external device data window. Some modules have a second
module-specific key next to this one, for example Zero.
4 Connector socket for patient cable/transducer
21
1 Introduction
22
1 Introduction
X2 Overview
1 On/Standby switch
2 Power and battery indicators (see “X2
Controls and Indicators” on page 23)
3 3.5-inch TFT LCD touchscreen QVGA
display
4 Alarm lamps (see “X2 Controls and
Indicators” on page 23)
5 Battery eject button
6 Hard keys (see “X2 Controls and
Indicators” on page 23)
7 Measurement connectors (see “X2 Patient
Connectors, Right Side” on page 24)
8 Battery compartment
1 External power LED. Green when monitor is powered from an external power source.
2 Battery status LED. Yellow when charging. Flashing red when battery is empty.
3 On/Standby LED. Green when monitor is on. Red indicates an error.
4 On/Standby switch. Disabled when X2 is connected to a host monitor
5 Main Screen key: closes all open menus/windows and returns to the main screen.
6 SmartKeys key: brings up SmartKeys on the screen.
7 Alarms key: turns alarms On/Off, or pauses them.
23
1 Introduction
8 Silence key
9 Active alarm lamp. Red or yellow, depending on alarm level. Blinks until active alarm is
acknowledged.
10 Active INOP alarm lamp in light blue. Blinks until active INOP is acknowledged.
11 Alarms off indicator. When alarms are suspended, the lamp is red (or yellow when yellow alarms
are suspended), and the alarms off symbol is shown.
1 Pressure (option)
2 Temperature (option)
3 Noninvasive blood pressure
4 SpO2
5 ECG sync pulse output
6 ECG/Respiration
7 CO2 (option in place of
Pressure and Temperature)
X2 Left Side
1 Loudspeaker
2 MSL Connector. Connects to the external power
supply or a host monitor via the MSL cable for AC
mains operation, battery charging, and
communication with a network.
24
1 Introduction
MMS Extensions
The MMS extensions connect to the X1 and X2 MMS and use the MMS settings and power. Trend
data and measurement settings from the measurements in the extensions are stored in the MMS.
WARNING
• The MMS extensions can only function when they are connected to an MMS. If the MMS is
removed during monitoring, the measurements from both the MMS and the extension are lost.
• Measurements from an MMS extension connected to an X2 are not available when the X2 is
running on its own battery power. They are only available when the X2 is powered from AC mains,
when connected to a host monitor or the external power supply (M8023A), or from the Battery
Extension.
To separate an extension from the MMS, press the release lever down, and push the MMS forward.
M3014A
25
1 Introduction
The optional M3015A Microstream CO2 extension adds microstream capnography and optionally
either pressure or temperature to the MMS. The optional M3015B Microstream CO2 extension adds
microstream capnography, two pressures and a temperature to the MMS.
M3015A M3015B
MX600/700/ The optional M3016A Mainstream CO2 extension adds mainstream capnography and optionally either
800 pressure or temperature to the MMS.
M3016A
When a capnography extension is connected to an X2 MMS with CO2, the CO2 from the extension
will be automatically deactivated in favor of the one in the X2. If you prefer to use the CO2
measurement on the extension, you can activate it via the measurement selection key (see “Resolving
Label Conflicts” on page 44).
26
1 Introduction
The cardiac output measurement in the M3014A is deactivated when the extension is used with an X2
MMS, even if the X2 is connected to an external power supply. The cardiac output measurement is
only available when the X2 is connected to a host monitor.
The M3012A Hemodynamic extension can be connected to the M3001A Multi-Measurement Module
to provide the following additional measurements: Temperature, Pressure, an additional Pressure or
Temperature, and C.O. and CCO measurements.
The cardiac output measurement is deactivated when the extension is used with an X2 MMS unless the
X2 is connected to a host monitor.
27
1 Introduction
MX400
On the MX400, the permanent keys and the key to access the SmartKeys are on the right of the screen.
28
1 Introduction
MX450/500/550/600/700/800
29
1 Introduction
Press and hold the Main Screen permanent key again to re-enable the touchscreen operation.
Moving Windows
You can move windows and menus using the Touchscreen or a mouse. To move a window,
1 Select the title of the window and keep your finger on the title, or the mouse button pressed.
2 Move your finger on the Touchscreen, or move the mouse, to move the window.
3 Take your finger off the screen, or release the mouse button, to place the window in the final
position.
The new position is only active until the window or menu is closed. Not all locations on the screen can
be a target position, a window cannot overlap the monitor info line, the alarms and INOPs or the
status line.
Using Keys
The monitor has four different types of keys:
Permanent Keys
A permanent key is a graphical key that remains on the screen all the time to give you fast access to
functions.
30
1 Introduction
SmartKeys
A SmartKey is a configurable graphical key, normally located at the bottom of the main screen. On
some models, a selection of SmartKeys is displayed by selecting the SmartKeys permanent key on the
right of the main screen.
SmartKeys give you fast access to functions. The selection of SmartKeys available on your monitor
depends on your monitor configuration and on the options purchased. If you have an integrated PC
(iPC) you may also see SmartKeys generated by applications on the iPC.
31
1 Introduction
start veni puncture (inflate cuff to set the NBP repeat time
subdiastolic pressure)
set wide automatic alarm limits set narrow automatic alarm limits
gas analyzer - exit standby mode suppress zero for all gas
measurements
32
1 Introduction
Hardkeys
A hardkey is a physical key on a monitoring device, such as the zero pressure key on the MMS or a
setup key on a module.
Pop-Up Keys
Pop-up keys are task-related graphical keys that appear automatically on the monitor screen when
required. For example, the Confirm pop-up key appears only when you need to confirm a change.
33
1 Introduction
The remote control provides you with direct access to five hard keys, a navigation knob and a numeric
keypad:
Hardkeys
1 Silence - acknowledges all active alarms by switching off audible alarm indicators and lamps.
Behavior follows the Silence permanent key configuration.
2 Alarms Off/Pause Alarms - pauses alarm indicators. Behavior follows the Pause Alarms
permanent key configuration.
3 Main Screen - close all open menus and windows and return to the main screen.
4 SmartKeys - display a block of SmartKeys specially configured for remote tasks (see below)
5 Back - go back one step to the previous menu.
6 Enter - mark the end of data entry
Keypad
7 Type numeric data on the keypad and press the Enter key to enter the data on the monitor.
Navigation knob
8 Rotate the knob to highlight screen elements, then press to select the highlighted element.
The remote control can be used with a USB cable connection to the monitor or without a cable using
short range radio. When used without a cable, the remote control must be assigned to the monitor.
The assignment is made in Configuration or Service mode.
34
1 Introduction
CAUTION
When using a remote control without a cable, it is important that the user knows which remote control
is assigned to which monitor. Use the tethering cable delivered with the remote control to attach it to a
bed rail or IV pole, or label the remote control with the bed or monitor ID.
Pressing the 1 key on the remote control selects the top left SmartKey, pressing the 8 key selects the
bottom center SmartKey. The . and the key can be used to select the arrow keys to page up and
down in the available SmartKeys.
The SmartKeys which appear can be configured so that you have the functions available which you
most often need when using the remote control. If no list of SmartKeys has been configured, the
standard SmartKeys will be displayed and you can page through to the key you want.
35
1 Introduction
Operating Modes
When you switch the monitor on, it starts up in monitoring mode. To change to a different mode:
1 Select the Main Setup menu.
2 Select Operating Modes and choose the mode you require.
Your monitor has four operating modes. Some are passcode protected.
• Monitoring Mode: This is the normal, every day working mode that you use for monitoring
patients. You can change elements such as alarm limits, patient category and so forth. When you
discharge the patient, these elements return to their default values. Changes can be stored
permanently only in Configuration Mode. You may see items, such as some menu options or the
altitude setting, that are visible but ‘grayed out’ so that you can neither select nor change them.
These are for your information and can be changed only in Configuration Mode.
• Demonstration Mode: Passcode protected, this is for demonstration purposes only. You must
not change into Demonstration Mode during monitoring. In Demonstration Mode, all stored
trend information is deleted from the monitor’s memory.
• Configuration Mode: Passcode protected, this mode is for personnel trained in configuration
tasks. These tasks are described in the Configuration Guide. During installation the monitor is
configured for use in your environment. This configuration defines the default settings you work
with when you switch on, the number of waves you see and so forth.
• Service Mode: Passcode protected, this is for trained service personnel.
When the monitor is in Demonstration Mode, Configuration Mode, or Service Mode, this is indicated
by a box with the mode name in the center of the Screen and a symbol in the bottom right-hand
corner. Select the mode box in the center of the screen to change to a different mode.
When an X2 is connected to a host monitor (Companion Mode is indicated):
• The monitor in companion mode will adopt the operating mode of the host monitor.
• You cannot change the operating mode at the monitor in companion mode.
Standby Mode
Standby mode can be used when you want to temporarily interrupt monitoring.
To enter Standby mode,
• select the Monitor Standby SmartKey or
• select Main Setup, followed by Monitor Standby.
The monitor enters Standby mode automatically after the End Case function is used to discharge a
patient.
Standby suspends patient monitoring. All waves and numerics disappear from the display but all
settings and patient data information are retained. A special Standby screen is displayed. This can be
configured to a moving image or a blank screen, or to your own custom image. If a temporary patient
location has been entered at the monitor or at the Information Center, this location will also be
displayed on the Standby screen.
To resume monitoring,
• Select anything on the screen or press any key.
When monitoring is resumed, alarms are paused for 1 minute to allow time to finish plugging in the
measurement cables.
36
1 Introduction
If you connect an X2 that is powered on (and not in Standby) to a host monitor in Standby mode, the
host will leave Standby mode. When connected to a host monitor, with both the host and the monitor
in companion mode in Standby mode, leaving Standby on the monitor in companion mode will also
make the host leave Standby.
Understanding Screens
Your monitor comes with a set of pre-configured Screens, optimized for common monitoring
scenarios such as "OR adult", or "ICU neonatal". A Screen defines the overall selection, size and
position of waves, numerics and other elements on the monitor screen when you switch on. You can
easily switch between different Screens during monitoring. Screens do NOT affect alarm settings,
patient category and so forth.
When you switch from a complex to a less complex Screen layout, some measurements may not be
visible but are still monitored in the background. If you switch to a more complex Screen with, for
example, four invasive pressure waves but you have only two pressures connected to the monitor, the
"missing" two pressures are either left blank or the available space is filled by another measurement.
37
1 Introduction
After a patient discharge, the monitor's default Screen is shown. Modified Screens are still available in
the Change Screen menu.
If the monitor is switched off and then on again, modified Screens are erased from the monitor's
memory and cannot be recalled. If a modified Screen was the last active Screen when the monitor was
switched off, it is retained (unless Automat. Default is configured to Yes).
Independent Displays
MX600/700/ If you have the optional independent display interface, you can connect a second display which can be
800 configured and operated individually using standard input devices.
For monitors with multiple displays and multiple input devices, the usage and behavior can be
configured according to specific requirements at installation (for example, use for two independent
operators or tracking of mouse input across two displays). For details refer to the Service Guide.
When two operators are using two displays, the scope of an action depends on the type of operation:
• Patient monitoring operations such as Silence or Pause alarms take effect for the monitor as a
whole, the results will be seen on both displays.
• Display operations such as the Main Screen key and Back hardkey will take effect only on the
display being operated.
If you are operating two displays with one remote control, to navigate from one display to another:
1 Move the highlight to the Main Screen key and then turn one click further.
The highlighting moves to a special "jump" field at the edge of the Screen.
2 Press the navigation knob on the remote control to confirm; the highlighting will automatically
move to the other display.
The content of each Screen can be changed individually as described in the previous section. If you are
operating two displays, you can choose Screens for both displays from one location:
1 Select Profiles in the monitor info line of the first display,
2 Select Display 1, or Display 2, then select the Screen you want to appear on that display from the
list of available Screens.
When two displays are mounted next to each other or one above the other, a special Screen can be
assigned which spans across both displays. The Screen content for these Tall and Wide Screens can
then use the increased area available with two displays. These Screens appear in the Screen list with a
special Tall Screen or Wide Screen symbol.
Certain windows (for example: cardiac output procedure) can only be shown on one display at a time.
If you try to open one of these windows when it is already shown on another display, you will see a
blank gray window with a cross through it.
38
1 Introduction
Understanding Profiles
Profiles are predefined monitor configurations. They let you change the configuration of the whole
monitor so you can adapt it to different monitoring situations. The changes that occur when you
change a complete profile are more far reaching than those made when you change a Screen. Screens
affect only what is shown on the display. Profiles affect all monitor and measurement settings.
The settings that are defined by Profiles are grouped into three categories. Each category offers a
choice of 'settings blocks' customized for specific monitoring situations. These categories are:
• Display (screens)
Each profile can have a choice of many different predefined screens.
If you are using a second display, each display can have its own individual screen selection. When
you change the profile, the screen selection configured for the new profile becomes active.
• Measurement Settings
Each profile can have a choice of different predefined measurement settings. These relate directly
to individual measurements, for example, measurement on/off, measurement color, alarms limits,
NBP alarm source, NBP repeat time, temperature unit (°F or °C), pressure unit (mmHg or kPa).
• Monitor Settings
Each profile can have a choice of different predefined monitor settings. These relate to the
monitor as a whole; for example, display brightness, alarms off/paused, alarm volume, QRS tone
volume, tone modulation, prompt tone volume, wave speed, resp wave speed, pulse source.
39
1 Introduction
You can change from one complete profile to another or swap individual settings blocks (display/
monitor settings/measurement settings) to change a subset of a profile. Changes you make to any
element within the settings blocks are not saved when you discharge the patient, unless you save them
in Configuration Mode.
Depending on your monitor configuration, when you switch on or discharge a patient the monitor
either continues with the previous profile, or resets to the default profile configured for that monitor.
WARNING
• If you switch to a different profile, the patient category and paced status normally change to the
setting specified in the new profile. However some profiles may be set up to leave the patient
category and paced status unchanged. Always check the patient category, paced status, and all
alarms and settings, when you change profiles.
• If your monitor is configured to show the profile name in the info line at the top of the screen, be
aware that individual settings may have been changed by other users or by settings synchronization
since the profile was loaded. Hence settings may be different than implied by the profile name.
When you leave Demonstration Mode, the monitor uses the default profile.
Default Profile
Your monitor has a default profile that it uses when you leave Demonstration mode, or when you
discharge a patient. This profile is indicated by a diamond shaped symbol.
Locked Profiles
Some profiles are locked, so that you cannot change them, even in Configuration Mode.
These are indicated by this lock symbol.
40
1 Introduction
Understanding Settings
Each aspect of how the monitor works and looks is defined by a setting. There are a number of
different categories of settings, including,
Screen Settings, to define the selection and appearance of elements on each individual Screen
Measurement settings, to define settings unique to each measurement, for example, high and low
alarm limits
Monitor settings, including settings that affect more than one measurement or Screen and define
general aspects of how the monitor works, for example, alarm volume, reports and recordings, and
display brightness.
You must be aware that, although many settings can be changed in Monitoring Mode, permanent
changes to settings can only be done in the monitor's Configuration Mode. All settings are reset to the
stored defaults:
• when you discharge a patient
• when you load a Profile
• when the monitor is switched off for more than one minute (if Automat. Default is set to Yes).
41
1 Introduction
Freezing Waves
You can freeze waves on the screen and measure parts of the wave using cursors. The waves are frozen
with a history of 20 seconds so that you can go back and measure what you have seen.
42
1 Introduction
43
1 Introduction
Using Labels
Every measurement associated with a monitor is identified by a unique label. You may have more than
one instance of some measurements, for example pressure, being used simultaneously. The monitor
uses the labels to distinguish between them. The default settings defined in the profile (such as
measurement color, wave scale, and alarm settings) are stored within each label. When you assign a
label to a measurement, the monitor automatically applies these default settings to the measurement.
The labels assigned are used throughout the monitor, in reports, recordings, and in trends.
44
1 Introduction
Any measurement labels causing a label conflict are shown in red. If a measurement device is
connected but currently unavailable, for example, because it was deactivated due to a label conflict, the
device is shown "grayed-out".
When an X2 is connected to a host monitor, the measurement selection window looks like this:
WARNING
When an X2 with an active measurement, say SpO2, is connected to a host monitor with the same
measurement already active, the SpO2 measurement on the X2 is deactivated and the Meas.
Deactivated INOP is displayed. The measurement can only be reactivated if the X2 is disconnected
from the host monitor. The label conflict can be resolved on the host monitor like any other label
conflict.
45
1 Introduction
Label Compatibility
When a new measurement is introduced, or new labels for an existing measurement, these labels will
not be shown on older Information Centers, and consequently not on the Overview screen sourced
from the Information Center.
When a patient is transferred from a monitor with these new labels to one with an older software
revision, the labels will be replaced with a generic label for that measurement. The settings for that
generic label will then be used.
If it is critical that the measurement labels are available at the Information Center and after transfers,
the older monitors and the Information Center must be upgraded to the appropriate software revision.
46
1 Introduction
Manually entered measurement values are marked with a * on the display, in trends, and so forth. Each
measurement has a defined measurement interval after which a value becomes invalid (no value is then
displayed). Values can be entered up to two hours after they have been measured or up to the
measurement interval, if this is shorter.
The list of measurement labels which appears in the Enter Measurement Values window is set in
Configuration Mode.
2 Select the appropriate setting for the screen brightness. 10 is the brightest, 1 is the least bright.
Optimum is suitable for most monitoring locations and optimizes power usage for battery
powered monitors.
The MX400/450/500/550 monitors can be configured to automatically adapt the screen brightness to
the ambient light conditions. The range within which this adaptation is made is determined by the
setting made with the Brightness SmartKey.
Your monitor may be configured with a lower brightness for Standby mode and also (for battery
powered monitors) for transport to conserve battery power. These settings can only be changed in the
monitor's Configuration Mode.
If you are using a monitor with an external display, the Brightness SmartKey does not adjust the
brightness of this display. Refer to the documentation supplied with the external display for
instructions.
47
1 Introduction
WARNING
Changing the date or time will affect the storage of trends and events.
Getting Started
Once you understand the basic operation principles, you can get ready for monitoring. We also
recommend working through the E-Learning Education Programs for self-training before using the
monitor (available at the Philips Learning Center, english only).
1 Before you start to make measurements, carry out the following checks on the monitor including
all connected MMSs, modules, or MMS extensions.
– Check for any mechanical damage.
– Check all the external cables, plug-ins and accessories.
2 Plug the power cord into the AC power source.
3 Check all the functions of the instrument that you need to monitor the patient, and ensure that the
instrument is in good working order.
Switching On
Press the power on/standby switch on the monitor for one second. The monitor performs a self test
during which all lamps will light up and a tone will be sounded, and is then ready to use. If you see a
message such as CO₂ Sens Warmup wait until it disappears before starting to monitor that
measurement.
48
1 Introduction
Connected devices usually take their power from the monitor. External devices such as gas monitors
and those connected via VueLink/IntelliBridge have their own power switches.
Starting Monitoring
After you switch on the monitor,
1 Admit your patient to the monitor.
2 Check that the profile, alarm limits, alarm and QRS volumes, patient category and paced status and
so forth are appropriate for your patient. Change them if necessary.
3 Refer to the appropriate measurement chapter for further details of how to perform the
measurements you require.
WARNING
During MR imaging, remove all transducers, sensors and cables from the patient. Induced currents
could cause burns.
49
1 Introduction
If Automat. Default is set to Yes, the default profile will be loaded when power is restored. If
Automat. Default is set to No, all active settings are retained, if power is restored within 48 hours. The
Automat. Default setting is made in Configuration Mode.
Networked Monitoring
You can connect your monitor to an Information Center on a network, using one of the optional
interfaces:
• Standard wired LAN
• Wireless LAN
• IntelliVue Instrument Telemetry System (IIT).
WARNING
Only connect patient monitors to networks that conform to the network installation instructions
provided by Philips for your system.
It is possible to assign additional monitoring equipment and a telemetry device to the same patient,
resulting in the information from multiple devices being combined in one sector at the Information
Center. (See “When Multiple Equipment is Used for One Patient” on page 122 and “Using a
Telemetry Device and a Monitor (PIIC only)” on page 307.)
If your monitor is connected to a network, a network symbol is displayed in the upper left corner next
to the bed label. To see details about the Care Group, the monitoring equipment, and technical
information about the network, select the bed label in the monitor info line.
Be aware that some network-based functions may be limited for monitors on wireless networks in
comparison to those on wired networks.
Printout functionality and data export are not guaranteed when using a standard hospital network.
WARNING
• Some clinical applications may show data from another patient. Be aware that some of the data on
your patient monitor display may not always be from your patient.
• Applications running on the iPC cannot act as a primary alarming device and cannot be relied
upon for alarm notification. There may be no audible or visible indications apart from what is
shown on the screen and any data shown may be delayed.
50
1 Introduction
CAUTION
When you install any software on the iPC you are responsible for ensuring that the resulting system
complies with all relevant local regulations.
WARNING
Always minimize the iPC audio volume or mute it completely when it is not needed for an iPC
application. This will avoid iPC tones distracting from or masking tones from the monitor.
51
1 Introduction
• The host monitor is the "master" of the system, and you have full control over all the system's
operation only via the host monitor.
• Functions you can operate on a monitor in companion mode are restricted to measurements
originating in that device. If you try to operate controls that are disabled, you are prompted by the
message Not available in Companion Mode.
• Depending on how it is configured, your host monitor can determine whether the user interface of
a connected monitor in companion mode is completely disabled or not, and what is displayed on
the screen (a standard main screen, or a blank screen indicating Companion Mode).
This is controlled by two monitor settings that are applied to the monitor in companion mode on
connection. You can change the settings in Configuration Mode.
• Global settings such as line frequency, QRS sound and ECG lead colors from the host monitor,
are applied to the monitor in companion mode on connection. When disconnected from the host,
the monitor in companion mode re-applies its own global settings.
• No audible alarms are available on a monitor in companion mode when connected to a host
monitor. The only visual alarm indication is provided by the alarm lamps which are controlled by
the host monitor. Alarms become active again as soon as the monitor in companion mode is
disconnected from the host monitor.
52
1 Introduction
• The host monitor is the master of all ADT information. ADT operations on the monitor in
companion mode are disabled, and any pending actions on the monitor in companion mode (for
example, admit or end case) are canceled.
• The date and time of the monitor in companion mode is synchronized with that of the host
monitor.
• An X2 switches on automatically when connected to a running host monitor.
• When an X2 is disconnected from a running host monitor, the X2 continues to run without
interruption on battery power.
• When an X2 is used with host monitors having different software revisions, be aware that
functionality set up in a monitor with a newer revision will disappear when the X2 is connected to
a monitor with an older revision without that functionality. For example, if an X2 is used with a
revision H monitor and has been set up to alarm on Afib, this alarm will no longer exist when the
X2 is connected to a revision G monitor. If you work in a mixed software environment, inform
yourself about the differences between revisions by referring to the What's New chapter.
• Event surveillance in the monitor in companion mode is disabled. Main Setup menu operations
and SmartKeys are disabled. While connected to a host monitor, no new events are detected in the
monitor in companion mode, and no events are deleted. There is no transfer of stored events from
the monitor in companion mode to the host monitor. After disconnection from the host monitor,
event surveillance is enabled again in the monitor in companion mode, and new events are
detected.
53
1 Introduction
54
2
2 What's New?
This section lists the most important new features and improvements to the monitor and its user
interface introduced with each release. Further information is provided in other sections of this book.
You may not have all of these features, depending on the monitor configuration purchased by your
hospital.
55
2 What's New?
56
2 What's New?
• Analysis results included in reports and additional report selections with ST Map.
• Remote operation of the 12-Lead Export function at the IntelliVue Information Center.
• Remote operation of the 12-Lead Lock/Unlock function at the IntelliVue Information Center.
• New filter settings that are used as a default for future 12-Lead captures.
57
2 What's New?
58
3
3 Alarms
The alarm information here applies to all measurements. Measurement-specific alarm information is
discussed in the sections on individual measurements.
The monitor has two different types of alarm: patient alarms and INOPs.
Patient Alarms
Patient Alarms are red and yellow alarms. A red alarm indicates a high priority patient alarm such as a
potentially life threatening situation (for example, asystole). A yellow alarm indicates a lower priority
patient alarm (for example, a respiration alarm limit violation). Additionally there are short yellow
alarms, most of which are specific to arrhythmia-related patient conditions (for example, ventricular
bigeminy).
INOPs
INOPs are technical alarms, they indicate that the monitor cannot measure or detect alarm conditions
reliably. If an INOP interrupts monitoring and alarm detection (for example, Leads Off), the monitor
places a question mark in place of the measurement numeric and an audible indicator tone will be
sounded. INOPs without this audible indicator indicate that there may be a problem with the reliability
of the data, but that monitoring is not interrupted.
Most INOPs are light blue, however there are a small number of INOPs which are always yellow or
red to indicate a severity corresponding to red and yellow alarms. The following INOPs can also be
configured as red or yellow INOPs to provide a severity indication:
• ECG Leads Off
• NBP Cuff Overpress
• Cuff Not Deflated
• Occlusion
• <SpO₂ Label> No Pulse
• <Press Label> No Pulse
• Tele Disconnected
• Battery Empty / Replace Battery
All monitors in a unit should have the same severity configured for these INOPs.
59
3 Alarms
Alarm Delays
There is a delay between a physiological event at the measurement site and the corresponding alarm
indication at the monitor. This delay has two components:
• The general measurement delay time is the time between the occurrence of the
physiological event and when this event is represented by the displayed numerical values.
This delay depends on the algorithmic processing and, for certain measurements (SpO2, EEG and
BIS), on the configured averaging time. The longer the averaging time configured, the longer the
time needed until the numerical values reflect the physiological event.
• The time between the displayed numerical values crossing an alarm limit and the alarm
indication on the monitor. This delay is the sum of the alarm delay configured for the specific
measurement plus the system alarm delay. The system alarm delay is the processing time the
system needs for any alarm on the monitor to be indicated after the measurement has triggered the
alarm. See the performance specifications in the Specifications chapter for the system alarm delay
specification.
The alarm delay configured for a specific measurement is normally a fixed time. For SpO2 it is also
possible to configure a Smart Alarm Delay that is calculated using an intelligent algorithm. See “Smart
Alarm Delays” on page 195 in the SpO2 chapter for more details.
Multiple Alarms
If more than one alarm is active, the alarm messages are shown in the alarm status area in succession.
An arrow symbol next to the alarm message informs you that more than one message is active.
The monitor sounds an audible indicator for the highest priority alarm. If more than one alarm
condition is active in the same measurement, the monitor announces the most severe. Your monitor
may be configured to increase alarm indicator volume automatically during the time when the alarm is
not acknowledged.
Alarm Message
An alarm message text appears in the alarm status area at the top of the screen indicating the source of
the alarm. If more than one measurement is in an alarm condition, the message changes every two
seconds, and has an arrow ( ) at the side. The background color of the alarm message matches the
alarm priority: red for red alarms, yellow for yellow alarms, light blue for standard INOPs, red for red
INOPs and yellow for yellow INOPs. The asterisk symbols (*) beside the alarm message match the
alarm priority: *** for red alarms, ** for yellow alarms, * for short yellow alarms. Standard INOPs do
not have a symbol, red and yellow INOPs have exclamation marks beside the alarm message: !!! for
red INOPs and !! for yellow INOPs.
60
3 Alarms
Depending on how your monitor is configured, it may display alarm limit violation messages:
• in text form, for example ** SpO₂ Low or
• in numeric form, for example **Pulse xxx>yyy where the first number shows the maximum
deviation from the alarm limit, and the second number shows the currently set limit.
Flashing Numeric
The numeric of the measurement in alarm flashes.
Alarm Lamp
A lamp on the monitor's front panel flashes. The alarm lamp is divided into two sections. The right
one flashes for a patient alarm, except for short yellow alarms where the lamp will light for
approximately six seconds. The color is yellow or red corresponding to the highest priority patient
alarm currently present. The left one lights continuously for a light blue INOP and flashes for yellow
or red INOPs as follows:
MX400/450/ If the screen brightness is configured to automatically adapt to ambient light conditions, the alarm
500/550 lamps will also adapt their brightness accordingly.
61
3 Alarms
WARNING
• Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of alarm
volume to a low level or off during patient monitoring may result in patient danger. Remember
that the most reliable method of patient monitoring combines close personal surveillance with
correct operation of monitoring equipment.
• No audible alarm indicators are available on an X2 when connected to a host monitor
(Companion Mode is indicated). Alarms become active again as soon as the X2 is disconnected
from the host monitor.
To change the volume, select the volume symbol and then select the required volume from the
pop-up selection.
62
3 Alarms
• If you want to see a numerical indication of the current alarm volume on a scale from zero to 10,
or change the setting, select the Alarm Volume SmartKey.
The volume scale pops up. The current setting is indented. To change the setting, select the
required number on the scale. Any settings that are inactive ("grayed out") have been disabled in
the monitor's Configuration Mode.
• When the alarm volume is set to zero (off), the alarm volume symbol reflects this.
If you switch the alarm volume off, you will not get any audible indication of alarm conditions.
There is no alarm volume indication on the screen of an X2 when connected to a host monitor
(Companion Mode is indicated).
Acknowledging Alarms
To acknowledge all active alarms and INOPs, select the Silence permanent key. This switches off the
audible alarm indicators and alarm lamps.
63
3 Alarms
Alternatively, you can acknowledge alarms by pressing the Silence hardkey on the MMS or on the
SpeedPoint. The hardkeys follow the behavior configured for the permanent key.
A check mark beside the alarm message indicates that the alarm has been acknowledged . If the
monitor is configured to re-alarm, a dashed check mark will be shown .
If the condition that triggered the alarm is still present after the alarm has been acknowledged, the
alarm message stays on the screen with a check mark symbol beside it, except for NBP alarms and
alarms from other intermittent measurements. When such an alarm is acknowledged the alarm
message disappears.
If the alarm condition is no longer present, all alarm indicators stop and the alarm is reset.
Switching off the alarms for the measurement in alarm, or switching off the measurement itself, also
stops alarm indication.
Alarm Reminder
If Alarm Reminder is configured on for your monitor, you will get an audible reminder of alarm
conditions that remain active after you have acknowledged the alarm. This reminder may take the form
of a repetition of the alarm tone for a limited time, or an unlimited repetition of the alarm tone (this is
the same as a new alarm). Alarm Reminder is not available for standard, light blue INOPs but for
yellow and red INOPs.
In Configuration Mode, you can set the interval between silencing the alarm and sounding the
reminder tone to one, two, or three minutes.
The alarm reminder behavior at the Information Center is different to that at the monitor. Refer to the
Information Center Instructions for Use for further information.
64
3 Alarms
There are some settings made in Configuration Mode that can affect the availability of the pause
alarms functionality.
• The Pause Alarms / Alarms Off permanent key can be removed from the screen to avoid
unintentional switching off of alarms. The corresponding hardkey on the SpeedPoint or
Navigation Point is then also disabled. In this case you can only pause alarms or switch alarms off
permanently in the Alarms menu, under Main Setup.
• The Pause Alarms / Alarms Off permanent key and the corresponding hardkey can be
configured to pause or switch off red and yellow alarms, yellow alarms only, or not to function at
all. If they are configured not to function, you cannot pause alarms or switch alarms off
permanently at all.
When the alarms off priority is set to Yellow Only, the Pause Alarms / Alarms Off key becomes
the Pause Yellow / Yellow Al. Off key.
• Select the Pause Alarms permanent key. If your monitor is configured to infinite pause time, the
permanent key is labeled Alarms Off, and selecting it switches alarms off.
• Or press the Alarms hardkey on the SpeedPoint or Navigation Point. The hardkey follows the
behavior configured for the permanent key.
Depending on the configuration, you may need to select Confirm to complete the change.
• Select the Alarms Off (or Yellow Al. Off) permanent key.
• Or press the Alarms hardkey on the SpeedPoint or Navigation Point. The hardkey follows the
behavior configured for the permanent key.
Depending on the configuration, you may need to select Confirm to complete the change.
65
3 Alarms
66
3 Alarms
Alarm Limits
The alarm limits you set determine the conditions that trigger yellow and red limit alarms. For some
measurements (for example, BIS and SpO2), where the value ranges from 100 to 0, setting the high
alarm limit to 100 switches the high alarm off, or setting the low alarm limit to 0 switches it off. In
these cases, the alarms off symbol is not displayed.
WARNING
Be aware that the monitors in your care area may each have different alarm settings, to suit different
patients. Always check that the alarm settings are appropriate for your patient before you start
monitoring.
If your monitor is not configured to show the alarm limits next to the numeric, or if the numeric is so
small that the limits cannot be displayed, you can see them in the appropriate measurement setup
menu. Select the measurement numeric to enter the menu and check the limits.
67
3 Alarms
To open the Alarm Limits window, either select any alarm field to open the Alarm Messages window,
then select the Alarm Limits pop-up key, or select the Alarm Limits SmartKey, if configured.
68
3 Alarms
1 Parameter label
2 High red alarm limit (view only)
3 High yellow alarm limit field. Select to open a pop-up list of high alarm limits
4 Alarms On/Off key - select to switch between alarms on or off
5 Preview Alarm AutoLimits for a measurement before applying
6 Select to apply wide AutoLimits
7 Select to apply narrow AutoLimits
8 Low yellow alarm field. Select to open a pop-up list of low alarm limits
9 Low red alarm (view only)
10 Graphic view of alarm limits with currently measured value
11 15-min trend, showing alarm limits and monitored measurement values
69
3 Alarms
If you set the yellow alarm limit outside the red alarm limit, the monitor will automatically adapt the
red alarm limit.
When an ST measurement is in the alarm limits window there are also two pop-up keys available
labeled All ST Narrow/All ST Wide. With these keys you can set Auto Limits for all ST Leads.
Alarm Limits can also be changed at the IntelliVue Information Center iX.
Limits Narrow sets limits close to the currently measured values for situations where it is critical for
you to be informed about small changes in your patient's vital signs.
Limits Wide sets limits further away from the currently measured values for situations where small
changes are not so critical.
Use the keys in the measurement alarm limits window to apply AutoLimits for individual
measurements. These keys are not available if AutoLimits have been disabled for the measurement in
the monitor's Configuration Mode.
AutoLimits are not available for all measurements. The list of measurements for which AutoLimits can
be used is defined in the monitor's Configuration mode.
Use the measurement alarm limits window to check AutoLimits before you apply them to ensure that
they are appropriate for your individual patient and their clinical condition. Once applied, AutoLimits
are shown on the monitor screen just like manually-set alarm limits. If the AutoLimits are not
appropriate for your patient, you must set alarm limits manually. The limits remain unchanged until
you set them again or change them manually.
70
3 Alarms
• Select the Print Limits pop-up key to print an overview of all alarm limits on a connected printer.
• Select the Record Limits pop-up key to send a recording of the alarm limits to a recorder.
Reviewing Alarms
You can see which alarms and INOPs are currently active in the respective alarms and INOPs fields at
the top of the screen.
To see the currently active alarms and INOPs listed in one place, select any of the alarm status areas on
the monitor screen. The Alarm Messages window pops up.
All alarms are erased from the Alarm Messages window when you discharge a patient, or if you
change to Demonstration Mode.
71
3 Alarms
The Review Alarms window pop-up keys appear when the window is opened. If alarm pause
extension is disabled, the pause pop-up keys are inactive. Selecting the Active Alarms pop-up key
opens the Alarm Messages window.
Latching Alarms
The alarm latching setting for your monitor defines how the alarm indicators behave when you do not
acknowledge them. When alarms are set to non-latching, their indicators end when the alarm condition
ends. Switching alarm latching on means that visual and/or audible alarm indications are still displayed
or announced by the monitor after the alarm condition ends. The indication lasts until you
acknowledge the alarm.
72
3 Alarms
All INOPs are non-latching. See “Yellow Arrhythmia Alarms” on page 160 for information on one-
star yellow alarms latching behavior.
Testing Alarms
When you switch the monitor on, a selftest is started. You must check that the alarms lamps light, one
after the other, and that you hear a single tone. This indicates that the visible and audible alarm
indicators are functioning correctly. For further testing of individual measurement alarms, perform the
measurement on yourself (for example SpO2 or CO2) or use a simulator. Adjust alarm limits and check
that appropriate alarm behavior is observed.
Alarm Recordings
You can set up your monitor so that it automatically triggers alarm recordings locally or at the
Information Center, or if configured, to a printer as a realtime report.
1 Press the Main Setup SmartKey.
2 Select Alarms from the Main Setup menu.
3 Select Alarm Recording from the Alarms menu to open the Alarm Recordings menu.
4 Select a measurement from those listed for which you want to change the alarm condition that
triggers an alarm recording. This opens a pop-up list.
5 For the desired measurement(s), choose the alarm condition to trigger an alarm recording:
Red Only: an alarm recording will automatically be triggered when the measurement enters a red
alarm condition.
Red&Yellow: both yellow and red alarms will trigger an alarm recording.
Off: disables automatic alarm recording.
Refer to the "Recording" chapter for details of how to set up a recording.
73
3 Alarms
74
4
75
4 Patient Alarms and INOPs
76
4 Patient Alarms and INOPs
77
4 Patient Alarms and INOPs
78
4 Patient Alarms and INOPs
79
4 Patient Alarms and INOPs
80
4 Patient Alarms and INOPs
Monitor INOPs
INOP Message, Indication What to do
Bad MSL 1) An MMS with an incompatible software revision is connected to the monitor. This combination
INOP tone does not allow monitoring, OR
2) You cannot use this combination of monitor, MMS and cable. Switch off the monitor and contact
your service personnel.
Central: Tele Only System connectivity via telemetry device is limited (No alarms, only local numerics) when in
INOP tone companion mode and host monitor does not have system connectivity. Only telemetry device
parameters can be displayed at central station.
Check Alarm Lamps Perform a visual check of the alarm lamp to establish whether there is a problem. Contact your service
INOP tone personnel to check the internal connections to the alarm lamps.
Check DrugSettings There was a problem loading the drug settings. Check that the settings are complete and correct.
INOP tone
!!Check ECG Source The telemetry device and the monitor both have valid ECG signals. Unpair the telemetry device and
Yellow tone the monitor if they are no longer used for the same patient.
!! Check Equipment There is an equipment status dispute relating to one or more of the devices assigned to this patient. See
INOP tone/Yellow tone the Equipment window for details.
Check Flex Texts Check the names of the monitor menus, for example the labels for screens, profiles, event or trend
INOP tone group names, before you resume monitoring. If they are unexpected, there may be a problem with the
monitor software. Contact your service personnel.
Check Keyboard Perform a visual and functional check of the keyboard. Contact your service personnel.
INOP tone
Check Main Board 2 There is a problem with the second main board in the monitor. Contact your service personnel.
INOP tone
Check Monitor Func Potential problem with alarm lamps, display or interfaces detected. Contact your service personnel.
INOP tone
Check Monitor Temp The temperature inside the monitor is too high. Check that the monitor ventilation is not obstructed.
INOP tone If the situation continues, contact your service personnel.
Check Mouse Perform a visual and functional check of the mouse input device. Contact your service personnel.
INOP tone
Check MSL Voltage There is a problem with the voltage of the Measurement Link (MSL). Contact your service personnel.
INOP tone
Check Network Conf The monitor is receiving network topology information from more than one source, e.g. the Database
INOP tone Server and an Application Server. Contact your service personnel.
!! Check Pairing There is a problem with device pairing. Check that the monitor and telemetry device are correctly
Yellow tone paired.
!!Check Patient ID There is a mismatch between patient data in two connected devices. Resolve the mismatch to allow
Yellow tone settings and data synchronization.
81
4 Patient Alarms and INOPs
82
4 Patient Alarms and INOPs
83
4 Patient Alarms and INOPs
Battery INOPs
INOP Message, Indication What to do
Batt 1 Missing The monitor requires two batteries but can detect only one battery. Insert the missing battery
Batt 2 Missing immediately.
INOP tone
During this INOP, alarms cannot
be paused or switched off.
Batt Empty The estimated remaining battery-powered operating time is less than 10 minutes. Replace the battery
!!Batt Empty immediately.
!!!Batt Empty If the condition persists and the monitor is not connected to mains power, this INOP is re-issued two
INOP tone, battery LED flashes minutes after you acknowledge it.
During this INOP, alarms cannot
be paused or switched off.
Batt Extensn Malf There is a hardware error in the Battery Extension. Contact your service personnel.
INOP tone
Batt Incompat The battery cannot be used with this monitor. Replace with the correct battery as specified in this
INOP tone book.
Batt Low The estimated battery-powered operating time remaining is less than 20 minutes.
INOP tone
Batt Malfunction The monitor cannot determine the battery status. If this INOP persists, replace the faulty battery. If
INOP tone, battery LED flashes the condition persists and the monitor is not connected to mains power, this INOP is re-issued two
During this INOP, alarms cannot minutes after you acknowledge it.
be paused or switched off unless Place the battery in a different monitor or in a battery charger. If the same INOP is shown, contact
the monitor is connected to your service personnel.
mains power.
Batteries Empty The estimated remaining battery-powered operating time of the indicated battery or batteries is less
!!Batteries Empty than 10 minutes. Replace the batteries immediately.
!!!Batteries Empty If the condition persists and the monitor is not connected to mains power, this INOP is re-issued two
Batt 1 Empty minutes after you acknowledge it.
!!Batt 1 Empty
!!!Batt 1 Empty
Batt 2 Empty
!!Batt 2 Empty
!!!Batt 2 Empty
INOP tone, battery LED flashes
During this INOP, alarms cannot
be paused or switched off.
Batteries Incompat The indicated battery or batteries cannot be used with this monitor. Replace with the correct battery or
Batt 1 Incompat batteries as specified in this book.
Batt 2 Incompat
INOP tone
Batteries Low The estimated battery-powered operating time remaining is less than 20 minutes.
Batt 1 Low
Batt 2 Low
INOP tone
84
4 Patient Alarms and INOPs
85
4 Patient Alarms and INOPs
Display INOPs
INOP Message, Indication What to do
Indep.Dsp Malfunc. There is a problem with the Independent Display. Check the MSL coupling cable then contact your
service personnel.
Indep.Dsp NotSupp. The monitor does not support a second main display. The monitor software is incompatible. Contact
your service personnel.
Intell.Dsp Malf. There is a problem with the Intelligent Display. Check the MSL coupling cable then contact your
service personnel.
Intell.Dsp Missing The monitor has lost contact with the connected Intelligent Display. Contact your service personnel.
Intell.Dsp Unsupp. The monitor does not support the connected Intelligent Display. The monitor software is
incompatible.
86
4 Patient Alarms and INOPs
87
4 Patient Alarms and INOPs
Pulse INOPs
INOP Message, Indication What to do
Pulse Not Alarming Pulse has no alarming because the system pulse is measured by an external device. Select another pulse
Numeric is replaced by -?- source to enable pulse alarming.
INOP tone
Resp INOPs
INOP Message, Indication What to do
Resp Equip Malf Contact your service personnel. The RESP hardware is faulty.
Numeric is replaced by -?-
INOP tone
Resp Erratic The monitor has detected too many artifacts in the measured Resp signal. Check that the RA and LL
Numeric is replaced by -?- electrodes are correctly attached and have not dried out.
