Us Mri Ct-Exam
Us Mri Ct-Exam
Us Mri Ct-Exam
A. Increases
B. Decreases
A. Increases
B. Decreases
A. 10,000
B. 20,000
C. 30,000
D. 40,000
C. The wave front is attenuated by converting the energy in the pressure wave to heat.
5. Which of the following types of tissue would have the lowest mean propagation velocity.
A. Air
B. Fat
C. Blood
D. Water
6. The ___ transducer and corresponding field of excitation excites molecules in the traditional "pie-
shaped" field.
A. 1D
B. 2D
C. 3D
D. 4D
8. When c2>c1, total reflection can occur when the angle of incidence is ______ than the critical
angle.
A. Less
B. Equal to
C. Greater
A. Aperture
B. Linear
C. Transducer
D. Propagation
A. Frequency (wavelength)
B. Transmit intensity
C. Pulse interval
D. Acquisition
Answer Key
1. A
2. A
3. B
4. D
5. A
6. B
7. D
8. C
9. A
10. D
If the speed of sound in a medium is 100m/sec, how long will it take sound to
travel 420m?
1. .42 sec
2. 4.2 sec
3. 42 sec
4. 420sec
5. 4200 sec
When the signal to noise ratio is low, the signal is __________ than the
noise, therefore the image quality is __________?
1. higher, better
2. lower, better
3, higher, worse
4. lower, worse
v=λf
• Which of the following is false about acoustic impedance?
• If two materials have the same acoustic impedance, their
boundary will not produce an echo
• If the difference in acoustic impedance is small, a weak echo
will be produced,
• If the difference in acoustic impedance is large, a strong
echo will be produced.
•
•
• Doppler shift is given by the following formula:
• FD = (V X FO ) / C
• where
• FD is the Doppler shift
• FO is the original frequency
V is the velocity of blood
C is the speed of sound in tissues
Therefore V = (FD X C) / FO
Transducer
Piezoelectric crystal
Emit sound after electric charge applied
Sound reflected from patient
Returning echo is converted to electric signal grayscale image
on monitor
Echo may be reflected, transmitted or refracted
Transmit 1% and receive 99% of the time
The wavelength of a sound wave is the physical distance between any two compressions
or decompressions (rarefactions.)
Ultrasound imaging makes use out of the reflections of high frequency sound waves.
The transducer head contains the piezoelectric material and the transducer is acoustically coupled to
the body by using a gel.
The gel minimizes reflections of the ultrasound at the skin’s surface and a beam of ultrasound
pulses are delivered to the body.
The ultrasound pulse has an intensity I that decreases exponentially as it passes through the
body’s tissues and organs.
Doppler Effect:
- For stationary targets the number of waves encountered by unity of time is equal the wave
frequency.
The number of additional wave fronts encountered is a function of the velocity of the moving target
and the velocity of sound.
However, if two materials have similar elastic properties then the US waves will produce small
reflections and most of the US will propagate into the second medium.
Advantages of US
Ultrasound examinations are non-invasive i.e. they do not require the body to be opened up, or
anything to be inserted into the body. This is a major advantage compared to fiber optic
endoscopy, for example, which may involve much more patient discomfort as the probe is
inserted.
Ultrasound methods are relatively inexpensive, quick and convenient, compared to
techniques such as X-rays or MRI scans. The equipment can be made portable, and the
images can be stored electronically.
No harmful effects have been detected, at the intensity levels used for examinations and
imaging. This contrasts with methods based on X-rays or on radioactive isotopes, which
have known risks associated with them, and ultrasound methods are preferred whenever
possible. This is particularly relevant to examination of expectant mothers.
Ultrasound is particularly suited to imaging soft tissues such as the eye, heart and other
internal organs, and examining blood vessels.
Disadvantages of US
The major disadvantage is that the resolution of images is often limited. This is being
overcome as time passes, but there are still many situations where X-rays produce a
much higher resolution.
Ultrasound is reflected very strongly on passing from tissue to gas, or vice versa. This means
that ultrasound cannot be used for examinations of areas of the body containing gas, such
as the lung and the digestive system.
