User Manual: Explanation of Symbols
User Manual: Explanation of Symbols
User Manual: Explanation of Symbols
Hazardous voltage Consult Instructions for Use Unit must be disposed at an appropriate facility.
Philips and the Philips Shield are registered trademarks of Koninklijke Philips Electronics N.V. All other trademarks are property
of their owners. ©2012 Discus Dental, LLC. All rights reserved. 19-2992 040312
1. Product overview 1.2 Intended Use
The Accelerator is a in-office teeth whitening light designed to
1.4 Accelerator Assembly
be used by dental professionals for use in whitening discolored 1. Remove all items from shipping carton.
1.1 Packaging Contents teeth. The Accelerator’s LED light source has been designed to 2. Snap the casters into the feet of the Accelerator Base and
Each Philips Zoom! WhiteSpeed Whitening LED Accelerator deliver the optimal spectrum of light energy to activate Philips lock the casters.
(Accelerator) has been thoroughly inspected and tested at Zoom! Light-Activated In-Office Whitening Gel to deliver up
to 6-8 shades* of teeth whitening. Wavelength emissions are 3. Place the Accelerator Base upright on the floor. Insert the
the factory. Upon arrival at your facility, inspect the shipping
between 400 to 505 nanometers(nm) and the energy intensity Accelerator Stand into the Accelerator Base by aligning the
container and the device for damage. Any damage should be
delivered to the teeth can be set at three levels depending on pin in the Accelerator Stand with the slot in the Accelerator
reported immediately to your Philips representative. In the case
Base and pushing the Stand securely into place. 2 3
of a damaged shipment, please save all packaging materials. patient and case type. The Accelerator must be used at all times
in accordance with the terms and conditions of the Letter of 4. Assemble the lock washer with the big teeth facing each
1. LED Accelerator Head (1) Agreement (LOA) provided and signed prior to delivery of the other and place it on the bolt, followed by the Fender (large)
2. Accelerator Boom (1) Accelerator. washer. Reaching under the Accelerator Base, screw the bolt
3. Accelerator Stand with Power Pack (1) into the bottom of the Base, securing it to the Accelerator
*Based on the Vitapan Classical shade guide arranged in value Stand.
4. Accelerator Base (1)
order. Vita and Vitapan are trademarks of Vita Zahnfabrik, AG
5. Casters (5) 5. To attach the Accelerator Head, loosen the knobs by the
Germany.
6. Safety Glasses (3) socket at the end of the Accelerator Boom by 4-5 turns.
7. Parts Bag 1.3 Safety Precautions Gently, but firmly, push the Accelerator Head into the socket.
Allen Wrench (1) Do not attempt to use the Accelerator without a thorough It will be held securely in place by the socket springs.
Screw (1) 4 5
understanding of this manual, and the manual within the Philips 6. Place the Accelerator Boom onto the Accelerator Stand
Bolt with Knob (1) Zoom! Light-Activated In-Office Whitening Kit (Kit). Failure to by carefully aligning the hole in the Boom with the hole in
Lock Washer in 2 parts (1) do so may result in injury to you or your patients, or damage the yoke. Insert the screw into the hole and tighten. Keep
Large Fender Washer (1) to the product. Only allow trained and qualified personnel to tightening the screw until it is completely in the pocket. The
8. User’s Manual (1) operate the system. screw should be flush or inset into the pocket. The screw
has glue on the threads (yellow in color) that releases as
Caution the screw is tightened. Therefore it is important to finish
tightening the screw upon insertion or the glue may stop the
3
screw from being fully inserted.
• Eye protection must be used during operation
to ensure the safety of everyone in the treatment 7. Connect the electrical cord from the Accelerator Boom to
2 area. The intense light can pose hazards for the eyes. the cords from the Accelerator Stand and the Accelerator 6
Protective safety goggles or spectacles that filter out Head. The Accelerator head connector must be turned until
wavelengths between 400-505 nm must be worn while you feel it lock into place.
1
operating this device. 8. When you are ready to test the Accelerator, activate the unit
4 • NEVER LOOK DIRECTLY INTO THE LIGHT. Never by turning on the main power switch on the right side of the
5 direct the light output of the Accelerator in any Power Pack.
6 direction other than the area being treated. Before
activating the light, confirm the delivery head is properly 1.5 Set up
8 positioned and aimed at the treatment area. Plug the Accelerator power cord protruding from the
• DO NOT ATTEMPT TO GAIN ACCESS TO THE Accelerator Base directly into a properly rated wall outlet.
7
INSIDE OF THE DEVICE. Service is limited to 7
authorized service representatives of Philips. Any Caution
attempt to gain access to the internal workings of the
device by unauthorized personnel or without proper Inlet (the end of the cord) must be accessible to disconnect
authorization is prohibited, may result in injury, and may power if a problem occurs.
void any warranty or service obligations by Philips.
