Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

BAM-1020-9800 Manual Rev F

Download as pdf or txt
Download as pdf or txt
You are on page 1of 81

Met One Instruments, Inc.

1600 NW Washington Blvd, Grants Pass, OR


TEL (541) 471-7111 Fax (541) 471-7116
metone@metone.com
Title:

BAM-1020 OPERATION MANUAL


Drawing Number: Revision:
BAM-1020-9800 F
Originator Approval: Checker Approval: Document Control:

REVISIONS
REV DESCRIPTION DATE BY
A EO 1624. This revision supersedes previous Rev 2.0. 03-29-01 -
Added discussion on BX-961 volumetric control.
Added discussion on Standard and Early Cycle Mode.
Updated Field Calibration of Flow System.
Effectivity: This revision manual required with 3236 Rev. C (Firmware 02.03).
B EO 1733: Modifies Caution Statement, Section 4.3, Section 6.3. 01-31-01 -
C Added this page (Title Block, etc...) 06-11-01 JJE
Section 4.14: Changed definition of FLOW TYPE.
Section 5.5: Added description on how the flow is reported.
D 1JAN02 MH
1. Section 2.3 – Details of the options for enclosures. Added siting for inlet radius etc. Steps 1-5
2. 2.4 added note on complete instruction for the BAM 1020.
3. Added Section 2.6
4. Equation 3.3 – units analysis shows the equation is incorrect. Added mu coefficient.
5. Section 3.2 – Clarified the span measurement purpose. Changed compensation to verification.
6. Section 3.2.3 added reference to Appendix B and spec’d the tolerances of the span
measurement
7. Section 3.2.4 Changed PM10 inlet filter to PM10 inlet head.
8. Section 3.3 added a reference to Appendix B for the factory values.
9. figure 6 – added flow screen, Added align screen
10. Added a heading to figure 7 BAM 1020 grounding terminals
11. Added Section 4.8 Leak Check
12. Section 4.13 minor grammar mistakes
13. Section 4.14 clarified flow type.
14. Section 4.16.1 added references to section 5.2 Appendix C.
15. Clarified sec5.2 note about function keys
16. Added details to cal data flag in section 5.2 and 5.6
17. Changed the wording and clarified flow type table in section 5.5
18. Added a section direction to 5.7
19. Reworked Section 6.3
20. Added Section 6.4
21. Added explanation of ABS in App B to Sect7.7
22. Added Parity switch info to Sec 8.4
23. Added Section 4.9 identifier in sect 9.1
24. Corrected the section identifier in section 9.1.2
25. Identified the Appendix in 9.1.4
26. Corrected the section identifier in section 9.1.5
27. Moved troubleshooting guide to sect 9.2
28. Added periodic maintenance to section 10.
29. Added Consumables and Optional Equip Sect 10.3

BAM-1020-9800 REV F
E 1. PG12 heater Kit info 25FEB02 MH
2. PG40 Heater in the setup screen with explanation below
3. PG40 Heater Control Details
4. PG46 Changed the Error Screen reflect new FW
5. PG56 HEATER section 4.20
6. PG82 HEATER SCREEN in TEST Menu Section 7.11
7. PG90 Section 8.6 Added 6 CSV DATA Output
8. PG99 Added Smart Heater to list of BAM1020 options BX-827 and BX-830
9. PG105 Changed 6 – none to 6 – CSV DATA Output.

F Major revision of manual. Updated all sections for accuracy and completeness from customer and 12-20-2006 DAH
Service dept feedback. Fixed countless mistakes, omissions, obsoletes. Reorganized section structure for
better flow and logical layout. Made operational changes for PM2.5 EPA equivalency testing and new 8- 3-27-2007
minute firmware. Effectivity: This revision is for all units with 3.2.4 or later firmware and all 2X
hardware improvements.
• Incorporated changes from 2X mechanical improvements.
• Revised introduction. Added brief beta description, left space for PM2.5 EPA designation.
• Revised Installation section. Site selection, enclosure selection, mounting options.
• Revised accessories section. Removed obsolete items, added items and pics, reorganized.
• Revised errors section. Greatly expanded description of all errors.
• Reworked specifications: updated for 2X units, clarified!! Moved to front.
• Revised 500 series sensor connection section. Was Appendix F.
• Moved the advanced theory of operation section to the back where it belongs.
• Revised flow calibration section, moved leak check and nozzle cleaning to flow cal section.
• Incorporated my flow type description sheet. Moved flow section forward.
• Revised maintenance schedule. Added items. Changed to table. Moved to Troubleshooting.
• Created user interface drawing and simplified the ridiculous keypad description!
• Rewrote SETUP section, moved important setups to front after install.
• Rewrote TEST section, moved to Troubleshooting section.
• Rewrote Communications section, added HyperTerminal setup, simplified command list.
• Moved Telemetry section to its own section, revised for clarity.
• Replaced HUGE photos, reduced manual file size from 30MB to 2.7MB!
• Simplified early cycle mode description, moved to telemetry section.
• Added Audit Sheet and notes page.
• Revised Troubleshooting information.
• Reduced from 113 pages to 80 pages.
• Moved all appendices to body text.
• Added new changes for 3.2.4 PM2.5 firmware.
• Added new count times, flow cal, flow stats, offsets, data outputs, flash upgrade from 3.2.4.
• Added field BKGD determination procedures.
• Added BX-596 AT/BP sensor information.
• Added Dr. Schloesser’s concentration STD/Actual conversion formulas to Theory section.
• Added suggestions from Mike Putnam on communication, accessories, etc.
• Added Jeff Elseth’s corrections for telemetry relays and errors.
• Added EPA PM2.5 sampling requirements.
• Refined CONC TYPE and FLOW TYPE sections.

Printing Instructions:
• Do not include the revision block pages as part of the printed manual (this page).
• Print double sided final from single sided master. Must be a printout or a high-quality
first generation copy. No copies of copies!!!
• Include standard Met One warranty sheet as last page.

BAM-1020-9800 REV F
BAM 1020
PARTICULATE MONITOR
OPERATION MANUAL

BAM-1020-9800 REV F

Met One Instruments, Inc.


1600 NW Washington Blvd.
Grants Pass, OR 97526
Telephone: (541) 471-7111
Facsimile: (541) 471-7116
www.metone.com

BAM-1020 Particulate Monitor Operation Manual - © Copyright 2001, 2007 Met One Instruments, Inc. All Rights
Reserved worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or
translated into any other language in any form without the express written permission of Met One Instruments, Inc.

BAM-1020-9800 REV F
Table of Contents
1 INTRODUCTION 4
1.1 About This Manual ..................................................................................................... 4
1.2 Technical Service....................................................................................................... 4
1.3 BAM: Beta Attenuation Monitor .................................................................................. 4
1.4 Beta Radiation Safety Statement ............................................................................... 5
1.5 Model BAM-1020 PM10 USEPA Equivalent Method................................................... 5
1.6 Model BAM-1020 PM2.5 Monitoring ............................................................................ 6
1.7 BAM-1020 Specifications ........................................................................................... 7
2 SITE SELECTION AND INSTALLATION 8
2.1 Unpacking and Inspection .......................................................................................... 8
2.2 Enclosure Selection ................................................................................................... 8
2.3 Site Selection ............................................................................................................. 9
2.4 Mounting Options in a Walk-In Shelter..................................................................... 10
2.5 Installation Instructions in a Walk-In Shelter ............................................................ 12
2.6 Electrical Connections.............................................................................................. 13
3 INITIAL SETUP OF YOUR BAM-1020 15
3.1 Power On ................................................................................................................. 15
3.2 Warm-up .................................................................................................................. 15
3.3 Using the Keypad and Display ................................................................................. 16
3.4 Filter Tape Loading .................................................................................................. 17
3.5 Self-Test................................................................................................................... 18
3.6 Initial SETUP Settings Considerations ..................................................................... 19
3.7 Initial Leak Check and Flow Check .......................................................................... 20
3.8 Starting a Measurement Cycle ................................................................................. 20
3.9 The Flow Statistics Screen....................................................................................... 20
3.10 The OPERATE Screen ............................................................................................ 21
3.11 The NORMAL Screen .............................................................................................. 21
3.12 The INSTANTANEOUS Screen ............................................................................... 22
3.13 The AVERAGE Screen ............................................................................................ 22
4 THE MEASUREMENT CYCLE 23
4.1 The One-Hour Cycle Timeline.................................................................................. 23
4.2 Automatic Span Check During The Cycle ................................................................ 24
4.3 Sample Period Description....................................................................................... 25
5 FLOW SYSTEM and FLOW CALIBRATIONS 26
5.1 Flow System Diagram .............................................................................................. 26
5.2 Flow Type Descriptions............................................................................................ 27
5.3 Leak Check Procedure............................................................................................. 28
5.4 Leak Check Nozzle Seal .......................................................................................... 29
5.5 Nozzle and Vane Cleaning....................................................................................... 29
5.6 Field Calibration of Flow System – Actual (Volumetric) Flow Mode ......................... 30
5.7 Field Calibration of Flow System – EPA Standard Flow Mode................................. 31
5.8 Field Calibration of Flow System – Metered Flow Mode ......................................... 32
6 SETUP MENU DESCRIPTIONS 34
6.1 CLOCK Screen ........................................................................................................ 35
6.2 SAMPLE Screen ...................................................................................................... 35
6.3 CALIBRATE Screen................................................................................................. 37
6.4 EXTRA1 Screen....................................................................................................... 39

Page 2 BAM-1020-9800 REV F


6.5 ERRORS Screen ..................................................................................................... 39
6.6 PASSWORD Screen................................................................................................ 42
6.7 INTERFACE Screen ................................................................................................ 42
6.8 SENSOR Screen...................................................................................................... 43
6.9 HEATER Screen ...................................................................................................... 44
7 MAINTENANCE, DIAGNOSTICS and TROUBLESHOOTING 46
7.1 Met One Suggested Periodic Maintenance .............................................................. 46
7.2 Audit Sheet and Test Records ................................................................................. 47
7.3 Self-Test Feature ..................................................................................................... 47
7.4 Power-Up Problems ................................................................................................. 47
7.5 Basic Problem and Cause/Solution Table ................................................................ 47
7.6 Nozzle Component Replacement............................................................................. 49
7.7 Field BKGD Zero Background Tests ........................................................................ 49
7.8 Test Menu System ................................................................................................... 50
7.9 COUNT test Screen ................................................................................................. 50
7.10 PUMP Test Menu..................................................................................................... 51
7.11 TAPE Test Menu...................................................................................................... 51
7.12 DAC Test Menu – Analog Output Test..................................................................... 52
7.13 CALIBRATE Test Menu ........................................................................................... 52
7.14 INTERFACE Test Menu........................................................................................... 53
7.15 FLOW Test Menu..................................................................................................... 53
7.16 ALIGN Test Menu – Photo Sensor Tests ................................................................. 53
7.17 HEATER Test Menu................................................................................................. 54
7.18 FILTER-T Test Menu – Filter Temperature Sensor.................................................. 54
7.19 RH Test Menu – Filter Humidity Sensor ................................................................... 55
8 EXTERNAL DATALOGGER INTERFACE SYSTEM 56
8.1 Analog Concentration Output Signal ........................................................................ 56
8.2 Early Cycle Mode Description .................................................................................. 57
8.3 Telemetry and Error Relays ..................................................................................... 58
9 RS-232 SERIAL COMMUNICATIONS – DATA RETRIEVAL 61
9.1 Serial Port Connections and Settings....................................................................... 61
9.2 Met One Communications Software......................................................................... 62
9.3 Downloading Data Using HyperTerminal ................................................................. 62
9.4 System Menu File Descriptions................................................................................ 63
9.5 Printer Output Port ................................................................................................... 68
9.6 Modem Option.......................................................................................................... 68
9.7 Flash Firmware Upgrades ........................................................................................ 69
10 ACCESSORIES and PARTS 70
10.1 Consumables, Replacement Parts, and Accessories............................................... 70
10.2 Series 500 Sensor Configurations............................................................................ 73
11 THEORY OF OPERATION and MATHEMATICAL ANALYSIS 75
11.1 Converting Data Between EPA Standard and Actual Conditions ............................. 77

BAM-1020-9800 REV F Page 3


1 INTRODUCTION
1.1 About This Manual
This document is organized with most of the important information toward the front of the
manual, such as site selection, installation, setups, and flow calibrations.

The sections and sub-sections marked with an Information Symbol contain critical
information which all BAM-1020 owners and operators should read and understand. Toward
the back are several sections that provide in-depth information on subjects such as theory,
diagnostics, accessories, and alternate settings. These sections are not as important for the
typical user to study, but provide valuable information which should be consulted as needed.

1.2 Technical Service


This manual is structured by customer feedback to provide the required information for setup,
operation, testing, maintaining, and troubleshooting your BAM-1020 unit. Should you still
require support after consulting your printed documentation, we encourage you to contact
one of our expert Technical Service representatives during normal business hours of 7:00
a.m. to 4:00 p.m. Pacific Standard Time, Monday through Friday. In addition, technical
information and service bulletins are often posted on our website. Please contact us and
obtain a Return Authorization (RA) number before sending any equipment back to the
factory. This allows us to track and schedule service work and expedite customer service.
Phone: (541) 471-7111 Fax: (541) 471-7116
E-Mail: service@metone.com Web: www.metone.com
Mail: Technical Services Department
Met One Instruments, Inc.
1600 NW Washington Blvd.
Grants Pass, OR 97526

1.3 BAM: Beta Attenuation Monitor


The Met One Instruments model BAM-1020 automatically measures and records airborne
particulate concentration levels using the principle of beta ray attenuation. This method
provides a simple determination of concentration in units of milligrams or micrograms of
particulate per cubic meter of air. A small 14C (Carbon 14) element emits a constant source of
high-energy electrons known as beta particles. These beta particles are detected and
counted by a sensitive scintillation detector. An external pump pulls a measured amount of
dust-laden air through a filter tape. After the filter tape is loaded with ambient dust, it is
automatically placed between the source and the detector thereby causing an attenuation of
the beta particle signal. The degree of attenuation of the beta particle signal is used to
determine the mass concentration of particulate matter on the filter tape, and hence the
volumetric concentration of particulate matter in ambient air. A complete description of the
measurement cycle is included in section 4. In addition, an in-depth scientific explanation of
the theory of operation and the related equations is included at the back of the manual.

Page 4 BAM-1020-9800 REV F


1.4 Beta Radiation Safety Statement
The Met One Instruments BAM-1020 contains a small 14C (Carbon 14) beta radiation-emitting
source. The activity of the source is 60 µCi ±15µCi (microcurries), which is below the “Exempt
Concentration Limit” as defined in 10 CFR Section 30.70 – Schedule A. The owner of a BAM-
1020 may elect to return the entire unit to Met One Instruments for recycling of the 14C source
when the unit has reached the end of its useable life, although the owner is under no
obligation to do so. Under no circumstances should anyone but factory technicians attempt to
remove or access the beta source. The beta source has a half-life of about 5730 years, and
should never need to be replaced. Neither the 14C source nor the beta particle detector are
serviceable in the field. Should these components require repair or replacement, the BAM-
1020 must be returned to the factory for servicing and recalibration.

1.5 Model BAM-1020 PM10 USEPA Equivalent Method


The Met One Instruments, Inc. Model BAM-1020 is designated as an equivalent method for
PM10 by the United States Environmental Protection Agency on August 3, 1998.

Designation Number: EQPM-0798-122

The EPA designation applies to the G, -1 and the G -1 BAM-1020 PM10 Beta Attenuation
Monitor, when used in conjunction with the following requirements. Users are advised that
configurations that deviate from this specific description may not meet the applicable
requirements of 40 CFR, parts 50 and 53.

• The BAM-1020 is operated for 24-hour average measurements with a filter change
frequency of one hour.
• The inlet must be equipped with the standard BX-802 EPA PM10 inlet head.
• The unit must be used with standard glass fiber filter tape.
• The unit may be operated with or without any of the following options: BX-823 inlet
tube extension, BX-825 heater kit, BX-826 230V heater kit, BX-828 roof tripod, BX-902
exterior enclosure, BX-903 exterior enclosure with temperature control, BX-961 mass
flow controller, BX-967 internal calibration device.
• The RANGE parameter must be set for 1.000 milligram/cubic meter.
• The SAMPLE TIME parameter must be set for 50 minutes.

BAM-1020-9800 REV F Page 5


1.6 Model BAM-1020 PM2.5 Monitoring
Starting in 2007, new BAM-1020 units built by Met One Instruments, Inc. have been updated
to meet the United States Environmental Protection Agency’s requirements for PM2.5
monitoring. All of the following settings and accessories must be incorporated when the BAM-
1020 is used to monitor PM2.5 particulate levels.

• The PM2.5 concentration is calculated (external to the BAM) as a 24 hour daily average
of the hourly concentration measurements made by the BAM-1020. At least 18 of the
24 data points must be valid (no error flags) in order to calculate a daily average.
• The unit must be equipped with the BX-596 ambient temperature and barometric
pressure combination sensor. This is used for flow control and flow statistics. The unit
can function with the BX-592 or CARB style temperature sensor instead, but the flow
statistics during the sample will not be available
• The unit must be equipped with the internal BX-961 automatic flow controller, and
must be operated in Actual (volumetric) flow control mode.
• The inlet must be equipped with the standard BX-802 EPA PM10 inlet head and a
PM2.5 very sharp cut (VSCC) cyclone.
• The unit must be equipped with a BX-827 (110V) or BX-830 (230V) Smart Inlet Heater,
with the heater RH regulation setpoint set to 35%.
• The BX-302 zero filter calibration kit is a required accessory. This must be used to
periodically audit the BKGD (background) value as described in this manual.
• The unit must be used with standard glass fiber filter tape.
• The COUNT TIME parameter must be set for 8 minutes.
• The SAMPLE TIME parameter must be set for 42 minutes.
• The unit must be equipped with firmware revision 3.2.4 or later.
• The unit must be equipped with the 8470-1 rev D or later tape control transport
assembly with close geometry beta source configuration. BAM-1020 units
manufactured after March 2007 have these features standard. Older units will have to
be factory upgraded and re-calibrated to the latest hardware specifications.
• The unit may be operated with or without an inlet tube extension (BX-823) and with or
without external enclosures BX-902 or BX-903.
• The BAM-1020 must be operated in proper accordance with this operation manual,
revision F or later.

Page 6 BAM-1020-9800 REV F


1.7 BAM-1020 Specifications
PARAMETER SPECIFICATION
Measurement Principle: Relative Particulate Concentration by Beta Attenuation.
Reference Method: Gravimetric Federal Reference Method. Designation EQPM-0798-122
Standard Range: 0 - 1000 µg/m3 (0 – 1.000 mg/m3)
Optional Ranges: 0 - 0.100, 0.200, 0.250, 0.500, 2.000, 5.000, 10.000 mg/m3
± 2 µg from 0.000 mg to 0.100 mg/m3
Accuracy (24 hours):
2% from 0.100 mg to 1.000 mg/m3
± 5 µg from 0.000 mg to 0.100 mg/m3
Accuracy (1 hour):
± 5% from 0.100 mg to 1.000 mg/m3
Resolution: ± 1 µg/m3
Lower Detection Limit: 3 µg
Measurement Cycle Time: 1 Hour
Flow Rate: 16.7 liters/minute, adjustable from 0-20 LPM. Volumetric or EPA Standard flow.
Filter Tape: Continuous glass fiber filter, 30mm wide, 21 meters long.
Internal membrane calibration, activated automatically every hour, deviations
Span Calibration:
from standard are logged.
Span Stability: ± 5%, verified with built-in calibrator.
Beta Source: 14
C, 60 µCi (< 2.22 X 106 Beq), Half-Life: 5730 years.
Beta Detector Type: Photomultiplier tube with organic plastic scintillator.
Operating Temp. Range: 0 to +40°C
Ambient Temp. Range: -30° to +60°C
Ambient Humidity Range: 0 to 90% RH, non-condensing.
Humidity Control: Actively controlled inlet heater module, 10 to 99% RH.
Approvals: US EPA, CE, NRC, TUV
Human Interface: Menu-driven character display with dynamic keypad.
Display Type: 8 x 40 character LCD with backlight and adjustable contrast.
Analog output: Isolated 0-1 VDC Standard.
0-10 VDC, 4-20mA, 0–16mA switch-selectable.
Serial Interface: RS-232 2-way serial port for PC or modem communications.
Output-only port may be used with an optional 80 column serial printer.
Printer Output:
Interface cable available for parallel printers.
Diagnostic Output: Printer output is configurable as a diagnostic output to a PC.
Clock Reset input (voltage or contact closure), Telemeter Fault input (contact
Telemetry:
closure). Isolated.
Alarm Contact Closures: Data Error, Tape Fault, Flow Error, Power Failure, Maintenance.
Compatible with HyperTerminal®, ProComm Plus®, and most Met One
Software:
Instruments, Inc. communication software.
Error Reporting: User-configurable. Available through serial port, display, and relay outputs.
Memory: 200 days (1-hour sample time). 3+ days (1-minute sample time).
Power Supply: 100/115/230 VAC, 50 or 60 Hz. Factory Configured.
Weight: 21 Kg (46.3 pounds) approximate, excludes pump.
Dimensions 12” (H) x 17” (W) x 18” (D) (31mm x 43mm x 45mm)

BAM-1020-9800 REV F Page 7


2 SITE SELECTION AND INSTALLATION
2.1 Unpacking and Inspection
If any damage to the shipment is noticed before unpacking, a claim must be filed with the
commercial carrier immediately. Notify Met One Instruments after notification of the
commercial carrier.

