Ocean Insight USB2000+ Operating Instructions
Ocean Insight USB2000+ Operating Instructions
Ocean Insight USB2000+ Operating Instructions
Regional Headquarters
Maybachstrasse 11
73760 Ostfildern
Phone 49-711 34 16 96-0
Fax 49-711 34 16 96-85
E-Mail Sales@Mikropack.de
270-00000-000-02-201107 i
Table of Contents
Appendix B: Specifications...............................................................17
Overview ....................................................................................................................... 17
How the USB2000+ Works ............................................................................................ 17
USB2000+ Components Table.................................................................................................. 18
USB2000+ Specifications .............................................................................................. 19
Accessory Connector Pinout.......................................................................................... 20
Accessory Connector Pinout Diagram ...................................................................................... 20
Accessory Connector – Pin Definitions and Descriptions ......................................................... 20
Index ...................................................................................................23
ii 270-00000-000-02-201107
About This Manual
Document Purpose and Intended Audience
This document provides the users of USB2000+ Spectrometers with instructions for setting up,
calibrating and performing experiments with their spectrometer.
Document Summary
Chapter Description
Appendix B: Specifications Contains technical specifications and connector pinouts for the
USB2000+ Spectrometer.
Product-Related Documentation
You can access documentation for Ocean Optics products by visiting our website at
http://www.oceanoptics.com. Select Technical , then choose the appropriate
document from the available drop-down lists. Or, use the Search by Model Number field at the bottom
of the web page.
Detailed instructions for SpectraSuite Spectrometer Operating Software are located at:
http://www.oceanoptics.com/technical/SpectraSuite.pdf.
Detailed instructions for the Breakout Box are located at:
http://www.oceanoptics.com/technical/HR4_breakout.pdf.
270-00000-000-02-201107 iii
About This Manual
Detailed instructions for External Triggering for firmware versions below 3.0 are located at:
http://www.oceanoptics.com/technical/external-triggering.pdf.
Detailed instructions for External Triggering for firmware versions 3.0 and above are located at:
http://www.oceanoptics.com/technical/external-triggering2.pdf.
Upgrades
Occasionally, you may find that you need Ocean Optics to make a change or an upgrade to your system.
To facilitate these changes, you must first contact Customer Support and obtain a Return Merchandise
Authorization (RMA) number. Please contact Ocean Optics for specific instructions when returning a
product.
Service
Our Total Technical Service protection plans protect your spectrometer for one year for any contingency -
- no matter what the circumstances, including total loss. Under these plans, Ocean Optics will repair or
replace your instrument with no deductible or any other out-of-pocket expense required from you. USB
service plans include the following:
TTS-USB: Protects your Ocean Optics USB spectrometer for any contingency for 1 year.
Provides for full spectrometer replacement, if necessary. Limited to 1 full replacement per year.
TTS-USB-E: This special plan for Ocean Optics USB spectrometers not previously covered
protects for any contingency for 1 year. Provides for full spectrometer replacement, if necessary.
Limited to 1 full replacement per year. This coverage starts 30 days after purchasing the plan.
In addition, we also have the WAVECAL-USB spectrometer maintenance and calibration service
available to recalibrate, clean and perform any adjustments necessary on your USB2000+ Spectrometer.
Contact us to learn more about these great service plans.
iv 270-00000-000-02-201107
Chapter 1
Introduction
Product Overview
The USB2000+ Miniature Fiber Optic Spectrometer is a unique combination of technologies a powerful
2-MHz analog-to-digital (A/D) converter, programmable electronics, a 2048-element CCD-array detector,
and a high-speed USB 2.0 port. This innovative combination produces our fastest spectrometer yet and
provides resolution to 0.35 nm (FWHM). The USB2000+ allows users to capture and store a full
spectrum into memory every millisecond (that's 1,000 full spectra every second) when the spectrometer is
interfaced to a computer via a USB 2.0 port. The USB2000+ is perfect for chemical, biochemical and
other applications where fast reactions need to be monitored.
The USB2000+ Spectrometer connects to a computer via the USB port or serial port. When connected
through a USB 2.0 or 1.1, the spectrometer draws power from the host computer, eliminating the need for
an external power supply. The USB2000+, like all USB devices, can be controlled by our SpectraSuite
software, a completely modular, Java-based spectroscopy software platform that operates on Windows,
Macintosh and Linux operating systems.
