Selectomat - PL - User Manual - ENG
Selectomat - PL - User Manual - ENG
Selectomat - PL - User Manual - ENG
Steam sterilizer
Selectomat PL
General information
Observe the following sections in particular of the original user manual:
■ Intended use
■ Rules of conduct during operation
■ Standards and directives
The following hazard warnings and symbols may appear in this user
manual:
Handling instruction
■ List
Danger
List of abbreviations
Protective ground
Device labelling
The specification of this steam sterilizer is indicated by the following device labeling:
Machine name Nominal chamber size Number Variant Position of the chamber
HWD of doors
Selectomat PL 9 6 12 2 H R
series
The chamber size (9 x 6 x 12) is based on the 1-door See the R Chamber right
dimensions for a sterilization unit 2-door table below Panel left
(1 STU = 3 x 3 x 6 dm).
L Chamber left
Example:
Panel right
Chamber size 9 6 12 = 12 STU
Variant Version
Machines with raised loading (variant H) are loaded and unloaded with a transport trolley and a loading
trolley positioned on it.
Machines with raised loading (variant V) are loaded and unloaded with a transport trolley and a loading
trolley positioned on it.
Floor-level, passable machines (variant G) are loaded and unloaded using trolleys.
Machines with raised loading (variant CL) are loaded and unloaded with a transport trolley and a loading
trolley positioned on it.
* Standard dimensions only. The dimensions may vary for each project.
Contents
1 Safety.........................................................................................11
1.1 Intended use.............................................................................. 11
1.2 Responsibilities of the operator................................................. 12
1.2.1 Commissioning - equipment requirements................................ 13
1.2.1.1 Connection to the electrical power supply................................. 14
1.2.1.2 Blow off line................................................................................14
1.2.1.3 Media..........................................................................................15
1.2.2 Installation and operation of the machine.................................. 16
1.2.3 Maintaining efficiency................................................................. 17
1.2.3.1 Process validation...................................................................... 18
1.2.3.2 Routine inspections.................................................................... 18
1.2.4 Operation of pressure equipment.............................................. 19
1.2.5 Disposal......................................................................................20
1.3 Users.......................................................................................... 21
1.3.1 Approved operators....................................................................22
1.3.2 System supervisor......................................................................22
1.3.3 Specialist.................................................................................... 22
1.4 Rules of Conduct during Operation........................................... 23
1.4.1 Personal protective equipment.................................................. 23
1.4.2 Starting a program..................................................................... 23
1.4.3 Working inside the chamber...................................................... 23
1.4.4 Working on the assemblies........................................................24
1.4.5 Working in the electrical control cabinet.................................... 24
1.4.6 Maintenance work...................................................................... 24
1.4.7 Automatic transport system (option).......................................... 24
1.5 Machine safety devices..............................................................25
1.5.1 Safety strip................................................................................. 25
1.5.2 Automatic program cancellation.................................................25
1.5.3 Program cancellation button...................................................... 25
1.5.4 Program stop switch.................................................................. 26
1.5.5 Master switch............................................................................. 26
1.5.6 Safety valve................................................................................27
1.5.7 Locking the chamber door and door release............................. 27
1.5.8 Monitoring the sterilization parameters...................................... 27
1.5.9 Access protection....................................................................... 28
1.6 Residual risks and measures to prevent them...........................28
1.7 Maintenance and Safety inspection........................................... 30
1.8 Equipment log............................................................................ 31
2 Machine description.................................................................33
2.1 Standard equipment................................................................... 33
2.2 Optional additional features....................................................... 35
2.2.1 Optional programs......................................................................35
2.2.1.1 Special programs....................................................................... 35
2.2.1.2 APPcycles® (option).................................................................. 35
2.2.2 Additional monitoring equipment................................................35
2.2.3 Reducing the consumption of media......................................... 37
2.2.4 Optional services........................................................................38
2.2.5 Automatic transport system....................................................... 40
2.2.6 Uninterruptible power supply (UPS) (option)............................. 40
2.3 Accessories Transport trolley.....................................................40
2.4 User administration.................................................................... 42
2.5 Batch documentation................................................................. 43
2.6 Automatic program selection using the barcode (option)...........44
2.7 Process description.................................................................... 45
2.7.1 Sterilization programs................................................................ 46
2.7.2 Heating and SteamSpy heating................................................. 49
2.7.3 Test programs............................................................................ 49
2.7.3.1 Vacuum test................................................................................49
2.7.3.2 Bowie-Dick test.......................................................................... 50
2.7.3.3 Air detector function test (option)...............................................51
2.7.4 Special programs....................................................................... 51
2.7.4.1 Alloplast...................................................................................... 51
2.7.4.2 Disinfection (option)................................................................... 51
2.7.4.3 Rack hatch................................................................................. 51
4 Operation...................................................................................61
4.1 Logging users in and out........................................................... 61
5 Machine configuration............................................................. 79
5.1 Configuring user administration (option).................................... 79
5.2 Setting batch print and save options......................................... 81
5.3 Print settings.............................................................................. 82
5.4 Setting the network addresses of the control panel................... 83
5.5 Activating auto start (option)...................................................... 84
5.6 Activating standby mode............................................................85
5.7 Setting the time.......................................................................... 85
5.8 Activating the steam manager (option)...................................... 86
5.9 Activating the cooling circuit vacuum pump...............................86
5.10 Calibration screen...................................................................... 88
7 Maintenance............................................................................111
7.1 Preparatory measures..............................................................112
7.2 Service module (option)........................................................... 113
7.3 Maintenance activities.............................................................. 113
7.4 Exchanging the door seal........................................................ 117
7.5 Thermal printer paper exchange (option).................................117
7.6 Periodic inspections of pressure vessels................................. 118
7.7 Cleaning................................................................................... 120
7.7.1 Chamber interior...................................................................... 121
7.7.2 Assemblies compartment......................................................... 122
7.7.3 External care............................................................................ 122
7.7.4 Cleaning of transport trolleys................................................... 123
10 Index........................................................................................ 135
1 Safety
Read this user manual carefully and save it for future reference.
Your personal safety and the correct functioning of the machine are
at stake.
The machine and accessories must not be modified. This especially applies
to the safety mechanisms.
This user manual contains information on the proper use of the machine.
For reasons of clarify, this user manual only contains the necessary instruc-
tions for normal operation and routine maintenance, as well as a description
of the essential functions of the machine.
For more information about your machine, please contact the manufacturer.
• Laundry
Disinfection of
• Utensils and anesthesia supplies
During the service life of the product, the properties and performance char-
acteristics are not influenced in a way that could endanger the safety of
patients, users and, if applicable, third parties.
