LECO 628Series-CHN PDF
LECO 628Series-CHN PDF
LECO 628Series-CHN PDF
Carbon/Hydrogen/Nitrogen
Determinator
Instruction Manual
Version 1.1x
Part Number 200-731
September 2012
Patents are Pending in U.S.A. and other countries
© 2012 LECO Corporation
Minor revisions may not be reflected in this manual.
1 Introduction
The Introduction chapter contains general information about the
CHN628 Series Determinators. This includes safety guidelines and
warranty terms. Reference this chapter for replacement parts,
operating supplies and optional accessories. To place an order by
phone, call our customer service department toll-free in the United
States at 1-800-292-6141 or 269-985-5496. Orders may also be sent
by fax to 269-982-8977.
1 Introduction
Illustrations .................................................................................. 1–7
WEEE......................................................................................... 1–11
Warranty.................................................................................... 1–16
LECO-Supplied Dell Computers ..................................................... 1–18
Interference to Other Devices ....................................................... 1–20
Declaration of Conformity ............................................................. 1–21
Software License ......................................................................... 1–22
Equipment Packages .................................................................... 1–26
Options ...................................................................................... 1–27
Components and Accessories List .................................................. 1–30
Specifications ............................................................................. 1–32
2 Installation
Illustrations .................................................................................. 2–2
Instrument Installation .................................................................. 2–3
Computer Installation .................................................................. 2–11
Data Transmit Installation ............................................................ 2–12
Apply Power ............................................................................... 2–13
3 Option Installation
Illustrations .................................................................................. 3–2
Balance Installation ....................................................................... 3–3
Stackable Carousels ...................................................................... 3–7
5 Operation
Illustrations .................................................................................. 5–4
First Time Analysis ........................................................................ 5–5
Shutdown Procedure .................................................................... 5–12
Sample Preparation ..................................................................... 5–13
Login ......................................................................................... 5–17
Correct for Moisture ..................................................................... 5–20
Correct for Atmospheric Blank ...................................................... 5–21
Enter a Mass from the Balance ...................................................... 5–22
Analyze a Sample ........................................................................ 5–23
Samples ..................................................................................... 5–39
Printing Reports .......................................................................... 5–45
Text Export Data ......................................................................... 5–49
Text Import Data ........................................................................ 5–59
Transmit Data ............................................................................. 5–60
Transmit Selected Samples .......................................................... 5–68
Data Backup ............................................................................... 5–69
Transfer Import .......................................................................... 5–71
Error Messages ........................................................................... 5–72
7 Theory of Operation
Illustrations .................................................................................. 7–2
Theory of Operation CHN628 .......................................................... 7–3
Method of Standard Additions ......................................................... 7–6
Calculations .................................................................................. 7–8
Barometric Pressure ...................................................................... 7–9
Thermal Conductivity Cell ............................................................. 7–10
Infrared Radiation Cell ................................................................. 7–12
Theory of Calibration ................................................................... 7–13
Comparator Level ........................................................................ 7–20
9 Service
Illustrations .................................................................................. 9–3
Installing LECO Software ................................................................ 9–4
Configuring the Ethernet Adapter .................................................. 9–13
Configuring Firewall Settings ........................................................ 9–18
Registering Software ................................................................... 9–19
Troubleshooting Copy Protection Key ............................................. 9–25
Loading Head Alignment and Adjustment ....................................... 9–26
Loading Head Interface Block Cleaning .......................................... 9–31
Carousel Alignment ..................................................................... 9–33
TC Cell Removal and Service ........................................................ 9–35
IR Cell Removal and Service ......................................................... 9–36
IR Detector Linearization .............................................................. 9–37
IR Cell Saturation Optimization ..................................................... 9–39
Troubleshooting .......................................................................... 9–43
10 Illustrations
11 Schematics
12 Index
Dell Support
All LECO supplied Dell PC's include Dell's ProSupport Service, which is
designed to provide a base level of customer security. This service
option can place a Dell-certified technician at your location the
following day, if necessary (after phone-based troubleshooting). In
many cases, remote troubleshooting can minimize overall system
downtime by getting the issue quickly fixed remotely or helping Dell to
dispatch the right part (if necessary).
The inclusion of Dell ProSupport with LECO supplied PC’s ensures that
any computer related service issues are handled directly by the
experts at Dell. This eliminates the unnecessary step of working
through the LECO service professionals for a resolution from Dell, and
provides PC support 24 hours a day, seven days a week.
Dell ProSupport provides expert assistance when you need rapid
problem resolution to help maximize your PC system's uptime and
productivity. ProSupport provides the following benefits:
• 24 x 7 Access to advanced-level North American-based phone
technicians
• Live Chat tech support
• DellConnect remote diagnosis and resolution
• Next Business Day On-Site Service after Remote Diagnostics
For more information, call 1-800-WWW-DELL (1-800-999-3355) or
access Dell's website by copying and pasting the following link into
your internet browser:
http://www.dell.com/content/topics/global.aspx/services
/prosupport/end_users?c=us&cs=555&l=en&s=biz
Please have your Service Tag and Express Service Code numbers
available when contacting Dell to expedite service. These can be
found on the back or side of the tower.
This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC Rules and
Regulations. These limits are designed to provide reasonable
protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the
instructions, may cause interference to radio communications.
However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined
by turning the equipment off and on, the user is encouraged to try to
correct the interference by one or more of the following methods:
• Reorient or relocate the radio or television receiving
antenna.
• Increase the separation between the equipment and the
radio or television receiver.
• Connect the equipment to an outlet on a circuit different
from that which the radio or television receiver is currently
connected.
• Consult the radio or television dealer or an experienced
radio/TV technician for help.
I. Definitions
"Licensed Software" shall mean the software supplied with this
Software License, which is software developed and distributed by
LECO.
"Instruction Manual" shall mean a guide describing the use and
operation of the Licensed Software, together with any related
supporting documentation.
"Licensee" shall mean and include any of the Licensee's divisions and
departments within Licensee's organization, but shall not include
wholly or partially owned subsidiaries, affiliates, successors or assigns,
or independent third parties.
VII. Confidentiality
All information and materials regarding the software, provided to
Licensee by LECO which are not generally known in the trade are
confidential, are proprietary, and such information and materials are
furnished solely for the use of Licensee under this Software License.
Licensee agrees not to use or disclose any such information or material
otherwise without prior written consent from LECO and Licensee
further agrees to take all reasonable steps requested by LECO, or
otherwise necessary to protect the confidentiality of all such
information and materials. The foregoing obligations will survive
termination of this Software License, and Licensee agrees to return all
such information and materials to LECO upon request or any non-
renewal or termination of this Software License.
Items listed below are repeated throughout this manual and are
subject to revision. Please consult the packing slip received with the
instrument.
*Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping regulations. These items are essential for operation.
Contact your LECO distributor for quotation and delivery.
*Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping regulations. These items are essential for operation.
Contact your LECO distributor for quotation and delivery.
*Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping regulations. These items are essential for operation.
Contact your LECO distributor for quotation and delivery.
Precision Range
Carbon: ......................................0.01 mg or 0.5% RSD, whichever is greater
Hydrogen: ...................................0.05 mg or 1.0% RSD, whichever is greater
Nitrogen: ....................................0.02 mg or 0.5% RSD, whichever is greater
Sample Mass
Up to 250 mg, 100 mg nominal
Readability
0.0001%
Analysis Time
4.5 minutes
Detection Method
Carbon/Hydrogen: ........................Non-Dispersive Infrared (IR) Absorption
Nitrogen: ....................................Thermal Conductivity (TC Cell) Detector
Gases Required
Carrier: .......................................Helium, 99.99% pure @ 35 psi (2.4 bar) ±10%
Combustion .................................Oxygen, 99.99% pure @ 35 psi (2.4 bar) ±10%
Pneumatic: ..................................Compressed Air @ 40 psi (2.8 bar) ±10%, source
must be oil and water free
Furnace
Resistance furnace; both primary and afterburner; up to 1050°C
Autoloader
30-position (stackable to 120 samples)
Environmental Conditions
Operating Temp: ..........................15°C to 35°C (59°F to 95°F)
Humidity: ....................................20% to 80%, non-condensing
Weight (approximate)
Instrument ..................................273 lb. (124 kg)
Shipping Weight ...........................324 lb. (147 kg)
**Allow a 6-inch (15 cm) minimum access area around all units.
The following procedure should be used for first time installation of the
instrument. This procedure assumes that the instrument has been
unpacked and positioned in its permanent location.
The Balance and Printer are not supplied with the instrument. They
must be purchased separately. Refer to Options, page 1–27.
Refer to the Facility Requirements diagram, page 2–6, during
installation of the instrument.
1. Provide a work surface with convenient electrical power and gas
supplies. Refer to the Facility Requirements diagram, page 2–6
for power and gas supply requirements.
2. Install the Instrument as shown in the Facility Requirements
diagram, page 2–8. Provide a minimum space of six inches
around the instrument for airflow.
International Installation
For international installation, a HAR approved line cord for use in the
end country must be obtained. This product requires a disconnect
device that may either be the plug of the power cord or a disconnect
device installed at the installation site if the equipment is permanently
wired.
When the plug on the power supply cord is intended to serve as the
disconnect device, the socket/outlet must be installed near the
equipment and shall be easily accessible. When the wiring is
permanently connected, a disconnect device (such as a circuit breaker
or switch) must be lockable, installed near the equipment, and be
easily accessible.
Combustion Tube
The combustion tube supplied in instruments purchased for domestic
use is packed before shipment from the factory. Instruments
purchased for international use may not have the combustion tube
packed. If the combustion tube is not packed, refer to Combustion
Tube Packing, page 6–29 and pack the combustion tube.
LANCE
0.25”
CRUCIBLE
Figure 2-2
Lance Installation
CRUCIBLE
LOADING HEAD
INTERFACE BLOCK
Figure 2-3
Crucible Installation
Removal
1. Slide reagent tube upward until the bottom end can swing free.
2. Tilt out the free reagent tube end.
3. Pull the reagent tube downward off the top port.
Packing
1. Insert a screen filter into the bottom of reagent tube. Position the
o-ring towards the bottom and push the screen filter up into the
tube 1/4-inch, Figure 2-4.
2. Pack 3/4-inch of glass wool into the bottom of reagent tube.
3. Fill the bottom 1/3 of reagent tube with Anhydrone.
4. Fill the top 2/3 of reagent tube with Lecosorb.
5. Firmly pack 1/2 to 2/3-inch of glass wool into the top of reagent
tube.
