Edited by J.W. Munch (1995)
Edited by J.W. Munch (1995)
Revision 1.1
506-1
METHOD 506
1.1 This method describes a procedure for the determination of certain phthalate and
adipate esters in drinking water by liquid/liquid or liquid/solid extraction. The
following compounds can be determined by this method:
1.2 This is a capillary column gas chromatographic (GC) method applicable to the
determination of the compounds listed above in ground water and finished
drinking water. When this method is used to analyze unfamiliar samples for any
or all of the compounds listed above, compound identifications should be
supported by at least one additional qualitative technique. Method 525.2 provides
gas chromatograph/mass spectrometer (GC/MS) conditions appropriate for the
qualitative and quantitative confirmation of results for all the analytes listed
above, using the extract produced by this method.
1.3 This method has been validated in a single laboratory, and method detection
limits (MDLs)1 have been determined for the analytes above (Table 2). Observed
detection limits may vary among waters, depending upon the nature of
interferences in the sample matrix and the specific instrumentation used.
506-2
2.0 SUMMARY OF METHOD
3.0 DEFINITIONS
3.1 Laboratory Reagent Blank (LRB) -- An aliquot of reagent water that is treated
exactly as a sample including exposure to all glassware, equipment, solvents,
reagents, internal standards, and surrogates that are used with other samples.
The LRB is used to determine if method analytes or other interferences are
present in the laboratory environment, the reagents, or the apparatus.
3.2 Field Reagent Blank (FRB) -- Reagent water placed in a sample container in the
laboratory and treated as a sample in all respects, including exposure to sampling
site conditions, storage, preservation and all analytical procedures. The purpose
of the FRB is to determine if method analytes or other interferences are present
in the field environment.
3.3 Laboratory Fortified Blank (LFB) -- An aliquot of reagent water to which known
quantities of the method analytes are added in the laboratory. The LFB is
analyzed exactly like a sample, and its purpose is to determine whether the
methodology is in control, and whether the laboratory is capable of making
accurate and precise measurements at the required method detection limit.
506-3
3.5 Stock Standard Solution -- A concentrated solution containing a single certified
standard that is a method analyte, or a concentrated solution of a single analyte
prepared in the laboratory with an assayed reference compound. Stock standard
solutions are used to prepare primary dilution standards.
3.7 Calibration Standard (CAL) -- A solution prepared from the primary dilution
standard solution and stock standard solutions of the internal standards and
surrogate analytes. The CAL solutions are used to calibrate the instrument
response with respect to analyte concentration.
3.8 Quality Control Sample (QCS) -- A sample matrix containing method analytes or
a solution of method analytes in a water miscible solvent which is used to fortify
reagent water or environmental samples. The QCS is obtained from a source
external to the laboratory, and is used to check laboratory performance with
externally prepared test materials.
4.0 INTERFERENCES
4.1.3 The use of high purity reagents and solvents helps to minimize
interference problems. Purification of solvents by distillation in an all
glass system may be required.
506-4
Warning: When a solvent is purified, stabilizers added by the
manufacturer are removed thus potentially making the solvent hazardous.
Also, when a solvent is purified, preservatives added by the manufacturer
are removed thus potentially reducing the shelf-life.
4.2 Matrix interferences may be caused by contaminants that are co-extracted from
the sample. The extent of matrix interferences will vary from source to source,
dependent upon the nature and diversity of the samples. Clean up procedures
can be used to overcome many of these interferences.
4.3 It is important that samples and working standards be contained in the same
solvent. The solvent for working standards must be the same as the final solvent
used in sample preparation. If this is not the case, chromatographic comparability
of standards to sample may be affected.
5.0 SAFETY
5.1 The toxicity or carcinogenicity of each reagent used in this method has not been
precisely defined; however, each chemical compound must be treated as a
potential health hazard. Accordingly, exposure to these chemicals must be
reduced to the lowest possible level. The laboratory is responsible for
maintaining a current awareness file of OSHA regulations regarding the safe
handling of the chemicals specified in this method. A reference file of material
safety data sheets should also be made available to all personnel involved in the
chemical analysis. Additional references to laboratory safety are available and
have been identified5-7 for the information of the analyst.
