Chapter 6 Troubleshooting: Warning
Chapter 6 Troubleshooting: Warning
Chapter 6 Troubleshooting: Warning
WARNING
When taking action on the primary side of the
breaker, be sure to turn off the main power supply
switch at your facility. Also, use caution not to
apply voltage accidentally.
Attempting to solve a problem with the power on can
result in electric shock and create a very dangerous
situation.
Use the supplied breaker handle stopper to prevent
the breaker from being turned on accidentally.
Shut the main breaker OFF before attempting to
access the electrical compartment.
124
Chapter 6 Troubleshooting
Alarm screen
Pressing the alarm event displays the following alarm help (details).
ALARM HELP screen
Notice
• If the error or warning buzzer is turned OFF, alarm status
cannot be notified with a buzzer sound, which may lead
to delays in acknowledging an error or warning. Do not
turn OFF the buzzer if at all possible.
If the buzzer is turned OFF, the occurrence of an error or
warning is indicated only by the red blinking pilot lamps and
on the Alarm display screen.
125
Chapter 6 Troubleshooting
Reference The alarm buzzer notifying a "warning/error" can be set using "Set
Sound" on the Configuration screen. This can be accessed from the
Chamber Setup screen.
Corrective action
Take the following action when an alarm is generated. There are 2 types
of alarms -“Error” and “Warning”- and they are dealt with differently in
some respects.
126
Chapter 6 Troubleshooting
127
Chapter 6 Troubleshooting
Date: To display the date and time the error or the warning
occurred.
: To select the previous/next page.
Seek No.: Input the alarm report number to jump to the corresponding
alarm or warning.
128
Chapter 6 Troubleshooting
129
Chapter 6 Troubleshooting
Alarm tables
Alarms are displayed as a pretext to the backup mode kicking-in. If the
mode is OFF when an error occurs, the entire system will shut down.
However, in the program mode, the operation will pause.
The following codes are used in these tables:
(If corrective actions fail to restore normal operation of the chamber,
please contact ESPEC CORP. or the place of purchase .
BU: Indicates the backup mode is available against the error in
question.
COM. OP: Alarm number of communication function is displayed.
(Optional)
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
Remedy on each
Error notification
REFRIGERATOR All refrigerators in error alarm occurrence
2320 Test system down already made for each
●
2056
2209
2219
REFRIG-□ 2096
TEMP ALM: COMP 2249 Compressor temperatu Turn [POWER] key
B
●U
Compressor thermal r wn
●U
HIGH PRESSURE
●U
130
Chapter 6 Troubleshooting
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
2058
2211
2221
2098
REFRIG.-□
2251 Refrigeration circuit lo
Turn [POWER] key
B
LOW PRESSURE
●U
2261 w pressure switch trip Evaporator is heavily
OFF.
2138 ped because of press frosted over.
□ Indicates refrigerator Perform defrosting.
2291 ure drop.
No.
2301
2178
2435
2445
2040
REFRIG □ 2226
BURN-OUT: DISCHG 2080 Burn-out of the refrige
Burn-out of the refriger Turn [POWER] key
B
●U
TEMP 2081
For single rator condenser therm ator condenser therma OFF.
2121
al sensor is detected. l sensor Call for service
□ indicates the 2161 refrigerator
refrigerator No. chambers
2042 Test system
2213 down
2224
REFRIG- 2082 For multiple
BURN-OUT: EVAP IN 2253 refrigerator Burn-out of the refrige
Burn-out of the refriger Turn [POWER] key
B
chambers.
●U
131
Chapter 6 Troubleshooting
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
2048
REFRIG- 2229 Temperature of the re Refrigerator breakdo
OUT-OF-RNG: DISCHG 2088 frigerator discharge pi wn Turn [POWER] key
B
●U
TEMP 2269 pe remains out of nor Condenser trouble OFF.
2128 mal range over the de Gas leak Cool down the
indicates the 2309 termined period of tim Overheating compressor.
refrigerator No. 2168 e. Open-phase
2453
When upper limit dete Turn [POWER] key
cted OFF.
Refrigerator breakdo When upper limit
REFRIG-
Temperature of the re wn detected
OUT-OF-RNG: COND 2049
Condenser failure Check cooling
B
frigerator condenser re
●U
TEMP 2089
mains out of normal r Gas leak water.
2129
ange over the determi Overheating Cool down the
indicates the 2169
ned period of time. Open-phase refrigerator.
refrigerator No.
When lower limit dete When lower limit
cted detected
Frosted cooler Perform defrosting.
2050
2215
2227
REFRIG- 2090 Temperature of the re
OUT-OF-RNG: EVAP IN 2255 frigerator evaporator in
B
●U
2105 rcuit error is detected Breakdown of the ref Turn [POWER] key
2145 on the refrigerator inv rigerator inverter OFF.
indicates the
2185 erter.
refrigerator No.
132
Chapter 6 Troubleshooting
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
REFRIG-
2066 Breakdown of the co
B
INVERTER ERROR T02 A current sensor alar
●U
2106 mpressor Turn [POWER] key
m is detected on the
2146 Breakdown of the ref OFF.
indicates the refrigerator inverter.
2186 rigerator inverter
refrigerator No. For single
REFRIG- refrigerator
2067 B chambers Breakdown of the co
INVERTER ERROR T03 A motor lock error is
●U
2107 Test system mpressor Turn [POWER] key
detected on the refrig
2147 Breakdown of the ref OFF.
indicates the down erator inverter.
