Panel Mount Load Cell Indicator: Operating Manual - English 2.00
Panel Mount Load Cell Indicator: Operating Manual - English 2.00
Panel Mount Load Cell Indicator: Operating Manual - English 2.00
0-10V 5VDC
Analog
0-20mA
Re-Transmission V Excitation
4-20mA
Introduction
The LT1240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.
The high bright 6-digit 14 segment LED displays make for easy setup and readability. A simple menu system with built in
help hints allows for easy configuration of display and load cell settings. The load cell calibration can be done directly from
the load cell calibration certificate or from using known weights / load cell simulator.
A universal mains switch mode power supply (85-264VAC) is provided as standard but an optional low voltage (10-
30VDC) isolated power supply or a high voltage (25-70VDC) isolated power supply can be installed.
The LT1240 contains precision front end circuitry for high accuracy and stability. The ratiometric ADC circuitry
automatically compensates for temperature drift and excitation voltage variances due to cable loss. The load cell
excitation voltage is 5VDC and can interface with both 4 wire and 6 wire load cells. The LT1240 can power up to 6x350Ω
load cells.
RS232 communications is supplied as standard with the MODBUS ™ RTU and MODBUS™ ASCII protocol. A simple
ASCII out protocol is also provided for serial printing and communicating to large displays. A second communication
RS485 interface can be added in conjunction with the standard RS232 interface.
The LT1240 also has an analog out or an isolated analog out option to generate a precise 0/4-20mA or 0-10V analog
output signal.
The LT1240 also includes advanced features such as auto-zero tracking, user input linearisation, max/min recording,
programmable front push buttons, programmable digital inputs, security menu lockout, motion indication, advanced digital
filtering, plus many more to provide a all in one precision load cell indicator.
LT1240 Panel Mount Load Cell Indicator Page 2
1 Features
High bright 6-digit 14 segment LED displays for easy setup and calibration
4 or 6 wire load cell / strain gauge input
Can power up to 6x350Ω load cells at +5VDC excitation voltage
High precision 24 bit ratiometric ADC front end circuitry
-199999 to +999999 display counts
Easy calibration either from the load cell calibration certificate or by using known weights / load cell
simulator
RS232 communications standard (MODBUS™ RTU/ASCII and an Infiniteq ASCII out protocol)
Type 4X, NEMA 4X front panel. 96X48 ABS/Polycarbonate enclosure
Universal mains switch mode power supply (85-264VAC) standard with built in EMI and fuse protection
2x Programmable digital inputs (pull up or pull down field jumper selectable)
3x Programmable front panel push buttons
16 Point lineariser
Auto-zero tracking function
Automatic offset calibration
Tare function
Gross/net function
Selectable/adjustable advanced digital filtering
Motion indication and Net front panel LED status
Up to 4 front panel LED indicators for alarm set point status (Mechanical or solid-state option required)
Maximum/Minimum recording
Built in menu help hints
Field upgradable firmware via the RS232 interface
1 Year Warranty
This instrument is marked with the international hazard symbol. It is important to read this manual before installing or
commissioning your panel meter as it contains important information relating to safety and Electromagnetic Compatibility
EMC.
ENSURE THAT ALL POWER IS SWITCHED OFF TO THE INSTRUMENT BEFORE INSTALLING OR DOING
MAINTENANCE WORK.
Do not place signal and power supply wiring in the same loom.
Make sure that all anti-static precautions are adhered to when handling the circuit boards.
Use screened cable for all signal inputs and attach to earth at one point only.
Use ferrules with all input connections for greater reliability.
The instrument may contain a battery for data retention purposes. The battery should be disposed of
correctly. Please contact your supplier or local council if in doubt.