6. Reinstall the reagent tube.
7. Clean excessive reagent from work area.
Installation
1. Push the top end of the reagent tube over the top port.
2. Tilt the reagent tube inward and pull it down, pushing it over the
bottom port.
619-268
ALIQUOT DOSE
REAGENT TUBE 502-174
LECOSORB
501-171
ANHYDRONE
617-548
INCOMING 763-265
HELIUM SCRUBBER GLASS WOOL
REAGENT TUBE
775-306
SCREEN FILTER
Figure 2-4
Reagent Tube Packing
Balance Setup
1. Unpack the Balance and assemble it as shown in the manual
supplied with the balance.
2. Plug the AC adapter, supplied with the balance, into facility
power. Connect the AC adapter cable to the power-input jack on
the rear of the balance.
3. Adjust the balance feet to level the balance. The level bubble
should be inside the circle when the balance is level. The level
bubble is located on the rear of the balance.
4. Connect one end of the balance interface cable to the serial
connector on the computer (COM 1). Connect the other end of
the cable to the balance connector on the rear of the balance.
5. To properly interface the balance with the analyzer, reset the
balance to the factory defaults. Refer to Resetting the Balance on
the next page.
6. After the balance is reset, do not change any of the configuration
codes. The balance codes can be changed to custom configure the
balance, if desired, after this procedure is completed and the
balance is operating with the determinator.
Figure 3-2
Carousel Open Position
Figure 3-3
Carousel Bottom Plate
Figure 3-4
Carousel Mounting Position
Figure 3-5
Stacked Carousels
10. To stack more carousels, repeat steps 7 through 9. For the third
carousel hole 28 should be open and for the fourth carousel hole
27 should be open. As the samples drop into the first carousel
they will be in the correct numerical order.
11. Enter the sample information into the spreadsheet. It will be
necessary to ignore the sample position numbers on the carousels
and number the samples in the spreadsheet from 1 to the highest
sample number in consecutive order.
If operation is intermittent with the weight of 4 carousels, perform
the carousel adjustment procedure on page 9–33.
Figure 4-1
Front Panel Controls and Functions
The following are the menu choices in the Windows® software. Clicking
on any of these will open another set of menu choices.
Database Menu
Edit Menu
Maintenance Menu
1. Click Font to change the font name, style, size, and color. The
Font configuration screen will appear.
2. From the Font configuration screen select the Font Name, Style
and Size. A sample of the selected Font will appear in the Sample
Text preview box.
2. Use the dialog box on the right to select the information shown in
the dashboard. Selections are:
Control Chart with Statistics—A chart with plotted sample
values in addition to the average value, standard deviation,
relative standard deviation and the number of samples
selected.
Control Chart—A chart with plotted sample values.
Statistics—The average value, standard deviation, relative
standard deviation and the number of samples selected.
3. Overlaid Plots—Plots of each selected sample in the same
dashboard.
4. Click the drop down arrow in the Stats to Display selection box
and select the desired statistics shown in the dashboard.
5. Click the Label Stats Points check box to label the sample points
on the chart. The numbers that appear with the point are the
sample numbers selected.
6. Click inside the Significant Digits selection box to select the
statistics number format.
Zoom—Position the mouse cursor over the area of a plot to zoom in.
Hold the left mouse button in and draw a box. The area inside the box
will be magnified. Draw another box to zoom in farther. Click Zoom
Out One Level to reduce the magnification by one level. Click Zoom
Out to zoom out all levels and return to the original plot.
Label Points—Label points will place the sample number near the point
on a control chart. Select multiple samples. Right click the desired
dashboard and select Label Points.
Statistics—Statistics will display a control chart or multiple plots.
Select multiple samples. Select statistics to display a control chart.
Deselect statistics to display multiple plots.
Auto Independent Scale—When multiple selected plots are scaled
differently, Auto Independent Scale rescales the plots so they can be
compared together. Select Auto Independent Scale to rescale plots.
Properties—Displays the Dashboard Properties dialog box for advanced
configuration or creation of a new dashboard. Refer to Create a
Dashboard on page 4–23.
Position a Dashboard
Move the mouse pointer over the dashboard tab and hold the left
mouse button down. Move the mouse to position the dashboard on the
main screen.
To size the dashboard, move the mouse over the splitter bars and hold
the left mouse button down. Move the mouse to move the splitter
bars.
Refer to Main Screen Overview on page 4–20.
3. Click Save and enter a unique name for the current screen layout.
4. Click Close to exit and save the screen layout.
From the View menu, click Toolbar. A check indicates the toolbar
will be displayed.
From the View menu, click Status Bar. A check indicates the
status bar will be displayed.
Log Bar
The log bar, located at the bottom of the screen, displays information
saved in the log file. It only appears when it displays information. This
information may be communication failures, severe errors or
operational procedures.
If Log Bar is not checked it will only display communications failures
and severe errors.
If Log Bar is checked it will display any information saved in the log
file. This includes communications failures, severe errors and
operational procedures.
Examples of operational procedures are sample data changes,
calibration changes, change to the sample mass and more.
Auto Widths
1. Select the Configuration menu
2. Click Auto Widths to adjust the size of each cell in the sample
display to fit the text inside the cell.
Test Commands
Send—Used to test the balance interface. Click Send to send a weight
from the balance to the system and display the weight in the
spreadsheet.
Tare—Used to test the balance interface. Click Tare to set the weight
displayed on the balance to zero.
2. Click inside the barometric Pressure value box, located to the left
of the Set Pressure button, and enter the current barometric
pressure.
Use values in mm Hg for barometric pressure.
Create a Method
1. From the Configuration menu, click Method. The Method dialog
box will appear.
NONE = 0 LPM
LOW = 0.3 LPM
MED
1 LPM MED = 1 LPM
HIGH = 4 LPM
MAX = ALL FLOWS ON
BURN 1 BURN 2
Figure 4-2
Burn Profile
H. Click the Macro Ballast tab to display the micro ballast
properties dialog box. For Macro Ballast definitions, refer to
the Glossary, page 12–1.
I. For Equilibrate Time, enter the time period for gases in the
ballast to mix after it is filled.
J. Enter the desired Macro Ballast property values.
K. Click OK when finished For Analysis and Element Parameter
definitions, refer to the Glossary, page 12–1.
Using this procedure you can also edit the method list.
1. From the Configuration menu, click Method. The Method screen
will appear.
2. Select the Method to be edited.
Import a Method
Import a Method transfers an exported method to an instrument and
adds it to the method list. A method can be imported that was
developed on the present instrument or another CHN628 instrument.
4. Click Open. The method will be added to the analysis method list.
2. Click the drop down arrow, in the sample name selection box, and
select a defined Standard from the list.
3. Enter the information required in the Login Screen.
4. Click OK to login the standard.
1. From the Samples menu, click Login. The Sample Login dialog
box will appear.
2. Click the drop down arrow to the right of sample name and enter
the desired sample name from the drop down list. The sample
name, method comments, and description will automatically be
entered into the spreadsheet.
3. Continue with sample login and enter the other necessary
information.
Refer to Sample Login for additional information.
Sample filters permits the operator to sort and display sample data
based on a predefined requirement. Fields can be selected to display a
certain name or value. Fields can also be mathematically compared
and displayed according to a requirement.
There are three predefined fields:
2. Select a sample filter and click OK. The dialog box will disappear
and the sample grid will reappear with the filter applied.
3. Click Cancel to exit without applying the filter.
2. Click Add to define and add a sample filter to the list. A new filter
with the filter name New Filter will be added to the list.
3. To define the new filter select New Filter and click Properties. The
Properties Dialog Box will appear.
4. From the Properties Dialog Box you can define the filter.
5. Click the Filter Name to enter and change the filter name. This is
the name that will appear in the Sample Filter Selection dialog
box.
Format Examples
*Format Type Number Format for Result Result
A 4.50360 A 4.50360000
0.123456 F4.5 0.1235
0.000789 F3.5 0.00079
Fn.m
0.000789 F3.2 0.00
321.8765 F3.5 322
0.123456 S4.5 1.235E-1
0.000789 S3.5 7.9E-4
Sn.m
0.000789 S3.2 0.0E-1
321.8765 S3.5 3.22E+2
0.123456 E4.5 123.5E-3
0.000789 E3.5 790E-6
En.m
0.000789 E3.2 000E-3
321.8765 E3.5 322E+0
In 321.8765 I2 322
* A = Default format
n = Total number of significant digits to be represented
m = Maximum number of digits after the decimal to define readability
I = Integer
S = Scientific notation: An exponential format (y.yyy x 10e) in which any number
is expressed as a number between 1 and 10 multiplied by a power of 10 that
indicates the correct position of the decimal in the original number.
E = Engineering format: An exponential format that displays the exponent in
multiples of 3.
F = Fixed point format: A format in which the decimal point is located at a single
unchanging position in a predetermined number of digits.
Formula Examples
* Function Type Example Function Result
[Column Heading of Analyte concentration
[Carbon %]
Desired Analyte Value] value
@ABS(exp) @ABS([Carbon %]-[Carbon Blank]) Absolute value
@Log10(exp) @Log10([Carbon ppm]) Log base 10
@Ln(exp) @Ln([Carbon %]) Natural log
@SQRT(exp) @SQRT([Carbon %]) Square root
@Maximum([Carbon %],[Carbon
@Maximum(exp1,exp2) Finds Maximum
Blank])
@Minimum([Carbon %],[Carbon
@Minimum(exp1,exp2) Finds Minimum
Blank])
If the Carbon ppm value
is > 5, the Carbon ppm
@IF(cond, true, false) @IF([Carbon ppm]>5,[Carbon ppm],0)
value will display,
otherwise 0 will display
If both Carbon and
@IF(@And([Carbon %]>0.01,[Nitrogen Nitrogen % are > 0.01,
@AND(cond1, cond2)
%]>0.01), 1, 0) 1 will display, otherwise
0 will display
If either Carbon or
@IF(@Or([Carbon %]>0.01,[Nitrogen Nitrogen % are > 0.01,
@OR(cond1, cond2)
%]>0.01), 1, 0) 1 will display, otherwise
0 will display
If JUST Carbon % or
JUST Nitrogen % is >
@IF(@XOr([Carbon %]>0.01,[Nitrogen
@XOR(cond1, cond2) 0.01 (not both), 1 will
%]>0.01), 1, 0)
display, otherwise 0 will
display
Create Backup
1. In the Start dialog box, highlight the database connection of the
database to back up. The file location of the corresponding
database will appear as shown below.
Validating a Database
During database validation, the database files will be reviewed for
corruption. If orphan objects are found, refer to Reviewing Orphan
Objects, page 4–68.
To validate a database,
1. Exit the instrument software.
2. Click Start on the Windows desktop, select Programs, select
LECO, select the instrument, and then click Database
Management. The Start dialog box will appear.
