BAM-1020-9800 Manual Rev F
BAM-1020-9800 Manual Rev F
BAM-1020-9800 Manual Rev F
REVISIONS
REV DESCRIPTION DATE BY
A EO 1624. This revision supersedes previous Rev 2.0. 03-29-01 -
Added discussion on BX-961 volumetric control.
Added discussion on Standard and Early Cycle Mode.
Updated Field Calibration of Flow System.
Effectivity: This revision manual required with 3236 Rev. C (Firmware 02.03).
B EO 1733: Modifies Caution Statement, Section 4.3, Section 6.3. 01-31-01 -
C Added this page (Title Block, etc...) 06-11-01 JJE
Section 4.14: Changed definition of FLOW TYPE.
Section 5.5: Added description on how the flow is reported.
D 1JAN02 MH
1. Section 2.3 – Details of the options for enclosures. Added siting for inlet radius etc. Steps 1-5
2. 2.4 added note on complete instruction for the BAM 1020.
3. Added Section 2.6
4. Equation 3.3 – units analysis shows the equation is incorrect. Added mu coefficient.
5. Section 3.2 – Clarified the span measurement purpose. Changed compensation to verification.
6. Section 3.2.3 added reference to Appendix B and spec’d the tolerances of the span
measurement
7. Section 3.2.4 Changed PM10 inlet filter to PM10 inlet head.
8. Section 3.3 added a reference to Appendix B for the factory values.
9. figure 6 – added flow screen, Added align screen
10. Added a heading to figure 7 BAM 1020 grounding terminals
11. Added Section 4.8 Leak Check
12. Section 4.13 minor grammar mistakes
13. Section 4.14 clarified flow type.
14. Section 4.16.1 added references to section 5.2 Appendix C.
15. Clarified sec5.2 note about function keys
16. Added details to cal data flag in section 5.2 and 5.6
17. Changed the wording and clarified flow type table in section 5.5
18. Added a section direction to 5.7
19. Reworked Section 6.3
20. Added Section 6.4
21. Added explanation of ABS in App B to Sect7.7
22. Added Parity switch info to Sec 8.4
23. Added Section 4.9 identifier in sect 9.1
24. Corrected the section identifier in section 9.1.2
25. Identified the Appendix in 9.1.4
26. Corrected the section identifier in section 9.1.5
27. Moved troubleshooting guide to sect 9.2
28. Added periodic maintenance to section 10.
29. Added Consumables and Optional Equip Sect 10.3
BAM-1020-9800 REV F
E 1. PG12 heater Kit info 25FEB02 MH
2. PG40 Heater in the setup screen with explanation below
3. PG40 Heater Control Details
4. PG46 Changed the Error Screen reflect new FW
5. PG56 HEATER section 4.20
6. PG82 HEATER SCREEN in TEST Menu Section 7.11
7. PG90 Section 8.6 Added 6 CSV DATA Output
8. PG99 Added Smart Heater to list of BAM1020 options BX-827 and BX-830
9. PG105 Changed 6 – none to 6 – CSV DATA Output.
F Major revision of manual. Updated all sections for accuracy and completeness from customer and 12-20-2006 DAH
Service dept feedback. Fixed countless mistakes, omissions, obsoletes. Reorganized section structure for
better flow and logical layout. Made operational changes for PM2.5 EPA equivalency testing and new 8- 3-27-2007
minute firmware. Effectivity: This revision is for all units with 3.2.4 or later firmware and all 2X
hardware improvements.
• Incorporated changes from 2X mechanical improvements.
• Revised introduction. Added brief beta description, left space for PM2.5 EPA designation.
• Revised Installation section. Site selection, enclosure selection, mounting options.
• Revised accessories section. Removed obsolete items, added items and pics, reorganized.
• Revised errors section. Greatly expanded description of all errors.
• Reworked specifications: updated for 2X units, clarified!! Moved to front.
• Revised 500 series sensor connection section. Was Appendix F.
• Moved the advanced theory of operation section to the back where it belongs.
• Revised flow calibration section, moved leak check and nozzle cleaning to flow cal section.
• Incorporated my flow type description sheet. Moved flow section forward.
• Revised maintenance schedule. Added items. Changed to table. Moved to Troubleshooting.
• Created user interface drawing and simplified the ridiculous keypad description!
• Rewrote SETUP section, moved important setups to front after install.
• Rewrote TEST section, moved to Troubleshooting section.
• Rewrote Communications section, added HyperTerminal setup, simplified command list.
• Moved Telemetry section to its own section, revised for clarity.
• Replaced HUGE photos, reduced manual file size from 30MB to 2.7MB!
• Simplified early cycle mode description, moved to telemetry section.
• Added Audit Sheet and notes page.
• Revised Troubleshooting information.
• Reduced from 113 pages to 80 pages.
• Moved all appendices to body text.
• Added new changes for 3.2.4 PM2.5 firmware.
• Added new count times, flow cal, flow stats, offsets, data outputs, flash upgrade from 3.2.4.
• Added field BKGD determination procedures.
• Added BX-596 AT/BP sensor information.
• Added Dr. Schloesser’s concentration STD/Actual conversion formulas to Theory section.
• Added suggestions from Mike Putnam on communication, accessories, etc.
• Added Jeff Elseth’s corrections for telemetry relays and errors.
• Added EPA PM2.5 sampling requirements.
• Refined CONC TYPE and FLOW TYPE sections.
Printing Instructions:
• Do not include the revision block pages as part of the printed manual (this page).
• Print double sided final from single sided master. Must be a printout or a high-quality
first generation copy. No copies of copies!!!
• Include standard Met One warranty sheet as last page.
BAM-1020-9800 REV F
BAM 1020
PARTICULATE MONITOR
OPERATION MANUAL
BAM-1020-9800 REV F
BAM-1020 Particulate Monitor Operation Manual - © Copyright 2001, 2007 Met One Instruments, Inc. All Rights
Reserved worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or
translated into any other language in any form without the express written permission of Met One Instruments, Inc.
BAM-1020-9800 REV F
Table of Contents
1 INTRODUCTION 4
1.1 About This Manual ..................................................................................................... 4
1.2 Technical Service....................................................................................................... 4
1.3 BAM: Beta Attenuation Monitor .................................................................................. 4
1.4 Beta Radiation Safety Statement ............................................................................... 5
1.5 Model BAM-1020 PM10 USEPA Equivalent Method................................................... 5
1.6 Model BAM-1020 PM2.5 Monitoring ............................................................................ 6
1.7 BAM-1020 Specifications ........................................................................................... 7
2 SITE SELECTION AND INSTALLATION 8
2.1 Unpacking and Inspection .......................................................................................... 8
2.2 Enclosure Selection ................................................................................................... 8
2.3 Site Selection ............................................................................................................. 9
2.4 Mounting Options in a Walk-In Shelter..................................................................... 10
2.5 Installation Instructions in a Walk-In Shelter ............................................................ 12
2.6 Electrical Connections.............................................................................................. 13
3 INITIAL SETUP OF YOUR BAM-1020 15
3.1 Power On ................................................................................................................. 15
3.2 Warm-up .................................................................................................................. 15
3.3 Using the Keypad and Display ................................................................................. 16
3.4 Filter Tape Loading .................................................................................................. 17
3.5 Self-Test................................................................................................................... 18
3.6 Initial SETUP Settings Considerations ..................................................................... 19
3.7 Initial Leak Check and Flow Check .......................................................................... 20
3.8 Starting a Measurement Cycle ................................................................................. 20
3.9 The Flow Statistics Screen....................................................................................... 20
3.10 The OPERATE Screen ............................................................................................ 21
3.11 The NORMAL Screen .............................................................................................. 21
3.12 The INSTANTANEOUS Screen ............................................................................... 22
3.13 The AVERAGE Screen ............................................................................................ 22
4 THE MEASUREMENT CYCLE 23
4.1 The One-Hour Cycle Timeline.................................................................................. 23
4.2 Automatic Span Check During The Cycle ................................................................ 24
4.3 Sample Period Description....................................................................................... 25
5 FLOW SYSTEM and FLOW CALIBRATIONS 26
5.1 Flow System Diagram .............................................................................................. 26
5.2 Flow Type Descriptions............................................................................................ 27
5.3 Leak Check Procedure............................................................................................. 28
5.4 Leak Check Nozzle Seal .......................................................................................... 29
5.5 Nozzle and Vane Cleaning....................................................................................... 29
5.6 Field Calibration of Flow System – Actual (Volumetric) Flow Mode ......................... 30
5.7 Field Calibration of Flow System – EPA Standard Flow Mode................................. 31
5.8 Field Calibration of Flow System – Metered Flow Mode ......................................... 32
6 SETUP MENU DESCRIPTIONS 34
6.1 CLOCK Screen ........................................................................................................ 35
6.2 SAMPLE Screen ...................................................................................................... 35
6.3 CALIBRATE Screen................................................................................................. 37
6.4 EXTRA1 Screen....................................................................................................... 39
The sections and sub-sections marked with an Information Symbol contain critical
information which all BAM-1020 owners and operators should read and understand. Toward
the back are several sections that provide in-depth information on subjects such as theory,
diagnostics, accessories, and alternate settings. These sections are not as important for the
typical user to study, but provide valuable information which should be consulted as needed.
The EPA designation applies to the G, -1 and the G -1 BAM-1020 PM10 Beta Attenuation
Monitor, when used in conjunction with the following requirements. Users are advised that
configurations that deviate from this specific description may not meet the applicable
requirements of 40 CFR, parts 50 and 53.
• The BAM-1020 is operated for 24-hour average measurements with a filter change
frequency of one hour.
• The inlet must be equipped with the standard BX-802 EPA PM10 inlet head.
• The unit must be used with standard glass fiber filter tape.
• The unit may be operated with or without any of the following options: BX-823 inlet
tube extension, BX-825 heater kit, BX-826 230V heater kit, BX-828 roof tripod, BX-902
exterior enclosure, BX-903 exterior enclosure with temperature control, BX-961 mass
flow controller, BX-967 internal calibration device.
• The RANGE parameter must be set for 1.000 milligram/cubic meter.
• The SAMPLE TIME parameter must be set for 50 minutes.
• The PM2.5 concentration is calculated (external to the BAM) as a 24 hour daily average
of the hourly concentration measurements made by the BAM-1020. At least 18 of the
24 data points must be valid (no error flags) in order to calculate a daily average.
• The unit must be equipped with the BX-596 ambient temperature and barometric
pressure combination sensor. This is used for flow control and flow statistics. The unit
can function with the BX-592 or CARB style temperature sensor instead, but the flow
statistics during the sample will not be available
• The unit must be equipped with the internal BX-961 automatic flow controller, and
must be operated in Actual (volumetric) flow control mode.