Ultrasound also does not pass well through bone, so that the method is of limited use in
diagnosing fractures. It is possible to obtain quite good ultrasound scans of the brain,
but much greater detail is obtained by an MRI scan.
confinement discomforting
In computerised tomography (CT) the X-ray source rotates around a plane of the
body, taking serial pictures with a detector (instead of a film) which are
synthesized by a computer. The resulting picture created by the computer is like
a section of the body and can be recorded on a film. CT pictures are therefore like
X-ray images.
Magnetic resonance imaging (MRI) uses the property of protons aligning
themselves in a magnetic field and their reaction to radio frequency waves. The
protons ‘resonate’ to the radio frequency and revert to normal (‘decay’) when the
radiation is stopped. Effectively it is the imaging of protons. The most commonly
imaged proton is a hydrogen nucleus. So far it is believed that this method does
not damage body tissues as X-rays do. MRI images are even more realistic than
CT images.
Ultrasound on the other hand uses mechanical waves of frequencies beyond the
audible range. These waves are reflected to various degrees from junctions of
tissues of different nature. Ultrasound pictures require considerable skill to
interpret. Ultrasound has a great advantage – it does not cause cellular damage
when used in quantities required for imaging.
(2) Which of the following has a Hounsfield unit value of approximately - 90?
(A) Fat
(D) Bone
(E) Lung
(4) The measured x-ray transmissions from a single CT fan beam through a
patient is called a:
(A) Filter
(B) Back-projection algorithm
(C) Tomographic slice
(D) Primary beam
(E) Projection
(6) If a CT display is set at a window width of 100 and a window center of 50, the:
(A) HU value of water changes to 50
(B) White matter will look gray
(C) Gray matter will look white
(D) Bone will look black
(E) Lung will look white
(9) CT anode heat loading increases with all of the following except an increase
in:
(A) Filtration
(B) Voltage
(C) Current
(D) Collimator thickness
(E) Field of view
(17) The theoretically best possible CT resolution for a 512 2 matrix and 25 cm FOV
is:
(A) Filtration
(B) mAs
(C) Matrix size
(D) Display window width
(E) Size of film image
(22) The difference in x-ray attenuation between white (40 HU) and gray matter (50
HU):
(A) Is 0.1%
(B) Is 1%
(C) Is 10%
(D) Is 25%
(28) Representative patient doses in CT are expected to include all the following
except:
(29) The dose to the fetus during an abdominal CT scan would not increase with
increasing:
(30) The scattered radiation dose 1 meter from a patient undergoing a head CT
scan is:
1. What imaging parameter is defined as the extent to which the x-ray helix is
stretched during a helical scan, determines volume covered, and affects
radiation exposure and image quality?
a. Pitch
b. Table increment
c. Slice thickness
d. kVp
4. What is the field of view that covers the area within the gantry from which the
raw data is acquired?
a. Scan field of view (SFOV)
b. Reconstruction field of view (RFOV)
c. Display field of view (DFOV)
d. None of the above
5. What post-processed image adjustment to the final displayed image determines
the range of CT numbers that occupy the full gray scale?
a. Window level
b. Window width
c. Image magnification
d. Image resolution
6. Which window center and window width would best display a lung window at
the viewing console?
a. C/W -50, 400
b. C/W -600, 1700
c. C/W 1000, 2500
d. C/W 700, -2000
7. What method or type of contrast administration may occur if the distal colon is
not properly opacified with timed oral contrast agents for a Pelvic CT exam?
a. IV contrast
b. Cystografin
c. Rectal contrast
d. Intrathecal contrast
9. When the scan plane is perpendicular to the subtalar joint and the gantry is
tilted towards the patient at around 25 degrees, what is the directly acquired
imaging plane?
a. Oblique coronal plane
b. Direct axial plane
c. Coronal plane
d. Direct sagittal plane
10. What is the optimal scan slice thickness when imaging all musculoskeletal
anatomy?
a. 5mm-7.5mm
b. 0.625mm-1.25mm
c. 2mm-3mm
d. 10mm
Front
Back
True or False
When two waves reach peaks and
False
cross the zero line at the exact same
Constructive
time, they are "in phase" or
destructive?