Resp Leads Off Not all the required leads for Resp monitoring are attached. Make sure that the RA and LL leads are
Numeric is replaced by -?- attached.
INOP tone
NBP INOPs
INOP Message, Indication What to do
!! Cuff Not Deflat Remove the cuff from the patient. Make sure that the tubing is not kinked or twisted and that the
!!!Cuff Not Deflat correct patient category is selected. Try repeating the measurement.
Numeric is displayed with a -?- You can silence the INOP, but the INOP message remains visible until the next NBP measurement is
Severe yellow/red INOP tone started or the Stop All SmartKey is selected.
During this INOP, alarms cannot [Adult or pediatric patients: The NBP cuff pressure has exceeded 15 mmHg (2 kPa) for more than
be paused or switched off. 3 minutes.
Neonatal patients: The NBP cuff pressure has exceeded 5 mmHg (0.7 kPa) for more than 90 seconds.]
!! Cuff Overpress The NBP cuff pressure exceeds the overpressure safety limits. Remove the cuff from the patient. Make
!!!Cuff Overpress sure that the tubing is not kinked or twisted and that the correct patient category is selected. Try
Numeric displayed with a -?- restarting the measurement.
Severe yellow/red INOP tone You can silence this INOP, but the INOP message remains visible until the next measurement is
During this INOP, alarms cannot started or the Stop All SmartKey is selected.
be paused or switched off.
NBP Deactivated The NBP measurement label in the measurement device has been deactivated by deactivating the label
INOP tone in the Measurement Selection window. The measurement automatically disappears from the
display. To switch the measurement on again, reactivate the measurement label in the Measurement
Selection window.
NBP Equip Malf Remove the cuff from the patient. The NBP hardware is faulty. Contact your service personnel.
Numeric is replaced by -?- You can silence this INOP, but the INOP message remains visible until the next measurement is
INOP tone started or the Stop All SmartKey is selected.
88
4 Patient Alarms and INOPs
Temperature INOPs
INOP Message, Indication What to do
T1, T2, T3, T4 INOPs See <Temp Label> INOPs
Tamb INOPs See <Temp Label> INOPs
Tart INOPs See <Temp Label> INOPs
Tcereb INOPs See <Temp Label> INOPs
Tcore INOPs See <Temp Label> INOPs
<Temp Label> A Temp measurement label in the measurement device has been deactivated, either by connecting a
Deactivated Pressure transducer in the shared Press/Temp socket, or by deactivating the label in the
INOP tone Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either reconnect a Temp transducer or reactivate the
measurement label in the Measurement Selection window.
<Temp Label> Equip Malf Contact your service personnel.
Numeric is replaced by -?- The temperature hardware is faulty.
INOP tone
<Temp Make sure the TEMP probe is connected to the MMS or module.
Label>NoTransducer If you silence this INOP, the measurement will be switched off.
Numeric is replaced by -?-
INOP tone
<Temp Label> Overrange Try changing the application site of the transducer.
Numeric is replaced by -?- [The temperature is less than -1°C, or greater than 45°C.]
INOP tone
<Temp Label> Unplugged A Temp measurement label has been deactivated, either by unplugging a module, or by deactivating
INOP tone the label in the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module or reactivate the measurement label in
the Measurement Selection window.
Tesoph INOPs See <Temp Label> INOPs
Tnaso INOPs See <Temp Label> INOPs
Trect INOPs See <Temp Label> INOPs
Tskin INOPs See <Temp Label> INOPs
Ttymp INOPs See <Temp Label> INOPs
89
4 Patient Alarms and INOPs
SpO2 INOPs
INOP Message, Indication What to do
<SpO₂ Label> Chk Sensor The condition of the signal at the SpO2 sensor is not as expected. Accuracy may be compromised.
Check proper application of the sensor in accordance with the Instructions for Use. If the INOP
persists, consider exchanging the sensor.
<SpO₂ Label> Deactivated The SpO2 measurement label in the measurement device has been deactivated by deactivating the label
INOP tone in the Measurement Selection window. The measurement automatically disappears from the
display. To switch the measurement on again, reactivate the measurement label in the Measurement
Selection window.
<SpO₂ Label> Equip Malf The MMS or module is faulty. Unplug and replug the MMS or module. If the INOP persists, contact
Numeric is replaced by -?- your service personnel.
INOP tone
<SpO₂ Label> Erratic Check the sensor placement. Try another adapter cable and sensor. If the INOP persists, contact your
Numeric is replaced by -?- service personnel.
INOP tone
<SpO₂ Label> The update period of displayed values is extended due to an NBP measurement on the same limb or
Extd.Update an excessively noisy signal.
Numeric is replaced by -?-
(questionable numeric)
<SpO₂ Label>Interference There is too much interference, caused by a high level of ambient light and/or electrical interference.
Numeric is replaced by -?- Cover the sensor to minimize ambient light. If the INOP persists, make sure that the sensor cable is
INOP tone not damaged or positioned too close to power cables.
<SpO₂ Label> Low Perf Accuracy may be compromised due to very low perfusion. Stimulate circulation at sensor site. If INOP
Numeric is replaced by -?- persists, change the measurement site.
(questionable numeric)
<SpO₂ Label> No Pulse Check the perfusion at measurement site. If necessary, stimulate circulation or change measurement
Numeric is replaced by -?- site. If the INOP is due to NBP measurement on the same limb, wait until the NBP measurement is
INOP tone finished.
<SpO₂ Label> No Sensor Make sure the SpO2 sensor is connected. If the INOP persists, try another adapter cable and sensor. If
Numeric is replaced by -?- you silence this INOP, the measurement will be switched off.
INOP tone
<SpO₂ Label> NoisySignal Excessive patient movement or electrical interference is causing irregular pulse patterns. Try to reduce
Numeric is replaced by -?- patient movement or to relieve the cable strain on the sensor.
INOP tone
<SpO₂ Label> Poor Signal The signal condition of the SpO2 measurement is poor and measurement accuracy may be
Numeric is replaced by -?- compromised. If the INOP persists, consider changing the application site or using another sensor.
(questionable numeric)
<SpO₂ Label> Pulse? The detectable pulsations of the SpO2 signal are outside the specified pulse rate range.
Numeric is replaced by -?-
INOP tone
<SpO₂ Label> Searching SpO2 is analyzing the patient signal to derive Pulse, SpO2 and Perf values. Please wait until the search
Numeric unavailable analysis is complete.
<SpO₂ Label> Sensor Malf The SpO2 sensor or adapter cable is faulty. Try another adapter cable and sensor. If the INOP persists,
Numeric is replaced by -?- contact your service personnel.
INOP tone
90
4 Patient Alarms and INOPs
Pressure INOPs
INOP Message, Indication What to do
ABP INOPs See <Press Label> INOPs
Ao INOPs See <Press Label> INOPs
ART INOPs See <Press Label> INOPs
BAP INOPs See <Press Label> INOPs
CPP Chk Sources Not all measurements or values required to perform the calculation are available. Check the
Numeric is replaced by -?- measurement sources.
INOP tone
CPP Chk Units The monitor has detected a conflict in the units used for this calculation. Check the unit settings.
Numeric is replaced by -?-
CPP Disabled CPP has been disabled, either in the setup in Configuration mode, or by loading settings with CPP
disabled. This can happen when changing from a pediatric or neonatal profile to an adult profile, as the
default profile settings have the CPP measurement disabled in the adult profile and enabled in the
pediatric and neonatal profiles. Enable CPP in the current profile to clear the INOP.
CVP INOPs See <Press Label> INOPs
FAP INOPs See <Press Label> INOPs
IC1 / IC2 INOPs See <Press Label> INOPs
ICP INOPs See <Press Label> INOPs
LAP INOPs See <Press Label> INOPs
P / P1 / P2 / P3 / P4 INOPs See <Press Label> INOPs
PAP INOPs See <Press Label> INOPs
PPV bad <Press Label> The arterial pressure source selected for PPV is not providing a pulsatile signal.
Signal
PPV bad Signal The arterial pressure source selected for PPV is not providing a pulsatile signal.
at Information Center
PPV Chk Sources The arterial pressure source selected for PPV is unplugged or switched off. When this INOP has
displayed for 1 minute PPV will be switched off.
<Press Label> Artifact A non-physiological event (flush or blood sample) is detected. A resulting limit alarm or non-pulsatile
Numeric questionable INOP will be suppressed.
91
4 Patient Alarms and INOPs
CO2 INOPs
INOP Message, Indication What to do
CO₂ Auto Zero The automatic zero calibration is in progress. This typically takes 10 seconds. During this time the CO2
Numeric is replaced by -?- values may not be updated, or they may be replaced by -?-. Wait until the zero calibration is complete
if the Autozero lasts >15 sec, to resume monitoring.
INOP tone sounds.
CO₂ Cal Failed Make sure that the Cal cell was changed between CAL1 and CAL2. Repeat the calibration. If the
Numeric is replaced by -?- INOP reappears, try another transducer. If the INOP persists, contact your service personnel.
INOP tone
92
4 Patient Alarms and INOPs
93
4 Patient Alarms and INOPs
SO2 INOPs
INOP Message, Indication What to do
<SO₂ Label> Cal Failed The calibration failed. Check the catheter-to-Optical-Module connection. Manually restart the
Numeric is replaced by -?- calibration. Try another catheter and Optical Module. If the catheter is already inserted, perform an in-
INOP tone vivo calibration.
<SO₂ Label>Cal Required There is no valid calibration data in the Optical Module. Perform either a pre-insertion or an in-vivo
Numeric is replaced by -?- calibration.
INOP tone
<SO₂ Label> Cal. Mode Pre-insertion calibration is complete, but the catheter tip is still inside the optical reference. The
Numeric is replaced by -?- catheter is now ready for insertion.
INOP tone
<SO₂ Label> Cannot Meas The signal is out of the normal range, and no oxygen saturation can be derived. Perform an in-vivo
Numeric is replaced by -?- calibration. If the INOP persists, try another Optical Module and catheter.
INOP tone
<SO₂ Label>Config Error The Optical Module has been configured to SaO2 Mode. Use Change to Venous in the setup
Numeric is replaced by -?- menu to reconfigure to venous saturation mode.
INOP tone
<SO₂ Label> Conn The Optical Module was disconnected during data storage. Reconnect the Optical Module for at least
OptMod 20 seconds.
Numeric is replaced by -?-
INOP tone
<SO₂ Label> Equip Malf The SO2/SvO2 Module or Optical Module is faulty. Unplug and replug the Optical Module and SO2/
Numeric is replaced by -?- SvO2 module. Exchange the modules. If the INOP persists, contact your service personnel.
INOP tone
<SO₂ Label> Incompat. The SO2 Module or Optical Module is not supported. Contact your service personnel.
INOP tone
<SO₂ Label> In-Vivo Cal The in-vivo calibration is not yet complete. Lab values must be stored to the Optical Module to
complete the calibration. Either continue with the next steps of the current calibration or recall the
previous calibration.
94
4 Patient Alarms and INOPs
C.O. INOPs
INOP Message, Indication What to do
C.O. Deactivated The Cardiac Output measurement label in the measurement device has been deactivated by
INOP tone deactivating the label in the Measurement Selection window. The measurement automatically
disappears from the display. To switch the measurement on again, reactivate the measurement label in
the Measurement Selection window.
C.O. Equip Malf There is a problem with the C.O. hardware. Contact your service personnel.
Numeric is replaced by -?-
INOP tone
C.O. Unplugged Plug in the C.O. module. Silencing this INOP switches off the measurement.
Numeric is replaced by -?-
INOP tone
CCI No BSA CCI cannot be calculated because the patient's body surface area is unknown. Enter the patient weight
CCI numeric unavailable and height to provide the BSA for CCI calculation.
INOP tone
CCO BadPressSignal The arterial pressure wave can currently not be used for pulse contour calculation for CCO or CCI
Numeric is replaced by -?- measurement. Possible causes are air bubbles in the tubing or a physiological condition, for example
INOP tone severe arrhythmia.
CCO No Calibration The CCO measurement is currently not calibrated.
Numeric is replaced by -?-
CCO No <Press Label> CCO/CCI cannot be calculated. Make sure that the pressure chosen in the Setup CCO menu under
Numeric is replaced by -?- CCO from matches the pressure measured with the arterial catheter for CCO measurement. A
INOP tone may sound pressure from an external device cannot be used. Select another pressure label, either ABP, Ao, ART,
BAP, FAP, or UAP.
95
4 Patient Alarms and INOPs
tcGas INOPs
INOP Message, Indication What to do
<tcGas Label> Cal Failed A calibration failed. Check the calibration unit, gas pressure, and tubing connections, then restart the
Numeric is replaced by -?- calibration. If the calibration has failed more than once, remembrane the transducer and restart the
INOP tone calibration. If this INOP persists, contact your service personnel.
<tcGas Label> Cal Wait until the tcpO2/tcpCO2 calibration is finished.
Running
Numeric displays first -?-, then
numeric is displayed with a ?
<tcGas Label> Calibration is required before applying the transducer to the patient.
CalRequired Insert a membraned transducer into the calibration chamber on the module, connect the calibration
Numeric is replaced by -?- unit to the calibration chamber, open the gas valve and start the calibration. If this INOP occurs
INOP tone during a calibration, there may be a module or transducer malfunction: contact your service personnel.
96
4 Patient Alarms and INOPs
EEG INOPs
INOP Message, Indication What to do
EEG Equip Malf The EEG hardware is faulty. Contact your service personnel.
INOP tone
EEG Impedance High The signal electrode in one or both channels exceeds the user-selected impedance limit, or the
EEG1 ImpedanceHigh impedance of a single electrode exceeds the limit. Check the impedance. If the impedance is too high,
EEG2 ImpedanceHigh reconnect the electrodes according to the EEG monitoring setup guidelines. If the INOP persists,
contact your service personnel.
EEG Line Noise Excessive line noise has been detected in either channel EEG1 or EEG2, or in both EEG channels.
EEG1 Line Noise Keep all cables together and away from metallic bodies, other cables & radiated fields.
EEG2 Line Noise
EEG Muscle Noise Too much power above 30 Hz has been detected in channel EEG1 or EEG2, or both.
EEG1Muscle Noise Check the Electrode-to-Skin Impedance and reposition the electrode away from possible muscle
EEG2Muscle Noise activity, if necessary.
EEG Unplugged Plug in module. Silencing this INOP switches off the measurement.
INOP tone
EEG1 Lead Off <n> Reconnect specified electrode.
EEG2 Lead Off <n>
[n = electrode]
EEG1 Lead Off One or more electrodes are not connected. Check in the EEG Impedance / Montage window on
EEG2 Lead Off the monitor which electrode(s) are affected and reconnect the electrodes.
at Information Center
EEG1 Leads Off Two or more electrodes are not connected. Check in the EEG Impedance / Montage window
EEG2 Leads Off which electrodes are affected and reconnect the electrodes.
EEG1 Overrange Input signal is too high in one or both channels. This is usually caused by interfering signals such as
EEG2 Overrange line noise or electrosurgery.
EEGNoTransducer The trunk cable is disconnected from the EEG plug-in module. Reconnect the trunk cable. Silencing
INOP tone this INOP switches the measurement off.
97
4 Patient Alarms and INOPs
BIS INOPs
INOP Message, Indication What to do
BIS Cable Incompat The semi-reusable sensor cable connected is unknown or not supported by your software revision.
INOP tone Replace it with a Philips-supported sensor cable.
BIS Cable Usage The semi-reusable sensor cable has exceeded the maximum number of uses. Replace the cable.
INOP tone
BIS DSC Disconnect DSC is not properly connected OR either DSC or BIS engine may be faulty.
INOP tone Make sure that the DSC is properly connected to the BIS Engine. If INOP persists, replace DSC with
a known good one of the same type.
If INOP persists replace BIS engine.
Silencing this INOP switches the measurement off.
BIS DSC Incompat. DSC is not supported by the BIS engine or new DSC connected to an old BIS engine. A software
INOP tone upgrade may be required. Contact your service personnel.
BIS DSC Malfunct Electrocautery used during self-test OR malfunction in the DSC hardware.
Make sure not to use electrocautery during the self-test procedure. Disconnect and reconnect the DSC
to the BIS engine. If the INOP persists, replace the DSC or contact your service personnel.
BIS DSC Upgrade DSC update currently being carried out. This INOP will disappear when the DSC update is finished.
INOP tone Do not disconnect the DSC during the update. No action is needed.
BIS Electr. Disc. One or more electrodes are not connected to the semi-reusable sensor cable. Check all electrode
INOP tone connections.
BIS Engine Disconn BIS engine not connected OR Module Cable defective.
INOP tone Make sure that the Module Cable is properly connected. If INOP persists, replace the Module Cable.
Silencing this INOP switches the measurement off.
BIS Engine Incomp. BIS engine software is not supported. A software upgrade may be required. Contact your service
INOP tone personnel.
MP20/30 - BIS engine not supported.
BIS Engine Malfunc Malfunction in the BIS engine hardware. Disconnect and reconnect the BIS engine. If the INOP
INOP tone persists, replace BIS engine.
BIS Equip Malf There is a malfunction in the BIS hardware. Unplug and replug the BIS module. If the INOP persists,
INOP tone contact your service personnel.
BIS High Impedance Impedance of one or more electrode(s) is above the valid range, most often caused by bad skin
INOP tone may sound preparation. Check the sensor montage and press the electrode pads firmly. If this INOP persists,
replace the sensor(s) in question using correct skin preparation.
If INOP persists, contact your service personnel.
BIS ImpedanceCheck The Cyclic Impedance check is running. It will stop automatically if all impedances are within the valid
INOP tone may sound range. If any electrodes do not pass the impedance test, check the sensor montage and press the
electrode pads firmly.
To manually stop the Cyclic Impedance Check, select Sensor Check off in the Setup BIS menu.
BIS IsoelectricEEG No discernible EEG activity is detected for longer than one minute.
Check the patient. Check that the electrodes are properly connected.
BIS Lead Off One or more electrodes have no skin contact and therefore impedances cannot be measured. Check
INOP tone may sound the sensor montage and press the electrode pads firmly.
If this INOP persists, replace the sensor(s) in question, using correct skin preparation.
BIS Overcurrent Unplug and replug the BIS module or, f or the MP20/MP30, disconnect and reconnect the BISx from
INOP tone the Interface board. If the INOP persists, contact your service personnel.
98
4 Patient Alarms and INOPs
Spirometry INOPs
INOP Message, Indication What to do
AWFChange Scale Airway flow signal exceeds range of selected scale. Adjust scale to display complete wave.
AWPChange Scale Airway pressure signal exceeds range of selected scale. Adjust scale to display complete wave.
AWVChange Scale Airway volume signal exceeds range of selected scale. Adjust scale to display complete wave.
Spiro Alarms Suppr Alarming is suppressed for the spirometry module.
SPIRO Cannot Meas Measurement is at its limit, e.g. ambient pressure out of range.
Spiro Gas Compens? Gas compensation is set to Gas Analyzer but not all gases necessary for compensation are
measured by a gas monitor. Some of the fall-back values provided by the user are used. Measurement
accuracy might be reduced.
Spiro Incompatible Module revision not compatible with the host monitor software revision. Contact your service
personnel.
Spiro Malfunction Module failure detected. Contact your service personnel.
Spiro No Breath No breath was detected for more than 25 seconds. Breath derived numerics are not available.
Spiro No Sensor No sensor detected. Make sure the correct sensor is attached to the breathing circuit.
99
4 Patient Alarms and INOPs
NMT INOPs
INOP Message, Indication Explanation
NMT Alarm Suppress NMT alarms are suppressed until the TOF count reaches zero for the first time.
NMT Cable Disconn No patient cable is connected to the NMT measurement module. Silencing this INOP switches the
INOP tone NMT measurement off.
NMT Cable Unknown The patient cable connected to the NMT module is not supported or cannot be identified. Replace it
INOP tone with a supported cable.
NMT Cal Failed The NMT calibration cycle failed. Check that the sensor and the electrodes are placed properly, then
start another calibration. Ensure that patient's arm is kept still before and during the calibration cycle.
If calibration fails again, replace the patient cable. If INOP persists, contact your service personnel.
NMT Cal Running An NMT calibration is running. Wait until the calibration is complete.
NMT Cannot Measure NMT is unable to reliably derive measurement values. Check that the sensor and the electrodes are
INOP tone placed properly and start another measurement. If this INOP persists, try another patient cable.
NMT Equip Malfunct There is a problem with the NMT hardware. Unplug the NMT measurement module then plug it in
INOP tone again. If the INOP persists contact your service personnel.
NMT Impedance High The impedance of the NMT stimulation electrodes has exceeded the allowed limit. Check that the
electrodes are placed properly, have firm skin contact and have not dried out.
NMT Incompatible This version of the NMT measurement module is not supported, or the module firmware is
INOP tone incomplete and needs to be upgraded. Contact your service personnel.
NMT Lead Off One or both NMT stimulation electrodes have become detached from the patient. Check that both
INOP tone electrodes have firm skin contact, that the electrodes are placed properly and that the lead wires are
connected.
NMT Neo Patient? The NMT measurement does not support neonatal patients. Check the patient category.
INOP tone
NMT Noisy Signal The NMT signal is too noisy to derive measurement values reliably. If electrosurgery is currently being
INOP tone used, repeat the NMT measurement after it has stopped.
NMT Overcurrent The current through the NMT stimulation electrodes is too high. Check that that the patient cable is
INOP tone not visibly damaged. Disconnect and reconnect the patient cable. If this INOP persists, replace the
cable and/or module.
NMT Overrange The signal from the acceleration sensor is out of the measurement range. Check that the acceleration
sensor is positioned properly. If this INOP persists, try another patient cable.
NMT Sensor Malfunc There is a problem with the acceleration sensor of the NMT patient cable. Disconnect and reconnect
INOP tone the patient cable. If this INOP persists, replace the cable.
NMT Upgrade An NMT module firmware upgrade is in progress. NMT monitoring is currently not possible. Wait
INOP tone until the upgrade has completed and this INOP is no longer displayed. Do not unplug the module or
switch off power.
100
4 Patient Alarms and INOPs
VueLink INOPs
INOP Message, Indication What to do
VueLink Alarm A technical alarm is present on the VueLink module. The INOP text on the monitor is defined by the
at Information Center VueLink device driver.
<VueLink Option> Check No cable or the wrong cable connected to the VueLink module, or incorrect device selected. Silencing
Cable this INOP switches the measurement off.
INOP tone VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> Check No information was received from the external device. The device may be switched off or
Setup disconnected.
INOP tone VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> Chk The wrong external device has been selected on the VueLink module, or the external device has not
Config been correctly setup, or the wrong cable has been used to connect the device to the VueLink module.
INOP tone VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> Equip Malfunction in the VueLink module. If this message appears repeatedly, the module must be replaced.
Malf Contact your service personnel.
INOP tone VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> No The VueLink module has not been configured during installation. The installation process should be
Config completed by either your biomedical engineering department or the Philips service engineer.
INOP tone VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> The VueLink module has been unplugged from the rack, or the whole rack has been disconnected.
Unplugged The measurement automatically disappears from the display. Silencing this INOP switches off the
INOP tone measurement.
VueLink INOP abbreviations may differ slightly depending on the device category.
IntelliBridge INOPs
INOP Message, Indication What to do
Device Check Setup Device identification completed, but communication could not be established due to timeout.
INOP tone IntelliBridge INOP abbreviations may differ slightly depending on the device category.
Device Chk Config Device identification completed, but communication could not be established due to error.
INOP tone IntelliBridge INOP abbreviations may differ slightly depending on the device category.
Device Demo Data The device connected to the IntelliBridge module reports demo data but the monitor is not in DEMO
INOP tone mode.
Device Real Data The monitor is in DEMO mode but the device connected to the IntelliBridge module reports data that
INOP tone are not flagged as demo data.
Device Unsupported Device identification completed, but no appropriate device driver installed.
INOP tone IntelliBridge INOP abbreviations may differ slightly depending on the device category.
<EC10 / EC40> Equip Malfunction in the IntelliBridge module. If this message appears repeatedly, the module must be
Malf replaced. Contact your service personnel.
INOP tone
EC10 INOP A technical alarm is present on the IntelliBridge EC10 module.
!! EC10 INOP On the monitor the indication is a red (!!!), yellow (!!) or cyan alarm lamp (as appropriate), an INOP
!!! EC10 INOP tone and an INOP text that is defined by the IntelliBridge EC10 device driver.
at Information Center
<External Device> The IntelliBridge module has been unplugged from the rack, or the whole rack has been disconnected.
Unplugged Silencing this INOP switches off the measurement.
INOP tone IntelliBridge INOP abbreviations may differ slightly depending on the device category.
No Device Data Communication with connected device has been lost.
101
4 Patient Alarms and INOPs
Telemetry INOPs
INOP Message, Indication What to do
Check ECG Settings Synchronization of ECG settings between the monitor and Information Center has failed. Check that
INOP tone the ECG settings in use are appropriate.
!!Check ECG Source Both the telemetry device and the monitor have valid ECG signals
Chk SpO₂T Settings Synchronization of SpO2T settings between the monitor and Information Center has failed. Check
INOP tone that the SpO2T settings in use are appropriate.
Invalid Leadset The leadset plugged in cannot be used with the telemetry device.
Leadset Unplugged The leadset has been unplugged from the telemetry device.
No ECG Source A telemetry device is paired with the monitor but the Information Center is not detecting a valid ECG
signal from either of them.
!! Repl. Tele Batt The battery in the telemetry device is almost empty and must be replaced.
!!!Repl. Tele Batt
Severe yellow/red INOP tone
During this INOP, alarms cannot
be paused or switched off.
Tele Battery Low The battery in the Telemetry device is low and must be replaced soon.
!!Tele Disconnect Telemetry transceiver was disconnected or short range radio link was lost.
!!!Tele Disconnect For cable connections; check Telemetry interface, cable connection and setup.
Severe yellow/red INOP tone For short range radio connections: if the telemetry transceiver has not moved out-of-range, check for
interference sources close to the monitor (bluetooth devices, DECT phones, cellular phones,
microwaves, etc.). If this INOP persists, ask your service personnel to survey the interference sources.
!! Tele INOP Check for further details at the Information Center or in the Telemetry Data window on the monitor.
!!! Tele INOP
Severe yellow/red INOP tone
Tele Sync Unsupp. The MMS in use does not support synchronization of ECG and SpO2 settings between the monitor
INOP tone and central station after a telemetry device has been paired. Use an MMS with revision E.0 or above.
ProtocolWatch INOPs
INOP Message, Indication What to do
PW: Check Settings Contact your service personnel. Settings could not be loaded or interpreted correctly.
PW: In Conflict There is a patient information mismatch which has not yet been resolved (>15 minutes).
PW:Action Required The protocol currently running requires a user response. Check which pop-up window is displayed and
provide the appropriate response.
102
4 Patient Alarms and INOPs
ΔTemp Chk Sources Not all measurements or values required to perform the calculation are available. Check measurement
Numeric is replaced by -?- sources.
ΔTemp Chk Units The monitor has detected a conflict in the units used for this calculation. Check the unit settings.
Numeric is replaced by -?-
103
4 Patient Alarms and INOPs
104
5
Equipment
When the monitor is used together with an IntelliVue Information Center, a variety of services are
provided to manage the interconnections between patients, equipment, hospital beds, caregivers, and
so on.
Many services can be used both at the monitor and at the Information Center, resulting in a more
efficient workflow.
Which services are available will depend on which Information Center you have. The Philips
IntelliVue Information Center iX (PIIC iX) provides more services and functionality than the
Philips IntelliVue Information Center (PIIC). In these Instructions for Use, information that is
only valid for one of the Information Centers will be marked with a corresponding side heading, for
example
PIIC iX For a list of the differences between the two Information Centers, see “Information Center
Compatibility” on page 132.
Patient Concepts
For a patient to be part of the system their identification data must be entered into the system. A
patient can then be assigned to a bed, have equipment assigned directly to him, and have a location set
when he is not currently in the bed.
Equipment Concepts
Depending on how equipment is used in your facility, there are various ways to associate devices with
patients, beds or monitors. Equipment usage must be configured appropriately at the Information
Center and the monitoring equipment.
105
5 Managing Patients and Equipment
Additionally there are mechanisms to automatically free up equipment that is no longer used. This
prevents unnecessary patient mismatches when the equipment is used for the next patient and avoids
data of different patients getting mixed.
PIIC iX When connected to PIIC iX, equipment states are tracked by the system. When the equipment is not
assigned to a patient or no patient is admitted at a monitor, the equipment is considered as free
equipment in the system. Equipment that is assigned or has a patient admitted is tracked as used (i.e.
not free) equipment.
The Information Center uses the equipment state (free/not free) to maintain and support workflow
and equipment lists.
Managing Patients
In order to attribute collected measurement data to a specific patient, or assign equipment or a bed to
a specific patient, each patient in the system must be identified. By admitting a patient, you identify
them for the system.
When equipment is freed up (for example by discharging the patient, removing a monitor from a
patient or using End Case) the collection of data is officially ended for this patient and important
settings on the monitor are reset to the defaults.
WARNING
Always perform a discharge or free up the monitor before starting monitoring for a new patient, even
if your previous patient was not admitted. Failure to do so can lead to data being attributed to the
wrong patient.
Admitting a Patient
The monitor displays physiological data and stores it in the trends as soon as a patient is connected.
This lets you monitor a patient who is not yet admitted. It is however important to admit patients
properly so that you can identify your patient on recordings, reports, and networked devices.
During admission you enter data that the monitor needs for safe and accurate operation. For example,
the patient category setting determines the algorithm the monitor uses to process and calculate some
measurements, the safety limits that apply for some measurements, and the alarm limit ranges.
NOTE
It is strongly recommended that the same patient data fields be configured to be mandatory at the
monitor and the Information Center.
106
5 Managing Patients and Equipment
The Enter Patient Demographics window appears. You can enter data as normal, using the
keyboard or a barcode scanner, or, with PIIC iX, use the Find Patient key to search for patient
data in connected systems as described in “Using "Find Patient" to Search for Patient Data” on
page 107.
3 Enter the patient information: select each field and use the keyboard or barcode scanner or choose
from the pop-up list of alternatives to input information.
PIIC iX When you enter data, the Information Center will automatically check for corresponding data and
will suggest an appropriate action based on what it finds. You can select Confirm to accept the
suggestion or Cancel to reject it.
– Last Name: Enter the patient's last name (family name).
– First Name: Enter the patient's first name.
– Middle Name (if configured to appear): Enter the patient's middle name.
– Lifetime ID, Encounter ID: Whether these fields appear and how they are labeled can be
configured for your hospital. One or both fields may be displayed and the labels may read:
MRN, Case ID, Visit Number, or other alternatives. Enter the appropriate data for the fields
displayed.
– Patient Cat.: Choose the patient category, either Adult, Pedi, or Neo.
– Paced Mode: Choose On or Off (You must use On if your patient has a pacemaker). With
PIIC iX there is a third choice: Unconfirmed. When the paced mode is Unconfirmed, the
algorithm for paced patients will be used.
– Height: Enter the patient's height.
– Weight: Enter the patient's weight.
– BSA: The monitor calculates the body surface area automatically.
– Date of Birth: Enter the patient's date of birth, in the form dd/mmm/yyyy.
– Age: The monitor calculates the patient age automatically.
– Gender: Choose Male or Female.
– Notes (1) / Notes (2): Enter any extra information about the patient or treatment.
When admission is complete, the patient's name appears on the monitor info line at the top of the
screen.
107
5 Managing Patients and Equipment
WARNING
Patient Cat. and Paced Mode will always contain a value, regardless of whether the patient is fully
admitted or not. If you do not specify settings for these fields, the monitor uses the default settings
from the current profile, which might not be correct for your patient.
Patient category
Changing the patient category may change the arrhythmia and NBP alarm limits. If possible, always
load a suitable profile after changing patient category, and always check alarm limits to make sure that
they are appropriate for your patient.
Paced mode
For paced patients, you must set Paced Mode to On. If it is incorrectly set to Off, the monitor could
mistake a pace pulse for a QRS and fail to alarm during asystole. With PIIC iX, the paced mode may be
Unconfirmed which will assume a pacemaker is present. You should still set the Paced Mode to On
for paced patients to make the status clear.
108
5 Managing Patients and Equipment
3 Select Enter.
4 In the confirmation window, select Confirm to stop monitoring for the previous patient or free up
the monitor (if confirmation is configured).
5 Check that patient category and paced status are correct for the new patient.
6 Check the current profile and, if necessary, load an appropriate profile.
If the monitor is connected to an Information Center and only the ID field is entered, the patient name
may be set to - - - at the Information Center, depending on the configuration. Complete the rest of the
demographic details as soon as possible to fully identify the patient on the network, on the monitor
and on printed reports. To complete the details, open the Patient Demographics window and
complete all required fields.
PIIC iX If data has been provided by a hospital information system, you will not be able to edit it on the
monitor.
WARNING
Always end monitoring for the previous patient (with Dischrge Patient, Remove Monitor, End Case
or New Patient) before starting monitoring for a new patient, even if your previous patient was not
admitted. Failure to do so can lead to data being attributed to the wrong patient.
Discharging a Patient
The discharge function is only available when the patient is monitored centrally at an Information
Center. A discharge transfers the patient out of the bed and frees all devices used for the patient. The
discharge function may be disabled at a monitor designated as a transport monitor, to ensure that a
patient cannot be accidentally discharged from the system when the transport monitor is used for
another patient.
A discharge:
– clears the patient demographics
– erases all patient measurement data (such as trend, event, and calculation data) from the
monitor, measurement modules and Information Center. This ensures that data from a
previous patient are not mixed with data from a new patient.
– resets patient category and paced settings to the settings defined in the default Profile
109
5 Managing Patients and Equipment
– resets all monitor and measurement settings as well as the active Screen to the settings defined
in the default Profile
– discharges the patient from the Information Center.
Make sure that you have printed out any required reports before discharging. Check that a functioning
local or central printer is available before you use End Case.
To Discharge a Patient
1 Select the patient name field or select the Patient Demogr. SmartKey to open the Patient
Demographics window and associated pop-up keys.
2 Select the pop-up key for either:
– End Case - to print any configured end case reports or vital signs recording, discharge the
patient and clear the patient database, then enter standby mode. If an End Case SmartKey is
configured for your monitor, you can also select this instead and then confirm.
To see which end case reports are set up for your monitor, select Main Setup, Reports, then
Auto Reports. For each auto report, if End Case Report is set to On, this report will be
printed when you select End Case. See “Setting Up Auto Reports” on page 372 for
information on setting up end case reports.
– Dischrge Patient - to discharge the patient without printing any reports.
To Remove a Monitor
1 Open the Equipment window - either directly in Main Setup or by selecting the bed label in the
information line.
2 Select the monitor.
3 Select the Remove Monitor pop-up key.
110
5 Managing Patients and Equipment
• when no basic vitals (HR, RR, Pulse, SpO2, NBP) have been measured for a specified period
The pop-up window is entitled Is this a New Patient?. The monitor offers a Yes key to stop
monitoring for the previous patient and begin monitoring a new patient and a No key to continue
monitoring with the current patient data and settings.
The time periods for the three conditions can be configured independently.
Transferring Patients
To save you from having to enter the same patient data multiple times and enable patient transfer
without loss of data, information can be shared between Multi-Measurement Modules (MMS), patient
monitors, and Information Centers.
– patient demographic information is shared between connected MMSs, patient monitors, and
Information Centers
– measurement settings and calibration data can be uploaded from an MMS to a patient
monitor, if configured
– trend information can be uploaded from an MMS to a patient monitor, if configured.
Different sets of patient and measurement-related data are stored in the monitor and the Multi-
Measurement Module. Understanding this will help you to understand what happens to patient data
when you transfer patients.
Patient Information Stored in Monitor Stored in X1 MMS and Stored in X2 MMS and
extensions extensions
Patient demographics (name, yes yes yes
DOB, patient IDs)
Monitor settings (alarm yes no no
pause time, alarm volume)
Measurement settings for all yes yes, for all MMS and MMS yes, for all MMS and
measurements (alarm limits, extension measurements MMS extension
measurement on/off, etc.) measurements
Trend data yes, for all MMS, MMS yes, most recent 8 hours of yes, most recent 8 hours
extension and measurement information, for all MMS of information. A
module measurements (up to and extensions standalone X2 stores all
a maximum of 24 or 32 or 50 measurements MMS and MMS extension
or 100, depending on your measurements (up to a
database configuration) maximum of 50
measurements).
During Companion Mode
the X2 stores all MMS and
MMS extension
measurements and
measurements collected
by the host monitor from
measurement modules (up
to a maximum of 50
measurements)
111
5 Managing Patients and Equipment
Patient Information Stored in Monitor Stored in X1 MMS and Stored in X2 MMS and
extensions extensions
Calculation data (HemoCalc yes no no
data)
Events data yes no no
WARNING
If the monitor is not battery-powered, you cannot monitor during transport.
Measurements from an MMS extension connected to an X2 are not available when the X2 is running
on its own battery power. They are only available when the X2 is powered by external power: when
connected to a host monitor, to the external power supply (M8023A) or to the Battery Extension
(865297).
The following sections describe the transfer procedures with a PIIC iX, a PIIC or with a standalone
monitor.
112
5 Managing Patients and Equipment
If the system is configured to support transfer with equipment, the next step is that you move the
patient together with the equipment, e.g. multi-measurement module, to the destination bed you
have selected. If the equipment is still connected to the source bed, the Transfer with Equipment
window opens requesting you to move the patient and equipment, but also offering the possibility
to keep the multi-measurement module at the current bed for use with the next patient (with the
pop-up key No Unplug).
WARNING
Before doing a transfer ensure that no other patient is using the destination bed and/or equipment
from the destination bed.
When a patient is transferred you need to check that the patient has all the required monitoring
equipment at the destination bed.
113
5 Managing Patients and Equipment
5 At the new location, connect the MMS to the monitor. If the monitor detects a patient mismatch,
a window will open showing your patient's data and asking whether to Complete transfer of this
patient?.
6 Select Yes to complete the transfer. This re-admits the patient from the transfer list to the new
monitor. This will upload the patient demographics, and, if configured, the measurement settings
and trend data stored in the MMS to the receiving monitor.
7 Verify that the settings for patient category and paced mode are correct.
WARNING
Measurements from an MMS extension connected to an X2 are not available when the X2 is running
on its own battery power. They are only available when the X2 is powered by external power: when
connected to a host monitor, to the external power supply (M8023A) or to the Battery Extension
(865297).
NOTE
The Transfer key is not available while the X2 is connected to a host monitor (Companion Mode is
indicated).
1 Remove the X2 and any connected extensions from the host monitor.
114
5 Managing Patients and Equipment
115
5 Managing Patients and Equipment
4 If prompted, re-admit the patient to the new monitor: in the Select Patient window of the new
monitor, select the patient in the X2 to retain the data in the X2. This will upload the patient
demographics, and, if configured, the measurement settings and trend data stored in the X2 to the
monitor. Verify that the settings for patient category and paced mode are correct.
WARNING
• It is important to resolve the mismatches as soon as they are identified. Failure to do so could
result in using incorrect/confusing data to make clinical decisions. Certain settings, for example
Paced Mode and Patient Cat., may not match between the Information Center and the monitor.
If the Paced Mode is set incorrectly, the system could mistake a pace pulse for a QRS and fail to
alarm in the case of asystole. It is important that the patient category is set correctly so the ECG
can be analyzed correctly and initial arrhythmia alarm limits set. A Check Patient ID INOP will
appear when a mismatch has not been resolved.
• As long as patient mismatch has not been resolved, data integration in the Information Center and
the own patient overview functionality might not work as equipment is not assigned correctly to
the patient.
116
5 Managing Patients and Equipment
PIIC • When a monitor is connected to an Information Center by the wireless IntelliVue Instrument
Telemetry interface, the patient data will automatically be merged in the case of a transfer. This
means there is no patient discharge at the monitor and settings and trend data will be retained. You
will see a message on the monitor and the Patient Demographics window will automatically
appear so that you can check the data and change it if necessary.
In the case where an X2 with an IntelliVue Instrument Telemetry interface is declared as a "telemetry
device" at the Information Center and is connected to a host monitor, it is important to resolve an
existing mismatch between the monitor and the Information Center before disconnecting the X2.
Failure to do so discharges the X2 and synchronizes the demographics and settings to the
Information Center.
WARNING
After resolving a patient mismatch, check that the monitor settings (especially patient category, paced
status and alarm limits) are correct for the patient.
117
5 Managing Patients and Equipment
1 Patient demographics
2 Patient category and paced status
3 Same Patient - see “Patient Mismatch - If Both Patient Data Sets Refer to the Same Patient” on
page 118.
4 New Patient - see “Patient Mismatch - If Neither Patient Data Set is Correct” on page 118.
After you resolve the mismatch, the monitor displays a confirmation window that shows the patient
that has been selected and where data will be erased, if applicable. Confirm your choice. The monitor
automatically displays the Patient Demographics window after confirmation. Verify that the settings
shown are correct for the patient.
Patient Mismatch - If Both Patient Data Sets Refer to the Same Patient
A patient mismatch where both sets of patient data are correct might occur if you admit a new patient
at the monitor (or Information Center) before the patient arrives at your unit and then connect the
MMS that was used during the patient transport to the monitor.
Select Same Patient if the patient information is different, but you are sure it is the same patient. This
merges the demographics and updates them in the Information Center, monitor, and MMS, according
to this table. Be aware that your monitor may be configured to merge trend data from the MMS and
the monitor, and to upload measurement settings from the MMS to the monitor.
118
5 Managing Patients and Equipment
Managing Equipment
Multiple pieces of monitoring equipment can be used for a patient. The association between a patient
and a device can be made by:
• identifying the patient at the device (by entering patient data or using Find Patient to get patient
data)
• "adding" equipment for an existing patient
The association between a patient and a device can be ended by:
• discharging the patient, or selecting End Case.
• "removing" the device from the patient.
• transferring the patient without the device.
• admitting a new patient to the device.
• automatic freeing of the device.
WARNING
In all cases listed above, when equipment is freed, all patient identification and measurement data are
deleted, all settings are reset to the defaults, and monitoring at the Information Center (if active) is
stopped. Any associated devices, e.g. multi measurement modules connected to a monitor, or cableless
measurement devices assigned to a monitor, may also be freed, depending on the configuration.
119
5 Managing Patients and Equipment
Removing Equipment
The association between a monitor and the patient can be ended by selecting Remove Monitor. With
PIIC iX, the Remove Monitor key is not available when the monitor is locked to a bed.
Other equipment such as telemetry devices or IntelliVue Cableless Measurements can be removed by
selecting the device then selecting Remove in the menu displayed.
120
5 Managing Patients and Equipment
PIIC iX • Select Caregiver - you can select a different caregiver from a list. When a new caregiver is
assigned, the patient will be automatically added to that caregiver's Care Group. The availability of
the function and the caregiver list depends on the Information Center configuration.
• My Patients - opens the My Patients window showing all patients in your Care Group.
121
5 Managing Patients and Equipment
When you select this area in the Equipment window, a menu opens giving access to the Enter Patient
Demographics window, the Transfer function (see “Transferring Patients” on page 111) and the End
Case function.
WARNING
When new equipment has been added for a patient, you will be asked to confirm that the device be
used for the patient. Always do this as soon as possible to avoid patient data mismatch.