Remove the unit and accessories from the shipping boxes and compare the received items to
the packing list. Make sure you have all of the required items for the type of installation you
plan to perform.

The BAM-1020 is shipped with two white foam rings and a white plastic shim inside the front
of the unit, which prevent the moving parts of the tape control assembly from being damaged
in transit. Do not remove the foam rings until the BAM-1020 is ready to be installed. The rings
must be replaced anytime the unit is being transported in order to avoid damaging the tape
control mechanism.

Please keep the special box and foam packing material that the BAM-1020 was shipped in, to
re-use in the event that you must return the unit to the factory. Met One is not responsible for
any damages to the unit if returned in non-original packaging or without the foam rings in
place. Contact Met One for replacement packing materials if necessary.

2.2 Enclosure Selection


The BAM-1020 is not weatherproof or water resistant, and is designed to be mounted in a
weatherproof, level, vibration free, dust free, and temperature controlled environment where
the operating temperature is between 0o C and +40o C, and where the relative humidity is not
condensing and does not exceed 90%. There are two standard configurations described
below for providing a weatherproof location in which to install the unit. Please contact Met
One if you have a non-standard mounting or enclosure configuration planned.

1. A walk-in building or mobile shelter with a flat roof: This could be a trailer, pre-
fabricated shelter, or a room in an existing building. The BAM is mounted on a bench-
top or in an equipment rack, often with a variety of other instruments installed in the
same shelter. The inlet tube of the BAM must extend up through the roof with
appropriate hardware. AC power must be available. Instructions for this type of
installation are included in this section of the manual.
2. An environmentally controlled “Dog House” enclosure: A small pre-fabricated
enclosure just big enough for the BAM and related accessories. They are available
with a heater only (BX-902), or with a heater and air conditioner (BX-903). These
enclosures are supplied by Met One Instruments, and are custom designed to accept
the BAM-1020. The BX-902 and BX-903 are supplied with a separate setup and
installation manual.

NOTE: The air temperature inside any enclosure in which a BAM-1020 is installed must be
held as constant as possible over the course of the hour. This is important because the unit
measures the beta particles through a small gap of air around the filter tape at the beginning
and the end of the hour. If the air temperature inside the enclosure has changed by more

Page 8 BAM-1020-9800 REV F


than about 2 degrees C during this time, the concentration measurement can be affected by
as much as several micrograms. Met One recommends logging air temperature inside the
enclosure to monitor this effect.

2.3 Site Selection


Selection of a proper site to install the BAM-1020 is critical for accurately measuring
particulate levels. In many cases, these items must be correctly addressed in order for the
collected data to be acceptable for regulatory requirements, such as EPA PM10 and PM2.5
equivalency methods. Specifications for the site selection can be found in EPA document
EPA-450/4-87-007 May 1987 “Ambient Monitoring Guidelines for Prevention of Significant
Deterioration”, as well as 40 CFR, Part 58.

Inlet Height:
• The inlet should be located in the “breathing zone”, between 2 and 15 meters above
ground level. If the BAM is to be installed in an enclosure at ground level, then the inlet
must be at least two meters above the ground.
• If the inlet is located on or through a rooftop, the total height should be no more than
15 meters from the ground level. The inlet should be located at least two meters above
roof surface of the building that the unit is installed in. This matches the specified inlet
height of most FRM samplers.
• If the BAM-1020 is to be co-located with other particulate instruments, such as FRM
filter-based samplers or other BAM units, then the air inlet must be the same height as
the inlet of the other samplers.
• The BX-902 and BX-903 environmental shelters are designed to locate the inlet two
meters above whatever surface they are placed on.
• Make sure to account for the height of the PM10 and/or PM2.5 heads when planning the
inlet tube length. Met One can supply a variety of tube lengths up to 8-feet long.
• The maximum allowable total inlet tube length is 16 feet between the BAM-1020 and
the bottom of the inlet head.

Inlet Radius Clearance:


• The BAM-1020 inlet must have a one meter radius free of any objects that may
influence airflow characteristics, including the airflow radius of another instrument. For
example, if a BAM-1020 is to be installed at a station with another BAM-1020 or a
PM2.5 FRM filter sampler, the inlets of each sampler must be no less than 2 meters
apart from each other.
• If installing near a PM10 SSI Hi-Volume sampler, then the distance between the inlet of
the BAM-1020 and the Hi-Vol must be no less than three meters.
• The BAM-1020 inlet must be located at least 2 meters from minor obstructions such as
short walls, fences, and penthouses.
• If located beside a major obstruction (such as a building) then the distance between
the unit and the building must be equal to twice the height of the building.
• The inlet must be at least 20 meters from the drip line of any overhanging trees.
• There must be at least a 270 degree arc of unrestricted airflow around the inlet. The
predominant direction of concentration movement during the highest concentration
season must be included in the 270 degree arc.

BAM-1020-9800 REV F Page 9


Particulate Sources: To avoid possible errors in the concentration measurements, the inlet
must be located as far as possible from any potential artificial sources of particulate, such as
blowers, vents, and air conditioners on a rooftop. Especially if any of these types of devices
blow air across the inlet of the BAM-1020. Even sources of filtered air must not blow across
the inlet.

Spacing from Roadways: The BAM-1020 should usually not be located directly next to a
major highway or arterial roadway, as vehicle exhaust will dominate the concentration
measurement. This effect can be difficult to predict accurately as shifting winds may direct the
plume toward or away from the BAM inlet.

• Roads with a daily traffic volume of less than 3,000 vehicles are generally not
considered major sources of pollutants, and in this case the BAM must be located at
least 5 meters from the nearest traffic lane.
• The BAM must be located at least 25 meters from any elevated roadway greater than
5 meters high.
• The unit should be located as far as possible from unpaved roadways, as these also
cause artificial measurements from fugitive dust.
• The unit should not be installed in unpaved areas unless year-round vegetative ground
cover is present, to avoid the affects of re-entrained fugitive dust.

2.4 Mounting Options in a Walk-In Shelter


When the BAM-1020 is to be located in a walk-in shelter, the unit will have to be installed on
either an equipment rack or a bench-top. Met One recommends using an equipment rack
when possible, as it does a better job of keeping the unit level and in the correct placement. A
rack also tends to be a cleaner installation and is more space-efficient. However, either
method may be used as long as the mounting is level and allows the inlet tube to be perfectly
vertical. Met One supplies brackets and a set of standard rack-mount screws with each unit.
Take the following into account when planning your mounting:

• Rear Access: It is important that your mounting leaves plenty of access to the rear of
the BAM-1020 unit for wiring connections and maintenance.
• Top Access: It is necessary to have a minimum of eight (8) inches clearance between
the top of the inlet receiver on the BAM and the bottom of the ceiling of the shelter to
accommodate the inlet heater which mounts on the inlet tube directly above the BAM.
• Rack Modifications: It is often necessary to modify the top of the equipment rack by
cutting a hole to allow the inlet tube to extend up through to the ceiling. The drawing
below shows the location of the hole. Note: The inlet heater is a cylinder which installs
on the inlet tube two inches above the top of the inlet receiver of the BAM-1020. If the
BAM unit is to be mounted high in the rack, it may be necessary to make the hole in
the top of the rack larger in order to clear the heater diameter. The heater is supplied
with an insulation tube cover which may be modified as needed. Make sure these
parts are going to fit before installing the BAM-1020.
• Mobile Shelters: If the BAM-1020 is being installed into an equipment rack in a
mobile shelter or van, the instrument must be supported on the bottom in addition to
the rack brackets due to additional strain. The foam rings must also be inserted any
time a mobile shelter is moved with the BAM-1020 inside.

Page 10 BAM-1020-9800 REV F


Rack Modifications for BAM-1020 Inlet

Typical Walk-In Shelter Installation

BAM-1020-9800 REV F Page 11


2.5 Installation Instructions in a Walk-In Shelter
Installation of the BAM-1020 should be performed by personnel familiar with environmental
monitoring equipment. There are no special precautions or handling concerns except for the
normal care required for handling scientific equipment.

1. Roof Modifications: Determine the exact location where the BAM inlet tube will pass
through the roof of the enclosure, and drill a 2 ¼” or 2 ½” diameter hole through the
roof at that location. Make sure the hole is directly above where the BAM inlet receiver
is to be located, as the inlet tube must be perfectly vertical. A plumb-bob is useful for
determining where to locate the hole. Note: The inlet receiver on the BAM is slightly
off of the center line of the unit.
2. Waterproof Flange: Apply plenty of all-weather caulking around the top of the hole,
and install the BX-801 roof flange onto the hole. It is usually best if the threaded barrel
of the flange assembly is installed downward, into the hole. Secure the flange in place
with four lag bolts or self-tapping screws (not supplied). Caulk around the screws to
prevent leaks. Apply Teflon tape to the threads of the gray plastic watertight fitting, and
screw it into the roof flange.
3. Inlet Tube: Remove the white cap and rubber seal from the flange assembly. This
makes it easier to install the inlet tube, as the rubber seal is a tight fit around the tube.
Lower the inlet tube through the flange assembly in the roof and into the inlet receiver
on the BAM Make sure the inlet tube is fully seated.
4. Inlet Alignment: It is very important for the inlet tube to be perpendicular to the top of
the BAM. The nozzle may not close properly if there is a bind caused by misalignment.
A simple way to check is by trying to rotate the inlet tube with your hand (before
tightening the roof flange seal or the BAM inlet set screws). If the inlet tube is installed
straight, then the tube should rotate fairly easily when installed into the BAM. If it does
not rotate, check the inlet tube for vertical alignment or move the BAM slightly.
5. Smart Heater: Before tightening the inlet tube in place, the BX-827 or BX-830 smart
inlet heater (used on most BAM-1020 units) must be installed on the tube. Pull the inlet
tube up out of the inlet receiver a few inches, and pass the tube through the hole in the
heater body (the cable comes out the bottom). Then re-insert the inlet tube into the
BAM. Position the bottom of the heater unit two inches above the top of the inlet
receiver on the BAM, and securely tighten the two set screws in the heater to fasten it
to the tube. Included with the smart heater is a 12” tube of insulation. The tube is split
down its length for easy application. Wrap the insulation around the heater body and
peel back the adhesive cover strip to secure in place. The insulation may be cut to fit if
needed.
6. Inlet Tightening: Slide the rubber seal and white cap down over the top of the inlet
tube and into the roof flange. It is easier if you wet the seal first. Tighten the white
plastic cap.
7. Support Struts: The inlet kit comes with two angled aluminum struts to support the
inlet tube above the roof. These struts are usually fastened (90 degrees apart) to the
inlet tube with the supplied hose clamp. The other end of the struts should be fastened
to the roof with lag bolts (not supplied). Note: Some installations require different
configurations for supporting the inlet tube.
8. Temperature Sensor: Most BAM-1020 units are supplied with a BX-592
(temperature) or BX-596 (temperature and pressure) probe, which is attached to the
inlet tube above the roof. The cable must feed into the shelter to be attached to the

Page 12 BAM-1020-9800 REV F


BAM. In some cases it is easiest to simply drill a 3/8” hole through the roof about six
inches away from the inlet tube, then feed the cable through the hole and caulk around
it to prevent leaks. In some applications there may be a better place to feed the cable
into the shelter. Route the cable into the shelter in the best manner. The BX-596
attaches directly to the inlet tube with a supplied U-bolt. If using a BX-592, fasten the
aluminum cross-arm to the inlet tube, and clip the temperature probe to the cross-arm.
9. Inlet Separator Heads: Install the PM2.5 cyclone (580314 if used) and the PM10 head
(BX-802) onto the top of the inlet tube.
10. Inlet Tube Grounding: Tighten the two set screws in the inlet receiver of the BAM to
the inlet tube. This is important for grounding, as static electricity can build up on the
inlet under certain atmospheric conditions and cause errors. This is also important in
areas near electromagnetic fields, high voltage power lines, or RF antennas. If
possible, scrape away a small spot of the clear anodizing near the bottom of the inlet
tube, and use a multimeter to measure the resistance between this spot and the
“CHASSIS” ground connection on the back of the BAM. It should measure just a few
Ohms if a good connection is made with the set screws. If not, we recommend
removing the set screws and running a ¼-20 tap through the holes. Then reinstall the
screws and check the electrical resistance again.

2.6 Electrical Connections


Each BAM-1020 is factory configured to run on either 120 or 230 volt AC power. The air
pump used will usually require the same voltage as the BAM. Your shelter must be wired for
power to run the BAM, the pump, and any other AC powered devices such as computers,
data loggers, other instruments, etc. Have a qualified electrical contractor provide your power
requirements according to all local codes. A good earth-ground point somewhere near the
BAM unit is highly recommended. After the unit is installed and power is provided, install and
connect the electrical accessories as follows:

1. BAM-1020: Plug the BAM-1020 into AC power with the provided power cord. Note:
Met One recommends plugging the BAM-1020 unit into a battery back-up UPS
(uninterruptible power supply) as even a momentary power outage will ruin the entire
hour worth of data collection. A small computer-style UPS is usually sufficient as the
BAM does not draw much power. The air pump does not need to be connected to the
UPS as the BAM can compensate for short pump outages.
2. Chassis Ground: Connect one of the terminals marked “CHASSIS” on the back of the
BAM to a ground point as close as possible to the instrument. Use the supplied
green/yellow ground wire supplied with the unit. A ground rod, cold water pipe, or
junction box safety ground are possible connection points. Be sure to follow electrical
codes. Note: the BAM-1020 also uses a standard safety ground in the power cord.
3. Pump Connection: Decide on a location to place the air pump. The best location is
often on the floor under the rack or bench, but may be up to 25 feet away if desired.
Route the air tubing from the pump to the back of the BAM unit, inserting it firmly into
the compression fittings on both ends. The tubing should be cut to the proper length
and the excess saved for replacements. The pump is also supplied with a 2-wire signal
cable which goes between the BAM-1020 and the pump to turn the pump on and off.
Connect this cable to the terminals on the back of the BAM marked “PUMP
CONTROL” The end of the cable with the square black ferrite filter goes to the BAM,

BAM-1020-9800 REV F Page 13


but the polarity of the wires is not important. Either the red or black wire can go to
either terminal. Connect the other end of the cable to the two terminals on the pump.
4. Temperature/Pressure Sensor: The BX-596 or BX-592 temperature sensor, if used,
should already be installed onto the inlet tube and the cable routed to the BAM-1020.
Connect the cable to the terminals on the back of the BAM as follows:

BX-596 AT/BP
Wire Color Terminal Name
Yellow Channel 6 SIG
Black/Shield Channel 6 COM
Red Channel 6 POWER
Green Channel 6 ID
White Channel 7 SIG

BX-592 AT
Wire Color Terminal Name
Yellow or White Channel 6 SIG
Black/Shield Channel 6 COM
Red Channel 6 POWER
Green Channel 6 ID

Additional Met One BX-500 series sensors may be connected to channels 1 through 5
to log various other parameters. Details on these sensors are given in Section 10.2 of
this manual.
5. Smart Heater: Plug the heater cable into the mating green metal connector on the
back of the BAM-1020.
6. Optional Data Logger Connection: The BAM-1020 has an analog output which may
be recorded by a separate data logger if required. Connect the terminals on the back
of the BAM marked ”VOLT OUT +, -“ to the data logger with 2-conductor shielded
cable (not supplied). Polarity must be observed. Information on configuring this analog
output is provided in Section 8 of this manual.
7. Other Connections: The BAM-1020 has a variety of telemetry I/O relays and error
relays located on the back of the unit. There are also RS-232 data connections. These
items are described in Section 8 and Section 9 of this manual.

Page 14 BAM-1020-9800 REV F


3 INITIAL SETUP OF YOUR BAM-1020
This section describes the processes for setting up and configuring your BAM-1020, as well
as the basic steps required to put the unit into operation. Some of the topics in this section
will direct you to other sections of this manual for more detailed information. It is assumed
that the unit is already installed and sited as described in Section 2. In some cases it is useful
to first set the BAM-1020 unit up on a test bench before deployment or installation in order to
explore the functions of the unit and perform setups. The following steps for starting up your
unit are described in this section:

1. Power on and warm up the unit.


2. Familiarize yourself with the user interface.
3. Load a roll of filter tape.
4. Perform a Self-Test.
5. Set the real-time clock, and review your SETUP parameters.
6. Perform a leak check and a flow check.
7. Return to the top-level menu and wait for automatic start at the top of the hour.
8. View the OPERATE menus during the cycle.

3.1 Power On
The BAM-1020 has a power switch located on the back of the unit directly above the power
cord. Verify that the unit is plugged in to the correct AC voltage, and that any electrical
accessories are correctly wired before turn the unit on. (Section 2.6) When power is switched
on the main menu screen should appear after a few seconds as shown below. The unit will
probably flash an error indicating that there is no filter tape installed.

3/28/2007 BAM 1020 16:08:29

LAST CONCENTRATION: 0.028 mg/m3


CURRENT FLOW: 16.7 LPM
STATUS: FILTER TAPE ERROR!
FIRMWARE: 3236-02 3.2.4

SETUP OPERATE TEST TAPE

The Main Menu Screen

3.2 Warm-up
The BAM-1020 must warm up for at least one hour before an operation cycle is started. This
is because the beta detector contains a vacuum tube which must stabilize every time the unit
is powered up. This also allows the electronics to stabilize for optimal operation. This applies
any time the unit is powered up after being off for more than a moment. Instrument setups
and filter tape installation can be performed during this time. Most agencies choose to discard
the first few hours of concentration data after the BAM is powered up.

BAM-1020-9800 REV F Page 15


3.3 Using the Keypad and Display
When the BAM-1020 is powered up it will display the main (top level) menu on the LCD
display. This menu is the starting point for all functions of the BAM-1020 user interface.

3/28/2007 BAM 1020 16:08:29

LAST CONCENTRATION: 0.028 mg/m3


CURRENT FLOW: 16.7 LPM
STATUS: ON
FIRMWARE: 3236-02 3.2.4

SETUP OPERATE TEST TAPE

F1 F2 F3

F4 F5 F6

The BAM-1020 User Interface

Soft Keys:
Directly beneath the display are four white buttons called “soft-keys” or “hot-keys”. These are
dynamic keys who’s function changes in response to a menu option displayed directly above
each key on the bottom row of the display. Whatever menu option is displayed above one of
these keys is the function which that key will perform in that particular menu. These are used
throughout the entire menu system for a wide variety of functions. Modifications made within
a menu are usually not saved unless a SAVE soft-key is pressed. EXIT is also another
common soft-key function.

Arrow (Cursor) Keys:


The four red arrow keys are used to scroll up, down, left, and right to navigate in the menu
system, and to select items or change fields on the screen. The arrow keys are also often
used to change parameters or increment/decrement values in the menu system.

Contrast Key:
The key with a circular symbol on it is for adjusting the light/dark contrast on the LCD display.
Press and hold the key until the desired contrast is achieved. It is possible to adjust the
contrast to make the entire display completely blank or completely dark, so be careful to set it
to a visible level or it may appear that the unit is not operating.

Function Keys F1 to F6:


The function keys serve as shortcuts to commonly used menu screens, and can be safely
pressed at almost any time without interrupting the sample cycle. The F keys are only
functional from the main menu screen or for entering passwords. The factory default
password is F1, F2, F3, F4.

Page 16 BAM-1020-9800 REV F


F1 Current: This key is a shortcut to the OPERATE > INST screen, used to display the
instantaneous data values that are being measured by the BAM-1020. See section 3.12. The
F1 key can be used without interrupting a sample cycle.
F2 Average: This key is a shortcut to the OPERATE > AVERAGE screen, used to display
the latest average of the data recorded by the BAM-1020. See Section 3.13. The F2 key can
be used without interrupting a sample cycle.
F3 Error Recall: This key allows the user to view any errors logged by the BAM-1020.
The errors are sorted by date. The F3 key can be used without interrupting a sample cycle.
F4 Data Recall: This key allows the user to view all of the data stored in the BAM-1020,
including concentrations, flow, and all six external channels. The data is sorted by date, and
the user can scroll through the data hour-by-hour using the soft-keys. The F4 key can be
used without interrupting a sample cycle.
F5 Transfer Module: This key is used to copy the memory contents to an optional
transfer storage module to retrieve the digital data without a computer. This function is rarely
used. Met One recommends downloading the data with a modern laptop computer.
F6 (Blank): This key is not assigned a data function.

3.4 Filter Tape Loading


A roll of filter tape must be loaded into the BAM-1020 for sampling. One roll of tape should
last more than 60 days under normal operation. It is important to have several spare rolls of
tape available to avoid data interruptions. Met One recommends wearing lint-free cotton
gloves when handling the tape. Some agencies save the used rolls of tape for post-sampling
analysis, although there is no guarantee that the sampled spots have not been contaminated.
Used filter tape should never be “flipped over” or re-used! This will result in measurement
problems. Loading a roll of filter tape is a simple matter using the following steps:

1. Turn the BAM-1020 on and enter the TAPE menu (Note: This is not the same as the
TEST > TAPE menu). If the nozzle is not in the UP position, press the TENSION soft-
key to raise the nozzle.
2. Lift the rubber pinch roller assembly and latch it in the UP position. Unscrew and
remove the clear plastic spool covers.
3. An empty core tube MUST be installed on the left (take-up) reel hub. This provides a
surface for the used tape to spool-up on. Met One supplies a plastic core tube to use
with the first roll of tape. After that, you can use the empty core tube left over from your
last roll to spool-up the new roll. Never fasten the filter tape to the aluminum hub.
4. Load the new roll of filter tape onto the right (supply) reel, and route the tape through
the transport assembly as shown in the drawing. Attach the loose end of the filter tape
to the empty core tube with cellophane tape or equivalent.
5. Rotate the tape roll by hand to remove excess slack, then install the clear plastic spool
covers. The covers will clamp the rolls to the hubs to prevent slipping.
6. Align the filter tape so that it is centered on each roller. Newer units have score marks
on the rollers to aide in visually centering the tape.
7. Lower the pinch roller assembly onto the tape. The BAM will not function if the pinch
rollers are latched up, and it has no way of automatically lowering the assembly!
8. Press the TENSION soft-key in the TAPE menu. The BAM-1020 will set the tape to the
correct tension and alert you if there was an error with the process. Exit the menu.

