Protector XL™
Laboratory Fume Hoods
Models
1110 Series
1111 Series
1112 Series
1113 Series
1114 Series
1115 Series
1116 Series
1117 Series
1118 Series
Original instructions
Copyright © 2018 Labconco Corporation. The information contained in this manual and the accompanying
products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves
the right to make periodic design changes without obligation to notify any person or entity of such change.
Warranty
Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and
reasonable labor as a result of normal and proper use of the equipment with compatible chemicals. Broken
glassware and maintenance items, such as filters, gaskets, light bulbs, finishes and lubrication are not
warranted. Excluded from warranty are products with improper installation, erratic electrical or utility supply,
unauthorized repair and products used with incompatible chemicals.
The warranty for Protector® XL™ Laboratory Fume Hoods will expire one year from date of installation or
two years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and only
applies to the owner (organization) of record.
Buyer is exclusively responsible for the set-up, installation, verification, decontamination or calibration of
equipment. This limited warranty covers parts and labor, but not transportation and insurance charges. If the
failure is determined to be covered under this warranty, the dealer or Labconco Corporation will authorize
repair or replacement of all defective parts to restore the unit to operation. Repairs may be completed by 3 rd
party service agents approved by Labconco Corporation. Labconco Corporation reserves the rights to limit this
warranty based on a service agent’s travel, working hours, the site’s entry restrictions and unobstructed access
to serviceable components of the product.
Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages
of any kind. This warranty is exclusive and in lieu of all other warranties whether oral, or implied.
TABLE OF CONTENTS
CHAPTER 1: INTRODUCTION 1
About This Manual 2
Typographical Conventions 3
(Conventions Typographiques)
Your Next Step 4
CHAPTER 2: PREREQUISITES 5
Location Requirements 6
Support Requirements 6
Exhaust Requirements 6
(Exigences électriques)
Electrical Requirements 8
Service Line Requirements 8
Space Requirements 8
CHAPTER 8: TROUBLESHOOTING 37
APPENDIX E: REFERENCES 53
CHAPTER 1
INTRODUCTION
Congratulations on your purchase of a Labconco Protector® XL™ Laboratory
Fume Hood. Your Protector Laboratory Fume Hood is designed to protect you. It
is the result of Labconco’s more than 60 years experience in manufacturing fume
hoods, and users like you suggested many of its features to us.
The Labconco Protector XL Fume Hood has been engineered to provide maximum
visibility in a laboratory, and effectively contain toxic, noxious, or other harmful
materials when properly installed. The Protector XL offers many unique features
to enhance safety, performance, and visibility. To take full advantage of them,
please acquaint yourself with this manual and keep it handy for future reference. If
you are unfamiliar with how fume hoods operate, please review Chapter 4:
Performance Features and Safety Precautions before you begin working in the
fume hood. Even if you are an experienced fume hood user, please review
Chapter 5: Using Your Protector XL, which describes your Protector Hood’s
features so that you can use the hood efficiently.
Figure 1-1
Chapter 2: Prerequisites explains what you need to do to prepare your site before
you install your laboratory fume hood. Electrical and service requirements are
discussed.
Chapter 3: Getting Started contains the information you need to properly unpack,
inspect, install, and certify your laboratory fume hood.
Chapter 5: Using Your Protector XL discusses the basic operation of your fume
hood. Information on how to prepare, use and shut down your Protector Hood are
included.
Chapter 7: Modifying Your Protector XL explains how to modify the fume hood or
add accessories.
Appendix D: Serial Number Tag Description provides current rating code used on
serial number tag.
Appendix E: References lists the various resources available that deal with
laboratory fume hoods.
Typographical Conventions
Recognizing the following typographical conventions will help you understand and
use this manual:
Book, chapter, and section titles are shown in italic type (e.g., Chapter 3:
Getting Started).
Steps required to perform a task are presented in a numbered format.
Comments located in the margins provide suggestions, reminders, and
references.
Critical information is presented in boldface type in paragraphs that are
! preceded by the exclamation icon. Failure to comply with the information
following an exclamation icon may result in injury to the user or permanent
damage to fume hood.
Les informations critiques sont présentées en gras dans les paragraphes qui
sont précédés par l'icône d'exclamation. Ne pas se conformer aux informations
qui suivent une icône d'exclamation peut résulter à la blessure de l'utilisateur
ou à des dommages irréversibles de la hotte aspirante.