2187 rigerator inverter
refrigerator No.
For multiple
REFRIG-
2068 refrigerator Breakdown of the co
B
In program mode,
system does not
pause.
System switches
over from temp.
& humidity Dry-boil temperature s Humidifier water sup
Turn [POWER] key
B
●U
133
Chapter 6 Troubleshooting
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
Operation
continues
uninterrupted.
In program mode,
system does not
Temperature switch in
HUM. SCALE BUILDUP pause. Too much scale inside
1801 side humidifier cylinde Clean humidifier.
●
WARNING * If left unattended cylinder
r tripped.
for 120 hours,
system interprets
trouble as error
and humidifier
stops operating.
System switches
over from temp.
Temperature switch in
& humidity
Turn [POWER] key
B
control to
HUM. DRAIN FAILURE 1803 does not drop when d Clogged drain line OFF.
temperature-only
raining. Humidifier float sens Clean humidifier.
control.
or trouble
In program mode,
system pauses.
Water supply valve
not reopened.
Clogged humidifier st Check valve.
rainer Clean strainer.
Operation
Water leak Check water supply
continues
Humidifier was not fill Low water pressure line.
HUM. WATER SUPPLY uninterrupted.
1804 ed within specified tim Humidifier float sens When water supply
●
134
Chapter 6 Troubleshooting
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
Turn [POWER] key
Air circulator (AC unit)
OFF.
AIR CIRCULATOR thermal relay tripped Overloaded air circulat
1816 Test system down Halt operations
●
FAILURE because of overcurren or motor
shortly to cool down
t.
air circulator motor.
Turn main power
Heater circuit breaker breaker OFF.
Short-circuit or current
tripped due to an incr Reset heater
HEATER FAILURE 1817 Test system down leakage in heater circ
●
135
Chapter 6 Troubleshooting
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
Remove heat-
generating
Operation specimens.
continues Heat generation from Check upper limit
uninterrupted. Temperature inside ch specimens setting and set
UPPER DEV. LIMIT: In program mode, amber exceeded uppe Upper limit setting to 10°C higher than
1834
●
TEMP. system pauses. r deviation limit alarm o low set point.
Heater and setting. Refrigerator is in oil Operation is
humidifier stop return operation. restored
operating. automatically when
temperature drops
below setting.
Check upper limit
Temporary rise in rel
alarm setting and
ative humidity during
Operation set 10%rh higher
step transition in pr
continued Humidity inside chamb than humidity set
UPPER DEV. LIMIT: ogram mode
uninterrupted er exceeded upper de point.
HUM. 1842 Relative humidity ala
●
136
Chapter 6 Troubleshooting
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
Loose terminal on th
e temperature contro
Burn-out detected at t
ller unit to which the Turn [POWER] key
he air conditioner tem
BURN-OUT (RTD2) 1911 Test system down temperature sensor OFF.
●
perature control senso
is connected. Call for service
r.
Burn-out of the temp
erature sensor
System error.
Turn main power
(Detected even while t
SYSTEM ERROR - Test system down System internal error breaker OFF and
●
For single
regrigerator
chambers
・Test sysem down Refrigerator breakdo
wn
REFRIG. □ For multiple Condenser trouble
2062 Condenser discharge Overheating
refrigerator Turn [POWER] key
B
CONDENSER ERROR
●U
2115 wn
TEMPERATURE mperature exceeds th OFF.
2155 Condenser trouble
ERROR (HIGH) □ e reference value. Call for service
2195
indicates the refrigerator
No.
REFRIG □
CONDENSER
SURFACE 2079
Turn [POWER] key
B
Condenser surface te
●U
mperature is outside t
●U
TEMPERATURE
●U
2116
ERROR rature exceeded the r alve breakdown OFF.
2156
2196
eference value. ・Feeler bulb contact f Call for service
□ Indicates the ailure
refrigerator No. Gas leak
137
Chapter 6 Troubleshooting
Type of
alarm
Displayed COM. Equipment Probable
Warning
Description Remedy
message OP response cause(s)
Error
REFRIG □
COOLING BYPASS
TEMPERATURE 2078 Disconnection of cooli ・Cooling bypass temp
Turn [POWER] key
B
●U
DISCONNECTION 2118 ng bypass temperatur erature sensor of reg OFF.
ERROR 2158 e sensor of refrigerato rigerator disconnecte Call for service
2198 r is detected. d.
□ Indicates the
refrigerator No.
Solenoid valve for cool
REFRIG □
ing bypass breakdow
COOLING BYPASS Cooling bypass tempr
2074 n.
Turn [POWER] key
B
138
Chapter 6 Troubleshooting
Warning
Description Remedy
message OP response cause(s)
Error
Dehumidifier module
●U
operating.
Failure 1850 le temperature sensor ule temperature senso OFF.
Operation
(Option) is burnt out. r is burnt out. Call for service
continues
uninterrupted.
Turn [POWER] key
A water leak is detect
OFF.
ed by the water leak
WATER LEAK ALARM Moisture adheres to th Eliminate the cause
1827 Test system down detector. (Detected ev
●
Turn [POWER]
(Option) he chamber is not run r some reason.
key OFF.
ning.)
139