6.0 EQUIPMENT AND SUPPLIES (All specifications are suggested, catalog numbers are
included for illustration only.)
6.1.1 Grab Sample Bottle -- 1 L or 1 qt amber glass, fitted with a screw cap
lined with Teflon. Protect samples from light if amber bottles are not
available. The bottle and cap liner must be washed, rinsed with acetone
or methylene chloride and dried before use in order to minimize
contamination (see Section 4.1.1.).
6.2 Glassware
506-5
6.2.4 Evaporative Flask -- Kuderna-Danish, 500 mL, attach to concentrator tube
with springs.
6.2.7 Vials -- 10-15 mL, amber glass with Teflon-lined screw cap.
6.2.8 Boiling Chips -- Approximately 10/40 mesh. Heat to 400°C for 30 minutes
or extract with methylene chloride in a Soxhlet apparatus.
6.3 Water Bath -- Heated (with concentric ring covers) capable of temperature control
(±2°C). The water bath should be used in a ventilating hood.
6.5.4 An automatic injector system is suggested, but was not used for the
development of this method.
506-6
7.1 Reagent Water -- Reagent water is defined as water in which an interfering
substance is not observed at the MDL of the parameters of interest. Reagent
water used to generate data in this method was distilled water obtained from the
Millipore L/A-7044 system comprised of prefiltration, organic adsorption,
deionization and Millipore filtration columnar units. Any system may be used
if it generates acceptable reagent water.
7.2 Acetone, Hexane, Methylene Chloride, Ethyl Acetate, Ethyl Ether and Iso-Octane -
- Pesticide quality or equivalent to distillation in glass quality.
7.3 Sodium Sulfate, ACS Granular, Anhydrous -- Several levels of purification may
be required in order to reduce background phthalate levels towards acceptance:
1) Heat four hours at 400°C in a shallow tray and 2) Soxhlet extract with
methylene chloride for 48 hours.
7.4 Florisil, PR Grade (60/100 Mesh) -- To prepare for use, place 100 g of Florisil into
a 500 mL beaker and heat for approximately 16 hours at 40°C. After heating
transfer to a 500 mL reagent bottle. Tightly seal and cool to room temperature.
When cool, add 3 mL of reagent water. Mix thoroughly by shaking or rolling for
10 minutes and let it stand for at least two hours. Store in the dark in glass
containers with ground glass stoppers or foil-lined screw caps.
7.5 Sodium Chloride, ACS Granular -- Heat four hours at 400°C in a shallow tray.
When cool, keep in tightly sealed glass (not plastic) bottle. This cleaning step is
required to minimize background contamination associated with this reagent.
7.8 Alumina, Neutral Activity Super I, W200 Series (ICN Life Sciences Group,
No. 404583) -- To prepare for use, place 100 g of alumina into a 500 mL beaker
and heat for approximately 16 hours at 400°C. After heating transfer to a 500 mL
reagent bottle. Tightly seal and cool to room temperature. When cool, add 3 mL
of reagent water. Mix thoroughly by shaking or rolling for 10 minutes and let it
stand for at least two hours. Keep the bottle sealed tightly.
506-7
7.10 Liquid-Solid Extraction Disks, C-18, 47 mm -- Disks are manufactured with Teflon
or other inert support and should contain very little contamination.
7.12 Stock Standard Solutions (SSS) (1.00 µg/µL) -- Stock standard solutions can be
prepared from pure standard materials or purchased as certified solutions.
7.12.2 Transfer the stock standard solutions into Teflon-sealed screw-cap bottles.
Store at 4°C and protect from light. Stock standard solutions should be
checked frequently for signs of degradation or evaporation, especially just
prior to preparing calibration standards from them.
7.12.3 Stock standard solutions must be replaced after six months, or sooner if
comparison with check standards indicates a problem. Butylbenzyl
phthalate is especially vulnerable to autoxidation.
8.1 Grab samples must be collected in amber glass containers (Section 6.1).
Conventional sampling practices should be followed8,9 ; however, the bottle must
not be prerinsed with sample before collection.
8.2.2 After the sample is collected, seal the bottle and swirl the sample until the
thiosulfate is dissolved.
