I. Principle of measurement
Fluorescence DO meter is based on the quenching principle of special matter against active
fluorescence in physics. The outside surface of fluorescent cap covers a layer of black light
insulation materials to prevent from the interference of sunlight and other fluorescent materials in
water, the internal surface of which covers florescent materials. When the blue ray radiated from
LED shines the fluorescent materials on internal surface of fluorescent cap, it will excite the
fluorescent materials to emit red light. Through detection of the phase difference between red light
and blue light, calculate the concentration of oxygen molecule, and obtain accurate data by means
of temperature, air pressure and salinity compensation.
DOD-2000 digital sensor directly outputs digital signal of RS485 interface, uses Modbus
RTU communication protocol to directly connect the computer, display and wireless remote
repeater to realize undistorted teletransmission, which is the best mode to conform to internet+
combination at present.
signal corrected interiorly, can be directly connected to PLC, mA conversion module, etc. to
switch data into system, achieving the wired teletransmission function. The external electronic box
between cables directly measures the atmosphere of environments to make the sensor have
one-click calibration operating function, convenient to use and operation.
The difference between fluorescence DO and the DO probe technique of polarography
electrochemistry principle lies in that the fluorescence method does not consume DO, the interior
not need electrolyte filling , and even not need the frequently repeated calibration or frequent
cleaning (except apply to adhesion liquid), which provides much longer service life and more
stable and accurate readings. The measuring indication is unrelated with water flow rate, so it can
be used in the situation at low flow rate and even in stagnant condition.
♦DOD-2000 digital sensor directly outputs the RS485 signal of Modbus RTU communication
protocol;
♦The electronic box between cables provides atmospheric pressure detection, salinity
compensation and signal processing;
♦With long service life(not less than one year) and rarely need of maintenance and slope
calibration;
♦Only fluorescent cap is replaceable consumables which is need of change once every year;
♦ Rarely no requirements of flow rate of water sample, suitable for a wide range of condition;
♦ Multiple mounting accessories for selection at will to meet the requirements of various
industries and conditions.
Recommended optional accessories:
MFC-2300 universal multifunctional controller
MCC-2300 universal multiline controller
Stainless steel sling SRS-100
ESB-100 3-DOF environmental protection sensor stand
Mounting bracket of environmental sensor with floating ball
Wireless relay module(GPRS/Zigbee/3G/WIFI, etc.)
Remarks: the service life of fluorescent cap at the front end of sensor will be affected by the direct
or reflected sunlight negatively. If not use (such as, during the period of water tank maintenance),
please cover a black plastic protection cap, and keep it in a dark place until repeat to install. If user
exposes the sensitive fluorescent cap of sensor directly to sunlight or reflected sunlight over one
week, the warranty of sensor will automatically expires.
1.3 Main parameters of transmitter
Power supply DC24V±15%
DO measuring range (0~20)mg/L, (0~100)%
Accuracy 0.1mg/L or ±1.0%of reading
Signal output (4~20)mA
Air pressure measuring range (51~112)kPa
Temperature measuring range (-5~50)°C
IV. Calibration
4.1 Basic introduction
The measuring principle of DOD/DOT-2000 fluorescent DO transmitter is to measure the
solubility of oxygen in water, and the solubility of oxygen depends on the temperature, pressure
and salinity of water;
Different from polarography DO electrode, it does not need frequent calibration or frequent
cleaning, but in order to reach measuring accuracy, it can be subject to the specific applications or
the requirements of competent department. When the reading of fluorescent DO transmitter
deviates, the slope of transmitter shall be calibrated; when the fluorescent cap of sensor in dry
place for a long time or changing it, it shall need hydration and calibration again.
4.2 DO measurement influencing factors
1) Influence of temperature: the oxygen saturation in water is different in different
temperature, which refers to table as follows:
The oxygen saturation list in water in different temperature
(The atmospheric pressure is 101.3252kPa, the humidity is 100%RH and salt content is .
The inside of this transmitter has thermistor to detect the temperature of solution in real time,
and performs temperature compensation to oxygen concentration, which does not require
calibration.
(2) Atmospheric pressure influence: according to Henry law, the oxygen saturation solubility
in water is proportional to partial pressure. The partial pressure of oxygen is related to the altitude
in the region. It installs an air pressure sensor in junction box of the sensor, which can realize
correction and compensation function during calibration, so the client does not resets the air
pressure;
(3) Salinity in water: the salinity refers to the salt content in water. As DO in saline water is
clearly lower than that in running water, in order to get an accurate measurement, it shall consider
the influence of salt content on DO; the salinity in water is inversely proportional to oxygen
saturation solubility. The sensor system does not have salinity compensation function, and instead
it can use the programmable controller and other host computers for handling.
4.3 Calibration slope
Depending on the fluorescent oxygen quenching principle in the back of penetrative DO
materials, fluorescent DO sensor gets digital signal after mathematical treatment. In order to get
accurate data, it is better calibrate the sensor before using. During calibration, please put the
electrode in calibration environment for calibration. After calibration, calibrate it comparing with
standard value. The calibration steps are as follows:
1. Before calibration of fluorescent DO sensor, it should be soaked in water for proper
activation. If the sensor film cap is dry, it requires of soaking the sensor in water for pretty wet all
night, and fully hydrating the fluorescent film cap of sensor;
2. Air calibration: air calibration is simple and reliable, which is the most common calibration
method for DO sensor. Fully polarize the film cap of sensor, use soft tissue to dry the water on
surface of film cap, and then put the sensor in the air at least for 15 minutes. After the value is
stable, conduct the air calibration.
3. Calibration of given DO concentration: the fluorescent DO sensor can be calibrated in water
sample with known saturated DO concentration. Put the activated fluorescent DO sensor in the
water sample with saturated DO concentration and wait for 5 minutes.
4. Calibration method: after fluorescent DO sensor putting in air or water with saturated DO
concentration and it becoming stable, put one needle in the hole marked "calibration (CAL)" on
transmitter box, then press it for 3 seconds and release it.
V. Typical applications
Typical applications: effluent, river, lake and fishpond.
The aeration tank, nitrifying tank, denitrifying tank, aerobic tank and anaerobic tank in
sewage treatment plant.
Measure DO in course of fresh water and seawater aquiculture.
Monitor ecology water resource for river and lake, and environment monitoring;
Monitor chemical deoxidization and thermal deaerator, etc.
On-line DO analysis of water tank, municipal tap water, tap water network and water in other
fields.
VII. Completeness
DOD/DOT-2000 fluorescent DO transmitter one set
DOD/DOT-2000 fluorescents DO transmitter instructions one volume
Cable 10m (bookable)