Keyop Functions 9F3715
Keyop Functions 9F3715
Keyop Functions 9F3715
CR Systems
Introduction
The Key Operator is responsible for setting the configuration defaults and
operating parameters for the CR System. CR System configuration is done at
installation to maximize workflow and System operation, and to optimize
workflow.
You may need to change the CR System configuration.
To access the Key Operator functions:
1. At the Login screen, enter your password for Key Operator and select
Login. See "Logging In" in the Kodak DirectView CR 850/950 or
CR 500 System User’s Guide.
Patient Worklist
Statistics
At the Statistics menu you can compare scan and erase cycles, view the
cassette, technologist and destination statistics and check the status of your
destinations.
Cassette Statistics 1. From the Key Operator Functions menu, select Statistics and then
select Cassette Statistics. Use the data to evaluate cassette
performance.
2. Select Clear Stats to clear the data fields.
Destination Statistics 1. At the Key Operator Functions menu, select Statistics, then select
Destination Statistics. The screen displays all network destinations,
the number of successful deliveries, number of failed deliveries, and the
successful delivery percentage. The data updates every time you deliver
an image.
2. Select Clear Stats to clear the data fields.
Technologist Statistics At the Key Operator Functions menu, select Statistics and then select
Tech Statistics.
This screen lists all the technologists in the system.
Technologist Statistics
Technologist Statistics 3
Scan Cycles
Destination Status At the Key Operator Functions menu select Statistics and then select
Summary Destination Status Summary for the status for each destination.
Destination Status
The screen displays the last time data was sent to the destinations listed and
the status of the devices at that time.
System Configuration
The Kodak CR System is highly configurable and integrates well into the
digital radiology department. You can configure, save and restore many
attributes of the CR System for your operation and optimize workflow:
• Restore Configuration —Due to the advanced printing features
incorporated in the CR System software, take care when configuring the
system. Generate test films when making changes that affect printed
output. This ensures that the changes made are compatible with the
printers you will be using. Once configured, you can save the settings for
backup or transfer them to another CR System.
• Option Registration —Install optional features easily.
• Workflow Optimization —Defining lists optimizes workflow by
reducing the amount of time it takes to create an exam and eliminating
typographical errors. There are lists for Departments, which can be used
to help deliver images to the correct locations, Physicians, and
Procedures. Workflow Optimization lets you map images to procedures,
that is, when you select a procedure from the list, or the RIS feeds the CR
System a Procedure Code, all the image icons are built automatically. You
can construct Trauma defaults for up to 35 patients.
• HIS/RIS Broker Configuration—Lets the CR System communicate with
your RIS, displaying the patient work list. You can also query the RIS for
individual patients.
• CR Display Configuration—Lets you define default query criteria and
how query results are displayed. You can disable the Right and Left
markers here, select Full Resolution viewing, and configure to show the
patient’s middle initial on the PEC Input screen.
• Profile Destination Configuration—Lets you tell the CR System where
to deliver images. You can deliver based on body part, projection, cassette
size, or department. This saves time by automatically delivering the images
where you want.
• Text Box Configuration—You can define what information appears on
the film, allowing you to see only the information that is important to you
and not taking up space with unnecessary data.
• Delivery Preferences —Lets you set CT Pixel spacing so that
measurements on workstations can be displayed in pixels or in mm. You
also set the name and address of your hospital here.
• External Devices —Lets you add ROPs, RPDES terminals, or a remote
Key Operator PC to your system.
• Bar Code Configuration —Lets you program the bar code scanner to
automatically recognize Tech IDs, Accession Numbers, Patient IDs, or
Cassette IDs, populating these fields with the information without the
Saving System 1. Insert a floppy disk into the CR System and select OK. If you do not insert
Configurations a floppy disk, the CR System prompts you to do so.
2. At the Key Operator Functions menu, select System Configuration,
then select Save all Configs.
Restoring You can save configurations to a floppy disk for backup purposes, then load
Configurations (restore) them. Using this technique you can copy settings from one
CR System and transfer them to other systems, reducing the time it takes to
change the configuration of multiple machines.
To restore system configuration information:
1. At the Key Operator Functions menu, select System Configuration,
then select Restore Configuration.
Restore Configuration
2. Insert the floppy disk that contains the configuration information into the
CR System.
3. Select each check box to select the options you want to load, or select
Select All.
NOTE: Do not select items below the line unless you are restoring settings to
the same machine.
4. Select Load Config.
You may have to restore the following items manually following the Restore
operation from floppy diskette:
At this Location Restore this Item
Required Fields screen Required to Submit Exam Information
Required Fields screen Required to Deliver Images to Mandatory
Destinations
Delivery Options screen Use a Unique Image for Every Image in Study
User Login Configuration User Login Active
screen
NOTE: Restore adds a default Profile automatically. You can delete this
profile with no affect to the system.
NOTE: In the second check box from the top on the
Restore Configurations screen, the following items are grouped
together:
Colors:
• The customized colors of buttons, statuses, etc.
Regional Data:
• Date Format
• Time Zone
• Military Time
• Language
CR Display:
• Japan Only - MultiByte name tab focus order
• Show ID Window on Image Viewer
• Show Cursor
• Allow Technologists to place digital markers
• Automatically populate technologist ID field with login name
• Volume
• Blue Tint Factor
Gamma LUT:
• Gamma LUT (monitor’s response curve and high/low ambient light) of
the CR and each ROP.