BAM-1020-9800 REV F Page 17


Filter Tape Loading Diagram

3.5 Self-Test
The BAM-1020 has a built-in self-test function which automatically tests most of the tape
control and flow systems of the unit. The self-test should be run right after each time the filter
tape is changed, and it can also be used if the operator suspects a problem with the unit.
More detailed diagnostic menus are also available in the BAM, and those are described in the
troubleshooting section.

The self-test feature is located in the TAPE menu. Press the SELF TEST soft-key to start the
test. The tests will take a couple of minutes, and the BAM-1020 will display the results of
each tested item with an OK or a FAIL tag. If all of the test items are OK, the status will show
SELF TEST PASSED as shown in the drawing below. If any item fails, the status will show
ERROR OCCURRED.

Page 18 BAM-1020-9800 REV F


02/08/1999 15:29:30
LATCH: OFF TAPE BREAK: OK
CAPSTAN: OK TAPE TENSION: OK
NOZZLE DN: OK SHUTTLE: OK
NOZZLE UP: OK REF EXTEND: OK
FLOW: OK REF WITHDRAW: OK
Status: SELF TEST PASSED
TENSION SELF TEST EXIT

Self-Test Status Screen

LATCH: This will show OFF if the photo interrupter senses that the pinch rollers
are unlatched as in normal operation. It will show ON if the roller assembly is latched in the
up position. The tape cannot move if the rollers are up!
CAPSTAN: The unit will rotate the capstan shaft forward and backwards and will
check if the photo interrupter sees the shaft rotating. The Capstan shaft is what moves the
filter tape back and forth.
NOZZLE DN: The unit will attempt to lower the nozzle, and will check if the nozzle
motor has moved to the down position with a photo interrupter. It is possible for the nozzle to
become stuck in the UP position, even if the nozzle motor has successfully moved to the
DOWN position. For this reason, proper inlet alignment and maintenance is necessary.
NOZZLE UP: The unit will attempt to raise the nozzle, and will check if the nozzle
motor has moved to the up position with a photo interrupter.
FLOW: The unit will attempt to turn the pump on, and will then look for output on
the flow sensor. This test takes about a minute and will fail if the pump is not connected.
TAPE BREAK: The unit will move the supply and take-up motors to create slack in the
filter tape, and look for proper operation of the tensioner photo interrupters.
TAPE TENSION: The unit will tension the filter tape, and then check the condition of the
tensioner photo interrupters.
SHUTTLE: The unit will attempt to move the shuttle beam left and right, and will
check the motion with a photo interrupter.
REF EXTEND: The unit will attempt to extend the reference membrane, and will check
the motion with a photo interrupter.
REF WITHDRAW: The unit will attempt to withdraw the reference membrane, and will check
the motion with a photo interrupter.

3.6 Initial SETUP Settings Considerations


The BAM-1020 comes pre-programmed with a wide array of default values for the settings
which govern the measurement and calibration. Many of these values will not be changed, as
the default values are accurate for the majority of applications. You will want to review the
Setup Menus in Section 6 of this manual and decide if any values need to be changed. At the
very least, review the following parameters:

1. Set the system clock in the SETUP > CLOCK menu. The BAM-1020 clock may drift as
much as a couple of minutes per month. It is important to check the clock at least once
per month to ensure the samples are performed at the correct times.
2. Review the BAM SAMPLE, COUNT TIME, MET SAMPLE, RANGE, and OFFSET
values in the SETUP > SAMPLE menu.

BAM-1020-9800 REV F Page 19


3. Review the FLOW RATE, FLOW TYPE, CONC TYPE, and HEATER CONTROL
settings in the SETUP > CALIBRATE menu.
4. Review the scaling of any external sensors in the SETUP > SENSORS menu.
5. Review the Smart Heater control settings in the SETUP > HEATER menu.

3.7 Initial Leak Check and Flow Check


Perform a leak check and flow check/calibration as described in Section 5. Become
comfortable with these processes, as they will be performed often.

3.8 Starting a Measurement Cycle


When the preceding steps of Section 3 have been completed, exit out to the main top level
menu. The “Status” line should display “ON” (no errors). If so, the unit will start at the top
(beginning) of the next hour, and will continuously operate until it is commanded to stop.
The unit will stop if the operator sets the Operation Mode to OFF or enters the SETUP or
TEST menus. The BAM-1020 will also stop itself if a non-correctable error is encountered,
such as broken filter tape or failed air flow.

3.9 The Flow Statistics Screen


In the main BAM-1020 menu screen a small arrow has been added to the bottom right
corner. When the DOWN ARROW button is pressed the BAM will display the FLOW
STATISTICS screen as shown below. This screen displays the flow, temperature and
pressure statistics for the current measurement cycle. Pressing the ARROW DOWN key
while in this screen will further scroll down to the remaining parameters below the viewable
area of the display. This screen will not interrupt the sample.

03/28/2007 FLOW STATISTICS 16:26:30


SAMPLE START: 2007/03/28 16:08:30
ELAPSED: 00:18:00
FLOW RATE: 16.7 LPM
AVERAGE FLOW: 16.7 LPM
FLOW CV: 0.2%
VOLUME: 0.834m3
EXIT
FLOW FLAG: OFF
AT: 23.0
MAX AT: 23.5
AVERAGE AT: 23.0
MIN AT: 22.5
BP: 760
MAX BP: 765
AVERAGE BP: 760
MIN BP: 755

The FLOW STATISTICS Screen

Page 20 BAM-1020-9800 REV F


3.10 The OPERATE Screen
Press OPERATE soft-key at the main menu to enter operate menu as shown below. This will
not interrupt the sample if already running.

11/15/2006 OPERATE MODE 14:13:07


= ON

= OFF
Operation Mode: ON
Status: ON

NORMAL INST AVERAGE EXIT

The OPERATE Menu

The DOWN arrow can be used to set the Operation Mode from ON to OFF. This will stop the
measurement cycle, but will not power-down the BAM-1020. NOTE: If the operator sets the
Operation Mode to OFF, or the unit stops itself due to an error, it will still automatically
set the mode back to ON at the top of the hour, and try to run a new cycle! The only
ways to prevent the unit from automatically starting a cycle are to power off the unit, leave the
unit in a TEST or SETUP menu, or leave the pinch rollers latched in the UP position.

The OPERATE menu has three soft-key options for viewing the operating status and sensor
measurements while the unit is operating: NORMAL, INST, and AVERAGE.

3.11 The NORMAL Screen


Normal Mode is the primary operation screen which displays most of the important
parameters of the sample progress in one place, as shown below. Many operators leave their
BAM-1020 in the NORMAL screen whenever the unit is operating, instead of the Main menu.

11/15/2006 Normal Mode 11:27:54


Flow(STD): 16.7 LPM
Flow(ACTUAL): 16.7 LPM
LAST C: 0.061 mg/m3 Press: 764 mmHg
LAST m: 0.806 mg/cm2 RH: 37 %
Heater: OFF
Delta-T: 4.2 C
STATUS: SAMPLING EXIT

The NORMAL Menu

The LAST C value indicates the last concentration recorded, updated at the end of the cycle.
The LAST m value indicates the last measured value of the reference membrane. The value
should be very close or equal to the expected value (ABS). The other values are
instantaneous measurements.

BAM-1020-9800 REV F Page 21


3.12 The INSTANTANEOUS Screen
The INST (Instantaneous) screen displays the instantaneous data values that are being
measured by the BAM-1020. This screen is useful for monitoring the current reading of any
optional sensors that may be connected to the BAM-1020. All values except Conc
(concentration) and Qtot (total flow volume) are current. The Conc represents the
concentration of the last period. Qtot represents total flow volume during the last period.

11/15/2006 CAL DATA FLAG: OFF 11:27:54


Eng Units Eng Units
1 Conc 0.010 mg 2 Qtot .834 m3
3 WS 0.000 4 WD 0.000
5 BP 0.000 6 RH 0.000
7 SR 0.000 8 AT 0.000
TOGGLE FLG VOLT/ENG EXIT

The Instantaneous Menu

The TOGGLE FLG soft-key in this menu allows the user to set the CAL DATA FLAG value
ON or OFF, which marks the data with an M flag to indicate a maintenance was performed
during that time, such as a flow check. This feature is rarely used, as most maintenance
requires stopping the sample anyway. The VOLT / ENG soft-key toggles the displayed values
between units and voltages, useful for diagnostic checks on external sensors.

3.13 The AVERAGE Screen


The AVERAGE screen is similar to the INST screen, except that the concentration and flow
are presented as the previous hour’s average values, and the six external datalogger
channels are average values over the average period of the data logger (set by the MET
SAMPLE value in the SETUP > SAMPLE menu - often also 60 minutes).

Page 22 BAM-1020-9800 REV F


4 THE MEASUREMENT CYCLE
This section describes the measurement and timing cycle of the BAM-1020 instrument. A
clear understanding of the measurement is helpful for the effective operation and
maintenance of the unit. For advanced information on the underlying theory and mathematics
of the measurement see Theory of Operation, Section 11.

Nozzle

Beta Source
Beta Source
Detector

Detector

BAM-1020 Sample and Measurement Stations

4.1 The One-Hour Cycle Timeline


The BAM-1020 is almost always configured to operate on 1-hour cycles. The unit has a real-
time clock which controls the cycle timing. You will see from the following timeline that the
unit makes two 8-minute beta measurements, and one 42-minute air sample, for a total of 58
minutes. The other two minutes are used for tape and nozzle movements during the cycle.

This example shows the timeline if the unit is set for a COUNT TIME of 8 minutes (required
for PM2.5). If the unit is set for 4 minutes, then the beta counts at the beginning and end of the
hour will be only 4 minutes long, with a 50-minute air sample in between. Again, the total
adds up to 58 minutes. Note: This cycle will be slightly altered if the unit is operated in the
Early Cycle mode with an external datalogger. See Section 8.

1. Minute 00: The beginning of an hour. The BAM-1020 immediately advances the
filter tape forward one “window” (the next fresh, unused spot on the tape). This takes a
few seconds. The new spot is positioned between the beta source and the detector,
and the BAM begins counting beta particles through this clean spot for exactly eight
minutes. (I0)
2. Minute 08: The BAM-1020 stops counting beta particles through the clean spot (I0),
and moves the tape exactly four windows forward, positioning that same spot directly
under the nozzle. This takes a few seconds. The unit then lowers the nozzle onto the
filter tape and turns the vacuum pump on, pulling particulate-laden air through the filter
tape (the spot in which I0 was just measured) for 42 minutes at 16.7 liters per minute.
3. Minute 50: The BAM-1020 turns the vacuum pump off, raises the nozzle, and moves
the filter tape backwards exactly four windows. This takes a few seconds, and puts the
spot that was just loaded with particulate back between the beta source and the

BAM-1020-9800 REV F Page 23


detector. The BAM begins counting beta particles through this (now dirty) spot of tape
for exactly eight minutes (I3).
4. Minute 58: The BAM-1020 stops counting beta particles through the dirty spot (I3).
The unit uses the I0 and I3 counts to calculate the mass of the deposited particulate on
the spot, and uses the total volume of air sampled to calculate the concentration of the
particulate in milligrams or micrograms per cubic meter of air. The BAM then sits idle
and waits a few moments for the remaining time in the hour to expire.
5. Minute 60: The beginning of the next hour. The BAM-1020 instantly records the just-
calculated concentration value to memory and sets the analog output voltage to
represent the previous hour’s concentration. The unit advances a new fresh spot of
tape to the beta measurement area and the whole cycle starts over…

4.2 Automatic Span Check During The Cycle


While the vacuum pump is on and pulling air through the filter tape, (as described above) the
BAM-1020 doesn’t have anything else to do, so it performs an automatic check of its
calibration (a span check), and checks for instrument drift caused by varying external
parameters such as temperature, barometric pressure, and relative humidity. This check is
performed every hour automatically as follows:

1. Minute 08: (When the BAM-1020 has just finished moving the clean spot to the
nozzle and turned the pump on) there is another clean spot of filter tape upstream four
windows, between the beta source and the detector. This same spot will stay there for
the entire time the pump is on (usually 42 minutes), as the tape cannot move with the
nozzle down. The BAM begins counting the beta particles through this spot for exactly
eight minutes (I1).
2. Minute 16: The BAM-1020 stops counting beta particles through this spot (I1), and
extends the Reference Membrane between the beta source and the detector, directly
above the spot of filter tape that was just measured. The Reference Membrane is an
extremely thin film of clear Mylar held in a metal tongue. The membrane usually has a
mass of about .800 mg. The BAM starts counting beta particles for eight minutes
again, this time through the membrane and the filter tape spot at the same time (I2).
3. Minute 24: The BAM-1020 stops counting beta particles through the membrane (I2),
withdraws the membrane assembly, and calculates the mass of the membrane “m”,
as if it were particulate on the filter tape spot.
4. Minute 42 (typical): (Eight minutes before the pump stops) The BAM-1020
counts the beta particles through the same spot again (without membrane) for another
eight minutes (called I1’ or I1 prime). This checks the ability of the unit to hold a
constant output when measuring blank filter tape, and is not otherwise used.

The mass density “m” (mg/cm2) of the reference membrane calculated during this automatic
process is compared to the known mass of the membrane (“ABS” value). During factory
calibration, the average value of “m” is determined over at least 24 hours and saved as the
“ABS” value. The percentage of the latest value of “m” from “ABS” should be within the limits
of known variability (+/-4%) If not, the unit records an error for that hour’s data. The value of
“m” provides a method of internal diagnostics and for the monitoring of external variables,
such as temperature swings or pressure changes. The “ABS” value is unique to each BAM-
1020, and can be found on the calibration sheet.

Page 24 BAM-1020-9800 REV F


4.3 Sample Period Description
The sample period is the time when the vacuum pump is pulling dust-laden air through the
BAM-1020. As the air enters the inlet, it first passes through the external PM10 head which
has a screen to keep out bugs and debris, and uses inertia to separate out and trap particle
larger than 10 microns in size. The air then immediately passes through the Optional PM2.5
Very Sharp Cut Cyclone (VSCC) which further separates out and traps particles larger than
2.5 microns in size.

The air then goes down the inlet and through the filter tape, where the remaining particles are
deposited. Some particles smaller than about 0.2 µm may pass through the filter tape and be
exhausted. After the sample period is completed and the particulate spot is measured, there
is almost always a clearly visible spot of dirt on the filter tape where the particulate was
deposited. The BAM-1020 will put the spots very close together on the tape. At exactly
midnight, the BAM will skip one spot, leaving a blank spot on the tape. This is a visual aid
which separates daily entries on the tape.

BAM-1020-9800 REV F Page 25


5 FLOW SYSTEM and FLOW CALIBRATIONS
5.1 Flow System Diagram
The BAM-1020 airflow control system is very simple and effective, consisting of a few rugged
components. Proper operation of the flow system is critical if accurate concentration data is to
be obtained from the unit. The key aspects of proper flow system maintenance are Leak
Checking, Flow Checking, and Nozzle Cleaning. These processes are described in this
section. Met One recommends performing a leak check and nozzle cleaning before flow
calibrations, as a leak will affect the flow. Many users perform an as-found leak check before
cleaning the nozzle, in order to validate already-collected data. Flow calibrations require a
reference flow meter and a reference standard for ambient temperature and barometric
pressure. NIST traceable standards are required in many applications. Met One suggests the
®
BGI DeltaCal brand (available from Met One as the BX-307 option). It includes flow,
temperature and pressure standards in one unit.
Barometric
Filter Temp and Pressure
RH sensors Sensor
(most units)

Inlet Mass Flow


Receiver Sensor

Automatic Flow
Beta Controller
Block
Debris
Filter

Outlet

Manual Flow Valve


(older units) BX-592
Temperature
Sensor

BAM-1020 Flow System

Page 26 BAM-1020-9800 REV F


5.2 Flow Type Descriptions
The BAM-1020 is designed to operate with an airflow rate of 16.7 liters per minute (lpm). This
is important, because the particle separators (PM10 inlet, sharp-cut cyclone, and WINS
impactors) require this flow rate in order to properly separate the correct sizes of particles
from the air stream. All of these separators use the inertia of the particles as they flow
through the inlet to sort out the ones above a certain size (cut point) so that they won’t be
measured by the instrument. If the airflow is more than about 10% above or below the ideal
rate of 16.7 lpm, then particles of the wrong size may be allowed through or sorted out.

This section describes the different types of flow control and regulation schemes used in the
BAM-1020. The unit can be set to any of three different flow types: Metered, Standard, or
Actual (Volumetric), depending on the hardware available and the desired reporting
conditions. All BAM-1020 units have a mass airflow sensor and a barometric pressure
sensor. The unit also has either a manual airflow valve on the back of the unit, or an
automatic flow control valve inside the unit. The unit may also have an optional BX-592 or
BX-596 ambient temperature sensor. Each flow type requires a different process for auditing
and calibrating the flow. To verify or set the flow type of the BAM, go to the SETUP >
CALIBRATE menu, and check the FLOW TYPE. NOTE: The concentration reporting
conditions can now be set independently of the flow type. See section 6.3.

METERED Flow Control:


Neither automatic flow control, nor flow correction for ambient conditions.
Metered flow control is used for BAM units that have a manual (hand-operated) air flow
control valve on the back of the unit. These units do not usually have an automatic flow
controller inside, so the unit cannot automatically adjust the flow to compensate for
temperature or barometric pressure changes, or for filter loading. The unit does have a mass
flow sensor inside. The flow reading from this sensor is stored in EPA conditions, meaning
that the volume of air is calculated with the assumption that the ambient temperature is 25
degrees C, and the barometric pressure is 760mmHg (one atmosphere), regardless of the
actual temperature and pressure, even if the unit is equipped with a temperature sensor. Due
to the lack of automatic flow control, metered units must be frequently flow calibrated and
audited, a process which involves a fair amount of math and takes much longer than
ACTUAL flow calibrations. Also, metered units must have the flow rate set at a point slightly
above the target rate of 16.7 LPM in order to compensate for the fact that the flow rate will
drop as the filter becomes loaded with particulate. If a BAM with an automatic flow controller
is set to METERED flow control, then the flow will be controlled to EPA conditions.

STD (EPA Standard) Flow Control:


Automatic flow control, but usually no flow correction for ambient conditions.
STD (Standard) flow type is often selected when required by specific EPA monitoring
regulations, or when no ambient temperature sensor is available. Standard flow control may
be selected on any units which have an automatic flow controller instead of the manual valve
(most BAM-1020 units have the automatic controller). The flow rate is automatically
controlled using EPA (standard) conditions, meaning that the volume of air (and thus the flow
rate) is calculated with the assumption that the ambient temperature is a standard value
(usually 25 degrees C), and the barometric pressure is 760mmHg (one atmosphere),
regardless of the actual temperature and pressure.

BAM-1020-9800 REV F Page 27


NOTE: At low altitudes and moderate temperature, EPA Standard flow will be very close to
the actual volumetric flow rate. However, at high altitudes the difference between Standard
and Actual flow will be quite significant, due to lower barometric pressure. Carefully consider
this effect when deciding on a flow type to implement.

ACTUAL (Volumetric) Flow Control:


Both automatic flow control, and flow correction for ambient conditions.
Actual (also known as volumetric) flow type is the most accurate flow control mode, and is
required for PM2.5 monitoring. The actual flow type is also the easiest and fastest to calibrate
and audit. The unit always uses actual ambient air temperature and barometric pressure to
correct the flow reading, and the flow rate is continuously and automatically adjusted to
correct for changes in ambient conditions and filter loading. The flow values will be stored
and displayed in actual volumetric conditions. To operate a BAM in actual flow mode, the unit
must have a BX-596 or BX-592 ambient temperature sensor attached to channel six.

5.3 Leak Check Procedure


Leak checks must be performed at least monthly and when the filter tape is changed. Almost
all air leaks in the BAM system occur at the nozzle where it contacts the filter tape. The BAM-
1020 has no way of automatically detecting a leak at this interface, because the airflow
sensor is located downstream of the filter tape. There will normally be a very small amount of
leakage at the tape, but an excessive leak lets an unknown amount of air enter the system
through the leak instead of the inlet. This will cause the total air volume calculation (and thus
the concentration) to be incorrect. Allowing a leak to persist may cause an unknown
amount of data to potentially be invalidated! Perform the following steps to check for
leaks:

1. Remove any PM10 and PM2.5 heads from the inlet tube. Install a BX-305 leak test valve
(or equivalent valve for auditing FRM samplers) onto the inlet tube. Turn the valve to
the OFF position to prevent any air from entering the inlet tube.
2. In the TEST > TAPE menu, advance the tape to a fresh, unused spot.
3. In the TEST > PUMP menu, turn on the pump. The flow rate should drop below 1.0
LPM. If the leak value is greater than 1.0 LPM, then the nozzle and vane may need
cleaning, or there may be another leak in the system.
4. Resolve the leak and perform the check again. A properly functioning BAM will usually
have a leak value of about 0.5 LPM with a clean nozzle and vane using this method.
5. Turn the pump off, remove the leak test valve, and re-install the inlet heads.

