Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

A5E02338368-ADen MAG 5000-6000 OI en-US

Download as pdf or txt
Download as pdf or txt
You are on page 1of 106

Introduction 1

Safety notes 2

Description 3
SITRANS F
Installing/Mounting 4
Electromagnetic flowmeters
SITRANS FM MAG 5000/6000 IP67 5
Connecting

Commissioning 6
Operating Instructions

Operating 7

Service and maintenance 8


Diagnostics and
Troubleshooting 9

Technical data 10

Spare parts/Accessories A

Menu diagrams B

Factory settings C
Product documentation and
support D
7ME6910 (SITRANS MAG 5000)
7ME6920 (SITRANS MAG 6000)

12/2019
A5E02338368-AD
Legal information
Warning notice system
This manual contains notices you have to observe in order to ensure your personal safety, as well as to prevent
damage to property. The notices referring to your personal safety are highlighted in the manual by a safety alert
symbol, notices referring only to property damage have no safety alert symbol. These notices shown below are
graded according to the degree of danger.

DANGER
indicates that death or severe personal injury will result if proper precautions are not taken.

WARNING
indicates that death or severe personal injury may result if proper precautions are not taken.

CAUTION
indicates that minor personal injury can result if proper precautions are not taken.

NOTICE
indicates that property damage can result if proper precautions are not taken.
If more than one degree of danger is present, the warning notice representing the highest degree of danger will be
used. A notice warning of injury to persons with a safety alert symbol may also include a warning relating to property
damage.
Qualified Personnel
The product/system described in this documentation may be operated only by personnel qualified for the specific
task in accordance with the relevant documentation, in particular its warning notices and safety instructions. Qualified
personnel are those who, based on their training and experience, are capable of identifying risks and avoiding
potential hazards when working with these products/systems.
Proper use of Siemens products
Note the following:

WARNING
Siemens products may only be used for the applications described in the catalog and in the relevant technical
documentation. If products and components from other manufacturers are used, these must be recommended or
approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and
maintenance are required to ensure that the products operate safely and without any problems. The permissible
ambient conditions must be complied with. The information in the relevant documentation must be observed.

Trademarks
All names identified by ® are registered trademarks of Siemens AG. The remaining trademarks in this publication
may be trademarks whose use by third parties for their own purposes could violate the rights of the owner.
Disclaimer of Liability
We have reviewed the contents of this publication to ensure consistency with the hardware and software described.
Since variance cannot be precluded entirely, we cannot guarantee full consistency. However, the information in this
publication is reviewed regularly and any necessary corrections are included in subsequent editions.

Siemens AG Document order number: A5E02338368 Copyright © Siemens AG 2019.


Digital Industries Ⓟ 02/2020 Subject to change All rights reserved
Postfach 48 48
90026 NÜRNBERG
GERMANY
Table of contents

1 Introduction...................................................................................................................................................7
1.1 Purpose of this documentation.................................................................................................7
1.2 Document history .....................................................................................................................7
1.3 Checking the consignment.......................................................................................................8
1.4 Items supplied ..........................................................................................................................8
1.5 Security information .................................................................................................................8
1.6 Transportation and storage ......................................................................................................9
1.7 Notes on warranty ....................................................................................................................9
2 Safety notes................................................................................................................................................11
2.1 Laws and directives................................................................................................................11
2.2 Conformity with European directives......................................................................................12
2.3 Installation in hazardous location ...........................................................................................12
3 Description..................................................................................................................................................15
3.1 System components...............................................................................................................15
3.2 Operating principle .................................................................................................................15
3.3 Applications............................................................................................................................15
3.4 Features .................................................................................................................................16
3.5 MAG 5000/MAG 6000 versions .............................................................................................16
3.6 Nameplate layout ...................................................................................................................17
4 Installing/Mounting......................................................................................................................................21
4.1 Introduction ............................................................................................................................21
4.2 Installation conditions.............................................................................................................22
4.3 MAG 5000/6000 compact ......................................................................................................24
4.4 Remote installation.................................................................................................................26
4.4.1 At sensor ................................................................................................................................26
4.5 MAG 5000/6000 CT ...............................................................................................................30
4.5.1 Installing hardware key ..........................................................................................................30
4.5.2 Sealing of MAG 5000/6000 CT ..............................................................................................31
4.5.2.1 Verification sealing .................................................................................................................31
4.5.2.2 User sealing ...........................................................................................................................32
4.5.3 Installation conditions.............................................................................................................32
4.5.3.1 MI-001 ....................................................................................................................................32
4.5.3.2 PTB K7.2................................................................................................................................33
4.6 Turning transmitter/keypad ....................................................................................................33

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 3
Table of contents

5 Connecting .................................................................................................................................................37
5.1 Electrical connection ..............................................................................................................38
5.2 Electrical connection PTB K7.2..............................................................................................40
5.3 Connection of add-on modules ..............................................................................................40
6 Commissioning ...........................................................................................................................................41
6.1 MAG 5000/6000 Blind ............................................................................................................41
6.2 Local user interface................................................................................................................42
6.3 Menu structure .......................................................................................................................43
6.4 Changing password ...............................................................................................................44
6.5 Changing basic settings .........................................................................................................44
6.6 Changing operator menu setup..............................................................................................47
6.7 Changing language ................................................................................................................48
7 Operating....................................................................................................................................................51
7.1 Output settings .......................................................................................................................51
7.2 External input .........................................................................................................................53
7.3 Sensor characteristics............................................................................................................53
7.4 Reset mode............................................................................................................................53
7.5 Service mode .........................................................................................................................55
7.6 MAG 5000 CT and MAG 6000 CT settings............................................................................55
7.7 MAG 6000 SV ........................................................................................................................56
8 Service and maintenance ...........................................................................................................................57
8.1 Transmitter check list .............................................................................................................57
8.2 Technical support...................................................................................................................58
8.3 Return procedure ...................................................................................................................59
8.4 Recalibration ..........................................................................................................................60
9 Diagnostics and Troubleshooting ...............................................................................................................61
9.1 Diagnostics.............................................................................................................................61
9.2 List of error numbers ..............................................................................................................63
10 Technical data ............................................................................................................................................67
10.1 Technical specifications .........................................................................................................67
10.2 Accuracy ................................................................................................................................70
10.3 Output characteristics ............................................................................................................72
10.4 Cable data..............................................................................................................................75
10.5 Cable requirements................................................................................................................76

SITRANS FM MAG 5000/6000 IP67


4 Operating Instructions, 12/2019, A5E02338368-AD
Table of contents

A Spare parts/Accessories.............................................................................................................................77
A.1 Ordering of spare parts ..........................................................................................................77
A.2 Spare parts.............................................................................................................................77
A.3 Sun shield ..............................................................................................................................78
B Menu diagrams...........................................................................................................................................79
B.1 Transmitter menu overview....................................................................................................79
B.2 Basic settings .........................................................................................................................80
B.3 Current output ........................................................................................................................83
B.4 Digital output - pulse ..............................................................................................................83
B.5 Digital output - frequency .......................................................................................................84
B.6 Digital output / Relay output - Error level ...............................................................................84
B.7 Digital output / Relay output - Error number...........................................................................84
B.8 Digital output / Relay output - Direction/limit ..........................................................................85
B.9 Relay output - Cleaning..........................................................................................................85
B.10 Digital output / Relay output - Batch.......................................................................................85
B.11 External input .........................................................................................................................87
B.12 Sensor characteristics............................................................................................................88
B.13 Reset mode............................................................................................................................89
B.14 Reset mode - MAG 6000 SV..................................................................................................90
B.15 Service mode .........................................................................................................................91
B.16 Operator menu setup .............................................................................................................92
B.17 Product identity ......................................................................................................................93
B.18 Add-on communication module..............................................................................................94
B.19 Change password ..................................................................................................................94
C Factory settings ..........................................................................................................................................95
C.1 Transmitter factory settings....................................................................................................95
C.2 50 Hz Dimension dependent Qmax .......................................................................................97
C.3 60 Hz Dimension dependent Qmax .......................................................................................98
C.4 50 Hz Dimension dependent volume/pulse and batch.........................................................100
C.5 60 Hz Dimension dependent volume/pulse and batch.........................................................101
D Product documentation and support.........................................................................................................103
D.1 Product documentation ........................................................................................................103
D.2 Technical support.................................................................................................................104
Index.........................................................................................................................................................105

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 5
Table of contents

SITRANS FM MAG 5000/6000 IP67


6 Operating Instructions, 12/2019, A5E02338368-AD
Introduction 1
1.1 Purpose of this documentation
These instructions contain all information required to commission and use the device. Read the
instructions carefully prior to installation and commissioning. In order to use the device
correctly, first review its principle of operation.
The instructions are aimed at persons mechanically installing the device, connecting it
electronically, configuring the parameters and commissioning it, as well as service and
maintenance engineers.

1.2 Document history


This document describes:
● SITRANS F MAG 5000 and MAG 6000 transmitters (standard version).
● Optional versions:
– MAG 5000 Blind and MAG 6000 Blind
– MAG 5000 CT and MAG 6000 CT
– MAG 6000 SV

Documentation history
The following table shows major changes in the documentation compared to the previous
edition.

Edition Remarks FW version


12/2019 ● BBL42 as default unit 4.09
● Improved operation without SENSORPROM
● Responsibility transfer to Siemens AG
12/2013 ● Customer defined unit 4.07
● Velocity value with unit
● Operational without SENSORPROM
● Signal suitability
01/2012 4.04
01/2010 First edition

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 7
Introduction
1.5 Security information

1.3 Checking the consignment


1. Check the packaging and the delivered items for visible damages.
2. Report any claims for damages immediately to the shipping company.
3. Retain damaged parts for clarification.
4. Check the scope of delivery by comparing your order to the shipping documents for
correctness and completeness.

WARNING
Using a damaged or incomplete device
Risk of explosion in hazardous areas.
● Do not use damaged or incomplete devices.

1.4 Items supplied

● SITRANS F M MAG 5000/6000 transmitter


● Siemens Process Instrumentation documentation disk
containing certificates, and manuals
e
Not
● Safety note Saf
ety

● Calibration report
p.
. re
Cal

1.5 Security information


Siemens provides products and solutions with industrial security functions that support the
secure operation of plants, systems, machines and networks.
In order to protect plants, systems, machines and networks against cyber threats, it is
necessary to implement – and continuously maintain – a holistic, state-of-the-art industrial
security concept. Siemens’ products and solutions constitute one element of such a concept.
Customers are responsible for preventing unauthorized access to their plants, systems,
machines and networks. Such systems, machines and components should only be connected
to an enterprise network or the internet if and to the extent such a connection is necessary and
only when appropriate security measures (e.g. firewalls and/or network segmentation) are in
place.
For additional information on industrial security measures that may be implemented, please
visit
https://www.siemens.com/industrialsecurity.
Siemens’ products and solutions undergo continuous development to make them more secure.
Siemens strongly recommends that product updates are applied as soon as they are available
and that the latest product versions are used. Use of product versions that are no longer

SITRANS FM MAG 5000/6000 IP67


8 Operating Instructions, 12/2019, A5E02338368-AD
Introduction
1.7 Notes on warranty

supported, and failure to apply the latest updates may increase customer’s exposure to cyber
threats.
To stay informed about product updates, subscribe to the Siemens Industrial Security RSS
Feed under
https://www.siemens.com/industrialsecurity.

1.6 Transportation and storage


To guarantee sufficient protection during transport and storage, observe the following:
● Keep the original packaging for subsequent transportation.
● Devices/replacement parts should be returned in their original packaging.
● If the original packaging is no longer available, ensure that all shipments are properly
packaged to provide sufficient protection during transport. Siemens cannot assume liability
for any costs associated with transportation damages.

NOTICE
Insufficient protection during storage
The packaging only provides limited protection against moisture and infiltration.
● Provide additional packaging as necessary.

Special conditions for storage and transportation of the device are listed in Technical data
(Page 67).

1.7 Notes on warranty


The contents of this manual shall not become part of or modify any prior or existing agreement,
commitment or legal relationship. The sales contract contains all obligations on the part of
Siemens as well as the complete and solely applicable warranty conditions. Any statements
regarding device versions described in the manual do not create new warranties or modify the
existing warranty.
The content reflects the technical status at the time of publishing. Siemens reserves the right
to make technical changes in the course of further development.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 9
Introduction
1.7 Notes on warranty

SITRANS FM MAG 5000/6000 IP67


10 Operating Instructions, 12/2019, A5E02338368-AD
Safety notes 2
CAUTION
Correct, reliable operation of the product requires proper transport, storage, positioning and
assembly as well as careful operation and maintenance.
Only qualified personnel should install or operate this instrument.

Note
Alterations to the product, including opening or improper modifications of the product are not
permitted.
If this requirement is not observed, the CE mark and the manufacturer's warranty will expire.

2.1 Laws and directives

General requirements
Installation of the equipment must comply with national regulations. For example EN 60079-14
for the European Community.

Instrument safety standards


The device has been tested at the factory, based on the safety requirements. In order to
maintain this condition over the expected life of the device the requirements described in these
Operating Instructions must be observed.

Environmental conditions according to IEC 61010-1 (2010)


● Indoor use
● Altitude up to 2000m
● Maximum relative humidity 80% for temperatures up to 31°C (88°F) decreasing linearly up
to 50% relative humidity from 40°C (104°F)
● Main supply voltage fluctuations up to ±10% of the nominal voltage (see Technical
specifications (Page 67))
● Overvoltage category II
● Pollution degree 2

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 11
Safety notes
2.3 Installation in hazardous location

Environmental conditions according MID (Directive 2014/32/EU)


● Environment class: E2 (electromagnetic), M1 (mechanical)
● Climatic class: -25°C - +55°C, condensing, closed

2.2 Conformity with European directives


The CE marking on the device symbolizes the conformity with the following European
directives:

Electromagnetic compatibili‐
Directive of the European Parliament and of the Council on the
ty EMC harmonisation of the laws of the Member States relating to elec‐
2014/30/EU tromagnetic compatibility
Low voltage directive LVD Directive of the European Parliament and of the Council on the
2014/35/EU harmonisation of the laws of the Member States relating to the
making available on the market of electrical equipment designed
for use within certain voltage limits
Atmosphère explosible Directive of the European Parliament and the Council on the
ATEX harmonisation of the laws of the Member States relating to
2014/34/EU equipment and protective systems intended for use in potentially
explosive atmospheres
CT: Measuring instruments Directive of the European Parliament and the Council on the
directive MID harmonisation of the laws of the Member States relating to the
2014/32/EU making available on the market of measuring instruments
The applicable directives can be found in the EU declaration of conformity of the specific device.

2.3 Installation in hazardous location

WARNING
Conditions for safe use
Equipment used in hazardous areas must be Ex-approved and marked accordingly. It is
required that the special conditions for safe use provided in the manual and in the Ex
certificate are followed!

Ex approvals
CSA Class I, Division 2, Groups A, B, C and D. Code T5 for an ambient temperature of +60 °C.
FM Class I, Division 2, Groups A, B, C and D and Class I, Zone 2, Group IIC indoor/outdoor
Type IP67 hazardous (classified) locations.

