4Sight
Service Manual
P/Ns : 24-8000
Summary Description
The 4sight device is designed to combine the functionalities of Accutome’s A-scan Plus (K123349),
B-scan Plus (K070943) and Accupach Pachymeter (K042752) devices into a combined, stand-alone
platform.
The device performs axial length measurements, corneal thickness measurements and images the
internal structure of the eye using these 3 modalities.
The 4Sight is a stand-alone system which runs on a Windows 8 platform consisting of a 4Sight
control unit, transducers, footswitch, and keyboard
Page 2 of 5
Software Context
The following system architecture diagram shows the architecture in the context of the 4Sight
system. The interaction between the complete 4Sight system and the larger environment is
beyond the scope of this section. The 4Sight application is written in Microsoft C# .NET version 4.0.
The WinUSB® device driver is used to connect USB based devices, such as the B-Scan device, with
the Windows OS.
Microsoft 4.0
Software Context Diagram
Mass Storage
Page 3 of 5
Troubleshooting Guide
Below are things to check for various Accutome 4Sight problems. If these suggestions
do not cure the problem, return the unit to Accutome, Inc for repair. Alteration, repair or
modification of any product that is performed by persons not authorized by Accutome,
Inc., will result in immediate loss of warranty.
Troubleshooting
Troubleshooting Guide
Message Cause
Error loading configuration file File may be corrupted. Contact Accutome for assistance.
“Can’t find the right type of probe for this Make sure you are using the correct probe for the modality
exam” selected. Re-check USB connection. Contact Accutome if
problem persists.
Error deleting snapshots Make sure snapshot files are not in use by another program or
open. Close if open and re-try.
Entered value out of range/ invalid Re-check values being entered to make sure they are correct.
Failed to initialize DICOM There may be an issue with the DICOM connection. Re-try
connecting the device to the server. If problem persists contact
Accutome support.
Could not locate the folder which Wrong file path may be specified in setup or folder may have
contains…. moved. Check to make sure folder is in the right location and make
sure file path is correct.
Failed to print to specified printer There may be an issue with the printer connection. Re-try
connecting printer to the control unit. Contact Accutome if
problem persists.
Page 4 of 5
Before returning instruments for service or repair, contact the Accutome Technical
service Group for a Return Authorization (RA) number.
After receiving authorization, print the RA number on the outside of the package and
send the instrument to:
Replacement Parts
The table below lists user replaceable items that are available from Accutome, Inc.
Please be sure to use the Accutome part number for the item when placing an order.
Standard Parts
Control Unit 24-8000
4Sight A-scan Probe 24-8000A
4Sight B-scan Probe 24-8000B
4Sight Pachymeter Probe 24-8000P
Footswitch 8000-0260
User Manual 24-8016
Keyboard 8000-0270
Mouse 6100-0020 6100-0020