Chapter 6 Troubleshooting: Warning

Download as pdf or txt
Download as pdf or txt
You are on page 1of 16

Chapter 6 Troubleshooting

This equipment performs self-checks when certain troubles occur. It


emits a buzzer and displays the status, cause and remedial procedure
on the screen in order to notify and assist the operator. It also inform the
operator when maintenance is necessary. These are collectively referred
to as “alarms”. This chapter explains about alarms and other possible
equipment troubles, their causes and how to remedy the situation. In the
following cases, contact ESPEC CORP. or the place if purchase:
When the equipment fails to operate properly after you have taken the prescribed remedial action.
Anywhere in this manual that you are instructed to “Call for service”.

WARNING
When taking action on the primary side of the
breaker, be sure to turn off the main power supply
switch at your facility. Also, use caution not to
apply voltage accidentally.
Attempting to solve a problem with the power on can
result in electric shock and create a very dangerous
situation.
Use the supplied breaker handle stopper to prevent
the breaker from being turned on accidentally.
Shut the main breaker OFF before attempting to
access the electrical compartment.

124
Chapter 6 Troubleshooting

6.1 Displayed alarms


When an alarm is generated, the alarm screen shown below
automatically appears and the buzzer is triggered. The ALARM icon
keeps blinking until the alarm is cleared.
Press the alarm event on the Alarm screen to display the status details.
Check the status details, and then press [Stop Beep] to stop the alarm
buzzer.

Alarm screen

Pressing the alarm event displays the following alarm help (details).
ALARM HELP screen

Notice
• If the error or warning buzzer is turned OFF, alarm status
cannot be notified with a buzzer sound, which may lead
to delays in acknowledging an error or warning. Do not
turn OFF the buzzer if at all possible.
If the buzzer is turned OFF, the occurrence of an error or
warning is indicated only by the red blinking pilot lamps and
on the Alarm display screen.

125
Chapter 6 Troubleshooting

Reference The alarm buzzer notifying a "warning/error" can be set using "Set
Sound" on the Configuration screen. This can be accessed from the
Chamber Setup screen.

Corrective action
Take the following action when an alarm is generated. There are 2 types
of alarms -“Error” and “Warning”- and they are dealt with differently in
some respects.

Error Generated when trouble occurs with the equipment or a


single unit. (A buzzer is emitted.)
Warning Generated when control is destabilized for reasons other
than trouble such as when maintenance is required.

Reference Operation may be continued in backup mode even when an


“Alarm” occurs. Refer to the section “Equipment response in
backup mode“in Chapter 6.

 When an error occurs

Procedure 1. Press [Stop Beep] to stop the alarm beep.

2. Check the troubleshooting procedure by following the instructions in


the manual or messages on the ALARM HELP screen.

 When a warning is generated

Procedure 1. Press [Stop Beep] to stop the warning beep.

2. Check the troubleshooting procedure by following the instructions in


the manual or messages on the ALARM HELP screen.

3. Press [CLR] on the ALARM HELP screen.


Even when a "warning" is generated, operation does not stop.
However, the warning notification will remain on the ALARM screen
until the power is turned OFF.

126
Chapter 6 Troubleshooting

Alarm Report screen


When an alarm has been generated, a report can be found by going to
the Chamber Setup menu. Report on alarms can be obtained while they
are still active, from the ALARM screen, but alarm entries are deleted the
moment the actual alarm is cleared. In any case, all alarms occurred to
date can be seen on the Alarm Report screen.
This screen reports other system information such as refrigerator
defrosting and humidifier cleaning.

Procedure 1. Select Chamber Setup mode from the menu.


Press the Chamber Setup tab.
Press [Alarm Report] on the Chamber Setup screen.

2. The alarm report is displayed.

127
Chapter 6 Troubleshooting

No.: To display the alarm report number (1 - 100).


Type: To display whether it is an error or a warning.
Alarm: To display the error event or warning that has occurred.
Pressing an alarm event displays the ALARM HELP screen
corresponding to that event.

Date: To display the date and time the error or the warning
occurred.
: To select the previous/next page.
Seek No.: Input the alarm report number to jump to the corresponding
alarm or warning.

Reference • Displays the last 10 alarm reports per screen in reverse


chronological order of occurrence.
• Up to 100 warning and error events can be stored. When the
number of events exceeds 100, the reports are deleted in
chronological order.