270-00000-000-02-201107 1
1: Introduction
Features
Sony ILX511 2048-element linear silicon CCD array detector
Responsive from 200 to 1100 nm
Sensitivity of up to 75 photons/count at 400 nm; 41 photons/count at 600 nm
An optical resolution of ~10.0 nm (FWHM)
Integration times from 1 ms to >60 seconds
EEPROM storage for
Wavelength Calibration Coefficients
Linearity Correction Coefficients
Absolute Irradiance Calibration (optional)
Low power consumption of only 250 mA @ 5 VDC
16 bit, 3MHz A/D Converter
4 triggering modes
22-pin connector for interfacing to external products
CE Certification
RoHS compliant
System Requirements
You can use the USB2000+’s USB connectivity with any computer that meets the following
requirements:
Microsoft Windows – Windows 2000, XP; 32-bit and 64-bit
Apple Macintosh – OS X version 10.0 or later
Linux – Red Hat 9 or later, Fedora (any version), Debian 3.1 or later (Sarge), SUSE (9.0 or later),
Centos (any version), and Ubuntu
EEPROM Utilization
An EEPROM memory chip in each USB2000+ contains wavelength calibration coefficients, linearity
coefficients, and a serial number unique to each individual spectrometer. The SpectraSuite software
application reads these values directly from the spectrometer, enabling the ability to “hot-swap”
spectrometers between computers without entering the spectrometer coefficients manually on each
computer.
About SpectraSuite
SpectraSuite is the latest generation of operating software for all Ocean Optics spectrometers. It is a
completely modular, Java-based spectroscopy software platform that operates on Windows, Macintosh
and Linux operating systems. The software can control any Ocean Optics USB spectrometer and device,
as well as any other manufacturer’s USB instrumentation using the appropriate drivers.
2 270-00000-000-02-201107
1: Introduction
SpectraSuite is a user-customizable, advanced acquisition and display program that provides a real-time
interface to a variety of signal-processing functions. With SpectraSuite, you have the ability to perform
spectroscopic measurements (such as absorbance, reflectance, and emission), control all system
parameters, collect and display data in real time, and perform reference monitoring and time acquisition
experiments. Consult the SpectraSuite manual for hardware requirements when using SpectraSuite (see
Product-Related Documentation).
Interface Options
The USB2000+ has both USB and serial port connectors (with the use of an adapter), enabling you to
connect the spectrometer to a computer via a USB or serial port. However, you must create custom
software if using the serial port. SpectraSuite software is available if you are connecting via the USB port.
Shipment Components
The following information and documentation ships with the USB2000+ Spectrometer:
Packing List
The packing list is inside a plastic bag attached to the outside of the shipment box (the invoice
arrives separately). It lists all items in the order, including customized components in the
spectrometer (such as the grating, detector collection lens, and slit). The packing list also includes
the shipping and billing addresses, as well as any items on back order.
USB Cable (USB-CBL-1)
Use this cable to connect your spectrometer to a computer running on a Windows, Mac or Linux
operating system.
270-00000-000- 02-201107 3
1: Introduction
Note
Please save the Wavelength Calibration Data Sheet for future reference.
4 270-00000-000-02-201107
1: Introduction
270-00000-000- 02-201107 5
Chapter 2
USB2000+ Installation
This section contains instructions for connecting the USB2000+ via both USB and serial modes.
USB Mode
Note
The USB port on a computer can power up to five USB2000+ spectrometer channels.
Systems with more than five channels require a powered USB hub.
Procedure
Follow the steps below to connect the USB2000+ to a computer via the USB port:
1. Install SpectraSuite on the destination computer.
2. Locate the USB cable (USB-CBL-1) provided with the USB2000+.
3. Insert the square end of the cable into the side of the USB2000+.
4. Insert the rectangular end of the cable into the USB port of the computer.
If you installed SpectraSuite prior to connecting the USB2000+, the SpectraSuite installs the USB2000+
drivers. If the drivers do not successfully install (or if you connected the USB2000+ to the computer
before installing SpectraSuite), consult Chapter 3: Troubleshooting.
