Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Mag5100 Mcerts

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

PRODUCT CONFORMITY CERTIFICATE

This is to certify that the

SITRANS F M MAGFLO MAG 5100W Electromagnetic


Flowmeter with MAG 5000, MAG 6000 or MAG 6000 Industry
Transmitter
manufactured by:

Siemens Flow Instruments Ltd


Magflow House Nordborgvej 81
Ebley Road 6430 Nordborg
Stonehouse Denmark
Glos.
GL10 2LU

has been assessed by Sira Certification Service


and for the conditions stated on this certificate complies with:

MCERTS Performance Standards for Water Monitoring


Equipment Part 3, Version 2.1, dated March 2008

Size range: DN 25-1200

Project No. : 674/0190


Certificate No : Sira MC080136/03
Initial Certification : 4 November 2008
This Certificate issued : 7 February 2014 R Cooper I Eng MInst MC
Renewal Date : 3 November 2018 Technical Director

MCERTS is operated on behalf of the Environment Agency by


Sira Certification Service
12 Acorn Industrial Park, Crayford Road, Crayford
Dartford, Kent, UK DA1 4AL
Tel: +44 (0)1322 520500 Fax: +44 (0)1322 520501

This certificate may only be reproduced in its entirety and without change
Registered Office: Rake Lane, Eccleston, Chester, UK CH4 9JN
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts

Page 1 of 5
Approved Site Application
The product may be used on all MCERTS applications including abstraction, effluent
discharge,
ultraviolet disinfection and industrial processing.

Any potential user should ensure, in consultation with the manufacturer, that the product is
suitable for the process on which it will be installed.

Field test Site


A three month field test was conducted on the final effluent discharge at a municipal waste
water treatment plant.

Basis of Certification

This certification is based on the following Test Report(s) and on Sira’s assessment and
ongoing surveillance of the product and the manufacturing process:

Sira Evaluation Report MAG 5100 674/0190 dated 04 November 2008

Product Certified
The measuring system consists of the DN (mm) Flow Rate unit
Min Max
following parts: 25 442.0 17671 l/h
40 1.2 45 m3/h
SITRANS F M MAGFLO MAG 5100W 50 1.6 63 m3/h
65 2.5 100 m3/h
Electromagnetic Flowmeter with MAG 5000, 80 4.0 160 m3/h
MAG 6000 or MAG 6000 Industry Transmitter 100 6.3 250 m3/h
125 10.0 400 m3/h
150 15.7 629 m3/h
This certificate applies to all instruments fitted 200 24.9 997 m3/h
with software version: 250 40.0 1600 m3/h
3.03 X 03 for standard MAG 5000 300 62.5 2500 m3/h
3.03 X 04 for MAG 5000 C with HART 350 86.6 3463 m3/h
400 113.1 4523 m3/h
3.03 X 05 for MAG 5000 CT 450 143.2 5725 m3/h
3.03 for standard MAG 600 500 176.8 7068 m3/h
3.03 X 02 for MAG 6000 CT 600 254.5 10178 m3/h
700 346.4 13854 m3/h
3.03 X 01 for MAG 6000 SV 750 397.7 15904 m3/h
3.04 for MAG 6000 Industry 800 452.4 18095 m3/h
900 573.0 22902 m3/h
100 707.0 28274 m3/h
Serial number (MLFB code) 7ME6520- 1100 855.3 34211 m3/h
XXX1X-2XXX-Z [Where X = any figure] 1200 1018.0 40715 m3/h
onwards

Certificate No: Sira M080136/03


This Certificate Issued: 7 February 2014
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts

Page 2 of 5
Certified Performance
The instrument was evaluated for use under the following conditions:
Ambient Temperature Range: -20ºC to +50ºC

The instrument meets MCERTS Class 1 requirements for the combined performance characteristic as specified in
Table 6 of the MCERTS performance standard. Details of individual performance characteristics are summarised
below:

Results are expressed as error % of certification range, unless otherwise stated

Test Results expressed as error % Other results MCERTS


of reading specification
<0.5 <1.0 <1.5 <2.0
Protection against unauthorised access Access to change mode is password protected Clause 3.1.2

Indicating device The flowmeter incorporates an indicating device, Clause 3.1.3


analogue and digital output signal

Units of measurement Various units of measurement are available. Clause 3.1.6

Bi-directional flow The sign (-) will stand in front of the flow reading when Clause 3.1.8
the reading is negative.

2% Class 1
Combined performance characteristic 0.806 Table 6

Clause 6.3.2
Mean error 0.15 ±1.5% Class 1

Clause 6.3.2
Repeatability 0.07 1% Class 1

Clause 6.3.3
Supply voltage 0.05 0.5% Class 1

Clause 6.3.4
Output impedance 0.15 0.5% Class 1

Clause 6.3.5
Fluid Temperature 0.13 0.5% Class 1

Clause 6.3.6
Ambient air temperature 0.60 0.5% Class 1

Clause 6.3.6
Relative humidity 0.04 0.5% Class 1

Clause 6.3.9
Stray currents 0.23 0.5% Class 1

Bi-directional flow Mean error


Mean error 0.18 ±1.5% Class 1

Repeatability 0.038 Repeatability


1% Class 1

Certificate No: Sira M080136/03


This Certificate Issued: 7 February 2014
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts

Page 3 of 5
Test Results expressed as error % Other results MCERTS
of reading specification
<0.5 <1.0 <1.5 <2.0
Clause 6.3.1
Loss of Power for electronic flowmeters No changes in pre set data to be reported

See Note 1 Clause 6.3.19


Response time 30 seconds

Field Test Results

The field test was conducted on two MAG 5100 sensors with MAG 6000 transmitters in series and is deemed
equivalent by the certification committee for the models stated on this certificate

Test Results expressed as error % Other results MCERTS


of reading specification
<0.5 <1.0 <1.5 <2.0
Error range 0.29% to 0.61% Clause 7.3
Error under field test conditions Field test error is <2% for 100% of readings 2% Class 1
5% Class 2
Clause 7.4
Up time 100% >95%

Clause 7.5
Maintenance none to be reported

Note 1: This test has not been conducted.

Note 2: The following tests are not applicable to the flowmeter:


6.3.7 Incident light 6.3.16 Effect of conduit material
6.3.8 Sensor location 6.3.17 Effect of conduit size
6.3.10 Sonic velocity compensation & response 6.3.18 Fill level
6.3.11 Accuracy of computation 6.3.20 Vibration
6.3.12 User defined stage-discharge equation

Certificate No: Sira M080136/03


This Certificate Issued: 7 February 2014
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts

Page 4 of 5
Description

Sitrans FM electromagnetic flow meters included in this certificate consist of a sensor type MAG 5100
W, in sizes from DN25 to DN1200; and a transmitter, type MAG 5000, MAG 6000 or MAG 6000
Industry. The plug-in transmitters can be integral to the sensor or remote mounted. MAG 5100W
sensors are designed specifically to meet water and waste water applications. IP68 versions can be
buried or submerged.

The measuring principle is based on Faraday’s law of electromagnetic induction. An electrode voltage,
proportional to velocity, is generated when a conductive liquid passes through the sensor’s magnetic
field.

Calibration data, sensor fingerprint, factory and customer settings are stored in a SENSORPROM
module, separate from the transmitter. Transmitters can, therefore, be freely exchanged. This
technology is proven in use; fitted in Sitrans FM meters for more than 10 years.

Transmitters use low noise high resolution digital signal processors which provide continuous self-
monitoring and adjustment of measurement circuits to maintain required accuracy. Plug-in modules for
digital communications, e.g. Profibus, can be added at any time during the life of the meter. Transmitter
dynamic range is better than 3000:1. Very high input impedance means measurement accuracy is
unaffected by liquid conductivity or cable length.

On site verification is achieved using the Siemens Magflo Verificator; a stand alone field test device,
independently calibrated every 12 months. It performs three tests, all referenced to original calibration:
Transmitter accuracy, Insulation of measurement circuits, and Sensor magnetism (fingerprint).

Approvals include the new EU directive for cold water custody transfer, MI 001, WRAS for potable
water, and OIML R49 pattern approval. Every Siemens flow meter is calibrated at facilities that are
individually accredited in accordance with ISO / IEC 17025 by UKAS, DANAK and traceable to NIST.

General Notes

1. This certificate is based upon the equipment tested. The Manufacturer is responsible for
ensuring that on-going production complies with the standard(s) and performance criteria defined
in this Certificate. The Manufacturer is required to maintain an approved quality management
system controlling the manufacture of the certified product. Both the product and the quality
management system shall be subject to regular surveillance according to ‘Regulations Applicable
to the Holders of Sira Certificates’. The design of the product certified is defined in the Sira
Design Schedule for certificate No. Sira MC 080136/02

2. If certified product is found not to comply, Sira Certification Service should be notified
immediately at the address shown on this certificate.

3. The Certification Marks that can be applied to the product or used in publicity material are
defined in ‘Regulations Applicable to the Holders of Sira Certificates’.

4. This document remains the property of Sira and shall be returned when requested by the
company.

Certificate No: Sira M080136/03


This Certificate Issued: 7 February 2014
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts

Page 5 of 5

You might also like