Ultrasonic Examination Procedure Megha Engineering & Infrastructures Limited
Ultrasonic Examination Procedure Megha Engineering & Infrastructures Limited
ULTRASONIC EXAMINATION
PROCEDURE
.
Page 2 of 10
Contents
1. SCOPE 3
2. REFERENCE DOCUMENTS 3
3. RESPONSIBILITIES 3
4. QUALIFICATION OF PERSONNEL 3
5. EQUIPMENT: 3
6. SEARCH UNITS: 4
7. SURFACECONDITION: 4
8. COUPLANT: 4
9. CALIBRATION OF EQUIPMENT: 5
10. CALIBRATIONTECHNIQUES: 6
12. SCANNINGSENSITIVITY: 8
13. EXAMINATIONCOVERAGE: 8
15. EVALUATION: 9
16. ACCEPTANCESTANDARD 9
19. REPORT: 9
20. REPORT: 9
.
Page 3 of 10
1. SCOPE
This procedure explains the requirements for manual A-scan, pulse echo,
contact technique of ultrasonic examination of weldments and materials of
carbon and low alloy steel in pressure vessels.
2. REFERENCE DOCUMENTS
The NDT Inspector shall be responsible for conducting and reporting the
results of inspection in accordance with project contract.
4. QUALIFICATION OF PERSONNEL
.
Page 4 of 10
6. SEARCH UNITS:
7. SURFACE CONDITION:
The finished contact surface shall be free from weld spatter and surface
irregularities which will impair the free movement of the search unit.
The weld reinforcement shall be ground flush smooth with adjacent
base material, if required
8. COUPLANT:
.
Page 5 of 10
9. CALIBRATION OF EQUIPMENT:
When two or more base material thicknesses are involved, the reference
block thickness shall be determined by the average thickness of the weld.
The thickness and the location and size of holes in the basic reference block
to be used shall be as indicated.
The finish on the surface of the reference block shall be representative of
the surface finish on the component. Variation in component
configuration and surface finish, require transfer correction to be
calculated.
The reference block material shall be completely examined with a straight
beam search unit. Areas that contain an indication exceeding the
remaining back reflection shall not be used as block material.
When the component has flat surfaces or curved surface with a diameter
greater than 20 inches, a flat basic reference block shall be used.
When the component is curved and has a diameter less than 20 inches, a
curved basic reference block shall be used. The components to be tested
can have a curvature between 0.9 to 1.5 times of basic reference block
diameter.
.
Page 6 of 10
For testing a weld thickness‘t’, the calibration block of the same thickness
will be ideal. However, the thickness of the calibration blocks acceptable
for use shall be
For weld thickness 25mm or less - 19mm
For weld thickness 25mm through 50mm - 38mm
For weld thickness 50mm through 100mm - 75mm
For weld thickness over 100mm - t±25mm
The reference blocks manufactured in house, shall be inspected and certified .
10. CALIBRATION TECHNIQUES:
The sweep range for straight beam is set on the equipment using the IIW
V1 block.
Distance - Amplitude Correction:
The probe is positioned for maximum response from the hole that gives
the highest amplitude and set to 80% (±5%) full screen height.
The peak of the indication is marked on the screen with a glass marking
pencil. Without disturbing the sensitivity controls of the unit, the probe is
positioned for maximum response from the other two hole indications and
their peaks marked on the screen.
The screen marks are connected and extended through the thickness to
provide the distance amplitude correction (DAC) curve using flat bottom
holes.
Technique for Angle Beam Calibration:
The basic calibration block shall be used for distance amplitude correction
and position calibration.
The sweep range is selected and the equipment is calibrated to the
selected range using IIW - V1 or V2 blocks.
Distance - Amplitude Correction (DAC):
To construct a DAC a basic calibration block having drilled holes at ¼T,
½T and ¾T shall be used. The gain is adjusted to provide an 80% ± 5% of
full screen indication from the ¼T hole.
.
Page 7 of 10
The peak of the indication is marked on the screen with a glass marking
pencil. Without disturbing the settings, the probe is positioned for maximum
response from the ½T, ¾T holes and T+¼T hole after the beam has bounced
from the opposite surface. The peak marks of the indications from the side
drilled holes are connected to provide DAC curve.
When calibrated using notched specimen, indication form the root notch
is brought to 80% of screen height and the sensitivity level is noted.
A calibration check on Precast one of the basic reflections in the basic
calibration block shall be made to ensure proper functioning of the
equipment, at the beginning and finish of each examination or series of
similar examinations, every 4 hours during the examination and when
examination personnel are changed.
The sensitivity (in dB) used for DAC shall be evaluation sensitivity or
reference level.
Transfer correction if any obtained, shall be added to test /
evaluation sensitivity.
11. EXAMINATION SURFACE:
The angle beam shall be directed at 90° to the weld axis, from two directions
where possible. The probes shall be manipulated so that the ultrasonic energy
passes through the required volumes of the weld and adjacent base material.
.
Page 8 of 10
For the angle beam examination the entire volume of the weld inclusive
of adjacent base metal shall be examined by moving the probe over the
examination surface.
During scanning, the probes shall be moved to and fro with swiveling
action of 10° to 15° on either side with overlap between successive scanning.
Each pass of the probes shall overlap a minimum of 15% of the crystal
element dimension perpendicular to the direction of the scan.
In any case the rate of probe movement for examination shall not
exceed 125mm per second.
14. CALIBRATION CONFIRMATION:
A calibration check shall verify the sweep range calibration and distance
amplitude correction.
SWEEP RANGE CORRECTION:
If a point on the DAC curve has moved on the sweep line more than 10% of
the sweep reading or 5% of full sweep whichever is greater, the sweep range
calibration shall be corrected. All recorded indications since the last valid
calibration or calibration check shall be re-examined with the corrected
calibration and the value changed.
.
Page 9 of 10
DAC CORRECTION:
If a point on the DAC curve has changed 20% or 2dB of its amplitude, all
recorded indications since the last valid calibration check shall be re-
examined with the corrected calibration.
15. EVALUATION:
After the examination is completed the couplant applied on the surface shall
be removed to the pre-inspection condition.
18. REPAIR & RETEST:
.
Page 10 of 10