128
Chapter 6 Troubleshooting

Equipment response in the backup mode


This chamber has a backup feature that can be turned ON/OFF from the
chamber setup mode.
If "ON" is selected for Backup Mode, the chamber issues a warning
when an error occur, but continue operation. (Refer to the program
response table below)
The backup mode can be turned OFF from the touch screen, but in this
situation, the entire system will shut down when trouble occurs.
Though in some cases test conditions cannot be met once the backup
mode kicks in, the mode itself has been added to protect specimens
against damage which might occur when the entire system is shut down,
as well as to minimize time loss as best possible when total shutdown
does in fact occur.
To better understand equipment behavior in the backup mode and use
the chamber more effectively, the below table describes how the
chamber responds when trouble occurs.
To check under which error statuses backup mode is applied, refer to
"Alarm tables" in "Chapter 6 Troubleshooting".

Program response in backup mode


Equipment response when When “OFF” is
Trouble backup mode is “ON” selected for
and kicks in. Backup Mode
Humidifier
Operation continues. Equipment
related trouble
suspends
Unaffected refrigerators continue operation.
running. However, the test system
Refrigerator (Status) is
shuts down if there is only one
related trouble “Program
refrigerator in the system or if all
refrigerators are affected. Suspended” or
“Constant-value
Other Test system shutdown Operation”.

Reference In the Alarm tables on the following


pages, “BU” is indicated where the
backup mode is available for use.
For information on setting backup
mode in the event of an alarm, refer to
section 10.2 “Setting the equipment
response during the operation” in
“Chapter 4 Management Setting” of the
Reference Manual.

129
Chapter 6 Troubleshooting

Alarm tables
Alarms are displayed as a pretext to the backup mode kicking-in. If the
mode is OFF when an error occurs, the entire system will shut down.
However, in the program mode, the operation will pause.
The following codes are used in these tables:
(If corrective actions fail to restore normal operation of the chamber,
please contact ESPEC CORP. or the place of purchase .
BU: Indicates the backup mode is available against the error in
question.
COM. OP: Alarm number of communication function is displayed.
(Optional)
Type of
alarm
Displayed COM. Equipment Probable
Warning

Description Remedy
message OP response cause(s)
Error

 Remedy on each
Error notification
REFRIGERATOR All refrigerators in error alarm occurrence
2320 Test system down already made for each

FAILURE state.  Turn [POWER] key


refrigerator separately
OFF.

2056
2209
2219
REFRIG-□ 2096
TEMP ALM: COMP 2249 Compressor temperatu  Turn [POWER] key

●U

SURFACE 2259 re switch tripped beca OFF.


2136 use of high temperatu  Cool down the
□ indicates the 2289 re on the surface.  Refrigerator breakdo compressor.
refrigerator No. 2299 wn
2176  Condenser trouble
2433  Gas leak
2443  Overheating
 Open-phase
REFRIG.-□
For single  Turn [POWER] key
UNIT FAILURE
2061 refrigerator Temperature rise in re OFF.

(For air-cooled spec.


●U

2101 chambers frigerator unit discharg  Check the cooling


only)
2141  Test system e pipe, or reverse pha water.
2181 down se in unit wiring.  Cool down the
□ indicates the
refrigerator.
refrigerator No.
For multiple
2059 refrigerator
2212 chambers.
2222  Refrigerator
REFRIG-□ 2099 changeover and  Turn [POWER] key
 Refrigerator breakdo
CURRENT VALUE ALM: 2252 continued OFF.

Compressor thermal r wn
●U

COMP 2262 operation.  Check the cooling


elay tripped because  Condenser trouble
2139 water.
of overcurrent.  Overheating
□ indicates the 2292  Cool down the
 Open-phase
refrigerator No. 2302 In program mode, refrigerator.
2179 system does not
2436 pause.
2446
2057
2210
2220
2097
REFRIG.-□
2250 Refrigeration circuit hi  Cooling tower fan st  Turn [POWER] key

HIGH PRESSURE
●U

2260 gh pressure switch tri opped OFF.