The Accelerator is now ready for use with a Philips Zoom!
Light-Activated In-Office Whitening Kit. Please refer to the Kit’s
manual for further set up and gel application instructions.
2 3
Other electronic products exceeding the limits defined in such
2. Philips Zoom! Light-Activated 5. Point out to the patient that progress of each session will be
tracked by 4 white lights on top of the Accelerator Head, EMC standards could, under unusual circumstances, affect the
be able to close the fuse door if the fuse rack is not inserted
correctly.
In-Office Whitening Treatment each indicating one quarter of a session. operation of the product.
10. Close the fuse door.
6. If a session must be interrupted, press the “Pause” button. The • Medical electrical products need special precautions regarding
2.1 Light Guide remaining session time will blink on the display. To reactivate EMC, and need to be installed and put into service according 11. Reinsert the power cord firmly into the housing.
The Light Guide is designed to ensure proper distance and to EMC information provided in this manual.
the session, press either the “Pause” or “Start” button. DO
alignment to maintain correct light exposure on the smile zone. 12. Reinsert the screw into the door and tighten with a
NOT PRESS THE “STOP” BUTTON AT ANY TIME DURING • The use of accessories and cables other than those specified Phillips head screwdriver.
Remove Light Guide from Kit and place on end of Accelerator THIS PROCESS OR AN ENTIRE 15-MINUTE SESSION may result in increased emission or decreased immunity levels.
Head. Turn on the Accelerator using power button located on WILL BE LOST.
• The product should not be used adjacent to other products
the side of the Power Pack. When the Light Guide is properly
and that if adjacent use is necessary, it should be observed to
Fuse Replacement
attached, the Power Pack display will read “Guide Status-Guide 7. At the end of each 15-minute session, the display will indicate
the number of remaining sessions left on the Light Guide. verify normal operation.
Attached” followed by “X of X Sessions”. Each new Light Guide
is loaded with four 15-minute activation sessions. Press “Start” to activate the light for the next session. • To avoid the risk of electrical shock, do not remove any System 1
8. When all sessions on a Light Guide are used, the display will panels.
2.2 Align Accelerator Head and Light Guide read “Sessions Complete” for three seconds, followed by • Refer servicing to qualified personnel.
to Retractor “Guide Status – Replace Guide”. Detach the Light Guide Fuse label
Once whitening gel application is complete (see Kit manual for from IsoPrep Retractor and pull Accelerator away from • Never operate the System with a damaged power cord.
instuctions), raise back of chair so patient is in a slightly reclined White power cord
patient. Storage
position (45 degree angle) and able to swallow comfortably The Accelerator Boom is designed to fold down to minimize
while maintaining light pressure on bite block. Slide Accelerator overall Accelerator footprint for convenient storage or to move
into place next to patient. Accelerator Boom should be movable, Tip the device. To do so, loosen Accelerator Boom knobs and fold
but stable. Accelerator Boom down and in towards the Accelerator Stand.
• Use High setting for maximum whitening, Low
To align the Accelerator Head: setting for maximum heat and sensitivity Disposal
1. Grasp the Accelerator Head with one hand and bring it close 2
management. However, please note that lower light Philips is concerned about the preservation of the environment.
to the patient’s mouth. Use your other hand to lift or lower intensity may reduce results. Please dispose of this product properly at the end of its useful
the Accelerator Boom as needed to make sure Accelerator life according to local laws and regulations for electronic Fuse door
Head is aimed directly at the patient’s teeth. • After each session, make sure the patient’s lips and gums equipment waste.
are consistently and thoroughly covered to Caution: Hazardous voltage
2. Use both hands to finalize positioning of Accelerator Head help prevent any soft tissue exposure to the LED
so that Light Guide and IsoPrep Retractor slots are aligned. lights and any whitening gel.
Tighten Accelerator Head and Accelerator Boom knobs (4). 4. Fuse replacement instructions Fuse label
Lock casters in position. While engaged, the Light Guide and
IsoPrep Retractor will allow the Accelerator Head to follow
minor movements of the patient’s head to ensure proper 3. Care and maintenance
Caution 3
alignment.
2.3 Running a whitening procedure cycle The Accelerator should be cleaned and disinfected after each 1. CAUTION: Unplug the Accelerator from the outlet.
Each Philips Zoom! Light-Activated In-Office Whitening whitening session. Do not spray or pour liquids directly onto Red fuse rack
Procedure consists of three 15-minute whitening sessions with the system. Apply cleaners to a soft cloth or paper towel before 2. Use a Phillips head screwdriver to remove the screw at the
a pause between each session to allow the operator to remove using. Do not use abrasive cleaners or solvents. The following bottom of the Power Supply housing. Set the screw aside for
and replace the whitening gel. Each Light Guide includes a fourth cleaners are recommended: Sanicloth, CaviWipes, SciCan, reassembly.
session for use on patients with shades A3 or darker. Birex, BioSurf, Cidex and IPA. Follow your cleaning products
3. Remove the door from the housing.
1. Ensure the patient is properly prepared according to the Kit instructions for use to ensure proper cleaning of the unit.
manual with whitening gel and all protective barriers in place. For optimum performance, ensure that the output window on 4. Pull out the white power cord from its connection and let it
the Accelerator Head remains clean and free of particles or hang freely.
2. Press the “Intensity” button to toggle through high, medium,
and low light intensity until the desired intensity is selected. streaks. 5. Look for the fuse label near the upper, right-hand side to iden-
tify the fuse location. (see figures 1- 2) 4 Tabs
This setting can be changed at any time during the whitening The Accelerator has no user-serviceable or user-replaceable
procedure. parts with the exception of two replaceable fuses that are
6. Using a small, thin bladed, flathead screwdriver, pry open the
3. Press the “Start” button to turn on light and activate session inside the Power Pack housing.
right-hand side of the fuse door. Then pry out the red fuse
timer. The Accelerator complies with relevant international and rack. (see figures 2-3)
national law and standards on EMC (electro-magnetic Fuses
4. Once light is on, a countdown will be displayed. The 7. Remove both fuses from the red fuse rack - note their loca-
compatibility) for this type of product when used as intended.
Accelerator will beep once when 3 minutes are remaining tion in the fuse rack. (see figures 4-5)
Such laws and standards define both the permissible
and beep three times in the final 3 seconds of the 15 minute electromagnetic emission levels from product and its required
session. When time reaches zero, light will turn off and a long 8. Snap in the new fuses at the top of the fuse rack, they should FRONT VIEW 5
immunity to electromagnetic interference from external
beep will sound. be closer to the side with the tabs (top). (see figures 4-5)
sources.
9. Reinsert the fuse rack into the fuse slot with the tabs first. BACK VIEW
Push the fuse rack in until it is securely in place. You will not
4 5
5. Trouble Shooting Guide Guidance and manufacturer’s declaration – electromagnetic emissions
The Zoom WhiteSpeed, Dental Teeth Whitening Lamp is intended for use in the electromagnetic environment specified below. The
Problem Causes Solution
customer or the user of the Zoom WhiteSpeed should assure that it is used in such an environment.
Display reads • One of the cables between power Make sure all 3 connectors are plugged in (Accelerator Emissions Compliance Electromagnetic environment – guidance
SYSTEM ERROR CALL pack and Accelerator head not plugged in head, back of the boom, into the power pack)
CUSTOMER SERVICE RF emissions Group 1 The Zoom WhiteSpeed uses RF energy only for its internal function. Therefore, its RF emissions are very low and are
If error code remains, call customer service CISPR 11 not likely to cause any interference in nearby electronic equipment.
The wireless transmitter in the foot pedal is battery powered only, and does not connect to public mains.
Display reads • New Light guide was first placed on Replace guide with a new guide that was not first
*GUIDE STATUS* Zoom Advanced Power or older system placed on a Zoom Advanced Power or older system RF emissions Class A
IMPROPER GUIDE CISPR 11 The Zoom WhiteSpeed is suitable for use in all establishments other than domestic and may be used in domestic
establishments and those directly connected to the public low-voltage power supply network that supplies buildings
Power line harmonics Class A used for domestic purposes, provided the following warning is heeded:
Display reads • One or more sessions was interrupted by If One or more sessions is interrupted by any of the IEC/EN 61000-3-2
*GUIDE STATUS* the off button 3 causes listed, the lost session can not be recovered Warning: This equipment/system is intended for use by healthcare professionals only. This equipment/system may
REPLACE GUIDE Power line flicker Complies cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation
before the guide is empty • One or more sessions was interrupted by measures, such as re-orienting or relocating the Zoom WhiteSpeed or shielding the location.
a power outage IEC/EN 61000-3-3
Display reads • Light guide is not fully seated Pull the light guide off and
*GUIDE STATUS* on the Accelerator head reinstall correctly
ATTACH GUIDE
When a guide is attached • Light guide was up on Place the light guide on the Accelerator head
upside down with the word TOP facing up
• The pins on the underside of the Call Customer service for a
Accelerator head are damaged Accelerator head replacement
6. Specifications
Philips Zoom! WhiteSpeed Whitening LED Accelerator
Classification Class 1 Type BF, IPXO, Equipment not suitable for use in the presence of flammable mixtures
Weight 11 kg / 24 lbs
6 7