TN3501 - SEC-MALS Noise Assessment Guide Rev A
TN3501 - SEC-MALS Noise Assessment Guide Rev A
TN3501 - SEC-MALS Noise Assessment Guide Rev A
or wavy light scattering baseline thereby decreasing into the MALS ........................................................................... 4
the signal-to-noise ratio and obscuring meaningful Check the noise of the autosampler ........................................ 5
signals. This guide will describe how to isolate the
Check the noise of the UV detector (if applicable) ............ 6
source of the noise in the HPLC/MALS system in
Flush the entire system (except for the columns) ............... 7
order to clean and remove contaminants.
Check the noise of the columns ................................................ 7
Related Technical Notes and Appendix 1: General System Cleaning Procedures ........ 9
Contents
Introduction ................................................................................... 2 Figure 1: An example of noisy light scattering signals from
Tools and Supplies needed ...................................................... 2 multiple scattering angles collected from a Bovine Serum
Albumin run through a dirty MALS instrument flow cell.
Procedure ........................................................................................ 3
Introduction
Before an SEC-MALS measurement can be started, your chromatography system must be equilibrated. When the
baseline is stable, a measurement can be performed. Ideal baseline and noise levels depend on your detector and
the system solvent:
Performing measurements with an increased baseline voltage or noise decreases the accuracy of the results.
Therefore, the root cause of the noise problem should be identified. Noise in the light scattering trace can be
caused by particles or contaminants in one or more of these components:
Note: Before you start, contact your HPLC vendor to make sure all components of your HPLC system are
compatible with nitric acid and any other solvent used in the cleaning protocols described below. Wyatt’s
MALS and RI detectors are compatible with 10% nitric acid. Do not flush your column with nitric acid.
TN3501 Rev A
SEC-MALS Noise Assessment and Cleaning Guide Page 3 of 10
Procedure
Note: You may substitute the 0.02 µm filter with a 0.1 µm syringe tip
filter if the smaller pore size is not available.
4) After flushing, check the noise again. If it is acceptable, it is Figure 3. The flow path used to check
the noise of the MALS instrument itself.
recommended that you continue checking each component of the
system using the steps below – the MALS detector may not be the
only cause of noise.
Check the noise of mobile phase with an offline measurement (aqueous only)
1) If the peak-to-peak noise determined in step 2 of the “Check the noise in the MALS detector with an offline
measurement” section was 30 µV or less in pure solvent, use the syringe pump to flush your instrument with
your mobile phase that was used during collections in which high noise levels were observed.
3) If the peak-to-peak noise is 30 µV or less, the mobile phase is clean; proceed to the next section. Otherwise,
replace mobile phase and see if the noise improves.
TN3501 Rev A
SEC-MALS Noise Assessment and Cleaning Guide Page 4 of 10
Check the noise of pump and inline filter flowing directly into the MALS
Figure 4. The flow path used for checking the noise of the pump, degasser, and inline filter.
Note: Avoid introducing air into the MALS detector. Make sure that the tubing used is filled with liquid. It is
recommended to flush a couple of drops out of the tubing when connecting to the MALS detector.
1) Connect the HPLC pump to the MALS detector bypassing all components between the pump/degasser and
the MALS detector (except the inline filter). See Figure 4.
2) Set the flow rate to 0.5-0.8 mL/min. It will take 5-10 min for the initial noise to decrease. Record 5-10 minutes
of data using your standard online ASTRA method. Uncheck the “Trigger on Auto-Inject” checkbox in the Basic
Collection window and hit the Run button.
3) If the peak-to-peak noise is 30 µV or less, the pump and inline filter are clean; proceed to the section below. If
it is higher, check the pump frit and the inline filter.
TN3501 Rev A
SEC-MALS Noise Assessment and Cleaning Guide Page 5 of 10
Figure 5. The flow path used for checking the noise of the autosampler.
2) Set the flow rate to 0.5-0.8 mL/min. It may take 5-10 min for the initial noise to come down. Record two
ASTRA collections using your standard online ASTRA method. Uncheck the “Trigger on Auto-Inject” checkbox
in the Basic Collection window and hit the Run button.
3) Check the noise with the injection valve in mainpass (flushing through the loop). If the noise increases
significantly when the system is in mainpass mode, the autosampler loop is dirty and needs to be cleaned.
4) To clean the loop, switch the system to 20% alcohol and make 5-10 injections every 5 minutes. Change back
to your mobile phase and check the new noise level. It should be less than 30 µV. If the noise is acceptable it
is recommended that you continue checking each component of the system using the steps below – the
autosampler may not be the only cause of noise.
TN3501 Rev A
SEC-MALS Noise Assessment and Cleaning Guide Page 6 of 10
Figure 6. Flow path used for checking the noise of the UV detector.
4) Set the flow rate to 0.5-0.8 mL/min. It may take 5-10 min for the initial noise to decrease. Record data using
your standard online ASTRA method. Uncheck the “Trigger on Auto-Inject” checkbox in the Basic Collection
window and hit the Run button.
5) If the peak-to-peak noise is 30 µV or less, the UV detector is clean. Otherwise, clean the UV detector
according to vendor instructions and check the noise level again. You may need to replace the UV detector’s
flow cell.