Critical information is presented in boldface type in paragraphs that are
preceded by the wrench icon. These operations should only be performed by a
trained certifier or contractor. Failure to comply with the information
following a wrench icon may result in injury to the user or permanent damage
to your hood.
Les informations critiques sont présentées en gras dans les paragraphes qui
sont précédés par l'icône de clé plate. Ces opérations devraient être seulement
exécutées par un professionnel agrée. L'échec pour se conformer aux
informations qui suivent une icône de clé plate peut résulter à la blessure de
l'utilisateur ou à des dommages irréversibles de la hotte.
Important information is presented in capitalized type in paragraphs that are
preceded by the pointer icon. It is imperative that the information contained in
these paragraphs be thoroughly read and understood by the user.
A The A icon indicates the text is specific to the A-Style Combination Sash
Model.
!
CAUTION – See Manual. When this symbol is on a fume hood it indicates a
caution that is detailed in this manual.
PRUDENCE – Consulter le Manuel. Quand ce symbole est sur une hotte
aspirante, il indique une prudence qui est détaillée dans ce manuel.
CAUTION – See Manual. This symbol on the fume hood indicates the
possibility of a pinch hazard.
PRUDENCE – Consulter le Manuel. Ce symbole sur la hotte indique la
possibilité d'un risque de pincement.
If you would like to review how laboratory fume hoods operate, go to Chapter 4:
Performance Features and Safety Precautions.
If your laboratory fume hood is installed and you need to perform routine
maintenance on the cabinet, proceed to Chapter 6: Maintaining Your Protector XL.
CHAPTER 2
PREREQUISITES
Before you install your laboratory fume hood, you need to prepare your site for
installation. Carefully examine the location where you intend to install your
hood. You must be certain that the area is level and of solid construction. In
addition, a dedicated source of electrical power must be located near the
installation site.
Carefully read this chapter to learn the requirements for your installation site:
Location Requirements
The fume hood should be located away from traffic patterns,
! doors, windows, fans, ventilation registers, and any other air-
handling device that could disrupt its airflow patterns. All
windows in the room should be closed.
Support Requirements
DO NOT install the fume hood on a cart, dolly, or mobile
! bench. ALL Protector Hood installations must be permanent
and stationary. The supporting structure usually consists of a
base cabinet and chemically resistant work surface.
Exhaust Requirements
The exhaust duct connection has been designed for 12" nominal duct (12.75" OD)
to allow for minimum static pressure loss while operating at 60 to 100 fpm face
velocities. The 12" diameter exhaust duct also allows for proper transport
velocities away from the hood in the 1000 fpm to 2500 fpm range. The proper
exhaust volume and static pressure loss are listed next for each hood model:
XL 100 160 725, 0.27" 955, 0.34" 1180, 0.46" 1640, 0.31" 760, 0.28" 1000, 0.36" 1235, 0.49" 1715, 0.33"
80 128 580, 0.17" 765, 0.22" 945, 0.29" 1310, 0.20" 605, 0.18" 800, 0.23" 990, 0.31" 1370, 0.21"
60 96 435, 0.10" 575, 0.12" 710, 0.17" 985, 0.11" 455, 0.10" 600, 0.13" 745, 0.18" 1030, 0.12"
50 80 365, 0.07" 480, 0.09" 590, 0.11" 820, 0.08" 380, 0.07" 500, 0.09" 620, 0.12" 860, 0.08"
40 64 290, 0.04" 380, 0.05" 470, 0.07" 655, 0.05" 305, 0.05" 400, 0.06" 495, 0.08" 685, 0.05"
XL 100 62 450, 0.11" 595, 0.13" 735, 0.18" 1025, 0.12" 485, 0.11" 640, 0.14" 790, 0.19" 1100, 0.13"
80 50 365, 0.07" 480, 0.09" 590, 0.11" 820, 0.08" 390, 0.07" 515, 0.08" 635, 0.11" 880, 0.08"
60 38 270, 0.04" 360, 0.05" 440, 0.07" 615, 0.04" 290, 0.04" 385, 0.05" 475, 0.07" 660, 0.05"
XL 100 160 705, 0.25" 930, 0.32" 1150, 0.44" 1600, 0.30" 740, 0.26" 975, 0.34" 1205, 0.47" 1675, 0.32"
W/Bypass 80 128 565, 0.16" 745, 0.21" 920, 0.28" 1280, 0.19" 590, 0.17" 780, 0.22" 965, 0.30" 1340, 0.21"
Block 60 96 425, 0.09" 560, 0.12" 690, 0.16" 960, 0.11" 445, 0.09" 585, 0.12" 725, 0.17" 1005, 0.12"