LT1240 Panel Mount Load Cell Indicator Page 4
2 Specifications
General:
Display 6-Digit, 13.8mm (0.543”) 14 segment high brightness red LED
Display range -199999 to +999999
Display decimal point 0 to 0.00000
Status LEDS 6 LEDs total (SP1 to SP4, Net & Motion)
Digital Inputs 2 Programmable digital inputs
Built in hysteresis, filter and input over voltage protection
Maximum input voltage <30VDC
Input logic is field jumper selectable
(Pull up, sinking inputs) - 10kΩ internal resistor to 5V
(Pull down, sourcing inputs) – 10kΩ internal resistor to common
Active/Non-Active input trigger: <1.9V
Non-Active/Active input trigger: >2.3V
Keypad 4 keys total, 3 programmable keys
Memory storage Non-volatile EEPROM, 100000 write cycles minimum
Warm up time 15 minutes
Power Requirements:
AC Power Supply 85-264VAC, 50/60Hz or 120-370VDC
Isolation: 3000VAC/1min
DC Power Supply, 10-30VDC (Optional) 10-30VDC input
Reverse and over voltage protected
Isolation: >1000V/1min
DC Power Supply, 20-70VDC (Optional) 25-70VDC input
Reverse and over voltage protected
Isolation: >1000V/1min
Power Consumption <10W (Depending on options selected)
Fuse (Built in) 2A Slow Blow (Wickmann 3721200000)
RS components part number 226-6599
Environmental:
Operating temperature -10°C to 50°C (14°F to 122°F)
Storage temperature -40°C to 80°C (-40°F to 176°F)
Operating and storage humidity <85% RH non-condensing
Enclosure:
Overall Dimensions 96x48x112mm (LxHxD) (3.78x1.89x4.41”) (Depth includes
connectors)
Mounting 92x45mm (3.62x1.77”)
Enclosure Material Rear ABS plastic, Front Polycarbonate
Front Facia Rating IP65, with o-ring supplied as standard
Wiring connections Removable terminal blocks
Input:
ADC Resolution 24 bit Delta-sigma, Ratiometric
Input range +-3.5mV/V
Conversion rate 10 updates/second
Filter Moving average digital filter with programmable input step detection
Increment size 1, 2, 5, 10, 20, 50, 100, 200
Input Impedance >100MΩ
CMRR >-110dB
Linearity <0.01% of full scale
Accuracy 0.05% of full scale
Calibration method From the load cell calibration certificate or from using known
weights / load cell simulator
LT1240 Panel Mount Load Cell Indicator Page 5
Communications:
Protocol MODBUS RTU
MODBUS ASCII
ASCII In (Infiniteq Protocol)
ASCII Out (Infiniteq Protocol)
RS232 Communications (Standard) Baud rate: 1200,2400,4800,9600,19200,38400,57600,115200
Data bits: 7 or 8 bits
Parity: Odd, Even or None
Stop bits: 1 or 2 stop bits
Non isolated
RS485 Communications (Optional) Baud rate: 1200,2400,4800,9600,19200,38400,57600,115200
Data bits: 7 or 8 bits
Parity: Odd, Even or None
Stop bits: 1 or 2 stop bits
Internal 120Ω field jumper selectable termination resistor
Max 32 instruments per line
3 Installation
3.1 Dimensions & Front panel layout
Gross / Net
Indication
Alarm Status
Indication
The supplied o-ring must be attached to the front cover to provide sealing between the indicator and the mounting
enclosure. The two supplied fastening metal side clips must be attached to either side as in the diagram below. Do not
over tighten the screws.
LT1240 Panel Mount Load Cell Indicator Page 8
To open the unit simply remove all the connectors from the rear and unscrew the 2 or 4 (Depending if an option board has
been installed) screws and simply slide out the instrument from the enclosure.
The front transparent face needs to be clipped off from the top and bottom circuit boards in order to add or remove the
option circuit board. When reassembling the unit, please make sure that the front push buttons are seated correctly
before clipping the front transparent cover onto the circuit boards.
Make sure full anti-static precautions are adhered to when handling the circuit boards.
Do not apply power to the instrument until the instrument has been carefully placed back in to its enclosure.
-Signal and control cables should be routed as far away as possible from contactors, DC motors etc.
-Never run signal or control cables in the same trunking as AC power lines or high current carrying conductors.
-Cables should be run in metal conduit that is grounded.
-Do not run cable near powerful radio transmitting devices eg. Two way radios.
-Keep cables as short as possible. Long cable runs are more susceptible to EMI then short run cables.