SITRANS FM MAG 5000/6000 IP67


12 Operating Instructions, 12/2019, A5E02338368-AD
Safety notes
2.3 Installation in hazardous location

Process temperature specifications for Ex use

Temperature class Ambient temperature [°C]


-40 to +40 -40 to +50 -40 to +60
T2 180 - -
T3 165 140 -
T4 100 100 80
T5 65 65 65
T6 50 50 50

EX requirements
It is required that:
● Electrical connections are in accordance with Elex V (VO in explosion hazardous areas) and
EN60079-14 (Installing Electrical Systems in Explosion Hazardous Areas).
● The protective cover over the power supply is properly installed. For intrinsically safe circuits
the connection area can be opened.
● Appropriate cable connectors are used for the output cicuits: intrinsically safe: blue, non-
intrinsically safe: black.
● Sensor and transmitter are connected to the potential equalization. For intrinsically safe
output circuits potential equalization must be maintained along the entire connection path.
● Sensor insulation thickness is max. 100mm (only insulated sensors).
● EN60079-31 is considered for installation in areas with combustible dust.
● When protective earth (PE) is connected, no potential difference between the protective
earth (PE) and the potential equalization (PA) can exist, even during a fault condition.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 13
Safety notes
2.3 Installation in hazardous location

SITRANS FM MAG 5000/6000 IP67


14 Operating Instructions, 12/2019, A5E02338368-AD
Description 3
3.1 System components
A SITRANS F M MAG 5000/6000 flowmeter system includes:
● Transmitter (type SITRANS F M MAG 5000/6000)
● Sensor (types: SITRANS F MAG 1100/1100F/3100/3100 P/5100 W)
● Communication module (optional) (types: HART, PROFIBUS PA/DP, MODBUS RTU RS
485, Foundation Fieldbus H1, Devicenet)
● SENSORPROM memory unit

Communication solutions
The SITRANS F USM II range of add on modules, presently including HART, Foundation
Fieldbus. MODBUS RTU RS 485, PROFIBUS PA / DP and Devicenet, are all applicable with
the SITRANS F M MAG 6000 transmitter.

3.2 Operating principle


The transmitters are microprocessor-based with a built-in alphanumeric display in several
languages. The flow measuring principle is based on Faraday's law of electromagnetic
induction. Magnet coils mounted diametrically on the measuring pipe generate a pulsed
electromagnetic field. The liquid flowing through this electromagnetic field induces a voltage.
The transmitters evaluate the signals from the associated electromagnetic sensors, convert the
signals into appropriate standard signals such as 4 to 20 mA, and also fulfil the task of a power
supply unit providing the magnet coils with a constant current.
The transmitter consists of a number of function blocks which convert the sensor voltage into
flow readings.

3.3 Applications
The pulsed DC-powered magnetic flowmeters are suitable for measuring the flow of almost all
electrically conductive liquids, pastes, and slurries with max. 40% solids.
The main applications can be found in the following sectors:
● Water and waste water
● Chemical and pharmaceutical industries
● Food & beverage industry
● Mining and cements industries
● Pulp and paper industry

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 15
Description
3.5 MAG 5000/MAG 6000 versions

● Steel industry
● Power generation; utility and chilled water industry

WARNING
This is a Class A product
In a domestic environment this product may cause radio interference in which case the user
may be required to take adequate measures.

3.4 Features

Power supply
2 different types of power supply are available. A 12 to 24 V AC/DC and a 115 to 230 V AC
switch mode type.
Coil current module generates a pulsating magnetizing current that drives the coils in the
sensor. The current is permanently monitored and corrected. Errors or cable faults are
registered by the self-monitoring circuit.
Input circuit amplifies the flow-proportional signal from the electrodes. The input impedance is
extremely high: >1014 Ω which allows flow measurements on fluids with conductivities aslow
as 5 μS/cm. Measuring errors due to cable capacitance are eliminated due to active cable
screening.
Digital signal processor converts the analog flow signal to a digital signal and suppresses
electrode noise through a digital filter. Inaccuracies in the transmitter as a result of long-term
drift and temperature drift are monitored and continuously compensated for via the
selfmonitoring circuit. The analog to digital conversion takes place in an ultra low noise ASIC
with 23 bit signal resolution. This has eliminated the need for range switching. The dynamic
range of the transmitter is therefore unsurpassed with a turn down ratio of minimum 3000:1.

Dialog module
The display unit consists of a 3-line display and a 6-key keypad. The display shows a flow rate
or a totalizer value as a primary reading.

Output module
The output module converts flow data to analog, digital and relay outputs. The outputs are
galvanically isolated and can be individually set to suit a particular application.

3.5 MAG 5000/MAG 6000 versions


The transmitters are designed in various versions and offer high performance and easy
installation, commissioning and maintenance.

SITRANS FM MAG 5000/6000 IP67


16 Operating Instructions, 12/2019, A5E02338368-AD
Description
3.6 Nameplate layout

Standard version
The standard version is an IP67 version for compact or remote installation. Its robust design
ensures a long lifetime if installed outdoors.

Blind version
This version carries all the normal MAG 5000/6000 features, except those associated with the
display and keypad.
Both current and digital outputs are available.
Factory setting of current output in unit is switched off when delivered.

CT version
The MAG 5000/6000 CT version is a custody transfer approved transmitter.
It is approved according to:
● Cold water approval:
– MI-001 (Tested according to OIML R 49)
● Other media than water:
– PTB K7.2
The above approval specifications apply at the time of publication. For the latest approval
updates, see: http://support.automation.siemens.com/WW/view/en/10806951/134200

SV version (MAG 6000 only)


This version is identical to the standard MAG 6000 transmitters except for the following
additional functions:
● Zero point adjustment
● Adjustable excitation frequency up to 44 Hz

3.6 Nameplate layout

Inspection
1. Check for mechanical damage due to possible improper handling during shipment. All
claims for damage are to be made promptly to the shipper.
2. Make sure the scope of delivery, and the information on the type plate corresponds to the
ordering information.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 17
Description
3.6 Nameplate layout

Identification


 6,75$16)00$*&7:
 2UGHU1R 0($$$$
 6HULDO1R +
 6XSSO\ 9$&+]9$
 ,31(0$
 7DPE rWRr&r)WRr)

&ODVV,'LYLVLRQ*URXS$%&'7
12',5(&7681/,*+7(;32685(


0HDVXULQJ(TXLSPHQW
APPROVED 1$
6LHPHQV$*'(.DUOVUXKH
 0DGHLQ)UDQFH

① Manufacturer
② Product name
③ Order number
④ Serial number
⑤ Power supply
⑥ Degree of protection
⑦ Ambient temperature
⑧ Conformity with country-specific directives
⑨ Place of manufacture
Figure 3-1 MAG 6000 transmitter nameplate example

SITRANS FM MAG 5000/6000 IP67


18 Operating Instructions, 12/2019, A5E02338368-AD
Description
3.6 Nameplate layout



















① Manufacturer
② Product name
③ System order number
④ System serial number
⑤ Transmitter serial number
⑥ Sensor serial number
⑦ Software version and hardware version
⑧ Ambient temperature
⑨ Uncertainty
⑩ Certification number
⑪ Conformity with country-specific directives
⑫ Place of manufacture
Figure 3-2 MAG 6000 CT system nameplate example

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 19
Description
3.6 Nameplate layout


 
 
 
 
 
 
 
 
 
 

① Manufacturer
② Product name
③ System order number
④ System serial number
⑤ Nominal diameter
⑥ Sensor material
⑦ Meter orientation
⑧ Environmental Class
⑨ Fluid group
⑩ Power supply
⑪ Certification number
⑫ Maximum allowable working pressures
⑬ Max. media temperature
⑭ Min. media temperature
⑮ Process connection
⑯ Year of manufacturing
⑰ Software version and hardware version
⑱ Dynamic range
⑲ Maximum flow rate
⑳ Conformity with country-specific directives
Place of manufacture
Figure 3-3 MAG 5000/6000 CT system sensor nameplate example

Note
The matched paired transmitter and sensor shall be mounted together
At installation, please check that the system nameplates of transmitter and sensor have the
same system serial number.

SITRANS FM MAG 5000/6000 IP67


20 Operating Instructions, 12/2019, A5E02338368-AD
Installing/Mounting 4
4.1 Introduction
● SITRANS F flowmeters are suitable for indoor and outdoor installations.

WARNING
Installation in hazardous location
Special requirements apply to the location and interconnection of sensor and transmitter.
See Installation in hazardous location (Page 12)

This chapter describes how to install the flowmeter in the compact version as well as in the
remote version.
The transmitter is delivered ready for mounting on the sensor. The transmitter is delivered with
a compression plate ready for mounting on the sensor. No further assembling is necessary.
The transmitter can be installed either compact on the sensor or remote.

Figure 4-1 Compact installation

Figure 4-2 Remote installation

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 21
Installing/Mounting
4.2 Installation conditions

CAUTION
See Cable requirements (Page 76) before installing transmitter

4.2 Installation conditions


Reading and operating the flowmeter is possible under almost any installation conditions
because the display can be oriented in relation to the sensor. To ensure optimum flow
measurement, attention should be paid to the following:

Vibrations

Figure 4-3 Avoid strong vibrations

SITRANS FM MAG 5000/6000 IP67


22 Operating Instructions, 12/2019, A5E02338368-AD
Installing/Mounting
4.2 Installation conditions

Compact installation

Medium temperature must be in accordance with the graphs showing max. ambient
temperature as a function of medium temperature.

$PELHQWWHPSHUDWXUH

0HGLXP
WHPSHUDWXUH

&7YHUVLRQ
6WDQGDUGEOLQGDQG69YHUVLRQV
Figure 4-4 Medium and ambient temperatures

Remote installation

Cable length and type (as described in Cable requirements (Page 76)) must be used.
For installation conditions for sensors, see relevant sensor Operating Instructions.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 23
Installing/Mounting
4.3 MAG 5000/6000 compact

4.3 MAG 5000/6000 compact

Install MAG 5000 / MAG 6000 compact version


1. Remove and discard terminal box lid of sensor.

2. Ensure SENSORPROM® memory unit is installed.


3. Fit M20 or ½" NPT cable glands for supply and output cables.
4. Unplug the two black plug assemblies for coil and electrode cables in terminal box.
5. Connect earth wire from connection board to bottom of terminal box.
6. Connect 2-pin connector and 3-pin connector as shown to their corresponding terminal
numbers on connection board as shown in Electrical connection (Page 38).
Note
System will not register flow if black plugs are not connected to connection board.
A

7. Fit supply and output cables through cable glands and connect to connection plate as
shown in Electrical connection (Page 38).

SITRANS FM MAG 5000/6000 IP67


24 Operating Instructions, 12/2019, A5E02338368-AD
Installing/Mounting
4.3 MAG 5000/6000 compact

8. Mount connection plate in terminal box.


Note
Check that your connection board lines up with SENSORPROM® unit, if not, move
SENSORPROM® unit to the other side of terminal box.
SENSORPROM® memory unit connections will be established automatically when
connection plate is mounted in terminal box.

9. Tighten cable glands to obtain optimum sealing.


10.Mount transmitter on terminal box. Recommended torque value: 2.5 Nm.

11.Transmitter is ready to be powered up.

NOTICE
Exposing transmitter to direct sunlight may increase operating temperature above its
specified limit, and decrease display visibility.
A sunshield is available as accessory.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 25
Installing/Mounting
4.4 Remote installation

4.4 Remote installation

4.4.1 At sensor
1. Remove terminal box lid.

2. Mount the two terminal blocks as shown and insert electrode cable plug ① (terminals 85
and 86) and coil cable plug ② (terminals 82, 0 and 83).


3. Remove SENSORPROM (to be mounted in transmitter terminal box). Ensure that the serial
no. on the SENSORPROM label is identical to the sensor serial no.

SITRANS FM MAG 5000/6000 IP67


26 Operating Instructions, 12/2019, A5E02338368-AD
Installing/Mounting
4.4 Remote installation

4. Connect electrode cable ③ and coil cable ④ in the corresponding terminals on the terminal
blocks.

5. Remount terminal box lid.


Note
Tightening torque
Tighten the bolts with 0.5 Nm.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 27
Installing/Mounting
4.4 Remote installation

Wall mounting
1. Mount bracket on a wall or on a horizontal or a vertical pipe using ordinary hose clips or duct
straps.

Figure 4-5 Wall mounting

Figure 4-6 Pipe mounting - vertical

Figure 4-7 Pipe mounting - horizontal

2. Ensure that correct SENSORPROM® memory unit is mounted in wall/pipe mounting unit.
3. Fit M20 or ½" NPT cable glands for cables from bottom or sides of terminal box.

SITRANS FM MAG 5000/6000 IP67


28 Operating Instructions, 12/2019, A5E02338368-AD
Installing/Mounting
4.4 Remote installation

4. Mount earth wire in bottom of terminal box.

3 1

2 4

① Connect electrode cable


② Connect coil cable - keep separate from electrode cable
③ Connect power supply
④ Connect output cable
⑤ Connect PE (ground) wire

5. Mount connection plate in terminal box.


B

3 1

2 4

① Electrode cable
② Coil cable
③ Power supply
④ Output cable

6. Fit coil, electrode, supply and output cables through cable glands and connect to connection
plate as shown in Electrical connection (Page 38).
7. Fix connection plate with the two diagonally opposite screws.
8. Tighten cable glands to obtain optimum sealing.

CAUTION
Coil cable shield
When remote mounted, power supply PE wire must be connected to PE terminal ( ).
Coil cable shield must be connected to SHIELD terminal.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 29
Installing/Mounting
4.5 MAG 5000/6000 CT

9. Mount transmitter on terminal box. Recommended torque value: 2.5 Nm.

10.Transmitter is ready to be powered up.

NOTICE
Direct sunlight
Exposing the transmitter to direct sunlight may increase the operating temperature above
its specified limit, and decrease display visibility.
A sun shield is available as accessory.

4.5 MAG 5000/6000 CT


Calibration sealing has been carried out at calibration.
MAG 6000 CT is installed like a Standard MAG 6000 except for the final sealing.

4.5.1 Installing hardware key

Use hardware key on non-verified transmitter

Note
Hardware key function
Setting of primary operating parameters is blocked during normal operation.
When key is mounted, access to all menu items is gained. When key is removed, primary
settings are blocked in accordance with requirements in authorisation.

SITRANS FM MAG 5000/6000 IP67


30 Operating Instructions, 12/2019, A5E02338368-AD
Installing/Mounting
4.5 MAG 5000/6000 CT

1. Mount hardware key on transmitter connection plate during setting of primary operating
parameters such as Qmax., low-flow cut-off, units, approvals, etc. in connection with
commissioning or calibration. See setup menus in appendix Transmitter menu overview
(Page 79).

2. Remove hardware key after setting up and calibrating unit.


This locks the menu structure and the selected settings.

4.5.2 Sealing of MAG 5000/6000 CT

4.5.2.1 Verification sealing

Note
The verification sealing is done at factory after initial verification.

nn
mnm
2

'.
20785



nn
nm
A5E34

mn nn
nmn
nmnm n
mnnnmnmn
m nm
mn mnn
mn

Under transmitter display frame On bottom of transmitter On side of transmitter

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 31
Installing/Mounting
4.5 MAG 5000/6000 CT

Note
For type-approved and verified MID MAG 5000/6000 CT products
The verification sealing may only be broken by an authorized person, with the acceptance and
under direction of the local authorities.

4.5.2.2 User sealing

Note
User sealing has to be done after commissioning by an authorized person.

1. Drill through marked drilling holes in terminal box and transmitter/lid.


2. Seal the transmitter on both sides with either one or two wires, as shown below.

Compact version sealed with Remote version sealed with Compact version sealed with
one wire one wire two wires

4.5.3 Installation conditions

4.5.3.1 MI-001
MAG 5000/6000 CT together with MAG 5100 W (7ME652) are approved for MI-001 under the
following installation conditions.
● DN 50 to 300 (2" to 12") in any orientation
● DN 350 to 1200 (14" to 48") only in horizontal installation
● Compact or remote with max. 500 m (1640 ft.) cable
● Power supply 115 to 230 V AC and 12 to 24 V DC

SITRANS FM MAG 5000/6000 IP67


32 Operating Instructions, 12/2019, A5E02338368-AD
Installing/Mounting
4.6 Turning transmitter/keypad

Other restrictions may apply (see certificate).

