To secure the space required for maintenance, install the IPU on the
right side of the XS main unit. Provide a distance of at least 50cm
behind the instrument.
XS-800i
XS-1000i
Fixing Tape
Transportation
Tape
(16) When main unit indicator located lower left become green, run
Setting sequence by selecting Controller( Service ( Setting
Sequence. It takes appoximately 24 minutes.
(14) Remove CD-ROM from the CD-ROM drive. Run IPU program by
!
selecting Start Æ All Programs Æ Startup ÆIPU. Log off IPU and
log on IPU with User name: sysmex and Password: “c”, “9”, “.”, “0”.
NOTE:
This uninstallation must be performed after logon to windows as
“Sysmex” and password as “c9.0”.
When IPU program is running, quit the IPU program by selecting file
Æ Exit. When confirmation screen appears, click OK.
When uninstalltion is performed, all files including setting files will be
deleted.
1
-1
1
00
SU
.11
OP
SN
Upper Cover
Fixing Screws
Cover Lock
Fixing Screws
Tie Wrap
Connect 6 connectors
3675 3677 Stamped wiring cord
3676 LED1_2 No. in Sampler side
ST1 Barcode Reader
M4X6 (348-3926-8)
M4X30 (348-3935-1)
Positioning Pin
M4X6 (348-3926-8)
10 200635
(1) Select Start Æ All Programs Æ Startup Æ IPU to start IPU program.
Logoff IPU program and logon with logon name “sysmex” and
password “C 9 . 0”.
(2) Turn on the main unit and start up option screen appears. Set
options as follows.
10 200635