8.2.3 The samples must be iced or refrigerated at 4°C free from light from the
time of collection until extraction. Limited holding studies have indicated
that the analytes thus stored are stable up to 14 days or longer. Analyte
stability may be affected by the matrix; therefore, the analyst should verify
that the preservation technique is applicable to the particular samples
under study.
506-8
8.3 Extract Storage -- Extracts should be stored at 4°C in absence of light. A 14-day
maximum extract storage time is recommended. The analyst should verify
appropriate extract holding times applicable to the samples under study.
9.2 Laboratory Reagent Blanks (LRB) -- Before processing any samples, the analyst
must demonstrate that all glassware and reagent interferences are under control.
Each time a set of samples is extracted or reagents are changed, a LRB must be
analyzed. If within the retention time window of any analyte of interest the LRB
produces a peak that would prevent the determination of that analyte using a
known standard, determine the source of contamination and eliminate the
interference before processing samples.
9.3.2 For each analyte the mean recovery value for these samples must fall in
the range of R ±30% using the values for R for reagent water in Tables 3
or 4. The precision of these measurements, expressed as RSD, must be
20% or less. For those compounds that meet the acceptance criteria,
performance is considered acceptable. For those compounds that fail these
criteria, this procedure must be repeated using fresh replicate samples
until satisfactory performance has been demonstrated.
9.3.3 For each analyte, determine the MDL. Prepare a minimum of seven LFBs
at a low concentration. Fortification concentration in Table 2 may be used
as a guide, or use calibration data obtained in Section 10.0 to estimate a
concentration for each analyte that will produce a peak with a three to five
times signal to noise response. Extract and analyze each replicate
according to Sections 11.0 and 12.0. It is recommended that these LFBs be
prepared and analyzed over a period of several days, so that day to day
variations are reflected in precision measurements. Calculate mean
recovery and standard deviation for each analyte. Use the standard
deviation and the equation given in Section 13.0 to calculate the MDL.
506-9
9.3.4 The initial demonstration of capability is used primarily to preclude a
laboratory from analyzing unknown samples via a new, unfamiliar
method prior to obtaining some experience with it. It is expected that as
laboratory personnel gain experience with this method the quality of data
will improve beyond those required here.
9.5.1 The laboratory must analyze at least one LFB sample with every twenty
samples or one per sample set (all samples extracted within a 24-hour
period) whichever is greater. Ideally, the fortified concentration of each
analyte in the LFB should be the same concentration selected in
Section 9.3.1. Calculate accuracy as percent recovery (Xi). If the recovery
of any analyte falls outside the control limits (see Section 9.5.2), that
analyte is judged out of control, and the source of the problem should be
identified and resolved before continuing analyses.
9.5.2 Until sufficient data become available from within their own laboratory,
usually a minimum of results from 20-30 analyses, the laboratory should
assess laboratory performance against the control limits in Section 9.3.2
that are derived from the data in Table 2. When sufficient internal
performance data becomes available, develop control limits from the mean
percent recovery ( ) and standard deviation (S) of the percent recovery.
These data are used to establish upper and lower control limits as follows:
9.6.1 The laboratory must fortify each analyte to a minimum of 10% of the
routine samples or one fortified sample per set, whichever is greater. The
fortified concentration should not be less than the background
concentration of the sample selected for fortifying. Ideally, this
concentration should be the same as that used for the laboratory fortified
blank (Section 9.5). Over time, samples from all routine sample sources
should be fortified.
506-10
9.6.2 Calculate the accuracy as percent recovery, R, for each analyte, corrected
for background concentrations measured in the unfortified sample. For
each analyte the mean recovery value for these samples must fall in the
range of R ±35% using the values for R for reagent water in Tables 3 or
4.
9.6.3 If the recovery of any such analyte falls outside the designated range, and
the laboratory performance for that analyte is shown to be in control
(Section 9.5), the recovery problem encountered with the dosed sample is
judged to be matrix related, not system related. The result for that analyte
in the unfortified sample is labeled suspect/matrix to inform the data user
that the results are suspect due to matrix effects.