50 80 350, 0.06" 465, 0.08" 575, 0.11" 800, 0.08" 370, 0.07" 490, 0.09" 605, 0.12" 840, 0.08"
40 64 280, 0.04" 370, 0.05" 460, 0.07" 640, 0.05" 295, 0.04" 390, 0.05" 480, 0.08" 670, 0.05"
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at Static Pressure (inches of water) Static Pressure (inches of water)
62.50% Full open 4' 5' 6' 8' 4' 5' 6' 8'
open at 18" at 28" Hood Hood Hood Hood Hood Hood Hood Hood
XL 100 62 440, 0.10" 580, 0.12" 720, 0.17" 1000, 0.12" 475, 0.10" 625, 0.13" 775, 0.18" 1075, 0.12"
W/Bypass 80 50 350, 0.06" 465, 0.08" 575, 0.11" 800, 0.08" 380, 0.07" 500, 0.09" 620, 0.12" 860, 0.08"
Block 60 38 265, 0.04" 350, 0.05" 430, 0.06" 600, 0.04" 285, 0.04" 375, 0.05" 465, 0.07" 645, 0.05"
Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. Contact Labconco Customer Service for
assistance in sizing a blower system.
Electrical Requirements
The Protector Hood models feature internal wiring for the fluorescent light
assembly and light switch. All internal wiring is terminated at the single point
wiring junction box for hook-up by a qualified electrician. The blower switch,
and light switch wires are also terminated at the single point wiring junction box
for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and
Appendix C: Protector XL Specifications for the wiring diagram for proper
electrical installation.
Space Requirements
The dimensions for the different models are shown in Appendix B: Protector XL
Dimensions.
CHAPTER 3
GETTING STARTED
Now that the site for your laboratory fume hood is properly prepared, you are
ready to unpack, inspect, install, and certify your unit. Read this chapter to learn
how to:
Unpack and move your Protector Hood.
Set up the fume hood with the supporting structure and work surface.
Connect to an exhaust system.
Connect the electrical supply source.
Connect the service lines.
Sealing the Protector Hood to the work surface.
Arrange certification of your Protector Hood.
Depending upon which model you are installing, you may need common
plumbing and electrical installation tools in addition to 5/16", 3/8", 7/16", and
1/2" wrenches, ratchets, sockets, a nut driver set, a flat-blade screwdriver, a
Phillips screwdriver, and a carpenter level to complete the instructions in the
chapter.
!
The Protector XL Hood models weigh between 400 to 1600 lbs.
(182-726 kg). The shipping skid allows for lifting with a
mechanical lift truck or floor jack. If you must lift the fume
hood manually, follow safe-lifting guidelines. Normally, the
fume hood can be slid off a hydraulic lift table and be placed
into position on top of the work surface. Do not lift by the
front air foil.
Les XL modèles de la Hotte Protecteur pèsent entre 400 à 1600
livres. (182-726 Kg). La palette bois d’envoi permet le
soulèvement par un camion muni d’un élévateur mécanique ou
par un cric rouleur. Si vous devez soulever manuellement la
hotte aspirante, respectez les règles de sécurité du soulèvement.
Normalement, la hotte aspirante peut être glissée d'une table
Product Service 1-800-522-7658 9
Original instructions
Chapter 3: Getting Started
that claims
AUTHORIZATION OF LABCONCO. UNAUTHORIZED be filed with
RETURNS WILL NOT BE ACCEPTED. the delivery
carrier
within fifteen
IF YOUR HOOD WAS DAMAGED IN TRANSIT, YOU MUST
(15) days of
FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER. delivery.
LABCONCO CORPORATION AND ITS DEALERS ARE NOT
RESPONSIBLE FOR SHIPPING DAMAGES.
Do not discard the shipping skid or packing material for your fume hood until you
have checked all of the components and installed and tested the unit. The XL
fume hood baffles are shipped loose behind the hood and do not discard. The
upper baffle is 16.5" tall, the middle baffle is 23.3" tall, and the lower
perforated baffle is 17.3" tall. See Figure 1-1, Figure 3-1 and Figure B-1 that
shows proper XL baffle installation. Do not remove the fume hood from its
shipping skid until it is ready to be placed into its final location. Move the unit by
placing a flat, low dolly under the shipping skid, or by using a floor jack.
LEAVE THE FUME HOOD ATTACHED TO ITS SHIPPING
SKID UNTIL IT IS AS CLOSE TO ITS FINAL LOCATION AS
POSSIBLE. MOVE THE HOOD BY USING A SUITABLE
FLOOR JACK, OR BY PLACING A FURNITURE DOLLY
UNDERNEATH THE SKID. DO NOT MOVE THE HOOD BY
TILTING IT ONTO A HAND TRUCK.
After you verify the fume hood components, move your hood to the location
where you want to install it. Should you require disassembly to move the hood,
then follow the instructions in Appendix F. Then, follow the steps listed next to
remove the shipping skid from your unit.
NOTE: THE SASH WEIGHT ITSELF WAS INDIVIDUALLY
MATCHED FOR THIS SPECIFIC HOOD AND SHOULD NOT
BE EXCHANGED ON ANY OTHER UNIT.
!
The Protector Hood is heavy! Use caution when lifting or
moving the unit.
When installing the Protector Fume Hood onto a chemically-resistant work surface
or benchtop, ensure that the structure can safely support the combined weight of the
fume hood and any related equipment. The work surface should be at least as wide
as the hood to properly support it. The work surface is aligned flush with the back
of the fume hood for good airflow: this will provide the correct spacing under
the air foil for proper bypass airflow. The lower base cabinets are placed flush
with the front of the work surface as shown in Figure 3-1.
!
WARNING: It is important to support the rear of the work
surface and fume hood. The cross support provides support
for the bottom of the work surface. Install the cross support
after the base cabinets and work surface are leveled and before
installing the hood.
Figure 3-1
1.7
Place Cross
Support Here
!
The selected exhaust duct material should match the hood
procedures and chemicals used to ensure compatibility.
!
Protector Hoods should include overload protection. A switch
or circuit breaker should be in close proximity to the
equipment and within easy reach of the operator. The switch
or circuit breaker is to be marked as the disconnecting device
for the equipment. Consult the NEC-2002 for proper
installation.
The identification plate, model number, serial number, and electrical connection
boxes are accessible from the front of the fume hood by removing the front panel.
The Protector XL Hood is normally wired for 115 Volt, 50/60 Hz, 20 Amp or
230Volt, 50/60 Hz, 10 Amp electrical service. Check the I.D. plate behind the
front panel for voltage verification. The number of circuits varies depending on
the model. All of the electrical connections are terminated at the field wiring
terminal box for hook-up by a qualified electrician. We recommend each circuit
be a dedicated branch circuit. However, if wired together the maximum load
allowed is the sum of individual outlets plus the rating of the unit (i.e. 2 Amps).
The single point internal junction box is used for the connection of the lights,
blower, and duplex outlets. Refer to the wiring diagram for your Protector XL in
Appendix C: Protector XL Specifications.
The fume hood is required to be grounded to the MAINS protective earthing ground
for safe operation. Using a ring terminal sized for a 10-24 machine screw, connect
the MAINS ground conductor to the grounding lug marked with the protective
earthing symbol, . Only MAINS ground conductors should be connected to the
protective earthing ground lug, no other conductors should be connected to this
grounding lug. Using wire nuts connect the MAINS supply conductors to the fume
hood supply wires. Insure that the wires are connected as per the appropriate wire
color codes for the input voltage. For 115V Phase (Hot) is black and Neutral is
white, for 230V Phase1 is brown and Phase2 is blue. Refer to the wiring diagram for
your Protector XL in Appendix C: Protector XL Specifications.
Internal Junction
Box
Figure 3-2
The fluorescent light has been mounted outside the top liner panel and is sealed
from vapors inside the hood structure. To change the fluorescent light bulbs in
your hood, you must first remove the front panel from the hood. Next remove the
knock out plugs holding the light fixture in place. Lift fixture up and replace any
deflective bulbs. Reverse order to reassemble.
!
NOTE: Inspect all fittings for leakage. Tighten the fittings
slightly if needed.
Should access to the hood plumbing fixture bodies be required, remove the
service access plate on the hood front corner posts by loosening their individual
screws (see item 11, Figure A-1 in Appendix A). The valve body will now be
fully exposed for any service work that may be necessary. The service fixtures
supplied on your laboratory hood are designed for use with the following services:
Your Protector Fume Hood has been tested at the factory per ASHRAE 110-1995.
All hoods achieve an “as manufactured rating” of less than 0.05 part per million
(ppm) at 4 liters per minute (lpm); AM<0.05 (consult Labconco for individual
fume hood ratings). For “field use” ASHRAE testing contact Labconco Sales
Engineering Team or Customer Service for a certified on-site contractor.
!
NOTE: Face velocity profiles and smoke testing should be
done periodically to ensure safe performance.
CHAPTER 4
PERFORMANCE FEATURES
AND SAFETY PRECAUTIONS
Performance Features:
The Protector XL Laboratory Hood is designed to meet the needs of the
laboratory scientist at OSHA approved velocities as low as 60 feet per minute.
The laboratory fume hood has been designed to effectively contain toxic, noxious,
or other harmful materials when properly installed. A fully featured by-pass hood
with baffle and air foil, this hood maintains safe airflow. Optional energy saving
A-Style Combination Sash models are also available. The hood features by-pass
airflow design that promotes full containment as the sash is moved. Airflow is
diverted behind the front panel and under the air foil to help control fluctuations
in face velocity, which occur as the sash is closed.
17
6
16
7
22
20
11
12
1
14
15
9
18
4 8
10 23
21
Figure 4-1
Safety Precautions
Although the laboratory hood has been engineered to maintain
! optimum operator safety, caution should always be used while
working in the hood. Prior to using the hood, check to make
sure that the exhaust blower is operating and that air is
entering the hood at its specified face velocity.
Avoid placing your head inside hood. Keep hands out of hood
as much as possible.
AVOID CROSS DRAFTS AND LIMIT TRAFFIC IN FRONT OF
THE HOOD. AIR DISTURBANCES CREATED MAY DRAW
FUMES OUT OF THE HOOD.
! Do not touch the blower motor. The surfaces of the motor can
become hot and could cause burns.
CHAPTER 5
USING YOUR PROTECTOR XL
Start-up
Turn on light and hood blower.
Slowly raise the sash.
Check the baffle air slots for obstructions.
Allow the hood to operate unobstructed for 5 minutes.
Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.
Wear a protective mask if appropriate.
Work Techniques
Keep all materials at least 6 inches inside of the sash, and perform all
contaminated operations as far to the rear of the work area as possible.
Segregate all clean and contaminated materials in the work area.
Avoid using techniques or procedures that disrupt the airflow patterns of
the hood.
Final Purging
Upon completion of work, the hood should be allowed to operate for two
to three minutes undisturbed, to purge airborne contaminants from the
work area before shutting down blower.
Shutdown
Turn off the light and hood blower, then close the sash.
CHAPTER 6
MAINTAINING YOUR
PROTECTOR XL
Now that you have an understanding of how to work in the fume hood, we will
review the suggested maintenance schedule and the common service operations
necessary to maintain your fume hood for peak performance.
Determine the actual face velocity through the sash opening of the hood
where the average reading should be at the specified velocity. (Use
calibrated thermal anemometer or other approved apparatus).
Déterminer l’actuel flux d’aspiration à travers le sas d’entrée de la hotte où
la valeur moyenne doit être égale à la valeur spécifiée. (Utiliser un
anémomètre thermique calibré ou d’autres appareils approuvés).
Using a damp cloth, clean the exterior surfaces of the hood, particularly
the front of the hood, to remove any accumulated dust.
En utilisant un chiffon humidifié, nettoyer les parties extérieures de la
hotte, en particulier le devant de la hotte pour enlever la poussière
accumulée.
Check all service valves, if so equipped, for proper operation.
Contrôler toutes les soupapes, si présentes, pour le bon fonctionnement.
The hood baffles should be checked for blockages behind them to ensure
that the hood is maintaining proper airflow.
Les déflecteurs de hotte devraient être contrôlés pour leur blocage arrière
afin d’assurer que la hotte maintient un flux d’air correct.
All weekly activities.
Toutes les activités hebdomadaires.
Annually
Annuellement
Biannually
The sash assembly should be checked to ensure proper operation and to
make sure there are no signs of abnormal wear on the sash pulleys, cables
and clamps.
