Section B Installation
Contents
1. INTRODUCTION .....................................................................................................................................2
3. DELIVERY INFORMATION.....................................................................................................................3
3.1. Shock indicators ...................................................................................................................................3
3.2. Dimensions and weights ......................................................................................................................3
3.3. Transportation ......................................................................................................................................4
1. INTRODUCTION
These installation instructions assume that the scanner room has been prepared as described in the
Philips CT Aura Planning Reference Manual.
The Planning Reference Manual (PRM) contains all the information that is necessary to prepare the
installation of a Philips CT Aura. It describes items like room layout, room construction requirements,
power requirements, radiation safety, insulation etc.
This PRM must be used to insure a proper preparation for installation.
Installation of a CT AURA requires two engineers for a period of three days. Prerequisite is a properly
prepared scannerroom.
All mounting hardware necessary for installation (except standard tools) is delivered with the system.
Mains cable, doorswitch and/or external X-ray ON indicator (including necessary wiring) must be
provided locally.
Tools, necessary to install and/or maintain the CT AURA are listed below: 12NC
(Not part of the system delivery)
• Transport kit 4522 501 52341
• Printer test tool 4522 980 38563
• Monitor adjustment tool 4522 980 38584
• Template gantry & table 4522 501 52541
• Service PC
• DSS + Electronic documentation 4522 210 42204
• Standard toolkit
• Electrostatic discharge (ESD) kit
Transport kit, printer test tool, monitor adjustment tool and the templates can be shared for several
systems.
Power cables can be cut to length if wanted. Cable tags and a set of labels is delivered with the
system, however the tools to attach the cable tags must be supplied locally.
Since the crates, in which the system is packed, contain quite some screws an electrical screwdriver is
advised.
3. DELIVERY INFORMATION
The four major component assemblies of the CT AURA (gantry, patient table, operator’s console and
isolator rectifier) are packed for transport via truck, air or rail. Each container is equipped with shock
indicators.
Shock indicators must be checked before the crates are unloaded from the truck in order to define
responsibilities in case of damage to the system.
NOTE
Crate / Box Length (mm) Width (mm) Height (mm) Weight (Newton)
Gantry 2251 1270 2232 11368
Patient table 3003 819 1073 4900
Isolator/rectifier 920 765 1283 3094
Integrated desk 1620 810 900 1882
Computer cabinet 870 670 850 461
Monitors 558 498 485 324
During unpacking gantry and patient table are rolled off their pallets via a ramp. This procedure
requires at least 4.5 m (15 feet) floorspace in addition to the footprint of the respective pallets.
3.3. TRANSPORTATION
Check the transport route for the equipment from the receiving area to the installation area with
respect to the size of:
- Doors
- Passage ways
- Ramps
- Elevators
Take into account the size of moving equipment. If an elevator is to be used, check its weight handling
capabilities.
The minimum required door width is 92cm (3 ft), in this case the passage width must be 230 cm (7.5 ft)
in order to manoeuvre the gantry through the door.
The minimum required passage width is 150 cm (5 ft), in this case the door width must be 150 cm(5 ft
4.1. INTRODUCTION
Since both, table and gantry use the same set of transport wheels, unpacking and siting is one,
combined operation.
First the gantry is sited then the table as described in keywords in chapter 4.4.
4.2. PREPARATION
The table must be secured to the floor. A template, indicating the position of the holes, is delivered
with the system. (In table crate)
To prevent dirt and damage the crates should be unpacked outside the scanner room. It is highly
recommended to have help from professional movers during unpacking and carrying-in the equipment
into the hospital.
The operators console and isolator/rectifier do not require a specific unpack procedure, the
unpackprocedure for gantry and patient table is described below.
Crate 1: Gantry.
Crate 2: Patient table, installation kit, and template.
Crate 3: Cables, service tools, and covers.
Crate 4: Isolator rectifier.
Crate 5: Operator console (Back-end) components and cables.
Cardboard crate: Operator console (Back-end) Table and patient table isolation kit.
Miscellaneous boxes: Chair, modem, modem cable and patient accessories.
NOTE
The gantry and patient table is rolled off their pallets via a ramp. This procedure requires at least 4.5 m
(15 feet) in addition to the footprint of the respective pallets.
________________
RAMP (2)
RAMP (2)
PALLET
TRANSPORT WHEELS (5)
SPACE NEEDED ADDITIONAL TO FOOTPRINT
OF GANTRY PALLET = 4.5 METER
TOTAL SPACE NEEDED = 7 METER
EP04.DRW
WARNING
• Check crate drop ‘n tell (10g & 25g) and other signs of damage.
• Remove the top of the gantry crate.
• Remove the crossbars on the top and bottom of the ramp end of the crate. (End without the
inspection hole) See figure 4-3.
• Remove the ramp.
• Remove the screws holding the two middle ramp supports in their shipping position.
• Fold down the two ramp supports and secure them with screws in the holes marked with
bulls eyes.
NOTE
The isolator rectifier requires a clearance of 60cm in front (Cable connection side) for service (Air filter)
The other sides require a clearance of 20cm for cooling.
________________
• Remove wheels.
CAUTION
• Slowly roll gantry off the pallet. Push only on end stands not the “donut” part of the gantry.
• Locate gantry near desired location.
NOTE
1 2
Figure 4-12 Two bottom transport blocks (1) and one of the three front transport locks (2)
If the table does not fit in the elevator as it is, it can be partly disassembled according the following
instructions. It reduces table length to 244 cm.
A Remove pallet.
Slide the carriage to its extended position.
Unfasten the two screws that connect the pallet to the carriage, located in the arms of
the carriage (1 each side).
Separate the pallet from the carriage. Slide the pallet slightly forward and lift up.
B Remove support frame cover.
Remove the two screws from the inside of the support frame, and the two screws from
the rear.
Slide the cover out from under the carriage and remove.
C Remove support frame extension.
Remove the tension screw. (t)
Slide the carriage to its extended position. (If not already there)
(2)
(4)
(r)
(s)
EP05.DRW
Remove the rear carriage stops and carriage stop washers (s).
Remove the round way extensions and the support rail extension (r).
Disconnect the limit switch (other side).
Remove the support frame extension.
Mount the rear carriage stops and carriage stop washers to the remaining ways.
(to prevent the carriage from accidentally sliding off during transportation of the patient
table)
Remove nose cover.
Remove pallet roller.
TABLE FRAME
4 MM
NOSE ROLLER
M8 SCREWS
EP06.DRW
Software package:
− System Software on CD-ROMs
− Release bulletin.
− One 3½” floppy disk with license key and Kermit communication software.
− Instructions for use (operator’s manual)
After unpacking the crates, the items delivered should be checked for the delivery completeness.
For the hospital table and Philips delivered computer cabinet (trolley):
Just one person can do the installation.
Installation time 2 hours without software loading.
NOTE
The main power for the Operators Console is received from the isolator rectifier.
________________
Monitor Monitor
Power Distribution
Unit
Image-Handler
SUN
Scan Controller Ultra 10
SUN
Ultra 10
Power Distribution
Unit
Peripheral 2
Peripheral 2 cabinet
cabinet
(CD
(Image Disk recordable)
Dial box
User control
interface
Isolator
rectifier
(Trolley box)
PHILIPS CT AURA
Technical Service Manual CT AURA
A schematic diagram for the signal distribution cables is given in figure 5-2 The Ethernet cables used
between image handler, scan controller and front-end are UTP (Unshielded Twisted Pair) category 5
CD-Recordable
User-Control CT-MR Drive
Monitor Interface Monitor Dial Box
(ScanController) (Desktop) Peripheral 2 cabinet (ImageHandling)
Video cable
Video cable
Interface cable
GANTRY SCSI cable RS232 Serial cable
(Trolley box)
ScanController ImageHandling Image Disk
RS 232 SCSI A SCSI cable
HME 1 Serial B
01.2
UTP Cable
HME 0 UTP Cable
Peripheral 2 Cabinet
HME 0
Serial A HME 1
UTP Cable
Modem
Keyboard Cable
Keyboard cable
Hospital
Network
Keyboard Keyboard
Mouse Mouse
Ethernet cables.
Section B
5.3.2.
B - 16
Figure 5-2 Signal cable connections operator console.
