<Contents>
6 Maintenance ............................................................................................................ 44
6-1. Maintenance intervals and items ............................................................................. 44
6-2. Gas calibration method............................................................................................ 45
6-3. How to clean............................................................................................................ 46
6-4. Parts replacement ................................................................................................... 46
8 Troubleshooting ....................................................................................................... 51
2
1
1-1. Preface
Thank you for choosing our portable gas monitor GX-8000. Please check that the model number of the
product you purchased is included in the specifications on this manual.
This manual explains how to use the gas monitor and its specifications. It contains information required for
using the gas monitor properly. Not only the first-time users but also the users who have already used the
product must read and understand the operating manual to enhance the knowledge and experience before
using the gas monitor.
The gas monitor comes in several types for different combinations of gases to be detected. Check the
specifications of the gas monitor before use and conduct gas detection properly in accordance with
purposes. (See the list of gases to be detected at the end of this operating manual.)
In addition to this operating manual, an operating manual for the data logger management program (option)
is available for the gas monitor. Contact RIKEN KEIKI if it is needed.
3
1-4. Method of confirmation for Standards and
Explosion proof specification
This instrument has some specification depends on standard and explosion proof certificate. Please
confirm the detector specification before using. Please refer Declaration of Conformity that is at the end of
this manual if you have CE marking type. You can confirm instrument specification to see name plate as
follows.
TIIS certificate
Ex marking
CE marking
4
2
Important Notices on
Safety
DANGER
Nameplate location
Main unit
Battery unit
5
DANGER
The specifications of main units are as follows:
Pump circuit: Allowable voltage of 4.95 V, allowable current of 1.12 A, and
allowable power of 1138 mW
Combustible gas sensor circuit: Allowable voltage of 4.95 V, allowable current of 0.834 A, and
allowable power of 853 mW
Buzzer circuit: Allowable voltage of 4.95 V, allowable current of 0.431 A, and
allowable power of 441 mW
Main circuit: Allowable voltage of 4.95 V, allowable current of 0.717 A, and
allowable power of 733 mW
Backup circuit: 3.0 VDC, 10μA
Ambient temperature: -20℃~+50℃
Explosion-proof grade of main unit is ExiaⅡCT4X.
6
DANGER
About combination
Make sure that the product model on the nameplate and combination of main unit and battery unit
is correct.
Inappropriate combinations of models deviate from the range of explosion-proof certification.
IP protection class: IP20
About use
While conducting measurement in a manhole or confined space, do not lean over or look into the
manhole or closed space. It may lead to dangers because oxygen-deficient air or other gases
may blow out.
Oxygen-deficient air or other gases may blow out from the gas exhausting outlet. Never inhale
the air or gases.
High-concentration (more than LEL) gases may blow out. Never use fire near the gas monitor.
7
2-2. Warning cases
WARNING
Sampling point pressure
The gas monitor is designed to draw gases around it under the atmospheric pressure. If
excessive pressure is applied to the gas inlet and outlet (GAS IN, GAS OUT) of the gas monitor,
detected gases may be leaked from its inside, thus leading to dangers. Be sure that excessive
pressure is not applied to the gas monitor while used.
Do not connect the gas sampling hose directly to a location with a pressure higher than the
atmospheric pressure. The internal piping system may be damaged.
Handling of sensor
Do not disassemble the electrochemical type sensor or galvanic cell type sensor because they
contain electrolyte. Electrolyte may cause severe skin burns if it contacts skin, while it may cause
blindness if it contacts eyes.
If electrolyte is adhered on your clothes, that part on your clothes is discolored or its material is
decomposed. If contact occurs, rinse the area immediately with a large quantity of water.
Others
Do not throw the gas monitor into fire.
Do not wash the gas monitor in a washing machine or ultrasonic cleaner.
Do not block the buzzer sound opening. No alarm sound can be heard.
Do not remove the battery unit while the power is ON.
8
2-3. Precautions
CAUTION
Do not use the gas monitor where it is exposed to oil, chemicals, etc. Do not submerge the gas
monitor under water on purpose.
Do not use in a place where the gas monitor is exposed to liquids such as oil and chemicals.
The gas monitor, being compliant to IP67, is not water-pressure-resistant. Do not use the gas
monitor where a high water pressure is applied to it (under a faucet, shower, etc.) or submerge it
under water for a long time. The gas monitor is water-proof only in fresh water and running
water, and not in hot water, salt water, detergent, chemicals, human sweat, etc.
The gas inlet and outlet are not water-proof. Be careful not to let water such as rainwater get into
these parts. Because this may cause trouble and gas cannot be detected.
Do not use the gas monitor where water or dirt gets accumulated. The gas monitor placed at
such a location may malfunction due to water or dirt that gets into the buzzer opening, gas inlet,
etc.
Note that drawing in dirty water, dust, metallic powder, etc. will significantly deteriorate the
sensor sensitivities. Be careful when the gas monitor is used in an environment where these
elements exist.
Do not use the gas monitor in a place where the temperature drops below -20ºC or rises over 50ºC.
The operating temperature of the gas monitor is -20 to 50ºC. Do not use the gas monitor at
higher temperatures, humidities, and pressures or at lower temperatures than the operating
range.
Avoid long-term use of the gas monitor in a place where it is exposed to direct sunlight.
Do not store the gas monitor in a sun-heated car.
Observe the operating restrictions to prevent condensation inside the gas monitor or gas sampling
hose.
Condensation formed inside the gas monitor or gas sampling hose causes clogging or gas
adsorption, which may disturb accurate gas detection. Thus, condensation must be avoided. In
addition to the operating environment, carefully monitor the temperature/humidity of the sampling
point to prevent condensation inside the gas monitor or gas sampling hose. Please observe the
operating restrictions.
Verify that the pump driving indicator is rotating before using the gas monitor.
If the pump driving indicator is not rotating, gas detection cannot be performed properly. Check
whether the flow rate is lost.
9
CAUTION
Others
Pressing buttons unnecessarily may change the settings, preventing alarms from activating
correctly. Operate the gas monitor using only the procedures described in this operating manual.
Do not drop or give shock to the gas monitor. The water-proof and explosion-proof properties
and accuracy may be deteriorated.
Do not use the gas monitor while recharging it.
Whereas the gas monitor can detect oxygen, combustible gases, carbon monoxide, and
hydrogen sulfide, the operating environment may include gases that have harmful effects on the
sensors of this unit. (Different gases can be defected depending on the type.)
The gas monitor cannot be used in the presence of the following gases:
(1) Sulfides (such as H2S and SO2) continuously existing in high concentrations
(2) Halogen gases (such as chloride compounds and chlorofluorocarbons)
(3) Silicone (Si compounds)
Do not use the gas monitor in the presence of the above gases (such as high-concentration
sulfides, halogen gases, and silicone), which may shorten the sensor life significantly or cause
malfunctions such as inaccurate readings.
