Operating Manual
www.wooyoungmed.com
Log
o
Important
2
Warnings
3
Warnings
4
Cautions
5
Table of Contents
Introduction ......................................................... 7
Features of the Accumate® 2300
– Front View & Key Pad ................................................................. 8
– Controls and Indicators ................................................................. 9
– Rear View ........................................................... 10
Directions for Operation
– Preparation of the IV-set ..............................................................12
– Turning ON the Power - Rear ............................................................13
– Turning ON the Power - Front ...........................................................14
– Insertion of the IV-set ................................................................15
– Setting the Flow Rate and Delivery Limit ...........................................17
– Clearing the Volume Delivered ..........................................................18
– Operating the Pump .............................................................18
– Purge Infusion (stop mode) ...............................................................19
– Bolus Infusion (start mode) ...............................................................19
– Titration .................................................................. 20
– Da ta L ock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
– Adjust Sound .................................................................. 22
– Occlusion Level ................................................................ 23
– Near End of Infusion ............................................................ 24
– End of Infusion ................................................................... 25
Alarms and Troubleshooting .....................................................26
Specifications .......................................................... 28
Maintenance
– Routine Maintenance Procedures ....................................................30
– Cleaning and Storage ..................................................................31
– Replacing the AC fuses ...................................................................32
– Battery operation ................................................................. 33
Occlusion Pressure Limits ........................................................34
Start-up & Trumpet Curves ......................................................35
Warranty ................................................................38
Service and Packaging ............................................. 39
Contact ................................................................ 40
6
Introduction
The Advantages
High precision.
Purge, automatic & manual bolus capability.
Easy to operate in only three steps.
– Power On > Select (rate, total, VTBI) > Start
Easy to read, large display.
History function.
– Historical data of infusion records.
(warnings, settings, starts and stops of infusion,
etc.)
Micro-infusion for neonatal and high risk application.
None dedicated sets, available from various
manufacturers.
Programmable, ultrasonic air bubble detector.
Multi-position clamp for vertical or horizontal fixation
to poles.
SAFETY STANDARD
The Accumate® 2300 meets all safety standards for
medical electrical devices, corresponding to IEC
60601-1 and IEC 60601-2-24.
7
Features of the Accumate® 2300
-Front View & Key Pad
a
n
m
b l
c k
d j
i
e
f h
g
o y
p
q x
r
s w
v
t u
8
Controls and Indicators
9
Controls and Indicators
10
Features of the Accumate® 2300
-Rear View
a
b l
d k
e j
f g h i
NO. Name Function
a Multi-position Clamp To fix the pump to vertical pole stand or horizontal bar of bed
c Drop Sensor Port To connect the drop sensor with the pump
d Nurse Call/ Comm. Port To connect the nurse call jack or communication cable
i AC Outlet/Fuse Socket To connect the AC 90~245V main power cord where the 1.25A fuse is
11
Direction for Operation
chamber
close
1/3
12
Direction for Operation
13
Direction for Operation
14
Direction for Operation
15
Direction for Operation
CAUTION : Make sure that IV tube is NOT loosely fastened into the internal
chamber of the pump. It should be straightforward.
16
Direction for Operation
17
Direction for Operation
18
Direction for Operation
19
Direction for Operation
CAUTION : The pump will not titration if the lock has been enabled.
20
Direction for Operation
21
Direction for Operation
22
Direction for Operation
23
Direction for Operation
caution : The end of infusion to change can be set from 0(no set)
to 15minutes.
24
Direction for Operation
25
Alarms and Troubleshooting
26
Alarms and Troubleshooting
Alarm
Indication & Possible Cause Action
Sound
Air-in-line Air bubble air-in-line sensor Remove the air from the tube
Air
bb-bb-bb IV tube not properly inserted Completely insert the IV
into the Bubble Sensor tube into the Bubble Sensor
Door Open Door opened during infusion Close the door
OPEn
bb-bb-bb
Occlusion Downstream tubing kinked Correct kinks
OccL Filter blocked Correct blocked filter
bb-bb-bb Closed roller clamp Open roller clamp
IV positional or infiltrated Check IV site
Catheter or vein too small
Occlusion limit too low Increase occlusion limit
Drive Error Motor malfunction Request A/S
Err Mechanical error Request A/S
bb-bb-bb
Low battery 30 minutes until battery is fully Connect pump to AC
BAtt discharged power
b-b-b
End of infusion Programmed infusion is Stop infusion and reset
End, conP complete
b-b-b
No Good Self-Test NG Request A/S
ng-? Internal Malfunction
silence
27
Specifications
28
Specifications
EMS DC 12V
Dimensions 140(W)*225(H)*140(D)mm
29
Maintenance
30
Maintenance
31
Maintenance
32
Maintenance
Battery Operation
The internal rechargeable battery allows continued
operation when the AC power is unavailable, for
example during patient transfer or AC power failure.
A fully charged battery will provide over 4 hours of
operation at typical infusion rates.
A low battery alert is activated approximately 30
minutes before the battery depleted alarm.
The battery is maintenance free, sealed Nickel Metal
Hydride and requires no routine servicing. However,
to achieve optimum operation, ensure that the battery
is fully recharged after full discharge, before storage,
and at regular 3 month intervals during storage.
Charge retention will eventually degrade. Where
retention is critical the internal battery should be
replaced every 3 years.
It is recommended that only a qualified service
engineer replaces the Battery. For further information
regarding the replacement of batteries refer to the
Technical Service Manual.
33
Occlusion Pressure Limits
14
12
10
8
6
4
2
0
0 1:00 2:00 3:00 4:00
min : sec
Bolus Volume
1.6
1.4
1.2
1
ml
0.8
0.6
0.4
0.2
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Occlusion Level(psi)
34
Start-up Curves
& Trumpet Curves
35
Start-up Curves
& Trumpet Curves
36
Start-up Curves
& Trumpet Curves
Start-up (25ml/hr)
30
Rate(ml/hr)
25
20
15
10
5
0
2m
5m
9m
s
m
17
13
17
31
Time
15.00%
10.00%
5.00%
Error(%)
0.00%
17s 1m 5m 11m 19m 31m
-5.00%
-10.00%
-15.00%
Observation Window (mins)
37
Warranty
38
Service & Packaging
Service
– If the pump fails to respond to the operating or
troubleshooting procedures listed in this manual
and the cause cannot be determined, discontinue
use and forward it to an authorized Woo Young
Medical's agency.
– With each complaint, please include pump serial
number and full description of difficulty, including
all setting, types of fluids, time, and alarm
messages.
– Return IV-set used if possible.
– Woo Young Medical cannot assume
responsibility for loss or damage to returned
pumps while they are in transit.
Packaging
– The pump packaging is reusable.
– Package Contents : Accumate® 2300 Infusion
Pump, power cable, operation manual
39