NOTE: The reason that there is a 1.0 LPM leak flow allowance is due to the test conditions.
With the inlet shut off the vacuum in the system is very high, about 21 inHg. This is about ten
times greater than the BAM-1020 will encounter during normal sampling. If the leak reading
during this test is 1.0 LPM or less, there will not be a significant leak during normal operation.
Note: If an as-found leak value is measured at 1.5 LPM or less, then previously collected data
is generally considered to be valid.

Page 28 BAM-1020-9800 REV F


5.4 Leak Check Nozzle Seal
Met One supplies a rubber nozzle seal (Part 7440) which can be used to test for leaks in
another manner. Remove the filter tape from the nozzle area and insert the rubber seal with
it’s hole centered under the nozzle. Lower the nozzle and perform a leak check. The hole in
the seal allows the leak check to be performed as before, but without any leakage caused by
the filter tape. With this method, the leak value should drop to 0.2 LPM or less. A leak can be
further isolated by turning the seal around to the side without a hole. This allows a leak test to
be performed on the system below the filter tape only.

5.5 Nozzle and Vane Cleaning


The nozzle and vane (under the nozzle) must be cleaned regularly to prevent leaks and
measurement errors. The cleaning must be done at least once per month, and some sites will
require more frequent cleaning (as often as weekly) as determined by the site administrator.
The worst environment for nozzle contamination seems to be hot, humid environments. This
is because damp filter tape fibers more easily stick to the nozzle and vane. The fibers can
quickly build up and dry out, creating air leaks or even punching small holes in the filter tape.
This will cause measurement errors. Use the following steps to clean the parts.

1. Raise the nozzle in the TEST > PUMP menu. Remove the filter tape (if installed) from
the nozzle area. It is not necessary to completely remove the tape from the unit.
2. With the nozzle up, use a small flashlight to inspect the cross-hair vane.
3. Clean the vane with a cotton-tipped applicator and isopropyl alcohol. Hardened
deposits may have to be carefully scraped off with a dental pick.
4. Lower the nozzle in the TEST > PUMP menu. Lift the nozzle with your finger and insert
another cotton swab with alcohol between the nozzle and the vane. Let the nozzle
press down onto the swab with its spring pressure.
5. Use your fingers to rotate the nozzle while keeping the swab in place. A few rotations
should clean the nozzle lip.
6. Repeat the nozzle cleaning until the swabs come out clean.
7. Inspect the nozzle lip and vane for any burrs which may cause leaks or tape damage.

Cotton
Applicator
Nozzle Lip

Vane

Nozzle Cleaning

BAM-1020-9800 REV F Page 29


5.6 Field Calibration of Flow System – Actual (Volumetric) Flow Mode
Actual (volumetric) flow calibration is very fast and easy. This type of calibration can only be
performed on BAM units which have an automatic flow controller and a BX-592 or BX-596
ambient temperature sensor on channel 6. The unit must also have the Flow Type set to
ACTUAL in the SETUP > CALIBRATE menu or the flow calibration screen will not be visible.

MULTIPOINT FLOW CALIBRATION


TARGET BAM STD
AT: 23.8 23.8 C
BP: 760 760 mmHg
<CAL> FLOW 1: 15.0 15.0 15.0 LPM
FLOW 2: 18.3 18.3 18.3 LPM
FLOW 3: 16.7 16.7 16.7 LPM
CAL NEXT DEFAULT EXIT
Actual Flow Calibration Screen

1. Enter the TEST > FLOW menu as shown above. The nozzle will lower automatically
when this screen is entered. The “BAM” column is what the BAM-1020 measures for
each parameter, and the “STD” column is where you will enter the correct values from
your reference standard. The <CAL> symbol will appear next to the parameter
selected for calibration. The ambient temperature (AT) and pressure (BP) must be
calibrated first, as the BAM uses these parameters to calculate the air flow rate in
actual mode.
2. Measure the ambient temperature with your reference standard positioned near the
BX-592 or BX-596 ambient temperature probe. Enter the value from your reference
standard into the STD field using the arrow keys. Press the CAL hot key to correct the
BAM reading. The BAM and STD values should now be the same.
3. Press the NEXT hot key to move the <CAL> indicator to the BP field, and repeat the
same steps for barometric pressure.
4. After the temperature and pressure readings are correct, remove the PM10 and PM2.5
heads from the inlet tube and install your reference flow meter onto the inlet. Press the
NEXT hot key to move the <CAL> indicator to the first flow point of 15.0 LPM. The
pump will turn on automatically. Allow the unit to regulate the flow until the BAM
reading stabilizes at the target flow rate. Enter the flow value from your standard into
the STD field using the arrow keys. Press the CAL hot key to correct the BAM reading.
NOTE: The BAM reading will not change to match the STD until after you have
entered all three calibration points.
5. Press the NEXT hot key to move the <CAL> indicator to the second flow point of 18.3
LPM and repeat the process.
6. Press the NEXT hot key to move the <CAL> indicator to the third flow point of 16.7
LPM and repeat the process.
7. When all of the calibrations are complete, the BAM readings should match the STD
reading at 16.7 LPM, +/- 0.1 LPM. Exit the calibration menu.

The DEFAULT hot key can be pressed to reset the user calibration from the selected
parameter and replace it with a factory setting. If any of the FLOW parameters are selected,
the DEFAULT key will reset the calibrations of all three flow points. This feature can be used
to start over with a calibration if difficulty is encountered.

Page 30 BAM-1020-9800 REV F


5.7 Field Calibration of Flow System – EPA Standard Flow Mode
Flow calibration on units operated in EPA STANDARD flow mode can be done a couple of
different ways. If the unit has a BX-592 or BX-596 ambient temperature probe installed, the
easiest way to calibrate the flow is to temporarily change the FLOW TYPE from STD to
ACTUAL in the SETUP > CALIBRATE menu, then perform an Actual (Volumetric) flow
calibration as described above. If this method is used, be sure to set the unit back to STD
flow when done.

If the unit does not have the temperature sensor then you will not have access to the TEST >
FLOW screen. Use the following steps to check the flow instead:

1. Attach your flow standard onto the BAM inlet while the pump cycle is running or turn
the pump on in the TEST > PUMP screen. Allow the BAM flow to stabilize.
2. If your reference flow meter has a STANDARD flow reading available, that value can
be directly compared to the BAM flow reading. If your flow meter only has a volumetric
flow reading, then convert the volumetric flow rate to standard flow Qs with the
following formula:
Qs = Qa * (Pa / Ta) * (298 / 760)
Ta = Ambient Temperature (Kelvin) (Kelvin = Celsius + 273)
Pa = Ambient Barometric Pressure (mmHg)
Qa = Actual Volumetric Flow from Reference Meter

3. Compare the reference flow (converted to STD conditions) to the BAM flow reading
(also in STD conditions). The two should match within 1% (about 0.17 LPM). If not, a
full flow calibration should be performed. Change the FLOW TYPE to METERED and
perform a flow calibration using the Cv and Q0 values as described in section 5.8. The
sections about the manual flow valve do not apply. Set the flow type back to STD
when finished.

BAM-1020-9800 REV F Page 31


5.8 Field Calibration of Flow System – Metered Flow Mode
Metered flow calibration is only performed on BAM-1020 units which have a hand-operated
manual flow valve on the back (mostly older units). Because these units do not have
automatic flow control, the calibration procedure is much more complicated. The flow must
also be more frequently checked due to changes in ambient conditions, which these type of
units cannot compensate for. Use the following steps for a full Metered flow calibration. The
full calibration only needs to be done once in a while, but the flow should be checked and
adjusted often.

1. Advance the filter tape to a fresh spot.


2. Enter the SETUP > CALIBRATE menu. Set the Cv (coefficient of variability) value to
1.000, and the Q0 (flow zero correction) value to 0.000.
3. Disconnect the pump tubing from the back of the BAM, and turn the pump on in the
TEST > PUMP menu. There will not be any air flowing through the unit. Record the
flow reading from the BAM display. This is the zero flow Zf value.
4. Re-enter the SETUP > CALIBRATE menu and set the Q0 value to equal the negative
of the zero flow value Zf.
5. Go back to the TEST > PUMP menu and turn the pump back on. Verify that the flow
reading on the BAM display now reads 0.0 LPM +/-0.1LPM.
6. Reconnect the pump tubing to the back of the BAM. Remove any PM10 and PM2.5
heads, and connect your reference flow meter to the inlet.
7. Record the ambient temperature Ta (Kelvin) from your reference standard, and record
the barometric pressure Pa (mmHg) from the OPERATE > NORMAL screen on the
BAM.
8. From the TEST > PUMP menu, turn the pump on and allow the flow to stabilize for 5
minutes. Then record the actual flow from your reference flow meter Qa, and record
the standard flow reading from the BAM display Qb.
8. Convert the volumetric flow rate Qa from your reference meter to EPA standard flow Qs
with the following formula:

Qs = Qa * (Pa / Ta) * (298 / 760)


Ta = Ambient Temperature (Kelvin) (Kelvin = Celsius + 273)
Pa = Ambient Barometric Pressure (mmHg)
Qa = Actual Volumetric Flow from Reference Meter

9. Calculate the final value for Cv:


Cv = Qs / Qb
10. Calculate the final value for Q0:
Q0 = -Cv * Zf
11. In the SETUP > CALIBRATE menu, enter the final values for Cv and Q0.
12. Turn the pump on in the TEST > PUMP menu again and verify that the flow reading
from the BAM display matches Qs within 1%. If not, repeat the entire flow calibration.
13. After the flow is calibrated, use the flow adjustment knob on the rear of the BAM-1020
and adjust the flow until the display reads 17.5 LPM. This level is within the
specification of the PM10 particle separator, and will allow for filter loading in high
concentration areas. In lower concentration areas the flow can be set at 16.7 LPM.
14. Exit the TEST menu.

Page 32 BAM-1020-9800 REV F


Quick Flow Check and Adjustment for Metered Flow:
These steps can be used to do a quick flow check on a Metered Flow BAM with a manual
valve while it is in operation.
4. Insert your volumetric flow standard onto the BAM inlet, and allow the BAM flow to
stabilize.
5. Record the ambient temperature, ambient pressure, and the volumetric flow from the
reference.
Ta = Ambient Temperature (Kelvin) (Kelvin = Celsius + 273)
Pa = Barometric Pressure (mmHg)
Qa = Actual Volumetric Flow

6. Convert the volumetric flow from the reference to EPA standard conditions:

Qs = Qa * (Pa / Ta) * (298 / 760)

7. Compare the reference flow (converted to STD conditions) to the BAM flow reading
(also in STD conditions). The two should match within 1%. If not, a full flow calibration
should be performed.
8. If the reference flow (converted to STD conditions) and BAM flow reading match, then
adjust the flow adjustment knob on the back of the unit until the BAM flow reading
equals 17.5 LPM.

Manual Flow Compensation for Seasonal Weather Changes:


These steps can be used to periodically adjust the flow rate on a Metered flow BAM with a
manual valve to compensate for changes in average local atmospheric conditions. Full flow
calibrations should still be performed on a regular basis.
1. Measure the ambient temperature Ta (Kelvin) at approximately 4:00 PM. This usually
represents an average daily temperature.
2. Stop the BAM pump and record the barometric pressure Pa (mmHg) from the
OPERATE > NORMAL screen.
3. Calculate a volume correction term by the following:

V = (Ta / Pa)*62.4

4. Divide V by 24.47 to determine the ratio of EPA flow to Ambient Flow and record as
CALNUM.
5. Turn the pump on in the TEST > PUMP menu and allow the flow to stabilize for 5
minutes. Then divide the displayed flow by CALNUM.
6. Adjust the flow adjustment knob on the back of the BAM-1020 until the BAM flow
reading equals 17.5 / CALNUM. Exit the TEST menu.

Example:
Temp = 300 Kelvin
Pressure = 710 mmHg
V = (300/710) * 62.4 = 26.4
CALNUM = 26.4/24.47 = 1.08
Adjusted Flow = 17.5/1.08 = 16.2 LPM
Adjust flow valve until BAM reads 16.2 LPM.

BAM-1020-9800 REV F Page 33


6 SETUP MENU DESCRIPTIONS
The BAM-1020 uses a comprehensive system of setup menus which contain all of the
settings and parameters needed to perform the measurement and operation of the unit.
Some of these settings are set at factory default values which are correct for most
applications, but may be altered by the operator to suit the specific needs of your monitoring
program. This section describes the SETUP menu in detail, and should be reviewed when
the instrument is put into service to ensure desired operation. Once set, most of the values in
the SETUP menus will not need to be changed by the site operator. The SETUP values will
not be lost if the unit is unplugged or powered down.

WARNING: Some of the settings in the SETUP menus are unit-specific calibration
constants which cannot be changed without affecting the accuracy and proper
operation of the unit.

WARNING: Entering the SETUP menu system will require stopping the sample cycle.
Older versions of firmware will not warn you before stopping the sample.

Press the SETUP soft-key to enter the menu as shown below. The Setup Menu provides a
choice of operations. Use the arrow keys to navigate to the desired field, then press the
SELECT soft-key to enter.

SETUP MODE SELECT


CLOCK SAMPLE CALIBRATE EXTRA1
ERRORS PASSWORD INTERFACE SENSOR
HEATER

SELECT EXIT

The SETUP Menu

A brief description of each sub-menu in shown in the table below. Detailed information is
provided in the following sub-sections.

Menu Settings
CLOCK Date and Time Settings.
SAMPLE Range, Offset, Sample Time, Count Time, Conc Units, Avg Period, Unit ID, and RS-232 settings.
CALIBRATE Factory Calibration Values, (Cv, Q0, ABS, µsw, K, BKGD) Flow rate, Flow type, Heater type.
EXTRA1 Low concentration clamp, e1 – e4, Rarely used.
ERRORS Analog error selections, Flow limits, Pressure drop limit.
PASSWORD Password change screen.
INTERFACE Cycle Mode early/standard, alarm relay polarity.
SENSOR Meteorological sensor scaling and configuration screens, Channels 1 – 6.
HEATER RH and Delta-T set-points for Smart Heater. Only visible if Heater Control is set to AUTO.

Page 34 BAM-1020-9800 REV F


6.1 CLOCK Screen
The SETUP > CLOCK screen allows for the setting of the time and date. Time is a 24-hour
clock only. Use the arrow keys to select and increment/decrement the desired field, then
press the SAVE soft-key. The BAM-1020 clock may drift as much as a minute or two per
month. Met One recommends checking the clock monthly to ensure correct sample timing.

6.2 SAMPLE Screen


The SETUP > SAMPLE screen is used to set the BAM-1020 sampling and averaging periods,
and some other important settings. Review each of these settings carefully. The SAMPLE
screen is shown below. The fields can be edited with the arrow keys, then saved with the
SAVE soft-key.

SETUP SAMPLE
RS232 9600 8N1 BAM SAMPLE 042 MIN

STATION # 01 MET SAMPLE 60 MIN


RANGE 1.000 mg OFFSET -0.015 mg
CONC UNITS mg/m3 COUNT TIME 8 MIN
SAVE EXIT

The SAMPLE Screen

RS-232: This allows you to set the baud rate of the RS-232 serial port. The available
values are 300, 600, 1200, 2400, 4800, 9600, 19200, and 38400 baud. Default is 9600. “8N1”
means 8 data bits, no parity, 1 stop bit. These cannot be edited.

BAM SAMPLE: This value sets the amount of time that the vacuum pump is on per cycle;
the actual sample period (see Section 4.1 for a description of the measurement cycle). The
BAM SAMPLE time must be set in response to the COUNT TIME value, since new versions
of the BAM-1020 allow the option of setting the count time to 4, 6, or 8 minutes. If the unit is
used for PM2.5 monitoring, the BAM SAMPLE must be set to 42 minutes with 8 minute count
time.

Count Time BAM Sample


4 min 50 min
6 min 46 min
8 min 42 min

The BAM SAMPLE value can be set from 0-200 minutes for custom applications. If set for
shorter period, such as 15 minutes, the unit will finish the sampling and then wait until the end
of the hour before beginning a new cycle. This may not leave time for the membrane span
check. Only one cycle per hour is allowed, regardless of duration. Setting the value too long
may cause the measurement to extend over into the next hour. Contact the Service
department before setting this to anything but the values shown in the table.

BAM-1020-9800 REV F Page 35


STATION #: This is a station identification number. This number has a range of 00-99,
and will be included in the data reports. When used in a network, every BAM-1020 should be
given a different station number. Default value is 01.

MET SAMPLE: This value is the averaging period for the data logger. It sets how often
data is written to memory, and can be set to 1,5,15, or 60 minutes. For example, if an
external wind speed sensor is attached to the BAM, the MET SAMPLE period could be set to
1 minute. This would cause the BAM to store an average of the WS reading every minute.
This value applies to all sensors attached to the unit. NOTE: This setting will affect how
long the memory will last before getting full. A 1 minute average period will use up the
memory in a little over 3 days, 60 times faster than a 60 minute average period. When the
memory gets full the unit over-writes the oldest data. Met One recommends leaving the MET
SAMPLE period set at the default value of 60 minutes unless required for a particular
application. The dust concentration value will always be an hourly average anyway.

RANGE: The Range setting sets the full-scale range of the analog voltage output. The
RANGE value is almost never changed from the factory setting of 1.000 mg (corresponding
to an analog output voltage of 0.000 to 1.000 volts = 0 to 1000µg). In special cases, this
value may be set to 0.100, 0.200, 0.250, 0.500, 2.000, 5.000, or 10.000 mg. Be sure to
account for this value if using a separate data logger to record the BAM-1020 analog output.
Note: Changing the range setting will affect future data stored to memory. Always download
any old data before changing settings, then clear the BAM memory. Firmware version 3.2.4
or later will prompt you to clear the memory before letting you change this setting.

OFFSET: The OFFSET value is used to set the lower end of the BAM-1020 measurement
range, and could more accurately be called a “range offset”. The new factory default value for
OFFSET is now -0.015 mg. This causes the entire range of the BAM-1020 to shift down
slightly so that it can read from -0.015 to 0.985 mg, instead of measuring from 0 to 1.000 mg
(if the RANGE is set to 1.000 mg). This allows the unit to measure slightly negative
concentration numbers near zero, which is helpful to differentiate between normal noise and
a failure such as punctured filter tape.

The previous default was -0.005 mg, and the value may be set to -0.005 if needed to work
with data logging systems structured around the old setting. Some BAM users choose to set
the OFFSET value to 0.000 to avoid confusion, at the expense of not being able to see the
true zero noise floor of the unit.

This value also affects the analog output, so that 0 to 1.000 volts equals -0.015 to 0.985 mg,
instead of 0.000 to 1.000 mg. This is because the voltage output cannot go negative. You
must take this scaling into account if an external data logger is recording the BAM-1020
analog output voltage. Contact the Service department if you plan to set the OFFSET to any
value other than -0.000, -0.005, or -0.015 mg.

Note: The OFFSET value is often misunderstood, and should not be confused with the
BKGD (factory calibration zero correction factor) or the “e1” (lower concentration limit clamp)
values. Be sure you understand all three of these settings!
Note: Changing the offset setting will affect future data stored to memory. Always download
any old data before changing settings, then clear the BAM memory. Firmware version 3.2.4
or later will prompt you to clear the memory before letting you change this setting.

Page 36 BAM-1020-9800 REV F


CONC UNITS: This setting determines the concentration units which the BAM-1020
displays and stores in memory. This can be set to ug/m3 (micrograms) or mg/m3 (milligrams)
per cubic meter. This is a new option for the BAM-1020. Past versions have always been set
for mg/m3. Note: 1 mg = 1000 µg.

COUNT TIME: This is the amount of time the unit takes to perform the I0 and I3 counts.
Past versions of the BAM-1020 have always been fixed at 4 minutes. The new options allow
the count time to be set for 4, 6 or 8 minutes. When used to monitor PM2.5, this must be set to
8 minutes. This increases the sensitivity of the unit in lower concentrations. Increasing the
count time will require decreasing the sample time. For example, a count time of 4 minutes
allows a sample time of 50 minutes, while a count time of 8 minutes allows a sample time of
only 42 minutes. The unit will prompt you to change the sample time if you change the count
time to an incompatible value.

6.3 CALIBRATE Screen


The SETUP > CALIBRATE screen is where most of the factory-determined calibration
parameters for the BAM-1020 are stored. These values are unit-specific, and can also be
found on the calibration certificate for the unit in case a setting is accidentally changed. Most
of these settings will never be changed without specific information from Met One
Instruments. It is good practice to periodically audit the calibration values to verify that they
have not been altered. The CALIBRATE screen is shown below.

CALIBRATE SETUP
FLOW RATE: 16.7
CONC TYPE: ACTUAL FLOW TYPE: ACTUAL
Cv: 1.047 Qo: 0.000
ABS: 0.822 µsw: 0.306
K: 1.005 BKGD: -0.0030
STD TEMP: 25C HEATER: AUTO
SAVE EXIT
The SETUP > CALIBRATE Screen

FLOW RATE: This sets the air flow rate for the BAM-1020 sample period. The BAM will
continuously regulate the flow to this value (except manual valve models). The flow rate is
almost always set for 16.7 liters per minute, as this is required for all PM2.5 and PM10
monitoring. The operator may change this value temporarily in order to test the ability of the
pump and flow controller to regulate the flow at different levels. The range of this setting is 10
to 20 LPM.