-Switching inductive loads cause high EMI. Use R-C Snubber networks or transient suppression devices across inductive
loads.
-The instrument should be mounted in a grounded metal enclosure.
-Use shielded cables for all connections to the instrument. Some applications could require that one side of the screen is
grounded.
-The use of external EMI suppression devices are recommended in high noise environments.
LT1240 Panel Mount Load Cell Indicator Page 11
A universal mains switch mode power supply (85-264VAC) is provided as standard but an optional low voltage (10-
30VDC) isolated power supply or a high voltage (25-70VDC) isolated power supply can be installed.
The instrument will consume a maximum of 10W with 4 load cells powered, all relays on, mA analog output fully loaded
and all segments illuminated.
WARNING - The instrument is designed for installation in an enclosure which provides adequate protection against
electric shock. Access to power terminals should be restricted to authorised skilled personnel only. Application of supply
voltages higher than those for which the instrument is intended may compromise safety and can cause permanent
damage.
An optional noise suppression device can be ordered. Install these devices as close to the load as possible.
Identify your display with one of the different engineering units. Simply select the appropriate label from the labeling sheet
and apply it to the right hand side of the display as in the diagram below.
LT1240 Panel Mount Load Cell Indicator Page 18
4 Menu System
The menu system can be entered by pressing and holding the menu button for 3 seconds. Use the up , left
, enter and back keys to navigate through the menu system. All the settings are saved in non-volatile
memory when exiting the menu system. The menu system has a 2 minute program timeout. If no key has been pressed
within this period then the instrument will save all settings and return to the normal display mode.
Wait 10 seconds
Example:
Press the “Menu” key for 3 seconds to access the menu system.
LT1240 Panel Mount Load Cell Indicator Page 19
Continue until the value has been set to the desired value.
Press the “Enter” key to accept the value and return to the menu
system.
LT1240 Panel Mount Load Cell Indicator Page 20
Alarm 1 setpoint (Only shown if 1 or more alarm options have been ordered)
Alarm 2 setpoint (Only shown if 2 or more alarm options have been ordered)
Alarm 3 setpoint (Only shown if 3 or more alarm options have been ordered)
Alarm configuration setup menu (Only shown if any of the alarm options have been
ordered)
Analog output setup menu (Only shown if the analog output option has been ordered)
RS485 communications setup menu (Only shown if the RS485 option has been ordered)
Exit menu. Settings are saved on menu exit and the instrument will return to the normal
display mode
Note: The menu system has a 2 minute program timeout. If no key has been pressed within this period
then the instrument will save all settings and return to the normal display mode.
LT1240 Panel Mount Load Cell Indicator Page 22
The alarm setpoints are only shown if any of the alarm options have been ordered.
Use the front panel push buttons to adjust the alarm setpoint value.
Select the engineering display units. The display units are used for the ASCII out printing
and modbus communications.
No unit
Grams
Kilograms
Metric Tonnes
Ounces
Pounds
Imperial Tons
Newtons
Select the display rounding in display counts. The round function rounds the display
value to the nearest rounding increment. Eg. With a rounding setting of 5, a display value
of 233 will be rounded up to 235. A setting of “10” will create a dummy zero. The display
rounding function can be used in conjunction with the digital filter settings to create a
more stable display in noisy environments.
The LT1240 instrument contains an advanced digital filter algorithm. The filter works by filtering small changes between
measurements but will react instantaneously to a large step response. There are 2 settings that are used to setup the
digital filter, namely the filter band and the filter time. The filter band is the threshold in counts that the value must change
by in order for the instrument to recognise it as a step response. The display will jump to this value immediately if a step
response is detected. The filter time is the time in seconds that the input signal will be filtered provided that the input
remains within the filter band setting. The filter is achieved by taking the moving average of the input signal for the filter
time setting.
An increase in filter time leads to a more stable display but with a reduced reaction time. Use the filter time in conjunction
with the filter band and display rounding settings to create a tradeoff between reaction time and display stability.
The diagram below illustrates the use of the filter time and the filter band.
Use the front panel push buttons to enter the filter band.