If multiple equipment is assigned to the patient, resolving a patient mismatch at the monitor may
resolve other pending patient mismatches or remove other associated equipment.
122
5 Managing Patients and Equipment
WARNING
All data presented in the own patient overview window are delayed for several seconds.
If you need realtime data, for example for defibrillation, always use the host monitor ECG instead of
telemetry or ECG from another monitoring device. As long as the ECG is being measured with
another device there will be no ECG signal available at the ECG analog output.
• If individual measurement alarms are switched off at any of the devices in use for the
patient, a crossed alarm symbol on a white background is shown beside the
measurement numeric
• If all alarms are switched off or paused for one of the devices in use for the patient, a
crossed alarm symbol on a dark gray background is shown beside all affected
measurement numerics (if so configured). If all red and yellow alarms are switched
off/paused, the crossed alarm symbol is red. If only yellow alarms are switched off/
paused, the crossed alarm symbol is yellow.
123
5 Managing Patients and Equipment
source is a patient monitor connected to the Information Center via IntelliVue Instrument Telemetry
(IIT), only the adjust HR alarms setting is available.
WARNING
Even when the other device data is not visible on the screen, you may be silencing monitor alarms and
other device alarms, if the Information Center and monitor are so configured.
Using Standby
When you select Standby mode at the monitor, the bedside goes into Standby mode but other assigned
devices may continue monitoring.
Refer to the Information Center Instructions for Use for details on how selecting Standby at the
Information Center affects the monitor and other devices.
124
5 Managing Patients and Equipment
The ECG measurement will be activated again at the monitor. [Note that in this case, as the screen
switches back to the monitor's own measurements, the SpO2T measurement (if present) will no longer
be displayed].
In the same way the source is tracked when a telemetry device is directly connected to a monitor, then
disconnected and vice versa.
In case of ambiguity, a yellow INOP message !!Check ECG Source indicates that more than one valid
ECG source is active.
Synchronized Settings
For some measurements, settings can be synchronized between the monitor and another measurement
device. For example, if ECG is measured at the monitor, and then the patient is connected to a
telemetry device for monitoring, the Information Center will use the monitor settings for the telemetry
device. In general, the following settings will be synchronized:
PIIC iX With the IntelliVue Information Center iX, the following additional measurement settings can be
synchronized:
125
5 Managing Patients and Equipment
SpO2 SpO2 Alarms on/off, SpO2 Alarm limits, Desat Alarm Limit, NBP Alarm
Suppression On/Off, Pulse(SpO2) On/Off, Measurement Mode1,
Repetition Time1
1
Measurement Mode and Repetition Time can only be synchronized if SpO2 comes from a Cableless SpO2 Pod or a
telemetry device.
WARNING
• Not all settings are synchronized; after changing the measurement source, always check that the
settings are appropriate.
• ECG: Va and Vb leads are reset to default (V2, V5) if the configured Va or Vb lead for the
telemetry device is not one of V1 through V6.
If later the patient is disconnected from the other device, and reconnected to the monitor again, any
changes in the settings made in the meantime will be passed on to the monitor. In this way, settings
continuity is preserved when the measurement source changes.
NOTE
Settings synchronization can be switched off at the monitor in Configuration mode.
In certain situations, you will be asked to confirm that synchronization of settings is appropriate.
Settings synchronization can only take place when there is no patient information mismatch between
the monitor and the Information Center.
PIIC If a Check ECG Settings or !! Check Pairing INOP appears always check that the ECG settings,
especially the paced setting, are appropriate for your patient.
PIIC iX If a Check ECG Settings, Chk SpO₂T Settings, or another check settings INOP appears, always check
that the settings, especially the paced setting for ECG, are appropriate for your patient.
Care Groups
If your monitor is connected to an Information Center, you can group bedside monitors into Care
Groups to facilitate caregiver assignment. This lets you:
• view information on the monitor screen from another bed in the same or in a different Care
Group.
• be notified of yellow or red alarm conditions at the other beds in the Care Group.
• see the alarm status of all the beds in the Care Group on each monitor screen.
The functions available with Care Groups depend on the Information Center revision your monitors
are connected to. See your Information Center Instructions for Use for further details.
126
5 Managing Patients and Equipment
PIIC iX Depending on the Information Center configuration, the Alarm Status overview bar may contain My
Patients (your own care group), My Unit (all beds from the unit) or it may not display at all.
WARNING
With PIIC iX, the alarm status overview bar is independent from the configured alarm pop-up
behavior.
127
5 Managing Patients and Equipment
128
5 Managing Patients and Equipment
PIIC Other Units lets you view a list of all the Care Units in your Care Domain. Select any Care Unit to
view a list of the Information Centers connected to it. Select an Information Center to see a list of
the monitors connected to it. Select any monitor to see the Other Bed window for that bed.
PIIC iX Other Patients lets you search for a patient or step through lists by unit and bed label to select
another patient.
Overview Bar next to the monitor's own bed label . Depending on the configuration, the
automatic pop-up is disabled for five minutes or until you set it to Enabled again.
• To open the Other Bed window, select the required bed label or patient name in the Alarm Status
Overview Bar. If you are in a Unit Group with many beds, the My Patients window may open for
you to select the bed.
129
5 Managing Patients and Equipment
The Other Bed window may be configured to display embedded in a specially designed Screen.
• To display the embedded Other Bed screen element, in the Change Screen menu, select a Screen
designed to show the Other Bed information permanently.
Changing the Screen may automatically change the bed shown in the other bed window. If you switch
to a different Screen, you should check that the correct other bed is displayed.
130
5 Managing Patients and Equipment
WARNING
The Silence Bed pop-up key in the Other Bed window silences alarms at a remote bed. Be aware that
accidental use of this key could silence alarms for the wrong patient.
To silence own bed alarms use the Silence permanent key on screen.
• If individual measurement alarms are switched off at any of the devices in use at the
other bed, a crossed alarm symbol in red on a white background is shown beside the
measurement numeric.
• If all alarms are switched off or paused for one of the devices in use at the other bed,
a crossed alarm symbol on a black background is shown beside all affected
measurement numerics (if so configured). If all red and yellow alarms are switched
off/paused, the crossed alarm symbol is red. If only yellow alarms are switched off/
paused, the crossed alarm symbol is yellow.
• In the embedded Other Bed window, the crossed speaker symbol in the upper right
hand corner indicates that the volume of the audible alarm status change notification
for the other beds in the care group is set to zero at the overview monitor.
131
5 Managing Patients and Equipment
• if configured, the Other Bed window, the Alarming Beds window or the My Patients window may
pop up on the Screen (if automatic alarm notification is enabled at the bedside monitor and at the
Information Center). The automatic alarm notification is suppressed when a window, menu or
pop-up keys are active.
• if configured, an audible status change notification is issued. The tone type and volume can be
configured.
PIIC iX Automatic alarm notification can also cause a window to pop-up when a patient from outside the Care
Group has been escalated to a higher attention status. Check that the detailed configuration at the
PIIC iX meets your unit's requirements.
Automatic alarm notification can be switched off permanently in the monitor's Configuration Mode or
at the Information Center.
132
6
Monitoring
The electrocardiogram (ECG) measures the electrical activity of the heart and displays it on the
monitor as a waveform and a numeric. This section also tells you about arrhythmia monitoring (see
“About Arrhythmia Monitoring” on page 154), ST monitoring (see “About ST Monitoring” on
page 164) and QT monitoring (see “About QT/QTc Interval Monitoring” on page 175).
CAUTION
To protect the monitor from damage during defibrillation, for accurate ECG information and to
protect against noise and other interference, use only ECG electrodes and cables specified by Philips.
Some non-authorized electrodes may be subject to large offset potentials due to polarization.
133
6 ECG, Arrhythmia, ST and QT Monitoring
WARNING
Pace pulse rejection must be switched on for paced patients by setting Paced Mode to On. Switching
pace pulse rejection off for paced patients may result in pace pulses being counted as regular QRS
complexes, which could prevent an asystole event from being detected. When changing profiles, and at
admission/discharge, always check that paced mode is correct for the patient.
Some pace pulses can be difficult to reject. When this happens, the pulses are counted as a QRS
complex, and could result in an incorrect HR and failure to detect cardiac arrest or some arrhythmias.
Make sure that pace pulses are detected correctly by checking the pace pulse markers on the display.
Keep pacemaker patients under close observation.
134
6 ECG, Arrhythmia, ST and QT Monitoring
1 Normal Beats
2 Pace Pulses/Pace Beats
You should choose a lead as primary or secondary lead that has these characteristics:
135
6 ECG, Arrhythmia, ST and QT Monitoring
• the normal QRS complex should be either completely above or below the baseline and it should
not be biphasic. For paced patients, the QRS complexes should be at least twice the height of pace
pulses.
• the QRS complex should be tall and narrow
• the P-waves and the T-waves should be less than 0.2 mV.
For ease of identification on the screen, the pacer spikes can be configured to have a fixed size. They
are then shown in the background as a dotted lines. The length of the dotted line is fixed to the wave
channel height and is independent of the actual pacer amplitude.
In the Setup ECG menu, select Paced Mode to switch between On, Off or Paced Mode On
Unconfirmed.
You can also change the paced mode in the Patient Demographics window.
When Paced Mode is set to On:
– Pace Pulse Rejection is switched on. This means that pacemaker pulses Paced Mode Off
are not counted as extra QRS complexes.
– pace pulse marks are shown on the ECG wave as a small dash (only
when the pacer spikes are not configured to have a fixed size).
– The paced symbol is displayed next to the HR numeric.
When Paced Mode is set to Off, pacer spikes are not shown in the ECG wave. Paced Mode
Be aware that switching pace pulse rejection off for paced patients may result Unconfirmed
in pace pulses being counted as regular QRS complexes, which could prevent
an asystole event from being detected.
When the paced mode has not yet been set for a patient, the paced symbol will
have a question mark indicating the Unconfirmed mode. When the paced
mode is Unconfirmed, pace pulse rejection is switched on.
136
6 ECG, Arrhythmia, ST and QT Monitoring
Changing the adjustment factor only changes the visual appearance of the ECG wave on the screen. It
does not affect the ECG signal analyzed by the monitor.
Comparing the wave size to the 1 mV calibration bar on the ECG wave segment can help you to get an
idea of the true ECG signal strength. If you choose a fixed adjustment factor, the 1 mV calibration bar
will be the same size for all the displayed ECG waves. If you choose Auto Size, or an individual size
using the Size Up/Size Down keys, the calibration bar may be a different size for each wave.
137
6 ECG, Arrhythmia, ST and QT Monitoring
– Filter: The filter reduces interference to the signal. It should be used if the signal is distorted
by high frequency or low frequency interference. High frequency interference usually results in
large amplitude spikes making the ECG signal look irregular. Low frequency interference
usually leads to a wandering or rough baseline. In the operating room, the Filter reduces
artifacts and interference from electrosurgical units. Under normal measurement conditions,
selecting Filter may suppress the QRS complexes too much and thus interfere with the clinical
evaluation of the ECG displayed on the monitor. This does not affect the ECG analysis
performed by the monitor.
If AutoFilter is set to On in Configuration Mode, the filter setting will automatically be set to
Filter if electromagnetic interference is detected.
– Diag: Use when diagnostic quality is required. The unfiltered ECG wave is displayed so that
changes such as R-wave notching or discrete elevation or depression of the ST segments are
visible.
The setting Diag selects the highest available ECG bandwidth which is 0.05 to 150 Hz for the
Adult, Pedi and Neo patient category. The term "diagnostic" relates only to the ECG
bandwidth requirements for diagnostic electrocardiographic devices as outlined in the ANSI/
AAMI standard EC11-1991.
138
6 ECG, Arrhythmia, ST and QT Monitoring
Especially in the precordial leads, which are close to the heart, QRS morphology can be greatly altered
if an electrode is moved away from its correct location.
139
6 ECG, Arrhythmia, ST and QT Monitoring
140
6 ECG, Arrhythmia, ST and QT Monitoring
6-Lead Placement
For a 6-lead placement use the positions from the 5-lead diagram above but with two chest leads. The
two chest leads, Va and Vb, can be positioned at any two of the V1 to V6 positions shown in the chest
electrode diagram below. The Va and Vb lead positions chosen must be selected in the Setup ECG
Menu to ensure correct labeling.
141
6 ECG, Arrhythmia, ST and QT Monitoring
For accurate chest electrode placement and measurement, it is important to locate the fourth
intercostal space.
1 Locate the second intercostal space by first palpating the Angle of Lewis (the little bony
protuberance where the body of the sternum joins the manubrium). This rise in the sternum is
where the second rib is attached, and the space just below this is the second intercostal space.
2 Palpate and count down the chest until you locate the fourth intercostal space.
10-Lead Placement
When monitoring 12-leads of ECG, using a 10-Electrode Lead Placement, it is important to correctly
place electrodes and to label all 12-lead ECG reports with the correct lead placement.
142
6 ECG, Arrhythmia, ST and QT Monitoring
1 V1 - V6
2 LA
3 RA
4 RL
5 LL
In conventional 12-Lead ECG using 10 electrodes, an electrode is placed on the right arm, left arm,
right leg, and left leg. Six V- electrodes are placed on the chest. The right leg electrode is the reference
electrode.
Limb electrodes:
– Place arm electrodes on the inside of each arm, between the wrist and the elbow.
– Place leg electrodes inside of each calf, between the knee and the ankle.
Chest electrodes:
V1 - on the 4th intercostal space at the right sternal border
V2 - on the 4th intercostal space at the left sternal border
V3 - midway between the V2 and V4 electrode positions
V4 - on the 5th intercostal space at the left midclavicular line
V5 - on the left anterior axillary line, horizontal with the V4 electrode position
V6 - on the left midaxillary line, horizontal with the V4 electrode position
143
6 ECG, Arrhythmia, ST and QT Monitoring
1 LA
2 V1-V6
3 LL
4 RL
5 RA
6 Angle of Lewis
If your institution uses modified 10-lead ECG electrode placement (the Mason-Likar Lead System),
place the four limb electrodes close to the shoulders and lower abdomen.
The six V electrodes are placed on the chest in the same position as the conventional 12-lead
placement.
WARNING
Do not use ECG analysis interpretation statements and measurements for 12-lead ECGs obtained
using the modified (Mason-Likar) limb electrode placement. This may lead to misdiagnosis since the
modified (Mason-Likar) limb electrode placement does not look the same as the conventional 12-lead
ECG and may mask inferior infarction due to calculated axis, R, P and T wave magnitudes shifts and
ST slope.
Do not export 12-lead ECGs obtained using the modified (Mason-Likar) limb electrode placement.
Captured 12-Lead ECGs using the modified (Mason-Likar) limb electrode placement exported from
the Information Center are not annotated with the Mason-Likar label.
144
6 ECG, Arrhythmia, ST and QT Monitoring
WARNING
EASI-derived 12-lead ECGs and their measurements are approximations to conventional 12-lead
ECGs. As the 12-lead ECG derived with EASI is not exactly identical to the 12-lead conventional
ECG obtained from an electrocardiograph, it should not be used for diagnostic interpretations.
Respiratory monitoring is also possible with the EASI placement; respiration is measured between the
I and A electrodes.
Place the electrodes as accurately as possible to obtain the best quality EASI measurements.
When EASI lead placement is selected, EASI is shown beside the 1 mV calibration bar on the ECG
wave on the display, and EASI is marked on any recorder strips and printouts.
EASI Monitoring During INOP Conditions
If one of the derived EASI leads has an INOP condition (for example, Lead Off), a flat line is
displayed. After 10 seconds, the directly acquired EASI AI, AS, or ES lead (depending on which is
available) is displayed with the corresponding lead label. This causes an arrhythmia relearn.
145
6 ECG, Arrhythmia, ST and QT Monitoring
Capture 12-Lead
You can view a 12-Lead ECG on the screen, capture a 12-Lead ECG episode, preview the captured
ECG data and then store it and send it to a connected Information Center for analysis. You can
download the resulting analysis from the Information Center and also other 12-Lead captures for
review at the monitor. A 12-Lead report can be printed that also includes the downloaded analysis
results and an ST Map. You can operate the 12-Lead Export function and the 12-Lead Lock/Unlock
function at the Information Center remotely from the monitor.
The monitor can store one 12-Lead ECG episode at any time.
146
6 ECG, Arrhythmia, ST and QT Monitoring
147
6 ECG, Arrhythmia, ST and QT Monitoring
148
6 ECG, Arrhythmia, ST and QT Monitoring
149
6 ECG, Arrhythmia, ST and QT Monitoring
150
6 ECG, Arrhythmia, ST and QT Monitoring
151
6 ECG, Arrhythmia, ST and QT Monitoring
You need to know which value has been configured for your monitor. Changing the high and low
alarm limits automatically changes the extreme alarm limits within the allowed range.
To see the extreme rate alarms set for your monitor, in the Setup ECG menu, see the menu items Δ
ExtrTachy and Δ ExtrBrady.
WARNING
Defibrillation and Electrosurgery:
Do not touch the patient, or table, or instruments, during defibrillation.
After defibrillation, the screen display recovers within 10 seconds if the correct electrodes are used and
applied in accordance with the manufacturers instructions.
ECG cables can be damaged when connected to a patient during defibrillation. Check cables for
functionality before using them again.
According to AAMI specifications the peak of the synchronized defibrillator discharge should be
delivered within 60 ms of the peak of the R wave. The signal at the ECG output on the IntelliVue
patient monitors is delayed by a maximum of 30 ms. Your biomedical engineer should verify that your
ECG/Defibrillator combination does not exceed the recommended maximum delay of 60 ms.
152
6 ECG, Arrhythmia, ST and QT Monitoring
When using electrosurgical (ES) equipment, never place ECG electrodes near to the grounding plate of
the ES device, as this can cause a lot of interference on the ECG signal.
General:
When you are connecting the electrodes or the patient cable, make sure that the connectors never
come into contact with other conductive parts, or with earth. In particular, make sure that all of the
ECG electrodes are attached to the patient, to prevent them from contacting conductive parts or earth.
During surgery:
Use the appropriate orange electrode ECG safety cable, or lead cable with an orange connector, for
measuring ECG in the operating room. These cables have extra circuitry to protect the patient from
burns during cautery, and they decrease electrical interference. This also reduces the hazard of burns in
case of a defective neutral electrode at the HF device. These cables cannot be used for measuring
respiration.
Pacemaker failure:
During complete heart block or pacemaker failure to pace/capture, tall P-waves (greater than 1/5 of
the average R-wave height) may be erroneously counted by the monitor, resulting in missed detection
of cardiac arrest.
Patients exhibiting intrinsic rhythm:
When monitoring paced patients who exhibit only intrinsic rhythm, the monitor may erroneously
count pace pulses as QRS complexes when the algorithm first encounters them, resulting in missed
detection of cardiac arrest.
The risk of missing cardiac arrest may be reduced by monitoring these patients with low heart rate limit
at or slightly above the basic/demand pacemaker rate. A low heart rate alarm alerts you when the
patient's heart rate drops to a level where pacing is needed. Proper detection and classification of the
paced rhythm can then be determined.
Filtered ECG signal from external instruments:
Instruments such as defibrillators or telemetry units produce a filtered ECG signal. When this signal is
used as an input to the bedside monitor, it is filtered again. If this twice-filtered signal is passed to the
arrhythmia algorithm, it may cause the algorithm to fail to detect pace pulses, pacemaker non-capture,
or asystole, thus compromising paced patient monitoring performance.
External pacing electrodes:
When a pacemaker with external pacing electrodes is being used on a patient, arrhythmia monitoring is
severely compromised due to the high energy level in the pacer pulse. This may result in the arrhythmia
algorithm's failure to detect pacemaker noncapture or asystole.
Fusion beat pacemakers:
Pacemakers that create fusion beats (pace pulse on top of the QRS complex) cannot be detected by the
monitor's QRS detector.
Rate adaptive pacemakers:
Implanted pacemakers which can adapt to the Minute Ventilation rate may occasionally react on the
Impedance measurement used by patient monitors for the determination of the Resp value and
execute pacing with the maximum programmed rate. Switching off the Resp measurement can prevent
this.
153
6 ECG, Arrhythmia, ST and QT Monitoring
Arrhythmia Options
Your monitor has either the basic or the enhanced arrhythmia option. Both options provide rhythm
and ectopic status messages and beat labeling. The number of rhythms being classified, events being
detected, and alarms generated differs according to the option. The alarms available with the different
options are listed in the section “ECG and Arrhythmia Alarm Overview” on page 149, the rhythm and
ectopic messages detected are listed in “Arrhythmia Status Messages” on page 157.
154
6 ECG, Arrhythmia, ST and QT Monitoring
155
6 ECG, Arrhythmia, ST and QT Monitoring
Aberrantly-Conducted Beats
As P-waves are not analyzed, it is difficult and sometimes impossible for the monitor to distinguish
between an aberrantly-conducted supraventricular beat and a ventricular beat. If the aberrant beat
resembles a ventricular beat, it is classified as ventricular. You should always select a lead where the
aberrantly-conducted beats have an R-wave that is as narrow as possible to minimize incorrect calls.
Ventricular beats should look different from these 'normal beats'. Instead of trying to select two leads
with a narrow R-wave, it may be easier to just select one lead and use single lead arrhythmia
monitoring. Extra vigilance is required by the clinician for this type of patient.
1 Beat label
2 Pace pulse marks
3 Rhythm status message
4 PVC Numeric
5 HR Numeric
6 Ectopic status message
7 Delayed arrhythmia wave
156
6 ECG, Arrhythmia, ST and QT Monitoring
157
6 ECG, Arrhythmia, ST and QT Monitoring
158
6 ECG, Arrhythmia, ST and QT Monitoring
Arrhythmia Relearning
During a learning phase:
• Alarm timeout periods are cleared
• Stored arrhythmia templates are cleared
• Asystole, Vfib, and HR alarms (when there are enough beats to compute the HR) are active. No
other alarms are active.
159
6 ECG, Arrhythmia, ST and QT Monitoring
WARNING
If arrhythmia learning takes place during ventricular rhythm, the ectopics may be incorrectly learned as
the normal QRS complex. This may result in missed detection of subsequent events of V-Tach and V-
Fib.
For this reason you should:
• take care to initiate arrhythmia relearning only during periods of predominantly normal rhythm
and when the ECG signal is relatively noise-free
• be aware that arrhythmia relearning can happen automatically
• respond to any INOP messages (for example, if you are prompted to reconnect electrodes)
• be aware that a disconnected EASI electrode triggers an arrhythmia relearn on all leads
• always ensure that the arrhythmia algorithm is labeling beats correctly.
Arrhythmia Alarms
Arrhythmia alarms can be switched on and off and the alarm settings changed just like other
measurement alarms, as described in the Alarms section. Special alarm features which apply only to
arrhythmia are described here.
The different alarms detected and generated by the monitor depend on the level of arrhythmia analysis
that is enabled. For a complete list of arrhythmia alarms and INOPs, see the “Alarms” chapter.
The monitor detects arrhythmia alarm conditions by comparing ECG data to a set of pre-defined
criteria. An alarm can be triggered by a rate exceeding a threshold (for example, HR >xx), an abnormal
rhythm (for example, Ventricular Bigeminy), or an ectopic event (for example, Pair PVCs).
WARNING
When arrhythmia analysis is on, all yellow ECG and arrhythmia alarms are short yellow alarms (one-
star). This means that the yellow alarm lamp and the tones are active for six seconds only, after which
the blinking numeric and the alarm message remain for up to three minutes. The only exception to this
are the HR High and Low alarms which can be configured as standard yellow alarms. Red alarms
behave as usual.
160
6 ECG, Arrhythmia, ST and QT Monitoring
161
6 ECG, Arrhythmia, ST and QT Monitoring
If you silence a yellow arrhythmia alarm and the alarm condition still exists, the visual indicators
continue until the condition stops. You will get an alarm reminder every time the configured timeout
period has expired.
If you silence a yellow arrhythmia alarm and the alarm condition has stopped, the visual
indicators are immediately cleared. Silencing an alarm does not reset its timeout period, so you will not
get a realarm for the same condition or lower on the chain until the timeout expires.
162
6 ECG, Arrhythmia, ST and QT Monitoring
If alarm conditions of equal severity from different chains are detected, the alarm condition that
occurred most recently is announced.
See “ECG and Arrhythmia Alarm Overview” on page 149 for information on which alarms are
included in the different arrhythmia options. See “Arrhythmia Alarm Timeout Periods” on page 161
for an explanation of how alarm timeouts work.
163
6 ECG, Arrhythmia, ST and QT Monitoring
About ST Monitoring
The monitor performs ST segment analysis on normal and atrially paced beats and calculates ST
segment elevations and depressions. This information can be displayed in the form of ST numerics and
snippets on the monitor.
The monitor also performs ST Elevation (STE) analysis using automated ISO and J point
determination and measuring the ST segment directly at the J point (J +0). This is based on the
recommendations for measuring ST Elevation published by the American Heart Association, the
American College of Cardiology and the European Society of Cardiology.
All available leads can be monitored continuously. The ECG waveform does not need to be displayed
on the monitor for ST Segment analysis.
ST analysis is always performed using a dedicated filter which ensures diagnostic quality. If you are
monitoring ECG using an ECG filter mode other than Diagnostic, the ST segment of the ECG wave
may look different from the ST segment of the ST snippet for the same wave. For diagnostic
evaluation of the ST segment, always switch to Diagnostic filter mode or use the ST snippet.
164
6 ECG, Arrhythmia, ST and QT Monitoring
WARNING
Some clinical conditions may make it difficult to achieve reliable ST monitoring, for example:
• if you are unable to get a lead that is not noisy
• if arrhythmias such as atrial fib/flutter are present, which may cause an irregular baseline
• if the patient is continuously ventricularly paced
• if the patient has left bundle branch block.
You should consider switching ST monitoring off if these conditions are present.
This monitor provides ST level change information; the clinical significance of the ST level change
information should be determined by a physician.
ST segment monitoring is intended for use with adult patients only and is not clinically validated for
use with neonatal and pediatric patients. For this reason, the recommended - and default - setting for
ST monitoring in neonatal and pediatric modes is ST Analysis: Off.
NOTE
The order in which ST leads are listed in the Setup ST Leads menu determines the order in which ST
leads are displayed on the monitor screen.
165
6 ECG, Arrhythmia, ST and QT Monitoring
1 ST numerics
2 Current HR alarm limits
3 Current heart rate
ST Numerics
Up to 12 ST numerics plus the ST index can be displayed on the monitor screen. They can be
configured to show beside the measurement numerics, beside the ECG wave, or beside the ST snippet.
A positive ST value indicates ST segment elevation; a negative value indicates depression.
ST numerics are displayed in the order in which you select ST leads for analysis. If there is additional
space in the field assigned to ST numerics, the monitor will display extra numerics in the order in
which they appear in the list in Setup ST Leads, in the Setup ST Analysis menu. Any ST leads
switched on for analysis that do not fit in the assigned numerics field are shown in succession in place
of the last ST numeric.
ST Index
The ST index numeric (STindx) is the sum of the absolute values for the ST leads V2, V5, aVF.
Because it is based on absolute values, it is always a positive number. If you haven't selected one of the
leads V2, V5, and aVF for ST analysis, the ST index numeric will display a question mark "?".
To switch the ST index numeric on or off for display, in the Setup ST Analysis menu, select ST-Index
to toggle between On and Off.
ST Snippets
ST snippets show a one second wave segment for each measured ST lead. The most recent snippet is
drawn in the same color as the ECG wave, usually green, superimposed over the stored baseline
snippet, drawn in a different color. The comparison shows any deviation in the measurement since the
baseline snippet was stored, for example as a result of a procedure carried out on the patient. The
information is updated once per minute.
If you do not see ST snippets on the Screen, select the Screen name in the Monitor Info Line and
select a Screen configured to show snippets from the pop-up list of available Screens.
166
6 ECG, Arrhythmia, ST and QT Monitoring
ST View Window
The ST View Window shows a current ST snippet and numeric with a baseline snippet and numeric.
The two snippets are in different colors, so that you can differentiate between them easily and see at a
glance which numerics belong to which snippet. In the upper left corner the current lead and the ISO/
J-Point mode for the current snippet and the baseline snippet is shown. On the right side all available
ST numerics are shown. The date and time of the baseline snippet is shown at the bottom of the
window.
The ST View Window opens with the ST pop-up keys Current / Baseline / Overlap (to select viewing
mode), Show Points /Hide Points, Update Baseline, Record ST, STE View, Adjust ST Points, ST
Map, Setup ST and arrow keys for scrolling through the available leads.
To view the ST View window, select any snippet on the Screen.
1 Current Lead
2 ISO/J-point mode for current snippet
3 ISO/J-point mode for baseline snippet
4 1 mV calibration bar
5 Current ST snippet and ST baseline snippet overlapped
6 Timestamp of most recently stored baseline snippet
7 All available ST numerics for current snippet and baseline snippet
The ST point, J-point and ISO point can be hidden using the Hide Points pop-up key.
STE View Window
The STE View window shows a current ST snippet and STE numeric with an ST baseline snippet. The
two snippets are in different colors, so that you can differentiate between them easily. In the upper left
corner the current lead is shown. On the right side all available STE numerics are shown. The date and
time of the ST baseline snippet is shown at the bottom of the window.
167
6 ECG, Arrhythmia, ST and QT Monitoring
The STE View window opens with the pop-up keys Current / Baseline / Overlap (to select viewing
mode), Show Points/Hide Points, Record STE, ST View, ST Map, Setup ST and arrow keys for
scrolling through the available leads.
The ST point, J-point and ISO point can be hidden using the Hide Points pop-up key.
Recording ST Segments
To record all currently available ST snippets and baselines, in the ST View window, select the pop-up
key Record ST.
168
6 ECG, Arrhythmia, ST and QT Monitoring
The isoelectric (ISO) point provides the baseline, the ST point is at the midpoint of the ST segment.
The J point is where the QRS complex changes its slope; as it is a fixed distance away from the ST
point, it can be useful to help you position the ST point correctly.
CAUTION
If using ST analysis, the ST measurement points need to be adjusted when you start monitoring, and if
the patient's heart rate or ECG morphology changes significantly, as this may affect the size of the QT
interval and thus the placement of the ST point. Artifactual ST segment depression or elevation may
occur if the isoelectric point or the ST point is incorrectly set.
Always ensure that ST measurement points are appropriate for your patient.
169
6 ECG, Arrhythmia, ST and QT Monitoring
3 Use the Select Point pop-up key to scroll through the points and activate the point you need to
adjust, then use the left and right arrow keys to move the measurement point. Each point is
highlighted while active.
1 ST label
2 Current ISO/J-point detection mode
3 1 mV calibration bar
4 Cursor for adjusting ISO points
5 Highlighted ISO point
6 All available ST numerics for current snippet
The ISO-point cursor (1) positions the isoelectric point relative to the R-wave peak. The relation is
shown beside the ISO-point in milliseconds. Position the ISO-point in the middle of the flattest
part of the baseline (between the P and Q waves or in front of the P wave).
The J-point cursor (2) positions the J-point relative to the R-wave peak. It helps you to correctly
position the ST-point. Position the J-point at the end of the QRS complex and the beginning of
the ST segment.
The J-point cursor is not available if your monitor is configured to let you set the ST point directly.
170
6 ECG, Arrhythmia, ST and QT Monitoring
ST Alarms
ST alarms are yellow alarms. Each ST lead has its own alarm limit. ST alarms are triggered when an ST
value exceeds its alarm limit for more than one minute. Switching ST alarms off switches off alarms for
all ST leads.
If more than one ST measurement is in alarm, the monitor only displays the alarm message of the ST
lead which is currently furthest from its set alarm limits.
STE Alarms
The STE alarm is a yellow alarm. It can be switched on and off in the Setup ST Elevation menu but
the limits can only be changed in Configuration mode. The STE alarm limits are gender specific and
can be set individually for limb leads, V2/V3 leads, and V1/V4/V5/V6 leads. The default values, for
example on V2 and V3 1.5 mm for females and 2.0 mm for males, are based on the recommendations
from the American Heart Association and American College of Cardiology.
The ST Elevation measurements with automated J-point determination generate ST Elevation alarms,
in addition to the ST measurements at the user-defined ST point (J+offset), which may be useful for
ST depression alarms. When ST and STE analysis are both in use, this may result in redundant alarms
for ST elevations. Because of the different measurement points, there may be different values
obtained. Thus there could be an ST alarm and an STE alarm but the STE alarm may announce sooner
based upon the values obtained.
171
6 ECG, Arrhythmia, ST and QT Monitoring
Viewing ST Maps
The monitor can derive a multi-axis portrait (map) from the ST analysis to help you detect changes in
ST values. It displays two planes obtained from a multilead ECG in a multi-axis diagram, where each
axis represents a lead. The ST value at the J point is given. The position of the axes within the diagram
correspond to the placement of the ECG leads. Each ST value is assigned to either a limb lead, or to a
chest lead. Every axis shows the polarity of the lead it represents. By joining adjacent ST values, the
monitor obtains the ST map. The contour line, and the map shading, is shown in the same color as the
ECG parameter.
Current View
In current view, the monitor displays an ST map that corresponds to the current ST values. Three or
more leads per plane are necessary to display a map.
The left of the following diagram shows leads I, II, III, aVR, aVL, and aVF on the limb leads. On the
right, the V-leads (V1, V2, V3, V4, V5, and V6) are on the chest leads.
172
6 ECG, Arrhythmia, ST and QT Monitoring
If a lead is in INOP (the value is being measured but is invalid or unavailable because, for example, the
corresponding ECG electrode is off), the area formed by the remaining ST leads is left open.
If there is insufficient information (for example, there are less than three chest leads) for a second ST
map to be displayed, the currently available ST values are displayed in place of the second ST map.
Trend View
In trend view, you can see up to four trended ST maps, and the current ST map, simultaneously. You
can configure the time interval between trended samples. The most recent map is shown in the same
color as the parameter itself. Past values change from white through dark gray. In the diagram below,
the time interval between trends is one minute. The first trended sample is white and is one minute old.
The second trended sample corresponds to the ST values two minutes ago and so forth. The ST values
on the diagrams show the current ST values.
If a lead is turned off, its axis is no longer shown. This has no impact on the presentation of trended
values that were recorded while the lead was still on.
1 Limb Leads
2 Chest Leads
3 current ST values
4 trending interval
173
6 ECG, Arrhythmia, ST and QT Monitoring
Viewing an ST Map
To display an ST map,
• In the Main Setup menu, select ST Map.
174
6 ECG, Arrhythmia, ST and QT Monitoring
The QT interval (1) has an inverse relationship to heart rate. Faster heart rates shorten the QT interval
and slower heart rates prolong the QT interval. Therefore there are several formulas used to correct
the QT interval for heart rate. The heart rate corrected QT interval is abbreviated as QTc. The monitor
uses as a default the Bazett correction formula and the alternative Fridericia formula can be selected in
Configuration Mode.
For QT interval monitoring to be effective, basic or enhanced arrhythmia monitoring should be turned
on.
QT Measurement Algorithm
The QT values are updated every five minutes except in the initial phase (first five minutes) where they
are updated once per minute. Normal or atrial paced beats and beats with a similar morphology are
averaged to form a representative waveform for further processing. Normal beats followed by a
premature QRS will be excluded from the measurements to prevent the premature beat from
obscuring the end of the T-wave. If the algorithm cannot form a representative waveform, for example
because the morphology of the beats is too varied, a Cannot Analyze QT INOP will be generated after
10 minutes. This is also the case if normal beats have been falsely labeled so that the algorithm does
not have enough valid beats to make QT measurements. No QT value is calculated if the QT-HR is
>150 bpm (Adult) or >180 bpm (Pedi/Neo).
175
6 ECG, Arrhythmia, ST and QT Monitoring
Because of the different algorithm approaches, a QT/QTc measurement from a diagnostic 12-lead
program may differ from the realtime measurement on the monitor.
WARNING
QT/QTc measurements should always be verified by a qualified clinician.
176
6 ECG, Arrhythmia, ST and QT Monitoring
• Single-Lead mode - a single lead selected from all available leads (except the augmented leads) will
be used for QT measurement. QT measurement will stop if the selected lead becomes unavailable.
To select the mode,
1 Select the QT numeric to enter the Setup QT Analysis window.
2 Select QT Lead and select All, Primary Lead or one of the available single leads.
When using the All Leads mode, make sure when you compare QT values that they are based on the
same set of leads.
Changing the lead(s) used for QT measurements will not cause the baseline to be reset.
QT View
In the QT View window you can verify that the QT algorithm detects correct Q and T points.
The current waves are shown in the upper half of the window and the baseline waves in a different
color below. The Q and T points are marked with a vertical line. By selecting one of the lead labels at
the top of the window you can highlight the corresponding wave; the other waves are shown in gray.
The underlined lead labels are the leads used for the QT calculation. By selecting the numeric area you
can highlight all underlined leads.
177
6 ECG, Arrhythmia, ST and QT Monitoring
QT Alarms
There are two QT alarms, QTc high limit alarm and ΔQTc high alarm. The QTc high limit alarm is
generated when the QTc value exceeds the set limit for more than 5 minutes. The ΔQTc alarm is
generated when the difference between the current value and the baseline value exceeds the set limit
for more than 5 minutes.
The Cannot Analyze QT INOP and the -?- will be displayed when no QT measurement could be
calculated for 10 minutes. Up to this time the previous valid value will be displayed. The following
additional messages on the cause of the invalid measurements may also be displayed.
178
6 ECG, Arrhythmia, ST and QT Monitoring
179
6 ECG, Arrhythmia, ST and QT Monitoring
180
7
181
7 Monitoring Pulse Rate
182
7 Monitoring Pulse Rate
WARNING
Selecting Pulse as the active alarm source for HR/Pulse switches off the arrhythmia alarms listed in the
section “ECG and Arrhythmia Alarm Overview” on page 149, including Asystole, Vfib and Vtach
alarms, and the heart rate alarms. This is indicated by the message ECG/Arrh AlarmsOff (unless this
has been configured off for your monitor), and the crossed-out alarm symbol beside the ECG heart
rate numeric. The message ECG/Arrh AlarmsOff can be configured off, or to switch to a yellow
(medium severity) INOP after a fixed number of hours.
High and low pulse rate and extreme bradycardia and extreme tachycardia alarms from pulse are active.
QRS Tone
The active alarm source is also used as a source for the QRS tone. You can change the tone volume in
the Setup SpO₂ and Setup ECG menus and the QRS tone modulation in the Setup SpO₂ menu.
WARNING
The audible QRS tone might be influenced by external interference and is not intended to be used as a
substitute for ECG based arrhythmia analysis.
If arrhythmia detection is needed, do not rely on the audible QRS tone.
183
7 Monitoring Pulse Rate
184
8
(Resp)
For the respiratory measurement (Resp), the monitor measures the thoracic impedance between two
ECG electrodes on the patient's chest. Changes in the impedance due to thoracic movement produce
the Resp waveform on the monitor screen. The monitor counts the waveform cycles to calculate the
respiration rate (RR).
Cardiac Overlay
Cardiac activity that affects the Resp waveform is called cardiac overlay. It happens when the Resp
electrodes pick up impedance changes caused by the rhythmic blood flow. Correct electrode
placement can help to reduce cardiac overlay: avoid the liver area and the ventricles of the heart in the
line between the respiratory electrodes. This is particularly important for neonates.
185
8 Monitoring Respiration Rate (Resp)
Some patients, especially neonates, expand their chests laterally. In these cases it is best to place the
two respiratory electrodes in the right midaxillary and left lateral chest areas at the patient's maximum
point of breathing movement to optimize the respiratory wave.
Abdominal Breathing
Some patients with restricted chest movement breathe mainly abdominally. In these cases, you may
need to place the left leg electrode on the left abdomen at the point of maximum abdominal expansion
to optimize the respiratory wave.
186
8 Monitoring Respiration Rate (Resp)
• In the Setup Resp menu, select Size Up to increase the size of the wave or Size Down to decrease
it.
187
8 Monitoring Respiration Rate (Resp)
188
8 Monitoring Respiration Rate (Resp)
This is because of the higher internal impedance of the OR cable set, required for use if electrosurgery
is being performed.
Rate adaptive pacemakers
Implanted pacemakers which can adapt to the Minute Ventilation rate may occasionally react on the
Impedance measurement used by patient monitors for the determination of the Resp value and
execute pacing with the maximum programmed rate. Switching off the Resp measurement can prevent
this.
189
8 Monitoring Respiration Rate (Resp)
190
9
9 Monitoring SpO2
Philips pulse oximetry uses a motion-tolerant signal processing algorithm, based on Fourier artifact
suppression technology (FAST). It provides four measurements:
• Oxygen saturation of arterial blood (SpO2) - percentage of oxygenated hemoglobin in relation to
the sum of oxyhemoglobin and deoxyhemoglobin (functional arterial oxygen saturation).
• Pleth waveform - visual indication of patient's pulse.
• Pulse rate (derived from pleth wave) - detected pulsations per minute.
• Perfusion indicator - numerical value for the pulsatile portion of the measured signal caused by
arterial pulsation.
The monitors are also compatible with SpO2 technologies from other manufacturers. Refer to the
instructions for use provided with these devices for further information.
SpO2 Sensors
Depending on the purchased SpO2 option, different sensors and adapter cables can be used. The
sensors for the different options are color-coded to match the connectors. See the “Accessories”
chapter for a compatibility table.
Familiarize yourself with the instructions for use supplied with your sensor before using it. In
particular, check that the sensor being used is appropriate for your patient category and application
site.
CAUTION
Do not use OxiCliq disposable sensors in a high humidity environment, such as in neonatal
incubators or in the presence of fluids, which may contaminate sensor and electrical connections
causing unreliable or intermittent measurements. Do not use disposable sensors when there is a
known allergic reaction to the adhesive.
Always use the MAX-FAST forehead sensor with the foam headband provided by Nellcor.
191
9 Monitoring SpO2
3 Apply the sensor to the patient. The application site should match the sensor size so that the
sensor can neither fall off, nor apply excessive pressure.
When using the M1195A Infant Finger Sensor, select a finger or toe with a diameter of between 7
and 8 mm (0.27" and 0.31"). When applying a M1193A neonatal sensor do not overtighten the
strap.
4 Check that the light emitter and the photodetector are directly opposite each other. All light from
the emitter must pass through the patient's tissue.
WARNING
Proper Sensor Fit: If a sensor is too loose, it might compromise the optical alignment or fall off. If it
is too tight, for example because the application site is too large or becomes too large due to edema,
excessive pressure may be applied. This can result in venous congestion distal from the application site,
leading to interstitial edema, hypoxemia and tissue malnutrition. Skin irritations or lacerations may
occur as a result of the sensor being attached to one location for too long. To avoid skin irritations and
lacerations, periodically inspect the sensor application site and change the application site regularly.
Venous Pulsation: Do not apply sensor too tightly as this results in venous pulsation which may
severely obstruct circulation and lead to inaccurate measurements.
Ambient Temperature: At elevated ambient temperatures be careful with measurement sites that are
not well perfused, because this can cause severe burns after prolonged application. All listed sensors
operate without risk of exceeding 41°C on the skin if the initial skin temperature does not exceed
35°C.
Extremities to Avoid: Avoid placing the sensor on extremities with an arterial catheter, an NBP cuff
or an intravascular venous infusion line.
CAUTION
Extension cables: Do not use more than one extension cable (M1941A). Do not use an extension
cable with Philips reusable sensors or adapter cables with part numbers ending in -L (indicates "long"
cable version).
Electrical Interference: Position the sensor cable and connector away from power cables, to avoid
electrical interference.
Humidity: For neonatal patients, make sure that all sensor connectors and adapter cable connectors
are outside the incubator. The humid atmosphere inside can cause inaccurate measurements.
Measuring SpO2
1 Select the correct patient category setting (adult/pediatric and neonatal), as this is used to optimize
the calculation of the SpO2 and pulse numerics.
192
9 Monitoring SpO2
WARNING
• For fully conscious pediatric or adult patients, who have a normal function of perfusion and
sensory perception at the measurement site:
To ensure skin quality and correct optical alignment of the sensor, inspect the application site
when the measurement results are suspicious or when the patient complains about pressure at
the application site, but at least every 24 hours. Correct the sensor alignment if necessary.
Move the sensor to another site, if the skin quality changes.
• For all other patients:
Inspect the application site every two to three hours to ensure skin quality and correct optical
alignment. Correct the sensor alignment if necessary. If the skin quality changes, move the
sensor to another site.
Change the application site at least every four hours.
• Injected dyes such as methylene blue, or intravascular dyshemoglobins such as methemoglobin
and carboxyhemoglobin may lead to inaccurate measurements.
• Inaccurate measurements may result when the application site for the sensor is deeply pigmented
or deeply colored, for example, with nail polish, artificial nails, dye or pigmented cream.
• Interference can be caused by:
– High levels of ambient light (including IR warmers) or strobe lights or flashing lights (such as
fire alarm lamps). (Hint: cover application site with opaque material.)
– Another SpO2 sensor in close proximity (e.g. when more than one SpO2 measurement is
performed on the same patient). Always cover both sensors with opaque material to reduce
cross-interference.
– Electromagnetic interference, especially at perfusion indicator values below 1.0 or signal
quality indicator below medium.
– Excessive patient movement and vibration.
193
9 Monitoring SpO2
WARNING
With pulse oximetry, sensor movement, ambient light (especially strobe lights or flashing lights) or
electromagnetic interference can give unexpected intermittent readings when the sensor is not
attached. Especially bandage-type sensor designs are sensitive to minimal sensor movement that might
occur when the sensor is dangling.
CAUTION
If you measure SpO2 on a limb that has an inflated noninvasive blood pressure cuff, a non-pulsatile
SpO2 INOP can occur. If the monitor is configured to suppress this alarm there may be a delay of up
to 60 seconds in indicating a critical status, such as sudden pulse loss or hypoxia.
194
9 Monitoring SpO2
Alarm Delays
There is a delay between a change in the oxygen saturation at the measurement site and the
corresponding alarm indication at the monitor. This delay has two components:
• The general measurement delay time is the time between the occurrence of the saturation
change and when the new value is represented by the displayed numerical values. This
delay depends on the algorithmic processing and the averaging time configured for SpO2. The
longer the averaging time configured, the longer the time needed until the numerical values reflect
the change in saturation.
• The time between the displayed numerical values crossing an alarm limit and the alarm
indication on the monitor. This delay is the sum of the alarm delay time configured for SpO2,
plus the system alarm delay. The system alarm delay is the processing time the system needs for
any alarm to be indicated on the monitor, after the measurement has triggered the alarm. See
“Monitor Performance Specifications” on page 464 for the system alarm delay specification.
For SpO2 high and low limit alarms, there are two different types of alarm delay time. The standard
alarm delay is set to a fixed value. The Smart Alarm Delay varies, based on an intelligent algorithm,
and can be used instead of the standard alarm delay.
The Desat alarm always uses the standard alarm delay.
When Smart Alarm Delays are used, the delay before the indication of an SpO2 high or low limit alarm
depends on the amount by which the limit is exceeded and for how long. This capability can be used to
avoid alarms when the SpO2 values show a pattern of recovering after a limit violation.
There are three modes available, Short, Medium and Long. All modes have a minimum delay of
10 seconds. The maximum delay is 25 seconds for Short mode, 50 seconds for Medium mode and
100 seconds for Long mode.
195
9 Monitoring SpO2
Alarm Limit
Violation
Alarm Limit
Deviation
from violated
Alarm Limit
Short Mode
Medium
Mode
Long Mode
This diagram shows the relationship between the alarm delay and the deviation from the alarm limit.
The shaded areas on the diagram show the area in which SpO2 values can violate the alarm limit
without causing an alarm to be indicated. The area is smallest for the Short mode, and is extended for
the Medium and Long modes by the corresponding areas shown above.
WARNING
Before using Smart Alarm Delays, make sure that you fully understand how the delay is applied and
what the consequences are.
196
9 Monitoring SpO2
Alarm Limit
Deviation from
violated Alarm
Limit
This diagram shows the area for Short mode, with two examples of hypoxia.
Progressive hypoxia scenario: SpO2 value (A) - the values drop steadily and after 10 seconds a value
leaves the shaded area. An alarm is indicated immediately.
Recovery scenario: SpO2 value (B) - the values stay within the shaded area for 24 seconds, deviating
from the alarm limit by 1% or 2%, before rising again above the alarm limit. No alarm is indicated
because the SpO2 values never leave the shaded area below the alarm limit.
On the basis of the two examples, you can see that the delay depends on how much the value exceeds
the limit, and for how long. Changes in the SpO2 value can be major, but very short, or minor and for
a longer time - as long as they stay within the shaded area there will be no alarm indicated.
197
9 Monitoring SpO2
Alarm Limit
Deviation from
violated Alarm
Limit
This diagram shows the area for Medium mode, with two examples of hypoxia.
Progressive hypoxia scenario: SpO2 value (C) - the values drop steadily and after 11 seconds a value
leaves the shaded area. An alarm is indicated immediately.
Recovery scenario: SpO2 value (D) - the values stay within the shaded area for 48 seconds, deviating
from the alarm limit by 1% or 2%, before rising again above the alarm limit. No alarm is indicated
because the SpO2 values never leave the shaded area below the alarm limit.
Alarm Limit
Deviation from
violated Alarm
Limit
This diagram shows the area for Long mode, with two examples of hypoxia.
198
9 Monitoring SpO2
Progressive hypoxia scenario: SpO2 value (E) - the values drop steadily and after 16 seconds a value
leaves the shaded area. An alarm is indicated immediately.
Recovery scenario: SpO2 value (F) - the values stay within the shaded area for 98 seconds, deviating
from the alarm limit by 1% to 3%, before rising again above the alarm limit. No alarm is indicated
because the SpO2 values never leave the shaded area below the alarm limit.
WARNING
High oxygen levels may predispose a premature infant to retrolental fibroplasia. If this is a
consideration do NOT set the high alarm limit to 100%, which is equivalent to switching the high
alarm off.
199
9 Monitoring SpO2
Pleth Wave
The Pleth wave is autoscaled to maximum display size. It decreases only when the signal quality
becomes marginal. It is NOT directly proportional to the pulse volume. If you need an indication of
change in pulse volume, use the perfusion indicator.
Perfusion Numeric
The perfusion numeric (Perf) gives a value for the pulsatile portion of the measured signal caused by
the pulsating arterial blood flow.
As pulse oximetry is based on the pulsatile nature of the signal, you can also use the perfusion numeric
as a quality indicator for the SpO2 measurement. Above 1 is optimal, between 0.3-1 is acceptable.
Below 0.3 is marginal; reposition the sensor or find a better site.
200
9 Monitoring SpO2
201
9 Monitoring SpO2
202
10
10 Monitoring NBP
This monitor uses the oscillometric method for measuring NBP.
The NBP measurement is suitable for use in the presence of electrosurgery and during the discharge of
a cardiac defibrillator according to IEC 601-2-30:1999/EN 60601-2-30:2000.
A physician must determine the clinical significance of the NBP information.
WARNING
Patient Category: Select the correct patient category setting for your patient. Do not apply the higher
adult inflation, overpressure limits and measurement duration to neonatal patients.
Intravenous infusion: Do not use the NBP cuff on a limb with an intravenous infusion or arterial
catheter in place. This could cause tissue damage around the catheter when the infusion is slowed or
blocked during cuff inflation.
Skin Damage: Do not measure NBP in cases of sickle-cell disease or any condition where skin
damage has occurred or is expected.
Existing Wounds: Do not apply the cuff over a wound as this can cause further injury.
Mastectomy: Avoid applying the cuff on the side of the mastectomy, as the pressure increases the risk
of lymphedema. For patients with a bilateral mastectomy, use clinical judgement to decide whether the
the benefit of the measurement outweighs the risk.
Unattended Measurement: Use clinical judgement to decide whether to perform frequent
unattended blood pressure measurements. Too frequent measurements can cause blood flow
interference potentially resulting in injury to the patient. In cases of severe blood clotting disorders
frequent measurements increase the risk of hematoma in the limb fitted with the cuff.
203
10 Monitoring NBP
Temporary Loss of Function: The pressurization of the cuff can temporarily cause loss of function
of monitoring equipment used simultaneously on the same limb.
CAUTION
If you spill liquid onto the equipment or accessories, particularly if there is a chance that it can get
inside the tubing or the measurement device, contact your service personnel.
Measurement Limitations
NBP readings can be affected by the position of the patient, their physiological condition, the
measurement site, and physical exercise. Thus a physician must determine the clinical significance of
the NBP information.
Measurements are impossible with heart rate extremes of less than 40 bpm or greater than 300 bpm, or
if the patient is on a heart-lung machine.
The measurement may be inaccurate or impossible:
• with excessive and continuous patient movement such as shivering or convulsions
• if a regular arterial pressure pulse is hard to detect with cardiac arrhythmias
• with rapid blood pressure changes
• with severe shock or hypothermia that reduces blood flow to the peripheries
• with obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming from
the artery
• on an edematous extremity.
The effectiveness of this sphygmomanometer has not been established in pregnant, including
pre-eclamptic patients.
When the accelerated measurement is used the minimum number of oscillations per deflation step
is 1, instead of the 2 in the standard measurement. This allows a faster measurement result but requires
that the patient keeps the limb in question still. The accelerated measurement is recommended for use
when very few or no artifacts are expected, for example with sedated patients.
You can see whether the accelerated measurement is in use by looking in the Setup NBP menu. The
Accelerated Msmt setting shows whether the accelerated measurement is Off, on for Manual
measurements or on for All measurements. This setting is view-only in Monitoring Mode and can be
changed in Configuration Mode. When the accelerated measurement is in use, no pulse rate is derived
from NBP.
Measurement Modes
There are four modes for measuring NBP:
• Manual - measurement on demand.
• Auto - continually repeated measurements (between one minute and 24 hours adjustable interval).
• Sequence - up to four measurement cycles which will run consecutively, with number of
measurements and interval between them configurable for each cycle.
• STAT - rapid series of measurements over a five minute period, then the monitor returns to the
previous mode. Use only on supervised patients.
204
10 Monitoring NBP
Reference Method
The measurement reference method can be Auscultatory (manual cuff) or Invasive (intra-arterial). For
further information, see the Application Note supplied on the monitor documentation DVD.
In Neonatal mode, to comply with safety standards, invasive is always used as the reference method.
This setting cannot be changed and is not visible in any operating mode.
In Adult and Pediatric mode, to check the current setting, select Main Setup then Measurements
followed by NBP, and check whether the Reference setting is set to Auscultatory or Invasive. This
setting can only be changed in Configuration Mode.
WARNING
Kinked or otherwise restricted tubing can lead to a continuous cuff pressure, causing blood flow
interference and potentially resulting in injury to the patient.
3 Make sure that you are using a Philips-approved correct sized cuff and that the bladder inside the
cover is not folded or twisted.
A wrong cuff size, and a folded or twisted bladder, can cause inaccurate measurements. The width
of the cuff should be in the range from 37% to 47% of the limb circumference. The inflatable part
of the cuff should be long enough to encircle at least 80% of the limb.
4 Apply the cuff to a limb at the same level as the heart. If it is not, you must use the measurement
correction formula to correct the measurement.
The marking on the cuff must match the artery location. Do not wrap the cuff too tightly around
the limb. It may cause discoloration, and ischemia of the extremities.
WARNING
Inspect the application site regularly to ensure skin quality and inspect the extremity of the cuffed
limb for normal color, warmth and sensitivity. If the skin quality changes, or if the extremity
circulation is being affected, move the cuff to another site or stop the blood pressure
measurements immediately. Check more frequently when making automatic or stat measurements.
205
10 Monitoring NBP
1 Alarm source
2 Measurement Mode
3 Timestamp/Timer
4 Mean pressure
5 Diastolic
6 Systolic
7 Alarm limits
Depending on the NBP numeric size, not all elements may be visible. Your monitor may be configured
to display only the systolic and diastolic values. If configured to do so, the pulse from NBP will display
with the NBP numeric.
The measured NBP value, together with the corresponding pulse rate if this is switched on, will be
displayed for one hour. After that the values are regarded as invalid and are no longer displayed.
During this hour, measurement values may be grayed out or disappear from the screen after a set time,
if configured to do so. This avoids older numerics being misinterpreted as current data. The time can
be set in Configuration mode.
In Auto mode the measurement values may disappear more quickly (to be replaced by new
measurement values), if the repeat time is set to less than one hour.
Alarm Sources
If you have parallel alarm sources, the sources are displayed instead of the alarm limits.
• The NBP timestamp will normally show the completion time of the NBP measurement.
206
10 Monitoring NBP
During Measurements
The cuff pressure is displayed instead of the units and the repeat time. An early systolic value gives you
a preliminary indication of the systolic blood pressure during measurement.
Start NBP
Start STAT
Start/ Stop
Stop NBP
207
10 Monitoring NBP
Stop NBP
Stop NBP
CAUTION
Use clinical judgment to decide whether to perform repeated series of STAT measurements because of
the risk of purpura, ischemia and neuropathy in the limb with the cuff.
208
10 Monitoring NBP
3 Select each sequence in turn and select the number of measurements and the time interval between
the measurements.
4 To have measurements continue after the sequence, set the number of measurements for your last
cycle to Continuous and this cycle will run indefinitely.
CAUTION
Be aware that, if none of the cycles are set to Continuous, NBP monitoring will end after the last
measurement of the cycle.
When the NBP measurement mode is set to Sequence, the repetition time for Auto mode cannot be
changed.
If Mean is not selected as alarm source (Sys., Dia., or Sys & Dia selected), but the monitor can only
derive a mean value, mean alarms will nevertheless be announced using the most recent mean alarm
limits. Check that the mean alarm limits are appropriate for the patient, even when not using mean as
the alarm source. When no value can be derived an NBP Measure Failed INOP will be displayed.
209
10 Monitoring NBP
NOTE
Performing a venous puncture while automatic or sequence NBP measurements are being made
suspends the measurement series for the duration of the venous puncture inflation and for three
minutes afterwards.
Calibrating NBP
NBP is not user-calibrated. NBP pressure transducers must be verified at least once every two years by
a qualified service professional, and calibrated, if necessary. See the Service Guide for details.
210
11
11 Monitoring Temperature
WARNING
Measurements from an MMS extension connected to an X2 are not available when the X2 is running
on its own battery power. They are only available when the X2 is powered by external power: when
connected to a host monitor, to the external power supply (M8023A) or to the Battery Extension
(865297).
You can measure temperature using an X1 or X2 Multi-Measurement Module (MMS), one of the
MMS extensions, or the temperature plug-in module - labeled as TEMP.
Temp measurement automatically switches on when you connect a probe. You can switch the
measurement off manually.
WARNING
Make sure you set alarm limits for the correct label. The alarm limits you set are stored for that
particular label only. Changing the label may change the alarm limits.
211
11 Monitoring Temperature
212
12
WARNING
Measurements from an MMS extension connected to an X2 are not available when the X2 is running
on its own battery power. They are only available when the X2 is powered by external power: when
connected to a host monitor, to the external power supply (M8023A) or to the Battery Extension
(865297).
CAUTION
Do not use a monitor with an M1006A pressure module as an additional monitor for your patient. This
may cause interference on the respiration or invasive pressure measurements.
You can measure pressure using an X1 or X2 Multi-Measurement Module (MMS), one of the MMS
extensions or the pressure plug-in module - labeled as PRESS. With the plug-in module you may see a
wave channel before the pressure cable is plugged in - with the MMS this will not be the case.
WARNING
If air bubbles appear in the tubing system, flush the system with the infusion solution again. Air
bubbles may lead to a wrong pressure reading.
213
12 Monitoring Invasive Pressure
WARNING
If measuring intracranial pressure (ICP, IC1 or IC2) with a sitting patient, level the transducer with the
top of the patient's ear. Incorrect leveling may give incorrect values.
Label Description
ABP Arterial blood pressure
ART Arterial blood pressure (alternative)
Ao Aortic pressure
CVP Central venous pressure
ICP Intracranial pressure
LAP Left atrial pressure
P Non-specific pressure label
PAP Pulmonary artery pressure
RAP Right atrial pressure
UAP Umbilical arterial pressure
UVP Umbilical venous pressure
Label Description
BAP Brachial arterial pressure
FAP Femoral arterial pressure
IC1, IC2 Alternative intracranial pressures
P1, P2, P3, P4 Alternative non-specific pressure labels
214
12 Monitoring Invasive Pressure
WARNING
If you select the label ICP (or IC1/IC2), the measurement device uses the most recently stored zero.
Therefore, make sure you zeroed the transducer correctly in accordance with the transducer
manufacturer's instructions and your hospital policy. When you use a transducer that you cannot
rezero after placement, ensure that you keep the measuring device with the patient so that you are
certain you have the correct zero data for this patient.
215
12 Monitoring Invasive Pressure
CAUTION
When using high frequency ventilation, ensure that the tubing from the ventilator does not touch the
arterial line, or connect with it indirectly, while zeroing the pressure. This could cause small pressure
variations which can interfere with the zero procedure.
If you are measuring pressures with more than one measuring device, using the Zero Press SmartKey
to initiate the zeroing calls up a list of all active pressures. Select the pressure you want to zero or select
All Press to zero all pressures simultaneously.
216
12 Monitoring Invasive Pressure
217
12 Monitoring Invasive Pressure
Using the pop-up keys you can also change the scale or the speed for the wave, freeze the wave or start
a printout or recording.
You can monitor for alarm conditions in systolic, diastolic and mean pressure, either singly or in
parallel. Only one alarm is given at a time, in this order of priority: mean, systolic, diastolic.
In the Setup <Press Label> menu, select Alarms from and choose the source.
Select and set the High Limit and Low Limit for the pressure(s) you have selected.
218
12 Monitoring Invasive Pressure
219
12 Monitoring Invasive Pressure
1 Tubing to manometer
2 Syringe with heparinised solution
3 To pressure connector on monitor
4 Patient connection stoppered
5 Off
3 Move the syringe barrel in and raise the mercury to 200 mmHg (30 kPa). 200 mmHg is the
recommended calibration pressure.
4 In the Setup <Press Label> menu, select Cal. Press.
5 Select the calibration pressure from the list, for example 200 mmHg.
6 Select Confirm to recalculate the calibration factor using the applied pressure.
7 When the monitor displays <Press Label> mercury calibr. done at <Date and Time>, remove
the manometer tubing, syringe and extra stopcock. We recommend you replace the transducer
dome and tubing with sterile ones.
8 Label the transducer with the calibration factor shown in the Cal. Factor field in the pressure's
setup menu.
9 Reconnect the patient and start measuring again.
220
12 Monitoring Invasive Pressure
WARNING
• This monitor can calculate PPV from beat-to-beat values of any arterial pulsatile pressure. The
circumstances under which the calculation of a PPV value is clinically meaningful, appropriate and
reliable must be determined by a physician.
• The clinical value of the derived PPV information must be determined by a physician. According
to recent scientific literature, the clinical relevance of PPV information is restricted to sedated
patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia.
• PPV calculation may lead to inaccurate values in the following situations:
– at respiration rates below 8 rpm
– during ventilation with tidal volumes lower than 8 ml/kg
– for patients with acute right ventricular dysfunction ("cor pulmonale").
• The PPV measurement has been validated only for adult patients
CAUTION
Older Multi-Measurement modules cannot supply a beat-to-beat arterial pressure value. In this case the
monitor shows a No PPV from MMS or No PPV from FMS INOP.
221
12 Monitoring Invasive Pressure
Measuring IAP
There are two methods to store IAP measurement readings in the monitor: with manual entry or using
a pressure waveform.
NOTE
For both methods, to be able to use the label IAP, it must be made available for manual entry in
Configuration Mode. For more details on manually entered measurements, see “Entering
Measurements Manually” on page 46.
Manual Entry
This method uses a reading from an independent IAP measurement device, for example a bladder
manometer; the user manually transcribes the IAP reading to the monitor.
To manually enter the measured IAP value:
1 Select the IAP numeric on the screen.
2 Enter the IAP value and select Enter using the numeric keypad.
3 If you need to change date and time, select Date/Time. The default is always the current date and
time. Select Enter to confirm.
4 Select Save.
222
12 Monitoring Invasive Pressure
occludes the artery allowing the monitor to record changes in the intrathoracic pressures that occur
throughout the respiration cycle. The pulmonary wedge pressure is the left ventricular end diastolic
pressure (preload).
The most accurate PAWP values are obtained at the end of the respiration cycle when the intrathoracic
pressure is fairly constant. You can use the respiration waveform as a reference when assessing the
PAWP waveform, to ensure constant measurement timing relative to the respiratory cycle. The
monitor displays the PAWP value for up to 24 hours or until you admit a new patient.
WARNING
The pressure receptor in the catheter records pressure changes that occur only in front of the
occlusion. Even though the catheter tip is in the pulmonary artery, the receptor records pressure
changes transmitted back through the pulmonary circulation from the left side of the heart.
While performing the wedge procedure, the monitor switches off the pressure alarms for pulmonary
artery pressure (PAP).
Due to a slight measurement delay, you should not use sidestream CO2 as a direct reference for
determining the end expiratory point in the pressure curve.
223
12 Monitoring Invasive Pressure
WARNING
Prolonged inflation can cause pulmonary hemorrhage, infarction or both. Inflate the balloon for the
minimum time necessary to get an accurate measurement.
If the pulmonary artery flotation catheter drifts into the wedge position without inflation of the
balloon, the pulmonary artery pressure waveform assumes a wedged appearance. Take appropriate
action, in accordance with standard procedures, to correct the situation.
If the PAWP (mean) is greater than the PAP (systolic), deflate the balloon and report the incident in
accordance with hospital policy, because the pulmonary artery could be accidentally ruptured, and the
wedge value derived will not reflect the patient's hemodynamic state, but will merely reflect the
pressure in the catheter or balloon.
224
13
225
13 Monitoring Cardiac Output
Hemodynamic Parameters
This table illustrates the hemodynamic parameters available with each method, whether they are
measured continuously, and whether they can be shown on the monitor's main screen or in the
Hemodynamic Calculations window.
* currently not available in the U.S.A. and territories relying on FDA market clearance.
NOTE
EVLWI is calculated using Ideal Body Weight (IBW). ITBVI and GEDVI are calculated using Ideal
Body Surface Area (IBSA). IBW and IBSA are based on the patient category, gender and height that
you have entered for the patient.
226
13 Monitoring Cardiac Output
227
13 Monitoring Cardiac Output
228
13 Monitoring Cardiac Output
WARNING
• The monitor calculates PPV from beat-to-beat values of the arterial pressure selected for CCO.
The circumstances under which the calculation of a PPV value is clinically meaningful, appropriate
and reliable must be determined by a physician.
• The clinical value of the derived PPV information must be determined by a physician. According
to recent scientific literature, the clinical relevance of PPV information is restricted to sedated
patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia.
• PPV calculation may lead to inaccurate values in the following situations:
– at respiration rates below 8 rpm
– during ventilation with tidal volumes lower than 8 ml/kg
– for patients with acute right ventricular dysfunction ("cor pulmonale").
• The PPV measurement has been validated only for adult patients
229
13 Monitoring Cardiac Output
1 Set up the arterial line using the arterial catheter (transpulmonary catheter) and the transducer kit
from PULSION Medical Systems. It must be placed in one of the bigger systemic arteries, for
example, the femoral or the axillary artery. You must use the approved catheters and puncture
locations.
2 Set up the central venous line.
3 Connect the injectate temperature probe housing to the venous line.
4 Plug the C.O. interface cable into the C.O. module or measurement extension module and connect
the following devices to the C.O. interface cable:
– Injectate temperature probe
– Thermistor connector
– Remote start switch (if used).
Follow your hospital standards to avoid unintentional extraction of the C.O. catheter. Secure the
cable using the mounting clip shipped with each C.O. interface cable. You may also find it helpful
to loop the C.O. interface cable, tape the loop, and attach it to the undersheet of the patient's bed
using a safety pin.
230
13 Monitoring Cardiac Output
5 If you are measuring CCO, set up the pressure measurement now. The CCO measurement requires
a minimally dampened invasive pressure setup. You must ensure that there are no air bubbles in
the pressure line or dome and use only specified accessories.
6 Check that the correct measurement method is selected.
If a catheter is already connected to the Cardiac Output Interface Cable, the monitor automatically
recognizes the method used. If not, in the Setup C.O. menu, select Method and then select
Transpulmnry.
7 Check that you are using the correct probe - the M1646. This is the only probe supported for use;
it can be used with room temperature injectate or with cold injectate.
8 Check that the correct arterial catheter constant is selected.
If the catheter is recognized by the monitor, the catheter constant is automatically displayed and
cannot be changed manually. If it is not recognized, in the procedure window, select CathCt and
use the pop-up keypad to enter the correct value. The catheter constant is usually written either on
the catheter or on the catheter packaging.
9 Make sure that the injectate volume setting matches the injectate volume you will use. To change
the volume, in the procedure window, select InjVol and select the correct injectate volume from
the pop-up list.
If there is a problem with the volume or temperature you have chosen, the monitor will issue a
curve alert message to inform you of this.
10 If you are measuring CCO or CCI, check that the correct pressure source is selected in the Setup
CCO menu. The pressure label under CCO from must match the pressure measured with the
arterial catheter. To change the pressure source, select CCO from to call up a list of available
pressure labels and select the correct label.
11 If you are measuring CCO or CCI, verify that the correct alarm source is selected in the menu item
Alarms from. To change the alarm source, select Alarms from and choose either CCO or CCI.
231
13 Monitoring Cardiac Output
WARNING
CCO calibration is patient-specific. When the C.O. module or measurement extension module is
plugged in after the patient has changed, make sure that the correct CCO calibration is used. When in
doubt perform a new CCO calibration first.
Cal A pressure signal for CCO was available during the measurement (valid for calibration)
232
13 Monitoring Cardiac Output
?Cal A disturbed pressure signal for CCO was available during the measurement (valid for
calibration)
N/A No adequate pressure signal for CCO was available during the measurement (no valid
calibration data)
Exp This trial is more than 15 minutes older than the most recent trial and has expired for
CCO calibration (no valid calibration data)
233
13 Monitoring Cardiac Output
234
13 Monitoring Cardiac Output
235
13 Monitoring Cardiac Output
High ETVI
The dilution of injectate is influenced by extravascular tissue and fluid. The accuracy of the PiCCO
method may be reduced in patients with high extra-vascular thermal volume index (ETVI) values. Use
the table below as a guide in selecting the correct injectate temperature. As shown in the table, if your
patient has an ETVI > 10, you must use cold injectate.
Weight
Use the table below as a guide in choosing an injectate volume appropriate for the patient weight.
236
13 Monitoring Cardiac Output
237
13 Monitoring Cardiac Output
238
13 Monitoring Cardiac Output
CAUTION
During the cardiac output measurement procedure the blood temperature alarms are inactive. This is
indicated by a crossed-out alarm symbol next to the temperature numeric. Making alarms inactive
during this procedure prevents false alarms. The alarms are automatically reactivated when you have
completed the measurement procedure.
239
13 Monitoring Cardiac Output
240
14
WARNING
Measurements from an MMS extension connected to an X2 are not available when the X2 is running
on its own battery power. They are only available when the X2 is powered by external power: when
connected to a host monitor, to the external power supply (M8023A) or to the Battery Extension
(865297).
Use the CO2 measurement to monitor the patient's respiratory status and to control patient ventilation.
There are two methods for measuring carbon dioxide in the patient's airway:
• Mainstream measurement uses a CO2 sensor attached to an airway adapter directly inserted into
the patient's breathing system. This method is available using the M3014A Capnography
Extension and the M3016A Mainstream CO2 Extension, or the X2 with the optional CO2
measurement.
• Sidestream measurement takes a sample of the respiratory gas with a constant sample flow from
the patient's airway and analyzes it with a remote CO2 sensor built into the measurement system.
Philips offers the sidestream CO2 measurement in the M3014A Capnography Extension or the X2
with the optional CO2 measurement, and the Microstream method of sidestream CO2
measurement in the M3015A/B Microstream CO2 Extension.
Refer to your Gas Module Instructions for Use, if you are using a Gas Module to monitor CO2.
WARNING
Correlation: The etCO2 readings do not always correlate closely with paCO2, especially in neonatal
patients and patients with pulmonary disease, pulmonary embolism or inappropriate ventilation.
Pharmaceuticals in aerosols: Do not measure CO2 in the presence of pharmaceuticals in aerosols.
Explosion Hazard: Do not use in the presence of flammable anesthetics or gases, such as a
flammable anesthetic mixture with air, oxygen or nitrous oxide. Use of the devices in such an
environment may present an explosion hazard.
Failure of operation: if the measurement or a sensor fails to respond as described, do not use it until
the situation has been corrected by qualified personnel.
241
14 Monitoring Carbon Dioxide
Low etCO2 values: Leakages in the breathing system or sampling system may cause the displayed
etCO2 values to be significantly too low. Always connect all components securely and check for leaks
according to standard clinical procedures. Displacement of the nasal or combined nasal oral cannulas
can cause lower than actual etCO2 readings. Even with combined nasal oral cannulas, the etCO2
readings may be slightly lower than actual in patients breathing through the mouth only.
Measurement Principles
For both mainstream and sidestream measurements, the measurement principle is infrared
transmission, where the intensity of infrared light passing the respiratory gas is measured with a photo
detector. As some of the infrared light is absorbed by the CO2 molecules, the amount of light passing
the gas probe depends on the concentration of the measured CO2.
When using a wet ventilator circuit, monitor mainstream CO2 if available, in preference to sidestream
CO2.
The partial pressure is calculated from the gas concentration by multiplying the concentration value
with the ambient pressure.
The measurement provides:
• a CO2 waveform.
• an end tidal CO2 (etCO2) value: the CO2 value measured at the end of the expiration phase.
• an inspired minimum CO2 (imCO2): the smallest value measured during inspiration.
• an airway respiration rate (awRR): the number of breaths per minute, calculated from the CO2
waveform.
Depending on the Max Hold setting configured for your monitor, the etCO2 numeric shows either the
highest CO2 value measured within the configured time period (Max Hold set to 10 sec or 20 sec) or
the etCO2 numeric shows breath-to-breath value (Max Hold set to Off).
The Microstream method also provides an Integrated Pulmonary Index (IPI) numeric, which is an
indication of the patient's overall ventilatory status based on four measurement parameters: etCO2,
awRR, pulse rate and SpO2. IPI can thus provide an early indication of a change in ventilatory status
which may not be shown by the current value of any of these four parameters individually. The IPI is
designed to provide additional information regarding patient status, possibly before etCO2, awRR,
SpO2, or pulse rate values reach levels of clinical concern.
The IPI is available for all three groups of pediatric patients (1-3 years, 3-6 years, and 6-12 years), and
for adult patients. It is displayed as a single value between 1 and 10.
242
14 Monitoring Carbon Dioxide
With intubated patients, a sample of the respiratory gas is drawn from the patient's breathing circuit
through an airway adapter and a gas sampling tube. With non-intubated patients, the gas sample is
drawn through a nasal or oral-nasal cannula.
WARNING
Altitude Setting: The monitor is not equipped with automatic barometric pressure compensation.
Before the CO2 measurement is used for the first time, the altitude must be set to the correct value. An
incorrect altitude setting will result in incorrect CO2 readings. The CO2 readings will typically deviate
5% for every 1000 m difference.
CAUTION
Use the CO2 measurement with Philips approved accessories only. Refer to the instructions for use
provided with the accessory.
3 Choose the appropriate airway adapter and connect it to the sensor head. The airway adapter clicks
into place when seated correctly. To zero the sensor:
– expose the sensor to room air and keep it away from all sources of CO2 including the
ventilator, the patient's breath and your own.
– in the setup menu for the CO2, select Start Zero Cal.
243
14 Monitoring Carbon Dioxide
– When you see the message CO₂ calibration done at <Date and Time> on the status line, the
zero calibration is finished and you can begin monitoring.
4 Install the airway adapter at the proximal end of the circuit between the elbow and the ventilator Y-
section.
WARNING
To prevent stress on the endotracheal tube, support the sensor and airway adapter.
Position sensor cables and tubing carefully to avoid entanglement or potential strangulation. Do not
apply excessive tension to any cable.
Replace the airway adapter, if excessive moisture or secretions are observed in the tubing or if the
CO2 waveform changes unexpectedly without a change in patient status.
To avoid infection, use only sterilized, disinfected or disposable airway adapters.
Inspect the airway adapters prior to use. Do not use if airway adapter appears to have been damaged
or broken. Observe airway adapter color coding for patient population.
2 Connect the cannula, airway adapter, or sample line as appropriate, to the sensor. It will click into
place when seated correctly.
244
14 Monitoring Carbon Dioxide
– expose the sensor to room air and keep it away from all sources of CO2 including the
ventilator, the patient's breath and your own.
– in the setup menu for the CO2, select Start Zero Cal.
– when you see the message CO₂ calibration done at <Date and Time> on the status line, the
zero calibration is finished and you can begin monitoring.
4 For intubated patients requiring an airway adapter: Install the airway adapter at the proximal end of the
circuit between the elbow and the ventilator Y-section.
For intubated patients with an integrated airway adapter in the breathing circuit.: Connect the male luer
connector on the straight sample line to the female port on the airway adapter.
For non-intubated patients: Place the nasal cannula onto the patient.
WARNING
Always connect the airway adapter to the sensor before inserting the airway adapter into the breathing
circuit. In reverse, always remove the airway adapter from the breathing circuit before removing the
sensor.
Make sure that you do not accidentally connect the luer connector of the gas sample line to an infusion
link or any other links in the patient vicinity.
245
14 Monitoring Carbon Dioxide
CAUTION
Always disconnect the cannula, airway adapter or sample line from the sensor when not in use.
1 Push the sensor into the holder until it clicks into position.
2 Clamp the holder onto an IV pole, a shelf or another appropriate location.
To remove the sensor from the holder, release the clip and pull the sensor out of the holder.
WARNING
Anesthetics: When using the sidestream CO2 measurement on patients who are receiving or have
recently received anesthetics, connect the outlet to a scavenging system, to avoid exposing medical
staff to anesthetics.
Use an exhaust tube to remove the sample gas to a scavenging system. Attach it to the sidestream
sensor at the outlet connector.
WARNING
Infra-red radiation: Do not expose the airway adapter or M1460A transducer to infra-red radiation
during use. This may cause incorrect readings.
246
14 Monitoring Carbon Dioxide
247
14 Monitoring Carbon Dioxide
CO₂ Sens Warmup message until the transducer reaches operating temperature. Wait until this
disappears before starting the measurement.
2 To remove the transducer from the airway adapter, open the latch and pull out the airway adapter.
WARNING
To prevent stress on the endotracheal tube, support the transducer and airway adapter.
To avoid infection, use only sterilized airway adapters.
248
14 Monitoring Carbon Dioxide
WARNING
When using the IPI numeric (available for adult and pediatric patients), always ensure that the patient's
date of birth is entered correctly and that the monitor is displaying the current date. These two pieces
of information are used to calculate the age of the patient, which affects the algorithm used to provide
the IPI numeric.
1 Attach the female Luer connector to the CO2 inlet connector by pushing the socket cover aside
and screwing the connector clockwise into the CO2 inlet until it can no longer be turned.
This will assure that there is no leak of gases at the connection point during measurement and that
measurement accuracy is not compromised. Following connection of the CO2 sampling line, check
that CO2 values appear on the monitor display.
2 Check that the FilterLine is not kinked.
3 Change the FilterLine if a CO₂ Occlusion INOP appears on the monitor or if the readings become
extremely erratic.
Disconnect the FilterLine during suctioning and nebulizing therapies or suppress sampling by turning
off the pump as described in “Suppressing Sampling (not Mainstream CO2)” on page 251.
249
14 Monitoring Carbon Dioxide
Check the table in the Microstream CO2 Accessories section of the Accessories chapter for typical
usage times for the different Microstream accessories.
CO2 values for non-intubated patients using Microstream accessories will always tend to be lower than
for intubated patients. If values appear extremely low, check whether the patient is breathing through
the mouth or whether one nostril is blocked.
Use an exhaust tube to remove the sample gas to a scavenging system. Attach it to the MMS Extension
at the outlet connector.
Correction
Altitude Altitude is set during installation. The monitor automatically applies an
(M3014A only) appropriate correction.
O2 In the Setup CO₂ menu, select Oxy. Corr and select a value between 20%
(M3014A only) and 100%, the default value is 20%. If you are not measuring the expired O2,
estimate it by subtracting 5% from the inspired O2 then select the nearest
value from the list.
250
14 Monitoring Carbon Dioxide
Correction
Humidity At installation, the monitor is configured to automatically apply either Body
Temperature Pressure Saturated (BTPS) or Ambient Temperature Pressure
Dry (ATPD). To see which, go to the Setup CO₂ menu, and scroll down to
look at Hum. Corr or Humidity Corr..
N2 O In the Setup CO₂ menu, select N₂O Corr. and turn on or off. If N2O is
(M3016A mainstream present in the ventilation gas mixture, you must turn this on.
and M3015A/B If the N2O correction is not available in the Setup CO₂ menu, the CO2
microstream only) measurement in your MMS Extension does not require N2O correction or it
is setup with Gas Corr. (see below).
Gas In the Setup CO₂ menu, select Gas Corr. and select Helium, N₂O or turn off.
(M3014A only) If Helium or N2O is present in the ventilation gas mixture, you must make
the appropriate selection.
If the Gas correction is not available in the Setup CO₂ menu, the CO2
measurement in your MMS Extension does not require N2O or Helium
correction or the N2O correction is setup with N₂O Corr. (see above).
Agent In the Setup CO₂ menu, select Agt. Corr and select the concentration of the
(M3014A only) anesthetic agent (between 0.0% and 20.0%). If an anesthetic agent is present
in the ventilation gas mixture, you must select the appropriate concentration.
251
14 Monitoring Carbon Dioxide
WARNING
Safety and effectiveness of the respiration measurement method in the detection of apnea,
particularly the apnea of prematurity and apnea of infancy, has not been established.
Prolonged delay: The selected apnea alarm delay may be prolonged by up to 17 seconds, if an apnea
occurs during the automatic zero process. This applies to the Microstream (M3015A/B) measurement
only.
252
14 Monitoring Carbon Dioxide
NOTE
The interpretation of the patient's IPI score may change in different clinical environments. For
example, patients with specific respiratory difficulties (in contrast to normally healthy patients who are
being monitored during sedation or pain management) may require a lower IPI Low Alarm threshold
to reflect their impaired respiratory capacity.
The IPI is available for all three groups of pediatric patients (1-3 years, 3-6 years, and 6-12 years), and
for adult patients.
WARNING
Ensure that the patient category and date of birth are set correctly before monitoring a patient's IPI.
An incorrect patient category or date of birth could produce incorrect IPI data.
253
14 Monitoring Carbon Dioxide
254
15
255
15 Monitoring Airway Flow, Volume and Pressure
WARNING
Electrical Shock Hazard: Do not open the module. Contact with exposed electrical components
may cause electrical shock. Always turn off and remove power before cleaning the module. Refer
servicing to qualified service personnel.
Select the appropriate flow sensor. Make sure that you are using the correct sensor for the respective
patient category. Otherwise accuracy may be reduced.
You can also use combined CO2/flow sensors. Note that the M3014A Capnography Extension is
required to measure CO2.
Pediatric/Adult Combined CO2/Flow Sensor (M2781A). Color: clear
For intubated patients with endotracheal tube diameters > 5.5 mm. Adds
8 cm3 of dead-space.
1 If you are using a combined CO2/Flow sensor, connect it to the CO2 sensor head first. The airway
adapter clicks into place when seated correctly.
2 Click the connector into place in the flow sensor receptacle on the monitor before connecting to
the breathing circuit.
256
15 Monitoring Airway Flow, Volume and Pressure
3 Install the flow sensor or the combined CO2/Flow sensor at the proximal end of the breathing
circuit between the elbow and the ventilator Y-piece. Make sure that the spirometry sensor is in a
horizontal position with its tubing pointing upwards. The correct position is also indicated by an
arrow on some of the sensors. (Graphic shows combined CO2/Flow Sensors).
1. Pediatric/adult sensor
2. Pediatric sensor
3. Infant/Neonatal sensor
WARNING
To prevent stress on the endotracheal tube, support the sensor and airway adapter.
Position sensor cables and tubing carefully to avoid entanglement or potential strangulation. Do not
apply excessive tension to any cable or tubing.
Replace the sensor, if excessive moisture or secretions are observed in the tubing and cannot be
removed by a purge cycle.
Reuse poses a patient hazard. Do not disassemble, clean, disinfect or sterilize the sensor.
A system leak may significantly affect readings of flow, volume, pressure and other respiratory
mechanics parameters.
Do not leave the sensor in the patient circuit when not connected to the monitor.
NOTE
• Do NOT place the airway adapter between the endotracheal tube and the elbow (pediatric/adult
circuit), as this may allow patient secretions to block adapter windows.
• The striped tubing of the flow sensor should always be proximal to the patient.
• Position the airway adapter with the spirometry tubing pointing upwards.
• To prevent "rain-out" and moisture from draining into the airway adapter, do NOT place the
airway adapter in a gravity dependent position.
• Periodically check the flow sensor and tubing for excessive moisture or secretion build up.
Perform a purge cycle if droplets are visible within the sensor or tubing. Repeat if necessary. If
purging does not remove the droplets, replace the sensor.
• For routine performance of airway care, separate the system between the endotracheal tube and
the airway adapter (neonatal circuit), or between the endotracheal tube and elbow (pediatric/adult
257
15 Monitoring Airway Flow, Volume and Pressure
circuit). Lavage and suctioning of the airway can then be performed without fluids and mucous
accumulating on the airway adapter windows.
• Measurement values provided by a ventilator may differ significantly from the values provided by
the spirometry module, due to different locations of the flow sensor.
• Incorrect entry of gas compensation parameters (i.e. temperature, gas composition) may reduce
the accuracy of the measured values.
• An abnormal volume waveform as displayed below may be an indicator of an air leak:
In general, if MVexp or TVexp are significantly smaller than MVin or TVin, the tubing should be
checked for leaks.
Zero Calibration
The zero calibration maintains the accuracy of the spirometry waves and numerics by regularly
compensating for drifts within the measurement section. It is performed automatically without user
interaction and takes about 2 seconds to complete. The automatic zero calibration is normally carried
out every ten minutes. During warm-up or when the ambient pressure changes, this interval may be
reduced to two minutes.
