Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Imaging Bio Molecules

Download as pdf or txt
Download as pdf or txt
You are on page 1of 48

Imaging of

Biomolecules
Bright/Dark Field
Phase Contrast Transmitted Light
Polarization
Differential Interference Contrast (DIC)

Fluorescence (epi, confocal, TIRF, super-resolution etc)

Friday 27 October 17
Microscopes are designed with 3 essential functions

Magnification (to make structural detail visible to the detector)

Resolution (limited by wavelength, not lens aberrations)

Contrast (most biological specimens are transparent at vis wavelengths)

Absorption based contrast:


- color dyes: histology
- epi-fluorescence (modern day fluo. probes)
Refractive Index Differences based contrast:
- in specimen structural details from background (media,
cytosol)

Friday 27 October 17
Specimens with a higher Refractive Index than the
background produce retardation of light

Friday 27 October 17
Retardations in Transparent Specimens Scatter
(Diffract) Illuminatng Light

Bright Field Microscopy

Friday 27 October 17
Dark Field Microscopy:
Objective Collects Only Scattered Light; Rejects
Illumination Light
Angle of the illuminating cone
of light (high numerical
aperture) is bigger than the
angle of the scattered cone of
light collected by the
objective (low numerical
aperture).

- Cardioid condenser

Friday 27 October 17
Friday 27 October 17
Dark Field Microscopy:
Objective Collects Only Scattered Light; Rejects
Illumination Light
Advantages:
- High sensitivity - black background provides high contrast to see scattered
light from small objects.
- excellent for low magnification outlines of individual cells (such as flagella).
- It can be effectively used at high magnifications to image living bacteria, or at
low magnifications to view and image cells, tissues, and whole mounts.

Disadvantages:
- Resolution limited by need for NAcondenser > NAobjective
- Scattered light in thick specimens lowers contrast of fine structural detail
- Poor depth of field
- Images of internal cellular structure often inaccurate coz of the missing
interference with the undiffracted light (loss of signal)

Friday 27 October 17
Phase Contrast Microscopy:
Objective Collects Only Scattered Light; Rejects
Illumination Light

Friday 27 October 17
Phase Contrast Microscopy:
Lateral shifts in phase of wavefronts passing through
sample and passing undiffracted
- Positioned within the objective rear focal
plane is a phase ring or plate that not only
attenuates the bright, direct light originating
from the phase stop in the condenser, but also
adds a constant phase shift to this light.

- If the specimen contains sub-structures that


have mixed refractive indices, these entities
guide the light from the direct waves into new
paths (red dotted line).

- Wavefronts diffracted by the specimen (in


effect, those containing structural
information) will not pass through the phase
ring in the objective, meaning that they will
not be attenuated or retarded.

- All of the wavefronts are ultimately


recombined to form the intermediate image by
the tube lens.

Friday 27 October 17
Polarization Contrast Microscopy:
Optical anisotropy based capture of scattered
wavefronts from the specimen
The polarizer beneath the condenser (near the
aperture diaphragm) ensures that the specimen is
illuminated with linearly polarized wavefronts
that pass through the condenser.

The analyzer (a second polarizer), which is


oriented at an angle of 90 degrees to that of the
polarizer, is located behind the objective.

If no specimen is placed on the microscope


stage, the scene observed through the eyepieces
will remain completely dark.

When illuminated, many specimens turn the


vibration direction of the polarized light out of
the plane produced by the polarizer and
eventually the diverted planes are allowed to pass
through the crossed polarizer (analyzer), showing
up as images.

Friday 27 October 17
Phase Polarized
- difference of n of
- difference of n
specimen in different
between specimen
directions
and background
- optical anisotropy

Friday 27 October 17
Differential Interference Contrast Microscopy:
A polarizing microscope with condensor and two
DIC prisms

Friday 27 October 17
Differential Interference Contrast Microscopy:
A polarizing microscope with condensor and two
DIC prisms

Friday 27 October 17
Differential Interference Contrast Microscopy:
A polarizing microscope with condensor and two
DIC prisms

Friday 27 October 17
DIC

Ph Con

Friday 27 October 17
Pros Cons

Friday 27 October 17
Fluorescence

Friday 27 October 17
Friday 27 October 17
Framework of Fluorescence
Microscopy

Friday 27 October 17
Friday 27 October 17
Excitation Filter

- The excitation filter is the


optical element that passes only
the wavelength of light necessary
for excitation from the excitation
light source (usually a mercury
lamp) to the fluorophore.
- As shown in Figure 2, only the
excitation wavelength passes
through the filter, usually a "band
pass filter".