1. Simply lift the front panel up and then away from the hood to provide access
to the top.
CHAPTER 7
MODIFYING YOUR
PROTECTOR XL
There are several ways to modify the fume hood for your individual requirements.
These include the addition of work surfaces, service fixtures, air monitor,
distillation grids, electrical duplex outlets, ceiling enclosures, and rear panels.
Figure 7-1
Configure Your Non-Adjustable Fixed Height Ceiling Enclosure Kit Catalog Number
Use this key to configure the seven digit catalog number to order your 3-Sided Fixed Height
Ceiling Enclosure Kit. For example, 9550511 is a 3-Sided Ceiling Enclosure Kit for a 5' x 37.7"
deep Protector XL Hood with a fixed height of 12".
Side Panel
(Ceiling Enclosure)
Rear Stiffener
Side Panel
(Ceiling Enclosure)
Figure 7-2
Figure 7-3 - Knob Figure 7-4 - Valve Figure 7-5 – Hose Connector
Figure 7-6
Figure 7-7
CHAPTER 8
TROUBLESHOOTING
Refer to the following table if your fume hood fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance.
Vertical sash no Cable is frayed or Inspect cable and replace cable if worn or
longer operates plastic protection is damaged immediately; otherwise injury
smoothly damaged. could result.
Service valves no Faulty building Inspect building supply shut off valves
longer operate supply. and appropriate pressures below the
recommended 40 PSI.
Supply line or outlet Inspect line for leaks and fix any
line has leaks. leaking plumbing connections.
APPENDIX A
PROTECTOR XL COMPONENTS
Illustration A-1 indicate the location of the following service parts:
15
16
14
18
22 8
11
13
5, 6
2
12
10, 21
Figure A-1
APPENDIX B
PROTECTOR XL DIMENSIONS
XL Model Widths
4' 5' 6' 8' 10' 12' 16'
A 48.00 60.00 72.00 96.00 120.00 144.00 192.00
External Depth
B 31.7, 37.7, or 43.7
C 38.25 50.25 62.25 86.25 110.25 134.25 182.24
Internal Depth
D 24.00, 30.00, or 36.00
Filler Panel
E 8.00, 14.00, or 20.00
Duct C/L C/L C/L 48.00 50.00 62.00 48.00 Four
Spacing One Duct One Duct One Duct Two Ducts Two Ducts Two Ducts Ducts
NOTE: See Chapter 3 for Dimension E.
Hood Size DIM “A” DIM “B” DIM “C” DIM “D”
24" Depth 31.7 24
3ft 30" Depth 36.00 37.7 26.25 30
36" Depth 43.7 36
24" Depth 31.7 24
4ft 30" Depth 48.00 37.7 38.25 30
36" Depth 43.7 36
24" Depth 31.7 24
5ft 30" Depth 60.00 37.7 50.25 30
36" Depth 43.7 36
24" Depth 31.7 24
6ft 30" Depth 72.00 37.7 62.25 30
36" Depth 43.7 36
24" Depth 31.7 24
7ft 30" Depth 84.00 37.7 74.25 30
36" Depth 43.7 36
24" Depth 31.7 24
8ft 30" Depth 96.00 37.7 86.25 30
36" Depth 43.7 36
Figure B-1
Product Service 1-800-522-7658 43
Original instructions
APPENDIX C
PROTECTOR XL
SPECIFICATIONS
Environmental Conditions
APPENDIX D
SERIAL NUMBER TAG
DESCRIPTION
Serial tag includes standard information with the following changes:
The portion of the tag for electrical information is labeled with the following
units: Volts AC, Phase, Cycle, AMP.
The X above will be one of several possible alpha characters as defined in the
manual as follows:
ANNEXE D
DESCRIPTION DE SÉRIE TAG NUMBER
Étiquette de série comprend des informations standard avec les modifications
suivantes:
APPENDIX E
REFERENCES
Many excellent reference texts and booklets are currently available. The
following is a brief listing:
Industrial Ventilation-ACGIH
Fume hood face velocities between 60-100 lfpm
Maximum of 125 lfpm for radioisotope hoods
Duct velocities of 1000-2000 fpm for vapors, gasses and smoke
Stack discharge height 1.3-2.0 x building height
Well designed fume hood containment loss, <0.10 ppm
Industrial Ventilation, A Manual of Recommended Practice.
