Electronic Total Station Series: Instruction Manual
Electronic Total Station Series: Instruction Manual
Electronic Total Station Series: Instruction Manual
GTS-720 SERIES
GTS-721
GTS-722
GTS-723
GTS-725
FOREWORD
Thank you for purchasing the TOPCON Electronic Total Station, GTS-720 series.
For the best performance of the instruments, please carefully read these instruc-
tions and keep them in a convenient location for future reference.
1
FOREWORD
If the above-mentioned operating procedure is not followed, then, the next time that power
is turned on, it will be necessary to reboot the GTS-720.
2
FOREWORD
Display Meaning
Ignoring or disregard of this display may lead to the danger of death or
WARNING
serious injury.
CAUTION Ignoring or disregard of this display may lead to personal injury or phys-
ical damage.
•Injury refers to hurt, burn, electric shock, etc.
•Physical damage refers to extensive damage to buildings or equipment and furniture.
Safety Cautions
WARNING
•There is a risk of fire, electric shock or physical harm if you attempt to disassemble or
repair the instrument yourself.
This is only to be carried out by TOPCON or an authorized dealer, only!
•Cause eye injury or blindness.
Do not look at the sun through a telescope.
•High temperature may cause fire.
Do not cover the charger while it is charging.
•Risk of fire or electric shock.
Do not use damaged power cable, plug and socket.
•Risk of fire or electric shock.
Do not use a wet battery or charger.
•May ignite explosively.
Never use an instrument near flammable gas, liquid matter, and do not use in a coal mine.
•Battery can cause explosion or injury.
Do not dispose in fire or heat.
•Risk of fire or electric shock.
Do not use any power voltage except the one given on manufacturers instructions.
•Battery can cause outbreak of fire.
Do not use any other type of charger other than the one specified.
•Risk of fire.
Do not use any other power cable other than the one specified.
•The short circuit of a battery can cause a fire.
Do not short circuit battery when storing it.
3
FOREWORD
CAUTION
•Do not connect or disconnect equipment with wet hands, you are at risk of electric shocks if you
do!
•Risk of injury by overturn the carrying case.
Do not stand or sit on the carrying cases.
•Please note that the tips of tripod can be hazardous, be aware of this when setting up or carry-
ing the tripod.
•Risk of injury by falling down the instrument or case.
Do not use a carrying case with a damaged which belts, grips or latches .
•Do not allow skin or clothing to come into contact with acid from the batteries, if this does occur
then wash off with copious amounts of water and seek medical advice.
•A plumb bob can cause an injury to a person if used incorrectly.
•It could be dangerous if the instrument falls over, please ensure you attach a handle to the in-
strument securely.
•Ensure that you mount the tribrach correctly, failing to do so may result in injury if the tribrach
were to fall over.
•It could be dangerous if the instrument falls over, please check that you fix the instrument to
the tripod correctly.
•Risk of injury by falling down a tripod and an instrument.
Always check that the screws of tripod are tightened.
4
FOREWORD
User
1)This product is for professional use only!
The user is required to be a qualified surveyor or have a good knowledge of surveying, in order to
understand the user and safety instructions, before operating, inspecting or adjusting.
2)Wear the required protectors (safety shoes, helmet, etc.) when operating.
5
FOREWORD
Labels
The following labels are found on the instruments which describe the GTS-720 series Laser plummet
type:Precautions and safety information about the laser beam.
If, at any time, any of these labels are damaged and become illegible, please replace these important
labels. Please the new labels in exactly the same position as the original labels.
Replacement labels can be obtained from Topcon or your authorized Topcon dealer.
Explanatory Label
CAUTION
LASER RADIATION-DO NOT
STARE INTO BEAM
WAVE LENGTH 633nm
1mW MAXIMUM OUTPUT
CLASS II LASER PRODUCT
Aperture Label
Beam aperture
Depending on the country where the instrument is sold, either of these labels may be
found on the GTS-720 series laser plummet type.
6
FOREWORD
Contents
FOREWORD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Handling Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Display for Safe Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
User. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Exceptions from Responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Standard for Laser Beam of Laser Plummet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Symbol Mark while the Laser is Emitting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Standard Set Composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1 NOMENCLATURE AND FUNCTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 Nomenclature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.2.1 Main Menu Contains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.2.2 Measurement Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.2.3 Display Marks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.2.4 Display keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.2.5 Shortcut Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1.3 Backlight Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.3.1 How to Adjust Reducing Time of Backlight . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.3.2 Adjust the Backlight Brightness by Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1.3.3 Selecting the Automatic Lighting Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
1.4 RAM Data Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
1.4.1 Execute the Backup Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
1.4.2 Set the Automatic Backup for Every Suspension . . . . . . . . . . . . . . . . . . . . . . . 1-12
1.4.3 Set the Restoration Disabled after Hardware Reset . . . . . . . . . . . . . . . . . . . . . 1-12
1.5 Hardware Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
1.6 Cover Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
1.7 Touch Panel Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
1.8 Operating Panel Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
1.8.1 Operating Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
1.9 Power OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
1.10 Function Key (Soft Key) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
1.11 Star Key Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
1.11.1 Setting by Using Star Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-22
1.12 Auto Power Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23
2 PREPARATION FOR MEASUREMENT . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1 Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 Setting Instrument Up For Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.3 Power Switch Key ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.4 Battery Power Remaining Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2.5 Vertical and Horizontal Angle Tilt Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.5.1 Setting Tilt Correction by Soft Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2.6 Compensation of Systematic Error of Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.7 How to Enter Numerals and Alphabet Letters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.8 Data Memory Card. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2.9 Active Sync . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2.9.1 Getting Connected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
3 STANDARD MEASUREMENT MODE. . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1 Angle Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1.1 Measuring Horizontal Angle Right and Vertical Angle. . . . . . . . . . . . . . . . . . . . . 3-1
3.1.2 Switching Horizontal Angle Right/Left . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.1.3 Measuring from the Required Horizontal Angle . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.4 Vertical Angle Percent Grade(%) Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.2 Distance Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.2.1 Setting of the Atmospheric Correction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.2.2 Setting of the Correction for Prism Constant. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.2.3 Distance Measurement (Continuous Measurement). . . . . . . . . . . . . . . . . . . . . . 3-5
3.2.4 Distance Measurement (Single/N-times Measurement) . . . . . . . . . . . . . . . . . . . 3-6
3.2.5 Fine/ Tracking / Coarse Measuring Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
7
FOREWORD
8
FOREWORD
(Make sure that all of the above items are with the instrument when purchased.)
