Test Report: The EMC Testing Has Been Performed On The Submitted Samples
Test Report: The EMC Testing Has Been Performed On The Submitted Samples
Test Report: The EMC Testing Has Been Performed On The Submitted Samples
Description Page
Test Report Declaration……………………………..…………………………………4
1. GENERAL INFORMATION ....................................................................................................... 5
1.1. Description of Device (EUT) ................................................................................................................ 5
1.2. Operational Mode(s) of EUT................................................................................................................. 5
1.3. Test Voltage(s) of EUT ......................................................................................................................... 5
1.4. Compliance with the EN 60601-1-2:2015 chapter 6 requires. .............................................................. 6
2. DESCRIPTION OF TEST STANDARD ................................................................................... 15
3. LABORATORY INFORMATION............................................................................................. 16
3.1. Laboratory Name ................................................................................................................................. 16
3.2. Location ............................................................................................................................................... 16
3.3. Test facility .......................................................................................................................................... 16
3.4. Measurement Uncertainty.................................................................................................................... 16
4. SUMMARY OF TEST RESULTS.............................................................................................. 17
5. BLOCK DIAGRAM OF TEST SETUP ..................................................................................... 18
5.1. Block Diagram of connection between EUT and simulation-EMS..................................................... 18
6. TEST INSTRUMENT USED ...................................................................................................... 19
6.1. For Conducted Disturbance at Mains Terminals Emission Test ......................................................... 19
6.2. For Radiation Test(In Anechoic Chamber)(Below 1000MHz) ........................................................... 19
6.3. For Radiation Test (In Anechoic Chamber)(Above 1000MHz) .......................................................... 19
6.4. For Harmonic / Flicker Test ................................................................................................................ 19
6.5. For Electrostatic Discharge Immunity Test ......................................................................................... 20
6.6. For RF Strength Susceptibility Test .................................................................................................... 20
6.7. For Electrical Fast Transient/Burst Immunity Test ............................................................................. 20
6.8. For Surge Test ..................................................................................................................................... 20
6.9. For Injected Currents Susceptibility Test ............................................................................................ 20
6.10. For Magnetic Field Immunity Test ...................................................................................................... 20
6.11. or Voltage Dips and Interruptions Test ............................................................................................... 21
7. CONDUCTED DISTURBANCE AT MAINS TERMINALS TEST ....................................... 22
7.1. Configuration of Test System .............................................................................................................. 22
7.2. Test Standard ....................................................................................................................................... 22
7.3. Power Line Conducted Disturbance at Mains Terminals Limit .......................................................... 22
7.4. Test Procedure ..................................................................................................................................... 22
7.5. Conducted Disturbance at Mains Terminals Test Results .................................................................. 23
8. RADIATED DISTURBANCE TEST ......................................................................................... 24
8.1. Configuration of Test System .............................................................................................................. 24
8.2. Test Standard ....................................................................................................................................... 24
8.3. Radiated Disturbance Limit ................................................................................................................. 24
8.4. Operating Condition of EUT ............................................................................................................... 24
8.5. Test Procedure ..................................................................................................................................... 25
8.6. Radiated Disturbance Test Results ...................................................................................................... 25
9. HARMONIC CURRENT TEST ................................................................................................. 26
9.1. Configuration of Test System .............................................................................................................. 26
9.2. Test Standard ....................................................................................................................................... 26
9.3. Test Limits ........................................................................................................................................... 26
9.4. Test Results ......................................................................................................................................... 26
10. VOLTAGE FLUCTUATIONS & FLICKER TEST ................................................................ 27
10.1. Configuration of Test System .............................................................................................................. 27
10.2. Test Standard ....................................................................................................................................... 27
10.3. Test Limits ........................................................................................................................................... 27
APPENDIX I (2 pages)
APPENDIX II (Test Photos) (1 page)
APPENDIX III (Photos of the EUT) (5 pages)
Test Result The EUT was found compliant with the requirement(s) of the standards.
EN 60601-1-2:2015
Standard (CISPR 11:2015+A1:2016+A2:2019, IEC 61000-4-2:2008,
IEC 61000-4-3:2006+A1:2007+A2:2010,IEC 61000-4-8:2009)
*Note
The above device has been tested by Shenzhen Most Technology Service Co., Ltd. To determine the
maximum emission levels emanating from the device and the severe levels of the device can endure and
its performance criterion. The test record, data evaluation & Equipment Under Test (EUT)
configurations represented are contained in this test report and Shenzhen Most Technology Service Co.,
Ltd. Is assumed full responsibility for the accuracy and completeness of test. Also, this report shows that
the EUT is technically compliant with the requirement of the above standards.
