Manual de Usuario de Centrifuge McKesson
Manual de Usuario de Centrifuge McKesson
Manual de Usuario de Centrifuge McKesson
OPERATION MANUAL
For use with MFR# 602.
McKesson Variable Speed Centrifuge
Thank you for choosing the McKesson Variable Speed Centrifuge. This precisely
built, durable centrifuge will give years of service to even the busiest office
practice. Our technical and customer support departments are ready to assist you
with any questions or comments you may have.
If ever you require an additional accessory or spare part, please contact your local
McKesson Medical-Surgical Inc. representative.
FEATURES
• Six place rotor for use with standard 5-10 ml tubes
Note: Can spin up to three 15 ml tubes
• Can spin 2 ml and 4-5 ml tubes with available adapters
• Linear speed control allows use from 500rpm- 4,000 rpm
• Dependable mechanical 30-minute timer with bell signal at end of run
• High Impact plastic housing for years of dependable use
• Safety switch in lid shuts off motor when lid is unlatched. Latch release
located on the right front side of unit
• Power indicator light tells when unit is powered on
CONSTRUCTION
The enclosures of the McKesson Variable Speed Centrifuge are made of high
impact plastic, which is acid and reagent resistant. The internal chamber is made
of stainless steel. A safety switch is built into the lid and disconnects the power to
the motor whenever the lid is unlatched. The motor is mounted on rubber mounts
to provide quieter running and less vibration. The rotor chamber can be easily
cleaned and disinfected.
ACCESSORIES
Each McKesson Variable Speed Centrifuge comes equipped with an fixed angle
head rotor with 6 place tube shields already installed. Optional 8-place rotor is
available on request.
2
UNPACKING AND INSTALLATION
The centrifuge is first packed in a Styrofoam and then put in a carton box for
shipping. Please examine the shipment upon receiving and retain all packing until
the unit has been tested. The warranty excludes any damage caused by shipping.
Inspect for any possible shipping damage. If you note any damage, please
contact your McKesson representative. Save the original packaging. If it becomes
necessary to return the instrument to us, we ask that it be returned in
the original packaging.
Unpack the shipping carton carefully and verify that the shipment is complete. It
should include the following items:
• Take the centrifuge from the Styrofoam and place it on a sturdy flat surface.
• Press the lid latch release button located on the right side and open the
centrifuge cover. A special packing Styrofoam disc is packed inside the lid of
the centrifuge to prevent the rotor or shields from moving during shipping.
REMOVE THIS PACKING DISC PRIOR TO OPERATION OF THE CENTRIFUGE
• The centrifuge rotor is pre-installed. No rotor installation and
calibration is required.
• The power inlet is located in the back of the centrifuge. Plug in one end of the
power cord to the inlet and the other end to a power outlet.
• Turn the speed control counterclockwise to “0” rpm.
Caution: Make sure the voltage of the power source matches the voltage
input of the centrifuge.
OPERATION
1. Set the timer control to “off”.
2. Plug the centrifuge into a power source outlet. Make sure the
voltage is correct.
3. Turn the power rocker switch in rear of unit to “on” position.
4. Open the lid by pressing the lid-opening button.
5. Place the specimen tubes into the tube shields of the rotor.
CAUTION!! Make sure that the specimen tubes are balanced; see safety
section below:
3
6. Close the lid firmly.
7. Set the speed control knob at or close to “0” rmp.
8. Set the timer clockwise to the desired time.
9. Rotate the speed control knob slowly to the appropriate speed.
10. After the centrifuge rotor has come to a complete stop (following the end of
the run bell), open the lid.
SAFETY
You’re PSS 602 centrifuge is a carefully designed and built instrument. When
operated properly, it presents no safety hazards, however, please take note of
these safety precautions.
LID
NEVER OPEN THE LID UNTIL THE ROTOR HAS COME TO A COMPLETE STOP!
LOAD BALANCE
Always make sure the load is balanced before operating the instrument.
It is very important that test tubes with equal loads are placed opposite each other
in the rotor. Also, when spinning a single specimen tube, another specimen tube
of the same size filled with water must be installed directly across from the
specimen tube to maintain balance. If the centrifuge is not properly balanced,
damage could occur. If excessive noise or vibration is noticed during a run, stop
the centrifuge and recheck the balance.
MAINTENANCE
This centrifuge is designed to be relatively maintenance free. The motor is a
brushless type and has permanently sealed bearings so oiling is not necessary. A
surface cleaner such as Formula 409 can be used to clean the outside of the
instrument. The rotor chamber and cover can be disinfected with a mild solution
of bleach or other disinfecting liquid and allowed to dry. Do not drench or
immerse the unit in any liquid! Be sure to disconnect the power before
attempting to clean.
4
CALIBRATION
According to Code of Federal Regulations Title 21, centrifuges require verification
or calibration as follows:
1. Before initial use.
2. After repair or adjustments.
3. 6 month after use.
BIOHAZARD
If a tube breaks, a biological hazard may exist. Unplug the instrument. Clean
using approved laboratory procedures. Always wear the proper protective clothing
and gloves. Use a suitable disinfectant and dispose of the sample properly. Refer
to the guidelines set forth by OSHA, CLIA, etc.
SERVICE
If you feel that your centrifuge needs factory service, please contact PSS
representative or speak to a customer service representative to determine the
nature of the problem. Simple problems may be resolvable over the telephone. If
the instrument needs to be returned, a manufacturer R.G.A. (return goods
authorization) number will be issued. Please reference the R.G.A. number on the
paperwork packed with the instrument, and on the outside of the carton
PROBLEM SOLVING
1. Centrifuge does not start
a) Check to determine if centrifuge is plugged in.
b) Check that timer has been set.
c) Check that lid is closed and latched.
d) Check that Power rocker switch in rear of unit is in the “on” position.
e) Check that the Power indicator light (on front panel) is lit. A failure of the
power light to illuminate could indicate a need for a new fuse.
5
2. Excessive noise when running. If the noise is irregular or different in tone, be
sure that there is no paper or debris in the rotor or rotor chamber. Make sure
that the load is balanced.
3. Indicator light fails to light.
a) Unit is not turned on
b) Lamp needs to be replaced
c) Fuse is blown
4. Fuse Replacement: Follow the steps below to replace the fuse
a) Unplug the power cord. Turn the on/off switch to “off”
b) Remove the power cord from the power inlet on the back of the centrifuge.
c) Locate the fuse holder. The fuse holder is a part of the power inlet.
d) Use a flat head screwdriver to pull out the snap-in holder
e) Remove the blown fuse and replace the same type and rating 250V/3A
f) Push the snap-in holder back in.
5. Intermittent motor operation; or no operation. Call your service department.
Specifications:
Speed 500 – 4,000 rpm
Speed accuracy ± 100 rpm
Width 10.2 inches
Depth 11.8 inches
Height 8 inches
Weight 18 lbs.