9
1 NOMENCLATURE AND FUNCTIONS
Point guide
Objective lens
Instrument
center mark
Optical plummet
telescope
Display window
(With touch panel)
GTS-721/722/723
Power supply
Operation keys connector
GTS-721/722/723
Communication port
(Serial signal connector RS-232C)
1-1
1 NOMENCLATURE AND FUNCTIONS
2speed way:
GTS-721/722/723
1speed way:
GTS-725
Power switch
Telescope focusing knob
Vertical motion clamp
Instrument
Telescope grip center mark
Card cover
Horizontal
tangent screw
2speed way:
Plate level
GTS-721/722/723
1speed way:
GTS-725
Display window
(With touch panel) Horizontal
motion clamp
Operation keys
Cover sensor
USB port
Tribrach fixing lever (Inside the cover)
Stylus pen
*1) The position of vertical motion clamp and tangent screw will differ depend on the markets.
1-2
1 NOMENCLATURE AND FUNCTIONS
1.2 Display
1.2.1 Main Menu Contains
The main menu contains as following items.
Select the menu by pressing icons.
Display icon
ADJUSTMENT MODE
This mode is used for checking and adjustment.
● Error of vertical angle 0 datum
● Setting instrument constant value
● Compensation systematic error of Instrument
● Setting FRQ check mode
(see Chapter 5 “CHECK AND ADJUSTMENT” .)
1-3
1 NOMENCLATURE AND FUNCTIONS
V-angle: 90°30’35”
H-angle:120°00’00”
Soft keys
✻ EDM working
Distance
To be distance measuring mode.
measuring key
Coordinate
measuring key To be coordinate measuring mode.
1-4
1 NOMENCLATURE AND FUNCTIONS
1-5
1 NOMENCLATURE AND FUNCTIONS
You can adjust the settings of this function to conserve more battery power or set your liking.
1-6
1 NOMENCLATURE AND FUNCTIONS
4 Press the [OK] key on title bar. After that "Power Properties" screen will close automatically.
1-7
1 NOMENCLATURE AND FUNCTIONS
3 Press the [OK] key on title bar. After that "Power Properties" screen will close automatically.
1-8
1 NOMENCLATURE AND FUNCTIONS
1 On the "Backlight" screen, select a Radio button from “Automatic lighting” column.
(Factory setting is “The light is switched on with an illuminometer.” as default)
2 Press the [OK] key on title bar. After that "Power Properties" screen will close automatically.
The “Time until it switches off a backlight.” time-menu is not activate if “The light is
switched on with an illuminometer.” option is selected.
1-9
1 NOMENCLATURE AND FUNCTIONS
You can use Backup function of the instrument in order to evade such kind of uneasiness.Your data will
be restored to latest condition* automatically when rebooting by using the Backup function.
The Backup function saves all data files of RAM (except for OS files), registry file and additionally
installed programs into named "Backup" folder in the "Internal Disk".
1-10
1 NOMENCLATURE AND FUNCTIONS
Return to "RAM Backup" screen automatically, when the data back up has been completed.
4 Press the [OK] key on title bar. After that "RAM Backup" screen will close automatically.
● Backing up data may be incomplete if remaining capacity of "Internal Disk" is not enough.
Please make sure the remaining capacity of "Internal Disk" before proceeding to the data
back up.
● Restoration will be impossible if you delete the "Backup" folder in the "Internal Disk".
1-11
1 NOMENCLATURE AND FUNCTIONS
2 Press the [OK] key on title bar. After that, "RAM Backup" screen will close automatically.
2 Press the [OK] key on title bar. After that, "RAM Backup" screen will close automatically.
1-12
1 NOMENCLATURE AND FUNCTIONS
You will lose all of data on the device other than that in the "Internal Disk" after hardware reset
and will need to reinstall the applications and the data you install on your instrument.
Stylus pen 2 Insert the stylus into the unit of hardware reset
switch.
Battery cover
Battery
Cover sensor
● If the battery cover is not completely closed, the GTS-720 will not operate normally,
regardless of whether the battery or the external power source is used.
● If the battery cover is opened while the GTS-720 is in operation, operation will
automatically be suspended.
1-13
1 NOMENCLATURE AND FUNCTIONS
1-14
1 NOMENCLATURE AND FUNCTIONS
1-15
1 NOMENCLATURE AND FUNCTIONS
To operate the touch panel, use either the stylus pen or your finger.
Do not use a ballpoint pen or a pencil.
Light sensor:
Microphone The stylus pen is stored beside the display.
adjust backlight
Star key Star key mode is used for each presetting or displaying.
B.S. Back space key When inputting numbers or characters, return the cursor to the left.
Shift Shift key Used with other keys. Refer to "1.2.5 Shortcut Keys".
Ctrl Control key Used with other keys. Refer to "1.2.5 Shortcut Keys".
Alt Alt key Used with other keys. Refer to "1.2.5 Shortcut Keys".
