04 - C208B - ND (Rev 3) - Part 6 - EDDY CURRENT
04 - C208B - ND (Rev 3) - Part 6 - EDDY CURRENT
04 - C208B - ND (Rev 3) - Part 6 - EDDY CURRENT
20-00-01-0 (Rev 3)
(5) An adjustable, self-expanding probe is recommended for semiautomatic inspection to ensure consistent coil contact
with the side of the hole.
(6) If a solid non-expanding probe is used it must fit the hole to allow for a minimum amount of lift-off. The difference
between the diameter of the probe and the diameter of the hole shall not exceed 0.005 inches either without tape
applied over the coil or after application of tape.
(7) Choose a probe that fits the hole snugly while still allowing for free probe travel.
(8) Teflon™ or similar tape may be used to help improve the signal to noise ratio and to decrease the wear on the eddy
current probe coil.
(9) The probe shall be capable of operating at the frequency required for the material being inspected. Refer to Table 1
for guidance on probe operating frequencies.
Table 1. Materials
Condition Material Type (NOTE 1) Frequency (NOTE 2)
Aluminum, Magnesium (15% or greater
Nonmagnetic, High conductivity 200kHz-500kHz
IACS)
Magnesium (< 15% IACS), Titanium,
Nonmagnetic, Low conductivity 1MHz-3MHz
Inconel and Nickel alloys
400 series Stainless, 17-7 TH1050,
High permeability 4130 or 4300 series Steels, in any heat 400kHz-1MHz
treat condition
301 series, 304 series stainless, 17-7
Low permeability or Nonmagnetic 1MHz-3MHz
PH annealed
NOTE 1: Material types listed are for information only, these are examples of common materials that
would fall into the ranges listed under the heading, "Condition"
NOTE 2: The frequencies listed above are generally recognized to provide the optimum sensitivity in
the materials listed. Use of frequencies outside of those listed are allowed providing that the
sensitivity can be verified on the correct size EDM notch on a reference standard of similar
material.
D. Reference Standards.
(1) Reference standards should be constructed of an alloy similar to the base metal of the test object and conform to the
requirements listed in Table 2.
(2) Reference standards shall have a minimum surface finish of 150 RHR or RMS 165.
(3) Reference standards used in accordance with this procedure shall at a minimum, have 1 corner EDM notch of 0.030
inch x 0.030 inch x 0.007 inch for the hole size being inspected.
(4) A 0.030 inch x 0.030 inch corner notch or 0.050 inch x 0.050 inch corner notch shall be used for calibration unless
otherwise specified. Refer to Table 3 and Table 4 in the calibration sections for sensitivity levels for the different sized
notches.
(5) The dimensional accuracy of EDM notches shall be documented and traceable to the National Institute of Standards
and Technology (NIST).
(6) A typical eddy current bolt hole reference standard is shown in Figure 1.
Table 2. Materials
Inspection Material Type Material for Reference Standard (Note 1)
All conductive non-ferromagnetic alloys with Non-ferromagnetic alloy with a conductivity that 2024-T3 27.5-32.5
conductivities of 15 percent IACS or greater. is within ±15 percent IACS of the test material %IACS 7075-T6 30.5-
range, but not lower than 15 percent IACS. 36 %IACS 7075-
T76XX 38-42 %IACS
6013-T6X 40-43
%IACS 6061-T6 40-50
%IACS
All conductive non-ferromagnetic alloys with a Non-ferromagnetic alloy with a conductivity that Titanium 6AL-4V 1.01
conductivity of 0.8 percent IACS to less than 15 is not more than 1.0 percent IACS greater than %IACS Inconel 1.7%
percent IACS. the test material range or less than 0.8 percent IACS Magnesium
IACS. AM100A-T4, AZ91C-
T4, AZ91E-T4 9.9
%IACS
High permeability steel 400 series Stainless, 17-7 TH1050, 4130 or 4300 series steels in any
heat treat condition.
400 series stainless steel alloys. Any 400 series stainless steel alloy.
Low permeability alloys. 304 stainless or 17-7 PH annealed.
NOTE 1: Materials listed are for information only. These are some common materials available and may be
used. Reference standard material needs to conform to the requirements of Table 2.
5. Inspection Preparation.
A. Hole bores shall be round and smooth enough to not cause eddy current signals that would exceed a 3:1 signal to noise
ratio. Refer to Figure 10 and Figure 11.
B. Visually inspect hole bores for nicks, burrs, scoring or other conditions that may cause an excessive noise signal.
C. Any hole condition that would cause damage to the probe is not suitable for inspection. The condition must be corrected
prior to inspection.
D. If necessary, clean up the holes prior to the inspection utilizing one or more of the following methods.
(1) Methods generally considered acceptable for cleaning are, (but are not limited to).
(a) Solvent moistened cotton swab.
(b) Scotch bright.
(c) Flex Hone.
E. Do not oversize the hole during the cleaning process unless specifically allowed per the Maintenance Manual, the
Structural Repair Manual, the Component Maintenance Manual, or on direction from Cessna Engineering.
F. If allowed per the Maintenance Manual, Structural Repair Manual, Component Maintenance Manual, or other maintenance
document or Engineering, a 1/64th inch, or the amount as directed, clean up ream may be performed if needed to achieve
a satisfactory surface condition of the hole.
G. If the hole cannot be satisfactorily cleaned up to allow for an inspection, contact Citation Customer Service for instructions.
Refer to the Introduction for contact information.
H. Continue to Section 6 for Manual Probe Calibration or Section 7 for Calibration for Semiautomatic Hole Inspection.
6. Manual Probe Calibration.
A. The instrument shall be calibrated and operated in accordance with this procedure and the manufacturer's instructions.
B. The minimum single layer material thickness that can be examined using a manual probe hole inspection is 0.063 inch.
C. Instrument calibration shall be performed prior to inspection. Calibration shall be checked periodically during continuous
use and at the conclusion of the inspection. If the calibration is below the levels originally established, all holes inspected
since the last known good calibration check shall be reinspected. The instrument calibration shall be verified if any part of
the system is replaced or if any calibrated control settings are changed.
D. Teflon™ Tape is highly recommended to be placed over the probe coil. It will act as a wear surface and on high
permeability steel will greatly improve signal to noise ratio.
E. Positioning the thumb screw in line with the coil will help to define the position of the coil when the coil is in the hole. Refer
to Figure 2.
F. The use of a Low Pass Filter will help to improve the signal to noise ratio. Set it to the highest value that does not cause
the signal to be unstable.
NOTE: Do not use the High Pass filter with the manual probe, regardless of the coil type.
G. Setting Lift-Off for Absolute Probe.
(1) Select the probe for the best fit in the hole to be inspected and based on the requirements of Table 1.
(2) Select the standard for the material being inspected based on the requirements of Table 2.
for Semiautomatic Hole Inspection Calibration after setting lift-off to ensure filter settings are optimized for the
inspection.
(l) Turn on the scanner.
NOTE: If the scanner rotation speed is adjustable, set it for a minimum of 1,000 RPM's. For larger
diameter holes scan speeds over 1500 to 2000 RPM may be necessary to ensure
complete coverage unless scan index speeds are kept very slow.
1 Slower scanner rotation speeds will require slower scan indexing speed in the hole.
2 For steel or other high permeable materials, a slower scan rotation is often helpful in reducing interference
from noise. If necessary, reduce scanner rotation speed until an acceptable signal to noise ratio is
achieved.
3 Do the Determining Maximum Scan Index Speed steps after final adjustment in scanner rotation speed.
(m) If the dot is not at the center of the screen, press the null button.
(n) Hold the scanner so the probe is parallel to the surface of the standard. Refer to Figure 9.
(o) Place the probe in contact with the standard using light pressure, keeping the coil away from edges and EDM
notches. This will generate a lift-off signal as the coil rotates. Refer to Figure 9.
(p) Adjust the phase or rotation to position the flat portion of the signal from lift-off horizontal, along the X axis. Refer
to Figure 9.
NOTE: The signal from lift-off should be a flattened figure 8 pattern. The less vertical spread there
is in the signal from the lift-off, the less base line noise there will be on the sweep trace. If
the signal from lift-off exceeds 2 major divisions in total vertical spread, reduce the gain
until the signal from lift-off is less than 2 major divisions vertically and all or mostly on
screen in the horizontal direction. Refer to Figure 9.
(2) Setting Filters
(a) Use the values in Table 4and or the following steps for initial filter settings:
1 With the scanner on and the probe in the appropriate hole in the reference standard.
2 Set the HPF to the lowest value.
3 Increase the HPF value until the signal sweep flattens out. Depending on where the LPF is set the signal
may look a little noisy at this point.
4 Set the LPF to it's highest value.
5 Decrease the LPF value to get a stable dot or until the spikes from the noise signals decrease and the
signal from the notch in the reference standard just starts to decrease.
Table 4. Probe Diameters
Probe Diameter (in inches) 5/32 to 7/32 7/32 to 5/16 5/16 to 7/16 7/16 to 3/4
LPF 500 500 700 1500
HPF 150 200 300 500
NOTE 1: The values in Table 4 are meant as a guide. They will generally provide good results.
Use them as a starting place and adjust them as needed to obtain the best signal to
noise ratio and the most uniform response from the EDM notch.
(3) Setting Inspection Gain.
(a) Follow the steps for Setting Lift-Off and Setting Filters prior to setting the gain.
(b) Follow equipment manufacturer's directions on how to turn the Sweep on, and set the scope to display the
sweep trace.
NOTE: Do not set the display to a Waterfall, Timed or Sync type display. Those displays are for
use with an automatic scanner and specialized equipment.
(c) Turn on the scanner.
(d) Insert the probe into the appropriate sized hole in the reference standard.
(e) Maximize the signal from the corner EDM notch at the surface of the standard by moving the probe in and out of
(2) If using a differential probe. It is optional to set the collar for a depth of 0.030 inches and index every 0.060 inches to
within 0.030 inches of the opposite surface.
NOTE: With a differential probe there will be very little or no impact from the interfaces of the different
layers.
(3) Calibrate manual probe, refer to Manual Probe Calibration.
(4) Set the collar for a depth as directed per Table 6 or section 6, Manual Probe Calibration for differential probe collar
depth.
(5) Insert the probe into the hole, hold the probe steady with the collar firmly seated on the surface of the structure.
(6) Press the Null/Balance button on the instrument.
(7) Rotate the probe through the entire diameter of the hole for each index depth required.
(8) Monitor the display through each rotation for any crack like indications.
NOTE: If there is dot movement not consistent with that of a crack. You may have nulled on or very
close to a crack. Position the probe in another spot in the hole at the same depth and re null,
and re inspect the hole at the same depth.
NOTE: As in a surface inspection, dot travel will follow the same direction relative to lift-off for a crack.
Dot movement in directions other than that which would be consistent for a crack in the
material being inspected need to be investigated for relevance before being rejected as a
crack.
(9) If an indication is noted, refer to the Evaluation section, for characterization instructions.
Table 6. Scan Index For Layer Thickness
Layer Thickness (in inches) Scan Index
.063 to 0.090 Middle of the layer.
0.090 to 0.120 .030 inch from each edge/interface.
Greater than 0.120 to 0.180 0.030 inch from each edge/interface plus the middle of the layer.
Greater than 0.180 0.030 inch from each edge/interface plus 0.060 inch index.
10. Semiautomatic Inspection.
A. Inspection steps for semiautomatic hole inspection.
(1) Calibrate per steps in the Calibration for Semiautomatic Hole Inspection section 7.
(2) Set the instrument to Sweep display.
(3) Visually inspect holes per the Inspection Preparation section, and correct any conditions that would prevent an
acceptable scan.
(4) Check fit the probe to the hole/holes being inspected to ensure that the correct probe size is being used. Adjust fit if
needed.
(5) With the probe scanner turned on, slowly insert the probe into the hole while monitoring the display.
(6) Make a complete scan of the entire length of the bore of the hole.
NOTE: A complete scan consists of monitoring the display as the probe is inserted all of the way into
and through the hole and then back out of the hole.
NOTE: Do not exceed the index speed established in the calibration section.
B. Any indication while in sweep mode that differentiates itself from baseline noise, needs to be investigated for the
possibility of it being a crack.
(1) With the probe in the hole and the scanner on, maximize the indication.
(1) Turn the Sweep mode off so the flying dot or impedance plane mode is on.
NOTE: As in a surface inspection, dot travel will follow the same direction relative to lift-off for a crack.
Dot movement in directions other than that which would be consistent for a crack in the
material being inspected need to be investigated for relevance before being rejected as a
crack. Refer to Figure 14, for examples of crack indications in different materials.
NOTE: If the indication in the display is traveling in a direction not consistent with a crack for the
material being tested, it is probably not a crack. It still may be a rejectable condition though.
Determine what the cause of the indication is to determine if it is rejectable.
C. If an indication is noted, refer to the Evaluation section, for characterization instructions.
11. Evaluation.
A. General
(1) Any vertical deflection that is distinguishable from the null position and lift-off, and is not caused by noise, lift-off or
part geometry is cause for further investigation regardless of signal amplitude. There are no acceptable cracks.
(2) Unless otherwise specified, all cracks are rejectable regardless of length or signal amplitude.
(3) Indications with a phase response different from the reference notch may indicate hole contamination or damage. If
indications exceed a 3:1 signal to noise ratio, additional hole preparation, reaming and/or cleaning, may be required.
(4) If horizontal lift-off is greater than 40% FSW on the impedance display check for out-of-round condition or interface
noise. If allowed, have the hole reamed and then reinspect.
NOTE: Any damage to the bore of the hole that cannot be corrected through limits established in the
Structural Repair Manual must be reported to Citation Customer Service Engineering, refer to
Introduction for contact information. Nicks and gouges are a rejectable condition as they
could easily develop into a crack. These should be reported the same as reporting a crack
indication.
(5) Do not report the depth of the crack radially (distance into the material from the edge of the hole). Due to the nature of
how a crack forms as compared to an EDM notch, depth measurements are very unreliable and should not be
reported.
(6) The use of a borescope, flashlight or other visual aids may be necessary to determine the cause of an indication.
(a) If a crack like indication is found:
1 Identify the circumferential position of the indication.
2 Using an approved marker, mark the position around the hole to define the orientation of the crack.
3 If the indication is repeatable, verify that it is not due to foreign material stuck to the side of the hole or
some other non relevant indication.
4 After the circumferential position has been established, inspect the hole again, if the circumferential
position of the indication changes, it is not a crack.
5 Look into the hole using any means suitable to determine if any foreign material in the hole could be
causing the indication.
6 If the fasteners were cadmium (cad) plated, pay particular attention to the possibility that some of the cad
plating could be stuck to the side of the hole. Cad plating will give a very strong crack like indication on
sweep mode but it will appear going near vertical when in the flying dot mode. Cad plating stuck to the side
of the hole will usually be small, darker grey specs and would be located in line with the circumferential
position of the crack indication.
NOTE: An X-ACTO™ blade or small dental pick can be useful to help remove a piece of cad
plating stuck to the side of the hole. Use care to not gouge or damage the bore of the
hole.
B. Manual Probe.
(1) For manual hole inspection, when the eddy current probe is centered over a crack, the signal will be at maximum and
a small movement of the probe will cause the signal to diminish, returning to the null point. Corrosion pits, foreign
material, gouges or other damage, will generally be apparent for a longer degree of probe rotation before the signal
from the indication returns to the null point.
(2) A crack indication will follow the same phase direction as a crack indication would in a surface inspection for the
material being inspected.
(3) If the indication in the display is traveling in a direction not consistent with a crack for the material being inspected, it
is probably not a crack. It still may be a rejectable condition though. Determine what the cause of the indication is to
determine if it is rejectable.
(4) Nicks and gouges may have a signal similar in phase to a crack but the indication on the screen will usually be wider
and more vertical and may apparent for a longer duration of probe rotation.
(5) Determining the length of a crack.
(a) Insert the probe into the hole to approximately the depth of the crack.
(b) Locate the crack and peak the amplitude from the crack.
(c) If the peak amplitude from the indication is off screen, reduce the gain until the whole indication is visible in the
display.
(d) Rotate the probe back and forth across the crack while moving the probe towards the end of the crack until a
point is reached where the crack signal peak amplitude has been reduced by 50 percent.
(e) Tighten the set screw.
(f) If the crack is located common to the near surface, remove the probe and measure from the center of the coil to
the edge of the collar. Report this dimension as the length of the crack.
(g) If the crack is common to an interface or the far edge, tighten the set screw on the collar and make a mark on the
probe shaft with a sharp pencil at the end of the collar.
(h) Loosen the set screw and move the probe towards the interface or far edge, once the interface or edge is
determined, tighten the set screw and make another mark on the probe shaft.
NOTE: The interface or far edge can be determined by rotating the probe so the coil is not
influenced by the crack and nulling the instrument. Slowly push the probe in towards the
interface or edge. Both will give an indication very similar to a crack. Use the same
method as finding the end of the crack to locate the edge of the interface or far edge of
the part.
(i) Measure the distance between the center of the two marks. Report this dimension as the length of the crack.
C. Semiautomatic Probe.
(1) Characterization to determine relevance of a crack indication should always be done with the display in the flying dot
or phase analysis mode.
(2) Maximize the signal from the indication.
(3) With the display set to the flying dot mode, observe the signal characteristics of the indication.
(a) A crack will follow the same phase direction as a crack would in a surface inspection for the material being
inspected, up and to the left for aluminum and vertical for steel. Semiautomatic inspection indications will have
both a positive and negative component although they may not be equal.
(b) Corrosion will have a response that is very random and will have vertical and horizontal components to the
indication. The signal will have the look of a birds nest on the display in the flying dot mode and on the sweep
mode will most likely have multiple indications along the sweep base line similar to a noisy hole.
(4) Determining the length of a crack.
(a) Set the display on the scope to Sweep mode.
(b) With scanner turned on, insert the probe into hole with crack indication.
(c) Peak the signal from the crack indication.
(d) If the peak amplitude from the indication is off screen, reduce the gain until the whole indication is visible in the
display.
(e) Move the probe in the hole until the signal from the crack is reduced by 50%.
(f) Hold the scanner steady while maintaining the signal from the crack at 50%.
(g) Using a sharp pencil or other fine point marker, lay it on the surface of the part so that the point is at the same
level as the surface of the part.
(h) Slowly move the pencil towards the shaft of the probe as it's rotating and make a mark on the probe shaft even
with the surface of the part. Refer to Figure 15.
(i) If the crack is located common to the near surface, remove the probe and measure from the center of the coil to
the center of the mark on the probe shaft. Report this dimension as the length of the crack.
(j) If the crack is common to an interface or the far edge, continue moving the probe in the hole until the interface or
edge is defined or by a reduction in the signal amplitude of the crack indication of 50%.
(k) Hold the scanner and probe steady in the hole and make another mark on the probe shaft. Refer to Figure 15.
(l) Remove the probe from the hole and measure the distance from the center of the two marks. Report this
dimension as the length of the crack.
12. Reporting Results.
A. If part disposition has not been specified, contact Citation Customer Service for disposition. Refer to Introduction for
contact information. Make sure to provide the following information:
(1) Location of the hole with observed discontinuity.
(2) Length of crack in the bore of the hole.
(3) If applicable, material layer or layers that are affected.
(4) Discontinuity orientation relative to hole. The 12 O'clock position is typically referred to as Up in a hole that is oriented
forward to aft or left to right and Forward in a hole that is oriented up and down.
ORIENT COLLAR
THUMB SCREW
0.030 INCH WITH COIL.
(0.762 mm)
DETAILA
0.030 INCH
(0.762 mm)
MANUAL PROBE CALIBRATION
PENCIL
MINIMUM
SENSITIVITY
LEVEL
MINIMUM
SENSITIVITY
LEVEL IS 3
DIVISIONS F ROM
20° PEAK TO PEAK
LIFT-OFF
A16317
(4) Surface Curvature: On concave surfaces, a curvature radius of 10 inches (250 mm) or more is required. On convex
surfaces, a curvature radius of 3 inches (8 mm) or larger is required.
(a) A correction factor must be applied for smaller radii.
(5) Probe Contact Area: The surface of the part to which the conductivity measuring probe is applied must be at least as
large as the outside diameter of the probe. If the part dimension is very nearly equal to the probe diameter, the probe
must be centered on the part.
6. Instrument Calibration
A. Calibration Reference Standards
(1) Each instrument must have a minimum of two aluminum alloy instrument conductivity standards which must always be
kept with the conductivity instrument. One standard must be in the range of 25% to 32% IACS and one in the range of
38% to 62% IACS. There must be a minimum spread of 10% IACS between the standard for the low range and the
standard for the high range. The instrument conductivity standards must be certified to be accurate within 0.85%
IACS by the comparison method to the laboratory conductivity standards in accordance with ASTM B193 procedure
in a system per ISO 10012-1 ANSI/NCSL Z540-1. Calibration must be made initially and at intervals such that they
coincide with the conductivity instrument recalibration cycle.
B. The instrument must be calibrated and operated as specified in this technique and the manufacturer's instructions.
C. Each time the conductivity instrument is used, it must be calibrated (standardized) using the instrument conductivity
standards before checks are made and rechecked at 15 minute intervals during continuous operation. If the instrument is
found to be out of calibration, all measurements done since the last calibration must be rechecked.
7. Instrument Calibration Standards
A. Nonferrous reference standards must be of an alloy that has the same major base metal, basic temper and the
approximate electrical conductivity of the material to be inspected.
B. Reference standards must have a minimum surface finish of 150 RHR or RMS 165.
8. Inspection
A. The purpose of the inspection is to collect information to let the responsible engineering group determine the material
properties in the affected area.
NOTE: Because variations in material properties, material stacking, and geometry cause changes of
conductivity readings, conductivity readings alone must not be used to determine if the affected
area is accepted. Reference must be made to the applicable material specifications and
engineering drawings.
B. Visual
(1) Clean the area to be examined with methods specified in the appropriate maintenance manual. Remove all dirt, grit,
soot, and other debris that will not let the probe maintain good contact with the structure.
(2) Examine the area visually for indications of possible heat damage. Some signs include paint/metal discoloration and
bubbled or peeled paint.
(3) Record the location and description of the heat damaged area. This description will be used along with the
conductivity readings to determine the type of repair. If photographs are used to describe the area, take the picture
before the conductivity test is done.
C. Eddy Current Conductivity
(1) Use the appropriate reference standards to make sure of proper instrument calibration within the general conductivity
range of aluminum structures .
(2) After the visual inspection, set a reference point. If there is visual evidence of possible heat damage, set the
reference point at the center of the area that appears to have been the most affected. If there is no visual evidence of
possible heat damage, set the reference point at the center of the area to be inspected. In either case, the reference
point will locate approximately the center of the area of interest.
NOTE: A detailed map will need to be made of the inspection area. It must include dimensions to
locate the reference point and enough information to let the engineering personnel that do the
review of the information locate the places the conductivity readings were taken.
(3) The total area to be examined, and the distance between readings will depend on the specific situation. It is
recommended that the distance between conductivity readings not exceed 1.0 inch (25 mm). If the visual evidence or
the conductivity readings suggest rapid changes in severity, decrease the distance between test points. It is
recommended that the total area to be examined exceed the area of visual evidence by at least 2.0 inches (50 mm). If
the conductivity readings continue to change, the area of examination should be expanded until readings remain fairly
constant to make sure a complete map of the area is prepared.
(4) Locate the reference point at the corner of a square, refer to Figure 1. Take conductivity readings as you work away
from the reference point in the increments and distance determined in the previous step. Enough information must be
recorded along with the conductivity reading to allow a person not familiar with the inspection to locate the data point.
NOTE: Structural considerations may make it necessary that the test points do not follow the pattern
of Figure 1. It is up to the inspector to set a pattern that best covers the area to be inspected.
A16319
(6) Move the probe over the notch. Monitor the display to make sure the signal moves up and toward the left as the probe
moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
B. Calibrate the 1 MHz Surface Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 1 MHz.
(3) Put the probe on the face of the steel standard away from the notch.
(4) Push the null (balance) button.
(5) Lift the probe away from the standard. Monitor the display.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(6) Move the probe over the notch. Monitor the display to make sure the signal moves up and toward the left as the probe
moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
C. Calibrate the 200 kHz Bolt Hole Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Put the probe in the 1/2 inch hole in the standard with the probe coil away from the notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Move the probe coil across the notch. Monitor the display to make sure the signal moves up and toward the left as the
probe moves across the notch.
(a) Adjust the instrument to get three major screen divisions between the null point and the notch point.
6. Inspection
A. Top and Bottom Outboard Mounts Surface Inspection (Refer to Figure 1 and Figure 2).
(1) Calibrate the 200 kHz surface probe and instrument. Refer to Calibrate the 200 kHz Surface Probe.
(2) Use the 200 kHz surface probe to examine all of the countersunk area of each of the top and bottom outboard
mounts.
B. Airplanes 20800330 and On Bottom Mount Bolt Hole Inspection
(1) Calibrate the bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.62 inch (15.75 mm), measurement is from the edge of the collar to the center of the coil.
(3) Put the inspection plate over one of the bottom mount holes.
NOTE: The plate will rest on the heads of the fasteners.
(a) Put the probe in the hole until the collar is against the inspection plate.
(b) Turn the probe a minimum of 360 degrees while you monitor the display for cracks.
(4) Adjust the collar of the probe so that it is at 0.74 inch (18.80 mm) away from the center of the coil.
(a) Put the probe in the hole until the collar is against the inspection plate.
(b) Turn the probe a minimum of 360 degrees while you monitor the display for cracks.
(5) Adjust the collar of the probe so that it is at 0.85 inch (21.59 mm) away from the center of the coil.
(a) Put the probe in the hole until the collar is against the inspection plate.
(b) Turn the probe a minimum of 360 degrees while you monitor the display for cracks.
(6) Do the inspection on the other bottom hole.
C. Airplanes 20800001 thru 20800329 and 208B Bottom Mount Bolt Hole Inspection
NOTE: Airplanes 20800001 thru 20800329 with a lower reinforcement assembly installed, use the
UPPER CENTER
ENGINE MOUNT
UPPER RIGHT
ENGINE MOUNT UPPER LEFT
ENGINE MOUNT
BOLT HOLE
COUNTERSINK INSPECTION
INSPECTION AREA
AREA
DETAIL A 2613T3007A
A2613T1016
A2683T1001
NOTE: Inspections must be done on both the left and right side of the airplane.
A. Forward Attach Fitting Bolt Hole Inspection (Refer to Figure 1).
(1) Calibrate the 7/8 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.050 inch (1.27 mm). Measure from the collar to the center of the coil.
(3) Put the probe in the hole from the forward side of the lugs.
(4) Turn the probe a minimum of 360 degrees while you monitor the display.
(5) Do the inspection at each of the probe depths in Table 1.
Table 2. Forward Lugs
Index Number Depth (Inches) Depth (mm)
1 0.050 1.27
2 0.150 3.81
3 0.250 6.35
4 0.862 21.89
5 0.962 24.43
6 1.062 26.97
(6) Do the inspection from the aft side of the lugs.
(7) Do the inspection for each of the attach fittings.
B. Aft Attach Fitting Bolt Hole Inspection (Refer to Figure 1).
(1) Calibrate the 15/16 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.050 inch (1.27 mm). Measure from the collar to the center of the coil.
(3) Put the probe in the hole from the forward side of the lugs.
(4) Turn the probe a minimum of 360 degrees while you monitor the display.
(5) Do the inspection at each of the probe depths in Table 1.
Table 3. Aft Lugs
Index Number Depth (Inches) Depth (mm)
1 0.050 1.27
2 0.140 3.56
3 0.230 5.84
4 0.778 19.76
5 0.868 22.05
6 0.958 24.33
(6) Do the inspection from the aft side of the lugs.
(7) Do the inspection for each of the attach fittings.
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
(6) Move the probe over the notch. Monitor the display to make sure the signal moves up and toward the left as the probe
moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
B. Calibrate the 200 kHz Bolt Hole Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Put the probe in the appropriate hole in the standard with the probe coil away from the notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Move the probe coil across the notch. Monitor the display to make sure the signal moves up and toward the left as the
probe moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
6. Inspection
NOTE: Inspections must be done on both the left and right side of the airplane.
A. Stiffener and Top Bulkhead Surface Inspection (Refer to Figure 1).
(1) Calibrate the 200 kHz surface probe and instrument. Refer to Calibrate the 200 kHz Surface Probe.
(2) Remove one bolt from the structure at a time.
(a) Remove nut from aft side.
(b) Remove the bolt.
NOTE: If there is not enough clearance to remove the bolt, it may be necessary to cut the head of
the bolt off.
1 Remove the bolt far enough to cut off the head.
2 Remove the shank of the bolt from the aft side.
(3) Examine around the circumference of the hole.
(4) Do a bolt hole inspection on the hole from the aft side. Refer to the appropriate Attach Fitting Bolt Hole Inspection.
(5) Install a new bolt from the aft side of the hole.
(6) Use the previous steps to examine each hole shown in Figure 1.
B. Attach Fitting Bolt Hole Inspection for Holes 4, 5, and 6
NOTE: Bolts 4, 5 and 6 may be removed at same time for Model 208B only.
(1) Calibrate the 1/4 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.201 inch measuring from the collar to the center of the coil.
(3) Put the probe in the hole.
(4) Turn the probe a minimum of 360 degrees while you monitor the display.
(5) Do the inspection at each of the probe depths of Table 1 for holes 4, 5, and 6.
Table 2. Holes 4, 5, and 6
Index Number Depth (Inches) Depth (mm)
1 0.201 5.11
2 0.321 8.15
3 0.441 11.20
4 0.561 14.25
5 0.801 20.35
6 0.921 23.39
7 1.041 26.44
8 1.061 26.95
9 1.226 31.14
10 1.329 33.76
11 1.449 36.80
12 1.569 39.85
13 1.689 42.90
14 1.809 45.95
15 1.929 49.00
16 2.049 52.04
17 2.169 55.09
18 2.289 58.14
19 2.354 59.79
C. Attach Fitting Bolt Hole Inspection for Holes 1, 2, and 3. (Refer to Figure 1).
(1) Calibrate the 3/16 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.211 inch (5.36 mm). Measure from the collar to the center of the coil.
(3) Put the probe in the hole.
(4) Turn the probe a minimum of 360 degrees while you monitor the display.
(5) Do the inspection for holes 1, 2, and 3.
NOTE: Inspection for holes 1, 2, and 3 must be done from the forward and aft sides of the bulkhead.
D. Attach Fitting Inspection for Holes 7, 8, and 9. (Refer to Figure 1).
(1) Calibrate the flexible surface probe and instrument. Refer to Calibrate the 200 kHz Surface Probe.
(2) Do a surface probe around the nuts of holes 7, 8, and 9.
(a) Insert the flexible surface probe through a lightning hole to do the inspection.
NOTE: Inspection for holes 7, 8, and 9 must be done on the forward and aft sides of the
bulkhead.
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
(a) Keep the probe as near to the edge of the hole as possible but try to avoid any edge effect.
(3) Examine around the circumference of the main landing gear forging slot. (Refer to Figure 3).
(a) Keep the probe as near to the edge of the hole as possible, but try to avoid any edge effect.
(4) Examine the inner radii around the circumference of the main landing gear forging slot. (Refer to Figure 4).
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Give the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
B. Calibrate the 200 kHz Bolt Hole Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Put the probe in the appropriate hole in the standard with the probe coil away from the notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Move the probe coil across the notch. Monitor the display to make sure the signal moves up and toward the left as the
probe moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
6. Inspection
NOTE: Inspections on the bottom fittings must be done on both the left and right and forward and aft sides of
the bulkhead.
A. Bottom Bulkhead Surface Inspection (Refer to Figure 1).
(1) Calibrate the 200 kHz surface probe and instrument. Refer to Calibrate the 200 kHz Surface Probe.
(2) Remove one bolt.
(3) Examine around the circumference of the hole.
(4) Do a bolt hole inspection on the hole. Use the procedure that applies to each hole.
(5) Install the bolt before removing the next bolt.
(6) Examine each hole using the previous steps.
B. Attach Fitting Bolt Hole Inspection for Holes 4, 5, and 6 (Refer to Figure 1).
NOTE: Holes 4, 5, and 6 go into the brace between the fittings.
(1) Calibrate the 1/4 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.180 inch (4.57 mm).
(3) Measure from the collar to the center of the coil.
(4) Put the probe in the hole.
NOTE: Due to lack of accessibility to perform the bolt hole eddy current inspection, it is permissible to
perform a surface eddy current inspection on any or all of these bolt holes. Refer to Bottom
Bulkhead Surface Inspection.
(5) Turn the probe a minimum of 360 degrees while you monitor the display.
(6) Install the bolt before removing the next bolt.
Index Number Depth (Inches) Depth (mm)
1 0.193 4.90
2 0.313 7.95
3 0.433 11.00
4 0.553 14.05
5 0.673 17.09
6 0.793 20.14
7 0.913 23.19
8 1.033 26.24
9 1.123 28.52
(7) Do the inspection at each of the depths for holes 4, 5, and 6, forward and aft, left and right..
C. Attach Fitting Bolt Hole Inspection for Holes 1, 2, and 3 (Refer to Figure 1).
(1) Calibrate the 3/16 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.203 inch (5.16 mm). Measure from the collar to the center of the coil.
(3) Put the probe in the hole.
(4) Turn the probe a minimum of 360 degrees while you monitor the display.
(5) Do the inspection for holes 1, 2, and 3.
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
B. Calibrate the 200 kHz Bolt Hole Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Put the probe in the appropriate hole in the standard with the probe coil away from the notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Move the probe coil across the notch. Monitor the display to make sure the signal moves up and toward the left as the
probe moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
6. Inspection
NOTE: Inspections must be done on both the left and right side of the airplane.
A. Top Bulkhead Surface Inspection (Refer to Figure 1).
(1) Calibrate the 200 kHz surface probe and instrument. Refer to Calibrate the 200 kHz Surface Probe.
(2) Remove one bolt.
(3) Examine around the circumference of the hole.
(4) Do a bolt hole inspection on the hole. Refer to the appropriate Attach Fitting Bolt Hole Inspection.
(5) Use the previous steps to examine each hole shown in Figure 1.
B. Forward Attach Fitting Bolt Hole Inspection For Bolt Holes 1, 2, 3 and 4 (Refer to Figure 1).
(1) Calibrate the 1/4 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.130 inch (3.30 mm). Measure from the collar to the center of the coil.
(3) Remove only one bolt.
(4) Put the probe in the hole from the aft side of the bulkhead.
(5) Turn the probe a minimum of 360 degrees while you monitor the display.
(6) Do the inspection at each of the probe depths of Table 2.
(7) Install the bolt.
(8) Remove the next bolt.
(9) Do the complete inspection of each of the bolt holes shown on the forward top carry-thru bulkhead.
Table 2. Forward Bolt Holes
Index Number Depth (Inches) Depth (mm)
1 0.130 3.30
2 0.210 5.33
3 0.290 7.37
4 0.370 9.40
5 0.450 11.43
6 0.530 13.46
7 0.610 15.49
8 0.690 17.53
9 0.770 19.56
10 0.850 21.59
11 0.930 23.62
12 1.010 25.65
13 1.090 27.69
14 1.155 29.34
15 1.255 31.88
16 1.335 33.91
17 1.415 35.94
18 1.495 37.91
19 1.575 40.01
20 1.655 42.04
21 1.735 44.07
22 1.815 46.10
23 1.895 48.13
24 1.975 50.17
25 2.055 52.20
26 2.135 54.23
27 2.215 56.26
28 2.280 57.91
C. Attach Fitting Bolt Hole Inspection for Holes 5A, 5B, 6A, 6B, 7A, 7B (Refer to Figure 1).
(1) Calibrate the adjustable bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.230 inch (5.84 mm). Measure from the collar to the center of the coil.
(3) Remove only one bolt.
(4) Put the probe in the hole.
(5) Turn the probe a minimum of 360 degrees while you monitor the display.
(6) Do the inspection for holes 5A, 5B, 6A, 6B, 7A, 7B.
D. Aft Attach Fitting Bolt Hole Inspection (Refer to Figure 1).
(1) Calibrate the 1/4 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.180 inch ( mm). Measure from the collar to the center of the coil.
(3) Remove only one bolt.
(4) Put the probe in the hole fron the forward side.
(5) Turn the probe a minimum of 360 degrees while you monitor the display.
(6) Do the inspection at each of the probe depths of Table 3.
(7) Install the bolt.
(8) Remove the next bolt.
(9) Do the complete inspection of each of the bolt holes shown on the aft top carry-thru bulkhead.
Table 3. Aft Bolt Holes
Index Number Depth (Inches) Depth (mm)
1 0.180 4.57
2 0.260 6.60
3 0.340 8.64
4 0.420 10.67
5 0.500 12.70
6 0.580 14.73
7 0.660 16.76
8 0.740 18.80
9 0.820 20.82
10 0.900 22.86
11 0.980 24.89
12 1.060 26.92
13 1.180 29.97
14 1.260 32.00
15 1.340 34.04
16 1.420 36.07
17 1.500 38.10
18 1.580 40.13
19 1.660 42.16
20 1.740 44.20
21 1.820 46.23
22 1.900 48.26
23 1.980 50.29
24 2.060 52.32
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
point.
6. Inspection
NOTE: Inspections must be done on both the left and right side of the airplane.
A. Attach Lug Bolt Hole Inspection. (Refer to Figure 1).
(1) Calibrate the 3/4 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.05 inch. Measure from the collar to the center of the coil.
(3) Put the probe in the hole from the forward side of the lugs.
(4) Turn the probe a minimum of 360 degrees while you monitor the display.
(5) Put the probe in the hole from the aft side of the lugs.
(6) Turn the probe a minimum of 360 degrees while you monitor the display.
(7) Do the inspection at each of the probe depths of Table 2 for each lug.
Table 2. Attach Lug
Index Number Depth (Inches) Depth (mm)
1 0.05 1.27
2 0.13 3.30
3 0.21 5.33
4 0.267 6.78
5 0.871 22.12
6 0.951 24.16
7 1.031 26.19
8 1.075 27.31
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(6) Move the probe over the notch. Monitor the display to make sure the signal moves up and toward the left as the probe
moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
B. Calibrate the 200 kHz Bolt Hole Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Put the probe in the 1/2 inch hole in the standard with the probe coil away from the notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Move the probe coil across the notch. Monitor the display to make sure the signal moves up and toward the left as the
probe moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
6. Inspection
NOTE: Inspections must be done on the left and right fittings.
A. Do the Forward Attach Fitting Surface Inspection (Refer to Figure 1).
(1) Calibrate the 200 kHz surface probe and instrument. Refer to Calibrate the 200 kHz Surface Probe.
(2) Use the 200 kHz surface probe to examine each of the countersink areas of the fittings.
B. Do an Inspection of the Bolt Hole for the Forward Attach Fitting (Refer to Figure 1 and Figure 2).
(1) Calibrate the 1/2 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe to a depth of 0.42 inch (10.67 mm).
(3) Measure from the collar to the center of the coil.
(4) Put the inspection plate on the probe.
(5) Put the probe in the hole until the collar is against the inspection plate.
(6) Turn the probe a minimum of 360 degrees while you monitor the display.
(7) Do the inspection at each of the depths in Table 1.
Table 2. Forward Attach Fitting
Index Number Depth (inches) Depth (mm)
1 0.42 10.67
2 0.52 13.21
3 0.62 15.75
4 0.72 18.29
C. Do an Inspection of the Bolt Hole for the Aft Attach Fitting (Refer to Figure 1)
(1) Calibrate the bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe to a depth of 0.165 inch (4.19 mm).
(3) Measure from the collar to the center of the coil.
(4) Put the probe in the hole.
(5) Turn the probe a minimum of 360 degrees while you monitor the display.
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
BOLT HOLE
INSPECTION AREA
SURFACE PROBE
INSPECT
VIEW A-A
CROSS SECTION OF FORWARD
FITTING AT THE ATTACH POINT
2683T1005
A2683T1001
VERTICAL STABILIZER
FORWARD BULKHEAD
2612T1008
LEFT TOP
ATTACH HOLE RIGHT TOP
ATTACH HOLE
RIGHT BOTTOM
LEFT BOTTOM ATTACH HOLE
ATTACH HOLE
AFT CANTED
BULKHEAD
2612T1007
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
B. Calibrate the 200 kHz Bolt Hole Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Put the probe in the appropriate hole in the standard with the probe coil away from the notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Move the probe coil across the notch. Monitor the display to make sure the signal moves up and toward the left as the
probe moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
6. Aft Attach Point Inspection
A. Do the Bolt Hole Inspection (Refer to Figure 1).
(1) Calibrate the bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Adjust the probe diameter to the diameter of the probe.
(a) Put a non-metallic material in the slot of the probe.
(3) Set the depth of the probe to 0.190 inch (4.83 mm) between the collar and the center of the coil.
(4) Put the probe into the hole and turn it 360 degrees while you watch the eddy current unit for crack signals.
(5) Do the inspection at each of the probe depths in Table 1.
(6) Do the inspection on each of the holes.
Table 2. Aft Attach Point
Index Depth (Inches) Depth (mm)
1 0.190 4.83
2 0.250 6.35
3 0.310 7.87
B. Do the Surface Inspection.
(1) Calibrate the 200 kHz surface probe and instrument. Refer to Calibrate the 200 kHz Surface Probe.
(2) Examine around the circumference of the attach holes.
7. Forward Attach Points Inspection
A. Do an Attach Mount Inspection (Refer to Figure 1).
(1) Calibrate the bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Adjust the probe diameter to the diameter of the hole with the use of nonmetallic material between the slot in the
probe.
(3) Set the probe to a depth of 0.050 inch (1.27 mm).
(4) Measure from the collar to the center of the coil.
(5) Turn the probe a minimum of 360 degrees while you monitor the display.
(6) Do the inspection at each of the probe depths in Table 2.
Table 3. Forward Attach Point
Index Depth (Inches) Depth (mm)
1 0.050 1.27
2 0.17 4.32
3 0.29 7.37
4 0.41 10.41
5 0.53 13.46
6 0.65 16.51
7 0.77 19.56
8 0.89 22.61
9 1.01 25.65
10 1.07 27.18
8. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
(1) Calibrate the ring probe and instrument. Refer to Calibrate the Ring Probe.
(2) Inspect each fastener on the top and bottom spar between RSS 9.90 and LSS 9.90.
B. Inspect the Aft Spar (Refer to Figure 1).
(1) Calibrate the ring probe and instrument. Refer to Calibrate the Ring Probe.
(2) Inspect each fastener on the top and bottom spar between RSS 10.60 and LSS 10.60.
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
(3) Put the probe in the applicable hole in the standard with the probe coil away from the notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Move the probe coil across the notch. Monitor the display to make sure the signal moves up and toward the left as the
probe moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
6. Forward Spar Fitting Inspection
A. Wing to Carry-Thru Forward Spar Attach Fitting Bolt Holes (Refer to Figure 1).
(1) Calibrate the 5/16 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Inspect attach fitting holes number 1, 2, 3, and 4.
NOTE: Figure 2 shows a typical hole cross section.
(3) Set the probe at a depth of 0.05 inch. Measure from the collar to the center of the coil.
(4) Put the probe in the hole.
(5) Turn the probe a minimum of 360 degrees while you monitor the display.
(6) Do the inspection at each of the probe depths of the applicable table for each hole.
Table 2. Hole 1
Index Structure Depth (inches) Depth (mm)
1 Fitting 0.05 1.27
2 Fitting 0.09 2.29
3 Fitting 0.13 3.30
4 Fitting 0.17 4.32
5 Angle 0.25 6.35
6 Spacer 0.32 8.13
7 Web 0.38 9.65
8 Spacer 0.44 11.18
9 Angle 0.51 12.95
10 Fitting 0.6 15.24
11 Fitting 0.64 16.26
12 Fitting 0.68 17.27
13 Fitting 0.72 18.29
Table 3. Hole 2
Index Structure Depth (inches) Depth (mm)
1 Fitting 0.05 1.27
2 Fitting 0.15 3.81
3 Fitting 0.29 7.37
4 Fitting 0.39 9.91
5 Angle 0.47 11.94
6 Spacer 0.54 13.72
7 Web 0.60 15.24
8 Spacer 0.66 16.76
9 Angle 0.73 18.54
(5) Turn the probe a minimum of 360 degrees while you monitor the display.
(6) Do the inspection at each of the probe depths of the applicable table for each hole.
Table 6. Hole 5
Index Structure Depth (inches) Depth (mm)
1 Fitting 0.05 1.27
2 Fitting 0.10 2.54
3 Fitting 0.15 3.81
4 Angle 0.23 5.84
5 Spacer 0.28 7.11
6 Spacer 0.38 9.65
7 Angle 0.43 10.92
8 Fitting 0.51 12.95
9 Fitting 0.56 14.22
10 Fitting 0.61 15.49
Table 7. Hole 6
Index Structure Depth (inches) Depth (mm)
1 Fitting 0.05 1.27
2 Fitting 0.18 4.57
3 Fitting 0.31 7.87
4 Angle 0.39 9.91
5 Spacer 0.44 11.18
6 Spacer 0.54 13.72
7 Angle 0.59 14.99
8 Fitting 0.67 17.02
9 Fitting 0.80 20.32
10 Fitting 0.92 23.37
Table 8. Hole 7
Index Structure Depth (inches) Depth (mm)
1 Fitting 0.05 1.27
2 Fitting 0.10 2.54
3 Fitting 0.15 3.81
4 Spacer 0.22 5.59
5 Cap 0.28 7.11
6 Cap 0.39 9.91
7 Spacer 0.45 11.43
8 Fitting 0.52 13.21
9 Fitting 0.57 14.48
10 Fitting 0.62 15.75
Table 9. Hole 8
Index Structure Depth (inches) Depth (mm)
1 Fitting 0.05 1.27
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
Figure 2 : Sheet 1 : Typical Hole Cross Section at Forward Spar Fitting Hole Locations
A29132
WEB
(0.040 INCH THICK) WEB
SPACER SPACER (0.040 INCH THICK)
(0.080 INCH THICK) ANGLE (0.080 INCH THICK)
(0.063 INCH THICK) ANGLE
ANGLE (0.063 INCH THICK)
(0.063 INCH THICK) ANGLE
(0.063 INCH THICK)
FITTING FITTING
(0.220 INCH THICK) (0.440 INCH THICK)
WEB
(0.040 INCH THICK) WEB
CAP CAP (0.040 INCH THICK)
(0.100 INCH THICK) (0.100 INCH THICK)
SPACER
(0.040 INCH THICK) SPACER
SPACER (0.040 INCH THICK)
(0.040 INCH THICK) SPACER
(0.040 INCH THICK)
FITTING FITTING
(0.220 INCH THICK) (0.440 INCH THICK)
2683T1003
Figure 3 : Sheet 1 : Typical Hole Cross Section at Aft Spar Fitting Hole Locations
A29133
WEB WEB
SPACER (0.032 INCH THICK) SPACER (0.032 INCH THICK)
(0.063 INCH THICK) (0.063 INCH THICK)
ANGLE ANGLE
ANGLE (0.050 INCH THICK) ANGLE
(0.050 INCH THICK) (0.050 INCH THICK) (0.050 INCH THICK)
FITTING FITTING
(0.200 INCH THICK) (0.360 INCH THICK)
HOLE 5 MATERIAL CROSS SECTION HOLE 6 MATERIAL CROSS SECTION
WEB WEB
CAP (0.032 INCH THICK) CAP (0.032 INCH THICK)
(0.080 INCH THICK) (0.080 INCH THICK)
SPACER SPACER SPACER
SPACER (0.040 INCH THICK) (0.040 INCH THICK)
(0.040 INCH THICK) (0.040 INCH THICK)
FITTING FITTING
(0.200 INCH THICK) (0.360 INCH THICK)
HOLE 7 MATERIAL CROSS SECTION HOLE 8 MATERIAL CROSS SECTION
2683T1004
B. Use the eddy current spot probe to inspect around each fastener common to the bottom spar and skin from the inboard
end of the wing to the rib at WS 126.50.
(1) Make sure the instrument and probe are calibrated. Refer to Calibrate the Spot Probe.
(2) The spar cap has two angles bonded to a center web.
(3) Do a calibration and inspection parallel to the forward and aft edges of each cap with a nonconductive straight edge
as a guide for the probe.
NOTE: If greater edge effect is sensed, use the Alternate Calibration for the Spot Probe to recalibrate
the equipment.
(4) Do a calibration and inspection with the spot probe around the remaining area of each of the fasteners from the
inboard end of the wing spar to WS 126.50.
7. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Provide the following information:
(1) Crack location.
(2) Crack length.
(3) Crack orientation.
FLAP TRACKS
1. General
A. Included in this section are instructions to do an eddy current inspection for fatigue cracks in the flap tracks.
B. This inspection satisfies requirements of Maintenance Manual, Chapter 5, (Refer to Table 1).
Table 1. Item Codes
AIRPLANE ITEM CODE TASK NUMBER
MODEL 208 A575002 57-10-00-254
MODEL 208 A575003 57-10-00-255
2. Personnel Requirements
A. Personnel certification requirements are found in Part 1of this manual.
3. Tools, Equipment and Materials
A. The equipment used in the development of this procedure is listed in the Table below.
NOTE: Equivalent eddy current equipment may be used provided it meets the sensitivity requirements of
this procedure.
NAME NUMBER MANUFACTURER USE
Eddy Current Instrument Nortec 2000 Olympus NDT To inspect flap tracks
48 Woerd Ave
Waltham, MA 02453
Web: http://www.olympus-imsg.com
Calibration Standard SRS-0824A Olympus NDT To calibrate equipment
NOTE: 1
Right Angle Surface MP902-60 Olympus NDT To inspect flap tracks
Probe
NOTE: 2
NOTE: 1. An equivalent calibration standard will be an aluminum alloy with a 0.020 inch deep or less,
surface EDM notch. Refer to Part 6, 20-00-02 for additional guidance.
2. Coil: 200 kHz, 1/8 inch diameter, shielded (ferrite core). Probe body: right angle, 0.2 inch drop, 6
inches long. Other lengths/styles of probes may be used provided the coil requirements are met.
4. Calibration
A. Calibrate the 200 kHz Surface Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Put the probe on the face of the standard away from the notch.
(4) Push the null (balance) button.
(5) Lift the probe away from the standard. Monitor the display.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(6) Move the probe over the 0.020 inch deep EDM notch. Monitor the display to make sure the signal moves up and
toward the left as the probe moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the peak notch
indication.
5. Flap Track Inspection
A. Do the Eddy Current Inspection of the Center, Inboard, and Outboard Flap Tracks (Refer to Figure 1).
(1) Lower the flaps to the fully extended 30 degree position.
(2) Disconnect all electrical power from the airplane. Attach maintenance warning tags to the battery connector(s) and
external power receptacle that have the following written on them:
RIGHT ANGLE
SURFACE PROBE
WING
DETAIL B
TYPICAL INTERNAL SURFACE
FLAP TRACK INSPECTION
RIGHT ANGLE
SURFACE PROBE
DETAIL C
TYPICAL UPPER AND LOWER EXTERNAL
SURFACE FLAP TRACK INSPECTION
B2622T1027
C2622T1029
RIGHT ANGLE
SURFACE PROBE
DETAIL D
TYPICAL INBOARD AND OUTBOARD
SURFACE FLAP TRACK INSPECTION
D2622T1028
(3) Put the probe on the face of the standard away from the notch.
(4) Push the null (balance) button.
(5) Lift the probe away from the standard. Monitor the display.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(6) Move the probe over the notch. Monitor the display to make sure the signal moves up and toward the left as the probe
moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
B. B. Calibrate the 200 kHz bolt hole probe
NOTE: This calibration is for either of the specified bolt hole probes.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Insert the probe into the appropriate hole of the reference standard.
NOTE: Make sure that the probe coil is positioned away from the EDM notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Rotate the probe coil across the notch.
NOTE: Monitor the display to make sure the signal moves up and toward the left as the probe moves
across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
6. Inspection Procedure
NOTE: Inspection procedures are typical for all strut fittings.
A. Do the inspection on the necessary areas of the attach fittings for cracks (Refer to Figure 1).
(1) Clean the areas to be inspected.
(2) Calibrate the 200 kHz surface probe and instrument. Refer to Calibrate the 200 kHz Surface Probe.
(3) Do the inspection on the surface of the fitting.
NOTE: The areas to be inspected are the edge of the fitting and the top of the lugs at the machined
step. Refer to Figure 1.
(4) Carefully do the inspection in the opposite direction of probe scanning to make sure of the crack indication.
(5) Do the inspection on the fittings (total of 8). Do the probe scanning in the direction of the arrows around first three
bolts as shown in Figure 1, Sheets 2 and 3, Details B and C
B. Bolt Hole Probe Inspection - Upper Fitting Attach Holes
NOTE: The upper fitting attach holes have a diameter of 0.875 inches (22.225 mm) so you must use the
BPMF-56 bolt hole probe. Each lug has a width of 0.70 inches (17.78 mm).
(1) Calibrate the BPMF-56 bolt hole probe. Refer to Calibrate the 200 kHz bolt hole probe.
(2) Position the collar of the bolt hole probe so it is 0.050 inch (1.270 mm) from the center of the probe coil.
(3) Insert the probe into an upper attach hole.
NOTE: It may be necessary to balance the instrument.
(4) Adjust lift-off so it moves horizontal and to the left.
(5) Monitor the instrument display while you rotate the probe around the entire circumference of the hole so it goes
beyond 360 degrees.
(6) Reposition the collar 0.10 inch (2.54 mm) from its position (0.15 inch (3.81 mm) from the coil center).
NOTE: Do the inspection again.
(7) Repeat the step above five more times.
(a) Reposition the collar 0.10” after each index until the lug as been inspected.
A
C
A A
TIEDOWN
(REFERENCE) B
VIEWA−A
MS21042L5
NUT
(REFERENCE)
B
OUTBD
DETAILA FWD
LOOKING UP AT LEFT
WING STRUT ASSEMBLY
UP (RIGHT SIDE OPPOSITE)
VIEWB−B
FWD
2610T7005
A2622T1022
AA2622T1024
BB2622T1023
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
B. Calibrate the 200 kHz Bolt Hole Probe.
(1) Connect the probe to the instrument.
(2) Set the eddy current instrument frequency to 200 kHz.
(3) Put the probe in the 7/8 inch hole in the standard with the probe coil away from the notch.
(4) Push the null (balance) button.
(a) Adjust the instrument so that the lift-off signal goes horizontal and to the left.
(5) Move the probe coil across the notch. Monitor the display to make sure the signal moves up and toward the left as the
probe moves across the notch.
(a) Adjust the instrument to get a minimum of three major screen divisions between the null point and the notch
point.
5. Inspection
A. Support Fitting Surface Inspection (Refer to Figure 1).
NOTE: This inspection is not necessary for Airplanes 20800001 thru 20800010 and Airplane 20800012.
NOTE: This inspection is necessary for both wings.
(1) Calibrate the surface probe. Refer to Calibrate the 200 kHz Surface Probe
(2) Holes 1 thru 4.
(a) Remove the nut from each bolt one at a time.
(b) Pull the bolt out of the hole far enough to allow you to do a surface inspection around the bolt shank.
(c) Do a surface scan inspection around the bolt on the fitting.
(d) Do a surface scan inspection around the bolt hole on the opposite fitting.
(e) Install the nut and torque to standard torque.
(f) Do the inspection at each of the holes.
(3) Holes 5 thru 7.
(a) Do a surface scan inspection around each of the fastener heads and collars. Do not remove collars.
B. Attach Lug Inspection (Refer to Figure 1).
NOTE: This inspection is for all airplanes.
(1) Calibrate the 7/8 inch bolt hole probe and instrument. Refer to Calibrate the 200 kHz Bolt Hole Probe.
(2) Set the probe at a depth of 0.05 inch (1.27 mm). Measure from the collar to the center of the coil.
(3) Put the probe in the hole.
(4) Turn the probe a minimum of 360 degrees while you monitor the display.
(5) Do the inspection at each of the depths in Table 1 for each of the lug holes.
Table 2. Attach Lugs
Index Number Depth (inches) Depth (mm)
1 0.05 2.03
2 0.17 4.32
3 0.29 7.37
4 0.41 10.14
5 0.53 13.46
6 0.65 16.51
6. Part Disposition
A. If a crack is found, contact Cessna Propeller Aircraft Product Support, P.O. Box 7706, Wichita, KS 67277 USA.
Telephone 316-517-5800. FAX 316-942-9006. Give the information that follows: