TestingMethodsForFertilizers2018 PDF
TestingMethodsForFertilizers2018 PDF
TestingMethodsForFertilizers2018 PDF
June , 2018
Incorporated Administrative Agency
Food and Agricultural Materials Inspection Center
General Director
Makoto Kimura
Testing Methods for Fertilizers (2018)
Contents
- vii -
Testing Methods for Fertilizers (2018)
1. General Rule±
1.1 Common Items
(1) Applicable range
The Testing Methods for Fertilizers stipulate the official method of analysis of fertilizers and
fertilizer materials. The type of samples in the tests is shown in the summary of respective test
items.
(2.2) General matters and terms that cite Japanese Industrial Standards (JIS)
a) General notices: General matters common to analysis are according to JIS K 0050.
b) Definition: The definitions of major terms used in the Testing Methods for Fertilizers are
according to JIS K 0211, JIS K 0214, JIS K 0215, JIS Z 8101-1, JIS Z 8101-2 or JIS Z
8101-3.
c) Laboratory sample: A sample transferred to a laboratory. Laboratory sample as specified in
JIS K 0211.
d) Test sample: A sample obtained from a laboratory sample after pretreatment such as grinding.
Test sample as specified in JIS K 0211.
e) Analytical sample: A sample measured from a laboratory sample or a test sample and to be
used in one test. Test sample or analytical sample as specified in JIS K 0211.
f) Sample: A sample in the testing methods indicates e) a laboratory sample, f) a test sample or
g) an analytical sample.
g) Rounding numbers: Methods of rounding the numbers are according to JIS Z 8401.
h) Absorptiometric analysis: General rules for absorptiometric analysis are according to JIS K
0115.
i) Atomic absorption spectrometry: Atomic absorption spectrometry includes flame atomic
absorption spectrometry, electrically heated atomic absorption spectrometry (hereinafter
1
Testing Methods for Fertilizers (2018)
(2.3) Description methods, procedures and terms in testing methods for fertilizers.
a) Reagent name: Unless otherwise specified, conform to the names by the chemical
nomenclature established by the Chemical Society of Japan [in accordance with the
International Union of Pure and Applicable Chemistry (IUPAC) nomenclature of inorganic
chemistry and nomenclature of organic chemistry] and the names of JIS reagents.
b) Organic matters: Fertilizers such as Organic fertilizers, sludge fertilizers and compost and
fertilizer materials. However, organic compounds such as urea and urea compounds are
excluded.
c) Actual article: A laboratory sample in original state.
d) Drying matter: The matter which remains after drying the actual article.
e) Notes, comments, figures, tables and formulae: Serial numbers for each test item should be
recorded in notes, comments, figures, tables and formulae.
f) Dilution of solution: “Transfer accurately a predetermined amount (to a vessel)” means the
procedure to measure any volume of solution with a measuring instrument specified in JIS K
0050 (into a vessel).
Also, “dilute accurately a predetermined amount (with solvent or solution)” means the
procedure to measure any volume of solution with a measuring instrument specified in JIS K
0050 into a volumetric flask of arbitrary volume and fill up to the marked line (with solvent or
solution) (1).
g) Description of mixture solution: Mixture solutions are described as shown in 1) - 4).
1) Reagent + reagent: Describe as reagent name 1−reagent name 2 (V1 + V2). In this case, V1
volume of reagent name 1 is mixed with V2 volume of reagent name 2.
Example: acetonitrile−water (1+1), hexane−ethyl acetate (2+1), methanol−buffer solution
(3+1)
2) Reagent + water: Describe as reagent name 1 (V1+V2). In the case of reagents described in
Table 1 in JIS K 0050, it means V1 volume of reagent name 1 is diluted by mixing with V2
volume of water.
Example: hydrochloric acid (1+1), sulfuric acid (1+2), ammonia solution (1+3)
2
Testing Methods for Fertilizers (2018)
Note (1) When the dilution factor is large, accuracy should be secured by procedures such as
repeating the dilution procedure.
(2) Set according to the specification and operation method of the measurement instrument
used. There is no need to include the minimum and the maximum values of the
calibration curb range described in the Testing Methods for Fertilizers.
(3) Water
a) Water: Water used in the Testing Methods for Fertilizers herein is water of A2 specified in JIS
K 0557 or water that is confirmed not to affect a quantitation value. However, when otherwise
specified in respective test items, use the specified water.
(4) Reagents
a) Reagents: When the reagent is JIS-specified, use one of highest quality among those marked
with the JIS symbol; when none of the reagent is marked with the JIS symbol, use one of
quality that will not cause a problem in the test. Use reference materials for volumetric
analysis specified in JIS K 8005 for the standardization of titration solutions.
b) Reference materials: The preparation of standard solutions or standardization of titration
solutions using reference materials 1) - 2) below other than materials specified in respective
testing items is possible.
3
Testing Methods for Fertilizers (2018)
(5) Apparatus
a) Glass apparatus: Unless otherwise specified, use glass apparatus specified in JIS R 3503 and
JIS R 3505. Also, when a heating procedure is involved, use borosilicate glass-1 specified in
JIS R 3503.
b) Non-glass apparatus: Unless otherwise specified, use plastic apparatus.
c) Desiccants for desiccators: Unless otherwise specified, use silica gel.
d) Porcelain crucibles and porcelain evaporating dishes: Use ones specified in JIS R 1301
and JIS R 1302.
e) Platinum crucibles and platinum evaporating dishes: Use ones specified in JIS H 6201 and
JIS H 6202.
f) Filter paper: Use that specified in JIS P 3801. However, the type of filter paper is specified in
respective test items.
g) Absorbance measurement (absorptiometric analysis) absorbance cells: Unless otherwise
specified, use ones of 10 mm in optical path length.
5
Testing Methods for Fertilizers (2018)
6
Testing Methods for Fertilizers (2018)
7
Testing Methods for Fertilizers (2018)
Note (1) Refer to the separate sheet: “Target of trueness and criteria of precision in respective
concentration levels” or the reproducibility of respective testing methods” in order to
determine if there is mutual agreement.
8
Testing Methods for Fertilizers (2018)
2. Handling of samples
2.1 Sampling
Sampling is according to “2.1 Sampling in the Official Methods of Analysis of Fertilizers (1992)”.
References
1) National Institute for Agro-Environmental Sciences, the Ministry of Agriculture,
Forestry and Fisheries: The Official Methods of Analysis of Fertilizers 1992, p.4 - 5,
Japan Fertilizers Analysis Association, Tokyo (1992)
2) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p. 12 - 17, Yokendo, Tokyo (1988)
9
Testing Methods for Fertilizers (2018)
(2) Apparatus and instruments: Apparatus and instruments are shown below.
a) Refrigerator: A refrigerator that can be adjusted to 1 ºC - 8 ºC.
b) Storage container for a sample: A storage container for a sample should be clean, strong and
completely sealed airtight. In particular, in case it contains sludge for raw materials, the
container should be made of non-degradable, non-absorbable material. Additionally, it should
be airtight, water-proof, vapor-proof and non-corrosive.
10
Testing Methods for Fertilizers (2018)
References
1) JIS M 8100: Particulate materials - General rules for methods of sampling (1992)
2) JIS K 0060: Sampling method of industrial wastes (1992)
11
Testing Methods for Fertilizers (2018)
2.3.1 Pre-drying
(1) Summary
This procedure is applicable to fertilizers whose laboratory sample is moist and hard to grind.
The symbol of the procedure is 2.3.1-2017 or PD.-1.
Conduct pre-drying using a drying apparatus, and measure the loss on drying in this procedure.
In addition, calculate a conversion factor (actual article), if necessary, to convert the component
content obtained in respective tests to the component content in a laboratory sample (actual article).
(2) Apparatus and instruments: Apparatus and instruments are shown below.
a) Drying apparatus: A drying apparatus that can be adjusted to the pre-drying temperature at
± 2 ºC.
b) Sample drying dish: Measure the mass to the order of 0.1g in advance. Additionally, use
materials of a quality that do not affect the measurement of test components.
Note (1) Examples of drying temperature and drying time: About 70 hours at 40 ºC, no less than
5 hours at 65 ºC.
(2) An example of time to leave at rest: About 20 minutes
Comment 1 When calculating major components in a laboratory sample (actual article) such as
compost and sludge fertilizers, etc. where the test sample is prepared conducting
pre-drying, convert component contents in the analytical sample obtained in
respective tests by the following formula.
References
1) Mariko AIZAWA, Yuji SHIRAI, Yasushi SUGIMURA, Yuichi TAKAHASHI, Jun OKI,
Yukio FUKUCHI and Norio HIKICHI: Evaluation of Pre-drying Procedure to Prepare
Test Samples from Sludge Fertilizer, Research Report of Fertilizer Vol.1, p. 122 - 128
(2008)
12
Testing Methods for Fertilizers (2018)
(4) Flow sheet for pre-drying The flow sheet for the pre-drying of a moist laboratory sample is
shown below.
13
Testing Methods for Fertilizers (2018)
(2) Apparatus
a) Scoop for increment reduction: A scoop for increment reduction specified in the attached
chart 1 of JIS M 8100.
b) Riffle sampler: A riffle sampler specified in the attached chart 3 of JIS M 8100.
14
Testing Methods for Fertilizers (2018)
2.3.3 Grinding
(1) Summary
This procedure is applicable to fertilizers. The symbol of the procedure is 2.3.3-2017 or GRD.-1.
In order to prepare a homogeneous test sample, grind a laboratory sample with an adequate grinder
until it completely passes through the designated granularity.
(2) Apparatus and instruments: Apparatus and instruments are shown below.
a) Grinder: Use a grinder (3) of a type and suitability for the granularity and the physical
characteristics (1) of a laboratory sample. In addition, the grinder apparatus which come into
contact with a laboratory sample should be made of materials (2) which do not affect the
analytical value.
b) Primary grinder: A grinder (3) that can primarily grind a large lump.
c) Cutter machine: A cutter that can cut long stems, etc.
d) Sieve: A sieve for the test specified in JIS Z 8801-1 or JIS Z 8801-2 or equivalents.
Note (1) The physical characteristics of a laboratory sample are defined by their solidity,
toughness, specific gravity and adhesiveness.
(2) (Ex.) Do not use stainless steel apparatus when preparing a test sample for chromium
or nickel.
(3) A centrifuging type grinder, a cutting mill, a vibration mill type grinder, etc.
(4) A blender with an attachable blade, etc.
Comment 1 If the sampling amount of an analytical sample is less than 1 g, use a test sample
which will completely pass through 500 µm aperture sieve. In addition, in case a test
sample which suits the aforementioned condition cannot be obtained due to a
deliquescent laboratory sample, etc., make one by crushing a test sample with a
mortar and pestle, which completely passes through 1 mm aperture sieve.
(3.2) Fused phosphate fertilizer, slag silicate fertilizer, etc.: Conduct as specified in 6.4 of JIS M
8100 and as shown below.
a) Grind a laboratory sample with a vibration mill type grinder.
b) Transfer the ground laboratory sample to a sieve of 212 µm apertures.
c) Incline the sieve about 20 degrees, supporting it with one hand or a bent arm, and tap the
sieve frame with the other hand at the rate of about 120 times per minute. During the
procedure, place the sieve in a horizontal position at the rate of 4 times per minute, rotate it 90
degrees and tap the sieve frame firmly one or two times.
d) When fine powder attaches to the back side of a sieve screen, remove it gently from the back
side to make minus sieve.
e) Regarding the plus sieve of a sample, make them pass through by repeating the procedure in
a) - d).
f) Combine and mix the sample passed to make the test sample.
15
Testing Methods for Fertilizers (2018)
Comment 2 Conduct the procedures in (3.2) to obtain observed value of citric acid-soluble main
components in a stable manner. For examples of applicable laboratory samples, fused
matters such as fused phosphate and slag silicate fertilizer, fertilizers made from
fused materials and calcined phosphate, etc. are listed.
Comment 3 The procedures in b) - d) are the procedures in 6.1.3 (1.4) of JIS Z 8815.
16
Testing Methods for Fertilizers (2018)
3. General tests
3.1 Moisture or moisture content
3.1.a Loss on drying method with drying apparatus
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type A (Def-M)
and its symbol is 3.1.a-2017 or Mois.a-1.
Use drying apparatus under conditions suitable for the kind of fertilizers to be measured to heat
analytical samples to measure loss on drying and obtain moisture in an analytical sample or the
moisture content of a quality labeling standard of a special fertilizer (herein after referred to as
“moisture”). Additionally, calculate a conversion factor (drying matter) to convert component
content obtained by respective tests to component content in a drying matter as necessary.
This testing method corresponds to loss on heating in the Official Methods of Analysis of Fertilizers
(1992).
(2) Apparatus and instruments: Apparatus and instruments are shown below.
a) Drying apparatus: Drying apparatus that can be adjusted to the test temperature ± 2 ºC.
b) Ground-in stoppered weighing bottles (1): Low-form weighing bottles, 50 mm × 30 mm,
specified in JIS R 3503. Dry by heating in advance in a drying apparatus at 75 ºC - 130 ºC,
stand to cool in a desiccator, and measure the mass to the order of 1 mg.
Note (1) Aluminum weighing dishes described in the Handbook of the Feed Analysis Standards
-2009- can also be used.
Note (2) Heat simultaneously the slightly moved or removed stopper of the ground-in stoppered
weighing bottle.
Comment 1 When pre-drying a laboratory sample such as compost and sludge fertilizers to
prepare a test sample, calculate the moisture of the laboratory sample (actual
article) by the following formula:
17
Testing Methods for Fertilizers (2018)
B: Loss on drying (% (mass fraction)) of the laboratory sample (actual article) by the
pre-drying procedure
C: Loss on drying (% (mass fraction)) in the analytical sample by moisture
measurement
Comment 2 When calculating harmful content in a drying matter of sludge fertilizers, etc.,
convert component content in a test sample obtained from respective tests by the
following formula.
Comment 3 Use drying conditions in Table 1 for fertilizers of the types shown below:
Comment 4 For samples containing volatile matters, subtract the volatile matter content by the
following a) and b) from loss on drying to obtain moisture.
a) Fertilizers containing guano or diammonium hydrogen phosphate, etc.: Determine
total nitrogen in the test sample, and in the analytical sample after the drying
procedure; convert the difference between the quantitation values into ammonia
(NH3) to make the volatile matter content.
b) Potassium hydrogen carbonate: Determine carbon dioxide in the test sample, and in
the analytical sample after the drying procedure; the difference between the
quantitation values is the volatile matter content.
18
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p. 20 - 23, Yokendo, Tokyo (1988)
2) Society for the Study of Feed Analysis Standards: Methods of Analysis of Feeds and
Feed Additives-2009-I, p. 37 - 39, Incorporated Administrative Agency Food and
Agricultural Materials Inspection Center, Saitama (2009)
3) JIS Z 0701: Silica gel desiccants for packaging (1997)
(4) Flow sheet for moisture: The flow sheet for moisture in fertilizers is shown below:
2 g -5 g Transfer to a ground-in stopper weighing bottle, spread so that the
analytical sample thickness is no more than 10 mm. Measure the mass to the order of 1 mg.
Figure Flow sheet for moisture in fertilizers by loss on drying with drying apparatus.
19
Testing Methods for Fertilizers (2018)
Note (1) There is a method to calibrate with calibration weights or a method to calibrate
automatically with built-in weights.
Note (2) The setup of the drying program and the determination parameter for the end of heating
(constant weight) is according to the specification and the operation method of the
moisture analyzer used.
Comment 1 When pre-drying is conducted, calculate the moisture of the laboratory sample
(actual article) by the following formula:
B: Loss on drying (% (mass fraction)) of the laboratory sample (actual article) by the
pre-drying procedure
C : Loss on drying (% (mass fraction)) in the analytical sample by moisture
20
Testing Methods for Fertilizers (2018)
measurement
Comment 2 When calculating harmful content in a drying matter of sludge fertilizers, etc.,
convert component content in a test sample obtained from respective tests by the
following formula.
Comment 3 Table 1 shows the results of the comparison of the measurement values by loss on
drying method with drying apparatus and the measurement values by loss on drying
method with a moisture analyzer using organic fertilizers, compost and sludge
fertilizers in order to evaluate trueness.
Table 2 shows results and analysis results from a collaborative study for testing
method validation.
21
Testing Methods for Fertilizers (2018)
(4) Flow sheet for moisture: The flow sheet for moisture in sludge fertilizers, compost, and
organic fertilizers, etc. is shown below:
22
Testing Methods for Fertilizers (2018)
(2) Apparatus and instruments: Apparatus and instruments are shown below.
a) Electric furnace: Electric furnace that can be adjusted to 550ºC ± 5 ºC.
b) Crucible: After heating a porcelain crucible for chemical analysis specified in JIS R 1301
with an electric furnace at 550 ºC ± 5 ºC, stand to cool in a desiccator in advance and measure
the mass to the order of 1 mg.
Note (1) Example of carbonizing and incineration procedure: After raising the temperature
from room temperature to about 250 °C in 30 minutes to 1 hour, continue heating
for about 1 hour and further raise to 550 °C in 1 to 2 hours.
(4) Flow sheet for ash content: The flow sheet for ash content in fertilizers is shown below:
23
Testing Methods for Fertilizers (2018)
3.3 pH
3.3.a Glass electrode method
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type A (Def-M)
and its symbol is 3.3.a-2017 or pH.a-1.
Measure the pH of fertilizers with a pH meter using a glass electrode.
Comment 1 Respective pH standard solutions stored for long time should not be used since the
pH value may change during storage period. In particular, note that borate pH
standard solution and carbonate pH standard solution easily absorb carbon dioxide in
the air, so that the pH values deteriorate.
The pH standard solution that was used once or left exposed to the air should not be
used.
Comment 2 Conduct the calibration of a pH meter as indicated in JIS Z 8802. Actual calibration
operation is according to the operation procedure of the pH meter used for
measurement.
When the pH of a sample solution is no more than 7, use neutral phosphate pH
standard solution and oxalate pH standard solution, or phthalate pH standard solution.
When it exceeds 7, use neutral phosphate pH standard solution and borate pH
standard solution, or carbonate pH standard solution.
Note (1) In the case of a moist laboratory sample, it is recommended to use a sample that is not
pre-dried.
24
Testing Methods for Fertilizers (2018)
Comment 3 The procedure in (4.1.1) is the same as 3.4.a (4.1). Additionally, the sample
solution prepared in 4.2.4.a (4.1) can be used instead of the sample solution
prepared by (4.1.2).
Note (2) It is necessary to transfer sufficient volume of sample solution to keep a measurement
value stable.
Comment 4 If a pH meter has a temperature correction dial or a digital switching, measure the pH
value after adjusting the graduation of the pH meter with the temperature of a
sample.
References
1) JIS Z 8802: Methods for determination of pH of aqueous solutions (2011)
(5) Flow sheet for pH value: The flow sheet for pH value in fertilizers is shown below.
Predetermined volume
Ground-in stopper flask.
of analytical sample
← Water 5 -10 times or 100 times the volume
Stirring
Measurement pH meter
25
Testing Methods for Fertilizers (2018)
Note (1) Store potassium chloride standard solution in a polyethylene or borosilicate glass bottle
and seal the bottle.
(2) The volume that is recommended for an instrument or a cell used.
Comment 1 Potassium chloride standard solution used once or left in the air should not be used.
Comment 2 Check the indicated value as shown in 6.2 in JIS K 0130 as necessary. Actual
procedure to check is according to the operation procedure of the electrical
conductivity meter used for measurement.
Note (3) In the case of a moist laboratory sample, it is recommended to use a sample that is not
pre-dried.
(4) If the sample solution becomes hard to measure because it is gelled by the influence of
flocculants in sludge fertilizer, etc., increase the volume of water added. However, this
fact should be expressed in the test result.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0130 and as shown below.
Actual measurement operation is according to the operation procedure of the electrical
conductivity meter used for measurement.
a) Wash the read station of an electrical conductivity meter repeatedly no less than 3 times with
water.
b) Transfer a sample solution into a beaker (5), dip the read station and measure electrical
26
Testing Methods for Fertilizers (2018)
conductivity.
Note (5) It is necessary to transfer sufficient volume of sample solution to keep a measurement
value stable.
References
1) JIS K 0130:General rules for electrical conductivity measuring method (2008)
(5) Flow sheet for electrical conductivity: The flow sheet for electrical conductivity is showed
below.
Predetermined volume
Ground-in stopper flask.
of analytical sample
← Water ,10 times the volume
Stirring
27
Testing Methods for Fertilizers (2018)
3.5 Granularity
3.5.a Dry-type sieving testing method
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type A (Def-M)
and its symbol is 3.5.a-2017 or P-size.a-1.
Measure the particle diameter distribution of solid fertilizers with a dry-type sieving analysis.
(3) Dry-type sieving analysis procedure: Conduct sieving analysis corresponding to the aperture
size of a sieve used as indicated in JIS Z 8815 and as shown below.
(3.1) More than 1mm and no more than 4mm
a) Stack a large aperture sieve on an acceptor so that the large sieve is on top.
b) Measure a laboratory sample to the order of 0.1g and transfer it to the sieve at the top section.
c) After putting a stopper on it, hold the stacked sieves with both hands, and vibrate (1) them
back and forth along a unidirectional and horizontal plane at about 60 times per minute with
about 70 mm amplitude.
d) Transfer respective plus and minus sieves to a weighing dish (2).
Note (1) Conduct more circular motion at the rate of about 3 revolutions per minute as
necessary.
(2) Turning over the back side of a sieve, remove the clogged particles from the sieve
screen with a clogging removal brush and combine them with the plus sieve.
Note (3) When fine powder attaches to the back side of a sieve screen, remove them gently from
the back side with a clogging removal brush and combine them with the minus sieve
b).
Mass percentage of plus sieve or minus sieve (%) (R) = (A/T) × 100
References
1) JIS Z 8815: Test sieving - General requirements (1994).
2) JIS K 0069: testing methods for sieving of chemical products (1992).
(5) Flow sheet for granularity: The flow sheet of granularity of solid fertilizers is shown below.
Laboratory sample
Measure the mass to the order of 0.1g
or Test sample
Dry-type sieving
Mass measurement of
respective plus sieve Measure the mass to the order of 0.1g
and minus sieve
29
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Drying apparatus: Drying apparatus that can be adjusted to the test temperature ± 2 ºC.
b) Soxhlet extractor: An inter-changeable Soxhlet extractor, cooling apparatus and weighing
bottles (Example: JIS R 3503, attached figure 71)
c) Water bath A water bath that can be adjusted to about 60 ºC
d) Weighing bottle: A flat bottle flask connectable to a Soxhlet extractor. After heating with a
drying apparatus at 100 ºC - 105 ºC in advance, stand to cool in a desiccator and measure the
mass to the order of 1mg.
e) Cylindrical filter paper: A cylindrical filter paper made of cellulose. Example: 22 mm
external diameter, 20 mm internal diameter, 90 mm total length (1).
30
Testing Methods for Fertilizers (2018)
(2) The purpose is to prevent overflow at the upper end of an analytical sample.
(3) The amount of diethyl ether depends on the volume of a weighing bottle.
(4) Adjust the temperature for diethyl ether to circulate 16 - 20 times per hour. (Target
temperature is about 60 ºC)
(5) Remove the cylindrical filter paper from the Soxhlet extractor. In the case of a cock
attached Soxhlet extractor, open the cock and recover it.
(6) It is dangerous if diethyl ether resides in a weighing bottle when the bottle is
transferred to a drying apparatus.
(7) Wipe the outside of a weighing bottle since there is a risk of garbage or stain sticking
to it.
References
1) Japan Oil Chemist ‘Society: Standard Method for the Analysis of Fats, Oils and Related
Materials (2003), 1.5 Oil content p. 1 - 2, Incorporated Foundation Japan Oil Chemist
‘Society, Tokyo (2009)
2) Feed Analysis Standard Task Force: Feed Analysis Method/Handbook -2009 - I, p. 37 -
39, Incorporated Administrative Agency, Food and Agriculture Materials Inspection
Center, Saitama (2009)
(5) Flow sheet for oil content: The flow sheet for oil content in organic fertilizers is shown
below:
2 g- 5 g
Weigh to the order of 1mg to a cylindrical filter paper
analytical sample
Soxhlet extractor
Weighing bottle
31
Testing Methods for Fertilizers (2018)
b) Sulfuric acid: A JIS Guaranteed Reagent specified in JIS K 8951 or a reagent of equivalent
quality.
c) 0.25 mol/L sulfuric acid (1) (2): Add about 14 mL of sulfuric acid to a beaker containing 100
mL of water in advance, stir well, and add water to make 1000 mL.
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL
- 300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
32
Testing Methods for Fertilizers (2018)
and titrate with a 0.1 mol/L -0.2 mol/L sodium hydroxide solution until the color of the
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
(1). Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L -0.2 mol/L sodium hydroxide solution equivalent to 1 mL of 0.25
mol/L sulfuric acid
= V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
d) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
e) Catalyst (5): Mix potassium sulfate specified in JIS K 8962 and copper (II) sulfate
pentahydrate (6) specified in JIS K 8983 in the ratio of 9 to 1.
f) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
g) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, add water to make 100 mL.
h) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
i) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
j) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
k) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
l) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100 mL),
add equal volume of bromocresol green solution (0.5 g/100 mL).
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) c) can be replaced with 0.25 mol/L sulfuric acid
33
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Steam distillation apparatus
b) Digestion flask: Kjeldahl flask
c) Distillation flask: A Kjeldahl flask or round bottom flask that can be connected to a steam
distillation apparatus.
(4.2) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (10) of 0.25 mol/L sulfuric acid to an acceptor (11), add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (10) of boric acid solution (40 g/L)
to an acceptor (11), add a few drops of methyl red−bromocresol green mixture solution, and
connect this acceptor to a steam distillation apparatus.
b) Transfer a predetermined amount of the digestion solution to a 300-mL distillation flask, add a
proper amount of sodium hydroxide solution (200 g/L - 500 g/L) (12), and immediately
connect this distillation flask to the steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
Note (10) 5 mL - 20 mL
(11) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
34
Testing Methods for Fertilizers (2018)
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(12) An amount sufficient to make the solution strong alkalinity. A blue color will appear.
Note (13) The endpoint is reached when the color changes from green to light red.
Comment 6 The titration procedures in (2) a) Standardization, (2) c) Standardization and (4.3)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
Comment 7 The nitrogen content in the analytical sample can be measured by using an automatic
35
Testing Methods for Fertilizers (2018)
nitrogen analyzer (Kjeldahl method) instead of the test procedure in (4). The setup of
the program and the parameter of the analyzer as well as vessels etc. are according to
the specification and the operation method of the automatic nitrogen analyzer used.
However, conduct in advance a comparative test with the test procedure in (4) using
fertilizers containing no nitrate nitrogen, to confirm that there is no difference in the
quantitation value of total nitrogen.
Comment 8 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) -
20 % (mass fraction) and 1 % (mass fraction) - 5 % (mass fraction) are 98.5 % -
100.6 % and 97.1%-99.2 % as total nitrogen (T-N) respectively.
The results of the collaborative study (limited to reported values with the Kjeldahl
method) to determine a certified reference material fertilizer were analyzed by using
a three-level nesting analysis of variance. Table 1 shows the calculation results of
reproducibility, intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.2 % (mass fraction) for solid fertilizers, and 0.02 % (mass fraction) for fluid
fertilizers.
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatabllity Intermediate precision Reproducibility
2) 4) 5) 6) 7) 8) 9)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-A-10 11 14.68 0.07 0.5 0.07 0.5 0.13 0.9
1) The number of laboaratories used 6) Intermediate standard deviation
for analysis conducting Kjeldahl method
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p. 27 - 31, Yokendo, Tokyo (1988)
2) Feed Analysis Standard Task Force: Feed Analysis Method/Handbook -2009 - I, p. 28 -
33, Incorporated Administrative Agency, Food and Agriculture Materials Inspection
Center, Saitama (2009)
3) Takashi KUBOTA, Tomoko OSHIDA, Kozue YANAI, Yuzuru INOUE, Seiji MATSUI,
Takaharu MATSUMOTO, Eiichi ISHIKURO and Akemi YASUI: Improvement of the
Conditions for the Determination of Total Nitrogen in Fish Meal in Kjeldahl Method
and Its Comparison with Dumas Method, Bunsekikagaku, 60, p. 67 - 74 (2011)
4) Kimie KATO, Masaki CHIDA and Erina WATANABE: Verification of Performance
Characteristics of Testing Method for Total Nitrogen Content in Fertilizer by Kjeldahl
method, Research Report of Fertilizer, Vol. 5, p. 156 - 166 (2012)
36
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total nitrogen: The flow sheet for total nitrogen in fertilizers is shown below:
0.5 g - 5 g
Weigh to the order of 1 mg into a 300-mL digestion flask.
analytical sample
← 5 g - 10 g catalyst
←20 mL - 40 mL sulfuric acid
Heating Gently
Standing to cool
← small amount of water
Transfer 250-mL - 500-mL volumetric flask, water
Cooling
← Water (up to the marked line)
Digestion
solution
Figure 1 The flow sheet for total nitrogen in fertilizers (Kjeldahl method procedure)
Digestion
solution
Figure 2 Flow sheet for total nitrogen in fertilizers (Distillation and mesurement procedure)
37
Testing Methods for Fertilizers (2018)
(3) Measurement: Conduct measurement as shown below. However, confirm in advance using
an analytical sample that there is no difference from the measured value of total nitrogen
obtained according to 4.1.1.a, 4.1.1.c, 4.1.1.d or 4.1.1.e.
a) Measurement conditions for the total nitrogen analyzer by the combustion method: Set
up the measurement conditions for the total nitrogen analyzer considering the following:
Combustion temperature: No less than 870 ºC
b) Calibration curve preparation
1) Turn on the total nitrogen analyzer by the combustion method (1), and adjust so that stable
indicated values can be obtained.
2) Weigh a predetermined amount of the standard for calibration curves (2) to the order of 0.1 mg
into a combustion vessel.
3) Insert the combustion vessel into the total nitrogen analyzer by the combustion method, and
read the indicated value.
4) Conduct the procedure in 3) for another combustion vessel for a blank test, and read the
indicated value.
5) Prepare a curve for the relationship between the nitrogen content and the indicated value of
the standard for calibration curves and the blank test for calibration curves.
c) Sample measurement
1) Weigh a predetermined amount of an analytical sample to the order of 0.1 mg into a
combustion vessel.
2) Insert the combustion vessel containing the analytical sample to the total nitrogen analyzer by
the combustion method, and read the indicated value.
3) Obtain the nitrogen content from the calibration curve, and calculate total nitrogen in the
analytical sample.
Note (1) The setup of the program and the parameter of the analyzer are according to the
specification and the operation method of the total nitrogen analyzer by the
combustion method used.
(2) Standard for calibration curves: DL-Aspartic acid (purity no less than 99 % (mass
fraction)), EDTA (purity no less than 99 % (mass fraction)), hippuric acid (purity no
less than 98 % (mass fraction)) or other reagents having equivalent purity
recommended by the total nitrogen analyzer by the combustion method used.
38
Testing Methods for Fertilizers (2018)
Comment 1 Sample an analytical sample from a test sample prepared in 2.3.3 Grinding (3.1) by
grinding with a mill until it completely passes through a sieve of 500 µm aperture or
from a test sample prepared in 2.3.3 Grinding Comment 1. Additionally, the
sampling amount of an analytical sample is as shown in Table 1. In addition, set the
sampling amounts of analytical samples considering the estimated content of total
nitrogen in the test sample and the measurement range of the total nitrogen analyzer
by the combustion method.
Table 1 Sampling amount of analytical samples
Type of fertilizers Sampling amount (g)
Mixed fertilizers and designated
0.02 - 0.5
blended fertilizers
Organic fertilizers and compost 0.05 - 0.5
Sludge fertilizers 0.05 - 0.5
Comment 2 Compound fertilizers, designated blended fertilizers and nitrolime may have high
contents of phosphoric acid (P2O5), alkali metals (Na, K), alkaline earth metals (Ca,
Mg), etc., causing contamination of packing or damage in quartz parts, etc. To avoid
their influences, it is recommended to add tungsten oxide (elemental analysis reagent
or heat-treated reagent) to completely cover the analytical sample.
Comment 3 When a sample with a low content of organic compounds, such as mixed fertilizers
and designated blended fertilizers etc., and thus with low combustion efficiency is
measured, it is recommended to add sucrose to the analytical sample so that the
carbon content will be comparable to the standard for calibration curves. Additionally,
confirm in advance that sucrose to be used has a nitrogen content that does not affect
the measured value of total nitrogen of the analytical sample.
Comment 4 Table 2, in order to evaluate trueness, shows the results of the comparison of
measurement values by the Combustion method and the Kjeldahl method with sludge
fertilizers, organic fertilizers, and inorganic fertilizers, etc.
Table 3 shows results and analysis results from a collaborative study for testing
method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.01 % (mass fraction) for fluid fertilizers for home gardening, and 0.05 % (mass
fraction) for the other fertilizers.
39
Testing Methods for Fertilizers (2018)
40
Testing Methods for Fertilizers (2018)
References
1) Mariko AIZAWA, Yasushi SUGIMURA, Yuichi TAKAHASHI, Jun OKI, Yukio
FUKUCHI, Yuji SHIRAI and Norio HIKICHI: Validation of a Combustion Method for
Determination of Total Nitrogen Content in Sludge Fertilizer. Research Report of
Fertilizer Vol.1, p. 12 - 17, (2008)
2) Mariko AIZAWA and Yuji SHIRAI: Determination of Total Nitrogen content in Sludge
Fertilizer by a Combustion Method: A Collaborative Study. Research Report of
Fertilizer Vol.1, p. 18 - 24, (2008)
3) Mariko AIZAWA and Yuji SHIRAI: Validation of a Combustion Method for
Determination of Total Nitrogen Content in Organic Fertilizer. Research Report of
Fertilizer Vol.2, p. 6 - 11, (2009)
4) Mariko AIZAWA and Yuji SHIRAI: Validation of a Combustion Method for
Determination of Total Nitrogen Content in Inorganic Fertilizer. Research Report of
Fertilizer Vol.3, p. 1 - 10, (2010)
5) Mariko AIZAWA, Yuko SEKINE and Yuji SHIRAI: Determination of Total Nitrogen
Content in Fertilizer by a Combustion Method: A Collaborative Study. Research Report
of Fertilizer Vol.3, p. 11 - 18, (2010)
6) Kazumi UCHIYAMA and Yoshio MAEBASHI: Effective Analysis of Organic Trace
Element, p. 99, Mimizuku-sha, Tokyo (2008)
(4) Flow sheet for total nitrogen: The flow sheet for total nitrogen in fertilizers is shown
below:
Figure Flow sheet for total nitrogen in fertilizers by the combustion method.
41
Testing Methods for Fertilizers (2018)
Reference: Chromatograms of the standard for calibration curves and an analytical sample are
shown below:
b) Sulfuric acid: A JIS Guaranteed Reagent specified in JIS K 8951 or a reagent of equivalent
quality.
c) 0.25 mol/L sulfuric acid (1) (2): Add about 14 mL of sulfuric acid to a beaker containing 100
mL of water in advance, stir well, and add water to make 1000 mL.
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL -
300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
and titrate with a 0.1 mol/L - 0.2 mol/L sodium hydroxide solution until the color of the
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
43
Testing Methods for Fertilizers (2018)
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula (1).
Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L -0.2 mol/L sodium hydroxide solution equivalent to 1 mL of 0.25
mol/L sulfuric acid
= V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
d) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
e) Hydrochloric acid: A JIS Guaranteed Reagent specified in JIS K 8180 or a reagent of
equivalent quality.
f) Tin (II) chloride dihydrate: A JIS Guaranteed Reagent specified in JIS K 8136 or a reagent
of mercury analysis grade or equivalent quality.
g) Devarda’s alloy: A reagent of nitrogen analysis grade specified in JIS K 8653 or a reagent of
equivalent quality.
h) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
i) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, add water to make 100 mL.
j) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
k) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
l) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
m) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
n) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100 mL),
add equal volume of bromocresol green solution (0.5 g/100 mL).
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Steam distillation apparatus
b) Digestion flask: Kjeldahl flask
44
Testing Methods for Fertilizers (2018)
c) Distillation flask: A Kjeldahl flask or round bottom flask that can be connected to a steam
distillation apparatus.
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) c) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
Note (5) In the case of direct distillation, a 500-mL Kjeldahl flask connectable to a steam
distillation apparatus is preferable.
(6) If the bubbles foam strongly and excessively, suspend heating for a little while.
(7) When the entire sample solution volume is used in measurement, it is not necessary to
precisely adjust.
(4.2) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (8) of 0.25 mol/L sulfuric acid to an acceptor, (9) add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (8) of boric acid solution (40 g/L) to
an acceptor (9), add a few drops of methyl red−bromocresol green mixture solution, and
connect this acceptor to a steam distillation apparatus.
b) Transfer a predetermined amount of the digestion solution to a 300-mL distillation flask, add a
proper amount of sodium hydroxide solution (200 g/L - 500 g/L) (10), and connect this
distillation flask to the steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
Note (8) 5 mL - 20 mL
(9) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
beaker with which the distillate outlet of the steam distillation apparatus can be
45
Testing Methods for Fertilizers (2018)
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(10) An amount sufficient to make the solution strong alkalinity. A blue color will appear.
V10: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
f2: Factor of 0.25 mol/L sulfuric acid
V11: Predetermined volume (mL) of the digestion solution in (4.1) g)
V12: Transferred amount (mL) of the digestion solution subjected to distillation in
(4.2) b)
W3: Mass (g) of the analytical sample
Note (11) The endpoint is reached when the color changes from green to light red.
Comment 3 The titration procedures in (2) a) Standardization, (2) c) Standardization and (4.3)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
References
46
Testing Methods for Fertilizers (2018)
47
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total nitrogen: The flow sheet for total nitrogen in fertilizers is shown
below:
0.5 g – 1 g
Weigh to the order of 1 mg into a 300-mL - 500-mL digestion flask.
analytical sample
← 60 mL of hydrochloric acid (1+1)
← 2 g of tin (II) chloride dihydrate
Leaving at rest About 20 minutes
← 3.5 g devardar's alloy
Leaving at rest About 40 minutes
← 70 mL sulfuric acid (1+1)
← 1 boiling stone
Heat at low temperature, and as soon as white smoke evolves,
Heating
strengthen heating gradually and further heat for about 90 minutes.
Standing to cool
← 100 mL - 200 mL Water
Transfer 250-mL - 500-mL volumetric flask, water
Cooling
← Water (up to the marked line)
Digedtion
Solution
Figure 1 The flow sheet for total nitrogen in fertilizers (Reduction and Kjeldahl method procedure)
48
Testing Methods for Fertilizers (2018)
Digestion
Solution
Figure 2 Flow sheet for total nitrogen in fertilizers (Distillation and measurement procedure).
49
Testing Methods for Fertilizers (2018)
b) Sulfuric acid: A JIS Guaranteed Reagent specified in JIS K 8951 or a reagent of equivalent
quality.
c) 0.25 mol/L sulfuric acid (1) (2): Add about 14 mL of sulfuric acid to a beaker containing 100
mL of water in advance, stir well, and add water to make 1000 mL.
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL
- 300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
and titrate with a 0.1 mol/L -0.2 mol/L sodium hydroxide solution until the color of the
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
50
Testing Methods for Fertilizers (2018)
(1). Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L -0.2 mol/L sodium hydroxide solution equivalent to 1 mL of 0.25
mol/L sulfuric acid
= V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
d) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
e) Reduced iron: Nitrogen content is no more than 0.005 % (mass fraction)
f) Catalyst (5): Mix potassium sulfate specified in JIS K 8962 and copper (II) sulfate
pentahydrate (6) specified in JIS K 8983 in the ratio of 9 to 1.
g) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
h) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8842, add water to make 100 mL.
i) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
j) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
k) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
l) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
m) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100
mL), add equal volume of bromocresol green solution (0.5 g/100 mL).
51
Testing Methods for Fertilizers (2018)
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) c) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
Note (7) A sudden reaction generates heat, and unreacted nitric acid vaporizes or digests to make
nitrogen oxide etc. through which process losses occur easily. Careful and efficient
operation should be taken
(8) Until a sudden reaction is settled.
(9) If the bubbles foam strongly and excessively, suspend heating for a little while.
(10) When the solution has finished changing color, heat further for no less than 2 hours.
(4.2) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (11) of 0.25 mol/L sulfuric acid to an acceptor (12), add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (11) of boric acid solution (40 g/L)
to an acceptor (12), add a few drops of methyl red−bromocresol green mixture solution, and
connect this acceptor to a steam distillation apparatus.
b) Transfer a predetermined amount of the digestion solution to a 300-mL distillation flask, add a
proper amount of sodium hydroxide solution (200 g/L - 500 g/L) (13), and connect this
distillation flask to the steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
Note (11) 5 mL - 20 mL
(12) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
52
Testing Methods for Fertilizers (2018)
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(13) An amount sufficient to make the solution strong alkalinity. A blue or reddish-brown
color will appear.
Note (14) The endpoint is reached when the color changes from green to light red.
Comment 3 The titration procedures in (2) a) Standardization, (2) c) Standardization and (4.3)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
53
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.33- 34, Yokendo, Tokyo (1988)
(5) Flow sheet for total nitrogen: The flow sheet for total nitrogen in fertilizers is shown below:
Standing to cool
← 5 g - 10 g catalyst
← 30 mL sulfuric acid
Heat gradually until water evaporates and the white smoke
Heating
of sulfuric acid evolves and ignite to digest completely
Standing to cool
← A small amount of water
Transfer 250-mL - 500-mL volumetric flask, water
Cooling
← Water (up to the marked line)
Digestion
Solution
Figure 1 The flow sheet for total nitrogen in fertilizers (Reduction and Kjeldahl method procedure)
54
Testing Methods for Fertilizers (2018)
Digestion
Solution
Aliquot
Distillation flask
(predetermined volume)
← Sodium hydroxide (200 g/L - 500 g/L)
Receiver: 200-mL - 300-mL Erlenmeyer flask or beaker
Steam distillation A predetermined amount of 0.25 mol/L sulfuric acid and a few drops of
apparatus methyl red-methylene blue mixture solution, or boric acid solution (40 g/L),
several drops of methyl red - bromocresol green mixture solution
Figure 2 Flow sheet for total nitrogen in fertilizers (Distillation and measurement procedure).
55
Testing Methods for Fertilizers (2018)
A-N: Ammoniac nitrogen (% (mass fraction)) in the analytical sample (1) obtained
by 4.1.2
N-N: Nitrate nitrogen (% (mass fraction)) in the analytical sample (1) obtained by
4.1.3
Note (1) A-N and N-N use raw data without rounding numerical value
56
Testing Methods for Fertilizers (2018)
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
(1). Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L -0.2 mol/L sodium hydroxide solution equivalent to 1 mL of 0.25
mol/L sulfuric acid
= V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
e) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
f) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
g) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, add water to make 100 mL.
h) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
i) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
j) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
k) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
l) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100
mL), add equal volume of bromocresol green solution (0.5 g/100 mL).
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) d) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Steam distillation apparatus
b) Distillation flask: A Kjeldahl flask or round bottom flask that can be connected to a steam
58
Testing Methods for Fertilizers (2018)
distillation apparatus.
Note (5) The sampling amount of the analytical sample is 5 g when there is less nitrogen content
in the fertilizers such as a home garden-use fertilizer.
Comment 3 When it is a fertilizer which contains uric ammonium acid, humus acid ammonium or
nitrate nitrogen, etc., or when it is not a fertilizer in which phosphate, ammonium and
magnesium coexist, conduct the procedure in (4.1.1) a) - c) in 4.2.4.a or the
procedure in (4.1.2) a) - c) in 3.2.4.a and transfer a predetermined amount of
suspension (the equivalents of 20 mg - 100 mg as N) to a 300-mL - 500-mL
distillation flask to make a sample solution.
(4.2) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (6) of 0.25 mol/L sulfuric acid to an acceptor (7), add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (6) of boric acid solution (40
(7)
g/L) to an acceptor , add a few drops of methyl red−bromocresol green mixture solution,
and connect this acceptor to a steam distillation apparatus.
b) Add (9) no less than 2 g of magnesium oxide (8) to the distillation flask, and connect this
distillation flask to a steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
Note (6) 5 mL - 20 mL
(7) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(8) An amount sufficient to make the solution strong alkalinity.
(9) Add a small amount of silicone oil as necessary.
Comment 4 When the sample does not contain organic matters or urea, a proper amount of
sodium hydroxide solution (200 g/L - 500 g/L) (8) can be added instead of
magnesium oxide.
59
Testing Methods for Fertilizers (2018)
V8: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration
C2 : Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
f2 : Factor of 0.25 mol/L sulfuric acid
W3: Mass (g) of the analytical sample
Note (10) The endpoint is reached when the color changes from green to light red.
Comment 5 If it is hard to confirm the endpoint due to the carbon dioxide resulting from
carbonate in the extract when magnesium oxide is used, it is recommended to boil
the extract for 1-2 minute(s) after distilling and cool, and then titrate.
Comment 6 The titration procedures in (2) a) Standardization, (2) d) Standardization and
(4.3) can be conducted by an automatic titrator. The setup of the titration program,
the determination parameter for the endpoint and vessels such as acceptors are
according to the specification and the operation method of the automatic titrator
used.
Comment 7 Recovery testing was conducted to evaluate trueness using a preparation sample. As
a result, the average rate of recovery at the content level of 10 % (mass fraction) -
21 % (mass fraction) and 1 % (mass fraction) are 102.2 % - 100.8 % and 102.5 %
as ammoniacal nitrogen (A-N) respectively.
The results of the collaborative study (limited to reported values with Distillation
method) to determine a certified reference material fertilizer were analyzed by
using a three-level nesting analysis of variance. Table 1 shows the calculation
results of reproducibility, intermediate precision and repeatability. Additionally, the
minimum limit of quantification of this testing method is about 0.1 % (mass
fraction) for solid fertilizers, and 0.01 % (mass fraction) for fluid fertilizers.
60
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 4) 5) 6) 7) 8) 9)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-B-10 11 8.38 0.09 1.0 0.11 1.3 0.15 1.8
FAMIC-B-14 11 8.06 0.03 0.4 0.05 0.6 0.07 0.9
1) The number of laboaratories used 6) Intermediate standard deviation
for analysis conducting Distillation method
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.36- 37, Yokendo, Tokyo (1988)
2) Kimie KATO, Masaki CHIDA and Erina WATANABE: Verification of Performance
Characteristics of Testing Method for Ammonia Nitrogen Content in Fertilizer by
Distillation Method, Research Report of Fertilizer, Vol. 6, p. 130 - 138 (2013)
61
Testing Methods for Fertilizers (2018)
(5) Flow sheet for ammoniac nitrogen: The flow sheet for ammoniac nitrogen in fertilizers is
shown below.
62
Testing Methods for Fertilizers (2018)
mL.
h) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
i) Bromothymol blue solution (1 g/100 mL): Dissolve 1 g of bromothymol blue specified in
JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, and add water to make 100 mL.
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 Thymol blue can be dissolved if it is sodium salt. The thymol blue specified in JIS K
8643 is slightly hard to dissolve in ethanol and hard to dissolve in water. Therefore,
add about 2.15 mL of sodium hydroxide solution (0.1 mol/L) per 0.1 g of thymol
blue to neutralize, and then prepare the thymol blue solution (1 g/100 mL) through
the same procedure as (2) i).
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.4.a.
Note (2) Add 3 mL of aluminum chloride per 0.04 g of P or 0.1 g of P2O5 in the sample
solution.
(3) Form precipitate of aluminum hydroxide and aluminum phosphate to separate the
phosphate.
Comment 4 In the procedure of (4.1.1) a) and (4.1.2) a), it is also allowed to weigh 2.5g of an
analytical sample to the order of 1 mg, and put it in a 250-mL volumetric flask
64
Testing Methods for Fertilizers (2018)
Comment 5 When it is not a fertilizer in which phosphate, ammonium and magnesium coexist,
about 400 mL of potassium chloride solution (1 mol/L) can be used instead of about
300 mL of hydrochloric acid (1+20) in the procedure of (4.1.2) b).
Comment 6 In the case of the mixed fertilizers containing bentonite, after shaking to mix using
about 300 mL of hydrochloric acid (1+20) or about 400 mL of potassium chloride
solution (1 mol/L) according to Comment 5 in (4.1.2) b), add water to the marked
line, filter with Type 3 filter paper and transfer 50 ml - 100 mL to a 250-mL
volumetric flask, and then conduct the procedure in (4.1.2) c) - e).
VS: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration in (4.2) e)
VB: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration of the blank test in (4.2) f)
C: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide
solution
f: Factor of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
V1: Predetermined volume (mL) of the sample solution in (4.1.1) c) or (4.1.2) d)
V2: Transferred amount (mL) of the sample solution in (4.2) a)
W2: Mass (g) of the analytical sample
Comment 7 The titration procedures in (2) a) Standardization and (4.2) e) - f) can be conducted
by an automatic titrator. The setup of the titration program, the determination
parameter for the endpoint and vessels such as acceptors are according to the
specification and the operation method of the automatic titrator used.
Comment 8 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) -
21 % (mass fraction) and 1 % (mass fraction) are 100.4 % - 101.0 % and 100.1 % as
ammoniacal nitrogen (A-N) respectively.
The results of the collaborative study (limited to reported values by Formaldehyde
65
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
certified reference laboratory Average2) sr
4)
RSD r
5)
s I(T)
6)
RSD I(T)
7)
sR
8)
RSD R
9)
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-A-10 10 10.66 0.07 0.7 0.09 0.8 0.16 1.5
FAMIC-A-13 9 10.36 0.06 0.5 0.08 0.8 0.21 2.0
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting Formaldehyde method
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.39- 42, Yokendo, Tokyo (1988)
2) Kimie KATO, Masaki CHIDA and Erina WATANABE: Verification of Performance
Characteristics of Testing Method for Ammonia Nitrogen Content in Fertilizer by
Formaldehyde Method, Research Report of Fertilizer Vol. 6, p. 139 - 147 (2013)
(5) Flow sheet for ammoniacal nitrogen: The flow sheet for ammoniacal nitrogen in fertilizers
is shown below.
5 g analytical sample
Weigh to the order of 1 mg into a 500-mL volumetric flask
(ammonium salts)
← About 400 mL of water
Shaking to mix Rotary shaker (30 - 40 revolutions/minute), 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 The flow sheet for ammoniac nitrogen in fertilizers (Extraction procedure (4.1.1))
66
Testing Methods for Fertilizers (2018)
5 g analytical sample
Weigh to the order of 1 mg into a 500-mL volumetric flask
(compound fertilizer)
← About 300 mL of hydrochloric acid (1 + 20)
Shaking to mix Rotary shaker (30 - 40 revolutions/minute), 30 minutes
← Aluminum chloride solution (1 mol/L)
← One or two drop(s) of methyl red
← Potassium hydroxide solution (170 g/L)
(Until the solution changes to light yellow)
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-2 The flow sheet for ammoniac nitrogen in fertilizers (Extraction procedure (4.1.2))
Sample solution
Figure 2 The flow sheet for ammoniac nitrogen in fertilizers (Measurement solution)
67
Testing Methods for Fertilizers (2018)
b) Sulfuric acid: A JIS Guaranteed Reagent specified in JIS K 8951 or a reagent of equivalent
quality.
c) 0.25 mol/L sulfuric acid (1) (2): Add about 14 mL of sulfuric acid to a beaker containing 100
mL of water in advance, stir well, and add water to make 1000 mL.
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL
- 300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
68
Testing Methods for Fertilizers (2018)
and titrate with a 0.1 mol/L -0.2 mol/L sodium hydroxide solution until the color of the
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
(1). Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L -0.2 mol/L sodium hydroxide solution equivalent to 1 mL of 0.25
mol/L sulfuric acid
= V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
d) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
e) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
f) Devarda’s alloy: A reagent of nitrogen analysis grade specified in JIS K 8653 or a reagent of
equivalent quality.
g) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, and add water to make 100
mL.
h) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
i) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
j) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
k) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
l) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100 mL),
add equal volume of bromocresol green solution (0.5 g/100 mL).
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) c) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
69
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Steam distillation apparatus
b) Distillation flask: A Kjeldahl flask or round bottom flask that can be connected to a steam
distillation apparatus.
Note (5) Conduct the procedure in Comment 3 when there is much nitrogen content in the
fertilizers such as simple salt fertilizers.
Comment 3 In the case of nitrate fertilizer, etc. containing much nitrogen content, weigh 2 g - 5g
of an analytical sample to the order of 1 mg, put it into a 250- mL volumetric flask,
dissolve it in water, and further add water up to the marked line. Put predetermined
volume of suspension (the equivalents of 20 mg -100 mg as N) into a 300- mL -
500-mL distillation flask.
(4.2) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (6) of 0.25 mol/L sulfuric acid to an acceptor (7), add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (6) of boric acid solution (40 g/L) to
an acceptor (7), add a few drops of methyl red−bromocresol green mixture solution, and
connect this acceptor to a steam distillation apparatus.
b) Add (10) no less than 3 mg of devarda’s alloy and adequate volume of sodium hydroxide (200
g/L - 500 g/L) (8) (9) and connect this distillation flask to the steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
Note (6) 5 mL - 20 mL
(7) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(8) Sudden reaction makes bubbles foam drastically and the bubbles overflow from a
distillation flask. Therefore, it is required to add an alkali solution gradually and mix
quietly.
(9) An amount sufficient to make the solution strong alkalinity.
(10) Add a small amount of silicone oil as necessary.
b) Calculate nitrogen content (N-N+A-N) in the analytical sample by the following formula:
c) Subtract the ammoniac nitrogen (A-N) separately obtained in 4.1.2 from the obtained nitrogen
content (N-N+A-N) to calculate nitrate nitrogen (N-N) (11) (12).
Note (11) The nitrogen content (N-N+A-N) and ammoniacal nitrogen (A-N) use raw data
without rounding the numerical value.
(12) When no ammoniacal nitrogen (A-N) is contained, the nitrogen content (N-N+A-N)
calculated in (4.3) b) is regarded as nitrate nitrogen (N-N).
V10: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration
C2 : Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
f2 : Factor of 0.25 mol/L sulfuric acid
W3: Mass (g) of the analytical sample
Note (13) The endpoint is reached when the color changes from green to light red.
Comment 4 The titration procedures in (2) a) Standardization, (2) c) Standardization and (4.3)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.49- 50, Yokendo, Tokyo (1988)
71
Testing Methods for Fertilizers (2018)
(5) Flow sheet for nitrate nitrogen: The flow sheet for nitrate nitrogen in fertilizers is shown
below.
72
Testing Methods for Fertilizers (2018)
b) Sulfuric acid: A JIS Guaranteed Reagent specified in JIS K 8951 or a reagent of equivalent
quality.
c) 0.25 mol/L sulfuric acid (1) (2): Add about 14 mL of sulfuric acid to a beaker containing 100
mL of water in advance, stir well, and add water to make 1000 mL.
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL
- 300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
and titrate with a 0.1 mol/L -0.2 mol/L sodium hydroxide solution until the color of the
73
Testing Methods for Fertilizers (2018)
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
(1). Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L -0.2 mol/L sodium hydroxide solution equivalent to 1 mL of 0.25
mol/L sulfuric acid
=V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
d) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
e) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
f) Reduced iron: Nitrogen content is no more than 0.005 % (mass fraction)
g) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, and add water to make 100
mL.
h) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
i) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
j) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
k) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
l) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100
mL), add equal volume of bromocresol green solution (0.5 g/100 mL).
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) c) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
Note (5) Conduct the procedure in Comment 3 when there is much nitrogen content in the
fertilizers such as simple salt fertilizers.
(6) Sudden reaction generates heat, and unreacted nitric acid vaporizes or digests to make
nitrogen oxide etc. through which process losses occur easily. Careful and efficient
operation should be taken
(7) Until a sudden reaction is settled.
Comment 3 In the case of nitrate fertilizer, etc. containing much nitrogen content, weigh 2 g - 5g
of an analytical sample to the order of 1 mg, put it into a 250- mL volumetric flask,
dissolve it in water, and further add water up to the marked line. Put predetermined
volume of suspension (the equivalents of 20 mg - 100 mg as N) into a 300-mL -
500-mL distillation flask.
(4.2) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (8) of 0.25 mol/L sulfuric acid to an acceptor (9), add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (8) of boric acid solution (40 g/L) to
an acceptor (9), add a few drops of methyl red−bromocresol green mixture solution, and
connect this acceptor to a steam distillation apparatus.
b) Add adequate volume of sodium hydroxide (200 g/L - 500 g/L) (10) and connect this
distillation flask to the steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
Note (8) 5 mL - 20 mL
(9) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(10) An amount sufficient to make the solution strong alkalinity.
Note (11) The nitrogen content (N-N+A-N) and ammoniacal nitrogen (A-N) use raw data
without rounding the numerical value.
(12) When no ammoniacal nitrogen (A-N) is contained, the nitrogen content (N-N+A-N)
calculated in (4.3) b) is regarded as nitrate nitrogen (N-N).
(4.3.2) When a boric acid solution (40 g/L) is used in (4.2) a):
a) Titrate the distillate with 0.25 mol/L sulfuric acid until the color of the solution becomes light
red (13).
b) Calculate nitrogen content (N-N+A-N) in the analytical sample by the following formula:
c) Subtract the ammoniac nitrogen (A-N) separately obtained in 4.1.2 from the obtained nitrogen
content (N-N+A-N) to calculate nitrate nitrogen (N-N) (11) (12).
V10: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration
C2 : Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
f2 : Factor of 0.25 mol/L sulfuric acid
W3: Mass (g) of the analytical sample
Note (13) The endpoint is reached when the color changes from green to light red.
Comment 4 The titration procedures in (2) a) Standardization, (2) c) Standardization and (4.3)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
76
Testing Methods for Fertilizers (2018)
(5) Flow sheet for nitrate nitrogen: The flow sheet for nitrate nitrogen in fertilizers is shown
below.
Standing to cool
←Sodium hydroxide (200 g/L - 500 g/L)
Receiver: 200-mL - 300-mL Erlenmeyer flask or beaker
A predetermined amount of 0.25 mol/L sulfuric acid
Steam distillation
and a few drops of methyl red-methylene blue mixture solution,
apparatus
or boric acid solution (40 g/L),
several drops of methyl red - bromocresol green mixture solution
77
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the nitrate standard solution in (2), a nitrate standard solution for a
calibration curve can be prepared using nitrate nitrogen (NO3-N 0.1 mg/mL or 1
mg/mL) traceable to National Metrology.
Note (3) As soon as the sample solution is prepared, conduct the procedure in (4.2) a).
Comment 2 If the filtrate of (4.1) e) is colored, add no more than 0.5 g of active carbon and
filter with Type 3 filter paper to make a sample solution.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 410 nm
b) Calibration curve preparation
1) Transfer 1 mL - 10 mL of nitrate standard solution (N-N 0.01 mg/mL) to small evaporating
dishes (4) step-by-step.
2) Conduct the same procedure as (4.2) b) - g) to make the nitrate standard solution for the
calibration curve preparation.
3) Transfer 40 mL of water to a 100-mL volumetric flask, and shake to mix while gently adding
2 mL of phenol sulfuric acid. Let it stand to cool and conduct the same procedure as (4.2) f) -
g) to make the blank test solution for calibration curve.
4) Measure absorbance at wavelength 410 nm of the nitrate standard solution for the calibration
curve preparation using the blank test solution for the calibration curve preparation as the
control.
5) Prepare the calibration curve of concentration of the nitrate nitrogen and absorbance of the
nitrate standard solutions for the calibration curve preparation.
c) Sample measurement
1) Regarding the solution in (4.2) g), measure absorbance by the same procedure as b) 4).
79
Testing Methods for Fertilizers (2018)
2) Obtain the nitrate nitrogen (N-N) content from the calibration curve and calculate nitrate
nitrogen (N-N) in the analytical sample.
Comment 3 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 16 % (mass fraction) and
1 % (mass fraction) - 3% (mass fraction) are 103.4 % and 101.1 % - 100.9 % as
nitrate nitrogen (N-N) respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.01 % (mass fraction) for solid fertilizers, and 0.002 % (mass fraction) for fluid
fertilizers.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.52- 55, Yokendo, Tokyo (1988)
2) Kimie KATO, Sakiko TAKAHASHI and Yuji SHIRAI: Validation of a Color metric
Method for Determination of Nitrogen, Phosphorus and Boron: Evaluation of
Calibration curve, Research Report of Fertilizer Vol. 2, p. 137 - 144 (2009)
3) Kimie KATO, Masaki CHIDA and Erina WATANABE: Verification of Performance
Characteristics of Testing Method for Nitrate Nitrogen Content in Fertilizer by Phenol
Sulfuric Acid Method, Research Report of Fertilizer Vol. 6, p. 148 - 155 (2013)
(5) Flow sheet for nitrate nitrogen: The flow sheet for nitrate nitrogen in fertilizers is shown
below.
1 mg
Weigh to the order of 1 mg to a 250-mL volumetric flask
analytical sample
← About 200 mL of copper sulfate - silver sulfate solution
Shaking to mix Rotary shaker (30 - 40 revolutions/min) for 20 minutes
← About 1 g of calcium hydroxide
← About 1 g of basic magnesium carbonate
Shaking to mix Rotary shaker (30 - 40 revolutions/min) for 10 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1 The flow sheet for nitrate nitrogen in fertilizers (Extraction procedure)
80
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined Small evaporating dish
volume)
Figure Flow sheet for nitrate nitrogen in fertilizers (Coloring and measurement procedure)
81
Testing Methods for Fertilizers (2018)
82
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the phosphoric acid standard solution in (2), a phosphoric acid standard
solution for a calibration curve can be prepared using a phosphoric acid standard
solution (P 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate the total phosphoric acid (T-P2O5) in an analytical sample by
multiplying the concentration (P) of a phosphoric acid standard solution for a
calibration curve or a measured value obtained in (4.3) by a conversion factor
(2.2914).
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Spectrophotometer: A spectrophotometer specified in JIS K 0115
b) Electric furnace: An electric furnace that can be adjusted to 550 ºC ± 5 ºC.
c) Hot plate or sand bath: A hot plate whose surface temperature can be adjusted up to 250 ºC.
Adjust the amounts of gas and silica sand of a sand bath so that the sand bath temperature can
be set to 250 ºC.
d) Digestion flask: Kjeldahl flask
Note (8) When the solution has finished changing color, heat further for no less than 2 hours.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1) in 4.2.1.b. Also, the
procedures in (4.1.1) a) - f) are the same as in (4.1) in 4.1.1.a.
Note (9) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
83
Testing Methods for Fertilizers (2018)
Comment 3 The procedures in (4.1.2) are the same as the procedures in (4.1.1) in 4.3.1.a, (4.1.1)
in 4.5.1.a, (4.1.1) in 4.6.1.a and (4.1) in 8.4.a.
Note (10) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for
about 1 hour and further raise to 450 °C in 1 to 2 hours.
(11) The watch glass can be removed.
(12) When drying and hardening it, phosphoric acid may not dissolve completely and the
concentration may become a low value in the procedure in g).
(13) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the
sample solution will be hydrochloric acid (1+23). For example, when a 100-mL
volumetric flask is used in the procedure in h), about 25 mL of hydrochloric acid
(1+5) should be added.
Comment 4 The procedures in (4.1.3) are the same as in (4.1.2) in 4.3.1.a, (4.1.2) in 4.5.1.a,
(4.1.2) in 4.6.1.a, (4.1) in 4.9.1.a, (4.1) in 4.9.1.b, (4.1) in 4.10.1.a and (4.1) in
4.10.1.b. And also, the same as (4.1) a) - h) in 5.3.a, (4.1) a) - h) in 5.3.b, (4.1) a) -
h) in 5.4.a, (4.1) a) - h) in 5.4.b, (4.1) a) - h) in 5.5.a, (4.1) a) - h) in 5.5.d, (4.1) a) -
h) in 5.6.a and (4.1) a) - h) in 5.6.b.
Note (14) It is not necessary to add a phenolphthalein solution (1 g/100 mL) when determination
can be done by the color change of copper ion (light blue → blue-purple).
(15) Without the addition of water, precipitate may be produced when a coloring reagent
84
Testing Methods for Fertilizers (2018)
solution is added.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 420 nm
b) Calibration curve preparation
1) Transfer 1 mL - 12 mL of phosphoric acid standard solution (P2O5 0.5 mg/mL) to 100-mL
volumetric flasks step-by-step.
2) Add a proper amount of water (15), and conduct the same procedure as (4.2) d) to make the
P2O5 0.5 mg/100 mL - 6 mg/100 mL phosphoric acid standard solutions for the calibration
curve preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
test solution for the calibration curve preparation.
4) Measure absorbance at a wavelength of 420 nm of the phosphoric acid standard solutions for
the calibration curve preparation using the blank test solution for the calibration curve
preparation as the control (16).
5) Prepare the calibration curve of the phosphoric acid concentration and absorbance of the
phosphoric acid standard solutions for the calibration curve preparation.
c) Sample measurement
1) Regarding the solution in (4.2) d), measure absorbance by the same procedure as b) 4) (16).
2) Obtain the phosphoric acid (P2O5) content from the calibration curve, and calculate total
phosphoric acid (T-P2O5) in the analytical sample.
Note (16) Measure within 6 hours after adding the coloring reagent solution in the procedure in
(4.2) d).
Comment 5 After the procedure in (4.2) a), it is also possible to measure soluble phosphoric acid
at the same time by adding 4 mL of nitrate acid (1+1) and 2 mL of Petermans citrate
solution and by conducting the procedures from (4.2) d) to (4.3) in 4.2.2.a (using b
reagent solution in the Official Methods of Analysis of Fertilizers (1992)).
Further after the procedure in (4.2) a), it is also possible to measure citrate soluble
phosphoric acid at the same time by adding 4 mL of nitrate acid (1+1) and 17 mL of
citrate solution and by conducting the procedures from (4.2) d) to (4.3) in 4.2.3.a
(using b reagent solution in the Official Methods of Analysis of Fertilizers (1992)).
Comment 6 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % - 20 % (mass
fraction) and 1 % (mass fraction) - 5 % (mass fraction) are 99.4 % - 100.2 % and
101.0 % - 105.7 % as total phosphoric acid (T-P2O5) respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.04 % (mass fraction) for solid fertilizers, and 0.01 % (mass fraction) for fluid
fertilizers.