270-00000-000-02-201107 6
2: Installing the USB2000+
If you have followed the previous steps and started SpectraSuite, the spectrometer is already acquiring
data. Even with no light in the spectrometer, there should be a dynamic trace displayed in the bottom of
the graph. If you allow light into the spectrometer, the graph trace should rise with increasing light
intensity. This means the software and hardware are correctly installed.
Note the spectrometer(s) that you have installed are listed in the Data Sources pane.
Once you install the software and hardware, and establish your sampling system, you are ready to take
measurements.
Procedure
Follow the steps below to connect the USB2000+ to the computer via serial port:
1. Connect the serial cable adapter block to the appropriate pins of the USB2000+’s 30-Pin
Accessory Connector.
2. Connect one end of the 9-pin serial cable to the adapter block on the USB2000+, and then
connect the other end to a serial port on the computer.
3. Note the number of the serial port (COM Port) to which you connected the USB2000+ (some
computers may not have numbered ports; handheld computers typically have only one serial
port).
4. Plug the 5 VDC external power supply into an outlet and connect it to the USB2000+.
USB2000+ Configuration
The USB2000+ can be used with either SpectraSuite or OOIBase32 software when connected to the USB
port. The configuration process differs, depending on whether you are running SpectraSuite or
OOIBase32 software.
270-00000-000- 02-201107 7
2: Installing the USB2000+
the graph. If you allow light into the spectrometer, the graph trace should rise with increasing light
intensity. This means the software and hardware are correctly installed.
Note the spectrometer(s) that you have installed are listed in the Data Sources pane.
8 270-00000-000-02-201107
Chapter 3
Troubleshooting
Overview
The following sections contain information on troubleshooting issues you may encounter when using the
USB2000+ Spectrometer.
Note
If these procedures do not correct your device driver problem, you must obtain the
Correcting Device Driver Issues document from the Ocean Optics website:
http://www.oceanoptics.com/technical/engineering/correctingdevicedriverissues.pdf.
270-00000-000-02-201107 9
3: Troubleshooting
Note
Improperly installed USB devices can also appear under the Universal Serial Bus
Controller option. Be sure to check this location if you cannot locate the unknown device.
3. Locate the unknown device (marked with a large question mark). Right-click on the Unknown
Device listing and select the Uninstall or Remove option.
4. Click the OK button to continue. A warning box appears confirming the removal of the Unknown
Device. Click the OK button to confirm the device removal.
5. Disconnect the USB2000+ from your computer.
6. Locate the section in this chapter that is appropriate to your operating system and perform the
steps in the following Remove Improperly Installed Files section.
Note
If the INF directory is not visible, you must disable the Hide System Files and Folders
and Hide File Extensions for Known File Types options in Windows Folder Options.
Access Windows Folder Options from Windows Explorer, under the Tools | Folder
Options menu selection.
3. Delete the OOI_USB.INF in the INF directory. If your computer is running either the Windows
2000 or XP operating system, you must also delete the OOI_USB.PNF file in the INF directory.
4. Navigate to the Windows | System32 | Drivers directory.
5. Delete the EZUSB.SYS file.
6. Reinstall your Ocean Optics application and reboot the system when prompted.
7. Plug in the USB device.
The system is now able to locate and install the correct drivers for the USB device.
10 270-00000-000-02-201107
3: Troubleshooting
270-00000-000- 02-201107 11
3: Troubleshooting
12 270-00000-000-02-201107
Appendix A
p = I + C1 p + C2 p2 + C3 p3
Where:
= the wavelength of pixel p
I = the wavelength of pixel 0
C1 = the first coefficient (nm/pixel)
C2 = the second coefficient (nm/pixel2)
C3 = the third coefficient (nm/pixel3)
You will be calculating the value for I and the three Cs.
270-00000-000-02-201107 13
A: Calibrating the Wavelength of the USB2000+
Note
Ocean Optics’ HG-1 Mercury-Argon lamp is ideal for recalibration. If you do not have an
HG-1, you need a light source that produces several (at least 4-6) spectral lines in the
wavelength region of your spectrometer.
A USB2000+ spectrometer
An optical fiber (for spectrometers without a built-in slit, a 50- m fiber works best)
A spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-
order linear regressions
Note
If you are using Microsoft Excel, choose Tools | Add-Ins and check AnalysisToolPak
and AnalysisTookPak-VBA.