In case the gas monitor is used for detection in the presence of silicone, etc., be sure to check
the gas sensitivities before using it again.
10
2-4. Safety information
The Portable Gas Monitor Model GX-8000 is a gas monitor designed to provide continuous exposure
monitoring of combustible gas, oxygen (O2), toxic gas such as carbon monoxide (CO) and hydrogen sulfide
(H2S) in hazardous environments. The gas sample is sucked in by build-in micro pump.
The battery can be selected either Li-ion battery or alkaline dry battery.
Li-ion battery unit is called BUL-8000,BUL-8000(G),BUL-8000(G1) and alkaline dry battery unit is called
BUD-8000(G),BUD-8000(G1).
The battery unit can be changed even by the end users.
・ Ⅱ 1 G Ex ia ⅡC T4 Ga
Electrical data
・Power supply of Li-ion battery unit : BUL-8000(G),BUL-8000(G1)
Two parallel connected Li-ion cells used in battery pack BP-8000,BP-8000(1) are from type Maxell
INR18650PB1. Um=250V.
・Power supply of alkaline battery unit : BUD-8000(G),BUD-8000(G1)
Powered by three series AA size alkaline batteries , model LR6 by TOSHIBA.
・Backup battery type CR1220 manufactured by Maxell.
Certificate numbers
・IECEx Certificate number :IECEx KEM 10.0038
・ATEX Certificate number :KEMA 10ATEX 0085
List of standards
・IEC 60079-0:2011 ・EN60079-0:2012
・IEC 60079-11:2011 ・EN60079-11:2012
WARNING
・Do not charge in hazardous location.
・Do not charge it expect by genuine charger.
・Do not replace battery unit in hazardous location.
・Do not replace dry batteries in hazardous location.
・Do not attempt to disassemble or alter the instrument.
・Use only battery unit type BUD-8000(G),BUD-8000(G1) with three series connected Alkaline
AA batteries, type LR6 manufactured by Toshiba, or use chargeable battery unit
type BUL-8000(G),BUL-8000(G1).
INST. No. 0 0 0 0 0 0 0 0 0 0 0
AB C D E
A: Manufacturing year (0-9)
B: Manufacturing month (1-9,XYZ for Oct.-Dec.)
C: Manufacturing lot
D: Serial number
E: Code of factory
Additional information
The performance of the Portable Gas Monitor GX-8000, as a safety device, is not covered by
this certificate.
11
3
Product Components
AC powered charger: 1
Shoulder strap: 1
Operating manual
Li-ion battery unit Product warranty
DANGER
About explosion-proof of main unit
Do not modify or change the circuit or structure, etc.
When measuring the oxygen concentration, do not measure anything but a mixture of air and
combustible gases or vapors and toxic gases.
When using this gas monitor in a hazardous area, take the following countermeasures for
preventing dangers resulting from electrostatic charges.
(1) Wear anti-static clothes and conductive shoes (anti-static work shoes).
(2) For indoor use, use the gas monitor while standing on a conductive work floor (with a
leakage resistance of 10 MΩ or less).
The battery unit that can be connected vary according to Explosion-proof approval No. of main
unit.
Model/Explosion-proof approval No. Model/Explosion-proof approval No.
of main unit of battery unit that can be connected
BUL-8000(G) / TC19437
GX-8000 / TC19439
BUD-8000(G) / TC19438
BUL-8000(G1) / TC21223
GX-8000 / TC21222
BUD-8000(G1) / TC21224
12
DANGER
The specifications of main units are as follows:
Pump circuit: Allowable voltage of 4.95 V, allowable current of 1.12 A, and
allowable power of 1138 mW
Combustible gas sensor circuit: Allowable voltage of 4.95 V, allowable current of 0.834 A, and
allowable power of 853 mW
Buzzer circuit: Allowable voltage of 4.95 V, allowable current of 0.431 A, and
allowable power of 441 mW
Main circuit: Allowable voltage of 4.95 V, allowable current of 0.717 A, and
allowable power of 733 mW
Backup circuit: 3.0 VDC, 10μA
Ambient temperature: -20℃~+50℃
Explosion-proof grade of main units are ExiaⅡCT4X.
NOTE
Nameplate
14
3-2. Names and functions for each part
<Outline Drawing>
15
(1) LCD display Displays gas concentrations, alarms, etc.
(2) Buzzer sound opening Emits a buzzer sound at an alarm. (Do not block it.)
(3) Alarm LED arrays The lamp blinks in response to an alarm.
(4) Infrared communication port Used to carry out data communications with a PC in data logger
mode.
(5) ▲/AIR switch Keep this switch pressed to perform fresh air adjustment.
(6) ▼/RESET switch When an alarm occurs, press this switch to reset the alarm.
(7) DISPLAY switch Press this switch to change between display modes.
(8) POWER/ENTER switch Turns on and off the power.
(9) Gas inlet Connect a sampling hose to this port.
(10) Gas outlet Exhausts the gas drawn into the gas monitor. (Do not block it.)
(11) Recharging jack cover Remove this cover to connect an AC powered charger and recharge
the batteries.
(12) Recharging indicator lamp Lights up in red during recharging and goes off when recharging is
completed.
(13) Sensor cover There are sensors inside. May be opened only when the sensor is to
be replaced.
(14) Activated carbon filter knob Turn this knob to detach and replace the activated carbon filter. No
activated carbon filter is used on a type that does not measure carbon
monoxide. Keep this knob tightened at all times.
(15) Battery unit screws Turn these screws to detach and replace the battery unit.
CAUTION
Do not jab the buzzer opening with a sharp-pointed item. The unit may malfunction or get
damaged, allowing water or foreign substance, etc. to get inside.
Do not remove the panel sheet on the display. The water-proof and dust-proof performances will
be deteriorated.
Do not affix a label on the infrared port. Infrared communications can no longer be conducted.
16
<LCD Display>
(1) Pilot indicator (5) Pump driving
indicator
(2) Combustible gas (6) Oxygen
concentration digital concentration digital
and bar display and bar display
(3) Carbon monoxide (7) Hydrogen sulfide
concentration digital concentration digital
and bar display and bar display
(1) Pilot indicator Displays the operating status in the detection mode. Normal: Blinking
(2) Combustible gas
Displays the gas concentration as a numeric value and a level in the bar
concentration
graph.
digital and bar display
(3) Carbon monoxide
Displays the gas concentration as a numeric value and a level in the bar
concentration
graph.
digital and bar display
(4) Displays the battery level. See the information below for the meanings of
Battery level display
battery level icons.
(5) Pump driving indicator Displays the suction status in the detection mode. Normal: Rotating
(6) Oxygen concentration Displays the gas concentration as a numeric value and a level in the bar
digital and bar display graph.
(7) Hydrogen sulfide
Displays the gas concentration as a numeric value and a level in the bar
concentration
graph.
digital and bar display
(8) Clock display Displays the current time.