Friday 27 October 17
Dichroic Mirror

- The dichroic mirror is the optical


element that separates the excitation
light from the fluorescence. Dichroic
mirrors are special mirrors that reflect
only a specific wavelength of light,
allowing all other wavelengths to pass
through.

- Dichroic mirrors used in epi-


fluorescence microscope filter blocks
are placed in a 45° incidence angle to
light, creating a "stop band" of
reflected light and a "pass band" of
transmitted light. Light passing
through the excitation filter is reflected
90° toward the objective and the
specimen

Friday 27 October 17
Emission Filter

Barrier filters are optical elements that


separate fluorescence emanating from
the fluorophore from other background
light. As shown in Figure 4, the barrier
filter transmits light of the fluorescence
wavelength which passes through the
dichroic mirror while blocking all other
light leaking from the excitation lamp
(reflected from the specimen or optical
elements)

This is necessary because the strength


of the fluorescent light is weaker than
the excitation light by a factor of
more than 100,000:1.

Friday 27 October 17
Friday 27 October 17
Organic dyes

Classics

Coumarin Fluorescein Rhodamine

332/456 490/520 550/575

Systems of conjugated bonds that share electrons


Larger systems >> longer wavelength
Friday 27 October 17
Mitotracker DAPI
Oxidized in Stains nucleic acid
mitochondria to a
fluorescent compound

Friday 27 October 17
Fluorescent Proteins

GFP

O Shimomura
DC Prasher
M Chalfie
R Tsien
Friday 27 October 17
Fluorescence

Epi Confocal

Friday 27 October 17
Fundamental problem with conventional epifluorescence
is that fluorescence is emitted along the entire
illumination cone in the sample and not just at focus

Friday 27 October 17
Fluorescence Illumination of a single point

Camera

Tube lens
Excitation light

Emission light

Objective lens

Sample

Widefield Illumination Point Illumination

Friday 27 October 17
The confocal microscope
Scan excitation spot point-
Detector
by-point to build up image

Pinhole

Tube lens
Excitation light

Emission light

Objective lens

Sample

The pinhole is conjugate to the focal point of the lens


Friday 27 October 17
Light sources

Excitation light must be focused to a


diffraction limited spot
Excitation light
Could be done with an arc lamp
and pinhole – but very inefficient

Enter the laser:


Perfectly collimated and Objective lens

high power
Sample

Friday 27 October 17
Scanning

Changing entrance
angle of illumination
moves illumination spot
on sample
Objective lens

The emission spot


Sample
moves, so we have to
make sure pinhole is
coincident with it

Friday 27 October 17
Confocal optical path

Dichroic Pinhole
Scanning mirrors

x
Detector
y

Laser
Relay lens

Objective

there never is a complete image of the sample -- at any given


instant, only one point of the sample is observed. The
detector is attached to a computer which builds up the image,
one pixel at a time. In practice, this can be done perhaps 3
times a second, for a 512x512 pixel image. The limitation is
in the scanning mirrors.

Friday 27 October 17
Downside - Slow

The confocal microscope


Scan excitation spot point-
Detector
by-point to build up image

Pinhole Problems:
Slow (~1 sec to acquire an
image)
Tube lens Low light efficiency (due to
Excitation light use of PMT as detector)

Solution:
Emission light
Use multiple pinholes and
a camera

Objective lens

Sample

Friday 27 October 17
Friday 27 October 17
Friday 27 October 17
Optical Sectioning and 3D reconstruction

x y

Friday 27 October 17
Optical Sectioning and 3D reconstruction

x y

Friday 27 October 17
Optical Sectioning and 3D reconstruction

x y

Friday 27 October 17
Friday 27 October 17
Some state-of-the-art variations of
fluorescence microscopy

Friday 27 October 17
Total Internal Reflection Fluorescence (TIRF microscopy) -
for problems at the membrane interfaces

Friday 27 October 17
Breaking the diffraction barrier

Friday 27 October 17
Breaking the diffraction barrier - Super-resolution

Friday 27 October 17
Stimulated Emission Depletion Microscopy

Friday 27 October 17
Friday 27 October 17
Which imaging technique should I use?

1-5 m TIRF (for samples at the coverslip)

Wide-field (+deconvolution)

Fast
1-20 m
Sample Thickness

Spinning Disk Confocal

Sensitivity
10-100 m Line-scanning confocal

>20 m Point scanning Confocal

Slow
>50-100 m 2-photon confocal

Friday 27 October 17

You might also like