Func Function key Used with other keys. Refer to "1.2.5 Shortcut Keys".
Cursor Moves the selected item or the cursor laterally and vertically.
1-16
1 NOMENCLATURE AND FUNCTIONS
If the above-mentioned operating procedure is not followed, then, the next time that power
is turned on, it will be necessary to reboot the GTS-720.
1-17
1 NOMENCLATURE AND FUNCTIONS
1-18
1 NOMENCLATURE AND FUNCTIONS
1-19
1 NOMENCLATURE AND FUNCTIONS
1-20
1 NOMENCLATURE AND FUNCTIONS
● Point guide ON/OFF
This feature is most useful when doing stake out work. The Point Guide's red LEDs on the GTS-720
Series telescope assist the rod person in getting on-line. The Point Guide feature is fast and simple
to use.
Instrument
The Point Guide should be used within a distance of
100 meters (328 ft.). The quality of its results will
depend on the weather conditions and the user's
eyesight.
The goal of the rod person is to look at both LEDs
on the instrument and move the prism on-line until
both LEDs become equally bright . If the solid LED
is brighter, move to the right. If the blinking LED is
brighter, move to the left.
Illuminate Blink
Prism
The received return signal level is displayed with bar graph as follows.
● Reticle illumination
Select the brightness by sliding [UP-DOWN] slider.
The brightness setting is stored in memory after power is turned off.
To turn on or off the reticle illumination, press the [reticle illumination] icon.
[OFF] Reticle illumination icon [ON] Reticle illumination icon
1-21
1 NOMENCLATURE AND FUNCTIONS
● Laser Plummet ( Only for Laser Plummet type)
Laser plummet option will help you to center the instrument easily onto the measurement point.
The following symbol mark will indicate that the laser is emitting.
1-22
1 NOMENCLATURE AND FUNCTIONS
1-23
1 NOMENCLATURE AND FUNCTIONS
While on external power, the auto power off function can be enabled too.
To set this function, please check it 'ON' the "Enable suspend while on external power” on
the "Power Off " screen, and select the auto power off time.
(Factory setting is 'OFF' as default)
1-24
2 PREPARATION FOR MEASUREMENT
BT-3L PC-6
PC-6
2-1
2 PREPARATION FOR MEASUREMENT
1. Setting up the Tripod 2 Rotate the instrument 90° (100g) around its
First, extend the extension legs to suitable lengths vertical axis and turn the remaining leveling
and tighten the screws on their midsections. screw or C to center the bubble once more.
2. Attaching the Instrument on the Tripod
Head Leveling screw C
Place the instrument carefully on the tripod head
and slide the instrument by loosening the tripod
screw. If the plumb bob is positioned right over the
center of the point, slightly tighten the tripod 90
screw.
3. Roughly Leveling the Instrument by Using
the Circular Level
1 Turn the leveling screws A and B to move the
bubble in the circular level. The bubble is now
located on a line perpendicular to a line
running through the centers of the two leveling 3 Repeat the procedures 1 and 2 for each 90°
screws being adjusted. (100g) rotation of the instrument and check
Leveling screw C whether the bubble is correctly centered for all
four points.
2-2
2 PREPARATION FOR MEASUREMENT
Main menu
The main menu will be displayed.
● Confirm the battery power remaining on the display. Replace with charged battery or charge when
battery level is low. see section 2.4“Battery Power Remaining Display” .
2-3
2 PREPARATION FOR MEASUREMENT
Measurement is possible.
Note:
1) The battery operating time will vary depending on the environmental conditions such as
ambient temperature, charging time, the number of times of charging and discharging etc. It
is recommended for safety to charge the battery beforehand or to prepare spare full charged
batteries.
2) For general usage of the battery, see Chapter 8 “POWER SOURCE AND CHARGING” .
3) The battery power remaining display shows the power level regarding to the measurement
mode now operating.
The safety condition indicated by the battery power remaining display in the angle
measurement mode does not necessarily assure the battery‘s ability to be used in the
distance measurement mode.
It may happen that the mode change from the angle mode to the distance mode will stop the
operation because of insufficient battery power for the distance mode which consumes more
power than angle mode.
4) When the measurement mode is changed, it rarely may happen that the Battery Power
Remaining Display will decrease or increase two steps momentarily because of the
accuracy of the battery checking system is rough. It is not trouble with the instrument.
2-4
2 PREPARATION FOR MEASUREMENT
Trunnion axis
Horizontal
NT
2-FRO
● GTS-720 compensates both the vertical angle and the horizontal angle readings due to inclination
of the standing axis in the X and Y directions .
● For more information about dual axis compensation, see Chapter “APPENDIX” .
Standing Axis in the X direction Standing Axis in the Y direction Standing Axis in the X and Y
out of range out of range directions out of range
● The display of Vertical or Horizontal angle is unstable when instrument is on an unstable stage or a
windy day. You can turn off the auto tilt correction function of V/H angle in this case.To set TILT
correction mode ON/OFF, refer to section 2.5.1 “Setting Tilt Correction by Soft Key”or Chapter 4
“PARAMETERS SETTING MODE”
2-5
2 PREPARATION FOR MEASUREMENT
● The setting performed here will be interlocked with setting in Chapter 4 “PARAMETERS SETTING
MODE”
2-6
2 PREPARATION FOR MEASUREMENT
● To adjust or reset the above compensation value, see Chapter 5 “CHECK AND ADJUSTMENT” .
● Enable you to stop this function, see Chapter 4 “PARAMETERS SETTING MODE” or Chapter 5
“CHECK AND ADJUSTMENT”
2-7
2 PREPARATION FOR MEASUREMENT
2-8
2 PREPARATION FOR MEASUREMENT
4 Enter Alphabets.
Input 'j',
Press [4](JKL)key. then the sub window
featuring 'j' character will appear on the display
which indicate a entering character.