Miscellaneous:
• Source to Bucky Front Distance
• Default Body Part IPL and ID
• Default View Position IPL ID
• Clustering (not used)
• Required Field Mode - required to submit or required to deliver
• System Maintenance Defaults (Disk High Water Mark, Disk Low Water
Mark)
Delivery Options:
• Upper Case all fields on delivery
• QA mode
• Create A Unique Image Number for Every Image
• Override Pass-through Mode
• Use CT Pixel Spacing
• Automatically Set Multi-format Sub Images to Delivered
Option Registration You activate upgrade options at the Option Registration Screen.
Option Registration
Changing Button The procedures for changing the names, colors, and position are consistent
Names, Colors, and for all the configuration setups. You may want to change the locations or
Position colors of buttons so they are grouped more logically for your hospital.
Body Part and Projection You can configure the body part and projection screens to meet your needs.
Configuration
• Change the color, name, and location of the button as it appears on the
pop-up keyboard.
• Hide any buttons that you do not want to be displayed.
• Re-organize the buttons into groups or sequence of appearance.
• Customize existing buttons.
NOTE: If you create a custom button, have your Applications Consultant set
up your Image Processing Preferences for the button. Use the
preference editor to configure the parameters that control the image
processing algorithms.
1. At the Key Operator Functions menu, select System
Configuration then select Workflow Optimization.
2. Select Body Part Configuration.
3. Select the button you want to change.
4. Select Change Name.
5. Enter the name you want using the virtual keyboard.
6. Change the color and position if necessary.
7. Select Save Changes.
Department and You can define a list of all departments, physicians, and procedures to
Physician List minimize the amount of data entry by the technologist and to eliminate
Configuration typographical errors. You can use Department lists for Profile Destination
Configuration.
When the lists are defined, select the field on the PEC Input screen to display
the list. Select the data you want to appear in the field.
3. Select a button.
Change Name
Procedure List and You can use these two features together or independently to automatically
Procedure Mapping create image icons for your department’s procedures.
Overview • Procedure List—lets you create 34 buttons that appear when you select
the Procedure field on the PEC input screen. Procedure Lists must have a
Name and a Procedure Code.
• Procedure Mapping—lets you define body parts, projections, positions,
etc., for the image icons displayed for the selected procedures for the 34
Procedure buttons you created using the Procedure List option. You can
also create image icons for hundreds of procedures from your RIS system,
determined by either the Procedure Code or the Procedure ID.
NOTE: See Appendix B “Default Procedure Codes” in the
Kodak DirectView CR 850/950 System User’s Guide or
Appendix A in the Kodak DirectView CR 500 System User’s Guide
for a listing of default procedure codes from Kodak.
Procedure List You can define up to 34 procedures by assigning names and procedure
Configuration codes. Subsequently, during the examination, after selecting a procedure
code and configuring the procedure mapping, the image icons appear
automatically.
1. At the Key Operator Functions menu select System Configuration,
then select Workflow Optimization.
2. Select Procedure List Configuration.
Procedure Mapping Procedure mapping associates the images required for a procedure. When
(Option) you select a procedure, the image icons appear automatically in the Study
Data area.
If you use the appropriate RIS codes, or select the Procedure ID in the
HIS/RIS - Broker Configuration, the image icons appear automatically when
the modality worklist is used.
NOTE: The Worklist SCP should use the Code Value element (0008,0100) of
the Requested Procedure Code Sequence (0032,1064), and
Requested Procedure ID element (0040, 1002).
You can easily map procedures directly from the CR System, or from
Microsoft Internet Explorer, using the keyboard and mouse on the Remote
Key Operator computer. See “Default Procedure Codes “in Appendix B in the
Kodak DirectView CR 850/950 System User’s Guide or Appendix A in the
Kodak DirectView CR 500 System User’s Guide. Rather than typing the
procedure code and name, you can select one from the list.
NOTE: If the Scheduled Workflow Option is enabled, the following fields
must be used: Code value element (0008, 0100) of the Requested
Protocol sequencer.
Procedure Mapping on 1. At the Key Operator Functions menu select System Configuration
the CR System then select Workflow Optimization.
2. Select Procedure Mapping. If the procedure is listed, select the
procedure and go to step 5.
NOTE: If the procedure code is listed in black, it has been assigned to a
button. Otherwise, it is listed in blue.
3. Select the Procedure Code field and select a Procedure Code from the
list.
4. Select the Procedure field and select a Procedure from the list.
5. Select View Exam.
View Images
– kV
– mAs
– Distance
6. Select Save Changes.
7. Repeat steps three through six for each additional image and select Save
Changes.
8. Select Back to return to this screen. Repeat these steps for each
procedure you want to map.
NOTE: You can map up to 30 images per procedure.
Mapping more than 34 You can add more than 34 procedure codes, but they won’t appear as
Procedures buttons. To do this, follow steps 3 - 8 of “Procedure Mapping on the CR
System” on page 26.
Editing Procedure Codes If you need to change the name or Procedure Code, for example, if you want
and Names to change the procedure code to one of the default procedures to a code that
is compatible with your RIS, do the following:
1. Enter a new code.
2. Enter a new name.
3. Select the Procedure Code or Procedure in the list that you want to
edit.
4. Select Update Procedure.
Using a HIS/RIS System To configure Procedure Mapping to work with your HIS/RIS System, make
sure that the code you enter for the Procedure or ID matches the code that
comes from the RIS. You can use any number for the Procedure Code,
however, it will only work automatically with the HIS/RIS when the code or ID
matches exactly. See “HIS/RIS Options” on page 40.
For example, if you had a HIS/RIS procedure “2-view Chest” code 102 and
you want to create a procedure on the CR System called 2-view Chest:
1. At the Key Operator Functions menu select System Configuration,
then select Workflow Optimization.
2. Select Procedure Mapping.
3. Create a procedure named 2-view Chest and enter a code of 102.
4. Select the icons you want to be displayed when a 2-view chest procedure
is ordered, such as an AP and LAT chest.