NOTE
The meanings of battery level icons are as follows:
: Sufficient / : Low / : Needs recharging
If the battery level is lower than the above, the inside of the battery icon starts to blink ( ).
17
4
How to Use
<Recharging of Batteries>
When the gas monitor is used for the first time, or when the battery level is low, be sure to use the
accessory AC powered charger to recharge the batteries.
CAUTION
Use the dedicated AC powered charger.
Recharge the batteries in a non-hazardous area.
Recharge the batteries at ambient temperatures between 0 to 40ºC.
Do not use the gas monitor while recharging it. Correct measurements cannot be obtained.
Furthermore, the batteries get deteriorated more quickly and may have a shorter life.
The AC powered charger is not water-proof or dust-proof. Do not recharge the batteries while the
gas monitor is wet.
The AC powered charger is not explosion-proof.
18
(1) Open the recharging jack cover of the gas monitor.
CAUTION
Do not pull the recharging jack cover too hard.
It may get damaged.
(2) Put the plug of the AC powered charger into the recharging jack of
the gas monitor.
(3) Connect the AC powered charger to the wall electric outlet.
When recharging is started, the recharging indicator lamp lights up
(red).
(Recharging time: Three hours at the maximum until the batteries are fully recharged)
(4) When recharging is completed, the recharging indicator lamp goes off.
(5) When recharging is completed, disconnect the AC powered charger from the wall electric outlet.
(6) Pull out the AC powered charger plug from the power jack of the gas monitor and reattach the
recharging jack cover. Put the recharging jack cover as far as it will go.
CAUTION
Do not use the gas monitor with the recharging jack cover detached. Dust or water may get into
the gas monitor, causing it to malfunction. Replace the recharging jack cover if it is damaged.
If the recharging jack cover is not completely closed, water may get in from the power jack. The
same thing occurs if a minute foreign substance is caught beneath the cover.
Disconnect the AC powered charger from the wall electric outlet while it is not in use.
NOTE
During recharging, the battery pack may get hot, but this is not abnormal.
The temperature of the gas monitor is high immediately after recharging is completed. Let it leave for
10 minutes or more before using it. Otherwise, correct measurements may not be obtained.
When fully recharged batteries are recharged again, the recharging indicator lamp does not go on.
19
<Attaching Batteries>
(when the optional unit BUD-8000(G) ,BUD-8000(G1) are used)
When the gas monitor is used for the first time, or when the battery level is low, attach new AA alkaline
batteries.
CAUTION
<Replacement>
Turn off the power of the gas monitor before replacing the batteries.
Replace the batteries in a non-hazardous area.
Replace all of the three batteries with new ones at one time.
Pay attention to the polarities of the batteries.
If the battery cover fixing screw is not completely tightened, the dry batteries may drop off or
water may get in through the clearance. Water may also get in if a minute foreign substance is
caught beneath the battery unit.
<Batteries>
Use AA alkaline batteries.
Rechargeable batteries cannot be used.
(3) Close the battery cover and tighten the battery cover
fixing screw.
20
<Detaching Battery Unit>
(1) Loosen the two battery unit screws.
(They need not be completely detached.)
(2) Detach the battery unit.
(3) Attach a new battery unit.
NOTE
Make sure that the battery unit is installed in
correct orientation by checking the locations of the
connection terminal and projection portions.
CAUTION
Turn off the power of the gas monitor before
replacing the battery unit.
Detach and reattach the battery unit in a
non-hazardous area.
If the battery unit screw is not completely
tightened, the battery unit may drop off or
water may get in through the clearance.
Water may also get in if a minute foreign
substance is caught beneath the battery
unit.
Do not damage the rubber seal.
To maintain the water-proof and dust-proof
performances, it is recommended to replace
the rubber seal every two years, whether or
not it has an abnormality.
Bottom view
21
<Connection of Gas Sampling Probe and Gas Sampling Hose>
Attach the gas sampling probe to the end of the gas sampling hose.
Connect the sampling hose securely to the gas inlet (GAS IN) of the gas monitor.
Insert the sampling hose to the gas inlet (GAS IN) until it clicks into
place to ensure connection.
CAUTION
Use only a gas sampling hose specified by RIKEN KEIKI.
Use the gas monitor with the gas sampling probe connected so that no foreign substance is
drawn into it.
Connect a gas sampling probe and a gas sampling hose by fastening them manually without
using any tool. If they are fastened too tightly using a tool, the plastic part of the gas sampling
probe may be broken.
22
4-3. Basic operating procedures
Normally, the detection mode is used for normal operations. (The detection mode is activated after the
power is turned on.)
<<Detection Mode>>
RESET
DISPLAY
<<Fault Alarm Status>>
(Self-latching) <<Display and Setting Mode>>
23
4-4. How to start the gas monitor
<<Start-up Procedure>>(About 30 seconds)
Keep the POWER switch pressed for three seconds or more.
↓
LCD
All LCDs light up. Alarm lamp lights up.
Buzzer sounds once. (Beep)
↓
Date/Time Display
↓
Battery Voltage Display
↓
Gas Name Display
↓
Full Scale Display
↓
First Alarm Setpoint Display
↓
Second Alarm Setpoint Display
24
↓
STEL Alarm Setpoint Display
↓
TWA Alarm Setpoint Display
↓
ID Display
↓
Detection Mode Buzzer sounds twice. (Beep, beep)
CAUTION
After start-up, perform air calibration before performing gas detection (air calibration mode).
NOTE
A sensor abnormality alarm is issued before the detection mode is entered if there is any abnormality
in the sensor. Press the RESET button. This will reset the sensor abnormality alarm temporarily, set
the gas concentration display that was abnormal on the sensor to ---, and start gas detection. However,
notify the abnormality to RIKEN KEIKI promptly. Gas for which there was an abnormality in the sensor
cannot be detected. However, the alarm cannot be reset if there is an abnormality in all the sensors.
If there is an abnormality in the built-in clock, a fault alarm FAIL CLOCK may be issued. Press the
RESET button. The fault alarm is temporarily reset, and measurement is started with incorrect clock
time.
25
4-5. How to detect
In the detection mode, put the gas sampling probe close to the detection area and take the reading on the
display.
Display example
DANGER
While conducting measurement in a manhole or confined space, do not lean over or look into the
manhole or closed space. It may lead to dangers because oxygen-deficient air or other gases
may blow out.
Oxygen-deficient air or other gases may blow out from the gas exhausting outlet. Never inhale
the air or gases.
High-concentration (more than LEL) gases may blow out. Never use fire near the gas monitor.
WARNING
The gas monitor is designed to draw gases around it under the atmospheric pressure. If
excessive pressure is applied to the gas inlet and outlet (GAS IN, GAS OUT) of the gas monitor,
detected gases may be leaked from its inside, thus leading to dangers. Be sure that excessive
pressure is not applied to the gas monitor while used.