85
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
certified reference laboratory Average2) sr
4)
RSD r
5)
s I(T)
6)
RSD I(T)
7)
sR
8)
RSD R
9)
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-C-12 9 8.62 0.03 0.4 0.04 0.4 0.08 0.9
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting Ammonium vanadomolybdate absorptiometric analysis
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.108- 114, Yokendo, Tokyo (1988)
2) Kimie KATO, Sakiko TAKAHASHI and Yuji SHIRAI: Validation of a Color metric
Method for Determination of Nitrogen, Phosphorus and Boron: Evaluation of
Calibration curve, Research Report of Fertilizer Vol. 2, p. 137 - 144 (2009)
3) Kimie KATO, Masayuki YOSHIMOTO and Yuji SHIRAI: Systematization of
Determination Methods of Major Components in Sludge Fertilizer, Compost and
Organic Fertilizer, Research Report of Fertilizer, Vol.3 p. 107 - 116 (2010)
4) Yoshiyuki SUNAGA, Yasushi SUGIMURA, Ichiro YOSHIDA and Hidenori KONISHI:
Verification of Performance Characteristics of Testing Methods for Phosphorus Content
in Fertilizer by Ammonium Vanadomolybdate Absorption Photometry, Research Report
of Fertilizer Vol. 5, p. 167 - 179 (2012)
86
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total phosphoric acid: The flow sheet for total phosphoric acid in fertilizers is
shown below:
0.5 g - 5 g
Weigh to the order of 1 mg into a 300-mL Kjeldahl flask.
analytical sample
← About 10 g catalyst
← 20 mL - 40 mL sulfuric acid
Heating Gently
Sample solution
87
Testing Methods for Fertilizers (2018)
5g
Weigh to the order of 1 mg into a 200-mL - 300-mL tall beaker.
analytical sample
Cooling
Sample solution
88
Testing Methods for Fertilizers (2018)
5g
Weigh to the order of 1 mg into a 200-mL - 300-mL tall beaker.
analytical sample
Standing to cool
Sample solution
89
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
90
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Water bath: Water bath that can be adjusted to 60 ºC - 65 ºC.
b) Drying apparatus: A drying apparatus that can be adjusted to 220 ºC ± 5 ºC.
c) Crucible type glass filter: A crucible type glass filter 1G4 specified in JIS R 3503. Let it
stand to cool in a desiccator after heating at 220 ºC ± 5 ºC in advance and measure the mass to
the order of 1 mg.
d) Digestion flask: Kjeldahl flask
Note (4) When the solution has finished changing color, heat further for no less than 2 hours.
Comment 1 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.1.a. In
addition, the sample solution prepared in (4.1.2) in 4.2.1.a and (4.1.3) in 4.2.1.a can
also be used.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.98- 106, Yokendo, Tokyo (1988)
92
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total phosphoric acid: The flow sheet for total phosphoric acid in fertilizers
is shown below:
0.5 g- 5 g
Weigh to the order of 1 mg into a 300-mL digestion flask.
analytical sample
← 5 g - 10 g catalyst
← 20 mL - 40 mL sulfuric acid
Heating Gently
Cooling
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
93
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 300-mL tall beaker
volume)
← 5 mL nitrate acid
← Water (to reach about 80 mL)
Cover with a watch glass for 3 minutes
Boiling
Wash a watch glass and inside the tall beaker with water
← Water (to be about 100 mL)
← Quimosiac solution 50 mL
Produce precipitate 60 ºC - 65 ºC, for 15 minutes, sometimes mixing
Filtration
under reduced Crucible type glass filtering apparatus 1G4, 3 times with water
pressure
Figure 2 The flow sheet for total phosphate in fertilizers (Measurement procedure)
94
Testing Methods for Fertilizers (2018)
Comment 1 The coloring reagent solution in d) can also be prepared by the following method.
Dissolve 2.24 g of ammonium vanadate (V) (3) specified in JIS K 8747 in water, add
300 mL of nitric acid, and add water to make 1000 mL. Separately, add 100 g of
95
Testing Methods for Fertilizers (2018)
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.2.2.b.
Comment 4 When the determination is affected by the coloring of the sample solution of d) and
h), transfer the predetermined volume (equivalent volume) (7) of the sample solution
(1) and the sample solution (2) to a 100-mL volumetric flask, add a few drops of
hydrochloric acid (1+1) to acidify, then add no more than 0.1 g of active carbon.
After leaving at rest for a little while, add water up to the marked line and filter. The
filtrate is to be the mixture solution for the sample solution of (4.2) a). Additionally,
as phosphorus contained in active carbon has the possibility to elute and affects the
determination value, a blank test is required.
Comment 5 The volumetric flask used in the procedure in a) should be distinguished as a flask to
be used for phosphate coloring operation and should not be used for other purposes.
Note (7) The transferred volume of the sample solution (1) and the sample solution (2) should be
equivalent.
(8) When the solution is muddled by adding nitric acid (1+1), conduct centrifugation after
the procedure in e).
(9) When it does not contain non-orthophosphate, the boiling operation is not necessary.
(10) Without the addition of water, precipitate may be produced when a coloring reagent
solution is added.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 420 nm
b) Calibration curve preparation
1) Transfer 1 mL - 12 mL of phosphoric acid standard solution (P2O5 0.5 mg/mL) to 100-mL
volumetric flasks step-by-step.
2) Add 2 mL of Petermans citrate solution, 4 mL of nitric acid (1+1) and a proper amount of
water (10), and conduct the same procedure as (4.2) e) to make the P2O5 0.5 mg/100 mL - 6
mg/100 mL phosphoric acid standard solutions for the calibration curve preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
test solution for the calibration curve preparation.
4) Measure absorbance at a wavelength of 420 nm of the phosphoric acid standard solutions for
the calibration curve preparation using the blank test solution for the calibration curve
preparation as the control (11).
5) Prepare the calibration curve of the phosphoric acid concentration and absorbance of the
phosphoric acid standard solutions for the calibration curve preparation.
c) Sample measurement
1) Regarding the solution in (4.2) e), measure absorbance by the same procedure as b) 4) (11).
2) Obtain the phosphoric acid (P2O5) content from the calibration curve, and calculate soluble
phosphoric acid (S-P2O5) in the analytical sample.
Note (11) Measure within 2 hours after adding the coloring reagent solution.
Comment 6 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) -
20 % (mass fraction) and 1 % (mass fraction) - 5 % (mass fraction) are 99.4 % -
100.6 % and 98.6 % - 103.0 % as soluble phosphoric acid (S-P2O5) respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
97
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
certified reference laboratory Average2) sr
4)
RSD r
5)
s I(T)
6)
RSD I(T)
7)
sR
8)
RSD R
9)
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-B-10 10 8.62 0.04 0.4 0.05 0.6 0.06 0.7
FAMIC-B-14 10 9.18 0.03 0.4 0.04 0.5 0.09 1.0
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting Ammonium vanadomolybdate absorptiometric analysis
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.108- 114, Yokendo, Tokyo (1988)
2) Kimie KATO, Sakiko TAKAHASHI and Yuji SHIRAI: Validation of a Color metric
Method for Determination of Nitrogen, Phosphorus and Boron: Evaluation of
Calibration curve, Research Report of Fertilizer Vol. 2, p. 137 - 144 (2009)
3) Akira SHIMIZU and Shin ABE: Verification of Performance Characteristics of Testing
Method for Citrate-Soluble Phosphorus Content by Ammonium Vanadomolybdate
Absorption Photometry, Research Report of Fertilizer Vol. 5, p. 180 - 189 (2012)
98
Testing Methods for Fertilizers (2018)
(5) Flow sheet for soluble phosphoric acid: The flow sheet for soluble phosphoric acid in
fertilizers is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 6 filter paper
Figure 1 The flow sheet for soluble phosphoric acid in fertilizers (Extraction procedure)
99
Testing Methods for Fertilizers (2018)
Cooling
← A proper amount of water
← 20 mL of coloring reagent solution
← Water (up to the marked line)
Leaving at rest For about 30 minutes
100
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Water bath: Water bath that can be adjusted to 65 °C ± 2 °C and 60 ºC - 65 ºC.
b) Drying apparatus: A drying apparatus that can be adjusted to 220 °C ± 5 ºC.
c) Crucible type glass filter: A crucible type glass filter 1G4 specified in JIS R 3503. Let it
stand to cool in a desiccator after heating at 220 ºC ± 5 ºC in advance and measure the mass to
the order of 1 mg.
250- mL (1) volumetric flask, and add 100 mL of Petermans citrate solution and stopple. Then
shake to mix until the filter paper breaks down completely.
f) Heat the volumetric flask in e) in water bath at 65 ºC ± 2 ºC for 1 hour while shaking to mix
every 15 minutes.
g) After immediate cooling is complete, add water up to the marked line
h) Filter with Type 6 filter paper to make a sample solution (2).
Comment 1 The procedure in (4.1) is the same as the procedure in (4.1) of 4.2.2.a.
Note (2) The transferred volume of the sample solution (1) and the sample solution (2) should be
equivalent.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.98- 106, Yokendo, Tokyo (1988)
102
Testing Methods for Fertilizers (2018)
(5) Flow sheet for soluble phosphoric acid: The flow sheet for soluble phosphoric acid in
fertilizers is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 6 filter paper
103
Testing Methods for Fertilizers (2018)
Filtration
under reduced Crucible type glass filtering apparatus 1G4, 3 times with water
pressure
104
Testing Methods for Fertilizers (2018)
Comment 1 The coloring reagent solution in c) can also be prepared by the following method.
Dissolve 2.24 g of ammonium vanadate (V) (3) specified in JIS K 8747 in water, add
300 mL of nitric acid, and add water to make 1000 mL. Separately, add 100 g of
hexaammonium heptamolybdate tetrahydrate (4) specified in JIS K 8905 while
dissolving in water, and further add water to make 1000 mL. In the case of usage,
mix equal volumes of the two solutions.
Comment 2 Instead of the phosphoric acid standard solution in (2), a phosphoric acid standard
solution for a calibration curve can be prepared using a phosphoric acid standard
solution (P 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
105
Testing Methods for Fertilizers (2018)
Note (6) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1) of 4.2.3.b, (4.1) of
4.2.3.c, (4.1) of 4.2.3.d, (4.1.1) of 4.3.2.a, (4.1) of 4.3.2.b, (4.1) of 4.3.2.c, (4.1.1) of
4.3.2.d, (4.1.1) of 4.6.3.a, (4.1.1) of 4.6.3.b, (4.1.1) of 4.7.2.a, (4.1.1) of 4.7.2.b,
(4.1.1) of 4.8.1.a and (4.1.1) of 4.8.1.b.
Note (7) Use a 250-mL flat-bottom volumetric flask to stabilize the shaking.
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.2) of 4.2.3.b, (4.1.2) of
4.2.3.d, (4.1.2) of 4.3.2.a, (4.1.2) of 4.3.2.d, (4.1.2) of 4.6.3.a, (4.1.2) of 4.6.3.b,
(4.1.2) of 4.7.2.a, (4.1.2) of 4.7.2.b, (4.1.2) of 4.8.1.a and (4.1.2) of 4.8.1.b.
106
Testing Methods for Fertilizers (2018)
Comment 5 For a by-product phosphate fertilizer, if the pH of the sample solution in (4.1.1) d)
and (4.1.2) d) is neutral or basic, prepare a sample solution anew by replacing “1 g of
an analytical sample” in the procedures in (4.1.1) a) and (4.1.2) a) with “0.5 g of an
analytical sample”.
Comment 6 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
Comment 7 When the determination is affected by the coloring of the sample solution of (4.1.1)
d) and (4.1.2) d), transfer the predetermined volume of the sample solution to a
100-mL volumetric flask, add a few drops of hydrochloric acid (1+1) to acidify, then
add no more than 0.1 g of active carbon. After leaving at rest for a little while, add
water up to the marked line and filter. The filtrate is prepared as the sample solution
of (4.2) a). Additionally, as phosphorus contained in active carbon has the possibility
to elute and affects the determination value, a blank test is required.
Comment 8 The volumetric flask used in the procedure in a) should be distinguished as a flask to
be used for phosphate coloring operation and should not be used for other purposes.
Note (8) When the solution is muddled by adding nitric acid (1+1), conduct centrifugation after
the procedure in e).
(9) When it does not contain non-orthophosphate, the boiling operation is not necessary.
(10) Without the addition of water, precipitate may be produced when a coloring reagent
solution is added.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 420 nm
b) Calibration curve preparation
1) Transfer 1 mL - 12 mL of phosphoric acid standard solution (P2O5 0.5 mg/mL) to 100-mL
volumetric flasks step-by-step.
2) Add 17 mL of a citric acid solution, then add 4 mL of nitric acid (1+1), further add a proper
amount of water (10) and conduct the same procedure as (4.2) e) to make the P2O5 0.5 mg/100
mL - 6 mg/100 mL phosphoric acid standard solution for the calibration curve preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
test solution for the calibration curve preparation.
4) Measure absorbance at a wavelength of 420 nm of the phosphoric acid standard solutions for
the calibration curve preparation using the blank test solution for the calibration curve
107
Testing Methods for Fertilizers (2018)
Note (11) Measure within 2 hours after adding the coloring reagent solution.
Comment 9 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) -
20 % (mass fraction) and 1 % (mass fraction) - 5 % (mass fraction) are 96.6 % -
103.4 % and 102.0 % - 103.8 % as citrate soluble phosphoric acid (C-P2O5)
respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.03 % (mass fraction) for solid fertilizers, and 0.01 % (mass fraction) for fluid
fertilizers.
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 4) 5) 6) 7) 8) 9)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-A-10 11 10.05 0.04 0.4 0.05 0.5 0.13 1.3
FAMIC-A-13 10 10.79 0.06 0.6 0.08 0.8 0.09 0.8
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting Ammonium vanadomolybdate absorptiometric analysis
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.108- 114, Yokendo, Tokyo (1988)
2) Kimie KATO, Sakiko TAKAHASHI and Yuji SHIRAI: Validation of a Color metric
Method for Determination of Nitrogen, Phosphorus and Boron: Evaluation of
Calibration curve, Research Report of Fertilizer Vol. 2, p. 137 - 144 (2009)
3) Yoshiyuki SUNAGA, Yasushi SUGIMURA, Ichiro YOSHIDA and Hidenori KONISHI:
Verification of Performance Characteristics of Testing Methods for Phosphorus Content
in Fertilizer by Ammonium Vanadomolybdate Absorption Photometry, Research Report
of Fertilizer Vol. 5, p. 167 - 179 (2012)
108
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate soluble phosphoric acid: The flow sheet for citrate soluble
phosphoric acid in fertilizers is shown below:
Sample solution
Figure 1-1 Flow sheet for citrate soluble phosphoric acid in fertilizers
(Extraction procedure (4.1.1))
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-2 Flow sheet for citrate soluble phosphoric acid in fertilizers
(Extraction procedure (4.1.2))
109
Testing Methods for Fertilizers (2018)
Sample solution
Cooling
← A proper amount of water
← 20 mL of coloring reagent solution
← Water (up to the marked line)
Leaving at rest For about 30 minutes
110
Testing Methods for Fertilizers (2018)
Comment 1 The coloring reagent solution in d) can also be prepared by the following method.
Dissolve 2.24 g of ammonium vanadate (V) (3) specified in JIS K 8747 in water, add
300 mL of nitric acid, and add water to make 1000 mL. Separately, add 100 g of
hexaammonium heptamolybdate tetrahydrate (4) specified in JIS K 8905 while
dissolving in water, and further add water to make 1000 mL. In the case of usage,
mix equal volumes of the two solutions.
Comment 2 Instead of the phosphoric acid standard solution in (2), a phosphoric acid standard
solution for a calibration curve can be prepared using a phosphoric acid standard
111
Testing Methods for Fertilizers (2018)
Note (6) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (7) Use a 250-mL flat-bottom volumetric flask to stabilize the shaking.
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 5 For a by-product phosphate fertilizer, if the pH of the sample solution in (4.1.1) d)
and (4.1.2) d) is neutral or basic, prepare a sample solution anew by replacing “1 g of
112
Testing Methods for Fertilizers (2018)
Note (8) A watch glass should not be uncovered when bubbles form while heating because they
may splash.
(9) It is recommended to transfer 2 mL of water to a 100-mL - 200-mL tall beaker in
advance and confirm the volume.
(10) The volume of the solution after transferring should be up to about 50 mL.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 420 nm
b) Calibration curve preparation
1) Transfer 1 mL - 12 mL of phosphoric acid standard solution (P2O5 0.5 mg/mL) to 100-mL
volumetric flasks step-by-step.
2) Add 17 mL of a citric acid solution, then add 4 mL of nitric acid (1+1), further add a proper
amount of water (11). Conduct the same procedure as (4.2) f) to make the P2O5 0.5 mg/100 mL
- 6 mg/100 mL phosphoric acid standard solutions for the calibration curve preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
test solution for the calibration curve preparation.
4) Measure absorbance at a wavelength of 420 nm of the phosphoric acid standard solutions for
the calibration curve preparation using the blank test solution for the calibration curve
preparation as the control (12).
5) Prepare the calibration curve of the phosphoric acid concentration and absorbance of the
phosphoric acid standard solutions for the calibration curve preparation.
c) Sample measurement
1) Regarding the solution in (4.2) f), measure absorbance by the same procedure as b) 4) (12).
2) Obtain the phosphoric acid (P2O5) content from the calibration curve, and calculate citrate
soluble phosphoric acid (C-P2O5) in the analytical sample.
Note (11) If water is not added, precipitate may form when adding the coloring reagent solution.
(12) Measure within 2 hours after adding the coloring reagent solution in the procedure in
113
Testing Methods for Fertilizers (2018)
(4.2) f).
Comment 7 Additive recovery testing was conducted to evaluate trueness using solid fertilizers
(10 samples) containing the equivalents of 1.03 % (mass fraction) - 51.04 % (mass
fraction) as citrate soluble phosphoric acid. As a result, the average recovery is 99 %
- 100 %.
The results of the repeatability tests on different days using a solid preparation
sample to evaluate precision were analyzed by one-way analysis of variance. Table 1
shows the calculation results of intermediate precision and repeatability.
Table 2 shows results and analysis results from a collaborative study for testing
method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.05 % (mass fraction).
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Preparation sample (solid) 1 5 51.01 0.12 0.2 0.16 0.3
Preparation sample (solid) 2 5 2.57 0.01 0.6 0.03 1.1
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
114
Testing Methods for Fertilizers (2018)
References
1) Toshiaki HIROI, Masayuki YAMANISHI Determination Method for Citric
Acid-Soluble Phosphorus in Solid Fertilizer Containing Phosphonate (Phosphite) using
Spectrophotometer, Research Report of Fertilizer, Vol.9, p. 43 - 58 (2016)
2) Masayuki YAMANISHI, Toshiaki HIROI and Fumika TAKATSU Determination
Method for Citric Acid-Soluble and Water-Soluble Phosphorus in Solid Fertilizer
Containing Phosphonic Acid or Phosphonate (Phosphite) using Spectrophotometer: A
Collaborative Study, Vol.9, p. 59 - 68 (2016)
115
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate-soluble phosphoric acid containing phosphorus acid, etc.: The flow
sheet for citrate-soluble phosphoric acid in fertilizers containing phosphorus acid, etc.is
shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
116
Testing Methods for Fertilizers (2018)
Sample solution
Standing to cool
117
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Constant-temperature rotary shaker: A constant-temperature rotary shaker that can rotate
a 250-mL volumetric flask, set up in a thermostat adjustable to 30 ºC ± 1 ºC, upside down at
30 - 40 revolutions/min.
b) Water bath: Water bath that can be adjusted to 60 ºC - 65 ºC.
c) Drying apparatus: A drying apparatus that can be adjusted to 220 ºC ± 5 ºC.
d) Crucible type glass filter: A crucible type glass filter 1G4 specified in JIS R 3503. Let it
stand to cool in a desiccator after heating at 220 ºC ± 5 ºC in advance and measure the mass to
the order of 1 mg.
Note (2) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
118
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.98- 106, Yokendo, Tokyo (1988)
119
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate soluble phosphoric acid: The flow sheet for citrate soluble
phosphoric acid in fertilizers is shown below:
Sample solution
Figure 1 Flow sheet for citrate soluble phosphoric acid in fertilizers (Extraction procedure)
120
Testing Methods for Fertilizers (2018)
Sample solution
Standing to cool
← Water (up to the marked line)
Filtration Type 3 filter paper
Aliquot
300-mL tall beaker
(predetermined volume)
← 5 mL nitric acid
← Water (to reach about 80 mL)
Cover with a watch glass for 3 minutes
Boiling
Wash a watch glass and the inside the tall beaker with water
← Water (to be about 100 mL)
← 50mL of quimosiac solution
Forming precipitation 60 ºC - 65 ºC, for 15 minutes, sometimes mixing
Filtration
under reduced Crucible type glass filtering apparatus 1G4, 3 times with water
pressure
Figure 2 Flow sheet for citrate soluble phosphoric acid in fertilizers (Measurement procedure)
121
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the phosphoric acid standard solution in (2), a phosphoric acid standard
solution for a calibration curve can be prepared using a phosphoric acid standard
solution (P 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate citrate-soluble phosphoric acid (C-P2O5) in an analytical sample
by multiplying the concentration (P) of a phosphoric acid standard solution for a
calibration curve or a measured value obtained in (4.2) by a conversion factor
(2.2914).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in f) and g) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a phosphoric acid
122
Testing Methods for Fertilizers (2018)
standard solution for the calibration curve in the concentration range which is
suitable to a device used.
Note (2) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (3) Use a 250-mL flat-bottom volumetric flask to stabilize the shaking.
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 5 For a by-product phosphate fertilizer, if the pH of the sample solution in (4.1.1) d)
and (4.1.2) d) is neutral or basic, prepare a sample solution anew by replacing “1 g of
an analytical sample” in the procedures in (4.1.1) a) and (4.1.2) a) with “0.5 g of an
analytical sample”.
Comment 6 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
123
Testing Methods for Fertilizers (2018)
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 178.287 nm
b) Calibration curve preparation
1) Spray the phosphoric acid standard solution for the calibration curve preparation and blank
test solution for the calibration curve preparation into inductively coupled plasma, and read
the indicated value at a wavelength of 178.287 nm.
2) Prepare a curve for the relationship between the phosphoric acid concentration and the
indicated value of the phosphoric acid standard solution for the calibration curve preparation
and the blank test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.5 mg - 40 mg as
P2O5) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the phosphoric acid (P2O5) content from the calibration curve, and calculate citrate
soluble phosphoric acid (C-P2O5) in the analytical sample.
124
Testing Methods for Fertilizers (2018)
Comment 8 The comparison of the measurement value (yi:1.74 % (mass fraction) - 49.04 %
(mass fraction )) of ICP Optical Emission Spectrometry and the measurement value
(xi) of Ammonium vanadomolybdate absorptiometric analysis was conducted to
evaluate trueness using processed phosphorus fertilizers (2 samples), compound
fertilizers (12 samples), home garden-use mixed fertilizers (1 sample), mixed
compost fertilizers (2 samples), mixed phosphate fertilizers (2 samples), designated
blended fertilizers (4samples), blended fertilizers (5 samples), byproduct mixed
fertilizers (1 sample), byproduct phosphorus fertilizers (2 samples), organic
compound fertilizers (1 sample) and fused phosphate fertilizers (1 sample). As a
result, a regression equation was y = − 0.0027 + 1.001x, and its correlation
coefficient (r) was 1.000. In addition, recovery testing was conducted using a
preparation sample. As a result, the average rate of recovery at the content level of
0.260 % (mass fraction) - 49.99 % (mass fraction) was 96.3 % - 100.8 %.
The results of the repeatability tests on different days using compound fertilizers and
blended fertilizer to evaluate precision were analyzed by one-way analysis of
variance. Table 2 shows the calculation results of intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.01 % (mass fraction).
125
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Simultaneous Determination of Water-Soluble Principal Ingredients
(W-P2O5, W-K2O, W-MgO, W-MnO and W-B2O3) in Liquid Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES), Research
Report of Fertilizer, Vol.8, p. 1 - 9 (2015)
(5) Flow sheet for citrate soluble phosphoric acid: The flow sheet for citrate soluble
phosphoric acid in fertilizers is shown below:
Sample solution
Figure 1-1 Flow sheet for citrate soluble phosphoric acid in fertilizers
(Extraction procedure (4.1.1))
126
Testing Methods for Fertilizers (2018)
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-2 Flow sheet for citrate soluble phosphoric acid in fertilizers
(Extraction procedure (4.1.2))
Sample solution
127
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the phosphoric acid standard solution in (2), a phosphoric acid standard
solution for a calibration curve can be prepared using a phosphoric acid standard
solution (P 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate water-soluble phosphoric acid (W-P2O5) in an analytical sample
by multiplying the concentration (P) of a phosphoric acid standard solution for a
calibration curve or a measured value obtained in (4.3) by a conversion factor
(2.2914).
128
Testing Methods for Fertilizers (2018)
Comment 2 In the procedure of (4.1.1.1) a), it is also allowed to weigh 2.5g of an analytical
sample to the order of 1 mg, and put it in a 250-mL volumetric flask
Comment 3 The procedure in (4.1.1.1) is the same as the procedure in (4.1.1) of 4.1.2.b, (4.1.1.1)
of 4.2.4.b, (4.1) of 4.2.4.c, (4.1) of 4.2.4.d, (4.1.2.1) of 4.3.3.a, (4.1.2) of 4.3.3.b,
(4.1.2) of 4.3.3.c, (4.1.1.1) of 4.7.3.a, (4.1.1) of 4.9.2.a, (4.1.1) of 4.10.2.a, (4.1.1) of
4.13.1.a and (4.1.1) of 4.14.1.a.
Comment 4 When the determination is affected by the coloring of the sample solution of (4.1.1.1)
d), (4.1.1.2) d) and (4.1.2) d), transfer the predetermined volume of the sample
solution to a 100-mL volumetric flask, add a few drops of hydrochloric acid (1+1) to
acidify, then add no more than 0.1 g of active carbon. After leaving at rest for a little
while, add water up to the marked line and filter. The filtrate is prepared as the
sample solution of (4.2) a). Additionally, as phosphorus contained in active carbon
has the possibility to elute and affects the determination value, a blank test is
required.
Comment 5 The procedure in (4.1.1.2) is the same as the procedures in (4.1.1.2) of 4.2.4.b,
(4.1.2.2) of 4.3.3.a and (4.1.1.2) of 4.7.3.a.
Comment 6 The procedure in (4.1.2) is the same as the procedure in (4.1.2) of 4.2.4.b, (4.1.3) of
4.3.3.a, (4.1) of 4.3.3.d, (4.1) of 4.5.3.b, (4.1) of 4.6.3.b, (4.1.2) of 4.7.3.a, (4.1) of
4.7.3.b, (4.1) of 4.8.2.b, (4.1.2) of 4.9.2.a, (4.1.2) of 4.9.2.b, (4.1.2) of 4.10.2.a, (4.1)
of 4.10.2.b, (4.1.2) of 4.13.1.a, (4.1) of 4.13.1.b, (4.1.2) of 4.14.1.a, (4.1) of 4.14.1.b,
(4.1.2) of 4.14.1.a, (4.1) of 4.14.1.b, (4.1) of 4.15.1.a and (4.1) of 4.15.1.b.
Comment 7 The volumetric flask used in the procedure in a) should be distinguished as a flask to
be used for phosphate coloring operation and should not be used for other purposes.
Note (6) When the solution is muddled by adding nitric acid (1+1), conduct centrifugation after
the procedure in e).
(7) When it does not contain non-orthophosphate, the procedure in b) is not necessary.
(8) Without the addition of water, precipitate may be produced when a coloring reagent
solution is added.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 420 nm
b) Calibration curve preparation
1) Transfer 1 mL - 12 mL of phosphoric acid standard solution (P2O5 0.5 mg/mL) to 100-mL
volumetric flasks step-by-step.
2) Add a proper amount of water (8), and conduct the same procedure as (4.2) e) to make the P2O5
0.5 mg/100 mL - 6 mg/100 mL phosphoric acid standard solutions for the calibration curve
preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
test solution for the calibration curve preparation.
4) Measure absorbance at a wavelength of 420 nm of the phosphoric acid standard solutions for
the calibration curve preparation using the blank test solution for the calibration curve
preparation as the control (9).
5) Prepare the calibration curve of the phosphoric acid concentration and absorbance of the
phosphoric acid standard solutions for the calibration curve preparation.
c) Sample measurement
130
Testing Methods for Fertilizers (2018)
1) Regarding the solution in (4.2) e), measure absorbance by the same procedure as b) 4) (9).
2) Obtain the phosphoric acid (P2O5) content from the calibration curve, and calculate
water-soluble phosphoric acid (W-P2O5) in the analytical sample.
Note (9) Measure within 6 hours after adding the coloring reagent solution in the procedure in
(4.2) e).
Comment 8 After the procedure in (4.2) a), it is also possible to measure soluble phosphoric acid
at the same time by adding 2 mL of Petermans citrate solution and by conducting the
procedures from (4.2) d) to (4.3) in 4.2.2.a (using b reagent solution in the Official
Methods of Analysis of Fertilizers (1992)).
Further after the procedure in (4.2) a), it is also possible to measure citrate soluble
phosphoric acid at the same time by adding 17 mL of citrate solution and by
conducting the procedures from (4.2) c) to (4.3) in 4.2.3.a (using b reagent solution
in the Official Methods of Analysis of Fertilizers (1992)).
Comment 9 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) -
20 % (mass fraction) and 1 % (mass fraction) - 5 % (mass fraction) are 100.5 % -
101.2 % and 99.0 % - 101.7 % as water-soluble phosphoric acid (W-P2O5)
respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
The comparison of the measurement value of extraction (yi:0.292 % (mass fraction)
- 40.40 % (mass fraction)) with a vertical reciprocating shaker and the measurement
value of extract with a rotary shaker (xi) was conducted to evaluate trueness of the
extraction of solid fertilizers using fertilizers (12 samples). As a result, a regression
equation was y = − 0.041 + 0.999x, and its correlation coefficient (r) was 1.000. Also,
the results of the repeatability tests on different days using compound fertilizers and
designated blended fertilizers to evaluate precision were analyzed by one-way
analysis of variance. Table 2 shows the calculation results of intermediate precision
and repeatability
The comparison of the measurement value of simple extraction (yi:1.92 % (mass
fraction) - 12.21 % (mass fraction )) and the measurement value of extract (xi) with a
rotary shaker was conducted to evaluate trueness of the extraction of fluid fertilizers
using fluid fertilizers (12 samples). As a result, a regression equation was y =
0.005+1.005x, and its correlation coefficient (r) was 0.999. Also, the results of the
repeatability tests on different days using fluid mixed fertilizers to evaluate the
precision of the extraction of fluid fertilizers were analyzed by the one-way analysis
of variance. Table 3 shows the calculation results of intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.03 % (mass fraction) for solid fertilizers, and 0.004 % (mass fraction) for fluid
fertilizers.
131
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
certified reference laboratory Average2) sr
4)
RSD r
5)
s I(T)
6)
RSD I(T)
7)
sR
8)
RSD R
9)
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-B-10 9 7.00 0.02 0.3 0.03 0.5 0.07 1.0
FAMIC-B-14 15 6.70 0.02 0.3 0.03 0.5 0.06 0.9
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting Ammonium vanadomolybdate absorptiometric analysis
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Designated blended
7 13.77 0.03 0.2 0.06 0.5
fertilizer
Compound fertilizer 7 1.19 0.01 0.5 0.01 0.5
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
132
Testing Methods for Fertilizers (2018)
133
Testing Methods for Fertilizers (2018)
(5) Flow sheet for water-soluble phosphoric acid: The flow sheet for water-soluble
phosphoric acid: in fertilizers is shown below:
5 g analytical sample
Weigh to the order of 1 mg into a 500-mL volumetric flask
(powdery)
← Water, about 400 mL
Shaking to mix Rotary shaker (30 - 40 revolutions/min) for 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
1 g analytical sample
Weigh to the order of 1 mg into a 100-mL volumetric flask
(fluid)
← Water, about 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
134
Testing Methods for Fertilizers (2018)
Sample solution
Cooling
← 1-2 drop(s) of phenolphthalein solution (1 g/100 mL)
← Ammonia solution (1+1) [neutralization]
← Nitric acid (1+10) [slightly acidic]
← A proper amount of water
← 20 mL of coloring reagent solution
← Water (up to the marked line)
Leaving at rest For about 30 minutes
135
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the phosphoric acid standard solution in (2), a phosphoric acid standard
solution for a calibration curve can be prepared using a phosphoric acid standard
solution (P 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate water-soluble phosphoric acid (W-P2O5) in an analytical sample
by multiplying the concentration (P) of a phosphoric acid standard solution for a
calibration curve or a measured value obtained in (4.3) by a conversion factor
(2.2914).
136
Testing Methods for Fertilizers (2018)
Comment 2 In the procedure of (4.1.1.1) a), it is also allowed to weigh 2.5g of an analytical
sample to the order of 1 mg, and put it in a 250-mL volumetric flask
Comment 3 The procedure in (4.1.1.1) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
Comment 4 The procedure in (4.1.1.2) is the same as the procedure in (4.1.1.2) in 4.2.4.a.
Comment 5 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.4.a.
c) Cover the tall beaker with a watch glass, heat on a hot plate or sand bath at 200 ºC - 250 ºC
and condense (7) until the solution volume becomes about 2 mL (6).
d) After standing to cool, transfer it with water to a 100- mL volumetric flask (8).
e) Add 1 - 2 drop (s) of phenolphthalein solution (1 g/100 mL), and neutralize by adding
ammonia solution (1+1) until the color of the solution becomes light red-purple.
f) Add nitric acid (1+10) until the light red-purple color of the solution disappears to make it
slightly acidic.
g) Add 20 mL of coloring reagent solution, and further add water up to the marked line, and then
leave at rest for about 30 minutes.
Note (6) It is recommended to transfer about 2 mL of water to a 100-mL - 200-mL tall beaker in
advance and confirm the volume.
(7) Care should be taken not to dry and harden it. When it dries completely, the
determined value becomes lower than usual in some cases.
(8) The volume of the solution after transferring should be up to about 50 mL.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 420 nm
b) Calibration curve preparation
1) Transfer 1 mL - 12 mL of phosphoric acid standard solution (P2O5 0.5 mg/mL) to 100-mL
volumetric flasks step-by-step.
2) Add a proper amount of water (9), and conduct the same procedure as (4.2) g) to make the
P2O5 0.5 mg/100 mL - 6 mg/100 mL phosphoric acid standard solutions for the calibration
curve preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
test solution for the calibration curve preparation.
4) Measure absorbance at a wavelength of 420 nm of the phosphoric acid standard solutions for
the calibration curve preparation using the blank test solution for the calibration curve
preparation as the control (10).
5) Prepare the calibration curve of the phosphoric acid concentration and absorbance of the
phosphoric acid standard solutions for the calibration curve preparation.
c) Sample measurement
1) Regarding the solution in (4.2) g), measure absorbance by the same procedure as b) 4) (10).
2) Obtain the phosphoric acid (P2O5) content from the calibration curve, and calculate
water-soluble phosphoric acid (W-P2O5) in the analytical sample.
Note (9) If water is not added, precipitate may form when adding the coloring reagent solution.
(10) Measure within 6 hours after adding the coloring reagent solution in the procedure in
(4.2) g).
Comment 6 Recovery testing was conducted to evaluate trueness using a fluid preparation sample.
As a result, the average rate of recovery at the content level of 30 % (mass fraction) -
50 % (mass fraction), 10 % (mass fraction) - 20 % (mass fraction), 4 % (mass
fraction) and 0.2 % (mass fraction) are 101.1 % - 101.8 %, 101.1 % - 101.5 %,
100.8 % and 102.5 % as water-soluble phosphoric acid (W-P2O5) respectively. In the
case of using a solid preparation sample, the average rate of recovery at the content
138
Testing Methods for Fertilizers (2018)
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Preparation sample (solid) 1 7 59.36 0.09 0.2 0.13 0.2
Preparation sample (solid) 2 7 5.90 0.07 1.2 0.07 1.2
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
139
Testing Methods for Fertilizers (2018)
140
Testing Methods for Fertilizers (2018)
References
1) Toshiaki HIROI, Masakazu SAIKI and Kimie KATO: Development and Validation of
Spectrophotometry for Determination of Water-Soluble Phosphoric acid in Liquid
Mixed Fertilizer Containing Phosphonic acid, Research Report of Fertilizer, Vol.1 P. 25
- 33 (2008)
2) Toshiaki HIROI, Masakazu SAIKI and Kimie KATO: Determination of Water-Soluble
Phosphoric Acid in Liquid Mixed Fertilizer Containing Phosphonic Acid by
Spectrophotometry: A Collaborative Study, Research Report of Fertilizer, Vol.1 p. 34 -
40 (2008)
3) Fumihiro ABE, Noriyuki SASAKI and Toshio HIRABARA: Determination of
Water-Soluble Phosphorus in Solid Fertilizer Containing Phosphonic Acid by
Spectrophotometry, Research Report of Fertilizer, Vol.8, p. 10 - 16 (2015)
4) Masayuki YAMANISHI, Toshiaki HIROI and Fumika TAKATSU Determination
Method for Citric Acid-Soluble and Water-Soluble Phosphorus in Solid Fertilizer
Containing Phosphonic Acid or Phosphonate (Phosphite) using Spectrophotometer: A
Collaborative Study, Vol.9, p. 59 - 68 (2016)
(5) Flow sheet for water-soluble phosphoric acid containing phosphorus acid, etc.: The flow
sheet for water-soluble phosphoric acid in fertilizers containing phosphorus acid, etc.is shown
below:
5 g analytical sample
Weigh to the order of 1 mg into a 500-mL volumetric flask
(powdery)
← Water, about 400 mL
Shaking to mix Rotary shaker (30 - 40 revolutions/min) for 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
141
Testing Methods for Fertilizers (2018)
1 g analytical sample
Weigh to the order of 1 mg into a 1000-mL volumetric flask
(fluid)
← Water, about 200 mL
Shaking to mix
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Standing to cool
142
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Rotary shaker: A rotary shaker that can rotate a 500-mL volumetric flask upside down at 30
- 40 revolutions/min.
b) Water bath: Water bath that can be adjusted to 60 ºC - 65 ºC.
c) Drying apparatus: A drying apparatus that can be adjusted to 220 ºC ± 5 ºC.
d) Crucible type glass filter: A crucible type glass filter 1G4 specified in JIS R 3503. Let it
stand to cool in a desiccator after heating at 220 ºC ± 5 ºC in advance and measure the mass to
the order of 1 mg.
Comment 1 In the procedure in a), it is also allowed to weigh 2.5 g of an analytical sample to the
order of 1 mg, and put it into a 250-mL volumetric flask.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 4.2.4.a.
inside the tall beaker with water and add water to make 100 mL.
d) Immediately, add 50mL of quimosiac solution, heat for about 15 minutes while sometimes
mixing in a water bath at 60 ºC - 65 ºC to produce the precipitate of quinolinium
phosphomolybdate.
e) After standing to cool down to room temperature while sometimes mixing, filter under
reduced pressure with a crucible type glass filter, wash the tall beaker 3 times with water and
transfer the whole precipitate into a crucible type glass filter, and further wash 7 - 8 times with
water.
f) Transfer the precipitate together with the crucible type glass filter into a drying apparatus and
heat at 220 ºC ± 5 ºC for about 30 minutes.
g) As soon as heating is complete, move it into a desiccator and let it stand to cool.
h) After standing to cool, remove the crucible type glass filter from the desiccator and measure
the mass to the order of 1 mg.
i) Calculate water-soluble phosphoric acid (W-P2O5) by the following formula.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.98- 114, Yokendo, Tokyo (1988)
(5) Flow sheet for water-soluble phosphoric acid: The flow sheet for water-soluble phosphoric
acid: in fertilizers is shown below:
Sample solution
144
Testing Methods for Fertilizers (2018)
Sample solution
Filtration
Crucible type glass filtering apparatus 1G4, 3 times with water
under reduced pressure
145
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the phosphoric acid standard solution in (2), a phosphoric acid standard
solution for a calibration curve can be prepared using a phosphoric acid standard
solution (P 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case,
calculate water-soluble phosphoric acid (W-P2O5) in an analytical sample by
multiplying the concentration (P) of a phosphoric acid standard solution for a
calibration curve or a measured value obtained in (4.2) by a conversion factor
(2.2914).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal
observation mode and an axial observation mode. The range of the concentration of
the standard solution for the calibration preparation curve in d) and e) applies to a
horizontal observation mode. While an axial observation mode can observe a
measurement component of low-level concentrations, it cannot gain the linearity of
a calibration curve in the range of high-level concentrations. Therefore, when using
the ICP-OES of an axial observation mode, it is recommended to prepare a
phosphoric acid standard solution for the calibration curve in the concentration
146
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: A spectrophotometer specified in JIS K 0115
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
Note (2) The sampling amount of the analytical sample is 10 g when there is less phosphoric
acid content in the fertilizers such as a home garden-use fertilizer.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the measurement
conditions for the ICP Atomic Emission Spectrometer considering the following:
Analytical line wavelength: 178.287 nm
b) Calibration curve preparation
1) Spray the phosphoric acid standard solution for the calibration curve preparation and blank
test solution for the calibration curve preparation into inductively coupled plasma, and read
the indicated value at a wavelength of 178.287 nm.
2) Prepare a curve for the relationship between the phosphoric acid concentration and the
indicated value of the phosphoric acid standard solution for the calibration curve preparation
and the blank test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.5 mg - 40 mg as
P2O5) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the phosphoric acid (P2O5) content from the calibration curve, and calculate
water-soluble phosphoric acid (W-P2O5) in the analytical sample.
Comment 5 The comparison of the measurement value (yi:0.179 % (mass fraction) - 10.88 %
(mass fraction )) of ICP Optical Emission Spectrometry and the measurement value
(xi) of Ammonium vanadomolybdate absorptiometric analysis was conducted to
evaluate trueness using fluid fertilizers (12 samples). As a result, a regression
equation was y = − 0.022 + 1.008x, and its correlation coefficient (r) was 0.999.
Additionally, additive recovery testing was conducted using a fluid mixed fertilizer
(1 brand) and a home garden-use mixed fertilizer (1 brand). As a result, the mean
recovery rates at additive level of 10 % (mass fraction) and 1 % (mass fraction) were
98.1 % and 101.9 % respectively.
The results of the repeatability tests on different days using a fluid mixed fertilizer
and a home garden-use mixed fertilizer to evaluate precision were analyzed by the
one-way analysis of variance. Table 2 shows the calculation results of intermediate
precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.02 % (mass fraction).
148
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Simultaneous Determination of Water-Soluble Principal
Ingredients (W-P2O5, W-K2O, W-MgO, W-MnO and W-B2O3) in Liquid Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES), Research
Report of Fertilizer, Vol.8, p. 1 - 9 (2015)
(5) Flow sheet: The flow sheet for water-soluble phosphoric acid of the fluid mixed fertilizers is
shown below:
Sample solution
Figure 1 The flow sheet for water-soluble phosphoric acid in fluid fertilizers
(Extraction procedure)
149
Testing Methods for Fertilizers (2018)
Sample solution
Figure 2 The flow sheet for water-soluble phosphoric acid in fluid fertilizers
(Measurement procedure)
150
Testing Methods for Fertilizers (2018)
4.3 Potassium
4.3.1 Total potassium
4.3.1.a Flame atomic absorption spectrometry or flame photometry
(1) Summary
This testing method is applicable to fertilizers containing organic matters. This testing method is
classified as Type C and its symbol is 4.3.1.a-2017 or T-K.a-1.
Pretreat an analytical sample with incineration and hydrochloric acid to convert the total potassium
into potassium ion, add an interference suppressor solution, and then spray in an acetylene-air flame,
and measure the atomic absorption with potassium at a wavelength of 766.5 nm or 769.9 nm to
quantify the total potassium. Or, determine the intensity of the emission line at a wavelength of
766.5 nm or 769.9 nm produced in flame to obtain the total potassium (T-K2O) in an analytical
sample. In addition, the performance of this testing method is shown in Comment 4.
Comment 1 Instead of the potassium standard solution in (2), a potassium standard solution for
the calibration curve preparation can be prepared by using a potassium standard
solution (K 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case,
calculate total potassium (T-K2O) in the analytical sample by multiplying the
concentration (K) of a potassium standard solution for calibration curve preparation
or a measurement value (K) obtained in (4.2) by a conversion factor (1.2046).
JIS K 0121.
1) Light source: A potassium hollow cathode lamp
2) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
ab) Flame photometer:
1) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
b) Electric furnace: An electric furnace that can be adjusted to 550 ºC ± 5 ºC.
c) Hot plate or sand bath: A hot plate whose surface temperature can be adjusted up to 250 ºC.
Adjust the amounts of gas and silica sand of a sand bath so that the sand bath temperature can
be set to 250 ºC.
Note (3) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 550 °C in 1 to 2 hours.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.2) in 4.2.1.a.
152
Testing Methods for Fertilizers (2018)
Note (4) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(5) The watch glass can be removed.
(6) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 3 The procedures in (4.1.2) are the same as in (4.1.3) in 4.2.1.a. and (4.1) a) - h) in
5.3.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer or flame photometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer or flame photometer:
Set up the measurement conditions for the atomic absorption spectrometer or flame
photometer considering the following:
Analytical line wavelength: 766.5 nm or 769.9 nm
b) Calibration curve preparation
1) Spray the potassium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 766.5 nm or 769.9 nm.
2) Prepare a curve for the relationship between the potassium concentration and the indicated
value of the potassium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.5 mg - 5 mg as
K2O) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (2), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the potassium content from the calibration curve, and calculate total potassium
(T-K2O) in the analytical sample.
Comment 4 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) -
20 % (mass fraction) and 1 % (mass fraction) - 5 % (mass fraction) are 97.8 % -
100.1 % and 100.9 % - 103.1 % as total potassium acid (T-K2O) respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.08 % (mass fraction) for solid fertilizers, and 0.03 % (mass fraction) for fluid
fertilizers.
153
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 4) 5) 6) 7) 8) 9)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-C-12 11 0.584 0.005 0.9 0.011 1.9 0.038 6.5
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting flame atomic absorbance spectrometry
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.132- 138, Yokendo, Tokyo (1988)
2) Kimie KATO, Masayuki YOSHIMOTO and Yuji SHIRAI: Systematization of
Determination Methods of Major Components in Sludge Fertilizer, Compost and
Organic Fertilizer, Research Report of Fertilizer, Vol.3 p. 107 - 116 (2010)
3) Yasuharu KIMURA and Hisanori ARAYA Verification of Performance Characteristics
of Testing Methods for Potassium Content in Fertilizer by Atomic Absorption
Spectrometry, Research Report of Fertilizer Vol. 5, p. 190 - 200 (2012)
154
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total potassium: The flow sheet for total potassium in fertilizers is shown
below:
5 g analytical sample Weigh to the order of 1 mg into a 200-mL - 300-mL tall beaker.
Sample solution
155
Testing Methods for Fertilizers (2018)
5g
Weigh to the order of 1 mg into a 200-mL - 300-mL tall beaker.
analytical sample
Sample solution
Sample solution
156
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Electric furnace: An electric furnace that can be adjusted to 550 ºC ± 5 ºC.
b) Drying apparatus: A drying apparatus that can be adjusted to 120 ºC ± 2 ºC.
c) Crucible type glass filter: A crucible type glass filter 1G4 specified in JIS R 3503. Let it
stand to cool in a desiccator after heating at 120 ºC ± 2 ºC in advance and measure the mass to
the order of 1 mg.
d) Hot plate or sand bath: A hot plate whose surface temperature can be adjusted up to 250 ºC.
Adjust the amounts of gas and silica sand of a sand bath so that the sand bath temperature can
be set to 250 ºC.
157
Testing Methods for Fertilizers (2018)
Note (2) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 550 °C in 1 to 2 hours.
Comment 1 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.1.a. In addition,
the sample solution prepared in (4.1) a) - h) in 4.9.1.a can also be used.
158
Testing Methods for Fertilizers (2018)
Note (3) About 3 ml of tetraphenylborate solution per 10 mg of K2O is required to form the
precipitate of potassium tetraphenylborate
Comment 2 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 25 % (mass fraction) -
30 % (mass fraction) and 10 % (mass fraction) - 20 % (mass fraction) are 99.5 % -
100.8 % and 99.5 % - 100.6 % as total potassium acid (T-K2O) respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.3 % (mass fraction).
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.122- 128, Yokendo, Tokyo (1988)
2) Keiji YAGI, Aiko YANO and Hideo SOETA: Verification of Performance
Characteristics of Testing Method for Nitrate Nitrogen Content in Fertilizer by Phenol
Sulfuric Acid Method, Research Report of Fertilizer Vol. 5, p. 201 - 211 (2012)
159
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total potassium: The flow sheet for total potassium in fertilizers is shown
below:
5 g analytical sample Weigh to the order of 1 mg into a 200-mL - 300-mL tall beaker.
Sample solution
160
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 100-mL tall beaker
amount)
← Water (add tetraphenylborate solution to reach 50 mL)
← Hydrochloric acid (1+9) (the equivalents of 0.2 mL of hydrochloric acid)
← 5 mL of formaldehyde solution
← 5 mL of ethylenediamine tetraacetate- sodium hydroxide solution
← Tetraphenylborate solution (necessary volume + 4 mL)
Forming precipitate For 30 minutes, sometimes mixing
161
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the potassium standard solution in (2), a potassium standard solution for
the calibration curve preparation can be prepared by using a potassium standard
solution (K 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case,
calculate citrate-soluble potassium (C-K2O) in the analytical sample by multiplying
the concentration (K) of a potassium standard solution for calibration curve
preparation or a measurement value (K) obtained in (4.2) by a conversion factor
(1.2046).
30 - 40 revolutions/min.
ab) Reciprocating water bath shaker: A reciprocating water bath shaker: that can be adjusted to
30 °C ± 1 °C and shake a 250-mL volumetric flask, set up vertically on the water surface
using a shaking rack, reciprocate horizontally at 160 times/ minute with amplitude of 25 mm -
40 mm
b) Analytical instrument: An atomic absorption spectrometer or a flame photometer as shown
below:
ba) Flame atomic absorption spectrometer: An atomic absorption spectrometer specified in JIS
K 0121.
1) Light source: A potassium hollow cathode lamp
2) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
bb) Flame photometer:
1) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
Note (3) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (4) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 3 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 4 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
163
Testing Methods for Fertilizers (2018)
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer or flame photometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer or flame photometer:
Set up the measurement conditions for the atomic absorption spectrometer or flame
photometer considering the following:
Analytical line wavelength: 766.5 nm or 769.9 nm
b) Calibration curve preparation
1) Spray the potassium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 766.5 nm or 769.9 nm.
2) Prepare a curve for the relationship between the potassium concentration and the indicated
value of the potassium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.5 mg - 5 mg as
K2O) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (2), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the potassium content from the calibration curve, and calculate citrate soluble
potassium (C-K2O) in the analytical sample.
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) -
20 % (mass fraction) and 1 % (mass fraction) - 5 % (mass fraction) are 100.2 % -
101.7 % and 100.4 % - 101.8 % as citrate soluble potassium (C-K2O) respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.05 % (mass fraction) for solid fertilizers, and 0.06 % (mass fraction) for fluid
fertilizers.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.136- 138, Yokendo, Tokyo (1988)
2) Yasuharu KIMURA and Hisanori ARAYA Verification of Performance Characteristics
of Testing Methods for Potassium Content in Fertilizer by Atomic Absorption
Spectrometry, Research Report of Fertilizer Vol. 5, p. 190 - 200 (2012)
164
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate soluble potassium: The flow sheet for citrate soluble potassium in
fertilizers is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 Flow sheet for citrate-soluble potassium in fertilizers (Extract procedure 4.1.1)
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-2 Flow sheet for citrate-soluble potassium in fertilizers (Extract procedure 4.1.2)
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← About 10 mL interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer or flame photometer
165
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Constant-temperature rotary shaker: A constant-temperature rotary shaker that can rotate
a 250-mL volumetric flask, set up in a thermostat adjustable to 30 ºC ± 1 ºC, upside down at
30 - 40 revolutions/min.
b) Drying apparatus: A drying apparatus that can be adjusted to 120 ºC ± 2 ºC.
c) Crucible type glass filter: A crucible type glass filter 1G4 specified in JIS R 3503. Let it
stand to cool in a desiccator after heating at 120 ºC ± 2 ºC in advance and measure the mass to
the order of 1 mg.
a) Weigh 1 g of an analytical sample to the order of 1 mg, and transfer to a 250-mL volumetric
flask.
b) Add 150 mL of citric acid solution heated up to about 30 ºC (2), and shake to mix at 30 - 40
revolutions/min (30 ºC ± 1 ºC) for 1 hour.
c) After immediate cooling is complete, add water up to the marked line
d) Filter with Type 3 filter paper to make a sample solution.
Note (2) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Note (2) About 3 ml of tetraphenylborate solution per 10 mg of K2O is required to form the
precipitate of potassium tetraphenylborate
Comment 3 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 25 % (mass fraction) -
30 % (mass fraction) and 10 % (mass fraction) - 20 % (mass fraction) are 98.6 % -
100.6 % and 100.6 % - 100.7 % as citrate soluble potassium (C-K2O) respectively.
167
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.122- 128, Yokendo, Tokyo (1988)
2) Keiji YAGI, Aiko YANO and Hideo SOETA: Verification of Performance
Characteristics of Testing Method for Nitrate Nitrogen Content in Fertilizer by Phenol
Sulfuric Acid Method, Research Report of Fertilizer Vol. 5, p. 201 - 211 (2012)
(5) Flow sheet for the testing method: The flow sheet for citrate soluble potassium in fertilizers
is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
168
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
100-mL tall beaker
(20 mL)
← Water (add tetraphenylborate solution to reach 50 mL)
← 5 mL of formaldehyde solution
← 5 mL of ethylenediamine tetraacetate - sodium hydroxide solution
← Tetraphenylborate solution (the equivalents of potassium + 4 mL)
Forming precipitate For 30 minutes, sometimes mixing
169
Testing Methods for Fertilizers (2018)
Note (2) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
170
Testing Methods for Fertilizers (2018)
Note (3) If the solution temperature is no more than 20 ºC, the reaction does not advance in
some cases. Therefore, it is recommended to heat the solution up to about 30 ºC.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.128- 132, Yokendo, Tokyo (1988)
171
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate soluble potassium: The flow sheet for citrate soluble potassium in
fertilizers is shown below:
Standing to cool
← Water (up to the marked line)
Filtration Type 3 filter paper
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Aliquot
100-mL volumetric flask
(5 mL - 15 mL)
← Water (the volume to reach about 30 mL)
← About 5 mL of formaldehyde solution
← 5 mL of sodium hydroxide solution (120 g/L)
← 25 mL tetraphenylborate solution
(one or two drop (s) per second while shaking to mix)
← Water (up to the marked line)
Leaving at rest 10 minutes
172
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the potassium standard solution in (2), a potassium standard solution for
the calibration curve preparation can be prepared by using a potassium standard
solution (K 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case,
calculate citrate-soluble potassium (C-K2O) in the analytical sample by multiplying
the concentration (K) of a potassium standard solution for calibration curve
preparation or a measurement value (K) obtained in (4.2) by a conversion factor
(1.2046).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The axial observation mode does not apply to
potassium since interference is serious.
- 40 revolutions/min.
bb) Reciprocating water bath shaker: A reciprocating water bath shaker that can be adjusted to
30 °C ± 1 °C and shake a 250-mL volumetric flask, set up vertically on the water surface
using a shaking rack, reciprocate horizontally at 160 times/ minute with amplitude of 25 mm -
40 mm.
Note (2) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (3) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 3 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 4 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 766.491 nm
b) Calibration curve preparation
1) Spray the potassium standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 766.491 nm.
2) Prepare a curve for the relationship between the potassium concentration and the indicated
value of the potassium standard solutions for the calibration curve preparation and the
174
Testing Methods for Fertilizers (2018)
Comment 6 The comparison of the measurement value (yi:3.57 % (mass fraction) - 34.24 %
(mass fraction )) of ICP Optical Emission Spectrometry and the measurement value
(xi) of Flame atomic absorption spectrometry analysis was conducted to evaluate
trueness using compound fertilizers (9 samples), mixed compost fertilizers (1
sample), designated blended fertilizers (1sample), blended fertilizers (4 samples),
and byproduct mixed fertilizers (1 sample). As a result, a regression equation was y =
− 0.0058 + 1.0027x, and its correlation coefficient (r) was 0.999. In addition,
recovery testing was conducted using a preparation sample. As a result, the average
rate of recovery at the content level of 0.329 % (mass fraction) - 63.18 % (mass
fraction) was 98.0 % - 100.3 %.
The results of the repeatability tests on different days using compound fertilizers and
blended fertilizers to evaluate precision were analyzed by one-way analysis of
variance. Table 1 shows the calculation results of intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.09 % (mass fraction).
175
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate soluble potassium: The flow sheet for citrate soluble potassium in
fertilizers is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 Flow sheet for citrate-soluble potassium in fertilizers (Extract procedure 4.1.1)
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-2 Flow sheet for citrate-soluble potassium in fertilizers (Extract procedure 4.1.2)
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← 25 mL hydrochloric acid (1+5)
← Water (up to the marked line)
Measurement ICP-OES (766.491 nm)
176
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the potassium standard solution in (2), a potassium standard solution for
the calibration curve preparation can be prepared by using a potassium standard
solution (K 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case,
calculate water-soluble potassium (W-K2O) in the analytical sample by multiplying
the concentration (K) of a potassium standard solution for calibration curve
preparation or a measurement value (K) obtained in (4.2) by a conversion factor
(1.2046).
40 mm).
b) Analytical instrument: An atomic absorption spectrometer or a flame photometer as shown
below:
ba) Flame atomic absorption spectrometer: An atomic absorption spectrometer specified in
JIS K 0121.
1) Light source: A potassium hollow cathode lamp
2) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
bb) Flame photometer:
1) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
c) Hot plate : A hot plate whose surface temperature can be adjusted up to 250 ºC.
Comment 2 In the procedure in a), a 250-mL volumetric flask can be used instead of a 300-mL
tall beaker. However, the volumetric flask used should be distinguished as an
extraction flask and should not be used for the other purposes. Additionally, “cover
with a watch glass” in b) is replaced by “place a funnel”. Skip “transfer to a 250-mL
volumetric flask with water” in the procedure in c).
Comment 3 The procedure in (4.1.1) is the same as the procedures in (4.1.1) of 4.3.3.b, (4.1.1) of
4.3.3.c and (4.1) of 4.8.2.a.
Comment 4 In the procedure of (4.1.2.1) a), it is also allowed to weigh 2.5g of an analytical
sample to the order of 1 mg, and put it in a 250-mL volumetric flask
Comment 5 The procedure in (4.1.2.1) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
Comment 6 The procedure in (4.1.2.2) is the same as the procedure in (4.1.1.2) in 4.2.4.a.
Comment 7 The procedure in (4.1.3) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer or flame photometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer or flame photometer:
Set up the measurement conditions for the atomic absorption spectrometer or flame
photometer considering the following:
Analytical line wavelength: 766.5 nm or 769.9 nm
b) Calibration curve preparation
1) Spray the potassium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 766.5 nm or 769.9 nm.
2) Prepare a curve for the relationship between the potassium concentration and the indicated
value of the potassium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.5 mg - 5 mg as
K2O) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (2), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the potassium content from the calibration curve, and calculate water soluble
potassium (W-K2O) in the analytical sample.
Comment 8 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) -
20 % (mass fraction) and 1 % (mass fraction) - 5 % (mass fraction) are 97.9 % -
100.2 % and 97.3 % - 100.6 % as water soluble potassium (W-K2O) respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
The comparison of the measurement value of extraction (yi:2.69 % (mass fraction) -
26.64 % (mass fraction )) with a vertical reciprocating shaker and the measurement
value of extract with a rotary shaker (xi) was conducted to evaluate trueness of the
extraction of solid fertilizers using fertilizers (12 samples). As a result, a regression
equation was y = 0.022 + 1.001x, and its correlation coefficient (r) was 1.000. Also,
the results of the repeatability tests on different days using compound fertilizers and
179
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
certified reference laboratory Average2) sr
4)
RSD r
5)
s I(T)
6)
RSD I(T)
7)
sR
8)
RSD R
9)
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-A-10 11 13.59 0.08 0.6 0.09 0.6 0.16 1.2
FAMIC-A-13 10 13.07 0.08 0.6 0.11 0.8 0.16 1.2
FAMIC-B-10 9 8.85 0.04 0.4 0.07 0.7 0.12 1.4
FAMIC-B-14 14 8.32 0.03 0.4 0.07 0.8 0.13 1.6
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting frame atomic absorption spectrometry
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
180
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.136- 138, Yokendo, Tokyo (1988)
2) Yasuharu KIMURA and Hisanori ARAYA Verification of Performance
Characteristics of Testing Methods for Potassium Content in Fertilizer by Atomic
Absorption Spectrometry, Research Report of Fertilizer Vol. 5, p. 190 - 200 (2012)
3) Shinji KAWAGUCHI: Simple Extraction Method for Water-Soluble Components
in Liquid Compound Fertilizers, Research Report of Fertilizer, Vol.9, p. 10 - 20
(2016)
4) Shinji KAWAGUCHI: Extraction Method for the Water-soluble Principal
Ingredients in the Solid Fertilizer using a General-purpose Equipment , Research
Report of Fertilizer, Vol.10, p. 1 - 8 (2017)
(5) Flow sheet for water soluble potassium: The flow sheet for water soluble potassium in
fertilizers is shown below:
2.5 g analytical sample Weigh to the order of 1 mg into a 300-mL tall beaker
(potassium salts, etc.)
← About 200 mL water
Heating Cover with a watch dish, boil for 15 minutes
Cooling Immediately
Sample solution
181
Testing Methods for Fertilizers (2018)
5 g analytical sample
Weigh to the order of 1 mg into a 500-mL volumetric flask
(compound fertilizer)
← About 400 mL water
Rotary shaker (30 - 40 revolutions/min)
Shaking to mix
For 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Analytical sample
Weigh to the order of 1 mg into a 100-mL volumetric flask
(fluid)
← About 50 mL water
Shaking to mix
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
182
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
volume)
← About 10 mL of interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer or flame photometer
183
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Rotary shaker: A rotary shaker that can rotate a 500-mL volumetric flask upside down at 30
- 40 revolutions/min.
b) Drying apparatus: A drying apparatus that can be adjusted to 120 ºC ± 2 ºC.
c) Crucible type glass filter: A crucible type glass filter 1G4 specified in JIS R 3503. Let it
stand to cool in a desiccator after heating at 120 ºC ± 2 ºC in advance and measure the mass to
the order of 1 mg.
d) Hot plate : A hot plate whose surface temperature can be adjusted up to 250 ºC.
(4.1.1) Compound fertilizers containing potassium salts and magnesium potassium sulfate
a) Weigh 2.5 g of an analytical sample to the order of 1 mg, and transfer to a 300-mL tall beaker.
b) Add about 200 mL of water, and cover with a watch glass and heat on a hot plate to boil for
about 15 minutes.
c) Immediately transfer it with water to a 250- mL volumetric flask.
d) After immediate cooling is complete, add water up to the marked line
e) Filter with Type 3 filter paper to make a sample solution.
Comment 1 In the procedure in a), a 250-mL volumetric flask can be used instead of a 300-mL
tall beaker. However, the volumetric flask used should be distinguished as an
extraction flask and should not be used for the other purposes. Additionally, “cover
with a watch glass” in b) is replaced by “place a funnel”. Skip “transfer to a 250-mL
volumetric flask with water” in the procedure in c).
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.3.3.a.
Comment 3 In the procedure in a), it is also allowed to weigh 2.5 g of an analytical sample to the
order of 1 mg, and put it into a 250-mL volumetric flask.
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
185
Testing Methods for Fertilizers (2018)
Note (2) About 3 ml of tetraphenylborate solution per 10 mg of K2O is required to form the
precipitate of potassium tetraphenylborate
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 30 % (mass fraction) -
50 % (mass fraction) and 10 % (mass fraction) - 20 % (mass fraction) are 100.2 % -
100.8 % and 99.3 % - 102.2 % as water soluble potassium (W-K2O) respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.7 % (mass fraction).
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.122- 128, Yokendo, Tokyo (1988)
2) Keiji YAGI, Aiko YANO and Hideo SOETA: Verification of Performance
Characteristics of Testing Method for Nitrate Nitrogen Content in Fertilizer by Phenol
Sulfuric Acid Method, Research Report of Fertilizer Vol. 5, p. 201 - 211 (2012)
(5) Flow sheet for water soluble potassium: The flow sheet for water soluble potassium in
fertilizers is shown below:
Cooling Immediately
Sample solution
186
Testing Methods for Fertilizers (2018)
5 g analytical sample
Weigh to the order of 1 mg into a 500-mL volumetric flask
(compound fertilizer)
← About 400 mL of water
Rotary shaker (30 - 40 revolutions/min)
Shake to mix
For 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Aliquot
(predetermined 100-mL tall beaker
amount)
← Water (add tetraphenylborate solution to reach 50 mL)
← 2 mL of hydrochloric acid (1+9)
← 5 mL of formaldehyde solution
← 5 mL of ethylenediamine tetraacetate-sodium hydroxide solution
← Tetraphenylborate solution (the equivalents of potassium + 4 mL)
Forming precipitate For 30 minutes, sometimes mixing
187
Testing Methods for Fertilizers (2018)
Comment 1 In the procedure in a), a 250-mL volumetric flask can be used instead of a 300-mL
tall beaker. However, the volumetric flask used should be distinguished as an
extraction flask and should not be used for the other purposes. Additionally, “cover
with a watch glass” in b) is replaced by “place a funnel”. Skip “transfer to a 250-mL
188
Testing Methods for Fertilizers (2018)
Comment 3 In the procedure in a), it is also allowed to weigh 2.5 g of an analytical sample to the
order of 1 mg, and put it into a 250-mL volumetric flask.