/
4.5.3.2 PTB K7.2
MAG 5000/6000 CT together with MAG 5100 W (7ME652) are approved for PTB K7.2 under
the following installation conditions.
SITRANS F M MAG 5100 W with MAG 5000/6000CT
● DN 15 to DN 300 (1/2" to 12") in any orientation
● DN 350 to DN 1200 (14" to 48") only in horizontal installation
● Compact or remote with max. 500 m (1640 ft.) cable
Other restrictions may apply (see certificate)

4.6 Turning transmitter/keypad

Note
Not allowed for MAG 5000/6000 CT
Altering the orientation of the transmitter or keypad is prohibited for non-CT versions.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 33
Installing/Mounting
4.6 Turning transmitter/keypad

Transmitter

Figure 4-8 Transmitter can be mounted with its front in either direction indicated by the arrows without
turning terminal box

Figure 4-9 Terminal box can be rotated ±90° in order to optimize viewing angle of transmitter display/
keypad

1. Unscrew the four screws in bottom of terminal box.


2. Turn terminal box to required position.
3. Retighten screws firmly.

SITRANS FM MAG 5000/6000 IP67


34 Operating Instructions, 12/2019, A5E02338368-AD
Installing/Mounting
4.6 Turning transmitter/keypad

Keypad
1. Remove outer frame using a screwdriver.

2. Loosen the four screws retaining keypad.

3. Withdraw keypad and turn it to required orientation.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 35
Installing/Mounting
4.6 Turning transmitter/keypad

4. Tighten the four screws until a mechanical stop is felt in order to obtain IP67 enclosure.

5. Snaplock outer frame onto keypad (click).

SITRANS FM MAG 5000/6000 IP67


36 Operating Instructions, 12/2019, A5E02338368-AD
Connecting 5
WARNING
The pertinent regulations must be observed for electrical installation.
● Never install the device with the mains voltage switched on!
● Danger of electric shock!
● The electrodes and magnetic current line may only be connected when the device is not
connected to the power supply.
● If the housing is under voltage (power supply), the cover may be unscrewed by qualified
personnel only.

WARNING
Mains supply from building installation Class II
A switch or circuit breaker (max. 15 A) must be installed in close proximity to the equipment
and within easy reach of the operator. It must be marked as the disconnecting device for the
equipment.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 37
Connecting
5.1 Electrical connection

5.1 Electrical connection

Power supply
Transmitter

115 to 230 V AC 11 to 30 V DC/11 to 24 V AC


L1 L 1
N N 2
PE PE

Outputs Current output


(Powered from transmitter)
31 0/4 - 20 mA
32 Load ≤ 800 Ω

Passive output
(External powered) Digital output
Vx 3 to 30 V
56 max. 110 mA
57 Menu setup
58 R PLC-Digital input
Negative: Positive:
Active output
(Powered from transmitter) 24 V max. 30 mA
56 Counter or PLC-Digital input R = Pull up/down resistor
1 to 10 KΩ may be
57 required - depending on
R Cables/Input resistance
58
Relay output
44 NO Relay
45 NC 24 V DC/1A
42 V AC/2A
46 Common

Digital input
77
11 to 30 V DC Input
78

91 92 93 94 95 96 97 Sensor
Reserved for communication modules

Sensor connection
81 Electrode cable 81
Brown
82 82
1)
Black
0 0
Blue
83 83
84 84

Coil cable
Blue
85 85
Brown
86 86

Shield

1) Note:
Special cable with individual wire shields (shown as dotted lines) are only required
when using empty pipe function or long cables.

Figure 5-1 Wiring diagram

Note
Terminals 81 and 84 are only to be connected if special electrode cable with double screening
is used, e.g. when empty pipe function or long cables are used.

SITRANS FM MAG 5000/6000 IP67


38 Operating Instructions, 12/2019, A5E02338368-AD
Connecting
5.1 Electrical connection

Mains supply
Mains supply 115 to 230 V AC from building installation Class II.

Note
For DC installations it is recommend to install an under voltage relay or protection circuit in the
application where there is a risk of low power supply below the specifications for more than 10
minutes.

WARNING
Grounding
Connect mains protective earth wire to PE terminal in accordance with diagram (due to class
1 power supply).

Mechanical counter
Connect a 1000 µF capacitor (capacitor+ to terminal 56 and capacitor- to terminal 58) if a
mechanical counter is connected to terminals 57 and 58 (active output).

Output cables
Use screened cables if long cables are used in noisy environments.

Digital output
If internal resistance of a load exceeds 10 kΩ, connect an external 10 kΩ load resistor in parallel
to this load.

WARNING
Intrinsically safe terminals
Always ensure that distance between cables/wires is minimum 50 mm in order to avoid that
wires/terminals of intrinsically safe circuits get into contact with wires of other cables.
Fasten cables/wires in a way that they cannot get into contact with each other, not even in
case of an error. Keep wire ends as short as possible.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 39
Connecting
5.3 Connection of add-on modules

5.2 Electrical connection PTB K7.2


Additional Electrical connection for PTB K7.2 approved MAG 5000/6000 CT with MAG 5100 W
(7ME652)

56 +
57 -

① MAG 5000/6000 CT pulse output


② Energy calculator

5.3 Connection of add-on modules


When the add-on module has been installed, the electrical connections are available on
terminal rows 91-97.

For more information


Refer to the relevant BUS communication Quick Start or Operating Instructions available on an
included DVD or on the internet, at: www.siemens.com/flowdocumentation
(www.siemens.com/flowdocumentation).

SITRANS FM MAG 5000/6000 IP67


40 Operating Instructions, 12/2019, A5E02338368-AD
Commissioning 6
In this chapter it is described how to commission the device via the local user interface (LUI).
The display is described in details in section Local user interface (Page 42).
Furthermore, the following functions are described in details:
● Changing password (Page 44)
● Changing basic settings (Page 44)
● Changing operator menu setup (Page 47)
● Changing language (Page 48)
Detailed diagrams concerning the specific menu are shown in appendix menu diagrams.
For factory settings, see Factory settings (Page 95).

6.1 MAG 5000/6000 Blind

Note
Does not have a display. All factory settings will be uploaded from the SENSORPROM® unit
after power-up.

For sensor dependent factory settings, see Transmitter factory settings (Page 95).

Changing settings
If other settings are required, a standard transmitter with display and similar power supply can
be used.
1. Unscrew and remove MAG 5000/6000 Blind.
2. Mount standard MAG 5000/6000 transmitter.
3. Change required settings via display and keypad.
All changed data will be stored in SENSORPROM® memory unit.
4. Remove standard transmitter and remount Blind transmitter.
5. Fasten screws holding transmitter.
New settings stored in SENSORPROM® memory unit will be uploaded in blind transmitter.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 41
Commissioning
6.2 Local user interface

6.2 Local user interface

S Sign field
P Primary field for numeric value flow rate, Totalizer 1 or Totalizer 2)
U Unit field
T Title line with individual information according to operator or setup menu selected.
ST Subtitle line which will either add information to the title line or keep individual information inde‐
pendent of the title line.
F Alarm field. Two flashing triangles will appear in case of a fault condition.
M Mode field
L Lock field
Figure 6-1 Local User Interface

Mode field symbols

Communication mode Language mode Sensor characteristics


Service mode Basic settings Reset mode
Operator menu Output Operator-active

Product identity External input Operator-inactive

Lock field symbols

Ready for change Access to submenu


Value locked RESET MODE: Zero setting of totalizers and initializa‐
tion of setting

SITRANS FM MAG 5000/6000 IP67


42 Operating Instructions, 12/2019, A5E02338368-AD
Commissioning
6.3 Menu structure

Keypad
The keypad is used to set the flowmeter. The keys function as follows:

TOP UP KEY This key (when held for 2 sec.) is used to switch between op‐
erator menu and setup menu. In transmitter setup menu, a
short press will cause a return to previous level.
FORWARD KEY This key is used to step forward through the menus. It is the
only key normally used by the operator.
BACKWARD KEY This key is used to step backwards through the menus.
CHANGE KEY With this key settings or numerical values are changed.
SELECT KEY With this key figures to be changed are selected.
LOCK/UNLOCK This key enables the operator to change settings and it gives
KEY access to submenus.

6.3 Menu structure

Note
Menus disabled on MAG 5000/6000 CT
Due to legal requirements, some parameters are disabled on MAG 5000/6000 CT.

The menu is built up of two parts. An operator menu and a setup menu, see Transmitter menu
overview (Page 79).

Operator menu
The operator menu is for daily operation. It is customized in the operator menu setup. The
transmitter always starts up in operator menu No. 1. The forward and the backward keys
are used to step through the operator menus.

Setup menu
The setup menu is for commissioning and service only. Access to the setup menu is gained by
pressing the top up key for 2 seconds. The setup menu operates in two modes:
● View mode
● Setup mode
View mode is a read-only mode. The pre-selected settings can only be scanned.
Setup mode is a read and write mode. The pre-selected settings can be scanned and changed.
Access to the setup mode is password-protected. The factory set password is 1000.
Access to a submenu in the setup menu is gained by pressing the lock key . Press the top
up key briefly to return to the previous menu. Press longer (2 sec.) to exit the setup menu and
return to operator menu No. 1.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 43
Commissioning
6.5 Changing basic settings

6.4 Changing password


The setup menu is password-protected in order to ensure that only authorized personnel can
make any changes in transmitter settings.
Change password as follows:
1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key or backward key to reach password menu.
4. Press lock/unlock key to unlock password.
5. Use select key and change key to change password.
6. Press lock/unlock key to confirm new password.
7. Press top up key two times to exit setup mode.

3DVVZRUG

&KDQJHSDVVZRUG

The factory-set password is 1000, but it can be changed to any value between 1000 and 9999.
Factory setting of password can be re-established as follows:
1. Switch off power supply.
2. While pressing top up key - switch on power supply.
3. Release top up key after 10 sec.

6.5 Changing basic settings


In the basic settings menu it is possible to set the following parameters:

Parameter Description
Main frequency Selection of main power supply frequency corresponding to the country in which
the flowmeter is installed (e.g. 60 Hz in America).
Flow direction Selection of correct flow direction in pipe.
Customer units Setting of user defined volume and time units.
Qmax Setting of measuring range, analog outputs and frequency output. Also individ‐
ual dimension-dependent setting of value, decimal point, unit and time.
Qmax 2 Setting of measuring range, analog outputs and frequency output. Also individ‐
ual dimension-dependent setting of value, decimal point, unit and time. This
menu is only visible if chosen as external digital input.
Totalizer Setting of unit and decimal point.

SITRANS FM MAG 5000/6000 IP67


44 Operating Instructions, 12/2019, A5E02338368-AD
Commissioning
6.5 Changing basic settings

Parameter Description
Low flow cut-off Setting of a percentage of selected Qmax. This filters noise in installation reducing
fluctuations in display and all outputs.
Empty pipe cut-off When set to "On" the alarm will indicate when sensor is running empty. All read‐
ings, display and outputs, will indicate zero.
Velocity unit Setting of velocity unit per time unit
Error level Selecting error level at which flowmeter will detect an error.

Note
Totalizer 2 is not visible when batch is selected as digital output.

Note
Qmax 2 is visible only when chosen as digital input.

Change basic settings as follows:


1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key to reach basic settings menu.
4. Press lock/unlock key to unlock settings.
5. Use forward key or backward key to reach relevant menu.
6. Press lock/unlock key to unlock settings.
7. Use select key and change key to change settings.
8. Press lock/unlock key to confirm new settings.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 45
Commissioning
6.5 Changing basic settings

9. Repeat steps 5-8 to change other settings.


10.Press top up key two times to exit setup mode.
%DVLF
VHWWLQJV

0DLQIUHTXHQF\

+]
+]

)ORZGLUHFWLRQ
3RVLWLYH
3RVLWLYH
1HJDWLYH

&XVWRPHUXQLWV &XVWYROXPHXQLW &XVWYROXPHXQLW


WH[W" HTXDO\\\\\\\P

&XVWWLPHXQLW &XVWWLPHXQLW
4PD[
PV− WH[W" HTXDO\\\\\\\\V
PV−
P V 36
PO 0LQ 30
O K 3+
KO G 3'
0O " "
IW
LQ
86*
86N*
%%/
%%/
8.*
8.0*
%%/

4PD[ QLJKW
PV−
PV−
P V 36
PO 0LQ 30
O K 3+
KO G 3'
0O " "
IW
LQ
86*
86N*
%%/
%%/
8.*
8.0*
%%/



7RWDOL]HU 7RWDOL]HU 7RWDOL]HU

/RZIORZ 7RWDOL]HUXQLW 7RWDOL]HUXQLW


FXW−RII [[[[[[, [[[[[[,
−  P P
PO PO
O O
KO KO
0O 0O
IW IW
LQ LQ
(PSW\SLSHFXW−RII 86* 86*
RII 86N* 86N*
%%/ %%/
RII
RQ %%/ %%/
8.* 8.*
8.0* 8.0*
%%/ "

9HORFLW\XQLW
\\\\\\PV 7RWDOL]HU 7RWDOL]HU
)RUZDUG 5HYHUVH
P V )RUZDUG )RUZDUG
PP PLQ 5HYHUVH 5HYHUVH
FP K 1HW 1HW
GP G
*

IW "
LQ

(UURUOHYHO
:DUQLQJ
)DWDO
3HUPDQHQW
:DUQLQJ

Decimal point can be positioned and units set individually for flow rate in totalizer 1 and totalizer
2.

SITRANS FM MAG 5000/6000 IP67


46 Operating Instructions, 12/2019, A5E02338368-AD
Commissioning
6.6 Changing operator menu setup

Changing decimal point position


1. Enter the respective totalizer menu.
2. Use select key to position cursor below decimal point.
3. Use change key to move decimal point to requested position.

Changing units

Note
Menus disabled on MAG 5000/6000 CT
Due to legal requirements, some parameters are disabled on MAG 5000/6000 CT. Only
available units are m3/h and m3.

1. Use select key to position cursor below unit.


2. Press change key until requested unit is displayed.

6.6 Changing operator menu setup


In the operator menu the menus required for daily operation of the flowmeter are shown. It is
possible to hide and change some of the menus in the operator menu. This is done in the
operator menu setup menu, see diagram Operator menu setup (Page 92).

Customizing menus in operator menu


To customize the menus in the operator menu perform the following steps:
1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key or backward key to reach operator menu.

Changing text in line 1


1. Press lock/unlock key to unlock setting.
2. Use change key to select desired text.
3. Press lock/unlock key to confirm selected text.
Note
If "Text" is selected in line 2, this line functions as a heading for the value shown in line 3.
Otherwise it shows the actual value of the reading selected.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 47
Commissioning
6.7 Changing language

Enabling two readings


1. Use forward key to reach requested menu.
2. Press lock/unlock key to unlock setting.
3. Use select key to move cursor to upper line.
4. Use change key to select requested reading.
5. Press lock/unlock key to confirm selection.
6. Use select key to move cursor to line 3.
7. Use change key to select desired setting.
8. Press lock/unlock key to confirm new setting.
9. Repeat steps 1-8 for each requested menu.

Showing/hiding menus in operator menu


1. Use forward key to reach requested menu.
2. Press lock/unlock key to unlock setting.
3. Use select key to move cursor to / symbol.
4. Press change key to select visible ( ) or hidden ( ).
5. Press lock/unlock key to confirm new setting.

6.7 Changing language


It is possible to change language in transmitter. Default language is English, but it can be
changed to various other languages.
Change language as follows:
1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key or backward key to reach language menu.
4. Press lock/unlock key to unlock language.
5. Use change key to select desired language.
6. Press lock/unlock key to confirm new language.
7. Press top up key two times to exit setup mode.

SITRANS FM MAG 5000/6000 IP67


48 Operating Instructions, 12/2019, A5E02338368-AD
Commissioning
6.7 Changing language

/DQJXDJH
PRGH

/DQJXDJH
(QJOLVK

(QJOLVK
'HXWVFK
)UDQFDLV
'DQVN
6YHQVND
6XRPHNVL
5XVVLDQ
(VSD³RO
,WDOLDQR
3RUWXJX¬V
3ROVNL

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 49
Commissioning
6.7 Changing language

SITRANS FM MAG 5000/6000 IP67


50 Operating Instructions, 12/2019, A5E02338368-AD
Operating 7
This chapter describes the various menus of the transmitter in details. The menu diagrams are
shown in appendix Menu diagrams.