CONC TYPE: This sets the way that the concentration values are reported. If set to
ACTUAL, then the concentration is calculated based on the volume of the air at ambient
conditions, and a BX-592 or BX-596 temperature sensor is required. If set to STD, the
concentration is calculated based on the standard values for temperature and pressure
(usually 25C and 760mmHg), even if a temperature sensor is available. This value is usually
set to match the FLOW TYPE setting, and must be set to ACTUAL for PM2.5 monitoring.

BAM-1020-9800 REV F Page 37


FLOW TYPE: This setting selects the flow control scheme used by the BAM-1020. The
three possible settings are METERED, STD, and ACTUAL.

• METERED: Usually used for BAM units with a manual flow valve on the back. Flow is
reported in EPA conditions.
• STD: EPA Standard flow. The flow is controlled and reported in EPA standard
conditions. Used where required by regulations.
• ACTUAL: Actual Volumetric flow is controlled and reported to ambient temperature
and pressure conditions. This is required on PM2.5 monitors, and is recommended by
Met One whenever possible. BX-596 or BX-592 sensor is required.

This is an important parameter to understand. At sea-level and moderate temperatures the


difference between these settings will be minimal, but at high elevations or varied
temperatures the flow rate can be greatly affected by this setting. NOTE: Section 5.2 contains
a detailed description of each of these flow types, and should be studied to ensure proper
operation of the unit.

Cv: This value is the factory-set Coefficient of Variability for the internal flow sensor. The
value of Cv is only changed by the user when performing a flow calibration on manual valve
(metered) or STD flow units. ACTUAL flow controlled units almost never need to have this
value altered.

Qo: This value is the factory-set zero correction factor for the internal flow sensor. The
value of Qo is only changed by the user when performing a flow calibration on manual valve
(metered) or STD flow units. ACTUAL flow controlled units almost never need to have this
value altered.

ABS: The ABS value is the factory-set mass of the reference membrane foil used during the
automatic span check. This value is compared to the measured value each hour (see section
4.2). Each unit’s ABS value is different, but is typically around 0.800 mg. The ABS value is
never changed by the operator unless the membrane is replaced due to damage.

µsw: This is called the Mu-switch value, and is the factory-set mass absorption coefficient
used by the BAM-1020 in the concentration calculations. Typical values are about 0.285 to
0.310. Warning: This is a unit-specific calibration value which may significantly affect
the accuracy of the unit. Never change this value without specific instruction from Met
One Instruments.

K: The K-factor is the factory-set slope correction (multiplier) for the BAM-1020
concentration. The K-factor value is determined by dynamic testing of the BAM-1020 in the
factory smoke chamber. This will always be a value between 0.9 to 1.1. All of the stored and
displayed data contains this correction. Warning: This is a unit-specific calibration value
which may significantly affect the accuracy of the unit. Never change this value
without specific instruction from Met One Instruments.

BKGD: The BACKGROUND value is the factory-set zero correction (slope offset) for
the BAM-1020 concentration. This is determined by running the unit for at least 48 hours with
a HEPA zero filter on the inlet. The concentration values over this time are averaged, and the
BKGD is the negative of this average. All of the stored and displayed data contains this
Page 38 BAM-1020-9800 REV F
correction. The BKGD value is typically between 0.000 and -0.005 mg/m3. Met One does offer
a zero filter inlet kit (BX-302) which may be used to audit this value. Warning: This is a unit-
specific calibration value which may significantly affect the accuracy of the unit. Note:
The BKGD value is not to be confused with the OFFSET (range offset) value in the SETUP >
SAMPLE menu. See section 6.2.

STD TEMP: This is the value of standard temperature, used for standard flow or
concentration calculations. In the U.S. the value of standard temperature is usually 25
degrees C as mandated by the USEPA. Some other areas use a standard temperature value
of 0C or 20C.

HEATER: This setting selects which mode the Smart Inlet Heater is used in. When set to
AUTO, the Smart Heater will use the filter RH and temperature sensors to control the inlet
tube heating. When set to MANUAL, the unit will simply turn the heater on all the time
regardless of filter conditions. The actual setup parameters for the Smart Heater are located
in the SETUP > HEATER menu, which will not appear unless this value is set to AUTO. The
operator may safely set this parameter as required. Met One recommends using the AUTO
setting. This value must be set to AUTO for PM2.5 monitoring.

6.4 EXTRA1 Screen


The settings in the EXTRA1 screen are special settings that have been installed for special
applications and generally will never be changed.

e1 Low Concentration Limit. The lowest concentration value the BAM-1020 is allowed to
store or display, despite what it measured. Any measurements below this value will be
clamped. The range is -0.015mg to +0.010mg, and the default value is -0.015mg.
Note: This value is not to be confused with the OFFSET value which sets the lower
limit of the measurement range, or the BKGD value which is the zero correction.

e2 Not Used.

e3 Membrane OFF Delay. Hysteresis timer, range is 0.000 to 5.000 seconds. Don’t
change this value unless instructed to do so by Met One.

e4 Membrane Time Out. The time the unit allows for the membrane assembly to move
before generating an error. Range is 10.00 to 20.00 seconds. Don’t change this value
unless instructed to do so by Met One.

6.5 ERRORS Screen


This screen allows the operator the option of reporting BAM-1020 errors with the analog
output signal. This type of error indication is used when the operator is limited to a single
voltage channel for particulate information, such as when the BAM is connected to certain
types of data loggers. In this case, the BAM sets the analog output to full scale voltage
(usually 1.000 volts) when an error occurs. At the beginning of the next hour, the errors are
reset and the output functions normally unless another error occurs. The operator can select

BAM-1020-9800 REV F Page 39


which errors, if any, are reported in this manner, by selecting each error from the list below
and enabling or disabling it (1=ON, 0=OFF) in the error setup screen.

Regardless if a particular error is enabled for the analog output in this manner or not, it will
always be reported in the BAM-1020 internal memory, and may be viewed with the display or
by downloading the data through the serial port. Some of these errors such as P, R, N, and E
may be set to cause the analog output to go full scale, even though there may be nothing
wrong with that hour’s data. In this case, the concentration data can still be downloaded from
the BAM. Errors which DO indicate that the hour’s data is invalid, cause the digital
concentration value to be set to full scale too, usually .995mg.

This scheme is used because it is rare for an actual valid concentration reading to measure
full-scale. However, concentrations at or near zero can be common, so leaving the data value
at 0.000 during an alarm could be mistaken for valid data.

SETUP MODE ERROR


EUMILRNFPDCT AP FRI FRh
111111111111 150 10 20
1=ON, 0=OFF
SAVE EXIT

The ERRORS Screen

E EXTERNAL RESET: This error indicates that the system clock time was unable
to reset when signaled by an external datalogger. If external reset is successful then
no error is logged (see Section 8.2). Sometimes called INTERFACE RESET.

U TELEMETRY FAULT: This error indicates that an external datalogger has sent an
error to the BAM-1020 (on the TELEM FAULT input) indicating that it has encountered
a problem. Check the datalogger.

M MAINTENANCE: This is a user-set data flag which indicates that calibration or


testing was performed during the flagged hour. The “M” flag may also be forced ON in
the SETUP > INTERFACE menu by setting “Force Maint” to ON, or in the OPERATE >
INST screen by pressing the TOGGLE FLG button.

I INTERNAL CPU: This indicates an error in the mass concentration calculation by


the central processor. Contact the Service department if these errors begin to occur
frequently.

L POWER FAIL: This error occurs any time power is cycled or lost, even
momentarily. Frequent “L” errors usually indicate poor-quality AC power. In some
cases these errors can be generated by electrical interference (such as large radio
antennas or motors) causing an internal reset in the BAM-1020.

R REFERENCE MEMBRANE: This error indicates that the reference membrane


assembly is not physically extending and retracting properly. The error is generated if

Page 40 BAM-1020-9800 REV F


photo sensors S2 and S3 never change state despite drive commands to the
membrane motor, and a timeout of the membrane motion occurred after 15 seconds.

N NOZZLE STUCK TIMEOUT (or Delta-T exceeded): This error indicates that the
nozzle motor is not operating. The error is triggered if photo sensors S4 and S5 never
change state despite drive commands to nozzle motor, and if the sensors do not see
the nozzle motor move within 12 seconds of it being turned on. NOTE: The nozzle
motor lifts the nozzle, but the nozzle is lowered only by its spring. So it is possible for
the nozzle to become stuck in the UP position without generating an error! Proper
maintenance and inlet alignment prevents this.

The “N” error is also used to indicate that the Delta-Temperature set-point was
exceeded. This occurs if the sample air temperature (measured below the filter tape) is
hotter than the ambient air by at least one degree above the set-point value. This is
due to the normal heating of the sample air by the smart heater. In this case, the error
is used to simply flag the data. Frequent errors may indicate that the set-point is set
too low. In most applications Delta-T control is disabled entirely. See the inlet heater
setup instructions in this manual.

F FLOW ERROR: This error occurs if the average air flow over the sample period
was out of the limits set by the FRl (low limit) and FRh (high limit) values. Momentary
changes in airflow do not trigger the error. This error may begin to occur if the vacuum
pump is wearing out, or due to a fault with the flow sensor, flow controller, or air tubing.

The “F” error is also used to indicate if the ambient temperature or barometric
pressure sensor has failed or is incorrectly connected (only if the BAM is set for
ACTUAL flow or concentration reporting). This applies to auto ID sensors BX-592 and
BX-596, the internal filter pressure sensor, and CARB style temperature sensors. The
sensor is considered failed if the 1 minute average reading is at or beyond the min or
max measurement range of the particular sensor.

P PRESSURE DROP EXCESSIVE: This error indicates that the vacuum beneath
the filter tape has exceeded the limit set by the AP value. This is almost always
caused by high concentrations, or certain types of particulate clogging the filter tape.
When this error occurs, the BAM stops the pump to prevent overheating, completes
the measurement early, then waits for the top of the next hour. To increase the amount
of particulate which can build up on the tape before this occurs, set the AP value
higher.

D DEVIANT MEMBRANE DENSITY: This error indicates that the reference


membrane span check measurement (m) for that hour was out of agreement with the
expected value (ABS) by more than ±5%. If these errors start to occur regularly, it
could indicate that the beta detector is beginning to wear out. It can also be caused by
a dirty or damaged membrane. Also sometimes called a BAM CAL error.

C COUNT ERROR: This error indicates that the beta particle counting system is not
operating properly, and is activated if the beta count rate falls below 10,000 per 4
minutes. The beta count rate through clean filter tape is usually more than 600,000 per

BAM-1020-9800 REV F Page 41


4 minutes. This error could occur if the beta detector has failed or if something is
blocking the beta particles, such as a stuck membrane assembly or debris.

T TAPE BREAK: This error indicates that the filter tape is broken or has run out.
The error is triggered if photo sensor S6 is ON continuously, despite drive commands
to motors M3-M5. Tape supply motor (M3) and tape take-up motor (M4) time out after
10 seconds. Capstan motor (M5) times out after 6 seconds. This error is also
generated if the pinch roller assembly has been left latched in the UP position when a
measurement cycle starts. Photo sensor S9 is ON any time the latch is set. The BAM-
1020 has no way of unlatching and lowering the pinch rollers. It must be done
manually.

AP Pressure-drop limit across the filter tape. The default setting is 150 mmHg, and the
range is 0-500 mmHg. See the PRESSURE DROP EXCESSIVE error definition above.

FRl Flow Rate Lower Limit. The default setting is 10 lpm, and the range is 0-30 lpm. See the
FLOW OUT OF LIMITS error definition above.

FRh Flow Rate Higher Limit. The default setting is 20 lpm, and the range is 1-38 lpm. See the
FLOW OUT OF LIMITS error definition above.

6.6 PASSWORD Screen


The SETUP > PASSWORD screen allows the program administrator to change the password
required to enter many of the SETUP menus. The password prevents untrained users from
accidentally changing critical settings on the unit. The password can be any 4-key
combination of the six function keys, F1 to F6. The default password is F1, F2, F3, F4. Met
One does not recommend changing the password unless truly necessary. Contact the Met
One Service department for instructions if the password is lost or forgotten.

6.7 INTERFACE Screen


The SETUP > INTERFACE screen is shown below. These settings are used to configure the
BAM-1020 for operation with an external data logger recording the analog output. Most of
these settings are rarely used, but the Cycle Mode setting must be reviewed if the analog
output is being used.

Interface Setup

Cycle Mode: STANDARD Force Maint: OFF


Fault Polarity: NORM Split DELTAP: 00300
Reset Polarity: NORM

SAVE EXIT
The INTERFACE screen

Page 42 BAM-1020-9800 REV F


Force Maint: This can be used to manually toggle the maintenance flag ON or OFF to
mark the data when the unit is being worked on, such as during a flow check. This also
toggles the Maintenance relay to an external data logger. Rarely used.

Fault Polarity: This sets the polarity of the Telemetry Fault Relay. NORM is normally
open, INV is normally closed. Rarely used.

Split DELTAP: Not used.

Reset Polarity: This tells the BAM-1020 the incoming polarity of an external clock reset
signal, if used. This signal is used to synchronize the BAM clock to an external data logger.
NORM is normally open, INV is normally closed.

Cycle Mode: The Cycle Mode can be set to STANDARD or EARLY. If you are not
using the analog output voltage of the BAM-1020, leave this set to STANDARD. See Section
8.2 for a description of the Cycle Modes.

6.8 SENSOR Screen


The SETUP > SENSOR menu is where configurations and setup parameters are located for
the six analog input channels used to log external meteorological sensors. Each channel
must be configured to accept the sensor before data can be acquired. Description for the
parameters are provided below. There is a separate configurable setup screen for each of the
six external sensor inputs in the SETUP > SENSOR menu. There are also two internal
channels (I1 concentration and I2 flow volume) which can be viewed but not modified.

Met One 500 series meteorological sensors have an Auto ID feature which allows the BAM to
automatically recognize the sensor and enter all of the setup parameters for any channel the
sensor is attached to. Each channel can also be manually configured by the user for other
sensors.

SETUP CHAN PARAMS

CH TYPE UNITS PREC MULT OFFSET


06 AT C 1 0100.0 -050.0
SENSOR FS VOLT: 1.000
INV SLOPE:N VECT/SCALAR:S MODE:AUTO ID

SAVE ID MODE EXIT


The SENSOR Menu

CH: This field selects the channel to be viewed or edited. Use the up/down arrow
keys to select the desired channel.

TYPE: This is the channel name. You can enter any desired name here by using the
arrow keys to scroll through the alphabet.

BAM-1020-9800 REV F Page 43


UNITS: This is the measurement units for the channel. You can enter a value here by
using the arrow keys to scroll through the alphabet.

PREC: This is the precision field, which sets the decimal place for the Multiplier and
Offset parameters.

MULT: This is the slope multiplier. Any input on the channel is multiplied by this
amount. The M factor in Y=MX+B.

OFFSET: This is the slope offset value. Any input on the channel has this amount added
to or subtracted from it, after the multiplier is applied. The B factor in Y=MX+B.

FS VOLT: This is the full-scale voltage output of the sensor. The maximum voltage range
that can be supplied by the sensor. This value is usually going to be either
1.000 or 2.500 volts. 2.500 is the maximum setting for this field.

INV SLOPE: This setting allows the channel to recognize a sensor with an inverse slope.
This is always set to N (no) except for use with thermistor temperature sensors
with resistance-only outputs.

VECT/SCALAR: This value sets the averaging method. S (scalar) is used for all
measurements except wind direction, which uses V (vector).

MODE: This field is toggled by pressing the ID MODE soft-key. The value can be set to
either MANUAL or AUTO ID. In MANUAL mode, the user can enter their own
setup parameters for the channel. AUTO ID mode is used with 500 series
sensors, and must be selected in order for the unit to recognize the sensor
automatically. NOTE: Any manually set parameters are lost when changing to
AUTO ID mode.

6.9 HEATER Screen


The SETUP > HEATER screen is only visible if the HEATER CONTROL value in the SETUP
> CALIBRATE menu is set to AUTO. This menu is used to configure the settings used by the
BAM-1020 to control the Smart Inlet Heater. The BAM uses an RH and temperature sensor
located below the filter tape in the sample air stream to monitor the conditions of the air as it
is being sampled. Tests have shown that as the relative humidity of the ambient air exceeds
about 55%, the particulate on the filter tape can begin to absorb moisture and the measured
mass will increase. The effect gets worse as the RH increases. The Smart Heater minimizes
this effect by actively heating the inlet tube to lower the humidity.

Page 44 BAM-1020-9800 REV F


Heater Setup
RH Control: YES
RH Setpoint: 45%
Datalog RH: YES (Chan 4)
Delta-T Control: NO
Delta-T Setpoint: 10 C
Datalog Delta-T: NO (Chan 5)
SAVE EXIT
The HEATER Setup Screen

RH Control: If YES is selected, the Smart Heater will be automatically turned on full
power whenever the humidity of the sample stream exceeds the RH Setpoint. When the RH
falls back below the set point, the heater turns down to a “low power” heat mode which
simulates the older style wrap-around heaters. If this is set to NO, The Smart Heater will stay
in low power mode and no further RH control will be performed.

RH Setpoint: The RH Setpoint can be a number from 10% to 99%. This is the relative
humidity level that will be maintained at the filter. Met One recommends setting this value at
35%, which is the equilibration value for FRM filters and is required when monitoring PM2.5.

Datalog RH: If YES is selected, the filter RH values will be logged on channel 4 of the
BAM-1020. Select YES if you do not have any external sensors attached to channel 4.

Delta-T Control: The BAM-1020 can compare the filter temperature to ambient and
calculate the difference (Delta-Temperature), if a BX-592 ambient temp sensor is attached to
channel 6. If YES is selected, the Smart Heater will be turned down to low power mode
whenever the Delta-T Setpoint is exceeded.

Volatile organic compounds (VOCs) may be vaporized if the Delta-T exceeds 5 degrees
Celsius. However, this is usually not a problem with the BAM-1020, because the spot of
particulate is sampled for less than an hour before being replaced. Met One does not
recommend using Delta-T control except in rare cases, as it overrides the RH control which
has a far greater effect on the concentration measurement. Note: Delta-T control must be set
to NO to maintain USEPA equivalent status for the data.

Delta-T Setpoint: This can be set from 1 to 99 degrees C. If the filter temperature exceeds
the ambient temperature by more than this amount, the Smart Heater will turn down to low
power mode, regardless of the RH level. An N error is logged in the data any time this
happens. Note: There is often a few degrees of Delta-T measured even if the heater is OFF,
due to mild heating effect of the unit itself. Frequent errors will be logged if the setpoint is too
low. Set this value to at least 8 or 10 degrees C if used.

Datalog Delta-T: If YES is selected, the Delta-T values will be logged on channel 5 of the
BAM-1020. Select YES if you do not have any external sensors attached to channel 5. Note:
The measured Delta-T may still be logged even if Delta-T control is set to NO. This is often a
useful parameter.

BAM-1020-9800 REV F Page 45


7 MAINTENANCE, DIAGNOSTICS and TROUBLESHOOTING
This section provides information for maintaining your BAM-1020, and performing diagnostic
tests if a problem is encountered. If the unit generates errors on the display or in the data
array, first check Section 6.5 to identify the error. Many times there is a simple solution.
Persistent errors often signify a failure or impending failure which will require investigation.

7.1 Met One Suggested Periodic Maintenance


Maintenance Item Suggested Period
Download Error Log Weekly
Download Data Log Weekly
Compare BAM-1020 Data with External Datalogger Weekly
Nozzle and Vane Cleaning (NOTE 1) Monthly
Leak Check (NOTE 1) Monthly
Flow Check (Calibration if necessary) Monthly
Clean BAM-1020 Enclosure Monthly
Clean Capstan Shaft and Pinch Roller Tires (NOTE 1) Monthly
Verify BAM-1020 Settings Monthly
Set Real-Time Clock Monthly
Replace Filter Tape 2 Months
Run SELF-TEST Function 2 Months
Full Flow Calibration 2 Months
Clean PM10 Head 2 Months
Clean PM2.5 Inlet 2 Months
Perform 72 hour BKGD (zero filter) test 6 months
Replace/Clean Pump Muffler 6 months
Test Flow Controller 6 months
Test Pump 6 months
Test Filter RH and Filter Temperature sensors 6 months
Test Smart Heater 6 months
Replace Nozzle O-Ring 12 Months
Clean Internal Debris Filter 12 Months
Clean Membrane Span Foil 12 Months
Beta Detector Count Rate Test 12 Months
Beta Detector Dark Count Test 12 Months
Test Analog Output (if used) 12 Months
Rebuild Vacuum Pump 24 months
Clean Inlet Tube 24 months
Replace Pump Tubing 24 Months
Factory recalibration. Not needed except as determined by Service Dept. As Required
NOTE 1: These items may have to be done more often as required.

Page 46 BAM-1020-9800 REV F


7.2 Audit Sheet and Test Records
The back of this manual contains a sample of a BAM-1020 Audit Sheet. This is a test record
which can be filled out as calibrations, checks, or audits are performed on the unit. The
operator is encouraged to make copies of this sample sheet to use as needed. Met One can
also supply the original document in a Microsoft Word® format to be modified as needed.
Keeping records of calibrations and maintenance is critical for any monitoring program.

7.3 Self-Test Feature


The primary method of identifying a malfunction in the BAM-1020 is the SELF TEST feature
in the TAPE menu. This can identify a large number of the possible mechanical failures in the
unit, and is a good place to start if a problem is suspected. See Section 3.5 for a description
of the self-test process.