Do not connect the sampling hose directly to a location with a pressure higher than the
atmospheric pressure. The internal piping system may be damaged.
When the fresh air adjustment is performed in the atmosphere, check the atmosphere for
freshness before beginning the adjustment. If other gases exist, the adjustment cannot be
performed properly, thus leading to dangers when the gas leaks.
Issuance of a gas alarm indicates that there are extreme dangers. Take proper actions based on
your judgment.
Before use, check that there remains sufficient battery power. When the gas monitor is used for
the first time or is not used for a long period, the batteries may be exhausted. Either fully
recharge the batteries or replace them with new ones before use.
If a low battery alarm occurs, gas detection cannot be conducted. If the alarm occurs during use,
turn off the power and promptly recharge the batteries in a non-hazardous area.
Do not block the buzzer sound opening. No alarm sound can be heard.
26
CAUTION
Before performing gas detection, attach the gas sampling probe provided with the gas monitor to
prevent disturbances by air dust.
When you measure concentrations of oxygen in inert gases for a long time, the carbon dioxide
concentration in the air must be 15% or less. When you use the gas monitor in the air with a
carbon dioxide concentration of 15% or higher, perform measurement in as short time as
possible. Using the gas monitor under high concentrations for a long time may shorten the life of
the oxygen sensor.
An oxygen concentration higher than a certain level is required for the combustible gas %LEL
sensor of the gas monitor to correctly detect gases and display concentrations.
During combustible gas detection (%LEL range), long-time detection of a high-concentration
combustible gas may adversely influence the sensor.
Use the gas monitor with the LCD display facing upward. The gas monitor, when used with the
LCD display in a tilted or flat status, may not display correct values.
NOTE
If the combustible gas reading exceeds 100% LEL, the CO reading rises temporarily but this is not
abnormal.
In a low-temperature environment, the operating time is shortened due to the battery performance
property.
At a low temperature, the response of the LCD display may get slow down.
If a combustible gas with a higher concentration than %LEL is drawn, some gas may remain in the gas
sampling hose due to adsorption in the hose, gas sampling probe, etc. After drawing a
high-concentration combustible gas, clean the gas monitor to remove the adsorbed gas (draw fresh air
and check that the reading becomes zero).
Performing fresh air adjustment before cleaning it completely will result in inaccurate adjustment,
giving adverse influence on measurement. In such a case, remove the gas sampling hose before
performing fresh air adjustment to avoid inaccurate adjustment.
Perform gas detection in the vol% range (Type-A and E only) in a place where the presence of a
high-concentration combustible gas is known.
NOTE
<Influence of Coexisting Gases on High-Concentration Combustible Gas Sensor> (TYPE-A and E Only)
A thermal-conductivity sensor that detects high-concentration combustible gases, based on the principle
of detection using a difference in thermal conductivities of gases, may display a reading that is disturbed
by a considerable change in the concentration of gases other than the combustible gases that coexist in
the atmosphere.
However, the influence of oxygen on readings can be automatically corrected by the sensor, which
detects oxygen at the same time, by feeding back a change in the oxygen concentration to the detection
result of high-concentration combustible gases.
If coexisting gases other than oxygen undergo a high concentration change, the influence from the
change cannot be automatically corrected. This sensor is adjusted to be able to detect high-concentration
combustible gases in the air as well as in an nitrogen atmosphere. If the composition of the atmosphere
other than nitrogen is known in advance, adjusting the sensor in accordance with the atmosphere can
alleviate influences on readings. For information on the adjustment procedure, see the separate
"Maintenance Manual."
27
<Manual Memory>
Any instantaneous value during measurement can be recorded.
Up to 256 points of data can be recorded. When the number of recorded data points reaches the maximum,
recorded data will be overwritten, starting from the oldest data.
(1) In the detection mode, keep the ▼/RESET switch pressed and press the ▲/AIR switch to prepare for
recording (about one second). The following screens are displayed in turn on the gas monitor.
NOTE
The screen displays the memory number, date, and instantaneous value in turn. Go to the next step to
execute recording. No value is recorded at this point yet. If you do not want to record a value, press the
DISPLAY switch to return to the detection mode.
(2) Press the ENTER switch. The date and the instantaneous value at the time when the ENTER switch is
pressed are recorded.
28
<Auto Range Switching Point> (TYPE-A and E Only)
If Auto Range is set on a type with the vol% range for combustible gases, the display is automatically
switched to the vol% range when the concentration of a detected combustible gas exceeds 100%LEL.
When the concentration drops, the display returns to the %LEL range again. The following shows an
example of switching timing.
Diagram of gas concentrations and range switching timing under Auto Range setting
CH4concentration
5 vol%
(100%LEL)
3 vol%
(60%LEL)
CAUTION
An oxygen concentration higher than a certain level is required for the combustible gas %LEL
sensor of the gas monitor to correctly detect gases and display concentrations. For the sake of more
accurate gas detection and concentration display, therefore, this gas sensor may perform detection
using the vol% sensor if the built-in oxygen sensor of this unit detects an oxygen concentration
lower than a certain level in the atmosphere.
In other words, the display changes at the timing shown above when the oxygen concentration is
equal to or higher than a certain level. If it is lower than a certain level, however, the vol% sensor is
used for detection even if the combustible gas concentration is lower than the switching point.
29
4-6. Modes
Details on each mode are provided as follows. (* Operations are slightly different depending on the type.)
Mode Item LCD display Details
Detection ― Concentration display Normal state
Mode
Alarm Setpoint ALARM-P Display the full scale and alarm setpoint
Display values and check the alarm operations for
Alarm Test the settings displayed.
30
4-7. Air calibration mode
Press the AIR switch.
WARNING
When air calibration is performed in the atmosphere, check the atmosphere for freshness before
beginning the calibration. If other gases exist, the adjustment cannot be performed properly, thus
leading to dangers when the gas leaks.
CAUTION
Perform air calibration under pressure and temperature/humidity conditions close to those in the
operating environment and in fresh air.
Perform air calibration after the reading is stabilized.
If there is a sudden temperature change of 15C or more between the storage and operation
locations, turn on the power of the gas monitor, let it leave for about 10 minutes in a similar
environment to the operation location, and perform air calibration in fresh air before using it.
31
NOTE
Air calibration can be performed even when there is a gas alarm.
If the air calibration fails, it displays "FAIL" - "AIR CAL" and which sensor
has become faulty. Press the RESET button to reset the fault alarm
(calibration failure). When the alarm is reset, the value before calibration is
displayed.
32
4-8. Display/setting mode
This mode allows you to change various displays and settings.
Every time the DISPLAY switch is pressed, various screens are displayed in turn.
(* Operations are slightly different depending on the type.)
LCD
Detection Mode
PEAK
Display the maximum
concentration (or minimum
concentration for oxygen)
detected during measurement
from power-on to the present.
STEL
Display the STEL value after
power-on.