During the zero calibration the waveform is flat but the numerics remain on the screen. Typically, a
zero calibration is started at the beginning of a respiration cycle, therefore a waveform may begin
normally and then immediately become flat for the time of the zero calibration.
1 Zero Calibration
Automatic Purging
A double lumen connecting line (tubing) connects the flow sensors to the patient monitor. The
M1014A Spirometry Module includes an automatic and manual purge feature which provides a flush
of room air to keep the sensor tubing free from water condensation and patient secretions. This
feature is available for the adult, pediatric, and neonatal modes. The purge will begin with the
exhalation portion of the ventilator cycle.
NOTE
• During the purge cycle the pump will be heard
258
15 Monitoring Airway Flow, Volume and Pressure
• If the purge does not sufficiently clear the flow tubing lines, the flow sensor should be replaced
• With each purge cycle, a zero calibration is automatically performed
Adult Mode
The system automatically purges the sensor tubing every ten minutes or less, depending on system
conditions. In adult mode, the system will purge both sides of the line, one at a time, during each purge
cycle. The higher the pressure, the more frequent the purging. This action anticipates increased
moisture migration into the sensor tubing due to the increase in circuit pressure.
Manual Purging
Occasionally, purging may be required in between the automatic purge cycles. The manual purge may
be used as often as needed. Check the orientation of the flow sensor if repeated manual purge cycles
become necessary.
To manually purge, press the Purge button on the module (marked by on international versions of
the module) or select Purge in the Setup Spirometry window to initiate a combined purge and zero
cycle.
Gas Compensation
Temperature and the proportions of individual gases influence the flow measurement. If values seem
inaccurately high or low, check that the monitor is using the appropriate gas compensation. Gas
compensation can be done using either manually entered gas concentrations or gas concentrations
from the gas analyzer.
NOTE
If you unintentionally switch the gas compensation to the incorrect setting, the accuracy of the
measured flow and volume values may be significantly reduced. The tables below show examples of
incorrect gas compensation settings and the resulting inaccuracies:
259
15 Monitoring Airway Flow, Volume and Pressure
Setting up Spirometry
The following settings can be accessed through the Setup Spirometry menu.
WARNING
• Safety and effectiveness of the respiration measurement method in the detection of apnea,
particularly the apnea of prematurity and apnea of infancy, has not been established.
• If an apnea occurs during a zero calibration or purge, the time delay between the start of apnea and
the activation of the apnea alarm could be up to 10 seconds plus the configured apnea delay time.
260
15 Monitoring Airway Flow, Volume and Pressure
NOTE
If MV and TV are set to off, there will be no alarming for these parameters.
NOTE
• Gas concentrations from the gas analyzer are only available for Philips gas analyzers, not for
devices connected via a Vuelink/IntelliBridge module.
• If gas concentrations from the gas analyzer are selected but not all data is available, the missing
data is taken from manually entered values. In case of invalid data or no data at all, the INOP
message Spiro Gas Compens? is displayed.
NOTE
Changing the temperature setting unintentionally will result in approximately 3% reduced accuracy per
10°C (18°F) temperature setting change.
261
15 Monitoring Airway Flow, Volume and Pressure
262
16
16 Monitoring tcGas
The tcGas module measures the partial pressure of the oxygen and carbon dioxide that diffuses
through the skin, thereby providing a measure of these gases in the capillary blood.
The monitor's settings for altitude and barometric pressure influence the measurement. The tcpO2/
tcpCO2 measurement is valid for an infant patient not under gas anesthesia. Anesthetic agents, such as
halothane, can cause incorrect or drifting readings.
Transcutaneous measurements cannot replace arterial blood gas monitoring. However, you can use
transcutaneous monitoring to reduce the frequency of arterial sampling. The values at tissue level will
not be the same as those measured arterially because the measurement is transcutaneous. They
correlate with (track closely) the arterial values. For example, a drop in transcutaneous values usually
indicates a corresponding drop in arterial values.
Transcutaneous values will not always correlate with blood samples taken from the capillary blood of
the heel (heelsticks or astrups).
263
16 Monitoring tcGas
WARNING
Prolonged continuous monitoring may increase the risk of undesirable changes in skin characteristics,
such as irritation, reddening, blistering or burns. If the site timer is disabled, the transducer will heat
indefinitely while on a patient. Change the site regularly, in accordance with medical procedures in your
hospital.
264
16 Monitoring tcGas
265
16 Monitoring tcGas
2 Plug the transducer cable into the module. Swing the calibration chamber cover open and insert
the transducer into the chamber. Close the cover to secure the transducer. Set the transducer
temperature at the monitor now.
3 On the 15210B calibration unit, turn the timer control clockwise as far as you can. On the
Radiometer calibration unit, press the button with the green arrow once.
4 Press CAL on the module until the light above the key comes on and wait (three - 20 minutes) for
the completion message to appear on the monitor. Alternatively, in the Setup tcGas menu, select
StartCalibration. To save gas on 15210B, if the timer control dial is not in the start position when
the monitor displays the completion message, turn the dial counter-clockwise to the start position.
For TCC3, if the green light is still flashing when INOP <tcGas Label> Cal Running disappears,
press the green arrow button again.
266
16 Monitoring tcGas
Calibration Failure
If calibration fails, the monitor displays <tcGas Label> transducer or cal. unit malfunction and the
<tcGas Label> Cal Failed INOP for the measurement.
2 Align the arrow on the transducer with the tab on the ring and fasten by turning a quarter-turn
clockwise. Wait 10-20 minutes for readings to stabilize.
267
16 Monitoring tcGas
3 Apply the transducer as soon as possible after you see the message indicating that calibration is
complete. If you wait longer than 30 minutes, the heat supply to the transducer switches off to
prevent the electrolyte from drying out and a new calibration is necessary.
Optimize the measurement by selecting a site with high capillary density and blood flow, thin
epidermis and no cardiovascular disorders. Most physicians use the abdomen, chest and back.
WARNING
You must either remove the transducer before defibrillating, or remembrane and calibrate the
transducer after defibrillating.
CAUTION
To avoid transducer damage, remove it from the patient during high frequency surgical procedures.
TcGas Corrections
Transcutaneous pCO2 values tend to be higher than arterial values due to the metabolic processes of
the skin and the effect of heating on the blood under the transducer. Depending on your monitor's
configuration, one or both of these corrections may automatically apply.
268
16 Monitoring tcGas
• In the Setup tcGas menu, look at the menu item CO₂ Correction. If correction is enabled, it is set
to On.
269
16 Monitoring tcGas
270
17
Monitoring Intravascular
17
Oxygen Saturation
Depending on the probe or catheter used and the measurement location, the two modules, M1011A
and M1021A, measure the central venous oxygen saturation or mixed venous oxygen saturation
continuously and invasively.
The two modules can be differentiated by their size and their labeling. The M1011A is a single-width
module (narrow) and is labeled SO2. The M1021A is a double-width module (wide) and is labeled
SvO2.
1 M1011A
2 M1021A
WARNING
Injected dyes, such as methylene blue, or intravascular dyshemoglobin may lead to inaccurate
measurements.
Do not monitor oxygen saturation during infusion of I.V. fat emulsion or other turbid substances
through the distal lumen of the OptiCath catheter. These liquids might temporarily modify the blood
scattering and absorption characteristics at the catheter tip. This interferes with the optical
measurement of oxygen saturation. After infusion is complete, you can again monitor oxygen
saturation accurately.
During injection of the bolus for thermodilution cardiac output measurements, the SvO2
measurement might be disturbed.
Explosion Hazard: Do not use in the presence of flammable anesthetics or gases, such as a
flammable anesthetic mixture with air, oxygen or nitrous oxide. Use of the devices in such an
environment may present an explosion hazard.
271
17 Monitoring Intravascular Oxygen Saturation
CAUTION
Use the modules with Philips approved accessories only. Refer to the instructions for use provided
with the accessory.
272
17 Monitoring Intravascular Oxygen Saturation
1 Optical module
2 Balloon inflation stopcock
3 Hospira fiber optic catheter
4 Optical reference
5 Enter setup/calibration
Connect the optical module (Hospira 50131) to the measurement module. Allow the optical module to
warm up before you perform a calibration. Although the warm up message disappears from the screen
after one minute, Hospira recommends letting the optical module warm up for 15 minutes for best
accuracy. Please refer to the instructions for the optical module.
To avoid false alarms during the pre-insertion calibration and insertion of the catheter into the patient,
the monitor automatically suspends alarms during the pre-insertion calibration, for up to three minutes
after you remove the catheter tip from the optical reference. After light intensity calibration, or after
three minutes (whichever comes first), the monitor returns to the alarm state it was in prior to pre-
insertion calibration.
Refer to the instructions for use that accompany the catheter. Do not use the catheter if the packaging
is damaged. If you have to disconnect the monitor from the patient (for example, when transferring
the patient from one location to another), you must disconnect at the SvO2 module. The catheter
should remain in the optical module, otherwise you need to recalibrate.
1 Remove outer wrapping from catheter tray to uncover optical connector.
2 Place the optical module on the catheter tray in the space provided and open the lid.
3 Place the optical connector into the optical module (with the label "TOP" facing upwards) and
close the lid.
4 In the Setup <SO₂ Label> menu, select Start Pre-InsCal. Ensure that the tip of the catheter is
still in the optical reference.
5 Insert the catheter when you see the message <SO₂ Label> calibration completed - catheter
ready for insertion. If the calibration fails, repeat the calibration before inserting the catheter. If it
fails a second time, replace the optical module.
273
17 Monitoring Intravascular Oxygen Saturation
274
17 Monitoring Intravascular Oxygen Saturation
275
17 Monitoring Intravascular Oxygen Saturation
After Insertion
The SO2 probe/catheter is thin and flexible, treat it carefully. Avoid kinking, bending or grasping the
probe/catheter with forceps or a hemostat. Damage to the fiber results in low intensity light and a
sudden decrease in intensity readings. Refer to the documentation provided with the fibre-optic
probe/catheter, paying special attention to any precautions, warnings or contraindications.
Secure the optical module directly attached or in close proximity to the patient, to avoid placing
excessive tension on the catheter, which would result in movement of the catheter tip from the optimal
position in the patient.
Position the optical module to avoid contact with liquids. Fluid entering the catheter-optical module
connection will impair measurement performance.
You must perform an in-vivo calibration once the probe/catheter is in place.
The probe/catheter should be replaced after it has been in place for 72 hours.
CAUTION
Do not apply excessive tension to any cable or part of the catheter.
276
17 Monitoring Intravascular Oxygen Saturation
4 Draw a blood sample from the distal port of the catheter and flush the line according to standard
hospital practice.
5 Obtain laboratory analysis of the sample using direct measurements.
6 Select CalibrationValue and select from the list the value received from the lab.
7 Select Hct [%] (or Hb [mmol/l] or Hb [g/dl] depending on the set up) and enter the corresponding
value from the laboratory analysis.
To change the setup for entering the Hb/Hct, see “Changing the Lab Value Required for Entry”
on page 277 below.
8 If you need to enter a new correction factor, select Catheter Factor.
9 Enter the correction factor.
10 Complete the calibration by selecting Store In-VivoCal (even if the stored calibration value did not
change) and select Confirm. This updates the data stored in the optical module.
Selecting Recall Last Cal recalls the previously stored calibration value.
If the calibration fails, check that the light intensity indicator is indicating a stable medium to high level.
Repeat the calibration.
277
17 Monitoring Intravascular Oxygen Saturation
278
18
18 Monitoring EEG
The Electroencephalograph (EEG) module monitors the patient's cerebral function by measuring the
electrical activity of the brain. It provides the monitor with two channels of realtime EEG waves, EEG
trend information in the form of Compressed Spectral Arrays (CSA), and up to eight of the following
numerics:
Spectral Edge Frequency - The SEF is the frequency below which a defined percentage of the Total
Power lies. The percentage is set in Configuration Mode.
Mean Dominant Frequency (MDF) - The MDF is the mean value of the frequency which
dominates the measured EEG.
Peak Power Frequency (PPF) - The PPF is the frequency with the highest measured amplitude.
Total Power (TP) - The TP numeric indicates the power in the measured frequency band.
Percentage of total power in each frequency band:
– Alpha waves (8 to 13 Hz)
– Beta waves (13 to 30 Hz)
– Theta waves (4 to 8 Hz)
– Delta waves (0.5 to 4 Hz).
279
18 Monitoring EEG
1 Electrode locations on the patient's head. The symbols represent the electrode-to-skin impedance.
2 Wiring and impedance values for the selected montage.
The electrode locations are labeled according to the international 10-20 electrode placement system.
280
18 Monitoring EEG
Impedance Indicators
Electrode/Skin Symbol Color Displayed Action
Impedance Impedance
Value
Electrode not connected red no value connect electrode
281
18 Monitoring EEG
EEG values are sampled at configured time intervals and displayed as color-coded trendlines.
Trendlines
Trendlines are available for the three frequency numerics (SEF, PPF, MDF)
INOP marker
282
18 Monitoring EEG
283
18 Monitoring EEG
EEG Reports
The content of EEG Reports is always the same and does not need to be configured.
To print an EEG Report, in the Setup EEG menu, select Print Report.
Alternatively, you can select the CSA and use the Print Report pop-up key to start the report.
To modify the buffer and trendline settings on the CSA Report, in the Reports menu, select CSA on
EEG Rep.. If you do not change these settings, the monitor will use the default settings with the
trendlines for the SEF numeric on and the buffer time from Buffer C.
284
19
19 Monitoring BIS
Bispectral Index monitoring helps to monitor the level of consciousness of a patient under general
anesthesia or sedation in the OR and ICU. The BIS sensor is placed on the patient's forehead to
capture electroencephalographic (EEG) signals from which several numerics are derived, including a
single BIS value representing the level of consciousness. See the chapter on Specifications for the BIS
intended use statement.
The BIS Module provides the monitor with an EEG wave and the following numerics:
Bispectral Index (BIS). The BIS numeric reflects the patient’s level of consciousness. It ranges from
100 (fully awake) to 0 (suppression; no electrical brain activity).
Signal Quality Index (SQI). The SQI numeric reflects signal quality and provides information about
the reliability of the BIS, SEF, TP, SR, and Bursts numerics during the last minute.
It ranges from 0 to 100%:
SQI < 15%: the numerics cannot be derived
SQI 15% to 50%: the numerics cannot be reliably derived
SQI 50% to 100%: the numerics are reliable.
Electromyographic Activity (EMG). The EMG numeric reflects the electrical power of muscle
activity and high frequency artifacts.
EMG < 55 dB: this is an acceptable EMG
EMG ≤ 30 dB: this is an optimal EMG
(note that the minimum possible EMG is approximately 25 dB).
Suppression Ratio (SR). The SR is the percentage of time over the last 63-second period during
which the EEG is considered to be in a suppressed state.
Spectral Edge Frequency (SEF). The SEF is the frequency below which 95% of the Total Power is
measured.
Total Power (TP). The TP numeric indicates the power in the frequency band 0.5 to 30 Hz. The
useful range is 30 - 100 dB.
Bursts (BISx used with the Extend sensor only). The Bursts numeric helps you quantify suppression
by measuring the number of EEG bursts per minute, where an EEG burst is defined as a period of
activity followed and preceded by inactivity (at least 0.5 second).
285
19 Monitoring BIS
1 If you are monitoring BIS with the DSC and BIS Engine,
a. Connect the BIS Engine to the BIS module using the BIS Engine Cable.
b. Connect the digital signal converter (DSC) to the digital signal converter port on the front of
the BIS Engine. Use the attachment clip to secure the digital signal converter near, but not
above the patient's head.
c. Attach the patient interface cable (PIC) to the digital signal converter (DSC).
2 Attach the BIS sensor to the patient following the instructions supplied with the sensor.
Make sure that the patient's skin is dry. Be aware that a wet sensor or a salt bridge may cause
erroneous BIS and impedance values.
A variety of sensors are available for use in the OR and ICU environments.
3 Connect the BIS sensor to the patient interface cable.
As soon as a valid sensor is detected, the impedances of all electrodes are measured automatically
and the results are shown in the BIS window.
286
19 Monitoring BIS
1 BIS Module
2 BISx
3 Patient Interface Cable
4 BIS Sensor
CAUTION
Ensure that the BISx does not come into prolonged contact with your patient's skin, as it may
generate heat and cause discomfort.
The BISx may remain connected to a patient during defibrillation as long as the sensor is not
located between the defibrillator pads.
Manufacturer's Information
BIS Engine, DSC, BISx, the Patient Interface Cable and the BIS Sensors are manufactured by
Covidien llc.
287
19 Monitoring BIS
CAUTION
Switching the continuous impedance check off will disable automatic notification to the user of
impedance value changes, which may lead to incorrect BIS values. Therefore, this should only be done
if the check interferes with or disturbs other measurements.
288
19 Monitoring BIS
BIS Window
To open the BIS window, in the Setup BIS menu, select Show Sensor.
The window may look slightly different on your monitor. The graphic in the BIS Window
automatically adapts to show the type of sensor you are using, showing three or four electrodes as
required. Each symbol in the graphic represents an electrode and illustrates the most recently-
measured impedance status of the electrodes. Although BIS may still be measured when the electrode
status is red or yellow, for best performance, all electrodes should be green.
In addition, if the measured electrode-to skin impedance of any electrode or electrode combination is
above the limit, or if disconnected electrodes are detected, an INOP will be issued, either BIS High
Impedance or BIS Lead Off.
1 Reference Electrode
2 Ground Electrode
3 Signal Electrode(s)
4 Time of the most recent cyclic check
Depending on your configuration, in addition to the symbols, the impedance value in kOhm may be
displayed.
289
19 Monitoring BIS
290
19 Monitoring BIS
WARNING
Conductive Parts: The conductive parts of sensors and connectors should not contact other
conductive parts, including earth.
High-frequency Surgery: To reduce the hazard of burns in the high-frequency surgical neutral
electrode connection, the BIS sensor should not be located between the surgical site and the
electrosurgical unit return electrode.
Defibrillation: The BIS sensor must not be located between defibrillator pads when a defibrillator is
used on a patient connected to the patient monitor.
Securing Cables: To minimize the risk of patient strangulation, the patient interface cable (PIC) must
be carefully placed and secured.
CAUTION
Revisions: The system will only function if all component revisions are compatible. Otherwise, an
incompatibility INOP is displayed.
If the DSC has an older software revision than the BIS Engine, the DSC will automatically be
upgraded by the BIS Engine. Do not disconnect the DSC from the BIS Engine, or disconnect the BIS
module from the monitor, or switch the monitor power off within the first ten seconds after
connection, as this will disrupt a possible software upgrade and cause damage to the DSC.
Impedance Checks: Impedance checks may influence data acquisition of other
electroencephalographic devices.
291
19 Monitoring BIS
292
20
20 Monitoring NMT
Neuro Muscular Transmission (NMT) and its measurement enables the evaluation of muscle
relaxation of patients under Neuromuscular Block by measuring the strength of muscle reaction after
electrically stimulating the dedicated motor nerve. The NMT Monitor electrodes are placed on the
patients skin over the ulnar nerve, a controllable current source delivers stimulation pulses to two skin
surface electrodes for the nerve stimulation, the muscle response is measured with an acceleration
sensor.
WARNING
The NMT measurement may not be used for neonatal patients and is therefore not supported in
the neonatal mode.
NMT stimulation current pulses may interfere with other sensitive equipment, for example,
implanted cardiac pacemakers. Do not use the NMT measurement on patients with implanted medical
devices unless so directed by a medical specialist.
Simultaneous use of the NMT with high frequency electrosurgical equipment may in unusual
circumstances result in burns at the stimulation site and can also adversely affect measurement
accuracy.
Do not use the NMT in close proximity to shortwave or microtherapy devices, there is a risk of
adversely affecting the NMT measurement.
Explosion Hazard: Do not use in the presence of flammable anesthetics or gases, such as a
flammable anesthetic mixture with air, oxygen or nitrous oxide. Use of the device in such an
environment may present an explosion hazard.
Never apply the electrodes trans-thoracically (across the chest) or transcerebrally (across the head),
there is a risk of seriously injuring the patient. Apply the stimulation electrodes close together as
described in the Instructions for Use.
Never apply electrodes to patients in areas where inflammation or injury is evident.
When you are connecting the electrodes or the patient cable, make sure that the connectors do not,
and cannot, come into contact with other conductive parts, or with earth.
Never touch the electrodes unless the stimulation has been stopped.
Patients with nerve damage or other neuromuscular problems may not respond properly to
stimulation. The NMT measurement may show unusual patterns when monitoring muscle paralysis in
these patients.
After repositioning the patient, check that the sensor is still applied and can move freely.
293
20 Monitoring NMT
CAUTION
NMT monitoring is intended as an adjunct in patient assessment and must be used in conjunction with
observation of clinical signs and symptoms.
NMT stimulation can be painful to a non-sedated patient. It is recommended not to stimulate before
the patient is adequately sedated.
Use only electrodes suitable for nerve stimulation deemed appropriate by the attending physician. Pay
special attention to current densities exceeding 2 mA r.m.s/cm2 for any electrodes.
Stimulation Modes
The NMT module provides four stimulation modes:
• Train-Of-Four (TOF)
• Single Twitch (Twitch)
• Post-Tetanic Count (PTC)
• Double-Burst Stimulation (DBS)
Train-Of-Four (TOF)
In this mode four stimuli are applied every 500 milliseconds (2 Hz). Each stimulation of the train
causes the muscle to contract. The fade in the individual response to each single stimulation provides a
basis for evaluation. The NMT Module then calculates a TOF ratio (TOFrat) value in % using the ratio
between the fourth and the first twitch. When fewer than four twitches are detected, the monitor
displays a TOF count (TOFcnt) value instead, this represents the number of responses to the four
TOF stimulation pulses.
294
20 Monitoring NMT
CAUTION
To avoid unintentional electrical shocks always make sure that the NMT stimulation has been stopped
before touching the electrodes.
NOTE
Ensure that the thumb can move freely before applying the NMT sensor.
Place the electrodes on the prepared site over the ulnar nerve on the palmar side of the wrist; refer to
the figure below.
295
20 Monitoring NMT
CAUTION
Make sure that the lead wires cannot come into contact with external pacemaker or catheter wires.
Arm Position
The arm used for the NMT measurement, should be kept immobile during the whole procedure.
NOTE
It is recommended that the patient be anesthetized before setting up the calibration twitch as nerve
stimulation can be painful.
Starting Calibration
1 Either press the NMT key on the module, or select the numeric on the screen to enter the Setup
NMT window.
2 Select Start Cal.
296
20 Monitoring NMT
3 Select Confirm.
This automatically determines the supramaximal stimulation current and takes the reference twitch
(100%) at this current. If the NMT Module is unable to establish a reference twitch, it uses an internal
reference value for single-mode measurements.
NOTE
This function is not available if a calibration is in progress or no calibration data is stored.
CAUTION
Changing the stimulation current or pulse width also clears stored references.
NOTE
To stop NMT stimulation in an emergency, disconnect the NMT patient cable from the NMT Module.
During an automatic measurement cycle the yellow Start/Stop LED on the NMT Module is lit. If the
automatic cycle is stopped, or the measurement mode is set to manual the LED is off.
The LED flashes (in both modes) when a stimulation pulse is given.
CAUTION
To avoid unwanted electrical shocks DO NOT touch the electrodes when the yellow Start/Stop LED
is flashing on the NMT module.
NOTE
Take care when removing the transducer from the patient. Do not pull on the cable.
297
20 Monitoring NMT
Measurement Mode
Allows the choice between manual and automatic NMT measurements.
In the Setup NMT window:
1 Select Mode to open the Mode window.
2 Select Auto or Manual mode.
NOTE
Auto mode is not available if the PTC or DBS stimulation mode is selected.
NOTE
This function is not available in the PTC and DBS mode.
CAUTION
Changing the stimulation current after calibration invalidates the stored reference data.
298
20 Monitoring NMT
CAUTION
Changing the stimulation pulse after calibration invalidates the stored reference data.
Value Lifetime
Displays the time after which an NMT numeric value disappears from the resting display.
The Value Lifetime setting is visible (grayed out) in the Setup NMT window. The setting can only be
changed in the Configuration Mode.
Alarms
TOFcnt High Limit Alarm
In the Setup NMT window:
1 Select the TOFcnt High Lim.
2 Select limit: 0, 1, 2, 3.
299
20 Monitoring NMT
Dependent upon the selected stimulation mode, the following information is provided:
Auto mode
The figure below shows the NMT measurement in Auto mode.
Manual mode
The figure below the NMT measurement in Manual mode.
Recovery Time
Most NMT stimulation modes require a minimum neurophysiological recovery time and during this
recovery phase no new stimulation can be started. Also, you cannot start a measurement or calibration.
300
21
Scoring
MX400/450/ Using an MX400/450/500/550 monitor with the appropriate option, you can use Guardian Early
500/550 Warning Scoring to get an early warning score based on vital signs and clinical observations collected at
intervals.
The goal of an early warning score is to help you recognize the early signs of deterioration in patients.
Depending on the score calculated, an Action List with appropriate recommendations is displayed.
There are no dedicated alarms associated with an Early Warning Score.
This functionality is highly configurable, to allow customization for the vital signs and observations
collected and the type of scoring used. Due to this configurability, your screens will probably not
exactly match the screens shown here. The workflow described here uses the special configuration
delivered with the monitor; if this configuration has been changed, you will not have all the
possibilities described.
WARNING
When using Guardian Early Warning Scoring, request a printout of the EWS scoring table and the
Action Lists from your configuration expert. Use them to verify that the configuration of the scoring
functionality matches your hospital’s requirements.
The IntelliVue Information Center will not display or trend any of the entered observations, the early
warning score, or the action list, but can pass these on to other connected systems.
301
21 Guardian Early Warning Scoring
3 Start collecting vital signs and entering observations. If the patient is continuously monitored, any
relevant currently measured vitals will appear automatically in the EWS Screen. Other vitals may
be measured by the monitor, or may need to be entered manually:
• If SpO2 is not already being measured, apply the SpO2 sensor and a short
time later the value will appear in the SpO2 box on the screen. Moving
dashes are displayed until the measured value is available.
• If the pulse rate is not already displayed from a continuously measured HR,
SpO2, or NBP, it will appear as soon as a blood pressure or SpO2
measurement has been made.
• Any vital sign or clinical observation you need to enter manually is indicated
by a keyboard symbol. Select the symbol to enter the missing data.
All vital signs and clinical observations with a white box are required, i.e. they must all be entered
for a complete SpotCheck record. If a keyboard symbol or start key is not in a white box, the
corresponding vital sign or clinical observation is optional.
When values have already been entered or measured, an existing white box will disappear.
302
21 Guardian Early Warning Scoring
One complete set of vitals and observations, together with the calculated score(s), is referred to as
a SpotCheck record. Once all the required vitals have been entered or measured, the SpotCheck
record will be automatically stored after the configured freeze time.
If you want to store values before the end of the freeze time, you can select Store Vitals, to store
the values and close data collection. The Store Vitals key can also be used to store an incomplete
set of vitals at any time.
If you want to break off the Early Warning Scoring process without storing the SpotCheck record,
use the Clear VS key.
4 To leave the EWS procedure and return to your standard monitoring screen, select the appropriate
screen from the screen list, or select the Previous Screen SmartKey (if configured).
WARNING
Alarms: There will be no alarm indications for measurements in the entry screen.
Values: The values shown in the entry screen do not have the same annotation as the measurement
values on other monitor screens. There is no special indication when a value has been manually
entered and no timestamp showing when the value was measured/entered. Be aware that values will
remain visible until the freeze time is at an end, even when the measurement itself is no longer
providing current values.
303
21 Guardian Early Warning Scoring
The MEWS score also uses the same color coding and risk levels, which determine the action that
is recommended.
How does MEWS work in Guardian Early Warning Scoring? MEWS results in a subscore
displayed in a circle with the corresponding color, next to each measured or entered value. Actions
are only then recommended when all required vitals have been measured or entered, and
the MEWS score has been calculated. The content of the action list depends upon the risk band
severity the MEWS score falls into. For each risk band severity (Normal, Observe - at risk,
Warning and Urgent) a separate action list can be configured.
1 This symbol indicates that more information is available than can be shown on the Screen. Select
the tabular trend to page through the remaining data columns.
Select the Graph/ Tabular pop-up key to view the data in graphical form. If any MEWS records are
present, they will be displayed in the color-coded risk bands (2) in the lower half of the graphical trend.
The MEWS score ranges corresponding to the colored bands are shown in the scale on the right.
Other record types, such as SPS, are displayed in the notification area (1). The symbols used are
explained in “Symbols Used in the Trend Views” on page 305.
Select the Detailed View pop-up key to see a combined graphical/tabular view for the currently
highlighted record. Use the cursor keys (right arrow and left arrow) to move from one record to the
next. The detailed data are shown above the graphical trend, with units and the corresponding SPS or
MEWS symbols. In the example below:
1 Colored circle for SPS data
2 MEWS score symbol (color coded according to the risk level)
3 MEWS subscore symbol (color coded according to the risk level)
304
21 Guardian Early Warning Scoring
Symbol Indicates
SPS SpotCheck record - normal range
Making Recordings
The trend data can be sent to a recorder using the recording pop-up keys.
• Record Selected makes a recording for the selected SpotCheck record.
• Record 8 h makes a recording for the last 8 hours of data.
• Record All makes a recording of all available SpotCheck records.
At the beginning of the recording, the title, the type of recording and the time will be printed. If
available, the patient's name, date of birth, ID and the type of scoring (protocol name) currently active
will also be printed.
305
21 Guardian Early Warning Scoring
2 Select Protocol.
3 Select the required protocol from the list.
or
1 Select Main Setup.
2 Select ProtocolWatch.
3 Select Protocol.
4 Select the required protocol from the list.
306
22
307
22 Using a Telemetry Device and a Monitor (PIIC only)
2 Select Telemetry
The Setup Telemetry menu will appear with only one entry Paired Equipment.
3 Enter here the equipment label of the telemetry device to be paired.
Pairing at the monitor is only possible when the monitor already has a connection to the Information
Center and the Information Center software version allows pairing at the monitor.
WARNING
• When ECG is being measured with a telemetry device directly connected to the monitor, there will
be no ECG signal available at the ECG analog output or ECG Sync Pulse output and no
synchronization marks on the ECG wave. A No ECG Out message will appear in the ECG wave
channel.
• When a telemetry device is connected to the monitor, arrhythmia relearning is initiated, and again
when the telemetry device is disconnected.
• Controls on the Telemetry Device (e.g. nurse call) will be inactive when the device is directly
connected to the monitor except in the case when the monitor has no network connection and
data are transferred via the telemetry device.
The X2 can also be manually paired to a host monitor without a direct connection, as described in
“Indirect Connection - Manual Pairing” on page 307.
Refer also to “Use Models With Telemetry” on page 309 for further related use modes.
308
22 Using a Telemetry Device and a Monitor (PIIC only)
• select the Unpair To Tele SmartKey to end pairing and have the Information Center receive the
measurement data from the telemetry device
or
use the Unpair function at the Information Center.
NOTE
The SmartKeys and pop-up keys for unpairing appear only on the monitor which is directly involved
in pairing.
309
22 Using a Telemetry Device and a Monitor (PIIC only)
310
23
23 Trends
Trends are patient data collected over time and displayed in graphic, tabular or histogram form to give
you a picture of how your patient's condition is developing. Trend information is stored in the trends
database for continuously-monitored measurements, such as ECG, as well as for aperiodically-
measured parameters, such as noninvasive blood pressure.
Viewing Trends
Trend information can be viewed embedded as a screen element on specially-designed Screens, or you
can open a trend window over the current Screen.
• To view trends embedded as a screen element, enter the Change Screen window, then select a
Screen designed to show an embedded trend window.
• To open the tabular trends window over the current Screen, select Main Setup, Trends, then
.
• To open the histogram trend window over the current screen, select Main Setup, Trends, then
Histogram, or select the Histo- gram SmartKey.
The trend windows open displaying the most recent data and are updated as new data is stored. A
timescale along the bottom or the top of the screen shows you where you are in the trends database.
The preview column on the right-hand side shows the latest available measurement value. The preview
column is updated every five minutes or whenever an NBP or other aperiodic measurement is
completed.
A question mark (?) beside a value means that the data may not be reliable, due perhaps to an INOP
condition in the measurement.
Your monitor screen may look slightly different to the examples shown in this chapter.
311
23 Trends
312
23 Trends
A cursor spanning all measurements in the trend group helps you to navigate through the trends
database and shows you your current position in the database. When the cursor is moved across the
time line, the values measured at the cursor time are shown in the right hand column.
In graphical trends, aperiodic measurement trends are shown as an asterisk, NBP has a special symbol.
To use the trend cursor to navigate in time through the trends database,
1 Select the graphical trend or the arrow pop-up keys to activate the cursor.
2 Use the arrow pop-up keys to move the trend cursor backwards and forwards in time, or
3 Place the cursor at a specific time by touching the graph.
313
23 Trends
To view the Vital Signs trend with one column for each NBP measurement (see “Defining the Column
Interval Using the NBP Measurement” on page 317):
1 With the Vital Signs trend open, select Interval.
2 Select NBP from the pop-up list.
The title line of the window shows the label of the trended measurement and the resolution of the
data; in the SpO2 histogram above, 1 second realtime samples.
The horizontal axis shows the range and unit of the displayed measurement. The vertical axis shows
the percentage of time.
The columns in the foreground show how much of the time the measured values fell into this range on
the scale. For example, in the histogram above, the SpO2 value was between 93 and 94% during 20%
of the last 12 hours. The display of the percentage value above each column, can be switched on or off
by selecting the histogram. The arrow mark over a column shows that the currently measured value is
also in this range. The columns are displayed in the same color as the measurement data.
The columns in the background show the cumulative percentage value: each of the foreground
columns is added to the sum of those columns to the left of it. The cumulative columns can be
switched off using the Curve On/Off pop-up key.
A question mark is displayed if less than two-thirds of the data are valid samples.
The pop-up keys in the window can be used to change the measurement, the time period and the
range. You can also print out a histogram report.
The SpO2 and System Pulse histograms can use trended data or realtime data with 1 second samples.
Using Cursors
You can set two cursors that span a "corridor", for example to divide a histogram into in-range and
out-of-range areas. Or you can set both cursors to the same value and use them as one cursor to divide
the histogram into an upper and a lower segment.
314
23 Trends
The cursors are represented by a vertical line from the x-axis and a dotted horizontal line from the y-
axis. When two cursors are used, they can be locked using Lock/ Unlock, so that the distance between
them is fixed. Moving a cursor will then move both together. Select Lock/ Unlock again to separate
the cursors and move them independently again.
A summary column on the y-axis shows the lower, middle and upper ranges as determined by the
cursor positions. When only one cursor is used there will be a lower and an upper range. Color shading
is used to differentiate the ranges in the cumulative column and to highlight the columns between the
cursors.
Setting Up Trends
Trends are set up in Configuration Mode. You can make temporary changes to trends settings such as
trend groups, priorities, or scales in Monitoring Mode. The general settings for all Trends are under
Main Setup, Trends.
Settings for a single segment in graphical trends or for the corresponding trend group can be made
more easily in that segment menu.
Expanded View
To expand the segment to fill the Graphical Trends window,
• in the Segment menu, select Expand to have that segment enlarged to fill the window.
In the expanded view, you can also highlight a measurement to make it more easily visible (for example
when multiple ST trends are in one segment). To highlight a measurement,
• in the Segment menu, select Highlight repeatedly until the highlight is on the required
measurement.
To return the segment to its original size,
• in the Segment menu, select Expand again.
Optimum Scale
To have the monitor automatically select an optimum scale for viewing, based on current values,
• in the Segment menu, select Optimum Scale.
This scale change is temporary. When the graphical trend window is closed the scale reverts back to
the setting in Parameter Scales.
315
23 Trends
Trend Group
To add or remove measurements for this trend group or change the order of the existing
measurements:
1 In the Segment menu, select Change Group.
2 Use the Add, Change, Sort Up and Sort Down pop-up keys to change or re-order the group as
required.
No. of Segments
In an embedded graphical trend window, you can select the number of segments to be displayed in the
Segment menu:
• In the Segment menu, select No. of Segments.
Trend Groups
The measurements grouped in trend groups define the trends displayed together in the Vital Signs or
Graphical Trends windows and printed in trends reports and recordings. The sequence of the
measurements in the group defines the order in which they are displayed. Measurements listed
between dotted line separators are displayed overlapping. The trend group All contains all available
measurements, you cannot change the order or selection of this group.
To change the selection of measurements in a trend group, either use the Change Group setting in the
Segment menu or:
1 Select Main Setup, Trends, then Trend Groups.
2 Select the Trend Group you want to change and use the pop-up keys to Add, Change, or Delete
the selection of measurements trended.
To temporarily change the order in which the measurements are displayed in a group,
1 Select Main Setup, Trends, then Trend Groups.
2 Select the Trend Group and then the measurement you want to move and use the Sort Up/Sort
Down pop-up keys.
Trend Interval
The trend interval defines the resolution of trend data shown on the Screen. High-resolution data is
especially suited for neonatal applications, where the clinical situation may change very quickly. In
adult monitoring, where the patient's status typically changes more gradually, a longer trend may be
more informative.
To set the trend resolution, in the Vital Signs or Graphical Trends window,
• Select the Select Interval pop-up key and then select the required interval from the list.
Trend Priority
The monitor stores trend information for all monitored measurements, if configured to do so. (Data
from VueLink or IntelliBridge modules cannot be included in trends when the label is a free-text label.)
316
23 Trends
If your configuration restricts the number of measurements trended, you must choose which
measurements will be included. A priority list is used to select the trended measurements.
To see the measurement priority list for trending,
1 In the Main Setup menu, select Trends.
2 Select Trend Priority.
To add measurements to the priority list,
1 Select the pop-up key Add and choose from the pop-up list of available measurements.
2 Use the Sort Up and Sort Down pop-up keys to change the priority order.
This sample ABP trend shows the continuously-measured values for the systolic, diastolic and mean
pressures displayed in band form.
317
23 Trends
• Trended Data and then select the required measurement from the trended data list.
318
23 Trends
Documenting Trends
To print a Vital Signs or Graphical Trends report,
• in the Vital Signs or Graph Trends window, select the pop-up key Print to print a report for the
trend group currently on the monitor screen.
Reports include the most recent information in the trends database and extend backwards in time
according to the selected trend interval and your monitor's configuration. Trends reports can be
printed on central or local printers.
To make a Vital Signs recording,
• in the Vital Signs window, select the Print/ Record pop-up key, then select the Record Vitals pop-
up key.
Vital Signs recordings print the trend data for the current group and trend period shown in the Vital
Signs window.
Trends Databases
Depending on the purchased options and the monitor's configuration, the trends databases store
information for up to 100 measurements for up to 96 hours.
The values in the trends database are stored as measured by the monitor, they are not averaged values.
The trend resolution defines how often a value is stored. In general, if more than one value is available
for storage in the trends database, the most recent value is used. Some values in the trends are marked
with arrows. This indicates that for this time period, more values were available and the most recent
one is shown.
Example database configuration
In this example, we see that the monitor stores the most recent data at the highest resolution, older
data are stored at a lower resolution.
319
23 Trends
"4 hours @ 12 second resolution" means that the monitor stores trend data every 12 seconds, for the
most recent four hours.
Screen Trends
Trends configured to display permanently on special monitor Screens are called screen trends. The
selection and number of measurement waves in the Screen Trend segment depends on your monitor
configuration. Screen trends are color-coded to match the measurement wave and numerics, and they
can be viewed in graphical, tabular, histogram or horizon format.
If you do not see screen trends on the monitor Screen, select a different Screen, one that is configured
to show screen trends. Screen trends are only available on specially designed Screens.
Screen Trend information is taken from the Trends database. If you do not see a Screen Trend for a
particular measurement on the Screen, check the trend priority list to ensure that this measurement is
being trended.
320
23 Trends
Tabular View
Aperiodic measurements such as NBP, C.O., C.I., and Wedge can be viewed as a screen trend in
tabular form. The measured values and their timestamps are shown, with the measurement label.
The trend time for tabular screen trends depends on the space available on the Screen. Up to 30
measurements or 12 hours information can be shown.
Histogram View
The histogram view presents a combination of graphical trend and histogram. The histogram is on the
right hand side in a horizontal presentation with vertical gridlines representing 25%, 50%, 75% and
100%. Each column shows the percentage of time that the measurement values were in a specific
range. This range is represented by the column's position in the graphical trend gridlines. The arrow
mark next to a column shows that the currently measured value is in the range covered by that column.
If less than two-thirds of the samples are valid, a question mark will be displayed with the histogram,
except when the histogram consists of intermittently measured data.
321
23 Trends
Horizon View
The horizon view presents trend information superimposed over a defined baseline or base range. This
helps you visualize changes in your patient's condition since the baseline was set.
The horizon view is made up of:
• a horizon, drawn in white, as a reference point or baseline to help you visualize changes in your
patient's condition. The horizon can be set to represent your patient's current condition, or a
target condition and can be a single value or a range.
• a graphical trend, displaying patient data for the set TrendTime (1).
• a trend indicator arrow, indicating how the patient trend has developed in the set time period
(10 minutes, 5 minutes or 2 minutes) (2).
• a deviation bar, showing how the currently measured value deviates from the set horizon (3). The
height of the deviation bar is an indication of the extent of the change in your patient's condition
relative to the (horizon) baseline.
Your monitor may not be configured to show all elements of the screen trend horizon view.
322
23 Trends
Be aware that changing the horizon trend scale can change the angle of the trend indicator, without the
patient's condition having changed.
If a measurement exceeds the outer limits of the scale, the wave will be clipped and you must either
reset the horizon or the horizon trend scale to display the values outside the scale limits.
323
23 Trends
324
24
24 Calculations
Calculations are patient data that is not directly measured but calculated by the monitor when you
provide it with the appropriate information.
Your monitor can perform the following hemodynamic, oxygenation, and ventilation calculations.
Pulmonary Vascular Resistance (PVR) Oxygen Consumption (VO2) Alveolar Ventilation (ALVENT)
The hemodynamic calculations available depend on the Cardiac Output measurement method being
used and the regulatory standards that apply for your hospital: see the “Monitoring Cardiac Output”
chapter for availability details.
325
24 Calculations
Viewing Calculations
Calculations Windows
This example calculations window shows the Hemodynamic Calculations window. The ventilation
and oxygenation windows are similar.
326
24 Calculations
Reviewing Calculations
To enter the calculations review window, select the Oxygen Review, Vent Review, or Hemo Review
pop-up key as required.
The review window lists all the input and output values for each measurement in the calculations
group. The timeline in the review window lists the times the calculations were performed.
To review individual calculations, select the calculation in the review window and then select the
Original Calc pop-up key.
The storage time for calculations is the same as the trend database storage time configured for the
monitor. So if trends are stored for 48 hours, the calculations will also be stored for 48 hours.
Performing Calculations
You must check that all input values are appropriate for your patient before performing calculations.
1 Select the Calcs SmartKey to open the Calculations window.
2 Check the calculation time in the Calc Time field.
When you enter the calculation window, this field will show either the current time or the time of
the most recent available C.O. measurement, depending on your monitor configuration.