14 270-00000-000-02-201107
A: Calibrating the Wavelength of the USB2000+
Predicted
True Wavelength (nm) Pixel # Pixel # 2 Pixel # 3 Difference
Wavelength
253.65 175 30625 5359375 253.56 0.09
296.73 296 87616 25934336 296.72 0.01
302.15 312 97344 30371328 302.40 -0.25
313.16 342 116964 40001688 313.02 0.13
334.15 402 161604 64964808 334.19 -0.05
365.02 490 240100 117649000 365.05 -0.04
404.66 604 364816 220348864 404.67 -0.01
407.78 613 375769 230346397 407.78 0.00
435.84 694 481636 334255384 435.65 0.19
546.07 1022 1044484 1067462648 546.13 -0.06
576.96 1116 1245456 1389928896 577.05 -0.09
579.07 1122 1258884 1412467848 579.01 0.06
696.54 1491 2223081 3314613771 696.70 -0.15
706.72 1523 2319529 3532642667 706.62 0.10
727.29 1590 2528100 4019679000 727.24 0.06
738.40 1627 2647129 4306878883 738.53 -0.13
751.47 1669 2785561 4649101309 751.27 0.19
5. Use the spreadsheet or calculator to calculate the wavelength calibration coefficients. In the
spreadsheet program, find the functions to perform linear regressions.
If using Quattro Pro, look under Tools | Advanced Math
If using Excel, look under Analysis ToolPak
6. Select the true wavelength as the dependent variable (Y). Select the pixel number, pixel number
squared, and the pixel number cubed as the independent variables (X). After executing the
regression, you will obtain an output similar to the one shown below. Numbers of importance are
noted.
Regression Statistics
Multiple R 0.999999831
R Square 0.999999663 R Squared
Adjusted R Square 0.999999607
Standard Error 0.125540214
Observations 22
Intercept
Third coefficient
270-00000-000- 02-201107 15
A: Calibrating the Wavelength of the USB2000+
7. Record the Intercept, as well as the First, Second, and Third Coefficients. Additionally, look at
the value for R squared. It should be very close to 1. If not, you have most likely assigned one of
your wavelengths incorrectly.
Keep these values at hand.
Procedure
To save wavelength calibration coefficients using the USB mode, perform the following steps:
1. Ensure that the USB2000+ is connected to the computer and that you have closed all other
applications.
2. Point your browser to http://www.oceanoptics.com/technical/softwaredownloads.asp and
scroll down to Microcode. Select USB EEPROM Programmer.
3. Save the setup file to your computer.
4. Run the Setup.exe file to install the software. The Welcome screen appears.
5. Click the Next button. The Destination Location screen appears.
6. Accept the default installation location, or click the Browse button to specify a directory. Then,
click the Next button. The Program Manager Group screen appears.
7. Click the Next button. The Start Installation screen appears.
8. Click the Next button to begin the installation. Once the installation finishes, the Installation
Complete screen appears.
9. Click the Finish button and reboot the computer when prompted.
10. Navigate to the USB EEPROM Programmer from the Start menu and run the software.
11. Click on the desired USB2000+ device displayed in the left pane of the USB Programmer
screen.
12. Double-click on each of the calibration coefficients displayed in the right pane of the USB
Programmer screen and enter the new values acquired in Steps 5 and 6 of the Calibrating the
Wavelength of the Spectrometer section in this appendix.
13. Repeat Step 12 for all of the new values.
14. Click on the Save All Values button to save the information, and then Exit the USB Programmer
software.
The new wavelength calibration coefficients are now loaded onto the EEPROM memory chip on the
USB2000+.
16 270-00000-000-02-201107
Appendix B
Specifications
Overview
This appendix contains information on spectrometer operation, specifications, and system compatibility.
It also includes accessory connector pinout diagrams and pin-specific information.
See USB2000+ Components Table on the following page for an explanation of the function of each
numbered component in the USB2000+ Spectrometer in this diagram.
270-00000-000-02-201107 17
B: Specifications
SMA 905 Secures the input fiber to the spectrometer. Light from the input fiber enters the
1
Connector optical bench through this connector.