True or False
Nyquist criteria is used to determine True
aliasing?
True or False
Logartithms provide a powerful tool to True
compress data non-linearly?
True or False
The speed of sound in soft tissue is True
assumed to be 1540 m/sec?
Kilo is equal to? thousand
1. 51 us
2. .51 m/s (ans. is with units of length)
3. .51 pascals
4. 51 watts
5. .51 mm
pressure
density
What are the 4 acoustic variables?
distance
temperture
True or False
False
If the power of a wave is halved, the
intensity is reduced to one-fourth it
Power is proportional to intensity
original value?
True or False
The effects of tissue on sound waves False
are called bioeffects?
Chapt. 4
Bandwidth
US pulses contain a range of
freguencies called the?
2.0 usec.
What is the pulse duration of a 4 cycle
pulse in a period of .5 usec?
(PD= period x # of cycles)
True or False
A range of the DF is encountered in
DMS because of various conditions True
chosen by the instrument and the
operator?
True or False
Pulsed wave ultrasound is described
False
by the additional terms frequency,
wavelength and propagational speed?
True or False
True
DF is unitless?
True or False
False
SPL is determined by the length of
the pulse and the propagation speed.
True or False
1. frequency x period
4. Period x # cycles in the pulse
2. period x wavelength
3. # of cycles in a pulse/wavelength
4. period x # of cycles in the pulse
1. Temporal
2. Contrast
3. Detail
4. Axial
True or False
If the penetration is increased the False
PRF must be increased
True or False
True
The DF is an indicator of mechanical
bioeffects?
True or False
A non scanned modality implies that
over time the acoustic beams are True
repeatedly transmitted in the same
direction
1024
How many gray levels can be
represented by 10 bits?
2 to 10 power.
True or False
Artifacts can help or hinder proper True
interpretation and diagnosis
True or False
Grating lobes are esential for the
False
proper operation of a linear phased
array transducer.
True or False
Enhancement weakens the amplitued False
of structures idstally.
True or False
Some artifacts are produce by True
improper equipment settings.
True or False
External influences can also produced True
artifiacts.
What is the hydrophone analysis used Safety, quality assureance and the
for? results used in the design process.
Stable
What are 2 types of cavitation?
Transient (Unstable)
-heat absorption
-for each 10 degrees C increased
How are thermal bioefftects caused?
there is a doubling in enzymatic
activity.
True or False
A peak is always greater than or True
equal to the average
True or False
Temporal average is almost always True
small than the pulse average.
-time exposure
What are thermal bioeffects
-Increased intensities above 40
dependent on
degree C.
True or False
One pulse of ultrasound generates a
True
single scan line as it travels through
tissue.
True or False
Pulsed ultrasound transducers can False
only generate ultrasound pulses.
1. period
2. propagation speed
3. wavelength
4. Frequency
True or False
Interfaces that scatter ultrasound
energy are usually considered to be True
those that are equal to or smaller than
the wavelength.
True or False
Spaltial pulse length is determined by
False
the length of teh pulse and the
propagation speed
True or False
The shorter the pulse, the broader teh True
bandwidth
True or False
Optimum apodization changes True
continually with focusing and steering.
True or False
Lateral resolution consistent at any False
depth.
True or False
Intensity variations are greatest in the False
far zone.
True or False
Many frequencies are present in short True
pulses.
True or False
As sound travels, the width of the
False
beam remains unchanged due to
focusing.
True or False
Even unfocused transducer have a True
natural focus.
True or False
Elevational resolution can be
True
considered a third aspect of detail
resolution.
True or False
Contrast and temporal resolutions False
relate more directly to the transducer.
True or False
Snell's law can be applied to the
surface of a transducer to determine False
the beam profile of the wave that
emanates from the surface.
True or False
Cermanics used as materials in the
False
production of modern US transducer
element are naturally piezoelectric.
In order to image deep structures
__________.
1. the TGC needs to be increased
2. the PRF needs to be decreased. 2. the PRF needs to be decreased.
3. the frame rate needs to be
increased.