If you wish to continue using the machine beyond the above period, please
contact the manufacturer or an authorised service company for inspection
of the machine to assess the wear, technical status and the possibility of
repair.
Before commissioning
The operator must ensure that the following requirements are met:
■ the device has been properly assembled and installed
■ the inspection and maintenance plan is available
■ the system supervisors and users have been trained
■ the operating permit has been issued (if required by national regula-
tions)
■ the operator must have given approval
■ the standard operating procedure is available
■ The validation has been performed ( see Chapter 1.2.3.1, "Process
validation")
■ unused programmes have been locked.
Network security
The sterilizer must be connected to the supply lines (e.g. drinking water
network) and secured to prevent pressure overshoots according to the
manufacturer's planning drawing and the dimension drawingpiping diagram
and the local regulations.
Make sure that the voltage, current type and frequency specified on
the type plate correspond to those at the site.
When working on the machine, turn off the master switch and secure
it with a lock to prevent it from being switched back on unintention-
ally.
Danger
Ensure that the line is unobstructed at all times, and that the steam can be
blown off safely.
1.2.1.3 Media
The media, particularly the sterilization steam, must meet the requirements
specified in the planning drawing or the technical dimension drawing and
the piping diagram (e.g. max. permissible pressure, steam quality etc.) to
prevent damage due to excess pressure and corrosion. The media must be
regularly checked in this regard.
Observe all relevant national regulations when installing and operating the
machine.
Observe the German Occupational Safety Act, and the Ordinance on Indus-
trial Safety and Health (BetrSichV). In accordance with BetrSichV, a risk
assessment must be performed regarding the operation within the company
of an installation subject to monitoring ( Chapter 1.2.4, "Operation of pres-
sure equipment").
Sound emissions
The operator must take the noise emission of the system into account when
performing a risk assessment.
At delivery at the latest, the staff entrusted with operating the machine
must be trained for this task by a qualified person ( see Chapter 1.3.3,
"Specialist" e.g. MMM customer service) and be familiar with the user
manual.
The names of persons authorized to use the machine as well as the system
supervisor ( see Chapter 1.3.2, "System supervisor ") must be recorded
in the medical products log / equipment log. When changes in staff occur,
new users must be inducted using this user manual and also recorded in
the medical products / equipment log.
Danger
Validation requirements are given in e.g. DIN EN ISO 17665 and the
recommendations of the Commission for Hospital Hygiene and Infection
Prevention (Kommission für Krankenhaushygiene und Infektionsprävention
- KRINKO) on the hygiene recommendations when preparing medical prod-
ucts (Anforderungen an die Hygiene bei der Aufbereitung von Medizinpro-
dukten).
In the case of changes which affect the process, e.g. changes to pack-
aging, introduction of new, design-critical instruments or after maintenance
activities, it is essential to perform a revalidation for a specific reason. The
content of the inspections depends on the change.
Data on the classification of the pressure vessel can be found on the pres-
sure vessel drawing or the test report according to Annex I of the Pressure
Equipment Directive.
The German Ordinance on Industrial Safety and Health (BetrSichV) and the
Technical Rules for Operating Safety (TRBS) are applicable. The BetrSichV
regulates the intervals for regular tests and the requirements for the testers
(qualified person as per TRBS 1203 or accredited inspection body (ZÜS)).
1.2.5 Disposal
The operator is responsible for disposing of the machine and the acces-
sories in the environmentally correct and appropriate manner in accordance
with the applicable local regulations.
The machine must not be disposed of with normal domestic waste. Please
contact the manufacturer if you have any questions about how to dispose of
individual components in an environmentally correct way.
1.3 Users
The machine comes with a user administration feature that can be used to
configure the authorizations of individual operators.
Danger
The supervisor must confirm receipt of the above with his or her signature.
1.3.3 Specialist
The term specialist within this user manual refers to persons who have the
required expertise due to their training, experience and recent employment,
who perform the assigned work and are capable of independently identi-
fying potential hazards and avoiding dangers.
The metal walls of the sterilization chamber and door become so hot during
operation that touching them with a bare hand may cause burns.
Risk of burns!
Before the program is started, it must be ensured that the chamber has
been correctly loaded and that all chamber doors and service doors are
closed.
Only trained personnel may work inside the chamber (e.g. 1.3.1
"Approved operators", Page 22).
The master switch can also be shut off and locked with a padlock to prevent
unauthorized reactivation.
When working on the assemblies, turn off the master switch and
Danger secure it with a lock to prevent it from being switched back on. Keep
the key with you when carrying out work to prevent unintentional
unlocking.
Risk of injury!
The self-closing chamber doors are fitted with a safety strip that stops door
movement as soon as the door encounters an obstacle (e.g. if an object or
a person is in the way of the door). The door is then opened again.
In the event of a fault that prevents the program sequence from completely
correctly, the control system cancels the program, e.g. faults with the
messages Pressure transducer controls defective or Motor protection
switch vacuum pump.
On machines with two doors, only the door on the loading side can be
opened.
On machines with two doors, only the door on the loading side can be
opened.
The program stop switch is on the control panel and must be used in case
of imminent danger, or if irregularities or malfunctions occur during the
program.
The program stop switch stops all movement of the machine by hardware,
i.e. the door close movement is stopped or prevented (this is also a safety
measure for work in an accessible chamber) and a program that has been
started is stopped.
For machines with two doors, only the door on the loading side can be
opened.
The program stop switch can only be unlocked with the specific key
supplied ( Chapter 1.3.2, "System supervisor " and Chapter 6.1, "Unex-
pected cancellation of the program sequence").
Turn off the master switch if there is a hazard and when working on
the device (e.g. rectifying faults, performing maintenance, etc.).
The steam sterilizer is optionally fitted with the following safety valve:
■ Compressed air (option): For certain variants, the compressed air
supply line is protected by a safety valve. The safety valve safely
discharges the compressed air into the assemblies compartment of the
machine in the event the permissible pressure is exceeded.
The control system prevents the door from being opened during the
program sequence. Consequently, the door buttons are not shown on the
display.
Danger
On completion of the program, the chamber door can only be opened when
the following conditions are met:
■ The pressure in the chamber is equal to the atmospheric pressure
■ The door seal has been released
Preset authorization levels with restricted rights are available for daily oper-
ation. These can be individually changed using the option User adminis-
tration ( Chapter 2.4, "User administration").
Residual risks
Use sterilized material barrier systems to protect the sterile goods from
recontamination. Only sterile goods packaging must be used which
complies with the national EN or DIN standards.
Perform the routine testing ( Chapter 1.2.3.2, "Routine inspections") on
daily start-up of the sterilizer.
Perform the daily vacuum tests ( Chapter 2.7.3.1, "Vacuum test") to
identify any leaks in the chamber.
Perform the daily Bowie-Dick tests ( Chapter 2.7.3.2) to prevent
an impermissibly high fraction of inert gas in the steam. See also
Chapter 4.2, "Daily start-up".
Only those goods for which a process validation exists may be treated.
Warranty advice
This machine complies with the following directives, regulations and stan-
dards:
Directives:
■ Directive 93/42/EEC on medical products (until May 25, 2020)
■ Directive 2017/745/EU on medical products (until May 26, 2020)
■ Directive 2014/68/EU on pressure vessels
■ Directive 2011/65/EU on restricting the use of certain hazardous
substances in electrical and electronic equipment (RoHS)
2 Machine description
What follows is a general description of the machine and the processes.
For the precise specifications of your machine, please refer to the technical
data sheet or the planning drawings as well as the type plate attached to
the machine and the machine configuration of the technical documentation.
Closure
The safety devices prevent the doors from opening during the program, and
the simultaneous opening of the loading and unloading sides (in the case
of two-door devices). To prevent jamming, the doors are fitted with a safety
strip.
Each sliding door is sealed pressure-tight using a special blow-out seal with
steam .
Steam supply
The steam is supplied either via the internal steam network (FD) or an
assigned or integrated electric steam generator (ED).
Vacuum facility
The machine is equipped with a vacuum facility which enables air extraction
during special programs and program phases, fractionated vacuum drying,
safety extraction and door seal release.
The vacuum pump unit comprises a powerful and quiet double stage water
ring vacuum pump with an upstream heat exchanger and water circulation
container.
Electrical equipment
High degree of connectivity with Ethernet and USB ports. Batches saved
locally to flash memory. Up to 20 000 batches can be saved.
Disinfection (option)
Sterilization program for processing medical products that have been used
for operations on high-risk tissues.
Alloplast (option)
The sterilizer is equipped with an air detector (air detector) for detection of
air leakages.
This monitors compliance with the sterilization conditions, since the effect
of sterilization is significantly impaired in the presence of non-condensable
gases, which tend to form air pockets in porous materials such as .
Air detector monitoring is only active if the air detector is installed and acti-
vated for the current program.
Additional function for creating a defined leak when performing a test using
an air-detection device.
The MMM SteamSpy is an integrated physical B&D test, which has been
tested as per ISO11140-4 by an independent test laboratory.
The integration of the electronic B&D test in the Autostart option automat-
ically starts the vacuum test, the heating program and the B&D test under
program control, which means that the machine is fully ready for operation
at the start of the shift.
If the device has the SteamSpy option, the routine cycle control is run via
SteamSpy. The sterilization process conditions tested by MMM SteamSpy
conform to the test body conditions as per DIN EN 867-5.
This optional equipment monitors the on-site compressed air supply and is
triggered if the supply pressure is too low.
This optional equipment monitors the on-site cold water pressure and is
triggered if the water pressure of the cold water is too low.
To reduce the amount of cold water used to cool the vacuum pump, it
is possible to connect the machine to an external cooling circuit ( see
Chapter 5.9, "Activating the cooling circuit vacuum pump").
After a prompt, the sterilizer indicates that it is ready to start by the green
lamp and the display. Only then can the machine be loaded and started. It
must not be loaded beforehand, as the sterilization materials are otherwise
impermissibly pre-heated and dried.
In machines with automatic loading this is done only after the steam
manager has sent the start release.
Standby mode
If the machine is not operated for a long time (no program active, screen is
not touched), it switches to standby mode after an adjustable time. Standby
mode is activated automatically after the set time has elapsed ( see
Chapter 5.6, "Activating standby mode ").
Jacket control is then paused and the machine cools down (energy saving).
Standby mode switches off as soon as the touch screen is touched. As the
machine cools down in standby mode, we recommend first running a warm-
up program after a long down time.
Whether and at what time the auto start should commence can be individ-
ually configured for each weekday. For example, a heating program can be
set from Monday to Friday, while no automatic program start takes place at
the weekend ( Chapter 5.5, "Activating auto start (option)").
■ Calibration/adjustment
■ Revalidation
The machine can send e-mails and/or SMS to the operator or MMM
customer service. This provides immediate information on faults and main-
tenance information (see separate instructions).
Barcodes are used to identify the sterilization material. They are located on
the goods to be treated and are scanned individually with a handscanner
per batch.
A maximum of 500 barcodes can be entered per batch. The barcode infor-
mation is transferred to the batch record.
The required program can also be stored in the barcodes to enable the
program to be selected automatically ( see Chapter 2.6, "Automatic
program selection using the barcode (option)").
A user ID that is written to the batch record can be entered via the touch-
screen or the barcode scanner. This makes it possible to establish at a later
time who started the program.
To simplify plant management, the device data from one or more machines
is monitored and displayed online by a centrally located, on-site network-
compatible PC.
Access to the remote maintenance function is via a VNC client. This service
can be used to operate the machine, view parameters, check the audit trail,
view the IO list, and much more.
Two door machines can be equipped with a Cart Logistik System for auto-
matic loading and unloading of the chamber (see the Cart Logistik user
manual).
The transport system is controlled, and supplied with power and if applic-
able compressed air, by the associated sterilizer.
Uninterruptible power supply for the control system. In the event of a power
failure, the control system remains active and cycle information is saved.
Process-relevant valves are closed during the power failure.
Both variants are available in various sizes to suit the loading trolleys you
are using with your sterilizer.
Transport trolley (TW) with a fixed loading height of 390 mm, 625 mm or
850 mm.
Read the original user manual from the manufacturer of the TW HUB
carefully and save it for future reference.
Observe the safety instructions. Your personal safety and the correct
functioning of the machine are at stake.
The left button moves the platform upwards, the right button moves the
platform downwards. The middle button between the handles moves the
platform to a preset height.
Danger
Also observe the cleaning and maintenance instructions and the internal
work instructions!
If the screen is not touched for 5 minutes (individually adjustable) after log-
in, the "Supervisor" user is automatically logged out and the "Anonymous"
user is logged in.
The user will only be able to perform the allowable actions after logging in.
When the device is commissioned, an administrator (S) is set up, who will
have the right to create or delete additional users and to assign specific
rights to users ( Chapter 5.1, "Configuring user administration (option)").
If the machine has a control panel on the unloading side, the Extended
user administration function can also be used to specify that the names
of the users who started a program, and who opened the door at the end
of the program, are recorded with the batch and are printed on the batch
record.
■ Course of the air detector temperature over time (for air detector option
only)
■ Evaluation of the integrated Bowie&Dick test (for SteamSpy option)
■ Evaluation of the integrated cycle control (for SteamSpy option)
■ Measurement record in plain text
■ All barcode numbers on labels read by the barcode scanner
The batch record saved in the device can be printed out at the end of the
batch ( Chapter 5.2, "Setting batch print and save options") and sent to an
external server, depending on the device equipment.
