ASTM D3230-13 (Reaprobada 2018) Salinidad en Crudo
ASTM D3230-13 (Reaprobada 2018) Salinidad en Crudo
ASTM D3230-13 (Reaprobada 2018) Salinidad en Crudo
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
--`,,,`,,,,`,`,`,`,``,,`-`-``,```,,,`---
of concentration covered is 0 mg ⁄kg to 500 mg ⁄kg or
0 lb ⁄1000 bbl to 150 lb ⁄1000 bbl as chloride concentration/ Crude Oils by Digital Density Analyzer
volume of crude oil.
3. Terminology
1.2 This test method measures conductivity in the crude oil
due to the presence of common chlorides, such as sodium, 3.1 Definitions of Terms Specific to This Standard:
calcium, and magnesium. Other conductive materials may also 3.1.1 PTB—lb/1000 bbl.
be present in the crude oil. 3.1.2 salts in crude oil—commonly, chlorides of sodium,
1.3 The values stated in SI units are to be regarded as calcium, and magnesium dissolved in crude oil. Other inor-
standard. No other units of measurement are included in this ganic chlorides may also be present.
standard.
1.3.1 Exception—Acceptable concentration units are g/m3 4. Summary of Test Method
or PTB (lb/1000 bbl).
4.1 This test method measures the conductivity of a solution
1.4 This standard does not purport to address all of the of crude oil in a mixed alcohol solvent when subjected to an
safety concerns, if any, associated with its use. It is the electrical stress. This test method measures conductivity due to
responsibility of the user of this standard to establish appro- the presence of inorganic chlorides, and other conductive
priate safety, health, and environmental practices and deter- material, in the crude oil. A homogenized test specimen is
mine the applicability of regulatory limitations prior to use. dissolved in a mixed alcohol solvent and placed in a test cell
For specific warning statements, see 7.3, 7.4, and 7.11. consisting of a beaker and a set of electrodes. A voltage is
1.5 This international standard was developed in accor- impressed on the electrodes, and the resulting current flow is
dance with internationally recognized principles on standard- measured. The chloride (salt) content is obtained by reference
ization established in the Decision on Principles for the to a calibration curve of current versus chloride concentration
Development of International Standards, Guides and Recom- of known mixtures. Calibration curves are based on standards
mendations issued by the World Trade Organization Technical prepared to approximate the type and concentration of chlo-
Barriers to Trade (TBT) Committee. rides in the crude oils being tested.
2. Referenced Documents
5. Significance and Use
2.1 ASTM Standards:2
D381 Test Method for Gum Content in Fuels by Jet Evapo- 5.1 This test method is used to determine the approximate
ration chloride content of crude oils, a knowledge of which is
important in deciding whether or not the crude oil needs
desalting. The efficiency of the process desalter can also be
1
This test method is under the jurisdiction of ASTM Committee D02 on evaluated.
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.03 on Elemental Analysis. 5.2 Excessive chloride left in the crude oil frequently results
Current edition approved July 1, 2018. Published August 2018. Originally in higher corrosion rates in refining units and also has
approved in 1973. Last previous edition approved in 2013 as D3230 – 13. DOI:
10.1520/D3230-13R18.
detrimental effects on catalysts used in these units.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
5.3 This test method provides a rapid and convenient means
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on of determining the approximate content of chlorides in crude
the ASTM website. oil and is useful to crude oil processors.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
--`,,,`,,,,`,`,`,`,``,,`-`-``,```,,,`---
7.3 Mixed Alcohol Solvent—Mix 63 volumes of 1-butanol
they are reasonably fluid before they are sampled; however, no
and 37 volumes of absolute methyl alcohol (anhydrous). To
sample shall be heated more than is necessary to lower the
each litre of this mixture, add 3 mL of water. (Warning—
viscosity to a manageable level.
Flammable. Liquid causes eye burns. Vapor harmful. May be
fatal or cause blindness if swallowed or inhaled.) 8.3 Samples of crude oil contain water and sediment and are
inhomogeneous by nature. The presence of water and sediment
NOTE 2—The mixed alcohol solvent is suitable for use if its conduc-
tivity is less than 0.25 mA at 125 V ac (or 2 microsiemens). High will influence the conductivity of the sample. The utmost care
conductivity can be due to excess water in the solvent and can indicate that shall be taken in obtaining homogenized representative
the methyl alcohol used is not anhydrous. samples.
7.4 Hexanes, Reagent Grade, (Warning—Extremely
9. Preparation of Apparatus
flammable, harmful if inhaled.)
9.1 Support the apparatus on a level, steady surface, such as
3
a table.
Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC. For Suggestions on the testing of reagents not 9.2 Prepare the apparatus for operation in accordance with
listed by the American Chemical Society, see Annual Standards for Laboratory
the manufacturer’s instructions for calibrating, checking, and
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, operating the equipment. (Warning—The voltage applied to
MD. the electrodes can be as great as 250 V ac, and hazardous.)
--`,,,`,,,,`,`,`,`,``,,`-`-``,```,,,`---
bration may be confirmed, if desired, by careful replicate analysis of
cylinder, add 15 mL of xylene. From a 10 mL pipet (total crude-oil samples by exhaustive extraction of salts with hot water,
delivery, see 6.3), or 10 mL graduated cylinder (see 6.3.1) followed by titration of the chlorides in the extract.
when applicable, add 10 mL of neutral oil. Rinse the 10 mL NOTE 8—In calibrating over a wide range of chloride concentrations, it
may be necessary to apply several voltages to obtain current readings
pipet or 10 mL graduated cylinder with xylene until free of oil. within the limit of the apparatus current level display (0 mA to 10 mA).
Make up to 50 mL with xylene. Stopper and shake the cylinder Higher voltages are applied for low concentrations and lower voltages are
vigorously for approximately 60 s to effect solution. Add a applied for high concentrations.
quantity of dilute mixed salts solution, in accordance with
Table 1, that is appropriate to the range of salt contents to be 11. Procedure
measured. Dilute to 100 mL with mixed alcohol solvent. Again 11.1 To a dry, 100 mL graduated, glass-stoppered cylinder,
shake the cylinder vigorously for approximately 30 s to effect add 15 mL of xylene and, using the same type of volume
solution, and allow the solution to stand approximately 5 min. transferring device used in 10.3 (that is, either a 10 mL pipet
Pour the solution into a dry test beaker. (total delivery, see 6.3), or 10 mL graduated cylinder (see
10.4 Immediately place the electrodes into the solution in 6.3.1) when applicable), transfer 10 mL of the crude oil sample
the beaker, making sure that the upper edge of the electrode into the 100 mL graduated, glass-stoppered cylinder. Rinse the
plates are below the surface of the solution. Adjust the 10 mL pipet or 10 mL graduated cylinder with xylene until free
indicated electrode voltage to a series of values, for example of oil. Make up to 50 mL with xylene. Stopper and shake the
25 V, 50 V, 125 V, 200 V, and 250 V ac. At each voltage, note cylinder vigorously for approximately 60 s. Dilute to 100 mL
the current reading and record the voltage displayed and the with mixed alcohol solvent, and again shake vigorously for
current to the nearest 0.01 mA. Remove the electrodes from the approximately 30 s. After allowing the solution to stand for
solution, rinse with xylene followed by naphtha, and allow approximately 5 min, pour it into the dry test beaker.
them to dry. 11.2 Follow the procedure in 10.4 to obtain voltage and
NOTE 5—With some apparatus, the detailed settings will not be required current readings. Record the indicated electrode current to the
nearest 0.01 mA and the nearest voltage (see Note 6).
TABLE 1 Standard Samples 11.3 Remove the electrodes from the sample solution, and
Salt, g/m 3
Salt, lb/1000 bbl Mixed Salts Solution clean the apparatus.
of Crude Oil of Crude Oil (dilute), mL
3 1.0 0.3 12. Calculation
9 3.0 1.0
15 5.0 1.5 12.1 Subtract the value obtained for the blank measurement
30 10.0 3.0 from the value obtained from the specimen measurement to
45 16.0 4.5 obtain the net current reading. From the calibration graph, read
60 21.0 6.0
75 26.0 8.0 the indicated salt concentration corresponding to the net
90 31.0 9.5 current (mA) reading of the sample (see Note 6).
115 40.0 12.0
145 51.0 15.0 12.2 Calculate the concentration in mg/kg by using the
190 66.0 20.0 appropriate equation given below:
215 75.0 22.5
245 86.00 25.5 1000X
290 101.0 30.5 Salt, mg/kg 5 (1)
d
430 151.0 45.0
Salt, mg/kg 5 2853 Y/d (2)
ANNEX
(Mandatory Information)
A1. APPARATUS
A1.1 Electrometric Chloride Apparatus (Custom Manu- A1.1.5 Transformer, plate supply 240 V, center tap, 50/60
facture) (see Fig. A1.1 and Note A1.1) Hz, 250 mA dc capacity.
A1.1.1 Milliammeter, 0 mA to 1 mA dc with 0 mA to 1 mA A1.1.6 Potentiometer, 25 Ω, ten turn.
--`,,,`,,,,`,`,`,`,``,,`-`-``,```,,,`---
A1.1.3 AC Voltmeter, rectifier type, 2000 Ω ⁄V, 0 V to 300 V A1.2 Test Cell Components (Custom Manufacture) (see
range. Fig. A1.2)
A1.1.4 Variable Voltage Autotransformer, input 105 V to A1.2.1 Beaker, 100 mL tall form without lip, as described
117 V, 50/60 Hz, output 0 V to 132 V, 1.75 A capacity. for use in Test Method D381.
--`,,,`,,,,`,`,`,`,``,,`-`-``,```,,,`---
--`,,,`,,,,`,`,`,`,``,,`-`-``,```,,,`---
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
--`,,,`,,,,`,`,`,`,``,,`-`-``,```,,,`---