Zoll MSeries
Zoll MSeries
Options
• Biphasic Defib/ Monitor/ Recorder/AED
• Defib/ Monitor/ Recorder/AED
• Defib/Monitor/Recorder/Pacer/AED/SpO2
• Defib/ Monitor/ Recorder/Pacer/AED
Specifications
ECG Monitoring Patient connection: 3-lead ECG cable, 5-lead ECG cable, 12-lead ECG
cable, paddles or MFE pads. Input protection: fully defibrillator protected.
Special circuit prevents distortion of ECG by pacer pulse. (Pacer version only).
Implanted pacemaker spike display: dedicated circuitry detects most implanted
pacemaker spikes and provides standard display marker of spike on ECG trace.
ECG size: 0.5, 1, 1.5, 2, 3 cm/mV. Heart rate: digital display 0-300 bpm ±5%.
Heart rate alarm: On/Off displayed on monitor. User selectable, tachycardia 60-
280 bpm, bradycardia 20-100 bpm
Display Screen type: high-resolution display. Screen size: 5” (12.7 cm) diagonally.
Information: heart rate, lead/pads, alarm on/off, NIBP, 12-lead ECG, SpO2,
EtCO2, AED functions and prompts, defibrillator test function, self test function,
error corrections and faults, pacer functions (optional), code markers, alarm
selection and limits, delivered energy.
Defibrillator Waveform: damped sinusoid or ZOLL rectilinear biphasic. Energy selection:
selectable at 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 75, 100, 150, 200, 300, 360
Joules selected using controls on sternum paddle or device front panel.
Rectilinear biphasic energy selection up to 200 joules maximum. Charge time:
less than 7 seconds with a new fully charged battery. Depleted batteries will
result in a longer defibrillator charge time. Energy display: monitor display
indicates both selected and delivered energy. Synchronized mode: synchronizes
defibrillator pulse to patient’s R-wave. “SYNC” message displays on monitor.
Marker on display and recorder paper identifies R-wave discharge points.
Advisory function: single analysis or programmable auto re-analyze x3 with
programmable auto energy level selection, screen prompts, and voice prompts.
Change controls on apex paddle and device front panel. . Paddles: external
anterior/anterior adult and pediatric paddles. Adult paddles slide off to expose
pediatric paddles. Defibrillator advisory: evaluates electrode connection and
patient ECG to determine if defibrillation is required.
Pacemaker (Pacer version only) Type: VVI demand; asynchronous (fixed rate) when used
without ECG leads or in ASYNC pacing mode. Pulse: rectilinear, constant