24th Edition, 2001
American Conference of Governmental Industrial Hygienists
1330 Kemper Meadow drive
Cincinnati, OH 45240-1634
(513) 742-2020
SEFA 1-2002
Fume hood face velocities based on toxicity levels of chemicals
Class A – 125 to 150 fpm
Class B – 80 to100 fpm
Class C – 75-to 80 fpm
Test method – face velocity profile and smoke generation
Scientific Equipment & Furniture Association
1028 Duchess Drive
McLean, VA 22102
(703) 538-6007
General References
DiBerardinis. L. et al. Guides for Laboratory Design, Health and Safety Considerations.
Wiley & Sons, 1987
Miller, Brinton M. et al. Laboratory Safety: Principles and Practices. American Society
for Microbiology, Washington, D.C.: 1986
NIH Guidelines for the Laboratory Use of Chemical Carcinogens. NIH Publication No.
81-2385.
Rayburn, Stephen R. The Foundation of Laboratory Safety, A Guide for the Biomedical
Laboratory. Springer-Verlag, New York: 1990
Sax, N. Irving and Lewis, JR., Richard J. Rapid Guide to Hazardous Chemicals in the
Workplace. Van Nostrand Reinhold, 1987.
Schilt, Alfred A. Perchloric Acid and Perchlorates. The G. Frederick Smith Chemical
Company, Columbus, OH: 1979.
Steere, Norman. CRC Handbook of Laboratory Safety, 2nd Edition. CRC Press, 1971.
APPENDIX F
DISASSEMBLY AND
REASSEMBLY INSTRUCTIONS
Required Tools
1. On the top of the hood, disconnect the Molex connector and remove the green ground
wire. All the removable wires are part of the wiring harness to the corner post
switches. Be sure to label all wires and refer to the wiring diagrams for proper
reassembly.
Header Removal
1. The header is fastened to the corner posts by four #12 screws. Simply reach behind
the corner posts from the side and remove the screws while another person supports
the header.
Sash Removal
There is one sash to be removed on the 4', 5', and 6' hoods. Please note that there are two
sashes that need to be removed for the 8', 10', and 12' hoods and four sashes on the 16'
hood. Please repeat the procedure for each sash.
1. For 8' through 16', remove the bumper from the middle sash slide by removing the
screw. Be careful not to misplace the loose parts. This part is a crucial component to
the operation of the sash.
2. For 8' through 16', remove the middle sash slide assemblies by sliding them up and
out.
3. Remove the sash end stops by removing the hardware.
4. Remove end cable and pulleys from sides.
5. Now the glass sash assembly can be removed by sliding it up and out.
6. Remove sash weights. The weights were shipped attached to the back.
Please note that the 4' through 8' hoods are designed with one continuous top liner.
Therefore, the top support is not required for the 4' through 8' hoods, but is used on the
10', 12', and 16' hoods.
1. Remove the top support by removing the 1/4 –20 hardware and screws.
2. Remove the anti-rack shaft assemblies by removing the screws.
3. Then remove the top liner assemblies by removing the screws.
Front Truss Removal (if applicable on larger hoods 8' and up)
1. Remove the front truss support from the side frames by removing the 1/4 –20 x 1.5"
hex head screws, washers, lockwashers, and nuts. Simple reach behind the side
frames while somebody supports the front truss and remove it.
Airfoil Removal
1. Remove the rear frames from the side frames by removing screws from each side
frame.
2. If desired, you can remove the right rear frame from the left rear frame by removing
the 1/4-20 hardware, angle support, and screws. This step is not always necessary
and depends on the installation and size of the hoods.
The various subassemblies are now ready to be transported and reassembled in the
opposite way the hood was disassembled. All hoods will require a minimum of five
people to lift the hood, two people on each side and one in front (for balance only). A
hydraulic lift table should be used as an aid in lifting. The 16' hood should be assembled
where placed. If in doubt about lifting the hood, always reassemble the hood where it
will be placed.
Reinstallation
Make sure the anti-rack shaft does not slide from left to right as this will be noisy and not
work effectively.
After the hood is assembled, the sash weights are easily installed before the hood is
placed against a rear wall. Please keep this in mind and install the sash weights now.
Baffles Installation
Install the upper baffle, middle baffle, and lower baffle. Be sure the baffles are resting in
the proper baffle mount supports.
The hood is now ready for placement and the HVAC, plumbing and electrical service can
be installed.