4. the PRF needs to be increased.
7.7cm
What is the maximum depth of soft
tissue for a PRF of 10kHz? PRP x ID x 13usec.
PRF 1/PRP = 1/10kHz
True or False
There are both transmission and True
reception channels.
True or False
If the penetration is increased the False
PRF must be increased.
Concept questions
Health Science Physics - Exam 6 - Q55
A B C D E
See the solution
Health Science Physics - Exam 2 - Q53
A B C D E
See the solution
Health Science Physics - Exam 2 - Q54
A B C D E
See the solution
Larmor Frequency
Health Science Physics - Exam 5 - Q55
CT does use x-rays and the way x-rays interact with tissue is simple as far as imaging is
concerned. x-rays pass through tissue and get attenuated as x-ray photons are scattered by
tissue atoms. The denser the tissue along the x-ray path, the more photons are scattered. So,
we send the x-ray through the sample and measure what comes out on the other side. The
attenuation measured at the receiver is the total attenuation along the entire line of the x-
ray beam.
Let's say you're getting a regular x-ray, and not a CT, and you're standing up straight, with
your body facing the x-ray source. Let's call the x-ray beam axis the z axis. The image we
obtain will be a value for each point on the x-y plane opposite your body, and this value is
directly proportional to the sum of tissue density along z. We say that's a projection of your
body density map onto the x-y plane where the z-axis gets compressed/summed.
So, what makes a CT different from an x-ray? Imagine keeping that x-ray source at a
constant distance from the center of your body and rotating it around your body slightly to
obtain another x-ray image from another angle or another perspective. Imagine we keep
doing this around 180 degrees. The result is many x-ray images. How do we combine them
to obtain a CT scan? We then take these projections, smear them along their respective z
axis (axis perpendicular to the projection plane), and then add the smeared images. It looks
like this:
MRI on the other hand works on a different principle: the basic idea is that we image
hydrogen in the body, which is amply present in water, and also in all the organic molecules
(fats, proteins, etc.). The hydrogen nucleus (a proton), aligns on average with the magnetic
field (which we supply - the MRI scanner consists of a huge magnet). Now, because there
are so many protons in the volume we measure, the ensemble average behaves classically.
So, we don't need to consider quantum mechanics for most imaging applications. Classically
speaking, we tip the proton magnetic moment away from its alignment using
radiofrequency energy, and the proton magnetic moment then precesses at its resonance
frequency, which is directly proportional to the field strength, emitting a signal that we
measure. In doing so, it loses energy, interacts with its environment and gives off a
radiofrequency signal. How the proton magnetic moments precess, the spread in their
precession frequency, and how long it takes them to go back to equilibrium, all these are
affected by various tissue properties.
We only measure from one antenna, and so the signal from the whole section being scanned
is added together into one measured signal; but how do we resolve which part of the signal
comes from which spatial location? Recall that we said that the resonance frequency of
precession of magnetic moments is proportional to the magnetic field we apply. So, we use
gradients (linear variations) in the field to vary the resonance frequency across the sample.
That way, we can resolve where each frequency is coming from. It's a bit more involved than
this, but this is the simple explanation.
This article explains MRI pretty well: Fundamental Physics of MR ImagingThe same link
lets you download a PDF version (which I prefer for its better placement of the
figures): Page on rsna.org
If you like videos, this video series is helpful. You'd likely need to watch a few of them to
start to get a good feel:
Also, this introductory free online book/tutorial by Prof. Joseph Hornak is a great intro for
someone who wishes to get better acquainted with MRI: The Basics of MRI . If you really
want to learn how it works, I'd start with this tutorial and then move on to one of the more
advanced textbooks (Magnetic Resonance Imaging: Physical Principles and Sequence
Design by Haacke et al. being the main one).
So, from this, you should be able to note one big difference between MRI and CT. CT only
gives you one parameter: density. MRI tells you a lot more. It's looking at hydrogen in
different contexts and different tissue environments, and we have multiple ways to probe
tissue properties, hydrogen density being only one of at least a half dozen properties you can
look at, and some of these properties give you interesting multidimensional information