(TYPE-A, B, C, and D only)
TWA
Display the TWA value after
power-on.
(TYPE-A, B, C, and D only)
33
Pump ON/OFF Setting Pump ON/OFF Setting
Used to turn on/off the pump ENTER P35
operations.
ID Display/Selection ID Display/Selection
Display an ID if it has been ENTER P36
registered in advance. Also used
to select an ID.
To Detection Mode
NOTE
The gas monitor automatically returns to the detection mode in about 20 seconds if the gas monitor is left
unoperated.
However, the gas monitor does not automatically return to the detection mode if the pump operation is set
to OFF.
34
<Combustible Gas Measurement Range Setting "HC RANGE"> (TYPE-A and E Only)
The models that can display combustible gas levels in two ways, "%LEL range" and "vol% range,"
automatically switch between these two displays according to the gas concentration or oxygen
concentration, from "%LEL range" to "vol% range" and vice versa. They also support manual selection of
measurement range.
(1) Press the DISPLAY switch and select the combustible gas measurement range setting from the
display/setting mode menu.
The following screens are displayed in turn on the gas monitor.
NOTE
If you do not want to make a change, press the DISPLAY switch to return to the display/setting mode
menu.
(3) Every time the ▲ or ▼ switch is pressed, the measurement range menus, AUTO RANGE (automatic
switching), ONLY VOL (fixed to the vol% range), and ONLY LEL (fixed to the %LEL range), are
displayed in turn.
Press the ▲ or ▼ switch to select a measurement range and press the ENTER switch.
(5) After completion, press the DISPLAY switch several times until it returns to the detection mode.
35
CAUTION
No gas alarm is triggered in the combustible gas vol% range-only setting.
In the vol% range-only setting, the screen displays [No ALARM] because no alarm is triggered.
Be careful because the %LEL range-only setting screen does not appear different from the auto
range setting screen.
Even if the concentration exceeds the full scale, the screen does not automatically switch to the
vol% range.
No ALARM display
NOTE
During vol% range-only measurement, "vol%" and "No ALARM" displays blink.
36
<Full Scale Display/Alarm Setpoint Display/Alarm Test "ALARM-P">
Display the full scale or alarm setpoint values and check the alarm operations for the settings displayed.
(1) Press the DISPLAY switch and select the "full scale display/alarm setpoint display/alarm test" from the
display/setting mode menu. The following screens are displayed in turn on the gas monitor.
(2) Press the ENTER switch to enter the alarm setpoint or other display.
NOTE
If you do not want to enter any display, press the DISPLAY switch to return to the display/setting mode
menu.
(3) Every time the ▲ or ▼ switch is pressed, the full scale and alarm setpoint menus, i.e., full scale display,
first alarm setpoint display, second alarm setpoint display, STEL alarm setpoint display, and TWA
alarm setpoint display, are displayed in turn.
Press either the ▲ or ▼ switch to select a setting that you want to check.
Full Scale Display First Alarm Setpoint Display Second Alarm Setpoint Display
(WARNING) (ALARM)
STEL Alarm Setpoint Display (*) TWA Alarm Setpoint Display(*) * TYPE-A, B, C, and D only
(4) Press the ENTER switch to perform alarm test. The alarm operation on this screen can be checked.
Press any switch to stop the alarm operation.
(5) Press the DISPLAY switch to exit the alarm setpoint display or alarm test. The display/setting mode
menu is displayed again.
(6) After completion, press the DISPLAY switch several times until it returns to the detection mode.
37
<Pump ON/OFF Setting "PUMP OFF">
Used to turn on/off the pump operations.
CAUTION
While the pump operation is set to OFF, no gas detection is performed because no gas is drawn.
The gas monitor does not automatically return to the detection mode if the pump operation is set
to OFF.
(1) Press the DISPLAY switch and select the pump ON/OFF setting from the display/setting mode menu.
The following screens are displayed in turn on the gas monitor.
(2) Press the ENTER switch to set the pump operation to ON or OFF.
NOTE
If you do not want to set the pump operation or ON or OFF, press the DISPLAY switch to return to the
display/setting mode menu.
(3) Every time the ENTER switch is pressed, the pump operation is turned ON or OFF.
NOTE
While the pump operation is set to OFF, only the ENTER switch is enabled.
(4) To return to the detection mode, press the ENTER switch to set the pump operation to ON.
(5) After completion, press the DISPLAY switch several times until it returns to the detection mode.
38
<ID Display/Selection "ID SELECT">
Display an ID if it has been registered in advance. Also used to select an ID.
(1) Press the DISPLAY switch and select the ID display/selection from the display/setting mode menu.
The following screens are displayed in turn on the gas monitor.
NOTE
If you do not want to set or select an ID, press the DISPLAY switch to return to the display/setting
mode menu.
On this gas monitor, either of the IDs from ST-ID000 to ST-ID255 has been registered, unless
otherwise specified.
The data logger management program (option) is required to register or change an ID. Please contact
RIKEN KEIKI.
(5) After completion, press the DISPLAY switch several times until it returns to the detection mode.
39
<Log Data Display "REC.DATA">
Display concentration data recorded to the manual memory.
(1) Press the DISPLAY switch and select the log data display from the display/setting mode menu.
The following screens are displayed in turn on the gas monitor.
NOTE
If you do not want to display the log data, press the DISPLAY switch to return to the display/setting mode
menu.
(3) Every time the ▲ or ▼ switch is pressed, the log data menus are displayed in turn.
Press either the ▲ or ▼ switch to select log data that you want to check. The log data menu displays
the year, month, day, time, and memory number.
(4) Press the ENTER switch to display the selected log data.
(5) If you want to display other log data, press the ENTER switch to return to the log data menu. Repeat
the steps (3) to (5).
(6) After completion, press the DISPLAY switch to return to the detection mode.
NOTE
If the display is not zero (or 20.9% for oxygen) when the power is turned off, a
purge operation may be performed for 30 seconds at the maximum to clean the
inside of the gas monitor.
40
5
41
<Display Operation>
Gas Concentration Display
In a gas alarm, the gas concentration display and the alarm type
display blink.
In case of over the detection range (Over Scale), "∩∩∩" is
displayed on the LCD.
Display example
Alarm Lamp
The alarm consists of two steps. Each of them is triggered when the respective alarm setpoint value is
reached to or exceeded.
Buzzer
The alarm consists of two steps. Each of them sounds when the respective alarm setpoint value is
reached to or exceeded.
Concentration
Time Time
WARNING
Issuance of a gas alarm indicates that there are extreme dangers. Take proper actions based on
your judgment.
42
5-2. Fault alarm activation
Fault alarm: Triggered when an abnormality is detected in the gas monitor. <<Self-latching>>
Alarm display: Notified by display of error messages, sounding of the buzzer, and lighting of the lamp.
Alarm types: Low flow rate, sensor abnormality, battery voltage low, system abnormality, and calibration
failure
<Display Operation>
CAUTION
To reset a low flow rate alarm (FAIL LOW FLOW), remove the cause of the low flow rate, and
then press the RESET switch.
When the low flow alarm has occurred, the instrument will stop the operation of pump for
protection of inside.
In that case, once, turn off the power to the instrument.
After surely treating the cause of the low flow alarm, press the RESET switch.
After the instrument sucked water, do not press the RESET switch without the treatment for
water.
Even if the low flow alarm is released without the treatment for water etc., the pump will
temporarily start working again.
In this way, the sucked water enters inside the instrument, it may cause a malfunction.
NOTE
For information on malfunctions (error messages), see "8. Troubleshooting".
NOTE
The data logger management program (option) is required to use the history and trend functions. Please
contact RIKEN KEIKI.
43
6
Maintenance
44
<About Maintenance Services>
We provide services on regular maintenance including span adjustment, other adjustments and
maintenance.
To make the calibration gas, dedicated tools, such as a gas cylinder of the specified concentration and
gas sampling bag must be used.
Our qualified service engineers have expertise and knowledge on the dedicated tools used for services,
along with other products. To maintain the safety operation of the gas monitor, please use our
maintenance service.
The followings are typical maintenance services. For more information, please contact RIKEN KEIKI.
Main Services
Battery Level : Checks the battery level.
Check
Concentration : Verifies that the concentration display value is zero (or 20.9 vol% on the oxygen deficiency
Display Check meter) by using the zero gas.
Performs the zero adjustment (fresh air adjustment) if the reading is incorrect.
Flow Rate Check : Checks the flow rate indicator to find abnormalities.
Checks the flow rate by using an external flow meter to verify the correctness of the flow rate
indicator on the gas monitor. If the flow rate is incorrect, performs the flow rate adjustment.
Filter Check : Checks the dust filter for dust or clogging.
Replaces a dirty or clogged dust filter.
Alarm Test : Checks the alarm lamp and buzzer for normal operation using the alarm test function.
Span Adjustment : Performs the span adjustment by using the calibration gas.
Gas Alarm Check : Checks the gas alarm by using the calibration gas.
Checks the alarm. (Checks the alarm activation when the alarm setpoint is reached.)
Checks the delay time. (Checks time to delay until the alarm is triggered.)
Checks the buzzer, lamp, and concentration display. (Check each activation of ALM1 and
ALM2.)
Cleaning and : Checks dust or damage on surface of the gas monitor, clean and repair such parts of the gas
Repair of Gas monitor.
Monitor Replaces parts which are cracked or damaged.
(visual diagnosis)
Gas Monitor : Uses the keys to check the operation of functions and parameters.
Operation Check
Replacement of : Replaces consumable parts, such as a sensor, filter and pump.
Consumable Parts
CAUTION
Do not use a lighter gas to check the sensitivity of the gas monitor. A constituent of the lighter gas
may deteriorate the sensor performances.
45
6-3. How to clean
Clean the gas monitor if it becomes extremely dirty. The gas monitor must be turned off while cleaning it.
Use a waste cloth to remove dust. Do not use water or organic solvent for cleaning because they may cause
malfunctions.
Because an extremely large amount of dust inside the gas sampling hose may disturb the gas detection, it
must be cleaned with dry AIR, etc.
CAUTION
When cleaning the gas monitor, do not splash water over it or use organic solvents such as alcohol
and benzene on it. The surface of the gas monitor may be discolored or damaged.
NOTE
When the gas monitor gets wet, water may remain in the buzzer sound opening or clearances. Drain
water as follows:
(1) Wipe away moisture on the gas monitor thoroughly using a dry towel, cloth, etc.
(2) While holding the gas monitor firmly, shake it about ten times with the buzzer sound opening facing
downward.
(3) Wipe away moisture coming out from the inside thoroughly using a towel, cloth, etc.
(4) Place the gas monitor on a dry towel, cloth, etc. and let it stand at normal temperatures.
NOTE
Use only the filters specified by RIKEN KEIKI.
46
How to replace the activated carbon filter
CAUTION
Turn off the power of the gas monitor before replacing the activated carbon filter.
Do not remove the activated carbon filter knob unless the activated carbon filter is to be
replaced. If the activated carbon filter knob is loose (not sufficiently tightened), accurate
measurement may not be possible due to leaks, or water may get inside.
No activated carbon filter is used on a type that does not measure carbon monoxide. Keep this
knob tightened at all times.
Use only an activated carbon filter dedicated to the gas monitor (GX-8000). Using a similar
product may have harmful effects on the gas detection performance.
(1) Using a coin, etc., turn the activated carbon filter knob
counterclockwise (in a direction indicated as OPEN in this figure) to
remove it.
(2) Pull out to remove the filter case from the activated carbon knob.
(3) Replace the two activated carbon filters in the filter case with new ones.
(4) Attach the filter case to the activated carbon filter knob and put it as far as
it will go.
(5) Attach the activated carbon filter knob to the main unit in an opposite
procedure to Step (1). Tighten it firmly using a coin, etc.
CAUTION
If the knob is not completely tightened, accurate gas measurement may not be possible due to
leaks, or water may get inside. The same thing occurs if a minute foreign substance is caught
beneath the knob.
47
<Replacement of Regular Replacement Parts>
List of recommended regular replacement parts
NOTE
The above replacement intervals are recommendation only. The intervals may change depending on the
operating conditions. These intervals do not mean the warranty periods either. The result of the regular
maintenance may determine when to replace the parts.
The operation of most of the periodical replacement parts must be checked after replacement by a qualified
service engineer.
For the stable operation of the gas monitor and safety, ask a qualified service engineer to take care of
replacement of the parts whose operation must be checked. Please contact RIKEN KEIKI.
48
7
Store the gas monitor in a shipping carton, if any, in which the product was delivered.
Store the gas monitor away from dust, etc. if the shipping carton is not available.
CAUTION
If the gas monitor is not used for a long time, turn on the power at least once every six months and
check that the pump draws in air (about three minutes). The gas monitor, when not activated for a
long time, may cease to work because of hardening of the grease in the pump motor.
NOTE
If the gas monitor with a Li-ion battery unit is not used for a long time, it is recommended to store it
after discharging the batteries until the battery level icon shows one battery mark or so. If the gas
monitor is stored with the batteries fully recharged, the batteries get deteriorated more quickly and may
have shorter life.
If the gas monitor with a dry battery unit is not used for a long time, store it after removing the batteries.
Battery leaks may result in fire or injury. If the gas monitor is not used for a short time, store it without
removing the batteries. While the power of the gas monitor is OFF, the sensor is energized at all times.
Therefore, it is necessary to store the gas monitor with the batteries in it.
CAUTION
When you use a stopped/stored gas monitor again, do not forget to perform a gas calibration. For
information on readjustment including gas calibration, please contact RIKEN KEIKI.
