Breathing system

Dräger Zeus® / Zeus® IE


Disassembly / Assembly

D-25240-2020
Disassembly
1 Unscrew the CLIC adapter (optional). 2 Insert cleaning disc into the adapter. Close adapter 3 Loosen and pull out the expiration port.
until it snaps.
D-25973-2020

D-25981-2020
D-7416-2014

Screw the CLIC adapter (optional). Remove cleaning disc from the adapter. Close adapter Check the grill in the expiration port*. Insert and
9 until it snaps. 8 tighten the expiration port. 7

4 Loosen and pull out the inspiration port. 5 Remove the flow sensors. 6 Turn the five screws using the hex key
counterclockwise about 90°.
D-25993-2020/D-7442-2014

D-26004-2020
D-25998-2020

91 02 537 | 21.02-2 | HQ | PP | Subject to modifications | © 2021 Drägerwerk AG & Co. KGaA


Check the grill in the inspiration port*. Insert and Insert the flow sensors. Turn the five screws using the hex key clockwise
tighten the inspiration port. 6 5 about 90°. 4

7 Lift off the upper part of the breathing system. 8 Screw off the APL valve. 9 Unscrew the additional air valve.
D-26006-2020

D-26009-2020

D-26014-2020

Place the upper part of the breathing system Align the APL valve correctly and tighten it with the Screw in the additional air valve.
on the lower part. 3 knurled nut. 2 1

Assembly
Reprocessing instruction
**

Disinfection by immersion Machine cleaning and disinfection Sterilization 134 °C (273.2 °F) Visual check
D-7442-2014*

***
D-7442-2014*
D-26241-2020a

D-26240.2020a
D-26240-2020
D-7350-2011

Only use disinfectants listed in the instructions for use of the Position components so that all interior spaces and surfaces Dräger SpiroLife® flow sensors are not suitable for The following parts must be free of deposits:
Dräger SpiroLife® flow sensors. are completely flushed and water can drain off freely. plasma sterilization or radiation sterilization. – Flow grills
– Non-return valves
If required, remove deposits on the valve plates of the
non-return valves with a soft cloth.

* In case the grill is damaged, contact DrägerService®.


** This document does not replace the instructions for use.