Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Scope Location 1 CC-2101

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD.

, HOSPITAL ROAD, VILLAGE-


WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 1 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
Permanent Facility

1 MECHANICAL- Digital/ Analog 100 rpm to 1000 rpm 1.5 % Using digital
ACCELERATION AND Tachometer (Non tachometer and
SPEED Contact Type) rotating disc of electric
motor as RPM Source

2 MECHANICAL- Digital/ Analog 20 rpm to 100 rpm 2.75 % Using Digital


ACCELERATION AND Tachometer (Non Tachometer and
SPEED Contact Type) Rotating Disc of
Electric Motor as RPM
Source

3 MECHANICAL- Digital/Analog 1000 rpm to 26000 rpm 0.2% Using digital


ACCELERATION AND Tachometer (Non tachometer and
SPEED Contact Type) rotating disc of electric
motor as RPM Source

4 MECHANICAL- Force/Force/Load Cell, 1 kN to 50 kN 0.05% IS 4169, ISO 376,


FORCE PROVING Proving Ring, ASTM E74/Dead
INSTRUMENTS Dynamometer Weight Force Machine

5 MECHANICAL- Force/Force/Load Cell, 200 N to 5 kN 0.05% IS 4169, ISO 376,


FORCE PROVING Proving Ring, ASTM E74/Dead
INSTRUMENTS Dynamometer Weight Force Machine

6 MECHANICAL- Push Pull Gauge 5 N to 500 N 1.0N Using Newtonian


MOBILE FORCE Weights & Fixture
MEASURING SYSTEM Frame based on
VDI/VDE 2624 (Part
2.1)

7 MECHANICAL- Hydraulic 0 to 100 bar 0.12 bar comparison method/


PRESSURE Pressure:Digital/ Standard digital
INDICATING DEVICES Analogue pressure pressure gauge / DKD
gauge, pressure R6-1
transducer/transmitter

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 2 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
8 MECHANICAL- Hydraulic 0 bar to 400 bar 0.48 bar comparison method/
PRESSURE Pressure:Digital/ Standard digital
INDICATING DEVICES Analogue pressure pressure gauge / DKD
gauge, pressure R6-1
transducer/transmitter

9 MECHANICAL- Hydraulic 0 bar to 700 bar 0.84 bar comparison method/


PRESSURE Pressure:Digital/ Standard digital
INDICATING DEVICES Analogue pressure pressure gauge / DKD
gauge, pressure R6-1
transducer/transmitter

10 MECHANICAL- Pneumatic pressure: 0 bar to 30 bar 0.036 bar comparison method/


PRESSURE Digital/Analogue Standard digital
INDICATING DEVICES pressure gauge, pressure gauge / DKD
pressure R6-1
transducer/transmitter

11 MECHANICAL- Pneumatic 0 to -0.9 bar, 0 to 6 bar 0.0022 bar, 0.0072 bar comparison method/
PRESSURE pressure:Digital / Standard digital
INDICATING DEVICES Analogue Vacuum pressure gauge / DKD
Gauge, Vacuum R6-1
Transducer/Transmitter

12 MECHANICAL- Glass Pipettes/Burettes 1 ml to 25 ml 1.05µl Using Semi Micro


VOLUME Balance and distilled
Water by gravimetric
method

13 MECHANICAL- Glass Pipettes/Burettes 25 ml to 100 ml 1.21µl Using Semi Micro


VOLUME Balance and Distilled
Water by Gravimetric
Method

14 MECHANICAL- Measuring 1 ml to 100 ml 1.5µl Using Semi Micro


VOLUME Cylinder/Volumetric balance and distilled
Flask/Conical water by gravimetric
Flask/Beaker etc method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 3 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
15 MECHANICAL- Measuring 1000 ml to 5000 ml 5ml Using 6200 g Balance
VOLUME Cylinder/Volumetric of Resolution 10 mg
Flask/Conical and distilled water by
Flask/Beaker etc gravimetric method

16 MECHANICAL- Measuring 100 ml to 1000 ml 1.95ml Using 1kg Balance of


VOLUME Cylinder/Volumetric Resolution 1 mg and
Flask/Conical distilled water by
Flask/Beaker, Kane etc gravimetric method

17 MECHANICAL- Piston Pipette 10 µl to 1000 µl 0.28µl Using digital balance of


VOLUME resolution 0.01 mg upto
200 g as per ISO 8655-
6

18 MECHANICAL- Piston Pipette 1000 µl to 10000 µl 13µl Using digital balance of


VOLUME 200 g with resolution of
0.01 mg as per ISO
8655-6

19 MECHANICAL- d= 10 mg and coarser 1 kg to 5 kg 10mg Using F1 Class


WEIGHING SCALE Weights for calibration
AND BALANCE of weighing scales as
per OIML R76-1

20 MECHANICAL- d= 5 g and coarser 5 kg to 50 kg 2.8g Using F1 Class


WEIGHING SCALE Weights for calibration
AND BALANCE of weighing scales as
per OIML R76-1

21 MECHANICAL- d=1 mg and coarser 200 g to 1 kg 1.7mg Using E2 and F1 Class


WEIGHING SCALE weights for calibration
AND BALANCE of weighing balances
as per OIML-R76-1

22 MECHANICAL- Micro Balance d=0.001 0 g to 50 g 0.02mg Using standard weights


WEIGHING SCALE mg and coarser of E1 Class as per
AND BALANCE OIML R 76-1

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 4 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
23 MECHANICAL- Micro Balance d=0.01 1 mg to 200 g 0.07mg Using E1 Class weights
WEIGHING SCALE mg and coarser for calibration of
AND BALANCE electronic weighing
balance as per OIML
R76-1

24 MECHANICAL- Mass / Weights 2 kg 20mg Using weights of


WEIGHTS accuracy class F1 and
digital balance of
resolution 10 mg by
substitution method
and ABBA weighing
cycle based on OIML
R111

25 MECHANICAL- Mass / Weights 200 g 0.15mg Using weights of


WEIGHTS accuracy class E2 and
digital balance of
resolution 0.01 mg by
substitution method
and ABBA weighing
cycle based on OIML
R111

26 MECHANICAL- Mass / Weights(Class 1 kg 2.97mg Using weights of


WEIGHTS F2 Accuracy and accuracy class F1 and
Coarser) digital balance of
resolution 1 mg by
substitution method
and ABBA weighing
cycle based on OIML
R111

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 5 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
27 MECHANICAL- Mass / Weights(Class 1 mg to 5 g 0.02mg Using weights of
WEIGHTS F2 Accuracy and accuracy class E2 and
Coarser) digital balance of
resolution 0.01 mg by
substitution method &
ABBA weighing cycle
based on OIML R111

28 MECHANICAL- Mass / Weights(Class 10 g to 20 g 0.03mg Using weights of


WEIGHTS F2 Accuracy and accuracy class E2 and
Coarser) digital balance of
resolution 0.01 mg by
substitution method
and ABBA weighing
cycle based on OIML
R111

29 MECHANICAL- Mass / Weights(Class 100 g 0.08mg Using weights of


WEIGHTS F2 Accuracy and accuracy class E2 and
Coarser) digital balance of
resolution 0.01 mg by
substitution method
and ABBA weighing
cycle based on OIML
R111

30 MECHANICAL- Mass / Weights(Class 5 kg 35mg Using weights of


WEIGHTS F2 Accuracy and accuracy class F1 and
Coarser) digital balance of
resolution 10 mg by
substitution method
and ABBA weighing
cycle based on OIML
R111

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 6 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
31 MECHANICAL- Mass / Weights(Class 50 g 0.05mg Using weights of
WEIGHTS F2 Accuracy and accuracy class E2 and
Coarser) digital balance of
resolution 0.01 mg by
substitution method
and ABBA weighing
cycle based on OIML
R111

32 MECHANICAL- Mass / Weights(Class 500 g 1.6mg Using weights of


WEIGHTS M1 Accuracy and accuracy class F1 and
Coarser) digital balance of
resolution 1 mg by
substitution method
and ABBA weighing
cycle based on OIML
R111

33 THERMAL- SPECIFIC RH Indicators with or In 4 %RH to 95 %RH@ 0.43%RH Using Rotronic RH


HEAT & HUMIDITY built/External Sensor, 25°C Sensor with Indicator
Thermo Hygrometers, by Kaymont Humidity
Data loggers with RH Generator
Sensors

34 THERMAL- SPECIFIC Temperature of RH 10 °C to 50 °C@ 50% 0.30°C Using Rotronic RH and


HEAT & HUMIDITY Indicators with or In RH Temperature Sensor
built/External Sensor , with Indicator by
Thermo hygrometer, Kaymont Humidity
data logger with RH Generator
sensor

35 THERMAL- RTD, Thermocuple with 100 °C to 600 °C 0.60°C Using RTD with Sensor
TEMPERATURE our without Indicator, with DMM by
data logger, comparison method
temperature gauge, using dry block
digital thermometer, calibrator
transmitter

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 7 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
36 THERMAL- RTD, Thermocuple with -35 °C to 100 °C 0.19°C Using 4 Wire RTD with
TEMPERATURE our without Indicator, Sensor with DMM by
data logger, comparison method
temperature gauge, using dry block
digital thermometer, calibrator
transmitter

37 THERMAL- Temperature indicator -80 °C to 300 °C 0.6°C Using RTD with


TEMPERATURE of liquid bath, freezer, Eurotherm Indicator by
refrigerator, incubator, at single specified
oven , furnace, position method
environmental
chamber, dry block
comparator etc

38 THERMAL- Temperature indicator 300 °C to 1200 °C 2.95°C Using R Type


TEMPERATURE of Oven , furnace, Dry Thermocouple with
block Comparator etc. Druck DPI Indicator

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 8 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
Site Facility

1 MECHANICAL- RPM Indicator/Indicator 100 rpm to 1000 rpm 1.5% Using digital
ACCELERATION AND of Centrifuge tachometer
SPEED

2 MECHANICAL- RPM Indicator/Indicator 1000 rpm to 26000 rpm 0.2% Using digital
ACCELERATION AND of Centrifuge tachometer
SPEED

3 MECHANICAL- RPM Indicator/Indicator 20 rpm to 100 rpm 2.75 % Using Digital


ACCELERATION AND of Centrifuge Tachometer
SPEED

4 MECHANICAL- Brinell Hardness HBW 10/3000 2.5% Using reference


HARDNESS TESTING Testing Machine hardness blocks as per
MACHINES IS 1500 -2 (Indirect
Method)

5 MECHANICAL- Brinell Hardness HBW 2.5/187.5 3.0% Using reference


HARDNESS TESTING Testing Machine hardness blocks as per
MACHINES IS 1500-2 (Indirect
Method)

6 MECHANICAL- BrinellI Hardness 5/750 HBW 3.0% Using reference


HARDNESS TESTING Testing Machine hardness blocks as per
MACHINES 1500(part-2), indirect
Method

7 MECHANICAL- Rockwell Hardness HRA 0.50HRA Using reference


HARDNESS TESTING Testing Machine hardness blocks as per
MACHINES IS 1586-2 (Indirect
Method)

8 MECHANICAL- Rockwell Hardness HRBW 0.87HRBW Using refrence


HARDNESS TESTING Testing Machine hardness blocks as per
MACHINES US 1586-2(Indirect
Method)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 9 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
9 MECHANICAL- Rockwell Hardness HRC 0.60HRC Using reference
HARDNESS TESTING Testing Machine hardness blocks as per
MACHINES IS 1586-2 (Indirect
Verification)

10 MECHANICAL- Vicker Testing Machine HV10 2.86% Using reference


HARDNESS TESTING hardness blocks as per
MACHINES IS 1501-2 (Indirect
Verification)

11 MECHANICAL- Vicker Testing Machine HV30 2.36% Using reference


HARDNESS TESTING hardness blocks as per
MACHINES IS 1501-2 (Indirect
Verification)

12 MECHANICAL- Vicker Testing Machine HV5 2.06% Using reference


HARDNESS TESTING hardness blocks as per
MACHINES IS 1501(Part-2)
(Indirect Method):

13 MECHANICAL- Hydraulic 0 to 100 bar 0.12 bar comparison method/


PRESSURE Pressure:Digital/ Standard digital
INDICATING DEVICES Analogue pressure pressure gauge / DKD
gauge, pressure R6-1
transducer/transmitter

14 MECHANICAL- Hydraulic 0 bar to 400 bar 0.48 bar comparison method/


PRESSURE Pressure:Digital/ Standard digital
INDICATING DEVICES Analogue pressure pressure gauge / DKD
gauge, pressure R6-1
transducer/transmitter

15 MECHANICAL- Hydraulic 0 bar to 700 bar 0.84 bar comparison method/


PRESSURE Pressure:Digital/ Standard digital
INDICATING DEVICES Analogue pressure pressure gauge / DKD
gauge, pressure R6-1
transducer/transmitter

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 10 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
16 MECHANICAL- Pneumatic pressure: 0 bar to 30 bar 0.036 bar comparison method/
PRESSURE Digital/Analogue Standard digital
INDICATING DEVICES pressure gauge, pressure gauge / DKD
pressure R6-1
transducer/transmitter

17 MECHANICAL- Pneumatic 0 to -0.9 bar, 0 to 6 bar 0.0022 bar, 0.0072 bar comparison method/
PRESSURE pressure:Digital / Standard digital
INDICATING DEVICES Analogue Vacuum pressure gauge / DKD
Gauge, Vacuum R6-1
Transducer/Transmitter

18 MECHANICAL- UTM, Force/Force/Uniaxial 1.0 kN to 50 kN 0.05% Compression &


TENSION CREEP Static Testing Tension mode upto
AND TORSION Machines Compression 50kN as per IS 1828 &
TESTING MACHINE & Tension IS0 7500

19 MECHANICAL- UTM, Force/Force/Uniaxial 1(compression)0.2 0.89% Compression Mode


TENSION CREEP Static Testing (tension) kN to 2000 (compression)0.89(tens Using force proving
AND TORSION Machines (compression)50 ion) instruments as per IS
TESTING MACHINE Compression/Tension (tension) kN 1828 & ISO 7500

20 MECHANICAL- d= 10 mg and coarser 1 kg to 5 kg 10mg Using F1 Class


WEIGHING SCALE Weights for calibration
AND BALANCE of weighing scales as
per OIML R76-1

21 MECHANICAL- d= 5 g and coarser 5 kg to 50 kg 2.8g Using F1 Class


WEIGHING SCALE Weights for calibration
AND BALANCE of weighing scales as
per OIML R76-1

22 MECHANICAL- d=1 mg and coarser 200 g to 1 kg 1.7mg Using E2 and F1 Class


WEIGHING SCALE weights for calibration
AND BALANCE of weighing balances
as per OIML-R76-1

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 11 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
23 MECHANICAL- Micro Balance d=0.001 0 g to 50 g 0.02mg Using standard weights
WEIGHING SCALE mg and coarser of E1 Class as per
AND BALANCE OIML R 76-1

24 MECHANICAL- Micro Balance d=0.01 1 mg to 200 g 0.07mg Using E1 Class weights


WEIGHING SCALE mg and coarser for calibration of
AND BALANCE electronic weighing
balance as per OIML
R76-1

25 THERMAL- SPECIFIC Relative Humidity 4 % RH to 95 % RH 0.66%RH Using RH Indicator with


HEAT & HUMIDITY Indicator of @25°C probe by single
Environmental specified position
Chambers

26 THERMAL- SPECIFIC RH Indicators with or In 4 %RH to 95 %RH@ 0.43%RH Using Rotronic RH


HEAT & HUMIDITY built/External Sensor, 25°C Sensor with Indicator
Thermo Hygrometers, by Kaymont Humidity
Data loggers with RH Generator
Sensors

27 THERMAL- SPECIFIC Temperature of RH 10 °C to 50 °C@ 50% 0.30°C Using Rotronic RH and


HEAT & HUMIDITY Indicators with or In RH Temperature Sensor
built/External Sensor , with Indicator by
Thermo hygrometer, Kaymont Humidity
data logger with RH Generator
sensor

28 THERMAL- Freezer, Oven, -35 °C to 300 °C 4.11°C Using RTD with


TEMPERATURE Incubator, Eurotherm Datalogger
Environmental by multiposition
Chamber etc calibration using 9
sensors and data
loggers(spatial
mapping)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DVG LABORATORIES AND CONSULTANTS PVT. LTD., HOSPITAL ROAD, VILLAGE-
WAZIRABAD, GURGAON, HARYANA, INDIA

Accreditation Standard ISO/IEC 17025:2017

Certificate Number CC-2101 Page No. : 12 / 12

Validity 23/02/2019 to 22/02/2021 Last Amended on 05/07/2019

S.No Discipline / Group Quantity Measured/ Range / Frequency * Calibration Remarks


Instrument Measurement
Capability(±)
29 THERMAL- RTD, Thermocuple with 100 °C to 600 °C 0.60°C Using RTD with Sensor
TEMPERATURE our without Indicator, with DMM by
data logger, comparison method
temperature gauge, using dry block
digital thermometer, calibrator
transmitter

30 THERMAL- RTD, Thermocuple with -35 °C to 100 °C 0.19°C Using 4 Wire RTD with
TEMPERATURE our without Indicator, Sensor with DMM by
data logger, comparison method
temperature gauge, using dry block
digital thermometer, calibrator
transmitter

31 THERMAL- Temperature indicator -80 °C to 300 °C 0.6°C Using RTD with


TEMPERATURE of liquid bath, freezer, Eurotherm Indicator by
refrigerator, incubator, at single specified
oven , furnace, position method
environmental
chamber, dry block
comparator etc

32 THERMAL- Temperature indicator 300 °C to 1200 °C 2.95°C Using R Type


TEMPERATURE of Oven , furnace, Dry Thermocouple with
block Comparator etc. Druck DPI Indicator

This is annexure to 'Certificate of Accreditation' and does not require any signature.

You might also like