LT1240 Panel Mount Load Cell Indicator Page 24
Select the filter time. See the paragraph above for an explanation of the filter time.
1 Second
3 Seconds
5 Seconds
10 Seconds
Enter the zero band setting. The auto-zero tracking and the manual gross zero will work
within the limits of this band.
Use the front panel push buttons to enter the zero band.
Select the minimum and maximum assignment. The instrument will use this variable for
the minimum and maximum comparison.
Gross weight.
Net weight.
Lineariser Sub-Menu
For non-linear processes, up to 16 scaling points may be used to provide a piece-wise linear approximation. The greater
the number of points the greater the accuracy. Each point has a real value and a corresponding display value. The real
value is the actual value of the input as it would be with the lineariser feature turned off, the display value is the desired
value.
Note:
If the measured value is above the last actual point then the
lineariser will use the last 2 points to calculate the slope and
similarly is the measured value is below the first actual point
then it will use the first 2 points to calculate the slope.
LT1240 Panel Mount Load Cell Indicator Page 25
Use the front panel push buttons to enter the actual or real
value.
Use the front panel push buttons to enter the display or desired
value.
The auto-zero tracking feature will zero the weight display at regular intervals as long as the measured weight is within
the zero band setting. The display will briefly flash “A.ZERO” when an auto-zero has been performed. When the
instrument restarts, the auto zero correction is lost, but it will start again with a new auto zero correction. Manual zeroing
can also be done via a front push button or via a rear digital input. The auto-zero tracking function can be used to
compensate for zero drift. Zero drift may be caused by changes in the electronics or accumulation of material on the
weight system. The auto-zero band should be set large enough to track normal zero drift, but small enough not to
interfere with normal measuring.
Auto-zero tracking delay time. This defines how often an auto zero is attempted.
Use the front panel push buttons to enter the auto-zero tracking
delay time in seconds.
Select the motion indication band. The display value must remain within the motion band
for 1 second in order for the motion LED to illuminate. Select the value in display counts.
The LT1240 can be calibrated by either using data from the load cell calibration certificate or by using known weights /
load cell simulator.
The following parameters must be entered into the LT1240 from the load cell calibration certificate.
Enter the value in engineering units of the low calibration point of the system. This is
normally zero.
Use the front panel push buttons to enter the low display value.
Enter the value in engineering units of the high calibration point of the system eg.
5000kg. This is normally the full rating value of the load cells.
Use the front panel push buttons to enter the high display value.
Enter the value in mV/V that corresponds to the low display value. This value is normally
found on the load cells calibration certificate. The load cells offset zero error is normally
expressed as uV (microvolts) at 10V excitation. Eg. A load cell zero error with no load is
560uV (0.56mV) at 10V excitation. To convert to mV/V divide by 10 which equals
0.056mV. This value must then be entered into the low mV/V setting.
Use the front panel push buttons to enter the low mV/V value.
Enter the value in mV/V that corresponds to the high display value. This value is normally
found on the load cells calibration certificate. The required value is the rated value of the
load cell in mV/V at full load.
Use the front panel push buttons to enter the high mV/V value.
This allows the user to enter and apply the low load cell calibration weight. The low
weight is normally zero.
Use the front panel push buttons to enter the display value that
corresponds to the low calibration weight.
Apply the low calibration weight to the scale and then press
enter.
The LT1240 will start to average and calculate the mV/V value
that corresponds to the low calibration weight.
Done! The low calibration weight can now be removed. The low
display value and its corresponding mV/V value will be saved
in the “LC CAL” menu settings.
This allows the user to enter and apply the high load cell calibration weight.
Use the front panel push buttons to enter the display value that
corresponds to the high calibration weight.
Apply the high calibration weight to the scale and then press
enter.
The LT1240 will start to average and calculate the mV/V value
that corresponds to the high calibration weight.
Select the alarm assignment. The alarm will use this value to compare against the
set point value.
Gross weight.
Net weight.
Enter the deviation low value. The low value of the band is the set point minus the
deviation low value. This menu option is only shown if the alarm mode is set to deviation.
Use the front panel push buttons to enter the deviation low
alarm.
LT1240 Panel Mount Load Cell Indicator Page 31
Enter the deviation high value. The high value of the band is the set point plus the
deviation high value. This menu option is only shown if the alarm mode is set to
deviation.
Use the front panel push buttons to enter the deviation high
alarm.
Enter the alarm hysteresis value. The hysteresis value is normally used to prevent an
alarm being activated and deactivated when a noisy measurement dithers around the
set point value.
Use the front panel push buttons to enter the hysteresis value.
Enter the alarm on delay value in seconds that the alarm condition must persist before
the alarm is activated.
Use the front panel push buttons to enter the alarm on delay.
Enter the alarm off delay value in seconds that the alarm condition must persist before
the alarm is de-activated.
Use the front panel push buttons to enter the alarm off delay.
The alarm can be set to remain activated even if the alarm condition has gone. When the
alarm condition has gone then the alarm latch can be reset by either a digital input or via
the front push buttons.
The above diagram illustrates the use of a high alarm with hysteresis and on/off delay.
This menu configures the analog output parameters. This menu is only shown if the analog output option has been
ordered.
Gross weight
Net weight
Use the front panel push buttons to enter the display value that
corresponds to the selected analog out low value E.G. 0.000
display counts = 4mA.
Use the front panel push buttons to enter the display value that
corresponds to the selected analog out low value E.G. 20.000
display counts = 20mA.
Enter what must happen to the analog output when an error occurs with the measured
weight. eg. Over-range, Under-Range etc.
The analog output will go to the analog low value when an error
condition occurs.
The analog output will go to the analog high value when an error
condition occurs.
Enter the communication address of the instrument. If more then one instrument is
connected via a multidrop network then the address of each instrument must be
unique. A unique address allows commands to be sent to an individual instrument as well
as it also prevents all the instruments on the bus replying simultaneously.
The ASCII out protocol address range is 0 to 255. The
Modbus address range is 001 to 247. Use the front panel push
buttons to enter the unit address.
1200 Baud.
2400 Baud.
LT1240 Panel Mount Load Cell Indicator Page 35
4800 Baud.
9600 Baud.
19200 Baud.
38.4k Baud.
57.6k Baud.
115.2k Baud.
7 data bits.
8 data bits.
No parity.
Even parity.
Odd parity.
1 Bit.
2 Bits.
Select the source for the communication data. This menu option is only shown if the
ASCII Out mode is selected
Gross weight
Net weight
LT1240 Panel Mount Load Cell Indicator Page 36
This menu option is only shown if the ASCII Out mode is selected.
This menu option enables the date and time to be included in the ASCII output string.
This menu option is only shown if the ASCII Out mode is selected and the RTC option
has been ordered.
The date and time information is not included in the ASCII output
string.
The date and time information is included in the ASCII output
string.
This menu option sets the communication timeout period. If no valid communications is
received within the timeout period then the display will flash to indicate an error. This
menu option is only shown if the ASCII In mode has been selected.
<*> = Decimal 42
<AAA><SPACE> (Only transmitted if address > 0)
<HH:MM:SS><SPACE><DD/MM/YYYY><SPACE> (Optional field if RTC selected) = Time & Date
<G/N><SPACE> = G=Gross, N=Net
<10 digits right justified, leading zero suppression, including decimal point and polarity>
<SPACE> = Decimal 32
<UNIT>=
<CR> = Decimal 13
<LF> = Decimal 10
LT1240 Panel Mount Load Cell Indicator Page 37
“MODBUS over Serial Line Specification and Implementation Guide V1.02” from Modbus-IDA.ORG.
And
“MODBUS Application Protocol Specification V1.1b” from Modbus-IDA.ORG.
Details of the Modbus protocol is described in these documents and is available for free download from the following
website URLs:
http://modbus-ida.org/docs/Modbus_over_serial_line_V1_02.pdf
http://www.infiniteq.co.za/manuals.aspx
Note: Broadcast read commands are ignored by the indicator, only broadcast write commands are processed.
3: Latch Reset
4: Gross/Net
5: Zero
6: Tare
7: Edit SP1
8: Edit SP2
9: Edit SP3
10: Edit SP4
11: Display Hold
182 8 bit unsigned R/W F3 Key Assignment
0: None
1: Min/Max Toggle
2: Min/Max Value Reset
3: Latch Reset
4: Gross/Net
5: Zero
6: Tare
7: Edit SP1
8: Edit SP2
9: Edit SP3
10: Edit SP4
11: Display Hold
This menu configures the front panel function key push buttons. Three of the front panel push buttons can be user
configured for specific functions as listed below.
The function key will toggle the display in the following order. The minimum recorded weight, the
maximum recorded weight and then the current measured weight value. The display will flash
either “MIN” or “MAX” to indicate that the displayed value is either the minimum or maximum
recorded weight.
The function key will reset the minimum and maximum recorded values to the current measured
weight value.
The function key will reset any of the latched alarms when the alarm condition has been
removed. This menu option is only displayed if any of the alarm options have been ordered and
the alarm latch function has been activated.
The function key will toggle the display between showing the gross and net weight. The net
status indicator will illuminate to indicate that the display is showing the net weight.
This function will perform a manual zero within the zero band setting. This is a useful function if
there is some residue material left on the scale.
The function key will tare the gross weight and it will then automatically swap the display to show
the net value. The net status indicator will illuminate to indicate that the display is showing the net
weight.
The function key will allow the user to edit the alarm 1 setpoint value. This menu option is only
displayed if any of the alarm options have been ordered.
The function key will allow the user to edit the alarm 1 setpoint value. This menu option is only
displayed if any of the alarm options have been ordered.
The function key will allow the user to edit the alarm 1 setpoint value. This menu option is only
displayed if any of the alarm options have been ordered.
The function key will allow the user to edit the alarm 1 setpoint value. This menu option is only
displayed if any of the alarm options have been ordered.
The function key will display hold the current measured weight value. The display will flash
“HOLD” to indicate that the displayed value is the display hold value. Press the function key
again to cancel the display hold function.
LT1240 Panel Mount Load Cell Indicator Page 47
This menu configures the two digital inputs. The digital inputs can be configured for specific functions as listed below.
The digital input will toggle the display in the following order. The minimum recorded weight, the
maximum recorded weight and then the current measured weight value. The display will flash
either “MIN” or “MAX” to indicate that the displayed value is either the minimum or maximum
recorded weight.
The digital input will reset the minimum and maximum recorded values to the current measured
weight value.
The digital input will reset any of the latched alarms when the alarm condition has been
removed. This menu option is only displayed if any of the alarm options have been ordered and
the alarm latch function has been activated.
The digital input will toggle the display between showing the gross and net weight. The net
status indicator will illuminate to indicate that the display is showing the net weight.
This function will perform a manual zero within the zero band setting. This is a useful function if
there is some residue material left on the scale.
The digital input will tare the gross weight and it will then automatically swap the display to show
the net value. The net status indicator will illuminate to indicate that the display is showing the net
weight.
The digital input will display hold the current measured weight value. The display will flash
“HOLD” to indicate that the displayed value is the display hold value. Activate the digital input
again to cancel the display hold function.
This menu option is only shown if either the RS232 or RS485 ASCII Out mode is selected. This
digital input allows the user to print the display value either via the RS232 or the RS485
interface. The display will briefly flash “PRINT” when the digital input is activated.
LT1240 Panel Mount Load Cell Indicator Page 48
Select this option if you want to password protect the menu system. Select “NONE” for
no menu protection, “FULL” for all menu options to be password protected, or
“ALM.VAL” for all menu options except the alarm setpoints to be password protected.
If a password has been set and one of the levels for access
control has been selected then the instrument will prompt the
user to enter the password. If the code is correct then it will allow
the user into the menu system otherwise it will return to the
normal display mode.
This menu option allows the user to set the RTC (Real Time Clock). This menu option is
only displayed if the RTC option has been ordered
This menu option will do a display test by turning all the segments on. Press enter or the
menu key to return to the miscellaneous menu option.
LT1240 Panel Mount Load Cell Indicator Page 49
This menu option allows access to technical functions such as input signal and analog
output calibration. These functions are accessed by the factory during the calibration of
the instrument. Please consult the factory for more information.
Use the front panel push buttons to enter the access code.
5 Error Messages
Hardware Under Range:
If the input to the ADC (analog to digital converter) exceeds its negative limit then
the hardware under range message is shown.
If the input to the ADC (analog to digital converter) exceeds its positive limit then the
hardware over range message is shown.
If the display value exceeds the negative display threshold of -199999 then the display
under range message is shown.
Calibration constants CRC error. The instrument could possibly have a non-volatile
memory failure in which case the instrument will then have to be returned to the factory.
Option board CRC error. The instrument has found an error with the top option PCB.
It could possibly be a non-volatile memory failure in which case the instrument will then
have to be returned to the factory.
Analog out calibration CRC error. Please recalibrate the analog out option. If the error
message still persists then it could possibly be a non-volatile memory failure in which
case the instrument will then have to be returned to the factory.
Menu list display error. Please contact the factory with diagnostic information.
7 Firmware Upgrading
The LT1240 Load Cell indicator can be upgraded in the field by connecting the RS232 port to a PC and running the
firmware update program. Note that only the RS232 port can be used to upgrade the firmware.
PC connections:
D9 Female Connector LT1240
Pin 2 Pin 9 (RS232 TXD)
Pin 3 Pin 10 (RS232 RXD)
Pin 5 Pin 11 (GND)
9 Cleaning
The unit should not be cleaned with any abrasive substances. The screen is very sensitive to certain cleaning materials
and should only be cleaned using a clean, damp cloth.
10 Ordering Information
Add option codes to suffix of model number separated by hyphens.
Example:
(LT1240 Weighing indicator with 2 mechanical relays, analog output and an additional RS485 interface)
LT1240-711-730-740
Option part numbers:
700 - Low voltage 10-30VDC isolated power supply
701 - High voltage 25-70VDC isolated power supply
710 - 1 Mechanical relay
711 - 2 Mechanical relays
712 - 3 Mechanical relays
713 - 4 Mechanical relays
720 - 1 Solid-state relay
721 - 2 Solid-state relays
722 - 3 Solid-state relays
723 - 4 Solid-state relays
730 - 16 Bit Analog Output (0/4-20mA, 0-10V)
731 - 16 Bit Isolated Analog Output (0/4-20mA, 0-10V)
740 - Second communication RS485 interface
750 - RTC (Real Time Clock)
760 - Panel mount engineering units
761 - Power connector protective cover
762 - 115VAC Inductive load suppressor
763 - 230VAC Inductive load suppressor
764 - 2A Slow blow replacement fuse
765 - R-C Snubber noise and arc suppressor
766 - Transparent protective front cover
11 Website
An electronic copy of this manual can be downloaded from www.loadtech.co.za.
LT1240 Panel Mount Load Cell Indicator Page 53
12 Notice
Specifications of the products displayed herein are subject to change without notice. Infiniteq cc, or anyone on its behalf,
assumes no responsibility or liability for any errors or inaccuracies.
Information contained herein is intended to provide a product description only. No license, express or implied, by estoppel
or otherwise, to any intellectual property rights is granted by this document. Except as provided in Infiniteq's terms and
conditions of sale for such products, Infiniteq assumes no liability whatsoever, and disclaims any express or implied
warranty, relating to sale and/or use of Infiniteq products including liability or warranties relating to fitness for a particular
purpose, merchantability, or infringement of any patent, copyright, or other intellectual property right.
The products shown herein are not designed for use in medical, life-saving, or life-sustaining applications. Customers
using or selling these products for use in such applications do so at their own risk and agree to fully indemnify Infiniteq for
any damages resulting from such improper use or sale.
13 Warranty
This product carries a warranty for a period of one year from date of purchase against faulty workmanship or defective
materials, provided there is no evidence that the unit has been mishandled or misused. Warranty is limited to the
replacement of faulty components and includes the cost of labor. Shipping costs are for the account of the purchaser.
Note: Product warranty excludes damages caused by unprotected, unsuitable or incorrectly wired
electrical supplies and or sensors, and damage caused by inductive loads.
DISTRIBUTED BY: