TOPLED-Service Manual
TOPLED-Service Manual
TOPLED-Service Manual
LED-V2.0 1
3.3.1.3 Replace the damper circle and the needle bearing of lamp holder.............21
3.3.1.4 Replace the damper circle and the needle bearing of electric-control handle.....
..22
3.3.2 System of light adjustment..
..............................................23
3.3.2.1 Replace the handle cover......................
.......................
..23
3.3.2.2 Replace the handle of the lamp cover......................................23
3.3.2.3 Replace the cover glass.................................................23
3.3.3Problem diagnosis and solution..........................................
..23
3.4 The electric control unit.................................................24
3.4.1 The circuit diagram of surgery shadowless lamp...........................
..24
3.4.2 The circuit connection of electric-control handle............................
..24
3.4.2.1 Replace the circuit board.............................................
..25
3.4.2.2 The circuit board principle chart.......................................
..25
3.4.3 Problem diagnosis and solution.........................................
...26
4 Maintain....................
..........................................
...26
LED-V2.0 2
1 Production introduction
The laminar flow type of TOPLED Series Surgery Shadowless Lamp is an ideal lighting
apparatus of the modern operating room. The airtight streamline lamp holder ensures that the
purified air can convect smoothly. TOPLED series surgery shadowless lamp adopts imported
LED as its light source, working with advanced focus lens and the tailormade lens holder,
which effectively focus the rays emitted by LED light source and meet the operation lighting
requirement of illumination intensity, shadowless degree and focus depth. Advanced design
idea used, line of beauty has been interpreted well. The spectrum of LED is close to the sun's
rays that effectively controlled the color temperature and coloration of the light. The LED has
a long service life and has the characteristics of cold light which provide favorable conditions
for the recovery of patients. The disassemble handle cover adopt engineering plastic, which
can make high-temperature autoclaving disinfect.
Model Code::
TOPLED-8080 double lamp-holder Surgery Shadowless Lamp
TOPLED-8000 single lamp-holder Surgery Shadowless Lamp
TOPLED series Surgery Shadowless Lamp consists of 4 parts: The rotation arm, the balance
arm, the lamp holder and the electric control unit. The lamp holder is used for lighting; The
balance arm is used for hanging lamp holder and making lamp holder keep equilibrium and
arbitrary to orientate within the range of activity; The rotation arm is used for hanging lamp
holder and balance arm and making them arbitrary to move and orientate within the range of
activity; The electric control unit is used for controlling the irradiation of lamp holder.
2 Installation
2.1 Condition and parameter
1) Installation personnel should understand whether users' installation terms accord with the
installation requirement for the surgery shadowless lamp. Usually the installation height
of the operating room is 2900-3100
2900-3100mm。
2) When the height of installation is lower than 2900mm or higher than 3100 mm, you should
contact manufacturer and put in the written requirement
requirement. And then manufacturer will
See the following list for details: The installation
customize according to user's request. (See
height and relevant size of the surgery shadowless lamp
lamp)
The installation height and relevant size list of of surgery Shadowless lamp
The height between the lamp and floor The lowest height
Mode code
(The installation height is 3000mm) of installation
TOPLED-8080 2030 2950
TOPLED-8000 2140 2850
3) Request the user to have the fixed embedded part at the installation place
place.
LED-V2.0 3
4) Request the user to save enough length of the power line at the installation place
place.
5) If decorative interlayer of operation ceiling used, request the user to elongate embedded
part to the inside of decorative interlayer and scoop up a 320mm hole at the installation
place.
Ex: Normal installation sketch map of elongating embedded part in the interlayer as follows
sketch map1
6) If the installation condition is special, please contact the manufacturer, the manufacturer
will provide installation plan.
2.2 Attentions
Installation of the surgery shadowless lamp has the strict request. When installing it, the user
should read carefully of the product manual.
1) Please do not scratch or knock the glass cover of the lamp.
2) If any unusual problem, please cut off power in time and apply for repair.
3) The surgery shadowless lamp is a hanging equipment. To avoid possible accident, please
check the fasteners frequently to make sure the fasteners are not flexible.
4) To avoid security accident, Personnel without the authorization of our company is NOT
allowed to disassemble or refit the surgery shadowless lamp.
The normal life span of the surgery shadowless lamp is ten years. And the equipment should
LED-V2.0 4
be discarded as useless mandatory when reaching the life span.