2137 pped because of pres  Foreign matter inside  Check the cooling
□ Indicates refrigerator
2290 sure rise. condenser water.
No.
2300
2177
2434
2444

130
Chapter 6 Troubleshooting

Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)

Error
2058
2211
2221
2098
REFRIG.-□
2251 Refrigeration circuit lo
 Turn [POWER] key

LOW PRESSURE
●U
2261 w pressure switch trip  Evaporator is heavily
OFF.
2138 ped because of press frosted over.
□ Indicates refrigerator  Perform defrosting.
2291 ure drop.
No.
2301
2178
2435
2445
2040
REFRIG □ 2226
BURN-OUT: DISCHG 2080 Burn-out of the refrige
Burn-out of the refriger  Turn [POWER] key

●U

TEMP 2266 rator discharge pipe th


ator discharge pipe th OFF.
2120 ermal sensor is detect
ermal sensor  Call for service
□ indicates the 2306 ed.
refrigerator No. 2160
2450
REFRIG-□
BURN-OUT: COND 2041
Burn-out of the refrige Burn-out of the refriger  Turn [POWER] key

●U

TEMP 2081
For single rator condenser therm ator condenser therma OFF.
2121
al sensor is detected. l sensor  Call for service
□ indicates the 2161 refrigerator
refrigerator No. chambers
2042  Test system
2213 down
2224
REFRIG- 2082 For multiple
BURN-OUT: EVAP IN 2253 refrigerator Burn-out of the refrige
Burn-out of the refriger  Turn [POWER] key

chambers.
●U

TEMP 2264 rator evaporator inlet t


 Refrigerator ator evaporator inlet th OFF.
2122 hermal sensor is dete
changeover and ermal sensor  Call for service
 indicates the 2293 cted.
refrigerator No. 2304 continued
2162 operation.
2437
2448
In program mode,
2043
system does not
2214
pause.
2225
REFRIG- 2083
BURN-OUT: EVAP OUT 2254 Burn-out of the refrige
Burn-out of the refriger  Turn [POWER] key

●U

TEMP 2265 rator evaporator outlet


ator evaporator outlet t OFF.
2123 thermal sensor is det
hermal sensor  Call for service
 indicates the 2294 ected.
refrigerator No. 2305
2163
2438
2449
REFRIG-
BURN-OUT: COMP 2044 Burn-out of the refrige
Burn-out of the refriger  Turn [POWER] key

●U

SUCTION TEMP 2084 rator compressor inlet


ator compressor inlet t OFF.
2124 thermal sensor is det
hermal sensor  Call for service
 indicates the 2164 ected.
refrigerator No.
2053
REFRIG- 2223  Refrigerator breakdo
BURN-OUT: DISCHG 2093 Temperature of the re wn  Turn [POWER] key

●U

TEMP 2263 frigerator discharge pi  Condenser trouble OFF.


2133 pe rises and an error  Gas leak  Cool down the
 indicates the 2303 is detected.  Overheating compressor.
refrigerator No. 2173  Open-phase
2447

131
Chapter 6 Troubleshooting

Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)

Error
2048
REFRIG- 2229 Temperature of the re  Refrigerator breakdo
OUT-OF-RNG: DISCHG 2088 frigerator discharge pi wn  Turn [POWER] key


●U
TEMP 2269 pe remains out of nor  Condenser trouble OFF.
2128 mal range over the de  Gas leak  Cool down the
 indicates the 2309 termined period of tim  Overheating compressor.
refrigerator No. 2168 e.  Open-phase
2453
When upper limit dete  Turn [POWER] key
cted OFF.
 Refrigerator breakdo When upper limit
REFRIG-
Temperature of the re wn detected
OUT-OF-RNG: COND 2049
 Condenser failure  Check cooling

frigerator condenser re
●U

TEMP 2089
mains out of normal r  Gas leak water.
2129
ange over the determi  Overheating  Cool down the
 indicates the 2169
ned period of time.  Open-phase refrigerator.
refrigerator No.
When lower limit dete When lower limit
cted detected
 Frosted cooler  Perform defrosting.
2050
2215
2227
REFRIG- 2090 Temperature of the re
OUT-OF-RNG: EVAP IN 2255 frigerator evaporator in

●U

TEMP 2267 let remains out of nor  Turn [POWER] key


 Gas leak
2130 For single mal range over the de OFF.
 indicates the 2295 refrigerator termined period of tim
refrigerator No. 2307 chambers e.
2170  Test system
2439 down
2451
2051 For multiple
2216 refrigerator
chambers.  Turn [POWER] key
2228
 Refrigerator When upper limit dete OFF.
REFRIG- 2091 Temperature of the re
changeover and cted When upper limit
OUT-OF-RNG: EVAP 2256 frigerator evaporator o
 Heat generation from detected