7.4 Power-Up Problems


The BAM-1020 must at least be able to power on before any further diagnosis can be
performed. There are only a few possible reasons that the unit will fail to even power on:
• Make sure that the unit is plugged into the correct AC voltage.
• Check or replace the two fuses (3.15A, 250V) inside the power switch housing. The
power cord MUST be removed before the fuse door can be opened, or you will break
it. Pry open the top edge of the power switch housing to access the fuses.
• It is possible for the display contrast to be set so light that it looks like the unit is OFF
when it is really ON. Try holding the contrast key for a few seconds. In rare cases the
display may fail completely. If the unit “beeps” when you press the keys, it is ON.
• If the above checks are all OK, there could be a failed power supply inside the unit.
Contact Met One for further instructions. Don’t open the power supply unless qualified.

7.5 Basic Problem and Cause/Solution Table


The following table contains information on some of the more common problems which may
be encountered, and some steps to identify the problems.

Problem: The BAM won’t start a measurement cycle.


Cause/Solution: • The unit is programmed not to start until the beginning of an hour. Make sure the
clock is set correctly.
• Don’t expect the pump to turn ON until the clean tape count is finished.
• The unit cannot start if the pinch rollers are latched UP! The unit cannot lower them.
• Make sure the filter tape is installed correctly.
• The unit will never start a cycle if the display is not on the main or OPERATE menu.
• The unit will usually display an error and “beep” if it cannot start a cycle

Problem: The analog output voltage and/or digital concentration reading are full-scale.
Cause/Solution: • The unit will force these values full-scale to indicate an error. Download the error
log. Whenever the unit is started or interrupted this will happen until the next hour.

BAM-1020-9800 REV F Page 47


Problem: The concentration is reading negative values.
Cause/Solution: • It is possible for the unit to occasionally read negative numbers if the actual
particulate concentration is very low, such as below 3 micrograms. This is because
the BAM has a “noise band” of several micrograms. This should not happen often.
• If the unit is reading negative numbers hour after hour, it is probably punching holes
in the filter tape. These holes can be very small and hard to see. This is almost
always caused by debris on the nozzle or vane. Clean the parts.
• Make sure the SETUP > CALIBRATE values match the calibration sheet, especially
the BKGD value.
• Met One supplies a BX-302 zero filter kit for auditing the zero reading of the unit.

Problem: The airflow rate is too low and won’t adjust up to 16.7 LPM.
Cause/Solution: • The gray plastic pump mufflers clog up after several months. Replace it or drill a
hole in the end of it for a temporary fix. The brass mufflers can often be cleaned.
• The pump may need to be rebuilt after about 2 years.
• Check the inlet and PM heads for obstructions.

Problem: The airflow is stuck at a particular rate, and will not change.
Cause/Solution: • The flow controller unit on some older units can become stuck. If your flow controller
does not have a small circuit board mounted directly on the motor, it needs to be
upgraded. Contact the Service dept.
• Set the flow rate in the SETUP > SAMPLE screen to 14.0 and 17.5 LPM, and turn
the pump on in the TEST > PUMP screen. The BAM should try to regulate to these
values. If the flow does not change, the flow controller is probably stuck. If the flow
regulates lower, but not higher than 16.7, the pump is probably worn out, or there is
a leak. Be sure to set the flow back to 16.7 LPM when done.

Problem: The nozzle gets stuck in the UP position, or won’t press down onto the tape fully.
Cause/Solution: • This is often caused by a misaligned inlet tube. Make sure it is straight up and
perpendicular to the top of the unit.
• The nozzle o-ring eventually breaks down and needs to be replaced. Contact Met
One for detailed instructions. BX-308 tool kit required.
• The brass nozzle bushings may have grit in them. Remove the nozzle and clean the
parts. Contact Met One for detailed instructions. BX-308 tool kit required.
• Lift the nozzle with your fingers and determine if it feels sticky or gritty.

Problem: The unit has flow leaks, even after cleaning the nozzle and vane.
Cause/Solution: • The nozzle may be sticking as described above.
• The inlet receiver o-rings may need to be replaced.
• The nozzle spring may need to be replaced.
• Check the internal flow system for split tubing.

Problem: The unit over-measures concentration compared to an FRM.


Cause/Solution: • The most common cause is moisture getting on the filter tape or being absorbed by
the particulate. Review the Smart Heater settings for proper operation.
• Log the filter RH on channel 4 if possible. RH should be controlled to below 50%.
• Verify the airflow rate and calibration.

Problem: The unit logs frequent “L” Power Failure errors.


Cause/Solution: • The 5 volt DC power supply must be set to 5.25 volts. Contact the Service dept for
instructions to check or adjust this.
• The CHASSIS terminal needs to be connected to an earth ground.
• Try plugging the BAM into a computer style UPS.
• Even a split second power failure will cause an “L” error.
• Local high power RF fields must be avoided if possible.

Page 48 BAM-1020-9800 REV F


Problem: Frequent “D” membrane density errors.
Cause/Solution: • This usually indicates the membrane foil surface is dirty or damaged. It can be
cleaned with water rinse. Damaged membranes must be replaced.

Problem: The clock settings disappear when the unit is powered down.
Cause/Solution: • The lithium battery on the 3230 circuit board may need to be replaced after about 10
years. It is normal for the clock to drift as much as 1 minute per month.

Problem: The display shows “MISSING TEMP PROBE” message.


Cause/Solution: • The unit requires a BX-596 or BX-592 ambient temperature probe if either the
CONC TYPE or FLOW TYPE are set to ACTUAL. If no probe is attached to the
BAM, this message will appear.

7.6 Nozzle Component Replacement


The nozzle components need periodic inspection, cleaning, and replacement. The nozzle o-
ring will need to be replaced every couple of years to prevent the nozzle from starting to stick.
The nozzle itself may need to be replaced if it becomes worn or damaged, and should be
removed and thoroughly cleaned annually. Complete instructions are available from Met One.
A BX-308 service tool kit is required for nozzle re-assembly.

Inlet Receiver

Nozzle O-ring
Nozzle Spring
Nozzle

7.7 Field BKGD Zero Background Tests


The Background value is a correction offset for the concentration data collected by the BAM-
1020 (see section 6.3 for a description of the BKGD value). This value is factory calibrated for
each unit under laboratory conditions, and is not changed for PM10 monitoring.

BAM-1020 units set up to monitor PM2.5 need to have this value field verified (and adjusted if
necessary) every six months using the BX-302 Zero Filter Calibration Kit. The test corrects
the BKGD value to compensate for minor variations caused by local conditions such as
grounding and shelter characteristics. The test involves running the BAM in its normal
operating environment with a zero filter on the inlet for at least 72 hours. The new BKGD will
then be calculated and entered into the BAM. The test should not be performed during a
period of rapidly changing weather. A complete set of instructions for the test are included
with the BX-302 kit.

BAM-1020-9800 REV F Page 49


7.8 Test Menu System
The following sub-sections provide information for performing diagnostic checks on the BAM-
1020 sub-systems using the TEST menus. Most of these tests will be used for
troubleshooting purposes only and are not necessary on properly functioning units. The
TEST menu system is accessed by the TEST soft-key from the main menu and is shown
below. These are used to perform calibrations and audits of various sensors, as well as some
advanced diagnostics to resolve failures and errors.

TEST MODE

COUNT PUMP TAPE DAC


CALIBRATE INTERFACE FLOW ALIGN
HEATER FILTER-T RH

SELECT EXIT
The TEST Menu

7.9 COUNT test Screen


The TEST > COUNT screen allows the user to check the function of the beta detector and
beta source separate from the rest of the mechanical or flow operations. Each count test will
take 4 minutes, and will show the number of beta particles counted as they accumulate. The
final count value will stay on the display after the counting is finished, and up to six count
tests can be displayed on the screen at once. Count tests are usually performed with a clean
section of filter tape between the source and detector, as in normal operation. The test also
allows the membrane to be extended between the source and detector as well, if desired.

TIME: This shows the time that the count test is started.

COUNT: This is the total number of beta particles counted during the four-minute test.
You will see this count rapidly accruing during the test. Typical four-minute
count values are between 600,000 and 1,000,000 counts through clean filter
tape. The count total will be lower if the membrane is extended, which
simulated adding particulate to the tape. If the count total is less than 500,000
then the beta detector is probably burning out.

M: Indicates if the membrane is extended (Y) or not extended (N).

MEMBRN: Press this soft-key to extend the membrane between the source and detector.

NO MEMBRN: This soft-key withdraws the membrane.

GO: This soft-key starts the four minute count test. The counting will immediately
begin. After four minutes the counting will stop and wait for the operator to
initiate another cycle or EXIT.

Page 50 BAM-1020-9800 REV F


7.10 PUMP Test Menu
The TEST > PUMP screen is very useful to test the pump and nozzle, and to perform leak
checks and nozzle cleaning. See section 5.3 for the leak check procedure. Note: The BAM
nozzle motor drives the nozzle UP, but the nozzle is lowered by its spring tension. It is
possible for the nozzle to become stuck in the UP position even if the motor is working and
no errors are generated. It is also possible for the nozzle to not fully seat against the filter
tape as well. These faults are usually caused by an inlet tube alignment problem, a
disintegrated nozzle o-ring, or grime in the nozzle bearings. A good indication of this problem
is an irregularly shaped dust spot on the tape, often with a “halo” around it.

NOZZLE/PUMP TEST MODE

NOZZLE: 7
Flow (STD): 16.7 LPM PUMP: ON

MOVE NOZZLE PUMP ON PUMP OFF EXIT


The PUMP Test screen

NOZZLE: Nozzle status. UP (5) or DOWN (6).

PUMP: Pump status. ON or OFF.

Flow (STD): The air flow rate, displayed in EPA Standard liters/minute.

MOVE NOZZLE: This soft-key will move the nozzle up or down. The test allows for checking
proper nozzle movement. Total elapsed time is about 5 seconds. If the
pump is ON this operation is disabled.

PUMP ON: This soft-key will turn ON the vacuum pump. The nozzle will automatically
be lowered if necessary.

PUMP OFF: This soft-key will turn the pump OFF.

7.11 TAPE Test Menu


The TEST > TAPE menu allows the user to manually move the filter tape forwards or
backwards in increments of 12.5mm “windows”. This is useful to test the tape transport
mechanism or to move fresh spots of tape for other tests, such as flow or count tests. The
nozzle will be automatically raised if necessary, and the tape will take a couple of seconds to
move each window.

X: This is the last number of windows moved. This number will be negative if
the last move was backwards.

BAM-1020-9800 REV F Page 51


FEED: This is the number of windows you want to move. Use the arrow up/down
keys to select up to 10 windows at a time.

FWD: This soft-key will move the filter tape forward amount of the FEED value.

BKWD: This soft-key will move the filter tape backward amount of the FEED value.

7.12 DAC Test Menu – Analog Output Test


The TEST > DAC screen is used to test the function of the analog concentration output
voltage and the DAC (digital-to-analog-converter) electronics. Use the up/down arrow keys to
set the voltage from 0.000 to 1.000 volts in 0.100 increments. Measure the VOLT OUT +/-
terminals on the back of the BAM-1020 with a high quality voltmeter and verify that the
voltage matches the BAM display within 0.002 volts across the full scale. Then attach the
voltmeter to the input of your datalogger and repeat the test to verify that the correct voltages
get to the input. If the analog output does not match the value on the TEST > DAC screen,
contact the Service department for instructions. Note: 1 millivolt = 1 microgram of
concentration in most applications.

7.13 CALIBRATE Test Menu


The TEST > CALIBRATE screen is used to perform tests of the reference membrane span
check which occurs automatically every sample cycle. This test can be run if the BAM-1020
has been logging D errors. Each BAM-1020 has an individually weighed membrane, and this
mass (m) is measured and displayed during this test. Compare the value from this test with
the ABS value on the calibration sheet for your unit. The values must match within 5%. If not,
the most common cause is a dirty membrane (dust or lubricant on the foil). The membrane
can be carefully cleaned with canned air or clean water rinse. Alcohol is not used because it
leaves a film. CD cleaner works well for badly soiled membranes. Caution: The membrane
foil is a thin sheet of polyester and is extremely fragile! It must be replaced if damaged.
Contact the Service department for replacement instructions.

CALIBRATION MODE
REF MBRN: <
COUNT (Io): 634000
COUNT (I): 556234
CAL MASS M: 0.801 mg/m3
START STOP EXIT

The CALIBRATE Test Screen

REF MBRN: This indicates if the reference membrane is extended (>) or withdrawn
(<) from the beta particle path.

COUNT (I0): The total 4-minute beta count through the filter tape only, no membrane.

COUNT (I): The total 4-minute beta count through both the filter and the membrane.

Page 52 BAM-1020-9800 REV F


CAL MASS M: This is the calculated calibration mass (m) derived from the two count
values, the mass which the unit has just measured for the membrane. This should match the
ABS value within 5%.

START: This soft-key starts the test cycle. Counting will immediately begin. After
4-minutes the I0 count will stop, the membrane will extend, and the I count will begin. At the
completion of the test the counting will stop and the mass of the membrane will be calculated.
The total elapsed time is about 8.1 minutes per test.

7.14 INTERFACE Test Menu


The TEST > INTERFACE screen is used to test the relay inputs and outputs on the back of
the BAM-1020. The two inputs (TELEM FAULT and EXT RESET) are tested by applying the
appropriate signal to the terminals on the BAM, then verifying that the value on the screen
changes in response. The five relay outputs (TAPE BREAK, FLOW ERROR, DATA ERROR,
MAINTENANCE and RANGE) are tested by turning them ON or OFF using the arrow keys,
then verifying that the outputs on the terminals respond accordingly. Note: RANGE not used.

7.15 FLOW Test Menu


The TEST > FLOW screen is where the very important flow calibrations are performed on
most BAM-1020 units. See section 5.6. This screen is also useful to check the ambient
temperature and barometric pressure sensors, and for pump and flow controller tests.

7.16 ALIGN Test Menu – Photo Sensor Tests


The TEST > ALIGN menu system is used primarily to factory-test the photo sensors which
monitor all of the mechanical movement in the BAM-1020. This is useful if the unit has failed
some of the Self-Test parameters. The function of the six ALIGN sub-menus are described in
this section. Note: Remove the filter tape for these tests, or it will break.

TEST MODE
NOZZLE SHUTTLE IDLER CAPSTAN
LATCH REF

SELECT EXIT

The ALIGN Menu

NOZZLE: This screen tests the two nozzle photo sensors. Use the UP and DOWN soft-
keys to move the nozzle, and monitor the status of the S4 and S5 photo sensors on the
screen.

SHUTTLE: This screen tests the photo sensor which monitors the position of the shuttle
beam (the two tape rollers that move together). The status of photo sensor S7 should only

BAM-1020-9800 REV F Page 53


change to ON when the beam is moved all the way to the right side. The shuttle must be
moved by hand for this test.

IDLER: This screen tests the photo sensor which monitors the position of the right-side
tape tensioner (the spring-loaded tape roller on the right). When the tensioner is in its left
position under its spring tension, both photo sensors S6 and S1 should be OFF. If the
tensioner is moved to the middle of its travel, photo sensor S1 should be ON and S6 OFF.
When the tensioner is at the rightmost position, S1 and S6 should both be ON. The tensioner
must be moved by hand. The left tensioner assembly has no sensors to check.

CAPSTAN: This screen tests the two photo sensors which watch the rotation of the
Capstan shaft. This is the shaft under the rubber pinch rollers which drives the filter tape
forwards and backwards. Press the ADVANCE soft-key to rotate the Capstan counter-
clockwise, and the BACKUP soft-key to rotate clockwise. The shaft should rotate one-half of
a rotation each time. Photo sensor S8 should turn ON to stop the shaft at each half-turn, and
will be OFF while the shaft to turning. It is helpful to put an ink mark on the end of the shaft to
view the rotation.

LATCH: This screen shows the status of the pinch roller latch. If the rollers are latched
in the UP position, then S9 should be ON. S9 should turn OFF if the latch is unhooked.

REF: This screen tests the two photo sensors which monitor the position of the
reference membrane assembly. When the EXTEND soft-key is pressed the membrane
should extend and the S2 photo sensor should be ON, and S3 OFF. When the WITHDRAW
soft-key is pressed the membrane should withdraw and the S2 photo sensor should be OFF
and S3 ON. It takes a few seconds for the membrane to move.

7.17 HEATER Test Menu


The TEST > HEATER screen is used to force the Smart Heater ON or OFF for testing
purposes. The heater takes several minutes to heat up or cool down noticeably. The heater
automatically turns back off upon exit from the screen.

7.18 FILTER-T Test Menu – Filter Temperature Sensor


The TEST > FILTER-T screen is used to calibrate the filter temperature sensor located in the
air stream beneath the filter tape. When this screen is entered, the BAM will automatically
raise the nozzle and turn the pump on. This allows ambient room air to equilibrate the filter
temperature sensor without the heating effects of the Smart Heater. Allow the pump to run for
at least 5 minutes to allow the sensor to equilibrate. Enter the ambient room temperature
from your reference standard into the REFERENCE field and press the CALIBRATE hot key.
The BAM reading should change to match within +/- 1 deg C. The RESET hot key can be
used to revert to default calibrations and start over.

Page 54 BAM-1020-9800 REV F


FILTER TEMPERATURE CALIBRATION

BAM: 26.1 C
REFERENCE: 26.1 C

CALIBRATE RESET Exit


The FILTER-T Test Screen

7.19 RH Test Menu – Filter Humidity Sensor


The TEST > RH screen is used to calibrate the filter relative humidity sensor located in the air
stream beneath the filter tape. This screen works just like the FILTER-T screen described
above. Allow the pump to run for at least 5 minutes to allow the sensor to equilibrate. Enter
the ambient room relative humidity from your reference standard into the REFERENCE field
and press the CALIBRATE hot key. The BAM reading should change to match within +/- 4%
RH. The RESET hot key can be used to revert to default calibrations and start over.

BAM-1020-9800 REV F Page 55


8 EXTERNAL DATALOGGER INTERFACE SYSTEM
This section describes the configuration of the BAM-1020 to work with a separate, external
datalogger. The BAM-1020 provides an analog concentration output voltage, as well as an
array of relay inputs and outputs. These allow the BAM-1020 to function as a sensor in a
larger array of data collection instruments. This configurations is not used in many BAM-1020
installations. There are a variety of dataloggers available which are compatible with the BAM-
1020 outputs, so consult the manual for your datalogger for the specific setup requirements.
Met One does supply a document showing the setup programming required for several types
of datalogger manufactured by under the ESC brand name. Contact the Service department.

8.1 Analog Concentration Output Signal


The primary link between the BAM-1020 and an external datalogger is the analog
concentration output signal. The analog output type is selectable between isolated voltage (0-
1 or 0-10 volt DC) and isolated current (4-20 or 0-16 mA). The rear panel dipswitches are
used to select the output as shown in the table below. The full-scale value of the output
voltage corresponds to the full-scale measurement of the BAM-1020, determined by the
RANGE and OFFSET setting. See Section 6.2. In the majority of applications, the analog
output is set for 0-1 volt = 0 to 1.000mg or 0-1 volt = -.005 to .995mg. If the OFFSET is set
to -0.015 then the analog output would be 0-1 volt = -.015 to .985mg. The analog output
should be tested as described in Section 7.12.

SWITCH ON OFF
SW1 0-10 vdc 0-1 vdc
SW2 4-20mA 0-16mA
SW3 Not used Not used
SW4 Not used Not used

In some cases the analog output is the only channel available between the BAM-1020 and
the datalogger, and any errors generated by the BAM must be reported using the same
voltage signal. The BAM-1020 will set the analog output to its full-scale reading when any of
the selected errors occur. The errors which cause this are selectable, and are described in
section 6.5. The external datalogger should be programmed to recognize a full-scale reading
as an error, not a valid concentration. This scheme is used because it is very rare for a
concentration reading to exceed the range of the BAM-1020. The data values collected by
the BAM are unchanged and may be viewed with the display or by downloading.

Page 56 BAM-1020-9800 REV F


8.2 Early Cycle Mode Description
During a standard BAM-1020 measurement cycle, the unit waits for the beginning of the new
hour before it sets the analog output to represent the just-finished hour’s concentration.
However, some types of dataloggers (such as ESC) must have the concentration value
available before the new hour starts, or the data will be stored in the wrong hour. The BAM-
1020 has a special EARLY cycle mode (in the SETUP > INTERFACE menu) which causes
the unit to start and finish the measurement a few minutes early in order to output the
concentration voltage for the last 5-minutes of the hour which was just sampled. The
datalogger must be programmed to read this value during the window. The BAM-1020 clock
and the datalogger clock will usually need to be synchronized because of the critical timing
involved. The following describes the timing of the STANDARD and EARLY modes.

STANDARD Cycle Example

EARLY Cycle Example

Analog Output Levels


C0 represents the concentration output level measured from time T0 to T1, where the T labels
represent the top (beginning) of an hour (such as 12:00:00). As you can see, the
concentration voltage C0 for the standard cycle is present for the whole next hour following
the measurement. In early mode the C0 voltage for the current hour is present for only the last
5 minutes of the hour just-sampled (minute 55 to 60), and all other times the concentration
output voltage is 0.920 volts.

External Reset Windows


An external reset signal may be used to synchronize the BAM-1020 clock to the datalogger.
In standard mode the external reset window is plus or minus 5 minutes around the beginning
of the hour, but in early mode the external reset window is between minute 50 and 60 only.
The clock will not reset if the cycle has not reached the I3 count. The error log will contain the

BAM-1020-9800 REV F Page 57


date and time of the reset attempt. If the I3 count is in progress, or the cycle is past the I3
count, then the measurement cycle is canceled. The error log will contain the date and time
of the reset. A canceled cycle will also force the analog output to the full-scale values (1.000
volts in standard mode, or 0.920 volts in early mode).

