Real Time Blood Sugar Monitoring: July 2017
Real Time Blood Sugar Monitoring: July 2017
Real Time Blood Sugar Monitoring: July 2017
net/publication/341878194
CITATIONS READS
0 9
2 authors:
Some of the authors of this publication are also working on these related projects:
PROTOTYPE OF VEHICLES POTHOLES DETECTION BASED BLOB DETECTION METHOD View project
Pelatihan Perakitan Sistem Filterisasi Air Minum Skala Rumah Tangga View project
All content following this page was uploaded by Dewiani Dewiani on 03 June 2020.
Abstract- This research aims to design an interface that connects the blood sugar measuring instrument with the mobile
phone as well as designing database systems and visual display for monitoring blood sugar levels in a patient's body.
Monitoring system that integrates glucometer, arduino, Bluetooth, and mobile phones, are used to transmit data from the
blood sugar measurement. Beside the position of patient there is glucometer along with android based mobile phone that will
be used to transmit the measured data to a web server that can be accessed by doctors to monitor the progress of patients'
health. Based on observations, we obtained error value of 23.33% for observation within 60 seconds, an error of 32% for
observation within 90 seconds, and the error of 31.11% for observation within 120 seconds.
114
International Journal of Industrial Electronics and Electrical Engineering, ISSN: 2393-2835 Volume-5, Issue-7, Jul.-2017
http://iraj.in
occurs when the body does not produce enough In glucometer, there is a blood strip containing the
insulin to maintain normal blood sugar levels or when blood sample as well serving as a sensor detecting
cells do not respond appropriately to insulin [7]. blood sugar. This sensor is equipped with glucose
reagent. When blood is dripped or glucometer test
strips attached to it, it will generate a potential
difference or voltage caused by the reaction between
blood glucose reagent on the transducer.[8] The
resulting voltage ranges from 0-5 volts. The voltage
in the form of electrical or analog quantity is entered
into the Analog to Digital Converter (ADC) and then
converted into a signal of 8 bits. ADC output signal
will go to the microcontroller for processing the data.
The results of data processing are then fed to the LCD
by the microcontroller to display the results. Keypad
on the glucometer is a key instruction for the
Fig. 1. Monitoring Blood SugarSystem Configuration microcontroller.[2] On the LCD there are 7 ports
(Seg1-Seg7) as a seven segment display and a 4 port
III. DESIGN MONITORING BLOOD SUGAR signal Com (COM1-COM4). Output from the ports
SYSTEM on this LCD glucometer in form of analog signals are
subsequently connected to arduino. The whole seven
The glucometer will be used in the subject (patient) to segment port together with Com signal port are
measure blood sugar levels. The readings in the form connected to the analog arduino input, starting from
of blood sugar levels (mg/dl) will then be sent to a analog input 0 to 10 (A0-A10). While the positive
receiving device contained on the medical side side of the battery on a glucometer port connected to
through the network telemedis. In general, schematic Vcc of 3.3 volts and the negative side of the battery is
diagram of a blood glucose monitoring system will be connected to GND port on arduino as a power supply.
designed is shown in Figure 1. In particular, the data In general, diagram LCD glucometer port that
transmission system used in this study are shown in connected to the port arduino, are shown in Figure 4.
Figure 2. The transmission system using the tool:
glucometer dr. test meter, arduino mega 2560, HC-05
bluetooth and mobile phone Samsung Galaxy Tab 2
GT-P3100.
115
International Journal of Industrial Electronics and Electrical Engineering, ISSN: 2393-2835 Volume-5, Issue-7, Jul.-2017
http://iraj.in
116
International Journal of Industrial Electronics and Electrical Engineering, ISSN: 2393-2835 Volume-5, Issue-7, Jul.-2017
http://iraj.in
Table 1. Error Value From Observations The doctor inserts
username and password in order to
access the login page. When the doctors choose icon
"PATIENT", then the names of some patients with
diabetes will appear, and then doctors can choose
which patients that they want to know his/her
development of the blood sugar condition. By
selecting the name of the patient, the screen then will
show a graph of measurement results of the patient's
blood in one patient as shown in Figure 11.