9.7 Quality Control Samples (QCS) -- Each quarter, the laboratory should analyze one
or more QCS (if available). If criteria provided with the QCS are not met,
corrective action should be taken and documented.
9.8 The laboratory may adopt additional quality control practices for use with this
method. The specific practices that are most productive depend upon the needs
of the laboratory and the nature of the samples. For example, field or laboratory
duplicates may be analyzed to assess the precision of the environmental
measurements.
506-11
10.2.2 Using injections of 1-2 µL, analyze each calibration standard according to
Section 11.5 and tabulate peak height or area responses against the mass
injected. The results can be used to prepare a calibration curve for each
compound. Alternatively, if the ratio of response to amount injected
(calibration factor) is a constant over the working range (<20% relative
standard deviation, RSD), linearity through the origin can be assumed and
the average ratio or calibration factor can be used in place of a calibration
curve.
11.0 PROCEDURE
11.1.1 Mark the water meniscus on the side of the sample bottle for later
determination of sample volume. Pour the entire sample into a 2 L
separatory funnel containing 50 g of NaCl.
11.1.2 Add 60 mL CH2Cl2, to the sample bottle. Seal, and shake gently to rinse
the inner walls of the bottle. Transfer the solvent to the separatory funnel.
Extract the sample by shaking the funnel for two minutes with initial and
periodic venting to release excess pressure. Allow the organic layer to
separate for a minimum of 10 minutes from the water phase. If the
emulsion interface between layers is more than one-third the volume of
the solvent layer, the analyst must employ mechanical techniques to
complete the phase separation. The optimum technique depends upon the
sample, but may include stirring, filtration of the emulsion through glass
wool, centrifugation, or other physical methods. Collect the solvent
extract in a 250 mL Erlenmeyer flask.
506-12
Then extract with 40 mL of hexane, which extract (top phase) is added to
the total.
11.1.5 Pour the combined extract through a drying column (Section 6.2.2)
containing about l0 cm of prerinsed anhydrous sodium sulfate, and collect
the extract in the K-D concentrator. Rinse the Erlenmeyer flask and
column with 20-30 mL of methylene chloride to complete the quantitative
transfer.
11.1.6 Add one or two clean boiling chips to the evaporative flask and attach a
three-ball Snyder column. Prewet the Snyder column by adding about
l mL of methylene chloride to the top. Place the K-D apparatus on a hot
water bath (60-65°C) so that the concentrator tube is partially immersed
in the hot water, and the entire lower rounded surface of the flask is
bathed with hot vapor. Adjust the vertical position of the apparatus and
the water temperature as required to complete the concentration in
40 minutes. At the proper rate of distillation the balls of the column will
actively chatter but the chambers will not flood with condensed solvent.
When the apparent volume of liquid reaches approximately 7 mL, remove
the K-D apparatus and allow it to drain and cool for at least l0 minutes.
11.1.7 Increase the temperature of the hot water bath to about 85°C. Remove the
Snyder column, rinse the column and the 500 mL evaporative flask with
1-2 mL of hexane. Replace with a micro column and evaporative flask.
Concentrate the extract as in Section 11.1.6 to 0.5-1 mL. The elapsed time
of concentration should be approximately 15 minutes.
11.1.8 Remove the micro Snyder column and rinse the column by flushing with
hexane using a 5 mL syringe. Concentrate to a volume of 1 mL by
purging the liquid surface with a gentle flow of nitrogen or clean air. If
an autosampler is to be used, transfer the extract to an autosampler vial
with a Pasteur pipet. Seal the vial with a threaded or crimp top cap.
Store in refrigerator if further processing will not be performed. If the
sample extract requires no further cleanup, proceed with gas
chromatographic analysis (Section 11.5). If the sample requires further
cleanup, proceed to Section 11.4.
11.1.9 Determine the original sample volume by refilling the sample bottle to the
mark and transferring the liquid to a 1000 mL graduated cylinder. Record
the sample volume to the nearest 5 mL.
506-13
11.2.1 This method is applicable to a wide range of organic compounds that are
efficiently partitioned from the water sample onto a C18 organic phase
chemically bonded to a solid inorganic matrix, and are sufficiently volatile
and thermally stable for gas chromatography 10. See Section 11.3 for the
disk option procedure. Particulate bound organic matter will not be
partitioned, and more than trace levels of particulates in the water may
disrupt the partitioning process. Single laboratory accuracy and precision
data have been determined at a single concentration for the analytes listed
in Section 1.1 fortified into reagent water and raw source water.
11.2.3 Mark the water meniscus on the side of the sample bottle (approximately
1 L) for later determination of sample volume. Pour the water sample
into the 2 L separatory funnel with the stopcock closed.
11.2.5 Periodically open the stopcock and drain a portion of the sample water
into the solvent reservoir. The water sample will drain into the cartridge,
and from the exit into the suction flask. Maintain the packing material in
the cartridge immersed in water at all times. After all of the sample has
passed through the LSE cartridge, wash the separatory funnel and
cartridge with 10 mL of reagent water, and draw air through the cartridge
for about 10 minutes.
11.2.6 Transfer the 125 mL solvent reservoir and LSE cartridge (from Figure 1A)
to the elution apparatus (Figure 1B). The same 125 mL solvent reservoir
is used for both apparatus. Wash the 2 L separatory funnel with 5 mL of
methylene chloride and collect the washings. Close the stopcock on the
100 mL separatory funnel of the elution apparatus, add the washings to
506-14
the reservoir and enough additional methylene chloride to bring the
volume back up to 5 mL and elute the LSE cartridge. Elute the LSE
cartridge with an additional 5 mL of methylene chloride (10 mL total). A
small amount of nitrogen positive pressure may be used to elute the
cartridge. Small amounts of residual water from the LSE cartridge will
form an immiscible layer with the methylene chloride in the 100 mL
separatory funnel. Open the stopcock and allow the methylene chloride
to pass through the drying column packed with anhydrous sodium sulfate
(1 in.) and into the collection vial. Do not allow the water layer to enter
the drying column. Remove the 100 mL separatory funnel and wash the
drying column with 2 mL of methylene chloride. Add this to the extract.
Concentrate the extract to 1 mL under a gentle stream of nitrogen. The
extract is now ready for gas chromatography (Section 11.4) or additional
cleanup (Section 11.3).
11.3.1.1 Insert the disk into the 47 mm filter apparatus. Wash the
disk with 5 mL methylene chloride (MeCl2) by adding the
MeCl2 to the disk, pulling about half through the disk and
allowing it to soak the disk for about a minute, then
pulling the remaining MeCl2 through the disk. With the
vacuum on, pull air through the disk for a minute.
11.3.3 Add the water sample to the reservoir and turn on the vacuum to begin
the filtration. Full aspirator vacuum may be used. Particulate-free water
may filter in as little as 10 minutes or less. Filter the entire sample,
draining as much water from the sample container as possible.
11.3.4 Remove the filtration top from the vacuum flask, but don't disassemble
the reservoir and fritted base. Empty the water from the flask and insert
506-15
a suitable sample tube to contain the eluant. The only constraint on the
sample tube is that it fit around the drip tip of the fritted base.
Reassemble the apparatus.
Repeat the above step twice, using MeCl 2 instead of CH3 CN. Pour the
combined eluates thru a small funnel with filter paper containing 3 grams
of anhydrous sodium sulfate. Rinse the test tube and sodium sulfate with
two 5 mL portions of MeCl2. Collect the filtrate in a concentrator tube.
11.3.5 With the concentrator tube in a 28°C heating block, evaporate the eluate
with a stream of N2 to 0.5 mL.
11.4 Extract Cleanup -- Cleanup procedures may not be necessary for a relatively clean
sample matrix, such as most drinking waters. If particular circumstances demand
the use of a cleanup procedure, the analyst may use either procedure below or
any other appropriate procedure. However, the analyst first must demonstrate
that the requirements of Section 9.0 can be met using the method as revised to
incorporate the cleanup procedure.
11.4.1.3 Next, elute the column with l00 mL of 20% ethyl ether in
hexane (V/V) into a 500 mL K-D flask equipped with a
l0 mL concentrator tube. Elute the column at a rate of
about 2 mL/min. for all fractions. Concentrate the
collected fraction as in Section 11.1. No solvent exchange
is necessary. Adjust the volume of the cleaned extract to
506-16
l mL in the concentrator tube and analyze by gas
chromatography.