When the work list on the CR System is updated and the RIS code 102 is
received, the AP chest and LAT chest image icons are automatically
displayed.
Deleting a Procedure 1. At the Key Operator Functions menu select System Configuration,
then select Workflow Optimization.
2. Select Procedure Mapping.
3. Select the procedure that you want to delete and select
Delete Procedure.
4. Select Yes on the pop-up screen.
Procedure Mapping You can use the Remote Key Operator to create or change Procedure Mapping
Using the Remote Key with these features:
Operator • Add—Enter a new Procedure Code (required) and Procedure Name and
use the pull-down menus to map the associated exams.
• Find—Search for a Procedure Code or Procedure Name.
• Update—Edit the Procedure Code or Procedure Name.
• Copy From—Copy from an existing Procedure Code to a new or existing
Procedure Mapping entry.
• Delete—Delete the procedure mapping for any existing procedure.
NOTE: There is a limit of 30 images you can map per procedure. See
Appendix B for a listing of default procedure codes from Kodak.
Adding a New Procedure 1. At the Key Operator Functions menu, select System Configuration,
Next, then select External Devices, and then select
Remote Key Operator.
2. Enter the IP address of the Remote Key Operator Workstation and select
Save Changes.
3. At the Remote Key Operator Workstation, open a Web browser.
4. Type the IP address of the CR System and press Enter.
NOTE: The Remote Key Operator Workstation can be any hospital computer
that has Microsoft Internet Explorer 5.5 or higher installed.
5. Log on with your Key Operator user name and password if applicable.
6. Select Procedure Mapping.
7. Enter a new Procedure Code (required) and Procedure Name and select
Add Exam.
Add Procedure
Locating Procedures • From the Remote Key Operator Procedure Mapping menu, enter
the Procedure Code and Procedure Name and select Find.
The procedure appears, or a message appears stating that no matches
were found.
Editing Procedures 1. From the Remote Key Operator Procedure Mapping menu, enter a
Procedure Code and Procedure Name, select Find or select Update
from the list of existing Procedure Codes.
2. Edit the procedure mapping from the drop-down lists.
3. Select Update and Back when finished.
Using an existing 1. At the Remote Key Operator Procedure Mapping screen, select
procedure to Create a Copy From on the procedure you want to use.
New Procedure
2. Enter the Procedure Code for an existing code or a new Procedure Code,
and select Copy.
Deleting a Procedure 1. Select the Delete button next to the procedure you want to delete and
select Enter.
2. The message “Do you really want to delete this procedure mapping?”
appears. Select Yes on the CR 850/950. Select OK on the CR 500.
HIS/RIS Broker
Configuration
Introduction The Kodak DirectView CR Systems utilize DICOM Worklist Management to
receive study information from the hospital’s HIS/RIS, usually via the Mitra
PACS Broker.
You can set up your CR System to communicate with the HIS/RIS in any
combination of the following:
• Polling—the CR System requests records at a defined interval, filtered by
modality, station name, A/E title, and date and time.
• Push—every time a record is created or changed, the Broker sends the
CR System an event notification. The CR System requests specific data
associated with the event. The record is transmitted once, and is sent as
soon as it is received by the Broker.
• Remote Patient Query—a single patient query, configured on the
Polling screen, queried from the Patient Query screen.
You can set up your system to operate in any combination, for example:
• Activate Push to reduce the workload impact on the HIS/RIS or PACS
Broker.
• Activate Polling and set the time interval to 24 hours to receive a
complete update once a day.
If polling is active, each time the CR is started the initial complete query is
started and the timer begins for the next query. For example, if the CR is
started at 9:07 and is set to poll every 20 minutes, the first query is at 9:07
and the second one is 20 minutes later, 9:27.
Polling Configuration 1. At the Key Operator Functions menu select System Configuration,
then select HIS/RIS Broker Configuration.
2. Select Polling Configuration.
Polling/Remote Configuration
3. Use the following table to make your selections and then select Save
Changes.
PACS Broker AE Title Broker Name, normally BROKER
PACS Broker IP Address PACS Broker IP Address
PACS Broker Port PACS Broker Port, normally 3320
PACS Poll Modality Normally CR
Station 1-5 ID Enter your station names or AE Title.
The system retrieves records that only
contain the names specified.
Poll Model Active Select to activate Polling Configuration.
Remote Query Active Select to turn on the Remote Patient Query
Option.
Push Configuration NOTE: The connection may be the Broker or direct HIS/RIS communication.
Configure the Broker to send the following event information:
• STUDY_CREATED
• STUDY_DELETED
• STUDY_UPDATED
• STUDY_SCHEDULED
1. At the Key Operator Functions menu select System Configuration,
and then select HIS/RIS Broker Configuration.
2. Select Push Configuration.
Push Configuration
3. Use the following table to make your selections and then select Save
Changes.
HIS/RIS Options The selection for the list on this screen is determined by the Procedure
Mapping configuration of the HIS/RIS Broker. To change selections, the
Broker must be reconfigured.
NOTE: If you enable your system for Scheduled Workflow, Protocol Code is
selected automatically, and is not configurable.
1. At the Key Operator Functions menu, select System Configuration,
HIS/RIS Broker Configuration and then select HIS/RIS Options.
HIS/RIS Options
CR Display This screen lets you optimize the way data is displayed on the CR System.
Configuration 1. At the Key Operator Functions menu, select System Configuration
and then select CR Display Configuration.
CR Display Configuration
2. Use the following table to make your selections then select Save
Changes.
In Japanese, Kanji Name Gets Tab Japanese only - changes the tab focus
Focus First on the Patient Input screen.
Reject Reason This option lets you build a list of reject reasons that a technologist can
Configuration choose from when rejecting an image. The following Reject Reasons are
(Option) factory defaults and can be used or deleted:
– Clipped Anatomy
– Patient Motion
– Positioning Error
You can define up to nine reject reasons. The last reason is always listed as
Other Reason.
To enter Reject Reasons:
1. At the Key Operator Functions menu, select System Configuration,
then select Workflow Optimization.
2. Select Next.
3. Select Reject Reason.
Reject Reason
Reject Reason
NOTE: Select the Track Rejects check box if you want the system to ask for
a reason why the technologist rejected an image. The data is stored in
the System database and can be viewed or downloaded to the Key
Operator PC.
Setting Trauma With this option you can define up to 35 default trauma buttons. Use this data
Defaults (Option) to pre-fill demographic data and procedures for creating a Trauma Study.
To create Trauma defaults:
1. At the Key Operator Functions menu, select System Configuration
then select Workflow Optimization.
2. Select Next.
3. Select Trauma Defaults.
Trauma Defaults
4. Select a button.
5. Select Change Name. Enter a name, such as John Doe.
6. Select Change Trauma Defaults.
7. Enter a Procedure Name and Procedure Code so that the Procedure
Mapping feature automatically creates image icons. You can use the
information from a previously defined procedure. See “Procedure
Mapping (Option)”.
NOTE: For optimal workflow, complete all required fields, so that no
additional information is needed before saving to the database or
routing the image.
8. Select Save Changes.
For example, create a procedure named MVA for a motor vehicle accident,
and enter this procedure for Trauma patient John Doe. When you select John
Doe, the image icons you defined for an MVA appear.
Using Unique Numbers The Unique Number feature lets you assign unique numbers to trauma
patients. For example if you have multiple trauma patients at once, each time
you select a new trauma patient, a unique number is assigned to the field you
select. This lets you keep separate exams for each patient.
You can enter a unique number in the following fields:
• Patient Last Name
• Patient First Name
• Patient ID
• Accession Number
The number increments each time you create a new Trauma PEC record.
1. When entering data for the field, select the Unique Number.
2. Select a number (1 - 4) that represents the number of digits you want to
use for the Unique Number. Percent (%) signs appear in the field.
For example, if you input Trauma %%% in the Accession Number field,
the first patient is numbered Trauma 001, the next Trauma 002, etc.
Profile Destination A profile is the set of destinations where images are sent. In the CR System,
Configuration Image Routing is controlled by profiles.
A profile has three components:
– Name
– Criteria (Cassette Size, Body Part, Projection, and Department)
– Destinations to which the data is sent when the criteria matches
NOTE: Define multiple profiles with criteria so you can assign the data’s
destinations.
The criteria fields are listed on the CR System by priority in the following
sequence (highest priority listed first):
– Cassette Size
– Body Part
– Projection
– Department
Therefore, a profile with a Cassette Size filter appears before any other profile.
When an exam is routed, the information contained in the Patient Record is
compared to the information contained in the criteria fields of all defined
profiles, similar to performing a query. The System assigns destinations based
on the first profile match it finds. If the image matches more than one profile,
the System assigns the first profile in the list that matches all criteria.
Default Profile A default profile is defined automatically by the System. It appears in the list of
Configuration profile destinations with an (*) asterisk next to the name. You can rename the
default profile, but you cannot delete it.
There must be only one default profile. If the data doesn’t match the criteria
for any of the existing profiles, the system uses the default profile destination.
Configuring Profiles 1. At the Key Operator Functions menu, select System Configuration,
then select Profile Destination Configuration.
NOTE: There is only one default profile, set automatically by the system. You
can rename the default, but it is the default profile used when all
other criteria fail.
3. To set up a profile that routes all images created with specific Cassette
Sizes to a specific destination:
a. Enter the name of the profile into the Profile Name field.
b. Enter a specific Cassette size into the Cassette Size field.
c. Leave the remaining fields blank.
d. Select the destinations you want.
e. Select Save Profile.
f. Repeat steps b and c for another Cassette size, if desired.
4. To set up profiles for Departments, repeat steps 2 and 3, completing only
the Department field. In this case, you can enter a unique name to
assign the department with the department names set up for the site.
5. To set up profiles for specific Body Part/Projection combinations, repeat
steps 2 and 3 completing both the Body Part and Projection fields with
the specific combinations for each different profile.
6. To set a single destination for all images, add a destination to the default
profile.
IMPORTANT: When selecting an Archive destination, Select
so patient data must go there. Incomplete or unassigned
data cannot go to mandatory locations. See “Required
Fields” on page 96 .
Text Box
Configuration
External Text Boxes IMPORTANT: Due to the complexity of configuring text boxes, use this
feature only with printers that have been tested and
approved by Kodak. If you are using this feature on a
printer with unknown characteristics, be sure to fully test
on site to avoid unexpected results.
The text in an External Text Box is printed outside the image area and does
not cover up anatomy. However, the box does require film area, thereby
reducing the overall size of the printed image. If True-size mode is selected,
then the system crops additional image data as compared to printing the
image without the External Text Box.
The CR System sends the External Text Box as a separate image file. The text
box you use depends on the capability of the printer.
An External Text Box:
• Shrinks the image to fit - may not work with all printers.
• Impacts True-size printing.
• Prints an External Text Box across the bottom of the film.
Single-Image External Multi-format Individual Multi-format Global
Text Box External Text Box Page External Text Box
External • Reduced image • Reduced image • Reduced image
Text Boxes size size size
• Located outside • Located outside • Located outside
of image area of image area of image area
• May not work • May not work • May not work
with all printers with all printers with all printers
• Applied per • Applied to a
image sheet of
multi-format
images
Internal Text Box The CR System creates the Internal text box as part of the image and sends one
file to the printer. The Internal text box works with every printer.
• You select the text box location during image review.
• The technologist can place the text box in one of eight locations during
image review.
Single Image Internal Multi-format Image
Text Box Internal Text Box
Internal • Larger area for • Larger area for
Text Boxes text text
• Works with every • Works with
printer every printer
• Located in the that supports
image area multi-format
printing
• Located in the
image area
External Text Box Use the following table to determine the capabilities of annotation text boxes
Characteristics supported by various Kodak printers.
Printers Single- Image Multi-Format Multi-Format
External Text Page External Image External
Box Text Box Text Box
Kodak PACS Link 9410 Acquisition System / Y Y N
DryView Laser Imagera 1 line of 64 characters 1 line of 64
maximum characters
DICOM annotation DICOM annotation
Box is located above
the image.
MIM / DryView V 5.x or later Y Y Y
1 or 2 lines of 120 1 or 2 lines of 120 1 line of 120
characters characters characters
Row symmetricb Row symmetric Row symmetric
MIM / Kodak Ektascan 2180 Laser Printer Y Y Y
V 5.x or later 1 or 2 lines of 120 1 or 2 lines of 120 1 or 2 lines of 120
characters characters characters
Row symmetric Row symmetric Row symmetric
MIM/KELI / Kodak Ektascan 160 Laser Y Y Y
Imager V 5.x or later 1 or 2 lines of 120 1 or 2 lines of 120 1 or 2 lines of 120
characters characters characters
Row symmetric Row symmetric Row symmetric
a. External text boxes should not be used with PACS Link products running less than Version 5.0 Software. Previous versions of
PACS Link Software may stop processing all print jobs if an image with external text is received from a CR system.
b. Row Symmetric: In a multi-format image, the images on one sheet must be the same height within each row, but each row
may be a different height.
Choosing a Text Box If you are going to use a single image, you can use either a Print Internal or a
Print External Text Box, or a combination of both.
For multi-format images, the text boxes appear on each image. Place
information that pertains to that image, such as Exposure Index, window and
level, projection, etc.
Place only study-specific data in Global text boxes.
Use Print External Text Boxes with PACS Link Medical Imaging products
running Version 5.0 or higher.
Magnification Factor The Key Operator can configure the system so that the image magnification
factors are included in Print Internal and Print External text boxes. If an
image is printed True-size, then the magnification factor is 100%. Minified
images have magnifications of less than 100%. Magnification factors are not
available for text boxes for Multi-format images.
Using the Text Box The text box is fully configurable; you can display only the parameter you
Editor wish.
There is a text box editor for Print Internal and Print External text boxes.
Internal Text Box All internal text boxes are burned into the image area.
Editor • Text Box Viewer—displays the contents of the text box.
• Field Editor—create new fields, edit the contents of existing fields.
• Navigation Controls—navigate to fields, add rows and columns, or
restore text box.
• Restore Large Text Box—provides a default for a large text box or
restores text box.
• Restore Small Text Box—provides a default for a small text box or
restores text box.
• Delete entry—deletes the highlighted entry when editing a text box.
Text Box Viewer Each field is represented by question marks. Each field can also have a label.
Each row is shown in the font size you select for the row.
When you select a field, the field background color changes to purple.
When you add a new field, the field background color changes to green.
Navigation Controls Use the Navigation Controls to move between fields, and add new rows and
columns to create new fields in the text box.
Adding a Column or a 1. Select the area where you want to add a new row or column in the Text
Row Box Viewer or use the arrows and navigate to the location.
2. Select Add New Row or Add New Column, or move all the way to the
right or bottom.
Deleting or Saving a Field • Delete a field—Select the field to delete and select Delete Entry. The
CR System automatically saves the change.
• Save a field—Select Save Entry after each field change.
External Text Box The External Text Box Editor is similar to the Internal Text Box Editor
Editor with the following exceptions:
• There is no Add Row button because the External text box is a fixed size.
Depending on the capability of the printer, use two lines x 120 characters
wide, or one line of 64 characters if the printer is not Row Symmetric.
• The top half of the Text Box Viewer shows the position of every character.
• The Font Size, Justification and Option fields are removed.
Saving and Restoring You can save your text box configuration options to a disk. See “Restoring
Configuration Options Configurations” on page 14.
Configuring a Text 1. Decide what type of text box you want to print.
Box 2. Decide what content you want in the text box.
• Page External—Study-specific data such as Patient ID, Accession
Number, etc.
• Image—Image-specific data such as Exposure Index, Window/Level,
etc.
NOTE: For Multi-format Images, place fields that have common information
in the Global External text box.
3. Select Key Operator Functions menu, select System Configuration,
select Next, then select Text Box Configuration.
NOTE: For the last three options in the table, the Key Operator sets the
default, which the technologist can override.
X X
Delivery Preferences
Delivery Preferences
Delivery Option 1. At the Key Operator Functions menu, select System Configuration,
Configuration Next, then Delivery Preferences.
2. Select Delivery Options.
Delivery Options
3. Use the following table to select from and select Save Changes.
Configuring Default 1. At the Key Operator Functions menu, select System Configuration,
Hospital Name and select Next, then Delivery Preferences, then Default Values.
Address 2. Enter the Hospital Name and Address and select Save Changes.
CAD Workstation 1. At the Key Operator Functions menu, select System Configuration,
Configuration select Next, then Delivery Preferences, then CAD Workstation
Configuration.
NOTE: All softcopy display stations appear on this screen. To define one as a
CAD Workstation, select the check box. Only unprocessed data is sent
to CAD Workstations.
2. Select the appropriate workstation and select Save Changes.
NOTE: If you have a CR 500 System with a DVD/CD Writer (also known as
LOCALHOST), you must check the LOCALHOST check box in order
for images to be archived in a format that can be retrieved by the
DirectView system.
CAD Workstation
Configuring External
Devices
Remote Operator Panel At the External Devices menu, you can configure up to 10 external devices
Expanded Connectivity such as Remote Operator Panels (ROPs) and computers for Remote Patient
Entry and Remote Key Operator access.
You can set up ROPs to communicate with multiple DirectView CR System
servers or to link to Web sites.
You can configure:
• multiple ROPs to communicate with a single CR System
• a single ROP to communicate with multiple CR Systems
• multiple ROPs to communicate with multiple CR Systems
ROP
IP 192.168.0.18
CR 3
CR 1 CR 2 (primary)
IP 192.168.0.15 IP 192.168.0.16 IP 192.168.0.17
ROP ROP
IP 192.168.0.18 IP 192.168.0.19
CR 3
CR 1 CR 2 (primary)
IP 192.168.0.15 IP 192.168.0.16 IP 192.168.0.17
Configuring a Remote In the following scenario, you could use CR3 (primary) to configure the ROP
Operator Panel to to communicate with CR 1, CR 2, and CR 3.
Multiple CR Systems
ROP
IP 192.168.0.18
CR 3
CR 1 CR 2 (primary)
IP 192.168.0.15 IP 192.168.0.16 IP 192.168.0.17
3. Enter the ROP IP address and select Save Changes. The Links button
appears.
4. Select Links. The ROP’s IP address appears at the top of the Remote
Operations Panel Link Configuration screen.
ROP IP Address
5. Select the http:// field and then use the virtual keyboard to enter the CR
System IP address. For a Web site, enter the URL, for example,
www.kodak.com.
6. Select the associated Text: field and use the virtual keyboard to enter the
CR System name or Web site name.
7. Repeat for other CR Systems or Web sites.
8. Select Update ROP to apply the changes to the ROP.
9. At the other CR Systems, enter the IP address of the ROP.
a. Log on as the Key Operator.
b. At the Key Operator Functions screen select System
Configuration, Next, External Devices, then Kodak Remote
Operation Panel Setup.
c. Enter the ROP IP address and select Save Changes.
d. Select Main Menu.
At the ROP, reload the Main Menu.
– If you are not at the Main Menu, select Main Menu.
– If you are at the Main Menu, select one of the menu selections to go to
another screen, then select Main Menu.
The following is an example of a ROP screen configured for multiple servers.
Configuring Multiple In the following scenario, you could use CR 3 (primary) to configure the
ROPs to Multiple CR ROPs to communicate with CR 1, CR 2 and CR 3.
Systems
ROP ROP
IP 192.168.0.18 IP 192.168.0.19
CR 3
CR 1 CR 2 (primary)
IP 192.168.0.15 IP 192.168.0.16 IP 192.168.0.17
4. Add the ROP IP addresses for all the ROPs and select Save Changes.
The Links button appears.
5. Select Links for the first ROP. The IP address at the top of the screen is
the IP address of the ROP.
ROP IP Address
6. Select the http:// field and then use the virtual keyboard to enter the CR
System IP address. For a Web site, enter the URL, for example,
www.kodak.com.
7. Select the associated Text field and enter the CR System name or Web site
name.
8. Enter the IP address and name for the other CR Systems or Web sites.
9. Select Apply To All ROPs. This copies the http:// link and link name to
each of the ROP configuration screens.
10. Select Update ROP.
11. Select Back.
12. Select Links for the next ROP.
13. Select Update ROP.
14. Repeat steps 11-13 for each ROP.
Remote Patient Data Remote Patient Data Entry Software is a software option for entering patient
Entry Software data at a workstation. It is installed on the CR System and enables
(RPDES) communication between a customer-supplied PC workstation and the CR
System.
IMPORTANT: A RPDES must have a fixed IP address. If the PC is
configured to use DHCP (dynamic IP address assignment
on boot), the link with the CR System may be lost when the
PC is rebooted.
You can add up to ten external devices.
CR
IP 192.168.0.30
Installation Instructions All software options are installed at the Option Registration screen.
At the Key Operator Functions menu, select System Configuration,
Option Registration, then Add Upgrade Options using the appropriate
floppy disk.
RPDES Configuration 1. At the Key Operator Functions menu, select System Configuration,
Instructions then select Next, then select External Devices.
2. Select the Remote Patient Data Entry Station (RPDES). The input
area for 10 RPDES appears.
Create a RPDES Shortcut In the following example, assume the CR System IP address is 129.126.6.62.
on the Desktop
1. On the desktop, right-click and point to New and select Shortcut.
2. Click Browse and locate IEXPLORE.exe executable file.
• Example: C:\Program Files\Plus!\MicrosoftInternet\IEXPLORE.EXE.
3. Add your CR System IP address to the end of the line:
4. Click Next.
5. Select a label for the new shortcut.
6. Click Finish.
7. Double-click the shortcut to confirm it operates correctly.
8. Press Ctrl-W to exit.
NOTE: If the Security Administrator has configured the system so that the
Technologist logs on with a user name and password, you must
always provide a Technologist (or greater) password to log on to
RPDES.
Logon
Main Menu
Downloading Statistics You can download technologist statistics and save the data.
1. Select Download Statistics.
2. Select a hyperlink time period. A file opens in Microsoft Excel.
3. Save the file.
Procedure Mapping See “Procedure Mapping Using the Remote Key Operator” on page 31.
Saving and Restoring You can save and restore a System Configuration file remotely. To save the
Configurations configuration file:
1. Click System Configuration.
2. Click Save Configuration.
3. Click Configuration.txt.
4. Save the file.
Save Configuration
Restore Configuration
Bar Code
Configuration
Overview If necessary, use the Bar Code Configuration menu to change the Patient
ID, Accession Number and Tech ID bar code characteristics for your site.
IMPORTANT: If you change the Cassette ID format, the country code, or
bar code reader prefix or suffix characters, then you must
program the Bar Code Scanner.
You can:
– change the data field size
– change the prefix and suffix characters
– strip a field when a bar code is read
– change the cassette ID format.
NOTE: You don’t need to make any changes unless the CR System Bar Code
Scanner settings are incompatible with your bar codes.
Changing the Bar Code 1. At the Key Operator Functions menu, select System Configuration,
Characteristics then select Next, then Bar Code Configuration.
2. Select Bar Code Configuration again.
NOTE: There are two bar code configuration screens: a Basic and an
Advanced Editor Screen. Select Advanced Bar Code Editor to use
the advanced screen.
3. For each Field, enter the desired Prefix and Suffix Character formats
and set the Data field size.
• Select a field and use the virtual keyboard to enter the data.
• Leave the Data Size field blank if you want to accept any size data.
• For the advanced screen, select Strip if you want to remove the field
when it is read by the bar code scanner.
4. Scan the bar code with the bar code scanner to test it. A message appears
with the test results.
5. When the codes have tested successfully, select Submit Changes.
NOTE: Test the cassette ID also to assure that there are no bar code conflicts.
Changing the Cassette ID If the Cassette ID Bar Code format (normally 10 digits in length) conflicts with
Bar Code Format other bar codes in use, you can change the CR System Bar Code format to an
11-digit format.
1. At the Bar Code Configuration screen select Modify Cassette Format.
2. Select the 11-digit Bar Code format check box.
Cassette ID Format
6. Use the bar code scanner and scan each bar code in numerical order
and then select Next.
IMPORTANT: If you have trouble scanning the bar codes from the
monitor, bar codes are available later in this section of the
manual. However, make sure the bar code names match
the names on the monitor.
7. Repeat for the next two screens.
8. When done scanning, select Next. A bar code appears to test the
configuration.
9. Scan the bar code. A message confirms whether or not the test was
successful.
NOTE: Do not scan a bar code more than once. If you scan twice, it will fail.
Changing the Country Use the Bar Code Configuration menu to program the country data. The
Code screens to configure the bar code scanner differ depending on the country.
To program the scanner, set the country code and then program the bar code
scanner.
1. At the Bar Code Configuration menu, select Set Country Code.
Changing the Prefix and NOTE: Only make this change if an [or] appears in the hospital’s bar cod.e.
Suffix This prefix/suffix does not appear in the actual bar code. It is added
by the scanner.
1. At the Bar Code Configuration menu, select Change Prefix & Suffix.
2. Enter two characters that do not appear in any of your bar codes.
3. Select Submit.
4. Select Program Bar Code Scanner.
5. Use the bar code scanner and scan each bar code in numerical order
and then select Next.
IMPORTANT: If you have trouble scanning the bar codes from the
monitor, use the bar codes that are available later in this
section of the manual. However, make sure the bar code
names match the names on the monitor.
6. Repeat for the next screen.
7. When done scanning, select Next. A bar code appears to test the
configuration.
8. Scan the bar code. A message confirms whether or not the test was
successful.
NOTE: Do not scan the test bar code more than once. If you scan it twice, it
will fail.
Save
Number “0”
Number “1”
Number “2”
Number “3”
Number “4”
Number “5”
Number “6”
Number “7”
Number “8”
Number “9”
Number “A”
Number “B”
Number “C”
Number “D”
Number “E”
Number “F”
Manual Trigger
Automatic Trigger
Test
Color Preferences You can change the default colors for field backgrounds and buttons.
1. At the Key Operator Functions menu select System Configuration,
then select Next.
2. Select Color Preferences.
3. Select the down arrow to select the button or text you want to change
color.
4. The button color is determined by the code values in the Red, Green
and Blue columns (0 = 0% of that color, 255 = 100% of that color).
Select a code value and a pop-up keypad appears.
Color R G B
White 255 255 255
Black 0 0 0
Yellow 255 255 0
Blue 0 0 255
5. Enter the new code value. Repeat for the other two code values if
required.
6. If you want to change other field background colors, repeat for each
field.
7. Select Submit to return to the Color Preferences screen.
8. At the Color Preferences screen, select Submit.
Configure Colors
Required Fields You can set required fields for delivery to mandatory destinations such as
archives, or for submitting exam/patient information. For example, you may
want to make Tech ID a required field if you are using Reject Reason logging.
Making a field Required to Submit guarantees that the technologist will enter
all required Exam/Patient Information at the time the patient record is
submitted to the database.
Using Required Fields for delivery lets you send images that have
incomplete patient demographic information to workstations and printers but
not to mandatory destinations. Once the required information is entered and
the image is accepted, the CR System forwards the image to the mandatory
destination.
1. At the Key Operator Functions menu, select System Configuration,
then select Next.
2. Select Required Fields.
Required Fields
NOTE:
• If you select Required to Submit Exam/Patient Information,
you can only submit the PEC data when all the required fields are
complete.
• If you select Required to Deliver Images to Mandatory
Destinations, you can submit the PEC data without the required
fields completed. You have to complete the fields before
submitting to mandatory destinations.
NOTE: The Contrast/Bolus and Date of Birth fields should not be mandatory
if the system is using enhanced trauma functionality. Doing so causes
added steps in the workflow. These two fields cannot be defined as
part of the setup for a trauma patient, and if they are required, the
technologist must complete these fields manually after selecting a
trauma patient but before the record is submitted to the database or
the image is routed to mandatory destinations.
System Maintenance This screen lets you determine when to erase images and when to delete
Defaults patients with no images from the system.
1. At the Key Operator Functions menu, select System Configuration,
then select Next and Next again.
2. Select System Maintenance Defaults.
Regional Settings This screen lets you set the CR System date, time, and language settings.
1. At the Key Operator Functions menu, select System Configuration,
then select Next, and Next, then select Regional Settings.
2. Make your selections and then select Save Changes.
Configure Monitor You can optimize the image display for the viewing conditions at the monitor’s
location. To do this, use the Configure Monitor screen to select the monitor
and ambient settings. Use the SMPTE Test Pattern to adjust the contrast and
brightness.
1. At the Key Operator Functions menu, select System Configuration,
Next, Next, and select Configure Monitor.
2. Select the monitor from the list. If your monitor does not appear in the
list, see your Kodak Service Representative.
3. Select the amount of ambient light in the CR environment.
NOTE: Ambient light is subjective. Typically, select High Ambient Light. Try
to shield as much ambient light as possible when making this
adjustment.
4. Select Submit.
5. To adjust the contrast and brightness, from this screen, select Back and
select SMPTE Test Pattern. (From the Key Operator Functions
6. Select the menu button on the monitor. Use the down arrow to select
brightness.
7. Adjust brightness with + and - controls to the lowest level while you can
still see contrast on either side of the circle in the dark (0 - 5%) target.
8. Use the down arrow key to select contrast.
9. Adjust contrast with + and - controls to the highest level while you can
still see contrast on either side of the circle in the bright (95 - 100%)
target.
NOTE: It may be necessary to adjust the screen several times before you
reach the optimum setting.
Miscellaneous Use the Miscellaneous screen to set your default Body Part and Projection.
1. At the Key Operator Functions menu, select System Configuration,
select Next and Next again, then select Miscellaneous.
SMPTE Test Pattern You can use a SMPTE Test Pattern to analyze image quality problems and to
change the contrast and brightness of the monitor. See “Configure Monitor”
on page 98.
1. Select Display SMPTE Test Pattern from the Key Operator
Functions menu.
2. Using the controls at the bottom of the screen, create a test pattern for
any one of these cassette sizes:
– 18 x 24 cm
– 24 x 30 cm
– 35 x 35 cm
– 35 x 43 cm
3. Select the appropriate button for the cassette size you wish to print.
4. Select the destination for the image. You can exit the screen when the
printing message appears.
5. You can graph the steps of the SMPTE pattern to determine if the image at
the destination is presented linearly.
Long Length Imaging This screen lets you configure the system for Long-length Imaging exams.
– Select the check box to use QA mode for all long-length composite
images.
– Select the check box to automatically set the status of all sub-images to
Delivered once the composite image is delivered.
– Set the default Source to Image Distance (SID.
1. At the Key Operator Functions menu, select System Configuration,
then select Next and Next again.
2. Select Long-Length Imaging.
Calibrate Touch 1. At the Key Operator Functions menu, select System Configuration,
Screen select Next, Next, and Next again.
2. Select Calibrate Touch Screen.
To calibrate the touch screen, select the red targets as they appear on the
screen. Then select the screen to check the cursor’s response.
CAUTION:
Selecting the screen in areas other than the red targets may
cause the select screen to behave in an unpredictable manner.
B R
bar code configuration reject image, 5
change prefix and suffix, 85 Remote Key Operator
changing, 81 color preferences, 95
modify cassette format, 83 create new procedure, 34
set country code, 84 deleting a procedure, 34
bar code samples, 88 download statistics, 78
editing procedures, 33
D login, 76
delivery preferences procedure mapping, 31, 33, 78
include CT pixel spacing, 60 set up, 76
multi-format, 60 RPDES, 72
option configuration, 59 configuration, 73
QA mode, 60 creating a desktop shortcut, 75
upper case DICOM fields, 60 installation instructions, 72
destination, 5 operation, 74
requirements, 72
E
S
EHR Scanning, 17
external devices scheduled workflow configuration, 103
multiple ROPs, 69 SMPTE Test Pattern, 101
Remote Key Operator, 76 statistics
ROP expanded connectivity, 63 cassette, 4
ROP to multiple CR systems, 64 destination, 5
destination status summary, 9
K erase cycles, erase lamps, 5
scan cycles, 8
Key Operator Functions
technologist, 6
managing patient exam records, 2
system configuration
system configuration, 10
bar code configuration, 81
P buttons, 19
CAD workstation, 62
procedure mapping, 25 calibrate touch screen, 104
creating a new procedure, 29 configuring profiles, 48
deleting a procedure, 30 CR display, 41
editing, 30 default hospital name, 61
more than 34 procedures, 29 default profile, 47
using a HIS/RIS, 30 delivery preferences, 59
using the CR System, 26 department, physician list, 19
using the Remote Key Operator, 31 external devices, 63
HIS/RIS Broker, 35
I-1
maintenance defaults, 97
menu, 11
option registration, 17
procedure list, mapping, 24
profile destination, 47
Push configuration, 38
regional settings, 98
reject reason, 43
required fields, 96
restoring, 14
saving, 13
trauma defaults,setting, 45
workflow optimization, 18
T
technologist, 5
text box editor
configuring, 56
external, 55
fields and definitions, 53
internal, 53
saving and restoring, 55
text box viewer, 53
add new column or row, 55
navigation controls, 55
save or delete a field, 55
text boxes
choosing, 52
external, 50
internal, 51
touch screen
calibrating, 104
U
unique numbers, using, 46
2