6) Before testing the last source of noise (the column), proceed to the next section to prepare the system for
attaching the column.
TN3501 Rev A
SEC-MALS Noise Assessment and Cleaning Guide Page 7 of 10
2) If you ran a nitric acid wash, check the pH of the mobile phase after it passes through the detectors; it should
no longer acidic from the nitric acid wash. Also change the filter membrane of the inline filter if present.
3) Record the noise in ASTRA. Uncheck the “Trigger on Auto-Inject” checkbox in the Basic Collection window and
hit the Run button.
4) If the peak-to-peak noise is 30 µV or less, proceed to the next section. If the noise is still high, consider
changing the tubing.
TN3501 Rev A
SEC-MALS Noise Assessment and Cleaning Guide Page 8 of 10
1) Disconnect the MALS detector from the flow path prior to connecting the column for its initial equilibration.
2) Decrease the flow rate of your system to 0.1 ml/min. Add the column to the flow path and increase the flow
2
rate slowly at a rate of 0.1 ml/min .
3) Once you’ve reached your standard system flow rate, let the column equilibrate per the manufacturer’s
instructions.
4) Once the column is equilibrated, reconnect the MALS detector to the flow path. Regular downstream HPLC
detectors can often be connected at the chromatography flow rate since they do not add significant back
pressure for typical flow rates and solvents. UHPLC detectors should not be connected at the regular flow rate
as they add significant back pressure; instead, they should be connected at zero or near zero flow rate (follow
your column manufacturer’s recommendations). Estimated back pressures in water at typical flow rates are
listed in Table 1 for HPLC detectors and in Table 2 for UHPLC detectors.
Table 1. Estimated back pressure for HPLC detectors in water at a flow rate of 1 mL/min.
Table 2. Estimated back pressure for UHPLC detectors in water at a flow rate of 0.5 mL/min.
5) After reconnecting the MALS detector, continue flushing the column. Once the initial noise of the column has
decreased after connecting the column to the MALS detector, purge the RI detector for at least 30 minutes.
6) Record an ASTRA collection (5-10 min) for your setup and note the noise. It should be 50 µV or lower. Up to
100 µV noise can be tolerated for sample injections with a high signal-to-noise ratio, e.g. high molecular
weight polymers or particles.
7) If the noise is still high, consider reconditioning or cleaning the column per manufacturer’s instructions. If
cleaning is not successful it may be necessary to replace the column.
TN3501 Rev A
SEC-MALS Noise Assessment and Cleaning Guide Page 9 of 10
1) 50 mL water
3) 50 mL water
5) 100 mL water
Note: If cleaning solutions such as Tergazyme or Hellmanex are compatible with the HPLC system according to
the manufacturer, then an additional cleaning step can be performed using these solutions as they can be
effective at removing proteins.
2) 50 mL water,
4) 100 mL water
Notes: Make sure to change the inline filter membrane after flushing the HPLC system with nitric acid before
reconnecting your column and running experiments.
If in-situ cleaning of the MALS detector as described above does not improve the cleanliness of the flow
cell, disassembling the flow cell and cleaning must be performed as described in the Service and
Maintenance section of the instrument’s User’s Guide. A detailed description of the cell cleaning
procedure is also provided in your LSU binder and a detailed video for cell cleaning can be found in the
Wyatt support center at www.wyatt.com/support (navigate to the Tutorials section).
TN3501 Rev A
SEC-MALS Noise Assessment and Cleaning Guide Page 10 of 10
Clean out the fan filters on the back of the MALS and Optilab instruments every three months. An
ambient MALS instruments should run at a temperature ~ 5 °C above room temperature if sufficient
space for ventilation around the instrument is provided.
Check the mobile phase daily with a laser pointer for particulates. Alternatively, you can use a batch DLS
instrument to detect particulates.
Switch off the MALS instrument laser (on the front panel when not in use), or use the Laser Saver Mode in
ASTRA (basic collection procedure) which accomplishes the same thing after running an acquisition. Install
the latest firmware for HELEOS II, TREOS, and μDAWN instruments. Switch off the MALS instrument
completely if you are not using for more than three days. It takes about 30 min to warm up the instrument
electronics. The laser warmup itself is completed within a few seconds.
Consider a column cleaning or regeneration step of your column according to the manufacturer’s
recommendation. Disconnect the MALS instrument when performing column cleaning.
If you need to open the top cover of the MALS or Optilab instruments and reach inside, make certain that
you ground yourself electrostatically. Disposable wrist straps are available in the MALS instrument
hardware kits.
When installing new loops or UV flow cells, pre-flush with alcohol (100% or 20% according to the
manufacturer’s specification) before installing.
When not using the HPLC system, flush into 20% alcohol and keep running at 0.1 mL/min.
Purchase a syringe pump for the lab for offline work (a basic model will do since the exact flow rate
delivery is not critical).
Perform maintenance on your HPLC system at regular intervals (according to your HPLC manufacturer’s
recommendations).
Note: Please see TN3502 – Guidelines for Improving the Quality of SEC-MALS Data in an Aqueous System for
more tips on performing successful SEC-MALS experiments.
TN3501 Rev A