Product Data

Endos ACP
Intra-oral X-ray unit

Tubehead Features
High voltage circuit type Single phase, self-rectifying, grid controlled
Rated output voltage 70 kVp ± 6% @ 230V
70 kVp ± 6% @ 120V
Tubehead current 8 mA
Tubehead power 430 VA
Total filtration 2 mm Al eq. @ 70 kV
Maximum deviation of output cur- ± 13%
rent
Pre-heating time 100 ms
Interval between exposures 1:32
(duty cycle)
Minimum focus to skin distance 20 cm
30 cm with optional collimator cone
X-ray field (at collimator tip) Circular, diameter 6 cm
Rectangular 35x45 mm with optional diaphragm
Leakage radiation @ 1m ≤ 0.1 mGy/h
Technical factors for leakage radia- 70 kV, 8 mA, 1 s, 1:32 duty cycle
tion
Tube insert type OCX/70-G
Focal spot 0.8 (IEC 336)
Anode material Tungsten
Anode angle 19°
Inherent filtration 0.5 mm Al eq. @ 70 kV
Anode thermal capacity 6 kJ

Scissors and Extension Arm Features


Material Aluminum
Type of painting Glossy
Mounting type - Wall mount with local timer
- Wall mount with remote timer
- Wall mount with remote X-ray pushbutton
- Ceiling mount
- Mobile unit on trolley
Choice of extension arm length - 30 cm
(only for wall mounted units) - 60 cm
- 80 cm

Product data Endos ACP – Rev.5 Page 1 of 5


Timer Features
Exposure times From 0.02 to 3.2s in 33 steps
Exposure times control Microprocessor controlled exposure times, with
automatic compensation of line voltage fluctuations
Timer accuracy ±10% or ±20 ms (whatever is greater) @ 230V
±10% or ±32.2 ms (whatever is greater) @ 120V
Exposure times visualization Digital display
Exposure button With extensible cable (3m)
Stand-by function The system enters a low-consumption state after 5
minutes of inactivity
Number of anatomic programs 30 pre-set times
Choice of patient size 3 choices (small, medium, large)
Choice of anatomic region - Incisors/canines (upper an lower)
- Premolars (upper an lower)
- Upper molars
- Lower molars
- Bite wing
Automatic dose reduction for Yes, with pushbutton on keyboard
videoradiography systems
Film speed setting Yes
CCD/ phosphor-plates setting Yes
Visualization of number of expo- Yes
sures
External connections Provisions for connection of:
- Remote x-ray pushbutton
- Unit powered signal lamp
- X-ray active signal lamp

Electrical Characteristics
Line voltage 230V ± 10% 120V ± 10%
Line frequency 50 Hz 60 Hz
Absorbed current 4A rms @ 230V 8A rms @ 120V
Absorbed power 920 VA @ 230V 1.1 kVA @ 120V
Line resistance ≤ 0.8 Ohm N/A
Max line voltage regulation N/A 3% @ 132V

Environmental Conditions
Operating conditions Temperature: +10° to +40° Celsius
Humidity: 30% to 75%
Conditions for transport and Temperature: -20° to +70° Celsius
storage Humidity: up to 95 % non condensing
Pressure: up to 630 hPa

Product data Endos ACP – Rev.5 Page 2 of 5


Standard and Regulations
This product is designed and produced to meet the following standards:
EN 60601-1, EN 60601-1-1, EN 60601-1-2, EN 60601-1-3, EN 60601-1-4, EN 60601-2-7,
EN 60601-2-28, EN 60601-2-32, EN 1441
CE symbol grants the product compliance to the European
Directive for Medical Devices 93/42/EEC and its revised
0051 versions as a class IIB device
c-ETL-us approval means that the product meets the
requirements of the applicable US and Canadian standards
(Only for 120V version)

FDA approval grants the product compliance to US Code of


Federal Regulations title 21 subchapter j
(Only for 120V version)
Health Canada Licence grants the product compliance to the
Canadian Medical Device Regulations SOR/98-282
(Only for 120V version)

Keyboard Functions

1. Ready for exposure signal lamp 10. Premolars (upper and lower) signal lamp
2. Manual decrease of exposure times 11. Incisors/canines (upper and lower)
3. Manual increase of exposure times signal lamp
4. X-ray emission warning lamp 12. Anatomic selection
5. Video radiography signal lamp 13. Large patient size signal lamp
6. Video radiography selection 14. Medium patient size signal lamp
7. Bite wing signal lamp 15. Small patient size signal lamp
8. Upper molars signal lamp 16. Patient size selection
9. Lower molars signal lamp 17. Exposure enable button

Product data Endos ACP – Rev.5 Page 3 of 5


Site Planning (quotes in mm and inches)
WALL MOUNT
120 (4-3/4")
FRONTAL VIEW

(14-1/2")
370

30 60
0
30
60
(89")
2262
290░ 200
130
(5-1/8") (7-7/8")
300 600 800

(57" SUGGESTED HEIGHT )


1450 ALTEZZA SUGGERITA
(11-3/4" 23-5/8" 31-1/2") 657 957 1157
785 1085 1285 (26") (37-1/2") (45")
(49-1/4")
1250

(31" 42-3/4" 50-5/8")

(42-1/2")
1080

(38-5/8")
1430 1730 1930 270

980
(56-5/16" 68-1/8" 76") (10-5/8")

CEILING MOUNT

120 (4-3/4")
70░

290░
300 (11-3/4")
1530 (60-1/4")

730 (28-3/4")
1430 (56-1/4")

1250 (49-1/4")

1370 (53-3/4") 270 (10-5/8")

Product data Endos ACP – Rev.5 Page 4 of 5


MOBILE UNIT ON TROLLEY
120 (4-3/4")

70░
2110 (83")

290░
430 (16-7/8")
1370 (54")

1070 (54") 270 (10-5/8")

1090 (43")
63 (2-1/2")

60
30
0
30

80 (3-1/8")

LARG. 20 (3/4")

30 0 30 60
360░
948 (37-5/16")

540 (21-1/4")

60░

30
0
30

130 (5 1/8")
60

743 (29 1/4")

Note: Products are continuously under review in the light of technical advancement. The actual specification may therefore
be subject to improvement or modification without notice.

Company with Quality System certified by


VILLA SISTEMI MEDICALI s.p.a.
20090 BUCCINASCO (MI) – ITALY,
Via delle Azalee, 3
Tel. +39-02-48859.1, Fax +39-02-4881844

ISO 9001:2008 ISO 13485: 2003