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
189
Testing Methods for Fertilizers (2018)
Note (2) If the solution temperature is no more than 20 ºC, the reaction does not advance in
some cases. Therefore, it is recommended to heat the solution up to about 30 ºC.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.128- 132, Yokendo, Tokyo (1988)
(5) Flow sheet for water soluble potassium: The flow sheet for water soluble potassium in
fertilizers is shown below:
Cooling Immediately
Sample solution
5 g analytical sample
Weigh to the order of 1 mg into a 500-mL volumetric flask
(compound fertilizer)
← About 400 mL of water
Rotary shaker (30 - 40 revolutions/min)
Shake to mix
For 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
190
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
100-mL volumetric flask
(5 mL to 15 mL)
← Water (the liquid volume to reach about 30 mL)
← About 5mL of formaldehyde solution
← 5 mL of sodium hydroxide solution (120 g/L)
← 25 mL tetraphenylborate solution
(one or two drop (s) per second while shaking to mix)
← Water (up to the marked line)
Leaving at rest 10 minutes
Aliquot
100-mL Erlenmeyer flask
(40 mL)
← A few drops of Titan Yellow solution
Benzaikonium chloride solution (3.3 g/500 mL)
Titration
(until the solution changes to light red)
191
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the potassium standard solution in (2), a potassium standard solution for
the calibration curve preparation can be prepared by using a potassium standard
solution (K 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case,
calculate water-soluble potassium (W-K2O) in the analytical sample by multiplying
the concentration (K) of a potassium standard solution for calibration curve
preparation or a measurement value (K) obtained in (4.2) by a conversion factor
(1.2046).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The axial observation mode does not apply to
potassium since interference is serious.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: An Optical Emission Spectrometer specified in JIS K
0116.
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
b) Add about 50 mL of water, shake to mix and add water up to the marked line.
c) Filter with Type 3 filter paper to make a sample solution.
Note (2) The sampling amount of the analytical sample is 10 g when there is less potassium
content in the fertilizers such as a home garden-use fertilizer.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct measurement according to JIS K 0116 and as shown below. Specific
measurement procedures are according to the operation method of the ICP Optical Emission
Spectrometry used in measurement.
a) Measurement conditions for the ICP Optical Emission Spectrometer: Set up the
measurement conditions for the ICP Optical Emission Spectrometer considering the
following:
Analytical line wavelength: 766.491 nm
b) Calibration curve preparation
1) Spray the potassium standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 766.491 nm.
2) Prepare a curve for the relationship between the potassium concentration and the indicated
value of the potassium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.2 mg - 16 mg as
K2O) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the potassium content from the calibration curve, and calculate water soluble
potassium (W-K2O) in the analytical sample.
193
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Simultaneous Determination of Water-Soluble Principal
Ingredients (W-P2O5, W-K2O, W-MgO, W-MnO and W-B2O3) in Liquid Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES), Research
Report of Fertilizer, Vol.8, p. 1 - 9 (2015)
(5) Flow sheet: The flow sheet for water-soluble potassium of the fluid mixed fertilizers is shown
below:
1 g analytical sample Weigh to the order of 1 mg to a 100-mL volumetric flask
← Water, about 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration
Sample solution
Sample solution
194
Testing Methods for Fertilizers (2018)
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Constant-temperature rotary shaker: A constant-temperature rotary shaker that can rotate
a 250-mL volumetric flask, set up in a thermostat adjustable to 30 ºC ± 1 ºC, upside down at
30 - 40 revolutions/min.
b) Hot plate: A hot plate whose surface temperature can be adjusted up to 250 ºC.
c) Beaker made of polymer: A beaker made of polyethylene, etc. of a quality of material
which prevents silicic acid from eluting in the extract procedure in (4.1).
d) Filter made of polymer: A Gooch crucible made of polymer (compatible filter diameter: 25
mm) or a funnel for reduced pressure filtering made of polymer (compatible filter diameter:
21 mm) A Gooch crucible made of polyethylene, etc. of a quality of material which prevents
silicic acid from eluting in the extract procedure in (4.1).
Comment 2 A funnel for reduced pressure filtering made of polymer (compatible filter diameter:
21 mm) is sold under the name Polyethylene Kiriyama Funnel SB-21.
Note (3) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.4.1.d.
Comment 4 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedure of (4.1) b).
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
C: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
f: Factor of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
V5: Predetermined volume (mL) of the extract in (4.1) c)
V6: Volume (mL) of the extract transferred in (4.2) a)
W2: Mass (g) of the analytical sample
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 25 % (mass fraction) -
40 % (mass fraction) and 10 % (mass fraction) are 98.4 % - 100.5 % and 101.0 % as
soluble silicic acid (S-SiO2) respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.3 % (mass fraction).
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.144- 146, Yokendo, Tokyo (1988)
2) Yasushi MIYASHITA Verification of Performance Characteristics of Testing Method
for Soluble Silicic Acid in Fertilizer by Potassium Fluoride Method, Research Report of
Fertilizer Vol. 7, p. 123 - 130 (2014)
197
Testing Methods for Fertilizers (2018)
5) Flow sheet for soluble silicic acid: The flow sheet for soluble silicic acid in fertilizers is
shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1 Flow sheet for soluble silicic acid in fertilizers (Extraction procedure)
Sample solution
Aliquot
(predetermined 200-mL beaker made of polymer
amount)
← About 10 mL hydrochloric acid
← About 15mL potassium fluoride solution
← About 2 g of potassium chloride
Cooling For 30 minutes or more in a refrigerator
Filtration under
reduced Filter made of polymer, Type 6 filter paper
pressure
198
Testing Methods for Fertilizers (2018)
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Water bath: A water bath that can be adjusted to 65 ºC ± 2 ºC.
b) Hot plate: A hot plate whose surface temperature can be adjusted up to 250 ºC.
c) Volumetric flask and beaker made of polymer: A flask and a beaker that are made of
polyethylene, etc. of a quality of material which prevents silicic acid from eluting in the
extract procedure in (4.1).
d) Gooch crucible made of polymer: A Gooch crucible made of polymer (compatible filter
diameter: 25 mm) or a funnel for reduced pressure filtering made of polymer (compatible
filter diameter: 21 mm) A Gooch crucible made of polyethylene, etc. of a quality of material
which prevents silicic acid from eluting in the extract procedure in (4.1).
Comment 2 A funnel for reduced pressure filtering made of polymer (compatible filter diameter:
21 mm) is sold under the name Polyethylene Kiriyama Funnel SB-21.
0.1 mol/L - 0.2 mol/L sodium hydroxide solution until the color of the solution becomes light
red.
f) Calculate soluble silicic acid (S-SiO2) by the following formula.
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
C: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
f: Factor of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
V5: Predetermined volume (mL) of the extract in (4.1) c)
V6: Volume (mL) of the extract transferred in (4.2) a)
W2: Mass (g) of the analytical sample
Comment 3 Table 1 shows results and analysis results from a collaborative study for testing
method validation.
References
1) Takeshi HASHIMOTO, Akira SHIMIZU and Kaori OKADA: Validation of a Method
of Potassium Fluoride for Determination of Sodium Hydroxide-Soluble Silicic Acid in
Silica Gel Fertilizer, Research Report of Fertilizer, Vol.3, p. 19 - 24 (2010)
2) Akira SHIMIZU, Shin ABE and Jun ITO: Determination of Solubility Silicic Acid in
Silica gel Fertilizer and Silica gel-including Fertilizer by Potassium Fluoride Method: A
201
Testing Methods for Fertilizers (2018)
5) Flow sheet for soluble silicic acid: The flow sheet for soluble silicic acid in fertilizers is
shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1 Flow sheet for soluble silicic acid in silica gel fertilizers (Extraction procedure)
Sample solution
Aliquot
(predetermined 200-mL beaker made of polymer
amount)
← About 10 mL hydrochloric acid
← About 15mL potassium fluoride solution
← About 2 g of potassium chloride
Cooling For 30 minutes or more in a refrigerator
Filtration under
reduced Gooch filter made of polymer, Type 6 filter paper
pressure
Figure 2 Flow sheet for soluble silicic acid in silica gel fertilizers (Measurement procedure)
202
Testing Methods for Fertilizers (2018)
f) Potassium fluoride solution (1): Dissolve 58 g of potassium fluoride specified in JIS K 8815
in 1000 mL of water (2).
g) Methyl red solution (0.1 g/100 mL): Dissolve 0.10 g of methyl red specified in JIS K 8896
in 100 mL of ethanol (95) specified in JIS K 8102.
h) Phenolphthalein solution (1 g/100 mL): Dissolve 1 g of phenolphthalein specified in JIS K
8799 in 100 mL of ethanol (95) specified in JIS K 8102.
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 A funnel for reduced pressure filtering made of polymer (compatible filter diameter:
21 mm) is sold under the name Polyethylene Kiriyama Funnel SB-21.
and wash the container 3 times with a potassium chloride solution, then move the whole
precipitate into the filter, and further wash 6 - 7 times with a small amount of potassium
chloride solution (6).
d) Move the precipitate on the filter together with the filter paper into a 300-mL tall beaker with
water, and further add water to make about 200 mL and heat it up to 70 ºC - 80 ºC on a hot
plate.
e) Add a few drops of phenolphthalein solution (1 g/100 mL) as an indicator and titrate with the
0.1 mol/L - 0.2 mol/L sodium hydroxide solution until the color of the solution becomes light
red.
f) Calculate soluble silicic acid (S-SiO2) by the following formula.
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
C: Estimated concentration (mol/L) of sodium hydroxide solution (0.1 mol/L - 0.2
mol/L)
f: Factor of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
V5: Predetermined volume (mL) of the sample solution in (4.1) c)
V6: Transferred amount (mL) of the sample solution in (4.2) a)
W2: Mass (g) of the analytical sample
Note (3) The transferred volume of the sample solution (1) and the sample solution (2) should
be equivalent.
(4) To be no more than 10 ºC
(5) Filter paper pulp can be stuffed to restrain precipitate from outflowing.
(6) Until the filtrate becomes neutral.
Comment 3 Table 1 shows results and analysis results from a collaborative study for testing
method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.6 % (mass fraction).
205
Testing Methods for Fertilizers (2018)
Table 1 Results and analysis results from a collaborative study for the test method validation
of soluble silicic acid in fertilizers including a silica gel fertilizer
2) 4) 5) 6) 7)
Number of Mean sr RSD r sR RSD R
Sample name 1) 3) 3) 3)
laboratories (%) (%) (%) (%) (%)
Mixed phosphate
8 24.99 0.16 0.6 0.33 1.3
fertilizer 1
Mixed phosphate
8 34.50 0.26 0.7 0.48 1.4
fertilizer 2
Compound
8 30.30 0.13 0.4 0.60 2.0
fertilizers 1
Compound
8 33.34 0.13 0.4 0.47 1.4
fertilizers 2
Compound
8 15.76 0.11 0.7 0.21 1.3
fertilizers 3
1) Number of laboratories used in analysis
2) Total mean (n = number of laboratories × number of replication(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
6) Reproducibility standard deviation
7) Reproducibility relative standard deviation
References
1) Akira SHIMIZU, Jun ITO and Shin ABE: Method Validation of Potassium Fluoride
Method for Determination of Acid-Soluble and Base-Soluble Silicic Acid in Fertilizer
containing Silica gel, Research Report of Fertilizer, Vol.4, p. 1 - 8 (2011)
2) Akira SHIMIZU: Method Validation of Potassium Fluoride Method for Determination
of Acid-Soluble and Base-Soluble Silicic Acid in Fertilizer containing Silica Gel,
Research Report of Fertilizer, Vol.6, p. 1 - 8 (2013)
3) Shinji KAWAGUCHI and Akira SHIMIZU: Determination of Soluble Silicic Acid in
Fertilizers Containing Silica Gel Fertilizer by Potassium Fluoride Method: A
Collaborative Study, Research Report of Fertilizer Vol. 7, p. 36 - 42 (2014)
206
Testing Methods for Fertilizers (2018)
(5) Flow sheet for soluble silicic acid: The flow sheet for soluble silicic acid in fertilizers
including silica gel fertilizers is shown below:
1g Weigh to the order of 1 mg
analytical sample into a 300-mL tall beaker
← About 150 mL hydrochloric acid (1+23) [30 ºC]
1 hour in water bath at 30 ºC ± 2 ºC
Warming
(Stir at every 10 minutes)
Cooling Immediately
207
Testing Methods for Fertilizers (2018)
208
Testing Methods for Fertilizers (2018)
(2) Reagents
a) Hydrochloric acid: A JIS Guaranteed Reagent specified in JIS K 8180 or a reagent of
equivalent quality.
b) Perchloric acid: A JIS Guaranteed Reagent specified in JIS K 8223 or a reagent of
equivalent quality.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Constant-temperature rotary shaker: A constant-temperature rotary shaker that can rotate
a 250-mL volumetric flask, set up in a thermostat adjustable to 30 ºC ± 1 ºC, upside down at
30 - 40 revolutions/min.
b) Hot plate: A hot plate whose surface temperature can be adjusted up to 250 ºC.
c) Electric furnace: An electric furnace that can be adjusted to 1000 ºC - 1100 ºC.
d) Crucible: After heating a chemical analysis porcelain crucible specified in JIS R 1301 in an
electric funnel at 1000 ºC - 1100 ºC, let it stand to cool in a desiccator and measure the mass
to the order of 1 mg.
Comment 1 The procedure in (4.1) is the same as the procedure in (4.1) of 4.4.1.a.
k) After ignition, move the crucible to a desiccator and let it stand to cool.
l) After standing to cool, remove the crucible from the desiccator and measure the mass to the
order of 1 mg.
m) Calculate soluble silicic acid (S-SiO2) by the following formula.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.143- 144, Yokendo, Tokyo (1988)
5) Flow sheet for soluble silicic acid: The flow sheet for soluble silicic acid in fertilizers is
shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1 Flow sheet for soluble silicic acid in fertilizers (Extraction procedure)
210
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 100-mL tall beaker
amount)
←About 10 mL of perchloric acid
When white smoke of perchloric acid starts evolving,
Heating
cover with a watch glass, heat for 15 - 20 minutes
Standing to cool
Figure 2 Flow sheet for soluble silicic acid in fertilizers (Measurement procedure)
211
Testing Methods for Fertilizers (2018)
f) Methyl red solution (0.1 g/100 mL): Dissolve 0.10 g of methyl red specified in JIS K 8896
in 100 mL of ethanol (95) specified in JIS K 8102.
g) Phenolphthalein solution (1 g/100 mL): Dissolve 1 g of phenolphthalein specified in JIS K
8799 in 100 mL of ethanol (95) specified in JIS K 8102.
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 A funnel for reduced pressure filtering made of polymer (compatible filter diameter:
21 mm) is sold under the name Polyethylene Kiriyama Funnel SB-21.
Comment 3 In the procedure in a), it is also allowed to weigh 2.5 g of an analytical sample to the
order of 1 mg, and put it into a 250-mL volumetric flask.
Comment 4 The procedure in (4.1) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
plate.
e) Add a few drops of phenolphthalein solution (1 g/100 mL) to the sample solution as an
indicator and titrate with the 0.1 mol/L - 0.2 mol/L sodium hydroxide solution until the color
of the solution becomes light red.
f) Calculate water soluble silicic acid (W-SiO2) by the following formula.
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
C: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
f: Factor of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
V5: Predetermined volume (mL) of the extract in (4.1) c)
V6: Volume (mL) of the extract transferred in (4.2) a)
W2: Mass (g) of the analytical sample
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 30 % (mass fraction) and
12 % (mass fraction) - 20% (mass fraction) are 100.7 % and 99.5 % - 100.5 as
water-soluble silicic acid (W-SiO2) respectively.
The results of the repeatability tests on different days using a liquid potassium
silicate fertilizer to evaluate precision were analyzed by the one-way analysis of
variance. Table 1 shows the calculation results of intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.2 % (mass fraction).
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.144- 146, Yokendo, Tokyo (1988)
2) Shinji KAWAGUCHI: Verification of Performance Characteristics of Testing Method
214
Testing Methods for Fertilizers (2018)
5) Flow sheet for water-soluble silicic acid: The flow sheet for water-soluble silicic acid in
fertilizers is shown below:
Sample solution
Figure 1 Flow sheet for water-soluble silicic acid in fertilizers (Extraction procedure)
Sample solution
Aliquot
(predetermined 200-mL beaker made of polymer
amount)
← About 10 mL hydrochloric acid
← About 15mL potassium fluoride solution
← About 2 g of potassium chloride
Cooling For 30 minutes or more in a refrigerator
Figure 1 Flow sheet for water-soluble silicic acid in fertilizers (Measurement procedure)
215
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the calcium standard solution in (2), a calcium standard solution for the
calibration curve preparation can be prepared by using a calcium standard solution
(Ca 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case, calculate
total lime (T-CaO) in the analytical sample by multiplying the concentration (Ca) of
the calcium standard solution for calibration curve preparation or a measurement
value (Ca) obtained in (4.2) by a conversion factor (1.3992).
Note (5) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 550 °C in 1 to 2 hours.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.1.a.
Note (6) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 550 °C in 1 to 2 hours.
(7) The watch glass can be removed.
(8) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
217
Testing Methods for Fertilizers (2018)
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 3 The procedures in (4.1.2) are the same as in (4.1.3) in 4.2.1.a. and (4.1) a) - h) in
5.3.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 422.7 nm
b) Calibration curve preparation
1) Spray the calcium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 422.7 nm.
2) Prepare a curve for the relationship between the calcium concentration and the indicated value
of the calcium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.5 mg - 5 mg as
CaO) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the calcium content from the calibration curve, and calculate total lime (T-CaO) in the
analytical sample.
Comment 4 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 15 % (mass fraction) and
1 % (mass fraction) are 101.8 % and 97.9% as total lime (T-CaO) respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.05 % (mass fraction).
218
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 4) 5) 6) 7) 8) 9)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-C-12 11 5.82 0.07 1.2 0.11 2.0 0.29 5.0
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting flame atomic absorbance spectrometry
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.156- 158, Yokendo, Tokyo (1988)
2) Kimie KATO, Masayuki YOSHIMOTO and Yuji SHIRAI: 3) Kimie KATO,
Masayuki YOSHIMOTO and Yuji SHIRAI: Systematization of Determination Methods
of Major Components in Sludge Fertilizer, Compost and Organic Fertilizer, Research
Report of Fertilizer, Vol.3 p. 107 - 116 (2010)
3) Souichi IGARASHI and Yasuharu KIMURA: Verification of Performance
Characteristics of Testing Methods for Calcium Content in Fertilizer by Atomic
Absorption Spectrometry: Research Report of Fertilizer Vol. 6, p. 183 - 192 (2013)
219
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total lime: The flow sheet for total lime in fertilizers is shown below:
5 g analytical sample Weigh to the order of 1 mg into a 200-mL- 300-mL tall beaker.
220
Testing Methods for Fertilizers (2018)
5g
Weigh to the order of 1 mg into a 200-mL - 300-mL tall beaker.
analytical sample
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
←About 10 mL interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer or frame photometer
Figure 2 Flow sheet for total lime in fertilizers (Measurement)
221
Testing Methods for Fertilizers (2018)
Note (1) 29 g of lanthanum oxide (atomic absorption analysis grade or equivalents) can also be
used.
(2) This is an example of preparation; prepare an amount as appropriate.
(3) Add an interference suppressor solution that is 1/10 volume of the volume to be
prepared.
(4) For storage, use a sealable container made of materials which are not likely to dissolve
calcium, such as borosilicate glass-1 specified in JIS R 3503 or Teflon.
Comment 1 Instead of the calcium standard solution in (2), a calcium standard solution for the
calibration curve preparation can be prepared by using a calcium standard solution
(Ca 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case, calculate
soluble lime (S-CaO) in the analytical sample by multiplying the concentration (Ca)
of the calcium standard solution for calibration curve preparation or a measurement
value (Ca) obtained in (4.2) by a conversion factor (1.3992).
Note (5) Be aware that an analytical sample should not solidify in the bottom of a beaker.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 422.7 nm
b) Calibration curve preparation
1) Spray the calcium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 422.7 nm.
2) Prepare a curve for the relationship between the calcium concentration and the indicated value
of the calcium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.5 mg - 5 mg as
CaO) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the calcium content from the calibration curve, and calculate soluble lime (S-CaO) in
the analytical sample.
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
223
Testing Methods for Fertilizers (2018)
result, the average rate of recovery at the content level of 20 % (mass fraction) and
1 % (mass fraction) are 100.9 % and 101.1% as soluble lime (S-CaO) respectively.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.167- 169, Yokendo, Tokyo (1988)
2) Souichi IGARASHI and Yasuharu KIMURA: Verification of Performance
Characteristics of Testing Methods for Calcium Content in Fertilizer by Atomic
Absorption Spectrometry: Research Report of Fertilizer Vol. 6, p. 183 - 192 (2013)
(5) Flow sheet for soluble lime: The flow sheet for soluble lime in fertilizers is shown below:
Cooling Immediately
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
←About 10 mL interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer
224
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the calcium standard solution in (2), a calcium standard solution for the
calibration curve preparation can be prepared by using a calcium standard solution
(Ca 10 mg/mL) traceable to National Metrology.
Comment 2 Instead of the calcium standard solution in (2), a calcium standard solution (CaO 5
µg/mL - 50 µg/mL) for the calibration curve preparation prepared in (2) in 4.5.1.a
can be used. In this case, calculate water-soluble calcium (W-Ca) in the analytical
sample by multiplying the concentration (CaO) of a calcium standard solution for
calibration curve preparation or a measurement value (CaO) obtained in (4.2) by a
conversion factor (0.7147).
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 422.7 nm
b) Calibration curve preparation
1) Spray the calcium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 422.7 nm.
2) Prepare a curve for the relationship between the calcium concentration and the indicated value
of the calcium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.5 mg - 5 mg as
Ca) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the calcium content from the calibration curve, and calculate water-soluble calcium
(W-Ca) in the analytical sample by the following formula.
Comment 3 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 1 % (mass fraction) - 5 %
(mass fraction) is 98.1 % - 101.1% as water-soluble calcium (W-Ca) respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.07 % (mass fraction) for solid fertilizers, and 0.04 % (mass fraction) for fluid
fertilizers.
References
1) Souichi IGARASHI and Yasuharu KIMURA: Verification of Performance
Characteristics of Testing Methods for Calcium Content in Fertilizer by Atomic
Absorption Spectrometry: Research Report of Fertilizer Vol. 6, p. 183 - 192 (2013)
226
Testing Methods for Fertilizers (2018)
(5) Flow sheet for water soluble calcium: The flow sheet for water soluble calcium in fertilizers
is shown below:
1.00 g
500-mL volumetric flask
analytical sample
← About 400 mL of water
Shaking to mix Rotary shaker (30 - 40 revolutions/min), for 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Aliquot
100-mL volumetric flask
(predetermined amount)
← About 10 mL interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (422.7 nm)
227
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the calcium standard solution in (2), a calcium standard solution for the
calibration curve preparation can be prepared by using a calcium standard solution
(Ca 10 mg/mL) traceable to National Metrology.
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in e) and f) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a calcium standard
solution for the calibration curve in the concentration range which is suitable to a
device used.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: A spectrophotometer specified in JIS K 0115
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
flask.
b) Add about 50 mL of water, shake to mix and add water up to the marked line.
c) Filter with Type 3 filter paper to make a sample solution.
Note (2) The sampling amount of the analytical sample is 10 g when the content in the
analytical sample is less than 0.01 % (mass fraction) as water-soluble calcium.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the measurement
conditions for the ICP Atomic Emission Spectrometer considering the following:
Analytical line wavelength: 393.366 nm
b) Calibration curve preparation
1) Spray the calcium standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 393.366 nm.
2) Prepare a curve for the relationship between the calcium concentration and the indicated value
of the calcium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 2 mg as
Ca) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the calcium content from the calibration curve, and calculate water-soluble calcium
(W-Ca) in the analytical sample by the following formula.
229
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Simultaneous Determination Method for Effect-Development
Promoting Agent (Ca, Fe, Co, Cu, Zn and Mo) in Liquid Compound Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometer (ICP-OES), Research
Report of Fertilizer, Vol.9, p. 1 - 9 (2016)
(5) Flow sheet: The flow sheet for water-soluble calcium of the fluid mixed fertilizers is shown
below:
Sample solution
Sample solution
230
Testing Methods for Fertilizers (2018)
4.5.4 Alkalinity
4.5.4.a Ethylenediamine tetraacetate method
(1) Summary
This testing method is applicable to fertilizers that guarantee alkalinity. This testing method is
classified as Type E and its symbol is 4.5.4.a-2017 or AL.a-1.
Add hydrochloric acid (1+23) to an analytical sample, boil to extract, and mask with
2,2’,2”-nitrilotriethanol and a potassium cyanide solution, add a 0.01 mol/L ethylenediamine
tetraacetate standard solution, and conduct a chelatometric titration with a 0.01 mol/L magnesium
standard solution to obtain alkalinity (AL). Or after masking, conduct a chelatometric titration with
an ethylenediamine tetraacetate standard solution to obtain alkalinity (AL) in an analytical sample.
h) 0.01 mol/L magnesium standard solution: Put 0.24 g of magnesium specified in JIS K
8875 into a 1000-mL beaker, add about 10 mL of hydrochloric acid to dissolve, add a proper
amount of water, and while adding a methyl red solution (0.1 g/100 mL) as an indicator,
neutralize with an ammonia solution (1+3) until the color of the solution becomes yellow, and
then add water to make 1000 mL.
Standardization: Transfer 25 mL of 0.01 mol/L magnesium standard solution to a 200-mL -
300-mL Erlenmeyer flask, add about 15 mL of water and about 5 mL of ammonium chloride
buffer solution, and while adding an Eriochrome Black T solution as an indicator, titrate
with 0.01 mol/L ethylenediamine tetraacetate standard solution until the color of the solution
becomes blue. Calculate the factor of a 0.01 mol/L magnesium standard solution by the
following formula.
Note (1) The reagent corresponds to triethanolamine in the Official Methods of Analysis of
Fertilizers (1992).
(2) This is an example of preparation; prepare an amount as appropriate.
Comment 1 Instead of a 0.01 mol/L ethylenediamine tetraacetate standard solution in (2) g), a 0.1
mol/L ethylenediaminetetraacetic acid dihydrogen disodium solution conforming to
ISO/IEC 17025 can be used.
Note (3) Be aware that an analytical sample should not solidify in the bottom of a beaker.
(4) Conduct the procedure in Comment 5 when there is much manganese content in the
fertilizers.
Comment 2 In the case of a by-product magnesia fertilizer, if the pH of the sample solution of d)
is neutral or basic, prepare a sample solution anew by replacing “2 g of an analytical
sample” in the procedure in a) with “1 g - 1.5 g of an analytical sample”.
Comment 3 In the procedure in a), a 500-mL volumetric flask can be used instead of a 500-mL
tall beaker. However, the volumetric flask used should be distinguished as an
extraction flask and should not be used for the other purposes. In addition, “cover
with a watch glass” in b) is replaced by “place a funnel”, and “transfer to a 250-mL -
500 mL volumetric flask with water” in the procedure in c) is skipped.
Comment 4 The procedure in (4.1) is the same as the procedure in (4.1) of 4.5.2.a.
Comment 5 Add a predetermined volume of filtrate in (4.1) e), add a drop of methyl red solution
as an indicator and drop ammonia water (28 % (mass fraction)) specified in JIS K
8085 until the color of the solution changes light yellowish red from purplish red.
Add 20 mL of ammonium peroxodisulfate to boil (5). Immediately transfer it to a
100-mL - 200-mL volumetric flask with water. After immediate cooling is complete,
add water up to the marked line. Filter with Type 3 filter paper to make a sample
solution.
(4.2) Measurement: Conduct measurement as shown below. Two examples of titration are shown
as follows.
(4.2.1) Measurement (A): Titration with a magnesium standard solution (0.01 mol/L)
a) Transfer a predetermined volume (the equivalents of 5 mg - 20 mg as CaO + MgO) of sample
solution to a 200-mL - 300-mL Erlenmeyer flask.
b) Add a proper amount of water, add a drop of methyl red solution as an indicator and drop
sodium hydroxide (5 g/100 mL) to neutralize until the color of the solution becomes yellow.
c) Add 0.1 g of ascorbic acid, 1 ml - 10 mL of 2,2’,2”-nitrilotriethanol−water (1+3) and 1 mL -
10 mL of potassium cyanide solution (6).
d) Add a predetermined volume of 0.01 mol/L ethylenediamine tetraacetate standard solution (7).
e) Add 20 mL of ammonium chloride solution or 2-aminoethanol solution.
f) Add several drops of Eriochrome Black T solution, and titrate with a 0.01 mol/L magnesium
standard solution until the color of the solution becomes red.
g) Calculate the alkalinity (AL) content in an analytical sample by the following formula.
(4.2.2) Measurement (B): Titration with an ethylenediamine tetraacetate standard solution (0.01
mol/L)
a) Transfer a predetermined volume (the equivalents of 5 mg - 20 mg as CaO+MgO) of sample
solution to a 200-mL - 300-mL Erlenmeyer flask.
b) Add a proper amount of water and 5 mL of citric acid solution (8), add a drop of methyl red
solution as an indicator and drop sodium hydroxide (5 g/100 mL) to neutralize until the color
of the solution becomes yellow.
c) Add 0.1 g of ascorbic acid, 1 ml - 10 mL of 2,2’,2”-nitrilotriethanol−water (1+3) and 1 mL -
10 mL of potassium cyanide solution (6).
d) Add 20 mL of ammonium chloride solution or 2-aminoethanol solution.
e) Add several drops of an Eriochrome Black T solution, and immediately titrate with a 0.01
mol/L ethylenediamine tetraacetate standard solution until the color of the solution becomes
blue-green.
f) Calculate the alkalinity (AL) content in an analytical sample by the following formula.
Note (8) When the sample solution does not contain phosphate, silicate, etc., it is not necessary
234
Testing Methods for Fertilizers (2018)
Comment 6 Care should be fully taken in the case of using potassium cyanide and its solution in
accordance with the Safety Data Sheet (SDS). In addition, observe laws and
ordinances concerned such as the Poisonous and Deleterious Substance Control Law.
Criteria of the abolition in the Poisonous and Deleterious Substance Control Law (for
reference): Add a sodium hydroxide solution to make it alkalinity more than pH 11,
and further add an oxidizer (sodium hypochlorite, bleaching powder) solution to
conduct oxidative degradation processes. After dissolving CN ingredient, neutralize
with sulfuric acid, and discard it after diluting it with a large amount of water. Take
enough time to dissolve the CN ingredient with alkalinity.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.162- 164, Yokendo, Tokyo (1988)
(5) Flow sheet for alkalinity: The flow sheet for alkalinity in fertilizers is shown below:
Cooling Immediately
Sample solution
235
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
200-mL - 300-mL Erlenmeyer Flask
(predetermined amount)
Figure 2-1 Flow sheet for alkalinity in fertilizers (Measurement procedure (4.2.1) (A))
Sample solution
Aliquot
200-mL - 300-mL Erlenmeyer Flask
(predetermined amount)
Figure 2-2 Flow sheet for alkalinity in fertilizers (Measurement procedure (4.2.2) (B))
236
Testing Methods for Fertilizers (2018)
Note (1) S-CaO and S-MgO use raw data without rounding numerical value
237
Testing Methods for Fertilizers (2018)
4.6 Magnesia
4.6.1 Total magnesia
4.6.1.a Flame atomic absorption spectrometry
(1) Summary
The testing method is applicable to compost, sludge fertilizers (except for calcined sludge fertilizer)
and poultry manure ash. This testing method is classified as Type D and its symbol is 4.6.1.a-2018
or T-Mg.a-1.
Pretreat an analytical sample with incineration and hydrochloric acid or nitric acid−hydrochloric
acid (3+1), add an interference suppressor solution, and then spray in an acetylene−air flame, and
measure the atomic absorption with magnesium at a wavelength of 285.2 nm to quantify total
magnesia (T-MgO) in an analytical sample.
In addition, the performance of this testing method is shown in Comment 4.
Comment 1 Instead of the magnesium standard solution in (2), a magnesium standard solution
for the calibration curve preparation can be prepared by using a magnesium
standard solution (Mg 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National
Metrology. In this case, calculate total magnesia (T-MgO) in the analytical sample
by multiplying the concentration (Mg) of the magnesium standard solution for
calibration curve preparation or a measurement value (Mg) obtained in (4.2) by a
238
Testing Methods for Fertilizers (2018)
Note (4) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 550 °C in 1 to 2 hours.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.2) in 4.2.1.a.
Note (5) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for
about 1 hour and further raise to 450 °C in 1 to 2 hours.
(6) The watch glass can be removed.
(7) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 3 The procedures in (4.1.2) are the same as in (4.1.3) in 4.2.1.a. and (4.1) a) - h) in
5.3.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 285.2 nm
b) Calibration curve preparation
1) Spray the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation into a flame, and read the indicated value at
a wavelength of 285.2 nm.
2) Prepare a curve for the relationship between the magnesium concentration and the indicated
value of the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.1 mg - 1 mg as
MgO) (8) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the magnesium content from the calibration curve, and calculate the total magnesia
(T-MgO) in the analytical sample.
Note (8) If there is a possibility that the concentration of total magnesia in the sample solution
exceeds the maximum limit of the calibration curve, transfer a predetermined amount
of the sample solution and dilute by adding water up to the marked line to make 0.1
mg - 1 mg as MgO.
Comment 4 Recovery testing was conducted to evaluate trueness using a preparation sample. As
a result, the average rate of recovery at the content level of 5 % (mass fraction),
1 % (mass fraction) and 0.2 % (mass fraction) are 102.4 %, 101.7 % and 103.0 % as
total magnesia (T-MgO) respectively.
The results of the repeatability tests on different days using swine manure compost,
composted sludge fertilizer and poultry manure ash (1 sample for each) to evaluate
precision were analyzed by the one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this test method is about 0.2 %
(mass fraction).
240
Testing Methods for Fertilizers (2018)
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Swine manure compost 5 3.14 0.03 1.1 0.05 1.5
Composted sludge fertilizer 5 0.84 0.01 1.2 0.01 1.3
Poultry manure ash 5 3.97 0.03 0.7 0.04 1.1
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
(5) Flow sheet for total magnesia: The flow sheet for total magnesia in fertilizers is shown
below:
5 g analytical sample Weigh to the order of 1 mg into a 200-mL- 300-mL tall beaker.
Sample solution
241
Testing Methods for Fertilizers (2018)
5g
Weigh to the order of 1 mg into a 200-mL - 300-mL tall beaker.
analytical sample
Charring Heat gently
Incineration Ignite at 450 ºC ± 5 ºC for 8 - 16 hours
Sample solution
Sample solution
Aliquot
100-mL volumetric flask
(predetermined amount)
←About 10 mL interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer
242
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the magnesium standard solution in (2), a magnesium standard solution for
the calibration curve preparation can be prepared by using a magnesium standard
solution (Mg 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate soluble magnesia (S-MgO) in the analytical sample by
multiplying the concentration (Mg) of a magnesium standard solution for calibration
curve preparation or a measurement value (Mg) obtained in (4.2) by a conversion
factor (1.6583).
JIS K 0121.
1) Light source: A magnesium hollow cathode lamp
2) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
b) Hot plate : A hot plate whose surface temperature can be adjusted up to 250 ºC.
Note (4) Be aware that an analytical sample should not solidify in the bottom of a beaker.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 285.2 nm
b) Calibration curve preparation
1) Spray the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation into a flame, and read the indicated value at
a wavelength of 285.2 nm.
2) Prepare a curve for the relationship between the magnesium concentration and the indicated
value of the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.1 mg - 1 mg as
MgO) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the magnesium content from the calibration curve, and calculate the soluble magnesia
(S-MgO) in the analytical sample.
244
Testing Methods for Fertilizers (2018)
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 15 % (mass fraction) and
1 % (mass fraction) are 101.7 % and 99.5% as soluble magnesia (S-MgO)
respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.2 % (mass fraction) for solid fertilizers, and 0.05 % (mass fraction) for fluid
fertilizers.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.167- 169, Yokendo, Tokyo (1988)
2) Souichi IGARASHI and Yasuharu KIMURA: Verification of Performance
Characteristics of Testing Methods for Magnesia Content in Fertilizer by Atomic
Absorption Spectrometry: Research Report of Fertilizer Vol. 6, p. 193 - 202 (2013)
(5) Flow sheet for soluble magnesia: The flow sheet for soluble magnesia in fertilizers is shown
below:
Cooling Immediately
Sample solution
Sample solution
Aliquot
100-mL volumetric flask
(predetermined amount)
← About 10 mL of interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (285.2 nm)
Figure 2 Flow sheet for soluble magnesia in fertilizers (Measurement procedure)
245
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the magnesium standard solution in (2), a magnesium standard solution for
the calibration curve preparation can be prepared by using a magnesium standard
solution (Mg 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate citrate-soluble magnesia (C-MgO) in the analytical sample by
multiplying the concentration (Mg) of a magnesium standard solution for calibration
curve preparation or a measurement value (Mg) obtained in (4.2) by a conversion
factor (1.6583).
described below.
aa) Constant-temperature rotary shaker: A constant-temperature rotary shaker that can rotate
a 250-mL volumetric flask, set up in a thermostat adjustable to 30 ºC ± 1 ºC, upside down at
30 - 40 revolutions/min.
ab) Reciprocating water bath shaker: A reciprocating water bath shaker: that can be adjusted
to 30 °C ± 1 °C and shake a 250-mL volumetric flask, set up vertically on the water surface
using a shaking rack, reciprocate horizontally at 160 times/ minute with amplitude of 25 mm -
40 mm.
b) Flame atomic absorption spectrometer: An atomic absorption spectrometer specified in JIS
K 0121.
1) Light source: A magnesium hollow cathode lamp
2) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
Note (4) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (5) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 3 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 4 For a by-product magnesia fertilizer, if the pH of the sample solution in (4.1.1) d) and
(4.1.2) d) is neutral or basic, prepare a sample solution anew by replacing “1 g of an
analytical sample” in the procedures in (4.1.1) a) and (4.1.2) a) with “0.5 g of an
analytical sample”.
Comment 5 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
Comment 6 In some slag silicate fertilizers, the variation of measurement value of citrate
magnesia (C-MgO) may be observed according to the time variation of heating state
after adding a citric acid solution. Therefore, in the case of slag silicate fertilizers, it
247
Testing Methods for Fertilizers (2018)
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 285.2 nm
b) Calibration curve preparation
1) Spray the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation into a flame, and read the indicated value at
a wavelength of 285.2 nm.
2) Prepare a curve for the relationship between the magnesium concentration and the indicated
value of the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.1 mg - 1 mg as
MgO) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the magnesium content from the calibration curve, and calculate the citrate soluble
magnesia (C-MgO) in the analytical sample.
Comment 8 Recovery testing was conducted to evaluate trueness using a preparation sample. As
a result, the average rate of recovery at the content level of 1 % (mass fraction) - 5 %
(mass fraction) are 98.9 % - 100.3 % as citrate-soluble magnesia (C-MgO)
respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this test method is about 0.06 %
(mass fraction) for solid fertilizers.
248
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 4) 5) 6) 7) 8) 9)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-A-10 11 3.28 0.07 2.0 0.08 2.5 0.11 3.3
FAMIC-A-13 9 3.18 0.03 1.0 0.04 1.4 0.12 3.8
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting flame atomic absorbance spectrometry
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.167- 169, Yokendo, Tokyo (1988)
2) Souichi IGARASHI and Yasuharu KIMURA: Verification of Performance
Characteristics of Testing Methods for Magnesia Content in Fertilizer by Atomic
Absorption Spectrometry: Research Report of Fertilizer Vol. 6, p. 193 - 202 (2013)
3) Souichi IGARASHI and Yasuharu KIMURA: Effect of Time during Extraction
Operation on Measurements of Citric Acid-Soluble Magnesium in Silicate Slag
Fertilizer Vol. 7, p. 145 - 156 (2014)
(5) Flow sheet for citrate-soluble magnesia: The flow sheet for citrate-soluble magnesia in
fertilizers is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
249
Testing Methods for Fertilizers (2018)
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-2 Flow sheet for citrate-soluble magnesia in fertilizers
(Extraction procedure (4.1.2))
Sample solution
Aliquot
100-mL volumetric flask
(predetermined amount)
← About 10 mL of interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (285.2 nm)
250
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the magnesium standard solution in (2), a magnesium standard solution for
the calibration curve preparation can be prepared by using a magnesium standard
solution (Mg 1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate citrate-soluble magnesia (C-MgO) in the analytical sample by
multiplying the concentration (Mg) of a magnesium standard solution for calibration
curve preparation or a measurement value (Mg) obtained in (4.2) by a conversion
factor (1.6583).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in f) and g) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a magnesium standard
solution for the calibration curve in the concentration range which is suitable to a
251
Testing Methods for Fertilizers (2018)
device used.
Note (2) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (3) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 3 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 4 For a by-product magnesia fertilizer, if the pH of the sample solution in (4.1.1) d)
and (4.1.2) d) is neutral or basic, prepare a sample solution anew by replacing “1 g of
an analytical sample” in the procedures in (4.1.1) a) and (4.1.2) a) with “0.5 g of an
analytical sample”.
Comment 5 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
Comment 6 In some slag silicate fertilizers, the variation of measurement value of citrate
magnesia (C-MgO) may be observed according to the time variation of heating state
after adding a citric acid solution. Therefore, in the case of slag silicate fertilizers, it
252
Testing Methods for Fertilizers (2018)
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the measurement
conditions for the ICP Atomic Emission Spectrometer considering the following:
Analytical line wavelength: 279.553 nm
b) Calibration curve preparation
1) Spray the magnesium standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 279.553 nm.
2) Prepare a curve for the relationship between the magnesium concentration and the indicated
value of the magnesium standard solutions for the calibration curve preparation and the
blank test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.02 mg - 1.6 mg
as MgO) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the magnesium content from the calibration curve, and calculate the citrate soluble
magnesia (C-MgO) in the analytical sample.
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Blended fertilizer 7 8.41 0.09 1.1 0.10 1.1
Compound
7 1.66 0.03 1.6 0.03 1.8
fertilizer
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days (T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
254
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate-soluble magnesia: The flow sheet for citrate-soluble magnesia in
fertilizers is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 Flow sheet for citrate-soluble magnesia in fertilizers (Extraction procedure (4.1.1))
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-2 Flow sheet for citrate-soluble magnesia in fertilizers (Extraction procedure (4.1.2))
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← About 10 mL of interference suppressor solution
← Water (up to the marked line)
Measurement ICP-OES (279.553 nm)
255
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the magnesium standard solution in (2), a magnesium standard solution for
the calibration curve preparation can be prepared by using a magnesium standard
solution (Mg 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate water-soluble magnesia (W-MgO) in the analytical sample by
multiplying the concentration (Mg) of a magnesium standard solution for calibration
curve preparation or a measurement value (Mg) obtained in (4.2) by a conversion
factor (1.6583).
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Flame atomic absorption spectrometer: An atomic absorption spectrometer specified in
JIS K 0121.
1) Light source: A magnesium hollow cathode lamp
256
Testing Methods for Fertilizers (2018)
Note (4) The volumetric flask used for extraction should be distinguished as an extraction flask
and should not be used for the other purposes.
Note (5) Be aware that an analytical sample should not solidify in the bottom of a beaker.
Comment 2 In the procedure in a), a 500-mL volumetric flask can be used instead of a 500-mL
tall beaker. However, the volumetric flask used should be distinguished as an
extraction flask and should not be used for the other purposes. Additionally, “cover
with a watch glass” in b) is replaced by “place a funnel”. Skip “transfer to a 500-mL
volumetric flask with water” in the procedure in c).
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 285.2 nm
b) Calibration curve preparation
1) Spray the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation into a flame, and read the indicated value at
a wavelength of 285.2 nm.
2) Prepare a curve for the relationship between the magnesium concentration and the indicated
value of the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.1 mg - 1 mg as
MgO) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the magnesium content from the calibration curve, and calculate the water-soluble
magnesia (W-MgO) in the analytical sample.
Comment 3 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 1 % (mass fraction) - 5 %
(mass fraction) are 100.4 % - 100.9 % as water-soluble magnesia (W-MgO)
257
Testing Methods for Fertilizers (2018)
respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.07 % (mass fraction).
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.167- 169, Yokendo, Tokyo (1988)
2) Souichi IGARASHI and Yasuharu KIMURA: Verification of Performance
Characteristics of Testing Methods for Magnesia Content in Fertilizer by Atomic
Absorption Spectrometry: Research Report of Fertilizer Vol. 6, p. 193 - 202 (2013)
(5) Flow sheet for water-soluble magnesia: The flow sheet for water-soluble magnesia in
fertilizers is shown below:
Cooling Immediately
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← About 10 mL of interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (285.2 nm)
258
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the magnesium standard solution in (2), a magnesium standard solution for
the calibration curve preparation can be prepared by using a magnesium standard
solution (Mg 1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate water-soluble magnesia (W-MgO) in the analytical sample by
multiplying the concentration (Mg) of a magnesium standard solution for calibration
curve preparation or a measurement value (Mg) obtained in (4.2) by a conversion
factor (1.6583).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in e) and f) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a magnesium standard
solution for the calibration curve in the concentration range which is suitable to a
device used.
259
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: An Optical Emission Spectrometer specified in JIS K
0116.
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
Note (2) The sampling amount of the analytical sample is 10 g when there is less magnesia
content in the fertilizers such as a home garden-use fertilizer.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Optical Emission Spectrometer: Set up the measurement
conditions for the ICP Optical Emission Spectrometer considering the following:
Analytical line wavelength: 279.553 nm
b) Calibration curve preparation
1) Spray the magnesium standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 279.553 nm.
2) Prepare a curve for the relationship between the magnesium concentration and the indicated
value of the magnesium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.02 mg - 1.6 mg
as MgO) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the magnesium content from the calibration curve, and calculate the water-soluble
magnesia (W-MgO) in the analytical sample.
The results of the repeatability tests on different days using a fluid mixed fertilizer
and a home garden-use mixed fertilizer to evaluate precision were analyzed by the
one-way analysis of variance. Table 1 shows the calculation results of intermediate
precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.002 % (mass fraction).
References
1) Keisuke AOYAMA: Simultaneous Determination of Water-Soluble Principal
Ingredients (W-P2O5, W-K2O, W-MgO, W-MnO and W-B2O3) in Liquid Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES), Research
Report of Fertilizer, Vol.8, p. 1 - 9 (2015)
(5) Flow sheet: The flow sheet for water-soluble magnesia in fluid fertilizers is shown below:
Sample solution
261
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
100-mL volmetric flask
(predetermined volume)
← 25 mL of hydrochloric acid (1+5)
← Water (up to the marked line)
262
Testing Methods for Fertilizers (2018)
4.7 Manganese
4.7.1 Soluble manganese
4.7.1.a Flame atomic absorption spectrometry
(1) Summary
This test method is applicable to fertilizers containing manganese carbonate fertilizers. This testing
method is classified as Type D and its symbol is 4.7.1.a-2017 or S-Mn.a-1.
Add hydrochloric acid (1+23) to an analytical sample, boil to extract and add an interference
suppressor solution, and then spray into an acetylene−air flame, and measure the atomic absorption
with manganese at a wavelength of 279.5 nm to obtain the hydrochloric acid (1+23) soluble
manganese (soluble manganese (S-MnO)) in an analytical sample. In addition, the performance of
this testing method is shown in Comment 4.
Comment 1 Instead of the manganese standard solution in (2), a manganese standard solution for
the calibration curve preparation can be prepared by using a manganese standard
solution (Mn 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate soluble manganese (S-MnO) in the analytical sample by
multiplying the concentration (Mn) of a manganese standard solution for calibration
curve preparation or a measurement value (Mn) obtained in (4.2) by a conversion
factor (1.2912).
JIS K 0121.
1) Light source: A manganese hollow cathode lamp
2) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
b) Hot plate : A hot plate whose surface temperature can be adjusted up to 250 ºC.
Note (4) Be aware that an analytical sample should not solidify in the bottom of a beaker.
Comment 2 In the procedure in a), a 500-mL volumetric flask can be used instead of a 500-mL
tall beaker. However, the volumetric flask used should be distinguished as an
extraction flask and should not be used for the other purposes. In addition, “cover
with a watch glass” in b) is replaced by “place a funnel”, and “transfer to a 250-mL -
500 mL volumetric flask with water” in the procedure in c) is skipped.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.5.2.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 279.5 nm
b) Calibration curve preparation
1) Spray the manganese standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation into a flame, and read the indicated value at
a wavelength of 279.5 nm.
2) Prepare a curve for the relationship between the manganese concentration and the indicated
value of the manganese standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.1 mg - 1 mg as
MnO) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the manganese content from the calibration curve, and calculate the soluble manganese
(S-MnO) in the analytical sample.
Comment 4 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 5 % (mass fraction) and
264
Testing Methods for Fertilizers (2018)
0.1 % (mass fraction) are 100.5 % and 101.3 % as soluble manganese (S-MnO)
respectively.
Additionally, the minimum limit of quantification of this testing method is about
0.006 % (mass fraction).
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.176- 177, Yokendo, Tokyo (1988)
2) Keiji YAGI, Natuki TOYODOME, Tokiya SUZUKI and Hideo SOETA: Verification of
Performance Characteristics of Testing Methods for Manganese Content in Fertilizer by
Atomic Absorption Spectrometry, Research Report of Fertilizer Vol. 6, p. 203 - 212
(2013)
(5) Flow sheet for soluble manganese: The flow sheet for soluble manganese in fertilizers is
shown below:
Cooling Immediately
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← About 10 mL of interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (279.5 nm)
265
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the manganese standard solution in (2), a manganese standard solution for
the calibration curve preparation can be prepared by using a manganese standard
solution (Mn 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate citrate-soluble manganese (C-MnO) in the analytical sample by a
multiplying the concentration (Mn) of manganese standard solution for calibration
curve preparation or a measurement value (Mn) obtained in (4.2) by a conversion
factor (1.2912).
described below.
aa) Constant-temperature rotary shaker: A constant-temperature rotary shaker that can rotate
a 250-mL volumetric flask, set up in a thermostat adjustable to 30 ºC ± 1 ºC, upside down at
30 - 40 revolutions/min.
ab) Reciprocating water bath shaker: A reciprocating water bath shaker: that can be adjusted to
30 °C ± 1 °C and shake a 250-mL volumetric flask, set up vertically on the water surface
using a shaking rack, reciprocate horizontally at 160 times/ minute with amplitude of 25 mm -
40 mm.
b) Flame atomic absorption spectrometer: An atomic absorption spectrometer specified in
JIS K 0121.
1) Light source: A manganese hollow cathode lamp
2) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
Note (4) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (5) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 3 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 4 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
267
Testing Methods for Fertilizers (2018)
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 5 % (mass fraction) and
0.1 % (mass fraction) are 101.9 % and 100.5 % as citrate-soluble manganese
(C-MnO) respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.006 % (mass fraction).
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 4) 5) 6) 7) 8) 9)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-A-10 9 0.403 0.004 1.1 0.01 1.3 0.010 2.4
FAMIC-A-13 10 0.356 0.010 2.7 0.01 3.9 0.018 4.9
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting flame atomic absorbance spectrometry
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.176- 177, Yokendo, Tokyo (1988)
2) Keiji YAGI, Natuki TOYODOME, Tokiya SUZUKI and Hideo SOETA: Verification of
Performance Characteristics of Testing Methods for Manganese Content in Fertilizer by
Atomic Absorption Spectrometry, Research Report of Fertilizer Vol. 6, p. 203 - 212
268
Testing Methods for Fertilizers (2018)
(2013)
(5) Flow sheet for citrate-soluble manganese: The flow sheet for citrate-soluble manganese in
fertilizers is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 Flow sheet for citrate-soluble manganese in fertilizers
(Extraction procedure (4.1.1))
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
269
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← About 10 mL of interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (279.5 nm)
Figure 2 Flow sheet for citrate-soluble manganese in fertilizers
(Measurement procedure)
270
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the manganese standard solution in (2), a manganese standard solution for
the calibration curve preparation can be prepared by using a manganese standard
solution (Mn 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case,
calculate citrate-soluble manganese (C-MnO) in the analytical sample by a
multiplying the concentration (Mn) of manganese standard solution for calibration
curve preparation or a measurement value (Mn) obtained in (4.2) by a conversion
factor (1.2912).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in f) and g) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a manganese standard
solution for the calibration curve in the concentration range which is suitable to a
271
Testing Methods for Fertilizers (2018)
device used.
Note (2) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (3) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 3 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 4 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
272
Testing Methods for Fertilizers (2018)
273
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate-soluble manganese: The flow sheet for citrate-soluble manganese in
fertilizers is shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 Flow sheet for citrate-soluble manganese in fertilizers
(Extraction procedure (4.1.1))
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← About 25 mL of hydrochloric acid (1+5)
← Water (up to the marked line)
Measurement ICP-OES (257.610 nm)
274
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the manganese standard solution in (2), a manganese standard solution for
the calibration curve preparation can be prepared by using a manganese standard
solution (Mn 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National Metrology. In
this case, calculate citrate-soluble manganese (C-MnO) in the analytical sample by a
multiplying the concentration (Mn) of manganese standard solution for calibration
curve preparation or a measurement value (Mn) obtained in (4.2) by a conversion
factor (1.2912).
ab) Vertical reciprocating shaker: A vertical reciprocating shaker that can shake a 250-mL
volumetric flask using an adapter for a flack to reciprocate vertically at 300 times/ minute
(amplitude of 40 mm).
b) Flame atomic absorption spectrometer: An atomic absorption spectrometer specified in
JIS K 0121.
1) Light source: A manganese hollow cathode lamp
2) Gas: Gas for heating by flame
(i) Fuel gas: acetylene
(ii) Auxiliary gas: Air sufficiently free of dust and moisture.
Comment 2 In the procedure of (4.1.1.1) a), it is also allowed to weigh 2.5g of an analytical
sample to the order of 1 mg, and put it in a 250-mL volumetric flask
Comment 3 The procedure in (4.1.1.1) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
Comment 4 The procedure in (4.1.1.2) is the same as the procedure in (4.1.1.2) in 4.2.4.a.
Comment 5 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 279.5 nm
b) Calibration curve preparation
276
Testing Methods for Fertilizers (2018)
1) Spray the manganese standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation into a flame, and read the indicated value at
a wavelength of 279.5 nm.
2) Prepare a curve for the relationship between the manganese concentration and the indicated
value of the manganese standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.1 mg - 1 mg as
MnO) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the manganese content from the calibration curve, and water-soluble manganese
(W-MnO) in the analytical sample.
Comment 6 Recovery testing was conducted to evaluate trueness using a preparation sample. As
a result, the average rate of recovery at the content level of 6 % (mass fraction) and
0.1 % (mass fraction) are 101.2 % and 101.1 % as water-soluble manganese
(W-MnO) respectively.
The comparison of the measurement value of extraction (yi:0.0330 % (mass fraction)
- 6.18 % (mass fraction )) by a vertical reciprocating shaker and the measurement
value of extract with a rotary shaker (xi) was conducted to evaluate trueness of the
extraction of solid fertilizers using fertilizers (12 samples). As a result, a regression
equation was y = − 0.009 + 1.011x, and its correlation coefficient (r) was 1.000. Also,
the results of the repeatability tests on different days using compound fertilizers and
mixed micro element fertilizers to evaluate precision were analyzed by one-way
analysis of variance. Table 1 shows the calculation results of intermediate precision
and repeatability
The comparison of the measurement value of simple extraction (yi:0.0590 % (mass
fraction) - 1.27 % (mass fraction )) and the measurement value of extract (xi) with a
rotary shaker was conducted to evaluate trueness of the extraction of fluid fertilizers
using fluid fertilizers (12 samples). As a result, a regression equation was y = − 0.001
+ 1.006x, and its correlation coefficient (r) was 1.000. Also, the results of the
repeatability tests on different days using fluid compound fertilizers and fluid mixed
micro element fertilizers to evaluate precision were analyzed by one-way analysis of
variance. Table 2 shows the calculation results of intermediate precision and
repeatability
Additionally, the minimum limit of quantification of this testing method is about
0.004 % (mass fraction).
277
Testing Methods for Fertilizers (2018)
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Mixed micro element
7 3.57 0.03 0.7 0.05 1.5
fertilizers
Blended fertilizer 7 0.226 0.002 1.0 0.004 1.7
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Fluid compound
7 1.28 0.01 0.4 0.02 1.3
fertilizers
Liquid micro element
7 0.230 0.001 0.5 0.003 1.5
compound fertilizers
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.176- 177, Yokendo, Tokyo (1988)
2) Keiji YAGI, Natuki TOYODOME, Tokiya SUZUKI and Hideo SOETA: Verification of
Performance Characteristics of Testing Methods for Manganese Content in Fertilizer by
Atomic Absorption Spectrometry, Research Report of Fertilizer Vol. 6, p. 203 - 212
(2013)
3) Shinji KAWAGUCHI: Simple Extraction Method for Water-Soluble Components in
Liquid Compound Fertilizers, Research Report of Fertilizer, Vol.9, p. 10 - 20 (2016)
4) Shinji KAWAGUCHI: Extraction Method for the Water-soluble Principal Ingredients
in the Solid Fertilizer using a Gneral-purpose Equipment , Research Report of Fertilizer,
Vol.10, p. 1 - 8 (2017)
278
Testing Methods for Fertilizers (2018)
(5) Flow sheet for water-soluble manganese: The flow sheet for water-soluble manganese in
fertilizers is shown below:
5 g analytical sample
Weigh to the order of 1 mg into a 500-mL volumetric flask.
(Powdery)
← About 400 mL of water
Shaking to mix Rotary shaker (30 - 40 revolutions/min), 30 minutes
Sample solution
1 g analytical sample
Weigh to the order of 1 mg into a 100-mL volumetric flask.
(Fluid)
← About 50 mL of water
Shaking to mix
Sample solution
279
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← About 10 mL of interference suppressor solution
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (279.5 nm)
280
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the manganese standard solution in (2), a manganese standard solution for
the calibration curve preparation can be prepared by using a manganese standard
solution (Mn 1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case,
calculate water-soluble manganese (W-MnO) in the analytical sample by a
multiplying the concentration (Mn) of manganese standard solution for calibration
curve preparation or a measurement value (Mn) obtained in (4.2) by a conversion
factor (1.2912).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in d) and e) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a manganese standard
solution for the calibration curve in the concentration range which is suitable to a
device used.
281
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: An Optical Emission Spectrometer specified in JIS K
0116.
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
Note (2) The sampling amount of the analytical sample is 10 g when there is less manganese
content in the fertilizers such as a home garden-use fertilizer.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Optical Emission Spectrometer: Set up the measurement
conditions for the ICP Optical Emission Spectrometer considering the following:
Analytical line wavelength: 257.610 nm
b) Calibration curve preparation
1) Spray the manganese standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 257.610 nm.
2) Prepare a curve for the relationship between the manganese concentration and the indicated
value of the manganese standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 0.8 mg
as MnO) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the manganese content from the calibration curve, and water-soluble manganese
(W-MnO) in the analytical sample.
respectively.
The results of the repeatability tests on different days using a fluid mixed fertilizer
and a home garden-use mixed fertilizer to evaluate precision were analyzed by the
one-way analysis of variance. Table 1 shows the calculation results of intermediate
precision and repeatability.。
Additionally, the minimum limit of quantification of this testing method is about
0.0002 % (mass fraction).
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Fluid compound
7 5.69 0.02 0.4 0.06 1.1
fertilzier
Home garden-use
mixed fertilizers 7 2.29 0.02 0.8 0.04 1.6
(Liquid)
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days (T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
Table 2 Analysis results of the repeatability tests on different days
References
1) Keisuke AOYAMA: Simultaneous Determination of Water-Soluble Principal
Ingredients (W-P2O5, W-K2O, W-MgO, W-MnO and W-B2O3) in Liquid Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES), Research
Report of Fertilizer, Vol.8, p. 1 - 9 (2015)
(5) Flow sheet: The flow sheet for water-soluble manganese in fluid fertilizers is shown below:
Sample solution
283
Testing Methods for Fertilizers (2018)
Sample solution
284
Testing Methods for Fertilizers (2018)
4.8 Boron
4.8.1 Citrate-soluble boron
4.8.1.a Azomethine-H method
(1) Summary
This testing method is applicable to fertilizers not containing organic matters. This testing method is
classified as Type C and its symbol is 4.8.1.a-2018 or C-B.a-2.
Extract by adding a citric acid solution to an analytical sample, mask co-existing copper, iron and
other salts with ethylenediamine tetraacetate and measure the absorbance with azomethine-H borate
formed by the reaction with azomethine-H to obtain citrate soluble boron (C-B2O3) in an analytical
sample. In addition, the performance of this testing method is shown in Comment 6.
Comment 1 Instead of the boron standard solution in (2), a boron standard solution for the
calibration curve preparation can be prepared by using a boron standard solution (B 1
mg/mL or 10 mg/mL) traceable to National Metrology. In this case, calculate
citrate-soluble boron (C-B2O3) in the analytical sample by multiplying the
concentration (B) of a boron standard solution for calibration curve preparation or a
measurement value (B) obtained in (4.3) by a conversion factor (3.2199).
Note (2) Shake to mix the volumetric flask gently and disperse an analytical sample into the
citric acid solution.
Comment 2 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (3) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 3 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 4 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
Comment 5 If formaldehyde processed urea, a large quantity of aluminum, copper, iron zinc,
organic matters, etc. coexists to affect quantification, transfer a predetermined
amount of the sample solution (the equivalents of 0.05 mg - 1 mg as B2O3, no more
than 10 mL of the solution) to a 100-mL separatory funnel, add 10 mL of
hydrochloric acid (1+3), add water to about 20 mL and add 20 mL of
2-ethyl-1,3-hexanediol – 4-methyl-2-pentanone (1+9) to shake to mix with a
shaking apparatus for about 1 minute. After allowing to stand still, remove the
lower layer (aqueous phase) and add 20 mL of sodium hydroxide (20 mg/L) to
shake to mix with a shaking apparatus for about 1 minute. After allowing to stand
still, transfer the lower layer (aqueous phase) to a 100-mL volumetric flask, add a
few drops of phenolphthalein solution (1 g/100mL) to neutralize with hydrochloric
acid (1+3) until the color of the solution becomes achromatic, and conduct the
286
Testing Methods for Fertilizers (2018)
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 415 nm
b) Calibration curve preparation
1) Transfer 1 mL - 20 mL of boron standard solution (B2O3 0.05 mg/mL) to 100-mL volumetric
flasks step-by-step.
2) Add 15 mL of citric acid solution and conduct the same procedure as (4.2) c) to make the
B2O3 0.05 mg/100 mL - 1 mg/100 mL boron standard solution for the calibration curve
preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
test solution for the calibration curve preparation.
4) Measure absorbance at a wavelength of 415 nm of the boron standard solutions for the
calibration curve preparation using the blank test solution for the calibration curve preparation
as the control.
5) Prepare a curve for the relationship between the boron concentration and absorbance of the
boron standard solutions for the calibration curve preparation.
c) Sample measurement
1) Regarding the solution in (4.2) c), measure absorbance by the same procedure as b) 4).
2) Obtain the boron (B2O3) content from the calibration curve, and calculate citrate soluble
boron (C-B2O3) in the analytical sample.
Comment 6 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 10 % (mass fraction) and
0.05 % (mass fraction) were 101.5 % and 95.7 % as citrate-soluble boron (C-B2O3)
respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.02 % (mass fraction).
287
Testing Methods for Fertilizers (2018)
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 4) 5) 6) 7) 8) 9)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-A-10 11 0.209 0.004 2.0 0.005 2.2 0.006 3.1
FAMIC-A-13 10 0.203 0.004 1.8 0.005 2.5 0.009 4.7
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting Azomethine-H method
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.184- 187, Yokendo, Tokyo (1988)
2) Kimie KATO, Sakiko TAKAHASHI and Yuji SHIRAI: Validation of a Color metric
Method for Determination of Nitrogen, Phosphorus and Boron: Evaluation of
Calibration curve, Research Report of Fertilizer Vol. 2, p. 137 - 144 (2009)
3) Akira SHIMIZU: Verification of Performance Characteristics of Testing Methods for
Boron Content by Azomethine H Absorption Photometry, Research Report of Fertilizer
Vol. 6, p. 174 -182 (2013)
288
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate-soluble boron: The flow sheet for citrate-soluble boron in fertilizers is
shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 Flow sheet for citrate-soluble boron in fertilizers
(Extraction procedure (4.1.1))
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
289
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← Citric acid solution, until the volume reaches
to the equivalents of 15 mL
← 25 mL of ethylenediamine tetraacetate solution
← 10 mL of ammonium acetate solution
← 10 mL of azomethine-H solution
← Water (up to the marked line)
Leaving at rest For about 2 hours
290
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the boron standard solution in (2), a boron standard solution for the
calibration curve preparation can be prepared by using a boron standard solution (B 1
mg/mL or 10 mg/mL) traceable to National Metrology. In this case, calculate
citrate-soluble boron (C-B2O3) in the analytical sample by multiplying the
concentration (B) of a boron standard solution for calibration curve preparation or a
measurement value (B) obtained in (4.2) by a conversion factor (3.2199).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in f) and g) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a boron standard solution
291
Testing Methods for Fertilizers (2018)
for the calibration curve in the concentration range which is suitable to a device used.
Note (3) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1) in 4.2.3.a.
Note (4) Use a 250-mL flat-bottom volumetric flask to stabilize the vibration.
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.3.a.
Comment 5 The determination may be affected if an analytical sample cakes on the bottom of a
250-mL volumetric flask. Therefore, confirm the status of the non-dissolved matters
after the procedures of (4.1.1) b) and (4.1.2) b).
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
292
Testing Methods for Fertilizers (2018)
Comment 6 Wash the sample injector of an ICP-OES sufficiently with water because boron easily
causes the memory effect.
Comment 7 The simultaneous measurement of multiple elements by an ICP-OES is available. In
that case, see 4.2.3.d Comment 7.
Comment 8 The comparison of the measurement value (yi:0.073 % (mass fraction) - 0.51 %
(mass fraction )) of ICP Optical Emission Spectrometry and the measurement value
(xi) of Flame atomic absorption spectrometry analysis was conducted to evaluate
trueness using compound fertilizers (7 samples), mixed phosphate fertilizers (1
sample), solid fertilizers (2 samples), blended fertilizers (3 samples) and organic
compound fertilizer (1 sample). As a result, a regression equation was y = − 0.0408 +
1.0456x, and its correlation coefficient (r) was 0.992. In addition, recovery testing
was conducted using a preparation sample. As a result, the average rate of recovery
at the content level of 0.601 % (mass fraction) - 35.51 % (mass fraction) was 97.0 %
- 102.0 %.
The results of the repeatability tests on different days using compound fertilizers and
blended fertilizers to evaluate precision were analyzed by one-way analysis of
variance. Table 1 shows the calculation results of intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.01 % (mass fraction).
293
Testing Methods for Fertilizers (2018)
(5) Flow sheet for citrate-soluble boron: The flow sheet for citrate-soluble boron in fertilizers is
shown below:
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 Flow sheet for citrate-soluble boron in fertilizers
(Extraction procedure (4.1.1))
Cooling Immediately
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
294
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the boron standard solution in (2), a boron standard solution for the
calibration curve preparation can be prepared by using a boron standard solution (B
1 mg/mL or 10 mg/mL) traceable to National Metrology. In this case, calculate
water-soluble boron (W-B2O3) in the analytical sample by multiplying the
concentration (B) of a boron standard solution for calibration curve preparation or a
measurement value (B) obtained in (4.3) by a conversion factor (3.2199).
Note (2) The sampling amount of the analytical sample is 1 g when there is a high amount of
boric acid content in the fertilizers such as a borate fertilizer and boric acid fertilizer.
295
Testing Methods for Fertilizers (2018)
Comment 2 In the procedure in (4.1.2) a) and (4.1.2) b), a 250-mL volumetric flask can be used
instead of a 300-mL tall beaker. However, the volumetric flask used should be
distinguished as an extraction flask and should not be used for the other purposes.
Additionally, “cover with a watch glass” in b) is replaced by “place a funnel”. Skip
“transfer to a 250-mL volumetric flask with water” in the procedure in c).
Comment 3 The procedure in (4.1.2) is the same as the procedure in (4.1.1) in 4.3.3.a.
Comment 4 If formaldehyde processed urea, a large quantity of aluminum, copper, iron zinc,
organic matters, etc. coexists to affect quantification, transfer a predetermined
amount of the sample solution (the equivalents of 0.05 mg - 1 mg as B2O3, no more
than 10 mL of the solution) to a 100-mL separatory funnel, add 10 mL of
hydrochloric acid (1+3), add water to about 20 mL and add 20 mL of
2-ethyl-1,3-hexanediol–4-methyl-2-pentanone (1+9) to shake to mix with a shaking
apparatus for about 1 minute. After allowing to stand still, remove the lower layer
(aqueous phase) and add 20 mL of sodium hydroxide (20 mg/L) to shake to mix
with a shaking apparatus for about 1 minute. After allowing to stand still, transfer
the lower layer (aqueous phase) to a 100-mL volumetric flask, add a few drops of
phenolphthalein solution (1 g/100mL) to neutralize with hydrochloric acid (1+3)
until the color of the solution becomes achromatic, and conduct the procedure in
(4.2) b).
Comment 5 Water-soluble boron can be measured simultaneously with citrate soluble boron by
adding 15 mL of citric acid solution before the procedure in (4.2) b).
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 415 nm
b) Calibration curve preparation
1) Transfer 1 mL - 20 mL of boron standard solution (B2O3 0.05 mg/mL) to 100-mL volumetric
flasks step-by-step.
2) Conduct the same procedures as (4.2) b) to make B2O3 0.05 mg/100 mL - 1 mg/100 mL of the
boron standard solutions for the calibration curve preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
test solution for the calibration curve preparation.
4) Measure absorbance at a wavelength of 415 nm of the boron standard solutions for the
calibration curve preparation using the blank test solution for the calibration curve preparation
as the control.
5) Prepare a curve for the relationship between the boron concentration and absorbance of the
boron standard solutions for the calibration curve preparation.
296
Testing Methods for Fertilizers (2018)
c) Sample measurement
1) Regarding the solution in (4.2) b), measure absorbance by the same procedure as b) 4).
2) Obtain the boron (B2O3) content from the calibration curve, and calculate water-soluble boron
(W-B2O3) in the analytical sample.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.184- 187, Yokendo, Tokyo (1988)
2) Kimie KATO, Sakiko TAKAHASHI and Yuji SHIRAI: Validation of a Color metric
Method for Determination of Nitrogen, Phosphorus and Boron: Evaluation of
Calibration curve, Research Report of Fertilizer Vol. 2, p. 137 - 144 (2009)
3) Akira SHIMIZU: Verification of Performance Characteristics of Testing Methods for
Boron Content by Azomethine H Absorption Photometry, Research Report of Fertilizer
Vol. 6, p. 174 -182 (2013)
(5) Flow sheet for water-soluble boron: The flow sheet for water-soluble boron in fertilizers is
shown below:
2.5 g analytical sample Weigh to the order of 1 mg into a 300-mL tall beaker.
← About 200 mL of Water
Heating Cover with a watch glass, boil for 15 minutes
Cooling Immediately
Sample solution
297
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← 25 mL of ethylenediamine tetraacetate solution
← 10 mL of ammonium acetate solution
← 10 mL of azomethine-H solution
← Water (up to the marked line)
Leaving at rest For about 2 hours
298
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the boron standard solution in (2), a boron standard solution for the
calibration curve preparation can be prepared by using a boron standard solution (B 1
mg/mL or 10 mg/mL) traceable to National Metrology. In this case, calculate
water-soluble boron (W-B2O3) in the analytical sample by multiplying the
concentration (B) of a boron standard solution for calibration curve preparation or a
measurement value (B) obtained in (4.2) by a conversion factor (3.2199).
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in d) and e) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a boron standard solution
for the calibration curve in the concentration range which is suitable to a device used.
299
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: An Optical Emission Spectrometer specified in JIS K
0116.
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5% (volume fraction) in purity
Note (3) The sampling amount of the analytical sample is 10 g when there is less boron content
in the fertilizers such as a home garden-use fertilizer.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Optical Emission Spectrometer: Set up the
measurement conditions for the ICP Optical Emission Spectrometer considering the
following:
Analytical line wavelength: 249.773 nm
b) Calibration curve preparation
1) Spray the boron standard solution for the calibration curve preparation and blank test solution
for the calibration curve preparation into inductively coupled plasma, and read the indicated
value at a wavelength of 249.773 nm.
2) Prepare a curve for the relationship between the boron concentration and the indicated value
of the boron standard solution for the calibration curve preparation and the blank test solution
for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.02 mg - 1.6 mg
as B2O3) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the boron content from the calibration curve, and calculate water-soluble boron
(W-B2O3) in the analytical sample.
Comment 4 Wash the sample injector of an ICP-OES sufficiently with water because boron easily
causes the memory effect.
Comment 5 The simultaneous measurement of multiple elements by an ICP-OES is available. In
that case, see 4.2.4.d Comment 4.
Comment 6 The comparison of the measurement value (yi:0.013 % (mass fraction) - 0.530 %
(mass fraction )) of ICP Optical Emission Spectrometry and the measurement value
(xi) of Azomethine-H method was conducted to evaluate trueness using fluid
fertilizers (12 samples). As a result, a regression equation was y = − 0.0041 + 0.986x,
and its correlation coefficient (r) was 0.999. Additionally, additive recovery testing
was conducted using a fluid mixed fertilizer (1 brand) and a home garden-use mixed
300
Testing Methods for Fertilizers (2018)
fertilizer (1 brand) and a liquid micro element mixed fertilizer (1 sample). As a result,
the mean recovery rates at additive level of 0.15 % (mass fraction) - 0.2 % and
0.01 % (mass fraction) were 95.5 % - 99.4 % and 96.5 % respectively.
The results of the repeatability tests on different days using a fluid mixed fertilizer
and a home garden-use mixed fertilizer to evaluate precision were analyzed by the
one-way analysis of variance. Table 1 shows the calculation results of intermediate
precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.0005 % (mass fraction).
References
1) Keisuke AOYAMA: Simultaneous Determination of Water-Soluble Principal
Ingredients (W-P2O5, W-K2O, W-MgO, W-MnO and W-B2O3) in Liquid Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES), Research
Report of Fertilizer, Vol.8, p. 1 - 9 (2015)
(5) Flow sheet: The flow sheet for water-soluble boron in fluid fertilizers is shown below:
Sample solution
Figure 1 Flow sheet for water-soluble boron in liquid fertilizers (Extraction procedure)
301
Testing Methods for Fertilizers (2018)
Sample solution
Figure 2 Flow sheet for water-soluble boron in liquid fertilizers (Measurement procedure)
302
Testing Methods for Fertilizers (2018)
4.9 Zinc
4.9.1 Total zinc
4.9.1.a Flame atomic absorption spectrometry
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type C and its
symbol is 4.9.1.a-2017 or T-Zn.a-1.
Pretreat an analytical sample with incineration and nitric acid-hydrochloric acid (1+3), and then
spray in an acetylene−air flame, and measure the atomic absorption with zinc at a wavelength of
213.9 nm to quantify the total zinc (T-Zn). In addition, the performance of this testing method is
shown in Comment 6.
Comment 1 Instead of the zinc standard solution in (2), a zinc standard solution for the
calibration curve preparation can be prepared by using a zinc standard solution (Zn 1
mg/mL or 10 mg/mL) traceable to National Metrology.
Note (2) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
Note (3) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 550 °C in 1 to 2 hours.
(4) The watch glass can be removed.
(5) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.b, (4.1) of
4.10.1.a, (4.1) of 4.10.1.b, (4.1) of 5.3.a, (4.1) of 5.3.b, (4.1) of 5.4.a, (4.1) of 5.4.b,
(4.1) of 5.5.a, (4.1) of 5.5.d, (4.1) of (5.6.a) and (4.1) of 5.6.b. In addition, the
sample solution can be used in 4.2.1.a, 4.2.1.b, 4.3.1.a, 4.3.1.b and 4.5.1.a.
Comment 4 The sample solution prepared in (4.1.3) in 4.2.1.a can also be used.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 213.9 nm
b) Calibration curve preparation
1) Spray the zinc standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 213.9 nm.
2) Prepare a curve for the relationship between the zinc concentration and the indicated value of
the zinc standard solutions for the calibration curve preparation and the blank test solution for
the calibration curve preparation.
c) Sample measurement
1) Subject the sample solution (6) to the same procedure as in b) 1) to read the indicated value.
2) Subject the blank test solution to the same procedure as in b) 1) to read the indicated value,
and correct the indicated value obtained for the sample solution.
3) Obtain the zinc content from the calibration curve, and calculate the total zinc (T-Zn) in the
analytical sample.
Note (6) If there is a possibility that the zinc concentration in the sample solution will exceed
the maximum limit of the calibration curve, dilute a predetermined amount with
304
Testing Methods for Fertilizers (2018)
Comment 5 The zinc concentration in the analytical sample can also be corrected by subjecting
the blank test solution to the same procedures as in 1) and 3) to obtain the zinc
content in the blank test solution.
Comment 6 Recovery testing was conducted to evaluate trueness using a preparation sample. As
a result, the average rate of recovery at the content level of 1.2 % (mass fraction) and
90 mg/kg are 99.5 % and 97.8 % as the total zinc (T-Zn) respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this test method is about 5
mg/kg for solid fertilizers.
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 3) 4) 5) 6) 7) 8)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1)
material fertilizer p (mg/kg) (mg/kg) (%) (mg/kg) (%) (mg/kg) (%)
FAMIC-C-12 12 992 11 1.1 17 1.7 32 3.3
1) The number of laboaratories used for analysis 5) Intermediate standard deviation
conducting Cold vapor atomic absorption spectrometry
2) Average (the number of laboratory(p )×test days (2) 6) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Repeatability standard deviation 7) Reproducibility standard deviation
4) Repeatability relative standard deviation 8) Reproducibility relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.193- 194, Yokendo, Tokyo (1988)
2) Kimie KATO, Masayuki YOSHIMOTO and Yuji SHIRAI: 3) Kimie KATO,
Masayuki YOSHIMOTO and Yuji SHIRAI: Systematization of Determination Methods
of Major Components in Sludge Fertilizer, Compost and Organic Fertilizer, Research
Report of Fertilizer, Vol.3 p. 107 - 116 (2010)
3) Shin ABE and Yoshiyuki SUNAGA: Verification of Performance Characteristics of
Testing Methods for Zinc in Fertilizer by Atomic Absorption Spectrometry, Research
Report of Fertilizer Vol. 6, p. 156 -164 (2013)
305
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total zinc: The flow sheet for total zinc in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
306
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the zinc standard solution in (2), a zinc standard solution for the
calibration curve preparation can be prepared by using a zinc standard solution (Zn 1
mg/mL or 10 mg/mL) traceable to National Metrology.
Note (2) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for
about 1 hour and further raise to 450 °C in 1 to 2 hours.
(3) The watch glass can be removed.
(4) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct measurement (Standard Addition Method) according to JIS K 0116
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Optical Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 206.191 nm
b) Calibration curve preparation and sample measurement
1) Put 5mL of sample solution to three 10-mL volumetric flasks respectively.
2) Add 2mL and 4 mL of zinc standard solution (0.25 μg/mL) to volumetric flasks of 1) above,
then add hydrochloric acid (1+23) up to the marked line to make a sample solution of
Standard Addition Method.
3) Add hydrochloric acid (1+23) to the marked line of the remaining volumetric flask of 1)
above to make a sample solution without a standard solution.
4) Spray the sample solution of Standard Addition Method and the sample solution without a
standard solution into the induction plasma, and read the indicated value at a wavelength of
206.191 nm.
5) Transfer 5 mL of blank test solution to a 10-mL volumetric flask, conduct the same
procedures as in 3) - 4) to read the indicated value, and correct the indicated value obtained
from the respective sample solutions.
6) Prepare a curve for the relationship between the added zinc concentration and the corrected
indicated value of the sample solution for Standard Addition Method and the sample solution
without a standard solution.
7) Obtain the zinc content from the intercept of the calibration curve to calculate the total zinc
(T-Zn) in the analytical sample.
Comment 4 The total zinc (T-Zn) in the analytical sample can also be corrected by subjecting the
blank test solution to the same procedures as in b) 1) - b) 4) and b) 6) - b) 7) to
obtain the zinc content in the blank test solution.
Comment 5 The simultaneous measurement of multiple elements by an ICP-OES is available. In
this case, transfer a pre-determined amount of copper standard solution (Cu 0.1
mg/mL, 1 mg/mL or 10 mg/mL), zinc standard solution (Zn 0.1 mg/mL, 1 mg/mL or
10 mg/mL), cadmium standard solution (Cd 0.1 mg/mL, 1 mg/mL or 10 mg/mL),
nickel standard solution (Ni 0.1 mg/mL, 1 mg/mL or 10 mg/mL), chromium standard
solution (Cr 0.1 mg/mL, 1 mg/mL or 10 mg/mL) and lead standard solution (Pb 0.1
308
Testing Methods for Fertilizers (2018)
Comment 6 The comparison of the measurement value (xi:65.0 mg/kg - 3310 mg/kg) of ICP
Optical Emission Spectrometry and the measurement value (yi) of Flame Atomic
Absorption Spectrometry was conducted to evaluate trueness using sludge fertilizers
(49 samples). As a result, a regression equation was y = −47.6 + 1.080x and its
correlation coefficient (r) was 0.995. Triplicates measurement for each one sample of
sewage sludge fertilizer, human waste sludge fertilizer, industrial sludge fertilizer,
mixed sludge fertilizer, calcined sludge fertilizers and composted sludge fertilizer
was conducted. As a result, a repeatability obtained was 0.1 % - 2.3 % as a relative
standard deviation.
Additionally, the minimum limit of quantification of this testing method is about 8
mg/kg.
References
1) Masahiro ECHI, Tomoe INOUE, Megumi TABUCHI and Tetuya NOMURA:
Simultaneous Determination of Cadmium, Lead, Nickel, Chromium, Copper and Zinc
in Sludge Fertilizer using Inductively Coupled Plasma-Atomic Emission Spectrometry
(ICP-AES), Research Report of Fertilizer, Vol.4, p. 30 - 35 (2011)
309
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total zinc: The flow sheet for total zinc in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
Sample solution
310
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the zinc standard solution in (2), a zinc standard solution for the calibration
curve preparation can be prepared by using a zinc standard solution (Zn 1 mg/mL or
10 mg/mL) traceable to National Metrology.
Note (2) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
Comment 2 In the procedure in (4.1.1) a), it is also allowed to weigh 2.50 g of the analytical
311
Testing Methods for Fertilizers (2018)
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 213.9 nm
b) Calibration curve preparation
1) Spray the zinc standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 213.9 nm.
2) Prepare a curve for the relationship between the zinc concentration and the indicated value of
the zinc standard solutions for the calibration curve preparation and the blank test solution for
the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.05 mg - 0.5 mg
as Zn) to a 100-mL volumetric flask.
2) Add about 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the zinc content from the calibration curve, and calculate the water-soluble zinc
(W-Zn) in the analytical sample.
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample
(solid). As a result, the average rate of recovery at the content level of 10 % (mass
fraction), 2 % (mass fraction) and 0.01 % (mass fraction) are 101.6 %, 101.9 % and
98.9 % as water-soluble zinc (W-Zn) respectively. In addition, recovery testing was
conducted using a preparation sample (fluid). As a result, the average rate of
recovery at the content level of 1 % (mass fraction), 0.05 % (mass fraction) and 20
mg/kg are 99.6 %, 100.4 % and 100.6 % as water-soluble zinc respectively.
The results of the repeatability tests on different days using a fluid mixed fertilizer
and a liquid microelement mixed fertilizer to evaluate the extract precision of fluid
fertilizers were analyzed by the one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about 10
mg/kg (solid fertilizers) and 0.9 mg/kg (fluid fertilizers).
312
Testing Methods for Fertilizers (2018)
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Fluid mixed fertilizer 7 1.28 0.01 0.4 0.02 1.3
Liquid microelement
7 0.230 0.001 0.5 0.003 1.5
mixed fertilizers
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.192- 194, Yokendo, Tokyo (1988)
2) Shin ABE and Yoshiyuki SUNAGA: Verification of Performance Characteristics of
Testing Methods for Zinc in Fertilizer by Atomic Absorption Spectrometry, Research
Report of Fertilizer Vol. 6, p. 156 -164 (2013)
3) Shinji KAWAGUCHI: Simple Extraction Method for Water-Soluble Components in
Liquid Compound Fertilizers, Research Report of Fertilizer, Vol.9, p. 10 - 20 (2016)
(5) Flow sheet for water-soluble zinc: The flow sheet for water-soluble zinc in fertilizers is
shown below:
5.00 g analytical
sample 500-mL volumetric flask.
(powdery)
← Water, About 400 mL
Shaking to mix Rotary shaker (30 - 40 revolutions/min), for 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
313
Testing Methods for Fertilizers (2018)
1.00 g analytical
sample 100-mL volumetric flask.
(fluid)
← Water, About 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-2 Flow sheet for water-soluble zinc in fertilizers
(Extraction procedure(4.1.2))
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← 25 mL of hydrochloric acid (1+5)
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (213.9 nm)
314
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the zinc standard solution in (2), a zinc standard solution for the calibration
curve preparation can be prepared by using a zinc standard solution (Zn 10 mg/mL)
traceable to National Metrology.
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in e) and f) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, in case of using the
ICP-OES of an axial observation mode, it is recommended to prepare a zinc standard
solution for the calibration curve in the concentration range which is suitable to a
device used.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: A spectrophotometer specified in JIS K 0115
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
Note (2) The sampling amount of the analytical sample is 10 g when the content in the sample
is less than 0.01 % (mass fraction) as water-soluble zinc.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 213.856 nm
b) Calibration curve preparation
1) Spray the zinc standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 213.856 nm.
2) Prepare a curve for the relationship between the zinc concentration and the indicated value of
the zinc standard solutions for the calibration curve preparation and the blank test solution for
the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 2 mg as
Zn) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the zinc content from the calibration curve, and calculate the water-soluble zinc
(W-Zn) in the analytical sample.
316
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Simultaneous Determination Method for Effect-Development
Promoting Agent (Ca, Fe, Co, Cu, Zn and Mo) in Liquid Compound Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometer (ICP-OES), Research
Report of Fertilizer, Vol.9, p. 1 - 9 (2016)
(5) Flow sheet for water-soluble zinc: The flow sheet for water-soluble zinc in fluid fertilizers is
shown below:
Sample solution
Figure 1 Flow sheet for water-soluble zinc in liquid fertilizers (Extraction procedure)
Sample solution
Figure 2 Flow sheet for water-soluble zinc in liquid fertilizers (Measurement procedure)
317
Testing Methods for Fertilizers (2018)
4.10 Copper
4.10.1 Total copper
4.10.1.a Flame atomic absorption spectrometry
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type C and its
symbol is 4.10.1.a-2017 or T-Cu.a-1.
Pretreat an analytical sample with incineration and nitric acid−hydrochloric acid (1+3), and then
spray in an acetylene−air flame, and measure the atomic absorption with copper at a wavelength of
324.8 nm to obtain the total copper (T-Cu) in an analytical sample. In addition, the performance of
this testing method is shown in Comment 6.
Comment 1 Instead of the copper standard solution in (2), a copper standard solution for the
calibration curve preparation can be prepared by using a copper standard solution
(Cu 1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (2) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
Note (3) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(4) The watch glass can be removed.
(5) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
Comment 4 The sample solution prepared in (4.1.2) in 4.2.1.a can also be used.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 324.8 nm
b) Calibration curve preparation
1) Spray the copper standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 324.8 nm.
2) Prepare a curve for the relationship between the copper concentration and the indicated value
of the copper standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Subject the sample solution (6) to the same procedure as in b) 1) to read the indicated value.
2) Subject the blank test solution to the same procedure as in b) 1) to read the indicated value,
and correct the indicated value obtained for the sample solution.
3) Obtain the copper content from the calibration curve, and calculate the total copper (T-Cu) in
the analytical sample.
Note (6) If there is a possibility that the copper concentration in the sample solution will exceed
the maximum limit of the calibration curve, dilute a predetermined amount with
hydrochloric acid (1+23).
319
Testing Methods for Fertilizers (2018)
Comment 5 The copper concentration in the analytical sample can also be corrected by subjecting
the blank test solution to the same procedures as in 1) and 3) to obtain the copper
content in the blank test solution.
Comment 6 Recovery testing was conducted to evaluate trueness using a preparation sample. As
a result, the average rate of recovery at the content level of 0.15 % (mass fraction)
and 0.03 % (mass fraction) are 100.4 % and 99.6 % as total copper (T-Cu)
respectively.
The results of the collaborative study to determine a certified reference material
fertilizer were analyzed by using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this test method is about 4
mg/kg.
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
2) 3) 4) 5) 6) 7) 8)
certified reference laboratory Average sr RSD r s I(T) RSD I(T) sR RSD R
1)
material fertilizer p (mg/kg) (mg/kg) (%) (mg/kg) (%) (mg/kg) (%)
FAMIC-C-12 11 583 7 1.1 11 1.9 22 3.8
1) The number of laboaratories used for analysis 5) Intermediate standard deviation
conducting Cold vapor atomic absorption spectrometry
2) Average (the number of laboratory(p )×test days (2) 6) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Repeatability standard deviation 7) Reproducibility standard deviation
4) Repeatability relative standard deviation 8) Reproducibility relative standard deviation
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.254- 255, Yokendo, Tokyo (1988)
2) Kimie KATO, Masayuki YOSHIMOTO and Yuji SHIRAI: Systematization of
Determination Methods of Major Components in Sludge Fertilizer, Compost and
Organic Fertilizer, Research Report of Fertilizer, Vol.3 p. 107 - 116 (2010)
3) Shin ABE and Yoshiyuki SUNAGA: Verification of Performance Characteristics of
Testing Methods for Copper in Fertilizer by Atomic Absorption Spectrometry, Research
Report of Fertilizer, Vol. 6, p. 165 - 173 (2013)
320
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total copper: The flow sheet for total copper in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
321
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the copper standard solution in (2), a copper standard solution for the
calibration curve preparation can be prepared by using a copper standard solution
(Cu 1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (2) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(3) The watch glass can be removed.
(4) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 324.754 nm
b) Calibration curve preparation and sample measurement
1) Put 5mL of sample solution to three 10-mL volumetric flasks respectively.
2) Add 2mL and 4 mL of copper standard solution (2.5 μg/mL) to volumetric flasks of 1) above,
then add hydrochloric acid (1+23) to the marked line to make a sample solution of Standard
Addition Method.
3) Add hydrochloric acid (1+23) to the marked line of the remaining volumetric flask of 1)
above to make a sample solution without a standard solution.
4) Spray the sample solution of Standard Addition Method and the sample solution without a
standard solution into the induction plasma, and read the indicated value at a wavelength of
324.754 nm.
5) Transfer 5 mL of blank test solution to a 10-mL volumetric flask, conduct the same
procedures as in 3) - 4) to read the indicated value, and correct the indicated value obtained
from the respective sample solutions.
6) Prepare a curve for the relationship between the added copper concentration and the corrected
indicated value of the sample solution for Standard Addition Method and the sample solution
without a standard solution.
7) Obtain the copper content from the intercept of the calibration curve to calculate the total
copper (T-Cu) in the analytical sample.
Comment 4 The copper concentration in the analytical sample can also be corrected by subjecting
the blank test solution to the same procedures as in b) 1) - b) 4) and b) 6) - b) 7) to
obtain the copper content in the blank test solution.
Comment 5 The simultaneous measurement of multiple elements by an ICP-OES is available. In
that case, see 4.9.1.b Comment 5.
Comment 6 The comparison of the measurement value (xi:12.0 mg/kg - 1400 mg/kg) of ICP
Optical Emission Spectrometry and the measurement value (yi) of Flame Atomic
Absorption Spectrometry was conducted to evaluate trueness using sludge fertilizers
(49 samples). As a result, a regression equation was y = − 5.5 + 1.062x and its
correlation coefficient (r) was 0.997. Triplicates measurement for each one sample of
323
Testing Methods for Fertilizers (2018)
sewage sludge fertilizer, human waste sludge fertilizer, industrial sludge fertilizer,
mixed sludge fertilizer, calcined sludge fertilizers and composted sludge fertilizer
was conducted. As a result, a repeatability obtained was 0.6 % - 1.8 % as a relative
standard deviation.
Additionally, the minimum limit of quantification of this testing method is about 3
mg/kg.
References
1) Masahiro ECHI, Tomoe INOUE, Megumi TABUCHI and Tetuya NOMURA:
Simultaneous Determination of Cadmium, Lead, Nickel, Chromium, Copper and Zinc
in Sludge Fertilizer using Inductively Coupled Plasma-Atomic Emission Spectrometry
(ICP-AES), Research Report of Fertilizer, Vol.4, p. 30 - 35 (2011)
(5) Flow sheet for total copper: The flow sheet for total copper in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
324
Testing Methods for Fertilizers (2018)
Sample solution
325
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the copper standard solution in (2), a copper standard solution for the
calibration curve preparation can be prepared by using a copper standard solution
(Cu 1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (2) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
326
Testing Methods for Fertilizers (2018)
Comment 2 In the procedure in (4.1.1) a), it is also allowed to weigh 2.50 g of the analytical
sample and transfer to a 250-mL volumetric flask.
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 324.8 nm
b) Calibration curve preparation
1) Spray the copper standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 324.8 nm.
2) Prepare a curve for the relationship between the copper concentration and the indicated value
of the copper standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.05 mg - 0.5 mg
as Cu) to a 100-mL volumetric flask.
2) Add about 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the copper content from the calibration curve, and calculate the water-soluble copper
(W-Cu).
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample
(solid). As a result, the average rate of recovery at the content level of 10 % (mass
fraction), 1 % (mass fraction) and 0.03 % (mass fraction) are 100.7 %, 99.4 % and
102.6 % as water-soluble copper (W-Cu) respectively. In addition, recovery testing
was conducted using a preparation sample (fluid). As a result, the average rate of
recovery at the content level of 1 % (mass fraction), 0.05 % (mass fraction) and 20
mg/kg are 98.8 %, 99.3 % and 101.4 % as water-soluble copper respectively.
The results of the repeatability tests on different days using a fluid mixed fertilizer
and a liquid microelement mixed fertilizer to evaluate the extract precision of fluid
fertilizers were analyzed by the one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about 10
mg/kg (solid fertilizers) and 3 mg/kg (fluid fertilizers).
327
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.254- 255, Yokendo, Tokyo (1988)
2) Shin ABE and Yoshiyuki SUNAGA: Verification of Performance Characteristics of
Testing Methods for Copper in Fertilizer by Atomic Absorption Spectrometry, Research
Report of Fertilizer, Vol. 6, p. 165 - 173 (2013)
3) Shinji KAWAGUCHI: Simple Extraction Method for Water-Soluble Components in
Liquid Compound Fertilizers, Research Report of Fertilizer, Vol.9, p. 10 - 20 (2016)
(5) Flow sheet for water-soluble copper: The flow sheet for water-soluble copper in fertilizers is
shown below:
5.00 g analytical
sample 500-mL volumetric flask.
(powdery)
← Water, About 400 mL
Shaking to mix Rotary shaker (30 - 40 revolutions/min), for 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
328
Testing Methods for Fertilizers (2018)
1.00 g analytical
sample 100-mL volumetric flask.
(fluid)
← Water, About 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← 25 mL of hydrochloric acid (1+5)
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (324.8 nm)
329
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the copper standard solution in (2), a copper standard solution for the
calibration curve preparation can be prepared by using a copper standard solution
(Cu 10 mg/mL) traceable to National Metrology.
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in e) and f) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, in case of using the
ICP-OES of an axial observation mode, it is recommended to prepare a copper
standard solution for the calibration curve in the concentration range which is
suitable to a device used.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: A spectrophotometer specified in JIS K 0115
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
flask.
b) Add about 50 mL of water, shake to mix and add water up to the marked line.
c) Filter with Type 3 filter paper to make a sample solution.
Note (2) The sampling amount of the analytical sample is 10 g when the content in the sample
is less than 0.01 % (mass fraction) as water-soluble copper.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 327.396 nm
b) Calibration curve preparation
1) Spray the copper standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 327.396 nm.
2) Prepare a curve for the relationship between the copper concentration and the indicated value
of the copper standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 2 mg as
Cu) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the copper content from the calibration curve, and calculate the water-soluble copper
(W-Cu).
331
Testing Methods for Fertilizers (2018)
Home garder-use
7 0.00976 0.00006 0.6 0.00033 3.4
mixed fertilizer(liquid)
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
References
1) Keisuke AOYAMA: Simultaneous Determination Method for Effect-Development
Promoting Agent (Ca, Fe, Co, Cu, Zn and Mo) in Liquid Compound Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometer (ICP-OES), Research
Report of Fertilizer, Vol.9, p. 1 - 9 (2016)
(5) Flow sheet for water-soluble copper: The flow sheet for water-soluble copper in fluid
fertilizers is shown below:
Sample solution
Sample solution
332
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Hot plate: A hot plate whose surface temperature can be adjusted up to 250 ºC.
b) Sample digestion flask (3): A 100-mL borosilicate glass volumetric flask 100 mL (180 mm
total height, 13 mm mouth diameter)
Note (3) Distinguish the volumetric flask used in digestion as a sample digestion flask and do
not use it for any other purposes.
Comment 1 Sample an analytical sample from a test sample prepared in 2.3.3 Grinding (3.1) by
grinding with a mill until it completely passes through a sieve of 500 µm aperture or
from a test sample prepared in 2.3.3 Grinding Comment 1.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
a) Transfer 20 mL of the sample solution to a 100-mL Erlenmeyer flask.
b) Add a 0.2 mol/L ammonium iron (II) sulfate solution drop-by-drop until the brown color of
dichromate ion almost disappears from the sample solution.
c) Add about a 0.25 mL of N-phenylanthranilic acid solution (6), and titrate with a 0.2 mol/L
ammonium iron (II) sulfate solution until the color of the solution changes from dark
red-purple to blue-green.
d) Transfer 20 mL of the blank test solution to a 100-mL Erlenmeyer flask, and conduct the
procedures in b) - c) to titrate.
e) Calculate the organic carbon (O-C) in the analytical sample by the following formula:
V4: Volume (mL) of 0.2 mol/L ammonium iron (II) sulfate solution needed for the
titration of the blank test solution
V5: Volume (mL) of 0.2 mol/L ammonium iron (II) sulfate solution needed for the
titration of the sample solution
C: Set concentration (0.2 mol/L) of 0.2 mol/L ammonium iron (II) sulfate solution
f: Factor of 0.2 mol/L ammonium iron (II) sulfate solution
V6: Predetermined volume (100 mL) of the sample solution and the blank test
solution in (4.1) d)
334
Testing Methods for Fertilizers (2018)
V7: Transferred volume (20 mL) of the sample solution and the blank test solution
subjected to titration in (4.2) a) and (4.2) d)
W2: Mass (g) of the analytical sample
Note (6) About 5 drops with a 1-mL - 2-mL Komagome pipette. Add the same volume to the
sample solution and the blank test solution.
Comment 2 The results of the collaborative study to determine a certified reference material
fertilizer were analyzed using a three-level nesting analysis of variance. Table 1
shows the calculation results of reproducibility, intermediate precision and
repeatability.
Additionally, the minimum limit of quantification of this testing method is about
1.5 % (mass fraction).
Table 1 Analysis results of the collaborative study to determine a certified reference material fertilizer
Name of Number of Repeatability Intermediate precision Reproducibility
certified reference laboratory Average2) sr
4)
RSD r
5) 6)
s I(T) RSD I(T)
7)
sR
8) 9)
RSD R
1) 3) 3) 3) 3)
material fertilizer p (%) (%) (%) (%) (%) (%) (%)
FAMIC-C-12 12 20.2 0.3 1.5 0.5 2.3 0.6 3.1
1) The number of laboaratories used for analysis 6) Intermediate standard deviation
conducting Dichromate oxidation
2) Average (the number of laboratory(p )×test days(2) 7) Intermediate relative standard deviation
×the number of replicate testing(3))
3) Mass fraction 8) Reproducibility standard deviation
4) Repeatability standard deviation 9) Reproducibility relative standard deviation
5) Repeatability relative standard deviation
References
1) Yuji SHIRAI, Yuko SEKINE, and Toshiaki HIROI: Validation of Determination Method
for Organic Form Carbon in Sludge Fertilizer and Compost, Research Report of
Fertilizer, Vol.3, p. 117 - 122 (2010)
(5) Flow sheet for organic carbon: The flow sheet for organic carbon in sludge fertilizers and
compost, etc. is shown below:
335
Testing Methods for Fertilizers (2018)
Cooling Immediately
← Water (fill up to100 mL)
20 mL aliquot 100-mL Erlenmeyer flask
Sample solution
Figure 1 Flow sheet for organic carbon in sludge fertilizers and compost, etc.
(Dichromate oxidation procedure)
Sample solution
Figure 2 Flow sheet for organic carbon in sludge fertilizers and compost, etc.
(Measurement procedure)
336
Testing Methods for Fertilizers (2018)
Comment 1 Sea sand (particle diameter 425 µm - 850 µm) is commercially sold by FUJIFILM
Wako Pure Chemical Co., Ltd. and YONEYAMA YAKUHIN KOGYO Co., Ltd.
Note (1) The setup of the program and the parameters of the analyzer are according to the
specification and the operation method of the total nitrogen-total carbon analyzer by the
combustion method used.
(4) Test procedures: Conduct measurement as shown below. However, confirm that there is no
difference from the measured value of organic carbon obtained in advance according to
4.11.1.a by using an analytical sample.
Note (2)The additive amount of hydrochloric acid (1+3) is merely a target. It is enough to make
337
Testing Methods for Fertilizers (2018)
the whole analytical sample come into contact with hydrochloric acid. Let it stand for a
short time in the case of producing bubbles.
(3) In some cases, it is not necessary to add water due to the capacity of a container.
(4) Shake calmly the container for combustion to make the analytical sample come into
complete contact with hydrochloric acid.
(5) Remove hydrochloric acid completely
Comment 2 Sample an analytical sample from a test sample prepared in 2.3.3 Grinding (3.1) by
grinding with a mill until it completely passes through a sieve of 500 µm aperture or
from a test sample prepared in 2.3.3 Grinding Comment 1.
Comment 3 When it is confirmed that, for example, the volatilization of hydrogen chloride is not
detected by using a test paper, etc. and hydrochloric is completely removed in the
procedure d), the procedure e) can be skipped.
(4.2) Measurement: Specific measurement procedures are according to the operation method of a
total nitrogen-total carbon analyzer by the combustion method.
a) Measurement conditions for the total nitrogen-total carbon analyzer by the combustion
method: Set up the measurement conditions for the total nitrogen-total carbon analyzer
considering the following:
Combustion temperature: No less than 870 ºC
b) Calibration curve preparation
1) Turn on the total nitrogen-total carbon analyzer by the combustion method (1), and adjust so
that stable indicated values can be obtained.
2) Weigh a predetermined amount of the standard for calibration curves (6) to the order of 0.1 mg
into a combustion vessel.
3) Insert the combustion vessel into the total nitrogen-total carbon analyzer by the combustion
method, and read the indicated value.
4) Conduct the procedure in 3) for another combustion vessel for a blank test, and read the
indicated value.
5) Prepare a curve for the relationship between the carbon content and the indicated value of the
standard for calibration curves and the blank test for calibration curves.
c) Sample measurement
1) Insert the combustion vessel containing the test sample to the total nitrogen-total carbon
analyzer by the combustion method, and read the indicated value.
2) Obtain the carbon content from the calibration curve, and calculate organic carbon in the
analytical sample.
Note (6) Standard for calibration curves: DL-Aspartic acid (purity no less than 99 % (mass
fraction)), EDTA (purity no less than 99 % (mass fraction)), hippuric acid (purity no less
than 98 % (mass fraction)) or other reagents having equivalent purity recommended by
the total nitrogen-total carbon analyzer by the combustion method used.
Comment 4 The comparison of the measurement value (yi:0.21 % (mass fraction) - 45.40 %
(mass fraction)) of Combustion method and the measurement value (xi) of
Dichromate oxidation was conducted to evaluate trueness using sludge fertilizers and
compost (total 25 samples). As a result, a regression equation was y = 0.004 + 1.009x
and its correlation coefficient (r) were 0.999.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of this testing method is about
338
Testing Methods for Fertilizers (2018)
References
1) Aiko YANO, Satono AKIMOTO, and Yuji SHIRAI: Determination of Organic Carbon
in Sludge Fertilizer and Compost by Hydrochloric Acid-Treated Combustion Method,
Research Report of Fertilizer, Vol.9, p. 9 - 19 (2013)
2) Aiko YANO and Yuji SHIRAI: Determination of Organic Carbon in Sludge Fertilizer
and Compost by Hydrochloric Acid-Treated Combustion Method: A Collaborative
Study, Research Report of Fertilizer, Vol. 7, p. 22 - 27 (2014)
339
Testing Methods for Fertilizers (2018)
(5) Flow sheet for organic carbon: The flow sheet for organic carbon in sludge fertilizers and
compost, etc. is shown below:
Test sample
340
Testing Methods for Fertilizers (2018)
Reference: Chromatograms of the standard for calibration curves and an analytical sample are
shown below:
341
Testing Methods for Fertilizers (2018)
O-C: Organic carbon (% (mass fraction)) (1) in the analytical sample obtained in
4.11.1
T-N: Total nitrogen (% (mass fraction)) (1) in the analytical sample obtained in
4.1.1
Note (1) O-C and T-N use raw data without rounding numerical value
342
Testing Methods for Fertilizers (2018)
4.12 Sulfur
4.12.1 Total sulfur content
4.12.1.a Potassium permanganate analysis
(1) Summary
This testing method is applicable to fertilizers mainly containing ferrous sulfate (iron (II) sulfate)
(FeSO4)) among sulfur and its compound. This testing method is classified as Type D and its
symbol is 4.12.1.a-2017 or T-S.a-1.
Dissolve an analytical sample in water and diluted sulfuric acid, add phosphoric acid and then titrate
ferrous sulfate (iron (II) sulfate) (FeSO4)) by oxidation-reduction with a potassium permanganate
solution to obtain the total sulfate (T-SO3). In addition, the performance of this testing method is
shown in Comment 1.
Note (1) The endpoint is reached when the color of solution keeps as it is for 30 seconds after
coloring
a) Weigh 0.5 g - 1 g of an analytical sample to the order of 0.1 mg, and transfer to a 200-mL tall
beaker.
b) Add about 50 mL of water and about 15 mL of sulfuric acid (1+5) and shake with a magnetic
stirrer to dissolve.
c) Immediately add about 1 mL of phosphate, and then titrate with a 0.02 mol/L potassium
permanganate solution until the color of the solution becomes light red (2).
d) As a blank test, conduct the procedures in b) - c) using another 200-mL tall beaker to titrate
(2)
.
e) Calculate the total sulfur content (T-SO3) in an analytical sample by the following formula.
Comment 1 Recovery testing was conducted using a reagent (ferrous sulfate heptahydrate); as a
result, the average rate of recovery was 29.1 % (mass fraction) as total sulfate
content (T-SO3). The recovery rate to a theoretical value was 101.0 %.
Additionally, the minimum limit of quantification of this testing method is about
0.04 % (mass fraction).
References
1) Yasushi SUGIMURA and Shinjiro IZUKA: Method validation of Redox Titration for
Determination of Sulfur content (as sulfur trioxide) in Fertilizers of Ferrous sulfate and
its mixture materials, Research Report of Fertilizer, Vol.3, 25 - 29 (2010).
2) JIS K 8978: Iron (II) sulfate heptahydrate (Reagent) (2008)
(5) Flow sheet for total sulfur content: The flow sheet for total sulfur content in fertilizers
mainly containing ferrous sulfate is shown below.
0.5 g - 1g
Weigh to the order of 0.1 mg into a 200-mL tall beaker
analytical sample
← About 50 mL of water
← About 15 mL of sulfuric acid (1+5)
Dissolution Shake to mix
← About 1 mL of phosphate
0.02 mol/L potassium permanganate solution
Titration
(until the solution becomes light red)
Figure Flow sheet for total sulfur content (raw material: ferrous sulfate)
344
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Hot plate: A hot plate whose surface temperature can be adjusted up to 250 ºC.
b) Water bath: Water bath that can be adjusted to 80 ºC - 90 ºC.
c) Crucible: After heating porcelain crucible or platinum crucible in an electric furnace at 800
ºC in advance, let it stand to cool in a desiccator and measure the mass to the order of 0.1 mg.
d) Drying apparatus: Drying apparatus that can be adjusted to 110 ºC - 120 ºC.
e) Electric furnace: An electric furnace that can be kept at 800 ºC ± 5 ºC.
Note (2) Omit extraction if fluid fertilizers are made from only sulfuric acid and all materials are
dissolved
(3) Until sulfur content is dissolved. About 5 minutes when raw materials, etc. are not
dissolved.
345
Testing Methods for Fertilizers (2018)
Note (5) Weigh 1g - 5g of an analytical sample to the order of 0.1 mg if the fluid fertilizers of the
analytical sample are made from only sulfuric acid and all materials are dissolved.
(6) It can be suspended after the procedures end.
(7) A pH meter can be used for neutralization.
(8) Omit the procedures in c) if fluid fertilizers are made from only sulfuric acid and all
materials are dissolved.
(9) Heated up to 70 ºC - 80 ºC in a water bath in advance.
(10) Add drop-by-drop.
(11) Until precipitate settles.
(12) Add a barium chloride solution slightly excessively to reduce the solubility of barium
sulfate.
(13) Wash the precipitate until a white turbidity is no longer formed when about 5 mL of
346
Testing Methods for Fertilizers (2018)
nitric acid (1+2) and about 1 mL of silver nitrate solution (2 g/100 mL) are added to
about 20 mL of washing.
(14) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 800 °C in 1 to 2 hours.
(15) To prevent a crucible from chipping, it is recommended to let it stand to cool gently in
an electric furnace until the temperature of an electric furnace falls to no more than
200 ºC.
(16) Time to let it stand to cool in a desiccator should be constant. In the case of a porcelain
crucible, it is about 45 - 60 minutes.
Comment 1 Testing was conducted using sulfur simple substance fertilizers containing no
materials (2 samples); as a result, the quantitative value of the total sulfur content
(T-SO3) was 99.9 % - 100.1 % to a theoretical value.
Table 1 shows results and analysis results from a collaborative study for testing
method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.4 % (mass fraction).
347
Testing Methods for Fertilizers (2018)
References
1) JIS K 8088: Sulfur (Regent) (2010)
2) JIS K 8217: Iron ores - Methods for determination of sulfur content (1994)
3) Edited by KANTO CHEMICAL CO., INC.: Technique on Chemical Analysis of Regent
(5) Flow sheet for total sulfur content: The flow sheet for total sulfur content in fertilizers
mainly containing sulfur and sulfuric acid is shown below.
1 g -5 g
Weigh to the order of 0.1 mg into a 200-mL tall beaker
analytical sample
← About 50 mL of potassium hydroxide/ethanol solution
Heating Cover with a watch glass and boil
Standing to cool
Sample solution
348
Testing Methods for Fertilizers (2018)
Sample solution
Standing to cool
← 1 - 2 drop(s) of phenolphthalein solution (1 g/100 mL)
Neutralization Hydrochloric acid (2+1) (until the solution becomes transparent)
← About 1mL of hydrochloric acid (2+1)
← Water (the liquid volume to reach about 100 mL)
Heating Boil, for 5 minutes
← About 6 mL of thermal barium chloride solution, while shaking
Leaving at rest For several minutes
← A few drops of thermal barium chloride solution
(Check that new precipitate is no longer formed)
← About 2 mL of thermal barium chloride solution, while shaking
Heating 80 ºC - 90 ºC, 2 hour
Standing to cool For no less than 4 hours, in a water bath whose heat source is stopped
Standing to cool
349
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Hot plate: A hot plate whose surface temperature can be adjusted up to 250 ºC.
b) Water bath: Water bath that can be adjusted to 30 °C ± 2 °C and 80 ºC - 90 ºC.
c) Magnetic stirrer:
d) Spectrophotometer: A spectrophotometer specified in JIS K 0115
c) After standing to cool, transfer to a 250-mL volumetric flask and add water up to the marked
line.
d) Filter with Type 3 filter paper (4) to make the extract.
Note (2) Omit extraction if fluid fertilizers are made from only sulfuric acid and all materials
are dissolved
(3) Until sulfur content is dissolved. About 5 minutes when raw materials, etc. are not
dissolved.
(4) Omit the procedures in d) when all materials are dissolved.
Note (5) Weigh 1g - 5g of an analytical sample to the order of 0.1 mg if the fluid fertilizers of
the analytical sample are made from only sulfuric acid and all materials are dissolved.
(6) It can be suspended after the procedures end.
(7) A pH meter can be used for neutralization.
(8) Omit the procedures in c) if fluid fertilizers are made from only sulfuric acid and all
materials are dissolved.
(4.4) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 450 nm
b) Calibration curve preparation
1) Transfer 50 mL of sulfate standard solution (SO3 0.02 mg/mL - 0.1 mg/mL) to a 100-mL
Erlenmeyer flask with screw cap and conduct the procedure (4.3) b) - f) to make the SO3 1
mg/65 mL - 5 mg/65 mL sulfate standard solution for the calibration curve preparation.
351
Testing Methods for Fertilizers (2018)
2) Transfer 50 mL of water to another 100-mL Erlenmeyer flask with screw cap and conduct the
same procedures as 1) to make the blank test solution for the calibration curve preparation.
3) Measure absorbance at a wavelength of 450 nm of the sulfuric acid standard solutions for the
calibration curve preparation using the blank test solution for the calibration curve preparation
as the control (8) (9).
4) Prepare the calibration curve of the sulfuric acid concentration and absorbance of the sulfuric
acid standard solutions for the calibration curve preparation.
c) Sample measurement
1) For the sample solution, conduct procedures similarly as in b) 3) to measure absorbance.
2) For the blank test solution, conduct procedures similarly as in 1) to measure absorbance, and
correct the absorbance obtained for the sample solution.
3) Obtain the sulfate (SO3) content from the calibration curve, and calculate the total sulfur
content (T-SO3) in the analytical sample.
Note (8) Measure right after stirring because barium sulfate easily precipitates.
(9) A spectrophotometer with automatic sample introducing device is preferable.
Comment 1 The range of calibration curve with linearity is SO3 1 mg/65 mL - 5 mg/65 mL and
the curve does not pass through the origin.
Comment 2 Testing was conducted using sulfur simple substance fertilizers containing no
materials (2 samples); as a result, the quantitative value of the total sulfur content
(T-SO3) was 98.4 % - 99.4 % to a theoretical value.
Additionally, the minimum limit of quantification of this testing method is about 1 %
(mass fraction).
References
1) JIS K 8001: General rule for test methods of reagents (2009)
2) JIS K 8088: Sulfur (Regent) (2010)
3) JAPAN Sewage Works Association: Sewage Sludge Analysis -2007-, p. 132 - 134 Tokyo
(1964)
4) Edited by KANTO CHEMICAL CO., INC.: Technique on Chemical Analysis of Regent -
Practical Basic Technique and Knowledge, p. 131 - 135 (2009)
5) Yasushi SUGIMURA: Method Validation of Turbidimetry for Determination of Sulfur
Content (as Sulfur Trioxide) in Fertilizers of Sulfur, Sulfuric acid and Its Mixture,
Research Report of Fertilizer Vol.6, p. 20 - 26 (2013)
352
Testing Methods for Fertilizers (2018)
(5) Flow sheet for total sulfur content: The flow sheet for total sulfur content in fertilizers
mainly containing sulfur and sulfuric acid is shown below.
1g-2g
analytical sample Weigh to the order of 0.1 mg into a 200-mL tall beaker
← About 50 mL of potassium hydroxide/ethanol solution
Heating Cover with a watch glass and boil
Standing to cool
Filtrate
Figure 1 Flow sheet for total sulfur content in fertilizers (Extraction procedure)
Filtrate
Standing to cool
← 1 - 2 drop(s) of phenolphthalein solution (1 g/100 mL)
Neutralization Hydrochloric acid (2+1) (until the color of solution becomes transparent)
Filtrate
Figure 2 Flow sheet for total sulfur content in fertilizers (Oxidation procedure)
353
Testing Methods for Fertilizers (2018)
Filtrate
Warming 30 ºC Warming 30 ºC
← Barium chloride 0.30 g
Stirring 2 minutes
Measurement Measurement
Spectrophotometer(450 nm) Spectrophotometer(450 nm)
354
Testing Methods for Fertilizers (2018)
4.13 Iron
4.13.1 Water-soluble iron
4.13.1.a Flame atomic absorption spectrometry
(1) Summary
This test method is applicable to fertilizers that indicate iron content as a response modifier. This
testing method is classified as Type D and its symbol is 4.13.1.a-2017 or W-Fe.a-1.
Extract by adding water to an analytical sample, spray in an acetylene−air flame and measure the
atomic absorption with iron at a wavelength of 248.3 nm to obtain water-soluble iron (W-Fe) in an
analytical sample. In addition, the performance of this testing method is shown in Comment 5.
Comment 1 Instead of the iron standard solution in (2), an iron standard solution for the
calibration curve preparation can be prepared by using an iron standard solution (Fe
1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (2) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
355
Testing Methods for Fertilizers (2018)
Comment 2 In the procedure in (4.1.1) a), it is also allowed to weigh 2.50 g of the analytical
sample and transfer to a 250-mL volumetric flask.
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
Comment 4 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 248.3 nm
b) Calibration curve preparation
1) Spray the iron standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 248.3 nm.
2) Prepare a curve for the relationship between the iron concentration and the indicated value of
the iron standard solutions for the calibration curve preparation and the blank test solution for
the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.05 mg - 0.5 mg
as Fe) to a 100-mL volumetric flask.
2) Add about 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the iron content from the calibration curve, and calculate the water-soluble iron (W-Fe)
in the analytical sample.
Comment 5 Recovery testing was conducted to evaluate trueness using a preparation sample
(solid). As a result, the average rate of recovery at the content level of 10 % (mass
fraction), 5 % (mass fraction) and 0.05 % (mass fraction) are 101.1 %, 102.8 % and
107.0 % as water-soluble iron (W-Fe) respectively. In addition, recovery testing was
conducted using a preparation sample (fluid). As a result, the average rate of
recovery at the content level of 1 % (mass fraction), 0.1 % (mass fraction) and
0.01 % (mass fraction) are 103.6 %, 105.7 % and 105.1 % as a water-soluble iron
(W-Fe) respectively.
The results of the repeatability tests on different days using a fluid mixed fertilizer
and a liquid microelement mixed fertilizer to evaluate the extract precision of fluid
fertilizers were analyzed by the one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about 40
mg/kg (solid fertilizers) and 4 mg/kg (fluid fertilizers).
356
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.252, Yokendo, Tokyo (1988)
2) Shinei TAKAHASHI, Chika SUZUKI and Noriyuki SASAKI: Verification of
Performance Characteristics of Testing Method for Iron in Fertilizer by Atomic
Absorption Spectrometry, Research Report of Fertilizer, Vol. 7, p. 131 - 137 (2014)
3) Shinji KAWAGUCHI: Simple Extraction Method for Water-Soluble Components in
Liquid Compound Fertilizers, Research Report of Fertilizer, Vol.9, p. 10 - 20 (2016)
(5) Flow sheet for water-soluble iron: The flow sheet for water-soluble iron in fertilizers is
shown below:
5.00 g analytical
sample 500-mL volumetric flask.
(powdery)
← Water, About 400 mL
Shaking to mix Rotary shaker (30 - 40 revolutions/min), for 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
357
Testing Methods for Fertilizers (2018)
1.00 g analytical
sample 100-mL volumetric flask.
(Fluid)
← Water, About 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← 25 mL of hydrochloric acid (1+5)
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (248.3 nm)
358
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the iron standard solution in (2), an iron standard solution for the
calibration curve preparation can be prepared by using an iron standard solution (Fe
10 mg/mL) traceable to National Metrology.
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in e) and f) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, in case of using the
ICP-OES of an axial observation mode, it is recommended to prepare an iron
standard solution for the calibration curve in the concentration range which is
suitable to a device used.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: A spectrophotometer specified in JIS K 0115
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
b) Add about 50 mL of water, shake to mix and add water up to the marked line.
c) Filter with Type 3 filter paper to make a sample solution.
Note (2) The sampling amount of the analytical sample is 10 g when the content in the sample
is less than 0.01 % (mass fraction) as water-soluble iron.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 259.940 nm
b) Calibration curve preparation
1) Spray the iron standard solution for the calibration curve preparation and blank test solution
for the calibration curve preparation into inductively coupled plasma, and read the indicated
value at a wavelength of 259.940 nm.
2) Prepare a curve for the relationship between the iron concentration and the indicated value of
the iron standard solutions for the calibration curve preparation and the blank test solution for
the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 2 mg as
Fe) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the iron content from the calibration curve, and calculate the water-soluble iron (W-Fe)
in the analytical sample.
360
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Simultaneous Determination Method for Effect-Development
Promoting Agent (Ca, Fe, Co, Cu, Zn and Mo) in Liquid Compound Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometer (ICP-OES), Research
Report of Fertilizer, Vol.9, p. 1 - 9 (2016)
(5) Flow sheet: The flow sheet for water-soluble iron in fluid fertilizers is shown below:
1 g analytical sample Weigh to the order of 1mg into a 100-mL volumetric flask
← Water, about 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration
Sample solution
Figure1 The flow sheet for water-soluble iron in fertilizers (Extraction procedure)
Sample solution
Figure 2 The flow sheet for water-soluble iron in fertilizers (Measurement procedure)
361
Testing Methods for Fertilizers (2018)
4.14 Molybdenum
4.14.1 Water-soluble molybdenum
4.14.1.a Sodium thiocyanate absorptiometric analysis
(1) Summary
This test method is applicable to fertilizers that indicate molybdenum content as a response modifier.
This testing method is classified as Type D and its symbol is 4.14.1.a-2017 or W-Mo.a-1.
Extract by adding water to an analytical sample, add sulfuric acid (1+1) and perchloric acid, further
add a sodium thiocyanate solution and a tin (II) chloride solution, and measure the absorbance with
thiocyanate complex formed by the reaction of reduced molybdenum (V) with thiocyanate ion to
obtain water-soluble molybdenum (W-Mo) in an analytical sample. In addition, the performance of
this testing method is shown in Comment 6.
Comment 1 Instead of the molybdenum standard solution in (2), a molybdenum standard solution
for the calibration curve preparation can be prepared by using a molybdenum
standard solution (Mo 0.1 mg/mL, 1 mg/mL or 10 mg/mL) traceable to National
Metrology.
362
Testing Methods for Fertilizers (2018)
Comment 2 In the procedure in (4.1.1) a), it is also allowed to weigh 2.50 g of the analytical
sample and transfer to a 250-mL volumetric flask.
Comment 3 The procedure in (4.1.1) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
Comment 4 When the sample solution in d) contains organic matters that affect the determination,
put a predetermined amount of the sample solution to a 100-mL tall beaker, add a
small amount of sulfuric acid and nitric acid to heat and digest the organic matters
until white smoke of sulfuric acid evolves. After standing to cool, transfer the
solution to a 100-mL volumetric flask and add water up to the marked line to filter.
The filtrate is prepared as the sample solution of (4.2) a).
Comment 5 The procedure in (4.1.2) is the same as the procedure in (4.1.2) in 4.2.4.a.
Note (3) When the solution becomes muddy, centrifuge after the procedure in c). However, if it
is presumed that copper (I) thiocyanate caused the muddying, centrifuge after leaving
at rest for 1 hour.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 460 nm
b) Calibration curve preparation
1) Transfer 1 mL - 30 mL of molybdenum standard solution (Mo 0.01 mg/mL) to 100-mL
volumetric flasks step-by-step.
2) Conduct the same procedure as (4.2) b) - c) to make the 0.01 mg/100 mL - 0.3 mg/100 mL
molybdenum standard solution for the calibration curve preparation.
3) Conduct the same procedures as 2) for another 100-mL volumetric flask to make the blank
363
Testing Methods for Fertilizers (2018)
Comment 6 Recovery testing was conducted to evaluate trueness using a preparation sample. As a
result, the average rate of recovery at the content level of 2.5 % (mass fraction) and
0.1 % (mass fraction) are 100.2 % and 100.8 % as water-soluble molybdenum
(W-Mo) respectively.
The results of the repeatability tests on different days using a fluid mixed fertilizer
and a liquid microelement mixed fertilizer to evaluate the extract precision of fluid
fertilizers were analyzed by the one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about 60
mg/kg (solid fertilizers) and 6 mg/kg (fluid fertilizers).
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.281- 283, Yokendo, Tokyo (1988)
2) Keiji YAGI, Natuki TOYODOME, Tokiya SUZUKI and Hideo SOETA: Verification of
Performance Characteristics of Testing Method for Water-soluble Molybdenum Content
in Fertilizer by Sodium Thiocyante Absorptiometric Analysis, Research Report of
Fertilizer Vol. 7, p. 138 - 144 (2014)
3) Shinji KAWAGUCHI: Simple Extraction Method for Water-Soluble Components in
Liquid Compound Fertilizers, Research Report of Fertilizer, Vol.9, p. 10 - 20 (2016)
364
Testing Methods for Fertilizers (2018)
(5) Flow sheet for water-soluble molybdenum: The flow sheet for water-soluble molybdenum
in fertilizers is shown below:
5 g analytical sample
500-mL volumetric flask.
(powdery)
← About 400 mL of water
Shaking to mix Rotary shaker (30 - 40 revolutions/min), for 30 minutes
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Figure 1-1 Flow sheet for water-soluble molybdenum in fertilizers
(Extraction procedure (4.1.1))
1.00 g analytical
sample 100-mL volumetric flask.
(fluid)
← About 50 mL of water
Shaking to mix
← Water (up to the marked line)
Filtration Type 3 filter paper
Sample solution
Sample solution
Aliquot
(predetermined 100-mL volumetric flask
amount)
← About 5 mL of sulfuric acid (1+1)
← About 5 mL of perchloric acid
← About 2 mL of iron (III) sulfate solution
←16 mL of sodium thiocyanate solution (while shaking to mix)
←About 10 mL of tin (II) chloride solution (while shaking to mix)
← Water (up to the marked line)
Leaving at rest Until the red color of iron disappears
Measurement Spectrophotometer (460 nm)
365
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the molybdenum standard solution in (2), a molybdenum standard solution
for the calibration curve preparation can be prepared by using a molybdenum
standard solution (Mo 10 mg/mL) traceable to National Metrology.
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in e) and f) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, when using the ICP-OES
of an axial observation mode, it is recommended to prepare a molybdenum standard
solution for the calibration curve in the concentration range which is suitable to a
device used.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: A spectrophotometer specified in JIS K 0115
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
flask.
b) Add about 50 mL of water, shake to mix and add water up to the marked line.
c) Filter with Type 3 filter paper to make a sample solution.
Note (2) The sampling amount of the analytical sample is 10 g when the content in the sample
is less than 0.01 % (mass fraction) as water-soluble molybdenum.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 202.030 nm
b) Calibration curve preparation
1) Spray the molybdenum standard solution for the calibration curve preparation and the blank
test solution for the calibration curve preparation into inductively coupled plasma, and read
the indicated value at a wavelength of 202.030 nm.
2) Prepare a curve for the relationship between the molybdenum concentration and the indicated
value of the molybdenum standard solution for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 2 mg as
Mo) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the molybdenum content from the calibration curve, and calculate the water-soluble
molybdenum (W- Mo).
367
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Simultaneous Determination Method for Effect-Development
Promoting Agent (Ca, Fe, Co, Cu, Zn and Mo) in Liquid Compound Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometer (ICP-OES), Research
Report of Fertilizer, Vol.9, p. 1 - 9 (2016)
(5) Flow sheet: The flow sheet for water-soluble molybdenum in fluid fertilizers is shown below:
1 g analytical sample Weigh to the order of 1mg into a 100-mL volumetric flask
← Water, about 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration
Sample solution
Sample solution
368
Testing Methods for Fertilizers (2018)
4.15 Cobalt
4.15.1 Water-soluble cobalt
4.15.1.a Flame atomic absorption spectrometry
(1) Summary
This testing method is applicable to fluid mixed fertilizers, liquid microelement mixed fertilizers
and the fluid fertilizers of home garden-use mixed fertilizers. This testing method is classified as
Type E and its symbol is 4.15.1.a-2017 or W-Co.a-1.
Extract by adding water to an analytical sample, spray in an acetylene−air flame and measure the
atomic absorption with cobalt at a wavelength of 240.7 nm to obtain water-soluble cobalt (W-Co) in
an analytical sample.
Comment 1 Instead of the cobalt standard solution in (2), a cobalt standard solution for the
calibration curve preparation can be prepared by using a cobalt standard solution (Co
1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (2) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
Note (3) The sampling amount of the analytical sample is 10 g when the content in the sample
is less than 0.01 % (mass fraction) as water-soluble cobalt.
369
Testing Methods for Fertilizers (2018)
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 240.7 nm
b) Calibration curve preparation
1) Spray the cobalt standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 240.7 nm.
2) Prepare a curve for the relationship between the cobalt concentration and the indicated value
of the cobalt standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 2 mg as
Co) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the cobalt content from the calibration curve, and calculate the water-soluble cobalt
(W-Co) in the analytical sample.
(5) Flow sheet: The flow sheet for water-soluble cobalt in fluid fertilizers is shown below:
1 g analytical sample Weigh to the order of 1mg into a 100-mL volumetric flask
← Water, about 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration
Sample solution
Sample solution
Aliquot 100-mL volmetric flask
(predetermined volume)
← 25 mL of hydrochloric acid (1+5)
← Water (up to the marked line)
Measurement Atomic absorption spectrometer (240.7 nm)
370
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the cobalt standard solution in (2), a cobalt standard solution for the
calibration curve preparation can be prepared by using a cobalt standard solution (Co
10 mg/mL) traceable to National Metrology.
Comment 2 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in e) and f) applies to a horizontal
observation mode. While an axial observation mode can observe a measurement
component of low-level concentrations, it cannot gain the linearity of a calibration
curve in the range of high-level concentrations. Therefore, in case of using the
ICP-OES of an axial observation mode, it is recommended to prepare a cobalt
standard solution for the calibration curve in the concentration range which is
suitable to a device used.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) ICP Optical Emission Spectrometry: A spectrophotometer specified in JIS K 0115
1) Gas: Argon gas specified in JIS K 1105 of no less than 99.5 % (volume fraction) in purity
flask.
b) Add about 50 mL of water, shake to mix and add water up to the marked line.
c) Filter with Type 3 filter paper to make a sample solution.
Note (2) The sampling amount of the analytical sample is 10 g when the content in the sample
is less than 0.01 % (mass fraction) as water-soluble cobalt.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1.2) of 4.2.4.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 228.616 nm
b) Calibration curve preparation
1) Spray the cobalt standard solution for the calibration curve preparation and blank test solution
for the calibration curve preparation into inductively coupled plasma, and read the indicated
value at a wavelength of 228.616 nm.
2) Prepare a curve for the relationship between the cobalt concentration and the indicated value
of the cobalt standard solution for the calibration curve preparation and the blank test solution
for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 2 mg as
Co) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+5), and add water up to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the cobalt content from the calibration curve, and calculate the water-soluble cobalt
(W-Co) in the analytical sample.
372
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Simultaneous Determination Method for Effect-Development
Promoting Agent (Ca, Fe, Co, Cu, Zn and Mo) in Liquid Compound Fertilizer using
Inductively Coupled Plasma-Optical Emission Spectrometer (ICP-OES), Research
Report of Fertilizer, Vol.9, p. 1 - 9 (2016)
(5) Flow sheet: The flow sheet for water-soluble cobalt in fluid fertilizers is shown below:
1 g analytical sample Weigh to the order of 1mg into a 100-mL volumetric flask
← Water, about 50 mL
Shaking to mix
← Water (up to the marked line)
Filtration
Sample solution
Figure 1 The flow sheet for water-soluble cobalt in liquid fertilizers
(Extraction procedure)
Sample solution
373
Testing Methods for Fertilizers (2018)
5. Harmful components
5.1 Mercury
5.1.a Cold vapor atomic absorption spectrometry
(1) Summary
The test method is applicable to fertilizers excluding fluid sludge fertilizers. This testing method is
classified as Type B and its symbol is 5.1.a-2017 or Hg.a-1.
Pretreat an analytical sample with nitric acid- perchloric acid, and then reduce mercury (II) in the
solution with tin (II) chloride. Aerate this solution, and measure the atomic absorption for generated
mercury vapor at a wavelength of 253.7 nm to obtain mercury (Hg) in an analytical sample. In
addition, the performance of this testing method is shown in Comment 3.
Note (1) Use a reagent with low mercury content, such as for mercury analysis or for harmful
metal analysis.
(2) Use as an anti-forming agent.
(3) This is an example of preparation; prepare an amount as appropriate.
(4) Store in a refrigerator, and do not use after 4 months after preparation.
(5) Store in a refrigerator, and do not use after 1 months after preparation.
Comment 1 Instead of the mercury standard solution in (2), a mercury standard solution for the
calibration curve preparation can be prepared by using a mercury standard solution
(Hg 1 mg/mL or 10 mg/mL) traceable to National Metrology.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Mercury atomic absorption spectrometer: Mercury atomic absorption spectrometer using
a method for producing the atomic vapor by reduction specified in JIS K 0121.
1) Light source: Low-pressure mercury lamp
b) Hot plate or sand bath: A hot plate whose surface temperature can be adjusted up to 250 ºC.
Adjust the amounts of gas and silica sand of a sand bath so that the sand bath temperature can
374
Testing Methods for Fertilizers (2018)
Note (6) Distinguish the volumetric flask used in digestion as a sample digestion flask and do
not use it for any other purposes.
(4.2) Measurement: Conduct measurement by cold vapor atomic absorption spectrometry specified
in JIS K 0121. Specific measurement procedures are according to the operation method of the
atomic absorption spectrometer used. An example of measurement using a mercury atomic
absorption spectrometer is shown below:
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 253.7 nm
b) Calibration curve preparation
1) Transfer 1 mL - 20 mL of mercury standard solution (Hg 0.1 µg/mL) to 100-mL volumetric
flasks step-by-step, and add water up to the marked line. Transfer 5 mL of these solutions to
respective reduction vessels, add 1 drop of tri-n-butyl phosphate (9), to make mercury standard
solutions for the calibration curve preparation.
2) Add 5 mL of water to another reduction vessel, and add 1 drop of tri-n-butyl phosphate (9), to
make the blank test solution for the calibration curve preparation.
3) Connect the reduction vessel to the mercury atomic absorption spectrometer, and introduce
sulfuric acid (1+1) and a tin (II) chloride solution, and circulate air.
4) Read the indicated value at a wavelength of 253.7 nm.
5) Prepare a curve for the relationship between the mercury content (µg) and the indicated value
of the mercury standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer 5 mL of a sample solution to respective reduction vessels, add 1 drop of tri-n-butyl
phosphate (9), and conduct similarly as in b) 3) - 4) to read the indicated value.
2) Transfer 5 mL of the blank test solution to a reduction vessel, add 1 drop of tri-n-butyl
phosphate (9), and conduct similarly as in b) 2) - 4) to read the indicated value, and correct the
indicated value obtained for the sample solution.
3) Obtain the mercury content (µg) from the calibration curve, and calculate mercury (Hg) in the
375
Testing Methods for Fertilizers (2018)
analytical sample.
Note (9) It is not required to add tri-n-butyl phosphate if not needed.
Comment 2 Instead of the correction method in c) 2), mercury (Hg) can be corrected by
obtaining the mercury content in the blank test.
Comment 3 Recovery testing was conducted to evaluate trueness using an industrial sludge
fertilizer (1 sample), composted sludge (3 samples) and a human waste sludge
fertilizer (1 sample). As a result, the average rate of recovery at the content level of
2 mg/kg and 0.2 mg/kg are 98.7 % - 101.6 % and 100.7 % - 105.4 % as mercury
(Hg) respectively. In addition, recovery testing was conducted using soybean meal,
rape seed meal, compound fertilizers (2 samples) and blended fertilizers. As a result,
the average rate of recovery at the content level of 40 mg/kg and 0.5 mg/kg are
98.5 % - 101.5 % and 100.4 % - 103.3 % as mercury (Hg) respectively.
Table 1 shows results and analysis results from a collaborative study for test
method validation.
Additionally, the minimum limit of quantification of this test method is about 0.01
mg/kg.
376
Testing Methods for Fertilizers (2018)
(5) Flow sheet for mercury: The flow sheet for mercury in fertilizers is shown below:
1.00 g
Sample digestion flask
analytical sample
← About 10 mL nitric acid
Heating For a short time
Standing to cool
← Water (fill up to 100 mL)
Sample solution
377
Testing Methods for Fertilizers (2018)
Note (1) Use a reagent with low mercury content, such as for mercury analysis or for harmful
metal analysis.
(2) This is an example of preparation; prepare an amount as appropriate.
(3) Store in a refrigerator, and do not use after 4 months after preparation.
(4) Store in a refrigerator, and do not use after 1 month after preparation.
Comment 1 Instead of the mercury standard solution in (2), a mercury standard solution for the
calibration curve preparation can be prepared by using a mercury standard solution
(Hg 1 mg/mL or 10 mg/mL) traceable to National Metrology.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Mercury atomic absorption spectrometer : Mercury atomic absorption spectrometer using
a method for producing the atomic vapor by reduction specified in JIS K 0121.
1) Light source: Low-pressure mercury lamp
b) Pressure vessel decomposing device: A device which pressurizes the inside of a vessel by
putting acid, etc.to heat in the airtight vessel, and decomposes a sample by the interaction of
heating, pressurizing and acid. The following requirements should be met.
1) The main part of a decomposing device: In the case of the microwave heating method, a
device should be able to produce a high-frequency wave using a frequency which is permitted
at an industrial frequency facility. It is desirable to be able to monitor pressure and
378
Testing Methods for Fertilizers (2018)
temperature, etc. in the airtight vessel with a sensor inside the device. The interior of a device
should be acid-resistant treated and should have a high temperature durability and a high level
of safety.
2) Exhaust system: A system which has an exhaust fan of acid-resistant specification, and has
an air-cool function inside the device to allow constant airflow, keeping operating temperature
below a certain temperature level.
3) Airtight vessel: A vessel which is heat-resistance and pressure-tight and has the durability
required to decompose particles, and has resistance to internal contamination. It should have
safety functions such as causing overheat prevention valve to work and internal pressure to
drop by emitting gas, and preventing gas bumping in the case of exceeding a pressure
limitation.
c) Centrifugal separator: A centrifugal separator that can work at about 1700 × g.
Note (5) The maximum limit of solid content, which is converted from moisture content, in the
sampling volume 20.0 g of an analytical sample is about 0.5 g. If solid content is likely
to exceed the limit, reduce sampling volume as necessary.
(6) Condition examples for a microwave decomposing device: 0 min (room temperature)
→ 10min (240 ºC) → 20 min (240 ºC) → 40 min (room temperature), initial output
1400 W
(7) When organic matters still remain, for example the digestion solution is colored,
repeat the procedures in (4.1) b) - c).
(8) The vessel should be made of polypropylene, etc. to not affect the measurement.
(9) 16.5-cm of radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) in 5.2.c, 5.3.c, 5.4.c, 5.5.e
and 5.6.c.
(4.2) Measurement: Conduct measurement by cold vapor atomic absorption spectrometry specified
in JIS K 0121. Specific measurement procedures are according to the operation method of the
atomic absorption spectrometer used. An example of measurement using a mercury atomic
absorption spectrometer is shown below:
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 253.7 nm
379
Testing Methods for Fertilizers (2018)
c) Sample measurement
1) Transfer 5 mL of a sample solution to respective reduction vessels, and conduct similarly as in
b) 3) - 4) to read the indicated value.
2) Transfer 5 mL of the blank test solution to a reduction vessel and conduct similarly as in b) 2)
- 4) to read the indicated value, and correct the indicated value obtained for the sample
solution.
3) Obtain the mercury content (µg) from the calibration curve, and calculate mercury (Hg) in the
analytical sample.
Comment 3 Instead of the correction method in c) 2), mercury (Hg) can be corrected by obtaining
the mercury content in the blank test.
Comment 4 Triplicates additive recovery testing was conducted to evaluate trueness using fluid
industrial sludge fertilizers (2 samples), and composted sludge fertilizers (6 samples).
As a result, the mean recovery at the concentration level of 0.2 mg/kg - 0.4 mg/kg,
0.01 mg/kg - 0.09 mg/kg and 0.7µg /kg - 7µg /kg are 100.0 % - 109.1 %, 99.0 % -
114.6 % and 100.4 % - 113.4 % as mercury (Hg) in an actual article respectively.
The results of the repeatability tests on different days using two kinds of fluid sludge
fertilizers to evaluate precision were analyzed by the one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.。
Additionally, the minimum limit of quantification of this testing method is about 0.2
µg/kg.
380
Testing Methods for Fertilizers (2018)
References
1) Toshiharu YAGI: Determination of Heavy Metals in Fluid Sludge Fertilizers by ICP-MS
and CV-AAS, Research Report of Fertilizer Vol. 8, p. 26 - 37 (2015)
(5) Flow sheet for mercury: The flow sheet for mercury in fluid sludge fertilizers is shown
below:
Sample solution
381
Testing Methods for Fertilizers (2018)
5.2 Arsenic
5.2.a Hydride generation atomic absorption spectrometry
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type B and its
symbol is 5.2.a-2017 or As.a-1.
Pretreat an analytical sample with nitric acid-sulfuric acid-perchloric acid, and then generate arsenic
hydride by the addition of sodium tetrahydroborate in the acidic condition with hydrochloric acid,
introduce it with argon gas to a heated absorption cell, and measure the atomic absorption with
arsenic at a wavelength of 193.7 nm to obtain arsenic (As) in an analytical sample. In addition, the
performance of this testing method is shown in Comment 7.
Note (1) The concentrations of a potassium iodide solution and a sodium tetrahydroborate
solution vary depending on the instrument used.
(2) This is an example of preparation; prepare an amount as appropriate.
(3) Store in a refrigerator, and do not use after 6 months after preparation.
(4) Store in a refrigerator, and do not use after 1 month after preparation.
Comment 1 Instead of the arsenic standard solution in (2), an arsenic standard solution for the
calibration curve preparation can be prepared by using an arsenic standard solution
(As 1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (5) For cell heating, there are a method of electric heating and a method of flame heating.
Note (6) Carbonization (degradation) of organic matters by sulfuric acid begins by heating when
nitric acid no longer remains. In this state, As5+ may be reduced to As3+ and
evaporate; therefore, stop heating immediately after the end of the generation of
nitroxide (yellow-brown smoke).
(7) Oxidation of organic matters by perchloric acid progresses extremely rapidly and
explosively. For that reason, add perchloric acid after fully degrading organic matters
with nitric acid to avoid danger.
(8) When the white smoke of perchloric acid is generated, if the solution is colored such
as black-brown or brown, stop heating immediately, and after standing to cool, add
nitric acid, and heat again to degrade remaining organic matters.
(9) The watch glass can be removed.
(10) The generation of arsenic hydride is inhibited by the presence of nitric acid; therefore,
remove nitric acid by sufficiently generating the white smoke of sulfuric acid.
Comment 2 The procedure in (4.1) is the same as in (4.1) in 5.2.b and 5.5.c. However, the
sampling amount of the analytical sample in (4.1) a) in 5.5.c is 1.00 g.
383
Testing Methods for Fertilizers (2018)
Comment 3 When the analytical sample solidifies in the procedure in (4.1) b), moisten the
analytical sample with a small amount of water as necessary in advance.
Comment 4 It is not necessary to conduct the procedure in (4.1) b) “leave at rest overnight” in the
case of fertilizers not containing organic matters.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used. Two examples of measurement procedures with a continuous
hydride generator are shown below.
(4.2.1) Measurement (A): Leaving at rest after adding a potassium iodide solution.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 193.7 nm
b) Calibration curve preparation
1) Transfer 2.5 mL - 10 mL of arsenic standard solution (As 0.1 µg/mL) to 50-mL volumetric
flasks step-by-step.
2) Add 5 mL of hydrochloric acid and 5 mL of potassium iodide solution, leave at rest for about
15 minutes, and then add water up to the marked line, to make 5 ng/mL -20 ng/mL arsenic
standard solutions for the calibration curve preparation.
3) Conduct the same procedures as 2) for another 50-mL volumetric flask to make the blank test
solution for the calibration curve preparation.
4) While letting argon flow, introduce the arsenic standard solution for the calibration curve
preparation for each step and the blank test solution for the calibration curve preparation
respectively, further introduce hydrochloric acid (1+1) and a sodium tetrahydroborate solution
to the hydride generator to generate arsenic hydride.
5) Separate arsenic hydride and liquid waste, and then introduce the gas containing arsenic
hydride to a heated absorption cell, and read the indicated value at a wavelength of 193.7 nm.
6) Prepare a curve for the relationship between the arsenic concentration and the indicated value
of the arsenic standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution to a 50-mL volumetric flask, and
conduct procedures similarly as in b) 2) and b) 4) - 5) to read the indicated value.
2) Transfer a predetermined amount of the blank test solution to a 50-mL volumetric flask, and
conduct procedures similarly as in b) 2) and b) 4) - 5) to read the indicated value, and correct
the indicated value obtained for the sample solution.
3) Obtain the arsenic content from the calibration curve, and calculate arsenic (As) in the
analytical sample.
further introduce a potassium iodide solution, hydrochloric acid (1+1) and a sodium
tetrahydroborate solution to the hydride generator to generate arsenic hydride.
3) Separate arsenic hydride and liquid waste, and then introduce the gas containing arsenic
hydride to a heated absorption cell, and read the indicated value at a wavelength of 193.7 nm.
4) Prepare a curve for the relationship between the arsenic concentration and the indicated value
of the arsenic standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution to a 50-mL volumetric flask, and
conduct procedures similarly as in b) 2) - 3) to read the indicated value.
2) Transfer a predetermined amount of the blank test solution to a 50-mL volumetric flask, add
water up to the marked line, and conduct similarly as in b) 2) - 3) to read the indicated value,
and correct the indicated value obtained for the sample solution.
3) Obtain the arsenic content from the calibration curve, and calculate arsenic (As) in the
analytical sample.
Comment 5 The coexistence of iron, nickel, and cobalt at over 5, 10, 80 folds amount of arsenic,
respectively, inhibits the generation of arsenic hydride. However, the inhibition of
arsenic hydride generation even in the coexistence of iron at 1000 folds amount can
be removed by adding or introducing a potassium iodide solution.
Comment 6 Instead of the correction method in c) 2), the arsenic (As) in the analytical sample
can also be corrected by obtaining the arsenic content in the blank test solution.
Comment 7 A recovery testing was conducted using industrial sludge fertilizer, composted sludge
fertilizer (3 samples) and human waste sludge fertilizer; as a result, the recovery at
the concentration level of 50 mg/kg and 5 mg/kg were 94.6 % - 100.6 % and 99.9 %
- 103.3 % as arsenic (As), respectively. Also, a recovery testing was conducted using
processed slug phosphate fertilizer, soybean meal, rape seed meal, compound
fertilizer and potassium-magnesium-sulfate fertilizer; as a result, the recovery at the
concentration level of 50 mg/kg and 5 mg/kg were 98.5 % - 109.8 % and 103.5 % -
108.6 %, respectively.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of this test method is about 0.1
mg/kg.
385
Testing Methods for Fertilizers (2018)
References
1) Naoki ASAO, Yukie ISHIDA, Shinjiro IZUKA and Masakazu SAIKI: Determination of
Arsenic in Sludge Fertilizer - Improved Decomposition Method - Research Report of
Fertilizer Vol.1, p. 74 - 81 (2008)
2) Naoki ASAO, Shinjiro IZUKA and Norio HIKICHI: Determination of Arsenic in
Sludge Fertilizer - Collaborative Test Results - Research Report of Fertilizer Vol.1, p. 82
- 89 (2008)
3) Yasushi SUGIMURA, Naoki ASAO and Shinjiro IZUKA: Determination of Arsenic in
Fertilizer- Expanding the scope of application of improved method -, Research Report
of Fertilizer Vol.2, p. 18 - 24 (2009)
386
Testing Methods for Fertilizers (2018)
(5) Flow sheet for arsenic: The flow sheet for arsenic in fertilizers is shown below:
1.00 g - 2.00 g
200-mL - 300-mL tall beaker
analytical sample
← A small amount of water to moisten analytical sample
(if needed).
← About 10 mL nitric acid
← About 5 mL sulfuric acid
Leaving at rest Cover with a watch glass
overnight
Sample solution
387
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the arsenic standard solution in (2), an arsenic standard solution for the
calibration curve preparation can be prepared by using an arsenic standard solution
(As 1 mg/mL or 10 mg/mL) traceable to National Metrology.
c) Hot plate or sand bath: A hot plate whose surface temperature can be adjusted up to 350 ºC.
Adjust the amounts of gas and silica sand of a sand bath so that the sand bath temperature can
be set to no less than 300 °C.
Note (3) Carbonization (degradation) of organic matters by sulfuric acid begins by heating
when nitric acid no longer remains. In this state, As+5 may be reduced to As+3 and
evaporate; therefore stop heating immediately after the end of the generation of
nitroxide (yellow-brown smoke).
(4) Oxidation of organic matters by perchloric acid progresses extremely rapidly and
explosively. For that reason, add perchloric acid after fully degrading organic matters
with nitric acid to avoid danger.
(5) When the white smoke of perchloric acid is generated, if the solution is colored such
as black-brown or brown, stop heating immediately, and after standing to cool, add
nitric acid, and heat again to degrade remaining organic matters.
(6) The watch glass can be removed.
(7) The generation of arsenic hydride is inhibited by the presence of nitric acid; therefore,
remove nitric acid by sufficiently generating the white smoke of sulfuric acid.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 5.2.a.
Comment 3 When the analytical sample solidifies in the procedure in (4.1) b), moisten the
analytical sample with a small amount of water as necessary in advance.
Comment 4 It is not necessary to conduct the procedure in (4.1) b) “leave at rest overnight” in
the case of fertilizers not containing organic matters.
e) Add about 1 mL of tin (II) chloride solution, shake and leave at rest for about 10 minutes (8).
f) Connect arsenic hydride generation bottle, glass tube lightly stuffed with lead acetate glass
wool in advance and 5 mL of silver diethyl dithiocarbamate solution (9), and quickly put 2.5g
of zinc into an arsenic hydride generation bottle.
g) Leave at rest at room temperature (15ºC - 25ºC) for about 45 minutes, and allow generated
arsenic hydride to absorb into the silver diethyl dithiocarbamate solution and color.
h) Transfer a predetermined amount of blank test solution into the arsenic hydride generation
bottle, conduct procedures similarly as in b) - g) to allow generated arsenic hydride to absorb
into the silver diethyl dithiocarbamate solution and color.
Note (8) When a large amount of iron is contained, add 1 g of ascorbic acid and 2mL of tin (II)
chloride solution and shake to mix to leave at rest for about 10 minutes instead of the
procedure in e).
(9) Apply a small amount of silicone grease, etc. to the connecting parts of an arsenic
hydride generation bottle, glass tube and arsenic hydride absorption tube to maintain
their airtightness.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used.
a) Measurement conditions of spectrophotometer: Set up the measurement conditions of
spectrophotometer considering the following.
Detection wavelength: 510 nm or 519 nm
b) Calibration curve preparation
1) Transfer 2.5 mL - 20 mL of arsenic standard solution (1 µg/mL) to arsenic hydride generation
bottles step-by-step.
2) Conduct procedures similarly as in (4.2) b) - g) and allow them to react.
3) For another arsenic hydride generation bottle, the silver diethyl dithiocarbamate solution
prepared similarly as in the procedure in 2) is used as the blank test solution for the calibration
curve preparation.
4) Measure the absorbance at wavelength 510 nm of the silver diethyl dithiocarbamate solution
of the arsenic standard solution for the calibration curve preparation using the blank test
solution for the calibration curve preparation as the control.
5) Prepare a curve for the relationship between the arsenic concentration and the indicated value
of the arsenic standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) For the silver diethyl dithiocarbamate solution of (4.2) g), conduct procedures similarly as in
b) 4) to measure absorbance.
2) For the silver diethyl dithiocarbamate solution of (4.2) h), conduct procedures similarly as in
b) 4) to measure absorbance, and correct the absorbance obtained for the sample solution.
3) Obtain the arsenic content from the calibration curve, and calculate arsenic (As) in the
analytical sample.
Comment 5 Instead of the correction method in c) 2), the arsenic (As) in the analytical sample
can also be corrected by obtaining the arsenic content in the blank test solution.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.270- 273, Yokendo, Tokyo (1988)
390
Testing Methods for Fertilizers (2018)
(5) Flow sheet for arsenic: The flow sheet for arsenic in fertilizers is shown below:
1.00 g -2.00 g
200-mL - 300-mL tall beaker
analytical sample
← A small amount of water to moisten analytical sample (if needed).
← About 10 mL nitric acid, about 5 mL sulfuric acid
Leaving at rest Cover with a watch glass
overnight
Sample solution
391
Testing Methods for Fertilizers (2018)
Sample solution
392
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the arsenic standard solution in (2), an arsenic standard solution for the
calibration curve preparation can be prepared by using an arsenic standard solution
(As 1 mg/mL or 10 mg/mL) traceable to National Metrology.
393
Testing Methods for Fertilizers (2018)
Note (6) The maximum limit of solid content, which is converted from moisture content, in the
sampling volume 20.0 g of an analytical sample is about 0.5g. If solid content is likely
to exceed the limit, reduce sampling volume as necessary.
(7) Condition examples for a microwave decomposing device: 0 min (room temperature)
→ 10min (240 ºC) → 20 min (240 ºC) → 40 min (room temperature), initial output
1400 W
(8) When organic matters still remain, for example the digestion solution is colored,
repeat the procedures in (4.1) b) - d).
(9) The vessel should be made of polypropylene, etc. to not affect the measurement.
(10) 16.5-cm of radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
394
Testing Methods for Fertilizers (2018)
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 5.1.b.
(4.2) Measurement: Conduct measurement (Internal Standard Method) according to JIS K 0113
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Mass Spectrometer used in measurement.
a) Measurement conditions for the ICP Mass Spectrometer: Set up the measurement
conditions for the ICP Mass Spectrometer considering the following:
Arsenic: monitor ion (m/z): 75
Rhodium: monitor ion (m/z): 103
b) Calibration curve preparation
1) Spray arsenic standard solution for the calibration curve preparation and blank test solution
for the calibration curve preparation into the inductively coupled plasma, and read the ratio of
the respective ion count number for the monitor ion of an element subjected to measurement
and an internal standard element.
2) Prepare a curve for the relationship between the concentration and the ratio of the ion count
number of an element subjected to measurement.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.05 µg - 5 µg as
As) to a 50-mL volumetric flask (9).
2) Add 5mL of internal standard solution (5) and add nitric acid (1+19) to the marked line.
3) Conduct procedures similarly as in b) 1) to read the ratio of the ion count numbers.
4) For blank test solution, conduct procedures similarly as in 1) - 3) to correct the ratio of the ion
count numbers obtained for the sample solution.
5) Obtain the arsenic content from the calibration curve, and calculate arsenic (As) in the
analytical sample.
Comment 3 Instead of the correction method in c) 4), the arsenic (As) in the analytical sample
can also be corrected by obtaining the arsenic content in the blank test solution.
Comment 4 Triplicates additive recovery testing was conducted to evaluate trueness using fluid
industrial sludge fertilizers (2 samples) and composted sludge fertilizers (6 samples).
As a result, the mean recovery at the concentration level of 1 mg/kg - 9 mg/kg, 0.1
mg/kg - 0.9 mg/kg and 0.02 mg /kg - 0.04 mg /kg are 85.0 % - 105.9 %, 90.6 % -
108.5 % and 95 % as arsenic (As) in an actual article respectively.
The results of the repeatability tests on different days using two kinds of fluid sludge
fertilizers to evaluate precision were analyzed by the one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.。
Additionally, the minimum limit of quantification of this test method is about 3 µg
/kg.
395
Testing Methods for Fertilizers (2018)
396
Testing Methods for Fertilizers (2018)
Comment 6 The presence and the extent of interference to the measurement mass number of an
element subjected to measurement and an internal standard element can be estimated
by qualitative analysis using an ICP-MS in advance to quantitation. Select a
measurement mass number considering the extent of interference. However, a mass
number cannot be changed in the measurement of arsenic. A magnetic sector
double-focusing mass spectrometer or a collision reaction cell specified in JIS K0133
can be used as a method to reduce spectrum interference.
References
1) Toshiharu YAGI: Determination of Heavy Metals in Fluid Sludge Fertilizers by ICP-MS
and CV-AAS, Research Report of Fertilizer Vol. 8, p. 26 - 37 (2015)
(5) Flow sheet for arsenic: The flow sheet for arsenic in fluid sludge fertilizers is shown below:
Sample solution
Sample solution
397
Testing Methods for Fertilizers (2018)
5.3 Cadmium
5.3.a Flame atomic absorption spectrometry
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type B and its
symbol is 5.3.a-2017 or Cd.a-1.
Pretreat an analytical sample with incineration and nitric acid-hydrochloric acid (1+3), spray into an
acetylene-air flame, and measure the atomic absorption with cadmium at a wavelength of 228.8 nm
to obtain cadmium (Cd) in an analytical sample. In addition, the performance of this testing method
is shown in Comment 5.
Comment 1 Instead of the cadmium standard solution in (2), a cadmium standard solution for the
calibration curve preparation can be prepared by using a cadmium standard solution
(Cd 1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (3) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
Note (4) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(5) The watch glass can be removed.
(6) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 228.8 nm
b) Calibration curve preparation
1) Spray the cadmium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 228.8 nm.
2) Prepare a curve for the relationship between the cadmium concentration and the indicated
value of the cadmium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Subject the sample solution (7) to the same procedure as in b) 1) to read the indicated value.
2) Subject the blank test solution to the same procedure as in b) 1) to read the indicated value,
and correct the indicated value obtained for the sample solution.
3) Obtain the cadmium content from the calibration curve, and calculate cadmium (Cd) in the
analytical sample.
Note (7) If there is a possibility that the cadmium concentration in the sample solution will
399
Testing Methods for Fertilizers (2018)
exceed the maximum limit of the calibration curve, dilute a predetermined amount
with hydrochloric acid (1+23).
Comment 4 Instead of the correction method in c) 2), the cadmium (Cd) in the analytical sample
can also be corrected by obtaining the cadmium content in the blank test solution.
Comment 5 Recovery testing was conducted using industrial sludge fertilizer and composted
sludge fertilizer (5 samples); as a result, the recovery at the concentration level of 5
mg/kg and 0.5 mg/kg were 97.5 % - 99.2 % and 96.7 % - 99.7 %, respectively.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of this test method is about 0.1
mg/kg.
References
1) Yoshinari SAKAKIBARA, Manabu MATSUZAKI and Tadao AMANO: Determination
of Cadmium, Lead, Nickel and Chromium in Sludge Fertilizer - Improved
Decomposition Method - Research Report of Fertilizer Vol.1, p. 41 - 49 (2008)
2) Yoshinari SAKAKIBARA and Manabu MATSUZAKI: Determination of Cadmium,
Lead, Nickel and Chromium in Sludge Fertilizer - Collaborative Test Results - Research
Report of Fertilizer Vol.1, p. 50 - 59 (2008)
3) Hisanori ARAYA and Yoshimi TAKEBA: Validation of atomic absorption spectrometry
for determination of cadmium, lead, nickel and chromium in calcined sludge fertilizer -
Research Report of Fertilizer Vol.3, p. 30 - 42 (2010)
400
Testing Methods for Fertilizers (2018)
(5) Flow sheet for cadmium: The flow sheet for cadmium in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
401
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the cadmium standard solution in (2), a cadmium standard solution for the
calibration curve preparation can be prepared by using a cadmium standard solution
(Cd 1 mg/mL or 10 mg/mL) traceable to National Metrology.
i) As a blank test, conduct the procedures in b) - h) using another tall beaker to prepare the
blank test solution.
Note (3) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(4) The watch glass can be removed.
(5) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct measurement (Standard Addition Method) according to JIS K 0116
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Optical Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 228.802 nm
b) Calibration curve preparation and sample measurement
1) Put 5mL of sample solution to three 10-mL volumetric flasks respectively.
2) Add 2mL and 4 mL of cadmium standard solution (0.25 μg/mL) to volumetric flasks of 1)
above, and further add hydrochloric acid (1+23) to the marked line to make a sample solution
of Standard Addition Method.
3) Add hydrochloric acid (1+23) to the marked line of the remaining volumetric flask of 1)
above to make a sample solution without a standard solution.
4) Spray the sample solution of Standard Addition Method and the sample solution without a
standard solution into the induction plasma, and read the indicated value at a wavelength of
228.802 nm.
5) Transfer 5 mL of blank test solution to a 10-mL volumetric flask, conduct the same
procedures as in 3) - 4) to read the indicated value, and correct the indicated value obtained
from the respective sample solutions.
6) Prepare a curve for the relationship between the added cadmium concentration and the
corrected indicated value of the sample solution for Standard Addition Method and sample
solution without a standard solution.
7) Obtain the cadmium content from the intercept of the calibration curve to calculate cadmium
(Cd) in the analytical sample
Comment 4 Instead of the correction method in c) 5), the cadmium (Cd) in the analytical sample
can also be corrected by obtaining the cadmium content in the blank test solution.
Comment 5 The simultaneous measurement of multiple elements by an ICP-OES is available. In
that case, see 4.9.1.b Comment 5.
Comment 6 The comparison of the measurement value (xi:0.003 mg/kg - 3.32 mg/kg) of ICP
Optical Emission Spectrometry and the measurement value (yi) of Flame Atomic
Absorption Spectrometry was conducted to evaluate trueness using sludge fertilizers
(49 samples). As a result, a regression equation was y = − 0.03 + 1.009x and its
403
Testing Methods for Fertilizers (2018)
correlation coefficient (r) was 0.996. Triplicates measurement for each one sample of
sewage sludge fertilizer, human waste sludge fertilizer, industrial sludge fertilizer,
mixed sludge fertilizer, calcined sludge fertilizers and composted sludge fertilizer
was conducted. As a result, a repeatability obtained was 0.8 % - 4.1 % as a relative
standard deviation.
Additionally, the minimum limit of quantification of this testing method is 0.2
mg/kg.
References
1) Masahiro ECHI, Tomoe INOUE, Megumi TABUCHI and Tetuya NOMURA:
Simultaneous Determination of Cadmium, Lead, Nickel, Chromium, Copper and Zinc
in Sludge Fertilizer using Inductively Coupled Plasma-Atomic Emission Spectrometry
(ICP-AES), Research Report of Fertilizer, Vol.4, p. 30 - 35 (2011)
(5) Flow sheet for cadmium: The flow sheet for cadmium in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
404
Testing Methods for Fertilizers (2018)
Sample solution
405
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the cadmium standard solution in (2), a cadmium standard solution for the
calibration curve preparation can be prepared by using a cadmium standard solution
406
Testing Methods for Fertilizers (2018)
Note (6) The maximum limit of solid content, which is converted from moisture content, in the
sampling volume 20.0 g of an analytical sample is about 0.5 g. If solid content is likely
to exceed the limit, reduce sampling volume as necessary.
(7) Condition examples for a microwave decomposing device: 0 min (room temperature)
→ 10min (240 ºC) → 20 min (240 ºC) → 40 min (room temperature), initial output
1400 W
(8) When organic matters still remain, for example the digestion solution is colored,
repeat the procedures in (4.1) b) - d).
(9) The vessel should be made of polypropylene, etc. to not affect the measurement.
(10) 16.5-cm of radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
407
Testing Methods for Fertilizers (2018)
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 5.1.b.
(4.2) Measurement: Conduct measurement (Internal Standard Method) according to JIS K 0113
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Mass Spectrometer used in measurement.
a) Measurement conditions for the ICP Mass Spectrometer: Set up the measurement
conditions for the ICP Mass Spectrometer considering the following:
Cadmium: monitor ion (m/z): 111, 114
Rhodium: monitor ion (m/z): 103
b) Calibration curve preparation
1) Spray cadmium standard solution for the calibration curve preparation and blank test solution
for the calibration curve preparation into the inductively coupled plasma, and read the ratio of
the respective ion count numbers for the monitor ion of an element subjected to measurement
and an internal standard element.
2) Prepare a curve for the relationship between the concentration and the ratio of the ion count
number of an element subjected to measurement.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.005 µg - 0.5 µg
as Cd) to a 50-mL volumetric flask (9).
2) Add 5mL of internal standard solution (5) and add nitric acid (1+19) to the marked line.
3) Conduct procedures similarly as in b) 1) to read the ratio of the ion count numbers.
4) For blank test solution, conduct procedures similarly as in 1) - 3) to correct the ratio of the ion
count number obtained for the sample solution.
5) Obtain the cadmium content from the calibration curve, and calculate cadmium (Cd) in the
analytical sample.
Comment 3 Instead of the correction method in c) 4), the cadmium (Cd) in the analytical sample
can also be corrected by obtaining the cadmium content in the blank test solution.
Comment 4 Triplicates additive recovery testing was conducted to evaluate trueness using fluid
industrial sludge fertilizers (2 samples), and composted sludge fertilizers (6 samples).
As a result, the mean recovery at the concentration level of 0.1 mg/kg - 0.9 mg/kg,
0.01 mg/kg - 0.09 mg/kg and 2 µg /kg - 4 µg /kg are 89.4 % - 108.5 %, 91.0 % -
112.0 % and 96.3 % - 108.5 % as cadmium (Cd) in an actual article respectively.
The results of the repeatability tests on different days using two kinds of fluid sludge
fertilizers to evaluate precision were analyzed by the one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this test method is about 2 µg
/kg.
408
Testing Methods for Fertilizers (2018)
References
1) Toshiharu YAGI: Determination of Heavy Metals in Fluid Sludge Fertilizers by ICP-MS
and CV-AAS, Research Report of Fertilizer Vol. 8, p. 26 - 37 (2015)
(5) Flow sheet for cadmium: The flow sheet for cadmium in fluid sludge fertilizers is shown
below:
Sample solution
409
Testing Methods for Fertilizers (2018)
Sample solution
410
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the cadmium standard solution in (2), a cadmium standard solution for the
calibration curve preparation can be prepared by using a cadmium standard solution
(Cd 1 mg/mL or 10 mg/mL) traceable to National Metrology.
putting acid, etc.to heat in the airtight vessel, and decomposes a sample by the interaction of
heating, pressurizing and acid. The following requirements should be met.
1) The main part of a decomposing device: In the case of the microwave heating method, a
device should be able to produce a high-frequency wave using a frequency which is permitted
at an industrial frequency facility. It is desirable to be able to monitor pressure and
temperature, etc. in the airtight vessel with a sensor inside the device. The interior of a device
should be acid-resistant treated and should have a high temperature durability and a high level
of safety.
2) Exhaust system: A system which has an exhaust fan of acid-resistant specification, and has
an air-cool function inside the device to allow constant airflow, keeping operating temperature
below a certain temperature level.
3) Airtight vessel: A vessel which is heat-resistance and pressure-tight and has the durability
required to decompose particles, and has resistance to internal contamination. It should have
safety functions such as causing overheat prevention valve to work and internal pressure to
drop by emitting gas, and preventing gas bumping in the case of exceeding a pressure
limitation.
c) Centrifugal separator: A centrifugal separator that can work at about 1700 × g.
(4.2) Measurement: Conduct measurement (Internal Standard Method) according to JIS K 0113
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Mass Spectrometer used in measurement.
a) Measurement conditions for the ICP Mass Spectrometer: Set up the measurement
conditions for the ICP Mass Spectrometer considering the following:
Cadmium: monitor ion (m/z): 111
Rhodium: monitor ion (m/z): 103
412
Testing Methods for Fertilizers (2018)
Comment 2 Instead of the correction method in c) 4), the cadmium (Cd) in the analytical sample
can also be corrected by obtaining the cadmium content in the blank test solution.
Comment 3 The comparison of the measurement value (xi:0.18 mg/kg - 3.02 mg/kg) of ICP-MS
and the measurement value (yi) of Flame Atomic Absorption Spectrometry was
conducted to evaluate trueness using sludge fertilizers (26 samples). As a result, a
regression equation was y = 0.0402 + 1.01x and its correlation coefficient (r) was
0.997.
The results of the repeatability tests on different days using two kinds of sludge
fertilizers to evaluate precision were analyzed by the one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this test method is about 0.1
mg/kg.
Comment 4 The presence and the extent of interference to the measurement mass number of an
element subjected to measurement and an internal standard element can be estimated
by qualitative analysis using an ICP-MS in advance to quantitation. Select a
measurement mass number considering the extent of interference. A magnetic sector
double-focusing mass spectrometer or a collision reaction cell specified in JIS K0133
413
Testing Methods for Fertilizers (2018)
(5) Flow sheet for cadmium: The flow sheet for cadmium in splid sludge fertilizers is shown
below:
Sample solution
Sample solution
414
Testing Methods for Fertilizers (2018)
5.4 Nickel
5.4.a Flame atomic absorption spectrometry
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type B and its
symbol is 5.4.a-2017 or Ni.a-1.
Pretreat an analytical sample with incineration and nitric acid−hydrochloric acid (1+3), spray into
an acetylene−air flame, and measure the atomic absorption with nickel at a wavelength of 232.0 nm
to obtain nickel (Ni) in an analytical sample. In addition, the performance of this testing method is
shown in Comment 5.
Comment 1 Instead of the nickel standard solution in (2), a nickel standard solution for the
calibration curve preparation can be prepared by using a nickel standard solution (Ni
1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (3) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
Note (4) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(5) The watch glass can be removed.
(6) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 232.0 nm
b) Calibration curve preparation
1) Spray the nickel standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 232.0 nm.
2) Prepare a curve for the relationship between the nickel concentration and the indicated value
of the nickel standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Subject the sample solution (7) to the same procedure as in b) 1) to read the indicated value.
2) Subject the blank test solution to the same procedure as in b) 1) to read the indicated value,
and correct the indicated value obtained for the sample solution.
3) Obtain the nickel content from the calibration curve, and calculate nickel (Ni) in the analytical
sample.
Note (7) If there is a possibility that the nickel concentration in the sample solution will exceed
the maximum limit of the calibration curve, dilute a predetermined amount with
hydrochloric acid (1+23).
416
Testing Methods for Fertilizers (2018)
Comment 4 Instead of the correction method in c) 2), the nickel (Ni) in the analytical sample can
also be corrected by obtaining the nickel content in the blank test solution.
Comment 5 Recovery testing was conducted using industrial sludge fertilizer and composted
sludge fertilizer (5 samples); as a result, the recovery at the concentration level of
300 mg/kg and 30 mg/kg were 98.5 % - 100.3 % and 97.1 % - 99.9 %, respectively.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of this test method is about 1
mg/kg.
References
1) Yoshinari SAKAKIBARA, Manabu MATSUZAKI and Tadao AMANO: Determination
of Cadmium, Lead, Nickel and Chromium in Sludge Fertilizer - Improved
Decomposition Method - Research Report of Fertilizer Vol.1, p. 41 - 49 (2008)
2) Yoshinari SAKAKIBARA and Manabu MATSUZAKI: Determination of Cadmium,
Lead, Nickel and Chromium in Sludge Fertilizer - Collaborative Test Results - Research
Report of Fertilizer Vol.1, p. 50 - 59 (2008)
3) Hisanori ARAYA and Yoshimi TAKEBA: Validation of atomic absorption spectrometry
for determination of cadmium, lead, nickel and chromium in calcined sludge fertilizer -
Research Report of Fertilizer Vol.3, p. 30 - 42 (2010)
417
Testing Methods for Fertilizers (2018)
(5) Flow sheet for nickel: The flow sheet for nickel in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
418
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the nickel standard solution in (2), a nickel standard solution for the
calibration curve preparation can be prepared by using a nickel standard solution (Ni
1 mg/mL or 10 mg/mL) traceable to National Metrology.
i) As a blank test, conduct the procedures in b) - h) using another tall beaker to prepare the
blank test solution.
Note (3) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(4) The watch glass can be removed.
(5) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 1 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct measurement (Standard Addition Method) according to JIS K 0116
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Optical Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 231.604 nm
b) Calibration curve preparation and sample measurement
1) Put 5mL of sample solution to three 10-mL volumetric flasks respectively.
2) Add 2 mL and 4 mL (0.25 μg/mL) of nickel standard solution to volumetric flasks of 1) above,
then add hydrochloric acid (1+23) to the marked line to make a sample solution of Standard
Addition Method.
3) Add hydrochloric acid (1+23) to the marked line of the remaining volumetric flask of 1)
above to make a sample solution without a standard solution.
4) Spray the sample solution of Standard Addition Method and the sample solution without a
standard solution into the induction plasma, and read the indicated value at a wavelength of
231.604 nm.
5) Transfer 5 mL of blank test solution to a 10-mL volumetric flask, conduct the same
procedures as in 3) - 4) to read the indicated value, and correct the indicated value obtained
from the respective sample solutions.
6) Prepare a curve for the relationship between the added nickel concentration and the corrected
indicated value of the sample solution for Standard Addition Method and the sample solution
without a standard solution.
7) Obtain the nickel content from the intercept of the calibration curve to calculate nickel (Ni)
concentration in the analytical sample.
Comment 3 Instead of the correction method in c) 5), the nickel (Ni) in the analytical sample can
also be corrected by obtaining the nickel content in the blank test solution.
Comment 4 The simultaneous measurement of multiple elements by an ICP-OES is available. In
that case, see 4.9.1.b Comment 5.
Comment 5 The comparison of the measurement value (xi:8.4 mg/kg - 129 mg/kg) of ICP
Optical Emission Spectrometry and the measurement value (yi) of Flame Atomic
Absorption Spectrometry was conducted to evaluate trueness using sludge fertilizers
(49 samples). As a result, a regression equation was y = − 0.96 + 1.010x and its
420
Testing Methods for Fertilizers (2018)
correlation coefficient (r) was 0.995. Triplicates measurement for each one sample of
sewage sludge fertilizer, human waste sludge fertilizer, industrial sludge fertilizer,
mixed sludge fertilizer, calcined sludge fertilizers and composted sludge fertilizer
was conducted. As a result, a repeatability obtained was1.0 % - 2.6 % as a relative
standard deviation.
Additionally, the minimum limit of quantification of this testing method is about 8
mg/kg.
References
1) Masahiro ECHI, Tomoe INOUE, Megumi TABUCHI and Tetuya NOMURA:
Simultaneous Determination of Cadmium, Lead, Nickel, Chromium, Copper and Zinc
in Sludge Fertilizer using Inductively Coupled Plasma-Atomic Emission Spectrometry
(ICP-AES), Research Report of Fertilizer, Vol.4, p. 30 - 35 (2011)
(5) Flow sheet for nickel: The flow sheet for nickel in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
421
Testing Methods for Fertilizers (2018)
Sample solution
422
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the nickel standard solution in (2), a nickel standard solution for the
calibration curve preparation can be prepared by using a nickel standard solution (Ni
1 mg/mL or 10 mg/mL) traceable to National Metrology.
423
Testing Methods for Fertilizers (2018)
Note (6) The maximum limit of solid content, which is converted from moisture content, in the
sampling volume 20.0 g of an analytical sample is about 0.5 g. If solid content is likely
to exceed the limit, reduce sampling volume as necessary.
(7) Condition examples for a microwave decomposing device: 0 min (room temperature)
→ 10min (240 ºC) → 20 min (240 ºC) → 40 min (room temperature), initial output
1400 W
(8) When organic matters still remain, for example the digestion solution is colored,
repeat the procedures in (4.1) b) - d).
(9) The vessel should be made of polypropylene, etc. to not affect the measurement.
(10) 16.5-cm of radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
424
Testing Methods for Fertilizers (2018)
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 5.1.b.
(4.2) Measurement: Conduct measurement (Internal Standard Method) according to JIS K 0113
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Mass Spectrometer used in measurement.
a) Measurement conditions for the ICP Mass Spectrometer: Set up the measurement
conditions for the ICP Mass Spectrometer considering the following:
Nickel: monitor ion (m/z): 60, 58
Rhodium: monitor ion (m/z): 103
b) Calibration curve preparation
1) Spray nickel standard solution for the calibration curve preparation and blank test solution for
the calibration curve preparation into the inductively coupled plasma, and read the ratio of the
respective ion count numbers for the monitor ion of an element subjected to measurement and
an internal standard element.
2) Prepare a curve for the relationship between the concentration and the ratio of the ion count
number of an element subjected to measurement.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.05 µg - 5 µg as
Ni) to a 50-mL volumetric flask (9).
2) Add 5mL of internal standard solution (5) and add nitric acid (1+19) to the marked line.
3) Conduct procedures similarly as in b) 1) to read the ratio of the ion count number.
4) For blank test solution, conduct procedures similarly as in 1) - 3) to correct the ratio of the ion
count number obtained for the sample solution.
5) Obtain the nickel content from the calibration curve, and calculate nickel (Ni) in the analytical
sample.
Comment 3 Instead of the correction method in c) 4), the nickel (Ni) in the analytical sample can
also be corrected by obtaining the nickel content in the blank test solution.
Comment 4 Triplicates additive recovery testing was conducted to evaluate trueness using fluid
industrial sludge fertilizers (2 samples) and composted sludge fertilizers (6 samples).
As a result, the mean recovery at the concentration level of 10 mg/kg - 60 mg/kg, 1
mg/kg - 9 mg/kg and 0.1 mg /kg - 0.9 mg /kg are 89.6 % - 99.2 %, 91.5 % - 114.7 %
and 96.1 - 103.7 % as nickel (Ni) in an actual article respectively.
The results of the repeatability tests on different days using two kinds of fluid sludge
fertilizers to evaluate precision were analyzed by the one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.。
Additionally, the minimum limit of quantification of this test method is about 6 µg
/kg.
425
Testing Methods for Fertilizers (2018)
References
1) Toshiharu YAGI: Determination of Heavy Metals in Fluid Sludge Fertilizers by ICP-MS
and CV-AAS, Research Report of Fertilizer Vol. 8, p. 26 - 37 (2015)
(5) Flow sheet for nickel: The flow sheet for nickel in fluid sludge fertilizers is shown below:
Sample solution
426
Testing Methods for Fertilizers (2018)
Sample solution
427
Testing Methods for Fertilizers (2018)
5.5 Chromium
5.5.a Flame atomic absorption spectrometry (Fertilizers containing organic matters)
(1) Summary
This testing method is applicable to fertilizers containing organic matters. This testing method is
classified as Type B and its symbol is 5.5.a-2017 or Cr.a-1.
Pretreat an analytical sample with incineration and nitric acid−hydrochloric acid (1+3), spray into
an acetylene−air flame, and measure the atomic absorption with chromium at a wavelength of 357.9
nm or 359.3 nm to obtain chromium (Cr) in an analytical sample. In addition, the performance of
this testing method is shown in Comment 5.
Comment 1 Instead of the chromium standard solution in (2), a chromium standard solution for
the calibration curve preparation can be prepared by using a chromium standard
solution (Cr 1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (4) There are the continuous source method, the Zeeman method, the non-resonance
428
Testing Methods for Fertilizers (2018)
Note (5) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(6) The watch glass can be removed.
(7) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 357.9 nm or 359.3 nm (8)
b) Calibration curve preparation
1) Spray the chromium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame (9), and read the indicated value at a
wavelength of 357.9 nm or 359.3 nm (8).
2) Prepare a curve for the relationship between the chromium concentration and the indicated
value of the chromium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer 25 mL of the sample solution (10) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add hydrochloric acid (1+23) to
the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Subject the blank test solution to the same procedure as in 1) - 2) and b) 1) to read the
indicated value, and correct the indicated value obtained for the sample solution.
429
Testing Methods for Fertilizers (2018)
5) Obtain the chromium content from the calibration curve, and calculate chromium (Cr) in the
analytical sample.
Note (8) When background correction is conducted by the Zeeman method, 359.3 nm is
recommended as the analytical line wavelength.
(9) Use low-fuel acetylene−air flame. Acetylene−nitrous oxide flame can also be used.
(10) If there is a possibility that the chromium concentration in the sample solution will
exceed the maximum limit of the calibration curve, decrease the amount to be
transferred.
References
1) Yoshinari SAKAKIBARA, Manabu MATSUZAKI and Tadao AMANO: Determination
of Cadmium, Lead, Nickel and Chromium in Sludge Fertilizer - Improved
Decomposition Method - Research Report of Fertilizer Vol.1, p. 41 - 49 (2008)
2) Yoshinari SAKAKIBARA and Manabu MATSUZAKI: Determination of Cadmium,
430
Testing Methods for Fertilizers (2018)
Lead, Nickel and Chromium in Sludge Fertilizer - Collaborative Test Results - Research
Report of Fertilizer Vol.1, p. 50 - 59 (2008)
3) Yoshinari SAKAKIBARA, Chie INOUE: Validation of determination methods for
chromium in sludge fertilizer - Evaluation of measurement procedure, Research Report
of Fertilizer Vol.2, p. 130 - 136 (2009)
(5) Flow sheet for chromium: The flow sheet for chromium in fertilizers containing organic
matters is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
431
Testing Methods for Fertilizers (2018)
5.5.b Flame atomic absorption spectrometry (Fertilizers mainly containing molten matters,
slag, etc.)
(1) Summary
This testing method is applicable to fertilizers mainly containing molten matters, slag, etc. This
testing method is classified as Type B and its symbol is 5.5.b-2017 or Cr.b-1.
Add ammonium sulfate to prevent the bumping of an analytical sample, pretreat the analytical
sample with nitric acid−sulfuric acid−perchloric acid, and then spray into an acetylene−air flame.
Measure the atomic absorption with chromium at a wavelength of 357.9 nm or 359.3 nm to obtain
chromium (Cr) in an analytical sample. The performance of this testing method is shown in
Comment 4.
In addition, fertilizers containing no organic matters can also be measured by the method of 5.5.c.
However, this method should be used for fertilizers that bump when heating. (Fertilizers made of
molten matters, slag, etc. may bump in many cases).
Comment 1 Instead of the chromium standard solution in (2), a chromium standard solution for
the calibration curve preparation can be prepared by using a chromium standard
solution (Cr 1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (4) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
Note (5) Sudden temperature rise may cause bumping. Therefore, increase the temperature
gradually.
(6) If there is no possibility of bumping, the watch glass can be removed.
(7) Drying and hardening it up may cause the chromium to become insoluble. Therefore,
care should be taken not to concentrate too much.
(8) Heating at high temperature may cause bumping. Therefore, it is recommended to
increase the temperature gradually from about 170 °C to the temperature at which it
boils slightly.
(9) If much nitric acid remains, care should be taken because bumping can occur easily
when heating temperature rises.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
433
Testing Methods for Fertilizers (2018)
Note (10) When background correction is conducted by the Zeeman method, 359.3 nm is
recommended as the analytical line wavelength.
(11) Acetylene−nitrous oxide flame can also be used.
(12) If there is a possibility that the concentration of chromium in the sample solution
exceeds the maximum limit of the calibration curve, transfer no more than 10 mL of
the sample solution to a 100-mL volumetric flask and add about 10 mL of interference
suppressor solution and about 67 mL of hydrochloric acid (1+17), and add water up to
the marked line.
434
Testing Methods for Fertilizers (2018)
References
1) Toshiaki HIROI, Fumika TAKATSU: Evaluation of Digest Method for Determination of
Chromium in Inorganic Fertilizer by Atomic Absorption Spectrometry, Research Report
of Fertilizer Vol.10, p. 9 - 28 (2017)
435
Testing Methods for Fertilizers (2018)
(5) Flow sheet for chromium: The flow sheet for chromium in fertilizers mainly containing
molten matters, slag is shown below:
1.00 g
300-mL tall beaker.
analytical sample
← 4 g of ammonium sulfate
← A small amount of water, moisten the analytical sample
← About 10 mL of nitric acid
← About 5 mL of sulfuric acid
← About 5 mL of perchloric acid
Cover with a watch glass, and heat mildly on a hot plate
Heating
or sand bath at 170 ºC - 220 ºC for no less than 1 hour
Sample solution
436
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
(no more than 25 mL) 100-mL volumetric flasks
← About 10 mL of interference suppressor solution
← Hydrochloric acid (1+17) ( up to the marked line)
Measurement Atomic absorption spectrometer (357.9 nm or 359.3 nm)
437
Testing Methods for Fertilizers (2018)
5.5.c Flame atomic absorption spectrometry (Fertilizers not containing organic matters)
(1) Summary
This testing method is applicable to fertilizers not containing organic matters (including calcined
sludge fertilizer). This testing method is classified as Type B and its symbol is 5.5.c-2017 or Cr.c-1.
In addition, for fertilizers mainly made of molten matters, slag, etc., care should be taken because
they may bump while heating. 5.5.b applies to fertilizers that may bump.
Pretreat an analytical sample with nitric acid−sulfuric acid−perchloric acid, and then spray into an
acetylene−air flame, and measure the atomic absorption with chromium at a wavelength of 357.9
nm or 359.3 nm to obtain chromium (Cr) in an analytical sample. In addition, the performance of
this testing method is shown in Comment 8.
Comment 1 Instead of the chromium standard solution in (2), a chromium standard solution for
the calibration curve preparation can be prepared by using a chromium standard
solution (Cr 1 mg/mL or 10 mg/mL) traceable to National Metrology.
Note (4) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 5.2.a.
Comment 3 When the analytical sample solidifies in the procedure in (4.1) b), moisten the
analytical sample with a small amount of water as necessary in advance.
Comment 4 It is not necessary to conduct the procedures in (4.1) b) “leave at rest overnight”
because the range subjected to analysis does not contain organic matters.
Comment 5 In some cases, about 10 minutes heating in procedure in (4.1) g) is required.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
439
Testing Methods for Fertilizers (2018)
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 357.9 nm or 359.3 nm (10)
b) Calibration curve preparation
1) Spray the chromium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a low-fuel acetylene−air flame (11), and read
the indicated value at a wavelength of 357.9 nm or 359.3 nm (10).
2) Prepare a curve for the relationship between the chromium concentration and the indicated
value of the chromium standard solutions for the calibration curve preparation and the blank
test solution for the calibration curve preparation.
c) Sample measurement
1) Transfer 25 mL of the sample solution (12) to a 100-mL volumetric flask.
2) Add about 10 mL of interference suppressor solution (3), and add hydrochloric acid (1+17) to
the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Subject the blank test solution to the same procedure as in 1) - 2) and b) 1) to read the
indicated value, and correct the indicated value obtained for the sample solution.
5) Obtain the chromium content from the calibration curve, and calculate chromium (Cr) in the
analytical sample.
Note (10) When background correction is conducted by the Zeeman method, 359.3 nm is
recommended as the analytical line wavelength.
(11) Acetylene−nitrous oxide flame can also be used.
(12) If there is a possibility that the concentration of chromium in the sample solution
exceeds the maximum limit of the calibration curve, transfer no more than 10 mL of
the sample solution to a 100-mL volumetric flask and add about 10 mL of interference
suppressor solution and about 67 mL of hydrochloric acid (1+17), and add water up to
the marked line.
440
Testing Methods for Fertilizers (2018)
References
1) Hisanori ARAYA, Yoshimi TAKEBA and Toshiaki HIROI: Evaluation of Digest
Method for Determination of Chromium in Calcined Sludge Fertilizer by Atomic
Absorption Spectrometry - Research Report of Fertilizer Vol.4, p. 23 - 29 (2011)
2) Takeshi UCHIYAMA and Yuji SHIRAI: Evaluation of Digest Method for Determination
of Chromium in Calcined Sludge Fertilizer by Atomic Absorption Spectrometry: A
Collaborative Study. Research Report of Fertilizer Vol.5, p. 41 - 47 (2012)
441
Testing Methods for Fertilizers (2018)
(5) Flow sheet for chromium: The flow sheet for chromium in fertilizers not containing calcined
sludge fertilizer and organic matters is shown below: However, fertilizers that may bump are
excluded from the scope of application.
1.00 g
200-mL - 300-mL tall beaker.
analytical sample
← A small amount of water, moisten the analytical sample (as necessary)
← About10 mL of nitric acid
← About 5 mL of sulfuric acid
Leaving at rest
Cover with a watch glass
overnight
Heating Heat on a hot plate or sand bath at no less than 300 ºC,
until yellow-brown smoke no longer evolves.
Sample solution
Figure 1 Flow sheet for chromium in fertilizers not containing calcined sludge fertilizer
and organic matters (Extraction procedure)
(Excluding fertilizers which may bump while heating)
442
Testing Methods for Fertilizers (2018)
Sample solution
Figure 2 Flow sheet for chromium in fertilizers not containing calcined sludge fertilizer
and organic matters (Measurement procedure)
(Excluding fertilizers which may bump while heating)
443
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the chromium standard solution in (2), a chromium standard solution for
the calibration curve preparation can be prepared by using a chromium standard
solution (Cr 1 mg/mL or 10 mg/mL) traceable to National Metrology.
i) As a blank test, conduct the procedures in b) - h) using another tall beaker to prepare the
blank test solution.
Note (3) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(4) The watch glass can be removed.
(5) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct measurement (Standard Addition Method) according to JIS K 0116
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Optical Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Optical Emission Spectrometer: Set up the
measurement conditions for the ICP Optical Emission Spectrometer considering the
following:
Analytical line wavelength: 205.552 nm
b) Calibration curve preparation and sample measurement
1) Put 5mL of sample solution to three 10-mL volumetric flasks respectively.
2) Add 2mL and 4 mL of chromium standard solution (2.5 μg/mL) to volumetric flasks of 1)
above, then add hydrochloric acid (1+23) to the marked line to make a sample solution of
Standard Addition Method.
3) Add hydrochloric acid (1+23) to the marked line of the remaining volumetric flask of 1)
above to make a sample solution without a standard solution.
4) Spray the sample solution of Standard Addition Method and the sample solution without a
standard solution into the induction plasma, and read the indicated value at a wavelength of
205.552 nm.
5) Transfer 5 mL of blank test solution to a 10-mL volumetric flask, conduct the same
procedures as in 3) - 4) to read the indicated value, and correct the indicated value obtained
from the respective sample solutions.
6) Prepare a curve for the relationship between the added chromium concentration and the
corrected indicated value of the sample solution for Standard Addition Method and the sample
solution without a standard solution.
7) Obtain the chromium content from the intercept of the calibration curve to calculate
chromium (Cr) in the analytical sample.
Comment 3 Instead of the correction method in c) 5), the chromium (Cr) in the analytical sample
can also be corrected by obtaining the chromium content in the blank test solution.
Comment 4 The simultaneous measurement of multiple elements by an ICP-OES is available. In
that case, see 4.9.1.b Comment 5.
Comment 5 The comparison of the measurement value (xi:12.9 mg/kg - 193 mg/kg) of ICP
Optical Emission Spectrometry and the measurement value (yi) of Flame Atomic
Absorption Spectrometry was conducted to evaluate trueness using sludge fertilizers
(49 samples). As a result, a regression equation was y = 1.74+0.971x and its
correlation coefficient (r) was 0.991. Triplicates measurement for each one sample of
sewage sludge fertilizer, human waste sludge fertilizer, industrial sludge fertilizer,
445
Testing Methods for Fertilizers (2018)
mixed sludge fertilizer and composted sludge fertilizer was conducted. As a result, a
repeatability obtained was 0.9 % - 2.5 % as a relative standard deviation.
Additionally, the minimum limit of quantification of this testing method is 4 mg/kg.
References
1) Masahiro ECHI, Tomoe INOUE, Megumi TABUCHI and Tetuya NOMURA:
Simultaneous Determination of Cadmium, Lead, Nickel, Chromium, Copper and Zinc
in Sludge Fertilizer using Inductively Coupled Plasma-Atomic Emission Spectrometry
(ICP-AES), Research Report of Fertilizer, Vol.4, p. 30 - 35 (2011)
(5) Flow sheet for cadmium: The flow sheet for cadmium in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
Aliquot 5 mL 10-mL volumetric flask, 3 flasks
← 0, 2 and 4 mL of chromium standard solution (2.5 μg / mL) respectively
← Hydrochloric acid (1+23) ( up to the marked line)
Measurement ICP Optical Emission Spectrometer (205.552 nm)
446
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the chromium standard solution in (2), a chromium standard solution for
447
Testing Methods for Fertilizers (2018)
Note (6) The maximum limit of solid content, which is converted from moisture content, in the
sampling volume 20.0 g of an analytical sample is about 0.5 g. If solid content is likely
to exceed the limit, reduce sampling volume as necessary.
(7) Condition examples for a microwave decomposing device: 0 min (room temperature)
→ 10min (240 ºC) → 20 min (240 ºC) → 40 min (room temperature), initial output
1400 W
(8) When organic matters still remain, for example the digestion solution is colored,
repeat the procedures in (4.1) b) - d).
(9) The vessel should be made of polypropylene, etc. to not affect the measurement.
(10) 16.5-cm of radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
448
Testing Methods for Fertilizers (2018)
force.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 5.1.b.
(4.2) Measurement: Conduct measurement (Internal Standard Method) according to JIS K 0113
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Mass Spectrometer used in measurement.
a) Measurement conditions for the ICP Mass Spectrometer: Set up the measurement
conditions for the ICP Mass Spectrometer considering the following:
Chromium: monitor ion (m/z):52、53、50
Rhodium: monitor ion (m/z): 103
b) Calibration curve preparation
1) Spray chromium standard solution for the calibration curve preparation and blank test solution
for the calibration curve preparation into the inductively coupled plasma, and read the ratio of
the respective ion count numbers for the monitor ion of an element subjected to measurement
and an internal standard element.
2) Prepare a curve for the relationship between the concentration and the ratio of the ion count
number of an element subjected to measurement.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.05 µg - 5 µg as
Cr) to a 50-mL volumetric flask (9).
2) Add 5mL of internal standard solution (5) and add nitric acid (1+19) to the marked line.
3) Conduct procedures similarly as in b) 1) to read the ratio of the ion count number.
4) For blank test solution, conduct procedures similarly as in 1) - 3) to correct the ratio of the ion
count number obtained for the sample solution.
5) Obtain the chromium content from the calibration curve, and calculate chromium (Cr) in the
analytical sample.
Comment 3 Instead of the correction method in c) 4), the chromium (Cr) in the analytical sample
can also be corrected by obtaining the chromium content in the blank test solution.
Comment 4 Triplicates additive recovery testing was conducted to evaluate trueness using fluid
industrial sludge fertilizers (2 samples) and composted sludge fertilizers (6 samples).
As a result, the mean recovery at the concentration level of 10 mg/kg - 90 mg/kg, 1
mg/kg - 9 mg/kg and 0.1 mg /kg - 0.9 mg /kg are 92.4 % - 108.8 %, 94.3 % -
115.4 % and 105.8 - 106.8 % as chromium (Cr) in an actual article respectively.
The results of the repeatability tests on different days using two kinds of fluid sludge
fertilizers to evaluate precision were analyzed by the one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.。
Additionally, the minimum limit of quantification of this test method is about 1 µg
/kg.
449
Testing Methods for Fertilizers (2018)
References
1) Toshiharu YAGI: Determination of Heavy Metals in Fluid Sludge Fertilizers by ICP-MS
and CV-AAS, Research Report of Fertilizer Vol. 8, p. 26 - 37 (2015)
450
Testing Methods for Fertilizers (2018)
(5) Flow sheet for chromium: The flow sheet for chromium in fluid sludge fertilizers is shown
below:
Sample solution
Sample solution
451
Testing Methods for Fertilizers (2018)
5.6 Lead
5.6.a Flame atomic absorption spectrometry
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type B and its
symbol is 5.6.a-2017 or Pb.a-1.
Pretreat an analytical sample with incineration and nitric acid−hydrochloric acid (1+3), spray into
an acetylene−air flame, and measure the atomic absorption with lead at a wavelength of 217.0 nm
or 283.3 nm to obtain the lead (Pb) in an analytical sample. In addition, the performance of this
testing method is shown in Comment 5.
Comment 1 Instead of the lead standard solution in (2), a lead standard solution for the calibration
curve preparation can be prepared by using a lead standard solution (Pb 1 mg/mL or
10 mg/mL) traceable to National Metrology.
Note (3) There are the continuous source method, the Zeeman method, the non-resonance
spectrum method, and the self-reversal method, etc.
452
Testing Methods for Fertilizers (2018)
Note (4) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250°C in 30 minutes to 1 hour, continue heating for about 1
hour and further raise to 450 °C in 1 to 2 hours.
(5) The watch glass can be removed.
(6) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 217.0 nm or 283.3 nm
b) Calibration curve preparation
1) Spray the lead standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 217.0 nm or 283.3 nm.
2) Prepare a curve for the relationship between the lead concentration and the indicated value of
the lead standard solutions for the calibration curve preparation and the blank test solution for
the calibration curve preparation.
c) Sample measurement
1) Subject the sample solution (7) to the same procedure as in b) 1) to read the indicated value.
2) Subject the blank test solution to the same procedure as in b) 1) to read the indicated value,
and correct the indicated value obtained for the sample solution.
3) Obtain the lead content from the calibration curve, and calculate lead (Pb) in the analytical
sample.
453
Testing Methods for Fertilizers (2018)
Note (7) If there is a possibility that the lead concentration in the sample solution will exceed
the maximum limit of the calibration curve, dilute a predetermined amount with
hydrochloric acid (1+23).
Comment 4 Instead of the correction method in c) 2), the lead (Pb) in the analytical sample can
also be corrected by obtaining the lead content in the blank test solution.
Comment 5 Recovery testing was conducted using industrial sludge fertilizer and composted
sludge fertilizer (5 samples); as a result, the recovery at the concentration level of
100 mg/kg and 10 mg/kg were 99.1 % - 100.6 % and 97.5 % - 99.6 %, respectively.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of this test method is about 1
mg/kg.
References
1) Yoshinari SAKAKIBARA, Manabu MATSUZAKI and Tadao AMANO: Determination
of Cadmium, Lead, Nickel and Chromium in Sludge Fertilizer - Improved
Decomposition Method - Research Report of Fertilizer Vol.1, p. 41 - 49 (2008)
2) Yoshinari SAKAKIBARA and Manabu MATSUZAKI: Determination of Cadmium,
Lead, Nickel and Chromium in Sludge Fertilizer - Collaborative Test Results - Research
Report of Fertilizer Vol.1, p. 50 - 59 (2008)
3) Hisanori ARAYA and Yoshimi TAKEBA: Validation of atomic absorption spectrometry
for determination of cadmium, lead, nickel and chromium in calcined sludge fertilizer -
Research Report of Fertilizer Vol.3, p. 30 - 42 (2010)
454
Testing Methods for Fertilizers (2018)
(5) Flow sheet for lead: The flow sheet for lead in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
455
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the lead standard solution in (2), a lead standard solution for the calibration
curve preparation can be prepared by using a lead standard solution (Pb 1 mg/mL or
10 mg/mL) traceable to National Metrology.
i) As a blank test, conduct the procedures in b) - h) using another tall beaker to prepare the
blank test solution.
Note (3) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 450 °C in 1 to 2 hours.
(4) The watch glass can be removed.
(5) Add hydrochloric acid (1+5) so that the hydrochloric acid concentration of the sample
solution will be hydrochloric acid (1+23). For example, when a 100-mL volumetric
flask is used in the procedure in h), about 25 mL of hydrochloric acid (1+5) should be
added.
Comment 2 Do not conduct the procedures in (4.1) b) - c) in the case of fertilizers not containing
organic matters.
Comment 3 The procedure in (4.1) is the same as the procedure in (4.1) of 4.9.1.a.
(4.2) Measurement: Conduct measurement (Standard Addition Method) according to JIS K 0116
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Optical Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Optical Emission Spectrometer: Set up the
measurement conditions for the ICP Optical Emission Spectrometer considering the
following:
Analytical line wavelength: 220.351 nm
b) Calibration curve preparation and sample measurement
1) Put 5mL of sample solution to three 10-mL volumetric flasks respectively.
2) Add 2 mL and 4 mL of lead standard (0.25 μg/mL) solution to volumetric flasks of 1) above,
then add hydrochloric acid (1+23) to the marked line to make a sample solution of Standard
Addition Method.
3) Add hydrochloric acid (1+23) to the marked line of the remaining volumetric flask of 1)
above to make a sample solution without a standard solution.
4) Spray the sample solution of Standard Addition Method and the sample solution without a
standard solution into the induction plasma, and read the indicated value at a wavelength of
220.351 nm.
5) Transfer 5 mL of blank test solution to a 10-mL volumetric flask, conduct the same
procedures as in 3) - 4) to read the indicated value, and correct the indicated value obtained
from the respective sample solutions.
6) Prepare a curve for the relationship between the added lead concentration and the corrected
indicated value of the sample solution for Standard Addition Method and the sample solution
without a standard solution.
7) Obtain the lead content from the intercept of the calibration curve, and calculate lead (Pb) in
the analytical sample.
Comment 4 Instead of the correction method in c) 5), the lead (Pb) in the analytical sample can
also be corrected by obtaining the lead content in the blank test solution.
Comment 5 The simultaneous measurement of multiple elements by an ICP-OES is available. In
that case, see 4.9.1.b Comment 5.
Comment 6 The comparison of the measurement value (xi:1.1 mg/kg - 69.0 mg/kg) of ICP
Optical Emission Spectrometry and the measurement value (yi) of Flame Atomic
Absorption Spectrometry was conducted to evaluate trueness using sludge fertilizers
(49 samples). As a result, a regression equation was y = − 0.31 + 1.045x and its
457
Testing Methods for Fertilizers (2018)
correlation coefficient (r) was 0.993. Triplicates measurement for each one sample of
sewage sludge fertilizer, human waste sludge fertilizer, industrial sludge fertilizer,
mixed sludge fertilizer, calcined sludge fertilizers and composted sludge fertilizer
was conducted. As a result, a repeatability obtained was 0.9 % - 3.3 % as a relative
standard deviation.
Additionally, the minimum limit of quantification of this testing method is about 5
mg/kg.
References
1) Masahiro ECHI, Tomoe INOUE, Megumi TABUCHI and Tetuya NOMURA:
Simultaneous Determination of Cadmium, Lead, Nickel, Chromium, Copper and Zinc
in Sludge Fertilizer using Inductively Coupled Plasma-Atomic Emission Spectrometry
(ICP-AES), Research Report of Fertilizer, Vol.4, p. 30 - 35 (2011)
(5) Flow sheet for lead: The flow sheet for lead in fertilizers is shown below:
5.00 g
200-mL - 300-mL tall beaker.
analytical sample
Sample solution
458
Testing Methods for Fertilizers (2018)
Sample solution
459
Testing Methods for Fertilizers (2018)
Comment 1 Instead of the lead standard solution in (2), a lead standard solution for the calibration
curve preparation can be prepared by using a lead standard solution (Pb 1 mg/mL or
460
Testing Methods for Fertilizers (2018)
Note (6) The maximum limit of solid content, which is converted from moisture content, in the
sampling volume 20.0 g of an analytical sample is about 0.5 g. If solid content is likely
to exceed the limit, reduce sampling volume as necessary.
(7) Condition examples for a microwave decomposing device: 0 min (room temperature)
→ 10min (240 ºC) → 20 min (240 ºC) → 40 min (room temperature), initial output
1400 W
(8) When organic matters still remain, for example the digestion solution is colored,
repeat the procedures in (4.1) b) - d).
(9) The vessel should be made of polypropylene, etc. to not affect the measurement.
(10) 16.5-cm of radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
461
Testing Methods for Fertilizers (2018)
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1) of 5.1.b.
(4.2) Measurement: Conduct measurement (Internal Standard Method) according to JIS K 0113
and as shown below. Specific measurement procedures are according to the operation method
of the ICP Mass Spectrometer used in measurement.
a) Measurement conditions for the ICP Mass Spectrometer: Set up the measurement
conditions for the ICP Mass Spectrometer considering the following:
Lead: monitor ion (m/z):208、206、207
Rhenium: monitor ion (m/z): 187
b) Calibration curve preparation
1) Spray lead standard solution for the calibration curve preparation and blank test solution for
the calibration curve preparation into the inductively coupled plasma, and read the ratio of the
respective ion count numbers for the monitor ion of an element subjected to measurement and
an internal standard element.
2) Prepare a curve for the relationship between the concentration and the ratio of the ion count
number of an element subjected to measurement.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.05 µg - 5 µg as
Pb) to a 50-mL volumetric flask (9).
2) Add 5mL of internal standard solution (5) and add nitric acid (1+19) to the marked line.
3) Conduct procedures similarly as in b) 1) to read the ratio of the ion count number.
4) For blank test solution, conduct procedures similarly as in 1) - 3) to correct the ratio of the ion
count number obtained for the sample solution.
5) Obtain the lead content from the calibration curve, and calculate lead (Pb) in the analytical
sample.
Comment 3 Instead of the correction method in c) 4), the lead (Pb) in the analytical sample can
also be corrected by obtaining the lead content in the blank test solution.
Comment 4 Triplicates additive recovery testing was conducted to evaluate trueness using fluid
industrial sludge fertilizers (2 samples) and composted sludge fertilizers (6 samples).
As a result, the mean recovery at the concentration level of 10 mg/kg - 20 mg/kg, 1
mg/kg - 5 mg/kg and 0.1 mg /kg - 0.9 mg /kg are 91.2 % - 103.1 %, 85.0 % -
113.9 % and 93.2 - 108.1 % as lead (Pb) in an actual article respectively.
The results of the repeatability tests on different days using two kinds of fluid sludge
fertilizers to evaluate precision were analyzed by the one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this test method is about 4 µg
/kg.
462
Testing Methods for Fertilizers (2018)
References
1) Toshiharu YAGI: Determination of Heavy Metals in Fluid Sludge Fertilizers by ICP-MS
and CV-AAS, Research Report of Fertilizer Vol. 8, p. 26 - 37 (2015)
(5) Flow sheet for lead: The flow sheet for lead in fluid sludge fertilizers is shown below:
Sample solution
Figure 1 Flow sheet for lead in liquid sludge fertilizers (Extraction procedure)
463
Testing Methods for Fertilizers (2018)
Sample solution
Figure 2 Flow sheet for lead in liquid sludge fertilizers (Measurement procedure)
464
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Ion Chromatograph (IC) or High-Performance Liquid Chromatograph (HPLC): IC
specified in JIS K 0127 or HPLC specified in JIS K 0124 that satisfies following
requirements.
1) Column: A 4-mm inner diameter 100-mm long stainless-steel column tube filled with
hydrophilic methacrylate-gel, to which 5-μm particle diameter class 4 ammonium group
chemically bonds (3).
2) Column bath: A column bath whose temperature can be adjusted to 55 ºC - 60 ºC.
465
Testing Methods for Fertilizers (2018)
Note (3) A column is commercially sold under the name Shodex IC NI-424, etc.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0127 and as shown below.
Specific measurement procedures are according to the operation method of the Ion
Chromatograph (IC) or the High-Performance Liquid Chromatograph (HPLC) used in
measurement.
a) Measurement conditions for the Ion Chromatograph (IC) or High-Performance Liquid
Chromatograph (HPLC): An example of measurement conditions is shown below. Set up
the measurement conditions considering it:
1) Column: A hydrophilic methacrylate-gel column (4-mm inner diameter, 100-mm long, 5-μm
particle diameter) to which quaternary ammonium group chemically bonds.
2) Column bath temperature: 58 °C
3) Eluent: Prepared by the procedures in (2) g)
4) Flow rate: 1 mL/min
5) Injection volume: 20 µL
6) Detection unit: An electric conductivity detector
Comment 1 In the measurement of a sample solution, there is a possibility that the recovery rate
becomes lower than actual due to the influence of matrix if the concentration is calculated with
peak height. Therefore, prepare a calibration curve using peak area.
c) Sample measurement
1) Subject 20 μL of sample solution to the same procedure as in b) 1)
2) Obtain the sulfamic acid content from calibration curve by peak area to calculate the sulfamic
acid (amidosulfuric acid) in the analytical sample.
Comment 2 Calculate the concentration by the peak area similarly as the calibration curve
preparation to prevent the influence of matrix.
Comment 3 Note that it takes time to stabilize the baseline due to the usage of the ion-pairing
reagent in the elute. It is recommended to take about 120 minutes for stabilization
time before starting measurement.
Comment 4 It is possible for the simultaneous measurement of sulfamic acid and ammonium
466
Testing Methods for Fertilizers (2018)
References
1) Toshiaki HIROI and Yuji SHIRAI: Simultaneous Determination of Sulfamic Acid and
Ammonium Thiocyanate in Ammonium Sulfate by Non-suppressed Ion
Chromatography, Research Report of Fertilizer, Vol.5, p. 1 - 23 (2012)
(5) Flow sheet for testing method: The flow sheet for sulfamic acid in ammonium sulfate is
shown below:
Sample solution
467
Testing Methods for Fertilizers (2018)
Reference: The IC chromatogram of sulfamic acid and thiocyanic acid of the standard solution
for the calibration curve preparation and sample solution (ammonium sulfate) are
shown below.
(A) Mixture standard solution (the equivalents of 60 ng as sulfamic acid and ammonium
thiocyanate (3 mg/L, 20 μL), respectively)
(B) Sample solution (the equivalents of 0.25 % (mass fraction) as sulfamic acid and
ammonium thiocyanate added in ammonium sulfate (2500 μg/g), respectively)
IC measurement conditions
Column: Shodex IC NI-424 (4.6-mm inner diameter, 100-mm long, 5μm-particle
diameter)
Other conditions are according to the example of HPLC measurement conditions in
(4.2) a)
468
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph/Mass Spectrometer (LC-MS): LC-MS
specified in JIS K 0136 that satisfies the following requirements.
1) High-Performance Liquid Chromatograph
(i) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45 ºC.
(ii) Column: A 2-mm - 3-mm inner diameter 100-mm - 150 mm long stainless-steel column
tube filled with silica gel to which 5-μm crosslink-type diol chemically bonds or
polyhydroxymethacrylate. This should comply with the specification of a Mass
Spectrometer.
2) Mass Spectrometer
(i) Ionization method: Electro-Spray Ionization (ESI) method
(ii) Ion detection method: Selected Ion Monitoring (SIM) method
b) Magnetic stirrer:
c) Centrifugal separator: A centrifugal separator that can work at about 1700 × g.
d) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g - 10000
× g.
469
Testing Methods for Fertilizers (2018)
Note (2) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
Comment 3 Instead of the procedures in (4.1.1) c) and d), it is allowed to filter with Type 3 filter
paper and the filtrate can be the extract.
Note (3) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(4) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 4 Instead of the procedures in (4.2) b) and c), it is allowed to filter with a membrane
filter (aperture diameter: no more than 0.5 µm) made of hydrophilic PTFE and the
filtrate can be the sample solution.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0136 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph Mass Spectrometer used in measurement.
470
Testing Methods for Fertilizers (2018)
c) Sample measurement
1) Subject 1 μL of sample solution to the same procedure as in b) 1)
2) Obtain the sulfamic acid content from calibration curve to calculate the sulfamic acid
(amidosulfuric acid) in the analytical sample.
Comment 5 Recovery testing was conducted using samples that sulfamic acid equivalent to 1/5 -
4 times of permissible content are added to ammonium sulfate fertilizer (1 brand),
by-product nitrogen fertilizer (1 brand), by-product mixed fertilizer (1 brand),
compound fertilizer (1 brand) and fluid mixed fertilizer (1 brand). As a result, the
mean recovery at the additive level of 0.1 % (mass fraction), 0.025 % (mass fraction)
and 0.005 % (mass fraction) are 97.6 % - 104.2 %, 95.2 % - 107.0 % and 96.4 % -
111.2 % respectively.
The results of the repeatability tests on different days using ammonium sulfate
fertilizer, by-product nitrogen fertilizer and by-product mixed fertilizer to evaluate
precision were analyzed by the one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability.
Table 2 shows results and analysis results from a collaborative study for testing
method validation. In addition, sufficient reproducibility by sulfamic acid
concentration 0.0116 % (mass fraction) was not obtainable. But sufficient
reproducibility was obtained in the scope of sulfamic acid concentration 0.0386 %
(mass fraction) - 0.401 % (mass fraction).
471
Testing Methods for Fertilizers (2018)
References
1) Kohei ITO, Mariko FUJITA, Yoshimi HASHIMOTO and Yuji SHIRAI: Determination
of Sulfamic Acid in Fertilizer by Liquid Chromatography/Mass Spectrometry (LC-MS),
Research Report of Fertilizer Vol. 8, p. 38 - 49 (2015)
2) Tomoharu NOZAKI Determination of Sulfamic Acid in Fertilizers using Liquid
Chromatography-Mass Spectrometer (LC-MS): A Collaborative Study, Research Report
of Fertilizer Vol. 9, p. 69 - 76 (2016)
472
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for sulfamic acid in fertilizers is shown
below:
Extract
1.00 g
100-mL volumetric flask
analytical sample (fluid)
← About 50 mL of water
Extraction Shake to mix
← Water (up to the marked line)
Extract
Extract
Sample solution
473
Testing Methods for Fertilizers (2018)
Reference: The chromatogram of sulfamic acid for the calibration curve preparation is shown
below.
Sulfamic acid
Sulfamic acid
474
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Ion Chromatograph (IC) or High-Performance Liquid Chromatograph (HPLC): IC
specified in JIS K 0127 or HPLC specified in JIS K 0124 that satisfies following
requirements.
1) Column: A 4-mm inner diameter 100-mm long stainless steel column tube filled with
hydrophilic methacrylate-gel, to which 5-μm particle diameter class 4 ammonium group
chemically bonds (4).
475
Testing Methods for Fertilizers (2018)
Note (4) A column is commercially sold under the name Shodex IC NI-424, etc.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0127 and as shown below.
Specific measurement procedures are according to the operation method of the Ion
Chromatograph (IC) or the High-Performance Liquid Chromatograph (HPLC) used in
measurement.
a) Measurement conditions for the Ion Chromatograph (IC) or High-Performance Liquid
Chromatograph (HPLC): An example of measurement conditions for an Ion
Chromatograph (IC) is shown below. Set up the measurement conditions considering it:
1) Column: A hydrophilic methacrylate-gel column (4-mm inner diameter, 100-mm long, 5-μm
particle diameter) to which quaternary ammonium group chemically bonds.
2) Column bath temperature: 58 °C
3) Eluent: Prepared by the procedures in (2) g)
4) Flow rate: 1 mL/min
5) Injection volume: 20 µL
6) Detection unit: An electric conductivity detector
Comment 1 In the measurement of a sample solution, there is a possibility that the recovery rate
becomes lower than actual due to the influence of matrix if the concentration is
calculated with peak height. Therefore, prepare a calibration curve using peak area.
c) Sample measurement
1) Subject 20 μL of sample solution to the same procedure as in b) 1)
2) Obtain the ammonium thiocyanatecontent from the calibration curve by peak area to calculate
ammonium thiocyanate (sulfurized cyanide) in the analytical sample.
Comment 2 Calculate the concentration by the peak area similarly as the calibration curve
preparation to prevent the influence of matrix.
Comment 3 Note that it takes time to stabilize the baseline due to the usage of the ion-pairing
reagent in the elute. It is recommended to take about 120 minutes for stabilization
time before starting measurement.
476
Testing Methods for Fertilizers (2018)
Comment 4 It is possible for the simultaneous measurement of sulfamic acid and ammonium
thiocyanate (sulfurized cyanide) in this testing method. In that case, mix a
predetermined amount of sulfamic acid standard solution (1000 mg/L) and
ammonium thiocyanate standard solution (1000 mg/L), dilute with water to prepare a
mixture standard solution (10 mg/L) (1) and use it instead of respective standard
solutions (10 mg/L) in (2) i). After that, conduct the same procedure in (4.2) b) to
calculate the respective concentrations of materials subjected to measurement in the
analytical sample.
Comment 5 A recovery testing of ammonium sulfate (3 brands) was conducted. As a result, the
mean recovery at additive level of 0.25 % (mass fraction) and 0.075 % (mass
fraction) was 101.8 % - 103.7 % and 93.9 % - 97.4 %.
Additionally, the minimum limit of quantification of this testing method is about
0.04 % (mass fraction).
References
1) Toshiaki HIROI and Yuji SHIRAI: Simultaneous Determination of Sulfamic Acid and
Ammonium Thiocyanate in Ammonium Sulfate by Non-suppressed Ion
Chromatography, Research Report of Fertilizer, Vol.5, p. 1 - 23 (2012)
(5) Flow sheet for testing method: The flow sheet for ammonium thiocyanate in ammonium
sulfate is shown below:
Sample solution
477
Testing Methods for Fertilizers (2018)
Reference: The IC chromatogram of sulfamic acid and thiocyanic acid of the standard solution
for the calibration curve preparation and sample solution (ammonium sulfate) are shown
below.
(A) Mixture standard solution (the equivalents of 60 ng as sulfamic acid and ammonium
thiocyanate (3 mg/L, 20 μL), respectively)
(B) Sample solution (the equivalents of 0.25 % (mass fraction) as sulfamic acid and
ammonium thiocyanate added in ammonium sulfate (2500 μg/g), respectively)
IC measurement conditions
Column: Shodex IC NI-424 (4.6-mm inner diameter, 100-mm long, 5μm-particle
diameter)
Other conditions are according to the example of HPLC measurement conditions in
(4.2) a)
478
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph (HPLC): HPLC specified in JIS K 0124 that
satisfies following requirements.
1) Column: A 4-mm to 6-mm inner diameter 150-mm to 250-mm long stainless steel column
tube filled with poly vinyl alcohol or silica gel (2) , to which 5-μm particle diameter amino
group chemically bonds.
2) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45 ºC.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 210 nm.
b) Magnetic stirrer:
c) Centrifugal separator: A centrifugal separator that can work at about 1700 × g.
d) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g - 10000
× g.
e) pH test paper: A pH test paper infilled with indicator and dried, which can measure the
479
Testing Methods for Fertilizers (2018)
value from pH 1 to pH 11 and a color change chart with the pH interval value 1 is attached.
Note (2) Remaining silanol group of silica gel affects the measurement of ion in some cases.
Therefore, use a column which does not affect the measurement of sodium thiocyanate
by treating the silanol group. As an example of the treatment, silica gel is to be entirely
covered with the uniform membrane of silicone polymer.
Comment 1 A column is sold under the names CAPCELL PAK NH2 UG80 or Asahipak
NH2P-50 4E.
Comment 2 pH test paper is sold under the name UNIV Test Paper, etc.
Note (3) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
Note (4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 3 Instead of procedures in (4.2) b) and e) - f), it is allowed to filter with a membrane
filter (aperture diameter: no more than 0.5-µm) made of hydrophilic PTFE and the
filtrate can be the sample solution.
480
Testing Methods for Fertilizers (2018)
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of High-Performance
Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for an Ion Chromatograph (IC) is shown below. Set up
the measurement conditions considering it:
1) Column: A vinyl alcohol polymer column (4-mm - 6-mm inner diameter, 150-mm - 250-mm
long, 5-µm particle diameter) to which amino group chemically bonds or a silica gel column
(4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm particle diameter) to which
amino group chemically bonds.
2) Column bath temperature: 30 °C - 40 °C
3) Eluent (1): Dissolve 1.79 g of disodium hydrogen phosphate dodecahydrate, 0.78 g of sodium
dihydrogen phosphate dihydrate and 14.04 g of sodium perchlorate monohydrate in water to
make 1000 mL. Filter with a membrane filter (aperture diameter: no more than 0.5-µm) made
of hydrophilic PTFE.
4) Flow rate: 0.9 mL/min - 1.0 mL/min
5) Injection volume: 10 µL
6) Detection unit: An absorptiometric detector, measurement wavelength: 210 nm
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the ammonium thiocyanate content from the calibration curve by peak area to
calculate ammonium thiocyanate (sulfurized cyanide) in the analytical sample.
Comment 4 This testing method enables the simultaneous measurement of nitrous acid and
ammonium thiocyanate (sulfurized cyanide). In this case, mix a predetermined
amount of nitrous acid standard solution (1 mg/mL) and ammonium thiocyanate
standard solution (1 mg/mL), dilute with water to prepare a mixture standard solution
(100 μg/mL) (1) and use it instead of (2) h) ammonium thiocyanate standard solution
(100 μg/mL). After that, conduct the same procedure in (4.3) b) to calculate the
respective concentrations of materials subjected to measurement in the analytical
sample.
Comment 5 A recovery testing was conducted using samples that ammonium thiocyanate
equivalent to 1/5 - 5 times of permissible content are added to an ammonium sulfate
fertilizer (1 brand), a coating nitrogen fertilizer (1 brand), a blended fertilizer (2
brands), a compound fertilizer (1 brand) and a fluid mixed fertilizer (1 brand). As a
result, the mean recovery at the additive level of 0.025 % (mass fraction), 0.01 %
(mass fraction), 0.005 % (mass fraction) and 0.0025 % (mass fraction) are 95.4 % -
100.5 %, 94.7 % - 103.8 %, 83.8 % - 109.0 % and 87.2 % - 103.3 % respectively.
Table 1 shows results and analysis results from a collaborative study for test method
validation. Ammonium thiocyanate had sufficient reproducibility in the range from
0.00476 % (mass fraction) to 0.204 %. (mass fraction).
Additionally, the minimum limit of quantification of this testing method is about
0.002 % (mass fraction).
481
Testing Methods for Fertilizers (2018)
References
1) Kohei ITO, Yasuharu KIMURA, Masanori HASEGAWA and Yuji SHIRAI:
Simultaneous Determination of Nitrous Acid and Thiocyanate in Fertilizer by
High-Performance Liquid Chromatography (HPLC), Japanese Journal of Soil Science
and Plant Nutrition, Vol. 87 (2), p.120 - 124 (2016)
2) Masanori HASEGAWA and Yasuharu KIMURA: Determination of Nitrous Acid and
Ammonium Thiocyanate in Fertilizer by High-Performance Liquid Chromatography
(HPLC): A Collaborative Study, Research Report of Fertilizer Vol. 8, p. 70 - 78 (2015)
482
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for ammonium thiocyanate in fertilizers is
shown below:
1.00 g analytical
sample 200-mL ground-in stopper Erlenmeyer flask
(powdery)
← 100 mL of water
Extraction Stir to mix, 10 minites
Filtrate
1.00 g analytical
sample 100-mL volumetric flask
(fluid)
← 50 mL - 70 mL of water
Extraction Shake to mix
← Water (up to the marked line)
Filtrate
483
Testing Methods for Fertilizers (2018)
Filtrate
pH adjustment
Sodium hydroxide solution (5 mg/mL)
(pH 5 - pH 7)
Sample
484
Testing Methods for Fertilizers (2018)
Reference: HPLC chromatogram of nitrous acid and ammonium thiocyanate are shown
below.
←1
2
↓
Retention time
1→
2
↓
Retention time
485
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph (HPLC): HPLC specified in JIS K 0124 that
satisfies following requirements.
1) Column: A 4-mm - 6-mm inner diameter 150-mm - 250-mm long stainless steel column tube
filled with poly vinyl alcohol or silica gel (2), to which 5-μm particle diameter amino group
chemically bonds.
2) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45 ºC.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 210 nm.
b) Magnetic stirrer:
c) Centrifugal separator: A centrifugal separator that can work at about 1700 × g.
d) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g -
10000 × g.
e) pH test paper: A pH test paper infilled with indicator and dried, which can measure the
value from pH 1 - pH 11 and a color change chart with the pH interval value 1 is attached.
486
Testing Methods for Fertilizers (2018)
Note (2) Remaining silanol group of silica gel affects the measurement of ion in some cases.
Therefore, use a column which does not affect the measurement of nitrous acid by
treating the silanol group. As an example of the treatment, silica gel is to be entirely
covered with the uniform membrane of silicone polymer.
Comment 1 A column is sold under the names CAPCELL PAK NH2 UG80 or Asahipak
NH2P-50 4E.
Comment 2 pH test paper is sold under the name UNIV Test Paper, etc.
Note (3) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
Note (4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 3 Instead of procedures in (4.2) b) and e) - f), it is allowed to filter with a membrane
filter (aperture diameter: no more than 0.5-µm) made of hydrophilic PTFE and the
filtrate can be the sample solution.
487
Testing Methods for Fertilizers (2018)
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of High-Performance
Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for an Ion Chromatograph (IC) is shown below. Set up
the measurement conditions considering it:
1) Column: A vinyl alcohol polymer column (4-mm - 6-mm inner diameter, 150-mm - 250-mm
long, 5-µm particle diameter) to which amino group chemically bonds or a silica gel column
(4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm particle diameter) to which
amino group chemically bonds.
2) Column bath temperature: 30 °C - 40 °C
3) Eluent (1): Dissolve 1.79 g of disodium hydrogen phosphate dodecahydrate, 0.78 g of sodium
dihydrogen phosphate dihydrate and 14.04 g of sodium perchlorate monohydrate in water to
make 1000 mL. Filter with a membrane filter (aperture diameter: no more than 0.5-µm) made
of hydrophilic PTFE.
4) Flow rate: 0.9 mL/min - 1.0 mL/min
5) Injection volume: 10 µL
6) Detection unit: An absorptiometric detector, measurement wavelength: 210 nm
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the nitrous acid content by the peak area using the calibration curve to calculate the
concentration of nitrous acid in the analytical sample.
Comment 4 This testing method enables the simultaneous measurement of nitrous acid and
ammonium thiocyanate (sulfurized cyanide). In this case, mix a predetermined
amount of nitrous acid standard solution (1 mg/mL) and ammonium thiocyanate
standard solution (1 mg/mL), dilute with water to prepare a mixture standard solution
(100 μg/mL) (1) and use it instead of (2) h) nitrous acid standard solution (100
μg/mL). After that, conduct the same procedure in (4.3) b) to calculate the respective
concentrations of materials subjected to measurement in the analytical sample.
Comment 5 Recovery testing was conducted using samples that nitrous acid equivalent to 1/5 - 5
times of permissible content are added to ammonium sulfate fertilizer (1 brand),
coating nitrogen fertilizer (1 brand), blended fertilizer (2brands), compound fertilizer
(1 brand) and fluid mixed fertilizer (1 brand). As a result, the mean recovery at the
additive level of 0.1 % (mass fraction), 0.04 % (mass fraction), 0.02 % (mass
fraction) and 0.01 % (mass fraction) are 99.0 % - 100.8 %, 100.4 % - 102.0 %,
103.1 % - 106.6 % and 101.2 % - 105.9 % respectively.
Table 1 shows results and analysis results from a collaborative study for test method
validation. Nitrous acid had sufficient reproducibility in the range from 0.0255 %
(mass fraction) to 0.291 %. (mass fraction).
Additionally, the minimum limit of quantification of this testing method is about
0.0003 % (mass fraction).
488
Testing Methods for Fertilizers (2018)
References
1) Kohei ITO, Yasuharu KIMURA, Masanori HASEGAWA and Yuji SHIRAI:
Simultaneous Determination of Nitrous Acid and Thiocyanate in Fertilizer by
High-Performance Liquid Chromatography (HPLC), Japanese Journal of Soil Science
and Plant Nutrition, Vol. 87 (2), p.120 - 124 (2016)
2) Masanori HASEGAWA and Yasuharu KIMURA: Determination of Nitrous Acid and
Ammonium Thiocyanate in Fertilizer by High-Performance Liquid Chromatography
(HPLC): A Collaborative Study, Research Report of Fertilizer Vol. 8, p. 70 - 78 (2015)
489
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for nitrous acid in fertilizers is shown below:
1.00 g analytical
sample 200-mL ground-in stopper Erlenmeyer flask
(powdery)
← 100 mL of water
Extraction Stir to mix, 10 minites
Filtrate
1.00 g analytical
sample 100-mL volumetric flask
(fluid)
← 50 mL - 70 mL of water
Extraction Shake to mix
← Water (up to the marked line)
Filtrate
Figure 1-2 Flow sheet for niyrous acid in fertilizers
(Extraction procedure (4.1.2))
490
Testing Methods for Fertilizers (2018)
Filtrate
pH adjustment
Sodium hydroxide solution (5 mg/mL)
(pH 5 - pH 7)
Sample
491
Testing Methods for Fertilizers (2018)
Reference: HPLC chromatogram of nitrous acid and ammonium thiocyanate are shown
below.
←1
2
↓
Retention time
1→
2
↓
Retention time
492
Testing Methods for Fertilizers (2018)
Comment 1 Biuret is sold by FUJIFILM Wako Pure Chemical Co., Ltd., Kanto Chemical Co., Inc.
and Tokyo Chemical Industry Co., Ltd.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph (HPLC): HPLC specified in JIS K 0124 that
satisfies following requirements.
1) Column: A 7.5-mm inner diameter 100-mm long stainless steel column tube filled with
weak acid ion-exchange resin.
2) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45ºC.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 190 nm.
b) Magnetic stirrer:
c) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g - 10000
493
Testing Methods for Fertilizers (2018)
× g.
Comment 2 A column is sold under the production name Asahipak ES-502C 7C.
Note (4) If there is a possibility that the biuret nitrogen (B-N) concentration in the sample
solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of supernatant with water.
(5) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(6) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Note (7) If there is a possibility that the biuret nitrogen (B-N) concentration in the sample
solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of precisely adjusted solution with water.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions is shown below. Set up the measurement conditions
considering it:
1) Column: A weak acid ion-exchange resin column (7.5-mm inner diameter, 100-mm long,
5-µm - 10-µm particle diameter)
2) Column bath temperature: 40 °C
3) Eluent (1): Dissolve 3.92 g of potassium dihydrogen phosphate and 0.12 g of phosphoric acid
in water to make 1000 mL. Filter with a membrane filter (aperture diameter: no more than
494
Testing Methods for Fertilizers (2018)
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the biuret nitrogen (B-N) content from the peak height using the calibration curve to
calculate the biuret nitrogen (B-N) in the analytical sample.
Comment 5 This testing method enables the simultaneous measurement of biuret nitrogen (B-N),
dicyandiamide nitrogen (Dd-N), urea nitrogen (U-N), guanidine nitrogen (Gd-N) and
a guanylurea nitrogen standard solution (GU-N). In this case, mix a predetermined
amount of biuret nitrogen (B-N 2 mg/mL), an urea nitrogen standard solution (U-N 2
mg/mL), a dicyandiamide nitrogen standard solution (Dd-N 2 mg/mL), a guanidine
nitrogen standard solution (Gd-N 2 mg/mL) and a guanylurea nitrogen standard
solution (GU-N 2 mg/mL), dilute with water to prepare a mixture standard solution
(200 μg/mL) (1) and use it instead of (2) e) a biuret nitrogen standard solution (B-N
200 μg/mL). After that, conduct the same procedure in (4.2) b) to calculate the
respective concentrations of materials subjected to measurement in the analytical
sample.
Comment 6 Additive recovery testing was conducted using one brand of an acetaldehyde
condensed urea fertilizer, a compound fertilizer, a blended fertilizer, a fluid
compound fertilizer and a home garden-use mixed fertilizer respectively. As a result,
the mean recovery at additive level of 0.2 % (mass fraction) and 0.1 % (mass
fraction) and 0.02 % (mass fraction) were 87.0 % - 95.1 %, 90.6 % - 101.1 % and
91.2 % - 105 5 % respectively.
The results of the repeatability tests on different days using a blended fertilizer, a
compound fertilizer and a home garden-use mixed fertilizer to evaluate precision
were analyzed by the one-way analysis of variance. Table 1 shows the calculation
results of intermediate precision and repeatability. Table 2 shows results and analysis
results from a collaborative study for testing method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.005 % (mass fraction).
495
Testing Methods for Fertilizers (2018)
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Blended fertilizer 5 0.204 0.0006 0.3 0.0017 0.9
Compound fertilzier 5 0.0969 0.0006 0.6 0.0016 1.6
Home garden-use
5 0.0103 0.0001 0.9 0.0001 0.9
mixed fertilizer
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
References
1) Masahiro ECHI, Yasuharu KIMURA and Yuji SHIRAI: Determination of Urea Nitrogen,
Biuret Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea
Nitrogen in Fertilizer by High-Performance Liquid Chromatography: A
Single-Laboratory Validation, Research Report of Fertilizer, Vol.10, p. 72 - 85 (2017)
2) Norio FUNAKI and Yasuharu KIMURA: Determination of Urea Nitrogen, Biuret
Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea Nitrogen in
Fertilizer by High-Performance Liquid Chromatography (HPLC): A Collaborative Study,
Research Report of Fertilizer, Vol.10, p. 86 - 100 (2017)
496
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for biuret nitrogen in fertilizers is shown
below:
1.00 g
200-mL ground-in stopper Erlenmeyer flask
analytical sample (powdery)
← 100 mL of water
Extraction Stir to mix, 10 minites
Stand to cool
1.00 g
100-mL volumetric flask
analytical sample (fluid)
←About 50 mL of water
Extraction Shake to mix
← Water (up to the marked line)
Ground-in stopper centrifugal precipitate tube,
Centrifugal separation
8000 ×g - 10000 ×g , 5 minutes
497
Testing Methods for Fertilizers (2018)
Reference: The chromatogram of the standard solution of biuret nitrogen for the calibration
curve preparation is shown below.
Reference diagram HPLC chromatogram of the mixture standard solutions for the calibration
curve preparation (10 mg/L for each)
Peak name
(1) Urea nitrogen (2) Biuret nitrogen (3) Dicyandiamide nitrogen
(4) Guanidine nitrogen (5) Guanylurea nitrogen
Measurement conditions for HPLC
Column: Asahipak ES-502C 7C (7.5-mm inner diameter, 100-mm long, 9-μm
particle diameter)
Other conditions are according to the example of HPLC measurement conditions
in (4.2) a)
498
Testing Methods for Fertilizers (2018)
5.11 Titanium
5.11.a ICP Optical Emission Spectrometry (1)
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type D and its
symbol is 5.11.a-2017 or Ti.a-1.
Pretreat an analytical sample with nitric acid - sulfuric acid - perchloric acid, introduce it to an ICP
Optical Emission Spectrometer (“ICP-OES”) and measure the emission with titanium at a
wavelength of 334.941 nm to obtain the titanium (Ti) in an analytical sample. In addition, the
performance of this testing method is shown in Comment 3.
Comment 1 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in h) applies to a horizontal observation
mode. While an axial observation mode can observe a measurement component of
low-level concentrations, it cannot gain the linearity of a calibration curve in the
range of high-level concentrations. Therefore, in case of using the ICP-OES of an
axial observation mode, it is recommended to prepare a titanium standard solution
for the calibration curve in the concentration range which is suitable to a device used.
b) Add about 10 mL of nitric acid and about 5 mL of sulfuric acid, cover the tall beaker with a
watch glass, and leave at rest overnight.
c) Heat mildly on a hot plate or sand bath at 170 ºC - 220 ºC for no less than 30 minutes. After
bubbles cease to form, set the temperature of the hot plate or sand bath to no less than 300 ºC,
and heat until nitroxide (yellow-brown smoke) is no longer generated (2).
d) After standing to cool, add about 5 mL of perchloric acid.
e) Cover the tall beaker with a watch glass, and heat on a hot plate or sand bath at no less than
300 ºC for 2 - 3 hours to digest (3).
f) Slightly move the watch glass (4), and keep on heating on the hot plate or sand bath to
concentrate until the liquid volume becomes no more than 2 mL.
g) After standing to cool, add about 5 mL of hydrochloric acid (1+10) and about 20 mL of water,
cover the tall beaker with a watch glass and heat mildly to dissolve.
h) After standing to cool, transfer to a 100-mL volumetric flask with water, add water up to the
marked line, and filter with Type 3 filter paper to make a sample solution.
i) As a blank test, conduct the procedures in b) - h) using another tall beaker to prepare the
blank test solution.
Note (2) Oxidation of organic matters by perchloric acid progresses extremely rapidly and
explosively. For that reason, add perchloric acid after fully degrading organic matters
with nitric acid to avoid danger.
(3) When the white smoke of perchloric acid is generated, if the solution is colored such
as black-brown or brown, stop heating immediately, and after standing to cool, add
nitric acid, and heat again to degrade remaining organic matters.
(4) The watch glass can be removed.
Comment 2 When the analytical sample solidifies in the procedure in (4.1) b), moisten the
analytical sample with a small amount of water as necessary in advance.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 334.941 nm
b) Calibration curve preparation
1) Spray the titanium standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 334.941 nm.
2) Prepare a curve for the relationship between the titanium concentration and the indicated
value of the titanium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.01 mg - 2 mg as
titanium) to a 100-mL volumetric flask.
2) Add 25 mL of hydrochloric acid (1+23) to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the titanium content from the calibration curve, and calculate titanium (Ti) in the
analytical sample.
500
Testing Methods for Fertilizers (2018)
Comment 3 Additive recovery testing was conducted using an autoclaved lightweight concrete
powdery fertilizer, a mixed phosphate fertilizer, a compound fertilizer, a slag
manganese fertilizer, a solid fertilizer, a fluid compound fertilizer, a mixed compost
compound fertilizer and a blended fertilizer respectively. As a result, the mean
recovery at additive level of 0.01 % (mass fraction) - 0.5 % (mass fraction) were
92.9 % - 99.5 % respectively.
The results of the repeatability tests on different days using a mixed phosphate
fertilizer and a compound fertilizer to evaluate precision were analyzed by the
one-way analysis of variance. Table 1 shows the calculation results of intermediate
precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.001 % (mass fraction).
References
1) Keisuke AOYAMA: Determination of Titanium in Fertilizer using Inductively Coupled
Plasma Optical Emission Spectrometry (ICP-OES) , Research Report of Fertilizer,
Vol.10, p. 29 - 40 (2017)
501
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for titanium in fertilizers is shown below:
1.00 - 2.00g
200-mL - 300-mL tall beaker.
analytical sample
← A small amount of water, moisten the analytical sample (as appropriate)
← About 10 mL of nitric acid
← About 5 mL of sulfuric acid
Leave at rest overnight Cover with a watch glass
Sample solution
Sample solution
Aliquot
100-mL volumetric flask
(Predetermined
← About 25 mL of hydrochloric acid (1+23)
← Water (up to the marked line)
Measurement ICP Optical Emission Spectrometer (334.941 nm)
Figure 2 Flow sheet for titanium in fertilizers (Measurement procedure)
502
Testing Methods for Fertilizers (2018)
Comment 1 There are two modes to observe emission from an ICP-OES, a horizontal observation
mode and an axial observation mode. The range of the concentration of the standard
solution for the calibration preparation curve in f) applies to a horizontal observation
mode. While an axial observation mode can observe a measurement component of
low-level concentrations, it cannot gain the linearity of a calibration curve in the
range of high-level concentrations. Therefore, in case of using the ICP-OES of an
axial observation mode, it is recommended to prepare a titanium standard solution
for the calibration curve in the concentration range which is suitable to a device used.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0116 and as shown below.
Specific measurement procedures are according to the operation method of the ICP Optical
Emission Spectrometry used in measurement.
a) Measurement conditions for the ICP Atomic Emission Spectrometer: Set up the
measurement conditions for the ICP Atomic Emission Spectrometer considering the
following:
Analytical line wavelength: 334.941 nm
b) Calibration curve preparation
1) Spray the titanium standard solution for the calibration curve preparation and blank test
solution for the calibration curve preparation into inductively coupled plasma, and read the
indicated value at a wavelength of 334.941 nm.
2) Prepare a curve for the relationship between the titanium concentration and the indicated
value of the titanium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Subject the sample solution to the same procedure as in b) 1) to read the indicated value.
2) Obtain the titanium content from the calibration curve, and calculate titanium (Ti) in the
analytical sample.
Comment 2 Recovery testing was conducted to evaluate trueness using slag silicate fertilizers (2
samples). As a result, the average rate of recovery at additive level of 0.1 % (mass
fraction) - 0.2 % (mass fraction) are 95.1 % - 98.2 %.
The results of the repeatability tests on different days using slag silicate fertilizers (2
samples) to evaluate precision were analyzed by the one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.02 % (mass fraction).
504
Testing Methods for Fertilizers (2018)
References
1) Keisuke AOYAMA: Determination of Titanium in Fertilizer using Inductively Coupled
Plasma Optical Emission Spectrometry (ICP-OES) , Research Report of Fertilizer,
Vol.10, p. 29 - 40 (2017)
505
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for titanium in fertilizers is shown below:
1.00 g
200-mL tall beaker.
analytical sample
← About 10 g of ammonium hydrogensulfate
Heat on a hot plate at no less than 350 ºC
Melting Make the melted ammonium hydrogensulfate
completely come into contact with the analytical sample
Sample solution
Sample solution
506
Testing Methods for Fertilizers (2018)
References
1) National Institute for Agro-Environmental Sciences, the Ministry of Agriculture,
Forestry and Fisheries: The Official Methods of Analysis of Fertilizers 1992, p.78 - 79,
Japan Fertilizers Analysis Association, Tokyo (1992)
2) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p. 196 - 197, Yokendo, Tokyo (1988)
507
Testing Methods for Fertilizers (2018)
Comment 2 A column is sold under production names such as Hibar LiChrosorb NH2, Inertsil
508
Testing Methods for Fertilizers (2018)
Note (3) Add methanol immediately as the determined value becomes higher than usual if it is
left in air.
(4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 3 Instead of the procedures in (4.1) c) - e), it is allowed to filter with a membrane filter
(aperture diameter: no more than 0.5-µm) made of PTFE and the filtrate can be the
sample solution.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter) to which amino or amino propyl chemically bonds.
2) Column bath temperature: 30 °C - 40 °C
3) Eluent: Acetonitrile - methanol (6+1)
4) Flow rate: 1 mL/min
5) Detection unit: An absorptiometric detector, measurement wavelength: 215 nm
b) Calibration curve preparation
1) Inject 10 µL of respective dicyandiamide standard solutions for the calibration curve
preparation to an HPLC, record a chromatogram at wavelength 215 nm, and obtain the peak
area or the height.
2) Prepare a curve for the relationship between the concentration and the peak area or the height
at wavelength 215 nm of the respective dicyandiamide standard solutions for the calibration
curve preparation.
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain dicyandiamide (Dd) content from the calibration curve to calculate the concentration
of dicyandiamide (Dd) in the analytical sample.
3) Calculate the dicyandiamide nitrogen (Dd-N) by the following formula.
509
Testing Methods for Fertilizers (2018)
Comment 4 Recovery testing was conducted using nitrolime (3 samples) and a blended fertilizer
containing nitrolime (2 samples), as a result, the recovery of dicyandiamide at
concentration level of 6 % (mass fraction) and 0.6 % (mass fraction) was 94.9 % -
105.1 % and 95.6 % - 103.5 % respectively.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of this testing method is about
0.01 % (mass fraction).
References
1) Masakazu SAIKI and Miyuki ASAO: Validation of High-Performance Liquid
Chromatography for Determination of Dicyandiamide in Nitrolime, Research Report of
Fertilizers, Vol.2, p. 25 - 31 (2009)
2) Masakazu SAIKI and Masayuki YOSHIMOTO: Determination of Dicyandiamide in
Nitrolime by High-Performance Liquid Chromatography: A Collaborative Study,
Research Report of Fertilizers, Vol.2, p. 32 - 37 (2009)
(5) Flow sheet for dicyandiamide nitrogen: The flow sheet for dicyandiamide nitrogen in
nitrolime and fertilizers containing nitrolime is shown below:
510
Testing Methods for Fertilizers (2018)
1.00 g
200-mL - 300-mL ground-in stopper Erlenmeyer flask
analytical sample
← 100 mL of methanol
Extraction Shaking, 10 minutes
0.11 0.11
Dd 8.458
Dd 8.541
0.10 0.10
0.09 0.09
0.08 0.08
0.07 0.07
0.06 0.06
AU
AU
0.05 0.05
0.04 0.04
0.03 0.03
0.02 0.02
0.01 0.01
0.00 0.00
0 1 2 3 4 5 6 7 8 9 10 11 0 1 2 3 4 5 6 7 8 9 10 11
min min
1) Dicyandiamide standard solution (the equivalents of 100 ng (10 µg/ mL, 10 µL) of
dicyandiamide)
2) Sample solution (nitrolime)
511
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph (HPLC): HPLC specified in JIS K 0124 that
satisfies following requirements.
1) Column: A 7.5-mm inner diameter 100-mm long stainless steel column tube filled with
weak acid ion-exchange resin.
2) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45ºC.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 190 nm.
b) Magnetic stirrer:
c) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g - 10000
× g.
512
Testing Methods for Fertilizers (2018)
Comment 2 A column is sold under the production name Asahipak ES-502C 7C.
Note (3) If there is a possibility that the dicyandiamide nitrogen (Dd-N) concentration in the
sample solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of supernatant with water.
(4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Note (6) If there is a possibility that the dicyandiamide nitrogen (Dd-N) concentration in the
sample solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of solution with water.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph: An example
of measurement conditions is shown below. Set up the measurement conditions considering it:
1) Column: A weak acid ion-exchange resin column (7.5-mm inner diameter, 100-mm long,
5-µm - 10-µm particle diameter)
2) Column bath temperature: 40 °C
3) Eluent (1): Dissolve 3.92 g of potassium dihydrogenphosphate and 0.12 g of phosphoric acid
in water to make 1000 mL. Filter with a membrane filter (aperture diameter: no more than
0.5-µm) made of hydrophilic PTFE.
4) Flow rate: 0.6 mL/min
513
Testing Methods for Fertilizers (2018)
5) Injection volume: 10 µL
6) Detection unit: An absorptiometric detector, measurement wavelength: 190 nm
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the dicyandiamide nitrogen (Dd-N) content from the peak height using the calibration
curve to calculate the dicyandiamide nitrogen (Dd-N) in the analytical sample.
Comment 5 This testing method enables the simultaneous measurement of biuret nitrogen (B-N),
urea nitrogen (U-N), dicyandiamide nitrogen (Dd-N), guanidine urea (Gd-N) and
guanylurea nitrogen (GU-N). In that case, see 5.10.a Comment 5.
Comment 6 Additive recovery testing was conducted using one brand of an acetaldehyde
condensed urea fertilizer, a compound fertilizer, a blended fertilizer, a fluid
compound fertilizer and a home garden-use mixed fertilizer respectively. As a result,
the mean recovery at additive level of 3 % (mass fraction) and 1.5 % (mass fraction)
and 0.3 % (mass fraction) were 96.3 % - 96.3 %, 94.5 % - 99.7 % and 88.9 % -
100.6 % respectively.
The results of the repeatability tests on different days using a blended fertilizer, a
compound fertilizer and a home garden-use mixed fertilizer to evaluate precision
were analyzed by the one-way analysis of variance. Table 1 shows the calculation
results of intermediate precision and repeatability. Table 2 shows results and analysis
results from a collaborative study for testing method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.01 % (mass fraction).
514
Testing Methods for Fertilizers (2018)
References
1) Masahiro ECHI, Yasuharu KIMURA and Yuji SHIRAI: Determination of Urea Nitrogen,
Biuret Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea
Nitrogen in Fertilizer by High-Performance Liquid Chromatography: A
Single-Laboratory Validation, Research Report of Fertilizer, Vol.10, p. 72 - 85 (2017)
2) Norio FUNAKI and Yasuharu KIMURA: Determination of Urea Nitrogen, Biuret
Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea Nitrogen in
Fertilizer by High-Performance Liquid Chromatography (HPLC): A Collaborative Study,
Research Report of Fertilizer, Vol.10, p. 86 - 100 (2017)
515
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for dicyandiamide nitrogen in fertilizers is
shown below:
Stand still
1.00 g
100-mL volumetric flask
analytical sample (fluid)
←About 50 mL of water
Extraction Shake to mix
← Water (up to the marked line)
Ground-in stopper centrifugal precipitate tube,
Centrifugal separation
8000 ×g - 10000 ×g, 5 minutes
516
Testing Methods for Fertilizers (2018)
Reference diagram HPLC chromatogram of the mixture standard solutions (10 mg/L for
each) for calibration curve preparation
Peak name
(1) Urea nitrogen (2) Biuret nitrogen (3) Dicyandiamide nitrogen
(4) Guanidine nitrogen (5) Guanylurea nitrogen
Measurement conditions for HPLC
Column: Asahipak ES-502C 7C (7.5-mm inner diameter, 100-mm long, 9-μm
particle diameter)
Other conditions are according to the example of HPLC measurement conditions
in (4.2) a)
517
Testing Methods for Fertilizers (2018)
6.2 Chlorine
6.2.a Ion Chromatography
(1) Summary
The testing method is applicable to potassium sulfate, potassium bicarbonate, magnesium potassium
sulfate, fish cakes powder, fish cakes and compost. This testing method is classified as Type D and
its symbol is 6.2.a-2017 or Cl.a-1.
Add water to an analytical sample to extract chloride ions, introduce them to an Ion Chromatograph
(IC) to isolate it with an ion exchange column, and then measure with an electric conductivity
detector to obtain chlorine (Cl) in an analytical sample. In addition, the performance of this testing
method is shown in Comment 3.
Comment 1 Instead of the chloride ion standard solution in (2), a chloride ion standard solution
for the calibration curve preparation can be prepared by using a chloride ion standard
solution (Cl- 0.1 mg/mL) traceable to National Metrology.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Magnetic stirrer:
b) Centrifugal separator: A centrifugal separator that can work at 1700 × g.
c) Ion Chromatograph (IC): IC specified in JIS K 0127 that satisfies following requirements.
1) Column: In case of a suppressor method, a 4-mm inner diameter 250-mm long 5-μm particle
diameter column tube filled with poly vinyl alcohol porous particles, to which quaternary
518
Testing Methods for Fertilizers (2018)
Note (4) A column is commercially sold under the name Shodex IC SI-52 4E, etc.
(5) A column is commercially sold under the name Shodex IC NI-424, etc.
Note (6) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
(7) In case of exceeding the calibration curve, dilute by a factor of more than 20.
Comment 2 Instead of the procedures in (4.1.1) c) and d), it is allowed to filter with Type 3 filter
paper and the filtrate can be the extract.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0127 and as shown below.
Specific measurement procedures are according to the operation method of an Ion
Chromatograph (IC) used in measurement.
a) Measurement conditions for Ion Chromatograph (IC): An example of measurement
conditions is shown below. Set up the measurement conditions considering it:
c) Sample measurement
1) Subject 20 μL of sample solution to the same procedure as in b) 1)
2) Obtain the chloride ion concentration from calibration curve by peak area to calculate the
chlorine (Cl) in the analytical sample.
Comment 3 Additive recovery testing was conducted with a suppressor method using samples
that 1.8 % (mass farction) – 33.4 % (mass fraction) of sodium chloride as chlorine is
added to potassium sulfate, magnesium potassium sulfate, potassium bicarbonate,
cow dung compost and fish cakes powder. As a result, the mean recovery at the
chlorine additive level of 33.4 % (mass fraction), 10 % (mass farction) - 13.4 %
(mass fraction) and 1.8 % (mass farction) - 9.1 % (mass fraction) are 100.8 %,
98.6 % - 101.1 % and 96.2 % - 103.2 % respectively. With a non-suppressor method,
they are 100.2 %, 96.4 % - 97.2 % and 93.3 % - 101.4 %.
The results of the repeatability tests on different days using potassium sulfate,
magnesium potassium sulfate, potassium bicarbonate, cow dung compost and fish
cakes powder to evaluate precision were analyzed by one-way analysis of variance.
Table 1 shows the calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.1 % (mass fraction).
520
Testing Methods for Fertilizers (2018)
References
1) Satoko SAKAIDA, Mariko FUJITA and Yuji SHIRAI: Determination of Chloride in
Fertilizers by Ion Chromatography (IC), Research Report of Fertilizers, Vol.8, p. 50 - 60
(2015)
(5) Flow sheet for testing method: The flow sheet for chlorine in fertilizers is shown below:
1.00 g
200-mL ground-in stopper Erlenmeyer flask
analytical sample
← 100 mL of water
Extraction Magnetic stirrer, 10 minutes
Sample solution
Measurement Ion Chromatograph
Figure Flow sheet for chlorine in fertilizers
521
Testing Methods for Fertilizers (2018)
Cl- Cl-
Cl- Cl-
(C) Chromatograph of fish cakes powder (D) Chromatograph of fish cakes powder
(Suppressor method) (Non-suppressor method)
522
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Magnetic stirrer:
b) pH test paper: A pH test paper infiltrated with indicator and dried, which can measure the
value from pH 1 to pH 11 and a color change chart with the pH interval value 1 is attached.
Comment 1 pH test paper is sold under the name UNIV Test Paper, etc.
flask.
b) Add 100 mL of water and stir it by using a magnetic stirrer for about 10 minutes.
c) Filter with Type 3 filter paper to make a sample solution.
V4: Volume (mL) of the 0.1 mol/L silver nitrate solution required for titration
C: Set concentration (0.1 mol/L) of 0.1 mol/L silver nitrate solution
f: Factor of 0.1 mol/L silver nitrate solution
V5: Liquid volume of water (100 mL) subjected to extraction in (4.1) b).
V6: Transferred volume (25 mL) of the sample solution subjected to titration in
(4.2) a).
W2: Mass (g) of the analytical sample
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.199- 201, Yokendo, Tokyo (1988)
524
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for chlorine in potassium sulfate is shown
below:
Sample solution
Sample solution
525
Testing Methods for Fertilizers (2018)
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL
- 300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
and titrate with a 0.1 mol/L -0.2 mol/L sodium hydroxide solution until the color of the
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
(1).
Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution equivalent to 1 mL of
0.25 mol/L sulfuric acid
= V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
e) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
f) Urease: A reagent which completely digests 0.25 g of urea by 0.5 g of urease.
g) Sodium hydroxide solution (5 g/L) (1): Dissolve 5 g of sodium hydroxide specified in JIS K
8576 in water to make 1000 mL.
h) Hydrochloric acid: A JIS Guaranteed Reagent specified in JIS K 8180 or a reagent of
equivalent quality.
i) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, add water to make 100 mL.
j) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
k) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
l) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
m) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
n) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100 mL),
add equal volume of bromocresol green solution (0.5 g/100 mL).
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
527
Testing Methods for Fertilizers (2018)
Comment 2 0.25 mol/L sulfuric acid in (2) d) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
Comment 3 Refined products of urease made from sword beans are commercially sold. The
degree of activation may degrade even if stored in a refrigerator. Therefore, it is
recommended to test similarly and confirm activation by using urea specified in JIS
K 8731 in advance.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Extractor: Rotary shaker or magnetic stirrer as described below
aa) Rotary shaker: A rotary shaker that can rotate a 500-mL volumetric flask upside down at 30 -
40 revolutions/min.
ab) Magnetic stirrer:
b) Steam distillation apparatus
c) Distillation flask: A Kjeldahl flask or round bottom flask that can be connected to a steam
distillation apparatus.
d) Water bath: Water bath that can be adjusted to 40 ºC - 45 ºC.
Comment 4 In the procedure in (4.1.1) a), it is also allowed to weigh 2.50 g of the analytical
sample and transfer to a 250-mL volumetric flask.
Comment 5 The procedure in (4.1.1) is the same as the procedure in (4.1.1.1) in 4.2.4.a.
Comment 6 The procedure in (4.1.2) is the same as the procedure in (4.1.1) in 6.3.c.
(4.3) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (11) of 0.25 mol/L sulfuric acid to an acceptor (12), add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (11) of boric acid solution (40 g/L)
to an acceptor (12), add a few drops of methyl red−bromocresol green mixture solution, and
connect this acceptor to a steam distillation apparatus.
b) Add 2 g - 3 g of magnesium oxide to a distillation flask containing the sample solution (13),
and connect this distillation flask to the steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
f) Subject the undigested test solution to the same procedure as in a) - e) to obtain the distillate
from the undigested test solution.
g) Subject the blank test solution to the same procedure as in a) - e) to obtain the distillate from
the blank test solution.
Note (11) 5 mL - 20 mL
(12) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(13) Add a small amount of silicone oil as necessary.
Comment 7 Immediately conduct the procedure in (4.3) b) not to emit ammonia gas in a vessel.
V6: Volume (mL) of 0.25 mol/L sulfuric acid transferred to the acceptor in (4.3) a)
V7: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration in (4.4.1) a)
V8: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration in (4.4.1) b)
V9: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration in (4.4.1) c)
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
f1: Factor of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
V10: Predetermined volume (mL) of the extract in (4.1.1) c) or additive volume of
water in (4.1.2) b)
V11: Transferred amount (mL) of the extract subjected to hydrolysis in (4.2) a)
W2: Mass (g) of the analytical sample in (4.1.1) a) or (4.1.2) a)
V12: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration in (4.4.2) a)
V13: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration in (4.4.2) b)
V14: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration in (4.4.2) c)
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
f2: Factor of 0.25 mol/L sulfuric acid
V10: Predetermined volume (mL) of the extract in (4.1.1) c) or additive volume of
water (mL) in (4.1.2) b)
V11: Transferred amount (mL) of the extract subjected to hydrolysis in (4.2) a)
W2: Mass (g) of the analytical sample in (4.1.1) a) or (4.1.2) a)
Note (14) The endpoint is reached when the color changes from green to light red.
Comment 8 If it is hard to confirm the endpoint due to the carbon dioxide resulting from
carbonate in the extract when magnesium oxide is used, it is recommended to boil
the extract for 1-2 minute(s) after distilling and cool, and then titrate.
Comment 9 The titration procedures in (2) a) Standardization, (2) d) Standardization and (4.4)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
Comment 10 An example of the additive volume and titter of urease is shown below.
If the urea content can be estimated, calculate the volume of urea in the transferred
volume of the extract in (4.2) a) by the following formula.
If the urea content can not be estimated, subtract ammoniacal nitrogen and nitrate
nitrogen from the total nitrogen of the permissible content or the display component
content of the urea nitrogen in a urea compound to obtain nitrogen content. The obtained
nitrogen content is calculated as around the maximum content of urea nitrogen (U-N). In
this case, calculate the volume of urea in the transferred volume of the extract in (4.1.1) or
(4.2) a) after the procedures in (4.1.1) by the following formula.
D2: Estimated urea nitrogen (U-N) content (% (mass fraction)) in the analytical
sample
V10: Predetermined volume (mL) of the extract in (4.1.1) c) or (4.1.2) c)
V11: Transferred amount (mL) of the extract subjected to hydrolysis in (4.2) a)
W2: Mass (g) of the analytical sample in (4.1.1) a) or (4.1.2) a)
The specification of urease shows that “No more than 0.5 g of urease completely digests
0.25 g of urea”. Therefore about 2 mg of urease is required to digest 1 mg of urea.
Assuming that the estimated urea content or the calculated urea content in the transferred
volume of the extract is about 43 mg (about 20 mg as urea nitrogen), about 86 mg of
urease is required. When adding 0.2 g of urease (potency 130 unit - 150 unit), the extract
containing the equivalents of 10 mg - 100 mg as urea nitrogen (U-N) can be hydrolyzed.
In addition, when about 20 mg is transferred as urea nitrogen, the differences between
the titer of the distillate from the sample solution ((4.4.1) a) or (4.4.2) a)) and the titer of
the distillate from the undigested test solution ((4.4.1) b) or (4.4.2) b)) are estimated as
follows: About 14 mL, when a 0.1 mol/L sodium hydroxide solution is used as a titrant,
about 7 mL, when 0.2 mol/L sodium hydroxide is used as a titrant and about 3 mL, when
0.25 mol/L sulfuric acid is used as a titrant.
Comment 11 Additive recovery testing was conducted to evaluate trueness using samples
prepared by adding urea to sulfur, a compound fertilizer, isobuttyraldehude
condensed urea, a mixed fertilizer containing formaldehyde processed urea (one
sample for each). As a result, the average rate of recovery at additive level of 1.58 %
(mass fraction) - 39.96 % (mass fraction) are 98.7 % to 103.7 %
The results of the repeatability tests on different days using urea and a compound
fertilizer containing UF (one sample for each), to evaluate precision were analyzed
by one-way analysis of variance. Table 1 shows the calculation results of
intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.4 % (mass fraction).
531
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.56- 59, Yokendo, Tokyo (1988)
2) Toshifumi FUJITA, Kimie KATO, Toyokazu NIIMI, Yasuharu KIMURA, Kohei ITO
and Yuji SHIRAI: Determination Method for Urea Nitrogen in Fertilizer by Urease,
Research Report of Fertilizer, Vol.10, p. 195 - 207 (2017)
(5) Flow sheet for urea nitrogen testing method: The flow sheets for urea nitrogen test method
in fertilizers are shown below:
5.00 g
Weigh into a 500-mL volumetric flask
Analytical sample
← About 400 mL of water
Shaking to mix Rotary shaker (30 - 40 revolutions/min), 1 hour
← Water (up to the marked line)
Filtration Type 3 filter paper
Extract
Figure 1-1 Flow sheet for urea nitrogen
(Extraction procedure (4.1.1))
532
Testing Methods for Fertilizers (2018)
1.00 g
Weigh into a 200-mL ground-in stopper Erlenmeyer flask
Analytical sample
← 100 mL of water
Shaking to mix Magnetic stirrer, 10 minures
Extract
Figure 1-1 Flow sheet for urea nitrogen
(Extraction procedure (4.1.2))
Extract
Aliquot
(predetermined 300-mL digestion flask
volume)
←Water (Until the solution reaches about 50 mL)
← A few drops of methyl red solution (0.1 g/100 mL)
← Hydrochloric acid (1+200) or sodium hydroxide solution (5 g/L)
(until the color of the solution becomes light yellowish red)
← Urease (a sufficient amount of urease to hydrolyze transferred urea)
Heating 40 ºC - 50 ºC, 1 hour
Standing to cool
Sample solution
← 2 g - 3 g of magnesium oxide
Receiver: 200-mL - 300-mL Erlenmeyer flask or
200-mL - 300-mL beaker. A predetermined amount of 0.25 mol/L sulfuric acid
Steam distillation
and a few drops of methyl red-methylene blue mixture solution,
apparatus
or boric acid solution (40 g/L),
several drops of methyl red - bromocresol green mixture solution
533
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph (HPLC): HPLC specified in JIS K 0124 that
satisfies following requirements.
1) Column: A 7.5-mm inner diameter 100-mm long stainless steel column tube filled with
weak acid ion-exchange resin.
2) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45ºC.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 190 nm.
b) Magnetic stirrer:
c) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g - 10000
× g.
Comment 1 A column is sold under the production name Asahipak ES-502C 7C.
flask.
b) Add 100 mL of water and stir it by using a magnetic stirrer for about 10 minutes.
c) After allowing to stand still, transfer a supernatant solution (2) to a 1.5-mL ground-in stopper
centrifugal precipitate tube (3).
d) Centrifuge it at 8000 × g - 10000 × g centrifugal force for about five minutes (4) and use the
supernatant as the extract.
Note (2) If there is a possibility that the urea nitrogen (U-N) concentration in the sample
solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of supernatant with water.
(3) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(4) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Note (5) If there is a possibility that the urea nitrogen (U-N) concentration in the sample solution
will exceed the maximum limit of the calibration curve, dilute a predetermined amount
of supernatant with water.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions is shown below. Set up the measurement conditions
considering it:
1) Column: A weak acid ion-exchange resin column (7.5-mm inner diameter, 100-mm long,
5-µm - 10-µm particle diameter)
2) Column bath temperature: 40 °C
3) Eluent (1): Dissolve 3.92 g of potassium dihydrogenphosphate and 0.12 g of phosphoric acid
in water to make 1000 mL. Filter with a membrane filter (aperture diameter: no more than
0.5-µm) made of hydrophilic PTFE.
4) Flow rate: 0.6 mL/min
5) Injection volume: 10 µL
6) Detection unit: An absorptiometric detector, measurement wavelength: 190 nm
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the urea nitrogen (U-N) content from the peak height using the calibration curve to
calculate the urea nitrogen (U-N) in the analytical sample.
Comment 4 This testing method enables the simultaneous measurement of biuret nitrogen (B-N),
urea nitrogen (U-N), dicyandiamide nitrogen (Dd-N), guanidine urea (Gd-N) and
guanylurea nitrogen (GU-N). In that case, see 5.10.a Comment 5.
When Asahipak ES-502C 7C is used as a column for HPLC to analyze urea nitrogen
(U-N) in a mixed fertilizer containing formaldehyde processed urea (UF), the peak of
the urea nitrogen (U-N) and the peak of the impurity components originating from
UF are not separated, so that urea nitrogen (U-N) cannot be measured. In this case,
the peak of the urea nitrogen (U-N) and the peak of the impurity components
originating from UF are separated by using PRP-X200 as a column for HPLC instead
to be able to analyze the urea nitrogen(U-N) in the compound fertilizer containing
UF. However, when the PRP-X200 column is used as a column for HPLC, the
simultaneous measurement of urea nitrogen (U-N) and biuret nitrogen (B-N), etc. is
impossible.
Comment 5 Additive recovery testing was conducted using one brand of an acetaldehyde
condensed urea fertilizer, a compound fertilizer, a blended fertilizer, a fluid
compound fertilizer and a home garden-use mixed fertilizer respectively. As a result,
the mean recovery at additive level of 6 % (mass fraction) and 3 % (mass fraction)
and 0.6 % (mass fraction) were 98.3 % - 102.9 %, 98.9 % - 105.2 % and 92.3 % -
99.9 % respectively.
The results of the repeatability tests on different days using a blended fertilizer, a
compound fertilizer and a home garden-use mixed fertilizer to evaluate precision
were analyzed by the one-way analysis of variance. Table 1 shows the calculation
results of intermediate precision and repeatability. Table 2 shows results and analysis
results from a collaborative study for test method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.03 % (mass fraction).
536
Testing Methods for Fertilizers (2018)
References
1) Masahiro ECHI, Yasuharu KIMURA and Yuji SHIRAI: Determination of Urea Nitrogen,
Biuret Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea
Nitrogen in Fertilizer by High-Performance Liquid Chromatography: A
Single-Laboratory Validation, Research Report of Fertilizer, Vol.10, p. 72 - 85 (2017)
2) Norio FUNAKI and Yasuharu KIMURA: Determination of Urea Nitrogen, Biuret
Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea Nitrogen in
Fertilizer by High-Performance Liquid Chromatography (HPLC): A Collaborative Study,
Research Report of Fertilizer, Vol.10, p. 86 - 100 (2017)
537
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for urea nitrogen in fertilizers is shown
below:
1.00 g analytical
sample 200-mL ground-in stopper Erlenmeyer flask
(powdery)
← 100 mL of water
Extraction Stir to mix, 10 minites
Stand still
1.00 g
100-mL volumetric flask
analytical sample (fluid)
←About 50 mL of water
Extraction Shake to mix
← Water (up to the marked line)
Ground-in stopper centrifugal precipitate tube,
Centrifugal separation
8000 ×g - 10000 ×g , 5 minutes
538
Testing Methods for Fertilizers (2018)
Reference: Chromatogram of the standard solution for calibration curve preparation of urea
nitrogen is shown below.
Reference diagram 1 HPLC chromatogram of the mixture standard solutions (10 mg/L for
each) for calibration curve preparation
Peak name
(1) Urea nitrogen (2) Biuret nitrogen (3) Dicyandiamide nitrogen
(4) Guanidine nitrogen (5) Guanylurea nitrogen
Measurement conditions for HPLC
Column: Asahipak ES-502C 7C (7.5-mm inner diameter, 100-mm long, 9-μm
particle diameter)
Other conditions are according to the example of HPLC measurement conditions
in (4.2) a)
539
Testing Methods for Fertilizers (2018)
Reference diagram 2 HPLC chromatogram of the urea nitrogen standard solutions (20 mg/L)
and a formaldehyde processed urea fertilizer for calibration curve
preparation
Peak name
(1) Urea nitrogen (2) Biuret nitrogen
Measurement conditions for HPLC
Column: PRP-X200 (4.1-mm inner diameter, 150-mm long, 10-µm particle
diameter)
Other conditions are according to the example of HPLC measurement conditions
in (4.2) a)
540
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Spectrophotometer: A spectrophotometer specified in JIS K 0115
b) Magnetic stirrer:
Comment 1 The procedure in (4.1) is the same as the procedure in (4.1.2) of 6.3.a.
Comment 2 When the determination is affected by the coloring of the sample solution of (4.1) c),
add about 0.5 g of active carbon and filter with Type 3 filter paper to make a sample
solution from which the coloring disappears.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0115 and as shown below.
Specific measurement procedures are according to the operation method of the
spectrophotometer used for the measurement.
541
Testing Methods for Fertilizers (2018)
Comment 3 Additive recovery testing was conducted using a compound fertilizer, crustacea
grade fertilizer powdery and a preparation sample. As a result, the mean recovery at
additive level of 20 % (mass fraction), 10 % (mass fraction) and 3 % (mass fraction)
were 100.0 % - 102.4 %, 100.5 % - 102.0 % and 98.0 % - 103.3 % as urea nitrogen
(U-N) respectively.
The results of the repeatability tests on different days using urea, a designated
blended fertilizer and a compound fertilizer to evaluate precision were analyzed by
the one-way analysis of variance. Table 1 shows the calculation results of
intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.2 % (mass fraction).
542
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.60- 62, Yokendo, Tokyo (1988)
(5) Flow sheet for testing method: The flow sheet for urea nitrogen in fertilizers is shown
below:
1.00 g
Analytical sample 200-mL Erlenmeyer flask
(powdery)
← 100 mL of water
Extract Stir for 10 minutes
Stand still
Sample solution
Sample solution
Aliquot
(predetermined 50-mL Volumetric flask
volume)
← 20 mL of coloring reagent solution
← Water (up to the marked line)
Leave at rest About 30 minutes
543
Testing Methods for Fertilizers (2018)
Comment 1 Guanidine sulfate is commercially sold by FUJIFILM Wako Pure Chemical Co., Ltd.,
Kanto Chemical Co., Inc. and Tokyo Chemical Industry Co., Ltd.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph (HPLC): HPLC specified in JIS K 0124 that
satisfies following requirements.
1) Column: A 7.5-mm inner diameter 100-mm long stainless steel column tube filled with
weak acid ion-exchange resin.
2) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45ºC.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 190 nm.
b) Magnetic stirrer:
c) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g - 10000
× g.
544
Testing Methods for Fertilizers (2018)
Comment 2 A column is sold under the production name Asahipak ES-502C 7C.
Note (3) If there is a possibility that the guanidine nitrogen (Gd-N) concentration in the sample
solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of supernatant with water.
(4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Note (6) If there is a possibility that the guanidine nitrogen (Gd-N) concentration in the sample
solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of supernatant with water.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions is shown below. Set up the measurement conditions
considering it:
1) Column: A weak acid ion-exchange resin column (7.5-mm inner diameter, 100-mm long,
5-µm - 10-µm particle diameter)
2) Column bath temperature: 40 °C
3) Eluent (1): Dissolve 3.92 g of potassium dihydrogenphosphate and 0.12 g of phosphoric acid
in water to make 1000 mL. Filter with a membrane filter (aperture diameter: no more than 0.5
µm) made of hydrophilic PTFE.
545
Testing Methods for Fertilizers (2018)
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the guanidine nitrogen (Gd-N) content from the peak height using the calibration curve
to calculate the guanidine nitrogen (Gd-N) in the analytical sample.
Comment 5 This testing method enables the simultaneous measurement of biuret nitrogen (B-N),
urea nitrogen (U-N), dicyandiamide nitrogen (Dd-N), guanidine urea (Gd-N) and
guanylurea nitrogen (GU-N). In that case, see 5.10.a Comment 5.
Comment 6 Additive recovery testing was conducted using a preparation sample for a guanylurea
fertilizer (one brand). As a result, the mean recovery at additive level of 3.71 %
(mass fraction), 1.85 % (mass fraction) and 0.371 % (mass fraction) were 91.2 %,
94.0 % and 100.0 % respectively.
The results of the repeatability tests on different days using a guanylurea fertilizer to
evaluate precision were analyzed by one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability. Table 2 shows results
and analysis results from a collaborative study for testing method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.02 % (mass fraction).
1) 3) 3) 3)
T (%) (%) (%) (%) (%)
Guanylurea fertilizer 5 1.81 0.01 0.8 0.04 2.0
1) The number of test days conducting a duplicate test 6) Intermediate standard deviation
2) Average (the number of test days(T ) 7) Intermediate relative standard deviation
×the number of duplicate testing(2))
3) Mass fraction
4) Repeatability standard deviation
5) Repeatability relative standard deviation
546
Testing Methods for Fertilizers (2018)
References
1) Masahiro ECHI, Yasuharu KIMURA and Yuji SHIRAI: Determination of Urea Nitrogen,
Biuret Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea
Nitrogen in Fertilizer by High-Performance Liquid Chromatography: A
Single-Laboratory Validation, Research Report of Fertilizer, Vol.10, p. 72 - 85 (2017)
2) Norio FUNAKI and Yasuharu KIMURA: Determination of Urea Nitrogen, Biuret
Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea Nitrogen in
Fertilizer by High-Performance Liquid Chromatography (HPLC): A Collaborative Study,
Research Report of Fertilizer, Vol.10, p. 86 - 100 (2017)
547
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for guanidine nitrogen in fertilizers is shown
below:
1.00 g
analytical sample 200-mL ground-in stopper Erlenmeyer flask
(powdery)
← 100 mL of water
Extraction Stir to mix, 10 minites
Stand still
1.00 g
100-mL volumetric flask
analytical sample (fluid)
←About 50 mL of water
Extraction Shake to mix
← Water (up to the marked line)
Ground-in stopper centrifugal precipitate,
Centrifugal separation
8000 ×g - 10000 ×g , 5 minutes
548
Testing Methods for Fertilizers (2018)
Reference: Chromatogram of the standard solution for calibration curve preparation of guanidine
nitrogen is shown below.
Reference diagram HPLC chromatogram of the mixture standard solutions (10 mg/L for
each)
Peak name
(1) Urea nitrogen (2) Biuret nitrogen (3) Dicyandiamide nitrogen
(4) Guanidine nitrogen (5) Guanylurea nitrogen
Measurement conditions for HPLC
Column: Asahipak ES-502C 7C (7.5-mm inner diameter, 100-mm long, 9-μm
particle diameter)
Other conditions are according to the example of HPLC measurement conditions
in (4.2) a)
549
Testing Methods for Fertilizers (2018)
b) Sulfuric acid: A JIS Guaranteed Reagent specified in JIS K 8951 or a reagent of equivalent
quality.
c) 0.25 mol/L sulfuric acid (1) (2): Add about 14 mL of sulfuric acid to a beaker containing 100
mL of water in advance, stir well, and add water to make 1000 mL.
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL
- 300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
and titrate with a 0.1 mol/L -0.2 mol/L sodium hydroxide solution until the color of the
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
550
Testing Methods for Fertilizers (2018)
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
(1). Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution equivalent to 1 mL of
0.25 mol/L sulfuric acid
=V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
d) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
e) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
f) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, add water to make 100 mL.
g) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
h) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
i) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
j) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
k) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100 mL),
add equal volume of bromocresol green solution (0.5 g/100 mL).
l) Phosphate solution: Dissolve 3.63 g of potassium dihydrogen phosphate specified in JIS K
9007 and 5.68 g of disodium hydrogen phosphate specified in JIS K 9020 in 1000 mL of
water (5). When it is used, adjust the liquid temperature at about 25 ºC (cold buffer solution).
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) c) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
551
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Rotary shaker: A rotary shaker that can rotate a 250-mL volumetric flask upside down at 30
- 40 revolutions/min.
b) Steam distillation apparatus
c) Digestion flask: A Kjeldahl flask which can be connected to a steam distillation apparatus.
Comment 3 Crush a laboratory sample with a mortar or a pestle, etc. until it completely passes
through an 850 µm aperture sieve.
Comment 4 If there is no possibility for an analytical sample to hydrolyze, it is allowed to use
water instead of a phosphoric acid solution.
Comment 5 The temperature of a solution in the procedure in (4.1) b) - d) should be no more
than 26 ºC.
(4.3) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (8) of 0.25 mol/L sulfuric acid to an acceptor, (9) add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (8) of boric acid solution (40 g/L) to
an acceptor, (9) add a few drops of methyl red−bromocresol green mixture solution, and
connect this acceptor to a steam distillation apparatus.
b) Add a proper amount of sodium hydroxide solution (200 g/L - 500 g/L) (10), and immediately
connect this digestion flask to the steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
552
Testing Methods for Fertilizers (2018)
Note (8) 5 mL - 20 mL
(9) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(10) An amount sufficient to make the solution strong alkalinity. A blue color will appear.
Cold buffer solution soluble nitrogen (% (mass fraction)) in the analytical sample
= (B × V6-V7) × C1 × f1 × (V8/V9) × (14.007/W2) × (100/1000)
= (B × V6-V7) × C1 × f1 × (V8/V9) × (1.4007/W2)
Cold buffer solution soluble nitrogen (% (mass fraction)) in the analytical sample
= V10 × C2 × 2 × f2 × (V11/V12) × (14.007/W2) × (100/1000)
= V10 × C2 × f2 × (V11/V12) × (2.8014/W2)
V10: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
f2: Factor of 0.25 mol/L sulfuric acid
V11: Predetermined volume (mL) of the extract in (4.1) c)
V12: Transferred amount (mL) of the extract subjected to the Kjeldahl method in
(4.2) a)
W2: Mass (g) of the analytical sample
Note (11) The endpoint is reached when the color changes from green to light red.
Comment 6 The titration procedures in (2) a) Standardization, (2) c) Standardization and (4.4)
553
Testing Methods for Fertilizers (2018)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.67- 68, Yokendo, Tokyo (1988)
(5) Flow sheet for cold buffer solution soluble nitrogen testing method: The flow sheets for
cold buffer solution soluble nitrogen testing method in formaldehyde processed urea
fertilizers are shown below:
1.00 g
Weigh into a 250-mL volumetric flask.
analytical sample
←150 mL phosphate solution
Shaking to mix Rotary shaker (30 - 40 revolutions/min), 1 hour
← small amount of water
Filtration Type 3 filter paper
Extract
Figure 1 Flow sheet for cold buffer solution soluble nitrogen testing method
in formaldehyde processed urea fertilizers
(Extraction procedure)
Extract
Aliquot 300-mL digestion flask
(predetermined
volume)
← 0.1 g copper (II) sulfate pentahydrate
← 5 mL sulfuric acid
Heating Gently
Standing to cool
← 1 g potassium sulfate
Heating After digestion, ignite further for 30 minutes
Standing to cool
← Water (until liquid volume reaches about 30 mL)
Cooling
Digestion solution
Figure 2 Flow sheet for cold buffer solution soluble nitrogen testing method
in formaldehyde processed urea fertilizers
(Kjeldahl method procedure)
554
Testing Methods for Fertilizers (2018)
Digestion solution
← Sodium hydroxide solution (200 g/L - 500 g/L)
Receiver: 200-mL - 300 - mL Erlenmeyer flask or beaker
A predetermined amount of 0.25 mol/L sulfuric acid and
Steam distillation
a few drops of methyl red-methylene blue mixture solution,
apparatus
or boric acid solution (40 g/L), several drops of
methyl red - bromocresol green mixture solution
Figure 3 Flow sheet for cold buffer solution soluble nitrogen testing method
in formaldehyde processed urea fertilizers
(Distillation and measurement procedure)
555
Testing Methods for Fertilizers (2018)
b) Sulfuric acid: A JIS Guaranteed Reagent specified in JIS K 8951 or a reagent of equivalent
quality.
c) 0.25 mol/L sulfuric acid (1) (2): Add about 14 mL of sulfuric acid to a beaker containing 100
mL of water in advance, stir well, and add water to make 1000 mL.
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL
556
Testing Methods for Fertilizers (2018)
- 300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
and titrate with a 0.1 mol/L -0.2 mol/L sodium hydroxide solution until the color of the
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
(1). Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution equivalent to 1 mL of
0.25 mol/L sulfuric acid
= V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
d) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
e) Catalyst (5): Mix potassium sulfate specified in JIS K 8962 and copper (II) sulfate
pentahydrate (6) specified in JIS K 8983 in the ratio of 9 to 1.
f) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
g) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, add water to make 100 mL.
h) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
i) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
j) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
k) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
l) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100 mL),
add equal volume of bromocresol green solution (0.5 g/100 mL).
m) Heat phosphate solution: Dissolve 3.63 g of potassium dihydrogen phosphate specified in
JIS K 9007 and 5.68 g of disodium hydrogen phosphate specified in JIS K 9020 in 1000 mL
of water (7). When it is used, heat until it reaches boiling point (heat buffer solution).
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) c) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Water bath: A water bath which can boil water.
b) Steam distillation apparatus
c) Digestion flask: A Kjeldahl flask which can be connected to a steam distillation apparatus.
d) Distillation flask: A Kjeldahl flask or round bottom flask that can be connected to a steam
distillation apparatus.
Comment 3 Crush a laboratory sample with a mortar or a pestle, etc. until it completely passes
through an 850 µm aperture sieve.
Note (9) When the solution has finished changing color, heat further for no less than 2 hours.
(10) When the entire sample solution volume is used in measurement, it is not necessary to
transfer it to a volumetric flask.
(4.3) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (11) of 0.25 mol/L sulfuric acid to an acceptor (12), add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (11) of boric acid solution (40 g/L)
to an acceptor (12), add a few drops of methyl red−bromocresol green mixture solution, and
558
Testing Methods for Fertilizers (2018)
Note (11) 5 mL - 20 mL
(12) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(13) An amount sufficient to make the solution strong alkalinity. A blue color will appear.
Heat buffer solution soluble nitrogen (% (mass fraction)) in the analytical sample
= (B × V6 - V7) × C1 × f1 × (V8/V9) × (14.007/W2) × (100/1000)
= (B × V6 - V7) × C1 × f1 × (V8/V9) × (1.4007/W2)
Note (14) The total nitrogen (T-N) and the nitrogen which is insoluble in a heat buffer solution
use raw data without rounding numerical value.
c) Subtract the nitrogen which is insoluble in a heat buffer solution from the separately obtained
total nitrogen (T-N) according to the method in 4.1.1 to calculate the nitrogen which is soluble
in a heat buffer solution
Heat buffer solution soluble nitrogen (% (mass fraction)) in the analytical sample
= V10 × C2 × 2 × f2 × (V11/V12) × (14.007/W2) × (100/1000)
= V10 × C2 × f2 × (V11/V12) × (2.8014/W2)
V10: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
f2: Factor of 0.25 mol/L sulfuric acid
V11: Predetermined volume (mL) of the digestion solution in (4.2) f)
V12: Transferred amount (mL) of the digestion solution subjected to distillation in
(4.3) b)
W2: Mass (g) of the analytical sample
Note (15) The endpoint is reached when the color changes from green to light red.
Comment 4 The titration procedures in (2) a) Standardization, (2) c) Standardization and (4.4)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.68- 69, Yokendo, Tokyo (1988)
(5) Flow sheet for heat buffer solution soluble nitrogen testing method: The flow sheets for
heat buffer solution soluble nitrogen testing method in methylolurea polymerization fertilizers
are shown below:
1.00 g
Weigh into a 300-mL tall beaker
analytical sample
←150 mL heat phosphate solution
Stirring Gently
Figure 1 Flow sheet for heat buffer solution soluble nitrogen testing method
in methylolurea polymerization fertilizers
(Extraction procedure)
560
Testing Methods for Fertilizers (2018)
Non-dissolved matter
Figure 2 Flow sheet for heat buffer solution soluble nitrogen testing method
in methylolurea polymerization fertilizers
( Kjeldahl method procedure)
Digestion solution
Figure 3 Flow sheet for heat buffer solution soluble nitrogen testing method
in methylolurea polymerization fertilizers
( Distillation and measurement procedure)
561
Testing Methods for Fertilizers (2018)
b) Sulfuric acid: A JIS Guaranteed Reagent specified in JIS K 8951 or a reagent of equivalent
quality.
c) 0.25 mol/L sulfuric acid (1) (2): Add about 14 mL of sulfuric acid to a beaker containing 100
562
Testing Methods for Fertilizers (2018)
mL of water in advance, stir well, and add water to make 1000 mL.
Standardization: Transfer a predetermined amount (3) of 0.25 mol/L sulfuric acid to a 200-mL
- 300-mL Erlenmeyer flask, add a few drops of methyl red−methylene blue mixture solution,
and titrate with a 0.1 mol/L - 0.2 mol/L sodium hydroxide solution until the color of the
solution becomes gray-green (4). Calculate the volume of a 0.1 mol/L - 0.2 mol/L sodium
hydroxide solution equivalent to 1 mL of 0.25 mol/L sulfuric acid by the following formula
(1)
. Or, calculate the factor of 0.25 mol/L sulfuric acid by the following formula (2):
Volume (B) of 0.1 mol/L -0.2 mol/L sodium hydroxide solution equivalent to 1 mL of 0.25
mol/L sulfuric acid
= V4/V5 ………. (1)
V4: Volume (mL) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution needed for
titration
V5: Volume (mL) of 0.25 mol/L sulfuric acid subjected to standardization
C1: Set concentration (mol/L) of 0.1 mol/L - 0.2 mol/L sodium hydroxide solution
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
d) Boric acid solution (40 g/L): Dissolve 40 g of boric acid specified in JIS K 8863 in water to
make 1000 mL.
e) Catalyst (5): Mix potassium sulfate specified in JIS K 8962 and copper (II) sulfate
pentahydrate (6) specified in JIS K 8983 in the ratio of 9 to 1.
f) Sodium hydroxide solution (200 g/L - 500 g/L) (1): Dissolve 100 g - 250 g of sodium
hydroxide specified in JIS K 8576 in water to make 500 mL.
g) Bromothymol blue solution (0.1 g/100 mL): Dissolve 0.1 g of bromothymol blue specified
in JIS K 8842 in 20 mL of ethanol (95) specified in JIS K 8102, add water to make 100 mL.
h) Methyl red solution (0.1 g/100 mL): Dissolve 0.1 g of methyl red specified in JIS K 8896 in
100 mL of ethanol (95) specified in JIS K 8102.
i) Methylene blue solution (0.1 g/100 mL): Dissolve 0.1 g of methylene blue specified in JIS K
8897 in 100 mL of ethanol (95) specified in JIS K 8102.
j) Methyl red−methylene blue mixture solution: To 2 volumes of methyl red solution (0.1
g/100 mL), add 1 volume of methylene blue solution (0.1 g/100 mL).
k) Bromocresol green solution (0.5 g/100 mL): Dissolve 0.5 g of bromocresol green specified
in JIS K 8840 in 100 mL of ethanol (95) specified in JIS K 8102.
l) Methyl red−bromocresol green mixture solution: To a methyl red solution (0.1 g/100 mL),
add equal volume of bromocresol green solution (0.5 g/100 mL).
m) Heat phosphate solution: Dissolve 1.43 g of potassium dihydrogen phosphate specified in
JIS K 9007 and 9.10 g of disodium hydrogen phosphate specified in JIS K 9020 in 1000 mL
of water (7). When it is used, heat until it reaches boiling point (heat buffer solution).
Comment 1 A 0.1 mol/L - 0.2 mol/L sodium hydroxide solution in (2) a) can be replaced with a
0.1 mol/L sodium hydroxide solution or a 0.2 mol/L sodium hydroxide solution
conforming to ISO/IEC 17025.
Comment 2 0.25 mol/L sulfuric acid in (2) c) can be replaced with 0.25 mol/L sulfuric acid
conforming to ISO/IEC 17025.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Water bath: A water bath which can boil water.
b) Steam distillation apparatus
c) Digestion flask: A Kjeldahl flask which can be connected to a steam distillation apparatus.
d) Distillation flask: A Kjeldahl flask or round bottom flask that can be connected to a steam
distillation apparatus.
Comment 3 Crush a laboratory sample with a mortar or a pestle, etc. until it completely passes
through an 850 µm aperture sieve.
Note (9) If it takes 4 minutes or more for filtrating, conduct extracting anew according to
Comment 4.
Comment 4 After conducting the procedures in (4.1.2) a) - c), add 1 g of diatomaceous earth to
stir and conduct the procedure in (4.1.2) d).
564
Testing Methods for Fertilizers (2018)
Note (10) When the solution has finished changing color, heat further for no less than 2 hours.
(11) When the entire sample solution volume is used in measurement, it is not necessary to
transfer it to a volumetric flask.
(4.3) Distillation: Conduct distillation as shown below. Specific distillation procedures are
according to the operation method of the steam distillation apparatus used in measurement.
a) Transfer a predetermined amount (12) of 0.25 mol/L sulfuric acid to an acceptor (13), add a few
drops of methyl red−methylene blue mixture solution, and connect this acceptor to a steam
distillation apparatus. Or, transfer a predetermined amount (12) of boric acid solution (40 g/L)
to an acceptor (13), add a few drops of methyl red−bromocresol green mixture solution, and
connect this acceptor to a steam distillation apparatus.
b) Transfer a predetermined amount of the digestion solution to a 300-mL distillation flask, add a
proper amount of sodium hydroxide solution (200 g/L - 500 g/L) (14), and immediately connect
this distillation flask to the steam distillation apparatus.
c) Send steam to the distillation flask to heat the solution in the distillation flask, and distill at a
distillation rate of 5 mL/min - 7 mL/min.
d) Stop distilling when the distillate has reached 120 mL - 160 mL.
e) Wash the part of the steam distillation apparatus that came in contact with the solution in the
acceptor with a small amount of water, and pool the washing with the distillate.
Note (12) 5 mL - 20 mL
(13) As an acceptor, use a 200-mL - 300-mL Erlenmeyer flask or a 200-mL - 300-mL
beaker with which the distillate outlet of the steam distillation apparatus can be
immersed in 0.25 mol/L sulfuric acid or a boric acid solution (40 g/L).
(14) An amount sufficient to make the solution strong alkalinity. A blue color will appear.
Cold water insoluble nitrogen (N1) or the nitrogen which is insoluble in a heat buffer
solution (N2) (% (mass fraction)) in the analytical sample
= (B × V6 - V7) × C1 × f1 × (V8/V9) × (14.007/W2) × (100/1000)
565
Testing Methods for Fertilizers (2018)
Note (15) Cold water insoluble nitrogen (N1) or the nitrogen which is insoluble in a heat buffer
solution (N2) uses raw data without rounding numerical value.
Cold water insoluble nitrogen (N1) or the nitrogen which is insoluble in a heat buffer
solution (N2) (% (mass fraction)) in the analytical sample
= V10 × C2 × 2 × f2 × (V11/V12) × (14.007/W2) × (100/1000)
=V10 × C2 × f2 × (V11/V12) × (2.8014/W2) ・・・・・ (3)
V10: Volume (mL) of 0.25 mol/L sulfuric acid needed for titration
C2: Set concentration (0.25 mol/L) of 0.25 mol/L sulfuric acid
f2: Factor of 0.25 mol/L sulfuric acid
V11: Predetermined volume (mL) of the digestion solution in (4.2) f)
V12: Transferred amount (mL) of the digestion solution subjected to distillation in
(4.3) b)
W2: Mass (g) of the analytical sample
Note (16) The endpoint is reached when the color changes from green to light red.
Comment 5 The titration procedures in (2) a) Standardization, (2) c) Standardization and (4.4)
can be conducted by an automatic titrator. The setup of the titration program, the
determination parameter for the endpoint and vessels such as acceptors are according
to the specification and the operation method of the automatic titrator used.
566
Testing Methods for Fertilizers (2018)
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.68- 69, Yokendo, Tokyo (1988)
(5) Flow sheet for the activity coefficient of nitrogen testing method: The flow sheets for the
activity coefficient of nitrogen testing method in formaldehyde processed urea fertilizers are
shown below:
1.00 g
Weigh into a 50-mL beaker
analytical sample
←A small amount of ethanol
←150 mL water [25ºC±2ºC]
Stirring
Analytical sample
(equivalent to 0.12 g
Weigh into a 200-mL tall beaker
of cold water soluble
nitrogen)
←150 mL heat phosphate solution
Stirring
567
Testing Methods for Fertilizers (2018)
Cold water
Subject Cold water non-dissolved mater or
non-dissolved mater or
heat buffer solution non-dissolved matter to
heat buffer solution non-
Kjeldahl method, disttilation and measurement respectively
dissolved matter
Transfer
(together with filter 300-mL - 500-mL digestion flask
paper)
←5 g - 10 g catalyst
←20 mL - 40 mL sulfuric acid
Heating Gently
Standing to cool
← A small amount of water
Transfer 250-mL - 500-mLvolumetric flask, water
Cooling
← Water (up to the marked line )
Digestion solution
Figure 2 Flow sheet for the activity coefficient of nitrogen testing method
in formaldehyde processed urea fertilizers
(Kjeldahl method procedure)
568
Testing Methods for Fertilizers (2018)
Digestion solution
Aliquot (predetermined
300-mL distillation flask
volume)
← Sodium hydroxide solution (200 g/L - 500 g/L)
Receiver: 200-mL - 300 - mL Erlenmeyer flask or beaker
Steam distillation A predetermined amount of 0.25 mol/L sulfuric acid and
a few drops of methyl red-methylene blue mixture solution,
apparatus or boric acid solution (40 g/L), several drops of
methyl red - bromocresol green mixture solution
569
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) Incubator: An incubator whose temperature is 30 ºC ± 1 ºC.
b) Total nitrogen: Apparatus and instruments corresponding to clauses in 4.1.1 when
determining total nitrogen.
c) Ammonium nitrogen: Apparatus and instruments corresponding to clauses in 4.1.2 when
determining ammoniacal nitrogen.
d) Nitrate nitrogen: Apparatus and instruments corresponding to clauses in 4.1.3 when
determining nitrate nitrogen.
e) Water-soluble phosphoric acid: Apparatus and instruments corresponding to clauses in
4.2.4 when determining water-soluble phosphoric acid.
f) Water-soluble potassium: Apparatus and instruments corresponding to clauses in 4.3.3
when determining water-soluble potassium.
g) Water-soluble magnesia: Apparatus and instruments corresponding to clauses in 4.6.3 when
determining water-soluble magnesia.
570
Testing Methods for Fertilizers (2018)
Note (1) Since no grinding is conducted and inhomogeneous laboratory samples are used, it is
desirable to heighten the reliability of determined values by conducting tests using 3 -
5 laboratory samples.
(2) Since an initial elution flow is estimated to be higher than usual if the laboratory
sample vibrates in water, water should be added gently. Do not shake to mix the
sample solution until the filtering in c) is completed.
(3) Filter most of the solution so that non-dissolved matter remains in the Erlenmeyer
flask.
(4.2) Measurement: Conduct the measurements of the initial elution flow of a target component as
specified in respective clauses in a) - f). In addition, specific measurement procedure for each
component is carried out according to a corresponding clause.
a) Total nitrogen: Transfer a predetermined amount of sample solution and quantitate total
nitrogen according to respective clauses in 4.1.1 to make the initial elution flow.
b) Ammonium nitrogen: Transfer a predetermined amount of sample solution and quantitate
ammoniacal nitrogen according to respective clauses in 4.1.2 to make the initial elution flow.
c) Nitrate nitrogen: Transfer a predetermined amount of sample solution and quantitate nitrate
nitrogen according to respective clauses in 4.1.3 to make the initial elution flow.
d) Water-soluble phosphoric acid: Transfer a predetermined amount of sample solution and
quantitate water-soluble phosphoric acid according to respective clauses in 4.2.4 to make the
initial elution flow.
e) Water-soluble potassium: Transfer a predetermined amount of sample solution and
quantitate water-soluble potassium according to respective clauses in 4.3.3 to make the initial
elution flow.
f) Water-soluble magnesia: Transfer a predetermined amount of sample solution and
quantitate water-soluble magnesia according to respective clauses in 4.6.3 to make the initial
elution flow.
Note (4) Determine total nitrogen (T-N), ammoniacal nitrogen (A-N), nitrate nitrogen (N-N),
water-soluble phosphoric acid (W-P2O5), water-soluble potassium (W-K2O) and
water-soluble magnesia (W-MgO) in 4.1.1, 4.1.2, 4.1.3, 4.2.4, 4.3.3 or 4.6.3 using test
samples prepared in 2.3 Preparation of test samples.
(5) Initial elution flow and corresponding component content use raw data without
rounding numerical value.
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.288- 290, Yokendo, Tokyo (1988)
571
Testing Methods for Fertilizers (2018)
(6) Initial elution rate testing method: The flow sheet for initial elution rate testing method in
coated fertilizers is shown below:
12.5 g
300 mL Erlenmeyer flask
laboratory sample
←250 mL water [30ºC±1ºC]
Extract 30ºC±1ºC, 24 hours
572
Testing Methods for Fertilizers (2018)
Note (1) Aluminum weighing dishes described in the Handbook of the Feed Analysis Standards
-2009- can also be used.
(2) A rotary shaker that can rotate a 100-mL ground-in stopper centrifugal precipitate tube
upside down at 30 - 40 revolutions/min may also be used.
(5) Stir with a glass rod, wash non-dissolved matter which may have adhered to the glass
rod with water and add wash to a centrifugal precipitate tube.
Note (6) Heat at the temperature to enable the procedure in (4.2.2) b).
References
1) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.316- 317, Yokendo, Tokyo (1988)
(6) Humic acid testing method: The flow sheet for humic acid testing method is shown below:
Removing supernatant
←Water
Stirring
Removing supernatant
Further repeat 3 times
575
Testing Methods for Fertilizers (2018)
Removing supernatant
←Water
Stirring
Removing supernatant
Further repeat 3 times
← 50 mL sodium hydroxide (10 g/L)
Shaking to mix Shaking, 1 hour
Removing supernatant
←Water
Stirring
Removing supernatant
Further repeat 3 times
Heating Drying
Transfer Ground-in stopper weighing bottle
Heating 105 ºC – 110 ºC, 3 hours
Standing to cool Desiccator
Measurement Weigh to the order of 1 mg
Figure 2 Flow sheet for humic acid testing method in humic acid salt fertilizers
(Testing procedure of acid insoluble - alkali non-dissolved matter (4.2))
576
Testing Methods for Fertilizers (2018)
6.10 Sulfate
This method is according to 5.29.2 Sulfate analysis in “The Official Methods of Analysis of
Fertilizers 1992”.
References
1) National Institute for Agro-Environmental Sciences, the Ministry of Agriculture,
Forestry and Fisheries: The Official Methods of Analysis of Fertilizers 1992, p.145 -
147, Japan Fertilizers Analysis Association, Tokyo (1992)
2) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.285- 286, Yokendo, Tokyo (1988)
577
Testing Methods for Fertilizers (2018)
References
1) National Institute for Agro-Environmental Sciences, the Ministry of Agriculture,
Forestry and Fisheries: The Official Methods of Analysis of Fertilizers 1992, p.121 -
123, Japan Fertilizers Analysis Association, Tokyo (1992)
2) Masayoshi KOSHINO: Second Revision of The Methods of Analysis of Fertilizers
(Details), p.259- 261, Yokendo, Tokyo (1988)
578
Testing Methods for Fertilizers (2018)
7. Nitrification inhibitor
7.1 2-amino-4-chloro-6-methylpyrimidine (AM)
7.1.a High-Performance Liquid Chromatography
(1) Summary
This testing method is applicable to fertilizers containing 2-amino-4-chloro-6-methylpyrimidine
(AM). This testing method is classified as Type C and its symbol is 7.1.a-2017 or AM.a-1.
Add methanol − water (1+1) to an analytical sample to extract
2-amino-4-chloro-6-methylpyrimidine, introduce it into a High-Performance Liquid Chromatograph
(HPLC), isolate with an octadecyl silylation silica gel column, and measure at wavelength 295 nm
to obtain 2-amino-4-chloro-6-methylpyrimidine (AM) in an analytical sample. In addition, the
performance of this testing method is shown in Comment 6.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 295 nm.
b) Magnetic stirrer:
c) Centrifugal separator: A centrifugal separator that can work at about 1700 × g.
d) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g -
10000 × g.
e) Acidic alumina cartridge column: Link a 10-mL cylinder to a column (3) that is filled with
500 mg - 1 g of acidic alumina, put 3 mL of methanol and let it flow down.
Note (3) A cartridge with a 3-mL - 6-mL column filled with 500 mg - 1 g of silica gel can be
used.
Comment 2 A column is sold under production names such as Inertsil ODS, Mightysil RP-18,
L-column ODS, Shim-pack VP-ODS, Silica C18M 4D, Puresil C18, COSMOSIL
5C18-MS-II, etc.
Comment 3 An acidic alumina cartridge is commercially sold under production names such as
Bond Elut AL-A, Sep-Pak Alumina-A, Supelclean LC-Alumina-A.
Note (4) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
(5) If there is a possibility that the 2-amino-4-chloro-6-methylpyrimidine in the sample
solution exceeds the maximum limit of the calibration curve, dilute a predetermined
amount of extract with methanol water (1+1).
Note (6) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(7) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 4 Instead of the procedures in (4.2) c) - d), it is allowed to filter with a membrane filter
(aperture diameter: no more than 0.5-µm) made of PTFE and the filtrate can be the
sample solution.
580
Testing Methods for Fertilizers (2018)
Comment 5 The test is possible by the following procedures in the case of fertilizers not
containing organic matters.
The procedures in (4.1) c) - d) and (4.2) a) - b) are omitted and “Transfer effluent” in
(4.2) c) is replaced with the “After allowing to stand still, transfer supernatant” to
operate.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter) to which octadecyl chemically bonds.
2) Column bath temperature: 30 °C - 40 °C
3) Eluent: Methanol - water (4+6)
4) Flow rate: 1 mL/min
5) Detection unit: An absorptiometric detector, measurement wavelength: 295 nm
b) Calibration curve preparation
1) Inject 10 µL of respective 2-amino-4-chloro-6-methylpyrimidine standard solutions for the
calibration curve preparation to an HPLC, record a chromatogram at wavelength 295 nm and
obtain the peak area or height.
2) Prepare a curve for the relationship between the concentration and the peak area or height at
wavelength 295 nm of the respective 2-amino-4-chloro-6-methylpyrimidine standard
solutions for the calibration curve preparation.
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain 2-amino-4-chloro-6-methylpyrimidine content from the calibration curve to calculate
2-amino-4-chloro-6-methylpyrimidine (AM) in the analytical sample.
Comment 6 A recovery testing was conducted using compound fertilizer (1 sample) and blended
fertilizer (2 samples), as a result, the mean recovery rate of
2-amino-4-chloro-6-methylpyrimidine at concentration level of 1.0 % (mass fraction),
0.4 % (mass fraction) and 0.1 % (mass fraction) were 99.1 % - 100.5 %, 99.3 % -
101.6 % and 100.2 % - 100.7 %.
Additionally, the minimum limit of quantification of this testing method is about
0.005 % (mass fraction).
References
1) Yuji SHIRAI: The volumetric analysis of 2-amino-4-chloro-6-methylpyrimidine in
fertilizers with High-Performance Liquid Chromatograph, Validation Report of
Fertilizers (in Japanese), 44 (3), p. 26 - 41(1991)
581
Testing Methods for Fertilizers (2018)
1.00 g
200-mL of ground-in stopper Erlenmeyer flask
analytical sample
← 100 mL of methanol-water (1+1)
Extraction Stirring, 30 minutes
Sample solution
582
Testing Methods for Fertilizers (2018)
583
Testing Methods for Fertilizers (2018)
Comment 2 A column is sold under production names such as Inertsil ODS, Mightysil RP-18,
L-column ODS, Shim-pack VP-ODS, Silica C18M 4D, Puresil C18, COSMOSIL
5C18-MS-II, etc.
Note (3) If there is a possibility that the 1-amidino-2-thiourea concentration in the sample
solution exceeds the maximum limit of the calibration curve, dilute a predetermined
amount of supernatant with water.
(4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 3 Instead of the procedures in (4.1) c) - d), it is allowed to filter with a membrane filter
(aperture diameter: no more than 0.5-µm) made of PTFE and the filtrate can be the
sample solution.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of High-Performance
Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter) to which octadecyl chemically bonds.
2) Column bath temperature: 30 °C - 45 °C
3) Eluent: Dissolve 0.94 g of sodium 1-hexasulfonic acid in 1000 mL of methanol water (2+8),
adjust to pH 3.15 with acetic acid and filter with a membrane filter (aperture diameter: no
more than 0.5-µm) made of hydrophilic PTFE (1).
4) Flow rate: 1 mL/min
5) Detection unit: An absorptiometric detector, measurement wavelength: 262 nm
Comment 4 A recovery testing with triplicates measurement was conducted using compound
fertilizer (2 samples), as a result, the mean recovery rate of 1-amidino-2-thiourea at
concentration level of 1.0 % (mass fraction), 0.5 % (mass fraction) and 0.25 % (mass
fraction) were 99.0 % - 104.3 %, 97.7 % - 100.7 % and 99.7 % - 101.3 %.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of this testing method is about
0.005 % (mass fraction).
References
1) Kazunori CHIBA: Analysis method of 1-amidino-2-thiourea (ASU) of Nitrification
inhibitor in fertilizers with High-Performance Liquid Chromatograph, Validation Report
of Fertilizers (in Japanese), 43 (4), p. 15 - 22 (1990)
2) Shigehiro KAI and Erina WATABE: Determination of 1-Amidino-2-thiourea as a
Nitrification Inhibitor in Compound Fertilizer by High-Performance Liquid
Chromatography: A Collaborative Study, Research Report of Fertilizers, Vol.6, p. 36 -
42 (2013)
585
Testing Methods for Fertilizers (2018)
(5) Flow sheet for 1-amidino-2-thiourea: The flow sheet for 1-amidino-2-thiourea in fertilizers
is shown below:
Sample solution
586
Testing Methods for Fertilizers (2018)
Comment 2 A column is sold under production names such as Hibar LiChrosorb NH2, Inertsil
NH2, Unison UK-Amino, Mightysil NH2, Shim-pack CLC-NH2, Shodex NH-5A,
Unisil Q NH2, etc.
Note (4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 3 Instead of the procedures in (4.1) c) - d), it is allowed to filter with a membrane filter
(aperture diameter: no more than 0.5-µm) made of PTFE and the filtrate can be the
sample solution.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter) to which amino or amino propyl chemically bonds.
2) Column bath temperature: 30 °C - 40 °C
3) Eluent: Acetonitrile - methanol (9+1)
4) Flow rate: 1 mL/min
5) Detection unit: An absorptiometric detector, measurement wavelength: 220 nm
b) Calibration curve preparation
1) Inject 10 µL of respective 4-amino-1,2,4-triazole standard solutions for the calibration curve
preparation to an HPLC, record a chromatogram at wavelength 220 nm and obtain the peak
area or height.
2) Prepare a curve for the relationship between the concentration and the peak area or height at
wavelength 220 nm of the 4-amino-1,2,4-triazole standard solutions for the respective
calibration curve preparation.
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
588
Testing Methods for Fertilizers (2018)
Comment 4 A recovery testing was conducted using compound fertilizer (2 samples), as a result,
the mean recovery rate of 4-amino-1,2,4-triazole hydrochloride at concentration level
of 0.5 % (mass fraction), 0.3 % (mass fraction) and 0.2 % (mass fraction) were
100.2 % - 104.9 %, 100.8 % - 103.0 % and 100.7 % - 104.2 %.
Additionally, the minimum limit of quantification of this testing method is about
0.005 % (mass fraction).
References
1) Koichi SAKAGAMI: Analysis methods of 4-amino-1,2,4-triazole hydrochloride with
High-Performance Liquid Chromatography, Validation Report of Fertilizers (in
Japanese), 40 (4), p.9 - 16 (1987)
(5) Flow sheet for 4-amino-1,2,4-triazole hydrochloride (ATC): The flow sheet for
4-amino-1,2,4-triazole hydrochloride in fertilizers is shown below:
Sample solution
589
Testing Methods for Fertilizers (2018)
590
Testing Methods for Fertilizers (2018)
Comment 1 A column is sold under production names such as Inertsil ODS, Mightysil RP-18,
L-column ODS, Shim-pack VP-ODS, Silica C18M 4D, Puresil C18, COSMOSIL
5C18-MS-II, etc.
Note (2) If there is a possibility that the N-2,5-dichlorophenyl succinamic acid concentration in
the sample solution exceeds the maximum limit of the calibration curve, dilute a
predetermined amount of the outflow solution with methanol.
(3) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement
(4) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 2 Instead of the procedures in (4.1) c) - d), it is allowed to filter with a membrane filter
(aperture diameter: no more than 0.5-µm) made of PTFE and the filtrate can be the
sample solution.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter) to which octadecyl chemically bonds.
2) Column bath temperature: 30 °C - 40 °C
3) Eluent: Methanol - water (5) (55 + 45)
4) Flow rate: 0.8 mL/min
5) Detection unit: An absorptiometric detector, measurement wavelength: 246 nm
Note (5) Adjust the water used to pH 3 with phosphpric acid in advance.
Comment 3 Recovery testing was conducted using compound fertilizer (2 samples) and blended
fertilizer (1 sample), as a result, the mean recovery rate of N-2,5-dichlorophenyl
succinamic acid at concentration level of 0.4 % (mass fraction), 0.2 % (mass
fraction) and 0.1 % (mass fraction) were 100.9 % - 101.4 %, 100.8 % - 101.4 % and
101.2 % - 103.4 %.
Additionally, the minimum limit of quantification of this testing method is about
0.005 % (mass fraction).
References
1) Akira KUBO: Analysis methods of N-2,5-dichlorophenyl succinamic acid (DCS) of
Nitrification suppression materials in fertilizers with High-Performance Liquid
Chromatograph, Validation Report of Fertilizers (in Japanese), 44 (4), p.25 - 36 (1991).
(5) Flow sheet for N-2,5-dichlorophenyl succinamic acid (DCS): The flow sheet for
N-2,5-dichlorophenyl succinamic acid (DCS) in fertilizers is shown below:
Sample solution
592
Testing Methods for Fertilizers (2018)
593
Testing Methods for Fertilizers (2018)
d) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g -
10000 × g.
e) Silica gel cartridge column: Link a 10-mL cylinder to the column (3) filled with 500 mg - 1
g of silica gel, add 3 mL of methanol to let it flow down.
Note (3) A cartridge with a 3-mL - 6-mL column filled with 500 mg – 1 g of silica gel can be
used.
Comment 2 A column is sold under production names such as Hibar LiChrosorb NH2, Inertsil
NH2, Unison UK-Amino, Mightysil NH2, Shim-pack CLC-NH2, Shodex NH-5A,
Unisil Q NH2, etc.
Comment 3 A silica gel cartridge column is commercially sold under production names such as
Sep-Pak Plus Silica, InertSep Si.
Note (4) Mix well until the whole sample comes in contact with water.
(5) About 5 g - 10 g.
(6) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
(7) If there is a possibility that the concentration of dicyandiamide in the sample solution
exceeds the maximum limit of the calibration curve, dilute a predetermined amount of
extract with methanol.
Note (8) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(9) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 4 Instead of the procedures in (4.2) c) - d), it is allowed to filter with a membrane filter
(aperture diameter: no more than 0.5-µm) made of PTFE and the filtrate can be the
sample solution.
595
Testing Methods for Fertilizers (2018)
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter) to which amino or amino propyl chemically bonds.
2) Column bath temperature: 30 °C - 40 °C
3) Eluent: Acetonitrile - methanol (6+1)
4) Flow rate: 0.5 mL/min - 1 mL/min
5) Detection unit: An absorptiometric detector, measurement wavelength: 215 nm
b) Calibration curve preparation
1) Inject 10 µL of respective dicyandiamide standard solutions for the calibration curve
preparation to an HPLC, record a chromatogram at wavelength 215 nm, and obtain the peak
area or the height.
2) Prepare a curve for the relationship between the concentration and the peak area or the height
at wavelength 215 nm of the respective dicyandiamide standard solutions for the calibration
curve preparation.
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain dicyandiamide content from the calibration curve to calculate dicyandiamide (Dd) in
the analytical sample.
Comment 5 Recovery testing was conducted using inorganic compound fertilizer (2 samples) and
organic compound fertilizer (3 samples), as a result, the mean recovery at the
concentration level of 2 % (mass fraction) and 0.2% (mass fraction) were 101.2 % -
102.6 % and 98.4 % - 100.6 %.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of this testing method is about
0.01 % (mass fraction).
596
Testing Methods for Fertilizers (2018)
References
1) Masakazu SAIKI: Development of High-Performance Liquid Chromatography for
Determination of Dicyandiamide as a Nitrification Inhibitor in Fertilizer. Research
Report of Fertilizer, Vol.3, p. 43 - 50 (2010)
2) Masakazu SAIKI: Determination of Dicyandiamide as a Nitrification Inhibitor in
Fertilizer by High-Performance Liquid Chromatography: Collaborative Study, Research
Report of Fertilizer, Vol.4, p. 16 - 22 (2011)
597
Testing Methods for Fertilizers (2018)
(5) Flow sheet for dicyandiamide: The flow sheet for dicyandiamide in fertilizers is shown
1.00 g
200-mL ground-in stopper Erlenmeyer flask
analytical sample
← 1 mL of water
Leave at rest 5 minutes
← 100 mL of methanol
Shaking 10 minutes
← An adequate amount of sodium sulfate
Ground-in stopper centrifugal precipitate tube,
Centrifugal separation
1700 × g , 5 minutes
Sample solution
598
Testing Methods for Fertilizers (2018)
Reference: HPLC chromatogram of dicyandiamide standard solution for the calibration curve
preparation and sample solution (compound fertilizer) are shown below.
0.20 0.20
0.18 0.18
8.607 Dd
0.16 0.16
0.14 0.14
AU
0.10 0.10
0.08 0.08
0.06 0.06
0.04 0.04
0.02 0.02
0.00 0.00
0 2 4 6 8 10 12 14 0 2 4 6 8 10 12 14
min min
1) Dicyandiamide standard solution (the equivalents of 100 ng (10 µg/ mL, 10 µL) of
dicyandiamide)
2) Sample solution (compound fertilizer)
599
Testing Methods for Fertilizers (2018)
Comment 1 2-sulfanilamide thiazole is sold under the production name sulfathiazole by Tokyo
Chemical Industry Co., Ltd, FUJIFILM Wako Pure Chemical Co., Ltd., and Kanto
Chemical Co., Inc.
10000 × g.
e) Acidic alumina cartridge column: Link a 10-mL cylinder to a column (3) that is filled with
500 mg - 1 g of acidic alumina, put 3 mL of methanol and let it flow down.
Note (3) A cartridge with a 3-mL - 6-mL column filled with 500 mg – 1 g of silica gel can be
used.
Comment 2 A column is sold under production names such as Inertsil ODS, Mightysil RP-18,
L-column ODS, Shim-pack VP-ODS, Silica C18M 4D, Puresil C18, COSMOSIL
5C18-MS-II, etc.
Comment 3 An acidic alumina cartridge is commercially sold under production names such as
Bond Elut AL-A, Sep-Pak Alumina-A, Supelclean LC-Alumina-A.
Note (4) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
(5) If there is a possibility that the 2-sulfanilamide thiazole concentration in the sample
solution exceeds the maximum limit of the calibration curve, dilute a predetermined
amount of the extract with methanol.
Note (6) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(7) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 4 Instead of the procedures in (4.2) c) - d), it is allowed to filter with a membrane filter
(aperture diameter: no more than 0.5-µm) made of PTFE and the filtrate can be the
sample solution.
Comment 5 The test is possible by the following procedures in the case of fertilizers not
containing organic matters.
The procedures in (4.1) c) - d) and (4.2) a) - b) are omitted and “Transfer effluent” in
(4.2) c) is replaced with the “After allowing to stand still, transfer supernatant” to
601
Testing Methods for Fertilizers (2018)
operate.
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter) to which octadecyl chemically bonds.
2) Column bath temperature: 30 °C - 40 °C
3) Eluent: Methanol−water (2+8)
4) Flow rate: 1 mL/min
5) Detection unit: An absorptiometric detector, measurement wavelength: 285 nm
b) Calibration curve preparation
1) Inject 10 µL of respective 2-sulfanilamide thiazole standard solutions for the calibration curve
preparation to an HPLC, record a chromatogram at wavelength 285 nm and obtain the peak
area or height.
2) Prepare a curve for the relationship between the concentration and the peak area or height at
wavelength 285 nm of the respective 2-sulfanilamide thiazole standard solutions for the
calibration curve preparation.
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the 2-sulfanilamide thiazole content from the calibration curve to calculate
2-sulfanilamide thiazole (ST) in the analytical sample.
Comment 6 Recovery testing was conducted using compound fertilizer (1 sample) and blended
fertilizer (2 sample), as a result, the mean recovery rate of 2-sulfanilamide thiazole at
the concentration level of 1.0 % (mass fraction), 0.4 % (mass fraction) and 0.1 %
(mass fraction) were 101.2 % - 102.1 %, 99.6 % - 101.7 % and 99.4 % - 101.0 %.
Additionally, the minimum limit of quantification of this testing method is about
0.005 % (mass fraction).
References
1) Yuji SHIRAI: The volumetric analysis of 2-sulfanilamide thiazole in fertilizers with
High-Performance Liquid Chromatograph, Validation Report of Fertilizers (in Japanese),
44 (1), p. 10 - 20 (1991)
602
Testing Methods for Fertilizers (2018)
(5) Flow sheet for 2-sulfanilamide thiazole: The flow sheet for 2-sulfanilamide thiazole in
fertilizers is shown below:
Sample solution
Reference: HPLC chromatogram of 2-sulfanilamide thiazole (ST) standard solution for the
calibration curve preparation is shown below.
603
Testing Methods for Fertilizers (2018)
8. Others
8.1 Melamine and its degradation products
8.1.a High-Performance Liquid Chromatography
(1) Summary
This testing method is applicable to fertilizers. This testing method is classified as Type D and its
symbol is 8.1.a-2017 or Mel.a-1.
Extract melamine and its derivative substances (hereinafter referred to as “melamine derivations”)
in organic matters and fertilizers containing organic matters with diethylamine−water−acetonitrile
(1+4+5) and derivatize with BSTFA−TMCS (99+1) and then measure with a gas
chromatography/mass spectrometer to obtain melamine deviations in an analytical sample. In
addition, the performance of this testing method is shown in Comment 8.
Comment 1 The structural formula of melamine and its degradation products is shown in the
figure 1. During the production process of melamine, a by-product that replaces
“-NH2” of R1 - R3 with “-OH” is formed in some cases.
R1 R1 R2 R3 MW
Melamine NH2 NH2 NH2 126.12
N N Ammeline OH NH2 NH2 127.10
Ammelide OH OH NH2 128.09
R2 N R3 Cyanuric acid OH OH OH 129.07
Note (1) After it is opened once, add a proper amount of sodium sulfate (anhydrous) and seal
tightly to store.
(2) A mixed derivatization reagent is commercially sold under the name BSTFA−TMCS
(99+1).
(3) The respective standard reagents of melamine, ammeline, ammelide and cyanuric acid
604
Testing Methods for Fertilizers (2018)
Comment 4 A capillary column is commercially sold under the names such as DB-5ms, Rtx-5ms,
HP-5ms, SLB-5ms, BPX-5, CP-Sil 8CB low Bleed/MS and TC -5HT for GC/MS.
Note (4) Confirm that the tube is made of polypropylene, etc. to not affect testing results.
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 5 Grind until it completely passes through a sieve of 500 µm aperture to prepare the
test sample.
Comment 6 Weigh 0.5 g of an analytical sample, extract with 200 mL of
diethylamine−water−acetonitrile (1+4+5). If it is diluted 50 times in the procedure in
d), the quantitative range of melamine derivations in the analytical sample is 0.2 %
605
Testing Methods for Fertilizers (2018)
(4.3) Measurement: Conduct the measurement as indicated in JIS K 0123 and as shown below.
Specific measurement procedures are according to the operation method of the gas
chromatograph/ mass spectrometer used in measurement.
a) Measurement conditions for the gas chromatograph/ mass spectrometer: Set up the
measurement conditions considering it:
1) Gas Chromatograph
(i) Sample injection method: split less injection method (1min)
(ii) Temperature of sample injector: 280 °C
(ii) Capillary column: A capillary column (0.25-mm - 0.32-mm inner diameter, 30-m long,
0.25 µm layer thickness) made of fused silica. 5 % phenyl 95 % methyl polysiloxane
chemically bonds to the inner surface of the capillary column.
(iv) Temperature of column bath: 100 °C (1 min) → (15 °C /min) → 320 °C (3 min)
(v) Temperature of GC/MS coupling portion: 250 °C
(vi) Carrier gas: helium, flow rate: 1.5 mL/min
2) Mass Spectrometer
(i) Ionization method: Electron-Impact ionization (EI) method
(ii) Ionization voltage: 70 V
(iii) Temperature of ion source: 230 °C
(iv) Ion detection method: Selected Ion Monitoring (SIM) method
(v) Measurement of ion: Shown in table 1
606
Testing Methods for Fertilizers (2018)
Note (9) Confirm that the ratio against the peak area ratio or height ratio of the standard
solution is within the range of about ± 30 %. In addition, the peak area ratio or height
ratio depends on the concentration.
607
Testing Methods for Fertilizers (2018)
Comment 8 Recovery testing of melamine derivations was conducted using soybean meal, fish
meal, fish waste processed fertilizer, mixed organic fertilizer, blended fertilizer and
compound fertilizer, as a result, the mean recovery rate at additive level of 10 %
(mass fraction) and 0.2 % (mass fraction) were 92.1 % - 102.9 % and 90.3 % -
102.2 %.
Additionally, the minimum limit of quantification of melamine derivations of the test
method is about 0.01 % (mass fraction).
References
1) Yuji SHIRAI, Jun OKI: Validation of Gas Chromatography/Mass Spectrometry for
Determination of Melamine and Its Degradation Products in Fertilizers, Research
Report of Fertilizer, Vol.1, p. 114 - 121 (2008)
608
Testing Methods for Fertilizers (2018)
(5) Flow sheet for melamine derivations: The flow sheet for melamine derivations in fertilizers
is shown below.
Extract
Extract
concentration
under reduced pressure Centrifugal evaporator (70 ºC ± 2 ºC)
/ Drying up
← 0.3 mL of pyridine (dehydration)
← 0.2 mL of derivatization reagent (BSTFA−TMCS (99+1))
Stopple and seal.
Derivatization 70 ºC ± 2 ºC, 45 minutes
Standing to cool
Sample solution
609
Testing Methods for Fertilizers (2018)
Reference: Total Ion Chromatogram (TIC) of GC/MS of mixture standard solution for
calibration curve preparation of melamine derivations is shown below.
↓a ↓b ↓c ↓d
610
Testing Methods for Fertilizers (2018)
8.1.b
611
Testing Methods for Fertilizers (2018)
Comment 1 The standard reagent of melamine, ammeline, ammelide and cyanuric acid are sold
by FUJIFILM Wako Pure Chemical Co., Ltd. and Kanto Chemical Co., Inc. and
Hayashi Pure Chemical Industries., Ltd.
Comment 2 Column is sold under the production name TSKgel Amide-80, etc. A column which
has actually isolated melamine, ammeline, ammelide and cyanuric acid should be
used.
Note (4) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
(5) If there is a possibility that the concentration of melamine derivations in the sample
solution exceeds the maximum limit of the calibration curb, the amount of a
supernatant solution to be transferred should be1 mL - 2.5 mL.
(6) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement
(7) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 3 Instead of the procedures in (4.1) f) - g), it is allowed to filter with a membrane filter
(pore size: no more than 0.5-µm) made of hydrophilic PTFE and the filtrate can be
the sample solution.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter column) to which carbamoyl chemically bonds.
613
Testing Methods for Fertilizers (2018)
Comment 4 Recovery testing was conducted using 3 brands of nitrolime, 1 brand of compound
fertilizer containing nitrolime, 2 brands of compound fertilizers not containing
nitrolime, 1 brand of ammonium sulfate and 1 brand of urea, as a result, the recovery
rate of melamine derivations at concentration level of 4 % (mass fraction) and 0.1 %
(mass fraction) were 90.5 % - 106.3 % and 92.2 % - 107.0 %.
Table 1 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification of the test method is about 0.02 %
(mass fraction) for melamine and cyanuric acid and about 0.01 % (mass fraction) for
ammeline and ammelide. In the case of ammelide and cyanuric acid, the sufficient
reproducibility was observed in the range of 0.188 % (mass fraction) - 1.10 % (mass
fraction) and 0.105 % (mass fraction) - 1.15 % (mass fraction) respectively.
614
Testing Methods for Fertilizers (2018)
References
1) Etsuko BANDO and Yuji SHIRAI: Validation of High-Performance Liquid
Chromatography (HPLC) for Determination of Melamine and Its Related Substances in
Fertilizer, Research Report of Fertilizer, Vol.6, p. 27 - 35 (2013)
2) Etsuko BANDO and Sigehiro KAI: Determination of Melamine and Its Related
Substances in Fertilizer by High-Performance Liquid Chromatography (HPLC): A
Collaborative Study, Research Report of Fertilizer, Vol.7, p. 10 - 21 (2014)
615
Testing Methods for Fertilizers (2018)
(5) Flow sheet for melamine derivations: The flow sheet for melamine derivations in fertilizers
is shown below.
Sample solution
616
Testing Methods for Fertilizers (2018)
Reference: HPLC chromatogram of the mixture standard solution for calibration curve
preparation of melamine derivations is shown below.
617
Testing Methods for Fertilizers (2018)
Comment 1 A standard reagent of melamine is sold by FUJIFILM Wako Pure Chemical Co., Ltd.,
Kanto Chemical Co., Inc. and Hayashi Pure Chemical Industries., Ltd.
Comment 2 Column is sold under the production name TSKgel Amide-80, etc. A column which
has actually isolated melamine, ammeline, ammelide and cyanuric acid should be
used.
Note (4) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
(5) If there is a possibility that the concentration of melamine derivations in the sample
solution exceeds the maximum limit of the calibration curb, the amount of a
supernatant solution to be transferred should be1 mL - 2.5 mL.
(6) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement
(7) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 3 Instead of the procedures in (4.1) f) - g), it is allowed to filter with a membrane filter
(pore size: no more than 0.5 µm) made of hydrophilic PTFE and the filtrate can be
the sample solution.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of High-Performance
Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions for a High-Performance Liquid Chromatograph (HPLC)
is shown below. Set up the measurement conditions considering it:
1) Column: A silica gel column (4-mm - 6-mm inner diameter, 150-mm - 250-mm long, 5-µm
particle diameter column) to which carbamoyl chemically bonds.
2) Column bath temperature: 40 °C ± 1 °C
3) Eluent: Acetonitrile - phosphate buffer solution (82+18)
4) Flow rate: 1 mL/min
619
Testing Methods for Fertilizers (2018)
Comment 4 Additive recovery testing was conducted to evaluate trueness using rape seed meal,
soybean meal, compound fertilizer containing lime nitrogen and organic matter,
compound fertilizer containing organic matter and blended fertilizer containing
organic matter (1 brand for each). As a result, the mean recovery rate at additive level
of 2 % (mass fraction), 0.4 % (mass fraction) and 0.1 % (mass fraction) were 94.6 %
- 99.8 %, 92.4 % - 98.5 % and 93.1 % - 98.4 % respectively.
The results of the repeatability tests on different days using soybean meal and
compound fertilizer containing organic matter, to evaluate precision were analyzed
by one-way analysis of variance. Table 1 shows the calculation results of
intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.02 % (mass fraction).
References
1) Etsuko FUNAMIZU: Determination Method for Melamine in Organic Fertilizer and
Fertilizer Containing Organic Matter using High-Performance Liquid Chromatograph
(HPLC), Research Report of Fertilizers, Vol.9, p. 33 - 42 (2016)
620
Testing Methods for Fertilizers (2018)
(5) Flow sheet for melamine derivations: The flow sheet for melamine derivations in fertilizers
is shown below.
Sample solution
←melamin
621
Testing Methods for Fertilizers (2018)
Comment 1 Structural formulas of clopyralid, aminopyralid and picloram are as shown in Figure
1.
622
Testing Methods for Fertilizers (2018)
acetonitrile, transfer to 100-mL volumetric flasks and add the solvent up to the marked line.
k) Mixture standard solution (100 ng/mL) (1): Dilute a predetermined amount of respective
agrichemical standard solutions (0.1 mg/mL) with formic acid (1+1000) to prepare mixture
standard solution (100 ng/mL).
l) Mixture standard solution for calibration curve preparation (5 ng/mL - 50 ng/mL) (1): In
the case of usage, transfer 2.5 mL - 25 mL of mixture standard solution (100 ng/mL) to 50 mL
volumetric flasks step-by-step, and add formic acid (1+1000) up to the marked line.
m) Mixture standard solution for calibration curve preparation (0.5 ng/mL - 5 ng/mL) (1): In
the case of usage, transfer 2.5 mL - 25 mL of mixture standard solution (10 ng/mL) to 50 mL
volumetric flasks step-by-step, and add formic acid (1+1000) up to the marked line.
Comment 2 Standard reagents of clopyralid, aminopyralid and picloram are sold by FUJIFILM
Wako Pure Chemical Co., Ltd., Kanto Chemical Co., Inc. and Hayashi Pure
Chemical Industries., Ltd.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph/Mass Spectrometer (LC-MS/MS):
LC-MS/MS specified in JIS K 0136 that satisfies the following requirements.
1) High-Performance Liquid Chromatograph
(i) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45 ºC.
(ii) Column: A 2-mm - 3-mm inner diameter 50-mm - 150-mm long 1.6-µm - 2.2-µm
particle diameter stainless steel column tube filled with silica gel to which octadecyl
chemically bonds. The specification is according to the mass spectrometer specification.
2) Mass Spectrometer
(i) Ionization method: Electro-Spray Ionization (ESI) method
(ii) Ion detection method: Selected Reaction Monitoring
b) Shaking apparatus
c) Manifold
d) Centrifugal separator: A centrifugal separator that can work at about 1700 × g.
e) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g -
10000 × g.
f) Concentrator: An evaporator that can adjust to 40 ºC ± 2 ºC.
g) Copolymer cartridge column: A divinylbenzene-N-vinylpyrrolidone copolymer mini
column (200 mg)
Comment 3 Column is sold under the production name ACQUITY UPLC HSS C18, etc.
Comment 4 A copolymer cartridge is sold under the production names such as Oasis HLB 6cc
(200 mg), Oasis PRiME HLB Plus Short Cartridge (225 mg).
(4)
d) Centrifuge it at about 1700 × g centrifugal force for about five minutes and use the
supernatant as the extract.
Note (3) It is also allowed to add 100 mL of sodium hydroxide solution (40 g/L)−methanol
[1+99]
(4) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
Comment 5 Grind until it completely passes through a sieve of 500 µm aperture to prepare the
test sample.
Note (5) Use a pressure reducing device in the procedure in (4.2) and (4.3) as appropriate.
(6) When pretreating many analytical samples, a free-standing type vessel that can contain
a solution with a liquid volume of 20 mL may be used. In this case, instead of
procedure d), put an effluent into a round-bottle flask, wash the vessel 2 times with 2.5
mL of methanol and add washing to the previous effluent.
(7) When using Oasis HLB 6cc (200 mg), add 5 mL of the extract 2 times.
Note (8) If there is a possibility that the clopyralid, aminopyralid and picloram concentration in
the sample solution exceed the maximum limit of the calibration curve, dilute a
predetermined amount of effluent with formic acid (1 +1000).
(9) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
624
Testing Methods for Fertilizers (2018)
(4.4) Measurement: Conduct the measurement as indicated in JIS K 0136 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph Mass Spectrometer used in measurement.
a) The measurement conditions of High-Performance Liquid Chromatograph/Mass
spectrometer: An example of measurement conditions for a High-Performance Liquid
Chromatograph/Mass Spectrometer is shown below. Set up the measurement conditions
considering it:
1) High-Performance Liquid Chromatograph
(i) Column: A silica gel column (2-mm - 3-mm inner diameter, 50-mm - 150-mm long,
1.6-µm - 2.2-µm particle diameter column) to which octadecyl chemically bonds.
(ii) Flow rate: 0.2 mL/min - 0.5 mL/min
(iii) Eluent: A: Formic acid (1+1000) B: Methanol:
(iv) Gradient: 0 min (5 %B) → 5 min (60 %B) → 6 min (95 %B) → 7 min (5 %B)
(v) Temperature of column bath: 40 °C
(vi) Injection volume: 5 µL
2) Mass Spectrometer
(i) Ionization method: Electro-Spray Ionization (ESI) method
(ii) Mode: Positive
(iii) Capillary voltage: 1.0 kV
(iv) Ion source temperature: 120 °C
(v) Desolvation temperature: 400 °C
(vi) Cone voltage: Shown in table 1
(vii) Collision energy: Shown in table 1
(viii) Monitor ion: Shown in table 1
625
Testing Methods for Fertilizers (2018)
c) Sample measurement
1) Subject 5 µL of the sample solution to the same procedure as in b) 2) - 3) (11).
2) Obtain the content of material subjected to measurement from the calibration curve to
calculate the material subjected to measurement in the analytical sample.
Note (11) Confirm that the ratio against the peak area ratio or height ratio of the standard
solution is within the range of about ± 30 %. In addition, the peak area ratio or height
ratio depends on the concentration.
Comment 6 Additive recovery testing of clopyralid, aminopyralid and picloram was conducted
using cow dung compost (2 kinds), composted sludge fertilizers containing cow
manure (2 kinds) and composted sludge fertilizers containing pig manure (1 kind), as
a result, the mean recovery rates at additive level of 1000 µg/kg, 400 µg/kg and 40
µg/kg were 78.1 % - 90.0 %, 81.0 % - 117.6 % and 71.2 % - 101.3 % respectively.
Table 2 shows results and analysis results from a collaborative study for testing
method validation.
Additionally, the minimum limits of quantification of clopyralid, aminopyralid and
picloram of the test method are about 10 µg/kg respectively.
626
Testing Methods for Fertilizers (2018)
628
Testing Methods for Fertilizers (2018)
(5) Flow sheet for clopyralid and its derivative substances: The flow sheet for clopyralid and
its derivative substances in compost and composted sludge fertilizer is shown below:
Extract
Figure 1 Flow sheet for clopyralid and its derivative substances in compost
and composted sludge fertilizers (Extraction procedure)
Extract
← 10 mL of extract
copolymer cartridge column (1)
Cleanup (1)
(Wash with about 5 mL of methanol, then 5 mL of water in advance)
[Collection, 100-mL eggplant-shaped flask]
← About 5 mL of sodium hydroxide solution (0.4 g/L)−methanol [1+1] × 2 times
[Collection, same as above vessel]
← About 5 mL of methanol
Concentration
under reduced 40 ºC, until the volume reaches about 5 mL
pressure
← 3 mL of hydrochloric acid (1+11)
copolymer cartridge column (2)
Cleanup (2) (Wash with about 5 mL of acetonitrile,
then 5 mL of hydrochloric acid (1+120) in advance)
← Wash the vessel 2 times with about 5mL of hydrochloric acid (1+120)
← Wash with about 5mL of hydrochloric acid (1+120)−acetonitrile [9+1]
← Wash with about 5mL of water
← 4 mL of ammonia solution (0.0028 w/v %)−acetonitrile [9+1]
[Elute, 10-mL centrifugal precipitate tube]
Elute
Figure 2 Flow sheet for clopyralid in compost and composted sludge fertilizers
(Cleanup (1) and cleanup (2) and measurement procedure)
629
Testing Methods for Fertilizers (2018)
Peak No.
1
Intensity / arb.units
0 5
Retention Time / min
630
Testing Methods for Fertilizers (2018)
(2)
in a weighing dish and measure the mass to the order of 0.1 mg. Dissolve with a small
amount of acetonitrile, transfer to 100-mL volumetric flasks and add the solvent up to the
marked line.
o) Clopyralid standard solution (250 ng/mL) (1): Dilute a predetermined amount of clopyralidl
standard solutions (0.1 mg/mL) with formic acid (1+1000) to prepare clopyralid standard
solution (250 ng/mL).
p) Clopyralid standard solution for calibration curve preparation (5 ng/mL - 50 ng/mL) (1):
In the case of usage, transfer 2.5 mL - 25 mL of clopyralid standard solution (100 ng/mL) to
50 mL volumetric flasks step-by-step, and add formic acid (1+1000) up to the marked line.
q) Clopyralid standard solution for calibration curve preparation (0.5 ng/mL - 5 ng/mL) (1):
In the case of usage, transfer 2.5 mL - 25 mL of clopyralid standard solution (10 ng/mL) to 50
mL volumetric flasks step-by-step, and add formic acid (1+1000) up to the marked line.
Comment 2 A standard reagent of clopyralid is sold by FUJIFILM Wako Pure Chemical Co., Ltd.,
Kanto Chemical Co., Inc. and Hayashi Pure Chemical Industries., Ltd.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph/Mass Spectrometer (LC-MS/MS):
LC-MS/MS specified in JIS K 0136 that satisfies the following requirements.
1) High-Performance Liquid Chromatograph
(i) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45 ºC.
(ii) Column: A 2-mm - 3-mm inner diameter 50-mm - 150-mm long 1.6-µm - 2.2-µm
particle diameter stainless steel column tube filled with silica gel to which octadecyl
chemically bonds. The specification is according to the mass spectrometer specification.
2) Mass Spectrometer
(i) Ionization method: Electro-Spray Ionization (ESI) method
(ii) Ion detection method: Selected Reaction Monitoring
b) Shaking apparatus
c) Manifold
d) Centrifugal separator: A centrifugal separator that can work at 700 × g - 2000 × g.
e) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g -
10000 × g.
f) Concentrator: An evaporator that can adjust to 40 ºC ± 2 ºC.
g) Copolymer cartridge column: A divinylbenzene-N-vinylpyrrolidone copolymer mini column
(200 mg or 335 mg)
h) Filter: A funnel for filtering under reduced pressure (compatible filter diameter: 60 mm)
i) Glass fiber filter paper: A filter paper made of glass fiber (filter diameter: 60 mm) that can
keep particle diameter 0.8 µm.
j) Test tube mixer: Vortex mixer
Comment 3 Column is sold under the production name ACQUITY UPLC HSS C18, etc.
Comment 4 A copolymer cartridge is sold under the production names such as Oasis HLB 6cc
(225 mg), Oasis PRiME HLB Plus Short Cartridge (225 mg).
Comment 5 A funnel for filtering under reduced pressure is sold under the production name
KIRIYAMA Funnel SB-60, KIRIYAMA Funnel SU-60, etc.
Comment 6 A glass fiber filter is sold under the production name Glass filter paper GFP-60, etc.
632
Testing Methods for Fertilizers (2018)
Note (3) A vessel used for an extract procedure should be made of glass or polypropylene and it
must enable vibration and centrifugation procedure.
(4) A 100-mL - 200-mL ground-in stopper or screw cap Erlenmeyer flask can also be
used. In this case, however, suspension should be transferred to a ground-in stopper or
screw cap centrifugal precipitate tube before the procedure c) and e).
(5) 16.5-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g centrifugal
force.
(6) An Erlenmeyer flask can also be used. In this case, however, filtrate should be
transferred to a 100-mL volumetric flask before the procedure h).
Comment 5 Grind until it completely passes through a sieve of 500 µm aperture to prepare the
test sample.
Note (7) Use a pressure reducing device in the procedure in (4.2) and (4.3) as appropriate.
(8) When making a pretreatment of many analytical samples, a free-standing type vessel
that can contain a solution with a liquid volume of 20 mL may be used. In this case,
instead of procedure d), put an effluent into a round-bottle flask, wash the vessel 2
times with 2.5 mL of methanol and add washing to the previous effluent.
(9) When using Oasis HLB 6cc (200 mg), add 5 mL of the extract 2 times.
b) After conducting vacuum concentration of the effluent in (4.2) c) until no more than 5 mL on
a water bath at no more than 40 °C, add 3 mL of hydrochloric acid (1 +11).
c) Put the concentrated effluent into the cartridge column and allow the effluent to overflow until
the surface of liquid reaches the top of packing materials.
d) Wash a round-bottom flask with about 5 mL of hydrochloric acid (1+120) 2 times and add
washing into the cartridge successively.
e) Then, add about 5 mL of hydrochloric acid (1+120)−acetonitrile (9+1) and about 5 mL of
water into the cartridge successively and allow the liquid to overflow.
f) Place a 10-mL cone shaped centrifugal precipitate tube with a screw cap (10) under the
cartridge column, add 4 mL of ammonia solution (0.0028 % (mass fraction))−acetonitrile
[9+1] to the cartridge column and allow clopyralid to elute.
Note (10) The part under 2 mL from the bottom forms a cone shape.
Note (11) 16.5-cm of rotor radius and 2000 rpm of revolutions makes about 740 × g centrifugal
force. Confirm the permissible range of centrifugal force of a 10-mL cone shaped
bottle centrifugal precipitate tube with a screw cap used.
(12) When using a Pasteur pipet, use the same Pasteur pipet through a series of procedures
in c) - d).
(13) Disperse dichloromethane sufficiently. If vibration occurs while dichloromethane layer
remains caked, the extraction efficiency of clopyralid deteriorates and mesurements
are affected.
(14) When using a Pasteur pipet, use the same Pasteur pipet through a series of procedures
in g) - h). Do not use the Pasteur pipet used in Note (9).
(15) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement
(16) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
634
Testing Methods for Fertilizers (2018)
(4.5) Measurement: Conduct the measurement as indicated in JIS K 0136 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph Mass Spectrometer used in measurement.
a) The measurement conditions of High-Performance Liquid Chromatograph/Mass
spectrometer: An example of measurement conditions for a High-Performance Liquid
Chromatograph/Mass Spectrometer is shown below. Set up the measurement conditions
considering it:
1) High-Performance Liquid Chromatograph
(i) Column: A silica gel column (2-mm - 3-mm inner diameter, 50-mm - 150-mm long,
1.6-µm - 2.2-µm particle diameter column) to which octadecyl chemically bonds.
(ii) Flow rate: 0.2 mL/min - 0.5 mL/min
(iii) Eluent: A: Formic acid (1+1000) B: Methanol:
(iv) Gradient: 0 min (5 %B) → 5 min (60 %B) → 6 min (95 %B) → 7 min (5 %B)
(v) Temperature of column bath: 40 °C
(vi) Injection volume: 5 µL
2) Mass Spectrometer
(i) Ionization method: Electro-Spray Ionization (ESI) method
(ii) Mode: Positive
(iii) Capillary voltage: 1.0 kV
(iv) Ion source temperature: 120 °C
(v) Desolvation temperature: 400 °C
(vi) Cone voltage: 20 V
(vii) Collision energy: 20 eV for determination, 30 eV for validation
(viii)Monitor ion: Precursor ion m/z 192
Product ion m/z 146 for determination, m/z 110 for validation
c) Sample measurement
1) Subject 5 µL of the sample solution to the same procedure as in b) 2) - 3) (18).
2) Obtain the clopyralid content from the calibration curve to calculate clopyralid in the
analytical sample.
Note (18) Confirm that the ratio against the peak area ratio or height ratio of the standard
solution is within the range of about ± 30 %. In addition, the peak area ratio or height
ratio may depend on the concentration.
Comment 8 Additive recovery testing of clopyralid was conducted using cow dung compost (1
kinds), as a result, the mean recovery rates at additive level of 50 µg/kg, 10 µg/kg
and 2 µg/kg were 78.9 %, 78.3 % and 71.5 % respectively. In addition, additive
recovery testing of clopyralid was conducted using swine manure compost, poultry
manure compost and composted sludge fertilizer and (1 sample for each), as a result,
the mean recovery rates at additive level of 200µg/kg, 2 µg/kg and 80µg/kg were
88.6 %, 81.2 % and 94.2 % respectively.
Table 1 shows results and analysis results from a collaborative study for testing
method validation.
Additionally, the minimum limit of quantification of clopyralid of this test method is
about 2 µg/kg.
(5) Flow sheet for clopyralid: The flow sheet for clopyralid in compost is shown below:
5.00 g
100-mL centrifugal precipitate tube with a screw cap
analytical sample
← 50 mL of sodium hydroxide solution (40 g/L)−methanol [1+99]
Shaking About 30 minutes
637
Testing Methods for Fertilizers (2018)
Extract
← 10 mL of extract
copolymer cartridge column (1)
Cleanup (1)
(Wash with about 5 mL of methanol, then 5 mL of water in advance)
[Collection, 100-mL eggplant-shaped flask]
← About 5 mL of sodium hydroxide solution (0.4 g/L)−methanol [1+1] × 2 times
[Collection, same as above vessel]
← About 5 mL of methanol
Concentration
under reduced 40 ºC, until the volume reaches about 5 mL
pressure
← 3 mL of hydrochloric acid (1+11)
copolymer cartridge column (2)
Cleanup (2) (Wash with about 5 mL of acetonitrile,
then 5 mL of hydrochloric acid (1+120) in advance)
← Wash the vessel 2 times with about 5mL of hydrochloric acid (1+120)
← Wash with about 5mL of hydrochloric acid (1+120)−acetonitrile [9+1]
← Wash with about 5mL of water
← 4 mL of ammonia solution (0.0028 w/v %)−acetonitrile [9+1]
[Elute, 10-mL centrifugal precipitate tube]
Elute
Figure 2 Flow sheet for clopyralid in compost and composted sludge fertilizers
(Cleanup (1) and cleanup (2) and measurement procedure)
638
Testing Methods for Fertilizers (2018)
Elute
Cleanup (3)
← 0.1 mL of sodium hydroxide (40 g/L)
Shake to mix Vortex
← 2 mL of dichloromethane
Shake to mix Vortex, 30 seconds
← 2 mL of dichloromethane
Shake to mix Vortex, 30 seconds
Exsiccation Nitrogen
← 1 mL of formic acid (1+1000)
Transfer 1.5 mL centrifugal precipitate tube made of polymer
Supernatant
Measurement LC-MS/MS
Figure 2 Flow sheet for clopyralid in compost and composted sludge fertilizers
(Cleanup (3) and measurement procedure)
639
Testing Methods for Fertilizers (2018)
Intensity /
Retention
640
Testing Methods for Fertilizers (2018)
(2) Summary
This testing method is applicable to fluid home garden-use mixed fertilizer and fluid mixed fertilizer.
This testing method is classified as Type B and its symbol is 8.3.1.a-2017 or AG-C-1.a-1.
Dissolve respective agricultural chemicals in fertilizers with acetonitrile and water, and extract.
Refine by using 2 kinds of cleanup cartridge, and then measure with a High-Performance Liquid
Chromatograph/Mass Spectrometer to obtain compounds subjected to analysis in an analytical
sample. In addition, the performance of this testing method is shown in Comment 3.
the case of usage, transfer 2.5 mL - 25 mL of mixture standard solution (1000 ng/ mL) to
50-mL volumetric flasks step-by-step, and add methanol up to the marked line.
q) Mixture standard solution for calibration curve preparation (5 ng/mL - 50 ng/mL): In
the case of usage, transfer 2.5 mL - 25 mL of mixture standard solution (100 ng/ mL) to
50-mL volumetric flasks step-by-step, and add methanol up to the marked line.
(4) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph/Mass Spectrometer (LC-MS/MS):
LC-MS/MS specified in JIS K 0136 that satisfies the following requirements.
1) High-Performance Liquid Chromatograph
(i) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45 ºC.
(ii) Column: A 2-mm - 3-mm inner diameter 50-mm - 150-mm long 1.6-µm - 3.0-µm
particle diameter stainless steel column tube filled with silica gel to which octadecyl
chemically bonds. The specification is according to the mass spectrometer specification
(3)
.
2) Mass Spectrometer
(i) Ionization method: Electro-Spray Ionization (ESI) method
(ii) Ion detection method: Selected Reaction Monitoring
b) Ultrasonic generator: An ultrasonic washer can be used.
c) Concentrator: An evaporator whose temperature can be adjusted up to 40 ºC.
d) Porous diatomaceous earth cartridge column: A column that is filled with the porous
diatomaceous earth (capacity: 5 mL) (4)
e) Graphite carbon-NH2 laminate cartridge column: A 6-mL cylinder on which 500 mg of
graphite carbon and 500 mg of aminopropyl silylation silica gel is laminated (5).
Note (3) The column is sold under the names ACQUITY UPLC HSS C18, etc.
(4) The column is sold under the names Chem Elut (5 mL), etc.
(5) The column is sold under the names Envi-carb/LC-NH2 (500 mg/500 mg, 6 mL), etc.
Note (6) After measuring the specific gravity of sample, calculate the concentration of materials
subjected to measurement in the analytical sample.
(7) Note that the volume of the solution may expand as a result of ultra-sonication. It is
recommended to leave it at room temperature for a while when it expands.
Comment 2 The specific gravity (density) can be calculated by placing a 10-mL volumetric flask
642
Testing Methods for Fertilizers (2018)
Note (8) Confirm the solution to elute before conducting the testing.
(9) There is a possibility for agricultural chemicals to vaporize if it is dried up excessively.
Note (10) There is a possibility for agricultural chemicals to vaporize if it is dried up excessively.
(11) If there is a possibility that the concentration of agricultural chemicals in the sample
solution exceeds the maximum limit of the calibration curve, dilute a predetermined
amount of the sample solution with methanol.
(5.4) Measurement: Conduct the measurement as indicated in JIS K 0136 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph Mass Spectrometer used in measurement.
a) The measurement conditions of High-Performance Liquid Chromatograph/Mass
spectrometer: An example of measurement conditions for a High-Performance Liquid
Chromatograph/Mass Spectrometer is shown below. Set up the measurement conditions
considering it:
1) High-Performance Liquid Chromatograph
(i) Column: A silica gel column (2-mm - 3-mm inner diameter, 50-mm - 150-mm long,
1.6-µm - 3.0-µm particle diameter column) to which octadecyl chemically bonds.
(ii) Flow rate: 0.2 mL/min - 0.5 mL/min
(iii) Eluent: A: Ammonium formate solution (0.1 mmol/L)−formic acid solution (0.1 v/v%)
[1+1]
B: Acetonitrile formate solution (0.1 v/v%)
(iv) Gradient: 0 min (50 %B) → 15 min (95 %B) → 20 min (98 %B) → 30 min (50 %B)
643
Testing Methods for Fertilizers (2018)
c) Sample measurement
1) Subject 5 µL of the sample solution to the same procedure as in b) 2) - 3) (12).
2) Obtain the content of materials subjected to measurement from the calibration curve of the
peak area or height to calculate materials subjected to measurement in the analytical sample.
Note (12) Confirm that the ratio against the peak area ratio or height ratio of the standard
solution is within the range of about ± 30 %. In addition, the peak area ratio or height
ratio depends on the concentration.
(5.5) Calculation
Calculate the respective concentration of agricultural chemicals in the analytical sample by the
following formula. in the analytical sample by the following formula:
644
Testing Methods for Fertilizers (2018)
= (A × B × 10) /C
Comment 3 A recovery testing was conducted using fluid home garden-use mixed fertilizer (3
kinds) and fluid mixed fertilizer (2 kinds), as a result, the mean recovery at additive
level of 4000 µg/kg and 400 µg/kg (However, 4000 µg/kg and 400 µg/kg of
pyrethrin contain total of pyrethrin I/II) were 77.0 % - 104.5 % and 85.6 % - 107.9 %
respectively.
Table 2 shows results and analysis results from a collaborative study for test method
validation.
Additionally, the minimum limit of quantification for respective agrichemicals of the
test method is about 10 µg/kg.
645
Testing Methods for Fertilizers (2018)
Table 2 Analysis results of a collaborative study for the testing method validation of multicomponent analysis of
2) Additive 3) 4)
Mean Recovery RSD r RSD R
Agrichemicals Sample name Number of amount
1)
laboratories (µg/kg) (µg/kg) (%) (%) (%)
Abamectin Home garden-use mixed fertilizer1 8 286.8 333.3 86.1 13.3 14.4
B1a Home garden-use mixed fertilizer2 8 358.9 416.7 86.1 13.4 14.8
Home garden-use mixed fertilizer3 8 425.8 500.0 85.2 8.6 11.6
Fluid mixed fertilizer1 8 288.6 333.3 86.6 7.1 8.5
Fluid mixed fertilizer2 8 405.5 500.0 81.1 7.1 7.2
Ivermectin Home garden-use mixed fertilizer1 8 298.9 333.3 89.7 14.9 15.0
B1a Home garden-use mixed fertilizer2 8 382.5 416.7 91.8 14.1 19.3
Home garden-use mixed fertilizer3 8 431.1 500.0 86.2 9.8 10.9
Fluid mixed fertilizer1 8 298.8 333.3 89.6 10.1 12.8
Fluid mixed fertilizer2 8 405.2 500.0 81.0 3.8 5.8
Eprinomectin Home garden-use mixed fertilizer1 8 293.5 333.3 88.1 7.0 10.4
B1a Home garden-use mixed fertilizer2 8 361.9 416.7 86.9 9.2 14.3
Home garden-use mixed fertilizer3 8 425.3 500.0 85.1 7.0 10.0
Fluid mixed fertilizer1 8 277.3 333.3 83.2 9.0 12.0
Fluid mixed fertilizer2 8 398.2 500.0 79.6 7.5 11.6
Rotenone Home garden-use mixed fertilizer1 8 276.8 333.3 83.1 5.7 7.8
Home garden-use mixed fertilizer2 8 353.5 416.7 84.8 9.8 12.5
Home garden-use mixed fertilizer3 8 426.6 500.0 85.3 6.6 8.5
Fluid mixed fertilizer1 8 263.5 333.3 79.1 11.0 12.3
Fluid mixed fertilizer2 8 385.2 500.0 77.0 5.7 12.1
Piperonyl Home garden-use mixed fertilizer1 8 318.2 333.3 95.5 8.1 13.2
butoxide Home garden-use mixed fertilizer2 8 395.6 416.7 94.9 8.4 13.6
Home garden-use mixed fertilizer3 8 450.3 500.0 90.1 4.6 9.3
Fluid mixed fertilizer1 8 299.7 333.3 89.9 7.4 11.0
Fluid mixed fertilizer2 8 435.8 500.0 87.2 5.8 7.4
1) Number of laboratories used in analysis
2) Gross mean (n = number of laboratories × number of repeated tests (2))
3) Repeatability (relative standard deviation)
4) Reproducibility (relative standard deviation)
Table 2 (Continued)
2) Additive 3) 4)
Mean Recovery RSD r RSD R
Agrichemicals Sample name Number of amount
1)
laboratories (µg/kg) (µg/kg) (%) (%) (%)
Pyrethrin I Home garden-use mixed fertilizer1 8 160.7 186.0 86.4 9.3 11.9
Home garden-use mixed fertilizer2 8 202.2 232.5 87.0 12.6 12.8
Home garden-use mixed fertilizer3 8 228.6 279.0 81.9 5.4 8.8
Fluid mixed fertilizer1 8 158.2 186.0 85.1 6.8 10.4
Fluid mixed fertilizer2 8 223.1 279.0 80.0 8.5 9.1
Pyrethrin II Home garden-use mixed fertilizer1 8 131.1 147.3 89.0 6.5 9.7
Home garden-use mixed fertilizer2 8 163.2 184.2 88.6 10.8 13.6
Home garden-use mixed fertilizer3 8 182.0 221.0 82.4 5.4 8.9
Fluid mixed fertilizer1 8 126.2 147.3 85.7 7.8 11.4
Fluid mixed fertilizer2 8 180.2 221.0 81.5 6.3 8.3
646
Testing Methods for Fertilizers (2018)
References
1) Toshiharu YAGI, Masayuki YAMANISHI, Yuji SHIRAI: Simultaneous Determination
of Agricultural chemicals in Fluid Fertilizer by Liquid Chromatography/Tandem Mass
Spectrometry, Research Report of Fertilizer, Vol.4, p. 36 - 48 (2011)
2) Toshiharu YAGI, Masayuki YAMANISHI, Yuji SHIRAI and Masato SHIBATA:
Simultaneous Determination of Six Kind of Agricultural Chemicals in Fluid Fertilizer
by Liquid Chromatograph-Tandem Mass Spectrometer (LC-MS/MS): A Collaborative
Study, Research Report of Fertilizer, Vol.5, p. 48 - 59 (2012)
(6) Flow sheet for simultaneous analysis of 6 kinds of agrichemicals: The flow sheet for
simultaneous analysis of 6 kinds of agrichemicals in fertilizer is shown below.
Extract
647
Testing Methods for Fertilizers (2018)
Extract
Aliquot (5 mL)
Sample solution
Measurement LC-MS/MS
648
Testing Methods for Fertilizers (2018)
Peak No.
1
4
Intensity / arb.units
7
0 Retention Time / min 20 0 0 Retention Time / min 20
649
Testing Methods for Fertilizers (2018)
(2) Summary
This testing method is applicable to compost and straw, raw materials of compost. This testing
method is classified as Type D and its symbol is 8.3.2.a-2017 or AG-C-2.a-1.
Extract respective agricultural chemicals in fertilizers or raw materials with acetonitrile and water,
refine by using a porous diatomaceous earth column, a gel permeation chromatograph and a
synthetic magnesium silicate cartridge column, and then measure with an electron capture detector
equipped gas chromatograph to obtain compounds subjected to analysis in an analytical sample. In
addition, the performance of this testing method is shown in Comment 7.
marked line.
(4) Apparatus and instruments: Apparatus and instruments are shown below.
a) Gas Chromatograph (GC): GC specified in JIS K 0114 that satisfies the following
requirements.
1) Sample injector: An injector that enables split less system.
2) Capillary column: A capillary column (0.25-mm inner diameter and 30-m long) made of
fused silica. 14 % cyanopropylphenyl -86 % dimethyl polysiloxane chemically bonds to the
inner surface of a capillary column with 0.25 µm thickness.
3) Detection unit: Electron capture detector (ECD)
b) Gel permeation chromatograph (GPC): Preparative liquid chromatograph specified in JIS
K 0135 that satisfies the following requirements. No detector is required.
1) Sample injector: A sample injector that can inject 5 mL of sample solution.
2) Column: A 20-mm inner diameter 300-mm long stainless-steel column tube filled with
styrendivynylbenzene copolymer system hard gel
3) Guard column: A 20-mm inner diameter 100-mm long stainless steel column tube filled
with styrendivynylbenzene copolymer system hard gel
4) Fraction collector: A fraction collector that can set up a fraction to which agrichemical
components elute.
c) Shaking apparatus
d) Concentrator: An evaporator whose temperature can be adjusted up to 40 ºC.
e) Filter: A funnel for filtering under reduced pressure (compatible filter diameter: 60 mm)
f) Porous diatomaceous earth cartridge column: A column that is filled with the porous
diatomaceous earth (capacity: 20 mL).
g) Synthetic magnesium silicate cartridge column: A cartridge column that is filled with 910
mg of synthetic magnesium silicate.
h) Membrane filters: Made of PTFE (pore size is no more than 0.5 μm)
Comment 2 Column for GC is sold under the production name DB-1701, Rtx-1701, SPB-1701,
etc. A column which has actually isolated compounds subjected to analysis should be
used.
Comment 3 Gel permeation chromatograph (GPC) is a preparative liquid chromatograph that
collects a fraction of a material subjected to measurement sieved and isolated by
packing materials of the column for GPC according to the size of the molecule of the
material. A column for GPC is sold under the production name Shodex CLNpak
EV-2000 AC, etc. In addition, a gurd column for GPC is sold under the production
name Shodex CLNpak EV-G AC, etc.
Comment 4 A funnel for filtering under educed pressure is sold under the production name
KIRIYAMA Funnel SB-60, KIRIYAMA Funnel SU-60, etc.
Comment 5 A porous diatomaceous earth cartridge is commercially sold under production name
Chem Elut (20 mL), etc.
Comment 6 Synthetic magnesium silicate is commercially sold under the production names such
as Sep-Pak Florisil Plus Long Cartridge (910 mg).
651
Testing Methods for Fertilizers (2018)
Comment 7 A membrane filter is sold under the production name HLC-DISK 25 Solvent system
(pore size: 0.45 µm), DISMIC 25JP050, Millex FH (diameter: 25 mm, pore size:
0.45 µm), etc.
Note (3) There is a possibility for agricultural chemicals to vaporize if it is dried up excessively.
most of the elute dries up, send a nitrogen gas to dry up the elute (3), and add 2 mL of hexane
to dissolve the residue.
Note (4) If there is a possibility that the concentration of respective agricultural chemicals in
the sample solution exceed the maximum limit of the calibration curve, dilute a
predetermined amount of the sample solution with 2,2,4-trimethylpentane−acetone
(4+1).
(5.4) Measurement: Conduct the measurement as indicated in JIS K 0114 and as shown below.
Specific measurement procedures are according to the operation method of a Gas
Chromatograph used in measurement.
a) The measurement conditions of Gas Chromatograph: An example of measurement
conditions for a Gas Chromatograph is shown below. Set up the measurement conditions
considering it:
1) Sample injection method: split less injection method (1min)
2) Temperature of sample injector: 250 °C
3) Capillary column: A capillary column (0.25-mm inner diameter, 30-m long, 0.25 µm layer
thickness) made of fused silica. 14 % cyanopropylphenyl−86 % dimethyl polysiloxane
chemically bonds to the inner surface of the capillary column with 0.25 µm thickness.
4) Column bath temperature: 60 °C (1 min) → (20 °C/min) → 180 °C→ (2 °C/min) →
260 °C→ (5 °C/min) → 275 °C (1 min)
5) Carrier gas: helium, Flow rate: 1.5 mL/min
6) Addition gas: Nitrogen, Flow rate: 60 mL/min
7) Detection unit: Electron capture detector (ECD)
8) Detector temperature: 280 °C
c) Sample measurement
1) Subject 1 μL of sample solution to the same procedure as in b) 1)
2) Obtain the content of materials subjected to measurement from the calibration curve of the
653
Testing Methods for Fertilizers (2018)
peak area or height to calculate materials subjected to measurement in the analytical sample.
References
1) Tomoharu NOZAKI: Determination of Organic Chloride Pesticides in Composts using
Gas Chromatography-Electron Capture Detector (GC-ECD), Research Report of
Fertilizers, Vol.10, p. 41 - 60 (2017)
654
Testing Methods for Fertilizers (2018)
(6) Flow sheet for simultaneous analysis of chloride pesticides: The flow sheet for
simultaneous analysis of chloride pesticides in fertilizer is shown below.
Filter
Filter paper (type 5 B), 300-mL round-bottom flask (1)
under reduced presuure
←Cleanup of the vessel and residue with 50 mL of acetonitrile
Extract Pool washing with the filtrate
Extract
Concentration
40 °C
under reduced presuure
← 20 mL of sodium chloride saturated solution,
Porous diatomaceous earth cartridge column
Cleanup (1)
Leave at rest for 5 minutes
[300-mL round-bottom flask (2)]
←Wash the 300-mL round-bottom flask (1) 3 times with about 20 mL
of hexane, add to the same cartridge column successively
←60 mL of hexane
Concentration
40 °C
under reduced presuure
Exsiccation Nitrogen gas
←10 mL of cyclohexane−acetone (4+1)
Filter Membrane filter (pore size: no more than 0.45 µm)
Filtrate
655
Testing Methods for Fertilizers (2018)
Concentration
40 °C
under reduced presuure
Exsiccation Nitrogen gas
←1 mL of 2,2,4-trimethylpentane−acetone (4+1)
Sample solution
Measurement GC(ECD)
656
Testing Methods for Fertilizers (2018)
8.4 Sodium
8.4.a Flame atomic absorption spectrometry
(1) Summary
This testing method is applicable to fertilizers containing organic matters. This testing method is
classified as Type D and its symbol is 8.4.a-2017 or Na.a-1.
Pretreat an analytical sample with incineration and hydrochloric acid, spray into an acetylene−air
flame, and measure the atomic absorption with sodium at a wavelength of 589.0 nm or 589.6 nm to
obtain sodium (Na) in an analytical sample. In addition, the performance of this testing method is
shown in Comment 3.
Comment 1 Instead of the sodium standard solution in (2) b), a sodium standard solution (Na 0.1
mg/mL 1 mg/mL or 10 mg/mL for Atomic Absorption Spectrometry traceable to
National Metrology can also be used.
657
Testing Methods for Fertilizers (2018)
Note (4) Example of charring and incineration procedure: After raising the temperature from
room temperature to about 250 °C in 30 minutes to 1 hour, continue heating for about
1 hour and further raise to 550 °C in 1 to 2 hours.
Comment 2 The procedure in (4.1) is the same as the procedure in (4.1.2) in 4.2.1.a.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0121 and as shown below.
Specific measurement procedures are according to the operation method of the atomic
absorption spectrometer used in measurement.
a) Measurement conditions for the atomic absorption spectrometer: Set up the measurement
conditions for the atomic absorption spectrometer considering the following:
Analytical line wavelength: 589.0 nm or 589.6 nm
b) Calibration curve preparation
1) Spray the sodium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation into a flame, and read the indicated value at a
wavelength of 589.0 nm or 589.6 nm.
2) Prepare a curve for the relationship between the sodium concentration and the indicated value
of the sodium standard solutions for the calibration curve preparation and the blank test
solution for the calibration curve preparation.
c) Sample measurement
1) Transfer a predetermined amount of the sample solution (the equivalents of 0.1 mg - 1 mg as
Na) (5) to a 100-mL volumetric flask.
2) Add hydrochloric acid (1+23) to the marked line.
3) Subject to the same procedure as in b) 1) to read the indicated value.
4) Obtain the sodium content from the calibration curve, and calculate the sodium (Na) in the
analytical sample.
Note (5) Sample a predetermined amount of sample solution according to the device model in
Note (2).
Comment 3 Additive recovery testing with triplicates measurement was conducted using fish
caked powder, fish waste processed fertilizers, rape seed meal and its powder,
composted sludge fertilizers and compost, as a result, the mean recovery rate at the
additive concentration of sodium in the range of 1 % (mass fraction) - 10 % (mass
fraction) was 97.0 % - 103 %.
The results of the repeatability tests on different days using fish caked powder
(sample to which sodium chloride is added) and compost to evaluate precision were
658
Testing Methods for Fertilizers (2018)
References
1) Kimie KATO, Masaki CHIDA and Toshifumi FUJITA: Method Validation for
Determination of Sodium in Fertilizer by Atomic Absorption Spectrometry , Research
Report of Fertilizers, Vol.8, p. 61 - 69 (2015)
(5) Flow sheet for sodium testing method: The flow sheet for sodium testing method in
fertilizers is shown below:
Sample solution
659
Testing Methods for Fertilizers (2018)
Sample solution
Aliquot
100-mL volumetric flask, water
(predetermined amount)
←Hydrochloric acid (1+23) (up to the marked line)
Measurement Atomic absorption spectrometer
(589.0 nm or 589.6 nm)
660
Testing Methods for Fertilizers (2018)
Comment 1 Guanylurea sulfate is commercially sold by Kanto Chemical Co., Inc. and Tokyo
Chemical Industry Co., Ltd.
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph (HPLC): HPLC specified in JIS K 0124 that
satisfies following requirements.
1) Column: A 7.5-mm inner diameter 100-mm long stainless steel column tube filled with
weak acid ion-exchange resin.
2) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45ºC.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 190 nm.
b) Magnetic stirrer:
c) High speed centrifugal separator: A centrifugal separator that can work at 8000 × g - 10000
× g.
661
Testing Methods for Fertilizers (2018)
Comment 2 A column is sold under the production name Asahipak ES-502C 7C.
Note (3) If there is a possibility that the guanyourea nitrogen (GU-N) concentration in the
sample solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of supernatant with water.
(4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement.
(5) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Note (6) If there is a possibility that the guanyourea nitrogen (GU-N) concentration in the
sample solution will exceed the maximum limit of the calibration curve, dilute a
predetermined amount of solution with water.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions is shown below. Set up the measurement conditions
considering it:
1) Column: A weak acid ion-exchange resin column (7.5-mm inner diameter, 100-mm long,
5-µm - 10-µm particle diameter)
2) Column bath temperature: 40 °C
3) Eluent (1): Dissolve 3.92 g of potassium dihydrogenphosphate and 0.12 g of phosphoric acid
in water to make 1000 mL. Filter with a membrane filter (aperture diameter: no more than
0.5-µm) made of hydrophilic PTFE.
662
Testing Methods for Fertilizers (2018)
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the guanylurea nitrogen (GU-N) content from the peak height using the calibration
curve to calculate guanylurea nitrogen (GU-N) in the analytical sample.
Comment 5 This testing method enables the simultaneous measurement of biuret nitrogen (B-N),
urea nitrogen (U-N), dicyandiamide nitrogen (Dd-N), guanidine urea (Gd-N) and
guanylurea nitrogen (GU-N). In that case, see 5.10.a Comment 5.
Comment 6 Additive recovery testing was conducted using a preparation sample for a guanylurea
fertilizer (one brand). As a result, the mean recovery at additive level of 36.7 %
(mass fraction), 35.2 % (mass fraction) and 33.4 % (mass fraction) were 103.8 %,
104.6 % and 105.6 % respectively.
The results of the repeatability tests on different days using a guanylurea fertilizer to
evaluate precision were analyzed by one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability. Table 2 shows results
and analysis results from a collaborative study for testing method validation.
Additionally, the minimum limit of quantification of this testing method is about
0.006 % (mass fraction).
663
Testing Methods for Fertilizers (2018)
References
1) Masahiro ECHI, Yasuharu KIMURA and Yuji SHIRAI: Determination of Urea Nitrogen,
Biuret Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea
Nitrogen in Fertilizer by High-Performance Liquid Chromatography: A
Single-Laboratory Validation, Research Report of Fertilizer, Vol.10, p. 72 - 85 (2017)
2) Norio FUNAKI and Yasuharu KIMURA: Determination of Urea Nitrogen, Biuret
Nitrogen, Dicyandiamide Nitrogen, Guanidine Nitrogen and Guanyl urea Nitrogen in
Fertilizer by High-Performance Liquid Chromatography (HPLC): A Collaborative Study,
Research Report of Fertilizer, Vol.10, p. 86 - 100 (2017)
664
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for guanylurea nitrogen in fertilizers is shown
below:
1.00 g
200-mL ground-in stopper Erlenmeyer flask
analytical sample (powdery)
← 100 mL of water
Extraction Stir to mix, 10 minites
Stand still
1.00 g
100-mL volumetric flask
analytical sample (fluid)
←About 50 mL of water
Extraction Shake to mix
← Water (up to the marked line)
Ground-in stopper centrifugal precipitate tube,
Centrifugal separation
8000 ×g - 10000 ×g , 5 minutes
665
Testing Methods for Fertilizers (2018)
Peak name
(1) Urea nitrogen (2) Biuret nitrogen (3) Dicyandiamide nitrogen
(4) Guanidine nitrogen (5) Guanylurea nitrogen
Measurement conditions for HPLC
Column: Asahipak ES-502C 7C (7.5-mm inner diameter, 100-mm long, 9-μm
particle diameter)
Other conditions are according to the example of HPLC measurement conditions
in (4.2) a)
666
Testing Methods for Fertilizers (2018)
(3) Apparatus and instruments: Apparatus and instruments are shown below.
a) High-Performance Liquid Chromatograph (HPLC): HPLC specified in JIS K 0124 that
satisfies following requirements.
1) Column: A 4.6-mm inner diameter 250-mm long stainless steel column tube filled with
silica gel, to which octadecyl, ion-exchange group for strong acidity and strong basic
anion-exchange group, chemically bond.
2) Column bath: A column bath whose temperature can be adjusted to 30 ºC - 45 ºC.
3) Detection unit: An absorptiometric detector that can measure at wavelength around 290 nm.
667
Testing Methods for Fertilizers (2018)
Note (2) Use a silicone stopper instead of a glass stopper as the solution is heated.
(3) Steam easily expels silicon stoppers, so while lightly holding down the stopper from
the top with your finger, shake it so that the water drops on the inside of the flask
come down as much as possible. In addition, conduct this procedure before and after
the heating procedure.
(4) The ground-in stopper centrifugal precipitate tube should be made of polypropylene,
etc. to not affect the measurement
(5) 7.2 cm - 18.9-cm of rotor radius and 3000 rpm of revolutions makes about 1700 × g
centrifugal force.
(6) If there is a possibility that the uric acid (U-acid) concentration in the sample solution
will exceed the maximum limit of the calibration curve, dilute a predetermined amount
of supernatant with phosphate solution.
(7) 7.2-cm - 8.9-cm of rotor radius and 10000 rpm of revolutions makes about 8100 × g -
10000 × g centrifugal force.
Comment 2 Instead of the procedures in (4.1) e), it is allowed to filter with a membrane filter
(pore size: no more than 0.5-µm) made of hydrophilic PTFE and the filtrate can be
the sample solution.
(4.2) Measurement: Conduct the measurement as indicated in JIS K 0124 and as shown below.
Specific measurement procedures are according to the operation method of a
High-Performance Liquid Chromatograph (HPLC) used in measurement.
a) Measurement conditions for a High-Performance Liquid Chromatograph (HPLC): An
example of measurement conditions is shown below. Set up the measurement conditions
considering it:
1) Column: A silica gel column, to which octadecyl, ion-exchange group for strong acidity and
strong basic anion-exchange group chemically bond (4.6-mm inner diameter 250-mm long 3-
μm particle diameter )
2) Column bath temperature: 40 °C
668
Testing Methods for Fertilizers (2018)
3) Eluent (1): Dissolve 1.54 g of ammonium acetate with water to make 1000 mL. Transfer 900
mL of the solution to mix with 100 mL of methanol. Filter with a membrane filter (aperture
diameter: no more than 0.5-µm) made of hydrophilic PTFE.
4) Flow rate: 0.4 mL/min
5) Injection volume: 10 µL
6) Detection unit: An absorptiometric detector, measurement wavelength: 290 nm
Comment 3 Eluent can be prepared as follows. Dissolve 15.4 g of ammonium acetate with water
to make 1000 mL and store in a refrigerator. At the time of usage, dilute a
predetermined volume of the solution by a factor of 10 to mix with methanol of
volume ratio 1/9 and filter with a membrane filter (aperture diameter: no more than
0.5-µm) made of hydrophilic PTFE.
c) Sample measurement
1) Subject 10 μL of sample solution to the same procedure as in b) 1)
2) Obtain the uric acid (U-A) content from the peak area or height using the calibration curve to
calculate the uric acid (U-A) in the analytical sample.
Comment 4 In addition to uric acid, allantoin and allantoic acid can be measured simultaneously
by this measurement method (when using a Scherzo SS-C18 column). Note that the
detection wavelength of allantoin and allantoic acid is 210 nm.
Comment 5 Additive recovery testing was conducted using one brand of a compound fertilizer, a
composted sludge fertilizer, a mixed compost fertilizer and compost. As a result, the
mean recovery at additive level of 0.1 % (mass fraction), 0.01 % (mass fraction) and
0.005 % (mass fraction) were 92.4 % - 101.8 %, 85.3 % - 105.0 % and 92.5 % -
114.1 % respectively.
The results of the repeatability tests on different days using one brand of a compound
fertilizer, a composted sludge fertilizer, a mixed compost fertilizer and compost to
evaluate precision were analyzed by one-way analysis of variance. Table 1 shows the
calculation results of intermediate precision and repeatability.
Additionally, the minimum limit of quantification of this testing method is about
0.0008 % (mass fraction).
669
Testing Methods for Fertilizers (2018)
(5) Flow sheet for testing method: The flow sheet for irea acid in fertilizers is shown below:
1.00 g
200-mL ground-in stopper Erlenmeyer flask
analytical sample (powdery)
← 100 mL of phosphate solution
Heat at 60 ºC ± 2 ºC for 30 minutes
Heating
while shaking to mix at every 10 minutes
670
Testing Methods for Fertilizers (2018)
Reference: Chromatogram of the standard solution for calibration curve preparation of uric acid is
shown below.
Reference diagram HPLC chromatogram of the uric acid standard solution (50
µg/mL) for calibration curve preparation
671
Testing Methods for Fertilizers (2018)
(1) Purposes
This article explains the procedure to validate characteristics of testing methods which will be listed
in the Testing Methods of Fertilizers. In addition, when testing institutes conduct a test which is
not included in the Testing Methods of Fertilizers, a procedure to evaluate the validity of the test
method should conform to a method stipulated in this article.
Additionally, this article targets chemical testing methods. However, this article is not applicable to
the extraction method of the content of effective figures (acid-, alkaline-, citrate- and water-soluble)
in a powdery sample or a solidified fertilizer.
Comment 1 The contents of effective figures (acid-, alkaline-, citrate- and water-soluble) are
stipulated in a notification of the Ministry of Agriculture, Forestry and Fisheries. In
addition, the change of measurement conditions such as an extraction temperature
may affect an observed value in some cases. Therefore, no changes will be
implemented in the extraction method of the contents of effective figures in a
powdery fertilizer and a solidified fertilizer for the present and the application of this
article is limited to the change of a measurement method (including refining of
extract, etc.).
attestation on which the characteristics of prescribed properties and their uncertainty and
metrological traceability are stated.
k) Blank sample: An analytical sample not containing components subjected to analysis (2).
l) Addition sample: An analytical sample the content of whose components subjected to
analysis is known, or an analytical sample to which reference materials are added (3) (4) or
compounded (3).
m) Natural contamination sample: A test sample prepared from fertilizers which naturally
contain the components subjected to analysis such as harmful components.
n) Distribution sample: An analytical sample prepared from fertilizers (5) which are
manufactured in a fertilizer production factory, etc.
o) Surrogate: A material which is added to an analytical sample in order to conduct a
pre-process operation, correct yields in respective steps of measurement procedure and
confirm recovery, whose chemical structure is identical or similar to a target component.
p) SN ratio: Intensity ratio of a signal (response value) S originating from the analysis target and
a signal (usually noise) N based on the other factors.
Note (1) In reality, the certified value of a certified reference material, the chemical composition
of a compound, the added content of a reference material, etc. and others.
(2) Reagents, etc. containing a target matrix can be used in the case that there is no
distribution fertilizer used as a blank sample for a recovery test and the confirmation
of the minimum limit of quantification, etc.
(3) Mix a component subjected to analysis with a mortar, etc. to sufficient uniformity
(4) In the case of adding a standard solution, vaporize the solvent sufficiently conducting
measures such as letting it stand for one night.
(5) A fertilizer containing components subjected to analysis whose formation or form
changed due to a chemical or physical process (a granulation process, etc.).
References
1) JIS K 0211: Technical terms for analytical chemistry (General part) (2013)
2) JIS K 0214: Technical terms for analytical chemistry (Chromatography part) (2013)
3) JIS Q 0035: Reference materials−General and statistical principles for certification (2008)
4) JIS Z 8101-2: Statistics−Vocabulary and symbols−Part 2: Statistical quality control terms
(1999)
5) JIS Z 8402-1: Accuracy (trueness and precision) of measurement methods and results -
Part 1: General principles and definitions (1999)
6) ALINORM 09/32/23 Joint FAO/WHO Food Standards Prorgamme: Repot of the Thirtieth
Session of the Codex Committee on Methods of Analysis and Sampling, Codex
Alimentarius Comission Thirty-second Session (2009)
7) ICH Harmonised Tripartite Guideline, Validation of Analytical Procedures: Text and
Methodology Q2(R1), International Conference on Harmonisation of Technical
Requirements for Registration of Pharmaceuticals for Human Use (ICH) (2005)
673
Testing Methods for Fertilizers (2018)
(3.2) Selectivity
(3.2.1) Case of Chromatography
Conduct a procedure for a blank sample and confirm that there is no peak (interference peak)
which affects the measurement of components subjected to analysis (6). In addition, in the case of
the simultaneous measurement of multi components, confirm that adjacent peaks are sufficiently
separated (6).
Note (6) Resolution (R) should be 1.0 or more at minimum though 1.5 or more is preferable.
Comment 2 Resolution (R) is used as a separation indicator of peaks. If Resolution (R) is 1.5 or
more, the adjacent two peaks are sufficiently separated and they do not affect a
measurement, whether a peak height or a peak area is used. If Resolution (R) is 1.0 or
more, the adjacent two peaks are sufficiently separated and they do not affect a
measurement, whether a peak height or a peak area is used.
Resolution (R) can be obtained using a peak width by the formula (1a). In addition, if
the peak is a normal distribution, it can be obtained using a peak width at half height
by the formula (1b). With the data processing device of a chromatograph, the formula
(1b) is often used to obtain Resolution (R).
−
Resolution ( ) = ・・・(1a)
1
( )
2× +
1.18 × ( − )
Resolution ( ) = ・・・(1b)
,
+ ,
,
: Peak width at half height of Peak 1
,
: Peak width at half height of Peak 2
674
Testing Methods for Fertilizers (2018)
Note (7) A test method such as Molecular absorption spectrometry, Atomic absorption
spectrometry or Titration analysis which does not isolate with a measurement
instrument.
(8) Absorbance, titer, etc.
References
1) AOAC Official Methods of Analysis Appendix K: Guidelines for Dietary Supplements
and Botanicals, AOAC INTERNATIONAL (2012)
2) JIS K 0114: General rules for gas chromatography (2012)
3) JIS K 0124: General rules for high performance liquid chromatography (2011)
4) The notification by the Director of Evaluation and Licensing Division, Pharmaceutical
and Food Safety Bureau, the Ministry of Health, Labour and Welfare: “Guideline on
Bioanalytical Method Validation in Pharmaceutical Development”, July 11, 2013,
Yaku-shoku-Sinsa-Hatsu-0711 No.1 (2013)
Note (9) The blank test solution for the calibration curve preparation can be included.
(10) In order to avoid nonlinear confusion due to the variation of sensitivity, etc., conduct
measurements randomly for each replicate determination.
(11) Absorbance, fluorescence intensity, peak height, peak area, etc.
(12) The difference between a signal obtained by measurement and a signal estimated using
a regression equation.
Comment 3 It is recommended that the 95% confidence interval of an intercept (a) includes the
origin (0).
Comment 4 Though it is usable if the coefficient of determination (r2) is 0.99 or more, it is
recommended that the coefficient of determination (r2) is 0.999 or more for a precise
analysis. If it is less than 0.99, use the equation of a higher order or study the
conversion of a numerical value.
Comment 5 The mean of residuals is 0 and the residuals indicate a random pattern.
References
1) AOAC Official Methods of Analysis Appendix K, Guidelines for Dietary Supplements
and Botanicals, AOAC INTERNATIONAL (2012)
2) Thompson, M., Ellison, S.L.R, Wood, R., Harmonized guidelines for
single-laboratoryvalidation of methods of analysis, Pure & Appl. Chem. 74 (5), 835-855
675
Testing Methods for Fertilizers (2018)
(2002)
3) CLSI EP9 A2 Ed. 2, Method Comparison and Bias Estimation Using Patient Samples,
Clinical and Laboratory Standards Institute (2002)
(3.4) Trueness
As the estimation method of trueness, the methods are recommended in the following order. (1)
Use of a certified reference material (3.4.1), (2) Comparison with an observed value by a validated
method (3.4.2) and (3) Recovery test (3.4.3).
In addition, if a surrogate is used, it is recommended that a recovery is about 40% or more.
Comment 6 A warning limit is given using the formula (2) which is obtained from a collaborative
study for the characterization of a certified reference material.
µ: Certified value
: Reproducibility standard deviation in a collaborative study
: Repeatability standard deviation in a collaborative study (14)
n: The number of analytical samples to repeatability test
: Pure between-laboratory standard deviation in a collaborative study
Note (13) The evaluation procedure of the difference between a measurement result and a
certified value (characteristic value) is shown in Reference 1 Procedure to compare
an observed value and a certified value.
(14) It may be expressed as within-laboratory standard deviation ( ) in some cases.
a) In case 12 or more samples are available: Conduct respective tests of 12 or more test
samples composed of addition samples, natural contamination samples or distribution samples
according to a new test method and a standard test method, create the correlation chart of
676
Testing Methods for Fertilizers (2018)
observed values with two methods for each sample and calculate the inclination (b) and the
intercept (a) of a regression line, and a correlation coefficient (r). Further confirm a prediction
interval.
However, in case that the width between the minimum and the maximum observed value is
small, conduct the paired samples t-test to confirm that a significant difference is not
observed.
Comment 7 It is recommended that the 95% confidence interval of an inclination (b) includes 1,
the 95% confidence interval of an intercept (a) includes the origin (0) and the
correlation coefficient (r) is no less than 0.99.
b) In case fewer samples are available: With regard to 3 or more test samples of different
concentration, conduct respective repeatability addition tests using 4 analytical samples
according to the new test method and a standard test method, confirm the homoscedasticity of
the results of 2 groups and conduct a t-test for each concentration to confirm that significant
difference is not observed under the two-sided significant level of 5%.
(3.4.3) In case neither certified reference material nor other validated test methods are usable
For 3 or more test samples of different concentration, conduct respective repeatability tests using
3 analytical samples and evaluate by obtaining the recovery using the mean of the observed values.
The criteria of the trueness are shown in Separate sheet: The target of trueness and the criteria
of precision in respective concentration levels.
References
1) AOAC Official Methods of Analysis Appendix K: Guidelines for Dietary Supplements
and Botanicals, AOAC INTERNATIONAL (2012)
2) Thompson, M., Ellison, S.L.R, Wood, R.,: Harmonized guidelines for
single-laboratoryvalidation of methods of analysis, Pure & Appl. Chem. 74 (5), 835-855
(2002)
3) Linsinger, T.,: Comparison of a measurement result with the certified value, European
Reference Materials' application note 1, European Commission - Joint Research Centre
Institute for Reference Materials and Measurements (IRMM) (2010)
4) Joint FAO/WHO Food Standards Programme: Procedural manual Twenty-second
edition, Codex Almentarius Comission (2013)
5) ICH Harmonised Tripartite Guideline: Validation of Analytical Procedures: Text and
Methodology Q2(R1), International Conference on Harmonisation of Technical
Requirements for Registration of Pharmaceuticals for Human Use (ICH) (2005)
(3.5) Precision
Evaluate reproducibility and repeatability by a collaborative study (3.5.1). Or evaluate an
intermediate precision and repeatability by a repeatability test (3.5.2).
Note (15) In case the number of laboratories which have required facility/instruments is limited,
677
Testing Methods for Fertilizers (2018)
References
1) AOAC Official Methods of Analysis Appendix K: Guidelines for Dietary Supplements
and Botanicals, AOAC INTERNATIONAL (2012)
2) Thompson, M., Ellison, S.L.R, Wood, R.,: Harmonized guidelines for
single-laboratoryvalidation of methods of analysis, Pure & Appl. Chem. 74 (5), 835-855
(2002)
3) AOAC Official Methods of Analysis Appendix D: Guidelines for Collaborative Study
Procedures To Validate Characteristics of a Method of Analysis, AOAC
INTERNATIONAL (2005)
4) Horwitz , W.: Protocol for the Design , Conduct and Interpretation of
Method-Performance Studies,Pure & Appl. Chem.,67 (2), 331 - 343 (1995)
Comment 8 In case permissible content and equivalent level is 1.0 mg/kg or more, the Minimum
Limit of Quantification (LOQ) of harmful components and restricted components,
etc. should be no more than 1/5 of the permissible content and equivalent level. In
case permissible content and equivalent level is no more than 1.0 mg/kg, the
Minimum Limit of Quantification (LOQ) should be no more than 2/5 of the
permissible content. Moreover, it is recommended that the Minimum Limit of
Quantification of main components/major components and material components
should be no more than 1/5 of minimum volume to be contained and the minimum
content of a distribution fertilizer. In addition, in case the Minimum Limit of
Quantification exceeds 1/5 of these minimum volumes, conduct the above-mentioned
repeatability test, confirm the Minimum Limit of Quantification and state clearly the
fact in the applicable range of a test method.
Comment 9 There are some methods to estimate Minimum Limit of Quantification. The methods
differ depending on whether they are based on an instrument analysis or not and
678
Testing Methods for Fertilizers (2018)
10 ×
Estimated value of Minimum Limit of Quantification (LOQ) = ・・・(4)
References
1) ICH Harmonised Tripartite Guideline, Validation of Analytical Procedures: Text and
Methodology Q2(R1), International Conference on Harmonisation of Technical
Requirements for Registration of Pharmaceuticals for Human Use (ICH) (2005)
2) The notification by the Director of Evaluation and Licensing Division, Pharmaceutical
and Food Safety Bureau, the Ministry of Health, Labour and Welfare: The text
(operation method) on Bioanalytical Method Validation”, October 28, 1997, Iyaku-Shin
No. 338 (1997)
Comment 10 There are some methods to estimate Minimum Limit of Detection. The methods
differ depending on whether they are based on an instrument analysis or not and
depending on instruments used. A method different from the methods shown in
(3.7.1) to (3.7.3) is allowed. However, the definition of a method and Minimum
679
Testing Methods for Fertilizers (2018)
Note (20) In case of a repeatability test using 7 analytical samples, the value is 1.94. In case of
using 10 analytical samples, the value is 1.83.
References
1) ICH Harmonised Tripartite Guideline: Validation of Analytical Procedures: Text and
Methodology Q2(R1), International Conference on Harmonisation of Technical
Requirements for Registration of Pharmaceuticals for Human Use (ICH) (2005)
2) The notification by the Director of Evaluation and Licensing Division, Pharmaceutical
and Food Safety Bureau, the Ministry of Health, Labour and Welfare: The text
(operation method) on Bioanalytical Method Validation”, October 28, 1997, Iyaku-Shin
680
Testing Methods for Fertilizers (2018)
(3.8) Robustness
Robustness should be studied when an analysis method is developed, and the estimation method
depends on the type of analysis method to be developed. Robustness expresses the reliability of an
analysis method when its analysis conditions are intentionally changed. If an observed value tends
to be easily affected by the variation of an analysis condition, it is necessary to consider a method to
control an analysis condition appropriately or to state the fact as a precaution in a testing method.
The evaluation of robustness enables the establishment of a series of parameters such as Resolution
related to system conformance. Similarly, the confirmation of these parameters ensures that the
validation of an analysis method is maintained in a daily analysis.
Typical variation factors are as follows.
(3.8.1) Common variation factors: Typical variation factors common to various kinds of test
methods are as follows.
a) Extraction time, extraction temperature
b) Stability of a test solution in respective steps
c) Reagent’s grade
References
1) ICH Harmonised Tripartite Guideline, Validation of Analytical Procedures: Text and
Methodology Q2(R1), International Conference on Harmonisation of Technical
Requirements for Registration of Pharmaceuticals for Human Use (ICH) (2005)
2) The notification by the Director of Evaluation and Licensing Division, Pharmaceutical
and Food Safety Bureau, the Ministry of Health, Labour and Welfare: The text
(operation method) on Bioanalytical Method Validation”, October 28, 1997, Iyaku-Shin
No. 338 (1997)
3) Thompson, M., Ellison, S.L.R, Wood, R., : Harmonized guidelines for
single-laboratoryvalidation of methods of analysis, Pure & Appl. Chem. 74 (5), 835-855
(2002)
681
Testing Methods for Fertilizers (2018)
Obtain the total mean (m) of the replication test results and the certified value (μ), and the
absolute value ( ) of the difference of the two values by the formula (R1.1). Next, obtain the
standard uncertainty ( ) of the certified value of a certified reference material by the formula
(R1.2), and obtain the standard uncertainty ( ) of the total mean by the formula (R1.3). Calculate
the combined standard uncertainty ( (∆ ) ) of by the formula (R1.4) using the obtained
and . Further, calculate an expanded uncertainty ( ∆ ) by the formula (R1.5) using the
coverage factor ( = 2).
Compare and ∆ to confirm that the criterion (the formula (R1.6)) is satisfied, that is,
is no more than ∆ .
The absolute value ( ) of the difference of the total mean of repeatability test
results and a certified value= | − | ・・・(R1.1)
%
The standard uncertainty ( ) of the certified value = ・・・(R1.2)
Criterion ≦ ∆ ・・・(R1.6)
682
Testing Methods for Fertilizers (2018)
: True value
683
Testing Methods for Fertilizers (2018)
684
Testing Methods for Fertilizers (2018)
: the total mean ( )of the effective data of collaborative study results
Note (1) In case < , assume = (that is, the pure between-laboratory variance
( )= 0 in the formula (R2.5)) and let the formula (R2.6) form = .
(2) The rounding of a numerical value is not executed in the middle of the calculation.
(3) The mean and the standard deviation are expressed rounding to the digit of the
observed value. The relative standard deviation is expressed rounding to the first
decimal place.
(3) Calculation procedure of intermediate precision and repeatability by the replicate test
results on different days
(3.1) Estimation of a true value and a variance
In an actual statistical analysis, a true value (μ), a true test day variance (σ(T)2) and a true
repeatability variance (σr2) are unknown. Therefore, they are replaced with estimated values
obtained from the eplicate test results on different days and are expressed as a mean (m), test days
variance (s(T)2) and a repeatability variance (sr2) respectively.
685
Testing Methods for Fertilizers (2018)
686
Testing Methods for Fertilizers (2018)
Note (4) In case < assume = (that is, the test days variance ( ( ) ) in the
formula (R2.12) = 0) and let the formula (R2.13) form ( ) = .
(4) Examples of the calculation of intermediate precision and repeatability by the replicate
test results on different days.
An example of repeatability test results on different days of citric acid-soluble phosphoric acid
using sample 1 and sample 2 containing phosphite is shown in Table 4. Conduct one-way analysis
of variance for the test results of respective samples to obtain the unbiased variance (V) of
respective variation factors (Table 5).
Examples of the calculation of intermediate precision and repeatability for the sample 1 and
sample 2 using the formula (R2.11) to the formula (R2.17) are shown in Table 6-1 and 6-2. In
addition, the results of respective standard deviation are expressed rounding to the digit of the
observed value and the results of the respective relative standard deviations are expressed rounding
to the first decimal place.
Table 4 Example of repeatability test results on different days (mass fraction (%))
Test day (factor) Total
Sample 1 2 3 4 5 6 7) mean
No. ( )1)
51.20 52.15 51.00 51.35 51.35 51.38 51.28
Sample 1 51.38
51.45 51.85 51.09 51.28 51.10 51.38 51.43
5.18 4.90 5.01 5.15 5.14 5.13 5.21
Sample 2 5.10
5.00 5.12 5.06 5.14 5.07 5.11 5.18
1) The mean is expressed rounding to the digit of the observed value.
687
Testing Methods for Fertilizers (2018)
Sample 1 Accidental
0.1253 7 0.01789
error (e)
Test days (T) 0.0478 6 0.00797 +2× ( )
Sample 2 Accidental
0.0448 7 0.00640
error (e)
Table 6-1 Calculation of intermediate precision and repeatability using the sample 1 replicate test results on
different days1)
Variation factor Unit Formula Calculation Result
Repeatability variance
= = 0.01789 0.01789
( )
Repeatability standard Mass
2) = = √0.01789 0.13
Deviation ( ) fraction (%)
Repeatability relative 0.1338
3)
% = × 100 = × 100 0.3
standard deviation( ) 51.38
− 0.17616 − 0.01789
Test days variance ( ) = = 0.07914
2
Intermediate variance ( ) = ( ) + = 0.07914+0.01789 0.09703
Intermediate standard Mass
2) = ( ) = √0.09703 0.31
Deviation ( ) fraction (%)
Intermediate relative ( )
= 0.3115
standard % = × 100 0.6
3)
51.38
deviation ( ) × 100
1) The rounding of a numerical value is not executed in the middle of the calculation.
2) The standard deviation is expressed rounding to the digit of the observed value.
3) The relative standard deviation is expressed rounding to the first decimal place.
688
Testing Methods for Fertilizers (2018)
Table 6-2 Calculation of intermediate precision and repeatability using the sample 2 replicate test results on
different days1)
Variation factor Unit Formula Calculation Result
Repeatability variance
= = 0.00640 0.00640
( )
Repeatability standard Mass fraction
2) = = √0.00640 0.08
Deviation ( ) (%)
Repeatability relative
0.0800
standard deviation % = × 100 = × 100 1.6
3)
5.10
( )
− 0.00797 − 0.00640
Test days variance ( ) = = 0.00078
2
Intermediate variance
= ( ) + = 0.00078+0.00640 0.00718
( )
689
Testing Methods for Fertilizers (2018)
Separate sheet: The target of trueness and the criteria of precision in respective concentration
levels
The target of trueness (recovery) and the criteria of precision in respective concentration levels to
evaluate Chromatography (1) and test methods other than Chromatography are shown in Table 1 and
Table 2. The target of trueness is generally within the recovery of Table 1. As for precision, the
permissible level may exceed them by a factor of 2.0.
690
Testing Methods for Fertilizers (2018)
691