5 Input 'o',
Press [5](MNO),[5],[5].
The character in the sub window will be altered
'm', 'n', 'o'.
m n
6 Input ‘b’,
o Press [7](ABC), [7].The character in the sub
window will be altered 'a', 'b'.
Then ‘b’ will be appended after ‘jo’.
7 Input ‘_’,
Press [3](YZ_), [3], [3].
The character in the sub window will be altered
'y', 'z', ‘_’.
Then ‘_’ will be appended after ‘job’.
2-9
2 PREPARATION FOR MEASUREMENT
9 Input ‘104’,
Press [1], [0], [4].
Then ‘104’ will be appended after ‘job_’.
2-10
2 PREPARATION FOR MEASUREMENT
● Invoke the software input panel.
2-11
2 PREPARATION FOR MEASUREMENT
Card cover
Card guide Card guide
2-12
2 PREPARATION FOR MEASUREMENT
Using ActiveSync, the GTS-720 can exchange data to a PC via USB cable.
To establish a connection between the GTS-720 and your PC, you first need to install ActiveSync in
your PC.
2 Connect the GTS-720 to your PC with an interface cable F-25 as shown below.
GTS-720
PC
7 Click the [Explore] icon. You will then see the GTS-720 file structure.
2-13
3 STANDARD MEASUREMENT MODE
3-1
3 STANDARD MEASUREMENT MODE
3-2
3 STANDARD MEASUREMENT MODE
For example:70°20'30"
Input 70.2030
*1)With wrong input value(for example 70'), setting will not be completed. Input again from step 3.
3-3
3 STANDARD MEASUREMENT MODE
*1) Every time pressing the [F2] key, the display mode switches.
3-4
3 STANDARD MEASUREMENT MODE
[Example]:
Horizontal distance / Relative elevation mode
*1)ÅA*
*1)The following characters will be shown on the 4th line right hand of the display to represent
measurement mode.
F=Fine; C=Coarse; T=Tracking; R=Continuous (Repeat); S=Single; N=N-times
*2)When EDM is working, the " *" mark appears in the display.
*3)The result is shown with buzzer sound.
*4)Measurement may repeat automatically if the result is affected by shimmer etc..
*5)To change single measuring, press the [F1] key.
*6)To change SD/HD&VD, press the [ ] key.
*7)To return to the angle measurement mode, press the [ANG] key.
3-5
3 STANDARD MEASUREMENT MODE
2)Measuring Method
Confirm the angle measurement mode.
*1)
*1)The following characters will be shown on the 4th line right hand of the display to represent
measurement mode.
R=Continuous (Repeat); S=Single; N=N-times
3-6
3 STANDARD MEASUREMENT MODE
•Coarse mode : This mode measures in shorter time than in fine mode.
Use this mode for the objects which may be slightly unstable.
Measurement time : approx. 0.7seconds
The unit to be displayed is 1mm or 10mm. (0.005ft or 0.02ft)
•Tracking mode : This mode measures in shorter time than in fine mode.
Use this mode for stake out measurement. It is very useful when tailing the moving
object or carrying out stake-out work.
Measurement time : approx. 0.4 seconds
The unit to be displayed is 10mm. (0.02ft)
*1)The following characters will be shown on the 4th line right hand of the display to represent
measurement mode.
F=Fine; C=Coarse; T=Tracking
*2)To cancel the setting, press the [ESC] key.
3-7
3 STANDARD MEASUREMENT MODE
● Stake out operation can be performed for horizontal distance (HD), relative elevation (VD) or slope
distance (SD)
[Example : Horizontal distance]
●To return to normal distance measurement mode, reset the standard distance to “ 0 ”.
3-8
3 STANDARD MEASUREMENT MODE
Reflector (n,e,z)
N
z
n
Inst.Point C
Origin(0,0,0)
E
e
3-9
3 STANDARD MEASUREMENT MODE
3-10
3 STANDARD MEASUREMENT MODE
3-11
3 STANDARD MEASUREMENT MODE
● When setting coordinate values of occupied point, see section 3.3.1“Setting Coordinate Values of
Occupied Point” .
● When setting the instrument height and reflector height, see section 3.3.2“Setting of the Instrument
Height / Reflector(Prism) Height” .
● The coordinates of the unknown point are calculated as shown below and displayed:
Coordinates of occupied point : (N0,E 0,Z0)
Instrument height : INST.HT
Reflector height : R.HT
Vertical distance(Relative elevation) : z
Coordinates of the center of the reflector,
originated from the center point of the instrument : (n,e,z)
Coordinates of unknown point : (N1,E 1,Z1)
N1=N0+n
E1=E0+e
Z1=Z0+INST.HT+z - R.HT
Coordinates of the center of the reflector, originated from the center point of the instrument (n,e,z)
Unknown point
(N 1, E1, Z1)
*1) In case the coordinate of occupied point is not entered, (0,0,0) will be used as the default for the
occupied point.
The instrument height will be calculated as 0 when the instrument height is not entered.
The reflector height will be calculated as 0 when the reflector height is not set.
*2) Refer to section 3.1.3“Measuring from the Required Horizontal Angle” .
3-12
3 STANDARD MEASUREMENT MODE
Mode Output
Angle mode ( V,HR or HL) ( V in percent) V, HR (or HL)
Horizontal distance mode (V,HR, HD, VD) V, HR, HD, VD
Slope distance mode (V, HR,SD) V, HR, SD,HD
Coordinate mode N, E, Z, HR
● The display and the output at the coarse mode are the same as the contents above.