49
7-3. Disposal of products
When the gas detector is disposed of, it must be treated properly as an industrial waste in accordance
with the local regulations.
WARNING
Do not disassemble the electrochemical type sensor or galvanic cell type sensor because they
contain electrolyte. Electrolyte may cause severe skin burns if it contacts skin, while it may cause
blindness if it contacts eyes.
If electrolyte is adhered on you clothes, that part on your clothes is discolored or its material is
decomposed. If contact occurs, rinse the area immediately with a large quantity of water.
Dispose of the batteries in accordance with procedure specified by the local authority.
When disposing of the gas detector in EU member states, sort the batteries as specified. Handle the
removed batteries according to the classified refuse collection system and recycling system based on
the regulations of EU member states.
Contact RIKEN KEIKI for disposal.
Removing batteries
See Section 4-2 "Preparation for start-up" and take out the batteries.
Model Type
BUL-8000(G) Lithium ion battery
BUL-8000(G1)
NOTE
BUL-8000(G), BUL-8000(G1) contains batteries.
Crossed-out recycle dustbin mark
This symbol mark is indicated on the products which contain the batteries which fall under EU Battery
Directive 2006/66/EC. Such batteries need to be disposed of as specified by the latest Directive.
This symbol mark indicates that the batteries need to be separated from the ordinary waste and disposed of
appropriately.
50
8
Troubleshooting
The troubleshooting does not explain the causes of all the malfunctions which occur on the gas monitor.
This simply helps to find the causes of malfunctions which frequently occur. If the gas monitor shows a
symptom which is not explained in this manual, or still has malfunctions even though remedial actions are
taken, please contact RIKEN KEIKI.
<Abnormalities on Unit>
Symptoms Causes Actions
The power cannot The battery level is too Li-ion battery unit: Recharge the batteries in a
be turned on. low. non-hazardous area.
Dry battery unit: Replace all the three dry batteries with
new ones in a non-hazardous area.
The power switch was For power-on, keep the POWER switch pressed until a
not pressed long enough. beep is heard (about two seconds).
Improper installation of Check whether battery unit is properly attached to the
battery unit main unit.
Abnormal Disturbances by sudden Turn off and restart the gas monitor.
operations surge noise, etc.
Key operations are Disturbances by sudden In a non-hazardous area, remove the battery unit once,
disabled. surge noise, etc. and reinstall the battery unit, and turn on the power to
perform operations.
System A circuit abnormality Request RIKEN KEIKI for repair.
abnormalities occurred.
FAIL SYSTEM
System
abnormalities
FAIL SYSTEM Abnormalities of internal
Error No.000 ROM
Abnormalities of internal
Error No.010 Request RIKEN KEIKI for repair.
RAM
Abnormalities of internal
Error No.021
FRAM
Abnormalities of internal
Error No.031
FLASH
Sensor A sensor has failed. Request RIKEN KEIKI to replace the sensor.
abnormalities (Only at power-on, press the RESET switch to continue
FAIL SENSOR the operation using only the normal sensors to detect
other gases.)
A low battery The battery level is low. Li-ion battery unit: Turn off the power and recharge it in a
voltage alarm is non-hazardous area.
displayed.
Dry battery unit: Turn off the power and replace the dry
FAIL BATTERY
batteries with new ones in a non-hazardous area.
51
Symptoms Causes Actions
A low flow rate Water or oil, etc. was Check the gas sampling hose for any damage or mark of
alarm is displayed. drawn in. drawn water or oil, etc.
FAIL LOW FLOW
The gas sampling hose Check the gas sampling hose for connections, clogging,
is clogged. twisting, etc.
The pump has Request RIKEN KEIKI to replace the pump.
deteriorated.
Fresh air Fresh air is not supplied Supply fresh air.
adjustment cannot around the gas monitor.
be performed.
FAIL AIR CAL
Clock Abnormalities of the Make a setting of Date/Time.
abnormalities internal clock If such a symptom is observed repeatedly, the built-in
FAIL CLOCK clock is seemingly malfunctioning. Thus, it must be
replaced. Please contact RIKEN KEIKI.
The batteries The charger is not Connect the AC powered charger to the wall electric
cannot be connected properly. outlet and jack properly.
recharged.
A recharging circuit Request RIKEN KEIKI for repair.
(Li-ion battery unit
abnormality occurred.
only)
The batteries have been When fully recharged batteries are recharged again, the
fully charged. recharging indicator lamp does not go on.
<Abnormalities of Readings>
Symptoms Causes Actions
The reading rises Drifting of sensor output Perform the zero adjustment (fresh air adjustment).
(drops) and it Presence of interference Disturbances by interference gases, such as solvents,
remains so. gas cannot be eliminated completely. For information on
actions, such as removal filter, please contact RIKEN
KEIKI.
Slow leak A very small amount of the gas to be detected may be
leaking (slow leak). Because ignoring it may cause
dangers, take a remedial measure, i.e., taking actions
the same as those for the gas alarm.
Environmental changes Perform the zero adjustment (fresh air adjustment).
In particular, the galvanic cell type is affected by the air
pressure.
A gas alarm is Presence of interference Disturbances by interference gases, such as solvents,
triggered despite gas cannot be eliminated completely. For information on
of no gas leak and actions, such as removal filter, please contact RIKEN
no other KEIKI.
abnormalities at Disturbance by noise Turn off and restart the gas monitor.
the detection If such a symptom is observed frequently, take
point. appropriate measures to eliminate the noise.
Slow response Clogged dust filter Replace the dust filter.
Bended or clogged Fix the defective parts.
suction tube or exhaust
tube
Condensation is formed Fix the defective parts.
inside the suction tube.
Deteriorated sensor Replace the sensor unit with a new one.
sensitivity
Span adjustment Improper calibration gas Use the proper calibration gas.
impossible concentration
Deteriorated sensor Replace the sensor unit with a new one.
sensitivity
52
9
Product Specifications
9-1. List of specifications
<TIIS specifications>
Model GX-8000
Detection principle Galvanic cell (OS) New ceramic catalytic(NC)/ Electrochemical (ES) Electrochemical (ES)
Thermal conductivity(TE)
Detectable gas Oxygen Combustible gas Hydrogen sulfide Carbon monoxide
(O2) (HC/CH4/H2/C2H2) (H2S) (CO)
Measuring range 0~25vol% 0~100%LEL/~100vol% 0~30ppm 0~150ppm
<Service range> <~40vol%> <~100ppm> <~500ppm>
Resolution 0.1vol% 1%LEL/1vol% 0.5ppm 1ppm
Alarm preset point 18vol%(L) 10%LEL(1st) 5ppm(1st) 25ppm(1st)
25vol%(H) 50%LEL(2nd) 30ppm(2nd) 50ppm(2nd)
40vol%(OVER) 100%LEL(OVER) 10ppm(TWA) 25ppm(TWA)
15ppm(STEL) 200ppm(STEL)
100ppm(OVER) 500ppm(OVER)
Indicate accuracy Less than±0.7vol% Less than±5%LEL(NC)/ Less than±1.5ppm Less than±15ppm
(under an identical condition) Less than±5vol%(TE)
Response time Less than 20sec(T90) Less than 30sec(T90) Less than 30sec(T90) Less than 30sec(T90)
(under an identical condition)
Alarm accuracy Less than±1vol% Less than±25% Less than±3ppm Less than±30%
(under an identical condition)
Alarm-delay time Less than 5sec Less than 30sec Less than 15sec Less than 30sec
(under an identical condition)
Gas concentration display LCD(digital)
Detection method Pump suction method
Suction flow Above 0.75L/min(open flow)
Various indications Clock/Battery condition/Pilot indicator/Flow indicator
Alarm sound Above 95dB(A)(30cm)
Gas alarm indication Lamp blinking/Intermittent buzzer/Gas concentration display blinking
Gas alarm action Latched
Trouble alarm・self diagnosis System failure/Sensor failure/Low battery/Calibration failure/Flow failure
Trouble alarm indication Lamp lighting/Intermittent buzzer/Content display
Trouble alarm action Auto-recover
Specification of transmission IrDA
Various functions LCD backlight/Data logger/Peak hold/Logged data display/Pump stop
Power supply Lithium ion battery unit 【BUL-8000(G), BUL-8000(G1)】
(Dedicated dry battery unit<three AA batteries>【BUD-8000(G) ,BUD-8000(G1)*1】 is available.)
Continuous operation time BUL-8000(G), BUL-8000(G1):Approx. 12 hours(25℃・Non alarm operation・After full charge)
BUD-8000(G), BUD-8000(G1):Approx. 6 hours(25℃・Non alarm operation)
Operating temperature -20~50℃
Operating humidity Less than 95%RH(non-condensing)
Operating pressure Atmospheric pressure(80kPa~110kPa)
Structure Dust and Water-proof Structure(IP67)
Explosion-proof structure Intrinsically safe structure
Explosion-proof grade Ex ia ⅡC T4 X(TIIS< Japan>)
Outer dimension Approx. 154(W)×81(H)×127(D)mm(projection excluding)
Weight Approx. 1.1kg(BUL-8000(G), BUL-8000(G1) when in use)
Approx. 1.0kg(BUD-8000(G), BUD-8000(G1) when in use)
*1 In order to meet requirement about explosion-proof, use dedicated dry cell battery when using dry battery unit.
53
<ATEX/IECEx specifications>
Model GX-8000
Detection principle Galvanic cell (OS) New ceramic catalytic(NC)/ Electrochemical (ES) Electrochemical (ES)
Thermal conductivity(TE)
Detectable gas Oxygen Combustible gas Hydrogen sulfide Carbon monoxide
(O2) (HC/CH4/H2/C2H2) (H2S) (CO)
Measuring range 0~25vol% 0~100%LEL/~100vol% 0~30ppm 0~150ppm
<Service range> <~40vol%> <~100ppm> <~500ppm>
Resolution 0.1vol% 1%LEL/1vol% 0.5ppm 1ppm
Alarm preset point 19.5vol%(L) 10%LEL(1st) 5ppm(1st) 25ppm(1st)
23.5vol%(H) 50%LEL(2nd) 30ppm(2nd) 50ppm(2nd)
40vol%(OVER) 100%LEL(OVER) 10ppm(TWA) 25ppm(TWA)
15ppm(STEL) 200ppm(STEL)
100ppm(OVER) 500ppm(OVER)
Indicate accuracy Less than±0.7vol% Less than±5%LEL(NC)/ Less than±1.5ppm Less than±15ppm
(under an identical condition) Less than±5vol%(TE)
Response time Less than 20sec(T90) Less than 30sec(T90) Less than 30sec(T90) Less than 30sec(T90)
(under an identical condition)
Gas concentration display LCD(digital)
Detection method Pump suction method
Suction flow Above 0.75L/min(open flow)
Various indications Clock/Battery condition/Pilot indicator/Flow indicator
Alarm sound Above 95dB(A)(30cm)
Gas alarm indication Lamp blinking/Intermittent buzzer/Gas concentration display blinking
Gas alarm action Latched
Trouble alarm・self diagnosis System failure/Sensor failure/Low battery/Calibration failure/Flow failure
Trouble alarm indication Lamp lighting/Intermittent buzzer/Content display
Trouble alarm action Auto-recover
Specification of transmission IrDA
Various functions LCD backlight/Data logger/Peak hold/Logged data display/Pump stop
Power supply Lithium ion battery unit 【BUL-8000(G), BUL-8000(G1)】
(Dedicated dry battery unit<three AA batteries>【BUD-8000(G), BUD-8000(G1)】 is available.)
Continuous operation time BUL-8000(G) BUL-8000(G1):Approx. 12 hours(25℃・Non alarm operation・After full charge)
BUD-8000(G) BUD-8000(G1):Approx. 6 hours(25℃・Non alarm operation)
Operating temperature -20~50℃
Operating humidity Less than 95%RH(non-condensing)
Operating pressure Atmospheric pressure(80kPa~110kPa)
Structure Dust and Water-proof Structure(IP67)
Explosion-proof structure Intrinsically safe structure
Explosion-proof grade Ⅱ1G Ex ia ⅡC T4 Ga(ATEX) / Ex ia ⅡC T4 Ga(IECEx)
Outer dimension Approx. 154(W)×81(H)×127(D)mm(projection excluding)
Weight Approx. 1.1kg(BUL-8000(G), BUL-8000(G1) when in use)
Approx. 1.0kg(BUD-8000(G), BUD-8000(G1) when in use)
54
9-2. List of accessories
Li-ion battery unit (BUL-8000 (G), BUL-8000 (G1))
AC powered charger
Standard
Shoulder strap
accessories
Gas sampling hose (1 m spiral)
Gas sampling probe
Dry battery unit (BUD-8000 (G), BUD-8000 (G1))
AA alkaline battery x 3
Waist strap
Waist strap fixing tool
Carrying case (leather)
Aluminum trunk case
Shipboard storage box (metal)
Sampling probe holder
Optional Gas sampling hose (with float probe, 8m)
accessories Gas sampling hose 30 m with plummet
Filter tube (CF-8385)
Filter tube fixing belt
Water trap
Relay tube
Set of gas bags (LEL/VOL)
Span cans (CALGAS 4 component mixed gas)
Demand flow valve
Data logger management program
CAUTION
In case of using gas sampling hose 30m with plummet (option), use CF-8385 type filter always.
(※CF-8385 type filter is to waterproof and dust removal applications. Relay tube and the filter fixing
belt will be required together too.)
In case of sucking water without CF-8385 type filter, the water enters to inside instrument, it may
cause a malfunction.
55
10