This report applies to above tested sample only. This report shall not be reproduced except in full,
without written approval of Shenzhen Most Technology Service Co., Ltd., this document may be altered
or revised by Shenzhen Most Technology Service Co., Ltd.., personal only, and shall be noted in the
revision of the document.
Prepared by
Lili Lu(Engineer)
Reviewed by
Sunny Deng(Engineer)
Approved by
Yvette Zhou(Manager)
Remark : N/A
6.1.201.3 Making on the outside of EQUIPMENT and SYSTEMS that are N/A
specified for use only in a shielded location
EQUIPMENT and SYSTEMS specified for use only in a shielded N/A
location shall be labeled with a warning that they should be used only
in the specified type of shielded location (see 6.8.3.201 c))
Passed ACCOMPANYING DOCUMENTS Passed
6.8.201
6.8.2.201 Instructions for use Passed
a) Requirements applicable to all EQUIPMENT and SYSTEMS Passed
Instructions for use shall included the following:
1) a statement that MEDICAL ELECTRICAL EQUIPMENT needs Passed
special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in the
ACCOMPANYING DOCUMENTS;
2) a statement that portable and mobile RF communications Passed
equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
Report No. MTET20040455 Page 6 of 46
b) Requirements applicable to EQUIPMENT and SYSTEM for which N/A
the connector testing exemption specified in 36.202.2. b) 3) is used
*3) Table 201, with the modifications specified below.3.4 The Passed
flowchart in Figure 201 is the requirement in step-by-step
Report No. MTET20040455 Page 7 of 46
graphical form for completion of Table 201 for CISPR 11
EQUIPMENT and YSTEMS. The flowchart in Figure 202 is the
requirement in step-by-step graphical form for completion of
Table 201 for CISPR 14 and CISPR 15 EQUIPMENT.
- For CISPR 11 EQUIPMENT and SYSTEMS, “[EQUIPMENT
or SYSTEMS]” shall be replaced with the MODEL OR TYPE
REFERECE of the EQUIPMENT or SYSTEMS.
- For CISPR 14 and CISPR 15 EQUIPMENT,
“[EQUIPMENT]” shall be replaced with the MODEL OR
TYPE REFERECE of the EQUIPMENT.
- For CISPR 11 Group 1 EQUIPMENT and SYSTEMS, rows 5,
11 and 12 shall be deleted.
- For CISPR 11 Group 2 EQUIPMENT and SYSTEMS, rows 4,
11 and 12 shall be deleted.
- For EQUIPMENT that complies with CISPR 14-1, rows 4
through 6 and row 12 shall be deleted.
- For EQUIPMENT that complies with CISPR 15, rows 4
through 6 and row 11 shall be deleted.
- For CISPR 11 EQUIPMENT and SYSTEMS that comply with
Class A, “[A or B]” in column 2 of row 6 shall be replaced with
“A”. For CISPR 11 EQUIPMENT and SYSTEMS that comply
with Class B, “[A or B]” in column 2 of row 6 shall be replaced
with “B”.
- For EQUIPMENT and SYSTEMS that comply with IEC
61000-3-2. “[Class A, B, C, D, or Not applicable]” in column 2
of row 7 shall be replaced with the class of the EQUIPMENT
and SYSTEMS according to IEC 61000-3-2. For EQUIPMENT
and SYSTEMS that comply with IEC 61000-3-2. “[Complies or
Not applicable]” in column 2 of row 8 shall be replaced with
“Complies”. For EQUIPMENT and SYSTEMS for which IEC
61000-3-2 and IEC 61000-3-3 are not applicable, “[Class A, B,
C, D, or Not applicable]” and “[Complies or Not applicable]”
shall each be replaced with “Not applicable”.
- For CISPR 11 EQUIPMENT and SYSTEMS, column 3 of
rows 6, 7 and 8 shall be merged into one cell. For CISPR 11
EQUIPMENT and SYSTEMS that comply with Class B and
with IEC 61000-3-2 and IEC 61000-3-3, the text in column 3 of
row 9 shall be moved into the merged cell. For CISPR 11
EQUIPMENT and SYSTEMS that comply with Class A or
which IEC 61000-3-2 and IEC 61000-3-3 are not applicable,
the text in column 3 of row 10 shall be moved into the merged
cell.
- For CISPR 14 or CISPR 15 EQUIPMENT, column 3 of rows 7
and 8 shall be merged into one cell. For CISPR 14 or CISPR 15
EQUIPMENT that comply with IEC 61000-3-2 and with IEC
61000-3-3, the text in column 3 of row 9 shall be moved into
the merged cell. For CISPR 14 or CISPR 15 EQUIPMENT for
which IEC 61000-3-2 and IEC 61000-3-3 are not applicable,
the text in column 3 of row 10 shall be moved into the merged
cell.
- For EQUIPMENT and SYSTEMS specified for use only in a
shielded location and for which the electromagnetic radiation
disturbance allowance or the mains terminal disturbance
NOTE There are six places in Tables 207 and 208 where “[EQUIPMENT
or SYSTEM]” must be replaced.
The following referenced standard are indispensable for the application of this report.
EN 60601-1-2:2015
Medical electrical equipment-Part 1-2: General requirements for basic safety and essential
performance-collateral standard: Electromagnetic compatibility-Reqirements and tests.
3.2. Location
No.5, 2nd Langshan Road, North District, Hi-tech Industrial Park, Nanshan, Shenzhen,
Guangdong, China
Accredited by TIMCO
Registration Number: Q1460
March 28, 2010
EN 61000-3-3:2013+
Voltage fluctuations & flicker --- N/A
A1:2019
IMMUNITY (EN 60601-1-2:2015)
Performance
Test Item Basic Standard Results
Criteria
Electrostatic discharge (ESD) IEC 61000-4-2:2008 --- PASS
Radio-frequency, IEC61000-4-3:2006+A1:2007
--- PASS
Continuous radiated disturbance +A2:2010
EUT
DC 3V by
Batteries
EUT
DC 3V by
Batteries
The frequency range from 150kHz to 30MHz is checked. The test result are reported on
Section 7.5.
7.5.2. If the average limit is met when using a quasi-peak detector, the EUT shall be deemed
to meet both limits and measurement with average detector is unnecessary.
8.4.4.Let the EUT work in test mode (Running) and test it.
The bandwidth setting on the test receiver is 120 kHz.(frequency range from 30MHz to
1000MHz) and 1MHz (frequency range from 1000MHz to 6000MHz).
The frequency range from 30MHz to 6000MHz is checked. The test result are reported on
Section 8.6
.
8.6.4. The test data and the scanning waveform are attached within Appendix I.
Note:All test modes are performed, only the worst case is recorded in this report.
10.2.Test Standard
EN 61000-3-3:2013+A1:2019
10.3.Test Limits
The limits shall be applicable to voltage fluctuations and flicker ata the supply
terminal s of the equipment under test, the following limites apply:
the value of Pst shall not be greater than 1.0;
the value of Plt shall not be greater than 0.65;
the value of d(t) during a voltage change shall not exceed 3.3% for more than 500ms;
the relative steady-state voltage change, dc, shall not exceed 3.3%;
the maximum relative voltage change dmax, shall not exceed
a) 4% without additional conditions;
b) 6% for equipment which is:
Switched manually, or
Swithced automatically more frequently than twice per day, and also has either a delayed
restart(the delay being not less than a few thens of seconds), or manual restart, after a
power supply interruption.
c) 7%for equipment which is
Attended whilst in use (for example: hair dryers, vacuum cleaners, kitchen equipment
such as mixers, garden equipment such as lawn mowers, portable tools such as electric
drills), or
switched on automatically, or is intended to be switched on manually, no more than twice
per day, and also has either a delayed restart(the delay being not less than a few tens of
seconds) or manual restart, after a power supply interruption.
10.4.Test Results
10.4.1.Test Results: N/A
For EQUIPMENT and SYSTEM with multiple FUNCTIONS, the criteria apply to each,
parameter, and channel.
All immunity test item on this report are compliance with EN 60601-1-2 chapter 36.201 *j)
require.
12.2.Test Standard
EN 60601-1-2:2015 (IEC 61000-4-2:2008)
(Severity Level 4 for Air Discharge at 15KV,
Severity Level 4 for Contact Discharge at 8KV)
12.4.Test Procedure
12.4.1.Air Discharge:
The test was applied on non-conductive surfaces of EUT. The round discharge tip of
the discharge electrode was approached as fast as possible to touch the EUT. After each
discharge, the discharge electrode was removed from the EUT. The generator was
re-triggered for a new single discharge and repeated 20 times for each pre-selected test
point. This procedure was repeated until all the air discharge completed
12.4.2.Contact Discharge:
All the procedure was same as Section 12.4.1. except that the generator was
re-triggered for a new single discharge and repeated 50 times for each pre-selected test
point. the tip of the discharge electrode was touch the EUT before the discharge switch
was operated.
12.5.Test Results
12.5.1.Test Results: PASS
Gaps A PASS
Switch A PASS
Keys A PASS
HCP C PASS
Discharge was considered on Contact and Air and Horizontal Coupling Plane (HCP) and Vertical Coupling Plane
(VCP).
Reviewer :
3m
3m EUT
1.5
0.8 m
m
7m
13.2.Test Standard
EN 60601-1-2:2015 (IEC 61000-4-3:2006+A1:2007+A2:2010)
(Severity Level: 1 at 3V/m)
13.5.Test Results
13.5.1.Test Results: PASS
Reviewer :
14.2.Test Standard
EN 60601-1-2:2015 (IEC 61000-4-4:2012)
(Severity Level 3 for power line at 2KV,
Severity Level 2 for input/output port at 1KV)
14.4.2.The EUT was connected to the power mains by using a coupling device which
couples the EFT interference signal to AC power lines. Both positive transients and
negative transients of test voltage was applied during compliance test and the duration
of the test can’t less than 2mins.
14.5.Test Results
14.5.1.Test Results: N/A
15.2.Test Standard
EN 60601-1-2:2015 (IEC 61000-4-5:2014+A1:2017+A1:2017)
(Severity Level : Line to Line was Level 2 at 1KV
Line to PE was Level 3 at 2KV)
15.4.2.For line to line coupling mode, provide a 1KV 1.2/50us voltage surge (at
open-circuit condition) and 8/20us current surge to EUT selected points, and for
active line / neutral line to ground are same except test level is 2KV.
15.4.3.At least 5 positive and 5 negative (polarity) tests with a maximum 1/min repetition
rate are applied during test.
15.4.5.Record the EUT operating situation during compliance test and decide the EUT
immunity criterion for above each test.
15.5.Test Results
15.5.1.Test Results: N/A
16.2.Test Standard
EN 60601-1-2:2015 (IEC61000-4-6:2013)
(Severity Level 2 at 3V (r.m.s.) and frequency is from 0.15MHz to 80MHz)
16.4.3.The EUT are placed on an insulating support 0.1m high above a ground reference
plane. CDN (coupling and decoupling device) is placed on the ground plane about
0.3m from EUT. Cables between CDN and EUT are as short as possible, and their
height above the ground reference plane shall be between 30 and 50 mm (where
possible).
16.4.5.The EUT operates within its operational mode(s) under intended climatic conditions
after power on.
16.4.6.The frequency range is swept from 0.150MHz to 80MHz using 3V signal level, and
with the disturbance signal 80% amplitude modulated with a 1KHz sine wave.
16.4.7.The rate of sweep shall not exceed 1.5*10-3decades/s. Where the frequency is swept
incrementally, the step size shall not exceed 1% of the start and thereafter 1% of the
preceding frequency value.
16.4.8.Recording the EUT operating situation during compliance testing and decide the EUT
immunity criterion.
16.5.Test Results
16.5.1.Test Results: N/A
EUT
0.1 m thickness
Insulating Support
Magnetic Field
Tester
17.2.Test Standard
EN 60601-1-2:2015 (IEC 61000-4-8:2009)
(Severity Level 4 at 30A/m)
1. 1
2. 3
3. 10
4. 30
5. 100
X. Special
17.5.Test Results
17.5.1.Test Results: PASS
Reviewer :
18.2.Test Standard
EN 60601-1-2:2015 (IEC 61000-4-11:2004+A1:2017)
(Severity level: 5% 250 periods
5% 0.5 period
40% 5 periods
70% 25 periods )
<5 >95 1
70 30 25
18.5.Test Results
18.5.1.Test Results: N/A
Compliance Criteria for Voltage Dips, Short Interruptions and Voltage Variations:
refer to EN 60601-1-2:2015 chapter 36.202.7.
Equipment and system with a rated input power of 1KVA or less and all
life-supporting equipment and systems shall comply with the requirements of
36.202.1j
Figure 2
General Appearance of the EUT
Figure 3
Inside of the EUT
Figure 4
Components side of the PCB
Figure 5
Components side of the PCB
Figure 6
Components side of the PCB
Figure 7
Components side of the PCB
Figure 8
Components side of the PCB
Figure 9
Components side of the PCB
Figure 10
Battery