●U

OUT TEMP 2268 continued utlet remains out of n


operation. specimens  Remove heat
2131 ormal range over the
When lower limit dete generating objects.
 indicates the 2296 determined period of t
cted When lower limit
refrigerator No. 2308 ime.
 Frosted cooler detected
2171 In program mode,
 Perform defrosting.
2440 system does not
2452 pause.
REFRIG- Temperature of the re
OUT-OF-RNG: COMP 2052 frigerator compressor i
 Turn [POWER] key

●U

SUCTION TEMP 2092 nlet remains out of no  Evaporator is heavily


OFF.
2132 rmal range over the d frosted over.
 Perform defrosting.
 indicates the 2172 etermined period of ti
refrigerator No. me.
REFRIG-
INVERTER
2072  Breakdown of the ref

COMMUNICATION A Communication erro


●U

2112 rigerator inverter  Turn [POWER] key


ERROR r is detected on the r
2152  Burn-out of the com OFF.
efrigerator inverter.
2192 munication cable
 indicates the
refrigerator No.
REFRIG-
2064
 Moisture adheres to t  Turn [POWER] key

INVERTER ERROR T00


●U

2104 Refrigerator inverter I


he IGBT of the refrig OFF.
2144 GBT shortcircuits.
 indicates the erator inverter.  Dry inverter.
2184
refrigerator No.
REFRIG-
2065 A position detecting ci

INVERTER ERROR T01


●U

2105 rcuit error is detected  Breakdown of the ref  Turn [POWER] key
2145 on the refrigerator inv rigerator inverter OFF.
 indicates the
2185 erter.
refrigerator No.

132
Chapter 6 Troubleshooting

Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)

Error
REFRIG-
2066  Breakdown of the co


INVERTER ERROR T02 A current sensor alar

●U
2106 mpressor  Turn [POWER] key
m is detected on the
2146  Breakdown of the ref OFF.
 indicates the refrigerator inverter.
2186 rigerator inverter
refrigerator No. For single
REFRIG- refrigerator
2067 B chambers  Breakdown of the co
INVERTER ERROR T03 A motor lock error is
●U
2107  Test system mpressor  Turn [POWER] key
detected on the refrig
2147  Breakdown of the ref OFF.
 indicates the down erator inverter.
2187 rigerator inverter
refrigerator No.
For multiple
REFRIG-
2068 refrigerator  Breakdown of the co

INVERTER ERROR T04 A breakdown error is


●U

2108 chambers. mpressor  Turn [POWER] key


detected on the refrig
2148  Refrigerator  Breakdown of the ref OFF.
 indicates the erator inverter.
2188 changeover and rigerator inverter
refrigerator No.
continued
REFRIG- operation. Temperature error (off-
2069

INVERTER ERROR T05 temp.) is detected by


●U

2109  Breakdown of the ref  Turn [POWER] key


the heatsink sensor o
2149 rigerator inverter OFF.
 indicates the In program mode, n the refrigerator inver
2189
refrigerator No. system does not ter.
REFRIG- pause.
2070 The Heatsink sensor  Burn-out of the heat
 Turn [POWER] key

INVERTER ERROR T06


●U

2110 on the refrigerator inv sink sensor on the r


OFF.
2150 erter has short-circuite efrigerator inverter
 indicates the  Dry inverter.
2190 d or burnt out.  Moisture is adhered.
refrigerator No.
REFRIG-
2071  Breakdown of the co

INVERTER ERROR T07 A case thermo action


●U

2111 mpressor  Turn [POWER] key


Test system down error is detected on th
2151  Breakdown of the ref OFF.
 indicates the e refrigerator inverter.
2191 rigerator inverter
refrigerator No.

 Turn main power


INVALID Incorrect input setting
Setting of the refrigera breaker OFF and
REFRIGERATOR 2322 Test system down of the refrigerator mod

tor model is wrong. ON again.


SETUP el
 Call for service

 Turn [POWER] key


 Stopped cooling tow
OFF.
REFRIG.: COOLING Water suspension rela er pump
 Check cooling
WATER FAILURE y tripped because of  Clogged cooling wat
2321 Test system down pump functions.

(For water cooled pressure drop in cooli er strainer


 Check for leaks in
spec. only) ng water line.  Water leak in coolin
cooling water line.
g water line
 Clean strainer.
For single
refrigerator
chambers
 Test system
down
REFRIG-
CURRENT VALUE ALM: For multiple The thermal relay trip
2060  Breakdown of the co  Turn [POWER] key

COND FAN refrigerator ped due to an increas


●U

2100 mpressor fan OFF.