7.1 Output settings


Three outputs are available:
● Current output (range and time constant); terminals 31 and 32.
● Digital output (pulse, frequency, error, limit, or batch settings); terminals 56, 57, and 58.
● Relay output (error, limit, and batch settings); terminals 44, 45, and 46.

Current output
In the current output menu it is possible to select current output direction, range and time
constant, see also Current output (Page 83).
If current output "4-20 mA + Alarm" is selected, then alarm level and alarm differentiation may
also be defined.
"Alarm level" defines if an alarm should be above 21 mA "High" or below 3.6 mA "Low".
"Alarm diff." defines whether or not the alarm should vary according to selected error level.
Error level "Fatal". "Permanent" or "Warning" is selected in "Basic settings".
If Alarm differentiation is set to "Yes", depending on the Alarm level setting, the current output
will show:

Alarm level Output / Error level


Fatal Permanent Warning
Low 1.3 mA 2 mA 3 mA
High 23 mA 22 mA 21.5 mA

If Alarm differentiation is set to "No", depending on the Alarm level setting, the current output
will show:

Alarm level Output


Low 3.5 mA
High 22.6 mA

For setting of error level, see Digital output / Relay output - Error level (Page 84).
If current output is not used, it must be set to "Off".

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 51
Operating
7.1 Output settings

Digital output
Digital output can be used to configure various settings:
● Pulse (volume/pulse, pulse output, pulse width, pulse polarity, and time constant),
see Digital output - pulse (Page 83).
● Frequency (frequency output, max frequency, and time constant), see Digital output -
frequency (Page 84).
● Error settings (level and number), see Digital output / Relay output - Error level (Page 84)
and Digital output / Relay output - Error number (Page 84).
● Limit settings (number of setpoints, setpoint settings, and hysteresis), see Digital output /
Relay output - Direction/limit (Page 85).
● Batch settings (quantity, time and counter settings, and time constant), see Digital output /
Relay output - Batch (Page 85).
Note
Batch settings
Only MAG 6000.
Not available in MAG 5000, MAG 5000 CT and MAG 6000 CT.

Note
When relay is set to batch function, pulse/frequency is not available on digital output.

Relay outputs
Relay output can be used to configure various settings:
● Error settings (level and number), see Digital output / Relay output - Error level (Page 84)
and Digital output / Relay output - Error number (Page 84).
● Limit settings (number of setpoints, setpoint settings, and hysteresis), see Digital output /
Relay output - Direction/limit (Page 85).
● Batch settings (quantity, time and counter settings, and time constant), see Digital output /
Relay output - Batch (Page 85).
● Cleaning (cycle time), see Relay output - Cleaning (Page 85).
Note
Batch settings
Only MAG 6000.
Not available in MAG 5000, MAG 5000 CT and MAG 6000 CT.

Note
Cleaning
If a cleaning unit is installed together with transmitter, relay output must always be used to
operate this unit. It cannot be used for other purposes.

SITRANS FM MAG 5000/6000 IP67


52 Operating Instructions, 12/2019, A5E02338368-AD
Operating
7.4 Reset mode

7.2 External input


By applying 11 to 30 V DC to terminals 77 and 78, it is possible to perform:
● Batch control (start, stop, hold/continue)
● Reset totalizer
● Force/freeze output
● Qmax 2 (night)
See External input (Page 87).

Note
Batch settings
Only MAG 6000.
Not available in MAG 5000, MAG 5000 CT and MAG 6000 CT.

Note
Manual cleaning
If the digital input is used for manual cleaning, the relay output also automatically changes to
"cleaning".

7.3 Sensor characteristics


The sensor characteristics menu shows:
● If a SENSORPROM® is installed or not
● Suppress error P 40 (SENSORPROM® not installed)
● Sensor size
● Calibration factor
● Correction factor
● Excitation
See Sensor characteristics (Page 88).

Note
If a SENSORPROM is not installed, check the sensor characteristics to see if they match the
product label and the previous customer settings.

7.4 Reset mode


The reset mode is used for resetting/presetting totalizers or for restoring the factory settings.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 53
Operating
7.4 Reset mode

Resetting
1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key or backward key to reach reset mode menu.
4. Press lock/unlock key to enter reset menu.
5. Press forward key to reach totalizer to be reset or default setting menu.
6. Press lock/unlock key to start resetting.
If restoring of factory settings is required:
1. Press lock/unlock key again to confirm destruction of customized settings.
See Reset mode (Page 89)

Zero point adjustment (MAG 6000 SV only)


Auto adjustment
Before auto zero point adjustment is carried out ensure that valves to and from flowmeter are
completely closed and that flow velocity in sensor is zero.
1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key or backward key to reach reset mode menu.
4. Press lock/unlock key to enter reset menu.
5. Press forward key to reach zero adjust menu.
6. Press lock/unlock key to enter the menu.
7. Use change key to select "auto".
8. Press forward key to view actual offset (lower line in display). Value will be zero after
adjustment has been performed.
9. Press lock/unlock key to start adjustment.
Manual adjustment.
1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key or backward key to reach reset mode menu.
4. Press lock/unlock key to enter reset menu.
5. Press forward key to reach zero adjust menu.
6. Press lock/unlock key to enter the menu.
7. Use change key to select "manual".
8. Press forward key and then select key and change key to key in offset value.
9. Press lock/unlock key to start adjustment.

SITRANS FM MAG 5000/6000 IP67


54 Operating Instructions, 12/2019, A5E02338368-AD
Operating
7.6 MAG 5000 CT and MAG 6000 CT settings

Zero point can be adjusted manually in range -1.000 to +1.000 m3/s. If value outside this range
is keyed in, zero point adjustment will not be implemented.
See Reset mode - MAG 6000 SV (Page 90).

7.5 Service mode


All outputs of the transmitter can be forced-controlled in the service mode menu, see Service
mode (Page 91).
It is possible to check whether the outputs are functioning.
Error pending and status log lists are also accessible from this menu and the operating time (in
days) can be read.
The forced control is stopped and all previous settings are reinitialized the moment the service
mode is left by pressing top up key .

7.6 MAG 5000 CT and MAG 6000 CT settings

Internal totallizers
Depending on the type of approval it is possible to reset the internal totalizers. The type of
approval is selected in the reset menu with the hardware key mounted. It is possible to choose
between:
● Hot/cold water
● Other liquids
Resetting of totalizers by electrical input is not possible.
Hot/cold water
● Totalizer 1 is allocated to forward flow (cannot be reset)
● Totalizer 2 is allocated to reverse flow (cannot be reset)
Other liquids
Both totalizer 1 and totalizer 2 are allocated to measure the net flow, i.e. any reverse flow will
make the totalizers count backwards.
● Totalizer 1 cannot be reset.
● Totalizer 2 can be reset if the flow velocity in the meter pipe is <0.25 m/s. When the totalizer
is reset, the pulse output register will also be reset.

Output
● When choosing hot water, changing the output settings is not allowed and the output setting
menus are not shown in display.
● When choosing cold water or other liquids, all output settings can be changed.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 55
Operating
7.7 MAG 6000 SV

7.7 MAG 6000 SV

Excitation frequency
The MAG 6000 SV excitation frequency can be changed in Sensor characteristics (Page 88)
to one of the following frequencies:
● 1 9/16 Hz
● 3 1/8 Hz
● 6 ¼ Hz
● 12½ Hz
● 25 Hz
● 44 Hz
Note
Calibration has been made with the frequency stored in SENSORPROM® memory unit. A
change in excitation frequency is not recommended and will always mean decreased
measuring accuracy. In some instances, however, it may be necessary to change frequency
due to pulsating flow from piston pumps or other resonance frequencies from surroundings.

It is highly recommended to carry out a External input (Page 53) after changing the excitation
frequency as the offset is affected by the frequency selected. When this is done, the decrease
in measuring accuracy can be kept below 1% o.r.
A too high frequency for the sensor used will cause a coil current alarm indication.

SITRANS FM MAG 5000/6000 IP67


56 Operating Instructions, 12/2019, A5E02338368-AD
Service and maintenance 8
The device is maintenance-free. However, a periodic inspection according to pertinent
directives and regulations must be carried out.
An inspection can include check of:
● Ambient conditions
● Seal integrity of the process connections, cable entries, and cover screws
● Reliability of power supply, lightning protection, and grounds

NOTICE
Repair and service must be carried out by Siemens authorized personnel only.

Note
Siemens defines flow sensors as non-repairable products.

Under ideal conditions the flowmeter will operate continuously with no manual adjustment or
intervention required.
The SITRANS F M Verificator is an external tool developed for verifying the SITRANS F M
system, installation, and application. It is a highly advanced instrument, which carries out the
complex verification of the entire flowmeter system according to unique SIEMENS patented
principles. The verification test is automated and the instrument easy to use, so no human error
or influence will affect the verification.

8.1 Transmitter check list


If unstable/wrong measurements occur, it is often due to insufficient/wrong earthing or potential
equalization. If earthing connection is OK, check transmitter as described below, and sensor as
described in sensor check lists (see relevant sensor Operating Instructions).
The easiest way to check the transmitter in a SITRANS F M installation is to replace the
transmitter with another transmitter with a similar power supply.
As all settings are stored in and downloaded from the SENSORPROM®, replacement is easily
done and no extra settings need to be made.

Check transmitter
If no replacement transmitter is available, check transmitter according to the following check
table.

Power on transmitter
0 Display light on? Yes ⇒ 1
No ⇒ 2

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 57
Service and maintenance
8.2 Technical support

Power on transmitter
1 Flashing error triangles? Yes ⇒ Check error table
No ⇒ 1.2
1.2 Output and display readings OK? Yes ⇒ 1.2.1
No ⇒ 1.2.2
1.2.1 Transmitter OK Check application
Check installation/sensor/earthing connec‐
tion etc.
1.2.2 Check cables/connections OK ⇒ 1.2.1
Check connection board Not OK ⇒ correct fault
Check pins in transmitter multiplug
2 Check cables/connections OK ⇒ 2.1
Check connection board Not OK ⇒ Correct fault
Check pins in transmitter multiplug
2.1 Output readings OK? Yes ⇒ 2.1.1
No ⇒ 2.1.2.
2.1.1 Display defective Replace display
2.1.2 Transmitter defective Replace transmitter

Note
Sensor check list
Check list for sensors are included in the relevant sensor Operating Instructions.

8.2 Technical support


If you have any technical questions about the device described in these Operating Instructions
and do not find the right answers, you can contact Customer Support:
● Via the Internet using the Support Request:
Support request (http://www.siemens.com/automation/support-request)
● Via Phone:
– Europe: +49 (0)911 895 7222
– America: +1 423 262 5710
– Asia-Pacific: +86 10 6475 7575
Further information about our technical support is available on the Internet at
Technical support (http://support.automation.siemens.com/WW/view/en/16604318)

Service & Support on the Internet


In addition to our documentation, we offer a comprehensive knowledge base online on the
Internet at:
Service and support (http://www.siemens.com/automation/service&support)

SITRANS FM MAG 5000/6000 IP67


58 Operating Instructions, 12/2019, A5E02338368-AD
Service and maintenance
8.3 Return procedure

There you will find:


● The latest product information, FAQs, downloads, tips and tricks.
● Our newsletter, providing you with the latest information about your products.
● Our bulletin board, where users and specialists share their knowledge worldwide.
● You can find your local contact partner for Industry Automation and Drives Technologies in
our partner database.
● Information about field service, repairs, spare parts and lots more under Services.

Additional Support
If you have additional questions about the device, please contact your local Siemens
representative and offices at:
Local contact person (http://www.automation.siemens.com/partner)

8.3 Return procedure


Enclose the bill of lading, return document and decontamination certificate in a clear plastic
pouch and attach it firmly to the outside of the packaging.

Required forms
● Delivery note
● Return goods delivery note (http://www.siemens.com/processinstrumentation/
returngoodsnote)
with the following information:
– Product (item description)
– Number of returned devices/replacement parts
– Reason for returning the item(s)
● Decontanimation declaration (http://www.siemens.com/sc/declarationofdecontamination)
With this declaration you warrant "that the device/replacement part has been carefully
cleaned and is free of residues. The device/replacement part does not pose a hazard for
humans and the environment."
If the returned device/replacement part has come into contact with poisonous, corrosive,
flammable or water-contaminating substances, you must thoroughly clean and
decontaminate the device/replacement part before returning it in order to ensure that all
hollow areas are free from hazardous substances. Check the item after it has been cleaned.
Any devices/replacement parts returned without a decontamination declaration will be
cleaned at your expense before further processing.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 59
Service and maintenance
8.4 Recalibration

8.4 Recalibration
Siemens AG offers to recalibrate the sensor. Please use the recalibration MLFB
9LA110-8Qxxx-xxxx, where xxx-xxxx indicate customer-specific configuration.

Note
For recalibration the SENSORPROM® memory unit must always be returned with the sensor.

SITRANS FM MAG 5000/6000 IP67


60 Operating Instructions, 12/2019, A5E02338368-AD
Diagnostics and Troubleshooting 9
9.1 Diagnostics

Error system
Transmitter system is equipped with an error and status log system with 4 groups of
information.
(I) Information - system will continue to measure as normal, relay and current outputs will not
be affected.
(W) Warning - system will continue to measure, but an event that may cause a system
malfunction and require operator attention has occurred. The cause of the error may disappear
on its own.
(P) Permanent error - may cause malfunction in the application and operator attention is
required.
(F) Fatal error - is essential for the operation of the flowmeter. Immediate operator attention is
required.
Two menus are available in service and operator menus for registration of information and
errors.
● Error pending
● Status log
Note
Registration of errors in different modes
● In setup mode (local dialog) errors are entered only to Error pending list and not to Error
log list, and not registered on physical outputs (current or relay).
● In service mode errors are entered to both Error pending and Error log lists, but not
registered on physical outputs (current or relay).

Note
Power-off
Both error pending and status logs are reset at power-off.

Error pending
The first 9 pending errors are stored in the error pending list. When the error is corrected, it is
removed from the error pending list.
The acceptance level for "error pending" can be individually configured to a particular
application.
The acceptance level is set in the basic settings menu (Page 44).

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 61
Diagnostics and Troubleshooting
9.1 Diagnostics

Acceptance levels
The following three acceptance levels are selectable.
● Fatal error: Only fatal errors are registered as errors
● Permanent error: Permanent and fatal errors are registered as errors
● Warning (Default value): Warnings, permanent and fatal errors are registered as errors
Error information is displayed in title and subtitle lines, see display layout (Page 42). Title line
will show time in days, hours and minutes since occurrence of error. Subtitle line will flash
between an error text and a remedy text. Error text will indicate type of error (I, W, P or F), error
number, and error text. Remedy text will inform operator of action to take to remove error.

3HQGLQJ
(UURUWH[W

Status log
The latest 9 errors are stored in the status log. Errors are stored in the status log for 180 days,
even if they are corrected.

Alarm field
The alarm field on the display will always flash when an error is pending.

Error output
The digital and relay output can be activated individually error by error (error level). The relay
output is default selected to error level. An output can also be selected to activate on a single
error number.
The alarm field, error output and error pending always operate together.

Operator menu
Error pending and status log are as default enabled ( ) in the operator menu.

SITRANS FM MAG 5000/6000 IP67


62 Operating Instructions, 12/2019, A5E02338368-AD
Diagnostics and Troubleshooting
9.2 List of error numbers

9.2 List of error numbers

Error Error text Comment Output Input


No. Remedy text status status
1 I1 - Power on
OK Device powered on Active Active
2 I2 - Add-on module
Applied A new module has been applied to the system Active Active
3 I3 - Add-on module
Install An add-on module is defect or has been removed. Active Active
This can be an internal add-on module
4 I4 - Param. corrected
OK A less vital parameter in the transmitter has been re‐ Active Active
placed by its default value
20 W20 - Totalizer 1
Reset manually During initialisation the check of the saved totalizer Active Active
value has failed. It is not possible to rely on the saved
totalizer value anymore. The totalizer value must be
reset manually in order to rely on future readings
20 W20 - Totalizer 2
Reset manually During initialisation the check of the saved totalizer Active Active
value has failed. It is not possible to rely on the saved
totalizer value anymore. The totalizer value must be
reset manually in order to rely on future readings
21 W21 - Pulse overflow
Adj. pulse settings Actual flow is too big compared with pulse width and Reduced Active
volume/pulse pulse width
22 W22 - Batch timeout
Check installation Duration of batching has exceeded a predefined max‐ Batch output Active
imum time on zero
23 W23 - Batch overrun
Check installation Batch volume has exceeded a predefined maximum Batch output Active
overrun volume on zero
24 W24 - Batch neg. flow
Check flow direction Negative flow direction during batch Active Active
30 W30 - Overflow
Adj. Qmax Flow is above Qmax settings Max. 120 % Active
31 W31 - Empty pipe
Pipe is empty Zero Active
40 P40 - SENSORPROM®
Insert/change SENSORPROM® unit not installed Active Active
41 P41 - Parameter range
Switch off and on A parameter is out of range. The parameter could not Active Active
be replaced by its default value. The error will disap‐
pear at the next power-on

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 63
Diagnostics and Troubleshooting
9.2 List of error numbers

Error Error text Comment Output Input


No. Remedy text status status
42 P42 - Current output
Check cables Current loop is disconnected or the loop resistance is Active Active
too big
43 P43 - Internal error
Switch off and on Too many errors occurred at the same time. Active Active
Some errors are not detected correctly
44 P44 - CT SENSORPROM®
Replace SENSORPROM® unit has been used as CT version Active Active
49 P49 - Protection violation Internal protection of the device has been violated. Active
Switch off and on
60 F60 - CAN comm. error
Transmitter/AOM CAN bus communication error. An add-on module, Zero Inactive
the display module or the transmitter is defective
61 F61 - SENSORPROM® error
Replace It is not possible to rely on the data in SENSOR‐ Active Active
PROM® unit anymore
62 F62 - SENSORPROM® ID
Replace The SENSORPROM® unit ID does not comply with Zero Inactive
the product ID.
The SENSORPROM® unit is from another type of
product SITRANS F C, SITRANS F US etc.
63 F63 - SENSORPROM®
Replace It is not possible to read from the SENSORPROM® Active Active
unit anymore
70 F70 - Coil current
Check cables Coil excitation has failed Active Active
71 F71 - Internal error
Replace transmitter Internal conversion error in ASIC Active Active

Symptom Output sig‐ Error code Cause Remedy


nals
Empty display Minimum 1. No power supply Power supply
Check MAG 5000/6000/6000 I for ben‐
ded pins on the connector
2. MAG 5000/6000/6000 I de‐ Replace MAG 5000/6000/6000 I
fective

SITRANS FM MAG 5000/6000 IP67


64 Operating Instructions, 12/2019, A5E02338368-AD
Diagnostics and Troubleshooting
9.2 List of error numbers

Symptom Output sig‐ Error code Cause Remedy


nals
No flow signal Minimum 1. Current output disabled Turn on current output
2. Digital output disabled Turn on digital output
3. Reverse flow direction Change direction
F70 Incorrect or no coil current Check cables/connections
W31 Measuring pipe empty Ensure that the measuring pipe is full
F60 Internal error Replace MAG 5000/6000/6000 I
Undefined P42 1. No load on current output Check cables/connections
2. MAG 5000/6000/6000 I de‐ Replace MAG 5000/6000/6000 I
fective
P41 Initializing error Switch off MAG 5000/6000/6000 I,
wait 5 sec. and switch on again
Indicates flow with no Undefined Measuring pipe empty Select empty pipe cut-off
flow in pipe Empty pipe cut-off is OFF Ensure that the measuring pipe is full
Electrode connection missing/ Ensure that electrode cable is connec‐
electrode cable is insufficiently ted and sufficiently screened
screened
Unstable flow signal Unstable 1. Pulsating flow Increase time constant
2. Conductivity of medium too Use special electrode cable
low
3. Electrical noise potential be‐ Ensure sufficient potential equalization
tween medium and sensor
4. Air bubbles in medium Ensure medium does not contain air
bubbles
5. High concentration of parti‐ Increase time constant
cles or fibres
Measuring error Undefined Incorrect installation Check installation
P40 No SENSORPROM unit ®
Install SENSORPROM® unit
P44 CT SENSORPROM® unit Replace SENSORPROM® unit or re‐
set SENSORPROM® unit with MAG
CT transmitter
P49 Protection violation Switch off MAG 5000/6000/6000I, wait
5 sec. and switch on again.
F61 Defective SENSORPROM® unit Replace SENSORPROM® unit
F62 Wrong type of SENSOR‐ Replace SENSORPROM® unit
PROM® unit
F63 Defective SENSORPROM® unit Replace SENSORPROM® unit
F71 Loss of internal data Replace MAG 5000/6000/6000 I
Maximum W30 Flow exceeds 100% of Qmax. Check Qmax (Basic Settings)
W21 Pulse overflow
Volume/pulse too small Change volume/pulse
Pulse width too large Change pulse width
Measuring approx. Missing one electrode connec‐ Check cables
50% tion
Loss of totalizer data OK W20 Initializing error Reset totalizer manually

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 65
Diagnostics and Troubleshooting
9.2 List of error numbers

Symptom Output sig‐ Error code Cause Remedy


nals
##### Signs in display OK Totalizer roll over Reset totalizer or increase totalizer unit
Empty pipe error mes‐ OK W31 Empty pipe error Switch off MAG 5000/6000/6000I, wait
sage when Empty 5 sec. and switch on again
pipe set to off

SITRANS FM MAG 5000/6000 IP67


66 Operating Instructions, 12/2019, A5E02338368-AD
Technical data 10
10.1 Technical specifications

Technical specifications for MAG 5000/6000


Mode of operation and design Measuring principle Electromagnetic with pulsed constant field
Empty pipe Detection of empty pipe (special cable required in remote
mounted installation)
Excitation frequency Depends on sensor size
Electrode input impedance > 1 x 1014 Ω
Input Digital input 11 to 30 V DC, Ri = 4.4 KΩ
Activation time 50 ms
Current IDC 11 V = 2.5 mA, IDC 30 V = 7 mA
Output Current output
Signal range 0 to 20 mA or 4 to 20 mA, Alarm
Load < 800 Ω
Time constant 0.1 to 30 s, adjustable (for batch: fixed at 0.1 s)
Digital output
Frequency 0 to 10 kHz, 50% duty cycle (uni/bidirectional)
Pulse (active) DC 24 V, 30 mA, 1 kΩ ≤ Ri ≤10 kΩ, short-circuit protected
(power supplied from flowmeter)
Pulse (passive) DC 3 to 30 V, max. 110 mA, 200 Ω ≤ Ri ≤10 kΩ (powered
from connected equipment)
Time constant 0.1 to 30 s, adjustable (for batch: fixed at 0.1 s)
Relay output
Time constant Changeover relay, same as current output
Load 42 V AC/2 A, 24 V DC/1 A
Low flow cut off 0 to 9.9% of maximum flow
Galvanic isolation All inputs and outputs are galvanically isolated
Max. measuring error MAG 5000 0.4% ± 1 mm/s (for v > 0.1 m/s)
(incl. sensor and zero point) MAG 6000 0.2% ± 1 mm/s (for v > 0.1 m/s)
Functions Flow rate, 2 totalizers, low-flow cut-off, empty pipe cut-off, flow direction, error system, oper‐
ating time, uni/bidirectional flow, limit switches, pulse output, control for cleaning and batch

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 67
Technical data
10.1 Technical specifications

Technical specifications for MAG 5000/6000


Rated operation conditions Ambient temperature
Operation Standard IP67, 19", blind and SV versions:
-20 to +60 °C (-4 to +140 °F)
CT version:
-25 to +55 °C (-13 to +131 °F)
Storage -40 to +70 °C (-40 to +158 °F)
Mechanical load 18 to 1000 Hz, 3.17 G rms, sinusoidal in all directions to IEC 68-2-36
Degree of protection IP67/NEMA 4X/6 to IEC 529 and DIN 40050 (1 mH2O 30 min.)
EMC performance EN 61326-1 (industrial environments)
EN 61326-2-5
Display and keypad Totalizer Two eight-digit counters for forward, net or reverse flow
Display Background illumination with alphanumeric text, 3 x 20 char‐
acters to indicate flow rate, totalized values, settings and
faults;
Reverse flow indicated by negative sign
Time constant Time constant as current output time constant
Design Enclosure material Fiber glass reinforced polyamide;
optional (IP67 only): AISI 316 stainless steel
Dimensions See dimensional drawings
Weight 0.75 kg (2 lb)
Power supply 115 to 230 V AC +10% -15%, 50 to 60 Hz, Fuse: 500 mA T
11 to 30 V DC or 11 to 24 V AC; Fuse 2 A T
Power consumption 115 to 230 V AC: 17 VA
24 V AC: 9 VA, IN = 380 mA, IST = 8 A (30 ms)
12 V DC: 11 W, IN = 920 mA, IST = 4 A (250 ms)
Certificates and approvals CE, C-UL US general purpose, C-tick, CSA/FM Class 1, div 2
Custody transfer approval Cold water approval: PTB K7.2, MI-001
(MAG 5000/6000 CT)
Communication MAG 5000 Without communication or HART as option
MAG 6000 / MAG 6000 CT Prepared for client mounted add-on modules:
HART, MODBUS RTU/RS485, FOUNDATION Fieldbus H1,
DeviceNet, PROFIBUS PA, PROFIBUS DP as add-on mod‐
ules

SITRANS FM MAG 5000/6000 IP67


68 Operating Instructions, 12/2019, A5E02338368-AD
Technical data
10.1 Technical specifications

Transmitter IP67/NEMA 4X/6 compact polyamide

Weight: MAG 5000/6000: 0.75 kg (1.65 lbs)

Transmitter IP67/NEMA 4X/6 wall-mounted polyamide

Weight (transmitter and wall mounting bracket): 1.65 kg (3.64 lbs)

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 69
Technical data
10.2 Accuracy

10.2 Accuracy
For accuracy reference conditions, see table Reference conditions for sensor calibration.

[±% E] )ORZPHWHUXQFHUWDLQW\
3,0
2,8
V: Flow velocity
2,6 E: Meter uncertainty as a percentage of measured value
2,4 v ≥ 0.1 m/s (0.3 ft/s) --> E: ± 0.4 ± 1 mm/s *
v < 0.1 m/s (0.3 ft/s) --> E: ± (0.25/v) % of measured value
2,2
2,0 * MAG 5100 W with DN 350 ... 1200
1,8 v ≥ 0.1 m/s (0.3 ft/s) --> E: ± 0.4 ± 2 mm/s
v < 0.1 m/s (0.3 ft/s) --> E: ± (0.25/v) % of measured value
1,6
1,4 * MAG 1100 with PFA liner DN 10 and DN 40
v ≥ 0.1 m/s (0.3 ft/s) --> E: ± 0.4 ± 2 mm/s
1,2 v < 0.1 m/s (0.3 ft/s) --> E: ± (0.25/v) % of measured value
1,0
0,8
0,6
0,4
0,2
0,0
0 1 2 3 4 5 6 7 8 9 10 11 12 [m/s]
0 3.3 6.6 9.9 13 16 20 23 26 30 33 36 39 [ft/s]

Figure 10-1 MAG 5000 with MAG 1100, MAG 1100 F, MAG 5100 W, MAG 3100 and MAG 3100 P and
MAG 6000 with MAG 1100 (PFA), MAG 1100 F (PFA)

[±% E] )ORZPHWHUXQFHUWDLQW\

1,4
V: Flow velocity
E: Meter uncertainty as a percentage of measured value
1,2 v ≥ 0.1 m/s (0.3 ft/s) --> E: ± 0.2 ± 1 mm/s *
v < 0.1 m/s (0.3 ft/s) --> E: ± (0.125/v) % of measured value *

1,0 * MAG 5100 W with DN 350 ... 1200


v ≥ 0.1 m/s (0.3 ft/s) --> E: ± 0.2 ± 2.5 mm/s
v < 0.1 m/s (0.3 ft/s) --> E: ± (0.25/v) % of measured value
0,8
* MAG 1100 and MAG 1100 F with ceramic liner DN 10 and DN 40
v ≥ 0.1 m/s (0.3 ft/s) --> E: ± 0.2 ± 2 mm/s
0,6 v < 0.1 m/s (0.3 ft/s) --> E: ± (0.125/v) % of measured value

0,4

0,2

0,0
0 1 2 3 4 5 6 7 8 9 10 11 12 [m/s]
0 3.3 6.6 9.9 13 16 20 23 26 30 33 36 39 [ft/s]

Figure 10-2 MAG 6000 with MAG 1100 (not PFA), MAG 1100 F (not PFA), MAG 5100 W, MAG 3100
and MAG 3100 P

SITRANS FM MAG 5000/6000 IP67


70 Operating Instructions, 12/2019, A5E02338368-AD
Technical data
10.2 Accuracy

[±% E] 5HIHUHQFH'1

15,0

10,0

5,0

0,0

-5,0

-10,0

-15,0

0,0
0,01 0,1 1 10 [m/s]
0,03 0,3 3,3 32,8 [ft/s]
Flow %Error MAG 6000
"R" (Q3/Q1) 63
"R" (Q3/Q1) 40

Figure 10-3 MID limits

Note
Uncertainties for MAG 5000/6000 CT systems under MID requirements
± 5% for flowrate between Q1 and Q2
± 2% for flowrate between Q2 and Q4 if media temperature ≤ 30°C
More detailed installation conditions can be found in the certificate.

Reference conditions
(ISO 9104 and DIN/EN 29104)
A calibration certificate is shipped with every sensor and calibration data is stored in
SENSORPROM memory unit.

Table 10-1 Reference conditions for sensor calibration

Medium temperature 20°C ± 5°C (68°F ± 9°C)


Ambient temperature 20°C ± 5°C (68°F ± 9°C)
Supply voltage Un ± 1%
Warming-up time 30 minutes

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 71
Technical data
10.3 Output characteristics

Incorporation in conductive pipe section


Inlet section 10 x DN (DN ≤ 1200/48")
5 x DN (DN > 1200/48")
Outlet section 5 x DN (DN ≤ 1200/48")
3 x DN (DN > 1200/48")
Flow conditions Developed flow profile

Table 10-2 Additions in the event of deviations from reference conditions

Current output As pulse output ± (0.1% of actual flow + 0.05% FSO)


Effect of ambient temperature
Display/frequency/pulse output < ± 0.003% / °C act.
Current output < ± 0.005% / °C act.
Effect of supply voltage < 0.005% of measuring value on 1% change
Repeatability ± 0.1% of actual flow for V ≥ 0.5 m/s (1.5 ft/s) and con‐
ductivity ≥ 10 μS/cm

10.3 Output characteristics

Current output

Output characteristics Bidirectional mode Unidirectional mode


0 to 20 mA P$ P$

 
*

*


&XW 4 &XW 4
RII RII

4 to 20 mA P$ P$

 

*
*

&XW 4 &XW 4
RII RII

Frequency output

Bidirectional mode Unidirectional mode


)>+]@ )>+]@

 
*
*

)PD[ )PD[

&XW 4 &XW 4
RII RII

SITRANS FM MAG 5000/6000 IP67


72 Operating Instructions, 12/2019, A5E02338368-AD
Technical data
10.3 Output characteristics

Pulse output

Bidirectional mode Unidirectional mode


([WHUQDO
FRXQWHU

&XW 4
RII

Relay output

Bidirectional mode Unidirectional mode


Power down Active

Error relay output

Bidirectional mode Unidirectional mode


No error Error

Limit switch (can be used as direction switch)

1 set point 2 set points

*E
*D
+\VWHUHVLV

+\VWHUHVLV

+\VWHUHVLV

VHWSRLQW VHWSRLQW

/LPLWVHWSRLQW /LPLWVHWSRLQW /LPLWVHWSRLQW


PLQLPXP PD[LPXP

Passive Digital output Passive Digital output

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 73
Technical data
10.3 Output characteristics

1 set point 2 set points



&RQQHFWHG

 &RQQHFWHG


+\VWHUHVLV

+\VWHUHVLV
VHWSRLQW

+\VWHUHVLV
VHWSRLQW


'LVFRQQHFWHG


*
 'LVFRQQHFWHG

*
/LPLWVHWSRLQW 
/LPLWVHWSRLQW /LPLWVHWSRLQW
PLQLPXP PD[LPXP

 
&RQQHFWHG &RQQHFWHG
 

+\VWHUHVLV

+\VWHUHVLV
+\VWHUHVLV
VHWSRLQW VHWSRLQW

 
'LVFRQQHFWHG 'LVFRQQHFWHG

*

*
 
/LPLWVHWSRLQW /LPLWVHWSRLQW /LPLWVHWSRLQW
PLQLPXP PD[LPXP

Limit/direction
'LJLWDORXWSXW
5HOD\RXWSXW

'LUHFWLRQOLPLW
/LPLWPRGH /LPLWVHWSRLQW

VHWSRLQW
(UURUOHYHO
(UURUQXPEHU VHWSRLQW
'LUHFWLRQOLPLW VHWSRLQWV

2II

/LPLWVHWSRLQW /LPLWVHWSRLQW +\VWHUHVLV


PLQLPXP PD[LPXP

Limit switches are available for both digital and relay outputs.
Direction mode: 1 set point at 0% flow; hysteresis 5%.
If 2 set points must activate 2 separate outputs, a single set point has to be selected individually
for digital as well as relay outputs.