Technical Service Manual CT AURA Section B
• Connect the analog Remote Service Modem to the serial port A of the Scan Control. The power
can be connected to the hospital mains (if the modem is delivered with a 110V adapter) or to the
power distribution unit (only for 230 Volts adapter!)
Regarding Physical planning and installation, the following basic issues must be taken into account:
• The location of the modem and related cable routing to telephone network and system.
• Because some types of hospital switchboards may cause problems in the application of modem
connections, we recommend to check whether a direct line to the public telephone network is
required.
CAUTION
Be sure that the input voltage for the remote service modem is correct.
(See modem for correct input voltage).
________________
NOTE
The scan controller keyboard can be recognized by stickers on the F9, F10, F11, “+” and “enter” keys.
In case there are no stickers on the keyboard, the keyboard must be modified using the stickerplate
(4522 142 3628x) F9 – Prep, F10 – Start, F11 – Stop, “+” – Move Table, “Enter” – Move Gantry
________________
PHILIPS CT AURA 01.2 B - 17
Copyright © 2000 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section B Technical Service Manual CT AURA
• Connect power cords between every unit and the mains distribution units as indicated in
figure 5-4. The power distribution unit consists out of two units connected to each other.
(See also figure 5-1)
NOTES
The power distribution unit is connected to the power outlet of the trolley box from the User Control
Interface.(See Figure 5-1)
________________
MONITOR MONITOR
TABLETOP
VIDEO SPLITTER
(OPTIONAL)
PERIPHERAL 2
CABINET
1 2 3 4 1 2 3 4
UCI
TROLLEYBOX
MAINS
DISTRIBUTION
GANTRY
• Connect a earth cable (yellow/green) from package 4522 142 36272 between the UCI box and the
major earth point on the back of the cabinet.
• Connect the signal cables between the units as indicated in figure 5-5. For details see also Z2-1
and Z2-2 (Section Z drawings)
NOTE
• The SCSI cabling must be terminated at the peripheral 2 cabinets. For this purpose terminators
are delivered with the system.
MONITOR MONITOR
TABLETOP
SCSI
PERIPHERAL 2
CABINET
1 2 3 4 1 2 3 4
UTP TO GANTRY
GANTRY
UCI
TROLLEYBOX
EP08.DRW
NOTES
The scan controller keyboard can be recognized by stickers on the F9, F10, F11, “+” and “enter” keys.
In case there are no stickers on the keyboard, the keyboard must be modified using the stickerplate
(4522 142 36281). F9 – Prep, F10 – Start, F11 – Stop, “+” – Move Table, “Enter” – Move Gantry
________________
• Connect power cords between every unit and the mains connection units as indicated in figure 5-7
NOTES
The PDU (Power Distribution Unit) is connected to the power outlet of the User Control Interface.
The on/off detection’s (PCI boards) in the scan controller and image handler) are connected to the
trolleybox of User Control Interface.
________________
MONITOR MONITOR
DIAL-BOX
(OPTION)
PERIPHERAL 2
CABINET UCI
DESKTOP
MOUSE MOUSE
VIDEO
SPLITTER
(OPTIONAL)
UCI
TROLLEYBOX
ISOLATOR
PERIPHERAL 2 RECTIFIER
CABINET
1 2 3 4 1 2 3 4
• Connect a short earth cable (yellow/green) from package 4522 142 36272 between frontcover and
the shelf of the metal cabinet.
• Connect a short earth cable (yellow/green) from package 4522 142 36272 between shelf and
major earth point at the back of the metal cabinet.
• Connect the long earth cable (yellow/green) from package 4522 142 36272 between UCI box and
the major earth point at the back of the metal cabinet.
• Connect the signal cables between the units as indicated in figure 5-8 For details see also Z2-1
and Z2-2 (Section Z drawings)
• A twisted pair Ethernet connection should be established between scancontrol and front-end and
between scan control system and image handler system. (Both cables must be crossed!)
NOTE
• The SCSI cabling must be terminated at the peripheral cabinets. For this purpose terminators are
delivered with the system.
MONITOR MONITOR
DIAL-BOX
(OPTION)
PERIPHERAL 2
CABINET UCI
DESKTOP
VIDEO VIDEO
MOUSE MOUSE
PERIPHERAL 2
CABINET GANTRY
1 2 3 4 1 2 3 4 UCI
TROLLEYBOX