11.4.2.2 Preelute the column with 40 mL of hexane. The rate for all
elutions should be about 2 mL/min. Discard the eluate
and just prior to exposure of the sodium sulfate layer to the
air, quantitatively transfer the sample extract (Section 11.1.8
or 11.2.6) onto the column, using an additional 2 mL of
hexane to complete the transfer. Just prior to exposure of
the sodium sulfate layer to the air, add 35 mL of hexane
and continue the elution of the column. Discard this
hexane eluate.
11.4.2.3 Next, elute the column with l40 mL of 20% ethyl ether in
hexane (V/V) into a 500 mL K-D flask equipped with a
l0 mL concentrator tube. Concentrate the collected fraction
as in Section 11.1. No solvent exchange is necessary.
Adjust the volume of the cleaned extract to l mL in the
concentrator tube and analyze by gas chromatography.
11.5.1 Table 1 summarizes the recommended operating conditions for the gas
chromatograph. Included are retention data for the primary and
confirmation columns. Other capillary columns, chromatographic
conditions may be used if the requirements of Section 9.0 are met.
11.5.3 Inject 1-2 µL of the sample extract or standard into the gas
chromatograph. Smaller (1.0 µL) volumes may be injected if automatic
devices are employed. For optimum reproducibility, an autoinjector is
recommended.
11.5.4 Identify the analytes in the sample by comparing the retention times of the
peaks in the sample chromatogram with those of the peaks in standard
chromatograms. The width of the retention time window used to make
identifications should be based upon measurements of actual retention
time variations of standards over the course of a day. Three times the
standard deviation of a retention time for a compound can be used to
calculate a suggested window size; however, the experience of the analyst
should weigh heavily in the interpretation of chromatograms.
506-17
11.5.5 If the response for a peak exceeds the working range of the system, dilute
the extract and reanalyze.
11.5.7 The calibration curves should be linear over the range of concentrations
in Tables 2-5. Do not extrapolate beyond the calibration range established
in Section 10.0. If analyte response is too high, dilute the extract and
reanalyze.
12.1 Calculate the amount of material injected from the peak response using the multi-
point calibration curve or calibration factor determined in Section 10.2.2. Do not
use the daily calibration verification standard to quantitate method analytes in
samples. The concentration in the sample can be calculated from the following
equation.
12.2 Report results in µg/L without correction for recovery data. All QC data
obtained should be reported with the sample results.
13.1 Single laboratory accuracy and precision data were obtained by replicate liquid-
liquid extraction analyses of reagent water fortified at two sets of concentrations
of method analytes. The data are given in Tables 2 and 3. Accuracy and
precision data by liquid-solid extraction of reagent water fortified at a single
concentration are given in Table 4. Finally, Method validation data obtained by
the analyses of fortified tap water and raw source water are given in Tables 5-7.
13.2 Demonstrated MDLs are given in Table 2. To calculate MDLs, the following
equations were used:
where: t(n-1,1-alpha = 0.99) = Student's t value for the 99% confidence level with
n-1 degrees of freedom
506-18
n = number of replicates
S = standard deviation of replicate analyses
14.1 One option of this method utilizes the new liquid-solid extraction (LSE)
technology to remove the analytes from water. It requires the use of very small
volumes of organic solvent and very small quantities of pure analytes, thereby
eliminating the potential hazards to both the analyst and the environment. The
other option in this method uses significant volumes of organic solvents. It is
highly recommended that laboratories use solvent recovery systems to recover
used solvent as sample extracts are being concentrated. Recovered solvents
should be recycled or properly disposed of.
14.2 For information about pollution prevention that may be applicable to laboratory
operations, consult "Less Is Better: Laboratory Chemical Management for Waste
Reduction" available from the American Chemical Society's Department of
Government Relations and Science Policy, 1155 16th Street N.W., Washington,
D.C. 20036.