● Output at the tracking mode is displayed as distance data only (HD,VD or SD).
3-13
3 STANDARD MEASUREMENT MODE
3-14
4 PARAMETERS SETTING MODE
V ANGLE Z0/H0 ZENITH/ Select the vertical angle reading for Zenith 0 or Horizontal 0.
HORIZONTAL
After turning on the power, select the same measuring mode
DIST MEAS
REPEAT/N TIME (REPEAT/N TIME) as when distance measurement was
NUMBER
initially carried out.
Select the display format in the coordinate measurement
NEZ / ENZ NEZ / ENZ
mode for NEZ or ENZ.
Select the coefficient correction for refraction and earth
W CORRECTION OFF/0.14/0.20 curvature. Selections for the refraction coefficient are; OFF
(No correction), K=0.14 or K=0.20.
4-1
4 PARAMETERS SETTING MODE
4.1.2 Communication
Factory default settings are indicated with underlines.
Menu Selecting Item Contents
1200 / 2400 / 4800 /
BAUD RATE Select the baud rate.
9600/19200
DATA LENGTH 7 BIT/ 8BIT Select the data length seven digits or eight digits.
PARITY BIT NON/EVEN/ODD Select the parity bit.
STOP BIT 1 BIT/ 2BIT Select the stop bit.
Select the option OFF or ON for carriage return and line feed
CR, LF OFF/ON
when collecting measurement data with a computer.
When communicating to an external device, the protocol for
handshaking can omit the [ACK] coming from the external
ACK MODE OFF/ON device so data is not sent again.
OFF : Omit the [ACK]
ON : Standard
4.1.4 Unit
Menu Selecting Item Contents
TEMP Celsius/Fahrenheit Select the temperature unit for the atmospheric correction.
PRESS hPa/mmHg/inHg Select the air pressure unit for the atmospheric correction .
Select degree(360°), gon(400G) or mil (6400M) for the
ANGLE deg / gon / mil
measuring angle unit to be shown on the display .
Select the distance measuring unit Meter or Feet shown on
DIST m/ft
the display.
Select the meter / feet conversion factor.
US survey feet
US/
FEET 1m=3.280833333333333 ft.
INTERNATIONAL
International feet
1m=3.280839895013123 ft.
4-2
4 PARAMETERS SETTING MODE
*1)To return to the SETUP MODE screen, press the [EXIT] key.
4-3
5 CHECK AND ADJUSTMENT
1) Provide point C on a straight line, connecting straight line AB which is almost level and about 100m
long. Measure straight lines AB, AC and BC.
A B
C
If using above procedure and no difference is found from the instrument constant at the factory or a
difference of over 5mm is found, contact TOPCON or your TOPCON dealer.
5-1
5 CHECK AND ADJUSTMENT
Prism
about 2m
2) Sight through the eyepiece and focus to the prism. Then center the prism on the cross hairs.
Prism
3) Set to the measure mode to distance measurement or set audio.
4) Sight through the eyepiece and focus the (blinking) red light spot by turning the focusing knob in the
direction of infinity (clockwise). If displacement of the reticle cross hairs is within one-fifth of the
diameter of the round red light spot both vertically and horizontally, adjustment will not be required.
If displacement is more than one-fifth in the above case, and still remains so after rechecking
the original line of sight, the instrument must be adjusted by competent technicians. Please
contact TOPCON or your TOPCON dealer to adjust the instrument.
5-2
5 CHECK AND ADJUSTMENT
5-3
5 CHECK AND ADJUSTMENT
Plate level
Leveling screw B
Leveling screw A
● Adjustment
1) Adjust the level adjustment capstan screw, with the accessory adjusting pin and return the bubble
towards the center of the plate level. Correct only one-half of the displacement by this method.
2) Correct the remaining amount of the bubble displacement with the leveling screws.
3) Rotate the instrument 180° or 200g around the vertical axis once more and check bubble
movement. If the bubble is still displaced, then repeat the adjustment.
Level adjustment
capstan screw
Half amount of displacement
Capstan adjustment
screws
5-4
5 CHECK AND ADJUSTMENT
● Adjustment
1) Unscrew the cross-hair adjustment section cover, by revolving it in the counterclockwise direction,
and take it off. This will expose four eyepiece section attachment screws.
Eyepiece section
Eyepiece section attachment screws
attachment screws
Eyepiece
2) Loosen all four attachment screws slightly with the accessory screw-drive (while taking note of the
number of revolutions).
Then revolve the eyepiece section so that the vertical cross-hair coincides to Point A’.
Finally, re-tighten the four screws by the amount that they were loosened.
3) Check once more and if the point travels the entire length of the vertical cross-hair, further
adjustment is not required.
5-5
5 CHECK AND ADJUSTMENT
● Adjustment
1) Unscrew the cross-hair adjustment section cover.
2) Find Point D at a point between Points C and B, which should
be equal to 1/4th the distance between Points B and C and
measured from Point C. This is because the apparent error
between Points B and C is four times the actual error since
the telescope has been reversed twice during the checking
operation.
Capstan Eyepiece
adjustment
screws
5-6
5 CHECK AND ADJUSTMENT
First, loosen the capstan adjustment screw on the side to which the vertical cross-hair line
must be moved. Then tighten the adjustment screw on the opposite side by an equal amount
which will leave the tension of the adjustment screws unchanged.
Revolve in the counterclockwise direction to loosen and in the clockwise direction to tighten,
but revolve as little as possible.
● Adjustment
1) Take off the adjustment section cover of the optical plummet telescope eyepiece. This will expose
four capstan adjustment screws which should be adjusted with the accessory adjusting pin to shift
the center mark to the point. However, correct only one-half of the displacement in this manner.
