Seer Light Ambulatory Recorder/Controller: Service Manual
Seer Light Ambulatory Recorder/Controller: Service Manual
Seer Light Ambulatory Recorder/Controller: Service Manual
Ambulatory Recorder/Controller
Service Manual
Software Version 1
2019818-008 Revision B
Marquette®, MARS®, MUSE®, and SEER® are trademarks owned by GE Medical Systems Information
Technologies, a General Electric Company going to market as GE Healthcare. All other marks are owned by
their respective owners.
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Self-Test Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
No Buzzer Sound After Installing the Batteries . . . . . . . . . . . . . . . . . . . . . . . 2-6
Beep Sound Will Not Stop After Installing the Batteries . . . . . . . . . . . . . . . . 2-7
ECG Cannot Be Previewed with Controller (I.R.) . . . . . . . . . . . . . . . . . . . . . 2-8
ECG Cannot be Previewed with Controller (Cable) . . . . . . . . . . . . . . . . . . . . 2-9
ECG Cannot Be Previewed with SEER Light Connect Device . . . . . . . . . . 2-10
Recording Cannot Start with the Controller (I.R.) . . . . . . . . . . . . . . . . . . . . 2-11
Recording Cannot Start with the SEER Light Connect Device . . . . . . . . . . 2-12
Cannot Start by Pressing the Start Button . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Cannot Record for 24 Hours . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Cannot Record for 48 Hours . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Cannot Transfer Data from the Controller . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
No Operation After Inserting Batteries and Pressing Power Button . . . . . . . . . . . 3-6
ECG Cannot be Previewed with the Controller (I.R.) . . . . . . . . . . . . . . . . . . . . . . . 3-7
ECG Cannot be Previewed by the Controller (Cable) . . . . . . . . . . . . . . . . . . . . . . 3-8
Cannot Start Recording (I.R.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Pressing the Appropriate Key(s) Does Not Start Operation . . . . . . . . . . . . . . . . . 3-10
Beep Sounds While the Flash Card is Inserted . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Data Cannot be Transferred from the Recorder . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Recording Start Date/Time is Not Correct . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Time is Not Correct . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
ECG Cannot Be Previewed with SEER Light Connect Device . . . . . . . . . . . 4-5
Recording Cannot Start with the SEER Light Connect Device . . . . . . . . . . . 4-6
Cannot Transfer Data from the Recorder to the
SEER Light Connect Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Recording Start Date/Time is Not Correct . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1
Manual Information
Revision History
Each page of the document has the document part number followed by a
revision letter at the bottom of the page. This letter identifies the
document’s update level. The latest letter of the alphabet corresponds to
the most current revision of the document.
Manual Purpose
This manual supplies technical information for service representative
and technical personnel so they can maintain the equipment. Use it as a
guide for maintenance and electrical repairs considered field repairable.
Where necessary the manual identifies additional sources of relevant
information and or technical assistance.
See the operator manual for the instructions necessary to operate the
equipment safely in accordance with its function and intended use.
Definitions
Items shown in Black text are keys on the keyboard, text to be
entered, or hardware items such as buttons or switches on the
equipment.
For example, “Press Ctrl + Esc” means to press and hold down the
Ctrl key while pressing the Esc key.
When instructions are given for typing a precise text string with one
or more spaces, the point where the spacebar must be pressed is
indicated as: <Space>. The purpose of the < > brackets is to ensure
you press the spacebar when required.
Related Manuals
Safety Information
Intended Use
The SEER Light recorder is a two and three channel digital Holter ECG
recorder that records the electrical signals associated with cardiac
activity for 24 or 48 hours. It is used in diagnosing cardiac abnormalities
and revealing trends or changes in heart function. This device is for the
use of trained personnel only.
This device is not intended for use on infants weighing less than 10 kg
(22 lbs).
Definitions
The terms danger, warning, and caution are used throughout this
manual to point out hazards and to designate a degree or level of
seriousness. Familiarize yourself with their definitions and significance.