Database Menu
Items on the Database menu in the instrument software will only
appear if logging was enabled when the database connection was
created. Refer to Enabling Logging for an Existing Database,
page 4–69.
2. Select the time and date for the presentation time as follows:
Select the Live check box to view current information or clear the
Live checkbox in order to select a different date and time.
When the Live check box is cleared, Date and Time are active.
View History
When logging is enabled for a database connection, View History
displays a log of database changes that could be used to identify the
time/date when some particular change took place. It can only be used
to identify who made a change because there is no description of what
each change represents.
1. Click the Database menu in the instrument software and select
View History. The View Log dialog box will appear.
2. The log items that appear in the View Log dialog box are
explained below.
Date/Time displays the date and time that the database change
occurred.
Entry ID displays a unique identifying number for the database
object that was changed. For example, each sample would have a
unique number.
Action displays Created, Modified or Deleted to indicate the type
of action that occurred.
User displays the username of the person who performed the
action.
3. Select the time zone to use for displaying Date/Time. Local
displays the local time zone. UT displays the universal time that
corresponds with the UTC (Universal Coordinated Time) as
defined by the National Bureau of Standards.
2. Click the drop-down arrow and enter the prior sample deletion
date.
3. If the prior samples to be deleted should be exported and saved
before they’re deleted, check the Transfer Export Samples to be
deleted check box.
4. Click Delete to delete the prior analyzed samples or Cancel to
cancel deletion.
The first time the instrument is operated, and any time the software
is updated, these steps must be performed in the order in which they
appear.
1. Perform a System check before operation to determine if the
instrument is operating properly.
A. Let the instrument warm up and stabilize.
B. From the Diagnostics menu, click System Check.
Use this procedure to safely power down and turn off the instrument.
If Cancel is clicked during this procedure shutdown will be aborted and
the furnace temperatures restored to their original settings.
During shutdown the gas is turned off.
INSTRUMENT DAMAGE
Failure to cool the furnaces prior to turning the power
switch OFF will result in overheating of the loading head
assembly and possible damage to the furnace.
1. Click Maintenance and select Instrument Shutdown. The
Instrument Shutdown screen will appear.
Each sample holder type will require its own blank value and should be
used in the main calibration.
Samples should be homogenized prior to encapsulation.
Sample presses may be used to maximize sample size.
Figure 5-1
Sample Preparation—Gel Capsule Method
For analysis of multiple samples, install and use the 619-180 Carousel
Assembly then load the samples in the proper order for analysis.
502-210
TUBE
(500/PK)
502-186
SMALL TIN FOIL CUP
(100/PK)
502-397
LARGE TIN FOIL CUP
(100/PK)
603-493
FOIL HOLDER
Figure 5-2
Sample Preparation—Quick Sampler Method
502-186
Small Tin Foil Cup
(100/pk)
502-397
Medium Tin Foil Cup
(100/pk)
Add
Sample
Twist
604-398
Foil
Sample Cup Holder
Figure 5-3
Sample Preparation—Manual Method
1. Place the Foil Capsule on the balance and tare the balance.
2. Place the Small Tin Foil Capsule in the foil holder.
3. Place the Funnel on the foil holder.
4. Add Sample.
5. Insert and press down the plunger to pack the sample.
6. Holding the plunger in place, lift the funnel up against the
plunger, and then remove both pieces.
7. Crimp the Foil closed with tweezers.
8. Remove and weigh the sample.
601-492
PLUNGER
601-491
FUNNEL
502-186
FOIL CUP
(100/pk)
601-495
FOIL
HOLDER
601-489
BASE
601-490
FOILER
Figure 5-4
Sample Preparation—Foiler Method
Blank Login
Before a blank is analyzed, a sample named Blank must be entered
into the spreadsheet. This procedure is called Blank Login.
An analysis method should be developed before a sample is logged in.
Refer to Method Configuration on page 4–36.
1. Click Samples and select Login. The Sample Login dialog box will
appear.
2. Click the drop down arrow, in the sample name selection box, and
select Blank from the list.
3. Type the number of blanks repetitions you require.
4. Enter the information required in the Login Screen. Refer to the
Glossary, page 12–1, for Blank Login definitions.
5. Click OK to login the blank.
1. From the Samples menu, click Login. The Sample Login dialog
box will appear.
1. From the Samples menu, click Login. The Sample dialog box will
appear.
1. From the Samples menu, click Login. The Sample Login Dialog
box will appear.
Analyze a Blank
Analyze a Blank permits the operator to run an analysis without
burning a sample. This is done in order to obtain data to set the blank
area. Refer to Blank Calibration, page 5–31, for additional information.
Before proceeding, Blank Login, page 5–17, must be completed.
1. Click Analyze (F5) analysis will automatically start with the first
blank.
2. Blank results will be shown in the spreadsheet and plot window.
3. Accept a blank value for all elements in the calibration.
Analyze a Standard
Analyze a Standard is used to analyze standard samples for
calibration.
Before proceeding, Define a Calibration Standard, page 4–47, and
Login using a Calibration Standard, page 4–48, must be completed.
For some applications it may be necessary to use a standard with the
same density and/or concentration of sulfur. Consult the LECO
Applications Laboratory for further assistance.
LECO recommends using certified reference materials for all
calibrations.
Abort an Analysis
An analysis in progress can be aborted. If the analysis is aborted,
"Abort" will appear under analysis date.
Before proceeding, Analyze a Sample, page 5–23, must be
completed.
From the Samples menu, click Abort. The analysis in progress will be
aborted.
3. In the highlighted field, enter the date and time to start the
delayed analysis. The date and time that appears may be edited.
The format for entering the date and time is [05/01/2000
4:30:00 PM]. A space should be inserted after the date and time
or select Manually if pausing for operator intervention.
4. Click OK to start the delayed analysis. A clock symbol will appear
in the spreadsheet next to the sample that will be analyzed at the
programmed delayed start time or if Manually was selected a red
octagon will appear in the spreadsheet. The analysis will begin
only when the operator clicks analyze.
Blank Calibration
Blank Calibration is used to calculate the blank or baseline. Sometimes
this is called the blank area. Blank calibration calculates the
instrument blank and adjusts the area of the analysis accordingly. The
system blank should be determined every day by performing a blank
calibration before analysis. A blank should be analyzed and the blank
calibration set prior to standard calibration or drift correction. Refer to
Blank Calibration on page 5–31.
Standard Calibration
Standard Calibration is used to calibrate the instrument with known
calibration standards. A single point calibration can be performed using
the curve type (single standard sample) (at a single mass) however;
LECO recommends multipoint calibrations that employ 1/certified
weighting. Additionally a blank calibration should be completed prior to
performing any calibration. Since Standard Calibration is saved with
each method, each method must be calibrated after it is created.
Standard Samples
During the calibration process only dried standards should be
employed. The analytical accuracy depends on the accuracy of the
standards. Drying the standards optimizes the accuracy. Standards
should be dried according to label instructions. Refer to Calibration on
page 5–32.
Drift Correction
Drift Correction is used to adjust the original calibration response to
match the current instrument response. Drift calibration should be
performed at the start of every day or when check standards fail to
return the proper values. This insures accurate calibration and analysis
results. Refer to Drift Correction on page 5–34. LECO recommends
that all check standards be independent of the drift standards. After
performing a drift correction, the result for the drift standard will be
the same result that the drift standard returned during the original
calibration (not the certified value).
5. Click the arrow in the drop down box and Select a Cell (detector)
to calculate a calibration for. A separate calibration can be
calculated for each cell.
Add Standards
Add standards permits the operator to add analyzed standards to the
list of standards used for calibration.
1. Select analyzed standard samples from the spreadsheet to be
added to the list of standards used for calibration.
2. Click Configuration and select Calibrations.
3. From the fly-out menu click Add Standards.
4. The highlighted Analyzed Standards will be added to the list of
samples used for calibration. To view the list of standard samples
click new from the calibrations fly-out menu.
View Calibration
Permits the operator to view the current calibration. The calibration
cannot be changed from this screen.
1. Click Configuration and select Calibrations.
2. From the fly-out menu click View Calibration.
3. The Calibration screen will appear. The screen will be the same as
new calibration except most of the selections will not be available.
Drift Correction
Drift Correction is used to adjust the original calibration response to
match the current instrument response. Drift calibration should be
performed at the start of every day or when the check standard
doesn’t return the correct result. This insures accurate calibration and
analysis results.
After calibration, a Standard in the center of the calibration range will
be automatically selected (with a tolerance of ±12.5%) to be the drift
standard.
A different Drift Standard can be manually selected from the standard
calibration screen. Refer to Standard Calibration on page 5–32 for
more information.
A Blank Calibration should be performed before every Drift Correction.
Refer to Blank Calibration on page 5–31 for more information.
The Drift Standard should be homogenous and return precise results.
Check Standards should be independent of the drift standard.
1. Click Configuration and select Drift. The fly out menu will appear.
2. From the fly-out menu click View History. The View History screen
will appear.
1. Define a new standard for the new lot. Refer to Add Standards on
page 5–33.
2. Perform an analysis with the new standard defined in step 1.
3. Select the standard sample analysis result from step 2.
4. Click Configuration and select Drift.
5. Click Replace Drift from the fly-out menu.
6. A dialog box will appear. Select the Drift standard to replace.
7. Select the drift standard to replace then click OK. The new drift
Standard is selected as the drift standard
8. Select the new drift standard just analyzed from the spreadsheet.
Insert a Sample
If a sample is going to be added to the last row of the spreadsheet,
sample login should be used. If a sample should be inserted out of
sequence, between unanalyzed samples, Insert a Sample should be
used.
An analysis method should be developed before a sample is logged
in. Refer to Method Configuration on page 4–36.
1. Click and select a row in the spreadsheet after the last analyzed
sample. Insert will insert a row before the selected row.
2. From the Edit menu, click Insert. A row will be inserted into the
spreadsheet before the selected row.
1. Click and drag the mouse pointer to select the cells to fill with the
same data.
In the following example the mouse pointer was dragged over
the Operator cells to be filled with the same operator’s name.
2. From the Edit menu, click Fill. The Fill screen will appear.
Recalculate Results
1. Select the samples in the spreadsheet to recalculate.
2. From the Samples menu, click Recalculate. The selected samples
will be recalculated using the new calibration.
Results are automatically recalculated when a field in the spreadsheet
is edited (ex. name, mass, method, analysis date, etc.).
If the calibration is changed, it can be reapplied by following the
procedure above.
View History
View History permits the operator or laboratory manager to look at a
history log of the actions performed on a selected sample.
2. Click the Samples menu and select View History. The View History
screen will appear. Refer to the following View History Definitions.
3. Click the Local Time button, in the Time screen, to display the
time in local 12-hour time.
4. Click the UTC button, in the Time screen, to display the time in
Universal Coordinated Time.
5. Click OK to exit view history.
Sample Signature
A signature confirms that a user has accepted the sample information
displayed in the spreadsheet. Any change to the sample information
after it is signed will clear the signature. The signature can be used to
ensure that the validity of the sample information has remained the
same since it was signed.
To view a sample signature, after it has been signed, refer to View
Signature on page 5–44.
2. Click the Samples menu and select Sign. The Sign screen will
appear.
View Signatures
View Signatures permits the operator or laboratory manager to view a
sample signature and determine if it is valid. A sample must be signed
before the signature can be viewed. Refer to Sample Signature on
page 5–43.
If information associated with the sample was changed after it was
signed, the signature will be cleared and the view signatures screen
will not appear in step 2.
1. Click on a row to select a sample from the spreadsheet to view.
2. Click the Samples menu and select View Signatures. The View
Signatures screen will appear.
When multiple columns are selected to appear in the report, the font
and page orientation can be adjusted to improve the readability of
the report. The font size can be changed using the font tab as
described in Configuring Report before Printing, page 5–45.
6. Click OK. The Print dialog box will appear.
7. Click OK to print.
Export Configuration
Filename
1. Select the Configuration menu and click Text Export Formats. The
Export Format Screen will appear.
7. Enter a filename for the exported file or click browse and search
for a file name.
8. Click Append to File, Automatically increment Filename or Prompt
for Filename as desired. Refer to Export Configuration Definitions
on page 5–50.
9. Click OK to save the changes and exit.
10. Click Cancel to exit this procedure without making changes.
Export Definitions
Append to File—When selected, the highlighted sample analysis data
will be attached to the previously saved file.
Automatically Increment Filename—If the filename contains numbers,
they will automatically be incremented by one when the file is saved.
Prompt for Filename—When selected, the export filename screen will
appear when exporting a file, prompting the operator to enter a
filename.
2. Position the cursor in the Export Format and click the mouse to
place the cursor at the point in the format to edit a field.
3. Select a Field from the Fields box and click Add Field to Export
Format to add a field.
4. Highlight a Field in the Export Format and then press Delete on
the keyboard to delete it.
5. Click OK to save the changes and exit.
6. Click Cancel to exit this procedure without making changes.
This is a status screen to view the export data format. The export
format cannot be changed from this screen.
Text Export Data permits the operator to select sample data and
export that data to a file. Sample data can be inserted into an Excel®
spreadsheet from the exported file.
1. Click and drag the mouse pointer to select the rows with sample
data to export to a file.
2. Click the Sample menu and select Text Export Data. The Export
Format selection screen will appear.
The format selection screen will not appear unless there is more than
one format selected.
Text Export Buffers permits the operator to select sample plot data
and export that data to a file. Sample plot data can be inserted into an
Excel® spreadsheet from the exported file.
1. Click and drag the mouse pointer to select the rows with sample
data to export to a file.
2. Click the Sample menu and select Text Export Buffers. The Export
Buffers Filename selection screen will appear.
File extensions of (.txt) or (.csv) are good choices to insert the saved
files into an Excel® spreadsheet.
3. Enter a filename and extension. Click Save to export and save the
file.
Text Import Data permits the operator to import sample data from
another application such as an Excel® spreadsheet.
1. Click the Sample menu and select Text Import Data. The Import
File selection screen will appear.
2. To add a new transmission format to the list, click Add and enter a
transmission format name in the row that appears at the bottom of
the list.
3. To move a selected transmission format up towards the top of the
list select it and click Promote.
4. To move the selected transmission format down towards the
bottom of the list select it and click Demote.
5. Select Ascending to list the transmission formats in ascending
alphabetical order.
6. Select Descending to list the transmission formats in descending
alphabetical order.
7. To delete a transmission format, select it and click Delete.
8. Click OK to save any changes and exit.
2. Click XML to export the text data in XML format. If XML is selected
no other parameters in the screen can be configured.
3. Click Column Header to transmit the column header.
4. Select or enter the Field, Start of Transmission, and End of Line
delimiters.
5. Click OK to save the changes and exit.
6. Click Cancel to exit this procedure without making changes.
2. Position the cursor in the Export Format and click the mouse to
place the cursor at the point in the format to edit a field.
3. Select a Field from the Fields box and click Add Field to Export
Format to add a field.
4. Highlight a Field in the Export Format and then press Delete on
the keyboard to delete it.
5. Click OK to save the changes and exit.
6. Click Cancel to exit this procedure without making changes.
Transmit Data permits the operator to select sample data and transmit
the data to a peripheral device. Sample data can be transmitted from
a serial port or over a network.
The transmission format selection screen will not appear unless there
is more than one format selected.
Transfer Export
1. Click and drag the mouse pointer to select the rows with sample
data to export to a file.
2. Click the Sample menu and select Transfer Export. The export file
selection screen will appear.
1. Click the Sample menu and select Transfer Import. The Import
File selection screen will appear.
3. Click Open. The sample data stored in the file will be restored as
the last samples in the spreadsheet.
Please report any errors that are not defined on page 5–72 to the
LECO Service Department.
Pause
Message Cause Actions
The date entered is not a Invalid data for Pause Enter a valid date for the
valid date pause
First select an unanalyzed Tried to Pause an analyzed Only use Pause on
sample before selecting sample unanalyzed samples
this command.
Analysis Warnings
Message Cause Actions
The <element name> Element’s results are Check the analysis for
answer is outside the\n outside the specified anomalies, possibly need
standard’s defined control control limits for the to drift the instrument.
limits.\n defined standard
Carousel Interaction
Message Cause Actions
To use Location, go to Number of carousels is set Enter the number of
system configuration and to 0 carousels to be used, or
select the number of do not try to enter a
carousels being used. location value
Current location range is Entered invalid location Enter a location value in
<minimum value> - value the proscribed range
<maximum value>
Application Software
Message Cause Actions
The application is already Trying to open another Only run one instance of
running. instance of CHN628 while the application at a time
the application is running
You must first exit the Session did not end, Shut down the application
application before shutting because you must shut and then try to end the
down. down the application first session
The selected language Language module missing Reinstall the software and
module could not be retry
loaded.
Missing IR Cell Missing linearization file Reinstall the software and
Linearization <file name> retry
Blank Calibration
Message Cause Actions
First select the blanks to No samples selected Select the samples to be
use for setting the new used
method blank area before
selecting this command.
The selected blanks must No method in one or more Select a method for each
all have a method before of the selected samples sample to be used
setting the new method
blank area.
The selected blanks must One or more of the Analyze all of the samples
all have been analyzed selected samples have not to be used
before setting the new been analyzed
method blank area.
The selected samples must Selection include some Select only blanks and try
all be blanks before sample other than blanks again
setting the new method
blank area.
The selected blanks must Selected samples refer to Select only samples with
all refer to the same multiple methods the same method
method before setting the
new method blank area.
Maintenance Counters
Message Cause Actions
This counter is required Tried to delete a required Do not try to delete
and cannot be deleted. counter required counters
A counter with name Trying to create a new Use a different name
<name> already exists counter
Balance
Message Cause Actions
The balance is sending Multiple mass entries from Check balance
more frequently than the balance in rapid communication setting at
allowed.\5–76either the succession the balance and the PC
balance is configured
incorrectly,\nor the print
key on the balance was
pressed twice.
Standards
Message Cause Actions
You cannot name a Tried to name a standard Do not name standards
standard “blank” ‘Blank’ ‘Blank’. It is reserved for
application use
You cannot delete Tried to delete a standard Delete all references to
standard “<standard still in use the standard before
name>”.\nIt is referenced deleting the standard
by a sample.
A standard with name Trying to create a new Use a different name
<name> already exists standard
Sample Template
Message Cause Actions
A sample template with Trying to create a new Use a different name
name <name> already sample template
exists
Drift Calibration
Message Cause Actions
Drift Mass Range Error : Entered a mass value Only enter mass ranges
Nominal Mass (<nominal outside the acceptable appropriate to the drift
mass>); Range range for the drift standard
(<acceptable mass standard
range>); Value Entered
(<entered mass value>)
Transmit
Message Cause Actions
There are no samples to No sample in the sample
transmit. log
First select the samples to No samples selected Select the samples to be
transmit before selecting used
this command.
A transmission format A data transmit format has Create a data transmit
must be defined to not been set up for this format for the method.
transmit a sample. method See setup transmission
formats
A communications error AN undefined error Verify the port exists, is
occurred while occurred during data active, and not is use by
transmitting. transmit other applications; If error
persists contact LECO
If the ballast portion of the oxygen leak check fails repeatedly and no
leaks are found in other areas, a dirty ballast tank could be the cause.
Ballast Tubing
Gas connections on the ballast and pinch valve assemblies are made
with flexible tubing. This tubing is pushed onto barb fitting and metal
tubing to form a gas tight seal. Refer to the figures on pages 6–11,
6–12, and 6–15.
In order to prevent leaks after maintenance, LECO recommends that
the flexible tubing be replaced when the ballast is cleaned or when the
tubing is removed from any fitting.
The flexible tubing is LECO part number 601-691-018 and is provided
in the component pack. Refer to page 1–31.
Figure 6-2
Ballast Top Assembly
Figure 6-3
Ballast Connections
Figure 6-4
C-Flex Tubing Connection
Figure 6-7
Carousel Bottom View
14. Insert the combustion tube into the furnace. The short end of the
tube should be inserted into the afterburner side of the furnace.
15. If the height of the combustion tube needs to be raised perform
the following steps. If the combustion tube is at the correct height
skip to step 16. If the furnace has been heated, the combustion
tube o-rings may be deformed and should be replaced.
16. Loosen the combustion tube stop on the bottom of the furnace
assembly, Figure 6-14, page 6–27.
17. Install the furnace bottom assembly.
Figure 6-11
Furnace Interface Blocks
Figure 6-13
Furnace Latch Pin
Figure 6-14
Combustion Furnace Assembly
190-423 601-442
SCREW O-RING
(4 REQ'D)
783-276
LANCE HOLDER
COVER
780-831
O-RING
(2 REQ'D)
783-275
LANCE HOLDER
BLOCK
616-146
LANCE
Figure 6-16
Lance Assembly
Figure 6-19
Lance Removal
Disassembly
1. Remove the three screws securing the loading head to the
furnace, Figure 6-18, page 6–33.
Figure 6-20
Slide Block Removal
Figure 6-23
Female Jaw and Loading Pin
Removal
1. Turn the Gas OFF.
2. Remove retaining clip above the aliquot dose tube.
3. Slide reagent tube upward until the bottom end can swing free.
4. Tilt out the free reagent tube end.
5. Pull the reagent tube downward off the top port.
Packing
1. Before repacking the reagent tube, remove all material from the
inside of the reagent tube, rinse the reagent tube with water, and
let it dry.
2. Insert a screen filter into the bottom of the reagent tube. Position
the o-ring towards the bottom and push the screen filter up into
the reagent tube 1/4-inch, Figure 6-24 on page 6–40.
3. Pack 3/4-inch of glass wool into the bottom of the reagent tube.
4. Fill the bottom 1/3 of the reagent tube with Anhydrone.
5. Fill the top 2/3 of the reagent tube with Lecosorb.
6. Firmly pack 1/2-inch to 2/3-inch of glass wool into the top of the
reagent tube.
7. Reinstall the reagent tube.
8. Reset the Reagent Tube Counter under Reset Maintenance
Counters, page 6–48.
9. Clean excessive reagent from work area.
Install
1. Push the top end of the reagent tube over the top port.
2. Tilt the reagent tube inward and pull it down, pushing it over the
bottom port.
3. Turn the gas on.
619-268
ALIQUOT DOSE
REAGENT TUBE 502-174
LECOSORB
501-171
ANHYDRONE
617-548
INCOMING 763-265
HELIUM SCRUBBER GLASS WOOL
REAGENT TUBE
775-306
SCREEN FILTER
Figure 6-24
Reagent Tube Packing
Removal
1. Turn the gas OFF.
2. Remove screw and open the door. Refer to Figure 6-25,
page 6–42.
3. Remove retaining clip above the glass tube on the right.
4. Slide the furnace filter tube upward until the bottom end can
swing free.
5. Tilt out the free furnace filter tube end.
6. Pull the furnace filter tube downward off the top port.
Cleaning
1. Remove the glass tube on right side, Figure 6-25,
page 6–42.
2. Clean and regrease o-rings.
3. Replace glass wool.
4. Reinsert glass tube.
5. Reinstall retaining clip removed is Step 3 above.
Install
1. Push the top end of the furnace filter tube over the top port.
2. Tilt the furnace filter tube inward and pull it down, pushing it over
the bottom port.
3. Turn the gas on.
Monitor the Secondary Wire Screen Filter regularly for corrosion. If the filter is
corroded replace with a new Secondary Wire Screen Filter.
Removal
1. Refer to Combustion Tube Removal and Replacement
(page 6–23) for instruction on accessing the Secondary Wire
Screen Filter.
Replacement:
1. When Secondary Wire Screen Filter is visibly dirty or corroded,
replace with a new one. Use a pair of needle nose pliers to
remove old filter and simply drop the new one in place.
2. Click the check box before the item that periodic maintenance
was performed on.
3. Click on the Comments box and enter any maintenance
comments.
4. Click OK.
The log file is a history of past events. Every time one of the following
procedures is performed, it will be recorded in the log file.
• Change of Mass by Operator
• Data Out of Range (Ambient Monitor parameter values)
• Date of last calibration, drift or blank
• Date scheduled maintenance is performed
• Network Errors
• Network Sign-on Errors
• Sample Aborted
• Sample Deleted
• Standard check out-of-range
1. From the Maintenance menu, click View Log Files. The Log File
window will appear.
S = aMs + b (1)
cx = (bcs/aMx)
The analyte concentration can then be calculated from the slope and
intercept of the standard addition calibration curve.
As an example, this technique can be used to determine the unknown
concentration of the nitrogen content in aqueous standards.
In the following calibration screen example, linear was selected and
the standard additions offset appears on the calibration screen at the
left.
Example
Barometric pressure is reported at 750 mmHg at sea level (0 feet) but
operation altitude is 1000 feet. Adjust the reported sea level pressure
to operation level pressure by using the formula:
733 mmHg X 750 mmHg = 723 mmHg
760 mmHg
In this example, the barometric pressure to be entered into the
instrument is 723 mmHg.
The thermal conductivity cell has the ability to detect differences in the
thermal conductivity of gases (see below). This TC cell is operated in a
nitrogen optimized mode. The cell consists of two pair of matched
filaments used in four legs of a Wheatstone bridge. The "reference"
filaments are maintained in a constant gas and gas flow environment
while the "measure" filaments are maintained in a constant gas flow
environment, but the gas composition is allowed to vary. All filaments
are mounted in an insulated metal block. A Filament Current Factor is
calculated by the software to compensate for changes in the filament
current due to TC cell temperature changes.
The Wheatstone Bridge is balanced with a specified bridge current
while both filaments are essentially in identical environments.
The bridge current causes self-heating of the filaments. The
temperature of the filaments is always much higher than the oven
temperature in which the cell is located.
As long as both filaments remain in the same environment under
which the bridge was balanced, the bridge output will remain at
approximately 0.3 volts. Any disturbance of this environment will
result in a change in bridge output. The bridge output is AC coupled to
amplifiers that have an output offset.
Once the bridge is balanced and stabilized under proper conditions, the
only variations in bridge output are due to variations in the type and
quantity of gas present at the "measure" filaments. The bridge is
balanced with Helium (He) flowing in the measure chamber and
reference chamber. The introduction of nitrogen causes the
temperature of the measure filaments to increase since nitrogen has a
lower thermal conductivity than helium. In this instance, the bridge
becomes unbalanced and an output becomes available to the preamp,
resulting in a positive reading. The amount of nitrogen determines the
magnitude of the readings. The sensitivity of the system is governed
by bridge current and the difference in thermal conductivity between
the analyze gas, nitrogen, and the carrier gas, helium.
Figure 7-3
Thermal Conductivity of Gases
4 4 4 4
3 3 3 3
2 2 2 2
1 1 1 1
0 0 0 0
1 9 17 25 33 1 10 19 28 37 1 10 19 28 37 1 10 19 28 37
Time Time Time Time
Raw Area = 19 Raw Area = 18 Raw Area = 19 Raw Area = 18
The area under the peak is calculated and stored as a raw area, Ar.
The current blank area, Ab, is subtracted from the raw area, Ar, of the
standard to yield the Area, A.
A = Ar - Ab
The area A, is adjusted for instrument drift by multiplying by the drift
factor. Each time a calibration is performed the drift factor is reset to
1.0. The units of the area, A, are then adjusted to reflect absolute
amount of analyte by multiplying by the sensitivity factor.
Aadj = A * Sensitivity Factor * Drift Factor
The resultant adjusted area, Aadj, of the standards are plotted against
the known grams Analyte * 100 of the analyzed standards. The best-
fit line, linear, quadratic, or cubic, is determined and taken as the
calibration curve for the method. The default calibration curve has a
slope of 1.0 and travels through the origin.
200 200
150 150
100 100
0 0
0 50 100 150 200 250 0 50 100 150 200 250
Calibration Definitions
1/Certified Weighting—A weighting that can be applied to the
calibration data points that will counteract a bias for higher
concentrations that comes naturally out of the curve fitting routines.
Adjusted Area (Aadj)—Adjusted Area = Area (A) times Correction
Factor times Sensitivity Factor times Drift Factor. Area (A) equals Raw
Area (Ar) minus Blank Area (Ab).
Analyte—The substance whose physical or chemical properties are
measured and correlated, directly or indirectly, to the desired
information.
Atmospheric Blank—In every encapsulated sample there are both
sample material and air (atmosphere). The amount of air trapped with
the sample, atmospheric blank, is dependant upon the volume of
sample (sample mass) and the physical state of the sample (pellet,
liquid, powder, crystal, grain, etc.). After the combustion of the sample
the nitrogen from sample and atmosphere are presented to the
detector and mass.
Blank—The signal obtained during an analysis that cannot be
attributed to the sample. The blank is due primarily to Argon
impurities in the oxygen. The blank should be determined before
calibration or drift correction.
20
10
0
0 50 100
Blank Analysis
Several blanks should be analyzed. The linearization table is applied to
the peak during the analysis. The area under the peak is calculated
and stored as a raw area (Ar).
4 4 4 4
3 3 3 3
2 2 2 2
1 1 1 1
0 0 0 0
1 9 17 25 33 1 10 19 28 37 1 10 19 28 37 1 10 19 28 37
Time Time Time Time
Raw Area = 19 Raw Area = 18 Raw Area = 19 Raw Area = 18
Standard Analysis
After the blank is defined, a calibration needs to be performed. At least
three replicates of the same standard need to be analyzed in the same
fashion as the blanks. The standard needs to be selected such that the
analyte concentration is in the middle of the operating range. The
linearization table is applied to the peak during the analysis.
The area under the peak is calculated and stored as a raw area, Ar.
The current blank area, Ab, is subtracted from the raw area, Ar, of the
standard to yield the Area, A.
A = Ar - Ab
10
8
grams Analyte * 100
0
0 2 4 6 8 10
Are a * Se nsitivity Factor
Slope = 1
The drift factor uses the calculated concentration from the calibration
curve and not the certified value.
DETECTOR OUTPUT
COLLECTED
DETECTOR OUTPUT
COLLECTED
COMPARATOR LEVEL
(SET TOO HIGH)
DETECTOR SIGNIFICANT
OUTPUT DETECTOR
COLLECTED OUTPUT
EXCLUDED
COMPARATOR LEVEL
(SET TOO LOW)
INSIGNIFICANT
DETECTOR
OUTPUT
DETECTOR
OUTPUT
COLLECTED
Ambient Chart
1. Click Diagnostics and select Chart to display the Ambient Monitor
Chart screen.
2. Check that all system parameter values fall within the Ambient
Monitor Ranges shown on page 8–6. If any values are out of
range, refer to the corresponding topic for additional information.
The instrument must be on for a least 2 hours and the oxygen flow
set to low before checking the following parameter values.
Set Bridge
Set Bridge sets the gain of the TC Cell.
The TC Cell gain has been set during the manufacturing process. It is
not necessary to reset the gain unless it has been serviced or
replaced.
1. Click Auto Set Bridge to automatically set the gain of the TC Cell.
The gain factor will appear in the edit box to the left of the set
bridge button.
2. To manually set the gain, enter the gain factor in the edit box and
click Set Bridge. The gain factor must be between 0 and 255. This
can be used for troubleshooting or servicing the instrument.
3. Enter the file name and folder location in the dialog box.
4. Click Save to save the backup file.
3. Select the folder and backup file from the dialog box.
4. Click Open to load the calibration values from the selected backup
file.
Reset Network
Click Reset Network to restart the network software. If something or
some operation causes an error this may correct it.
Upgrade a Device
1. To upgrade a device with a new version of application firmware,
click the Device to upgrade. Then click Download to download the
new firmware.
2. Click Reset Board to restart firmware application program on the
circuit board.
Network Definitions
Device—The name of the electronic assembly connected to the
network. Example: TC Cell.
MAC-id—The network ID number assigned to the electronic assembly
connected to the network. This number is assigned at power up and
can vary.
Class—The type number of the electronic assembly signed on to the
network. Example: 21 is the class number assigned to the TC Cell.
Instance—A unique number assigned to an electronic assembly. If two
or more of the same type electronic assemblies are installed in the
instrument this number will be different for each assembly and is used
by the controller to differentiate one from the other.
Serial—The hardware ID number assigned to the electronic assembly.
This number is assigned during manufacturing and is unique to each
electronic assembly. The serial number is stored on the circuit board in
a PROM.
Version—The version number of the application software located on
the electronic assembly. This software can be upgraded from the
network procedure.
Checksum—A number unique to the application software to verify if
the software was properly installed. The will be a checksum number
supplied with any software upgrades. Verify that the number
appearing in the checksum cell is the same as the number supplied
with the software upgrade.
NI Version—The version number of the network control software
installed in the electronic assembly. This software is located in a PROM
and cannot be upgraded from the network procedure.
Hardware ID—Additional circuit board hardware identification.
FPGA Version—The version of the FPGA (Field Programmable Gate
Array) software.
The Furnace diagnostics screen is used to test the furnace and verify
proper operation. The meters on the right side of the screen can be
used to monitor Furnace Temperature.
When testing the furnace, set the temperature and observe the target
value. The furnace should ramp to the set value and remain there.
Refer to Furnace Diagnostics Definitions on page 8–15.
Click Diagnostics and select Furnace. The Furnace diagnostics
screen will appear.
2. Click a button in the upper left of the screen to select the leak
check procedure to perform. Only one procedure can be selected
at a time.
3. Click Start to perform a leak check. System pressure is monitored
in the chart at the middle of the screen.
4. Click Stop to abort a leak check in progress.
5. Click Print to print the results of a completed leak check on the
system printer.
Figure 8-1
Analytical Function Diagram
2. Click Start to system check and test the systems listed on the
screen.
3. The results of system check will be displayed to the right of the
system being checked under the results column. An X inside the
circle to the left of the system also indicates the system checked
failed.
4. Click Print to print the results of the system check on the system
printer.
2. Click the Check Box next to the desired solenoid valve or switch
to activate or deactivate it. A check mark in the box indicates it’s
activated.
3. Check Solenoids to highlight and locate the solenoid valves on the
flow diagram.
4. Check Gauges to locate the system pressure gauges and display
their current pressure reading on the flow diagram.
5. Check Cells to locate the IR and TC cells and display their current
cell voltage on the flow diagram.
The following section explains the required settings, below, and the
installation procedure, page 9–6, for LECO software. Additional
information can be found in Configuring the Ethernet Adaptor, page
9–13, Configuring Firewall Settings, page 9–18, and Registering the
Software, page 9–19.
Antivirus Software
LECO recommends that antivirus checking be turned off
for the files listed below to avoid interference with data
collection or storage.
Do not install the Copy Protection Key until after the software has
been installed.
10. Determine the folder location for Adobe Reader. Select Next for
the default location, or select Browse, select a different location,
and then select Next.
12. When the "Setup Completed" dialog box appears, click Finish.
Windows 7
9. Click OK.
The following steps apply only when a firewall program other than
Microsoft Windows is used.
Do not insert the copy protection key into the USB port until
the LECO software is installed and the PC has been
rebooted.
4. Click Copy.
5. Open a blank email message and press Ctrl and V on the
keyboard to paste the license information into the email message.
6. Send the email or just the Key number to the LECO designee.
The Copy Protection Key Not Found or License Not Valid dialog box
may appear if the copy protection key is not installed, the software is
not registered, or the software is already running. Follow the steps
below.
• Verify the copy protection key is installed. If not,
1. Click OK, and then insert the copy protection key.
2. On the Licenses dialog box, click Close. The software login will
appear.
• If the copy protection key is installed, the software may not be
registered.
1. Click OK.
2. Insert the copy protection key. The Licenses dialog box will
appear.
3. Click on the old version and then click Change. Follow the
instructions in Registering Software, beginning with step 5.
Initial Alignment
Before the load head assembly is ever placed onto the load head
interface block, the following procedure must be followed to insure
easy removal/installation of the load head assembly.
1. Loosen the following screws to allow alignment the loading head
assembly, to the load head interface block on the dual furnace
assembly. Refer to Figure 10-32 for screw part numbers.
A. Loosen (2) screws which secure the male mounting block to
the loading head assembly. Refer to Figure 9-1.
B. Loosen the screw which secures the connector assembly to
the male mounting block.
C. Loosen the (2) setscrews in the male mounting block. Refer
to Figure 9-2.
D. Loosen the (2) screws that secures the female mounting
block to load head interface block on the furnace assembly.
Refer to Figure 9-3.
E. Loosen the screw which secures the connector assembly to
the female mounting block.
F. Back off the setscrew in the female mounting block.
2. With everything above loosened, carefully align loading head
assembly and the load head interface block on the furnace
assembly.
3. Grab the loading head at each side, carefully line-up the male
electrical connector on the loading head and the female connector
on the loading head interface block. At the same time, verify the
three nipples from the loading head are lined up with the bores
on the load head interface block. Once aligned, push down. You
may hear a hissing sound.
4. Tighten the captive screws in the loading head to secure it to the
loading head interface block. Any hissing sound should disappear.
If a hissing sound continues, remove load head and refer to the
loading head interface block maintenance on page 9–31.
Figure 9-1
Male Mounting Block Top Screws
Figure 9-3
Loading Heading Alignment—Interface Block
Figure 9-4
Slide Block Speed Adjustments
This procedure should be used to clean and service the Loading Head
Interface Block.
Perform the Instrument Shutdown procedure before removing the
Loading Head. Refer to Instrument Shutdown on page 5–12.
Disassembly
1. Turn off the pneumatic gas supply.
2. Remove the loading head. Refer to the load head removal
procedure on page 9–26.
3. Remove top furnace plate.
4. Remove lance assembly. Refer to the lance assembly removal
procedure.
5. Remove and inspect the loading head block interface seal o-ring,
601-504. Lightly apply a thin coating of vacuum grease to the
o-ring.
6. From Diagnostics access the Solenoids and Switches screen.
A. Toggle the Set Slide Block Seal (SV18) solenoid to relieve
pressure in lines.
B. Toggle Slide Block Open and Close (SV13/SV14) solenoid to
relieve pressure in lines.
7. Remove the (2) 8-32 socket head screws, 190-867, which secure
the valve cap.
8. Carefully remove the valve cap. Verify the three o-rings
underneath the valve cap remain on the interface block.
9. Inspect and clean valve cap.
10. Inspect and clean three o-rings. Lightly apply a thin coating of
vacuum grease to the o-rings.
11. Remove (2) balls, 805-079, and (2) springs, 617-212, from two
outside bores.
12. Inspect and clean the balls, springs, and holes.
PINCH HAZARD
Keep your hands away from the carousel and loading head
during operation.
Figure 9-6
Carousel Alignment
SHOCK HAZARD
Turn the AC power OFF and disconnect the instrument from
the facility electrical power source.
Removal
1. Remove the right side panel from the instrument.
2. Remove the plastic cover from the oven assembly.
3. Loosen the TC cell Thumbscrew.
4. Remove the TC cell cover.
5. Disconnect the network electrical connection to the TC cell.
6. Loosen the Captive Screw.
7. Pull the TC cell away and out of the oven.
Service
1. Check the condition of the o-rings. If they are cracked or worn
replace them.
2. If you having a problem with the TC Cell, ambient values, a gas
leak, or measurement results, contact the LECO service
department for further information.
SHOCK HAZARD
Turn the AC power OFF and disconnect the instrument from
the facility electrical power source.
Removal
1. Remove the right side panel from the instrument.
2. Remove the plastic cover from the oven assembly.
3. Remove the foam insulation covering the IR cells.
4. Remove the two IR cell mounting screws.
5. Pull the IR cell away and out of the oven.
Service
1. Check the condition of the o-rings. If they are cracked or worn
replace them.
2. If you having a problem with the IR Cell, IR Cell ambient values,
a gas leak, or measurement results, contact the LECO service
department for further information.
2. All the IR Detectors for the system are listed. If the IR detector
has been replaced, enter the reference and saturation that are
provided on the tag attached to the replacement IR detector.
3. Refer to the IR Detector Linearization Definitions, page 9–38, for
additional information.
After values have been changed, existing methods must be
recalibrated.
10. Select the sample data run in step 9 and click Configuration, then
select Calibration.
13. Within the grid of the calibration screen, select the results of the
EDTA analyses.
14. Click Edit and select Optimize Saturation.
Use the troubleshooting table to locate and correct problems that may
develop with the instrument.
Troubleshooting Table
A WIRING HARNESSES A
B ECLIPSE_ETHERNET
J512I
ECLIPSE 54" H9 ECL_PNEUMATIC_FLOW
3-A3
B
618-045
ECLIPSE
ECLIPSE 36" H9 ECLIPSE_OVEN
J512J F2
618-040
J512K
622-001-307 622-001-308 625-601-621 622-001-261 J573A
DC POWER IN
WIRING_HARNESS WIRING_HARNESS WIRING_HARNESS WIRING_HARNESS
J512P
AC_POWER DC_POWER PV_PLATE_BALLAST JUMPER_FAN
+ J512AL
J512Q
J512A J512A
ECLIPSE RJ45-DSUB 625-099 A88
+24VDC 1 1 JF6
C +24VDC 2 2
H6
1 +
RED 615-784
FURNACE_BLOWER
C
622-001-347 622-001-368 619-110 619-230 H6
J512AJ 24GND 3 3 2 - FAN6
WIRING_HARNESS WIRING_HARNESS WIRING_HARNESS WIRING_HARNESS PS1 V+
V+
1
2
H2 1 +24VDC 24GND 4 4
BL
ECLIPSE BALLST_FAN TE_COOLER LOAD_HEAD
2 24GND
V+ 3
J512F J512F
V- 4
J512AC +24VDC 1 1 JF7 RED
H6 616-140
H13 H14 H15 H16 619-288
V- 5
H2 1 +24VDC_IN +24VDC 2 2
H6
1 + FURNACE_ENCLOSURE_FAN_A
V- 6
2 NC 24GND 3 3 2 - BL FAN7
3 +24VDC_RET 24GND 4 4
H1
G 7
J512AF J512G J512G
N 8 1 +24VDC_IN +24VDC 1 1 JF8 RED
H6 616-140
622-001-369 622-001-375 622-001-388 622-001-018 24V L 9
2 NC +24VDC 2 2
H6
1 + FURNACE_ENCLOSURE_FAN_B
D J512AN J512M D
1 RSENSE+ +24VDC 1
2 NC +24VDC 2
3 RSENSE- 24GND 3
24GND 4
J512N J512N
H6 TEC-24PWR1
+24VDC 1 1 3-G1
H6 TEC-24PWR2
+24VDC 2 2 3-G1
H6 TEC-24GND1
J512L J512L 24GND 3 3 3-G1
H1 H6 TEC-24GND2
1 1 AC_HI_A 24GND 4 4 3-G1
2 2 GND
DETECTOR OVEN H1
3 3 AC_LO_A
+24VDC
J512V
1
J512V
1
H6
JF4 RED 614-724
J512T +24VDC 2 2
H6
1 + TOP_CABINET_FAN
MOV1 1 1 AC_HI_A 24GND 3 3 2 - BL FAN4
E J313D J313D JF3 RED 622-001-166
619-001 2 2 GND 24GND 4 4 E
A3A FAN_1+ 1 1 1 + CELL_OVEN_FAN
ANALYZER MOV 3 3 AC_LO_A
FAN_1- 2 2 2 - BL FAN3 J512W J512W
J512U J512U +24VDC 1 1 JF5 RED
A7 H5 H6 614-724
666-310X120 J313F
RED FAN CELL_OVEN
4-C7 1 1 AC_HI_A +24VDC 2 2
H6
1 + BOTTOM_CABINET_FAN
ECLIPSE_DAUGHTER FAN_2+ 1 + 619-249 DUAL_HTR_V7 H5
2 2 GND 24GND 3 3 2 - BL FAN5
FAN_2- 2 -
BL FAN10 4-C7 3 3 AC_LO_A 24GND 4 4
LINE_CORD
POWER ENTRY
615-574
RECEPTACLE
H1
LINE_FILTER
H1
CB1 15A
H1
2
3
1
GND
AC_LO_B
J512AI
AC_HI_B
+24VDC
+24VDC
24GND
J512AE
1
2
3
LECO
Description
St. Joseph,
Michigan USA
J51
AC2
GND
H1 H1
LINE2
GND 20A
LOAD2
610-052
LINE2
MAIN_BREAKER
LOAD2
789-320
J512AM
1 AC_HI_B
+24VDC
+24VDC
J512AH
1
2
CHN628 G
2 GND 24GND 3 Part Number Rev
PROTECTIVE EARTH GROUND.
THIS GRAPHIC IS TO BE
H1 622-001-433 3 AC_LO_B
ECLIPSE_DISTRIBUTION
24GND 4
151-035 D
STAMPED NEXT TO
GND
FURNACE
CONNECTION. CHASSIS GROUND BASE GROUND HOUSING BASE GROUND
Date: 04-09-2012_14:25 Type M
GND 8 8 JSV7
H3
SOL15 9 9 1 + 618-119
A4A GND 10 10
H3
2 - SV7 HELIUM_MICRO_CARRIER_VALVE
SOL16 11 11
GND 12 12
JSV10
H3
1 + 618-119
666-310X120 SOL17 13 13 H3
2 - SV10 HELIUM_LEAK_CHECK_VALVE
1
J516U
1 GND
SOL19 17 17
H3
2 - SV11 HELIUM_ANALYZE_FLOW_VALVE
SENSOR BLUE PS- GND 18 18 JPV1
B 2 2 AIN_1- H3 B
PNEUMATIC_PRESSURE MT4 BLK OUT
3 3 AIN_1+ J516
SOL20 19 19
H3
1 +
PV1
621-557-101
622-001-596 BRWN PS+
4 4 +24V_OUT
GND 20 20 2 - COMBUSTION_ISOLATION_VALVE
SOL21 21 21
J93 J93 JSW13 616-583
GND 22 22 SW13_C H7
J516W J516W D4 1 1 1 SW13
MT01-SPLY H3 N/C 23 23 SW13_E H7
C8 1 1 +5V_ANA D4 2 2 2 SWITCH
MT01-OUT H3 KEYED 24 24 H3 H7
C8 2 2 AIN_2+ 3 3 PISTON_HOME
H3 H7 JPV3
3 3 AIN_2- J516H J516H 4 4 621-557-101
MT01-GND H3 H3 H7 1 +
D8 4 4 GND SOL22
GND
1
2
1
2
H3
5
6
5
6
H7 2 - PV3 VALVE
BALLAST_EXHAUST
H3 H7
J516V SOL23 3 3 7 7 JPV4
H3 H7 621-557-101
1 GND GND 4 4 8 8 1 +
2 AIN_3- SOL24 5 5
H3
9 9
H7
2 - PV4 VALVE
H3 H7 BALLAST_INPUT
3 AIN_3+ GND 6 6 10 10
4 +24V_OUT N/C 7 7 11 11
H7
1
JPV5
621-557-101
H7 1 +
JMT2
H3
KEYED 8 8
MT01-SPLY
12 12
H7
2
2 - PV5 VALVE
MTA HEADER
619-043 H3 MT01-OUT H7
3 2 2 1 1 +5V_ANA GND 10 10 C4 14 14 JPV6
COMBUSTION_PRESSURE MT2 2 3 3
H3
2 2 AIN_4+ A4 SOL26 11 11 C4
MT01-GND
15 15
H7
1 +
621-557-101
SENSOR
1 4 4
H3
3 3 AIN_4- 666-742X GND 12 12 16 16 2 - PV6 VALVE
BALLAST_BY-PASS
4 4 GND VALVE1+ 13 13
VALVE1- 14 14 RED SPLY
SENSOR
J516AA J516AA VALVE2+ 15 15 GREEN OUT
622-001-606
OUT+ H3
1 1 +24V_OUT VALVE2- 16 16 BLACK GND MT1 BALLAST_PRESSURE
BROWN 625-601-790
SW14
OXYGEN_PRESSURE_SWITCH
BLUE
2
3
2
3
SW1 +24V
+24V
17
18
17
18
JPV7
GND H3
OUT- H3 1 + 621-557-101
24GND
24GND
19
20
19
20
H3
2 - PV7 BALLAST_SAMPLE_VALVE
J514Y J516Y
SW13_C H3 JLH JLH JHLI JHLI J20
C8 1 1 +24V_LIM1 J516J J516J H3 H12 H4
1 1 + + 1 + 617-434
C8
SW13_E H3
2
3
2
3
+5V_LIM1 SOL1
GND
1
2
1
2
H3
2 2
H12
- -
H4
2 - SV17 SAMPLE_DUMP_CLOSED_SOLENOID
OPTO_SW1
J516K
JSV12 B B
D J516Z C C 619-590-509 D
1 +24V_LIM2
SOL2
GND
1
2
A A SV12 PISTON_RETURN_VALVE
2 622-001-597
+5V_LIM2
3 OPTO_SW2 J516L J516L JSV19 B B
SOL3 1 1 C C 619-590-509
GND 2 2 A A SV19 PNEUMATIC_PRESSURE_SELECT_VALVE
622-001-604
J516M J516M
SOL4 1 1
JSV18 B B
C C 619-590-509
GND 2 2
A A SV18 SLIDE_BLOCK_SEAL_VALVE
622-001-603
J516N J516N
SOL5 1 1 JSV14 B B
GND 2 2 C C 619-590-509
JSV13 B B
A A SV14 SLIDE_BLOCK_CLOSED_VALVE
F F
J386 +
J386E J386E
H11
JHP1 JHP1 HP1
POWER_OUT1+ 1 1 1 1 619-225 TEC
H11
J386A POWER_OUT1- 2 2 2 2 HEAT_PUMP
ECLIPSE_TEC ECLIPSE H11 -
2-B11 ECLIPSE VALVE- 3 3
H11
JMT9 JMT9 T MT9
THERMISTOR1+ 4 4 1 1
H11 619-708
THERMISTOR1- 5 5 2 2
H11 TEC_THERMISTOR
VALVE+ 6 6 R
LECO
5000
POWER_OUT2- 7 7
JPV2 JPV2A JPV2 St. Joseph,
J386B J386B POWER_OUT2+ 8 8 H15 621-557-101
TEC-24PWR1 1 1 1 +
2-D10
TEC-24PWR2
1 1 AUX_+24VDC
2 2
H15
2 - PV2 VALVE Michigan USA
2-D10 2 2 AUX_+24VDC J386F J386F TEC_COOLER_DRAIN
TEC-24GND1 H11
2-E10 3 3 GND FAN+ 1 1
TEC-24GND2 H11 Description
2-E10 4 4 GND FAN- 2 2 CABINET
Schematic
J386C
G 2-F10
2-F10
TEC-24PWR3
TEC-24PWR4
1
2
J386C
1 AUX_+24VDC
2 AUX_+24VDC
J386D
SPARE1+ 1
SPARE1- 2
JF2
1 +
RED FAN
619-249
TEC_EXHAUST CHN628 G
2-F10
TEC-24GND3
3 3 GND THERMISTOR2+ 3
2 -
BL FAN2 Part Number Rev
2-F10
TEC-24GND4
4 4 GND THERMISTOR2- 4
SPARE2- 5
151-035 D
SPARE2+ 6
Date: 04-09-2012_14:25 Type M
Schematic Sheet 3 OF 4
1 2 3 4 5 6 7 8 9 10 11 Rev G
1 2 3 4 5 6 7 8 9 10 11
Name: 151035
A6
+5VANA 1 1
GREEN OUT
30psi A7 666-514X170
PRESSURE 2 2
BLACK GND MT6 TC_PRESSURE
DUAL_HEATER_VERSION-7
A 666-382X140 GND 3 3 625-601-833
A
TC_CELL_CARD_V4 J291H J291H J634A
H16
SENSE+ 1 1 3 3 TEMP
H16 HEATER
SENSE- 2 2 2 2 N/C 625-510-182
A21 666-554X
IR_EMITTER_CTRL
A22
IR_INTERCONNECT_6.00_INCH
666-260X A23
THERMOPILE_PREAMP
666-553X120
(CHN ONLY)
J381E
J284E
D 2-B11
ECLIPSE_IR ECLIPSE
ECLIPSE
D
J284C J241B J241A J259B J259 J240 J240A
J5
1 SOL0+
A25 666-554X
IR_EMITTER_CTRL
A26
IR_INTERCONNECT_0.725_INCH
666-258X
A27 666-240X (CN & CHN ONLY)
666-381X THERMOPILE_PREAMP
2 SOL0-
3 SOL1+ IR_CONTROL_CARD
4 SOL1-
5
6
SOL2+
SOL2-
A9 J284D J241B J241A J260B J260 J240 J240A
7
8
SOL3+
SOL3-
A? 666-554X A? 666-260X A? 666-553X
(S ONLY)
IR_EMITTER_CTRL IR_INTERCONNECT_6.00_INCH THERMOPILE_PREAMP
FURNACE CONTROL
E E
A8 666-514X180
IR CELLS IN DUAL_HEATER_VERSION-8
J284A UPPER
LEFT NOT USED NOT USED NOT USED NOT USED NOT USED NOT USED 2 2 HTR6
622-001-002
F
J284B LOWER
LEFT
A20
625-509
4
3
4
AFTERBURNER
F
J284C LOWER
RIGHT
A24
625-348
IR_CELL_0.725"_CO2
A24
625-348
IR_CELL_0.725"_CO2
A24
625-348
IR_CELL_6"_SO2
A?
625-345
IR_CELL_6"_SO2
A?
625-345
IR_CELL_6"_SO2
J514E
1 EXT_RELAY0
HEATER_0_LO 2 2
H5
2 COIL_0- TB1A
3 GND K_TC_0+ 1 + 660-010-130
4 SOLENOID_0 K_TC_0- 2 - MT10 COMBUSTION_FURNACE R
J514A
ECL_DUAL_HTR_V8 ECLIPSE
2-A11 ECLIPSE Michigan USA
J514F J514F
Description
A8
2-F7 1 1 AC_HI J514B J514B
H5
CABINET
DUAL_HTR_V8
2 2 CH_GND HEATER_1_HI 1 1
H5
Schematic
G 2-F7 3 3 AC_LO
J514D
HEATER_1_LO 2 2
CHN628 G
1 EXT_RELAY1 TB3A Part Number Rev
2 COIL_1-
3 GND
K_TC_1+ 1
K_TC_1- 2
+
- MT12 660-010-130
AFTERBURNER
151-035 D
4 SOLENOID_1 THERMOCOUPLE
Date: 04-09-2012_14:25 Type M
Schematic Sheet 4 OF 4
1 2 3 4 5 6 7 8 9 10 11 Rev G
This page intentionally left blank.
A
Add as Next Sample to Run: Click to enter the logged in standard
after the last sample that was analyzed.
Add to End of List: Check to enter the logged in standard in the last
row of the spreadsheet.
Aliquot Dose Reagent Tube: Purifies the gas leaving the aliquot
doser and catalyst heater by removing CO2 and moisture.
Analysis Date: The date and time the analysis was completed.
B
Balance Mode: If a balance is connected to the system set this
parameter to Normal. If there isn’t a balance connected to the system
set this parameter to Disabled.
Baud Rate: Sets the data transmission speed in bits per second.
Selections: 1200, 2400, 4800, or 9600 bps. Default: 1200 bps
Class: A two digit listing of all different groups of circuit boards that
are signed on to the network. Each group will have a distinct class
assigned.
E
Endline Time: The amount of time used to collect baseline value
after the peak. Collection begins after the comparator level has been
reached. Endline Time is used in conjunction with the baseline taken
before the peak. Default: Nitrogen – 2 seconds.
Equilibrate Time: The time period for gases in the ballast to mix
after it is filled.
Equilibrate Pressure Time: The time period to wait once the aliquot
loop is filled and isolated to let pressure in the aliquot loop to stabilize.
Furnace Filter Tube: Filters the gases leaving the furnace and
entering the thermoelectric cooler. Any particles leaving the furnace
are trapped in the filter.
Furnace Flow: The furnace gas flow rate. Selections are None, Low,
Medium, High and Maximum. Click to enter the desired flow rate.
Furnace Particle Filter: Removes dirt and particles that may not
have been trapped in the furnace filter.
Keep Login Samples until Cancel is Pressed: Check this box to log
consecutive standards. The dialog box will reappear after OK is clicked.
This is an optional entry.
M
Mass: The mass of the standard. Click Balance to enter a mass from
an external balance.
Method: The method used for analysis. Click the Down Arrow to
select from the list of methods.
Not Filled Time Out: If the ballast is not filled by the entered time
period, analysis will abort. This could be caused by a leak in the
system. This time period should be longer than the combined time of
the burn steps.
P
Parity: Sets the method of error checking used. Selections: Even,
Odd, or None. Default: Even
S
Sample Moisture: No entry should be made for standard login.
Send Command: The command sent from the instrument that the
balance reads to send and enter the weight in the spreadsheet. Default
<ESC>P<CR><LF>
Stop Bits: Used to select the number of bits that are used to signal
the end of a transmitted data byte. Selections: 1 or 2. Default: 1
Tare Command: The command sent from the instrument that the
balance reads to set the weight displayed on the balance to zero.
Default: <ESC> T
Time: The amount of time in seconds that the step will burn before
the instrument switches to the next step. Click to enter the desired
time.
A
Accessory Pack ..................................................................................................1–30
Add Standards for Calibration ..............................................................................5–33
Air Filter Cleaning...............................................................................................6–6
Aliquot Dosing Valve Cleaning ..............................................................................6–7
Ambient Chart ...................................................................................................8–3
Ambient Monitor ................................................................................................8–3
Ranges .............................................................................................................8–6
Analysis
Abort an Analysis .....................................................................................5–24
Cancel Pause ...........................................................................................5–29
Delay ......................................................................................................5–27
First Time ................................................................................................5–5
Pause .....................................................................................................5–28
Multiple Samples ......................................................................................5–24
Standard .................................................................................................5–23
Auto Widths.......................................................................................................4–30
B
Backup (Hardware) Calibration ............................................................................8–8
Balance Configuration .........................................................................................4–33
Balance Installation ............................................................................................3–3
Reset ......................................................................................................3–4
Setup .....................................................................................................3–3
Test Procedure .........................................................................................3–4
Ballast Oven Temperature Control ........................................................................8–15
Ballast Tank Cleaning .........................................................................................6–9
Barometric Calibration ........................................................................................4–35
Barometric Pressure ...........................................................................................7–9
Blank Analysis ...................................................................................................5–23
Blank Analysis, Theory ........................................................................................7–17
Blank Calibration ................................................................................................5–31
Blank Login .......................................................................................................5–17
Burn Profile .......................................................................................................4–36
D
Data
Backup ...................................................................................................5–69
Fill Cells with same ...................................................................................5–39
Transmit Installation.................................................................................2–11
Define Calibration Standards ...............................................................................4–47
Deleting Analyzed Samples..................................................................................4–78
Display Configuration ..........................................................................................4–20
Drift Correction ..................................................................................................5–34
Drift Factor, Theory ............................................................................................7–19
F
Facility Requirements .........................................................................................1–32
Furnace
Diagnostics..............................................................................................8–13
Temperature Adjustment ..........................................................................8–14
Temperature Control ................................................................................8–14
G
Gel Capsule .......................................................................................................5–13
H
Hardware Calibrations .........................................................................................8–7
Heating Element Installation ................................................................................2–3
How to Perform a Leak Check ..............................................................................8–16
I
Illustrations List .................................................................................................1–7
Import Text Data ...............................................................................................5–59
Infrared Radiation Cell, Theory ............................................................................7–12
Insert a Sample .................................................................................................5–39
Installation ........................................................................................................2–3
Installing LECO Software .....................................................................................9–4
Interference to Other Devices ..............................................................................1–20
International Installation .....................................................................................2–5
IR Cell Removal and Service ................................................................................9–36
IR Detector Linearization .....................................................................................9–37
M
Main Screen Overview ........................................................................................4–20
Maintenance Counters
Configure ................................................................................................6–46
Reset ......................................................................................................6–48
Maintenance Menu .............................................................................................4–13
Managing Databases ...........................................................................................4–59
Menu
Configuration ...........................................................................................4–11
Database ................................................................................................4–7
Diagnostics..............................................................................................4–13
Edit ........................................................................................................4–7
Maintenance ............................................................................................4–13
Overview.................................................................................................4–7
View .......................................................................................................4–8
Method
Clone ......................................................................................................4–42
Configuration ...........................................................................................4–36
Delete.....................................................................................................4–46
Edit ........................................................................................................4–41
Export ....................................................................................................4–43
Import ....................................................................................................4–44
Standard Additions ...................................................................................7–6
Micro Line Temperature Control ...........................................................................8–15
O
Operation ..........................................................................................................5–1
Options .............................................................................................................1–27
O-rings .............................................................................................................6–38
P
PC Tower Assembly ............................................................................................1–31
Periodic Maintenance Schedule ............................................................................6–4
Poll Devices .......................................................................................................8–12
Power ...............................................................................................................2–13
Primary Dust Filters Cleaning ...............................................................................6–41
Printer Installation..............................................................................................3–6
Printing Reports .................................................................................................5–45
Protein Results Display .......................................................................................5–41
R
Reagent Tube Packing.........................................................................................2–9
Reagent Tube Repacking .....................................................................................6–39
Recalculate Results.............................................................................................5–40
Reduction Heater Tube Packing ............................................................................6–19
Reduction Tube Temperature Control ....................................................................8–14
Replace Drift Standard ........................................................................................5–37
Reset Network ...................................................................................................8–12
S
Sample
Analysis ..................................................................................................5–24
Filters .....................................................................................................4–52
History ....................................................................................................5–42
Login ......................................................................................................5–18
Sample Preparation ..................................................................................5–13
Sample Signature .....................................................................................5–43
Sample Template ...............................................................................................4–49
Delete a Sample Template.........................................................................4–50
Login ......................................................................................................4–51
Samples ............................................................................................................5–39
Delete.....................................................................................................5–40
Menu ......................................................................................................4–9
Set Barometric Pressure ......................................................................................8–7
T
TC Cell Removal and Service ...............................................................................9–35
Theory of Operation............................................................................................7–1
Tin Foil..............................................................................................................5–14
Toolbar .............................................................................................................4–29
Transfer Export ..................................................................................................5–69
Transfer Import .................................................................................................5–71
Transmit Data ....................................................................................................5–60
Transmit Format Configuration.............................................................................5–60
Transmit Selected Samples .................................................................................5–68
U
Upgrade a Device ...............................................................................................8–12
User Defined Fields.............................................................................................4–56
User Permissions ................................................................................................4–16
V
View Calibration .................................................................................................5–33
View History ......................................................................................................5–36
View Log File .....................................................................................................6–45
View Signatures .................................................................................................5–44
W
Warranty...........................................................................................................1–16