Installation Guide
Notice
Congratulations on your purchase of the KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging
System. Thank you for your confidence in our products and we will do all in our power to ensure your complete
satisfaction.
The Installation Guide for the KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System
includes information on the cephalometric features. For the panoramic features, see the KODAK 8000 Extraoral
Imaging System (SM721) Installation Guide. We recommend that you thoroughly familiarize yourself with this
Guide in order to make the most effective use of your system.
WARNING: We recommend that you consult the “Safety, Regulatory and the Technical
Specification User Guide” before using the KODAK 8000C Extraoral Imaging Systems.
The information contained in this Guide may be subject to modification without notice, justification or notification
to the persons concerned.
No part of this Guide may be reproduced without the express permission of Carestream Health, Inc.
The US Federal law restricts this device to sale by or on the order of a physician.
Manual Name: KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation
Guide
Part Number: SM736
Revision Number: 02
Print Date: 03/2010
The brand names and logos reproduced in this Guide are copyright.
KODAK is a trademark of KODAK used under License.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System, complies with Directive
93/42/CEE relating to medical equipment.
.
0086
Manufacturer
Carestream Hea lth, Inc.
150 Verona Street
Rochester NY 14 608
TROPHY
4, Rue F. Pelloutier, Croissy-Beaubourg
77435 Marne la Vallée Cedex 2, France
Contents
1—About This Guide
Conventions in this Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 iii
Contents
6—Maintenance
Annual Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–1
iv
Chapter 1
About This Guide
WARNING
Warns you to avoid injury to yourself or others by following the
safety instructions precisely.
CAUTION
Alerts you to a condition that might cause serious damage.
IMPORTANT
Alerts you to a condition that might cause problems.
NOTE
Emphasizes important information.
TIP
Provides extra information and hints.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 1–1
Conventions in this Guide
The KODAK 8000C digital panoramic and cephalometric unit is designed to carry out the
following radiological examinations:
• Panoramic
• Maxillary Sinus
• Temporomandibular Joints (TMJ)
• Lateral cephalometric
• Frontal (PA or AP) cephalometric
• Oblique cephalometric
• Submento-vertex cephalometric
• Carpus cephalometric
General Overview
The KODAK 8000C digital panoramic and cephalometric unit is composed of the
following functional components:
The following figures illustrate the general overview of the KODAK 8000C digital
panoramic and cephalometric units.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 2–1
General Overview
IMPORTANT
The Cephalostat can be positioned either on the right or the
left side of the KODAK 8000 unit.
Mobile Components
Figure 2-1 illustrates the up and down movement of the KODAK 8000C digital
panoramic and cephalometric units mobile component and the rotation of the rotative
arm.
Figure 2–1 KODAK 8000 and KODAK 8000C Units Mobile Components
Figure 2-2 illustrates the general functional components of the KODAK 8000C digital
panoramic and cephalometric units.
Figure 2–2 KODAK 8000 and KODAK 8000C Units Functional Components
20
19
21
22
18
23
24
16
15
17
5
1
25
4a
14
29
2
4
28
3 3a 26
8 27
12
10
RJ45/1
RJ45/2
9
6 7 13
11
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 2–3
General Overview
Figure 2-3 illustrates the locations of the digital panoramic and digital cephalometric
sensors of the KODAK 8000C digital panoramic and cephalometric units.
Figure 2–3 KODAK 8000 and KODAK 8000C Units Digital Sensor Locations
system
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 2–5
General Overview
Laser Locations
Figure 2-4 illustrates the location of the lasers of the KODAK 8000C digital panoramic
and cephalometric units.
Figure 2–4 KODAK 8000 and KODAK 8000C Units Laser Beam Locations
3 2
Control Panel
The control panel is an alphanumeric, digital soft touch console. It allows the operator to
control certain unit functions. It also displays the operating parameters and error
messages.
kV mA S
1 2 3 4
1 X-Ray emission LED: Yellow, indicates the x-rays are being emitted.
2 Display Screen: Displays the current acquisition parameters and the error messages.
3 Reset button: Resets the unit arm to the initial position to enable the patient to enter
and exit the unit.
4 Laser beam button: Activates the laser positioning beams to correctly position the
patient.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 2–7
X-Ray Remote Control Overview
The x- ray remote control enables you to launch a radiological image acquisition via the
exposure button from outside the x-ray room. You must press and hold the exposure
button until the end of acquisition. Premature release of the exposure button interrupts
the acquisition.
The following accessories are used when positioning a patient. They are delivered with
the KODAK 8000 digital panoramic and KODAK 8000C digital panoramic and
cephalometric unit.
Table 2-1 and Table 2-2 list the panoramic and cephalometric positioning accessories.
Accessory Description
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 2–9
Positioning Accessories and Replacement Parts
Accessory Description
x2
Nasion support
• Cephalostat head
• Cephalostat arm
• Cephalostat covers 1200mm (L) x 800mm (D) x 1180mm (H) 80 kg
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 3–1
Standard Packaging
IMPORTANT
Prior to placing the order and before installation, carefully check the
following requirements for the x-ray room.
Standard Compliance
Install the unit in an x-ray room compliant with all official regulations applicable to
protection against radiation. This room must reduce the frequency interferences of the
30MHz to 1GHz band to at least 12db.
Environmental Requirements
Check the following ambient operating condition requirements of the x-ray room before
installing the unit:
• Temperatures: 5 ~35 °C
• Relative humidity: 30 ~ 85%
• Atmospheric pressure: 700 ~ 1060 hpa
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 4–1
Unit Dimensions
Unit Dimensions
406 mm 1261 mm
(15.98") (49.63")
306 mm
(12.05")
10 mm
(0.38")
1810 mm 440 mm
(71.3") (17.3")
WARNING
If you need to add a base plate you must add 15mm to the height of the unit.
500
600
Electrical Requirements
WARNING*
You MUST select the operating voltage when placing an order. The
operating voltage CANNOT be modified on site.
• 100/110/130 V 50/60 Hz
• 230/240 V 50/60 Hz
CAUTION
The power supply line must be equipped with a connection box that
ensures a constant connection. It must not be possible to connect the
unit to the power supply without using a tool. The unit must be
protected against any accidental disconnection.
If other units are installed on the same line, interference and voltage fluctuations can
cause the radiological unit to operate abnormally. We strongly recommend that a
separate electrical line be dedicated to supply power to the KODAK 8000C unit.
This line should be protected by a circuit breaker with a maximum current of:
• 16 A at 230/240 V
• 20 A at 110/130 V
• A differential circuit breaker of 30 mA
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 4–3
Electrical Requirements
G L N
13
4
L
1A
6 5
RJ45
K8000C
8
L N
12
10
7bis 9
L N
11
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 4–5
Electrical Requirements
Contactor:
Manufacturer: Telemecanique
Ref: LC1 D4011 • F6 (110/130 V - 60 Hz)
• P5 (230/240 V - 50 Hz)
• U6 (230/240 V - 60 Hz)
IMPORTANT
If you cannot use the above manufacturer devices, use an equivalent
emergency switch and contactor in the UL list with the same
specifications. For more information, see UL 2601.1 clause 22.7.
2x
2x
16 mm 16 mm
2x
2x
16 mm 16 mm
IMPORTANT
Install the unit where a minimum amount of space is provided to allow
easy access for the patient or the maintenance technician (see Figure
4-2).
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 4–7
X-Ray Room Requirement
Figure 4–2 Minimum X-Ray Room Space Configuration for K8000 and K8000C
RJ45/1
RJ45/2
MINI 2350 mm
(92.52")
MINI 500 mm
(19.68")
13
12 9 10 6 5 7bis 11
4 9 10
5 7
WARNING
You MUST prepare the appropriate electrical requirements and
configurations of the x-ray room before installing the unit. You MUST
locate separately the following low voltage and high voltage
connections:
- Low Voltage: (12), (9), (10)
- High Voltage: (6), (5), (7bis)
To keep you always updated on the latest evolution of the minimum computer system
requirements for KODAK 8000C system software, this section is posted on site
MyTrophytec: http://www.mytrophytec.com, as a separate document with edition
number. Before any intervention on the client’s site, check first the MyTrophytec site for
the latest edition.
IMPORTANT
It is MANDATORY to check that the computer system configuration is
compatible with the computer system requirements for the KODAK
8000C software. If necessary you MUST update your computer system
configuration. KODAK 8000C MUST be connected to the computer via
a point-to-point Ethernet link and not via a LAN. DO NOT place the
computer and the peripheral equipment connected to it in the
immediate vicinity of the patient in the unit. Leave at least 1.5 m
distance from the unit. The computer and the peripheral equipment
must conform to the IEC 60950 standard.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 4–9
Computer System Requirements
Tool Requirements
The installer must supply the following tools:
• Power drill
• Screws and heavy duty fixings
• Spirit level
• Measuring tape
• Cutter
• Metric Allen keys
• Metric spanners
IMPORTANT
The tool references mentioned in this manual are ISO tool
references.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–1
Opening the Box
Before opening the cephalostat assembly box make sure that you have the required tool.
1. Remove the upper cardboard from the packaging. DO NOT cut through the
cardboard.
5. Remove gently the cephalostat head from the transport support and place it on the
ground.
IMPORTANT
Make sure that the wall mounting bracket’s hole (A) is the furthest
from the wall before installing the cephalostat unit.
1. With the help of 1 person, position the cephalostat arm on the rear of the mobile
component of the column. To attach the arm, insert the screws on the column on the
lower holes (A). Insert and then tighten the fixing screws (C).
13
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–3
Installing the Cephalostat Unit
2. Make sure that the tube (A) is for the appropriate side (by default it is for the right side
position of the ceph), if not, unscrew the fixing screws and reposition it for the
appropriate side. Insert all the way tube (A) inside the tube clamps (B) and screw in
the fixing nut.Tigten the fixing screws of the tube clamps (B).
4 mm
6 mm
7 mm
22 mm
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–5
Installing the Cephalostat Unit
5. Correctly position the cephalostat head using a spirit level (C) placed as shown in the
figure. Tighten the fixing screws (B) to adjust with the spirit level (C).
4 mm
C B
6. Correctly position the cephalostat head using a spirit level (D) placed as shown in the
figure. Tighten the fixing screws (C) to adjust with to the spirit level (D).
2.5 mm
4 mm
6 mm
A
x1
C
B
C
x1
7. To adjust the height of the Cephalostat head, using a measurement tape, measure the
height from the floor to the middle of the collimator outlet (A). Measure the height
from the floor to the center of the ear cone ball (B). The measurements must be the
same (A = B). To adjust the height use the tube fixing nut (C), see step 2 of the
procedure.
A = B
8. Insert the cables (B) inside the elbow tube (A) tube.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–7
Installing the Cephalostat Unit
9. Pull the cables out of the elbow tube (A). Remove the adhesive tape. Connect the
following wiring in the following order:
D
A
2.5 mm
B
39 40
CJ653
J4
E "R" J6
J10 A
J5
1
J9
CJ699
40 39
B
C
1
J9
CJ699
E "L"
J1 1
CJ664
J1 0
D
1
J21
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–9
Post-Installation Control
Post-Installation Control
3. Launch an x-ray using the x-ray remote control. The head clamps must be correctly
and vertically positioned on the acquired x-ray image.
4. If needed, adjust the cephalostat head mechanically using the elbow tube for
horizontal inclination of the cepahlostat head. Untighten the fixing screws (A) slightly.
Use the fixing screws (B) to adjust the horizontal inclination. Tighten the fixing screws
(A) when you have reached the desired adjustment.
2.5 mm
4 mm
x4
A
x1
B
A
B
x1
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–11
Post-Installation Control
4. In the Imaging Window, on the drawing toolbar, click . On the acquired image,
draw a line from the bottom of the image to the center of the ball (ear cone). The
resulting value must be:
30 mm (± 5 mm)
110 mm (± 5 mm)
5. If needed, adjust the height, using the cephalostat arm screws (A). Untighten slightly
the screws (A) and lift the cephalostat head to the correct height. Tighten the
cephalostat arm screws (A).
x4
A
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–13
Post-Installation Control
3. Launch an x-ray using the x-ray remote control. The ear cone ball (A) must be at the
center of the head clamps circle (B).
4. In the Imaging Window, on the drawing toolbar, click . On the acquired image,
draw a line from the center of the ball (A) to the center of the circle (B). The resulting
value must be ±1.5mm.
+1.5 mm
4 mm
3. Launch an x-ray using the x-ray remote control. The nasion support must be
perpendicular to the ear cones.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–15
Post-Installation Control
4. In the Imaging Window, on the drawing toolbar, click . On the acquired image,
draw a perpendicular line between the center of the ear cones and the nasion support
holes to check that the nasion support is positioned vertically. The resulting value
must be A = B ±3mm.
A B
3 mm
5. If needed, adjust the nasion support, using the nasion support fixing screws (A).
2 mm
2. Turn on the unit. On the control panel, press to turn on the positioning laser
beams.
3. Make sure that the cephalometric Frankfurt positioning laser beam is projected in the
middle of the ear cone.
4. If needed, adjust the laser using the fixing screw of the laser socket.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–17
Post-Installation Control
3. Position the image quality phantom (A) at the center of the digital sensor.
• 64 Kv
• 15 mA
• 0.4 s
7. In the Imaging Window, on the acquired image, zoom on the value grid (B) of the
image quality phantom (A). The resulting value clearly visible must be minimum 2.5
(right, left, center and bottom).
Check that there is no splitting lines of the grid of the image quality phantom.
8. Save the image. Use this image as reference for a later image quality verification.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–19
Fitting the Covers
• All the screws in the wall and the floor have been tightened.
• All the screws attaching the head and the fixed arm to the column have been
tightened.
D
2x M5x20
B
2x M5x50
2x M5x15
C
E
2x M5x50
2x
M5x12
3 mm
1. Fit the upper cover (A) on the column. Fit the cover on the wall-bracket (B). Fit the
cover on the column connecting terminal (C) and tighten the fixing screws.
2. Fit the head cover (D) of the unit to the head chassis and tighten the fixing screws.
3. Fit under the head cover (E) and tighten the fixing screws.
4. Fit the inside column cover (F) and position and attach securely the ON/OFF button
to the cover with the fixing screws. Screw and tighten the fixing screws of the column
cover.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–21
Fitting the Covers
1. Fit the cephalostat end of the arm cover (A) and screw the fixing screws.
2. Fit the cephalostat head side covers (B) and screw the fixing screws.
3. Fit the cephalostat head top cover (C) and screw the fixing screws.
C
M5x12
4x
2x
M5x16
M5x12
5x
Annual Maintenance
We recommend that a general inspection of the unit should be carried out every year by
an approved dental systems technician.
• Make an image acquisition with the test tools and check the image.
If the results of any of these inspections are unsatisfactory, refer to the Service Manual in
order to rectify any problems.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 6–1
Annual Maintenance
6–2 Maintenance