Capstan
adjustment
screws
1/2 of displacement
Capstan
Eyepiece
adjustment
screws
Plummet telescope
2) Use the leveling screws and coincide the point and center mark.
3) Rotate the instrument 180° or 200g around the vertical axis once more and check the center mark.
If it is coincided to the point, then further adjustment is not required. Otherwise, repeat the
adjustment.
First, loosen the capstan adjustment screw on the side to which the center mark must be
moved. Then tighten the adjustment screw on the opposite side by an equal amount which
will leave the tension of the adjustment screws unchanged.
Revolve in the counterclockwise direction to loosen and in the clockwise direction to
tighten, but revolve as little as possible.
5-7
5 CHECK AND ADJUSTMENT
5-8
5 CHECK AND ADJUSTMENT
3 Input value. * 1)
4 Press the [SET] key.
5-9
5 CHECK AND ADJUSTMENT
5-10
5 CHECK AND ADJUSTMENT
13 Collimate target B.
*1)Pressing the [SKIP] key enables to set next step without changing the last compensated value.
5-11
5 CHECK AND ADJUSTMENT
5-12
5 CHECK AND ADJUSTMENT
5-13
6 SETTING ATMOSPHERIC CORRECTION
6-1
6 SETTING ATMOSPHERIC CORRECTION
3 Press the
[Prism constant value, Atmospheric correction]
icon.
*1)Range :
Temp.
-30.0 °C to +60.0 °C (0.1 °C step) ,
-22.0 °F to +140.0 °F (0.1 °F step)
Press.
560.0 to 1066.0hPa (0.1hPa step) ,
420.0 to 800.0mmHg (0.1mmHg step) ,
16.5 to 31.5inHg (0.1inHg step)
*2)When the atmospheric correction value, which is calculated from the input temperature and pressure
values, exceeds the range ± 999.9ppm, the operating procedure returns to step 4 automatically. Input
values again.
6-2
6 SETTING ATMOSPHERIC CORRECTION
● How to Set the Atmospheric Correction Value Directly
Measure the temperature and air pressure to find atmospheric correction value(PPM) from the chart or
correction formula.
3 Press the
[Prism constant value, Atmospheric correction]
icon.
[PPM] icon
6-3
6 SETTING ATMOSPHERIC CORRECTION
Atmospheric Correction Chart (For your reference)
The atmospheric correction value is obtained easily with the atmospheric correction chart. Find the
measured temperature in horizontal, and pressure in vertical on the chart.
Read the value from the diagonal line, which represents the required atmospheric correction value.
Example:
The measured temperature is +26°C
The measured pressure is 1013 hPa
There fore,
The correction value is +10ppm
6-4
6 SETTING ATMOSPHERIC CORRECTION
6-5
6 SETTING ATMOSPHERIC CORRECTION
6-6
7 CORRECTION FOR REFRACTION AND EARTH CURVATURE
● The conversion formula for horizontal and vertical distances is as follows when correction for
refraction and earth curvature is not applied.
D=L•cosα
Z=L•sinα
The coefficient of the instrument has been set at 0.14 before shipment (K=0.14).
if the "K" value is to be changed, refer to 4 “PARAMETERS SETTING MODE”.
7-1
8 POWER SOURCE AND CHARGING
BT-61Q
Battery cover lever
AC-cable
BC-30
Battery
● To charge
2 Plug the AC-Plug into the outlet. (The POWER LED will light.)
3 Attach the battery in the charger. Charging will start. (The CHARGE LED will light.)
Charging will take approximately 4 hours. (The CHARGE LED will go out.)
● To install
8-1
8 POWER SOURCE AND CHARGING
● Do not charge continuously, otherwise the battery and the charger may be deteriorated.
If charging is necessary, use the charger after stopping charge for approximately 30 minutes.
● Do not charge the battery in right after the battery is charged, it causes deterioration of the battery
in rare cases.
● The charger may develop heat while charging, there is no problem of it.
1 Recharging should take place in a room with an ambient temperature range of 10°C to
40°C (50°F to 104°F).
2 If charging is done at high temperature, charging time of the battery may take longer.
3 Exceeding the specified charging time may shorten the life of the battery and should be
avoided if possible.
4 The battery source will discharge when stored and should be checked before using with
instrument .
5 If you allow the battery to be completely discharged, it will have an effect on the overall
performance for proper charging in the future.
Keep batteries charged at all times..
8-2
9 DETACH/ATTACH OF TRIBRACH
9 DETACH/ATTACH OF TRIBRACH
The instrument is easily detached or attached to the tribrach, with a tribrach locking lever loosened or
tightened for this purpose.
● Detachment
1) Loosen the tribrach locking lever, by revolving it 180° or 200g in the counterclockwise direction
(which will point the triangle mark upwards).
2) Grip the carrying handle firmly with one hand while holding the tribrach with the other. Then lift the
instrument straight upwards and off.
● Attachment
1) Hold the instrument by the carrying handle, with one hand, and carefully lower it on top of the
tribrach while, at the same time, coinciding the alignment piece with the tribrach alignment groove
on the instrument and tribrach respectively.
2) When fully seated, revolve the tribrach locking lever 180° or 200g clockwise ( which will point the
triangle mark downwards again).
Alignment piece
Alignment groove
Securing screw
Tribrach locking lever
9-1
10 SPECIAL ACCESSORIES
10 SPECIAL ACCESSORIES
Diagonal eyepiece, Model 10 Power cord PC-6 (For AC-6 and BT-3L)
Observation in an easy posture will be provided up to the ● L-shape plug provided
zenith position ● Cord length: 2m approx.