If a record has not been added to the electronic archive, either the corre-
sponding printout must be retained or the batch in question discarded.
The last 20 batches stored on the device can be shown on the display and
reprinted.
■ The first barcode to be scanned determines the barcode group. All other
scanned barcodes must belong to the same group.
■ If barcodes are scanned that are assigned to different sterilization
temperatures, the lowest temperature is selected.
■ A newly scanned barcode is checked to see whether a program with
the same sterilization temperature but different sterilization time has
already been scanned. If this is the case, the new barcode is rejected.
This means that all programs with the same sterilization temperature
have to have the same sterilization time.
■ If barcodes are scanned that have been allocated the SuperDry (SD)
drying method, the SD program with the lowest previously selected
temperature is selected. If there is no SD program with this temperature,
the barcode is rejected.
■ If barcodes are scanned that have been assigned a program number for
a program that has not been activated, the barcode is rejected.
■ Heating program
■ 1 rack hatch program (for two-door machines)
The programs defined and set with the operator on commissioning are
displayed in the program selection menu. A maximum of 13 programs can
be activated simultaneously.
1. Pulsed evacuation of the chamber with interim steam surges for air
removal and steam penetration
2. Heating phase up to the sterilization setpoint pressure
3. Sterilization for a specific time at a specific temperature. The pressure
container's design pressure, the installed measurement technology and
the media supply allows sterilization temperatures from 105°C to 134°C
to be set. Sterilization time variably adjustable.
4. Chamber pressure relief
5. Drying of the material through evacuation of the chamber for a specific
time.
6. Chamber aeration to atmospheric pressure
7. Release of the door seal
Program selection:
Universal
Standard program
■ Sterilization temperature: 134°C
■ Exposure time: 5 min
■ Packed instruments
■ Materials that are difficult to ventilate
Universal SD
For goods that are difficult to dry (e.g. excessively heavy packaging), the
drying effect is improved by alternating between aeration and evacuation in
the Drying program phase.
■ Sterilization temperature: 134°C
■ Exposure time: 5 min
Plastics
■ Steam-resistant plastic
For the preparation of medical products that have been used for operations
on high-risk tissues and for which it was not possible to undergo the appro-
priate precleaning (see the hygiene recommendations when preparing
medical products (Anforderungen an die Hygiene bei der Aufbereitung von
Medizinprodukten) of the Commission for Hospital Hygiene and Infection
Prevention (Kommission für Krankenhaushygiene und Infektionsprävention
- KRINKO), 2012).
A program for heating the chamber and the optional SteamSpy equipment
(option) to the operating temperature.
On machines with two doors, only the door on the loading side can be
opened.
The "Vacuum test" program class does not involve a sterilization program.
The sequence is characterized by the following elements:
1. Evacuation of the
chamber
2. Waiting time 5 minutes
3. Test time 10 minutes
4. Chamber aeration to
atmospheric pressure
5. Release of the door seal
On failure of the vacuum test, the message Last vacuum test not passed is
shown and the red indicator lamp lights.
The test must be performed once per day pursuant to DIN 17665-1 at
operating temperature (e.g. following an empty sterilization or a heating
program). The following test methods are available:
■ Original test package comprising folded cotton cloths. The test is
performed and evaluated according to the manufacturer's instructions.
■ Alternative Bowie-Dick test in accordance with DIN EN ISO 11140-4,
e.g. using chemical indicators or SteamSpy. The test is performed and
evaluated according to the manufacturer's instructions.
When using the Bowie-Dick test with SteamSpy option, the SteamSpy
Heating program must be used.
The air detector function test enables one or more programs to be set up
to test the function of the air detector. During an air detector function test
program a defined leak is generated during the entire preliminary phase
when the chamber is at underpressure.
2.7.4.1 Alloplast
For redelivery of empty loading trolleys from the unloading to the loading
side through the machine chamber. Only one door can be opened at one
time.
This program can also be used for service purposes (e.g. for cleaning the
chamber).
1. Indicators
2. Chamber pressure
measurement device
3. Program stop
4. Door open
5. Door closed
For floor level loading machines with automatic unloading, the controls for
the automatic unloading are located on the control panel of the sterilizer.
Please refer to the separate user manual for your Cart Logistik
System.
1. Chamber pressure
measurement device
2. Indicators
3. Emergency stop
(automatic unloading
only)
4. Program stop
5. Door open
6. Door closed
7. Reset (automatic
unloading only)
8. Transport release
(automatic unloading
only)
EMERGENCY STOP
For floor level loading machines with automatic unloading, the EMER-
GENCY STOP for the transport mechanism is located on the machine
enclosure of the sterilizer.
Floor level loading machines with automatic unloading are equipped with an
“all clear” button.
This is located on the machine enclosure on the unloading side and lights
up when the machine is ready for unloading.
Reset button
Floor level loading machines with automatic unloading are equipped with a
reset button.
This is located on the machine enclosure on the unloading side, and lights
up when a fault arises or the EMERGENCY STOP switch is pressed.
Pressing the reset button returns the transport cylinder back to the starting
position. The conveying system will remain in fault mode until the reset is
performed.
Yellow indicator
Floor level loading machines with automatic unloading are equipped with an
indicator.
This is located in a well visible position in the unloading area, and blinks
when unloading is active.
Risk of injury!
It is located on the other side of the chamber door in relation to the control
panel. Before closing the chamber door, the user is therefore compelled to
look into the chamber and ensure that there is no one inside the chamber.
Function overview
9 Close door
Selection of saved Opens the list of the last 20 saved batch records.
batch records
Scroll button If a menu level covers more than one screen, use
this button to scroll through the screens.
Scroll back button If a menu level covers more than one screen, use
this button to scroll back through the screens.
4 Operation
Observe the safety instructions in Chapter 1, “Safety”.
User log-in
Press the log-in button.
Touch the "Name" entry field. The alphanumerical input block appears.
Enter the corresponding user name (e.g. "S" for supervisor).
Confirm the entry with the tick button.
Touch the "Password" entry field. The alphanumerical or numerical input
block appears.
Enter the corresponding password.
Confirm the entry with the tick button.
Press the OK button.
After log-in, the corresponding user name (e.g. "S") appears in the user bar.
The user can carry out any action for which he is authorised.
If a user does not log out actively, this occurs automatically within the set
time limit if no button on the touch screen has been pressed within this
period of time.
User log-out
Press the log-out button.
If the configured morning autostart cannot be run because the device was
switched off on the previous evening, or a door of the device was left open,
the autostart can be initiated manually.
Open the door on the loading side once the program has
finished.
