Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                
0% found this document useful (0 votes)
140 views

Ultrasonic Examination Procedure Megha Engineering & Infrastructures Limited

This document outlines the procedure for conducting ultrasonic examinations using manual A-scan pulse echo contact technique. It details responsibilities, personnel qualifications, required equipment including search units and calibration blocks, surface preparation, couplant use, calibration techniques, examination parameters, evaluation standards, and reporting requirements. Calibration must be performed using reference blocks of the same material and thickness as the components under examination. Distance-amplitude correction curves are established to compensate for sound attenuation over the examination range. Personnel must be properly certified and equipment regularly calibrated and validated.

Uploaded by

Murali Y
Copyright
© © All Rights Reserved
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
140 views

Ultrasonic Examination Procedure Megha Engineering & Infrastructures Limited

This document outlines the procedure for conducting ultrasonic examinations using manual A-scan pulse echo contact technique. It details responsibilities, personnel qualifications, required equipment including search units and calibration blocks, surface preparation, couplant use, calibration techniques, examination parameters, evaluation standards, and reporting requirements. Calibration must be performed using reference blocks of the same material and thickness as the components under examination. Distance-amplitude correction curves are established to compensate for sound attenuation over the examination range. Personnel must be properly certified and equipment regularly calibrated and validated.

Uploaded by

Murali Y
Copyright
© © All Rights Reserved
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 10

Page 1 of 10

Doc. No. MEIL/UT/01.REV 0


ULTRASONIC EXAMINATION Page: 10 of 10
Date: 20.02.2022
PROCEDURE

MEGHA ENGINEERING & INFRASTRUCTURES LIMITED

ULTRASONIC EXAMINATION
PROCEDURE

PREPARED REVIEWED BY CERTIFIED BY

.
Page 2 of 10

Contents

1. SCOPE 3

2. REFERENCE DOCUMENTS 3

3. RESPONSIBILITIES 3

4. QUALIFICATION OF PERSONNEL 3

5. EQUIPMENT: 3

6. SEARCH UNITS: 4

7. SURFACECONDITION: 4

8. COUPLANT: 4

9. CALIBRATION OF EQUIPMENT: 5

10. CALIBRATIONTECHNIQUES: 6

11. EXAMINATION SURFACE: 7

12. SCANNINGSENSITIVITY: 8

13. EXAMINATIONCOVERAGE: 8

14. CALIBRATION CONFIRMATION: 8

15. EVALUATION: 9

16. ACCEPTANCESTANDARD 9

17. POST EXAMINATION CLEANING: 9

18. REPAIR & RETEST: 9

19. REPORT: 9

20. REPORT: 9

.
Page 3 of 10

1. SCOPE

 This procedure explains the requirements for manual A-scan, pulse echo,
contact technique of ultrasonic examination of weldments and materials of
carbon and low alloy steel in pressure vessels.
2. REFERENCE DOCUMENTS

 ASME Boiler and Pressure Vessel Code Section V, Edition 2010.


 ASME Boiler and Pressure Vessel Code Section VIII Division-1, Edition 2010.
 ASME Boiler and Pressure Vessel Code Section VIII Division-2, Edition 2010.
 ASME Boiler and Pressure Vessel Code Section IX Edition 2010.
3. RESPONSIBILITIES

 The NDT Inspector shall be responsible for conducting and reporting the
results of inspection in accordance with project contract.
4. QUALIFICATION OF PERSONNEL

 The personnel involved in Ultrasonic examination, interpretation and


evaluation shall be certified as per ASNT-SNT-TC-1A and written
practice for NDE Personnel Qualification.
 When the written practice is revised, the certification of NDE personnel
remains valid to the requirements of the previous revision until the expiry
date of the personnel qualification certificate; then, recertification to the
requirements of the new revision is required.
5. EQUIPMENT:

 The examination shall be conducted with A scan type pulse echo


ultrasonic equipment that generates frequencies over the range of 1MHz
to 5MHz and shall be equipped with a stepped gain control in units of 2
dB or less.
 The reject control shall be in the ‘off’ position for all examination.
 The equipment shall have valid calibration and stickered mentioning the date
of expiry.

.
Page 4 of 10

6. SEARCH UNITS:

 Search units may contain single crystal/dual crystal transducers.


 For straight beam examination longitudinal wave probes (0°) shall be used.
For angle beam examination shear wave probes of 45° shall generally be
used and other angles up to 70° shall be used appropriate to the configuration
being examined. Groove angle and thickness shall be considered in selecting
the probe angle other than 45°
 For straight beam examination, the crystal dimensions shall be selected in the
range of 5 mm to 25 mm. For angle beam examination, the crystal
dimensions shall be selected appropriate to the contact available with
examination surface and will generally be 8x9 mm or specified
 Probes of frequencies between 2.5 MHz and 4 MHz shall be used for
both straight beam and angle beam examination.

7. SURFACE CONDITION:

 The finished contact surface shall be free from weld spatter and surface
irregularities which will impair the free movement of the search unit.
 The weld reinforcement shall be ground flush smooth with adjacent
base material, if required

8. COUPLANT:

 Grease/oil may be used as couplant. The couplant used for the


examination shall be the same as that used for calibration.
 Couplant used on nickel base alloys shall not contain more than 250 ppm
of sulfur.
 Couplants used on austenitic stainless steel or titanium shall not contain
more than 250 ppm of halides (Chlorides plus fluorides).

.
Page 5 of 10

9. CALIBRATION OF EQUIPMENT:

 The basic reference reflectors shall be used to establish a primary reference


response of the equipment. The material from which the block is fabricated
shall be of the same product form and material specifications or equivalent P
- Number grouping (P. Nos. 1, 3, 4 and 5 , 5A through 5C and 15A through
15F are considered equivalent) as one of the material being examined.
 Reference blocks for dissimilar metal welds, the material selection shall be
based on the material on the side of the weld from which the examination
will be conducted. If the examination will be conducted from both sides,
calibration reflectors shall be provided in both materials.
 Reference blocks shall receive Precast the minimum tempering
treatment required by the material specification for the type and
grade.

 When two or more base material thicknesses are involved, the reference
block thickness shall be determined by the average thickness of the weld.
 The thickness and the location and size of holes in the basic reference block
to be used shall be as indicated.
 The finish on the surface of the reference block shall be representative of
the surface finish on the component. Variation in component
configuration and surface finish, require transfer correction to be
calculated.
 The reference block material shall be completely examined with a straight
beam search unit. Areas that contain an indication exceeding the
remaining back reflection shall not be used as block material.
 When the component has flat surfaces or curved surface with a diameter
greater than 20 inches, a flat basic reference block shall be used.
 When the component is curved and has a diameter less than 20 inches, a
curved basic reference block shall be used. The components to be tested
can have a curvature between 0.9 to 1.5 times of basic reference block
diameter.

.
Page 6 of 10

 For testing a weld thickness‘t’, the calibration block of the same thickness
will be ideal. However, the thickness of the calibration blocks acceptable
for use shall be
For weld thickness 25mm or less - 19mm
For weld thickness 25mm through 50mm - 38mm
For weld thickness 50mm through 100mm - 75mm
For weld thickness over 100mm - t±25mm
 The reference blocks manufactured in house, shall be inspected and certified .
10. CALIBRATION TECHNIQUES:

Technique for Straight Beam Calibration:

 The sweep range for straight beam is set on the equipment using the IIW
V1 block.
Distance - Amplitude Correction:
 The probe is positioned for maximum response from the hole that gives
the highest amplitude and set to 80% (±5%) full screen height.
 The peak of the indication is marked on the screen with a glass marking
pencil. Without disturbing the sensitivity controls of the unit, the probe is
positioned for maximum response from the other two hole indications and
their peaks marked on the screen.
 The screen marks are connected and extended through the thickness to
provide the distance amplitude correction (DAC) curve using flat bottom
holes.
Technique for Angle Beam Calibration:
 The basic calibration block shall be used for distance amplitude correction
and position calibration.
 The sweep range is selected and the equipment is calibrated to the
selected range using IIW - V1 or V2 blocks.
Distance - Amplitude Correction (DAC):
 To construct a DAC a basic calibration block having drilled holes at ¼T,
½T and ¾T shall be used. The gain is adjusted to provide an 80% ± 5% of
full screen indication from the ¼T hole.

.
Page 7 of 10

 The peak of the indication is marked on the screen with a glass marking
pencil. Without disturbing the settings, the probe is positioned for maximum
response from the ½T, ¾T holes and T+¼T hole after the beam has bounced
from the opposite surface. The peak marks of the indications from the side
drilled holes are connected to provide DAC curve.
 When calibrated using notched specimen, indication form the root notch
is brought to 80% of screen height and the sensitivity level is noted.
 A calibration check on Precast one of the basic reflections in the basic
calibration block shall be made to ensure proper functioning of the
equipment, at the beginning and finish of each examination or series of
similar examinations, every 4 hours during the examination and when
examination personnel are changed.
 The sensitivity (in dB) used for DAC shall be evaluation sensitivity or
reference level.
 Transfer correction if any obtained, shall be added to test /
evaluation sensitivity.
11. EXAMINATION SURFACE:

STRAIGHT BEAM EXAMINATION


 For straight beam examination the weld and base material shall be
scanned progressively along and across the weld surface to the extent
possible from at least one surface of the part. Each pass of the search unit
shall overlap a minimum of 15% of the transducer dimension clausellel
to the direction of scan indexing.
ANGLE BEAM SCANNING:

Defects parallel to the Weld:

 The angle beam shall be directed at 90° to the weld axis, from two directions
where possible. The probes shall be manipulated so that the ultrasonic energy
passes through the required volumes of the weld and adjacent base material.

.
Page 8 of 10

Defects Transverse to Weld:


 For detection of transverse defects the weld shall be examined using 45 deg.
angle beam probe.
 For butt joints the angle beam shall be directed essentially clausellel to the
weld axis. The probe shall be manipulated so that the angle beam passes
through required volumes of the weld and adjacent base metal. The probe
shall be rotated at 180° and examination shall be repeated.
12. SCANNING SENSITIVITY:

 Scanning sensitivity shall be a minimum of +6dB above the reference


level. However evaluation of indications shall be performed at the
primary reference level.
13. EXAMINATION COVERAGE:

 For the angle beam examination the entire volume of the weld inclusive
of adjacent base metal shall be examined by moving the probe over the
examination surface.
 During scanning, the probes shall be moved to and fro with swiveling
action of 10° to 15° on either side with overlap between successive scanning.

 Each pass of the probes shall overlap a minimum of 15% of the crystal
element dimension perpendicular to the direction of the scan.
 In any case the rate of probe movement for examination shall not
exceed 125mm per second.
14. CALIBRATION CONFIRMATION:

 A calibration check shall verify the sweep range calibration and distance
amplitude correction.
SWEEP RANGE CORRECTION:
 If a point on the DAC curve has moved on the sweep line more than 10% of
the sweep reading or 5% of full sweep whichever is greater, the sweep range
calibration shall be corrected. All recorded indications since the last valid
calibration or calibration check shall be re-examined with the corrected
calibration and the value changed.

.
Page 9 of 10

DAC CORRECTION:
 If a point on the DAC curve has changed 20% or 2dB of its amplitude, all
recorded indications since the last valid calibration check shall be re-
examined with the corrected calibration.
15. EVALUATION:

 Imperfections that cause an indication greater than 20% of the


reference level shall be investigated to the extent that the ultrasonic
examination personnel can determine their shape, identity, and location,
and evaluate them
 Discrimination of flaw indication from geometric indications shall be
carried out by manual finger damping.
 Sizing of the defect shall be carried out by 6 dB drop method.
16. ACCEPTANCE STANDARD

 ASME SEC. VIII DIV.1


 APPENDIX - 12 ASME SEC.VIIIDIV.2
 ASME SEC. IX
17. POST EXAMINATION CLEANING:

 After the examination is completed the couplant applied on the surface shall
be removed to the pre-inspection condition.
18. REPAIR & RETEST:

 Repaired weld regions shall be ultrasonically re-examined by the same


procedure used for original testing.
19. REPORT:

 Details of inspection of original weld and repaired area shall be recorded as


per Exhibit No: MEIL/IREP/UT/001
 Recording all reflections from uncorrected areas having responses that
exceed 50% of reference level shall be made.
20. CALIBRARTATION BLOCK:

.
Page 10 of 10

You might also like