726 Calibrador de Procesos
726 Calibrador de Procesos
726 Calibrador de Procesos
Change #1
On page 2, replace the Pulse row with the following:
1-100,000 1-10,000
Pulse
Frequency Max 15 kHz Frequency Range 2 CPM to 15 kHz
On page 12, Table 4, replace the Description for Number with the following:
Cycles through:
Slow repeating 0 % - 100 % - 0 % ramp
Configurable repeating 0 % - 100 % - 0% ramp
Configurable repeating 0 % - 100 % - 0 % ramp in 25 % steps
Used for the pulse train and totalizer functions.
9/20 1
Manual Supplement 726 Users
2 7/15
726 Users Manual Supplement
On page 14, under Configuration Menus, add the following under the last bullet:
• Step time
• Ramp time
On page 15, under Shut Down Mode, replace numbered steps with the following:
1. Press until SHUT DOWN appears on the display.
2. Use and to increase or decrease the time.
3. Use and to turn on and off.
4. Press to save the setting.
On page 16, under Hart Resistor ON/OFF, replace step 2 with the following:
2. Use or and to toggle ON and OFF.
Add the following below the Note:
Step Time
Step Time sets the ramp step time from 1 sec to 99 sec.
1. Press until STEP TIME appears on the display.
2. Use and to set the step time.
3. Press to save the setting.
9/20 3
Manual Supplement 726 Users
Ramp Time
Ramp Time sets the ramp time from 5 sec to 99 sec.
1. Press until RAMP TIME appears on the display.
2. Use and to set the ramp time.
3. Press to save the setting.
On page 25, Table 6, add the following:
Cu10 10 Ω Copper 0.0042 Ω/°C -100 to 250
On page 42, under Auto Ramping the Output, replace the 2nd and 3rd bullets with the following:
• 0 % - 100 % - 0 % configurable time smooth ramp. Set ramp time using configuration
menu.
• 0 % - 100 % - 0 % Stair-step ramp in 25 % steps, pausing at each step. Set ramp
time using configuration menu. Steps are listed in Table 7.
On page 60, under Frequency Measurement, add the following to the bottom of the table:
Sensitivity: 1 V peak to peak minimum
Waveform: Squarewave
4 7/15
726 Users Manual Supplement
On page 62, change the Range entries for “Thermocouple in mV read” and “Thermocouple in
mV source”:
From: -10 °C to 75 °C
To: -10 mV to 75 mV
On page 63, under the RTD Accuacy (Read and Source) (ITRS-90) table, replace the CU10 row
with the following:
Cu10 -100.0 250.00 1.8
On page 64, under Pulse Read and Pulse Source, replace the Frequency entry with the
following:
2 CPM to 15 kHz
9/20 5
Manual Supplement 726 Users
6 7/15
726 Users Manual Supplement
Change #3
On page 54, Table 8:
Change:
Test lead, red 688051 1
8
Test lead, black 688066 1
To:
8 Fluke-7XX Test Lead Set 3397308 1
7/15 7
Manual Supplement 726 Users
On page 65, under General Specifications, replace the Safety section and add:
Electromagnetic Compatibility
(EMC)
International IEC 61326-1: Portable Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio
frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and
those directly connected to a low-voltage power supply network that supplies buildings
used for domestic purposes. There may be potential difficulties in ensuring
electromagnetic compatibility in other environments due to conducted and radiated
disturbances.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment
is connected to a test object.
8 7/15
726 Users Manual Supplement
To this (remove the third ‘0’ after the decimal point in each Minimum and Maximum cell):
Range Minimum Maximum Accuracy (% of Reading + Floor)
100 mV (Source) 0.00 100.00 0.010 % + 10 µV
90 mV (Read) 0.00 90.00 0.010 % + 10 µV
Maximum current output in voltage ranges is 1 mA with an output impedance of ≤ 1 Ω
7/15 7