969 HQT
969 HQT
969 HQT
Section 4. Disassembly/Reassembly
Options
S Additional memory modules are avaiIable.
Up to four 16 megabyte memory boards or
four 32 megabyte memory boards can be
installed. Since board types cannot be
mixed, maximum total image memory is
either 64 or 128 megabytes.
S Imation Cleaning Kit
S Imation/OEM Interfaces
S Imation IR Safelight
S Imation Footswitch
969 HQT Laser Imager
Service Manual Addendum
Installation EImation 1998 Rev. A 2-1
COMM 1
Keypad ....
S System configuration (parameter settings) Keypad Extension .
for the IMS. Cable Cable
....
COMM 0
2-2. Cable Connections ....
.
2-2-1. AC Source Input Wiring
! Caution Keypad
Before power is applied to the imager, the
input wires from the ac source must be
line–tapped appropriately at the TS1 terminal Figure 2-1. User Keypad to TDB Cabling
strip for the T1 power transformer.
2-2-3. Local Keypad to TDB
Connect the keypad cable to the upper
! WARNING communication port on the rear of the IMS
BEFORE REMOVING THE MACHINE enclosure. (A cable inside the IMS enclosure
LOWER REAR PANEL, MAKE SURE THAT leads from this point to J4 on the TDB.)
THE POWER CORD IS REMOVED FROM
THE AC OUTLET. FAILURE TO DO SO MAY ! Caution
CAUSE SEVERE INJURY OR DEATH IF
CONTACT WITH PRIMARY COMPONENTS The user and local keypads must be
OCCURS. connected to the correct ports to avoid
possible damage to the keypads or TDB.
1. Measure the AC voltage at the wall outlet.
2. Remove the machine lower rear panel. 2-2-4. Video Source to VIB
1. Use an analog cable to connect the video
3. Connect the line and jumper wires at the
signal from the modality to the appropriate
TS1 terminal strip appropriately for the ac
Video In connector on the VIB in the IMS
source voltage. Refer to the label on the
Enclosure. Run the analog cable through
power transformer. (Jumper wires are
the ferrite core as shown on the machine
provided in the shipping package.)
label. Refer to Figure 2-2.
969 HQT Laser Imager
Service Manual Addendum
Installation EImation 1998 Rev. A 2-2
2. If the modality provides a pixel clock signal, 2-3. Switch and Jumper Settings
use an analog cable to connect it to the
appropriate Ext Clock In connector on the 2-3-1. TDB
VIB in the IMS Enclosure. Refer to Jumpers W1 and W4
Figure 2-2.
Jumpers W1 and W4 configure J1 and J2 for
VIB RS422 or RS232 input. W1 configures J1
(Comm 0) and W4 configures J2 (Comm 1). Set
VIDEO VIDEO
the jumper to the left for RS232 and to the right
IN
PORT 0
for RS422. Refer to Figure 2-4.
OUT
Video
VIDEO VIDEO
IN
W4 SW2 J2 (Comm 1)
PORT 1
Modality TDB/C
OUT
Pixel Clock W1 SW1 J1 (Comm 0)
PORT 0 PORT 1
Figure 2-4. TDB Jumpers and Switches
EXT CLOCK IN
Digital Image
PORT 0
Modality
E8 E7
(2) Siemens
external fiber
E6
E5
E4
E3
E2
E1
optic converter
(RS232
jumpered)
Figure 2-5. VIB Jumpers
2-3-2. VIB
2-4. System Configuration
Pixel Clock Source Note
Two jumpers on the VIB specify the pixel clock The following procedures are performed
source. For an internal clock (generated by a with the MPC connected to the 969 HQT,
Phase Lock Loop module installed on the VIB), and require the HQT MPC for Windows
set the jumpers to E1-E2 and E4-E5. For an program. Connect the MPC
external clock (provided by the source communication cable to the bottom
modality), set the jumpers to E2-E3 and E5-E6. communication port at the rear of the
The jumpers are factory set, and should not IMS enclosure. (See paragraph 8-3-1 in
need to be changed unless a PLL module is Troubleshooting.) No hardlock key is
added or removed. Refer to Figure 2-5. required for the HQT MPC for Windows
application.