The explanation of the signs on the equipment
Embedded part
Chassis dust cover
(2)
Rotation arm
Fix board
(3) (6)
(4) A
Decorative hoop
(1:4)
(5) (4)
Protective cover
Curving cannula
Set screw of
protective cover
Locking ring
Lamp holder
Sketch map 2
Detailed installation steps as follows:
1) Put the fix board to horizontal position and fix it onto the embedded part (expansion
screws prohibited). (See Picture 1)
LED-V2.0 5
Picture 1
2) Connect the rotation arm firmly onto the fix board by using four groups of installation
bolts, and peel the red, blue and yellow green wires 10mm respectively, which are on the
outer side of the rotation arm. (See Picture 2)
Picture 2
LED-V2.0 6
3) Peel the red, blue and yellow green wires 10mm, which are on the balance arm side. And
use junction caps to connect the red, blue and yellow green wires on the balance arm side
to the red, blue and yellow green wires on the outer side of the rotation arm respectively.
Fix the balance arm onto the outer side of the rotation arm by using three M5*10 bolts.
(See Picture 3)
Picture 3
4) The decorative hoop drill through the balance arm, and append the transom of rotation arm.
(See picture 4)
Pre-cover Part
Chassis
Rotation Arm
Spring Arm
Decorative Hoop
Picture 4
5) Remove the bolts on the outer axis cover of the balance arm, push the cover upwards, and
remove the fixers. Plug the curving cannula of the lamp holder into the balance arm,
replace the fixers, pull back the cover, replace and screw the bolts. (See picture 5)
LED-V2.0 7
Picture 5
6) According to the circuit diagram(See Picture 37) , connect the external power supply line
with the terminal in the electrical box , cover the lid, and coat the dustproof cover. Replace
the decorative hoop on the dust cover, its top is near to the ceiling. The power line should
be fixed on the wall or the ceiling by the cable clamp or line slot.
Notice: The external power supply line should conform to requirement of GB5013.1 (The
ordinary wear-resisting rubber protective covering cord or the ordinary PVC protective
0.75mm²² and goes through a fixed main
covering cord); the line diameter should be 0.75mm
power switch which can pass and break simultaneously; the switch should conform to the
appliance””(1972) (Creepage distance and
requirement of IEC328 "Switch of the Electric appliance
Air clearance); The movement direction of switch should conform to requirement of GB
4205 "Standard movement direction of handling parts for controlling electrical
ontrolling
equipment " and symbol should conform to the requirement of GB9706.1 appendix
D(List D1)
D(List
7) Using the Allen key to adjust the braking bolt and the functioning range of the balance arm,
until the rotation arm and the lamp holder can both be positioned arbitrarily and smoothly
within functioning range and feel light and handy.
2.4 Examine and debug
1) After connected to power, the power indicator on the touch screen on the electric-control
handle will be lighted. Touch“ ” to turn on the lamp, indicator light turn from yellow to
green, touch the irradiation adjusting button to adjust the irradiation. When the surgery
LED-V2.0 8
finished, touch “ ” to turn off the lamp and then turn off the main power.
2) Rotate the light-focus handle at the center of the lamp-holder to focus and adjust the
lighting angle; the handle cover can be removed for sterilization, and the sterilized handle
cover is for manipulation by the surgeon. To avoid contamination, no other personnel
should touch the handle cover. (to remove the handle cover, push the button on the upper
part of the handle.)
3) If the fuse is damaged, please shut off the main power switch and remove the dustproof
cover and the cap of the electric box for replacing. (Should choose the fuse of specification
indicated on the inner label .)
5) If the surgery shadowless lamp is found unable to be positioned, please refer to
Installation and debugging---- step 7”
“Installation
6) Suggestion: a clean and wet duster cloth can be used to clean the exterior of the surgery
shadowless lamp, but liquid should not enter the interior, and formaldehyde or ozonated
water can be used to fumigate the equipment for sterilization, and use ultraviolet radiation to
sterilize; the handle cover can be sterilized in glutaraldehyde or being steamed at 134℃ for
20 minutes. Since users may be in a different country or region where the regulations for
medical sterilization might be different, please act according to the local regulation.
2.5Working condition of the surgery shadowless lamp
2.5Working
Normal use condition of the surgery shadowless lamp:
Voltage of the power supply AC 220V±4V 50Hz±1Hz
Working environment temperatur +10~+30℃
Relative humidity range ≤75%
Atmosphere range 700~1060hpa
This equipment cannot be used where there’s an air mixture of flammable anaesthetic gas and air
or oxygen and nitrous oxide.
Transportation condition: environment temperature: -40~+55℃ relative humidity: ≤93%
atmosphere: 500~1060 hpa
3 Function introduce, Accessory replace and Problem diagnosis
Here we introduce the functions of each parts and the partial components replacement
operation of TOPLED series surgery shadowless lamp (Use the double lamp as the example)
3.1 The rotation arm part
3.1.1 Chassis subassembly (See picture 6)
LED-V2.0 9
3
4 6
2 5
Picture 6
Fittings list
No. Name Purchase code
1 The electric box F7060-020301
2 The cover of electric box F7060-020302
3 The connection holder TB-1503
4 The fuse holder BZ001
5 Fuse RT1-32 *5A RT1-32 *1A
6 The switching power supply SP-240-30
3.1.1.1 Replace the fuse (See picture 7)
1) First turn off the power. If only the fuse outside of the electric box is damaged, remove
it and then replace it.
2) If the fuse inside of electric box is damaged, remove the cover of electric box and then
you can replace the fuse inside of it. (Notice: The fuse inside of the electric box is 1A)
Picture 7
3.1.1.2 Replace the switching power supply (See picture 8)
1) When the switching power supply is damaged, screw off the junction cap that connect
LED-V2.0 10
with the power wire and remove each wire that joins with the switching power supply.
2) Remove the bolt that fixes the switching power supply and then you can replace it.
Picture 8
3.1.1.3 Replace the connection holder in the electric box (See picture 9)
1) Remove the cover of electric box.
2) Remove the wire that connects with the connection holder.
3) Remove the bolt that fixes the connection holder and then you can replace it.
Picture 9
5 1
6 7
3
4
Picture 10
LED-V2.0 11
Fittings list
No. Name Purchase code
1 The cover of transom. F7060-0215
2 Central gasket F7060-0211
3 The braking bolt GB78-M12×20-1Cr
4 Cover of the bearing pedestal F7060-0212
5 The brush subassembly F7060-021400
6 Transom subassembly (long) F7060-020400
7 Transom subassembly (short) F7060-021000
3.1.2.1 Replace the brush suit (See picture 11)
1) Remove the cover of transom.
2) When inspecting the brush suit is damaged, loosen the bolt that fixes the brush suit and
then you can replace it.
Picture 11
LED-V2.0 12
Picture 12
3.1.3 The central axis group of transom:(See picture 13)
Picture 13
3.1.3.1 Replace the damper block, the conductive body and the scroll
bearing:(See picture 14)
LED-V2.0 13
1) Remove the cover of transom. Please refer to step-3.1.2.1.
2) Remove the wire from electric subassembly which connect to the conductive circle
3) Remove the cover from the bearing pedestal and the central gasket.
4) Remove the stop gasket 48 and orbicular nut. And then you can replace the damper
block.
5) Take the transom subassembly (long and short) away. And then you can replace the
conductive body and the scroll bearing.
6) When replacing the conductive subassembly, please pay attention to the suitable place
of the red, blue and yellow green wires.
5
9 3
6
· 7 2
4
8
Picture 14
Fittings list
No. Name Purchase code
1 Gasket 48 GB858-48-A3
2 Orbicular nut GB812-M48*1.5-45
3 Damper block F7060-0207
4 Cardo F7060-0208
5 Conductive subassembly F7060-020500
6 Insulated circle F7060-0206
7 Insulated pedestal F7060-0213
8 Scroll bearing 2007110 GB297-2007110
Double lamp:F7060-0209
9 Central axis
Single lamp:F7000-0103
LED-V2.0 14
3.1.3.2Replace the central axis:(See picture 15)
1) After remove scroll bearing, turn the central axis so as to another side.
2) Remove fixed bolt. And then you can replace the central axis.
Picture 15
3.1.4Problem diagnosis and solution
3.1.4Problem
Problem Possible diagnosis Solution
1. Connect the equipment to
1.The equipment is not connected to
the power supply
the power supply
2. Replace the fuse
2.The fuse is damaged
The POWER indicator 3. Replace the brush suit
3.The brush suit is damaged
4.Replace the conductive
on the touch screen 4.The conductive subassembly is
subassembly.
cannot be turned on damaged.
5. Replace the switching
5.The switching power supply is
power supply.
damaged
6. Replace the connection
6.The connection holder is damaged.
holder.
1.Use a multi-meter to check
1.Contact problem with the brush and
the 24V exporting circuit to
connection point
Irregular blinking of the bulb
2.There’re high-frequency emitter
2.Separate the surgery
the surgery shadowless equipments (such as
shadowless lamp’s power
lamp high-frequency electric bistoury,
supply and the power supply
microwave equipments) in the same
of other high-frequency
power supply network..
equipments.
1.The fix board was not adjusted to 1.Use a gradienter to check if
horizontal position when being the fix board is horizontal.
installed. 2.Reinforce the embedded
The rotation arm
2.Embedded part is lacking in part.
cannot be positioned inflexibility. 3.Adjust the braking bolt
arbitrarily 3.The braking bolt becomes flexible 4.Replace the damper block.
4.The damper block is damaged. 5.Replace the scroll bearing.
5.The scroll bearing is damaged.
LED-V2.0 15
3.2 The balance arm part
3.2.1 Adjust the balance arm (See picture 16)
Picture 16
Fittings list
No. Name Purchase code
1 The fixed holder of three-core plug F7060-0301
2 three-core plug Female 3-pole
3 The connective socket of the transom F7060-0302
4 The joint arm ACROBAT-2000N-18
5 The braking bolt GB77-M6×6-1Cr
LED-V2.0 16
Picture 17
3.2.1.2 Replace the damper circle and the needle bearing of the balance arm:(See
picture 18)
1) Remove the block circle 32 of axis and the braking bolt.
2) Remove the connective socket of the transom. And then you can replace the damper
circle and the needle bearing.
Picture 18
Fittings list
No. Name Purchase code
1 The block circle 32 GB894-32-65Mn
2 The damper circle F7060-0303
3 The needle bearing GB5801
3.2.1.3 Adjust the balance arm
1) The balance arm can rotate 95° in the vertical direction and 360° in the horizontal
direction.
2) Using the Allen key to adjust the braking bolt insures that the balance arm can both be
positioned arbitrarily and smoothly within functioning range and feel light and handy.
3) Using the Allen key to adjust the weight hanging bolt can control the lamp holder up or
down (See picture 19)
4) Using the Allen key to adjust the rotation bolt of the balance arm causes the lamp holder
in vertical place to achieve the highest position. (See picture 20)
LED-V2.0 17
Picture 19
Picture 20
LED-V2.0 18
3.2.2 Problem diagnosis and solution
Problem Possible diagnosis Solution
The balance arm can not be 1.The braking bolt becomes 1.Adjust braking bolt.
positioned arbitrarily flexible 2.Replace the damper circle.
2.. The damper circle is 3.Replace the needle bearing
damaged
3. The needle bearing is
damaged
The lamp holder up or down The bolt that control the Adjust the bolt on the balance
weight hanging is not well. arm that control the weight
hanging. (See picture 19)
The balance arm can only be up The bolt that control the Adjust the up and down rotation
rotated in the limit angle. rotation angle is not well. bolts.( See picture 20)
The balance arm electrified The three-core plug is Replace the three-core plug
instability or none of power damaged
3.3The lamp holder part
3.3The
3.3.1 The siphon part (See picture 21)
3 4
Picture 21
LED-V2.0 19
Fittings list
Picture 22
3.3.1.2 Replace the three-core plug (See picture 23)
1) Open the electric-control handle box and remove the wire on the circuit board that
connects with the three-core plug.
2) Carefully pull the three-core plug and then you can replace it.
LED-V2.0 20
Picture 23
3.3.1.3 Replace the damper circle of the lamp holder and the needle
bearing :(See picture 24)
1) First remove the upper cover and the viewfinder. Please find the step of how to replace
the viewfinder for details.
2) Loosen the wire which connect the curving cannula and the lamp holder, Remove the
bolt which connect the axis of rotation of the lamp holder and the lamp holder.
3) Remove the braking bolt and the orientation bolt
4) Remove the block circle 30 of axis and the axis of rotation of the lamp holder, then it
can be replaced.
2 3
Picture 24
LED-V2.0 21
Fittings list
No. Name Purchase code
1 The block circle 30 of axis GB894-30-65Mn
2 Damper circle of the lamp holder F7060-040103
3 Needle bearing GB5801-6905
4 Axis of rotation of the lamp holder TOPLED-80000202
3.3.1.4 Replace the damper circle and the needle bearing of electric-control
handle (See Picture 25)
1) First open the electric-control handle box. Please find the step one and two of how to
replace the circuit board for details.
2) Remove the radiator piece and then take the lock piece out.
3) Remove the braking bolt, the junction caps of wire connection and the long curving
cannula. Then you can replace the damper circle and the needle bearing.
2 3
Picture 25
Fittings list
No. Name Purchase code
1 Lock piece F7060-040104
2 Damper circle of the lamp holder F7060-040103
3 Needle bearing GB5801-6905
LED-V2.0 22
3.3.2 System of light adjustment
3.3.2.1 Replace the handle cover (See picture 27)
1) Press the button to remove the handle cover, and then sterilize it or replace it.
2) Rotating the handle can adjust the size of facula. Thus the product can meet the
requirement of the user.
Picture 27
LED-V2.0 23
3.4 The electric control part
3.4.1 The circuit diagram of surgery shadowless lamp (See picture 37) (Use the
double lamp as the example)
GND
LED+
RT1-30 5A
L +
AC220V RT1-30 1A LED
50HZ SP-240-30
DC30V
-
N LED-
GND
LED+
RT1-30 5A
+
RT1-30 1A DC30V LED
SP-240-30 -
LED-
Control panel
Picture 37
Picture 38
LED-V2.0 24
Fittings list
No. Name Purchase code
1 The touch panel TOPLED-80000303
2 The circuit board FHLED
3 The electric-control handle cover TOPLED-80000301
4 The electric-control handle box TOPLED-80000302
3.4.2.1 Replace the circuit board (See picture 39)
1) Turn off the power before disassembling the circuit board. And then open the
electric-control handle box.
2) Disassemble individual electrical wire and connectors that connect with the circuit board.
When operating, you should hold down the connection spot. Thus can avoid the circuit
board excessively distorting. When installing connectors, please avoid the similar
phenomenon as well.
3) Loosen the bolt that fixes circuit board and then replace the circuit board. When installing
the new circuit board, you should pay attention to the accuracy of connection.
Picture 39
3.4.2.2 The circuit board principle chart (See picture 40)
LED-V2.0 25
1 2 3 4 5 6
D C7 D1 D
R1
D2 202
18.432M 33PF
Y1
D3 202 R2 LD2
R15
C6 202 R3 9 8
+ + 101
VCC + + R14
R5 33PF GND 12 5 R13
302 13 4 101
VCC + + 101
C5
IC2 + + R12
1
Vcc 28
16 1
P2.2 101
10/16V 2
P2.3 P2.1 27
j1 3 26PWM2
RST P2.0/PCA2/PWM2
4 25 LD1
1 RxD/P3.0 P1.7/SCLK/ADC7 R11
5 24
2 TxD/P3.1 P1.6/MIS0/ADC6 9 8
6 23 101
3 XTAL2 P1.5/MOS1/ADC5 + + R10
7 22
4 XTAL1 P1.4/SS/ADC4 + +
8 21
5 INT0/P3.2 P1.3/ADC3 12 5 R9
9 20 101
6 INT1/P3.3 P1.2/ADC2 13 4
10 19 101
K1 EC1/T0/P3.4 P1.1/ADC1 + +
C 11 18 C
PWM1/T 1/P3.5 P1.0/ADC0 + + R8
C4 K2 PWM1
12 17PWM0
PWM3/P 2.4 P3.7/PWM0 16 1
104 13 16 101
P2.5 P2.7 15 Q1
14 102 R6
GND P2.6
SP1 D4
1 STC12C5604 R7
2 R4 202
102
3
4
IC1
LM2575 VCC
IN D5
FD1 4
1 1 FEEDBACK
B Vin B
2 4007
L1
C1 OUT 2
DC 221
GND 3
ON/OFF
D6
OUT
5819 C2
220/50V
5
1 470/16V C3
2 0.01
LED
R18
104
Q3
Q2 44N
PWM1 R16 8050
104
R17
104
A A
Title
Picture 40
3.4.3Problem diagnosis and solution
3.4.3Problem
Problem Possible diagnosis Solution
The surgery 1. The ill-contact problem with the 1. Use a multi-meter to check
shadowless lamp can circuit the 30V exporting circuit to the
not be turned on. 2. The circuit board problem circuit board
2. Replace the circuit board
4 Maintain
The user should carry on inspecting to the surgery shadowless lamp at least every three
months. Mainly inspect whether the fasteners become flexible or not; whether the spare bulb is
good or not; whether the lamp holder can balance locates or not; if problem found, please
promptly solve the problem according to the manual. If has the problem which cannot remove,
please promptly contact our company to avoid mistake.
LED-V2.0 26
Agent::
Global Exclusive Agent
Nison Instrument (Shanghai) Ltd.
Floor 15th, No.2 Hua Shan Road, Shanghai 200040, P. R. China
Tel:021-62728646
Fax:021-62728538
Zip:200040
Http://www.nison-shanghai.com/
Customer service hotline:800-820-6044
Manufacturer::
Manufacturer
SHANGHAI FUHONG SCIENCE & TECHNOLOGY DEVELOPMENT CO., LTD.
NO.8 Zhenchuan warehouse, Duimian Street, East Huaxia Road, Pudong New Area, Shanghai
Tel:021-58374808
Fax:021-58376324
Zip:201200
LED-V2.0 27