15.1 It is the laboratory's responsibility to comply with all federal, state, and local
regulations governing waste management, particularly the hazardous waste
identification rules and land disposal restrictions. The laboratory using this
method has the responsibility to protect the air, water, and land by minimizing
and controlling all releases from fume hoods and bench operations. Compliance
is also required with any sewage discharge permits and regulations. For further
information on waste management, consult "The Waste Management Manual for
Laboratory Personnel", also available from the American Chemical Society at the
address in Section 14.2.
16.0 REFERENCES
1. Glaser, J.V., D.L. Foerst, G.D. McKee, S.A. Quave, and W.L. Budde. "Trace
Analysis for Waste Waters", Environ. Sci. Technol. 15, 1426, 1981.
3. Giam, C.S., Chan, H.S. and Nef, G.S. "Sensitive Method for Determination of
Phthalate Ester Plasticizers in Open-Ocean Biota Samples", Anal. Chem., 47, 2225
(1975).
4. Giam, C.S., and Chan, H.S. "Control of Blanks in the Analysis of Phthalates in Air
and Ocean Biota Samples", U.S. National Bureau of Standards, Special Publication
442, pp. 70l-708, l976.
506-19
5. "Carcinogens - Working with Carcinogens", Department of Health, Education, and
Welfare, Public Health Service, Center for Disease Control, National Institute for
Occupational Safety and Health, Publication No. 77-206, August l977.
6. "OSHA Safety and Health Standards, General Industry", (29 CFR l9l0),
Occupational Safety and Health Administration, OSHA 2206 (Revised,
January l976).
8. ASTM Annual Book of Standards, Part 3l, D3694-78. "Standard Practices for
Preparation of Sample Containers and for Preservation of Organic Constituents",
American Society for Testing and Materials, Philadelphia.
9. ASTM Annual Book of Standards, Part 3l, D3370. "Standard Practices for
Sampling Water", American Society for Testing and Materials, Philadelphia.
506-20
17.0 TABLES, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA
Column 2: DB-1, fused silica capillary, 30 m x 0.32 mm I.D., 0.25 micron film
thickness, Helium linear velocity = 30 cm/s.
506-21
TABLE 2. ACCURACY, PRECISION, AND METHOD DETECTION LIMIT DATA
FROM SIX LIQUID-LIQUID EXTRACTION ANALYSES
OF FORTIFIED REAGENT WATER
Mean Mean
Meas. Accuracy
True Conc. Std. % of
Conc. µg/L Dev. True MDL
Analyte µg/L µg/L Conc. µg/L
Dimethyl Phthalate 2.02 1.42 0.38 70.3 1.14
Diethyl Phthalate 15 71 16
Di-n-butyl Phthalate 15 68 15
Bis(2-Ethylhexyl) Adipate 30 69 18
Bis(2-Ethylhexyl) Phthalate 30 67 21
Di-n-octyl Phthalate 30 62 23
506-22
TABLE 4. ACCURACY AND PRECISION DATA FROM SIX LIQUID-SOLID
EXTRACTION ANALYSES OF FORTIFIED REAGENT WATER
Relative
True Mean Accuracy Standard
Concentration % of True Deviation
Analyte µg/L Conconcentration %
Dimethyl phthalate 15 74 11
Diethyl Phthalate 15 85 10
Di-n-butyl Phthalate 15 74 11
Bis(2-Ethylhexyl) Adipate 30 84 11
Di-n-octyl Phthalate 30 85 13
506-23
TABLE 6. ACCURACY AND PRECISION DATA FROM SIX LIQUID-LIQUID
EXTRACTION ANALYSES OF FORTIFIED RAW SOURCE WATER
Relative
True Mean Accuracy Standard
Concentration % of True Deviation
Analyte µg/L Conconcentration %
Dimethyl Phthalate 5 59 51
Diethyl Phthalate 5 78 45
Di-n-butyl Phthalate 5 99 29
Bis(2-Ethylhexyl) Phthalate 5 91 35
Di-n-octyl Phthalate 5 54 24
Di-n-butyl Phthalate 5 95 30
Bis(2-Ethylhexyl) Adipate 5 65 24
Bis(2-Ethylhexyl) Phthalate 5 60 21
Di-n-octyl Phthalate 5 53 15
506-24
506-25
506-26
THIS PAGE LEFT BLANK INTENTIONALLY
506-27