• The inlet must be equipped with the standard BX-802 EPA PM10 inlet head and a
PM2.5 very sharp cut (VSCC) cyclone.
• The unit must be equipped with a BX-827 (110V) or BX-830 (230V) Smart Inlet Heater,
with the heater RH regulation setpoint set to 35%.
• The BX-302 zero filter calibration kit is a required accessory. This must be used to
periodically audit the BKGD (background) value as described in this manual.
• The unit must be used with standard glass fiber filter tape.
• The COUNT TIME parameter must be set for 8 minutes.
• The SAMPLE TIME parameter must be set for 42 minutes.
• The unit must be equipped with firmware revision 3.2.4 or later.
• The unit must be equipped with the 8470-1 rev D or later tape control transport
assembly with close geometry beta source configuration. BAM-1020 units
manufactured after March 2007 have these features standard. Older units will have to
be factory upgraded and re-calibrated to the latest hardware specifications.
• The unit may be operated with or without an inlet tube extension (BX-823) and with or
without external enclosures BX-902 or BX-903.
• The BAM-1020 must be operated in proper accordance with this operation manual,
revision F or later.
Remove the unit and accessories from the shipping boxes and compare the received items to
the packing list. Make sure you have all of the required items for the type of installation you
plan to perform.
The BAM-1020 is shipped with two white foam rings and a white plastic shim inside the front
of the unit, which prevent the moving parts of the tape control assembly from being damaged
in transit. Do not remove the foam rings until the BAM-1020 is ready to be installed. The rings
must be replaced anytime the unit is being transported in order to avoid damaging the tape
control mechanism.
Please keep the special box and foam packing material that the BAM-1020 was shipped in, to
re-use in the event that you must return the unit to the factory. Met One is not responsible for
any damages to the unit if returned in non-original packaging or without the foam rings in
place. Contact Met One for replacement packing materials if necessary.
1. A walk-in building or mobile shelter with a flat roof: This could be a trailer, pre-
fabricated shelter, or a room in an existing building. The BAM is mounted on a bench-
top or in an equipment rack, often with a variety of other instruments installed in the
same shelter. The inlet tube of the BAM must extend up through the roof with
appropriate hardware. AC power must be available. Instructions for this type of
installation are included in this section of the manual.
2. An environmentally controlled “Dog House” enclosure: A small pre-fabricated
enclosure just big enough for the BAM and related accessories. They are available
with a heater only (BX-902), or with a heater and air conditioner (BX-903). These
enclosures are supplied by Met One Instruments, and are custom designed to accept
the BAM-1020. The BX-902 and BX-903 are supplied with a separate setup and
installation manual.
NOTE: The air temperature inside any enclosure in which a BAM-1020 is installed must be
held as constant as possible over the course of the hour. This is important because the unit
measures the beta particles through a small gap of air around the filter tape at the beginning
and the end of the hour. If the air temperature inside the enclosure has changed by more
Inlet Height:
• The inlet should be located in the “breathing zone”, between 2 and 15 meters above
ground level. If the BAM is to be installed in an enclosure at ground level, then the inlet
must be at least two meters above the ground.
• If the inlet is located on or through a rooftop, the total height should be no more than
15 meters from the ground level. The inlet should be located at least two meters above
roof surface of the building that the unit is installed in. This matches the specified inlet
height of most FRM samplers.
• If the BAM-1020 is to be co-located with other particulate instruments, such as FRM
filter-based samplers or other BAM units, then the air inlet must be the same height as
the inlet of the other samplers.
• The BX-902 and BX-903 environmental shelters are designed to locate the inlet two
meters above whatever surface they are placed on.
• Make sure to account for the height of the PM10 and/or PM2.5 heads when planning the
inlet tube length. Met One can supply a variety of tube lengths up to 8-feet long.
• The maximum allowable total inlet tube length is 16 feet between the BAM-1020 and
the bottom of the inlet head.
Spacing from Roadways: The BAM-1020 should usually not be located directly next to a
major highway or arterial roadway, as vehicle exhaust will dominate the concentration
measurement. This effect can be difficult to predict accurately as shifting winds may direct the
plume toward or away from the BAM inlet.
• Roads with a daily traffic volume of less than 3,000 vehicles are generally not
considered major sources of pollutants, and in this case the BAM must be located at
least 5 meters from the nearest traffic lane.
• The BAM must be located at least 25 meters from any elevated roadway greater than
5 meters high.
• The unit should be located as far as possible from unpaved roadways, as these also
cause artificial measurements from fugitive dust.
• The unit should not be installed in unpaved areas unless year-round vegetative ground
cover is present, to avoid the affects of re-entrained fugitive dust.
• Rear Access: It is important that your mounting leaves plenty of access to the rear of
the BAM-1020 unit for wiring connections and maintenance.
• Top Access: It is necessary to have a minimum of eight (8) inches clearance between
the top of the inlet receiver on the BAM and the bottom of the ceiling of the shelter to
accommodate the inlet heater which mounts on the inlet tube directly above the BAM.
• Rack Modifications: It is often necessary to modify the top of the equipment rack by
cutting a hole to allow the inlet tube to extend up through to the ceiling. The drawing
below shows the location of the hole. Note: The inlet heater is a cylinder which installs
on the inlet tube two inches above the top of the inlet receiver of the BAM-1020. If the
BAM unit is to be mounted high in the rack, it may be necessary to make the hole in
the top of the rack larger in order to clear the heater diameter. The heater is supplied
with an insulation tube cover which may be modified as needed. Make sure these
parts are going to fit before installing the BAM-1020.
• Mobile Shelters: If the BAM-1020 is being installed into an equipment rack in a
mobile shelter or van, the instrument must be supported on the bottom in addition to
the rack brackets due to additional strain. The foam rings must also be inserted any
time a mobile shelter is moved with the BAM-1020 inside.
1. Roof Modifications: Determine the exact location where the BAM inlet tube will pass
through the roof of the enclosure, and drill a 2 ¼” or 2 ½” diameter hole through the
roof at that location. Make sure the hole is directly above where the BAM inlet receiver
is to be located, as the inlet tube must be perfectly vertical. A plumb-bob is useful for
determining where to locate the hole. Note: The inlet receiver on the BAM is slightly
off of the center line of the unit.
2. Waterproof Flange: Apply plenty of all-weather caulking around the top of the hole,
and install the BX-801 roof flange onto the hole. It is usually best if the threaded barrel
of the flange assembly is installed downward, into the hole. Secure the flange in place
with four lag bolts or self-tapping screws (not supplied). Caulk around the screws to
prevent leaks. Apply Teflon tape to the threads of the gray plastic watertight fitting, and
screw it into the roof flange.
3. Inlet Tube: Remove the white cap and rubber seal from the flange assembly. This
makes it easier to install the inlet tube, as the rubber seal is a tight fit around the tube.
Lower the inlet tube through the flange assembly in the roof and into the inlet receiver
on the BAM Make sure the inlet tube is fully seated.
4. Inlet Alignment: It is very important for the inlet tube to be perpendicular to the top of
the BAM. The nozzle may not close properly if there is a bind caused by misalignment.
A simple way to check is by trying to rotate the inlet tube with your hand (before
tightening the roof flange seal or the BAM inlet set screws). If the inlet tube is installed
straight, then the tube should rotate fairly easily when installed into the BAM. If it does
not rotate, check the inlet tube for vertical alignment or move the BAM slightly.
5. Smart Heater: Before tightening the inlet tube in place, the BX-827 or BX-830 smart
inlet heater (used on most BAM-1020 units) must be installed on the tube. Pull the inlet
tube up out of the inlet receiver a few inches, and pass the tube through the hole in the
heater body (the cable comes out the bottom). Then re-insert the inlet tube into the
BAM. Position the bottom of the heater unit two inches above the top of the inlet
receiver on the BAM, and securely tighten the two set screws in the heater to fasten it
to the tube. Included with the smart heater is a 12” tube of insulation. The tube is split
down its length for easy application. Wrap the insulation around the heater body and
peel back the adhesive cover strip to secure in place. The insulation may be cut to fit if
needed.
6. Inlet Tightening: Slide the rubber seal and white cap down over the top of the inlet
tube and into the roof flange. It is easier if you wet the seal first. Tighten the white
plastic cap.
7. Support Struts: The inlet kit comes with two angled aluminum struts to support the
inlet tube above the roof. These struts are usually fastened (90 degrees apart) to the
inlet tube with the supplied hose clamp. The other end of the struts should be fastened
to the roof with lag bolts (not supplied). Note: Some installations require different
configurations for supporting the inlet tube.
8. Temperature Sensor: Most BAM-1020 units are supplied with a BX-592
(temperature) or BX-596 (temperature and pressure) probe, which is attached to the
inlet tube above the roof. The cable must feed into the shelter to be attached to the
1. BAM-1020: Plug the BAM-1020 into AC power with the provided power cord. Note:
Met One recommends plugging the BAM-1020 unit into a battery back-up UPS
(uninterruptible power supply) as even a momentary power outage will ruin the entire
hour worth of data collection. A small computer-style UPS is usually sufficient as the
BAM does not draw much power. The air pump does not need to be connected to the
UPS as the BAM can compensate for short pump outages.
2. Chassis Ground: Connect one of the terminals marked “CHASSIS” on the back of the
BAM to a ground point as close as possible to the instrument. Use the supplied
green/yellow ground wire supplied with the unit. A ground rod, cold water pipe, or
junction box safety ground are possible connection points. Be sure to follow electrical
codes. Note: the BAM-1020 also uses a standard safety ground in the power cord.
3. Pump Connection: Decide on a location to place the air pump. The best location is
often on the floor under the rack or bench, but may be up to 25 feet away if desired.
Route the air tubing from the pump to the back of the BAM unit, inserting it firmly into
the compression fittings on both ends. The tubing should be cut to the proper length
and the excess saved for replacements. The pump is also supplied with a 2-wire signal
cable which goes between the BAM-1020 and the pump to turn the pump on and off.
Connect this cable to the terminals on the back of the BAM marked “PUMP
CONTROL” The end of the cable with the square black ferrite filter goes to the BAM,
BX-596 AT/BP
Wire Color Terminal Name
Yellow Channel 6 SIG
Black/Shield Channel 6 COM
Red Channel 6 POWER
Green Channel 6 ID
White Channel 7 SIG
BX-592 AT
Wire Color Terminal Name
Yellow or White Channel 6 SIG
Black/Shield Channel 6 COM
Red Channel 6 POWER
Green Channel 6 ID
Additional Met One BX-500 series sensors may be connected to channels 1 through 5
to log various other parameters. Details on these sensors are given in Section 10.2 of
this manual.
5. Smart Heater: Plug the heater cable into the mating green metal connector on the
back of the BAM-1020.
6. Optional Data Logger Connection: The BAM-1020 has an analog output which may
be recorded by a separate data logger if required. Connect the terminals on the back
of the BAM marked ”VOLT OUT +, -“ to the data logger with 2-conductor shielded
cable (not supplied). Polarity must be observed. Information on configuring this analog
output is provided in Section 8 of this manual.
7. Other Connections: The BAM-1020 has a variety of telemetry I/O relays and error
relays located on the back of the unit. There are also RS-232 data connections. These
items are described in Section 8 and Section 9 of this manual.
3.1 Power On
The BAM-1020 has a power switch located on the back of the unit directly above the power
cord. Verify that the unit is plugged in to the correct AC voltage, and that any electrical
accessories are correctly wired before turn the unit on. (Section 2.6) When power is switched
on the main menu screen should appear after a few seconds as shown below. The unit will
probably flash an error indicating that there is no filter tape installed.
3.2 Warm-up
The BAM-1020 must warm up for at least one hour before an operation cycle is started. This
is because the beta detector contains a vacuum tube which must stabilize every time the unit
is powered up. This also allows the electronics to stabilize for optimal operation. This applies
any time the unit is powered up after being off for more than a moment. Instrument setups
and filter tape installation can be performed during this time. Most agencies choose to discard
the first few hours of concentration data after the BAM is powered up.
F1 F2 F3
F4 F5 F6
Soft Keys:
Directly beneath the display are four white buttons called “soft-keys” or “hot-keys”. These are
dynamic keys who’s function changes in response to a menu option displayed directly above
each key on the bottom row of the display. Whatever menu option is displayed above one of
these keys is the function which that key will perform in that particular menu. These are used
throughout the entire menu system for a wide variety of functions. Modifications made within
a menu are usually not saved unless a SAVE soft-key is pressed. EXIT is also another
common soft-key function.
Contrast Key:
The key with a circular symbol on it is for adjusting the light/dark contrast on the LCD display.
Press and hold the key until the desired contrast is achieved. It is possible to adjust the
contrast to make the entire display completely blank or completely dark, so be careful to set it
to a visible level or it may appear that the unit is not operating.
1. Turn the BAM-1020 on and enter the TAPE menu (Note: This is not the same as the
TEST > TAPE menu). If the nozzle is not in the UP position, press the TENSION soft-
key to raise the nozzle.
2. Lift the rubber pinch roller assembly and latch it in the UP position. Unscrew and
remove the clear plastic spool covers.
3. An empty core tube MUST be installed on the left (take-up) reel hub. This provides a
surface for the used tape to spool-up on. Met One supplies a plastic core tube to use
with the first roll of tape. After that, you can use the empty core tube left over from your
last roll to spool-up the new roll. Never fasten the filter tape to the aluminum hub.
4. Load the new roll of filter tape onto the right (supply) reel, and route the tape through
the transport assembly as shown in the drawing. Attach the loose end of the filter tape
to the empty core tube with cellophane tape or equivalent.
5. Rotate the tape roll by hand to remove excess slack, then install the clear plastic spool
covers. The covers will clamp the rolls to the hubs to prevent slipping.
6. Align the filter tape so that it is centered on each roller. Newer units have score marks
on the rollers to aide in visually centering the tape.
7. Lower the pinch roller assembly onto the tape. The BAM will not function if the pinch
rollers are latched up, and it has no way of automatically lowering the assembly!
8. Press the TENSION soft-key in the TAPE menu. The BAM-1020 will set the tape to the
correct tension and alert you if there was an error with the process. Exit the menu.
3.5 Self-Test
The BAM-1020 has a built-in self-test function which automatically tests most of the tape
control and flow systems of the unit. The self-test should be run right after each time the filter
tape is changed, and it can also be used if the operator suspects a problem with the unit.
More detailed diagnostic menus are also available in the BAM, and those are described in the
troubleshooting section.
The self-test feature is located in the TAPE menu. Press the SELF TEST soft-key to start the
test. The tests will take a couple of minutes, and the BAM-1020 will display the results of
each tested item with an OK or a FAIL tag. If all of the test items are OK, the status will show
SELF TEST PASSED as shown in the drawing below. If any item fails, the status will show
ERROR OCCURRED.
LATCH: This will show OFF if the photo interrupter senses that the pinch rollers
are unlatched as in normal operation. It will show ON if the roller assembly is latched in the
up position. The tape cannot move if the rollers are up!
CAPSTAN: The unit will rotate the capstan shaft forward and backwards and will
check if the photo interrupter sees the shaft rotating. The Capstan shaft is what moves the
filter tape back and forth.
NOZZLE DN: The unit will attempt to lower the nozzle, and will check if the nozzle
motor has moved to the down position with a photo interrupter. It is possible for the nozzle to
become stuck in the UP position, even if the nozzle motor has successfully moved to the
DOWN position. For this reason, proper inlet alignment and maintenance is necessary.
NOZZLE UP: The unit will attempt to raise the nozzle, and will check if the nozzle
motor has moved to the up position with a photo interrupter.
FLOW: The unit will attempt to turn the pump on, and will then look for output on
the flow sensor. This test takes about a minute and will fail if the pump is not connected.
TAPE BREAK: The unit will move the supply and take-up motors to create slack in the
filter tape, and look for proper operation of the tensioner photo interrupters.
TAPE TENSION: The unit will tension the filter tape, and then check the condition of the
tensioner photo interrupters.
SHUTTLE: The unit will attempt to move the shuttle beam left and right, and will
check the motion with a photo interrupter.
REF EXTEND: The unit will attempt to extend the reference membrane, and will check
the motion with a photo interrupter.
REF WITHDRAW: The unit will attempt to withdraw the reference membrane, and will check
the motion with a photo interrupter.
1. Set the system clock in the SETUP > CLOCK menu. The BAM-1020 clock may drift as
much as a couple of minutes per month. It is important to check the clock at least once
per month to ensure the samples are performed at the correct times.
2. Review the BAM SAMPLE, COUNT TIME, MET SAMPLE, RANGE, and OFFSET
values in the SETUP > SAMPLE menu.
↑
= ON
↓
= OFF
Operation Mode: ON
Status: ON
The DOWN arrow can be used to set the Operation Mode from ON to OFF. This will stop the
measurement cycle, but will not power-down the BAM-1020. NOTE: If the operator sets the
Operation Mode to OFF, or the unit stops itself due to an error, it will still automatically
set the mode back to ON at the top of the hour, and try to run a new cycle! The only
ways to prevent the unit from automatically starting a cycle are to power off the unit, leave the
unit in a TEST or SETUP menu, or leave the pinch rollers latched in the UP position.
The OPERATE menu has three soft-key options for viewing the operating status and sensor
measurements while the unit is operating: NORMAL, INST, and AVERAGE.
The LAST C value indicates the last concentration recorded, updated at the end of the cycle.
The LAST m value indicates the last measured value of the reference membrane. The value
should be very close or equal to the expected value (ABS). The other values are
instantaneous measurements.
The TOGGLE FLG soft-key in this menu allows the user to set the CAL DATA FLAG value
ON or OFF, which marks the data with an M flag to indicate a maintenance was performed
during that time, such as a flow check. This feature is rarely used, as most maintenance
requires stopping the sample anyway. The VOLT / ENG soft-key toggles the displayed values
between units and voltages, useful for diagnostic checks on external sensors.
Nozzle
Beta Source
Beta Source
Detector
Detector
This example shows the timeline if the unit is set for a COUNT TIME of 8 minutes (required
for PM2.5). If the unit is set for 4 minutes, then the beta counts at the beginning and end of the
hour will be only 4 minutes long, with a 50-minute air sample in between. Again, the total
adds up to 58 minutes. Note: This cycle will be slightly altered if the unit is operated in the
Early Cycle mode with an external datalogger. See Section 8.
1. Minute 00: The beginning of an hour. The BAM-1020 immediately advances the
filter tape forward one “window” (the next fresh, unused spot on the tape). This takes a
few seconds. The new spot is positioned between the beta source and the detector,
and the BAM begins counting beta particles through this clean spot for exactly eight
minutes. (I0)
2. Minute 08: The BAM-1020 stops counting beta particles through the clean spot (I0),
and moves the tape exactly four windows forward, positioning that same spot directly
under the nozzle. This takes a few seconds. The unit then lowers the nozzle onto the
filter tape and turns the vacuum pump on, pulling particulate-laden air through the filter
tape (the spot in which I0 was just measured) for 42 minutes at 16.7 liters per minute.
3. Minute 50: The BAM-1020 turns the vacuum pump off, raises the nozzle, and moves
the filter tape backwards exactly four windows. This takes a few seconds, and puts the
spot that was just loaded with particulate back between the beta source and the
1. Minute 08: (When the BAM-1020 has just finished moving the clean spot to the
nozzle and turned the pump on) there is another clean spot of filter tape upstream four
windows, between the beta source and the detector. This same spot will stay there for
the entire time the pump is on (usually 42 minutes), as the tape cannot move with the
nozzle down. The BAM begins counting the beta particles through this spot for exactly
eight minutes (I1).
2. Minute 16: The BAM-1020 stops counting beta particles through this spot (I1), and
extends the Reference Membrane between the beta source and the detector, directly
above the spot of filter tape that was just measured. The Reference Membrane is an
extremely thin film of clear Mylar held in a metal tongue. The membrane usually has a
mass of about .800 mg. The BAM starts counting beta particles for eight minutes
again, this time through the membrane and the filter tape spot at the same time (I2).
3. Minute 24: The BAM-1020 stops counting beta particles through the membrane (I2),
withdraws the membrane assembly, and calculates the mass of the membrane “m”,
as if it were particulate on the filter tape spot.
4. Minute 42 (typical): (Eight minutes before the pump stops) The BAM-1020
counts the beta particles through the same spot again (without membrane) for another
eight minutes (called I1’ or I1 prime). This checks the ability of the unit to hold a
constant output when measuring blank filter tape, and is not otherwise used.
The mass density “m” (mg/cm2) of the reference membrane calculated during this automatic
process is compared to the known mass of the membrane (“ABS” value). During factory
calibration, the average value of “m” is determined over at least 24 hours and saved as the
“ABS” value. The percentage of the latest value of “m” from “ABS” should be within the limits
of known variability (+/-4%) If not, the unit records an error for that hour’s data. The value of
“m” provides a method of internal diagnostics and for the monitoring of external variables,
such as temperature swings or pressure changes. The “ABS” value is unique to each BAM-
1020, and can be found on the calibration sheet.
The air then goes down the inlet and through the filter tape, where the remaining particles are
deposited. Some particles smaller than about 0.2 µm may pass through the filter tape and be
exhausted. After the sample period is completed and the particulate spot is measured, there
is almost always a clearly visible spot of dirt on the filter tape where the particulate was
deposited. The BAM-1020 will put the spots very close together on the tape. At exactly
midnight, the BAM will skip one spot, leaving a blank spot on the tape. This is a visual aid
which separates daily entries on the tape.
Automatic Flow
Beta Controller
Block
Debris
Filter
Outlet
This section describes the different types of flow control and regulation schemes used in the
BAM-1020. The unit can be set to any of three different flow types: Metered, Standard, or
Actual (Volumetric), depending on the hardware available and the desired reporting
conditions. All BAM-1020 units have a mass airflow sensor and a barometric pressure
sensor. The unit also has either a manual airflow valve on the back of the unit, or an
automatic flow control valve inside the unit. The unit may also have an optional BX-592 or
BX-596 ambient temperature sensor. Each flow type requires a different process for auditing
and calibrating the flow. To verify or set the flow type of the BAM, go to the SETUP >
CALIBRATE menu, and check the FLOW TYPE. NOTE: The concentration reporting
conditions can now be set independently of the flow type. See section 6.3.
1. Remove any PM10 and PM2.5 heads from the inlet tube. Install a BX-305 leak test valve
(or equivalent valve for auditing FRM samplers) onto the inlet tube. Turn the valve to
the OFF position to prevent any air from entering the inlet tube.
2. In the TEST > TAPE menu, advance the tape to a fresh, unused spot.
3. In the TEST > PUMP menu, turn on the pump. The flow rate should drop below 1.0
LPM. If the leak value is greater than 1.0 LPM, then the nozzle and vane may need
cleaning, or there may be another leak in the system.
4. Resolve the leak and perform the check again. A properly functioning BAM will usually
have a leak value of about 0.5 LPM with a clean nozzle and vane using this method.
5. Turn the pump off, remove the leak test valve, and re-install the inlet heads.
NOTE: The reason that there is a 1.0 LPM leak flow allowance is due to the test conditions.
With the inlet shut off the vacuum in the system is very high, about 21 inHg. This is about ten
times greater than the BAM-1020 will encounter during normal sampling. If the leak reading
during this test is 1.0 LPM or less, there will not be a significant leak during normal operation.
Note: If an as-found leak value is measured at 1.5 LPM or less, then previously collected data
is generally considered to be valid.
1. Raise the nozzle in the TEST > PUMP menu. Remove the filter tape (if installed) from
the nozzle area. It is not necessary to completely remove the tape from the unit.
2. With the nozzle up, use a small flashlight to inspect the cross-hair vane.
3. Clean the vane with a cotton-tipped applicator and isopropyl alcohol. Hardened
deposits may have to be carefully scraped off with a dental pick.
4. Lower the nozzle in the TEST > PUMP menu. Lift the nozzle with your finger and insert
another cotton swab with alcohol between the nozzle and the vane. Let the nozzle
press down onto the swab with its spring pressure.
5. Use your fingers to rotate the nozzle while keeping the swab in place. A few rotations
should clean the nozzle lip.
6. Repeat the nozzle cleaning until the swabs come out clean.
7. Inspect the nozzle lip and vane for any burrs which may cause leaks or tape damage.
Cotton
Applicator
Nozzle Lip
Vane
Nozzle Cleaning
1. Enter the TEST > FLOW menu as shown above. The nozzle will lower automatically
when this screen is entered. The “BAM” column is what the BAM-1020 measures for
each parameter, and the “STD” column is where you will enter the correct values from
your reference standard. The <CAL> symbol will appear next to the parameter
selected for calibration. The ambient temperature (AT) and pressure (BP) must be
calibrated first, as the BAM uses these parameters to calculate the air flow rate in
actual mode.
2. Measure the ambient temperature with your reference standard positioned near the
BX-592 or BX-596 ambient temperature probe. Enter the value from your reference
standard into the STD field using the arrow keys. Press the CAL hot key to correct the
BAM reading. The BAM and STD values should now be the same.
3. Press the NEXT hot key to move the <CAL> indicator to the BP field, and repeat the
same steps for barometric pressure.
4. After the temperature and pressure readings are correct, remove the PM10 and PM2.5
heads from the inlet tube and install your reference flow meter onto the inlet. Press the
NEXT hot key to move the <CAL> indicator to the first flow point of 15.0 LPM. The
pump will turn on automatically. Allow the unit to regulate the flow until the BAM
reading stabilizes at the target flow rate. Enter the flow value from your standard into
the STD field using the arrow keys. Press the CAL hot key to correct the BAM reading.
NOTE: The BAM reading will not change to match the STD until after you have
entered all three calibration points.
5. Press the NEXT hot key to move the <CAL> indicator to the second flow point of 18.3
LPM and repeat the process.
6. Press the NEXT hot key to move the <CAL> indicator to the third flow point of 16.7
LPM and repeat the process.
7. When all of the calibrations are complete, the BAM readings should match the STD
reading at 16.7 LPM, +/- 0.1 LPM. Exit the calibration menu.
The DEFAULT hot key can be pressed to reset the user calibration from the selected
parameter and replace it with a factory setting. If any of the FLOW parameters are selected,
the DEFAULT key will reset the calibrations of all three flow points. This feature can be used
to start over with a calibration if difficulty is encountered.
If the unit does not have the temperature sensor then you will not have access to the TEST >
FLOW screen. Use the following steps to check the flow instead:
1. Attach your flow standard onto the BAM inlet while the pump cycle is running or turn
the pump on in the TEST > PUMP screen. Allow the BAM flow to stabilize.
2. If your reference flow meter has a STANDARD flow reading available, that value can
be directly compared to the BAM flow reading. If your flow meter only has a volumetric
flow reading, then convert the volumetric flow rate to standard flow Qs with the
following formula:
Qs = Qa * (Pa / Ta) * (298 / 760)
Ta = Ambient Temperature (Kelvin) (Kelvin = Celsius + 273)
Pa = Ambient Barometric Pressure (mmHg)
Qa = Actual Volumetric Flow from Reference Meter
3. Compare the reference flow (converted to STD conditions) to the BAM flow reading
(also in STD conditions). The two should match within 1% (about 0.17 LPM). If not, a
full flow calibration should be performed. Change the FLOW TYPE to METERED and
perform a flow calibration using the Cv and Q0 values as described in section 5.8. The
sections about the manual flow valve do not apply. Set the flow type back to STD
when finished.
6. Convert the volumetric flow from the reference to EPA standard conditions:
7. Compare the reference flow (converted to STD conditions) to the BAM flow reading
(also in STD conditions). The two should match within 1%. If not, a full flow calibration
should be performed.
8. If the reference flow (converted to STD conditions) and BAM flow reading match, then
adjust the flow adjustment knob on the back of the unit until the BAM flow reading
equals 17.5 LPM.
V = (Ta / Pa)*62.4
4. Divide V by 24.47 to determine the ratio of EPA flow to Ambient Flow and record as
CALNUM.
5. Turn the pump on in the TEST > PUMP menu and allow the flow to stabilize for 5
minutes. Then divide the displayed flow by CALNUM.
6. Adjust the flow adjustment knob on the back of the BAM-1020 until the BAM flow
reading equals 17.5 / CALNUM. Exit the TEST menu.
Example:
Temp = 300 Kelvin
Pressure = 710 mmHg
V = (300/710) * 62.4 = 26.4
CALNUM = 26.4/24.47 = 1.08
Adjusted Flow = 17.5/1.08 = 16.2 LPM
Adjust flow valve until BAM reads 16.2 LPM.
WARNING: Some of the settings in the SETUP menus are unit-specific calibration
constants which cannot be changed without affecting the accuracy and proper
operation of the unit.
WARNING: Entering the SETUP menu system will require stopping the sample cycle.
Older versions of firmware will not warn you before stopping the sample.
Press the SETUP soft-key to enter the menu as shown below. The Setup Menu provides a
choice of operations. Use the arrow keys to navigate to the desired field, then press the
SELECT soft-key to enter.
SELECT EXIT
A brief description of each sub-menu in shown in the table below. Detailed information is
provided in the following sub-sections.
Menu Settings
CLOCK Date and Time Settings.
SAMPLE Range, Offset, Sample Time, Count Time, Conc Units, Avg Period, Unit ID, and RS-232 settings.
CALIBRATE Factory Calibration Values, (Cv, Q0, ABS, µsw, K, BKGD) Flow rate, Flow type, Heater type.
EXTRA1 Low concentration clamp, e1 – e4, Rarely used.
ERRORS Analog error selections, Flow limits, Pressure drop limit.
PASSWORD Password change screen.
INTERFACE Cycle Mode early/standard, alarm relay polarity.
SENSOR Meteorological sensor scaling and configuration screens, Channels 1 – 6.
HEATER RH and Delta-T set-points for Smart Heater. Only visible if Heater Control is set to AUTO.
SETUP SAMPLE
RS232 9600 8N1 BAM SAMPLE 042 MIN
RS-232: This allows you to set the baud rate of the RS-232 serial port. The available
values are 300, 600, 1200, 2400, 4800, 9600, 19200, and 38400 baud. Default is 9600. “8N1”
means 8 data bits, no parity, 1 stop bit. These cannot be edited.
BAM SAMPLE: This value sets the amount of time that the vacuum pump is on per cycle;
the actual sample period (see Section 4.1 for a description of the measurement cycle). The
BAM SAMPLE time must be set in response to the COUNT TIME value, since new versions
of the BAM-1020 allow the option of setting the count time to 4, 6, or 8 minutes. If the unit is
used for PM2.5 monitoring, the BAM SAMPLE must be set to 42 minutes with 8 minute count
time.
The BAM SAMPLE value can be set from 0-200 minutes for custom applications. If set for
shorter period, such as 15 minutes, the unit will finish the sampling and then wait until the end
of the hour before beginning a new cycle. This may not leave time for the membrane span
check. Only one cycle per hour is allowed, regardless of duration. Setting the value too long
may cause the measurement to extend over into the next hour. Contact the Service
department before setting this to anything but the values shown in the table.
MET SAMPLE: This value is the averaging period for the data logger. It sets how often
data is written to memory, and can be set to 1,5,15, or 60 minutes. For example, if an
external wind speed sensor is attached to the BAM, the MET SAMPLE period could be set to
1 minute. This would cause the BAM to store an average of the WS reading every minute.
This value applies to all sensors attached to the unit. NOTE: This setting will affect how
long the memory will last before getting full. A 1 minute average period will use up the
memory in a little over 3 days, 60 times faster than a 60 minute average period. When the
memory gets full the unit over-writes the oldest data. Met One recommends leaving the MET
SAMPLE period set at the default value of 60 minutes unless required for a particular
application. The dust concentration value will always be an hourly average anyway.
RANGE: The Range setting sets the full-scale range of the analog voltage output. The
RANGE value is almost never changed from the factory setting of 1.000 mg (corresponding
to an analog output voltage of 0.000 to 1.000 volts = 0 to 1000µg). In special cases, this
value may be set to 0.100, 0.200, 0.250, 0.500, 2.000, 5.000, or 10.000 mg. Be sure to
account for this value if using a separate data logger to record the BAM-1020 analog output.
Note: Changing the range setting will affect future data stored to memory. Always download
any old data before changing settings, then clear the BAM memory. Firmware version 3.2.4
or later will prompt you to clear the memory before letting you change this setting.
OFFSET: The OFFSET value is used to set the lower end of the BAM-1020 measurement
range, and could more accurately be called a “range offset”. The new factory default value for
OFFSET is now -0.015 mg. This causes the entire range of the BAM-1020 to shift down
slightly so that it can read from -0.015 to 0.985 mg, instead of measuring from 0 to 1.000 mg
(if the RANGE is set to 1.000 mg). This allows the unit to measure slightly negative
concentration numbers near zero, which is helpful to differentiate between normal noise and
a failure such as punctured filter tape.
The previous default was -0.005 mg, and the value may be set to -0.005 if needed to work
with data logging systems structured around the old setting. Some BAM users choose to set
the OFFSET value to 0.000 to avoid confusion, at the expense of not being able to see the
true zero noise floor of the unit.
This value also affects the analog output, so that 0 to 1.000 volts equals -0.015 to 0.985 mg,
instead of 0.000 to 1.000 mg. This is because the voltage output cannot go negative. You
must take this scaling into account if an external data logger is recording the BAM-1020
analog output voltage. Contact the Service department if you plan to set the OFFSET to any
value other than -0.000, -0.005, or -0.015 mg.
Note: The OFFSET value is often misunderstood, and should not be confused with the
BKGD (factory calibration zero correction factor) or the “e1” (lower concentration limit clamp)
values. Be sure you understand all three of these settings!
Note: Changing the offset setting will affect future data stored to memory. Always download
any old data before changing settings, then clear the BAM memory. Firmware version 3.2.4
or later will prompt you to clear the memory before letting you change this setting.
COUNT TIME: This is the amount of time the unit takes to perform the I0 and I3 counts.
Past versions of the BAM-1020 have always been fixed at 4 minutes. The new options allow
the count time to be set for 4, 6 or 8 minutes. When used to monitor PM2.5, this must be set to
8 minutes. This increases the sensitivity of the unit in lower concentrations. Increasing the
count time will require decreasing the sample time. For example, a count time of 4 minutes
allows a sample time of 50 minutes, while a count time of 8 minutes allows a sample time of
only 42 minutes. The unit will prompt you to change the sample time if you change the count
time to an incompatible value.
CALIBRATE SETUP
FLOW RATE: 16.7
CONC TYPE: ACTUAL FLOW TYPE: ACTUAL
Cv: 1.047 Qo: 0.000
ABS: 0.822 µsw: 0.306
K: 1.005 BKGD: -0.0030
STD TEMP: 25C HEATER: AUTO
SAVE EXIT
The SETUP > CALIBRATE Screen
FLOW RATE: This sets the air flow rate for the BAM-1020 sample period. The BAM will
continuously regulate the flow to this value (except manual valve models). The flow rate is
almost always set for 16.7 liters per minute, as this is required for all PM2.5 and PM10
monitoring. The operator may change this value temporarily in order to test the ability of the
pump and flow controller to regulate the flow at different levels. The range of this setting is 10
to 20 LPM.
CONC TYPE: This sets the way that the concentration values are reported. If set to
ACTUAL, then the concentration is calculated based on the volume of the air at ambient
conditions, and a BX-592 or BX-596 temperature sensor is required. If set to STD, the
concentration is calculated based on the standard values for temperature and pressure
(usually 25C and 760mmHg), even if a temperature sensor is available. This value is usually
set to match the FLOW TYPE setting, and must be set to ACTUAL for PM2.5 monitoring.
• METERED: Usually used for BAM units with a manual flow valve on the back. Flow is
reported in EPA conditions.
• STD: EPA Standard flow. The flow is controlled and reported in EPA standard
conditions. Used where required by regulations.
• ACTUAL: Actual Volumetric flow is controlled and reported to ambient temperature
and pressure conditions. This is required on PM2.5 monitors, and is recommended by
Met One whenever possible. BX-596 or BX-592 sensor is required.
Cv: This value is the factory-set Coefficient of Variability for the internal flow sensor. The
value of Cv is only changed by the user when performing a flow calibration on manual valve
(metered) or STD flow units. ACTUAL flow controlled units almost never need to have this
value altered.
Qo: This value is the factory-set zero correction factor for the internal flow sensor. The
value of Qo is only changed by the user when performing a flow calibration on manual valve
(metered) or STD flow units. ACTUAL flow controlled units almost never need to have this
value altered.
ABS: The ABS value is the factory-set mass of the reference membrane foil used during the
automatic span check. This value is compared to the measured value each hour (see section
4.2). Each unit’s ABS value is different, but is typically around 0.800 mg. The ABS value is
never changed by the operator unless the membrane is replaced due to damage.
µsw: This is called the Mu-switch value, and is the factory-set mass absorption coefficient
used by the BAM-1020 in the concentration calculations. Typical values are about 0.285 to
0.310. Warning: This is a unit-specific calibration value which may significantly affect
the accuracy of the unit. Never change this value without specific instruction from Met
One Instruments.
K: The K-factor is the factory-set slope correction (multiplier) for the BAM-1020
concentration. The K-factor value is determined by dynamic testing of the BAM-1020 in the
factory smoke chamber. This will always be a value between 0.9 to 1.1. All of the stored and
displayed data contains this correction. Warning: This is a unit-specific calibration value
which may significantly affect the accuracy of the unit. Never change this value
without specific instruction from Met One Instruments.
BKGD: The BACKGROUND value is the factory-set zero correction (slope offset) for
the BAM-1020 concentration. This is determined by running the unit for at least 48 hours with
a HEPA zero filter on the inlet. The concentration values over this time are averaged, and the
BKGD is the negative of this average. All of the stored and displayed data contains this
Page 38 BAM-1020-9800 REV F
correction. The BKGD value is typically between 0.000 and -0.005 mg/m3. Met One does offer
a zero filter inlet kit (BX-302) which may be used to audit this value. Warning: This is a unit-
specific calibration value which may significantly affect the accuracy of the unit. Note:
The BKGD value is not to be confused with the OFFSET (range offset) value in the SETUP >
SAMPLE menu. See section 6.2.
STD TEMP: This is the value of standard temperature, used for standard flow or
concentration calculations. In the U.S. the value of standard temperature is usually 25
degrees C as mandated by the USEPA. Some other areas use a standard temperature value
of 0C or 20C.
HEATER: This setting selects which mode the Smart Inlet Heater is used in. When set to
AUTO, the Smart Heater will use the filter RH and temperature sensors to control the inlet
tube heating. When set to MANUAL, the unit will simply turn the heater on all the time
regardless of filter conditions. The actual setup parameters for the Smart Heater are located
in the SETUP > HEATER menu, which will not appear unless this value is set to AUTO. The
operator may safely set this parameter as required. Met One recommends using the AUTO
setting. This value must be set to AUTO for PM2.5 monitoring.
e1 Low Concentration Limit. The lowest concentration value the BAM-1020 is allowed to
store or display, despite what it measured. Any measurements below this value will be
clamped. The range is -0.015mg to +0.010mg, and the default value is -0.015mg.
Note: This value is not to be confused with the OFFSET value which sets the lower
limit of the measurement range, or the BKGD value which is the zero correction.
e2 Not Used.
e3 Membrane OFF Delay. Hysteresis timer, range is 0.000 to 5.000 seconds. Don’t
change this value unless instructed to do so by Met One.
e4 Membrane Time Out. The time the unit allows for the membrane assembly to move
before generating an error. Range is 10.00 to 20.00 seconds. Don’t change this value
unless instructed to do so by Met One.
Regardless if a particular error is enabled for the analog output in this manner or not, it will
always be reported in the BAM-1020 internal memory, and may be viewed with the display or
by downloading the data through the serial port. Some of these errors such as P, R, N, and E
may be set to cause the analog output to go full scale, even though there may be nothing
wrong with that hour’s data. In this case, the concentration data can still be downloaded from
the BAM. Errors which DO indicate that the hour’s data is invalid, cause the digital
concentration value to be set to full scale too, usually .995mg.
This scheme is used because it is rare for an actual valid concentration reading to measure
full-scale. However, concentrations at or near zero can be common, so leaving the data value
at 0.000 during an alarm could be mistaken for valid data.
E EXTERNAL RESET: This error indicates that the system clock time was unable
to reset when signaled by an external datalogger. If external reset is successful then
no error is logged (see Section 8.2). Sometimes called INTERFACE RESET.
U TELEMETRY FAULT: This error indicates that an external datalogger has sent an
error to the BAM-1020 (on the TELEM FAULT input) indicating that it has encountered
a problem. Check the datalogger.
L POWER FAIL: This error occurs any time power is cycled or lost, even
momentarily. Frequent “L” errors usually indicate poor-quality AC power. In some
cases these errors can be generated by electrical interference (such as large radio
antennas or motors) causing an internal reset in the BAM-1020.
N NOZZLE STUCK TIMEOUT (or Delta-T exceeded): This error indicates that the
nozzle motor is not operating. The error is triggered if photo sensors S4 and S5 never
change state despite drive commands to nozzle motor, and if the sensors do not see
the nozzle motor move within 12 seconds of it being turned on. NOTE: The nozzle
motor lifts the nozzle, but the nozzle is lowered only by its spring. So it is possible for
the nozzle to become stuck in the UP position without generating an error! Proper
maintenance and inlet alignment prevents this.
The “N” error is also used to indicate that the Delta-Temperature set-point was
exceeded. This occurs if the sample air temperature (measured below the filter tape) is
hotter than the ambient air by at least one degree above the set-point value. This is
due to the normal heating of the sample air by the smart heater. In this case, the error
is used to simply flag the data. Frequent errors may indicate that the set-point is set
too low. In most applications Delta-T control is disabled entirely. See the inlet heater
setup instructions in this manual.
