Calibration

Certificate
Certificate Number: B08-1020966 Page 1 (2)
Customer: Sample Inc
123 Sample Rd.
Sample, MA 01234
Manufacturer: Vaisala Oyj
Instrument: VL-2000-20R Humidity and Temperature Logger
Serial Number: 10062253

The internal temperature sensor of the logger was calibrated by comparing the logger readings to a reference thermometer in Vaisala Boston
Calibration and Repair Services (CRS).
The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k = 2,
which for a normal distribution corresponds to a coverage probability of approximately 95 %.
The measurement results are traceable to the international system of units (SI) through national metrology institutes (NIST USA, MIKES Finland,
or equivalent) or accredited calibration laboratories. Pass: Error less than Limit minus Uncertainty; Pass*: Error less than or equal to Limit; Fail:
Error more than Limit.
Measurement results before adjustment:
Reference Channel 1
Temperature Observed Error Uncertainty Limit Pass/Fail

°C °C °C °C °C
-25.03 -25.06 -0.03 0.08 ±0.25 Pass
10.01 9.99 -0.02 0.06 ±0.25 Pass
25.01 24.99 -0.02 0.06 ±0.15 Pass
30.03 30.03 0.00 0.06 ±0.25 Pass
40.04 40.03 -0.01 0.06 ±0.25 Pass
45.03 45.02 -0.01 0.07 ±0.25 Pass
69.89 69.92 0.03 0.08 ±0.25 Pass

Measurement results after adjustment:


Reference Channel 1
Temperature Observed Error Uncertainty Limit Pass/Fail

°C °C °C °C °C
-25.03 -25.03 0.00 0.08 ±0.15 Pass
10.01 10.01 0.00 0.06 ±0.15 Pass
25.01 25.00 -0.01 0.06 ±0.1 Pass
30.03 30.04 0.01 0.06 ±0.15 Pass
40.04 40.04 0.00 0.06 ±0.15 Pass
45.03 45.03 0.00 0.07 ±0.15 Pass
69.89 69.89 0.00 0.08 ±0.15 Pass

Reference(s): Instrument Number Calibration Date Certificate Next Calibration


Fluke 1560/2560/5606 1011-0342-1 2016-05-04 K008-Z01358 2017-05-31

Note(s):
Service report as an attachment

Calibration Date: February 22, 2017 Next Calibration: February 22, 2018

Date: March 09, 2017 Ambient Condition(s):


Digitally signed by JOY
Date: 2017.03.09 10:22:14 -05:00 22.5 °C ±1.3 °C
Reason: Calibration responsible
Signature: Location: Vaisala Inc 40.4 %RH ±8.8 %RH
Joey Yang 1007.8 hPa ±3.9 hPa
Service Technician

This Certificate may only be reproduced in full, except with the prior written permission by the issuing Laboratory.
This instrument was calibrated in compliance with ISO/IEC 17025:2005.
Vaisala USA 10-D Gill Street Woburn, MA 01801, USA | Phone: 800-408-9456 | Fax: +1 781 933 8029 |
Email helpdesk@vaisala.com | www.vaisala.com
Calibration
Certificate
Certificate Number: B08-1020966 Page 2 (2)
The humidity sensor of the logger was calibrated by comparing the logger readings to the generated reference humidity readings in Vaisala
Boston Calibration and Repair Services (CRS).
The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k = 2,
which for a normal distribution corresponds to a coverage probability of approximately 95 %.
The measurement results are traceable to the international system of units (SI) through national metrology institutes (NIST USA, MIKES Finland,
or equivalent) or accredited calibration laboratories. Pass: Error less than Limit minus Uncertainty; Pass*: Error less than or equal to Limit; Fail:
Error more than Limit.
Measurement results before adjustment:
Reference Channel 2
Humidity Temperature Observed Error Uncertainty Limit Pass/Fail

%RH °C %RH %RH %RH %RH


44.98 10.01 43.80 -1.18 0.80 ±3 Pass
10.99 24.99 11.13 0.14 0.70 ±2 Pass
45.03 25.01 45.38 0.35 0.80 ±2 Pass
59.82 25.02 60.03 0.21 0.80 ±2 Pass
79.79 24.98 79.59 -0.20 0.80 ±2 Pass
65.06 30.03 64.85 -0.21 0.80 ±3 Pass
74.94 40.04 74.57 -0.37 0.90 ±3 Pass
44.94 45.03 44.75 -0.19 0.80 ±3 Pass

Measurement results after adjustment:


Reference Channel 2
Humidity Temperature Observed Error Uncertainty Limit Pass/Fail

%RH °C %RH %RH %RH %RH


44.98 10.01 45.02 0.04 0.80 ±2 Pass
10.99 24.99 11.05 0.06 0.70 ±1 Pass
45.03 25.01 45.13 0.10 0.80 ±1 Pass
59.82 25.02 59.86 0.04 0.80 ±1 Pass
79.79 24.98 79.88 0.09 0.80 ±1 Pass
65.06 30.03 65.06 0.00 0.80 ±2 Pass
74.94 40.04 75.01 0.07 0.90 ±2 Pass
44.94 45.03 44.97 0.03 0.80 ±2 Pass

Reference(s): Instrument Number Calibration Date Certificate Next Calibration


Thunder 2500 5011-0088 2016-12-16 161216-TS2500- 2017-12-31
1007799
Fluke 1560/2560/5606 1011-0342-1 2016-05-04 K008-Z01358 2017-05-31
PTU303 3011-0400 2016-11-09 161109-PTU303- 2017-05-31
L1330477
Note(s):
Service report as an attachment

Calibration Date: February 22, 2017 Next Calibration: February 22, 2018

Ambient Condition(s):
22.5 °C ±1.3 °C
40.4 %RH ±8.8 %RH
1007.8 hPa ±3.9 hPa

This Certificate may only be reproduced in full, except with the prior written permission by the issuing Laboratory.
Vaisala USA 10-D Gill Street Woburn, MA 01801, USA | Phone: 800-408-9456 | Fax: +1 781 933 8029 |
Email helpdesk@vaisala.com | www.vaisala.com