A SAG+, Ag-coated mirror focuses light entering the optical bench towards the
Collimating Grating of the spectrometer.
4
Mirror Light enters the spectrometer, passes through the SMA Connector, Slit, and Filter,
and then reflects off the Collimating Mirror onto the Grating.
A #3 (600 lines per millimeter, blazed at 500 nm) grating diffracts light from the
5 Grating
Collimating Mirror and directs the diffracted light onto the Focusing Mirror.
Focusing A SAG+, Ag-coated mirror receives light reflected from the Grating and focuses
6
Mirror first-order spectra onto the detector plane.
Collects the light received from the Focusing Mirror or L2 Detector Collection Lens
Detector and converts the optical signal to a digital signal. Each pixel on the Detector
8
responds to the wavelength of light that strikes it, creating a digital response. The
spectrometer then transmits the digital signal to the SpectraSuite application.
Optional Linear Variable Filters (LVF) construct systems with excellent separation of
excitation and fluorescence energy. LVF-L Linear low-pass filters fine tune the
9 LVF Filters
excitation source for maximum signal with minimum overlap. LVF-H high-pass filters
are available for the detection side. These filters are optional.
18 270-00000-000-02-201107
B: Specifications
USB2000+ Specifications
The following table provides specification information for the USB2000+.
Specification Value
Spectrometer USB2000+
Entrance aperture 5, 10, 25, 50, 100 or 200 µm wide slits or fiber (no slit)
AgPlus Mirrors Provide reflectivity >95% over the visible and NIR wavelength range and
over a wide range of angles of incidence
Data transfer rate Full scans (2048 wavelengths) into memory every 2 milliseconds
Time acquisition approximately every 25 milliseconds
Weight 190 g
270-00000-000- 02-201107 19
B: Specifications
A1 SPI_CLK Output The SPI Clock signal for communications to other SPI peripherals
1 Input power pin for USB2000+ – When operating via USB, this pin
VCC, VUSB, or Input or
can power other peripherals – Ensure that peripherals comply with
20 270-00000-000-02-201107
B: Specifications
TTL signal driven Active HIGH when the Lamp Enable command
4 Lamp Enable Output
is sent to the spectrometer
Continuous TTL output signal used to pulse a strobe – Divided down from the
5 Output
Strobe master clock signal
The SPI Master Out Slave In (MOSI) signal for communications to other
11 MOSI Output
SPI peripherals
The SPI Master In Slave Out (MISO) signal for communications to other
12 MISO Input SPI peripherals
270-00000-000- 02-201107 21
B: Specifications
22 270-00000-000-02-201107
Index
A F
accessories, 4, 7 features, 2
accessory connector filter, 18
diagram, 20 focusing mirror, 18
pin definitions, 20
pinout, 20
additional equipment, 4
G
Adobe Acrobat Reader, 4
grating, 18
C I
Calibrating, iii, 13
calibration, 13 Installation, 6
preparing for, 14 Serial Port mode, 7
procedure, 14 USB mode, 6
calibration coefficients installed filter, 18
saving in USB mode, 16 Interface, 3
CCD Detector, 18
collimating mirror, 18
Components Table, 18
L
configuration, 7
L2 Detector Collection Lens, 18
configuring, 8
Lens, 18
D M
Detector Collection Lens, 18
memory chip, 2
document
mirror, 18
audience, iii
O
purpose, iii
summary, iii
E OOIBase32, 8
Options
Interface, 3
EEPROM, 2
P
External Triggering, 8
packing list, 3
passwords, 4
product-related documentation, iii
270-00000-000-02-201107 23
Index
S U
Sampling upgrades, iv
System, 3 USB mode, 6
Serial Port mode, 7
service, iv
setup, 6
W
shipment components, 3
slit, 18 Wavelength Calibration
SMA Connector, 18 about, 13
Software and Resources Library CD, 4 Wavelength Calibration Data File, 4
specifications, 17, 19 Wavelength Calibration Data Sheet, 4
SpectraSuite, 2 what's new, iii
spectroscopic accessories, 7
System Requirements, 2
T
Triggering, 8
troubleshooting
Linux systems, 11
Mac systems, 10
Windows systems, 9
Troubleshooting, 9
24 270-00000-000-02-201107