– To choose a different calculation time, select the Calc Time field. This calls up a list showing
the timestamps of calculations performed earlier. Select a time from this list, or select Select
Time to enter a time of your choice. The values from the Vital Signs database from the 30
minute period before the selected time will be used.
– To enter the current time, select the Resample Vitals pop-up key. If you choose the current
time, the monitor will resample all the required values that are monitored.
3 Enter any values that must be entered or edited manually. Select the value field and then use the
pop-up keypad to enter the required values. Select Enter to confirm each entered value. Manually-
entered values are marked with an asterisk (*).
327
24 Calculations
BSA Formula
Your monitor provides both the Boyd and Dubois formulas for the calculation of body surface area
(BSA). For calculations, the monitor uses the setting defined in the Patient Demographics menu. All
calculation results that use BSA are indexed to the selected formula.
• To check the current setting, select the patient name to enter the Patient Demographics menu.
BSA(B) indicates that the Boyd formula is used; BSA(D) indicates that the Dubois formula is used.
328
24 Calculations
Documenting Calculations
To send a Calculations recording to a connected recorder, in the Calculations window, select the
Print/ Record pop-up key, then select the Record Calc pop-up key.
Calculations recordings print the patient demographic information and the content of the current
Calculations window on the recorder strip.
To print a report for the calculation group currently on the monitor screen, select the pop-up key Print
Calc. To print the review window, select the pop-up key Print in the review window. All the
calculations in the current group will be printed in the report.
Calculation reports can be printed on central or local printers.
This example report shows the oxygen calculation group. Ventilation and hemodynamic calculation
reports are similar.
1 Patient information
2 Calculation group
3 Three columns of calculations input and output values, with times, units and ranges, where
appropriate
329
24 Calculations
330
25
331
25 High Resolution Trend Waves
For each hi-res trend wave, a maximum of six minutes of measurement data will be printed.
See the “Printing Patient Reports” chapter for more information on setting up reports.
332
26
26 Event Surveillance
Events are electronic records of episodes in your patient's condition. You can trigger them manually,
or set them to trigger automatically during pre-defined clinical situations.
Depending on the level of event surveillance available on your monitor, the information stored for
each event episode can include:
• waveforms for up to four measurements of your choice (depending on episode type, see “Events
Pop-Up Keys” on page 335 for more details)
• numeric vital signs for all the measurements monitored
• any alarm conditions active when the event episode was triggered
• any annotations connected with the event.
You can navigate through the event database to view events retrospectively, and you can document
events on a recording or report marked with the patient name, bed label, and the data and time.
Trigger types Simple ("at least one") Combined ("at least two") Simple ("at least one")
333
26 Event Surveillance
Event Groups
The event group
• monitors the patient's signals to detect event triggers
• defines which waveforms are recorded in the event data.
In basic event surveillance only one event group can be active at a time, with Advanced Event
Surveillance all six groups can be active simultaneously. Active event groups monitor for event triggers.
Event groups are defined in Configuration Mode. In monitoring mode the groups can be adapted to
current conditions, for example episode types and threshold levels can be changed.
Event Episodes
When an event occurs, information for a predefined duration is stored. This is the event episode. It
includes information from a defined period before the trigger, called the event pre-time. The episode
time after the event is called the event post-time. If a further event occurs during the event post-time it
changes a single event to a combined event (combi-event).
Manually-triggered event episodes document patient information from the time leading up to the event
trigger; they do not have a post-time.
The episode type defines the level of detail captured in an event episode. The higher the data
resolution, the shorter the period that the monitor can store in its memory. High-resolution data is
334
26 Event Surveillance
suited for neonatal applications, where the clinical situation may change very quickly. In adult
monitoring, where the patient's status typically changes more gradually, a longer trend may be more
informative.
335
26 Event Surveillance
Event Triggers
You can trigger event capture manually, for example, if you want to record a patient's condition before
a procedure. You can also set events to trigger automatically, when the patient's values cross a
predefined threshold value, or when a particular measurement or procedure is carried out.
If more than one trigger is available for the measurements in the active event group, the trigger
condition may be At least one Param., At least two Param., At least three Par., or All four
Parameter. If the trigger is At least one Param. (this is short for "at least one measurement
parameter"), the monitor starts an event capture if a trigger occurs in any of this event group's
measurements. If the trigger is At least three Par., the monitor captures events when three or more
trigger thresholds from this event group's measurements are violated. With Enhanced Condition you
can not only select a minimum number of triggers to trigger an event but define which specific
measurement triggers these must be. For example, At least two Param. will cause an event to be
captured if a trigger occurs in any two of this event group's measurements - with Enhanced Condition
you can select that only when triggers are in HR and SpO2 an event will be captured.
The trigger condition for event groups is set in the monitor's Configuration Mode.
336
26 Event Surveillance
Event Retriggering
If a condition that triggered an event persists and the values remain beyond the trigger threshold, a
new event will not be triggered.
For a new event to be triggered by the same condition, the measured values for at least one of the
triggers must cross back into the normal range and then recross the trigger threshold.
Event Notification
Advanced You can be notified when an event is detected. For each event group you can define a type of
Event notification depending on the severity of the event conditions. The notification can be a status
Surveillance message with a prompt tone or a standard *, **, or *** alarm notification. These event alarms are
Only handled exactly like measurement alarms; they can be silenced and are also suspended when all alarms
are suspended. You should only use alarm notification for events which are comparable in severity to
standard measurement alarms to avoid potential confusion due to too many alarms. Notification in the
form of an alarm is not available when the trigger condition is At least one Param.. Selecting None
switches event notification off.
Setting the type of notification, or switching notification off, is done in Configuration Mode.
337
26 Event Surveillance
1 Select Main Setup then Event Surveill. followed by Setup Events to enter the Event Setup
window.
2 Select the name of the current episode type to set the episode pre/post time.
3 Set the event trigger for each measurement. Select each trigger name and select, if available, either
an alarm trigger, or a user-defined trigger from the pop-up trigger list. If you select a user-defined
trigger, set the required threshold level and delay time.
4 Set the trigger status to Activated to start event triggering. If the status is Deactivated event
surveillance is effectively switched off.
5 Select Confirm to confirm your changes.
1 Select the group name to enter the setup window for that group.
2 Set the trigger status to Activated, to have this event group trigger events, or Deactivated.
3 Set the episode type.
Select the name of the current episode type and select an episode type from the pop-up list. The
pre/post episode time for the selected episode type is displayed.
338
26 Event Surveillance
4 Set the trigger for each measurement. Select each trigger name and select either an alarm trigger or
a user-defined trigger from the pop-up trigger list. If you select a user-defined trigger, set the
required threshold level and delay time. If a deviation trigger is configured, set the deviation and
duration.
5 Select Confirm to confirm your changes.
339
26 Event Surveillance
Viewing Events
• To see a summary of all the events in every group in the event database, use the Event Summary
window.
• To review all the events in a particular event group, use the Event Review window.
• To review individual event episodes in detail, use the Event Episode window.
To start viewing events, either:
• in the Main Setup menu, select Event Surveill. and then select the event view you require from the
list, or
• select the Event Surveill SmartKey
and then select the event view you require from the list.
Vertical bars mark events in the Event Summary window. The timeline shows the position of the
stored events in the event database. Selecting this view activates a cursor that lets you navigate across
the timeline. Use the Show Episode pop-up key to select individual events for review in the Event
Episode window. It also calls up the events pop-up keys.
The event counter counts the total number of events in the database. If more than one event group
was set to trigger events within the event history, the event counter also counts the event group totals.
Counting Combi-Events
If one or more events occur during the same Episode Time, the monitor combines them and displays
them as distinct events in one event episode, called a combi-event. The first event is the trigger event,
and the others are follow-up events. For example, if an apnea event is followed 40 seconds later by a
brady event, the brady event is not counted as a single event but as part of the apnea event.
340
26 Event Surveillance
symbol meaning
group activated, notification set to ***alarm
Selecting the Event Summary window calls up the events pop-up keys.
Parts of the Event Summary window can also be embedded in Screens so that they are always visible,
for example the Total row showing the total number of events with the bars on the timeline or the
column showing all the groups with the activation and notification status.
When you open the Event Review window, it automatically shows the event group with the most
recent event.
341
26 Event Surveillance
• To view other event groups, select the pop-up key Review Group and select the group from the
list.
The event values to the left of the measurement channels show the trigger threshold set and the
maximum amount by which this limit was exceeded. In this example, Desat 71<85 tells you that 71
was the lowest SpO2 value measured during the event time and that the Desat trigger threshold was set
to 85 when the event was triggered. If the event was manually triggered, the event value boxes display
"manual".
Annotating Events
1 To annotate an event, in the Event Episode window, select the pop-up key Select Annotatn.
2 Select the required annotation from the pop-up list of available annotations for the currently active
event group.
Up to 20 annotations can be configured to let you add commonly-used clinical notes to event episodes
for documentation purposes. To see the complete list of available annotations, in the Event Setup
menu, select Event Annotation.
Documenting Events
You can print a report or make a recording of the events history stored in the database or of individual
event episodes or a Car Seat Assessment Record.
342
26 Event Surveillance
Event Recordings
Event recordings can be sent to a locally-connected M1116B/C recorder module.
343
26 Event Surveillance
In tabular event review recordings, the events stored in the event database are shown in chronological
order, with a number and time-stamp.
The measurements in the event group are shown in the next columns, marked "Parameter 1, Parameter
2...", along with the event values measured at the time of the event. For each event, the trigger values
are shown.
This section of the recording is A4 or letter size, so that it fits in a patient file.
344
26 Event Surveillance
The second section (2 below) shows the waveforms recorded during the episode. The trigger moment
is marked with a triangle and divides the episode into the pre/post time. Any calibration marks and
grid marks on the screen are automatically printed on the recording.
If there are four measurements in the event group being recorded, two waveforms will be recorded in
two separate waveforms segments.
The third section (3 above) shows the most important vital signs information, including numerics,
active alarms, and any annotations made on the event episode.
The fourth section (4 above) shows the numerics for all the currently monitored vital signs and any
alarm conditions or INOPs active at the time the event was triggered.
Event Reports
Event reports can be printed on A4 and letter size paper on a printer connected locally or centrally to
your monitor.
345
26 Event Surveillance
346
26 Event Surveillance
347
26 Event Surveillance
348
27
27 ProtocolWatch
ProtocolWatch (PW) is a clinical decision support tool. It allows you to run a clinical protocol which
can monitor developments in the patient's condition, taking into account:
• measured values from the monitor
• values manually entered by you (for example manual temperature measurements, lab values)
• your assessment of patient status
ProtocolWatch notifies you when certain conditions or combinations of conditions occur and it
documents developments in a log which can be printed.
349
27 ProtocolWatch
Sepsis Management Bundle - the recommendations to maintain patient status are listed and can be
checked off as they are implemented. The Sepsis Management Bundle ends when 24 hours have
passed since the Sepsis Resuscitation Bundle began.
The SSC Sepsis protocol may have been configured, to customize it for the specific treatment
measures used in your facility. This can include changes to limits for measured or manually entered
values, changes to the Severe Sepsis Screening criteria, and changes to the recommendations in the
Sepsis Resuscitation Bundle and the Sepsis Management Bundle. If configuration changes have been
made, your screens may not exactly match the screens shown here.
Depending on the option your monitor has, it may be that only the Severe Sepsis Screening phase is
implemented. If, at the end of this phase sepsis has been confirmed, and the patient is transferred for
treatment in the resuscitation and management phases, you can transfer the current state, the settings
and the log from the SSC Sepsis Protocol in the MMS to another monitor with the full version of the
SSC Sepsis Protocol. The transfer of data and settings must be enabled in Configuration mode.
WARNING
Always consider the specific clinical context for your patient, before following SSC Sepsis Protocol
treatment recommendations.
The SSC Sepsis Protocol and this description are based on the SSC Guidelines for Management of
Severe Sepsis and Septic Shock from January 2008 (including the Amendment from October 2011)
and the SSC Bundle definitions from January 2005. To check the version of the Guidelines and Bundle
definitions on the monitor,
• Select Main Setup then Revision then ProtocolWatch followed by SSC Sepsis
You can also place the ProtocolWatch symbol, together with the currently active phase, on the main
screen, in one of the numeric positions. This allows you to see at a glance which phase is currently
active, even when no SSC Sepsis Protocol window is currently displayed.
If you need to enter data or perform an action in an SSC Sepsis Protocol window, the symbol will turn
into a SmartKey.
350
27 ProtocolWatch
CAUTION
It is not possible to transfer SSC Sepsis Protocol data from a monitor with release F.0 software to
another monitor with release G.0 software or above, and vice versa.
351
27 ProtocolWatch
Changing Conditions
The window reappears if:
• a previously fulfilled criteria is again fulfilled
• a new HR or RR criteria is fulfilled,
• the HR or RR value which previously fulfilled the criteria now triggers a ** RR High, ** HR High
alarm or a ***Tachy xxx > yyy alarm,
• after 8 hours (configurable to 12 hours) if at least one infection sign is still present.
Lactate Measurement
To enter the Lactate value, select the Enter Lact key. If the value entered is > 4 mmol/l the check box
is automatically checked.
Select Confirm when a value has been entered. What follows depends on the Lactate value:
• If Lactate is > 4 mmol/l: the value meets the criteria for severe sepsis and a window appears
recommending authorized clinician review before entering the Sepsis Resuscitation Bundle.
• If Lactate is ≤ 4 mmol/l: a further window appears requiring your input to determine on the
basis of blood pressure values whether the patient meets the criteria for severe sepsis.
Hypotension Evaluation
If the Lactate value was not above 4 mmol/l, the next window asks whether the patient has persistent
hypotension.
The protocol defines Hypotension as:
• SBP < 90mmHg (12.0kPa), or
• MAP < 70mmHg (9.3kPa), or
• SBP decrease > 40mmHg (5.0kPa) below baseline.
You can see this definition in the window by selecting Show Details.
If the patient has persistent hypotension as defined, select Yes. This acknowledges that the patient
meets the criteria for severe sepsis and a final window appears recommending authorized clinician
review before entering the Sepsis Resuscitation Bundle.
If the patient does not have persistent hypotension, select No. After an hour a screen appears asking Is
the previously acknowledged infection still present?.
If you select Yes, the window for the Lactate measurement will reappear. If you select No, screening
continues - comparing heart rate, temperature and respiration rate values against the screening criteria.
352
27 ProtocolWatch
353
27 ProtocolWatch
354
27 ProtocolWatch
355
27 ProtocolWatch
356
28
28 Recording
There are two ways to record data using your monitor. If a paper-strip recorder is available, you can
make paper recordings, selecting from a variety of recording types. If no paper-strip recorder is
available, or if you want to record data in an electronic form, you can use electronic recording. See
“Paper-Strip Recording” on page 357 and “Electronic Recording” on page 365 for further details
Paper-Strip Recording
MX500/550/ The M1116B/C plug-in recorders record numerics for all active measurements and up to three
600/700/800 waveforms. You can use them for local recording mounted in the monitor's FMS.
357
28 Recording
M1116C
1 Continue LED ( ) - lights if a continuous
recording is ongoing.
Integrated Recorder
MX400/450 The optional integrated recorder records numerics for all active measurements and up to three
waveforms.
358
28 Recording
Central Recording
For central recording from the bedside, your monitor must be connected via a network to an
Information Center. You can use either the M1116B 2-Channel Recorder or the standalone M3160A
4-Channel Recorder. Recordings made on the M3160A may look slightly different to those described
here. See the documentation supplied with the Information Center for information on the 4-Channel
Recorder.
Quickstarting Recordings
To quickstart any type of recording using a pre-configured recordings template,
• Select the Recor- dings SmartKey and then select the pop-up key of the recording type you want
to start.
Alternatively, you can
• Select the Main Setup SmartKey, select Recordings, then select the recording type.
To quickstart a delayed recording,
• Select the SmartKey Delayed Record to immediately start a delayed recording.
M1116B/C You can also start a delayed recording by pressing the Run/Cont key on the recorder module.
only
Extending Recordings
Timed (non-continuous) recordings stop when their runtime is over. Continuous recordings continue
until stopped manually or by an INOP condition.
• To extend an ongoing recording by its runtime, reselect its Start pop-up key once.
• To make an ongoing recording continuous, reselect its Start pop-up key twice within 5 seconds.
M1116B/C You can also make an ongoing recording continuous by pressing the Run/Cont key on the recorder
only module.
359
28 Recording
Stopping Recordings
Recordings stop automatically when the preset runtime is over, when the recorder runs out of paper,
when you open the recorder door or when the recorder has an INOP condition.
• To manually stop a recording, select the Recor- dings SmartKey and then select the pop-up key
Stop All Recordng.
M1116B/C You can also stop a recording by pressing the Stop key on the recorder module.
only
360
28 Recording
361
28 Recording
• Overlap: define whether the recorded waveforms will be printed overlapping or beside each other.
• Speed: choose the recording print speed.
• Delay Time: Delayed recordings start documenting on the recorder strip from a pre-set time
before the recording is started. This interval is called the "Delay Time" and can be set as specified
in “Overview of Recording Types” on page 360. This setting can only be changed in Configuration
Mode.
• Run Time: see how long this type of recording is configured to run. This setting can only be
changed in Configuration Mode. Continuous recordings run indefinitely.
Central Config: if available in the General menu, select this setting to use the recording settings made
for the centrally-connected recorder.
Recording Priorities
Manually-started recordings have priority over automatically-started recordings. If an
automatically-triggered alarm recording is running, and a realtime or delayed recording is manually
started, the alarm recording is stopped and the manually-requested recording is started.
More recent manually-started recordings have priority over older manually-started recordings.
If a manually-started recording is running, and another manually-started recording is triggered, then
the older recording is stopped and the more recent manually-started recording is started.
Alarm recordings are prioritized according to alarm priority. If an alarm recording triggered by a
yellow alarm is running and a new alarm recording is triggered by a red alarm, the yellow alarm
recording is stopped and the red alarm recording is started.
Recording Strip
The information printed on the recording strip includes the patient name and MRN, bed number, date
and time of recording, recording speed, and (except when the information is printed above the waves)
recording code. Active alarm and INOP messages as well as numerics for all currently monitored
measurements are also printed.
Recording strip annotations are printed at the beginning of the recording strip, either before or above
the waves depending on the configuration. The annotations are updated at regular intervals, every 15
minutes for recordings made at speeds lower than 6.25 mm/s, and every 60 seconds for recordings
made at speeds greater or equal to 6.25 mm/s. When annotations are configured to appear above the
wave, it may not be possible to print them all when the runtime of the recording is too short.
362
28 Recording
Code Meaning
Recording type 90 Realtime
8A Delayed
0B Alarm
91 Context (Procedures)
Operating mode M Monitoring
D Demo
C Configuration
S Service
Application area I ICU
O OR
C CCU
N NICU
Patient category A Adult
P Pediatric
N Neonatal
Recorded Waveforms
A selection of up to three waveforms is recorded, marked with wave labels and wave scale information.
Wave scale information can be in the form of a calibration bar, like the 1 mV calibration bar for ECG,
or calibration steps before the waveform starts.
363
28 Recording
Reloading Paper
M1116B (MX500/550/600/700/800)
1 Use the latch on the right side of the recorder door to pull the door
open.
2 Remove the empty core.
3 Insert a new roll so that it fits snugly into its housing and the paper
end is feeding from the top. Recommended paper: 40477A and
40477B.
4 Pull out some paper and fold along the front edge at a 45° angle. This
makes it easier to feed the paper under the roller as shown.
5 Feed the paper through and pull some paper out from the top of the
roller.
6 Close the recorder door.
7 To test if paper is loaded correctly, start a recording. If no printing
appears, paper may be loaded backwards. Try reloading the paper.
CAUTION
When the recorder is disabled (for example, door open, or out of paper), any alarm recordings will be
sent to the central station recorder, if there is one. If no recorder is available, alarm recordings may be
lost during the time the recorder is disabled. The message No alarm recording available will be
displayed. This message is not shown if Printer is configured as the alarm recording destination.
364
28 Recording
Electronic Recording
Electronic recording allows alarm-triggered reports to be captured electronically in the monitor
database. They can then be printed when a printer is available. The printer can be locally connected at
the USB interface, a central printer connected to the Information Center or the XDS Printing Service
that is part of the IntelliVue XDS Application. With an XDS print server, the reports can be printed to
a standard off-the-shelf printer and can also be stored as files (e.g. .jpg, .bmp or .pdf) on the associated
PC.
All settings must be The output device for alarm recordings is set to "printer".
made in Configuration This automatically makes the alarm event documentation a realtime
mode. report.
The target device for realtime reports is set to "database"
The alarms to trigger the report are set up.
The database is set to print automatically when a printer is available.
When one of the selected alarms occurs, the monitor automatically captures the alarm and creates a
realtime report which is stored in the database.
365
28 Recording
As soon as the monitor is connected to a printer, or to a PC or network with the IntelliVue XDS
Application software, it will automatically print the report, or send it to a patient-specific folder as an
electronic file.
NOTE
The monitor cannot check for printer errors. If a report is sent to a printer but cannot be printed due
to an error, it may be lost. Use an XDS print server set up to save the reports as electronic files, to
ensure that no reports are lost.
366
28 Recording
The maximum number of reports that can be stored in the database varies depending on the database
configuration and the size of the reports. When the database is full, no new reports can be stored. In
this case, print or delete reports to make space available in the database.
367
28 Recording
368
29
Printout Location
Depending on availability, reports can be printed:
• on locally connected printers,
• on printers connected to the Information Center or the Application server,
• to an IntelliVue XDS Application print service,
• to an external PC-based print server with Philips server software,
• into the print database.
Print jobs stored in the database are automatically printed when a print device with a matching paper
size becomes available.
Print jobs in the print database are not deleted after a patient discharge. A new patient can be
admitted and their reports are saved in addition to the previous reports. Always admit your patient so
that reports can be clearly assigned to a patient.
369
29 Printing Patient Reports
When an X2 is connected to a host monitor, reports requested or stored on the X2 can be printed via
the host monitor. Reports requested on the X2 when connected to a host monitor will not show any
patient alarms or INOPs. In order to see alarms, reports should be requested on the host monitor.
370
29 Printing Patient Reports
Setting Up Reports
As the content of context-linked reports, such as Cardiac Output, Calculations, and Wedge, is defined
by the content of the procedure window these reports do not need to be set up, however a target
printer can be configured in the Setup Reports menu.
The content you define in the individual Setup Reports menus will be used in all reports of this type:
for example, if you set a print speed of 50 mm/sec in the ECG Reports menu, every ECG report will
be printed at this speed, irrespective of format.
ECG report layout and Auto Reports settings must be defined in Configuration Mode.
371
29 Printing Patient Reports
372
29 Printing Patient Reports
2 Select Report Type and then select the reports template you want the report selected in Step 2 to
use. The selection of templates listed depends on the report type selected.
Each template includes patient demographic data, and each Realtime Report also includes alarm
and INOP information. In addition, the templates include:
VisibleWaves: all waves currently visible, in the order they appear on the screen.
All Waves: all measured waves.
RT Waves: all currently measured realtime waves, according to the monitor's priority list.
HiRes Waves: all measured HiRes waves.
OxyCRG Waves: the OxyCRG/Neonatal Event Review waves.
Vital Signs: trend information in tabular form.
Graph Trend: trend information in graphic form.
ECG reports: ECG3x4, ECG6x2, ECG12x1, ECG4x2, ECG8x1,ECG12x1 (2P)
EEG reports: EEG Report.
Episode: a single patient event episode.
Review: an overview of patient events
Alarm Limits: a list of all currently set alarm limits.
3 Select Report Size to set the paper size to be used for the report: Unspecified to use the default
size for the template chosen, Universal, A4, Letter, LrgUniversal, A3, or Ledger. The list of
available sizes depends on the report type selected.
4 Select Orientation to set the orientation of the report printout: Unspecified to use the default size
for the template chosen, Landscape or Portrait.
5 Select Target Device and choose which printer the print job will be sent to: Unspecified to use
the default printer, or choose from the list of printer names defined at the Information Center or
in the monitor's Configuration Mode (for example, Remote 1 or Database).
Some settings may be inactive ("grayed-out") in this menu for reports that can only be started in a
special window.
373
29 Printing Patient Reports
If a section of a wave on a report is drawn with dashed lines, this tells you that a setting that affects the
appearance of the wave on the screen was changed while the report was printing.
For example, if you change the wave scale while a report is printing, the wave scale and wave size are
changed immediately on the monitor screen and on the report. To indicate that the scale information
printed at the beginning of the report no longer matches the currently used scale, the wave will be
printed using dashed lines, starting from the moment the change took place.
Some examples of settings that cause dashed lines in reports if changed during printing are: Filter
mode, ECG lead placement, wave scale, measurement unit, paced/non-paced setting, and
measurement mode. Note that as ECG waves are drawn with a fixed gain on reports (either 10 mm/
mV or 20 mm/mV), changing the ECG wave scale will not cause dashed-line reports.
To avoid dashed lines on reports, wait 15 seconds after changing a setting before you trigger a report.
374
29 Printing Patient Reports
• Enable the new target printer by selecting it in the Setup Printers menu and toggling to Enabled.
As the monitor tries to send the report to the printers in the order they are listed, you must make
sure that all the printers above the new target printer are disabled.
If the target device of the print job was set to a specific printer, re-routing is not possible.
CAUTION
The Reports Job List includes privacy information, in the form of the patient name with the related
report title and date. It is advisable to provide controlled access to this data to ensure confidentiality.
If an X2 is connected to a host monitor, all print requests which are stored in the print database are
shown "grayed out" on the Reports Job List and will not be deleted with the Delete All Repts key.
Printing Manually
Those jobs shown in black will be printed automatically when a matching printer is available. If the
Auto Print Dbs setting is Host Only or Never, some or all reports will not be printed automatically and
will be shown in gray. Any jobs shown in gray must be printed manually; to do this,
1 In the Reports Job List, select the required report
2 Select Print Report.
Selecting Print All Reps will send all reports to the printer.
375
29 Printing Patient Reports
376
29 Printing Patient Reports
The monitor may be configured to leave a space on the top left or right of the report printout to enable
you to stick a patient address label on it. This setting is called the Addressograph and it can only be
changed in the monitor's Configuration Mode.
1 Measurement labels, with alarms off symbol where alarms are switched off
2 Graphic view of current alarm limits in relation to currently monitored measurement value
377
29 Printing Patient Reports
Realtime Report
1 Patient information
2 Numbered trial curves
3 Trial information in tabular form
378
29 Printing Patient Reports
ECG Reports
1 Patient information
2 Numeric block
3 Wave area
Below the header on ECG Reports, the numeric block shows the current HR, PVC, and ST values.
The wave area shows the printed waves for all available ECG leads. A 1 mV calibration bar is printed
at the beginning of each wave. With the 3X4, 6X2, and 2X4 formats, a rhythm stripe prints a longer
section of the ECG wave from the primary ECG lead for ECG rhythm evaluation. The ECG signal
bandwidth, the patient's paced status, the ECG gain, and the print speed are printed at the bottom of
the wave area. Pace pulse marks are automatically printed beside the wave for paced patients. Beat
labels can be set to print on the rhythm stripe. The 12X1 (2P) format prints the report over two pages.
379
29 Printing Patient Reports
1 Patient data
2 Analysis data from the Information Center (PIIC iX only)
3 12-Lead wave area
4 Data related to the wave presentation
5 ST Map with related ST numerics (optional)
6 Administrative data - including optional custom text fields
7 Hospital ID data - Institution, facility and department, if configured
Other Reports
See the sections on Trends and Calculations and the chapter on Event Surveillance for other example
reports.
380
30
*Be aware that your hospital may use either 'µg' or 'mcg' as an abbreviation for microgram. These
abbreviations are equivalent.
WARNING
Before you administer any drug, always check that the correct calculation units and patient category are
selected. Consult your pharmacy if you have questions.
Decisions on the choice and dosage of drugs administered to patients must always be made by the
physician in charge. The Drug Calculator performs calculations based on the values input during use, it
does not check the plausibility of the calculations performed.
381
30 Using the Drug Calculator
CAUTION
JCAHO (Joint Commission of Accredited Healthcare Organizations) recommends disabling the Rule
of Six. The configuration is not JCAHO-compliant if Rule of Six is enabled.
The Drug Calculator uses the following formula for Rule of Six calculations, based on the patient's
weight:
• For a target dose of 0.1 mcg/kg/min, the Drug Calculator multiplies 0.6 x patient weight to
calculate the amount you need to add to the IV solution to equal a total of 100 ml.
• For a target dose of 1.0 mcg/kg/min, the Drug Calculator multiplies 6.0 x patient weight to
calculate the amount you need to add to the IV solution to equal a total of 100 ml.
382
30 Using the Drug Calculator
If a list of drugs has not been configured for your monitor, you can use the Drug Calculator to
calculate drug doses for a single, generic drug called ANY DRUG. Selecting the arrow beside the drug
name in the Drug Calculator window shows that are no other drugs configured.
1 Enter three of these four values: dose, amount, volume, rate of the infusion solution.
To enter values, select the correct unit, then select each value field and use the pop-up keypad to
enter the correct value.
2 If you have chosen a weight-dependent dose unit, you must enter the patient weight now or
choose a different unit.
If available, the patient weight from the Patient Demographics window is entered automatically in
the Drug Calculator window when the Drug Calculator is accessed. To change the patient weight,
select the Weight key then use the on-screen keypad to enter the correct value. This will not
change the patient weight stored in the patient demographic information.
3 When you have entered three values, the Drug Calculator automatically calculates the unknown
fourth value and shows it in the highlighted field. Standardized rate and concentration are also
calculated.
383
30 Using the Drug Calculator
1 Select the arrow beside ANY DRUG, then select the required drug from the list of drugs.
This opens the window for the selected drug.
2 Select the correct patient category for your patient.
If you have a neonatal or pediatric patient, the Rule of 6 choice may be available. Select if required.
3 Enter the patient weight, if necessary.
If available, the patient weight from the Patient Demographics window is entered automatically in
the Drug Calculator window. To change the patient weight, select the Weight key then use the on-
screen keypad to enter the correct value. This will not change the patient weight stored in the
patient demographic information. Select the weight unit shown to change between lb and kg.
4 When a specific drug is selected, the initial values for Dose, Amount and Volume are the
configured Start values for this drug. The Rate is then calculated. If other values are required you
can calculate any value by entering the remaining three values as described in “Performing
Calculations for a Non-Specific Drug” on page 383.
The column on the right of the window shows either drug dose or dosage rate, whichever is the
calculated value. The current calculated value is shown on a scale with the recommended range in
green. If the current calculated value lies outside the recommended range it is shown in red.
If you have changed values in the Drug Calculator and you want to revert to the configured values,
select Reset Values key at any time.
384
30 Using the Drug Calculator
The Drip Table shows you at a glance how much of the infusion has been administered to your patient
and how much time is left.
• To see the Drip Table, in the Drug Calculator window, select the Drip Table pop-up key.
If the Drip Time exceeds 24 hours, the Drip Time timestamp shows:
- - :- - : - - .
1 Currently set values for Dose, Rate, Amount, Volume, and Weight.
2 Minimum dose, maximum dose and increments used for the table.
3 Shows whether Dose increments or Rate increments are the basis of the table.
Use the Titration Table to see at a glance what dose your patient would receive of a drug at different
infusion rates. By clicking on the blue title row of the table you can switch between the Dose
increments and Rate increments view.
Values outside of the recommended range are shown in red.
To see the Titration Table,
• in the Drug Calculator window, select the pop-up key Titr. Table.
The Titration Table is configured with the service Support Tool.
385
30 Using the Drug Calculator
386
31
31 VueLink Modules
A VueLink module transmits information from a connected external device to your monitor. Each
module can be connected to one of up to three preselected external devices, and supports alarms from
the external device. Data from VueLink modules cannot be included in monitor trends when the label
is a free-text label.
The external device may show more information than is available on the monitor. The number of
waves and numerics you can view simultaneously on your monitor's main screen depends on the
module type. Type A modules support one wave and two numerics, type B modules support two
waves and six numerics.
1 module name
2 device label
3 selection LED
4 setup key
5 external device cable connector
6 setup indicator LED
7 module type (A or B)
The device labels (2) on the module indicate for which external devices the module is configured. The
selection LED (3) shows which device is currently active. The device label text may differ slightly from
the labels on the external devices.
387
31 VueLink Modules
See the documentation supplied with the VueLink module for a list of supported devices and
accessories, and for configuration information.
CAUTION
Selecting the wrong device can cause unpredictable system behavior. Rectify this by switching off the
external device when it is safe to do so, and selecting the correct device.
388
31 VueLink Modules
389
31 VueLink Modules
When Device Alarms Ignored is set, alarms from the external device will not be displayed on the
monitor and will not be transmitted to the Information Center. External device alarms status symbols
precede some, but not all, measurement labels.
the monitor is configured to accept external device alarms, but the alarms
are switched off at the external device.
390
32
32 IntelliBridge EC10
An IntelliBridge EC10 module or Interface board (for MX400/450/500/550) transmits information
from a connected external device to your monitor. The IntelliBridge EC5 ID module is used to
provide identification information from the external device. Data imported from the external device,
for example waveforms, measurement numerics, settings and alarms, can be displayed on the monitor,
passed on to an Information Center and included in trends as appropriate. The external device may
show more information than is available on the monitor. Data from IntelliBridge EC10 cannot be
included in monitor trends when the label is a free-text label.
See the documentation supplied with the IntelliBridge EC5 ID module for a list of supported devices
and accessories, and for configuration information.
391
32 IntelliBridge EC10
6 The connection status LED (4) will flash green until it has correctly identified the external device
and started communication. Check that the connection status LED then lights green continuously
indicating that communication has been established.
392
32 IntelliBridge EC10
the monitor is configured to accept external device alarms, but the alarms are
switched off at the external device.
Alarms from external devices are transmitted to the monitor. For all numerics configured in the setup
menu, an alarm condition is announced at the monitor. For one or more measurements not configured
in the setup menu, an alarm is announced as a text message for the highest priority alarm. Priority is
determined at the external device.
Refer to the documentation supplied with the IntelliBridge EC5 ID module for details of how alarms
are announced on the monitor and the Information Center.
393
32 IntelliBridge EC10
394
33
33 Using Timers
With the Timers function you can set timers to notify you when a specific time period has expired.
CAUTION
Do not use the timers to schedule critical patient-related tasks.
Viewing Timers
You can view currently active timers in the Timers window or directly on the Main Screen (if a timer
has been substituted for a numeric).
To open the Timers window:
Timer Types
There are four types of timer: Basic, Enhanced, Cyclic and No Limit. A Basic timer has a single,
defined run time and progress is shown in the progress bar. An Enhanced timer is like a Basic timer
but the progress bar shows progress beyond the end of the run time. A Cyclic timer is like a Basic
timer but restarts automatically when the run time is expired. A No Limit timer has no run time or
progress bar and shows the time elapsed since the timer was started.
The type of timer is set in the monitor's Configuration Mode and is associated with the timer label.
395
33 Using Timers
Setting Up Timers
The timer type, display color of the timer and whether a window automatically pops up on expiry are
settings made in Configuration Mode for each timer label.
The remaining timer settings can be made in Monitoring Mode. If you change settings when a timer is
running it will not be stopped. The timer will continue to count but the new settings will be applied.
To display the setup window for a timer,
• in the Timers window, select the required timer, then select the setup key or
• select a timer on the Screen then select Setup Timers.
Timer Label
You can select from a variety of specific labels, for example Tourniquet, Infus, Docu or from four
non-specific labels TimerA, TimerB, TimerC, and TimerD. When you assign a label to a timer, the
monitor automatically applies the associated configuration settings to this timer, but the timer
continues counting and is not reset.
To select a label, in the Timers window:
1 Select the required timer and display the Setup window.
2 Select Label.
3 Select a specific or non-specific label from the list.
Run Time
The run time can be set between 1 minute and 96 hours. No Limit timers have no run time.
To set the run time, in the Timers window:
1 Select the required timer and display the Setup window.
2 Select Run Time.
3 Select a run time from the list.
396
33 Using Timers
Notification
When a timer expires (except a No Limit timer), the color changes to red and a message appears in the
status line on the Main Screen. Be aware that if a timer has been configured to display in light red, the
color will not visibly change when the timer expires.
Additional notification is also available and can be set to:
Alarm: An INOP alarm indicates the expiry of the timer
Sound: A single tone indicates the expiry of the timer
No Sound: No additional notification
To set the additional notification, in the Timers window:
1 Select the required timer and display the Setup window.
2 Select Notification and select the notification level required from the list.
Timer Volume
For all Timers with the notification setting Sound, you can set the volume of the tone.
To set the volume, in the Timers window:
1 Select the required timer and display the Setup window.
2 Select Timer Volume and select a volume setting from the list.
397
33 Using Timers
398
34
34 Respiratory Loops
Using a Spirometry module or a VueLink/IntelliBridge module connected to a ventilator, you can
measure and store graphic representations of realtime respiratory loops. Respiratory loops can help
you recognize changes in your patient's lung function, and they can also indicate a fault in the airway
tubing (for example, if the respiratory loop does not close).
You can measure:
• Pressure-volume loops
• Pressure-flow loops
• Volume-flow loops.
Note that you cannot store loops from different patients and different source devices in the same list
as they are patient and device specific. This prevents you from inadvertently comparing information
from different patients.
Viewing Loops
In the loops display, the current loop is drawn in white, and up to six stored loops are color-coded to
match their timestamps. The currently-used source device is shown in the window title.
To view respiratory loops permanently on your Screen,
• select the current Screen name to open the Change Screen menu, then select a Screen configured
to display the loops screen element.
Select the loops screen element to access the loops pop-up keys.
399
34 Respiratory Loops
Showing/Hiding Loops
Colored rectangles beside the loops timestamps and color-coded with the loops tell you whether each
loop is currently displayed or not:
– A filled-in rectangle marks loops currently shown in the Loops window
– A rectangle outline marks loops not currently shown.
Selecting the timestamp of the currently-displayed loop hides it and shows the next in the list; selecting
the timestamp of a currently hidden loop displays it.
400
34 Respiratory Loops
Documenting Loops
1 In the Loops window, select the pop-up key Print Loop
2 From the list of available loops, select an individual loop, or select Print All to print a report of all
For each loop, the report prints
• the currently-displayed loop, with the loop capture timestamp
• up to six realtime numerics provided by the loop source device
• SpO2, etCO2, PO2, and PCO2 numerics from the patient monitor, if available.
401
34 Respiratory Loops
402
35
35 Laboratory Data
Laboratory data can be entered manually at the Information Center or at the monitor. For details on
manual entry at the monitor, see “Entering Measurements Manually” on page 46. Selected laboratory
data can be integrated on the monitor - values are then stored in the database and included in trends
and reports and can be displayed as a numeric on the main screen. Which data can be stored and used
on the monitor in this way is selected in Configuration Mode.
403
35 Laboratory Data
404
36
36 Using Batteries
MX400/ The Lithium Ion batteries used in your monitor store a large amount of energy in a small package. This
MX450/ allows reliable battery-operated monitoring but also requires care in use and handling of the batteries.
MX500/ Follow the instructions in this chapter and refer for further details to the Service Guide.
MX550 One Philips M4605A rechargeable Lithium Ion battery must be inserted into the battery compartment
on the right side of the monitor to use the monitor with battery power.
1 Battery compartment
You can switch between battery-powered and mains-powered (AC) operation without interrupting
monitoring.
The battery recharges automatically when the monitor is connected to mains power.
405
36 Using Batteries
Battery LED
The battery LED on the front panel of the monitor is indicated by a battery symbol.
1 Battery LED
This shows the remaining battery power. It is divided into sections, each representing 20% of the total
power. If three sections are filled, as in this example, this indicates that 60% battery power remains. If
no battery is detected, a blank battery gauge marked with a flashing red X is displayed. If no data is
available from the battery, a question mark is shown in the gauge.
If there are problems or changes in the status of the battery this is indicated by a blank battery gauge
containing a symbol. If the symbol is red, this indicates a critical situation.
406
36 Using Batteries
Monitoring Time Available: While the monitor is running on battery power, a time is displayed
below the battery power gauge. No time is displayed when the monitor is running on external power.
This is the estimated monitoring time available with the current battery power. Note that this time
fluctuates depending on the system load (the display brightness and how many measurements you
carry out).
Replacing a Battery
407
36 Using Batteries
Charging a Battery
A battery can be charged in monitors used to monitor patients. Charging is quicker in monitors which
are switched off. You can also use the 865432 Smart Battery Conditioner (previously M8043A) to
charge batteries.
1 Insert the battery into a monitor connected to mains power. The battery LED will light yellow to
indicate that charging is in process (it can take up to three minutes before charging begins).
2 Charge the battery until it is full, the battery LED is green, and the battery power gauge is filled.
In certain situations, where many measurements are in use plus the recorder, the load on the monitor
may be so high that the batteries will not charge. In this case you must use the 865432 Smart Battery
Conditioner (previously M8043A Smart Battery Conditioner) to charge the battery.
Conditioning a Battery
You must condition a battery when its "battery requires maintenance" symbol shows on the Screen.
Do not interrupt the charge or discharge cycle during conditioning. You can condition a battery in a
monitor.
CAUTION
Do not use a monitor that is being used to condition batteries to monitor patients. The monitor
switches off automatically when there is no battery power left.
It is, however, preferable to use the 865432 Smart Battery Conditioner (previously M8043A) for
external battery conditioning. The Smart Battery Conditioner automatically performs the correct
charge or conditioning process and evaluates the capacity when fully charged. For details please see the
Instructions for Use for the Smart Battery Conditioner. Do not use any other battery chargers or
conditioners.
To condition a battery using a monitor,
1 Insert the battery into a monitor connected to mains power.
2 Switch the monitor power off.
3 Charge the battery until it is completely full. Open the Battery Status window and check that the
Batt fully charged message is displayed.
4 Disconnect the monitor from mains power, and let the monitor run until there is no battery power
left and the monitor switches itself off.
5 Reconnect the monitor to mains power and charge the battery until it is full for use or charge to
50% for storage.
408
36 Using Batteries
Storing a Battery
WARNING
Remove the battery from the monitor when it is not used for a longer period of time, to avoid
potential damage caused by battery leakage.
409
36 Using Batteries
410
37
General Points
Keep your monitor, modules, Multi-Measurement Module, MMS extensions, Flexible Module Rack,
cables and accessories free of dust and dirt.
After cleaning and disinfection, check the equipment carefully. Do not use if you see signs of
deterioration or damage. If you need to return any equipment to Philips, decontaminate it first.
Observe the following general precautions:
• Always dilute cleaning agents according to the manufacturer's instructions or use lowest possible
concentration.
• Do not allow liquid to enter the case.
• Do not immerse any part of the equipment or any accessories in liquid.
• Do not pour liquid onto the system.
• Never use abrasive material (such as steel wool or silver polish).
• Never use bleach except in an approved product listed in this chapter.
WARNING
If you spill liquid on the equipment, battery, or accessories, or they are accidentally immersed in liquid,
contact your service personnel or Philips service engineer. Do not operate the equipment before it has
been tested and approved for further use.