(For air-cooled type) chambers. e in the current value
2140  Overheating  Cool down the
 Refrigerator of the compressor fa
2180  Open-phase compressor fan.
 indicates the changeover and n.
refrigerator No. continued
operation.

In program mode,
system does not
pause.
 System switches
over from temp.
& humidity Dry-boil temperature s  Humidifier water sup
 Turn [POWER] key

●U

HUMIDIFIER control to ensor tripped because ply circuit trouble


1800 OFF.
BOIL-DRY temperature-only of humidifier cylinder  Dirt inside the humid
 Clean humidifier.
control. overheating. ifier cylinder
In program mode,
system pauses.

133
Chapter 6 Troubleshooting

Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)

Error
Operation
continues
uninterrupted.
In program mode,
system does not
Temperature switch in
HUM. SCALE BUILDUP pause. Too much scale inside
1801 side humidifier cylinde Clean humidifier.


WARNING * If left unattended cylinder
r tripped.
for 120 hours,
system interprets
trouble as error
and humidifier
stops operating.
 System switches
over from temp.
Temperature switch in
& humidity
 Turn [POWER] key

side humidifier cylinde


●U

HUM. SCALE BUILDUP control to Too much scale inside


1802 r has been in error st OFF.
ALARM temperature-only cylinder
atus for 120 consecuti  Clean humidifier.
control.
ve hours.
In program mode,
system pauses.
 System switches
over from temp.
 Drainage pump malf
& humidity
 Turn [POWER] key

Humidifier water level unction


●U

control to
HUM. DRAIN FAILURE 1803 does not drop when d  Clogged drain line OFF.
temperature-only
raining.  Humidifier float sens  Clean humidifier.
control.
or trouble
In program mode,
system pauses.
 Water supply valve
not reopened.
 Clogged humidifier st  Check valve.
rainer  Clean strainer.
 Operation
 Water leak  Check water supply
continues
Humidifier was not fill  Low water pressure line.
HUM. WATER SUPPLY uninterrupted.
1804 ed within specified tim  Humidifier float sens  When water supply

WARNING In program mode,


e limit. or trouble is restored, normal
system does not
 Water supply suspen operation is
pause.
ded because safety automatically
device in RO water resumed.
purifier (option) tripp
ed.
 Turn [POWER] key
 System switches
OFF.
over from temp.
 Check valve.
& humidity Humidifier was not fill
 Clean strainer.

●U

HUM. WATER SUPPLY control to ed within specified tim


1805  Check water supply
ALARM temperature-only e from when “WARNI
line.
control. NG” was issued.
 Check the safety
In program mode,
devices of water
system pauses.
purifier.
 System switches
over from temp.  Turn main power
& humidity breaker OFF.
Humidifier breaker trip  Short-circuit or curre

●U

control to  Reset humidifier


HUMIDIFIER FAILURE 1806 ped because of overc nt leakage in humidif
temperature-only breaker inside
urrent. ier circuit
control. electrical
In program mode, compartment.
system pauses.
The humidifier is now
being cleaned or the s
witch was left in OFF
 Operation position after cleaning
"Humidifier ON/OFF s
continued of the humidifier. Note Set the humidifier
witch" installed on hu
CHECK HUMIDIFIER uninterrupted that water supply star ON/OFF switch to
1808 midifier is set to OFF

SWITCH In program mode, ts if this switch is retu "ON" position after


during humidity-only r
system pauses. rned to the ON positio cleaning is complete.
un.
 Humidifier stops. n or the [CLR] key on
the instrumentation s
creen is pressed durin
g humidifier cleaning.

134
Chapter 6 Troubleshooting

Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)

Error
 Turn [POWER] key
Air circulator (AC unit)
OFF.
AIR CIRCULATOR thermal relay tripped Overloaded air circulat
1816 Test system down  Halt operations


FAILURE because of overcurren or motor
shortly to cool down
t.
air circulator motor.
 Turn main power
Heater circuit breaker breaker OFF.
Short-circuit or current
tripped due to an incr  Reset heater
HEATER FAILURE 1817 Test system down leakage in heater circ

ease in the current va breaker inside


uit
lue of the heater. electrical
compartment.
 Turn main power
1818 Output circuit breaker
breaker OFF.
tripped due to an incr Short-circuit, ground-fa
OUTPUT CIRCUIT  Reset the output
Test system down ease in the current va ult or overcurrent of th