Standard Mode Clock Resets:


Minute 0 to 5: An external reset signal will change the BAM clock back to the 00:00 of
the current hour. If a cycle has already started, it will continue. No error
occurs since there is adequate time to complete the cycle.
Minute 5 to 55: An external reset signal has no effect. The error log will contain the date
and time of the reset attempt.
Minute 55 to 60: If an external reset occurs after a completed cycle (idle condition), then
no error occurs. The clock will be set forward to 00:00 of the next hour
and a new measurement cycle will start.

EARLY Mode Clock Resets:


Minute 55 to 60: The external reset signal changes the clock back to minute 55:00 of the
current hour. A new measurement cycle will start at that moment. If a
cycle has already started, it will continue. No error occurs since there is
adequate time to complete the cycle.
Minute 0 to 50: The external reset signal has no effect. The error log will contain the date
and time of the reset attempt.
Minute 50 to 55: If an external reset occurs after a completed cycle (idle condition), then
no error occurs. The clock will be set forward to minute 55:00 of the
current hour and a new measurement cycle will start.

8.3 Telemetry and Error Relays


In addition to the analog output, input and output relay connections are provided on rear
panel of the BAM-1020 to allow the unit to be used with an external data logger in a
synchronous mode of operation. The function of each input and output is described below.
Many of the relay outputs described below are related to BAM-1020 error conditions
described in section 6.5.

Note: A contact-closure input to the BAM-1020 is achieved by shorting the two terminals
on that particular input together, usually with a relay on the external datalogger. The
datalogger should not apply any voltage or current to the terminals. Contact-closure
outputs from the BAM-1020 are provided by the unit shorting the two terminals together with
an internal relay, without applying any voltage or current to them. The external datalogger
must then sense the closure. Normally-Open means that the relay contacts are not shorted
together unless a certain condition occurs, while Normally-Closed means that the relay
contacts are shorted until the condition occurs, then they open.

Page 58 BAM-1020-9800 REV F


Sensor 1
Inputs
2

10
Analog
Output
11 Selection
Switches
12

13

14

BAM-1020 Back Panel Relay Connections

1. TELEM FAULT N.V. Telemetry Fault Non-Voltage. This input can be used to
signal the BAM-1020 that the external telemetry system (datalogger) is not
operational. This is a contact-closure input which must be activated for a minimum of
2-seconds. If activated, the BAM will continue to function and will log a “U” error (see
section 6.5) and activate the DATA ERROR relay output. This input can be set to
normally-open or normally-closed in the SETUP > INTERFACE menu.
2. EXT RESET N.V. External Reset Non-Voltage. This input is can be used to
synchronize the BAM-1020 clock to the external datalogger, and is often used in
EARLY cycle mode (see section 8.2). This is a contact-closure input which must be
activated for a minimum of 2-seconds. The input can be set to normally-open or
normally-closed in the SETUP > INTERFACE menu.

BAM-1020-9800 REV F Page 59


3. EXT RESET V External Reset Voltage. This input is the same as above
except the input is activated by a voltage logic level instead of a contact-closure.
4. TAPE FAULT This is a contact-closure output which will be activated
whenever a “T” or tape error is generated (see section 6.5). Polarity is normally-open.
5. FLOW FAULT This is a contact-closure output which will be activated
whenever an “F” or flow error is generated (see section 6.5). Polarity is normally-open.
6. INVALID DATA This is a contact-closure output which will be activated
whenever a C, P, N, R, L, I, M, or U error is generated by the BAM (see section 6.5).
Polarity is normally-open.
7. MAINT This is a contact-closure output which will be activated
whenever a maintenance “M” flag is generated (see section 6.5). Polarity is normally-
open.
8. RANGE X1 This contact-closure output is no longer supported. The
relay for this channel is now used by the Smart Heater.
9. RANGE X10 This contact-closure output is no longer supported. The
relay for this channel is now used by the Smart Heater.
10. PUMP CONTROL This is the low-voltage output which signals the vacuum
pump to turn on.
11. POWER FAIL This is a contact-closure output which will be activated
(closed) whenever a power failure of the 5 volt DC system or an “L” error occurs (see
section 6.5).
12. VOLT OUT This is the analog concentration output voltage terminal.
Typically 0-1 VDC equals 0 to 1000 micrograms (1.000 mg). See section 8.1.
13. CHASSIS These are the earth-ground terminals. These must be
attached to a solid ground point for best operation of the unit.
14. CL OUT Current Loop Output. This is used when the analog output
is needed in current loop form instead of voltage. Typically only used if there is a long
distance between the BAM and the datalogger. Output is selectable between 4-20mA
or 0-16mA.

Page 60 BAM-1020-9800 REV F


9 RS-232 SERIAL COMMUNICATIONS – DATA RETRIEVAL
This section describes the methods used to retrieve data files through the RS-232 serial
communications system on the BAM-1020. The unit has a two-way serial port which may be
used with a computer, laptop, modem, or data transfer module. There is also a configurable
output-only serial printer port. Access to the data through the serial port is a simple menu
driven interface.

9.1 Serial Port Connections and Settings


The RS-232 serial port on the back of the BAM-1020 handles data transfer and may also be
used for instrument setup and operation status checks. The serial port may also be used with
an optional modem for remote communications through a phone line (See Section 9.6).

Desktop Computer Connections: The BAM-1020 can be connected to almost any


standard PC that has an RS-232 serial port available (COM1 to COM4). Connect the RS-232
port on the back of the BAM-1020 to the COM port connector on the computer with a female-
to-female 9-pin null RS-232 cable. (Belkin F3B207-06 is recommended and available from
Met One). CAUTION: Do not confuse the parallel printer port or video adapter port on your
computer with a serial port. Connecting the BAM-1020 to these may cause damage to your
computer and the BAM. If in doubt, consult the computer manual before connecting.

Laptop Computer Connections: The BAM-1020 can be connected to most laptop


computers. Most old laptops have a regular 9-pin RS-232 serial port, just like a desktop
computer. Modern laptops do not have RS-232 ports, so a converter will have to be obtained.
The easiest and cheapest type is a USB-to-RS232 serial adapter. Met One recommends the
Belkin F5U109, available from Met One or a local electronics store. You will still need the
Female-to-Female 9-pin RS-232 cable. Certain laptops occasionally have difficulty
communicating through this type of adapter. Met One does not recommend converters sold
under the Radio Shack brand name. Another option is an RS-232 serial PCMCIA card, such
as the Quatech SSP-100 which installs in an expansion card slot in the laptop and provides a
serial port for the BAM. This type of adapter is very reliable, but more expensive and takes
longer to install and configure. See www.quatech.com for more information.

Communication Settings: The BAM-1020 communicates at 9600 Baud, 8 data bit, no


parity, one stop bit. 9600 baud is the default setting which may be changed. The BAM-1020
settings must match these in the SETUP > SAMPLE menu. If unable to communicate, try
changing the RS-232 Polarity switch on the back of the BAM-1020. This swaps the polarity of
the TX and RX lines (2&3) and functions as a null modem. NOTE: The BAM-1020 user
interface must be in the main top-level menu or OPERATE menu before serial port
communication can be established. The serial port is disabled in all other menus. Also, the
LCD display and keypad on the BAM-1020 are disabled whenever RS-232 communication is
in progress.

BAM-1020-9800 REV F Page 61


9.2 Met One Communications Software
The BAM-1020 is compatible with several communications software programs available from
Met One Instruments.

MicroMet Plus: A powerful and comprehensive data logging program for meteorological
applications which is configurable to collect and manage all data from the BAM-1020.
Comet: A simple and easy to use communications terminal program which can retrieve
data from Met One data loggers, including the BAM-1020. This program replaces the old TUS
(Terminal Utility Software) program.

9.3 Downloading Data Using HyperTerminal


The BAM-1020 data can be easily downloaded through the serial port using HyperTerminal®
or other simple terminal programs. Nearly all PCs running Microsoft Windows 95® or later
operating systems have the HyperTerminal program included. This section describes how to
set up this program for communication with the BAM-1020. Note: The BAM-1020 display
must be in the Main Menu to establish communications.

1. Connect the RS-232 port on the back of the BAM to your computer or laptop using the
appropriate cable. Connect to the Com1 serial port if available.
2. Open HyperTerminal. (Usually located in the Programs\Accessories\Communications
directory). The program will ask you to type a name for the connection. Type “BAM-
1020” or a name of your choice, then click “OK”.
3. The “Connect To” window will open. Select COM1 (or another port if used) from the
drop-down menu in the “Connect using:” field. Click “OK”. Note: You could also set up
the program to dial the BAM through a modem in this window.
4. The “COM1 Properties” window will open. Set the following values in the drop-down
menus, then click “Apply” and “OK”.
Bits per second: 9600
Data bits: 8
Parity: None
Stop bits: 1
Flow control: None

5. The main HyperTerminal connection window should now be open. Press the ENTER
key a couple of times. The window should respond with an asterisk (*) indicating that
the program has established communication with the BAM-1020.
6. Once communication is established, press the h key. This should cause the BAM-
1020 System Menu to appear on the window as shown below. You can now send any
of the characters in the menu to retrieve the desired files. The menu options are
described in the following sections.
7. HyperTerminal will only display 100 lines of data in the window. To capture larger files
(such as All Data), first select Transfer > Capture Text from the drop-down menu.
Select a location for the file, then click the “Start” button. Retrieve the desired files, and
HyperTerminal will automatically store them to the text file.
8. When you exit HyperTerminal, it will ask if you want to save your connection. Click
“Yes” and a file named BAM-1020.ht will be created in the HyperTerminal folder, which
will have all of the settings saved. Use this for future communications with the BAM.

Page 62 BAM-1020-9800 REV F


Windows 95® and HyperTerminal® are registered trademarks of their respective corporations.

9.4 System Menu File Descriptions


Once a serial connection between the computer and the BAM-1020 has been established as
shown above, you will have access to the main BAM-1020 System Menu. Each number 0-9
represents a different data file you can download from the unit. Each file is described below.
To get the desired file, simply press the appropriate number on your keyboard. After a few
minutes, the BAM will stop waiting for a command and you will have to send another “h” to
reestablish the connection.

File 1: Display Current Day Data


This file will include the current day’s data only. An example of the data format is shown
below. The first column is the time, followed by a series of dashes. Each dash represents a
possible error. If an error occurred, a letter representing the error will appear in this field. For
this example, at 7:00 am an “L” error (power failure) occurred. Then at 8:00 an “M” error
was logged, indicating that the operator was performing maintenance that hour.

The next column is the concentration. Notice that the values went full-scale during the two
hours that errors were logged. This indicates invalid data. The Qtot column is total flow
volume for the hour. With a flow rate of 16.67 LPM, and a sample time of 50 minutes, this
value will be about .834 m3 per hour. (16.7 * 50 / 1000 = .8335). The remaining six columns
are the six datalogger inputs on the BAM. In this example RH was logged on channel 4, and
Ambient Temperature was logged on channel 6. The other four channels had nothing
attached, but will appear in the array anyway. (The data on those channels is only noise.)

BAM-1020-9800 REV F Page 63


Report for 04/22/2005 - Day 112 > BAM 1020 < Station ID: 1

Channel 01 02 03 04 05 06
Sensor Conc Qtot WS no WS RH WS AT
Units mg/m3 m3 KPH V MPS % KPH C
=====================================================
01:00 ------------ 0.010 0.834 019.6 0.012 000.3 00017 132.2 008.7
02:00 ------------ 0.009 0.834 019.9 0.012 000.3 00018 132.1 007.4
03:00 ------------ 0.011 0.834 019.8 0.012 000.3 00018 132.1 006.5
04:00 ------------ 0.011 0.833 020.0 0.012 000.3 00018 132.1 006.1
05:00 ------------ 0.012 0.833 019.8 0.012 000.3 00018 132.1 005.3
06:00 ------------ 0.011 0.834 020.1 0.012 000.3 00018 132.0 005.6
07:00 ----L------ 0.995 0.000 020.3 0.012 000.3 00018 132.0 007.4
08:00 ---M------ 0.995 0.000 019.8 0.012 000.3 00017 132.1 009.4
09:00 ------------ 0.008 0.833 019.9 0.012 000.3 00015 132.2 012.5
10:00 ------------ 0.003 0.834 019.5 0.012 000.3 00014 132.2 016.2
11:00 ------------ 0.007 0.833 019.5 0.012 000.3 00013 132.2 019.7
12:00 ------------ 0.011 0.833 019.5 0.012 000.3 00012 132.0 020.7
13:00 ------------ 0.008 0.833 019.1 0.011 000.3 00010 132.0 021.9
14:00 ------------ 0.010 0.833 019.2 0.011 000.3 00010 131.9 022.3
15:00 ------------ 0.020 0.833 019.1 0.011 000.3 00011 132.0 020.9
16:00 ------------ 0.011 0.834 019.3 0.011 000.3 00012 132.1 018.7
17:00 ------------ 0.010 0.833 019.5 0.012 000.3 00012 132.2 017.9
18:00 ------------ 0.010 0.833 019.4 0.012 000.3 00012 132.1 017.1
19:00 ------------ 0.010 0.834 019.4 0.012 000.3 00014 132.2 015.3
20:00 ------------ 0.007 0.833 019.6 0.012 000.3 00015 132.1 014.4
21:00 ------------ 0.006 0.834 019.5 0.012 000.3 00017 132.1 013.3
22:00 ------------ 0.006 0.834 019.7 0.012 000.3 00021 132.0 011.2
23:00 ------------ 0.005 0.833 019.6 0.012 000.3 00023 132.0 010.0
00:00 ------------ 0.011 0.834 019.9 0.012 000.3 00017 132.2 009.5

Savg 0.009 0.833 019.7 0.012 000.3 00015 132.1 013.2


Vavg 0.000 0.000 000.0 0.000 000.0 00000 000.0 000.0
Data Recovery 100.0 %

File 2: Display All Data


This file will download all of the data stored in the BAM-1020 memory, in the same format as
the above example. Be sure to capture text (section 9.3) if downloading this file using
HyperTerminal, as it can be fairly large.

File 3: Display New data


This file will contain all of the data stored by the BAM-1020 since the last time the data was
downloaded. Useful to avoid duplicate data in your database.

File 4: Display System Configuration


This file will contain a list of most of the BAM-1020 settings and calibration values as shown
below. This is useful for verifying the setup on a remote BAM, and to send to the factory if
service is required. The setting report has been updated and reformatted. Following is an
example of the new settings report.

Page 64 BAM-1020-9800 REV F


BAM 1020 Settings Report
06/07/2006 14:19:45
Station ID, 1
Firmware, 3236-02 3.1.0
K, 01.000
BKGD, 00.000
usw, 00.285
ABS, 00.805
Range, 1.000
Offset, -0.005
Clamp, -0.005
Conc Units, mg/m3
Conc Type, STD
Cv, 01.000
Qo, 00.000
Flow Type, METERED
Flow Setpt, 0016.7
Std Temp, 25
Temp Mult, 1.0000
Pres Mult, 1.0000
Flow Mult, 1.0000
High Flow Alarm, 20
Low Flow Alarm, 10
Heat Mode, MANUAL
Heat OFF, 20
RH Ctrl, NO
RH SetPt, 99
RH Log, NO
DT Ctrl, NO
DT SetPt, 99
DT Log, NO
BAM Sample, 50
MET Sample, 60
Cycle Mode, STANDARD
Fault Polarity, NORM
Reset Polarity, NORM
Maintenance, OFF
EUMILRNFPDCT
000000000000
AP, 000150
Baud Rate, 9600
Printer Report, 2
e3, 00.000
e4, 15.000

Channel, 1, 2, 3, 4, 5, 6,
Sensor ID, 255, 255, 255, 255, 255, 255,
Channel ID, 255, 255, 255, 255, 255, 255,
Name, XXXXX, XXXXX, XXXXX, XXXXX, XXXXX, XXXXX,
Units, XXX, XXX, XXX, XXX, XXX, XXX,
Prec, 0, 0, 0, 0, 0, 0,
FS Volts, 1.000, 1.000, 1.000, 1.000, 1.000, 1.000,
Mult, 1.000, 1.000, 1.000, 1.000, 1.000, 1.000,
Offset, 0.000, 0.000, 0.000, 0.000, 0.000, 0.000,
Vect/Scalar, S, S, S, S, S, S,
Inv Slope, N, N, N, N, N, N,

BAM-1020-9800 REV F Page 65


File 5: Display Date / Time
This file will show the date, time and serial number of the BAM-1020.

File 6: CSV Type Report


This command will give you three Comma-Separated-Value options which you can select by
sending the appropriate number below. Each of the data files are the same as above, except
the values in each column are separated by commas (,). This allows the text file to be opened
directly by Microsoft Excel® or other spreadsheets. This is the recommended data retrieval
method. Be sure to capture text for large files if using HyperTerminal.

2 – Display All Data (All data)


3 – Display New Data (Data since last download)
4 – Display Last Data (Previous hour’s data only)

5 – Display All Flow Stats (All flow stats)


6 – Display New Flow Stats (Stats since last download)
7 – Display All 5 Min Flow (5 minute averages of all flow stats)
8 – Display New 5-Min Flow (5 min averages of flow stats since last download)

The flow statistics fields are described below.

Field Description
Start Start time of BAM sample period.
Elapsed Elapsed BAM sample time.
Flow Average flow rate for the BAM sample period.
CV Flow rate coefficient of variance for the BAM sample period.
Volume Sample volume for the BAM sample period.
Flag Flow regulation out of range warning flag.
AT Average ambient temperature for the BAM sample period.
AT Min Minimum ambient temperature for the BAM sample period.
AT Max Maximum ambient temperature for the BAM sample period.
BP Average ambient pressure for the BAM sample period.
AT Min Minimum ambient pressure for the BAM sample period.
AT Max Maximum ambient pressure for the BAM sample period.

The 5 minute flow statistics averages are described below.

Field Description
Time Event time stamp in seconds since January 1, 1970 00:00:00
Flow Minute average flow rate for the BAM sample period.
AT Minute average ambient temperature for the BAM sample period.
BP Minute average ambient pressure for the BAM sample period.

Page 66 BAM-1020-9800 REV F


File 7: Display Last 100 Errors
This file will contain the date, time, and a description of each of the last 100 errors logged by
the BAM-1020. Useful for troubleshooting and to send to the factory if service is required.

File 8: Display > BAM-1020 < Utility Commands


This file contains a list of the ASCII commands can be sent to the BAM-1020 through the
serial port to configure certain parameters or to perform advanced diagnostics. Most of these
commands will not be used by the typical operator unless instructed by a factory technician.
Some of these commands require a password to access. The password is the same as the F-
key sequence used to enter SETUP screens (default password is 1 2 3 4).

Command Command Function


Printer Port Output Configuration. This sets what is output on the 2nd serial port. Sending this
command will prompt the following sub-menu:
1 – Printer Port (default)
a 2 – Standard Diagnostic Port
3 – Factory Diagnostic Port
4 – Comma Separated Data Output Port
c Clear Data Memory. This command erases all stored data from memory! Password required.
d Set Date. This sets the date on the unit. Password required.
Display Hex EEPROM Setup Values. This displays the special memory locations where the
e setup values are stored. Diagnostic only.
f Factory Calibration Test. This is used for factory calibration only!
Display System Menu. This is the command used to access the data downloading menu
h or ? options. Become familiar with this command.
Display ID Values. This command displays the ID codes of the met sensors for diagnostic
i purposes.
Display Hex Data Memory Values. This command displays the data memory locations for
m diagnostic purposes.
p Modify Modem Pointer. Factory use only.
q Display Station ID. This command displays the preset station ID number.
t Set Time. This command sets the time on the unit. Password required.
XMODEM Data Download. This command allows binary data transfer of the unit memory.
b Download only. Requires software handshaking. For use with special software only, not terminal
programs. Advanced use only.
XMODEM Real-Time Value Download. This command is only used by special software to scan
r instantaneous values of sensors, alarms and settings. Requires software handshaking. Advanced
use only.
XMODEM EEPROM Value Download. This command allows quick scanning of non-volatile
x memory for diagnostic purposed. Advanced use only.
Enable concentration report to PRINTER output. This command configures the printer port to
z output a fixed-width concentration report at the end of the sample period. For external loggers.

File 9: Display Pointers


This file is a display of the current status of the data storage memory. The current pointer
position and number of full memory locations is shown. Rarely used.

BAM-1020-9800 REV F Page 67


9.5 Printer Output Port
The Printer port on the back of the BAM-1020 is an output-only RS-232 serial interface which
may be used with serial printer or as a diagnostic output to a computer. The printer port
output can be configured by using the “a” utility command through the other RS-232 port.
(See section 9.4) The output may be set for data paper printouts, comma-separated data
output, or one of two diagnostic modes. Diagnostic modes are generally not used except
under the direction of a factory technician.

A new configuration has been added for the printer port which enables it to output a fixed-
width concentration report at the end of the sample period, which can be used to interface to
a serial data logger. This output is enabled by using the “z” utility command through the
serial port. The output format is date, time, concentration, and flow volume as shown below.

Format in mg/m3 is: mm/dd/yy hh:mm:ss,+99.999,+9.999


Format in µg/m3 is: mm/dd/yy hh:mm:ss,+999999,+9.999

If the BAM is set to STANDARD cycle mode, the output will occur at the top of the next hour.
For example, if a measurement is made over hour 2, then the format would be:

03/28/07 03:00:00, +00.027,+0.834

If the BAM is set to EARLY cycle mode, the output will occur at minute 55:00 for the current
hour. For example, if a measurement is made over hour 2, then the format would be:

03/28/07 02:55:00, +00.027,+0.834

9.6 Modem Option


Connect the Met One Instruments Modem to the RS-232 serial. Note: the RS-232 Polarity
switch on the back of the BAM-1020 may need to be set to REVERSE polarity for
communication using the modem. The Met One Instrument Modem is recommended for use
with the BAM, as it is designed to reliably communicate when other modems will not. If you
are using one of the Met One Instruments data acquisition programs such as MicroMet Plus
or MicroMet AQ you need only enter the telephone number of the site in the system setup
menu of the program. Multiple telephone numbers can be entered for connection to multiple
remote sites.

If you are communicating with a terminal program such as HyperTerminal® or ProComm


Plus® you will need to define the serial port configuration in the setup of the program. Set the
baud rate to 9600, with 8 data bits, no parity, and 1 stop bit. Use the terminal program’s
internal dialing command sequence to dial up the BAM-1020. Verify the connection to the
BAM-1020 by pressing the <Enter> key until the command prompt asterisk (*) appears. If not,
verify the cabling and communications settings. Once connected, the access to the BAM-
1020 is the same menu driven interface as used for the direct PC connection.

Page 68 BAM-1020-9800 REV F


9.7 Flash Firmware Upgrades
The BAM-1020 now has the capability for flash firmware upgrades. This allows the field
operator to reprogram the flash EEPROM through the serial port using the Flash Update
Utility. Units with a firmware revision of 3.0 or higher can be flash upgraded. If the unit
currently has a revision lower than 3.0, the EEPROM will have to be replaced with a flash
compatible chip. The following tasks must be performed whenever firmware is upgraded or
the EEPROM is replaced:

1. Download and save the data and error logs before proceeding. These will be cleared
during the upgrade process.
2. Record the OFFSET value from the SETUP > SAMPLE screen, and the BKGD value
from the SETUP > CALIBRATE screen.
3. Update the firmware.
4. The baud rate will default to 38400. Reset as appropriate.
5. Recalibrate the filter temperature and filter RH sensors.
6. Set the values of OFFSET, CONC UNITS, and COUNT TIME in the SETUP >
SAMPLE screen.
7. Set the values of CONC TYPE, FLOW TYPE, and BKGD in the SETUP > CALIBRATE
screen.

BAM-1020-9800 REV F Page 69


10 ACCESSORIES and PARTS
10.1 Consumables, Replacement Parts, and Accessories
The following parts are available from Met One for maintenance, replacement, service, and
upgrades. If unsure about a part you need, please contact the Service department. Some of
these parts require technical skills or special considerations before use or installation.

Description Part Number Note Graphic


Consumables
Filter Tape Roll, Glass Fiber, 60 day supply 460130
1

Cotton-Tipped Applicators, nozzle cleaning, 100 pack 995217


Solon #362 1

Tools
BAM-1020 Service Tool Kit: (nozzle shims, spring BX-308
2
scales, reel spacer, filter sensor removal tool)
Leak Test Rubber Nozzle Seal 7440 2
Mass Flow Calibration Kit, 0-20 SLPM BX-303
Flow Inlet Adapter Kit (Leak Test Valve) BX-305
2

Zero Filter Calibration Kit, with valve BX-302

Volumetric Flow Calibration Kit (BGI Delta Cal) BX-307


Flow, Temp, and Pressure Reference Standard
2

Pumps and Pump Parts


Pump, Medo, 115 VAC, 50/60 Hz, Low Noise BX-126 2
Pump, Medo, 230 VAC, 50/60 Hz, Low Noise BX-127

Pump, Gast, Rotary Vane, 100 VAC, 60 Hz BX-123


Pump, Gast, Rotary Vane, 115 VAC, 50/60 Hz BX-121 2
Pump, Gast, Rotary Vane, 230 VAC, 50 Hz BX-122
Pump, Gast, Rotary Vane, 230 VAC, 60 Hz BX-124
Muffler, Medo/Gast Pump 580293 1
Pump Rebuild Kit, Gast 680828 1
Pump Rebuild Kit, Medo 680839 1
Pump Service Kit, Filter Replacement, Medo 8588
Pump Controller (Relay Module) Medo/Gast BX-839

Page 70 BAM-1020-9800 REV F


Flow Components
Flow Sensor, Mass, 0-20 LPM, Internal 970608

Flow Controller Assembly 8776 1


Automatic Flow Controller Upgrade Kit BX-961

Filter Assembly, Pisco In-line 580291


Filter Element Only, Pisco In-line 580292 1
Filter Temperature and RH Sensor Kit BX-962 2
Filter RH Sensor 9278 1
Filter Temperature Sensor 9279
Nozzle, Replacement 8009 1
Nozzle Spring, Replacement 2998 1
O-Ring, Nozzle 720066 1
O-Ring, Inlet Receiver, 2 required 720069 1
O-Ring Kit, BAM 9122 1
Pump Tubing, Clear, 10mm O.D. 960025 1

Electrical and Electronic Parts


Display, LCD, Front Panel 2823 3
Circuit Board, CPU 3230-8 3
Circuit Board, Interface 3250-1 3
Fuse, BAM-1020, 3.15A, 250V, 5x20mm, 2 Req’d 590811 1
Motor, with gear box, 4 RPM 8105 3
Motor, with gear box, 10 RPM 8106 3

Power Supply, 115 VAC, 60 Hz BX-115


Power Supply, 115 VAC, 50 Hz BX-116
Power Supply, 230 VAC, 60 Hz BX-230
Power Supply, 230 VAC, 50 Hz BX-231
Power Supply, 100 VAC, 60 Hz BX-100
Power Supply, 100 VAC, 50 Hz BX-101

Inlet Components
PM10 Inlet Head BX-802
2

TSP Sampling Inlet Cap BX-803

PM2.5 Sharp Cut Cyclone BX-807 2


PM2.5 Very Sharp Cut Cyclone, Required for PM2.5 580314
2

PM2.5 WINS Impactor BX-804

BAM-1020-9800 REV F Page 71


Inlet Roof Mounting Kit, With Inlet Tube and Braces, BX-801
Specify Tube Length

Inlet Tube Extension, 8 foot, with coupler BX-823


Inlet Tube, 8 foot 8112
Inlet Tube, 4 foot 8112-4
Inlet Tube, 2 foot 8112-2
Inlet Tube Coupler, F/F 8777
Smart Heater Option, 115 VAC BX-827 2
Smart Heater Option, 230 VAC BX-830 2
Smart Heater Upgrade Kit, 115VAC 9307
Smart Heater Upgrade Kit, 220VAC 9308
Inlet Tube Cleaning Kit BX-341
PM10 Head Cleaning Kit BX-344
O-Rings, Cyclone, set of 6 720097 1
O-Rings, PM10 Head, set of 3 8965 1

Meteorological Sensors
590 Wind Direction Sensor BX-590
591 Wind Speed Sensor BX-591
592 Ambient Temperature Sensor BX-592

593 Ambient Relative Humidity Sensor BX-593


594 Ambient Barometric Pressure Sensor BX-594
595 Solar Radiation Sensor BX-595
596 AT/BP Combo Sensor, Required for PM2.5 BX-596
2

Real-Time Module (RTM), BAM Inlet Particle Sensor BX-894

Communications Components and Misc. Accessories


Serial Printer Kit BX-601
Converter for Parallel Printers BX-602
Modem Kit AX-996
Serial Cable, 6’, DB-9 female/female, BAM to PC 400658 1
Belkin F5U109 USB-to-RS-232 Adapter 550067
Enclosure, Exterior, with heater only BX-902
Enclosure, Exterior, with heater and air conditioner BX-903
Membrane Assembly, replacement BX-301 3

Note 1: Met One recommended spare part.


Note 2: Met One recommended accessory.
Note 3: Part requires special considerations or technical skills.

Page 72 BAM-1020-9800 REV F


10.2 Series 500 Sensor Configurations
The BAM-1020 has six channels of inputs available on the back of the unit for data logging
external sensors. The 500 Series sensors are a set of meteorological sensors designed for
direct compatibility with these channels. The sensors each have an auto-identification (ID)
signal wire with a voltage unique to that particular type of sensor. When one of these sensors
is attached to the BAM, the unit senses this ID voltage and automatically configures the
channel to read the sensor with all the correct scaling parameters.

Many BAM-1020 units are equipped with at least the ambient temperature sensor BX-592
because ambient temperature is required on channel six for actual flow control. If the BAM is
used for PM2.5 monitoring, then the BX-596 sensor is required instead. This is a combination
ambient temperature and barometric pressure sensor which attaches to channels six and
seven, and provides the EPA required AT/BP measurements for actual flow control and flow
statistics.

The scaling and setups values of the series 500 Sensors are provided in the chart below. The
unit should automatically set these values in the SETUP > SENSORS menu. The ID MODE
must be set to AUTO to identify the sensors, or set to MANUAL to change the parameters.

500 Series Sensor Setup Parameters


Model Type Units Range Mult Offset FS VOLT S/V Inv Slope
BX-590 WD Deg 0 to 360 360 0 1.0 V N
mph 0 to 100 100 0 1.0 S N
BX-591 WS
m/s 0 to 44.704 44.70 0 1.0 S N
0
F -22 to +122 144 -22 1.0 S N
BX-592 AT 0
C -30 to +50 80 -30 1.0 S N
BX-593 RH % 0 to 100 100 0 1.0 S N
inHg 26 to 32 6 26 1.0 S N
BX-594 BP mmHg 660.4 to 812.8 152.40 660.40 1.0 S N
mbar 880.5 to 1083.7 203.19 880.46 1.0 S N
Ly/ min 0 to 2 2 0 1.0 S N
BX-595 SR
W/M2 0 to 2000 2000 0 1.0 S N
0
C -40 to +55 95 -40 2.5 S N
BX-596 AT/BP
mmHg 525 to 825 - - 2.5 S N

SIG
CHAN X

COM

POWER

ID

BAM-1020 Back Panel Sensor Input Terminal

BAM-1020-9800 REV F Page 73


500 Series Sensor Wiring Connections for BAM-1020

BX-590 Wind Direction Sensor BX-591 Wind Speed Sensor


Terminal Block Cable Wire Color Terminal Block Cable Wire Color
SIG Yellow SIG Yellow
COM Black/Shield COM Black/Shield
POWER Red POWER Red
ID Green ID Green

BX-592 Ambient Temp Sensor BX-593 Relative Humidity Sensor


Terminal Block Cable Wire Color Terminal Block Cable Wire Color
SIG Yellow SIG Yellow
COM Black/Shield COM Green/Shield
POWER Red POWER White
ID Green ID Red

BX-594 Barometric Pressure Sensor BX-595 Solar Radiation Sensor


Terminal Block Cable Wire Color Terminal Block Cable Wire Color
SIG White SIG Yellow
COM Black/Shield COM Black/Shield
POWER Red POWER Red
ID Yellow ID Green

BX-596 Temperature/Pressure Combo Sensor


Terminal Block Cable Wire Color
Channel 6 SIG Yellow
Channel 6 COM Black/Shield
Channel 6 POWER Red
Channel 6 ID Green
Channel 7 SIG White

Notes:
• BX-592 is almost always connected to channel 6 when used with a BAM-1020.
• BX-592 or BX-596 is required for actual flow control.
• BX-596 is required for PM2.5 monitoring, effective March 2007.

Page 74 BAM-1020-9800 REV F


11 THEORY OF OPERATION and MATHEMATICAL ANALYSIS
When the high-energy electrons emanating from the radioactive decay of 14C (carbon-14)
interact with nearby matter they loose their energy and, in some cases, are absorbed by the
matter. These high-energy electrons emitted through radioactive decay are known as beta
rays and the process is known as beta-ray attenuation. When matter is placed between the
radioactive 14C source and a device designed to detect beta rays, the beta rays are absorbed
and/or their energy diminished. This results in a reduction in the number of beta particles
detected. The magnitude of the reduction in detected beta particles is a function of the mass
of the absorbing matter between the 14C beta source and the detector.

The number of beta particles passing through absorbing matter, such as dust deposited on a
filter tape, decrease nearly exponentially with the mass through which they much pass.
Equation 1 shows this relationship.
Equation 1

I = I 0e−µx

In Equation 1, I is the measured beta ray intensity (counts per unit time), of the attenuated
beta ray (dust laden filter tape), I0 is the measured beta ray intensity of the un-attenuated
beta ray (clean filter tape), µ is the absorption cross section of the material absorbing the beta
rays (cm2/g), and x is the mass density of the absorbing matter (g/cm2).

Equation 1 very closely resembles the Lambert-Beers Law, which is used in spectrometric
analysis. Just as the Lambert-Beers Law is an idealization of what is actually observed,
Equation 1 is also an idealized simplification of the true processes occurring meant to simplify
the corresponding mathematics. However, experimental measurement shows that in properly
designed monitors, such as the BAM-1020, the use of this equation introduces no substantial
error.

Equation 1 may be rearranged to solve for x, the mass density of the absorbing matter. This
is shown in Equation 2.
Equation 2
1 ⎡ I ⎤ 1 ⎡ I0 ⎤
− ln ⎢ ⎥ = ln =x
µ ⎣ I 0 ⎦ µ ⎢⎣ I ⎥⎦

In practice, the absorption cross section is experimentally determined during the calibration
process. Once I and I0 are experimentally measured, it is a simple matter to calculate x, the
predicted mass density.

In practice, ambient air is sampled at a constant flow rate (Q) for a specified time ∆t. This
sampled air is passed through a filter of surface area A. Once x, the mass density of collected
particles, has been determined, it is possible to calculate the ambient concentration of
particulate matter (µg/m3) with Equation 3.

BAM-1020-9800 REV F Page 75


Equation 3
⎛ µg ⎞ 10 6 A(cm 2 )
c⎜ 3 ⎟ =
⎝m ⎠ ⎛ liter ⎞ ⎛ cm 2 ⎞
Q⎜ ⎟∆t(min)µ⎜⎜ ⎟⎟
⎝ min ⎠ ⎝ g ⎠

In Equation 3, c is the ambient particulate concentration (µg/m3), A is the cross sectional area
on the tape over which dust is being deposited (cm2), Q is the rate at which particulate matter
is being collected on the filter tape (liters/minute), and ∆t is the sampling time (minutes).
Combining these equations yields to the final expression for the ambient particulate
concentration in terms of measured quantities. This is shown in Equation 4.
Equation 4
⎛ µg ⎞ 10 6 A(cm 2 ) ⎛I ⎞
c⎜ 3 ⎟ = ln⎜ 0 ⎟
⎝m ⎠ ⎛ liter ⎞ ⎛ cm 2 ⎞ ⎝ I ⎠
Q⎜ ⎟ ∆t(min)µ ⎜⎜ ⎟⎟
⎝ min ⎠ ⎝ g ⎠

The key to the success of the beta attenuation monitor is due in part to the fact that µ, the
absorption cross-section, is almost insensitive to the nature of the matter being measured.
This makes the BAM-1020 very insensitive to the chemical composition of the material being
collected.

It is instructive to perform a conventional propagation of errors analysis on Equation 4. Doing


so, one can develop an equation for the relative measurement error (σc/c) as a function of the
uncertainty in each of the parameters comprising Equation 4. This leads to Equation 5.
Equation 5

σc
2 2
σ 2A σ Q σ 2t σ µ σ 2I σ 2I0
= + + + + −
c A2 Q2 t 2 µ 2 2 2
I ln ⎡ I ⎤ I 0 ln ⎡ I ⎤
2 2
⎢⎣ I 0 ⎥⎦ ⎢⎣ I 0 ⎥⎦

Inspection of Equation 5 reveals several things. The relative uncertainty of the measurement
(σc/c) is decreased (improved) by increasing the cross sectional area of the filter tape (A), the
flow rate (Q), the sampling time (t), the absorption cross-section (µ), I and I0.

In practice, the uncertainty associated with the filter area (σA/A), may be minimized by
ensuring that the tape is in exactly the same position during the I0 measurement as in the I
measurement phase. Careful design of the shuttle and tape control mechanisms inside of the
BAM-1020 results in minimal error here.

The uncertainty in the flow rate (σQ/Q) may be minimized by properly controlling the flow of
the instrument. For BAM-1020 units with a manual flow valve, this value is on the order of ±
3%. For BAM-1020 units equipped with the mass flow controller device, (σQ/Q) decreases to
± 1%.

Page 76 BAM-1020-9800 REV F


The relative error due to the uncertainly in the absorption cross section (σµ/µ), is due to its
slight variation as a function of the chemical composition of the matter being monitored.
Generally, this relative error is on the order of ± 2-3%, with judicious selection of the
calibrated value of µ.

The uncertainty associated with the measurement of I and I0 has to do with the physical
nature of the process leading to the emission of beta particles from the decay of 14C. This
process follows Poisson statistics. Poisson statistics show the uncertainty in the
measurement of I (σI/I) and I0 (σI0/I0) are minimized by increasing the sampling time.
Mathematical analysis shows that doubling the sampling time and hence the measured
intensity of I or I0 will reduce the uncertainty of the measurement by a factor of 1.41 (square
root of 2).

11.1 Converting Data Between EPA Standard and Actual Conditions


As described in this manual, the BAM-1020 can obtain concentration data using either actual
or standard values for ambient temperature and pressure. In some cases, it is necessary to
convert past concentration data collected in standard conditions to actual conditions, or the
other way around. Note: temperature is in degrees Kelvin (C+273) and pressure is in mmHg.

Equation 6
Cstd = Camb * (Pamb / Pstd) * (Tstd / Tamb)

Equation 6 can be used to calculate the standard concentration (Cstd) from the ambient
concentration (Camb) data using ambient barometric pressure and temperature data (Pamb and
Tamb) from the same time period in which the ambient concentration was recorded. Pstd and
Tstd are the values of standard barometric pressure and standard ambient temperature.
These values are usually the EPA mandated 760 mmHg and 298 degrees Kelvin (25 C). Note:
Some other countries use different values for standard temperature and pressure.

Equation 7
Camb = Cstd * (Pstd / Pamb) * (Tamb / Tstd)

Equation 7 can be used to calculate the ambient concentration (Camb) from the standard
concentration (Cstd) data using the ambient temperature and pressure. It is necessary to have
access to valid data for the ambient temperature and pressure for the desired sample hour in
order to be able to make the calculations.

Example: You have a data value of 27µg from a BAM which was configured to report data in
EPA Standard conditions (298K and 760 mmHg), but you need to know what the concentration
would have been in actual conditions. The actual average temperature for the hour in
question was 303K and the average pressure was 720mmHg.

Camb = Cstd * (Pstd / Pamb) * (Tamb / Tstd)


Camb = 27 * (760/720) * (303/298)
Camb = 27 * 1.0556 * 1.0168
Camb = 28.98 µg

BAM-1020-9800 REV F Page 77


BAM-1020 Audit Sheet
Model: BAM-1020 Serial Number:
Audit Date: Audited By:

Flow Audits
Flow Reference Standard Used: Model: Serial No: Calibration Date:
Temperature Standard Used: Model: Serial No: Calibration Date:
Barometric Pressure Standard Used: Model: Serial No: Calibration Date:
Leak Check Value: as found: lpm as left: lpm
BAM Ref. Std. BAM Ref. Std.
Ambient Temperature: as found: C C as left: C C N/A
Barometric Pressure: as found: mmHg mmHg as left: mmHg mmHg
Flow Rate (Actual Volumetric): as found: lpm lpm as left: lpm lpm N/A
Flow Rate (EPA Standard): as found: lpm lpm as left: lpm lpm N/A

Mechanical Audits
Pump muffler unclogged: as found as left PM10 particle trap clean: as found as left N/A
Sample nozzle clean: as found as left PM10 drip jar empty: as found as left N/A
Tape support vane clean: as found as left PM10 bug screen clear: as found as left N/A
Capstan shaft clean: as found as left PM2.5 particle trap clean: as found as left N/A
Rubber pinch rollers clean: as found as left Inlet tube water-tight seal OK: as found as left
Chassis ground wire installed: as found as left Inlet tube perpendicular to BAM: as found as left

Analog Voltage Output Audit N/A Membrane Audit Flow Control Range
DAC Test Screen BAM Voltage Output Logger Voltage Input LAST m (mg): Flow Setpoint BAM Flow
0.000 Volts Volts Volts ABS (mg): 14.0 LPM
0.500 Volts Volts Volts Difference (mg): 16.7 LPM
1.000 Volts Volts Volts % Difference: 17.5 LPM

Setup and Calibration Values


Parameter Expected Found Parameter Expected Found Parameter Expected Found
Clock Time/Date FLOW TYPE AP
RS232 Cv FRl
STATION # Qo FRh
RANGE ABS Password F1 F2 F3 F4
BAM SAMPLE µ sw Cycle Mode
MET SAMPLE K Factor RH Control
OFFSET BKGD RH Setpoint
CONC UNITS STD TEMP Datalog RH
COUNT TIME HEATER Delta-T Control
FLOW RATE e1 Delta-T Setpoint
CONC TYPE Errors Datalog Delta-T

Last 6 Errors in BAM-1020 Error Log


Error Date Time Error Date Time
1 4
2 5
3 6

Audit Notes:

Page 78 BAM-1020-9800 REV F


OPERATOR NOTES:

BAM-1020-9800 REV F Page 79

You might also like