The general care and cleaning information given here meets the requirements of Covidien llc for their
BIS measurement devices.
411
37 Care and Cleaning
CAUTION
To clean the touch-enabled display, disable the touch operation by switching off the monitor during
the cleaning procedure, or by selecting and holding the Main Screen key until the padlock symbol
appears on it, indicating that touch operation is disabled. Select and hold again to re-enable touch
operation.
Unplug a mouse, keyboard or other connected device before cleaning it.
412
37 Care and Cleaning
413
37 Care and Cleaning
4 Thread the cloth cleaning strip instead of paper around the rubber roller until approximately two
inches of the strip come out from the top of the roller.
5 Close the recorder door, aligning both ends of the strip over the top of the door.
6 Holding the top end of the cleaning strip between your thumb and forefinger, pull the strip
through and out of the recorder.
7 Open the door and ensure that the paper cavity is dust-free. Re-thread the paper and replace the
recorder.
414
38
Maintenance and
38
Troubleshooting
WARNING
Schedule: Failure on the part of the responsible individual hospital or institution employing the use of
this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure
and possible health hazards.
Contact: If you discover a problem with any of the equipment, contact your service personnel, Philips,
or your authorized supplier.
WARNING
Electrical Shock Hazard: Do not open the monitor or measurement device. Contact with exposed
electrical components may cause electrical shock. Always turn off and remove power before cleaning
the sensor, monitor or measurement device. Do not use a damaged sensor or one with exposed
electrical contacts. Refer servicing to qualified service personnel.
415
38 Maintenance and Troubleshooting
416
38 Maintenance and Troubleshooting
Troubleshooting
If you suspect a problem with an individual measurement, read the Instructions for Use and double-
check that you have set up the measurement correctly.
If you suspect an intermittent, system-wide problem call your service personnel. You may be asked for
information from the status log. To view the status log,
You can disassemble the monitor, MMS, FMS and modules as described in the Service Guide.
You will find detailed disposal information on the following web page:
http://www.healthcare.philips.com/main/about/Sustainability/Recycling/pm.wpd
The Recycling Passports located there contain information on the material content of the equipment,
including potentially dangerous materials which must be removed before recycling (for example,
batteries and parts containing mercury or magnesium).
Do not dispose of electrical and electronic equipment as unsorted municipal waste. Collect it
separately, so that it can be safely and properly reused, treated, recycled, or recovered.
417
38 Maintenance and Troubleshooting
WARNING
Ensure that the cylinder is completely empty before trying to remove the valve stem or drill a hole in
the cylinder.
418
39
39 Accessories
This chapter lists Philips-approved accessories for use with patient monitors as described in this
document. You can order parts and accessories from Philips supplies at www.medical.philips.com or
consult your local Philips representative for details. Some accessories may not be available in all
countries.
WARNING
Reuse: Never reuse disposable transducers, sensors, accessories and so forth that are intended for
single use, or single patient use only. Reuse may compromise device functionality and system
performance and cause a potential hazard.
Philips' approval: Use only Philips-approved accessories. Using non-Philips-approved accessories
may compromise device functionality and system performance and cause a potential hazard.
Packaging: Do not use a sterilized accessory if its packaging is damaged.
ECG/Resp Accessories
This symbol indicates that the cables and accessories are designed to have special protection
against electric shocks (particularly regarding allowable leakage currents), and are defibrillator proof.
The following cables may not all be available in all countries. Please check availability with your local
Philips supplier.
Recommended Cables
Trunk Cables
419
39 Accessories
420
39 Accessories
One-piece Cables
Radio-translucent Cables
Pack of five single wires, radio-translucent, 0.9 m, M1649A
421
39 Accessories
Supported Cables
Trunk Cables
422
39 Accessories
NBP Accessories
These cuffs and tubings are designed to have special protection against electric shocks
(particularly regarding allowable leakage currents), and are defibrillator proof. You can use them during
electrosurgery.
423
39 Accessories
424
39 Accessories
425
39 Accessories
426
39 Accessories
427
39 Accessories
Primary Extension
Flush Drip Quantity
Description Stopcocks Tubing Tubing Part No.
Device Chamber (per box)
Length length
Single DPT TP4, (152 cm, Two 3-way 122 cm (48 in) 31 cm (12 in) 3 ml/hr, Macrodrip 20 989803177901
60 in) Red dash tubing Squeeze
Dual DPT TP4, (183 cm, Four 3-way 152 cm (60 in) 31 cm (12 in) Two 3 ml/hr, Macrodrip 10 989803177911
72 in) Red/Blue dash Squeeze
tubing
Single DPT TP4, (213 cm, Two 3-way 183 cm (72 in) 31 cm (12 in) 3 ml/hr, Macrodrip 20 989803179721
84 in) Red dash tubing Squeeze
Triple DPT TP4, (152 cm, Six 3-way 122 cm (48 in) 31 cm (12 in) 3 ml/hr, Macrodrip 10 989803179731
60 in) Squeeze
Single DPT TP4, (183 cm, Two 3-way 152 cm (60 in) 31 cm (12 in) 3 ml/hr, Macrodrip 20 989803179771
72 in) Red dash tubing Squeeze
Single DPT TP4, (23 cm, One 3-way 23 cm (9 in) 3 ml/hr, Macrodrip 20 989803179871
9 in) Patient Mount Squeeze
Single DPT TP4, (61 cm, One 4-way 61 cm (24 in) No flush 20 989803181141
24 in) Compartmental One 3-way device
Pressure
Single DPT TP4, (152 cm, Two 3-way 122 cm (48 in) 31 cm (12 in) 3 ml/hr, Macrodrip 20 989803181211
60 in), Premium Stripe Squeeze
Single DPT TP4, (183 cm, Two 3-way 152 cm (60 in) 31 cm (12 in) 3 ml/hr, Macrodrip 20 989803181221
72 in), Premium Stripe Squeeze
Single DPT TP4, (213 cm, Two 3-way 183 cm (72 in) 31 cm (12 in) 3 ml/hr, Macrodrip 20 989803181231
84 in), Premium Stripe Squeeze
Dual DPT TP4, (183 cm, Four 3-way 152 cm (60 in) 31 cm (12 in) 3 ml/hr, Macrodrip 10 989803181241
72 in), Premium Stripe Squeeze
Triple DPT TP4, (152 cm, Six 3-way 122 cm (48 in) 31 cm (12 in) 3 ml/hr, Macrodrip 10 989803181251
60 in), Premium Stripe Squeeze
428
39 Accessories
Primary Extension
Flush Drip Quantity
Description Stopcocks Tubing Tubing Part No.
Device Chamber (per box)
Length length
Single Neonatal DPT TP4 Two 3-way 61 cm (24 in) 30 ml/hr 20 989803179841
(61 cm, 24 in) Squeeze
Single Neonatal DPT TP4 Three 3- 31 cm (12 in) 30 ml/hr 20 989803179851
(31 cm, 12 in) way Squeeze
Single Neonatal DPT TP4 Three 46 cm (18 in) 30 ml/hr 20 989803179881
(46 cm, 18 in) one in-line 3-way Squeeze
Luer sampling port &
10 ml In-line reservoir
SpO2 Accessories
This section lists accessories for use with Philips SpO2 technology. For accessory lists for other SpO2
technologies, refer to the instructions for use provided with these devices.
Some Nellcor sensors contain natural rubber latex which may cause allergic reactions. See the
Instructions for Use supplied with the sensors for more information.
Do not use more than one extension cable with any sensors or adapter cables. Do not use an extension
cable with:
• Masimo adapter cables,
• Philips reusable sensors or adapter cables with part numbers ending in -L (indicates "Long"
version) or with a cable longer than 2 m.
All listed sensors operate without risk of exceeding 41°C on the skin, if the initial skin temperature
does not exceed 35°C.
Make sure that you use only the accessories that are specified for use with this device, otherwise patient
injury can result.
Always use the MAXFAST forehead sensor with the foam headband provided by Nellcor.
Option A01 is the Philips FAST-SpO2 version; Option A02 is the Nellcor OxiMax-compatible
version.
429
39 Accessories
The SpO2 Option A02 may not be available in all countries. Some sensors may not be available in all
countries.
430
39 Accessories
Nellcor Sensors
Nellcor sensors must be ordered from Nellcor/Covidien.
431
39 Accessories
432
39 Accessories
433
39 Accessories
434
39 Accessories
Temperature Accessories
Temperature Probes Part No. Minimum measurement
time for accurate
readings
Reusable
General purpose probe 21075A 90 sec
Small flexible vinyl probe (Infant/Pediatric) 21076A 60 sec
Attachable surface probe 21078A 60 sec
Disposable
General purpose probe M1837A 90 sec
Skin probe 21091A 60 sec
Esophageal/Stethoscope Probe (French 12) 21093A 180 sec
Esophageal/Stethoscope Probe (French 18) 21094A 210 sec
Esophageal/Stethoscope Probe (French 24) 21095A 310 sec
Esophageal/Rectal Probe (French 12) 21090A 90 sec
Foley Catheter Probe (12 French) M2255A 180 sec
Foley Catheter Probe (16 French) 21096A 180 sec
Foley Catheter Probe (18 French) 21097A 180 sec
Adapter cable 1.5 m 21082B --
Adapter cable 3.0 m 21082A --
435
39 Accessories
436
39 Accessories
437
39 Accessories
NOTE
Not all accessories are available in all countries.
438
39 Accessories
Spirometry Accessories
Description Part No.
Adult/Pediatric Flow Sensor M2785A
Neonatal Flow Sensor M2786A
Adult/Pediatric CO2/Flow Sensor M2781A
Neonatal CO2/Flow Sensor M2782A
Pediatric CO2/Flow Sensor M2783A
tcGas Accessories
This symbol indicates that the specified transducer (but not its membranes) is designed to have
special protection against electric shocks (particularly regarding allowable leakage currents), and is
defibrillator proof.
439
39 Accessories
EEG Accessories
Description Part No.
Trunk Cable 2.7 m M2268A
Trunk Cable, 1.0 m M2269A
Reusable 80-cm-long 5-lead cables with 10 mm silver/silverchloride leadwired cup electrodes M1931A
(Adult)
Reusable 80 cm 5-lead cables with 6 mm silver/silverchloride leadwired cup electrodes M1932A
(Pediatric/Neonatal)
Reusable 80 cm 5-lead cables with clip M1934A
Disposable EEG electrodes M1935A
EC2™ Electrode Cream (conductive paste) M1937A
BIS Accessories
To re-order sensors, contact Covidien llc. BIS sensors are not available from Philips.
440
39 Accessories
NMT Accessories
Description Product Number (Order Number)
IntelliVue NMT Patient Cable (reusable) 989803174581
Symbol Meaning
Temperature limitations for storage
Contents of the packaging are not made with natural rubber latex.
Manufacturer information
441
39 Accessories
Recorder Accessories
Description Part No.
For the M1116B recorder:
10 rolls of paper 40477A
80 rolls of paper 40477B
For the integrated recorder (MX400/MX450) and the M1116C recorder
10 rolls of paper M4816A
80 rolls of paper M4817A
Battery Accessories
Description Part No.
Lithium Ion Smart Battery 1Ah (internal battery for X2) M4607A
Lithium Ion Smart Battery 6Ah (for MX400/MX450/MX500/MX550) M4605A
Battery Charger and Conditioner (requires size adapter 451261017451 to charge M4607A 865432
battery)
Size adapter for M4607A battery 451261017451
442
40
40 Specifications
The specifications in this section apply to the MX 400/MX450, MX500/MX550, MX600, MX700 and
MX800 patient monitors.
The monitors are not user installable. They must be installed by qualified service personnel.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for,
multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for
use by trained healthcare professionals in a hospital environment.
The MX400/MX450/MX500/MX550 monitors are additionally intended for use in transport
situations within hospital environments.
The monitors are only for use on one patient at a time. They are not intended for home use. Not
therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed
morphology of complex cardiac complexes (according to AAMI EC 11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for
use with neonatal and pediatric patients.
The transcutaneous gas measurement (tcGas) is restricted to neonatal patients only.
BIS is intended for use under the direct supervision of a licensed health care practitioner or by
personnel trained in its proper use. It is intended for use on adult and pediatric patients within a
hospital or medical facility providing patient care to monitor the state of the brain by data acquisition
of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents.
Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of
the incidence of awareness with recall in adults during general anesthesia and sedation.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with
adult patients only.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years)
patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients
receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV
measurement has been validated only for adult patients.
443
40 Specifications
Restricted Availability
Following new features and functionality may not be available in all geographies:
• Smart Alarm delays
• IntelliVue Information Center iX
Use Environment
Hospital Environment
The monitors are suitable for use in all medically used rooms which fulfill the requirements regarding
electrical installation according to IEC 60364-7-710 "Requirements for special installations or locations
- Medical locations", or corresponding local regulations.
WARNING
The monitors are not intended for use in an MRI environment or in an oxygen-enriched environment
(for example, hyperbaric chambers).
Manufacturer's Information
You can write to Philips at this address
Philips Medizin Systeme Boeblingen GmbH
Hewlett-Packard-Str. 2
71034 Boeblingen
Germany
Visit our website at: www.healthcare.philips.com/us/.
© Copyright 2013. Koninklijke Philips N.V. All Rights Reserved.
444
40 Specifications
Symbols
These symbols can appear on the monitor and its associated equipment.
Symbols
Caution, refer to Protective earth Equipotential
accompanying grounding
documents
445
40 Specifications
Symbols
Electrical input Electrical output Connector has special
indicator indicator protection against
(In some cases gas (In some cases gas electric shocks and is
input indicator) output indicator) defibrillator proof (Type
CF according to
IEC 60601-1)
Gas input indicator Gas output indicator Quick mount release
Purge flow tubing with Zero the invasive Enter the measurement
air pressure transducer setup menu
ECG Sync Output/ Followed by two Indicates location of
Analog ECG Output alphanumeric catalog number
characters, indicates
ingress protection
grade
Indicates location of Rechargeable battery Battery LED
service number symbol
Indication for battery Enter the
placement in battery measurement setup
compartment menu to start a
calibration
446
40 Specifications
Grounding The monitor must be grounded during operation. If a three-wire receptacle is not available, consult the
hospital electrician. Never use a three-wire to two-wire adapter.
Equipotential Grounding If the monitor is used in internal examinations on the heart or brain, ensure that the room incorporates an
equipotential grounding system to which the monitor has a separate connection.
Combining equipment Combinations of medical equipment with non-medical equipment must comply with IEC 60601-1-1. Never
use a multiple portable socket-outlet or extension cord when combining equipment unless the socket outlet is
supplied specifically for use with that equipment.
Fusing The monitor uses double fusing (line and neutral).
Network Cables All network cables must be unshielded.
Connectors
The actual placement of boards and configuration of connections for your monitor depends on how
your hardware has been configured. See the symbols table (“Symbols” on page 445) to see which
symbols are used to mark the connections.
WARNING
Connect only medical devices to the ECG output connector socket.
447
40 Specifications
MX400/450
Rear of the Monitor
Shown here without the standard spillage cover that protects the connectors.
448
40 Specifications
Optional Interfaces
449
40 Specifications
450
40 Specifications
MX500/550
Rear of the Monitor
Optional Interfaces
451
40 Specifications
452
40 Specifications
453
40 Specifications
MX600/700/800
Rear of the Monitor
1 Serial/MIB (RS232) interface cards, type RJ45 or Flexible Nurse Call interface card or
combination of both (optional)
2 AC power input
3 Protective earth screw hole
4 Equipotential ground connector
5 Measurement link connectors (MSL)
(Two standard in MX800, one standard in MX600/700)
6 USB rear connectors (for remote control, keyboard, pointing devices, printer)
7 Serial RS232 connector
8 Nurse Call
9 Wired network connector
10 Video out connector (digital/analog)
The following connectors are only present with the iPC
11 USB rear connectors (for keyboard, pointing devices, printer)
12 Audio in/out
13 Local Area Network
14 Video out connector (digital/analog)
The following connectors are only present with the Independent Display Interface
15 Serial RS232 connector for touch
16 Video out connector (digital/analog)
454
40 Specifications
455
40 Specifications
CAUTION
• Never carry a monitor by the handle when equipment is attached below it with the Quick Mount.
To transport such equipment combinations, always support the equipment from below.
• Only use the bed hanger for the MX400/450/500/550 on beds with a support below the rail, to
ensure that the monitor does not tilt.
WARNING
Not adhering to these instructions when repositioning the monitor can cause damage to the monitor.
In extreme cases, when force is applied to the monitor, it can result in the monitor falling from the
mounting arm.
If, at any time, the monitor appears to be loose or insecurely mounted, contact your service personnel.
Altitude Setting
Altitude affects tcGas and CO2 measurements. The monitor must be configured at installation to the
correct altitude.
The monitors, together with the Multi-Measurement Modules (M3001A/M3002A), the Flexible
Module Racks FMS-8 (M8048A) and FMS-4 (865243), and all modules and MMS extensions, comply
with the Medical Device Directive 93/42/EEC.
Covidien
Covidien BISx, Covidien BIS Engine and Covidien DSC comply with the Medical Device Directive
93/42/EEC of 14 June 1993.
In addition, the product complies with:
IEC 60601-1:1988 + A1:1991 + A2:1995; EN 60601-1:1990 + A1:1993 + A2:1995; UL 60601-1:2003;
CAN/CSA C22.2#601.1-M90 +Suppl. No. 1-94 + Am. 2; IEC 60601-1-1:2000; EN 60601-1-1:2001;
IEC 60601-1-2:2001 + A1:2004; EN 60601-1-2:2001 + A1:2006.
The possibility of hazards arising from software errors was minimized in compliance with
ISO 14971:2007, EN 60601-1-4:1996 + A1:1999 and IEC 60601-1-4:1996 + A1:1999.
Classification (according to IEC 60601-1): Class 1, Type CF, Continuous Operation. The BIS and
NMT measurement use a Type BF applied part.
456
40 Specifications
Physical Specifications
Product Maximum Weight WxHxD Comments
MX800 Monitor <12 kg (<26.4 lb) <478 x 364 x 152 mm without cable
<18.82 x 14.33 x 5.98 in management
MX600/MX700 Monitor <9.5 kg (<20.9 lb) <392 x 321 x 163 mm without cable
<15.43 x 12.64 x 6.42 in management
MX550 Monitor <7.3 kg (<16.1 lb) <404 x 308 x 184 mm
<15.91 x 12.13 x 7.24 in
MX500 Monitor <6.0 kg (<13.1 lb) <327 x 288 x 182 mm
<12.87 x 11.34 x 7.17 in
MX450 Monitor <5.9 kg (<12.9lb) <327 x 288 x 182mm
<12.87 x 11.34 x 7.17 in
MX400 Monitor <5.2 kg (<11.5lb) <274 x 288 x 178 mm
<10.79 x 11.34 x 7.01 in
M3001A <650 g (<1.4 lb) 188 x 96.5 x 51.5 mm
Multi-Measurement Module (MMS) 7.4 x 3.8 x 2 in
M3002A <1.25 kg (<2.8 lb) 188 x 99 x 86 mm including battery,
Multi-Measurement Module (MMS) 7.4 x 3.9 x 3.4 in without handle or
options
M3012A <550 g (<1.2 lb) <190 x 98 x 40 mm
Hemodynamic MMS Extension <7.5 x 4 x 1.6 in
M3014A <500 g (<0.99 lb) <190 x 98 x 40 mm
Capnography MMS Extension <7.5 x 4 x 1.6 in
M3015A <550 g (<1.21 lb) <190 x 98 x 40 mm
Microstream CO2 MMS Extension <7.5 x 4 x 1.6 in
M3016A <450 g (<0.99 lb) <190 x 98 x 40 mm
Mainstream CO2 MMS Extension <7.5 x 4 x 1.6 in
M8048A <3500 g (<7.7 lb) <320 x 135 x 120 mm without plug-in
8-Slot Flexible Module Rack (FMS-8) 12.6 x 5.3 x 4.7 in modules
865243 <950 g (<2.09 lb) <194 x 139 x 110 mm without MMS Mount
4-Slot Flexible Module Rack (FMS-4) <7.64 x 5.47 x 4.33 in
M8025A <300 g (<0.7 lb) 62 x 125 x 63 mm
Remote Alarm Device 2.4 x 5 x 2.5 in
M1006B 190 g (6.7 oz) 36 x 99.6 x 97.5 mm
Invasive Press Module Option #C01: 225 g (7.9 oz) 1.4 x 3.9 x 3.8 in
M1029A 215 g (7.6 oz) 36 x 99.6 x 97.5 mm
Temperature Module 1.4 x 3.9 x 3.8 in
M1012A 225 g (7.9 oz) 36 x 99.6 x 97.5 mm
Cardiac Output Module 1.4 x 3.9 x 3.8 in
M1014A 250 g (8.8 oz) 36 x 99.6 x 97.5 mm
Spirometry Module 1.4 x 3.9 x 3.8 in
M1018A 350 g (11.3 oz) 72.5 x 99.6 x 97.5 mm
Transcutaneous Gas Module 2.9 x 3.9 x 3.8 in
M1020B <250 g (<0.55 lb) 36 x 99.6 x 97.5 mm
SpO2 Module 1.4 x 3.9 x 3.8 in
M1021A 460 g (13.04 oz) 72.5 x 99.6 x 97.5 mm
Mixed Venous Oxygen Saturation Module 2.9 x 3.9 x 3.8 in
457
40 Specifications
Environmental Specifications
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all products.
458
40 Specifications
MX400/450/500/550 Monitor
MX400/450/500/550 Monitor
Item Condition Range
Temperature Range Operating 0 to 40 ºC (32 to 104 ºF)
0 to 35°C (32 to 95°F) when charging the
battery
0 to 35°C (32 to 95°F) if an X2 is mounted
on the rear.
0 to 35°C (32 to 95°F) if IntelliVue
Instrument Telemetry Wireless Network is
connected.
Storage -20 to 60°C (-4 to 140°F
Humidity Range Operating 15% to 95% Relative Humidity (RH) (non
condensing)
Storage 5% to 95% Relative Humidity (RH) (non
condensing)
Altitude Range Operating -500 m to 3000 m (10000 ft)
Storage -500 m to 4600 m (15000 ft)
Ingress Protection IP21
MX600/700/800 Monitor
MX600/700/800 Monitor
Item Condition Range
Temperature Range Operating 0 to 40°C (32 to 100°F)
0 to 35°C (32 to 95°F) when equipped with the iPC
Storage -20 to 60°C (-4 to 140°F)
Humidity Range Operating 15% to 95% Relative Humidity (RH) (non condensing)
Storage 5% to 95% Relative Humidity (RH)
Altitude Range Operating -500 m to 3000 m (-1600 to 10000 ft)
Storage -500 m to 4600 m (-1600 to 15000 ft)
Ingress Protection IPX1
459
40 Specifications
460
40 Specifications
461
40 Specifications
Remote Control
Remote Control 865244
Item Condition Range
Temperature Range Operating 0 to 40°C (32 to 100°F)
Storage -20 to 60°C (-4 to 140°F)
Humidity Range Operating 15% to 95% Relative Humidity (RH) (non condensing)
Storage 5% to 95% Relative Humidity (RH)
Altitude Range Operating -500 m to 3000 m (-1600 to 10000 ft)
Storage -500 m to 4600 m (-1600 to 15000 ft)
The radio component contained in this device is in compliance with the essential requirements
and other relevant provisions of Council Directive 1999/5/EC (Radio Equipment and
Telecommunications Terminal Equipment Directive)
In addition the product complies with: ETSI EN 300 328; AS/NZS 4771+A1; ARIB STD-T66.
Class 1 radio equipment.
To obtain a copy of the original Declaration of Conformity, please contact Philips at the address given
in the “Manufacturer's Information” section of these Instructions for Use.
462
40 Specifications
The device for the band 5150-5250 MHz is only for indoor usage to reduce potential for harmful
interference to co-channel mobile satellite systems.
CAUTION
High power radars are allocated as primary users (meaning they have priority) of 5250-5350 MHz and
5650-5850 MHz and these radars could cause interference and /or damage to LE-LAN devices.
This device is compliant to Council Directive 2006/95/EC (Low voltage directive) & 2004/
108/EC (EMC directive) & 1999/5/EC (Radio Equipment and Telecommunications Terminal
Equipment Directive)
The radio device used in this product is in compliance with the essential requirements and
other relevant provisions of Directive 1999/5/EC (Radio Equipment and Telecommunications
Terminal Equipment Directive). Class 2 radio equipment. Member states may apply restrictions on
putting this device into service or placing it on the market. This product is intended to be connected to
the Publicly Available Interfaces (PAI) and used throughout the EEA.
The radio device used in this product is in compliance with the essential requirements and
other relevant provisions of Directive 1999/5/EC (Radio Equipment and Telecommunications
Terminal Equipment Directive). Class 2 radio equipment. Member states may apply restrictions on
putting this device into service or placing it on the market. This product is intended to be connected to
the Publicly Available Interfaces (PAI) and used throughout the EEA.
463
40 Specifications
464
40 Specifications
MX400/450/500/550
Sounds Audible feedback for user input
Prompt tone
QRS tone, or SpO2 modulation tone
4 different alarm sounds
Remote tone for alarms on other beds in network
Tone for Timer expired
Display Wave Speeds Available for standard waves 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
with ±5% accuracy (guaranteed only for integrated displays)
Available for EEG and BIS waves 6.25 mm/s, 12.5 mm/s, 15 mm/s, 25 mm/s, 30 mm/s, 50
mm/s
with ±5% accuracy (guaranteed only for integrated displays)
Trends Resolution 12, 16, 24 or 32 numerics @ 12 sec, 1 minute, 5 minute
resolution
Information Multiple choices of number of numerics, resolution and
duration depending on trend option and application area.
For example:
neonatal extended 12 numerics, 24 hours @ 12 secs or
32 numerics 32 hours @ 1 minute
intensive care extended: 16 numerics 120 hours @ 5 minutes
anesthesia extended 32 numerics 9 hours @ 12 seconds
High Res Trend Waves Measurements available HR, SpO2, Resp, tcpO2, Pulse, Perf, tcpCO2, CO2, ABP, PAP,
CVP, ICP, CPP, BIS, CCO, AWP, Anesthetic Agents, Delta
SpO2, inO2
Resolution Measurement samples are taken at a resolution of four samples
per second
Update speed Waves are drawn at a speed of 3 cm/minute
Events Information Trigger condition and time, event classification and associated
detailed view of episode data
Episode data Configurable, either:
4 minutes of high resolution trend or
20 minutes of numerics trend @ 12 sec. resolution or
15 seconds of 4 waves @ 125 samples/sec. (Snapshot)
including all current numerics, alarms and inops
Capacity (max.) 25 or 50 events for either 8 or 24 hours
465
40 Specifications
MX400/450/500/550
Alarm Signal System alarm delay less than 4 seconds
The system alarm delay is the
processing time the system needs
for any alarm to be indicated on the
monitor, after the measurement has
triggered the alarm.
Delay for alarm availability on the less than 5 seconds
network
This is the time needed after alarm
indication on the monitor until the
alarm signal is available on the
network, to the IntelliVue
Information Center or for
transmission to other systems.
Pause duration 1,2,3 minutes or infinite, depending on configuration
Extended alarm pause 5 or 10 minutes
Sound pressure range min. 0 dB(A)
max. 45-85 dB(A)
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity 300 items
Real Time Clock Range from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy better than 4 seconds per day
Hold Time infinite if powered by AC; otherwise at least 24 hours
Buffered Memory Hold Time if powered by AC: infinite
without power: at least 8 hours
Contents Active settings, trends, patient data, realtime reports, events,
review alarms
466
40 Specifications
MX600/700/800
Trends Resolution 12, 16, 24 or 32 numerics @ 12 sec, 1 minute, 5 minute
resolution
Information Multiple choices of number of numerics, resolution and
duration depending on trend option and application area.
For example:
neonatal extended 12 numerics, 24 hours @ 12 secs or
32 numerics 32 hours @ 1 minute
intensive care extended: 16 numerics 120 hours @ 5 minutes
anesthesia extended 32 numerics 9 hours @ 12 seconds
High Res Trend Waves Measurements available HR, SpO2, Resp, tcpO2, Pulse, Perf, tcpCO2, CO2, ABP, PAP,
CVP, ICP, CPP, BIS, CCO, AWP, Anesthetic Agents, Delta
SpO2, inO2
Resolution Measurement samples are taken at a resolution of four samples
per second
Update speed waves are drawn at a speed of 3 cm/minute
Events Information trigger condition and time, event classification and associated
detailed view of episode data
Episode data configurable, either:
4 minutes of high resolution trend or
20 minutes of numerics trend @ 12 sec. resolution or
15 seconds of 4 waves @ 125 samples/sec. (Snapshot)
including all current numerics, alarms and inops
Capacity (max.) 25 or 50 events for either 8 or 24 hours
Alarm Signal System alarm delay less than 4 seconds
The system alarm delay is the
processing time the system needs
for any alarm to be indicated on
the monitor, after the
measurement has triggered the
alarm.
Delay for alarm availability on the less than 5 seconds
network
This is the time needed after alarm
indication on the monitor until the
alarm signal is available on the
network, to the IntelliVue
Information Center or for
transmission to other systems.
Pause duration 1,2,3 minutes or infinite, depending on configuration
Extended alarm pause 5 or 10 minutes
Sound pressure range min. 0 dB(A)
max. 45-85 dB(A)
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity 300 items
Real Time Clock Range from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy better than 4 seconds per day
Hold Time infinite if powered by AC; otherwise at least 24 hours
467
40 Specifications
MX600/700/800
Buffered Memory Hold Time if powered by AC: infinite
without power: at least 8 hours
Contents Active settings, trends, patient data, realtime reports, events,
review alarms
X2 Performance Specifications
X2 (M3002A) Performance Specifications
Power Specifications Power consumption <12 W average
<30 W while battery is loading
Operating Voltage 36 to 60 V DC floating
Battery Specifications Operating Time Basic monitoring configuration: 2.5 hours
(with new, fully charged battery at (Brightness set to optimum, ECG/Resp, SpO2 measurements
25°C) in use, NBP measurement every 15 minutes)
Charge Time When X2 is off: 2 hours
When X2 is in use, and connected to a monitor, without
extensions: 12 hours approx.
When X2 is in use, and connected to the external power supply
(M8023A), without extensions: 12 hours approx.
Indicators Alarms Off red or yellow LED with crossed-out alarm symbol
Alarms red/yellow/light blue (cyan) LED
On/Standby / Error green / red LED
AC Power green LED
Battery yellow (charging)/red blinking (empty) LED
External Power green LED
Sounds Audible feedback for user input
Prompt tone
QRS tone, or SpO2 modulation tone
4 different alarm sounds
Trends Resolution 12 or 16 numerics @ 12 sec, 1 minute, 5 minute resolution.
Information Multiple choices of number of numerics, resolution and
duration depending on trend option and application area.
For example:
For neonatal, you can choose between 12 and 16 numerics.
Alarm signal System alarm delay less than 4 seconds
Pause duration 1,2,3 minutes or infinite, depending on configuration
Extended alarm pause 5 or 10 minutes
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity 300 items
Real Time Clock Range from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy <4 seconds per day (typically)
Hold Time infinite if powered by host monitor or external power supply;
otherwise at least 48 hours
468
40 Specifications
Integrated PC (iPC)
Integrated PC (iPC) Performance Specifications
PC Components Processor Intel Core 2 Duo SP9300/SP9400
Solid-state drive 100 GB or above
RAM 4 GB
469
40 Specifications
Interface Specifications
MX400/450/500/550/600/700/800 Interface Specifications
MX400/450/500/550/600/700/800
Network Standard 100-Base-TX (IEEE 802.3 Clause 25)
Connector RJ45 (8 pin)
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
MIB/RS232 Standard IEEE 1073-3.2-2000
Connectors RJ45 (8 pin)
Mode Software-controllable
BCC (RxD/TxD cross over) or
DCC (RxD/TxD straight through) not available for the
advanced system interface MX400/450/500/550
Power 5 V ±5 %, 100 mA (max.)
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
USB Interface Standard USB 2.0 full-speed (embedded host)
Connector USB series "Standard A" receptacle
Power Low power port 4.4V min; max. load for all ports together
500 mA
Isolation none
RS232 (Standard) Connector RJ45 (8-pin)
Power none
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
RS232 (Independent display Connector RJ45 (8-pin)
interface option) Power none
Isolation none
Basic Nurse Call Relay Connector modular Jack 6P6C, active open and closed contact
Contact <=100 mA, <=24 V DC
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
Delay <[Configured Latency +0.5] sec
Flexible Nurse Call Relay Connector 20 pin MDR (Mini D-Ribbon), active open and closed contacts
Contact <=100 mA, <=24 V DC
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
Delay <[Configured Latency +0.5] sec
IntelliBridge EC10 Interface Connector RJ45 (8-pin)
board Speed 1200 Bd, 2400 Bd, 4800 Bd, 9600 Bd, 19200 Bd, 38400 Bd,
57600 Bd, or 115200 Bd signaling speed
Character bits 5-8
Stop bits 1 or 2
Parity Even or odd parity generation and checking
Power 5 V±5 % @ 0-100 mA
Isolation double insulation (reference voltage: 250 V; test voltage: 4000
V)
470
40 Specifications
MX400/450/500/550/600/700/800
IntelliVue Instrument Type Internal WMTS Adapter
Telemetry Wireless Network Technology compatible with Philips IntelliVue Telemetry System (ITS),
(USA only) cellular infrastructure
Frequency Band WMTS, 1395-1400 MHz and 1427-1432 MHz
IntelliVue Instrument Type Internal ISM Adapter
Telemetry Wireless Network Technology compatible with Philips IntelliVue Telemetry System (ITS),
(except USA) cellular infrastructure
Frequency Band 2.4 GHz ISM
IntelliVue 802.11 Bedside Type Internal Wireless Adapter
Adapter (Wireless Network Technology IEEE 802.11a/b/g
Adapter)
Frequency Band 2.4 GHz and 5 GHz Band
Modulation technique DSSS (CCK, DQPSK, DBPSK), OFDM (BPSK, QPSK, 16-
QAM, 64-QAM)
Effective radiated power (max.) MX400-MX550: 16 dBm (35 mW)
MX600-MX800: 17 dBm (50 mW)
Short Range Radio Interface Type Internal SRR interface
Technology IEEE 802.15.4
Frequency band 2.4 GHz ISM (2.400 - 2.483 GHz)
Modulation technique DSSS (O -QPSK)
Effective radiated power max. 0 dBm (1 mW)
Measurement Link (MSL) Connectors ODU out (Proprietary)
MX600/700/800 Voltage 56 V ±10 %
Power 45 W
Power Sync. 5 V CMOS Level; 78.125 kHz (typ.)
LAN signals IEEE 802.3 10-Base-T compliant
Serial signals RS-422 compliant
Measurement Link (MSL) Connectors ODU out (Proprietary)
MX400/450/500/550 Voltage 48 V ±10 %
Power 12 W
Power Sync. 5 V CMOS Level; 78.125 kHz (typ.)
LAN signals IEEE 802.3 10-Base-T compliant
Serial signals RS-422 compliant
Video Interface (standard) Connector DVI (digital and analog, single link)
Digital video signals single link TMDS
Analog video signals 0.7 Vpp@75 Ω
HSYNC/VSYNC signals TTL
DDC signals none (except for MX400/450/500/550)
DDC power 5 V +/-5% @0-55 mA
471
40 Specifications
MX400/450/500/550/600/700/800
Video Interface (Independent Connector DVI (digital and analog, single link)
display interface option) Pixel clock frequency 31.5 MHz - 119 MHz
Digital video signals single link TMDS
Analog video signals 0.7 Vpp@75 Ω
HSYNC/VSYNC signals TTL
DDC signals none
DDC power 5 V +/-5% @0-55 mA
ECG Sync Output/Analog ECG Output (1/4" stereo phone jack with tip, ring, sleeve)
General Connector 1/4" phone with tip, ring, sleeve
Isolation functional isolation
Analog ECG Output Full scale on display signal gain x measured ECG voltage
(ring, tip) Gain error <15 %
Baseline offset <100 mV
Bandwidth 1 to 100 Hz
Output voltage swing ±4 V (min)
Signal delay <22 ms
Signal delay with older versions of <30 ms
the M3001A MMS
[identifiable with the serial number
prefix DE227 or DE441 and
option string #A01]
Pacemaker Pulse filtered and included in ECG output signal
Digital Pulse Output (ring) Output low voltage level <0.4 V @ I=-1 mA
Output high voltage level >2.4 V @ I=1 mA
Pulse Width 100 ms±10 ms (active high)
Pulse Rise Time <1 ms (from 0.4 V to 2.4 V)
Signal delay <25ms per AAMI EC13
Signal delay with older versions of <35ms per AAMI EC13
the M3001A MMS
[identifiable with the serial number
prefix DE227 or DE441 and
option string #A01]
472
40 Specifications
X2 Interface Specifications
X2 (M3002A) Interface Specifications
Measurement Link (MSL) Connectors Female ODU (Proprietary)
Power 30 V to 60 V input
Power Sync. RS-422 compliant input 78.125 kHz (typical)
LAN signals IEEE 802.3 10-Base-T compliant
Serial signals RS-422 compliant
Local signals Provided for connecting MMS extensions
ECG Sync Pulse Output Cable detection Yes
(rectangular pulse) Marker In No
Wave Output No
Connector Binder Series 709/719
Output levels Output low <0.8 V @ I = -4 mA
Output high >2.4 V @ I = 4 mA
Isolation None
Pulse Width 100 +/- 10 ms (high)
Delay from R-wave peak to start of 20 ms maximum per AAMI EC13
pulse
Minimum required R-wave 0.5 V
amplitude
Wireless Network Device Signals RD+/-, TD+/-: IEEE 802.3 10Base-T, PWR, GND
Interface 12.5 V ±20%, 3.5 W continuous
Short Range Radio Interface1 Type Internal SRR interface
Technology IEEE 802.15.4
Frequency Band 2.4 GHz ISM (2.400 - 2.483 GHz)
Modulation Technique DSSS (O -QPSK)
Effective radiated power max. 0 dBm (1 mW)
1The short range radio interface is compatible with the following devices: TRx4841A/TRx4851A Intellivue Telemetry
System Transceiver and MX40 Wearable Patient Monitor
473
40 Specifications
Display Specifications
Display Specifications
19" WSXGA+ Type 482 mm active matrix color LCD (TFT)
MX600/700/800 Resolution 1680 x 1050 (WSXGA+)
Refresh frequency 58 Hz
Useful screen 409.5 mm x 255.9 mm
Pixel size 0.244 mm x 0.244 mm
15" WXGA Type 390 mm active matrix color LCD (TFT)
MX550/600/700/800 Resolution 1280 x 768 (WXGA)
Refresh frequency 59.9 Hz
Useful screen 334.1 mm x 200.5 mm
Pixel size 0.261 mm x 0.261 mm
12" WXGA (16:10) Type 310 mm active matrix color LCD (TFT)
MX450/500 Resolution 1280 x 800 (WXGA 16:10)
Refresh frequency 59.9 Hz
Useful screen 261.1 mm x 163.2 mm
Pixel size 0.204 mm x 0.204 mm
9" WVGA Type 230 mm active matrix color LCD (TFT)
MX400 Resolution 800 x 400 (WVGA)
Refresh frequency 60 Hz
Useful screen 196.8 mm x 118.1 mm
Pixel size 0.246 mm x 0.246 mm
474
40 Specifications
Compatible Devices
Compatible Devices
Displays (must be approved for medical use)
M8031B XGA color 15" LCD touchscreen
M8033C SXGA color 17" LCD touchscreen
865299 SXGA color 19" LCD touchscreen
Measurement Specifications
See the Appendix on Default Settings for a list of the settings the monitor is initially shipped with.
ECG/Arrhythmia/ST/QT
Complies with IEC 60601-2-25:1993 + A1:1999 /EN 60601-2-25:1995 + A1:1999,
IEC 60601-2-27:2005/EN 60601-2-27:2006, IEC 60601-2-51:2003 /EN 60601-2-51:2003 and AAMI
EC11/EC13:1991/2002.
475
40 Specifications
476
40 Specifications
ECG/Arrhythmia/ST/QT
Range Adjustment
Alarm Specifications
HR 15 to 300 bpm Adult:1 bpm steps (15 to 40 bpm)
maximum delay: 10 seconds 5 bpm steps (40 to 300 bpm)
according to AAMI EC 13-1992 Pedi/Neo:1 bpm steps (15 to 50 bpm)
standard 5 bpm steps (50 to 300 bpm)
477
40 Specifications
Respiration
Respiration Performance Specifications
Respiration Rate Range Adult/pedi: 0 to 120 rpm
Neo: 0 to 170 rpm
Accuracy at 0 to 120 rpm ±1 rpm
at 120 to 170 rpm ±2 rpm
Resolution 1 rpm
Bandwidth 0.3 to 2.5 Hz (-6 dB)
Noise Less than 25 mΩ (rms) referred to the input
Respiration Alarm
Range Adjustment Delay
Specifications
High Adult/pedi: 10 to 100 rpm under 20 rpm: 1 rpm steps max. 14 seconds
Neo: 30 to 150 rpm over 20 rpm: 5 rpm steps
Low Adult/pedi: 0 to 95 rpm under 20 rpm: 1 rpm steps for limits from 0 to 20 rpm:
Neo: 0 to 145 rpm over 20 rpm: 5 rpm steps max. 4 seconds
for limits above 20 rpm: max.
14 seconds
Apnea Alarm 10 to 40 seconds 5 second steps
478
40 Specifications
SpO2
Unless otherwise specified, this information is valid for SpO2 measured using the M3001A and
M3002A Multi-measurement modules and the M1020B measurement module. The SpO2 Performance
Specifications in this section apply to devices with Philips SpO2 technology. For SpO2 Performance
Specifications valid for other SpO2 technologies, refer to the instructions for use provided with these
devices.
Complies with ISO 9919:2005 / EN ISO 9919:2009 (except alarm system; alarm system complies with
IEC 60601-2-49:2001).
Measurement Validation: The SpO2 accuracy has been validated in human studies against arterial
blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically
distributed, only about two-thirds of the measurements can be expected to fall within the specified
accuracy compared to CO-oximeter measurements.
Display Update Period: Typical: 2 seconds, Maximum: 30 seconds. Maximum with NBP INOP
suppression on: 60 seconds.
479
40 Specifications
480
40 Specifications
1. Refer to “Monitor Performance Specifications” on page 464 for system alarm delay specification.
481
40 Specifications
NBP
Complies with IEC 60601-2-30:1999/EN 60601-2-30:2000.
482
40 Specifications
Measurement Validation: In adult and pediatric mode, the blood pressure measurements determined
with this device comply with the American National Standard for Electronic or Automated
Sphygmomanometers (ANSI/AAMI SP10 - 1992) in relation to mean error and standard deviation,
when compared to intra-arterial or auscultatory measurements (depending on the configuration) in a
representative patient population. For the auscultatory reference the 5th Korotkoff sound was used to
determine the diastolic pressure.
In neonatal mode, the blood pressure measurements determined with this device comply with the
American National Standard for Electronic or Automated Sphygmomanometers (ANSI/AAMI SP10 -
1992 and AAMI/ANSI SP10A -1996) in relation to mean error and standard deviation, when
compared to intra-arterial measurements in a representative patient population.
483
40 Specifications
484
40 Specifications
Temp
Complies with EN 12470-4:2000 and IEC 60601-2-49:2001. Specified without transducer.
CO2
The CO2 measurement in M3014A and M3015A/B complies with EN ISO 21647:2004 + Cor.1:2005
(except alarm system; alarm system complies with IEC 60601-2-49:2001).
485
40 Specifications
where:
PBTPS = partial pressure at body temperature and pressure, saturated
PATPD = partial pressure at ambient temperature and pressure, dry
486
40 Specifications
where:
PBTPS = partial pressure at body temperature and pressure,
saturated
PATPD = partial pressure at ambient temperature and pressure,
dry
Pabs = absolute pressure
PH2O = 42 mmHg @35°C and 100% RH
487
40 Specifications
where:
PBTPS = partial pressure at body temperature and pressure,
saturated
PATPD = partial pressure at ambient temperature and pressure,
dry
Pabs = absolute pressure
PH2O = 42 mmHg @35°C and 100% RH
CO2 Alarm
Range Adjustment Delay
Specifications
etCO2 High 20 to 95 mmHg (2 to 13 kPa) 1 mmHg (0.1 kPa) M3002A/M3014A/M3016A: less
etCO2 Low 10 to 90 mmHg (1 to 12 kPa) than 14 seconds
M3015A: less than 21 seconds.
imCO2 High 2 to 20 mmHg steps of 1 mmHg (0.1 kPa) M3002A/M3014A/M3016A: less
(0.3 to 3.0 kPa) than 14 seconds
M3015A: less than 21 seconds.
awRR High Adult/pedi: 10 to 100 rpm under 20 rpm: 1 rpm steps M3002A/M3014A/M3016A: less
Neo: 30 to 150 rpm over 20 rpm:5 rpm steps than 14 seconds
M3015A: less than 21 seconds.
awRR Low Adult/pedi: 0 to 95 rpm M3015A:
Neo: 0 to 145 rpm settings <20 rpm: less than
8 seconds
>20 rpm: less than 21 seconds
M3002A/M3014A/M3016A
settings <20 rpm: less than
4 seconds
>20 rpm: less than 14 seconds
IPI Low (M3015A/B Adult/pedi: 2 to 9 steps of 1 Maximum 14 seconds after
only) displayed value goes below the low
alarm limit setting.
Apnea delay 10 to 40 seconds 5 second steps set apnea delay time + 4 seconds
(M3002A/M3014A/M3016A) or
8 seconds (M3015A with 2 m
sample lines) or 11 seconds
(M3015A with 4 m sample lines).
488
40 Specifications
Gas Level
Gas or Vapor M3002A/M3014A M3015A
(% volume fraction)
Spirometry
Complies with IEC 60601-2-49:2001. The following specifications apply for 760 mmHg ambient
pressure and patient gas: room air at 35°C, unless otherwise noted.
489
40 Specifications
490
40 Specifications
491
40 Specifications
C.O./CCO Alarm
Range Adjustment Delay
Specifications
TBlood 17 to 43°C Steps of 0.5°C (17 to 35°C) 10 seconds after the value exceeds
Steps of 0.1°C (35 to 43°C) the set limit range
Steps of 1°F (63 to 95°F)
Steps of 0.2°F (95 to 109°F)
CCO 0.1 to 25.0 l/min 0.1 l/min (0.1 to 10.0 l/min) 10 seconds after the value exceeds
0.5 l/min (10.0 to 25.0 l/min) the set limit range
tcGas
Complies with IEC 60601-2-23:1999/EN 60601-2-23:2000.
492
40 Specifications
tcGas Alarm
Range Adjustment Alarm Delay
Specifications
tcpO2 10 to 745 mmHg 10 to 30 mmHg: 1 mmHg 10 seconds after the value exceeds
1.0 to 99.5 kPa 1.0 to 4.0 kPa: 0.1 kPa the set limit range.
32 to 100 mmHg: 2 mmHg
4.2 to 13 kPa: 0.2 kPa
105 to 745 mmHg: 5 mmHg
13.5 to 99.5 kPa: 0.5 kPa
tcpCO2 10 to 195 mmHg 10 to 30 mmHg: 1 mmHg
1.0 to 26 kPa 1.0 to 4.0 kPa: 0.1 kPa
32 to 100 mmHg: 2 mmHg
4.2 to 13 kPa: 0.2 kPa
105 to 195 mmHg: 5 mmHg
13.5 to 26 kPa: 0.5 kPa
Performance Specifications
SO2, SvO2, ScvO2 Range 10% to 100%
Accuracy ±2% (i.e. ±2 units), 1 standard deviation over 40% to 100%1 range.
Resolution 1%
Stability (system) Drift <2% over 24 hours
Response Time (10 % to 90 %) 5 seconds
1 The accuracy specification for Edwards Lifesciences’ Philips-compatible catheters has been verified
in the saturation range 40%-95%.
EEG
Complies with IEC 60601-2-26:1994/EN 60601-2-26:1994.
493
40 Specifications
BIS
Complies with IEC 60601-2-26:1994.
NMT
Complies with IEC 60601-2-49:2001
494
40 Specifications
** TOFcnt High 0, 1, 2, 3 TOF Counts Max. 4 sec system delay after value goes
above the high alarm limit setting
WARNING
Using accessories other than those specified may result in increased electromagnetic emission or
decreased electromagnetic immunity of the monitoring equipment.
Electromagnetic Emissions
The monitor is suitable for use in the electromagnetic environment specified in the table below. You
must ensure that it is used in such an environment.
495
40 Specifications
WARNING
The monitor should not be used next to or stacked with other equipment. If you must stack the
monitor, you must check that normal operation is possible in the necessary configuration before you
start monitoring patients.
Electromagnetic Immunity
The monitor is suitable for use in the specified electromagnetic environment. The user must ensure
that it is used in the appropriate environment as described below.
IEC 60601-1-2
Immunity test Compliance level Electromagnetic environment guidance
test level
Electrostatic discharge ±6 kV contact ±6 kV contact Floors should be wood, concrete, or
(ESD) ±8 kV air ±8 kV air ceramic tile. If floors are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/ ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality should be that of a
burst ±1 kV for input/output lines ±1 kV for input/output lines typical commercial and/or hospital
IEC 61000-4-4 environment
Surge ±1 kV differential mode ±1 kV differential mode Mains power quality should be that of a
IEC 61000-4-5 ±2 kV common mode ±2 kV common mode typical commercial and/or hospital
environment
Voltage dips, short <5% UT (>95% dip in UT) for <5% UT (>95% dip in UT) for Mains power quality should be that of a
interruptions and voltage 0.5 cycles 0.5 cycles typical commercial and/or hospital
variations on power supply 40% UT (60% dip in UT) for 40% UT (60% dip in UT) for environment. If the user of the monitor
input lines 5 cycles 5 cycles requires continued operation during power
IEC 61000-4-11 mains interruptions, it is recommended
70% UT (30% dip in UT) for 70% UT (30% dip in UT) for that the monitor is equipped with an
25 cycles 25 cycles internal battery or is powered from an
<5% UT (>95% dip in UT) for <5% UT (>95% dip in UT) for uninterruptible power supply.
5 sec 5 sec
496
40 Specifications
IEC 60601-1-2
Immunity test Compliance level Electromagnetic environment guidance
test level
Power frequency (50/ 3 A/m 3 A/m Power frequency magnetic fields should be
60 Hz) magnetic field a t levels characteristic of a typical location
IEC 61000-4-8 in a typical commercial and/or hospital
environment
In this table, UT is the a.c. mains voltage prior to application of the test level.
In the following table, P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer and d is the recommended separation distance in meters (m). The
values given in brackets are for respiration and BIS.
Portable and mobile RF communications equipment should be used no closer to any part of the
monitor, including cables, than the recommended separation distance calculated from the equation
appropriate for the frequency of the transmitter.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with this symbol:
IEC 60601-1-2
Immunity test Compliance level Electromagnetic environment guidance
test level
Conducted RF 3 Vrms 3 Vrms Recommended separation distance:
IEC 61000-4-6 150 kHz to 80 MHz (1 Vrms for respiration d = 1.2√P
and BIS) for respiration and BIS:
d = 3.5√P
497
40 Specifications
IEC 60601-1-2
Immunity test Compliance level Electromagnetic environment guidance
test level
Radiated RF 3 V/m 3 V/m Recommended separation distance:
IEC 61000-4-3 80 MHz to 2.5 GHz (1 V/m for respiration 80 MHz to 800 MHz
and BIS) d = 1.2√P
80 MHz to 800 MHz for respiration and BIS
d = 3.5√P
800 MHz to 2,5 GHz
d = 2.3√P
800 MHz to 2,5 GHz for respiration and BIS
d = 7.0√P
2.0 to 2,3 GHz for short range radio
d = 7.0√P
Field strengths from fixed transmitters, such as base stations for radio (cellular, cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the monitor is used exceeds the applicable RF compliance level above, the monitor
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the monitor.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
Frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
d = 1.2√P d = 1.2√P d = 2.3√P
Equation for respiration and BIS: for respiration and BIS: for respiration and BIS:
d = 3.5√P d = 3.5√P d = 7.0√P
Rated max. output
Separation distance Separation distance Separation distance
power of transmitter
0.01 W 0.1 (0.4) m 0.1 (0.4) m 0.2 (0.7) m
0.1 W 0.4 (1.1) m 0.4 (1.1) m 0.7 (2.2) m
1W 1.3 (3.5) m 1.3 (3.5) m 2.3 (7.0) m
10 W 3.8 (11.1) m 3.8 (11.1) m 7.3 (22.1) m
100 W 12.0 (35.0) m 12.0 (35.0) m 23.0 (70.0) m
498
40 Specifications
Fast Transients/Bursts
The equipment will return to the previous operating mode within 10 seconds without loss of any
stored data (M1032A within 30 seconds). If any user interaction is required, the monitor indicates this
with a technical alarm (INOP).
Restart time
After a power interruption, an ECG wave will be shown on the display after 30 seconds maximum. If
an X2 is connected to a host monitor, it can take up to a maximum of 45 seconds until the ECG wave
is shown.
499
40 Specifications
500
41
501
41 Default Settings Appendix
502
41 Default Settings Appendix
503
41 Default Settings Appendix
504
41 Default Settings Appendix
505
41 Default Settings Appendix
506
41 Default Settings Appendix
507
41 Default Settings Appendix
508
41 Default Settings Appendix
509
41 Default Settings Appendix
510
41 Default Settings Appendix
511
41 Default Settings Appendix
Factory Defaults
QT Settings
Factory Adult Factory Pedi Factory Neo
QT Lead All All All
QTc High Limit 500 ms 480 ms 460 ms
ΔQTc High Limit 60 ms 60 ms 60 ms
QTc High Alarm On On On
ΔQTc High Alarm On On On
QT Analysis Off Off Off
QTc Formula Bazett Bazett Bazett
512
41 Default Settings Appendix
513
41 Default Settings Appendix
514
41 Default Settings Appendix
CVP, RAP, LAP, UVP Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Settings Adult Pedi Neo Adult Pedi Neo
Alarms from Mean Mean Mean
High Limit 14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High 5 5 5
Δ Extreme Low 5 5 5
Sys. High Clamp/ Dia. 20/10 (15) 15/5 (10) 15/5 (10)
High Clamp (Mean
High Clamp)
Sys. Low Clamp/Dia. 0/-5 (-5) 0/-5 (-5) 0/-5 (-5)
Low Clamp (Mean Low
Clamp)
Scale 30 30 30
515
41 Default Settings Appendix
CVP, RAP, LAP, UVP Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Settings Adult Pedi Neo Adult Pedi Neo
Mean only Yes Yes Yes
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr. 60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Cyan (light Cyan (light Cyan (light Blue Blue Blue
blue) blue) blue)
Factory Defaults
PAP Settings
Adult Pedi Neo
Alarms from Dia. Dia. Dia.
High Limit 34/16 (20) 60/4 (26) 60/4 (26)
Low Limit 10/0 (0) 24/-4 (12) 24/-4 (12)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High 5 5 5
Δ Extreme Low 5 5 5
Sys. High Clamp/Dia. High Clamp (Mean 45/20 (25) 65/5 (35) 65/5 (35)
High Clamp)
Sys. Low Clamp/Dia. Low Clamp (Mean 5/-5 (-5) 15/-5 (5) 15/-5 (5)
Low Clamp)
Scale 30 30 30
Mean only No No No
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr. 60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Yellow Yellow Yellow
Factory Defaults
ICP, IC1, IC2 Settings
Adult Pedi Neo
Alarms from Mean Mean Mean
High Limit 14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High 10 10 10
Δ Extreme Low 10 10 10
Sys. High Clamp/Dia. High Clamp (Mean 20/10 (15) 15/5 (10) 15/5 (10)
High Clamp)
516
41 Default Settings Appendix
Factory Defaults
ICP, IC1, IC2 Settings
Adult Pedi Neo
Sys. Low Clamp/Dia. Low Clamp (Mean 0/-5 (-5) 0/-5 (-5) 0/-5 (-5)
Low Clamp)
Scale 30 30 30
Mean only Yes Yes Yes
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr. 60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Magenta Magenta Magenta
517
41 Default Settings Appendix
518
41 Default Settings Appendix
519
41 Default Settings Appendix
520
41 Default Settings Appendix
521
41 Default Settings Appendix
522
1 Index
# addressograph (printer configuration physiological 75
setting) 376 pleth as source 200
10-lead placement (ECG) 142 adjusting ST measurement points 168 recordings 73
12-lead placement (ECG) 142 adjusting wave scale (pressure) 217 red 59
3-lead placement (ECG) 140 restarting 66
adjusting wave size (CO2) 250, 260
4-slot flexible module rack 19 reviewing 71
admitting a patient reviewing messages 71
5-lead placement (ECG) 141 editing information 109 reviewing window 71
8-slot flexible module rack 19 quick admit 108 selftest 73
advanced event surveillance 333 silencing 63
A
airway adapter 249 SpO2 high and low limits 199
AAMI ECG lead labels 139 CO2, microstream accessory 249 SpO2 specific 194
abdominal breathing 186 alarm latching 72, 161 ST 171
and Resp electrode placement 186 alarm limits 67, 68, 70, 377 suspended symbol 66
changing 37 SvO2, behavior during
aberrantly conducted beats 156 measurement 272
accessories 242, 256 checking 67
switching on and off 41
10-electrode cable sets 421 manually adjusting 68
temperature 211
3-electrode cable sets 420, 422 narrow 70
testing 73
5-electrode cable sets 420, 422 report 312
tone configuration 62
6-electrode cable sets 420 setting 67
traditional 62
C.O. 228, 230, 233, 247, 436 ST 171
volume, changing 62
CO2 243, 247, 248, 249 switching auto limits on/off 70
VueLink and external devices 389
CO2 (mainstream) 437, 438 using automatic limits 70
yellow 59
CO2 (microstream) 438 wide 70
window 68 alphabetical listing of alarms 75
ECG 133
alarm recording 360 analog output 447
Hospira Inc. 440
ECG 133
NBP 423 alarm source selection, disabled 183
pressure 213
adult cuffs 425 alarms 59
comfort cuffs 423 annotating events 342
acknowledging 63
disposable cuffs 425 active 59 annotation 342
multi-patient comfort cuff kits 423 active SpO2 source 201 apnea alarm delay 251, 260
neonatal/infant cuffs (single alphabetical listing 75 apnea alarm delay time (RESP) 188
patient) 426 apnea delay 251, 260 apnea alarms 188
reusable cuffs 423 apnea delay time (Resp) 188 and Resp detection modes 187
single patient, adult/pediatric soft arrhythmia 154
cuffs 425 arrhythmia 154
audible indicators 61
Nellcor 429 aberrantly conducted beats 156
awrr limits 252
pressure 426 analysis, how it works 154
chaining 162
Pulsion 426 atrial fibrillation and flutter 155
CO2 specific 251
resp 419 beat labels 157
CO2, apnea delay 251, 260
set combiners and organizers 421, 423 initiating learning 159, 160
CO2, awRR 252
SpO2 191, 192, 429 intermittent bundle branch block 156
desat, SpO2 200
Nellcor adhesive sensors learning during ventricular rhythm 160
effect on pressure alarms during
(disposable) 429 levels of analysis 149, 154
zero 215
Philips sensors (disposable) 429 monitoring non-paced patients 155
extending pause time 66
Philips sensors (reusable) 429 monitoring paced patients 155
high priority 59
SvO2 273, 274, 441 options 154
INOP 59
temperature 435 relearning 159
ISO/IEC standard 62
trunk cables 419, 422 relearning and lead fallback 160
limit, SpO2 479
active alarms 59 status messages 157
NBP source 209
switching on/off 70
address, Philips 444 patient messages 75
understanding the display 156
pausing 64
523
arrhythmia alarms 59, 160, 161, 162, 164, 475 impedance indicators 289 calibration 210, 219, 273, 274, 276
adjusting alarm limits 161 monitoring 36 NBP 210
all yellow on/off 161 monitoring setup 286 pressure 219
chaining 162 numerics 206 pressure transducer 219
latching 72 on/off 290 SvO2 light intensity 274
multiple 162 safety information 152, 188, 239 SvO2, pre-insertion 273
PVC-related alarms 164 smoothing rate 290 calibration status indicators (C.O) 232
sinus and SV rhythm ranges 478, 479 changing 37 capnography 25
switching on/off 70 window 68 mainstream 25
timeout periods 161 blood pressure. See also NBP (non- Microstream 25
arrhythmia monitoring 152 invasive) or PRESS (invasive) 203
capturing loops 400
and defibrillation 152 BSA formula (trends) 328
carbon dioxide, see CO2 241, 255
arrhythmia options 149 buffer (EEG) on reports 284
cardiac output 225, 378
arrhythmia relearning 145 buffer (EEG) on screen 282 how the measurement works 203
with EASI INOP 145 report 312
arterial catheter constant (PiCCO) 230 C
cardiac overlay 185, 187
arterial pressure source 221 C.I. 226 and Resp detection modes 187
arterial pulsation 191 C.O. 225 when measuring Resp 185
arterial values 263 accessories 242, 256 cardiotach alarms 149
artifact suppression (pressure) 218 calibrating measurements (PiCCO) 232 care groups 131
Aspect Medical Systems Inc 445 curve alert messages 236 alarm notification 131
documenting measurements 235 pop-up window 131
atrial fibrillation and flutter 155
editing measurements (PiCCO) 232 catheter 274
audible latching (arrhythmia alarms) 161 editing measurements (RH method) 235 SvO2, insertion 274
auto alarm limits 70 flow-through method 233
switching on/off 70 catheter constant, setting (PiCCO) 230
hemocalc window 228
auto detection mode (Resp) 186 hemodynamic parameters available 226 catheter preparation 273
injectate guidelines 235 SvO2 271
auto ECG wave gain (recordings) 362
PiCCO method 228 CCO 225, 228, 230, 232
auto window (care groups) 131
Procedure window 226 calibration status indicators 232
autofilter 137 prompt messages 238 choosing the correct pressure
automatic arrhythmia relearn 159 results table 226 source 230
automatic default setting 41 right heart thermodilution method 233 how the measurement works 203
automatic NBP 208 saving measurements (PiCCO) 232 PiCCO method 228
repeat time 206 setup for PiCCO method 230 cerebral perfusion 221
autosize 137 setup for RH method 233 CFI 226
ECG wave 136 temperature unit 226 chaining 162
warning messages 238
average trend events 334 change screen menu 37
C.O./CCO safety information 239
awRR alarms 252 changing 37
limits 252 Calc Type 326
changing ECG lead sets 139
calculating 381
B changing EEG wave scale 283
drug infusions 381
changing EEG wave speed 283
calculating cerebral perfusion 221
balloon inflation, wedge measurement 222 changing Resp detection mode 186
calculating oxygen extraction 277
baseline 168, 174 changing Resp wave size 187
ST map, updating 174 calculating pulse pressure variation 221, 229
changing Resp wave speed 188
ST, updating 168 calculating temperature difference 212
changing screen content 37
basic arrhythmia option 149, 154 calculations 325, 326
pop-up keys 33, 335 changing screens 37
basic event surveillance 333, 338
setup 230 calculations reports 329 changing wave scale 283
beat labels 157 calculations review 327 checking battery charge 407
arrhythmia 154 calculator 35 checking paced status 134
BIS 285 calibrating 210 cleaning 411
Continuous Impedance Check 288 CO2 transducer 247 infection control 411
Cyclic Impedance Check 288 tcGas transducer 265 method 412
filters 290 monitoring accessories 413
calibrating C.O. measurements 232
Ground Check 288 recommended substances 412
524
recorder printhead 413 Cyclic Impedance Check (BIS) 288 ECG 133
clock 397 accessories 242, 256
D alarms off (Config Mode) 152
displaying on main screen 397
CO2 241 changing lead sets 139
database 339
airway adapter 249 choosing electrode sites 138
events 334
alarms, apnea delay 251, 260 conventional 12-lead 143
trends 319
alarms, awRR 252 external pacing electrodes 152
date, setting 47 filter settings 137
alarms, specific 251
default profile 40 fusion beat pacemakers 152
awRR alarm limits 252
checking transducer accuracy 247 default settings 501 intrinsic rhythm 152
correction, humidity 250 defibrillation 152, 284 modified 12-lead 143
correction, N2O 250 and arrhythmia monitoring 152 New Lead Setup 139
corrections 250, 260 and ECG monitoring 152 pacemaker failure 152
FilterLine 249 during EEG monitoring 284 rate adaptive pacemakers 152
mainstream accessories 242, 256 delayed recording 360 unfiltered 137
measuring mainstream 243, 247 wave size 136
deleting events 339
measuring microstream 249 ECG analog output 447
demonstration mode 36
method, mainstream 241 ECG cable 152
desat alarm, SpO2 200
method, microstream 241 for operating room 152
method, sidestream 241 detection modes (Resp) 186
ECG cables, connecting 133
microstream accessories 249 device driver 390, 393
ECG connector 133
microstream extension 248 language conflict with monitor 390, 393
ECG electrode colors 139
removing exhaust gases 246, 250 diagnostic (ECG filter setting) 137
transducer, calibrating 247 ECG electrode placement 152
disabling touch operation 30
transducer, using 247 during electrosurgery 152
disconnect INOPs 64
troubleshooting 417 ECG gain 362, 371
silencing 63
wave scale, adjusting 250, 260 in reports 332
disinfecting 411
CO2 (mainstream) 437, 438 ECG lead labels 139
infection control 411
accessories 242, 256 ECG lead placement 138
recommended substances 412
CO2 (microstream) 438 choosing EASI/Standard 138
display 156, 166, 186, 206
accessories 242, 256 ECG leads monitored 139
ECG 134
combi-events 340 NBP 206 ECG report 371, 379
computation constant (RH) 234 Resp 186 lead layout 371
configuration 36 ST 166 ECG safety information 152
drug calculator 381 display settings 39 ECG source tracking 124
trends database 319 disposal 417 ECG wave 136
configuration mode 36 gas cylinder 417 autosize 137
conflict 44 parts and accessories 417 calibration bar 137
label 44 documenting events 342 ectopic status messages (arrhythmia
connecting BIS accessories 286 dosemeter (drug calculator) 384 monitoring) 158
connecting temperature probe 211 dPmax 226 editing C.O. measurements (PiCCO
method) 232
connection direction symbol 445 drip table (drug calculator) 385
editing C.O. measurements (RH
connectors 447 drug calculator 381 methods) 235
continuous cardiac output 225 dual SpO2 201 EEG
Continuous Impedance Check (BIS) 288 dual Temp measurement 212 accessories 242, 256
conventional 12-lead ECG 143 dyshemoglobins 192 monitor configuration and upgrade 284
correcting the NBP measurement 205 intravascular (SpO2) 192 EEG monitoring
counting events 340 changing filter frequencies 283
E
CPAP (RESP) 186 changing wave scale 283
early systolic blood pressure, NBP 206 changing wave speed 283
cuff 205, 206
electrode-to-skin impedance 281
pressure, NBP 206 EASI 138, 145
gridlines 283
selection, NBP 205 activating 138
impedance quality indicators 281
current view 172 ECG monitoring 145
safety information 152, 188, 239
ST map 172 lead placement 138, 185
setting report buffer time 284
curve alert messages (C.O.) 236 EASI ECG lead labels 139 setting screen buffer time 282
525
setup 230 exhaust gases, removing 246, 250 extension 27
skin preparation 133 extension cable for SpO2 192 hemodynamic parameters 226
switching numerics on and off 283 external devices 387, 388, 389 high filter (EEG) 283
EEG wave 290 alarms and INOPS 389 high pass filter (BIS) 290
changing scale 290 connecting to VueLink 387 high resolution recording 360
electrical interference connecting via VueLink 388
high-res trend event episodes 334
during EEG monitoring 284 external pacing electrodes 152
high-res trend waves 331
electrode placement (ECG) 133, 143 and ECG monitoring 152
about 41
conventional 12-lead 143 extreme bradycardia alarm 151, 183 OxyCRG 331
modified 12-lead 143 extreme pressure alarms 218 high-res waves 332
electrode placement (Resp) 185, 186 extreme rate alarms 151, 183 in reports 332
with abdominal breathing 186
extreme tachycardia alarm 151, 183 recordings 73
with lateral chest expansion 186
HiResTrnd 334
electrode-to-skin impedance (BIS) 289 F
see high-res trend 334
electrode-to-skin impedance (EEG) 281
fallback (ECG) 139 horizon trend 320
electrosurgery 152 trend time 320
and ECG 152 FAST 191
Fourier artefact suppression HR = RR (Resp) 186
EMC interference 188 technology 191 HR alarms 152
Resp 186
filter (ECG filter setting) 137 when arrhythmia off 152
EMI filter for ECG 137
filter (ECG) 137 HR alarms off (Config Mode) 152
end case reports 372
filter frequencies (EEG) 283 HR and pulse alarm source selection 183
setup 230
FilterLine 249 HR from (heart rate source) 182
enhanced arrhythmia option 149, 154
CO2, microstream accessory 249 humidity correction 250
entering values 328
filters 290 hypotension evaluation 352
calculations 325, 326
BIS 290
event database 339 I
flexible module rack 19
event episode reports 372
flushing invasive pressure accessories 213 IEC ECG lead labels 139
event groups 334
FMS-4 19 impedance indicators (BIS) 289
event report 346
FMS-8 19 impedance quality indicators 281
event review reports 372
functional arterial oxygen saturation 191 IMV (Resp) 186
event snapshots 334
fusion beat pacemakers 152 infection control 411
event surveillance 333 and ECG monitoring 152
options 154 cleaning 411
disinfecting 411
events 333 G
sterilizing 411
annotation 342
combi-events 340 gas cylinder 417 Information Center 113, 114
counting oxyCRG/NER events 340 empty, disposing of 418 transferring patients 113, 114
event counter 340 GEDV/GEDVI 226 transferring patients using IIT 114
event episode 334 getting started 48 injectate guidelines for C.O. 235
event episode recording 344 global trend time 320 injectate volume setting (C.O.) 230
event episode types 334 graphic trend 312 INOPs 59, 64, 389
event episode window 342 report 312 from external devices 389
event post-time 334 indicators 59
graphic trends 313
event pre-time 334 silencing 63
event retriggering 337 graphic trends report 319
installation 447
event review recording 343 gridlines (EEG) 283, 290
connectors 447
event review window 341 Ground Check (BIS) 288
integrated PC 50
event summary view 340
event time 334 H intermittent bundle branch block 156
event triggers 336 intermittent mandatory ventilation
hardkeys 33 (Resp) 186
event values 342
manual event triggers 339 help 71 intravascular dyshemoglobins (SpO2) 192
setting up NER 338 INOPS 71
intrinsic rhythm 152
events pop-up keys 335 hemodynamic calculations 325
invasive pressure 213
EVLW/EVLWI 226 hemodynamic measurement module
in-vivo calibration, SvO2 274, 276
526
iPC 50 capnography 25 monitor settings 39
desktop 51 mainstream CO2 242, 247, 256 changing 37
input devices 51 accessories 242, 256 monitoring 36
shutting down 52 measuring 243, 247 monitoring mode 36
ISO point (ST) 169 maintenance 415 mouse 30
ITBV/ITBVI 226 schedule 415 using 30
visual inspection 415 multi-measurement module 133
J
management bundle recommendations 354 ECG connector 133
J point (ST) 169 manual detection mode (Resp) 187, 188
and apnea alarms 188 N
K manually triggering events 339 N2O correction 250
keys 30, 33 manufacturer's information 444 narrow alarm limits 70
hardkeys 33 map 172 navigating 30
permanent keys 30 ST 172 mouse 30
pop-up 33 Mason-Likar lead system 143 permanent keys 30
max hold setting (CO2) 241 trackball 30
L
mean pressure calculation (trends) 328 NBP 203
label conflict resolution 44 measurement 41, 42 adult cuffs 425
labels 44 setting up 41 alarm source 209
resolving conflict 44 switching on and off 41 ANSI/AAMI SP10-1992 203
lactate measurement 352 wave speed, changing 42 automatic mode, enabling 208
calibrating 210
language conflicts with device driver 390 measurement module extension 25, 27
comfort cuff kits 423
latching 72 M3012A 27
comfort cuffs 423
alarms, behavior 72 M3015A 25
cuff pressure 206
M3016A 25
latching arrhythmia alarms 161 cuff, applying 205
measurement modules 48 cuff, selecting 205
lateral chest expansion (neonates) 186
monitoring Resp 186 measurement points, ST 168 cuff, tightness 205
LCW/LCWI 226 measurement selection window 44 disposable cuffs 425
measurement settings 39 how the measurement works 203
lead fallback 160
merging patient data 118 measurement correction 205
and arrhythmia relearning 160
measurement limitations 204
lead fallback (ECG) 139 methemoglobin (SpO2) 192
measurement methods, auto 204
lead labels (ECG) 139 Microstream capnography 25 measurement methods, manual 204
lead placement 138, 185 microstream CO2 248, 249 measurement methods, sequence 204
activating EASI/Standard 138 accessories 242, 256 measurement methods, stat 204
for Resp measurement 185 measuring 243, 247 measurement, starting 207
leads monitored (ECG) 139 mismatch 117 measurement, stopping 207
Leads Off INOP (ECG) 139 patient data, resolving 117 neonatal cuffs (disposable) 426
MMS 115 numerics 206
levels of arrhythmia analysis 154
using for patient transfer 115 oscillometric method 203
levels of event surveillance 333 pediatric cuffs 425
modified 12-lead ECG 143
line frequency interference (BIS) 290 preparing to measure 205
modified screen history 37 repeat time 206
loops 399, 400, 401
capturing 400 modifying 37 repeat time for automatic 208
loop size in window 400 screens 37, 39 repetition time, setting 208
loops report 401 module 263, 387, 391 reusable cuffs 423
source device 401 tcGas 263 single patient, adult/pediatric soft
volume-flow 399 VueLink 387 cuffs 425
low filter (EEG) 283 VueLink Type A and Type B 387 site inspection 192
monitor 49 time of last measurement 206
low pass filter (BIS) 290
starting monitoring 49 units 206
LVCI 226 venous puncture 210
monitor (ECG filter setting) 137
M neonatal event counting 340
monitor defaults 501
neonates 186
M8048A 19 monitor revision 48
Resp electrode placement 186
how to find 48
mainstream 25 NER setup 338
527
non-invasive blood pressure. See NBP 203 patient mismatch 117 selecting 134
non-paced patients 155 patient reports 376 print job 374
arrhythmia monitoring 152 contents 376 suspended 374
notch filter (BIS) 290 patient trends 311 printer 373, 374, 375
numerics 206 viewing 311 disabling 30
explanation of NBP display 206 paused alarms 64, 66 settings 41, 125, 501
nurse call 66 extending time 66 status messages 157
restarting 66 unavailable 374
O PAWP 222 printing 369
performance specifications 475 C.O. measurements 235
on screen calculator 35 calculations reports 329
operating 30 performance test 495
event reports 342
mouse 30 performing calculations 327 ST map reports 175
permanent keys 30 perfusion indicator 191, 192, 200 status log 417
trackball 30 pf loops 399 trends reports 319
operating modes 36 Philips contact information 444 wedge 222
configuration 36 priority list for trends 316
physiological alarms 59
demonstration 36 probes 211
monitoring 36 PiCCO method 228
C.O. 228 disposable temperature 211
passcode protection 36
CCO 228 profiles 39, 40
service 36
setup 230 default profile 40
operating room ECG cable 152 patient category 39
pleth alarm source 200
orange ECG cable 152 swapping 40
pleth wave 200
organizers 421, 423 swapping setting block 40
pleth waveform 191
Oridion Systems Ltd 445 protocol log 355
pop-up keys 33, 335
Original Calc pop-up key 327 ProtocolWatch 349
events 334
oscillometric NBP measurement SSC sepsis 349
power 49
method 203 pulse 181, 182
disconnecting from 49
overlapping screen trends 321 system pulse source 181
mains power 49
oxyCRG 331, 340 disconnecting from 49 pulse numerics for SpO2 192
event counting 340 PPV 221, 229 pulse pressure variation 221, 229
OxyCRG event episodes 334 preparing skin 133 Pulsion Medical Systems AG 445
oxygen extraction 277 for ECG 133 pv loops 399
oxygenation calculations 325 pressure 213 PVC-related alarms 164
alarms during zero 215 PVR/PVRI 226
P
arterial source 221
calibration pressure 219 Q
pace pulse rejection (ECG) 134, 136
about 41 cerebral perfusion, calculating 221
QRS tone pitch, SpO2 201
switching on/off 70 performance specifications 475
wave scale 217 QRS volume, changing 137
paced patients 136, 152, 155
wave size 136 QT alarms 178
arrhythmia monitoring 152
wedge 222 QT baseline 177
repolarization tails 136
wedge, editing 223 QT measurement algorithm 175
safety information 152, 188, 239
zeroing the transducer 215
setting status 152 QT monitoring 176
pressure accessories 426 limitations 176
paced status 134
checking 67 pressure analog output 224 QT/QTc monitoring 175
pacemaker failure 152 pressure artifact suppression 218 quick admit 108
paper size 372 pressure of NBP cuff 206
pressure transducer 215 R
for reports 372
parameter scales 317 calibration 219 radiated field immunity 188
trends 317 zeroing 215 Resp 186
passcode protection 36 pressure-flow loops 399 rate adaptive pacemakers 152
patient 203 pressure-volume loops 399 and ECG monitoring 152
category, NBP 203 previous screen 37 ratemeter (drug calculator) 384
patient alarm messages 75 primary lead (ECG) 134
528
RCW/RCWI 226 scheduled 372 safety test 495
realtime recording 360 setting up 41 same patient data merge 118
realtime report 378 ST map 175 saving C.O. measurements (RH
stopping printouts 371 methods) 235
realtime reports 372
titration table 386
content 372 scale 136, 187
trends 319
rear connectors 448 ECG wave 136
re-routing reports 374 Resp wave 187, 188
Recorder 357
resampling vitals 326 scales 317
reloading paper 364
resolving patient mismatch 117 for trends waveforms 317
recorder 413
cleaning the print head 413 resp accessories 419 scheduled reports 372
recorder status messages 365 Resp alarms 188 screen 30
apnea alarm delay time 188 disabling touch operation 30
recording 357
alarm 62 Resp detection level 188 screen trend 320
annotation 342 and apnea detection 188 trend time 320
beat-to-beat 360 Resp detection modes 186, 187 screen trends 320
C.O. measurements 235 and cardiac overlay 185 screening criteria 351
context 360 changing 37
screens 37, 39
delayed 360 Resp display 186 changing 37
drug calculations 386 Resp monitoring 185 changing content 37
ECG gain 362, 371 and cardiac overlay 185 understanding 37
extending 359 Resp safety information 188 visitor screen 39
high resolution 360
Resp wave 187, 188 secondary lead (ECG) 134
preventing fading ink 363
changing size 187 selecting 134
procedure 360
changing speed 42 SEF numeric (BIS) 290
realtime 360
recording strip 362 respiratory loops 401 on/off 290
recording strip code 363 source device 401 selecting the primary lead (ECG) 134
ST segments 168 restarting paused alarms 66 selecting the secondary lead (ECG) 134
starting and stopping 359 results table (C.O.) 226 selftest 73
types 360 resuscitation bundle recommendations 353 sensor 191
wave scale 217 disposable SpO2 191
re-triggering events 337
waveforms recorded 363
retrolental fibroplasia (SpO2) 199 sensor temperature 264
wedge 222
review 327 tcGas 263
recording alarms 73
calculations 325, 326 sepsis management bundle 349, 354
recording events 342 recommendations 353
reviewing alarm messages 71
recording strip code 363 sepsis resuscitation bundle 349, 353
reviewing alarms 71
recycling 417 recommendations 353
reviewing alarms window 71
reference waves, wedge measurement 222 sequence mode 208
RH method 234
rejecting pace pulses 134 service mode 36
computation constant 234
relearning arrhythmia 159 set combiners 421, 423
rhythm status messages (arrhythmia
repolarization tails 136 monitoring) 157 setting the arterial catheter constant
report 312 right heart thermodilution method (PiCCO) 230
cardiac output 225, 378 (C.O.) 233 setting the computation constant (RH) 234
event review 346 right heart thermodilution setup (C.O.) 233 setting up 41
reports 369 rule of six 382 trends 311
alarm limits 377 setting up reports 371
RVSW/RVSWI 226
calculations 325, 326
setting up the sequence 208
choosing paper size 372 S
contents 376 settings 41, 125, 501
drip table 386 safety 456 about 41
drug calculator 381 monitor 49 default 501
ECG 133 measurement settings 39
safety information 152, 188, 239
loops 399, 400, 401 monitor settings 39
BIS 291
patient trends 311 screen settings 41
C.O./CCO 239
realtime report 378 synchronized telemetry and
ECG 133
monitor 125
re-routing 374 Resp 186
529
settings blocks 39 SSC guidelines 349 Tamb 212
severe sepsis screening 349, 351 SSC sepsis 349 task window for ST map 174
criteria 351 ST 164 Tcereb 212
short yellow alarms on/off 161 adjusting alarm limits 161 tcGas 263
signal quality of SpO2 194 baseline, updating 168 sensor temperature 264
sinus and SV rhythm ranges 478, 479 ensuring diagnostic quality 164 transducer, calibrating 247
filtering 164 tcGas site timer 264
site timer 264
measurement points, adjusting 168
tcGas 263 technical alarms messages 81
snippets 166
skin preparation 133 see INOPs 81
ST display 166
ECG 133 telemetry alarms 124
ST map 172 suspending at bedside 124
smartkeys key 35
baseline, updating 174
smoothing rate (BIS) 290 temperature 211
current view 172
snapshots 334 accessories 435
report, printing (ST map) 175
events 334 alarm settings 211
scale, changing 174
connecting probe to monitor 211
source device 401 task window 174
difference, calculating 212
loops 399, 400, 401 trend view 173
dual Temp measurement 212
source tracking 124 trending interval, changing 175
extended label set 212
ECG 133 ST maps 172 first 212
specifications 475 ST point 169 label 44
arrhythmia 154 standard 10-lead placement 142 making a measurement 211
speed 42 standard 3-lead placement 140 probe, disposable 211
recording 360 standard 5-lead placement (ECG) 141 probe, selecting 211
wave speed, changing 42 second 212
standardized rate 381
SpO2 191 tcGas sensor 264
starting monitoring 49
accessories 429 temperature probe 211
active alarm source 200 status log 417
testing alarms 73
alarms specific to SpO2 194 printing 417
time, setting 47
arterial pulsation 191 status messages 157
printer 373, 374, 375 timers 395, 396, 397
assessing suspicious reading 194
recorder 413 counting direction 396
calculating difference between
displaying on main screen 397
values 201 status messages (arrhythmia) 157, 158
notification 397
connecting the cables 192 ectopic 158
selecting label 396
disposable sensors 191 rhythm 157
viewing 311
dual SpO2 201 sterilizing 411
extension cable 192 Tinj Probe Type 230
infection control 411
FAST technology 191 titration table (drug calculator) 385
stopping reports printouts 371
Nellcor adhesive sensors tone configuration, alarm 62
(disposable) 429 surgical ECG cable 152
tone mod (SpO2) 201
perfusion indicator 191, 192, 200 surviving sepsis campaign 349
tone modulation 201
Philips sensors (disposable) 429 suspended alarm 66
Philips sensors (reusable) 429 touchscreen 30
suspicious SpO2 reading 194
pleth as alarm source 200 disabling 30
SV/SI 226
pleth wave 200 TP numeric (BIS) 290
pleth waveform 191 SvO2 271 on/off 290
pulse numerics 192 accessories 242, 256
trackball 30
signal quality 194 catheter insertion 274
catheter preparation 273 trackball, using 30
site inspection 192
in-vivo calibration 274, 276 trademarks 445
tone modulation 201
light intensity calibration 274 transcutaneous gas measurements 263
SpO2 desat alarm 200
monitoring 36 transducer 215, 247, 265
SpO2 limit alarms 199
SVR/SVRI 226 CO2, accuracy 247
Sp-vO2 277 CO2, calibrating 247
system pulse 181
SQI numeric 290 pressure, zeroing 215
systolic blood pressure, NBP, early 206
on/off 290 tcGas, calibrating 265
SR numeric (BIS) 290 T transferring centrally-monitored
on/off 290 patients 113, 114
tabular trends 313 using IIT 114
530
transferring patients with MMS 115 language conflicts 390
trend time 320 module 387
global 320 module options 387
trend view 173 module setup 388
ST map 172 visible waves and numerics 387
trending interval 175 W
ST map 172
trends 311 wave 42, 217, 250, 260
automatic unit conversion 328 changing speed 42
automatic value substitution 328 scale (CO2) 250, 260
database configuration 319 scale (pressure) 217
resolution 316 size (CO2) 250, 260
screen trends 320 size (pressure) 217
setting parameter scales 317 wave channel speed 42
setup 230 wave group speed 42
viewing 311 wave scale (EEG) 290
trends pop-up keys 312 changing 37
trends priority list 316 wave size 136
ST map 172 Resp 186
trigger conditions 337 wave size (ECG) 136
events 337 wave speed 42
triggers 336 eeg speed 42
for events 336 wave speed 42
troubleshooting 417 global speed 42
CO2 247, 248, 249 respiratory speed 42
trunk cables 419, 422 wave speed (EEG) 283
Ttymp 212 wave speed (Resp) 188
Tvesic 212 wedge 222
balloon inflation 222
U editing 223
pulmonary artery 222
unfiltered ECG signal 137
reference waves 222
upgrading 284
wide alarms limits 70
effect on EEG configuration 284
Z
V
zero 215
V electrode placement (ECG) 142
effect on pressure alarms 215
venous puncture 210
zeroing 215
ventilation calculations 325 pressure transducer 215
viewing arrhythmia waves 156
viewing trends 311
visible waves report 372
visitor screen 39
visual latching (arrhythmia alarms) 161
vital signs 312
recording 319
report 319
vital signs report 319
volume 62
alarm 62
volume-flow loops 399
VueLink 387
alarm messages 389
device driver conflicts 390
531
532
Part Number 453564443141
Published in Germany 11/2013
*453564443141*