TROUBLE circuit breaker


lue of the chamber co e output circuit
inside the electrical
ntrol circuit.
compartment.
 Turn [POWER] key
OFF.
Overheat protector trip  Heat generation from
 Remove heat-
ped because of abnor specimens
OVERHEATING 1819 Test system down generating

mal temperature rise i  Low overheat protect


specimens.
nside chamber. or setting
 Check overheat
protector setting.
 Heat generation from
 Turn [POWER] key
The heater temperatur specimens
OFF.
e switch tripped becau  Heater control error
AIR CONDITIONER  Cool down air
1821 Test system down se of abnormal tempe  Air circulator trouble

OVERHEATING conditioner to max.


rature rise inside air c  Power breaker was s
50°C by forcing air
onditioner. hut off during high t
inside.
emperature testing.
Reverse phase or ope
n phase detected on  Turn main power
POWER PHASE primary side Wrong connection in p breaker OFF.
1822 Test system down

FAILURE (Detected even while t rimary power line  Check power


he chamber is not run supply connections.
ning. )
 Turn [POWER] key
OFF.
Temperature inside ch  Heat generation from
 Remove heat-
ABS. HIGH LIMIT: amber exceeded absol specimens
1832 Test system down generating

TEMP. ute upper limit alarm  Upper limit setting to


specimens.
setting. o low
 Check upper limit
alarm setting.
 Operation  Temporary rise in R  Check upper limit
continues H during step transiti alarm setting.
Humidity inside chamb
uninterrupted. on in program mode  Operation is
er exceeded absolute
ABS. HIGH LIMIT: HUM. 1840 In program mode,  Upper limit setting to restored

upper limit alarm setti


system pauses. o low automatically when
ng.
 Humidifier stops  Refrigerator is in oil humidity drops
operating. return operation. below setting.
 Turn [POWER] key
 Excessive refrigeratio
OFF.
Temperature inside ch n
 Remove
ABS. LOW LIMIT: amber dropped below  Cooling caused by s
1833 Test system down trouble-causing

TEMP. absolute lower limit al pecimens


specimens.
arm setting.  Lower limit setting to
 Check lower limit
o high
alarm setting.
 Temporary drop in R  Check lower limit
 Operation H during step transiti alarm setting.
Humidity inside chamb
continues on in program mode  Operation is
er dropped below abs
ABS. LOW LIMIT: HUM. 1841 uninterrupted.  Lower limit setting to restored

olute lower limit alarm


In program mode, o high automatically when
setting.
system pauses.  Refrigerator is in oil humidity rises
return operation. abouve setting.

135
Chapter 6 Troubleshooting

Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)

Error
 Remove heat-
generating
 Operation specimens.
continues  Heat generation from  Check upper limit
uninterrupted. Temperature inside ch specimens setting and set
UPPER DEV. LIMIT: In program mode, amber exceeded uppe  Upper limit setting to 10°C higher than
1834


TEMP. system pauses. r deviation limit alarm o low set point.
 Heater and setting.  Refrigerator is in oil  Operation is
humidifier stop return operation. restored
operating. automatically when
temperature drops
below setting.
 Check upper limit
 Temporary rise in rel
alarm setting and
ative humidity during
 Operation set 10%rh higher
step transition in pr
continued Humidity inside chamb than humidity set
UPPER DEV. LIMIT: ogram mode
uninterrupted er exceeded upper de point.
HUM. 1842  Relative humidity ala

In program mode, viation limit alarm setti  Operation is


(Option) rm setting is too lo
system pauses. ng. restored
w.
 Humidifier stops. automatically when
 Refrigerator is in oil
humidity drops
return operation.
below setting.
 Chamber door is op  Check door and
en. ventilation fan.
 Ventilation fan is run  Remove
 Operation
Temperature inside ch ning. trouble-causing
LOWER DEV. LIMIT: continues
amber dropped below  Excessive refrigeratio specimens.
TEMP. 1835 uninterrupted.

lower deviation limit al n  Operation is


(Option) In program mode,
arm setting.  Cooling caused by s restored
system pauses.
pecimens automatically when
 Refrigerator is in oil humidity rises
return operation. above setting.
 Check lower limit
alarm setting and
 Temporary drop in R
set 10%rh lower
Operation H during step transiti
Humidity inside chamb than humidity set
LOWER DEV. LIMIT: continues on in program mode
er dropped below low point.
HUM. 1843 uninterrupted.  Lower limit setting to

er limit deviation alar  Operation is


(Option) In program mode, o high
m setting. restored
system pauses.  Refrigerator is in oil
automatically when
return operation.
humidity rises
above setting.
 Loose terminal on th
e temperature contro
Burn-out detected at t
ller unit to which the  Turn [POWER] key
BURN-OUT: AIR CON he air conditioner tem
1848 Test system down temperature sensor OFF.

TEMP SENSOR perature control senso


is connected.  Call for service
r.
 Burn-out of the temp
erature sensor
 Loose terminal on th
e temperature contro
BURN-OUT: AIR Burn-out detected at t ller unit to which the  Turn [POWER] key
CONDITIONER HUM 1849 Test system down he air conditioner hum humidity sensor is OFF.

SENSOR idity control sensor. connected.  Call for service


 Burn-out of the humi
dity sensor
 Loose terminal on th
e temperature contro
Burn-out detected at ller unit to which the  Turn [POWER] key
BURN-OUT (RTD) 1880 Test system down the instrumentation te temperature sensor OFF.

mperature sensor. is connected.  Call for service


 Burn-out of the temp
erature sensor
 Loose terminal on th
e temperature contro
Burn-out detected at t
ller unit to which the  Turn [POWER] key
he air conditioner tem
BURN-OUT (RTD1) 1896 Test system down temperature sensor OFF.

perature control senso


is connected.  Call for service
r.
 Burn-out of the temp
erature sensor

136
Chapter 6 Troubleshooting

Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)

Error
 Loose terminal on th
e temperature contro
Burn-out detected at t
ller unit to which the  Turn [POWER] key
he air conditioner tem
BURN-OUT (RTD2) 1911 Test system down temperature sensor OFF.


perature control senso
is connected.  Call for service
r.
 Burn-out of the temp
erature sensor

System error.
 Turn main power
(Detected even while t
SYSTEM ERROR - Test system down System internal error breaker OFF and

he chamber is not run


ON again.
ning.)

For single
regrigerator
chambers
・Test sysem down  Refrigerator breakdo
wn
REFRIG. □ For multiple  Condenser trouble
2062 Condenser discharge  Overheating
refrigerator  Turn [POWER] key

CONDENSER ERROR
●U

2102 pipe temerature switch  Open-phase


chambers OFF.
2142 or motor protector is
□ Indicates the ・Refrigerator ・Reversed-phase oper  Call for service
2182 activated.
refrigerator No. changeover ation
・Control continues Ambient temp. is too
high
In program mode,
system does not
pause.
REFRIG. □
CONDENSER
2075  Refrigerator breakdo
 Turn [POWER] key

SURFACE Condenser surface te


●U

2115 wn
TEMPERATURE mperature exceeds th OFF.
2155  Condenser trouble
ERROR (HIGH) □ e reference value.  Call for service
2195
indicates the refrigerator
No.
REFRIG □
CONDENSER
SURFACE 2079
 Turn [POWER] key

Condenser surface te
●U

TEMPERATURE 2119  Evaporator is heavily


mperature is below t OFF.
ERROR (LOW) 2159 frosted over.
he reference value.  Perform defrosting.
2199
□ Indicates the
refrigerator No.
REFRIG □
CONDENSER
SURFACE 2077 Disconnection of cond ・Condenser surface te
 Turn [POWER] key

●U

TEMPERATURE 2117 enser surfacetempera mperature sensor of OFF.


DISCONNECTION 2157 ture sensor is detect refrigerator is disconn  Call for service
ERROR 2197 ed. ected.
□ Indicates the
refrigerator No.
REFRIG □
CONDENSER  Refrigerator breakdo
Condenser surface te wn
SURFACE 2073  Turn [POWER] key
 Condenser trouble

mperature is outside t
●U

TEMPERATURE 2113 OFF.


he normal range for l  Gas leak
RANGE ERROR 2153  Cool down the
onger than predetermi  Overheating
2193 compressor.
ned period of time.  Open-phase
□ Indicates the
refrigerator No.
REFRIG □ Solenoid valve for cool
COOLING BYBASS ing bypass breakdown
2076
Cooling bypass tempe ・Thermal expansion v  Turn [POWER] key