Batch on digital output

Unidirectional mode (forward flow only)


&RQQHFWHG 
*


&RQWLQXH
%DWFK

+ROG


'LVFRQQHFWHG


 4 W

SITRANS FM MAG 5000/6000 IP67


74 Operating Instructions, 12/2019, A5E02338368-AD
Technical data
10.4 Cable data

Totalizer lap

/DSYDOXH

'LVSOD\HGWRWDOL]HUYDOXH

*
/DSSHULRGH

Batch on relay output

Unidirectional mode (forward flow only)


Hold Batch

10.4 Cable data

Description

Electrode or coil cable (standard)

Electrode cable, double shielded (for empty pipe detection or low


conductivity fluids)
Cable kit with standard coil cable and electrode cable double
shielded (also available as low noise cable for MAG 1100 sensor)

Technical data

Standard cable Double-shielded cable


(electrode/coil) (electrode)
Basic data No. of conductors 3 3
Sqr. area 1.5 mm2 0.25 mm2
Screen Yes Double
Color code Brown, blue, black Brown, blue, black
Outside color Grey Grey
Ext. diameter 7.8 mm 8.1 mm
Conductor Flexible CU Flexible CU
Isolation material PVC PVC

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 75
Technical data
10.5 Cable requirements

Standard cable Double-shielded cable


(electrode/coil) (electrode)
Ambient temperature Flexible installation -5 to +70°C -5 to +70°C
(23 to 158°F) (23 to 158°F)
Non-flexible installation -30 to +70°C -30 to +70°C
(-22 to 158°F) (-22 to 158°F)
Cable parameter Capacity 161.50 pF/m -
Inductance 0.583 μH/m -
L/R 43.83 µH/Ω -

For more information on cable lengths, empty pipe detection, and conductivity, see Operating
Instructions for relevant sensor.

10.5 Cable requirements

Coil cable Electrode cable


Basic data No. of conductors 2 3
Min. sqr. area 0.5 mm 2
0.2 mm2
Screen Yes Yes
Max. capacitance N/A 350 pF/m
Max. cable loop resistance Media temperature:
< 100 °C 40 Ω N/A
> 200 °C 6Ω N/A
Cable glands on sensor M20x1.5 gland - Cable ø 5 to 13 mm (0.20 to 0.51 inches)
½ NPT gland - cable ø 5 to 9 mm (0.20 to 0.35 inches)

WARNING
Cable glands
For Ex zone 1 installations only certified cable glands with protection type "e" can be used for
the power supply and the coil cable. The cable glands must be approved for the actual
temperature and cable dimension.

SITRANS FM MAG 5000/6000 IP67


76 Operating Instructions, 12/2019, A5E02338368-AD
Spare parts/Accessories A
A.1 Ordering of spare parts
Ensure that your ordering data is not outdated. The latest ordering data is always available on
the Internet: SIOS catalog (https://support.industry.siemens.com/cs/products?
dtp=Catalog&mfn=ps&pnid=17318&lc=en-US)

A.2 Spare parts

Description
Connection plate

SENSORPROM® memory unit

Display unit

Communication modules for MAG 6000

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 77
Spare parts/Accessories
A.3 Sun shield

A.3 Sun shield

Description
Sun shield

SITRANS FM MAG 5000/6000 IP67


78 Operating Instructions, 12/2019, A5E02338368-AD
Menu diagrams B
B.1 Transmitter menu overview
The menu diagrams shown on the following pages apply to MAG 5000/6000 as well as MAG
6000 I.

0HQXKLGGHQIRU 0HQXORFNHGIRU
VRPH 0$*
FRQILJXUDWLRQV &7

VHF

)ORZUDWH 3DVVZRUG %DVLF 2XWSXW ([WHUQDO 6HQVRU


 VHWWLQJV LQSXW FKDUDFWHULVWLFV

6(16253520
&XUUUHQWRXWSXW 6WDUWEDWFK
0DLQIUHTXHQF\ LQVWDOOHG
QRWLQVWDOOHG

)ORZGLUHFWLRQ 6XSSUHVV3
'LJLWDORXWSXW +ROGFRQWLQXH
6HQVRUSURP

&XVWRPHUXQLWV 6WRSEDWFK 6HQVRUVL]H


5HOD\RXWSXW

4PD[ 7RWDOLO]HU &DOIDFWRU


UHVHW

7RWDOL]HU &RUUHFWLRQ
7RWDOL]HUODS
IDFWRU

([FLWDWLRQ
/RZIORZFXWRII )RUFHRXWSXW IUHTXHQF\

(PSW\SLSH
)UHH]HRXWSXW

9HORFLW\XQLW 4PD[
1LJKW

(UURUOHYHO 0DQXDO
FOHDQLQJ

2II

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 79
Menu diagrams
B.2 Basic settings

5HVHW 6HUYLFH 2SHUDWRU 3URGXFW 3DVVZRUG /DQJXDJH


PRGH PRGH PHQXVHWXS LGHQWLW\ PRGH

8SSHUOLQH
&7UHVHW )ORZUDWH &KDQJH
)ORZYHORFLW\ &RQYHUWHU (QJOLVK
PRGH WRWDOL]HU SDVVZRUG
WRWDOL]HU
7RWDOL]HU /RZIORZ 7H[W 7UDQVPLWWHU 'HXWVFK
UHVHW FXWRII 7RWDOL]HU FRGHQR

7RWDOL]HU ([FLWDWLRQ 7H[W 7UDQVPLWWHU


RQRII 7RWDOL]HU )UDQFDLV
UHVHW VHULDOQR

%DWFKF\FOH 6LJQDOVXLWDELOLW\ 7H[W 7UDQVPLWWHU


'DQVN
FRXQWHUUHVHW OHYHO 4PD[ 6:YHUVLRQ

6HWGHIDXOW &XUUHQWRXWSXW 7H[W 6HQVRUW\SH 6YHQVND


VHWWLQJ QRUPDOIRUFHG )ORZUDWH

7RWDOL]HU 'LJLWDORXWSXW 7H[W 6HQVRU 6XRPHNVL


SUHVHW QRUPDOIRUFHG )ORZYHORFLW\ FRGHQR

7RWDOL]HU 5HOD\ 7RWDOL]HU 6HQVRU


VHULDOQR 5XVVLDQ
SUHVHW QRUPDOIRUFHG UHVHW

2SHUDWLRQWLPH 7RWDOL]HU 6HQVRUVL]H (VSDQ´O


UHVHW

(UURUSHQGLQJ %DWFK 7DJQR ,WDOLDQR


TXDQWLW\

6WDWXVORJ %DWFKVWDUW 3RUWXJXHV


SDXVHGUHVHW

'LVSOD\WHVW %DWFKDPRXQW
3ROVNL
GRQH

7H[W
2SHUDWLQJWLPH

%DWFKF\FOHFQW
UHVHW

7H[W
6HQVRUVL]H

7H[W
PHQX

(UURUSHQGLQJ

6WDWXVORJ

B.2 Basic settings

Note
Locked or hidden menus
Depending on your configuration some menus might be locked or hidden.

SITRANS FM MAG 5000/6000 IP67


80 Operating Instructions, 12/2019, A5E02338368-AD
Menu diagrams
B.2 Basic settings

%DVLF
VHWWLQJV

0DLQIUHTXHQF\

+]
+]

)ORZGLUHFWLRQ
3RVLWLYH
3RVLWLYH
1HJDWLYH

&XVWRPHUXQLWV &XVWYROXPHXQLW &XVWYROXPHXQLW


WH[W" HTXDO\\\\\\\P

&XVWWLPHXQLW &XVWWLPHXQLW
4PD[
PV− WH[W" HTXDO\\\\\\\\V
PV−
P V 36
PO 0LQ 30
O K 3+
KO G 3'
0O " "
IW
LQ
86*
86N*
%%/
%%/
8.*
8.0*
%%/

4PD[ QLJKW
PV−
PV−
P V 36
PO 0LQ 30
O K 3+
KO G 3'
0O " "
IW
LQ
86*
86N*
%%/
%%/
8.*
8.0*
%%/

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 81
Menu diagrams
B.2 Basic settings


7RWDOL]HU 7RWDOL]HU 7RWDOL]HU

/RZIORZ 7RWDOL]HUXQLW 7RWDOL]HUXQLW


FXW−RII [[[[[[, [[[[[[,
−  P P
PO PO
O O
KO KO
0O 0O
IW IW
LQ LQ
(PSW\SLSHFXW−RII 86* 86*
RII 86N* 86N*
%%/ %%/
RII
RQ %%/ %%/
8.* 8.*
8.0* 8.0*
%%/ "

9HORFLW\XQLW
\\\\\\PV 7RWDOL]HU 7RWDOL]HU
)RUZDUG 5HYHUVH
P V )RUZDUG )RUZDUG
PP PLQ 5HYHUVH 5HYHUVH
FP K 1HW 1HW
GP G
IW "
LQ

(UURUOHYHO
:DUQLQJ
)DWDO
3HUPDQHQW
:DUQLQJ

1)
When batch is selected on digital output or relay, Totalizer 2 is not shown because it is
controlled by the batch function.

SITRANS FM MAG 5000/6000 IP67


82 Operating Instructions, 12/2019, A5E02338368-AD
Menu diagrams
B.4 Digital output - pulse

B.3 Current output

*
Current output Current output Current output
Off Unidirectional
On Unidirectional
Off Bidirectional

Current output
Alarm 

Current output Current output
Alarm level: Low Alarm diff: No
Alarm 
Low No
High Yes

Current output
Time const.

1) If HART communication is installed, it is not possible to set the output for 0-20 mA (even though
the option is visible in the display). This is due to the fact that HART does not work if the output
falls below 2-3 mA.
2) 4-20 mA + Alarm is the default setting for MAG 6000 I. For all other variants, the default setting
is 4-20 mA.
3) For MAG 6000 I only: The controlling of alarm levels does not recognize if the jumper is mounted
for passive output. Do not combine differentiation and low alarm level together with passive
output. The output will try to pull down the level to 1.3 mA at fatal errors which is not possible for
passive output.

B.4 Digital output - pulse


*

'LJLWDORXWSXW 9ROXPHSXOVH 3XOVHRXWSXW 3XOVHZLGWK


3XOVH 8QLGLUHFWLRQDO
3XOVH P 8QLGLUHFWLRQDO
)UHTXHQF\ PO %LGLUHFWLRQDO
 O
2II KO
0O
IW
LQ
86*
86N*
%%/
%%/
8.*
8.0*
"

3XOVHRXWSXW 3XOVHSRODULW\
WLPHFRQVW 3RVLWLYH
3RVLWLYH
1HJDWLYH

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 83
Menu diagrams
B.7 Digital output / Relay output - Error number

B.5 Digital output - frequency


'LJLWDORXWSXW )UHTXHQF\RXWSXW )UHTXHQF\RXWSXW )UHTXHQF\RXWSXW
)UHTXHQF\ 8QLGLUHFWLRQDO )PD[ 7LPHFRQVW
3XOVH 8QLGLUHFWLRQDO
)UHTXHQF\ %LGLUHFWLRQDO
(UURUOHYHO
(UURUQXPEHU
'LUHFWLRQOLPLW
%DWFK
2II

B.6 Digital output / Relay output - Error level


'LJLWDORXWSXW
5HOD\RXWSXW

(UURUOHYHO
(UURUOHYHO
(UURUQXPEHU
'LUHFWLRQ/LPLW

2II

B.7 Digital output / Relay output - Error number


'LJLWDORXWSXW
5HOD\RXWSXW
(UURUQXPEHU
(UURUQXPEHU
(UURUOHYHO
(UURUQXPEHU
'LUHFWLRQOLPLW

2II

SITRANS FM MAG 5000/6000 IP67


84 Operating Instructions, 12/2019, A5E02338368-AD
Menu diagrams
B.10 Digital output / Relay output - Batch

B.8 Digital output / Relay output - Direction/limit


'LJLWDORXWSXW
5HOD\RXWSXW

'LUHFWLRQOLPLW
/LPLWPRGH /LPLWVHWSRLQW

VHWSRLQW
(UURUOHYHO
(UURUQXPEHU VHWSRLQW
'LUHFWLRQOLPLW VHWSRLQWV

2II

/LPLWVHWSRLQW /LPLWVHWSRLQW +\VWHUHVLV


PLQLPXP PD[LPXP

B.9 Relay output - Cleaning

5HOD\RXWSXW &OHDQLQJ
&OHDQLQJ F\FOHWLPH
(UURUOHYHO
(UURUQXPEHU
'LUHFWLRQ/LPLW
&OHDQLQJ
2II

Note
Relay outputs
If cleaning unit is installed, relay outputs must always be used to operate cleaning.
Relay outputs cannot be used for other purposes

B.10 Digital output / Relay output - Batch

Note
This menu does not exist on MAG 5000/6000 CT

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 85
Menu diagrams
B.10 Digital output / Relay output - Batch

5HOD\RXWSXW
'LJLWDORXWSXW

*
%DWFKTXDQWLW\ %DWFKFRPS %DWFKWLPHHUURU
%DWFK
2II
3XOVH P 2II
)UHTXHQF\ PO 2Q
(UURUOHYHO O
(UURUQXPEHU KO
'LUHFWLRQOLPLW 0O
IW
LQ
2II
86*
86N*
%%/
%%/
8.*
8.0*
"

0D[EDWFKWLPH %DWFKRYHUUXQHUU %DWFKRYHUUXQHUU %DWFKFRXQWHU


K 2II 'RZQ
− V 2II 8S
− 2Q 'RZQ
PLQ
− K
− G

1) Visible only on Digital output.

Note
If batch function is chosen on the relay output, the digital output will be turned off if it has been
set up for pulse, frequency or batch.
If digital output is set up for pulse, frequency or batch, then the relay output will be turned off if
it has been set up for batch.

SITRANS FM MAG 5000/6000 IP67


86 Operating Instructions, 12/2019, A5E02338368-AD
Menu diagrams
B.11 External input

B.11 External input

*
([WHUQDOLQSXW

([WHUQDOLQSXW
7RWDOL]HUUHVHW
2II
7RWDOL]HU
6WDUWEDWFK 7RWDOL]HU
+ROGFRQWLQXH 7RWDOL]HU
6WRSEDWFK 7RWDOL]HU
7RWDOL]HUUHVHW
7RWDOL]HUODS
)RUFHRXWSXW
)UHH]HRXWSXW
4PD[ QLJKW )RUFHRXWSXW
0DQXDOFOHDQLQJ
2II

1) The value showing totalizer 1 on the display is frozen for as long as the digital output is activated.
However, totalizer 1 continues counting, and when the digital input is released, the value on the
display again follows totalizer 1.
2) MAG 6000 I cannot be equipped with cleaning unit. The cleaning option for relay output is
however possible. When selecting function for MAG 6000 I relay output, the relay output has the
same behavior as if cleaning unit was installed.
The relay output of the transmitter determines when the relay is on by applying voltage for
approximately 60 seconds. The metering is resumed after another 60 seconds when the cycle is
complete. (The display is locked during this time). The time cycle can be set at 1 to 240 hours.
If the cycle is set at for example three hours, the transmitter will be active every three hours.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 87
Menu diagrams
B.12 Sensor characteristics

B.12 Sensor characteristics

*
6HQVRU
FKDUDFWHULVWLFV

6HQVRUSURP
,QVWDOOHG

,QVWDOOHG
1RWLQVWDOOHG

6XSSUHVV3
6(162535201R

<HV
1R

6HQVRUVL]H

&DOLEUDWLRQIDFWRU

&RUUHFWLRQIDFWRU

([FLWDWLRQ




1) Error status (level or number) on an output is updated only at the time the error status changes
(occurs or disappears). If P40 is suppressed after it has been detected (at power up), the output
does not change state. In this case the power must be switched off/on to suppress the P40 error
on the output.
2) The frequency can be set to 44 Hz in the MAG 6000 SV transmitter only.

SITRANS FM MAG 5000/6000 IP67


88 Operating Instructions, 12/2019, A5E02338368-AD
Menu diagrams
B.13 Reset mode

B.13 Reset mode

5HVHW
PRGH

7RWDOL]HUUHVHW

7RWDOL]HUUHVHW

%DWFKF\FOHFQW
UHVHW

6HWGHIDXOW 'HVWUR\RZQ
VHWWLQJ VHWWLQJV"

7RWDOL]HUSUHVHW

7RWDOL]HUSUHVHW

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 89
Menu diagrams
B.14 Reset mode - MAG 6000 SV

B.14 Reset mode - MAG 6000 SV

5HVHW
PRGH

7RWDOL]HUUHVHW

7RWDOL]HUUHVHW

%DWFKF\FOHFQW
UHVHW

=HURDGMXVW =HURDGMXVW =HURDGMXVW


=HURDGMXVW
EXV\
$XWR
ILQLVKHG

$XWR
0DQXDO

=HURDGMXVW

6HWGHIDXOW 'HVWUR\RZQ 6HWGHIDXOW 6HWGHIDXOW


VHWWLQJ VHWWLQJV" EXV\ ILQLVKHG

7RWDOL]HU
SUHVHW

7RWDOL]HU
SUHVHW

SITRANS FM MAG 5000/6000 IP67


90 Operating Instructions, 12/2019, A5E02338368-AD
Menu diagrams
B.15 Service mode

B.15 Service mode

*
6HUYLFH
PRGH

)ORZYHORFLW\
\\\\\\PV

/RZIORZ
&XWRII


([FLWDWLRQ
2Q
2Q
2II

6LJQDOVXLWDELOLW\
/HYHO\

&XUUHQWRXWSXW &XUUHQWRXWSXW &XUUHQWRXWSXW


1RUPDO )RUFHGP$
1RUPDO P$
)RUFHG

'LJLWDORXWSXW 'LJLWDORXWSXW 'LJLWDORXWSXW


1RUPDO )RUFHG+LJK
1RUPDO +LJK
)RUFHG /RZ


'LJLWDORXWSXW
)RUFHG+]
+]

5HOD\RXWSXW 5HOD\RXWSXW
1RUPDO
1RUPDO
)RUFHGRQ
)RUFHGRII

2SHUDWLQJ
WLPH[[[[GD\V

(UURUSHQGLQJ 3HQGLQJ[[[G[[K[[P

3HQGLQJ

6WDWXVORJ /RJ[[[G[[K[[P

/RJ

1) Signal suitability is a level from 0 to 9 of the electrode measured voltage. Level 0 is equal to the
limit value that is set for empty pipe error detection, and level 9 is the best signal measured.
2) If digital output is set to pulse (standard).
3) If digital output is set to frequency.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 91
Menu diagrams
B.16 Operator menu setup

B.16 Operator menu setup

*
2SHUDWRUPHQX

8SSHUOLQH 7H[W
)ORZUDWH )ORZYHORFLW\
)ORZUDWH 7H[W
7RWDOL]HU )ORZUDWH
7RWDOL]HU )ORZYHORFLW\
%DWFKDPRXQWGRQH 4PD[
4PD[WH[W
7RWDOL]HU
7RWDOL]HU
7H[W %DWFKF\FOHFQW
6HQVRUVL]H
0HQX 7RWDOL]HU
6HQVRUW\SH
7$*QXPEHU
2SHUDWLQJWLPH

7H[W
0HQX 7RWDOL]HU

7H[W
0HQX 4PD[

7H[W
0HQX )ORZUDWH

7H[W
7H[W 0HQX
2SHUDWLQJWLPH
0HQX )ORZYHORFLW\

%DWFKF\FOHFQW
7RWDOL]HUUHVHW
UHVHW
,I%DWFKLVFKRVHQ

7RWDOL]HUUHVHW 7H[W
0HQX 6HQVRUVL]H
,I%DWFKLVQRWFKRVHQ

%DWFKTXDQWLW\ 7H[W
0HQX 6HQVRUW\SH
,I%DWFKLVFKRVHQ

%DWFKVWDUW (UURU
SDXVHGUHVHW SHQGLQJ
,I%DWFKLVFKRVHQ

%DWFKDPRXQW 6WDWXVORJ
GRQH
,I%DWFKLVFKRVHQ

1) When selecting Batch amount for upper line, the upper line is initially blank. The amount done
may not appear until the batch is started.
2) 'Text' means that the text for the chosen measured value is shown. For example, if textis chosen
in line 2 and flow velocity is chosen in line 3, the text "Flow velocity " is shown in line2 and the
measured flow velocity is shown in line 3.

SITRANS FM MAG 5000/6000 IP67


92 Operating Instructions, 12/2019, A5E02338368-AD
Menu diagrams
B.17 Product identity

B.17 Product identity

3URGXFW
LGHQWLW\

7UDQVPLWWHUW\SH
0$*

7UDQVPLWWHUFRGHQR
0(

7UDQVPLWWHUVHUQR
1

7UDQVPLWWHU6:YHUVLRQ
]]]]

6HQVRUW\SH
0$*

6HQVRUFRGHQR
0(

6HQVRUVHULDOQR
1

6HQVRUVL]H
'1,1

7$*QXPEHU
;;;;;

$GGRQPRGXOH
6:YHUVLRQ]]]]

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 93
Menu diagrams
B.19 Change password

B.18 Add-on communication module

Example: HART

+$57PRGXOH

+$57VKRUWDGGUHVV

7$*QDPH

'HVFULSWLRQ

'DWH

+$576:YHUVLRQ

Note
Burst mode is not available with HART communication

B.19 Change password

3DVVZRUG

&KDQJHSDVVZRUG

SITRANS FM MAG 5000/6000 IP67


94 Operating Instructions, 12/2019, A5E02338368-AD
Factory settings C
C.1 Transmitter factory settings
The factory settings shown on the following pages apply to MAG 5000/6000 as well as
MAG 6000 I.

Note
1)
Due to legal requirements the only units available for MAG 5000/6000 CT are m3 and m3/h.

Menu item Parameter Factory settings Options More info


Password Password 1000 1000 to 9999 Changing pass‐
word (Page 44)
Basic settings Flow direction Positive Positive, negative Changing basic
Qmax Sensor size de‐ Sensor size dependent settings
pendent (Page 44)

● volume unit 1) Sensor size de‐ m 3, ml, l, hl, Ml, ft 3, in 3, US G, US kG, BBL32,
pendent BBL42, UK G, UK MG, ? (customer unit)
● time unit Sensor size de‐ Sec., min., hour, day, ? (customer unit)
pendent
Totalizer 1 Forward Forward, reverse, net
● Totalizer 1 unit Sensor size de‐ m 3, ml, l, hl, Ml, ft 3, in 3, US G, US kG, BBL32,
1) pendent BBL42, UK G, UK MG, ? (customer unit)
Totalizer 2 Reverse Forward, reverse, net
● Totalizer 2 unit Sensor size de‐ m 3, ml, l, hl, Ml, ft 3, in 3, US G, US kG, BBL32,
1) pendent BBL42, UK G, UK MG, ? (customer unit)
Low flow cut-off 1.5% 0 to 9.9%
Empty pipe Off On, Off
Velocity unit m/s m, mm, cm, dm, ft , in
per s, min, h, d, ? (customer unit)
Error level Warning Fatal, permanent, warning

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 95
Factory settings
C.1 Transmitter factory settings

Menu item Parameter Factory settings Options More info


Output Current output Off for On/off, Unidirectional/bidirectional, 0 to 20 mA/4 Output settings
MAG5000/ to 20 mA/4 to 20 mA + Alarm (Page 51)
6000
● Alarm level Low High/Low
● Alarm diff. No Yes/No
● Time constant 5s 0.1 to 30 s
Digital output Pulse Error, direction/limit, batch, frequency, pulse, er‐ Digital output -
ror number, off pulse (Page 83)
Relay output Error level Error, direction/limit, cleaning, error number, off Digital output /
Relay output -
Error level
(Page 84)
Direction/limit Off 1 setpoint, 2 setpoints Digital output /
switch Relay output -
Direction/limit
● Setpoints 0% -100 to +100% (Page 85)
● Hysteresis 5% 0.0 to 100%
Batch Off Digital output /
Relay output -
Batch (Page 85)
● Batch quantity 0 Sensor size dependent

● Batch compen‐ 0 -100 to +100 m3


sation
● Batch counter Down Up, down
Frequency Off 500 Hz, 1 kHz, 5 kHz, 10 kHz Digital output -
● Time constant 5s 0.1 to 30 s frequency
(Page 84)
Pulse On Digital output -
● Pulse polarity Positive Positive, negative pulse (Page 83)

● Pulse width 66 ms 64 μs 130 μs, 260 μs, 510 μs, 1.0 ms, 2.0 ms, 4.1
ms, 8.2 ms, 16 ms, 33, ms, 66 ms, 130 ms, 260
ms, 520 ms, 1.0 s, 2.1 s, 4.2 s
● Volume/pulse Sensor size de‐ Dimension-dependent
pendent
● Time constant 0.1 s 0.1 to 30 s
External input External input Off Batch, reset totalizer, freeze output, forced out‐ External input
put, off (Page 87)
● Batch Start Start, hold/continue, stop, Qmax 2
Sensor charac‐ Correction factor 1 0.85 to 2.00 Sensor charac‐
teristics teristics
(Page 88)
Language Language English English, German, French, Danish, Swedish, Fin‐ Changing lan‐
nish, Spanish, Russian, Italian, Portuguese, Pol‐ guage (Page 48)
ish

SITRANS FM MAG 5000/6000 IP67


96 Operating Instructions, 12/2019, A5E02338368-AD
Factory settings
C.2 50 Hz Dimension dependent Qmax

Menu item Parameter Factory settings Options More info


Operator menu Primary field Flow rate Flow rate, Totalizer 1, Totalizer 2 Changing opera‐
Title/subtitle lines Flow rate Flow rate, Flow velocity, Qmax, Totalizer 1, To‐ tor menu setup
talizer 2, Totalizer 1 reset, Totalizer 2 reset, (Page 47)
Batch start/paused/stop, Batch cycle counter,
Batch cycle counter reset, Sensor size, Sensor
type, Error pending, Status log, Tag No.

See also
Change password (Page 94)

C.2 50 Hz Dimension dependent Qmax

Table C-1 MAG 1100, MAG 1100 F, MAG 3100, MAG 3100 P and MAG 5100 W with M20 cable glands

DN Qmax*
Factory sett‐ MAG 5100 W (Order no. 7ME6520) MAG 1100, MAG 1100 F, 5100 W unit
ting (Order no. 7ME6580), MAG 3100,
3100 P
mm (inch) min. max. min. max.
2 ( /12)
1
30 - - 3.903623 156.1448 l/h
3 (1/8) 70 - - 6.361726 254.469 l/h
6 ( /4)
1
300 - - 25.44691 1017.876 l/h
10 ( /8)
3
900 - - 70.68584 2827.433 l/h
15 (½) 2000 - - 159.0432 6361.725 l/h
25 (1) 5000 441.7865 17671.45 441.7865 17671.45 l/h
40 (1½) 12 1.130974 45.23893 1.130974 45.23893 m3/h
50 (2) 20 1.574527 62.98107 1.767146 70.68583 m3/h
65 (2½) 30 2.499681 99.98723 2.986477 119.459 m3/h
80 (3) 50 4.003646 160.1458 4.523894 180.9557 m3/h
100(4) 120 6.252163 250.0864 7.068584 282.7433 m3/h
125 (5) 180 10.00647 400.2585 11.04467 441.7864 m3/h
150 (6) 250 15.74527 629.8107 15.90432 636.1725 m3/h
200(8) 400 24.93797 997.5184 28.27434 1130.973 m3/h
250(10) 700 40.00377 1600.15 44.17865 1767.145 m3/h
300 (12) 1000 62.50395 2500.157 63.61726 254469 m3/h
350 (14) 1200 86.59015 3463.605 86.59015 3463.605 m3/h
400 (16) 1800 113.0974 4523.893 113.0974 4523.893 m3/h
450 (18) 2000 143.1389 5725.552 143.1389 5725.552 m3/h
500 (20) 3000 176.7146 7068.583 176.7146 7068.583 m3/h
600 (24) 4000 254.4691 10178.76 254.4691 10178.76 m3/h
700 (28) 4500 346.3606 13854,42 346.3606 13854.42 m3/h
750 (30) 5000 397.6079 15904.31 397.6079 15904.31 m3/h

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 97
Factory settings
C.3 60 Hz Dimension dependent Qmax

DN Qmax*
Factory sett‐ MAG 5100 W (Order no. 7ME6520) MAG 1100, MAG 1100 F, 5100 W unit
ting (Order no. 7ME6580), MAG 3100,
3100 P
mm (inch) min. max. min. max.
800 (32) 7000 452.3894 18095,57 452.3894 18095.57 m3/h
900 (36) 9000 572.5553 22902,21 572.5553 22902.21 m3/h
1000 (40) 12000 706.8584 28274.33 706.8584 28274.33 m3/h
1050 (42) 12000 706.8584 28274.33 706.8584 28274.33 m3/h
1100 (44) 14000 855.986 34211.94 855.2986 3421194 m3/h
1200 (48) 15000 1017.877 40715.04 1017.877 40715.04 m3/h
1400 (54) 25000 - - 1385.443 55417.69 m3/h
1500 (60) 30000 - - 1590.432 63617.25 m3/h
1600 (66) 35000 - - 1809.558 72382.29 m3/h
1800 (72) 40000 - - 2290.222 91608.84 m3/h
2000 (78) 45000 - - 2827.434 113097.3 m3/h
2200 (90) 50000 - - 3421,195 136847.7 m3/h
2400 (96) 55000 - - 4071.505 162860.1 m3/h
2600 (102) 60000 - - 4778.363 191134.4 m3/h
280 (114) 65000 - - 5541.77 221670.7 m3/h
3000 (120) 70000 - - 6361.726 254469 m3/h

* The min. and max. amount values show mathematical values and do not indicate
measurement accuracy

C.3 60 Hz Dimension dependent Qmax

Table C-2 MAG 1100, MAG 1100 F, MAG 3100, MAG 3100 P and MAG 5100 W with ½" NPT cable glands

DN Qmax.
Factory set‐ MAG 5100 W (Order no. 7ME6520) MAG 1100, MAG 1100 F, 5100 W unit
ting* (Order no. 7ME6580), MAG 3100,
3100 P
mm (inch) min. max. min. max.
2 ( /12)
1
0.14 - - 0.01718714 0.6874852 US GPM
3 (1/8) 0.31 - - 0.02800984 1.120393 US GPM
6 ( /4)
1
1.4 - - 0.1120394 4.481573 US GPM
10 (3/8) 4 - - 0.3112204 12.44881 US GPM
15 (½) 9 - - 0.7002459 28.0 US GPM
25 (1) 23 1.945128 77.80509 1.945128 77.80509 US GPM
40 (1½) 53 4.979526 199.181 4.979526 199.181 US GPM
50 (2) 89 6.932434 277.2973 7.78051 311.2203 US GPM
65 (2½) 133 11.00577 440.2305 13.14907 525.9624 US GPM
80 (3) 221 17.62753 705.1008 19.91811 796.7241 US GPM

SITRANS FM MAG 5000/6000 IP67


98 Operating Instructions, 12/2019, A5E02338368-AD
Factory settings
C.3 60 Hz Dimension dependent Qmax

DN Qmax.
Factory set‐ MAG 5100 W (Order no. 7ME6520) MAG 1100, MAG 1100 F, 5100 W unit
ting* (Order no. 7ME6580), MAG 3100,
3100 P
mm (inch) min. max. min. max.
100(4) 529 27.52745 1101.097 31.12204 1244.881 US GPM
125 (5) 793 44.05714 1762.285 48.62819 1945.127 US GPM
150 (6) 1101 69.32434 2772.973 70.02459 2800.984 US GPM
200 (8) 1762 109.7986 4391.941 124.48819 4979.525 US GPM
250 (10) 3083 176.1313 7045.251 194.5128 7780.507 US GPM
300 (12) 4403 275.1967 11007.86 280.0984 11203.93 US GPM
350 (14) 5284 381.245 15249.79 381.245 15249.79 US GPM
400 (16) 7926 497.9526 19918.1 497.9526 19918.1 US GPM
450 (18) 8806 630.2213 25208.84 630.2213 25208.84 US GPM
500 (20) 13209 778.051 31122.03 778.051 31122.03 US GPM
600 (24) 17612 1120.394 44815.73 1120.394 44815.73 US GPM
700 (28) 19813 1524.98 60999.19 1524.98 60999.19 US GPM
750 (30) 22015 1750.615 70024.58 1750.615 70024.58 US GPM
800 (32) 3082 1991.811 79672.4 1991.811 79672.41 US GPM
900 (36) 39626 2520.885 100835.3 2520.885 100835.3 US GPM
1000 (40) 52835 3112.204 124488.1 3112.204 124488.1 US GPM
1050 (42) 52835 3112.204 137248.1 3112.204 124488.1 US GPM
1100 (44) 61641 3765.767 150630.6 3765.767 150630.6 US GPM
1200 (48) 66044 4481.574 179262.9 4481.574 179262.9 US GPM
1400 (54) 110072 - - 6099.92 243996.7 US GPM
1500 (60) 1320867 - - 7002.459 280098.3 US GPM
1600 (66) 154101 - - 7967.242 318689.6 US GPM
1800 (72) 176115 - - 10083.54 403341.5 US GPM
2000 (78) 198130 - - 12448.82 497952.5 US GPM
2200 (90) 220144 - - 15063.07 602522.6 US GPM
2400 (96) 242158 - - 17926.3 717051.7 US GPM
2600 (102) 264173 - - 21038.5 841539.8 US GPM
2800 (114) 286187 - - 24399.68 975987 US GPM
3000 (120) 308201 - - 28009.84 1120393 US GPM

* Factory setting sets Qmax to a metric unit (see previous table). The values here are converted
to rounded off US GPM.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 99
Factory settings
C.4 50 Hz Dimension dependent volume/pulse and batch

C.4 50 Hz Dimension dependent volume/pulse and batch

Table C-3 MAG 1100, MAG 1100 F, MAG 3100, MAG 3100 P and MAG 5100 W with ½" NPT cable glands

DN Volume/pulse or batch quantity* Factory setting


MAG 5100 W (Order no. MAG 1100, MAG 1100 F, 5100 W (Order Volume/ Pulse & Totaliz‐
7ME6520) no. 7ME6580), MAG 3100, 3100 P pulse & batch er
mm (inch) min. max. min. max. batch unit unit
amount
2 ( /12)
1
- - 3.61466 μl 94.75103 l 0.1 ml ml
3 (1/8) - - 5.890487 μl 154.4155 l 0.1 ml ml
6 ( /4)
1
- - 23.56195 μl 617.6622 l 1 l l
10 ( /8)
3
- - 65.44985 μl 1.715728 m 3
1 l l
15 (½) - - 147.2622 μl 3.860389 m3 1 l l
25 (1) 409.0616 μl 10.7233 m3 409.0616 μl 10.7233 m3 10 l l**
40 (1½) 1.047198 ml 27.45165 m3 1.047198 ml 27.45165 m3 10 l l**
50 (2) 1.457896 ml 38.21785 m 3
1.636247 ml 42.89321 m 3
10 l l**
65 (2½) 2.31452 ml 60.67373 m3 2.765257 ml 72.48952 m3 100 l l**
80 (3) 3.70708 ml 97.17886 m 3
4.188791 ml 109.8066 m 3
100 l l**
100(4) 5.789039 ml 151.7561 m 3
6.544985 ml 171.5728 m 3
100 l l**
125 (5) 9.265244 ml 242.8828 m3 10.22654 ml 268.0825 m3 100 l m3
150 (6) 14.57896 ml 382.1785 m3 14.72622 ml 386.0389 m3 100 l m3
200 (8) 23.09071 ml 605.309 m3 26.17994 ml 686.2913 m3 1 m3 m3
250 (10) 37.04053 ml 970.995 m 3
40.90616 ml 1072.33 m 3
1 m 3
m3
300 (12) 57.87403 ml 1517.132 m3 58.90487 ml 1544.155 m3 1 m3 m3
350 (14) 80.17607 ml 2101.767 m 3
80.17607 ml 210.7671 m 3
1 m 3
m3
400 (16) 104.7198 ml 2745.165 m3 104.7198 ml 2745.165 m3 1 m3 m3
450 (18) 132.536 ml 3474.35 m3 132.536 ml 3474.35 m3 1 m3 m3
500 (20) 163.6247 ml 4289.321 m 3
163.6247 ml 4289.321 m 3
10 m 3
m3
600 (24) 235.6195 ml 6176.622 m3 235.6195 ml 6176.622 m3 10 m3 m3
700 (28) 320.7043 ml 8407.069 m 3
320.7143 ml 8407.069 m 3
10 m 3
m3
750 (30) 368.1554 ml 9650.972 m3 368.1554 ml 9650.972 m3 10 m3 m3
800 (32) 418.8791 ml 10980.66 m 3
418.8791 ml 10980.66 m 3
10 m 3
m3
900 (36) 530.1438 ml 13897.4 m3 530.1438 ml 13897.4 m3 10 m3 m3
1000 (40) 654.4985 ml 17157.28 m3 654.4985 ml 17157.28 m3 10 m3 m3
1050 (42) 654.4985 ml 17157.28 m 3
654.4985 ml 17157.28 m 3
10 m 3
m3
1100 (44) 79.94321 ml 20760.31 m3 791.9432 ml 20760.31 m3 10 m3 m3
1200 (48) 942.4778 ml 24706.48 m 3
942.4778 ml 24706.48 m 3
10 m 3
m3
1400 (54) - - 1.282817 l 33628.27 m3 10 m3 m3
1500 (60) - - 1.472622 l 38603.89 m3 10 m3 m3
1600 (66) - - 1.675517 l 43922.64 m 3
10 m 3
m3
1800 (72) - - 2.120576 l 55589.6 m3 10 m3 m3
2000 (78) - - 2.617994 l 68629.13 m 3
10 m 3
m3
2200 (90) - - 3.167773 l 83041.25 m3 10 m3 m3

SITRANS FM MAG 5000/6000 IP67


100 Operating Instructions, 12/2019, A5E02338368-AD
Factory settings
C.5 60 Hz Dimension dependent volume/pulse and batch

DN Volume/pulse or batch quantity* Factory setting


MAG 5100 W (Order no. MAG 1100, MAG 1100 F, 5100 W (Order Volume/ Pulse & Totaliz‐
7ME6520) no. 7ME6580), MAG 3100, 3100 P pulse & batch er
mm (inch) min. max. min. max. batch unit unit
amount
2400 (96) - - 3.769912 l 98825.9 m 3
10 m3 m3
2600 (102) - - 4.4241 l 115983. m 3
10 m 3
m3
2800 (114) - - 5.131268 l 134513.1 m3 10 m3 m3
3000 (120) - - 5.890487 l 154415.5 m3 10 m3 m3

* The min. and max. amount values show mathematical values and do not indicate
measurement accuracy.
** For CT devices the totalizer 1 unit is in m3

C.5 60 Hz Dimension dependent volume/pulse and batch

Table C-4 MAG 1100, MAG 1100 F, MAG 3100, MAG 3100 P and MAG 5100 W with ½" NPT cable glands

DN Volume/pulse or batch quantity


MAG 5100 W (Order no. 7ME6520) MAG 1100, MAG 1100 F, 5100 W (Order no. 7ME6580),
MAG 3100, 3100 P
mm (inch) US G min. US G max. US G min. US G max.
2 ( /12)
1
- - 0.00000095484069 25.03057
3 (1/8) - - 0.000001556102 40.79227
6 ( /4)
1
- - 0.000006224408 163.1691
10 ( /8)
3
- - 0.00001729003 453.2475
15 (½) - - 0.00003890255 1019.806
25 (1) 0.0001080627 2832.796 0.0001080627 2832.796
40 (1½) 0.0002766404 7251.96 0.0002766404 7251.96
50 (2) 0.0003851353 10096.08 0.0004322506 11331.18
65 (2½) 0.0006114314 16028.3 0.0007305034 19149.7
80 (3) 0.0009793068 25671.93 0.001106562 29007.84
100(4) 0.001529303 40089.74 0.001729003 45324.75
125 (5) 0.002447619 64162.85 0.002701566 70819.92
150 (6) 0.003851353 100960.8 0.003890255 101980.6
200 (8) 0.00609992 159905.7 0.006916009 181299
250 (10) 0.009785071 256509.7 0.01080627 283279.6
300 (12) 0.01528871 400784.1 0.01556102 407922.7
350 (14) 0.02118028 555228.2 0.02118028 555228.2
400 (16) 0.02766404 725196 0.02766404 725196
450 (18) 0.0350123 917826.2 0.0350123 917826.2
500 (20) 0.04322506 1133118 0.04322506 1133118
600 (24) 0.06224408 1631691 0.06224408 1631691
700 (28) 0.0847211 2220912 0.0847211 2220912

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 101
Factory settings
C.5 60 Hz Dimension dependent volume/pulse and batch

DN Volume/pulse or batch quantity


MAG 5100 W (Order no. 7ME6520) MAG 1100, MAG 1100 F, 5100 W (Order no. 7ME6580),
MAG 3100, 3100 P
mm (inch) US G min. US G max. US G min. US G max.
750 (30) 0.09725637 2549517 0.09725637 2549517
800 (32) 0.1106562 2900784 0.1106562 2900784
900 (36) 0.1400492 3671304 0.1400492 3671304
1000 (40) 0.1729003 4532475 0.1729003 4532475
1050 (42) 0.1729003 4532475 0.1729003 4532475
1100 (44) 0.2092093 5484294 0.2092093 5484294
1200 (48) 0.2489763 6526764 0.2489763 6526764
1400 (54) - - 0.3388844 8883651
1500 (60) - - 0.3890255 10198060
1600 (66) - - 0.4426246 11603130
1800 (72) - - 0.5601967 14685210
2000 (78) - - 0.6916009 18129900
2200 (90) 0.836837 21937170
2400 (96) 0.995906 26107050
2600 (102) 1.168806 30639530
2800 (114) 1.355538 35534600
3000 (120) 1.556102 40792270

SITRANS FM MAG 5000/6000 IP67


102 Operating Instructions, 12/2019, A5E02338368-AD
Product documentation and support D
D.1 Product documentation
Process instrumentation product documentation is available in the following formats:
● Certificates (http://www.siemens.com/processinstrumentation/certificates)
● Downloads (firmware, EDDs, software) (http://www.siemens.com/processinstrumentation/
downloads)
● Catalog and catalog sheets (http://www.siemens.com/processinstrumentation/catalogs)
● Manuals (http://www.siemens.com/processinstrumentation/documentation)
You have the option to show, open, save, or configure the manual.
– "Display": Open the manual in HTML5 format
– "Configure": Register and configure the documentation specific to your plant
– "Download": Open or save the manual in PDF format
– "Download as html5, only PC": Open or save the manual in the HTML5 view on your PC
You can also find manuals with the Mobile app at Industry Online Support (https://
support.industry.siemens.com/cs/ww/en/sc/2067). Download the app to your mobile device
and scan the device QR code.

Product documentation by serial number


Using the PIA Life Cycle Portal, you can access the serial number-specific product information
including technical specifications, spare parts, calibration data, or factory certificates.

Entering a serial number


1. Open the PIA Life Cycle Portal (https://www.pia-portal.automation.siemens.com).
2. Select the desired language.
3. Enter the serial number of your device. The product documentation relevant for your device
is displayed and can be downloaded.
To display factory certificates, if available, log in to the PIA Life Cycle Portal using your login or
register.

Scanning a QR code
1. Scan the QR code on your device with a mobile device.
2. Click "PIA Portal".
To display factory certificates, if available, log in to the PIA Life Cycle Portal using your login or
register.

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 103
Product documentation and support
D.2 Technical support

D.2 Technical support

Technical support
If this documentation does not completely answer your technical questions, you can enter a
Support Request (http://www.siemens.com/automation/support-request).
Additional information on our technical support can be found at Technical Support (http://
www.siemens.com/automation/csi/service).

Service & support on the Internet


In addition to our technical support, Siemens offers comprehensive online services at Service
& Support (http://www.siemens.com/automation/service&support).

Contact
If you have further questions about the device, contact your local Siemens representative at
Personal Contact (http://www.automation.siemens.com/partner).
To find the contact for your product, go to "all products and branches" and select "Products &
Services > Industrial automation > Process instrumentation".
Contact address for business unit:
Siemens AG
Digital Industries
Process Automation
Östliche Rheinbrückenstr. 50
76187 Karlsruhe, Germany

SITRANS FM MAG 5000/6000 IP67


104 Operating Instructions, 12/2019, A5E02338368-AD
Index

A L
Add-on module, 15 Laws and directives, 11
Add-on modules
Electrical connection:Add-on modules, 40
Alarm differentiation, 51 M
Alarm level, 51
Mains supply, 37
Maintenance, 57
Manuals, 103
C Menu structure, 51
Catalog
catalog sheets, 103
Certificates, 103 R
Communication module, (See Add-on module)
Repair, 57
Compliance, 11
Return procedure, 59
Customer Support, (Refer to Technical support)
Customer Support Hotline, 58
S
D Safety, 11
Instrument safety standards, 11
Device
Safety instructions
Inspection, 18
Electrical connection, 37
Diagnostic functions, 61
Scope of delivery, 8
Downloads, 103
Service, 57, 58, 104
Service and support
Internet, 104
E Settings, 51
Electrical connection, 40 Support, 58, 104
Safety instructions, 37 Support request, 104
Error handling, 61 System components, 15

H T
Hotline, 58, (Refer to Support request) Technical support, 104
partner, 104
personal contact, 104
I Troubleshooting, 64
Installation
Indoor/outdoor, 21
Internet
W
Contact person, 59 Warranty, 9
Support, 58

SITRANS FM MAG 5000/6000 IP67


Operating Instructions, 12/2019, A5E02338368-AD 105
Index

SITRANS FM MAG 5000/6000 IP67


106 Operating Instructions, 12/2019, A5E02338368-AD

You might also like