Warnings
WARNINGS
Cautions
CAUTIONS
Serial Number
Every GE Medical Systems Information Technologies device has a
unique serial number for identification. The serial number appears on
the device label similar to the one shown below.
50A
General
Refer equipment servicing to GE authorized service personnel only. Any
unauthorized attempt to repair equipment under warranty voids that
warranty.
This device is intended for use under the direct supervision of a licensed
health care practitioner.
Parts and accessories used must meet the requirements of the applicable
IEC 60601 series safety standards, and/or the system configuration must
meet the requirements of the IEC 60601-1-1 medical electrical systems
standard.
Equipment Symbols
Event.
This symbol indicates the polarity orientation that each battery should have when you insert it
into the unit. This unit requires you to insert the batteries so that the polarities are oriented in
alternating directions.
Power.
Stop.
Input connector.
Output connector.
This symbol indicates that the waste of electrical and electronic equipment must not be
disposed as unsorted municipal waste and must be collected separately. Please contact an
authorized representative of the manufacturer for information concerning the
decommissioning of your equipment.
Structure
The SEER Light ambulatory recorder is shown and described below.
A B
C J
F E D 017A
C battery compartment cover Slide the cover to open and set the batteries in the battery compartment.
D For transfer of data to the SEER Light controller or to the SEER Light Connect device.
A data output
connector
E access LED Will flash during communication with the SEER Light controller.
F infrared terminal (I.R. Window) Used to receive the signal from the SEER Light controller to begin ECG recording.
Used to receive patient information and ECG recording starting time.
Used to confirm the ECG waveform recorded by the recorder (ECG preview).
G Not used.
B output connector
H DATA LED The LED flashes while transferring data to the SEER Light controller.
I patient cable connector Used to connect patient cable for ECG input.
Troubleshooting
Self-Test Mode
A self-test can be done on the following functions.
1. Self-test I (pacemaker detection check mode)
After setting the batteries, press the STOP button three times
within a second.
Call GE service.
Call GE service.
Yes
Position the controller
within 4.9 ft (1.5 m).
Is the I.R.
window facing the No
controller properly?
Has recording
Yes
already started?
Is cable connection No
proper?
No Replace with a
new cable.
Call GE service.
Call GE Service.
Yes
Is the I.R.
window facing the No
controller properly?
Has recording
Yes
already started?
Is the distance
from the controller No
within 4.9 ft
(1.5 m?)
Call GE Service.
Yes
Does
the controller display Yes
“Already started?”
Is the distance
from the Connect No
within 4.9 ft
(1.5 m?)
Adjust the position
within 4.9 ft. (1.5 m)
Yes from the Connect.
Was
the START button
pressed over 1.5 No
seconds?
Is it within
one hour of setting the No
batteries?
Yes
Reset the batteries.
Call GE Service
Were alkaline No
batteries used?
Are the
battery electrodes dirty Yes
or improperly
soldered?
Call GE Service.
Is the recorder
labeled “SEER Light No
Extend?”
Were alkaline No
batteries used?
Are the
battery electrodes dirty Yes
or improperly
soldered?
Call GE Service.
Is the transfer
cable properly connected? No
Does the
controller display “No data Yes
in the recorder?”
Does the
controller display Yes
“Recorder contains data
error?”
No Call GE Service.
No
Call GE Service.
Call GE Service.
Cannot Transfer Data from the Recorder to the SEER Light Connect Device
Cannot transfer data to
the Connect.
Plug it in.
Yes
Is the transfer
cable properly No
connected?
Connect the
Yes cable properly.
Does the
controller display “No data Yes
in the recorder?”
No data in the recorder.
No Record ECG again.
Call GE Service.
No
Call GE Service.
Call GE Service.
Artifact on ECG
recording.
Are electrodes
connected properly on No
the patient?
Is the patient
cable connector No
clean?
No Call GE Service.
Call GE Service.
Cannot detect
pacemaker pulse.
Structure
Below are the names of each part on the SEER Light/SEER Light Extend
Controller.
A B
1.2 in.
(30 mm)
3.5 in. (90 mm)
C
D
G
7.9 in. (200 mm)
037A
C data transfer cable Used to transfer data from the SEER Light recorder. When not in use, store the cable in
the guide on the backside panel.
G F1 button Used to enter the preview mode (wired or wireless) to confirm the quality of ECG
recording.
H F2 button Used to enter the data transfer mode to transmit data to a Holter card.
I F3 button Used to start recording of the SEER Light recorder and to start transferring data to a
Holter card.
L battery cover In direction indicated, slide the cover open and place four new alkaline AAA type batteries
in the battery box.
Troubleshooting
Software Version
The software version of the SEER Light/SEER Light Extend Controller
appears in the lower right hand corner of the display upon start up.
Self-Test
A self-test can be done on the following functions.
The name of the button that was pressed will be displayed, and
an audible sound will be heard when the respective button is
pressed once.
NOTE
If the ENTER button is pressed three times the unit will
enter into the LCD check mode.
Do the
batteries have sufficient No
power of >4.5 V?
Did the
beep sound when the Yes
POWER button was
pressed?
Call GE service.
No
Call GE service.
Is the I.R.
window facing the No
recorder properly?
Is cable connection No
proper?
No Call GE service.
Call GE service.
Does
the controller display Yes
“Already started?”
Can
the operation panel No
key(s) be pressed
properly?
Call GE service.
Yes
Is there
debris between PC Yes
board and key sheet
panel?
No Call GE service.
Call GE service.
Is the
card inserted in the Yes
opposite direction?
Call GE service.
No Replace the
connector.
Has
the compact flash card No
been formatted?
No Replace with a
new card.
No Transfer the
data to the PC.
Are
there distorted card Yes
connector pins?
No Call GE service.
Call GE service.
Is date within
No
one month?
Yes
Adjust date.
Is it within 5 years No
of use?
Are the
battery electrodes No
clean or properly
solder?
Inspection
Production process inspection sheet and inspection procedure.
M. No.
Product Model Name SEER Light Controller Date (Start Inspection) Approval Audit
Serial No. Board No.
Production Lot No.
Program Version Production Specification
High temperature heat run (40 to 48 hours) Start and End Date
Inspection Before Heat-Run Staff
No. Item Inspection Procedures Result Accept Reject
1 Appearance Check paint condition, no scratches or stains on LCD and cases, clear wording,
etc. Check connector pin of memory card for breaks, bends, etc.
2 Power switch Turn ON the power switch. Ensure there are buzzer sounds and the factory reset
menu appears on the LCD. Confirm program version number.
3 Set the time Set year, month, date, and time. Turn off the Auto Power function for heat-run
inspection.
4 Key/LCD check Under the inspection mode, ensure the key input and LCD display are normal,
LCD backlight is lit, and the keys should not stick when pressed.
5 Leakage current Patient leakage current is less than 10 µ A. µA
6 Power ON Insert the batteries and confirm buzzer one time. Press start switch, confirm to buzzer and
blink red LED. Then press stop switch and check buzzer sound and LED blinking are
stopped.
7 Start test Ensure that the SEER Light is starting (straight line: 1.5 m).
8 IrDA preview Use SEER Light, which is pre-installed with ECG wave, and ensure that the
check displayed waveform of 1 ch, 2 ch, and 3 ch are normal under the preview function.
9 Preview check Use SEER Light, which is pre-installed with ECG wave, and ensure that the
with cable displayed analog waveform of 1 ch, 2 ch, and 3 ch are normal.
10 Analog output Using inspection controller with cable, confirm analog waveforms 1 ch, 2 ch, and 3 ch are
normal.
11 Transfer test Connect the SEER Light. Ensure that data is transferred and that the data can be
installed onto the Holter card.
12 Check settings Check the year, month, date, and time (with ±1 minute).
13 Reset menu Used to initialize the settings (menu).
14 Auto Power Off Turn ON the power switch. Check that the power is automatically switched off if no
function operation for 15 minutes continuously.
15 Appearance Check painting conditions such as no scratches stains on LCD cases, clear wording,
Ver. No.
etc. Check serial plate and battery mark seal (version number) is securely affixed.
16 Supply current Start at 6 V. The supply current is 60 ±6 mA. mA
No. Visual Inspection Accept Reject No. Visual Inspection Accept Reject
1 Battery mark 1 place 4 Electrode welding 2 places
2 Ribbon 1 place 5 LCD boards, connector cutting 1 place
3 Board installation 1 place 6 Main body assembly 4 places
Notes:
Structure
Below are the names of each part on the SEER Light Connect.
084A
Name Function
A data transfer cable Used to transfer data from the SEER Light recorder
to the SEER Light Connect.
C USB Connection Uses a USB patch cord to transfer data from the
SEER Light Connect to the Holter analysis system.
Troubleshooting
WARNING
Software Version
To view the software version of the SEER Light Connect device, click
About in the SEER Light Hookup window.
Call GE Service.
Yes
Is the I.R.
window facing the No
controller properly?
Has recording
Yes
already started?
Call GE Service.
Yes
Does
the controller display Yes
“Already started?”
Is the distance
from the Connect No
within 4.9 ft
(1.5 m?)
Adjust the position
within 4.9 ft. (1.5 m)
Yes from the Connect.
Cannot Transfer Data from the Recorder to the SEER Light Connect Device
Cannot transfer data to
the Connect.
Plug it in.
Yes
Is the transfer
cable properly No
connected?
Connect the
Yes cable properly.
Does the
controller display “No data Yes
in the recorder?”
No data in the recorder.
No Record ECG again.
Call GE Service.
No
Call GE Service.
Is date correct on
No
the PC?
Yes
Adjust date/time on the PC.
Call GE Service.
Maintenance
Visual Inspection
Perform a visual inspection daily. If you notice any items that need
repair, contact an authorized GE Medical Systems Information
Technologies service person to make the repairs.
Check the case and display screen for cracks or other damage.
Regularly inspect all cords and cables for fraying or other damage.
Inspect all plugs, cables, and connectors for bent prongs or pins.
Precautions
Do not immerse any part of the equipment in water.
NOTE
Cleaning
Clean the exterior surfaces with a clean, soft cloth and a mild
dishwashing detergent diluted in water.
Wring the excess water from the cloth. Avoid contact with open
vents, plugs, or connectors.
Cleaning Frequency
With each new patient use, clean the SEER Light Recorder/Controller
and the SEER Light recorder pouch, leadwires, and patient cable.
When the SEER Light Recorder/Controller will not be in use for a period
of time:
Maintenance/Repair Log
Institution Name:
Date Maintenance/Repair Technician
Maintenance/Repair Log
Institution Name:
Date Maintenance/Repair Technician
Technical Specifications
Description Specification
Dimensions Height: 54 mm (2.1 in)
Width: 85 mm (3.35 in)
Depth: 15 mm (0.6 in)
Weight 72 g (2.5 oz) including batteries
Operating temperature 0 to 45°C (32 to 113°F)
Operating humidity 10 to 95% relative humidity (no condensation
allowed)
Storage temperature –20 to 65°C (-4 to 149°F)
Storage humidity 5 to 90% relative humidity (no condensation
allowed)
Material Aluminum ABSPC (main body)
Recording time within 24 hours — SEER Light
within 48 hours — SEER Light Extend
Recording method Digital memory
Power supply 2 x AAA Alkaline battery
Recording channel ECG: 3 channels
Movement level: 1 channel
Pacemaker pulse: 1 channel
Pacemaker detection channel CH1
Memory 32 Mbyte
Input level 16 mV p-p
A/D converter 10 bit, 8 ms sampling
Frequency response 0.05 to 40 Hz
Safety Type B
Input impedance Over 10 M Ohms
Time data backup Within 1 week (when started by the Cardy 301
recorder)
Description Specification
Dimensions Height: 200 mm (7.9 in)
Width: 90 mm (3.5 in)
Depth: 25 mm (1.2 in)
Weight 285 g (9.9 oz) including batteries
Operating temperature 10 to 35°C (50 to 95°F)
Operating humidity 10 to 95% relative humidity (no condensation
allowed)
Storage temperature –20 to 65°C (-4 to 149°F)
Storage humidity 5 to 90% relative humidity (no condensation
allowed)
Power supply 4 x AAA Alkaline battery
Vibration endurance Operation: 0.5 G (10 to 20 Hz)
Non operation: 3.0 G (100 to 300 Hz)
Data transfer medium Compact flash card
Time accuracy + 60 seconds per month
Description Specification
Dimensions Height: 13 mm (.51 in)
Width: 90 mm (3.54 in)
Depth: 51 mm (2.01 in)
Cable length: 260 mm (10.24 in)
Weight 87 g (.19 lbs)
Operating temperature 10 to 35°C (50 to 95°F)
Operating humidity 10 to 95% relative humidity (no condensation
allowed)
Storage temperature –20 to 65°C (-4 to 149°F)
Storage humidity 5 to 90% relative humidity (no condensation
allowed)
Power supply DC 5V
Recorders SEER Light Recorder
SEER Light Extend Recorder
WARNING
Use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause
unexpected or adverse operation.
WARNING
The equipment or system should not be used adjacent to,
or stacked with, other equipment. If adjacent or stacked
use is necessary, the equipment or system should be
tested to verify normal operation in the configuration in
which it is being used.
RF emissions
Class A
CISPR11 The equipment is suitable for use in all
establishments including domestic
Harmonic Emissions
Class A establishments and those directly connected to
EN 61000-3-2
the public low-voltage power supply network
Voltage fluctuations/ that supplies buildings used for domestic
Flicker emissions Complies purposes.
EN 61000-3-3
Electromagnetic Environment -
Immunity Test EN 60601 Test Level Compliance Level
Guidance
Electrical fast ± 2 kV for power supply lines ± 2 kV for power supply lines Not applicable. The SEER Light recorder
transient/burst ±1 kV for input/output lines ±1 kV for input/output lines and controller are not powered by Mains
EN 61000-4-4 power.
Surge ± 1 kV differential mode ± 1 kV differential mode Not applicable. The SEER Light recorder
EN 61000-4-5 ± 2 kV common mode ± 2 kV common mode and controller are not powered by Mains
power.
Voltage dips, short <5% Ut (>95% dip in Ut) <5% Ut (>95% dip in Ut) Not applicable. The SEER Light recorder
interruptions and for 0.5 cycles for 0.5 cycles and controller are not powered by Mains
voltage variations on 40% Ut (60% dip in Ut) 40% Ut (60% dip in Ut) power.
power supply input for 5 cycles for 5 cycles
lines 70% Ut (30% dip in Ut) 70% Ut (30% dip in Ut)
EN 61000-4-11
for 25 cycles for 25 cycles
<5% Ut (>95% dip in Ut) <5% Ut (>95% dip in Ut)
for 5 sec for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be
(50/60 Hz) magnetic at levels characteristics of a typical location
field in a typical commercial or hospital
EN 61000-4-8 environment.
NOTE
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from structures, objects,
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, AM
and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the equipment is used exceeds the applicable RF compliance level above, the equipment should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the equipment.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
The SEER Light recorder and controller are intended for use in the
electromagnetic environment on which radiated RF disturbances are
controlled. The customer or the user of the SEER Light recorder and
controller are can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the SEER Light recorder
and controller as recommended below, according to the maximum output
power of the communications equipment.
100 12 12 12 23
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can estimated using
the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (w)
according to the transmitter manufacturer.
NOTE
WARNING
The use of accessories, transducers and cables other than
those specified may result in increased emissions or
decreased immunity performance of the equipment or
system.
NOTE
Any supplied accessories that do not affect EMC compliance are not
included.