TEMPERATURE
●U

2116
ERROR rature exceeded the r alve breakdown OFF.
2156
2196
eference value. ・Feeler bulb contact f  Call for service
□ Indicates the ailure
refrigerator No.  Gas leak

137
Chapter 6 Troubleshooting

Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)

Error
REFRIG □
COOLING BYPASS
TEMPERATURE 2078 Disconnection of cooli ・Cooling bypass temp
 Turn [POWER] key


●U
DISCONNECTION 2118 ng bypass temperatur erature sensor of reg OFF.
ERROR 2158 e sensor of refrigerato rigerator disconnecte  Call for service
2198 r is detected. d.
□ Indicates the
refrigerator No.
Solenoid valve for cool
REFRIG □
ing bypass breakdow
COOLING BYPASS Cooling bypass tempr
2074 n.
 Turn [POWER] key

TEMPERATURE ature exceeds the ran


●U

2114 ・Thermal expansion v


RANGE ERROR ge for longer than the OFF.
2154
predetermined period alve breakdown.  Call for service
2194 ・Feeler bulb connecti
□ Indicates the of time.
refrigerator No. on failure.
 Gas leak

138
Chapter 6 Troubleshooting

Alarms related to optional equipment


BU: Indicates the backup mode is available against the error in
question.
COM. OP: Alarm number of communication function is displayed.
(Optional)
Type of
alarm
Displayed COM. Equipment Probable

Warning
Description Remedy
message OP response cause(s)
Error

・ Excessive  Turn [POWER] key


OFF.
Overcool protector trip refrigeration.
 Remove
OVERCOOLING ped because of abnor ・Cooling caused by s
1820 Test system down trouble-causing

(Option) mal temperature drop i pecimens.


specimens.
nside chamber. ・Overcool protector se  Check overcool
tting too high. protector setting.
 Press reset [BZ.
OFF] key on
Operator safety switch
operation panel and
 Test system inside the chamber w
OPERATOR SAFETY Someone inside the c check inside
down as triggered.
SWITCH TRIP 1824 hamber triggered the chamber.

 Alarm buzzer (Detected even while t


(Option) operator safety switch.  Free personnel
emitted. he chamber is not run
from inside.
ning.)
 Turn [POWER] key
OFF.
 Dehumidifier
 Turn main power
only stops See dehumidifier’s inst

DEHUMIDIFIER Safety device for exter breaker OFF.


●U

operating. ruction manual for err


FAILURE 1810 nal dehumidifier trippe  Eliminate cause of
 Operation or description and cau
(Option) d. error inside
continues se.
dehumidifier.
uninterrupted.
 Dehumidifier
only stops
The dehumidifier modu The dehumidifier mod  Turn [POWER] key

Dehumidifier module
●U

operating.
Failure 1850 le temperature sensor ule temperature senso OFF.
 Operation
(Option) is burnt out. r is burnt out.  Call for service
continues
uninterrupted.
 Turn [POWER] key
A water leak is detect
OFF.
ed by the water leak
WATER LEAK ALARM Moisture adheres to th  Eliminate the cause
1827 Test system down detector. (Detected ev

(Option) e water leak sensor. of the water leak.


en while the test syste
 Dry water leak
m is down.)
sensor.
 Turn [POWER] key
Variable velocity devic
OFF
INVERTER FAILURE e inverter failure is det Breakdown of the inve
1825 Test system down  In case of

(Option) ected on the air circul rter


recurrence, call for
ator fan.
service
EXTERNAL
1856 Error detected in conn  Turn [POWER] key
EQUIPMENT(): See external unit’s ins
1857 ected external unit. OFF.
ALARM truction manual for err
1858 Test system down (Detected even while t  Eliminate cause of

(Option) or description and cau


1859 he chamber is not run error inside external
 indicates the external 1860 ning.)
se.
unit.
equipment No.
 Eliminate cause of
EXTERNAL Warning condition dete error inside external
1864
EQUIPMENT(): cted in connected exte See external unit’s ins unit.
1865
WARNING rnal unit. truction manual for err  Operation is
1866 Test system down

(Option) (Detected even while t or description and cau restored


1867
 indicates external 1868
he chamber is not run se. automatically when
equipment No. ning.) warning state is
cleared.
Emergency stop pushb
 Check why switch
EMERGENCY STOP utton triggered Emergency stop push
was pressed.
ALARM 1826 Test system down (Detected even while t button was pressed fo

 Turn [POWER]
(Option) he chamber is not run r some reason.
key OFF.
ning.)