10-1
10 SPECIAL ACCESSORIES
Prism sets
See the description on Chapter 12 “PRISM SYSTEM”
Prism unit case, Model 5
1 prisms unit or fixed 3 prisms unit can be stored in
this case. Especially, this is a very easy case to
carry. Soft material is used.
● External dimensions:
200(L)×200(W)×350(H) mm
● Weight:0.5kg
10-2
10 SPECIAL ACCESSORIES
Aluminum extension leg tripod, Type E Wide-frame extension leg tripod, Type E (Wood)
● Flat head 5/8" × 11 threads with adjustable legs. ● Flat head 5/8" × 11 threads with adjustable legs.
10-3
11 BATTERY SYSTEM
11 BATTERY SYSTEM
BT-3L
PC-6
PC-5
BT-61Q PC-6
AC-6
BC-30 Quick
For AC100V~240V use
Approx. 4h
BT-61Q
11-1
12 PRISM SYSTEM
12 PRISM SYSTEM
Arrangement according to your needs is possible.
Target pole-2
(not used with 9 prisms)
Prism-2
Tilting prism Tilting prism Single prism Triple prism Tilting triple
holder-2 holder-3 holder-2 holder-2 prism holder-1 9 prism
with target holder-2
plate-2
Tribrach Tribrach
adaptor-2 adaptor-S2
Pole adaptor- A
Tilt single prism unit Fixed 3 prisms unit Fixed 9 prisms unit
Tripod
Use the above prisms after setting them at the same height as the instruments.To adjust the height of
prism set, change the position of 4 fixing screws.
12-1
13 PRECAUTIONS
13 PRECAUTIONS
1) For transportation, hold by the handle or yoke of the instrument. Never hold by the lens barrel as it
can affect the fixing bracket inside and reduce the accuracy of the instrument.
2) Never expose the instrument without a filter to direct sunlight. It may damage the components
inside the instrument.
3) Never leave the instrument unprotected in high temperature. The temperature inside instrument
may easily reach up to 70°C or above and will reduce the service life.
4) When a high degree of precision is required for measurement, provide shade against direct sunlight
for the instrument and tripod.
5) Any sudden change of temperature to the instrument or prism may result in a reduction of
measuring distance range, i.e. when taking the instrument out from a heated vehicle.
6) When opening the carrying case and taking out the instrument, place the case horizontally, then
open the case.
7) When returning the instrument to its case, be sure to match the white positioning marks provided
with the case and place the instrument.
9) For cleaning the instrument after use, remove dust using a cleaning brush, then wipe off with a
cloth.
10) For cleaning the lens surface, use a cleaning brush to remove the dust, then use a clean lintless
cotton cloth. Moisten it with alcohol ( or mixture with ether ) to wipe gently in a rotational motion
from the center out.
11) Even if any abnormality occurs, never attempt to disassemble or lubricate the instrument yourself.
Always consult with TOPCON or your dealer.
12) To remove the dust on the case, never use thinner or benzine. Use a clean cloth moistened with
neutral detergent.
13) Check each part of the tripod after extended use. Parts (screws or clamps) may work themselves
free.
13-1
14 ERROR DISPLAYS
14 ERROR DISPLAYS
14.1 Message
Description Countermeasures
[Please Input Value!] Input numerical value.
[Please Input Exact Value!] Input exact numerical value.
[Please Input Exact Command!] Input exact command.
[V Angle 0Set Error(Step1)] V angle 0-set is over. (in normal telescope)
[V Angle 0Set Error(Step2)] V angle 0-set is over. (in reverse telescope)
V angle 0-set is over.
[V Angle 0Set Over(Total)]
(in normal and reverse telescope)
Press the [Yes] key : Data is outputted
[DATA OUT?]
Press the [No] key : Cancel
[V Angle Range Over] V angle range is over.
[V Angle Offset Range Over] V angle offset range is over.
[V Angle Tilt Offset Range Over] V angle tilt offset range is over.
[Collimation constant Range Over] Collimation constant range is over.
[Horizontal angle axis constant Range Over] Horizontal angle axis constant range is over.
[The unit of an angle is wrong] The unit of a set angle is wrong.
● If error still persist after attempting to clear them, contact your local Topcon dealer or Topcon head
office.
14.2 Error
Description Contents
[Data Read Error 01~19] The data could not be loaded.
[Data Set Error 01~10] The data could not be set.
[EDM Offset Read Error] The EDM offset could not be loaded.
[EDM Offset Set Error] The EDM offset could not be set.
The external communication could not be
[Ext Communication Retry Error]
accomplished.
[XTILT OVER] X tilt is over. ( ±4’ )
[YTILT OVER] Y tilt is over. ( ±4’ )
Displayed when the instrument rotated too
[H-Angle Error] fast or any abnormality occurs in angle
measuring system .
Displayed when the telescope rotated too
[V-Angle Error] fast or any abnormality occurs in angle
measuring system.
Any abnormality occurs with Tilt sensor.
[Tilt Error]
Repair is required.
Any abnormality occurs with EDM (distance
[E-60’] measuring system).
Repair is required.
Restart and confirm the operation procedure
[E-86]
is correct.
[E-99] Repair is required.
14-1
15 SPECIFICATIONS
15 SPECIFICATIONS
Telescope
Length : 150mm
Objective lens : 45mm (EDM 50mm)
Magnification : 30×
Image : Erect
Field of view : 1°30'
Resolving power : 2.5"
Minimum focus : 1.3m
Reticle illumination : Provided
Distance measurement
Measurement range
Atmospheric conditions
Model Prism
Condition 1 Condition 2
Condition 1: Sight haze with visibility about 20km (12.5miles) moderate sunlight with light heat
shimmer.
Condition 2: No haze with visibility about 40km(25 miles), overcast with no heat shimmer.
Measurement accuracy
: ±(2mm +2ppm × D) m.s.e.
D:Measuring distance (mm)
Least Count in Measurement
Fine measurement mode : 1mm (0.005ft.) / 0.2mm (0.001ft.)
Coarse measurement mode : 1mm (0.005ft.) /10mm (0.02ft.)
Tracking measurement mode : 10mm (0.02ft.)
Measurement Time
Fine measurement mode : 1mm : Approx. 1.2sec. (Initial 4 sec.)
0.2mm : Approx. 2.8 sec. (Initial 5 sec.)
Coarse measurement mode : Approx. 0.7sec. (Initial 3 sec.)
Tracking measurement mode : Approx. 0.4sec. (Initial 3 sec.)
(The initial time will be different by a condition and
setting EDM off time. )
Atmospheric Correction Range : –999.9 ppm to +999.9 ppm, in 0.1 ppm increments
Prism Constant Correction Range : –99.9 mm to +99.9 mm, in 0.1 mm increments
Coefficient Factor : Meter / Feet
International feet 1meter = 3.2808398501 ft.
US SURVEY feet 1meter = 3.2808333333 ft.
15-1
15 SPECIFICATIONS
Electronic Angle Measurement
Method : Absolute reading
Detecting system :
Horizontal : 2 sides
Vertical : 2 sides
Minimum reading
GTS-721 : 1"/0.5" (0.5mgon/0.1mgon, 5mmil/2mmil) reading
GTS-722/723/725 : 5"/1" (1mgon/0.2mgon, 20mmil/ 5mmil) reading
Diameter of circle : 71 mm
Tilt Correction
Type : Automatic vertical and Horizontal index
Method : Liquid type
Compensating Range : ±4'
Correction unit : 1"(0.1mgon)
Computer unit
Microprocessor : SH7750R
Processor speed : 177MHz
Operating system : Microsoft Windows CE.NET 4.2
Memory : 64MB / RAM
64MB Flash disk (Internal)
Display
LCD : 3.5 inch TFT color LCD
(240x320 dots )
Touch panel : Electrical analog resistor thin film system
Others
Instrument height : 182mm (7.2in) Base unit detachable
(Height from the tribrach dish to the center of
telescope)
Level sensitivity
Circular level : 10'/2mm
Plate level : 30"/2 mm
15-2
15 SPECIFICATIONS
Dimension
(GTS-721, 722, 723) : 343 (H) × 245 (W)× 219 (L) mm
(13.5 (H) × 9.6 (W) × 8.6 (L) in)
Weight
Instrument (GTS-721, 722, 723) : 6.2 kg (13.6 lbs) (With battery)
Durability
Protection against water and dust : IP54 (Based on the standard IEC60529)
Ambient Temperature Range : –20°C to +50°C (–4°F to +122°F)
● Battery using time will vary depending on environmental conditions and operations done with
GTS-720 series.
15-3
APPENDIX
APPENDIX
1 Dual Axis Compensation
Inclination of the vertical axis with respect to true vertical will result in incorrectly measured horizontal
angles. The extent of the error in horizontal angle measurement due to axis tilt depends on three
factors :
Example: When the vertical axis is tilted by 30 arcseconds, the target is 10° above the horizon and
rotated 90 in azimuth from the direction of the vertical axis error.
From the above example it can be seen that horizontal angle errors will increase with steeper vertical
sights (tangent will increase as vertical angle increases) and will be at a maximum when the target is at
right angles (sin 90°=1) to the direction of the vertical axis error. Errors will be at a minimum when the
sights are nearly horizontal (h=0, tan0=0) and in the same direction as the vertical axis error ( α =0,
sin0=0). Please refer to the table below to see the relationship between axis tilt (v) and elevation (h) and
the error in horizontal angles which results from these factors.
APPENDIX-1
APPENDIX
It is clear from the table that dual axis compensation has the most benefit when the elevation
of the target is greater then 30° and the axis is tilted more than 10". The entries indicated in
bold in the table show, in fact, that for many common surveying applications i.e. target elevation
<30° and axis error <10", virtually no correction would be required. Dual axis compensation is
especially suited then for applications where the sights are very steep.
Even though the compensators can correct horizontal angles for vertical axis errors,
it is still important to use care in setting up the instrument.
Centering error, for instance, cannot be corrected by the compensators. If the vertical axis is
tilted by 1’ with the instrument 1.4meters above the ground, a centering error of approx. 0.4mm
will result. The maximum effect of this error at 10m is about 8" of horizontal angle error.
In order to maintain the increased accuracy possible through dual axis compensation, it is
necessary to keep the compensators in proper adjustment.The compensators must agree with
the actual level condition of the instrument. Through various environmental stresses, the
agreement between the level condition sensed by the compensators and the true level condition
of the instrument may be disturbed. In order to reestablish the correct relationship between the
compensator and the true level condition of the instrument, it is necessary to carry out the
vertical indexing procedure listed on Section 5.3.6“Adjustment of Vertical Angle 0 Datum” . This
adjustment will both reset the vertical index (cause a direct + indirect zenith reading to the same
elevation to equal 360°) and zero the level reference for the horizontal compensator. While
correct vertical angles can be obtained by averaging direct and indirect reading even when the
index is improperly adjusted, the same is not true for horizontal angles. Since the vertical axis
error is fixed for a given setup, its effect cannot be removed by averaging two readings.
For this reason, it is extremely important to maintain the vertical indexing adjustment to
insure proper correction of the horizontal angles.
APPENDIX-2
APPENDIX
APPENDIX-3
ELECTRONIC TOTAL STATION
GTS-720 series
TOPCON CORPORATION