F FLOW ERROR: This error occurs if the average air flow over the sample period
was out of the limits set by the FRl (low limit) and FRh (high limit) values. Momentary
changes in airflow do not trigger the error. This error may begin to occur if the vacuum
pump is wearing out, or due to a fault with the flow sensor, flow controller, or air tubing.
The “F” error is also used to indicate if the ambient temperature or barometric
pressure sensor has failed or is incorrectly connected (only if the BAM is set for
ACTUAL flow or concentration reporting). This applies to auto ID sensors BX-592 and
BX-596, the internal filter pressure sensor, and CARB style temperature sensors. The
sensor is considered failed if the 1 minute average reading is at or beyond the min or
max measurement range of the particular sensor.
P PRESSURE DROP EXCESSIVE: This error indicates that the vacuum beneath
the filter tape has exceeded the limit set by the AP value. This is almost always
caused by high concentrations, or certain types of particulate clogging the filter tape.
When this error occurs, the BAM stops the pump to prevent overheating, completes
the measurement early, then waits for the top of the next hour. To increase the amount
of particulate which can build up on the tape before this occurs, set the AP value
higher.
C COUNT ERROR: This error indicates that the beta particle counting system is not
operating properly, and is activated if the beta count rate falls below 10,000 per 4
minutes. The beta count rate through clean filter tape is usually more than 600,000 per
T TAPE BREAK: This error indicates that the filter tape is broken or has run out.
The error is triggered if photo sensor S6 is ON continuously, despite drive commands
to motors M3-M5. Tape supply motor (M3) and tape take-up motor (M4) time out after
10 seconds. Capstan motor (M5) times out after 6 seconds. This error is also
generated if the pinch roller assembly has been left latched in the UP position when a
measurement cycle starts. Photo sensor S9 is ON any time the latch is set. The BAM-
1020 has no way of unlatching and lowering the pinch rollers. It must be done
manually.
AP Pressure-drop limit across the filter tape. The default setting is 150 mmHg, and the
range is 0-500 mmHg. See the PRESSURE DROP EXCESSIVE error definition above.
FRl Flow Rate Lower Limit. The default setting is 10 lpm, and the range is 0-30 lpm. See the
FLOW OUT OF LIMITS error definition above.
FRh Flow Rate Higher Limit. The default setting is 20 lpm, and the range is 1-38 lpm. See the
FLOW OUT OF LIMITS error definition above.
Interface Setup
SAVE EXIT
The INTERFACE screen
Fault Polarity: This sets the polarity of the Telemetry Fault Relay. NORM is normally
open, INV is normally closed. Rarely used.
Reset Polarity: This tells the BAM-1020 the incoming polarity of an external clock reset
signal, if used. This signal is used to synchronize the BAM clock to an external data logger.
NORM is normally open, INV is normally closed.
Cycle Mode: The Cycle Mode can be set to STANDARD or EARLY. If you are not
using the analog output voltage of the BAM-1020, leave this set to STANDARD. See Section
8.2 for a description of the Cycle Modes.
Met One 500 series meteorological sensors have an Auto ID feature which allows the BAM to
automatically recognize the sensor and enter all of the setup parameters for any channel the
sensor is attached to. Each channel can also be manually configured by the user for other
sensors.
CH: This field selects the channel to be viewed or edited. Use the up/down arrow
keys to select the desired channel.
TYPE: This is the channel name. You can enter any desired name here by using the
arrow keys to scroll through the alphabet.
PREC: This is the precision field, which sets the decimal place for the Multiplier and
Offset parameters.
MULT: This is the slope multiplier. Any input on the channel is multiplied by this
amount. The M factor in Y=MX+B.
OFFSET: This is the slope offset value. Any input on the channel has this amount added
to or subtracted from it, after the multiplier is applied. The B factor in Y=MX+B.
FS VOLT: This is the full-scale voltage output of the sensor. The maximum voltage range
that can be supplied by the sensor. This value is usually going to be either
1.000 or 2.500 volts. 2.500 is the maximum setting for this field.
INV SLOPE: This setting allows the channel to recognize a sensor with an inverse slope.
This is always set to N (no) except for use with thermistor temperature sensors
with resistance-only outputs.
VECT/SCALAR: This value sets the averaging method. S (scalar) is used for all
measurements except wind direction, which uses V (vector).
MODE: This field is toggled by pressing the ID MODE soft-key. The value can be set to
either MANUAL or AUTO ID. In MANUAL mode, the user can enter their own
setup parameters for the channel. AUTO ID mode is used with 500 series
sensors, and must be selected in order for the unit to recognize the sensor
automatically. NOTE: Any manually set parameters are lost when changing to
AUTO ID mode.
RH Control: If YES is selected, the Smart Heater will be automatically turned on full
power whenever the humidity of the sample stream exceeds the RH Setpoint. When the RH
falls back below the set point, the heater turns down to a “low power” heat mode which
simulates the older style wrap-around heaters. If this is set to NO, The Smart Heater will stay
in low power mode and no further RH control will be performed.
RH Setpoint: The RH Setpoint can be a number from 10% to 99%. This is the relative
humidity level that will be maintained at the filter. Met One recommends setting this value at
35%, which is the equilibration value for FRM filters and is required when monitoring PM2.5.
Datalog RH: If YES is selected, the filter RH values will be logged on channel 4 of the
BAM-1020. Select YES if you do not have any external sensors attached to channel 4.
Delta-T Control: The BAM-1020 can compare the filter temperature to ambient and
calculate the difference (Delta-Temperature), if a BX-592 ambient temp sensor is attached to
channel 6. If YES is selected, the Smart Heater will be turned down to low power mode
whenever the Delta-T Setpoint is exceeded.
Volatile organic compounds (VOCs) may be vaporized if the Delta-T exceeds 5 degrees
Celsius. However, this is usually not a problem with the BAM-1020, because the spot of
particulate is sampled for less than an hour before being replaced. Met One does not
recommend using Delta-T control except in rare cases, as it overrides the RH control which
has a far greater effect on the concentration measurement. Note: Delta-T control must be set
to NO to maintain USEPA equivalent status for the data.
Delta-T Setpoint: This can be set from 1 to 99 degrees C. If the filter temperature exceeds
the ambient temperature by more than this amount, the Smart Heater will turn down to low
power mode, regardless of the RH level. An N error is logged in the data any time this
happens. Note: There is often a few degrees of Delta-T measured even if the heater is OFF,
due to mild heating effect of the unit itself. Frequent errors will be logged if the setpoint is too
low. Set this value to at least 8 or 10 degrees C if used.
Datalog Delta-T: If YES is selected, the Delta-T values will be logged on channel 5 of the
BAM-1020. Select YES if you do not have any external sensors attached to channel 5. Note:
The measured Delta-T may still be logged even if Delta-T control is set to NO. This is often a
useful parameter.
Problem: The analog output voltage and/or digital concentration reading are full-scale.
Cause/Solution: • The unit will force these values full-scale to indicate an error. Download the error
log. Whenever the unit is started or interrupted this will happen until the next hour.
Problem: The airflow rate is too low and won’t adjust up to 16.7 LPM.
Cause/Solution: • The gray plastic pump mufflers clog up after several months. Replace it or drill a
hole in the end of it for a temporary fix. The brass mufflers can often be cleaned.
• The pump may need to be rebuilt after about 2 years.
• Check the inlet and PM heads for obstructions.
Problem: The airflow is stuck at a particular rate, and will not change.
Cause/Solution: • The flow controller unit on some older units can become stuck. If your flow controller
does not have a small circuit board mounted directly on the motor, it needs to be
upgraded. Contact the Service dept.
• Set the flow rate in the SETUP > SAMPLE screen to 14.0 and 17.5 LPM, and turn
the pump on in the TEST > PUMP screen. The BAM should try to regulate to these
values. If the flow does not change, the flow controller is probably stuck. If the flow
regulates lower, but not higher than 16.7, the pump is probably worn out, or there is
a leak. Be sure to set the flow back to 16.7 LPM when done.
Problem: The nozzle gets stuck in the UP position, or won’t press down onto the tape fully.
Cause/Solution: • This is often caused by a misaligned inlet tube. Make sure it is straight up and
perpendicular to the top of the unit.
• The nozzle o-ring eventually breaks down and needs to be replaced. Contact Met
One for detailed instructions. BX-308 tool kit required.
• The brass nozzle bushings may have grit in them. Remove the nozzle and clean the
parts. Contact Met One for detailed instructions. BX-308 tool kit required.
• Lift the nozzle with your fingers and determine if it feels sticky or gritty.
Problem: The unit has flow leaks, even after cleaning the nozzle and vane.
Cause/Solution: • The nozzle may be sticking as described above.
• The inlet receiver o-rings may need to be replaced.
• The nozzle spring may need to be replaced.
• Check the internal flow system for split tubing.
Problem: The clock settings disappear when the unit is powered down.
Cause/Solution: • The lithium battery on the 3230 circuit board may need to be replaced after about 10
years. It is normal for the clock to drift as much as 1 minute per month.
Inlet Receiver
Nozzle O-ring
Nozzle Spring
Nozzle
BAM-1020 units set up to monitor PM2.5 need to have this value field verified (and adjusted if
necessary) every six months using the BX-302 Zero Filter Calibration Kit. The test corrects
the BKGD value to compensate for minor variations caused by local conditions such as
grounding and shelter characteristics. The test involves running the BAM in its normal
operating environment with a zero filter on the inlet for at least 72 hours. The new BKGD will
then be calculated and entered into the BAM. The test should not be performed during a
period of rapidly changing weather. A complete set of instructions for the test are included
with the BX-302 kit.
TEST MODE
SELECT EXIT
The TEST Menu
TIME: This shows the time that the count test is started.
COUNT: This is the total number of beta particles counted during the four-minute test.
You will see this count rapidly accruing during the test. Typical four-minute
count values are between 600,000 and 1,000,000 counts through clean filter
tape. The count total will be lower if the membrane is extended, which
simulated adding particulate to the tape. If the count total is less than 500,000
then the beta detector is probably burning out.
MEMBRN: Press this soft-key to extend the membrane between the source and detector.
GO: This soft-key starts the four minute count test. The counting will immediately
begin. After four minutes the counting will stop and wait for the operator to
initiate another cycle or EXIT.
NOZZLE: 7
Flow (STD): 16.7 LPM PUMP: ON
Flow (STD): The air flow rate, displayed in EPA Standard liters/minute.
MOVE NOZZLE: This soft-key will move the nozzle up or down. The test allows for checking
proper nozzle movement. Total elapsed time is about 5 seconds. If the
pump is ON this operation is disabled.
PUMP ON: This soft-key will turn ON the vacuum pump. The nozzle will automatically
be lowered if necessary.
X: This is the last number of windows moved. This number will be negative if
the last move was backwards.
FWD: This soft-key will move the filter tape forward amount of the FEED value.
BKWD: This soft-key will move the filter tape backward amount of the FEED value.
CALIBRATION MODE
REF MBRN: <
COUNT (Io): 634000
COUNT (I): 556234
CAL MASS M: 0.801 mg/m3
START STOP EXIT
REF MBRN: This indicates if the reference membrane is extended (>) or withdrawn
(<) from the beta particle path.
COUNT (I0): The total 4-minute beta count through the filter tape only, no membrane.
COUNT (I): The total 4-minute beta count through both the filter and the membrane.
START: This soft-key starts the test cycle. Counting will immediately begin. After
4-minutes the I0 count will stop, the membrane will extend, and the I count will begin. At the
completion of the test the counting will stop and the mass of the membrane will be calculated.
The total elapsed time is about 8.1 minutes per test.
TEST MODE
NOZZLE SHUTTLE IDLER CAPSTAN
LATCH REF
SELECT EXIT
NOZZLE: This screen tests the two nozzle photo sensors. Use the UP and DOWN soft-
keys to move the nozzle, and monitor the status of the S4 and S5 photo sensors on the
screen.
SHUTTLE: This screen tests the photo sensor which monitors the position of the shuttle
beam (the two tape rollers that move together). The status of photo sensor S7 should only
IDLER: This screen tests the photo sensor which monitors the position of the right-side
tape tensioner (the spring-loaded tape roller on the right). When the tensioner is in its left
position under its spring tension, both photo sensors S6 and S1 should be OFF. If the
tensioner is moved to the middle of its travel, photo sensor S1 should be ON and S6 OFF.
When the tensioner is at the rightmost position, S1 and S6 should both be ON. The tensioner
must be moved by hand. The left tensioner assembly has no sensors to check.
CAPSTAN: This screen tests the two photo sensors which watch the rotation of the
Capstan shaft. This is the shaft under the rubber pinch rollers which drives the filter tape
forwards and backwards. Press the ADVANCE soft-key to rotate the Capstan counter-
clockwise, and the BACKUP soft-key to rotate clockwise. The shaft should rotate one-half of
a rotation each time. Photo sensor S8 should turn ON to stop the shaft at each half-turn, and
will be OFF while the shaft to turning. It is helpful to put an ink mark on the end of the shaft to
view the rotation.
LATCH: This screen shows the status of the pinch roller latch. If the rollers are latched
in the UP position, then S9 should be ON. S9 should turn OFF if the latch is unhooked.
REF: This screen tests the two photo sensors which monitor the position of the
reference membrane assembly. When the EXTEND soft-key is pressed the membrane
should extend and the S2 photo sensor should be ON, and S3 OFF. When the WITHDRAW
soft-key is pressed the membrane should withdraw and the S2 photo sensor should be OFF
and S3 ON. It takes a few seconds for the membrane to move.
BAM: 26.1 C
REFERENCE: 26.1 C
SWITCH ON OFF
SW1 0-10 vdc 0-1 vdc
SW2 4-20mA 0-16mA
SW3 Not used Not used
SW4 Not used Not used
In some cases the analog output is the only channel available between the BAM-1020 and
the datalogger, and any errors generated by the BAM must be reported using the same
voltage signal. The BAM-1020 will set the analog output to its full-scale reading when any of
the selected errors occur. The errors which cause this are selectable, and are described in
section 6.5. The external datalogger should be programmed to recognize a full-scale reading
as an error, not a valid concentration. This scheme is used because it is very rare for a
concentration reading to exceed the range of the BAM-1020. The data values collected by
the BAM are unchanged and may be viewed with the display or by downloading.
Note: A contact-closure input to the BAM-1020 is achieved by shorting the two terminals
on that particular input together, usually with a relay on the external datalogger. The
datalogger should not apply any voltage or current to the terminals. Contact-closure
outputs from the BAM-1020 are provided by the unit shorting the two terminals together with
an internal relay, without applying any voltage or current to them. The external datalogger
must then sense the closure. Normally-Open means that the relay contacts are not shorted
together unless a certain condition occurs, while Normally-Closed means that the relay
contacts are shorted until the condition occurs, then they open.
10
Analog
Output
11 Selection
Switches
12
13
14
1. TELEM FAULT N.V. Telemetry Fault Non-Voltage. This input can be used to
signal the BAM-1020 that the external telemetry system (datalogger) is not
operational. This is a contact-closure input which must be activated for a minimum of
2-seconds. If activated, the BAM will continue to function and will log a “U” error (see
section 6.5) and activate the DATA ERROR relay output. This input can be set to
normally-open or normally-closed in the SETUP > INTERFACE menu.
2. EXT RESET N.V. External Reset Non-Voltage. This input is can be used to
synchronize the BAM-1020 clock to the external datalogger, and is often used in
EARLY cycle mode (see section 8.2). This is a contact-closure input which must be
activated for a minimum of 2-seconds. The input can be set to normally-open or
normally-closed in the SETUP > INTERFACE menu.
MicroMet Plus: A powerful and comprehensive data logging program for meteorological
applications which is configurable to collect and manage all data from the BAM-1020.
Comet: A simple and easy to use communications terminal program which can retrieve
data from Met One data loggers, including the BAM-1020. This program replaces the old TUS
(Terminal Utility Software) program.
1. Connect the RS-232 port on the back of the BAM to your computer or laptop using the
appropriate cable. Connect to the Com1 serial port if available.
2. Open HyperTerminal. (Usually located in the Programs\Accessories\Communications
directory). The program will ask you to type a name for the connection. Type “BAM-
1020” or a name of your choice, then click “OK”.
3. The “Connect To” window will open. Select COM1 (or another port if used) from the
drop-down menu in the “Connect using:” field. Click “OK”. Note: You could also set up
the program to dial the BAM through a modem in this window.
4. The “COM1 Properties” window will open. Set the following values in the drop-down
menus, then click “Apply” and “OK”.
Bits per second: 9600
Data bits: 8
Parity: None
Stop bits: 1
Flow control: None
5. The main HyperTerminal connection window should now be open. Press the ENTER
key a couple of times. The window should respond with an asterisk (*) indicating that
the program has established communication with the BAM-1020.
6. Once communication is established, press the h key. This should cause the BAM-
1020 System Menu to appear on the window as shown below. You can now send any
of the characters in the menu to retrieve the desired files. The menu options are
described in the following sections.
7. HyperTerminal will only display 100 lines of data in the window. To capture larger files
(such as All Data), first select Transfer > Capture Text from the drop-down menu.
Select a location for the file, then click the “Start” button. Retrieve the desired files, and
HyperTerminal will automatically store them to the text file.
8. When you exit HyperTerminal, it will ask if you want to save your connection. Click
“Yes” and a file named BAM-1020.ht will be created in the HyperTerminal folder, which
will have all of the settings saved. Use this for future communications with the BAM.
The next column is the concentration. Notice that the values went full-scale during the two
hours that errors were logged. This indicates invalid data. The Qtot column is total flow
volume for the hour. With a flow rate of 16.67 LPM, and a sample time of 50 minutes, this
value will be about .834 m3 per hour. (16.7 * 50 / 1000 = .8335). The remaining six columns
are the six datalogger inputs on the BAM. In this example RH was logged on channel 4, and
Ambient Temperature was logged on channel 6. The other four channels had nothing
attached, but will appear in the array anyway. (The data on those channels is only noise.)
Channel 01 02 03 04 05 06
Sensor Conc Qtot WS no WS RH WS AT
Units mg/m3 m3 KPH V MPS % KPH C
=====================================================
01:00 ------------ 0.010 0.834 019.6 0.012 000.3 00017 132.2 008.7
02:00 ------------ 0.009 0.834 019.9 0.012 000.3 00018 132.1 007.4
03:00 ------------ 0.011 0.834 019.8 0.012 000.3 00018 132.1 006.5
04:00 ------------ 0.011 0.833 020.0 0.012 000.3 00018 132.1 006.1
05:00 ------------ 0.012 0.833 019.8 0.012 000.3 00018 132.1 005.3
06:00 ------------ 0.011 0.834 020.1 0.012 000.3 00018 132.0 005.6
07:00 ----L------ 0.995 0.000 020.3 0.012 000.3 00018 132.0 007.4
08:00 ---M------ 0.995 0.000 019.8 0.012 000.3 00017 132.1 009.4
09:00 ------------ 0.008 0.833 019.9 0.012 000.3 00015 132.2 012.5
10:00 ------------ 0.003 0.834 019.5 0.012 000.3 00014 132.2 016.2
11:00 ------------ 0.007 0.833 019.5 0.012 000.3 00013 132.2 019.7
12:00 ------------ 0.011 0.833 019.5 0.012 000.3 00012 132.0 020.7
13:00 ------------ 0.008 0.833 019.1 0.011 000.3 00010 132.0 021.9
14:00 ------------ 0.010 0.833 019.2 0.011 000.3 00010 131.9 022.3
15:00 ------------ 0.020 0.833 019.1 0.011 000.3 00011 132.0 020.9
16:00 ------------ 0.011 0.834 019.3 0.011 000.3 00012 132.1 018.7
17:00 ------------ 0.010 0.833 019.5 0.012 000.3 00012 132.2 017.9
18:00 ------------ 0.010 0.833 019.4 0.012 000.3 00012 132.1 017.1
19:00 ------------ 0.010 0.834 019.4 0.012 000.3 00014 132.2 015.3
20:00 ------------ 0.007 0.833 019.6 0.012 000.3 00015 132.1 014.4
21:00 ------------ 0.006 0.834 019.5 0.012 000.3 00017 132.1 013.3
22:00 ------------ 0.006 0.834 019.7 0.012 000.3 00021 132.0 011.2
23:00 ------------ 0.005 0.833 019.6 0.012 000.3 00023 132.0 010.0
00:00 ------------ 0.011 0.834 019.9 0.012 000.3 00017 132.2 009.5
Channel, 1, 2, 3, 4, 5, 6,
Sensor ID, 255, 255, 255, 255, 255, 255,
Channel ID, 255, 255, 255, 255, 255, 255,
Name, XXXXX, XXXXX, XXXXX, XXXXX, XXXXX, XXXXX,
Units, XXX, XXX, XXX, XXX, XXX, XXX,
Prec, 0, 0, 0, 0, 0, 0,
FS Volts, 1.000, 1.000, 1.000, 1.000, 1.000, 1.000,
Mult, 1.000, 1.000, 1.000, 1.000, 1.000, 1.000,
Offset, 0.000, 0.000, 0.000, 0.000, 0.000, 0.000,
Vect/Scalar, S, S, S, S, S, S,
Inv Slope, N, N, N, N, N, N,
Field Description
Start Start time of BAM sample period.
Elapsed Elapsed BAM sample time.
Flow Average flow rate for the BAM sample period.
CV Flow rate coefficient of variance for the BAM sample period.
Volume Sample volume for the BAM sample period.
Flag Flow regulation out of range warning flag.
AT Average ambient temperature for the BAM sample period.
AT Min Minimum ambient temperature for the BAM sample period.
AT Max Maximum ambient temperature for the BAM sample period.
BP Average ambient pressure for the BAM sample period.
AT Min Minimum ambient pressure for the BAM sample period.
AT Max Maximum ambient pressure for the BAM sample period.
Field Description
Time Event time stamp in seconds since January 1, 1970 00:00:00
Flow Minute average flow rate for the BAM sample period.
AT Minute average ambient temperature for the BAM sample period.
BP Minute average ambient pressure for the BAM sample period.
A new configuration has been added for the printer port which enables it to output a fixed-
width concentration report at the end of the sample period, which can be used to interface to
a serial data logger. This output is enabled by using the “z” utility command through the
serial port. The output format is date, time, concentration, and flow volume as shown below.
If the BAM is set to STANDARD cycle mode, the output will occur at the top of the next hour.
For example, if a measurement is made over hour 2, then the format would be:
If the BAM is set to EARLY cycle mode, the output will occur at minute 55:00 for the current
hour. For example, if a measurement is made over hour 2, then the format would be:
1. Download and save the data and error logs before proceeding. These will be cleared
during the upgrade process.
2. Record the OFFSET value from the SETUP > SAMPLE screen, and the BKGD value
from the SETUP > CALIBRATE screen.
3. Update the firmware.
4. The baud rate will default to 38400. Reset as appropriate.
5. Recalibrate the filter temperature and filter RH sensors.
6. Set the values of OFFSET, CONC UNITS, and COUNT TIME in the SETUP >
SAMPLE screen.
7. Set the values of CONC TYPE, FLOW TYPE, and BKGD in the SETUP > CALIBRATE
screen.
Tools
BAM-1020 Service Tool Kit: (nozzle shims, spring BX-308
2
scales, reel spacer, filter sensor removal tool)
Leak Test Rubber Nozzle Seal 7440 2
Mass Flow Calibration Kit, 0-20 SLPM BX-303
Flow Inlet Adapter Kit (Leak Test Valve) BX-305
2
Inlet Components
PM10 Inlet Head BX-802
2
Meteorological Sensors
590 Wind Direction Sensor BX-590
591 Wind Speed Sensor BX-591
592 Ambient Temperature Sensor BX-592
Many BAM-1020 units are equipped with at least the ambient temperature sensor BX-592
because ambient temperature is required on channel six for actual flow control. If the BAM is
used for PM2.5 monitoring, then the BX-596 sensor is required instead. This is a combination
ambient temperature and barometric pressure sensor which attaches to channels six and
seven, and provides the EPA required AT/BP measurements for actual flow control and flow
statistics.
The scaling and setups values of the series 500 Sensors are provided in the chart below. The
unit should automatically set these values in the SETUP > SENSORS menu. The ID MODE
must be set to AUTO to identify the sensors, or set to MANUAL to change the parameters.
SIG
CHAN X
COM
POWER
ID
Notes:
• BX-592 is almost always connected to channel 6 when used with a BAM-1020.
• BX-592 or BX-596 is required for actual flow control.
• BX-596 is required for PM2.5 monitoring, effective March 2007.
The number of beta particles passing through absorbing matter, such as dust deposited on a
filter tape, decrease nearly exponentially with the mass through which they much pass.
Equation 1 shows this relationship.
Equation 1
I = I 0e−µx
In Equation 1, I is the measured beta ray intensity (counts per unit time), of the attenuated
beta ray (dust laden filter tape), I0 is the measured beta ray intensity of the un-attenuated
beta ray (clean filter tape), µ is the absorption cross section of the material absorbing the beta
rays (cm2/g), and x is the mass density of the absorbing matter (g/cm2).
Equation 1 very closely resembles the Lambert-Beers Law, which is used in spectrometric
analysis. Just as the Lambert-Beers Law is an idealization of what is actually observed,
Equation 1 is also an idealized simplification of the true processes occurring meant to simplify
the corresponding mathematics. However, experimental measurement shows that in properly
designed monitors, such as the BAM-1020, the use of this equation introduces no substantial
error.
Equation 1 may be rearranged to solve for x, the mass density of the absorbing matter. This
is shown in Equation 2.
Equation 2
1 ⎡ I ⎤ 1 ⎡ I0 ⎤
− ln ⎢ ⎥ = ln =x
µ ⎣ I 0 ⎦ µ ⎢⎣ I ⎥⎦
In practice, the absorption cross section is experimentally determined during the calibration
process. Once I and I0 are experimentally measured, it is a simple matter to calculate x, the
predicted mass density.
In practice, ambient air is sampled at a constant flow rate (Q) for a specified time ∆t. This
sampled air is passed through a filter of surface area A. Once x, the mass density of collected
particles, has been determined, it is possible to calculate the ambient concentration of
particulate matter (µg/m3) with Equation 3.
In Equation 3, c is the ambient particulate concentration (µg/m3), A is the cross sectional area
on the tape over which dust is being deposited (cm2), Q is the rate at which particulate matter
is being collected on the filter tape (liters/minute), and ∆t is the sampling time (minutes).
Combining these equations yields to the final expression for the ambient particulate
concentration in terms of measured quantities. This is shown in Equation 4.
Equation 4
⎛ µg ⎞ 10 6 A(cm 2 ) ⎛I ⎞
c⎜ 3 ⎟ = ln⎜ 0 ⎟
⎝m ⎠ ⎛ liter ⎞ ⎛ cm 2 ⎞ ⎝ I ⎠
Q⎜ ⎟ ∆t(min)µ ⎜⎜ ⎟⎟
⎝ min ⎠ ⎝ g ⎠
The key to the success of the beta attenuation monitor is due in part to the fact that µ, the
absorption cross-section, is almost insensitive to the nature of the matter being measured.
This makes the BAM-1020 very insensitive to the chemical composition of the material being
collected.
σc
2 2
σ 2A σ Q σ 2t σ µ σ 2I σ 2I0
= + + + + −
c A2 Q2 t 2 µ 2 2 2
I ln ⎡ I ⎤ I 0 ln ⎡ I ⎤
2 2
⎢⎣ I 0 ⎥⎦ ⎢⎣ I 0 ⎥⎦
Inspection of Equation 5 reveals several things. The relative uncertainty of the measurement
(σc/c) is decreased (improved) by increasing the cross sectional area of the filter tape (A), the
flow rate (Q), the sampling time (t), the absorption cross-section (µ), I and I0.
In practice, the uncertainty associated with the filter area (σA/A), may be minimized by
ensuring that the tape is in exactly the same position during the I0 measurement as in the I
measurement phase. Careful design of the shuttle and tape control mechanisms inside of the
BAM-1020 results in minimal error here.
The uncertainty in the flow rate (σQ/Q) may be minimized by properly controlling the flow of
the instrument. For BAM-1020 units with a manual flow valve, this value is on the order of ±
3%. For BAM-1020 units equipped with the mass flow controller device, (σQ/Q) decreases to
± 1%.
The uncertainty associated with the measurement of I and I0 has to do with the physical
nature of the process leading to the emission of beta particles from the decay of 14C. This
process follows Poisson statistics. Poisson statistics show the uncertainty in the
measurement of I (σI/I) and I0 (σI0/I0) are minimized by increasing the sampling time.
Mathematical analysis shows that doubling the sampling time and hence the measured
intensity of I or I0 will reduce the uncertainty of the measurement by a factor of 1.41 (square
root of 2).
Equation 6
Cstd = Camb * (Pamb / Pstd) * (Tstd / Tamb)
Equation 6 can be used to calculate the standard concentration (Cstd) from the ambient
concentration (Camb) data using ambient barometric pressure and temperature data (Pamb and
Tamb) from the same time period in which the ambient concentration was recorded. Pstd and
Tstd are the values of standard barometric pressure and standard ambient temperature.
These values are usually the EPA mandated 760 mmHg and 298 degrees Kelvin (25 C). Note:
Some other countries use different values for standard temperature and pressure.
Equation 7
Camb = Cstd * (Pstd / Pamb) * (Tamb / Tstd)
Equation 7 can be used to calculate the ambient concentration (Camb) from the standard
concentration (Cstd) data using the ambient temperature and pressure. It is necessary to have
access to valid data for the ambient temperature and pressure for the desired sample hour in
order to be able to make the calculations.
Example: You have a data value of 27µg from a BAM which was configured to report data in
EPA Standard conditions (298K and 760 mmHg), but you need to know what the concentration
would have been in actual conditions. The actual average temperature for the hour in
question was 303K and the average pressure was 720mmHg.
Flow Audits
Flow Reference Standard Used: Model: Serial No: Calibration Date:
Temperature Standard Used: Model: Serial No: Calibration Date:
Barometric Pressure Standard Used: Model: Serial No: Calibration Date:
Leak Check Value: as found: lpm as left: lpm
BAM Ref. Std. BAM Ref. Std.
Ambient Temperature: as found: C C as left: C C N/A
Barometric Pressure: as found: mmHg mmHg as left: mmHg mmHg
Flow Rate (Actual Volumetric): as found: lpm lpm as left: lpm lpm N/A
Flow Rate (EPA Standard): as found: lpm lpm as left: lpm lpm N/A
Mechanical Audits
Pump muffler unclogged: as found as left PM10 particle trap clean: as found as left N/A
Sample nozzle clean: as found as left PM10 drip jar empty: as found as left N/A
Tape support vane clean: as found as left PM10 bug screen clear: as found as left N/A
Capstan shaft clean: as found as left PM2.5 particle trap clean: as found as left N/A
Rubber pinch rollers clean: as found as left Inlet tube water-tight seal OK: as found as left
Chassis ground wire installed: as found as left Inlet tube perpendicular to BAM: as found as left
Analog Voltage Output Audit N/A Membrane Audit Flow Control Range
DAC Test Screen BAM Voltage Output Logger Voltage Input LAST m (mg): Flow Setpoint BAM Flow
0.000 Volts Volts Volts ABS (mg): 14.0 LPM
0.500 Volts Volts Volts Difference (mg): 16.7 LPM
1.000 Volts Volts Volts % Difference: 17.5 LPM
Audit Notes: