Product introduction

1.1 Product introduction


DermaShine® PRO (dermasha third generation) is an innovative digital syringe system (with
suction screw and multiple needles) using
Dermashine inhales. The multi-screw needle consists of 9 needles, making it easier to insert into
hard injection sites. The third generation of Martha inhales through the lift
Strain the injection site to make it easier to inject. The process takes place simultaneously,
making injection surprisingly easy and fast.

Advantages of DermaShine® PRO


The DermaShine® PRO syringe will inject the same amount of medicine at the depth you need.
(0-2 mm) injection points are evenly distributed.
① DDS (Drug Delivery System) medical equipment that efficiently injects drugs into the 1~2mm
injection site.
② After more than 2 years of development process, a safer and more detailed high-quality
water-light machine was produced.
③ 9-pin improves the efficiency of processing. Spend less time & effort.
④ Shorten the recovery time & relieve pain to maximize patient satisfaction.
⑤ The treatment time should not exceed 7 minutes.
⑥ Improve the convenience of the operator through the 8ms fine injection mode. (Obtained a
patent)

1.2 Precautions when using

1. Please read the instructions carefully before use

2. When the drug syringe is connected to the sterile syringe, if the connection is too tight, the
sterile syringe may occur during the injection process

Separation from the medicine syringe or leakage of medicine liquid

3. When using an aspirator, the medicine enters the syringe through the straw due to the loss of
medicine and breeding

Inside, this situation may cause the product to malfunction, please check the hydraulic damper at
any time during or after use.

4. If the syringe is injected upwards, the lost medicine will flow into the device, which may cause
the product to malfunction

5. When unplugging the power cord, please hold the power plug and pull it out. Do not pull the
power cord directly

6. The sound of the end of an injection-in Auto-sensing/Auto-dose/dose


When the drug injection ends in the (dose) mode, the syringe itself will make a sound and stop
the injection. Start from

Auto-Dose/Dose mode until the bell rings, press the foot switch (Foot Switch), the injection will
only be used once

Quantitative drugs. In order to prevent over-injection, the user needs to interrupt the foot switch
before the drug injection is completed.

(Foot Switch), injection

7. It will also stop.

8. According to the Shot setting value reaching 50% of the total number of times, the machine
will emit a beep, and the beep will be heard when the drug injection is over

It will sound and the injection will end.

9. When the pump on is turned on, if the power supply is temporarily interrupted due to a power
failure, in order to notify the situation

For personnel, the machine will beep for 3 minutes and the indicator light will flash for 5 minutes.

10. Do not move the device or disconnect the interface when the machine temporarily stops
operating, and be sure to unplug the power

Then proceed.

1.3 Symbols and Labels

1. Product label pattern


1.4 Machine composition and installation method

No Description Function

1. LED Lamp Warning conditions, please refer to the warning content


below for details

2. LCD Display LCD screen

3. POWER Power on: Press POWER BUTTON

BUTTON (Check the blue boot light)

Shut down: Press POWER BUTTON

4. INJECTOR The part connected to the data line of the handle

POWER
CONNECTOR

5. SUCTION The part connected to the plastic tube of the handle

HOSE

CONNECTOR

6. AIR OUTLET Outlet for suction air

7. UPGRADE Connection part of software upgrade

CONNECTOR

8. FOOT SWITCH The part connected to the pedal cord

CONNECTOR

9. POWER The part connected to the power cord

CONNECTOR

10. Injector Handle (black is data line, plastic tube is negative pressure
line)
(Include Data

Cable &

Suction hose)

2.1 Information

Brand Name - DermaShine® PRO System

Product Name - Infusion pump, electrically-powered


REF - Panace-DS-50

Input : 12V , 1A, BPM040S12F10

Type of protection & Degree against electric Class I, Type BF

Dermashine III Cold Compress

1. Product Introduction

Dermashine Cool is to cool the skin surface through a freezing circuit system, even if there is
strong electricity

The skin is not easy to burn, and the drug penetration ability is better than other instruments. To
make the drug deeper

Even if there is a strong current, it will not feel painful.

1. Apply directly to the skin

2. Does not irritate the skin and damage the skin at the same time deeply replenish nutrients

3. 80-90% high efficiency drug penetration ability

4. Can use polymer functional cosmetics

5. Form the best skin temperature to allow the drug to penetrate into the dermis

2. The effect of De Martha III cold compress

1) Efficient medicine injection

2) Various effects such as skin lifting, calming, whitening, and wrinkle improvement
Injection effect of other Machine Injection effect of the Derma shine machine

Drug penetration rate is 15-20% Drug penetration rate is 80-90%

Unequal penetration of drugs Drugs are equally penetrated

3. How to use

1) Connect with the power cord

2) Press for 1~2 seconds to turn on and off, the LED light will automatically turn on
after a sound

3) Use , to select the mode and press , and use , to adjust the intensity
after starting the mode.

4) If you need to change the mode, press . After the LED is turned on , the mode can be
changed again.
MODE 1 2 3 4 HOT COOL

EFFECT Nutrition Lift Whitening Wrinkle Clean Calm

Mode 1 Nutrition Mode

Mode 2 Flexible Mode

Mode 3 Whitening Mode

Mode 4 Wrinkle Mode

Mode 5 Hot Mode

Mode 6 Hot Mode


No. Description

1 Adjust the suction intensity

2 Adjust the syringe head (up and down) on the handle

3 Save the settings, (1~5)

4 For specific settings, refer to the content below

5 Set the language (Chinese is available)

6 Start operation (preparation for negative pressure)

2.2 Set the main page

2.3 Adjust the syringe head on the handle (up and down)
After installing the syringe (medicine) on the handle, adjust the syringe head of the handle.

The syringe (medicine) touches the injection start on the handle

2.4 Set injection mode


Select injection mode 1. Auto-Sensing mode 2. Auto-Dose mode 3.

Dose-mode 4. Continuous-mode (Normal &


Slow)

Auto-Sensing mode Auto-Sensing mode is automatic injection


mode, the needle is put into the skin

When entering, the machine automatically


turns on the negative pressure system and
automatically injects medicines.

After the medicine is injected, the machine


automatically disappears the negative
pressure.

Auto-Dose mode Auto-Dose mode

The machine repeatedly injects the set


amount of medicine when the foot pedal is
pressed.

The injection speed can be adjusted on the


host.

(When pressing the pedal, the medicine is


injected repeatedly)
Dose-mode Dose-mode is the amount of medicine that
has been injected when the pedal is pressed

(Press the foot pedal once to inject medicine


once).

Continuous mode (Normal & slow) Continuous mode (Normal & slow) is the
positive

Inject the medicine repeatedly at a constant


speed or a slower speed.

(Normal=Normal toxin, Slow=Slower speed)

2.5 Choose the type of syringe

After clicking "Syringe Type", select the type of syringe (medicine). If the prepared medicine is 3
ml

Then, please choose 3 ml.

No. Description Function

1. 1 ㎖/1 ㎖ The syringe (medicine) for dispensing is 1 ml. The total


number of injections according to the setting content

Automatic decision.
2. (Hi)2 ㎖/3 ㎖ There is 3 ml of medicine in a 2 ml syringe. Total number of
injections by setting content

Determined automatically.

3. (Hi)2.5 ㎖/3 ㎖ There is 3 ml of medicine in a 2.5 ml syringe. Total injection


times by setting content

The number is determined automatically.

4. (Hi)3 ㎖/3 ㎖ There is 3 ml of medicine in a 3 ml syringe. Total number of


injections by setting content

Determined automatically.

5. 3 ㎖/3 ㎖ There is 3 ml of medicine in a 3 ml syringe. Total number of


injections by setting content

Determined automatically.

6. 4 ㎖/5 ㎖ There is 4 ml of medicine in a 5 ml syringe. Total number of


injections by setting content

Determined automatically.

7. 5 ㎖/5 ㎖ There is 4 ml of medicine in a 5 ml syringe. Total number of


injections by setting content

Determined automatically.

2.6 Dose / Shot-One shot volume setting

After pressing the Onedose/shot button, the following prompt window will appear. After setting
the setting value button below, press save and close
Close the key to complete the setting

* With the change of the set value, the Syringe and volume values in the main screen will also
automatically display and change.

* With the change of the set value, the injection value in the main screen will automatically
display and change.

*Note: If the user arbitrarily modify the injection volume during use, the number of injections
and the set value of injection volume will also be changed accordingly

Change, so be sure to confirm the residual amount before performing surgery. And with the
arbitrary change of the initial shot value calculated by the program, 50% of the total number of
times reaches the alarm line, and the alarm prompt at the end (as shown in the figure) will also
change.
2.7 Adjust flow

Click "Flow Rate" and a window will appear.

2.8 Set the total number of injections

The total number of injections can be selected from 10 to 140.

2.9 Adjust suction intensity

Press Power Level, Up / Down appears. When you press the Power level button, it will pop up like

The window shown below. Press the button to set the value down and then press the save and
close button to complete the setting. Depending on the injection site

Health needs to adjust the standard of inhalation device, and then perform surgery.

3. Operation method

3.1 Before use

① Sterilization gloves

② Connect the syringe (prepared) to the needle

3.2 Operation method and sequence

① Turn on

② Please confirm the normal up and down movement of the injection head
③ Install the syringe with the needle connected to the handle

④ Confirm that the needle is installed normally

⑤ Confirm the installation of the syringe and the buckle


⑦ After the syringe is installed, the syringe and the syringe head of the handle touch

⑧ Set up on the main page

⑨ Click "SUCTION READY"

⑩ Confirm the settings on the screen before operation

⑪ Inject with the foot pedal

In Auto-Sensing mode, there is no need to use a foot switch (Foot Switch).

Inject according to each injection volume

3.3 Storage method after use...

1. Do not place this product in a place with a strong magnetic field

2. Do not place in a place with high temperature or high humidity

3. Do not place it in a high place or a place easily broken

4. In the event of a malfunction, please stop using it immediately and contact the experts

4. Repair Repair

4.1Product warranty period

The product warranty period starts from the product purchase date. If you are not sure about the
purchase date, please consult Gu

Customer Service Center

The product warranty period is for the warranty period of each product, and for the product
under normal conditions

Free repairs for defects in quality, performance, and functionality

The warranty period of the product is effective from the day the product is purchased. This
certificate can be cleared through the purchase receipt or transaction.

Single to confirm. If the transaction list (or other supporting materials) is lost, it will no longer be
covered by the warranty

Inside.

Free warranty period: 1 year from the day of purchase, 6 months for other accessories (data
cables, etc.)

Free after-sales service processing

During the free warranty period, the product has defects in quality, performance and efficacy
under normal use

question. Accept after-sales service due to the same problem within one year from the date of
acceptance of the free warranty

The free warranty period for accessories (such as data cables, dampers, etc.) is 6 months

5. Warning

5.1 Warning

Description Situation Approach

Power This will appear when the respirator Press the switch (yellow LED will
is turned off while running remain for 5 minutes)
Shut down
Warning (yellow LED)

Encoder error Encoder problem 1. Pull out the handle connection part

2. Continue to adjust the injection


mode

3. Press the foot pedal

4. After confirming the "Encoder error"


text, click "OK"

Over Current Excessive current is injected into the 1. The syringe head on the handle
motive when running the motive moves back
error
2. Fix the syringe head

3. Change the injection mode to the


dosage mode

4. Press the pedal to make an injection

5. Confirm the "Over Current Error"


text and click "OK".

50% Continuous warning sound (2 times). N/A


completion 50% after injection

beep sound

100% Continuous warning sound (2 times). N/A


After the total time injection is
completion
completed
beep sound
One time Warning sound (1 time). After one N/A
injection is completed
Injection

beep sound

Touch Warning sound (1 time). N/A

the displayed Adjust screen

on LCD Menu

6. After sales service

1. Warranty for main machine is 1 year, 3 months for accessories and life-long maintenance!

Within warranty, any problems, free repair.

After warranty, we can help repair, and there will be some cost charges.

2. Following personal reasons, not provide free warranty service:

A, unauthorized disassemble.

B, Drop on the ground by personal reason.

C, Lack of reasonable maintenance .

D, Not operate as per user manual

3. The company has right to change the product design & specifications without notice, if in
doubt, please contact the company.