9092/9092C CHLORINE/pH ANALYZER

Introducing the new Waltron 9092/9092C


Chlorine Analyzer for automatic, continuous
measurement of free chlorine, active chlorine,
ozone, chlorine dioxide and bromine in
aqueous media. Measurement is based on a
microprocessor based potentiostatic three
electrode system.

The Waltron 9092/9093 series instruments are used for


industrial process monitoring and process control.
Applications include drinking water treatment, cooling
water treatment, water management in swimming pools, or
the control of residual concentration Cl2 of chlorine-
treated waste water. The potentiostatic technology allows
for a much faster response time than membrane or
amperometric systems, ensuring short start-up times, easy
operation, and low maintenance.

FEATURES
Portable system – Model 9092C available
Automatic calibration option
Wide range analysis: 0-1000 μg/L, 0-10 mg/L
Automatic temperature and flow compensation

Fast response time (t90 = 30 sec)


No zero point adjustment required

Withstands pressures up to 10 bar (145 psi)


Analog and digital interface; data logging option
Minimal maintenance required

Automatic pH compensation for free chlorine reading

See reverse side for specifications.


9092/9092C CHLORINE/pH ANALYZER | GENERAL SPECIFICATIONS
PERFORMANCE
Range 0-1000 μg/L, 0-10 mg/L
Accuracy +/- 3% of reading
Response Time t90 = 30 sec
Temperature Compensation Automatic
pH Compensation Automatic
Calibration Automatic or manual
Analog Outputs 0-20mA, 4-20mA (shunt max 500 Ohm)
Digital Outputs Serial interface RS232
Alarms Flow, calibration, power
Power 100VAC - 240 VAC, 50/60 Hz

SAMPLE DELIVERY OPERATING CONDITIONS


Temperature: Sample Flow 3-18 L/h
Sample Temperature 32-140° F (0-60° C)
Ambient Temperature 32-131° F (0-55° C)
Pressure ≤ 10 bar (145psig)
Sample Conductivity ≥ 10 μS/cm; salt cell is required if conductivity is less

MECHANICAL
Dimensions (Full size 9091) Transmitter: Height=11” (27.7 cm), Width=7” (17.8 cm), Depth=7.75” (19.7 cm)
with wall-mount bracket, Depth = 4.5” (11.4 cm) without wall-mount bracket
Wet Section: Height=6” (15 cm), Width=4” (10 cm), Depth=6” (15 cm)
Dimensions (Portable 9091C) Transmitter: Height=5.5” (14 cm), Width=7” (18 cm), Depth=7” (18 cm)
Wet Section: Height=6” (15 cm), Width=4” (10cm), Depth=6” (15 cm)
Protection Class IP54
Piping Swagelok® tube diameter = 6mm

Materials used Measuring Electrode: gold


Counter Electrode: stainless steel 1.4571 (314)
Reference Electrode: Ag/AgCl in saturated KCl-solution