The program selection menu then appears.
The chamber and jacket must be heated before a sterilization program can
be started.
Select the Heating or Heating SteamSpyprogram (if
installed) or another program suitable for heating the
chamber.
Open the door on the loading side once the program has
finished.
The Bowie-Dick test is performed once a day when the machine is warm.
Evaluate the test results and enter the results in the daily record.
See also Chapter 1.6, "Residual risks and measures to prevent them".
If scanning with the handscanner fails and the barcode does not appear
in the list (e.g. because the barcode is dirty), the barcodes can be entered
manually).
Press the Tool selection button.
Deleting barcodes
Load the chamber according to the process specified in the process valida-
tion. This is the only way to guarantee a satisfactory sterilization result.
Place heavy containers at the bottom or with neutral impact on the centre of
gravity.
Deviations are possible if they have been tested as part of a validation and
authorized.
Risk of injury!
Make sure that the loading trolley is firmly locked on the transport
Danger trolley.
Before loading the chamber, consider any waiting times that need to be
adhered to. A program with a lower temperature must not be used immedi-
ately on completion of another vacuum program, as the chamber would be
too hot for this program sequence.
Store the sterilization material outside the sterilizer during waiting times,
and only transport it into the chamber directly before starting the actual
program.
Risk of injury!
Make sure that the loading trolley is firmly locked on the transport
Danger trolley.
Do not adjust the height when the TW HUB is docked and locked!
To load the sterilization material into the chamber, push lever 2 on the
transport trolley to release the loading trolley from it.
Push the loading trolley into the chamber.
Risk of injury!
The loading trolley can fall off while being drawn into the sterilizer.
Danger Ensure that the transport trolley is docked correctly to the sterilizer
before the loading trolley is pushed into the chamber.
To release the transport trolley from the sterilizer, push lever 3 on the
transport trolley.
Risk of injury!
Ensure that the loading trolley is located fully within the chamber
Danger before releasing the transport trolley.
Program selection
Select the desired program by touching
the name of the program on the touch
screen.
Program start
The color of the diamond indicates which MMM BMS PCD test needs to be
used for the selected program.
Observe the instructions for use included with the PCD test.
Program start
Press the start button and if necessary enter the user ID (see
below).
User identification
Program sequence
During the program sequence, the on-line process flow chart appears. The
pressure, temperature and, if applicable, air detector curves are displayed
in real time on the screen.
Remaining time
The program name, batch number of the
current batch, current process step, time
remaining to the end of the batch, chamber
pressure (registration) and chamber
temperature (registration) are also shown
on the remaining time display.
The user can switch back to the screen with the online process curve by
touching the screen.
If a program has been accidentally started that must not be used for the
material to be treated, the “Cancel program” should be pressed to prevent
any damage to the material or the machine.
After acknowledging the alarm message, the door on the loading side can
be opened.
On two-door devices, the door on the unloading side can be opened and
closed as many times as desired, until the loading door has been opened
once.
The door on the loading side can only be opened once the door on the
unloading side has been opened and closed again.
Once the following programs are complete, only the door on the loading
side can be opened at the end of the program:
■ B&D Test
■ Vacuum test
■ Heating and SteamSpy heating
■ Disinfection program
■ Failed program
For your personal protection, stand to the side of the chamber when
Danger opening the chamber door.
Hot surfaces!
Do not adjust the height when the TW HUB is docked and locked!
Risk of injury!
The loading trolley can fall off while being drawn into the transport
Danger trolley.
Ensure that the transport trolley is coupled firmly to the sterilizer
before the loading cart is pulled out of the chamber.
To release the transport trolley from the sterilizer, push lever 3 on the
transport trolley.
Risk of injury!
Ensure that the loading trolley is located fully on the transport trolley
Danger before releasing the transport trolley from the locking mechanism.
The pass through-program for the redelivery of empty loading carts from
the unloading side to the loading side is started from the loading side.
Select the Pass through program
The door on the unloading side can now be opened and closed as many
times as desired, until the door on the loading side has been opened once.
However, the door on the loading side can only be opened once the door
on the unloading side has been shut again.
The batch is released based on a visual inspection and the batch record.
Check that the program selection corresponds to the prepared goods.
Check that the process-related data on the batch record, e.g. tempera-
ture, pressure and time, were maintained throughout the entire process
and correspond to the data specified in the validation .
Check the packaging and, if necessary, the seals for integrity (visual
inspection).
Check the packaging for condensate residue (visual inspection).
Check the results of the process indicators (if present). The operator
must document the criteria for evaluating the process indicators in a
work instruction.
When using the MMM SteamSpy: Check the results of the cycle control
given on the batch record.
Release documentation
The release results must be documented. The goods must only be released
by a person named by the operator. The batch and the release personnel
must be clearly identifiable in the release documentation.
The last 20 batch records stored on the device can be shown on the display
and reprinted.
Press the Tool selection button.
The batch name and date the selected batch was saved are shown in the
info field (right side of screen).
The selected batch record is printed depending on the setting of the batch
print option ( Chapter 5.2, "Setting batch print and save options").
The last 20 batch files stored in the machine controller can be copied to a
USB stick.
Press the Save button to copy the last 20 batch files to the
USB stick.
The device immediately switches to standby mode. The MMM logo and the
following message appears in the display: Standby mode active.
The jacket control is paused in standby mode.
4.6 Shut-off
The device is switched off using the mains switch. In normal operation, the
master switch remains switched on.
Press the mains switch
Keep the chamber doors closed to reduce heat output and prevent
chamber interior soiling when the machine is not in use.
5 Machine configuration
The machine configuration described below can be performed by the
system supervisor via the control menu.
Process-related changes can only be carried out by persons who have the
master passport (e.g. MMM customer service).
Users can also be deleted and passwords changed on this menu screen.
Scroll to page 2.
Programs that the currently logged-in user is not permitted to start are
greyed-out in the program selection menu and the user cannot start the
programs.
Depending on the model, one or more of the following options for printing
and saving the data of the batch record will be available.
■ Thermal printer option: Print-out to integrated thermal printer.
■ Direct printing option: Direct printing to a DIN A4
printer on the network. One printer per sterilizer is required for multiple
sterilizers.
■ SimServ option: Transfer of batch data to a PC installed in the network.
Data can be saved and printed here. The SimServ program must be
installed on the PC. Only one PC with SimServ and a printer is required
for multiple devices.
Thermal printer
Activate this printer by placing a green tick in the Thermal printer
section.
Printing via PC
Activate this printer by placing a green tick in the SimServ batch printing
section.
Enter the IP address of the network computer on which SimServ is
installed.
The selected parameters are printed out to the connected and activated
printer.
Each panel has a node number, which is set with the rotary switch on the
back of the panel. The node numbers per machine must match. Duplicate
node numbers are not permitted in the network.
This screen shows the MAC address, which will be required with service
questions.
Because the machine must not be shut off if auto start is enabled, this
function should be combined with the standby mode function.
In the energy-saving standby mode, the jacket heating is paused and the
machine cools down.
Enable the
steam manager parameter.
Switching between operation with cold water and operation with an external
cooling circuit is carried out using manually actuated ball valves and by acti-
vating the corresponding parameter.
On operation with cold water, the levers of three-way ball valves must be
moved manually to the vertical position and the lever of two-way ball valve
must be closed. The parameter is deactivated.
The function for automatically switching from the cooling circuit to cold
water can be activated in the global parameters of the machine.
If a fault occurs, the program is canceled and the machine enters a safe
state. The door on the loading side can only be opened after the fault has
been confirmed. The door on the unloading side is locked.
After the fault has been eliminated, the interrupted program must be
restarted manually.
Danger
Unloading the chamber after the program has not completed properly
On machines with two doors, only the door on the loading side can be
opened.
Steam may escape. Stand to the side of the chamber when opening
the door.
Wear gloves to protect your hands when removing the material.
Danger
Unload the chamber manually by pulling the loading trolley out of the
chamber on the loading side.
Risk of injury!
The resuction of the door seal can be repeated manually. To do this, the
machine must be set to Operational or Unloading.
The door seal is suctioned backfor a specific, previously set time; the blue
lamp flashes.
If the door seal cannot, or cannot be fully suctioned back, automatic move-
ment of the door is not possible.
If, due to a defect, the door seal repeatedly cannot be suctioned back
and the door cannot be opened (e.g. if the blue indicator is lit up
continuously, i.e. the door seal pressure cannot be reduced), inform
MMM customer service.
The service door has been opened. The relevant message appears:
Service door open on loading side or Service door open on unloading side.
Close the service door.
During the preceding program sequence, the program stop was pressed.
The Program stop actuated message appears.
According to the door logic, only one door can normally ever be opened.
The message Door BS open or Door ES open appears.
Close this door.
The chamber door cannot be opened whilst the chamber is not recognised
as depressurised.
Inform MMM customer service.
Unload the chamber manually by pulling the loading trolley out of the
chamber on the loading side.
Danger
If the green lamp next to the hardware name is lit, the input is active or the
output actuated.
You can scroll between the individual pages using the arrow buttons.
The names of the inputs and outputs can be assigned to the corresponding
hardware elements based on the electrical plan.
If the ISA - Intelligent Service Advisor option is activated, the user can
send a message with diagnostic information to MMM customer service after
a fault.
Open the control menu on the loading side.
Press the Service button.
For configuration of the ISA option observe the ISA Installation and
Configuration Manual.
The status display on the screen and the info bar (if installed) switch on and
off alternately.
The cold water valves are opened and closed several times.
Loading is started.
Risk of injury!
Since the unloading function check can only be started from the
Danger loading side, a second person must observe the procedure on the
unloading side to ensure the movement zone of the transport cylinder
remains clear.
When using a floor level loading machine, simultaneously press the All
clear button and the Reset button for 10 seconds.
7 Maintenance
The preventative maintenance counter displays how many cycles are left
until the next preventative maintenance procedure. If the preventative main-
tenance counter displays zero, the info message Perform preventative
maintenance appears on the display after each cycle and is printed on the
batch record.
When working on the machine, turn off the master switch and secure
it with a lock to prevent it from being switched back on unintention-
Danger ally.
Work in the area of the door mechanism or closure mechanism must only
ever be performed by qualified persons ( Chapter 1.3.3, "Specialist") as
specified by the manufacturer.
When performing maintenance work underneath the door, the door must be
secured by screwing the safety screw into the blind hole in the door guide
rail to prevent it from moving. The safety screw must be secured in place
with a lock nut.
Risk of injury!
If the door is not secured, there is a risk of injury from the door drive
Danger chain or door spring.
Risk of crushing!
Vacuum pump
When the machine is idle for long periods (longer than four weeks) or if
there is a risk of frost, the vacuum pump and circulation containers must
be drained by removing all lock screws at the bottom and preserved with
antifreeze (MMM article no. 010495501).
Loading side:
Press the Service button.
a medical products log / equipment log must be maintained; current reports such as batch documen-
tation, assignment of the temperature and pressure curves (if available) and the tests, checks, mainte-
nance and repairs performed must be entered in this each day.
Daily:
Check door seals The seal is automatically tested for vacuum tightness by the Beware of steam emissions!
vacuum test program. In the event of a leak, i.e. when
During subsequent empty sterilization (heating program), steam escapes from the door area,
check the seal for leak tightness against excess pressure. replace the seal ( Chapter 7.4,
If the SteamSpy option is installed, the door seal only need be "Exchanging the door seal").
checked for leak-tightness with the vacuum test program every
four weeks.
Check dirt filter Check the dirt filter in the chamber floor; remove and clean it if
necessary.
Weekly:
Risk of burns!
Allow the chamber to cool sufficiently
beforehand.
Door seal and closure Check the chamber's door seal and closure mechanism for
wear and damage.
To prolong the service life, we recommend cleaning the seal groove regularly.
Clean seal groove Remove the seal from the groove using a suitable tool. Observe the instructions in
Chapter 7.4, "Exchanging the door
Clean seal groove. seal", to avoid damaging the seal.
Grease the sides of the seal groove (with MMM special
lubricant, article no. 0613047).
Due to reasons of operating safety, reliability and economy, the machine must be regularly checked by a
specialist (preferably MMM customer service). The result must be entered in the medical products log /
equipment log.
Maintenance is required after 12 months or 800 batches at the latest. Technical data subject to change
without notice.
Check pilot valves Y771 Check safety shut-off of media valves to the chamber when door end Y771 (and with optional bulk-
(and Y772 in the case switch S11 or S13 signals "Door open". head Y772) must close.
of optional bulkhead) During the program sequence:
Manually actuate the S11 (S13) roller lever
Safety valve Check the function of the safety valve (compressed air).
(if present) The function of the valve can be checked by removing it, and then
testing it on a suitable test stand or using suitable equipment.
To be performed by a
specialist only!
Check thermal water Check setting of thermal water valve and adjust if necessary
valve
Check valve Y03 Check the valve function by starting a program (except vacuum test).
Steam pressure reducer Check setting of steam and air pressure reducers, adjust if neces-
(if present) and air pres- sary
sure reducer
Temperature measuring Check temperature measuring facility
facility
Test run Perform test run and check program sequence for correct time
setting. Check pressure and temperature measurement devices and
their displays.
Check jacket pressure compare the pressure transducer pressure value (screen) with the
control manometer value.
Check program stop When operational: Door movement not possible.
function Reset with key.
Press program stop
Function of SteamSpy Check the function of SteamSpy using a Bowie-Dick test or cycle The error message "Failed
(if present) control. integrated B&D test" or "Failed
cycle control" appears.
Loading trolley Clean the steering and fixed castors of the loading trolleys of Recommended lubricant:
Clean rollers machines that can be entered at floor level, and check they are suffi- Klübersynth VH1 64-1302
ciently lubricated. (MMM article no. 0634570)
Transport trolleys Check the function of the locking bolt and parking brake.
Danger
Then a new inspection is required as per DIN EN ISO 17665 T1, P.12.5.
Use thermal paper 112 x 46 x 12, which can be archived for up to 10 years.
Open flap.
On opening the paper repository, the printer
roller is automatically guided out of the
printing mechanism with the flap.
For further information, see the pressure vessel drawing or the acceptance
test report as per Annex I of the Pressure Equipment Directive.
Check safety- Check that they are installed All safety-relevant compo-
relevant compo- nents must be installed
nents Visual inspection: Check status for and in functional condi-
tion.
visible faults
Function test
Visual inspections of the outside of the machine can be replaced with other
suitable equivalent procedures if visual inspection is not possible due to the
structure of the pressure vessel or the operating conditions. The pressure
vessel or system component is not usually shut down for the inspection.
The vessel must be checked for cracks (e.g. surface cracking, ultrasonic
tests) as part of the inspection program specified by the operator with refer-
ence to the number of cycles during operation.
Check pres- Visual inspection: Visually assess The system parts must be
sure-bearing free from damage.
relevant system components. Pres-
walls
sure-bearing walls can be inspected Highly stressed locations
must not have any cracks.
with aids such as mirrors or camera
systems.
Function test
Documents Check that the system is installed as The system must conform
to the data in the inspec-
specified by the information in the
tion documentation.
inspection documentation.
7.7 Cleaning
Regularly cleaning the sterilization chamber, the assemblies compartment
and the external surfaces prolongs your sterilizer's service life.
The inner chamber walls are manufactured from stainless steel, which is
corrosion-resistant to many media and environmental conditions. Regular
cleaning is recommended to maintain the chamber's corrosion resistance
and durability ( see Chapter 7, "Maintenance ").
This cleaning can be carried out by washing out with deionised water and/
or using commercially available stainless steel cleaning agents (observe
the manufacturer's instructions for use!). Cleaning can be supported with
brushes, textiles, non-woven cloths or abrasive and polishing agents.
When using cleaning agents, always make sure they are suitable for
stainless steel!
This applies to both chemical cleaning agents and aids such as cloths and
brushes, etc.
Normal steel wool or steel brushes, whose abrasion can cause extra-
neous rust, and cleaning agents containing halogens (chloride and
fluoride ions), i.e. hydrochloric or hydrofluoric acid, must not be used
under any circumstances.
Danger
All parts of the system must then be disinfected with a commercially avail-
able disinfectant.
To care for the stainless steel front cover, we recommend Ultrapur D steel
cleaning agent (MMM Art. No.: 0544308-3921). It is sprayed sparingly onto
the metal surfaces or a cloth and distributed in the direction of the grinding
surface patterns.
The touch screen can be cleaned with a soft cloth and a gentle cleaning
agent or a standard disinfectant.
If this is done with the machine switched on, the touch screen's sensitivity
to touch must be temporarily deactivated to prevent unintentionally pressing
buttons.
Open the control menu.
Press the Clean touch button.
The touch screen no longer reacts to inputs for 30 seconds. The time
remaining for this function is decremented on the screen.
The following cleaning instructions apply to all MMM transport trolleys (incl.
TW HUB and TW HV).
Clean and disinfect the rollers and platform using a cleaning cloth and suit-
able surface disinfectant.
You must not use the software if you do not agree to the terms of this
Licence Agreement. If that is the case, you must inform MMM Münch-
ener Medizin Mechanik GmbH and return the unused software and the
accompanying documentation within ten (10) days of purchase.
Section 1 Definition
"Patches" shall mean the removal of program errors (bugs) or faulty func-
tions of the software or of the corresponding software code.
"Product" shall refer to (a) one or more machines of this series or other
machines that you may have bought for use with the software from MMM
Münchener Medizin Mechanik GmbH or (b) the software itself if you have
only purchased the software.
"Updates" shall mean software that corrects faults in an older version of the
software or that, without implying any obligation under this contract to do
so, enhances the functionality of the software by providing additional func-
tions or any other increases in performance.
Provided that the following restrictions are legally acceptable, the licensee
is allowed to use the software only in such a way, as it is allowed according
to the licence agreement described above; the licenser is not allowed
(a) to modify the software or parts of the software in any way whatso-
ever (including, but without limitation by modifying, adapting, translating
or secSoftware Licence Agreement REV1.0 / 2014-09 45 / 49 ond-hand
version), (b) to decompile the software or parts of the software, (c) to
reverse engineer or disassemble the software or parts of the software or
to transform the software in any other way into a human readable format,
This software licence does not apply to software of any third party and does
not include the use of such software. The right of the licensee to use such
software is subject to regulations stipulated by such third parties.
Section 3 Warranty
Express warranty. The Licensor warrants to the original Licensee that (a)
the data carrier on which the software is stored is free from any manufac-
turing and material defects when delivered to the Licensee and that (b) the
software (but not updates) shall, at a fundamental level, function in compli-
ance with the Specification of Services for a warranty period of one (1)
year from the date of purchase, provided that the software is used in accor-
dance with the foregoing licence, on the Defined Computer System and in
compliance with the terms stipulated in the product description relating to its
installation, use and operation.
The Licensor does not warrant that the software shall be free from defects,
run without interruption, meet the expectations of the Licensee or function
in combination with third-party hardware or software products, or that all
program errors shall be removed. In addition to the foregoing, in order for
a defect in the software to be sufficiently material as to violate the warranty
set forth in letter (b) above of this paragraph, the defect must cause the
software – whilst being used as permitted by the foregoing license – to
function in a way so divergent from the Specification of Services that it
is unsuitable for the purpose described in the product description. If the
required functionality can be achieved by the Licensee indirectly (through a
"work-around"), then the applicable impairment shall not constitute a defect
giving rise to obligations under the foregoing warranty. The Licensor's sole
obligation under the foregoing warranty shall be – at its own discretion and
cost – to either (a) replace the data carrier and/or software so that it corre-
sponds, at a fundamental level, to the Specification of Services (including
replacement with a newer version or an equivalent piece of software) or
(b) repair the software by supplying correction code, work-around solu-
tions and/or updates, including updated documentation and other docu-
ments, or (c) to terminate this Licence Agreement and refund the licence
fee upon return of the software, if the fee was charged separately when
the product was handed over, in accordance with Section 6 below. The
foregoing warranty shall apply to any replaced data carriers and software
until expiry of the original warranty period. The costs and risk of delivering
software to the service point defined by the Licensor shall be borne by the
Licensee.
If the Licensee identifies a defect in the Software that may give rise to an
obligation under the foregoing warranty, it shall cease using the Software
immediately and notify the Licensor in writing of the defect within the defect
notification period while also providing sufficient supporting documentation.
The defect notification period is ninety (90) days from delivery of the data
carrier (for a defect in the data carrier) and six (6) months from delivery of
the software (for a defect in the software). The written supporting docu-
mentation providing proof of the defect shall be sufficient if the Licensor is
able to reproduce the defect identified by the Licensee. The Licensee shall
enclose the relevant purchase receipt so that the Licensor may determine
whether the defect notification period has been observed in respect of the
defect concerned. The Licensee shall not carry out any modifications or
repairs itself or allow modifications or repairs to be carried out by unautho-
rised third parties. If the Licensor so desires, the Licensee shall support the
Licensor in the analysis of the causes and circumstances giving rise to the
defect, as well as in the development and testing of correction codes or a
work-around solution.
Exclusivity of warranty.
The Licensee's sole remedy for software defects is set forth in the fore-
going express warranty. The software is licensed with its current features
"as is". Other than the foregoing express warranty, the Licensor provides
no other warranty or guarantee whatsoever. Such express warranty is in
lieu of all other warranties, express or implied, either in fact or by opera-
tion of law, including assurances, terms or conditions of merchantability,
fitness for a particular purpose, satisfactory quality and non-infringement,
all of which are expressly disclaimed. It shall be the responsibility of the
Licensee to select the software that fulfils its requirements. The Licensee
shall bear the full risk for the performance of and results achieved by the
software and for its suitability for the use that Licensee has planned for it,
even if the Licensor has been informed of the planned use of the software
in advance. The Licensor shall be relieved of its obligations under the fore-
The Licensee shall only hold those rights to the Software that are expressly
described in Section 2 of this Licence Agreement. The Licensor shall
remain the sole holder of any other rights with regard to the software,
including, without limitation, ownership rights and patent, copyright, trade-
mark, trade secret and other intellectual property rights. The Licensee must
not remove, obscure or modify any references on the software to copy-
rights, trademarks or other ownership rights of the Licensor. The Licensee
shall take all reasonable steps to prevent any unauthorised use, repro-
duction, sale or publication of the software or the unauthorised provi-
sion of access thereto. The Licensee shall indemnify and hold harmless
the Licensor against any loss, damage, claims and expenses (including,
without limitation, reasonable legal expenses) relating to any infringe-
ment of the rights of the Licensor caused by the Licensee, the Licensee's
breach of this Licence Agreement or the Licensee's use of the software in a
manner not permitted under this Licence Agreement.
To the extent permitted under applicable law, the Licensor shall not be
liable for any direct, indirect or consequential loss or damage, including,
without limitation, loss of profits, unrealised cost reductions, loss of data or
increased costs on the part of the Licensee or any other financial losses
resulting from or in connection with the purchase, licensing, use, or failure
of the software, or disturbances when operating the software. The fore-
going limitation of liability shall also apply in the event that the Licensor has
been notified in advance of the possibility of such loss or damage being
incurred. The Licensor shall only be liable for loss or damage caused by the
Licensor's gross negligence or wilful misconduct. This limitation of liability
shall apply to all claims for damages irrespective of their legal grounds,
including, without limitation, claims based on tort, contract, pre-contract
or quasi-contract. This limitation of liability shall also apply to any officers,
directors or employees of the Licensor and any representatives or agents of
the Licensor that are involved in the development, marketing or delivery of
the software. It is the sole obligation of the Licensee to ensure that it and its
employees possess the required knowledge to properly install and/or use
the software. The Licensor shall not be liable for problems and defects that
arise from insufficient knowledge on the part of the users of the software.
This licence agreement shall become effective on the date, when the
licensee accepts and acknowledges the terms of this licence agreement.
The beginning and end of the warranty period is defined in the contractual
agreements.
10 Index
D Maintenance activities............................. 113
.................................................................. 95
Manual advance........................................ 93
debug e-mail............................................. 95
A Master switch............................................ 26
Deleting barcodes..................................... 67
Accessories............................................... 40 Media......................................................... 15
Disinfection.......................................... 35, 51
Access protection...................................... 28
display....................................................... 56 N
Acknowledgement button.......................... 55
Disposal..................................................... 20
network addresses.................................... 83
Acknowledging messages......................... 91
door movement......................................... 91
Network security........................................ 13
Air detector................................... 36, 36, 51
door release.............................................. 27
All clear button.......................................... 55 P
Alloplast.............................................. 35, 51 E
Personal protective equipment.................. 23
Analysis options in the event of faults....... 95 Electrical equipment.................................. 34
Plastics...................................................... 48
APPcycles................................................. 71 EMERGENCY STOP................................ 54
Pressure container.................................... 33
Approved operators................................... 22
F printing the batch...................................... 81
assemblies................................................ 24
Print settings............................................. 82
Function buttons on the color touch
autostart.................................................... 63
Prion prevention.................................. 35, 48
display....................................................... 58
Auto start............................................ 38, 84
Process validation..................................... 18
Function check.......................................... 96
B program cancellation........................... 25, 89
H Program cancellation................................ 25
barcode............................................... 44, 65
Heating program....................................... 63 Program end............................................. 73
Barcode administration............................. 39
Height-adjustable transport trolley............. 41 Program selection............................... 47, 70
Batch documentation................................ 43
Program sequence.................................... 71
batch record.............................................. 76 I
Program start............................................ 71
Batch release............................................ 75
inspection intervals.................................. 118
Program stop pressed............................... 89
Batch release with SteamSpy................... 75
inspections.............................................. 118
Program stop switch................................. 26
Bowie-Dick test......................................... 50
Instrumentation and control equipment..... 34
pulsed vacuum process............................ 46
C Integrated Bowie-Dick test with SteamSpy 36
steam generator........................................ 62
SteamSpy........................................... 36, 75
Steam supply............................................ 34
Sterilization programs............................... 46
sterilization steam..................................... 15
System supervisor..................................... 22
T
TCP/IP configuration................................. 83
time........................................................... 85
Time taken................................................ 91
transport trolley......................................... 40
TW HUB.................................................... 41
U
Universal................................................... 47
Universal SD............................................. 47
unloading side........................................... 53
V
Vacuum facility.......................................... 34
check......................................................... 96