Due to the instability of glucometer delivery to the
mobile phone, the program is installed in the
mobile phone side. Hence, the mobile phone
functions as a direct transmitter to the server.
Measurement Blood Sugar Value Result Display on
the phone are shown in Figure 9. Through the
program interface,, the real value received by the
phone can be stopped and thensent to the server.
Data sent to the server is accessible to the doctoe. Fig. 12. Patient Data Table
Hence, the doctor can observe the patient's condition.
Login display for the doctor seen in Figure 10. Similar to the chart display, the display data from a
patient's blood sugar measurements in the form of
this table provide more detailed information related to
the value of accurate blood glucose measurement
results, with date and time of the measurement of
blood sugar. With the information in the form of
charts and tables, the doctor can analyze the patient's
condition, so that they can do a subsequent medical
action if necessary.
CONCLUSIONS
117
International Journal of Industrial Electronics and Electrical Engineering, ISSN: 2393-2835 Volume-5, Issue-7, Jul.-2017
http://iraj.in
due to the frequency of glucometer ADC at up to 200 Data Medis Pada Jaringan Telemedis Interaktif), Electrical
Engineering Study Program, Informatics Engineering sub-
KHz while the processor frequency is 125 kHz,
Study Program, Engineering Faculty Universitas Hasanuddin.
which does not meet the criterion of Nyquist theory. Makassar, 2014.
To overcome this problem, the program that can [2] Rizha Tamridho, “Construction of Blood Sugar Tester
received real value and then be stored and sent to the Instrumentation System” (Rancangbangun Alat Pengukur
server was made on android phone. Kadar Gula Darah), Universitas Indonesia, Depok, 2010.
[3] Shyr-Kuen Chen, Tsair Kao, Chia-Tai Chan, Chih-Ning
Huang, Chih-Yen Chiang, Chin-Yu Lai, Tse-Hua Tung, and
ACKNOWLEDGMENTS Pi-Chung Wang. A Reliable Transmission Protocol for
ZigBee-Based Wireless Patient Monitoring. IEEE
transactions on information technology in biomedicine, vol.
This work was part of the Hasanuddin University
16, no. 1, January 2012.
PUPT Research and Development (R&D) Project [4] B.G.Sudharshan, S.C Prasanna Kumar, Shashiraj Yadav R V.
funded from DGHE (Directorate General of Higher A Review of Various Technologies And Transmission Modes
Education), Ministry of Research, Technology and For Design And Development of A Transmission Enabled
Glucometer. International Journal of Research in Engineering
Higher Education, Republic of Indonesia. The R&D
and Technology. Volume: 03 Issue: 06, June-2014.
grant itself was granted for the financial year 2016. [5] Mala Lavanya, and Ashok Shigli. Reliable Transmission
The authors would like to deliver the sincere thanks Protocal for Gsm And Http Based on Wireless Patient
to all parties who involved and supported the Monitoring. International Journal of Advance Engineering
implementation of the research project. The research and Research Development,Volume 1, Issue 12, December –
2014.
works were also the research collaboration between [6] Poedjiadi, Anna, “Fundamental of Biochemistry” (Dasar-
the Electrical Engineering Department and the Dasar Biokimia), Jakarta: Universitas Indonesia Press, 1994.
Medical Science Faculty, Hasanuddin University [7] Amatsier, Yunita. “Fundamental of Nutrition Science”
(Prinsip Dasar Ilmu Gizi). Jakarta : Gramedia Pustaka Utama,
(UNHAS) Makassar Indonesia
2013.
[8] Ponco Indrianto, “Measurement Tool for Glucose Quantity
REFERENCES Containing in Blood” (Alat Pengukur Glukosa Dalam
Darah), Department of Electrical Engineering, Industrial
[1] Amil Ahmad Ilham, Elyas Palantei, and Santi, “Development Engineering Faculty, Universitas Katolik Soegijapranata,
of Medical Data Visual Display for The Interactive Semarang, 2004.
Telemedicine Network” (Pengembangan Tampilan Visual
118