Continuous Acquire Mode
When a jumper is connecting pins E7 and E8, 2-4-1. Load IMS Parameters
the VIB operates in continuous-acquire mode. 1. From MPC Operations Setup, select the IMS
This may be useful when attempting to examine subsystem.
the source video signal. However, for normal
operation the two pins must not be connected 2. Select Output 0 Comm 0 from the
(place the jumper block on a single pin; this is component select dropdown list box (to the
the factory setting). right of the subsystem buttons).
3. Select the Comm window display button.
4. Select the Printer Defaults button, then
select the Save button.
5. Select the first user from the component
select dropdown list box.
969 HQT Laser Imager
Service Manual Addendum
Installation EImation 1998 Rev. A 2-4
7. Select the Comm window display button. 2-4-5. Digital Modality Setup
Select the appropriate Defaults button,
Note
modify the parameter settings as needed for
Load digital parameters from a script file
host control users, then select the Save
whenever possible (refer to Step 6 of
button.
paragraph 2-4-1). If a script file is not
Note available for the modality, enter
The following default settings should be parameters manually. To determine the
used, depending on your system correct settings, refer to the OEM
configuration: specifications, contact the OEM site
engineer, or contact the Technical
969 – Used for Superset Host Control Assistance Center.
users.
1. From MPC Operations Setup, select the EIB
subsystem.
959 – Used for 831/952 Host Control
users. 2. Select the first DIB user from the component
select dropdown list box.
HQT – Used for all Keypad users.
3. Select the Config window display button.
Modify the parameter settings as needed,
Siemens – Used for all Siemens Host
then select the Save button.
Control users.
Note
8. Select the Image window display button.
Modify the parameter settings as needed, When the Save button is selected, digital
then select the Save button. parameters are loaded into the DIB.
9. Select the Host window display button. 4. Verify that the modality is generating an
Modify the parameter settings as needed, image.
then select the Save button. 5. Select the Acquire button.
10. Select the System window display button.
Modify the parameter settings as needed,
then select the Save button.
969 HQT Laser Imager
Service Manual Addendum
Installation EImation 1998 Rev. A 2-5
6. Connect the host video cable to the 10. Try acquiring (be sure the Sync, Enable, and
oscilloscope (use a T connector with a 75 Acquire LEDs are all on).
ohm terminator attached). Measure the
S If successful, proceed to Step 11.
video signal from sync tip to maximum white
level. If less than 1 volt, set the Double Gain S If unsuccessful, try acquiring two or three
parameter to 0.5 to 1.0. If more than 1 volt, more times. If still unsuccessful, reset
set the Double Gain parameter to 1.0 to 2.0 the 969 HQT and the MPC, and try two
volts. or three more times. If still unsuccessful,
double check Steps 1 through 9.
7. Set the following parameters to midpoints as
indicated: 11. Adjust the following framing parameters to
obtain the full frame (rough draft only; this
S Clock Delay to 6.
will be done again during fine tuning):
S Fine Pixel Delay to 180.
S Vertical Delay and Image Lines
S Black Level to 100.
S Horizontal Delay and Horiz Active Pixels
8. Select the Save button.
12. Perform video fine tuning.
9. Verify that the Enable LED is on. This
indicates that the parameter set has been 2-4-6-3. Fine Tune Video Parameters
loaded and the VIB is in a ready state. Refer Note
to Figure 2-6. When printing images from MPC, the
images are replicated. Replication
reproduces the image from the host
exactly, without any smoothing. This is
ENABLE
INSYNC
ACQUIRE
CONFIG
OVRFLW
UNDFLW
14. Select the Print 1-Up button in the Video 21. To ensure that image Dmin will be
Fine Tuning window. acceptable, increase the gain as follows:
15. Select an image to print in the Video Fine S Add one to the displayed Digital Gain
Tuning – Print Full Size window (image value, then select the Manual Adjust –
numbers are printed above each image on Once button. The intent now is to
the 4:1 prints), then select the OK button. produce a maximum value of 1023.
Select the OK button when the Successful
22. Increasing the gain decreases Dmax as well
Print message is displayed. Examine the
as Dmin. To offset the gain increase,
image on the film.
increase the displayed Black Level value by
16. Repeat Step 15 as needed. When the single one, then select the Manual Adjust – Once
best image has been identified, select the button.
Cancel button in the Video Fine Tuning –
23. Select the Close button to accept the
Print Full Size window.
displayed values.
17. Select the Config window display button.
24. Acquire and print a SMPTE test pattern
Enter the clock delay and fine pixel delay
using the lowest available contrast setting
settings that are printed above the
(usually contrast #1). Check the black and
previously identified image. Select the Save
white levels on the film. If the levels are set
button.
correctly, the 95% and 5% patches on the
18. Repeat Steps 2 through 7. film will be equally visible.
19. At the MPC, if the adjustment is correct, the If a SMPTE pattern is not available, use a
following values will be displayed in the Gain customer image and check the grey scale
& Black Level Fine Tune window: steps. Again, using the lowest available
contrast setting, inspect the film for visual
S The maximum value will be below 1023,
distinction of the first two and last two steps
and should be 1016 for 8-bit pixels and
of the grey scale. ( Note: A low contrast
1022 for 12-bit pixels.
setting produces low contrast in the middle
S The minimum value will be above 0, and of the scale and high contrast at the ends of
should be 4 for 8-bit pixels and 1 for the scale.) Step 2 should not blend into step
12-bit pixels. 1, and step 15 should not blend into step 16.
If the OEM monitor is adjusted properly
20. When the values listed in the previous step (refer to paragraph 2-4-6-5), its grey scale
are displayed, select the Auto Adjust – Stop can be used for comparison purposes.
button to stop the continuous sampling of
the video. If the values are close, but not If the VIB is set up CORRECTLY: What is
right on, select the Auto Adjust – Once normally seen (using a contrast test) is that
button to perform a single sample and adjust the contrast test #1 Dmax (0% square) is
cycle (this can be repeated as required). lighter than a dark border and Dmin (100%
square) is equal to or slightly darker than a
Note clear border. It is not until later contrast tests
The sampling circuitry in the VIB is that image Dmax and Dmin equal film Dmax
susceptible to video noise which can and Dmin. This is acceptable, gives good
affect image Dmin values. To ensure that quality images, and results in the 95% and
image Dmin will be acceptable, the 5% patches on a SMPTE test film being
following step will increase the gain equally visible. If desired, the image Dmax
slightly to clip the noise from the video can be forced to match the border Dmax.
signal. This will decrease image Dmin There are two ways to accomplish this:
(and Dmax) slightly.
969 HQT Laser Imager
Service Manual Addendum
Installation EImation 1998 Rev. A 2-9
S In the Image Parameters window, set the 25. If the black or white level is unacceptable,
Match Border parameter to Yes. try any of the following:
S Increase the Black Level value by one or S Assuming that fine tuning has been
two digits. This will darken up the image done, try selecting a different sampling
Dmax, and will also darken up the image area (use different row and column
Dmin slightly. Do not go past the point of settings) when performing the MPC gain
just matching image Dmax to border adjustment.
Dmax or the 5% patch will start to S Try a different image (see Step 1 of this
disappear. procedure).
If the VIB is set up INCORRECTLY: Using S The Black Level setting can be
contrast #1, either the 95% or 5% patch will increased by one or two digits to darken
be gone or faint. If using a customer image, up the image Dmax. This will also
step 2 of the grey scale will blend with or be darken up the image Dmin slightly.
only faintly darker than step 1, or step 15 will
S If still having problems, contact TAC, a
blend with or be only faintly lighter than step
PST member, or the local video expert.
16. (See Step 25 of this procedure for
possible solutions.) 26. Make final adjustments to framing
parameters to obtain the full frame. The
The other possible incorrect setup of the VIB preceding fine tuning steps may have
is that image Dmax is significantly lighter shifted the horizontal delay so that a pixel is
than film Dmax or that image Dmin is lost on the right or left side of the image. To
significantly darker than film Dmin. Using a check for this, print a film with clear borders
contrast test, this can be seen by the fact (change the Border setting in the Image
that image Dmax and Dmin do not match Parameters window to 4095). View the
film Dmax and Dmin until the last few image. If a pixel is missing from the left or
contrast levels, or they never reach film right side, add or subtract one pixel from the
Dmax and Dmin. (See Step 25 of this Horizontal Delay setting. Be sure to change
procedure for possible solutions.) the Border setting back to 0 when fine
If the VIB is set up PERFECTLY: All of the tuning is complete.
following will be true (with Match Border set 27. Proceed to paragraph 2-4-6-4.
to Yes):
S Contrast #1 Dmax and Dmin will match 2-4-6-4. Set Customer Preferences
the border Dmax and Dmin. When performing the following procedure, note
S Increasing the final Black Level setting that the method of setting customer preferences
by one will make the contrast #1 Dmin varies depending on the control source.
slightly darker than a clear border. For Keypad and host control users:
S Decreasing the final Black Level setting S Density and contrast are set at the Local
by one will make the contrast #1 Dmax Keypad of the 969 HQT. Refer to the 969
slightly lighter than a dark border. HQT Operator’s Guide for details.
This would be the perfect situation.
However, in most cases, the VIB does not S Smooth/sharp image processing is set via
seek black and white level this accurately. MPC, based on interpolation settings. Refer
to the MPC for Windows Help file for details.
969 HQT Laser Imager
Service Manual Addendum
Installation EImation 1998 Rev. A 2-10
1. Have the customer select a typical image time the monitor will tend to drift, and the
that contains the range of contrasts they will brightness and contrast may have to be
be looking for. adjusted by the OEM.)
2. Acquire the image and print a contrast test. Check
3. If everything looks too light or too dark, With the SMPTE pattern at OEM defined
adjust the density setting. If the density window and level values, the 5% patches (both
setting looks okay, have the customer select black and white) should be visible, and should
a contrast setting. have equal contrast to the enveloping 100%
and 0% boxes around them. If the 5% patches
4. If the customer finds the image
are not visible, or are not equally visible, the
unacceptable at any combination of density
OEM monitor should be adjusted.
and contrast levels, it is possible that the
gain/black level adjustment needs to be S Ask the OEM to adjust the monitor.
redone using a different image, or the OEM
S Be there when the OEM adjusts the monitor,
monitor may be misadjusted (refer to
and ensure that the ambient lighting is the
paragraph 2-4-6-5).
same as the normal lighting the customer
5. Have the customer select smooth or sharp uses when filming.
image processing. If neither smooth nor
Adjustment
sharp is acceptable to the customer, the
interpolation settings may need to be 1. Adjust the white first. Looking at the text,
changed. Refer to the MPC for Windows turn the OEM contrast knob until the whites
Help file for details on how to change just start to smear. Then back off the
interpolation settings. contrast just to the threshold of smearing.
The 95%/100% patch in the SMPTE should
2-4-6-5. OEM Monitor Adjustment now be visible.
If the customer likes the images there is no
2. Adjust the black. Turn the brightness knob
need to adjust the OEM monitor even though it
until the image starts to fill in. Continue until
may be slightly off. However, if the customer is
the black in the image starts to turn grey. At
unable to get an acceptable contrast on the
this point, the 5%/0% patch should be
images, it may be necessary to make the
visible. Back off the brightness until there is
following checks and/or adjustments.
an equal balance in the visual ratio of 0% to
Theory 5% as 100% to 95%. These two patches
are the key.
The OEM’s video generator board outputs a
video signal to the OEM monitor. This same
video signal is sent to the VIB. The window and
level controls adjust this video signal, which
affects both the image displayed on the OEM
monitor and the image printed on the film. If the
monitor’s brightness and contrast are out of
adjustment, the customer will compensate by
adjusting the window and level controls until the
image looks good on the monitor. The result of
this is that the image on the film and the image
on the monitor do not match. Therefore, it is
important that after the brightness and contrast
have been set correctly, the customer does not
turn the brightness and contrast knobs. (Over
969 HQT Laser Imager
Service Manual Addendum
Adjustments EImation 1998 Rev. A 3-1
3-1. General
The adjustments in the 969 HQ/HQS Service
Manual apply to the 969 HQT except as follows:
Paragraph 3-9, IMS Printed Wiring Assemblies,
does not apply.
The only adjustments required for the 969 HQT
IMS are:
S Jumper and switch settings on the
Translator Daughter Board (TDB).
S Jumper settings on the Video Interface
Board (VIB).
Note
See paragraph 2-3 in Section 2,
Installation, for procedures for setting
these jumpers and switches.
969 HQT Laser Imager
Service Manual Addendum
Disassembly/Reassembly EImation 1998 Rev. A 4-1
4-3. Replacing the IMS AC Power 4. Pry loose and remove the red fuseholder.
Fuses 5. Remove the blown fuse(s).
Reassembly
! WARNING
1. Place one end of the new fuse on top of the
When the power cord is plugged in,
fuse cartridge end terminal (see side view in
hazardous voltages are present in some
Figure 4–2).
areas of the 969 HQT. These voltages can
cause severe injury or death. 2. Place the other end of the fuse on the
center set of clips. (It is not necessary to
Disassembly force it down into the clips.)
1. Power down the 969 HQT and unplug the 3. Position the other fuse as described in steps
power cord. 1 and 2 above.
2. Remove the lower back access panel from 4. While holding the fuses in place, insert the
the 969 HQT. fuse cartridge into the fuse and switch
assembly. The fuses will snap into the
3. At the top of the IMS enclosure (see Figure fuseholder clips as the cartridge is inserted.
4-2), use a small flat-blade screwdriver to
pry open the door to the fuse cartridge.
Figure 4-2.
969 HQT Laser Imager
Service Manual Addendum
Disassembly/Reassembly EImation 1998 Rev. A 4-3
1. Power down the 969 HQT and unplug the 4. When replacing an MDB, make sure it is
power cord. positioned in the correct location (see Figure
4-3). If only one board is used, for example,
2. Extend the IMS enclosure from the 969 HQT it must be placed in the MDB 1 location.
cabinet as described in paragraph 4-2.
5-1. General
The entire Section 5, Additional Information, of
the 969 HQ/HQS Service Manual applies to the
969 HQT.
969 HQT Laser Imager
Service Manual Addendum
Theory of Operation EImation 1998 Rev. A 6-1
Reference
Name Function
Designation
IPB Image Processor Board Image processing
MDB Memory Daughter Board Image memory – up to 4 boards per system
TDB Translator Daughter Keypad or host interface – any command set
Board
DIB Digital Interface Board Digital host image interface
VIB Video Interface Board Video host image interface
MCS-IB Machine Control System Interfaces the IMS with the 969 HQT machine control
Interface Board system
969 HQT Laser Imager
Service Manual Addendum
Theory of Operation EImation 1998 Rev. A 6-2
Block Function
Microprocessor Controls the functions of the IPB and interface boards.
Image Interface Connects the IPB to one of the four possible image interface boards.
Scale LUT Maps the input pixel width to the memory width.
Acquisition Controls image acquisition modes and generates the timing needed for the
Control & image memory.
Memory Timing
Image Memory Consists of image memory control, an address generator, and connectors for
Control the four MDBs.
Interpolation Performs image interpolation and filtering.
Contrast Table Provides a 10-bit to 12-bit output conversion and a linear 12-bit output path.
(Located in IPB EPROM.)
Format Control Controls the output to the Laser Optics interface and switches between image
and border.
Printer Interface Outputs the formatted image data to the Laser Opticsvia the MCS.
Parameter RAM Stores acquisition and print parameters and performs parameter initialization
between various logic blocks.
Communication Provides an interface to the TDB.
Microprocessor
969 HQT Laser Imager
Service Manual Addendum
Theory of Operation EImation 1998 Rev. A 6-5
Comm Micro
Micro Block
Block
TDB
Interface
Acquisition Format
Control & Memory Parameter
RAM Block
Timing Block
Image acquisition will support a line burst rate of The memory banks can be depopulated for an
4 megapixels per second and have a maximum 8-bit image plus 1-bit parity configuration. The
average transfer rate of 3 megapixels per four least significant bits (0–3) are depopulated.
second line to line. Acquisition of video images A depopulated board provides for an 8-bit data
having a 110 MHz pixel frequency will be path and has 36 DRAMs.
supported by allowing a maximum of 32 The address for each memory bank is provided
acquisition passes. through one of two address multiplexers which
also double as the address drivers. One
multiplexer drives banks 0 and 1 and the other
drives banks 2 and 3. They take the lower 22
bits of the input address and convert these to
row and column addresses to the DRAM. A
single control signal from the IPB switches the
multiplexers between the two DRAM addresses.
The presence or absence of a pull-down resistor
at the IPB connector is used to indicate to the
IPB whether a MDB is an 8 or 12-bit version.
The IPB masks the four least significant bits for
an 8-bit MDB. A pin is grounded at the IPB
connector to indicate to the IPB whether a MDB
is installed at that connector location. Another
pin is grounded at the IPB connector to indicate
to the IPB whether a MDB is a 16 or 32 MP
version.
969 HQT Laser Imager
Service Manual Addendum
Theory of Operation EImation 1998 Rev. A 6-7
D IN INPUT
BUFFER
BANK BANK BANK BANK
D OUT 0 1 2 3
OUTPUT
LATCH
IPB CONNECTOR
ADDR 1
MUX 1 1
ADDR 1
MUX 2 1
3
3
CNTL CNTL 3
PAL DRIVERS 3
6-2-4-2. Video Interface Board (VIB) input, video output and a pixel clock input. The
The VIB converts video image data to a digital VIB interfaces to the IPB through the image
format and transfers it to the IPB over the image source interface connector (CN1107). It
source interface. The video image must be a provides a 9-bit image output to the IPB along
composite video signal meeting the RS170/343 with a strobe. the VIB receives its parameters
or equivalent standard and is usually the output over a 10-bit parameter bus and returns
of a D/A converter. The VIB also provides an command status over a 2-bit command bus.
input for the clock that samples the D/A Both functions provide a strobe signal. The VIB
converter. The video is digitized to 9-bits/pixel. receives its power, clock and power-up reset
The VIB supports external clock sampling up to from the IPB through CN1107.
110 MHz by allowing a maximum of 32 passes
for one acquisition. A Phase Locked Loop (PLL)
6-2-5. New Modem Board (NMB)
module can be plugged into the VIB to provide Refer to Figure 6-5 for a block diagram of the
the clock for synchronous digitization when an NMB. This board provides an RS232 serial
external clock is not available. The VIB can communications interface to the IPB via a nine-
accept video signals with interlaced or pin connector. The MPC can be connected to
progressive scan modes. this port for performaqnce of system setup and
diagnostics. Or an external modem can be
The VIB has two video interfaces for image connected to the port from an approved
acquisition only one of which may acquire at ”medical” isolation transformer (refer to
any one time. Each interface consists of a video Figure 6-6).
DB9 Female
1
6
2 RCV 21 21 TX2
7 CTS 25 25
3 TXD 19 19 RX2
8 RTS 23 23
4 DTR 30 30 DTR
9
5
New Modem Board Image Processor Board
Phone
”Medical”
Isolation
Isolated External
Transformer Power Modem
OR DB9 Female
Input Power
From Line MPC New Modem Board
Figure 6-6. Implementation of an External Modem Via the New Modem Board
969 HQT Laser Imager
Service Manual Addendum
Theory of Operation EImation 1998 Rev. A 6-10
6-2-6. MCS Interface Board The optics interface supports the following
This board interfaces the IMS with the 969 HQT functions:
machine control system and laser optics. It S Accepts start-of-line requests from the
provides both an image data interface and a optics, and supports up to two requests for
communications interface, connecting with the each stored line.
IPB on one side (IMS data interface) and the
MCS CPU Board (optics interface) on the other S Transmits pixels to the optics at a 2.5
(see Figure 6-7). megapixel per second rate, and outputs
4096 pixels per line.
The IMS data interface supports the following
functions: S Generates odd parity for each output pixel.
S Generates start-of-line requests to the IMS. S Provides a passive serial communication link
between the IMS and the optics.
S Accepts image data from the IMS at a
varying rate from 4.167 to 1.042 megapixels The data buffer supports the following functions:
per second. S Stores two lines of data consisting of 4096
S Provides a 12-bit input data path. pixels per line and 12 bits per pixel.
S Provides a passive serial communication link S Provides time interleaving of write and read
between the IMS and the optics. operations.
S Provides write rates up to 4.167 pixels per
second.
S Provides read rates of 2.5 megapixels per
second.
Message
Local Panel Message Corrective Action
Number
–– Film Processor Info Status message; no action required.
Developer Fixer Dryer
{Temp} {Temp} {Temp}
Idle
–– Film Processor Info Status message; no action required.
Developer Fixer Dryer
{Temp} {Temp} {Temp}
Processing
–– Laser Imager Status message; no action required.
Printing
{Media Size} {Media Type}
Supply XXX Receive XXX
–– Laser Imager Status message; no action required.
Ready
{Media Size} {Media Type}
Supply XXX Receive XXX
–– LCD, LED, and BEEPER Status message; no action required.
––– Testing –––
LASER IMAGER
–– MCS: Selftesting Status message; no action required.
–– Printer is currently performing selftest. Status message; no action required.
P112 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P113 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P114 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P115 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P116 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P117 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P118 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P119 Media pickup error. This error should automatically clear. See flowchart in 8-3 of 969 HQ
Service Manual.
P121 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P122 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-5
Message
Local Panel Message Corrective Action
Number
P123 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P125 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P126 Service mode. Optics ignored. Set bit 2 of S3 on PWA-A to OFF.
P130 Close supply drawer. See flowchart in 8-3 of 969 HQ
Service Manual.
P131 The media installed is not supported in this printer. Insert Load a supply cartridge containing a
another type. supported media type.
P132 Insert supply cartridge. See flowchart in 8-3 of 969 HQ
Service Manual.
P133 Supply cartridge has less than 20 sheets remaining. Status message; no action required.
P134 Supply cartridge is empty. See flowchart in 8-3 of 969 HQ
Service Manual.
P135 Opening media supply cartridge. Status message; no action required.
P136 Closing media supply cartridge. Status message; no action required.
P137 Printing in progress. When printing is complete, supply Status message; no action required.
drawer will open.
P138 Press supply button for 5 seconds to open the drawer. Self-explanatory; alternates with P177.
This will expose XXX sheets in the supply cartridge.
P139 Could not read barcode on supply cartridge. Replace See flowchart in 8-3 of 969 HQ
cartridge. Service Manual.
P140 Close receive door. See flowchart in 8-3 of 969 HQ
Service Manual.
P141 Insert empty receive magazine. See flowchart in 8-3 of 969 HQ
Service Manual.
P142 Receive magazine can hold media of only one size and Self-explanatory.
type. Please empty receive magazine.
P144 Please empty receive magazine. See flowchart in 8-3 of 969 HQ
Service Manual.
P147 Printing in progress. When printing is complete, receive Status message; no action required.
door will open.
P148 Press receive button for 5 seconds to open the door. Self-explanatory; alternates with P174.
This will expose XX sheets in the receive magazine.
P152 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P154 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
P155 Printer malfunction. See flowchart in 8-3 of 969 HQ
Service Manual.
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-6
Message
Local Panel Message Corrective Action
Number
P160 Media jam. Attempting to clear jam. Status message; no action required.
P161 Media jam at supply cartridge. Remove jam. See flowchart in 8-3 of 969 HQ
Service Manual.
P162 Media jam near supply cartridge. Remove jam. See flowchart in 8-3 of 969 HQ
Service Manual.
P163 Media jam near exposure area. Remove jam. See flowchart in 8-3 of 969 HQ
Service Manual.
P164 Media jam at exposure area. Remove jam. See flowchart in 8-3 of 969 HQ
Service Manual.
P165 Media jam behind receive door. Remove jam. See flowchart in 8-3 of 969 HQ
Service Manual.
P166 Media pickup error. This error should automatically clear. See flowchart in 8-3 of 969 HQ
Service Manual.
P171 Supply drawer could not be opened. See flowchart in 8-3 of 969 HQ
Service Manual.
P172 Receive door could not be opened. See flowchart in 8-3 of 969 HQ
Service Manual.
P174 Receive magazine could not be closed. See flowchart in 8-3 of 969 HQ
Service Manual.
P175 Receive magazine could not be opened. See flowchart in 8-3 of 969 HQ
Service Manual.
P176 Supply cartridge could not be opened. Replace See flowchart in 8-3 of 969 HQ
cartridge. Service Manual.
P177 Supply cartridge could not be closed. Opening drawer See flowchart in 8-3 of 969 HQ
will expose XXX sheets. Service Manual.
P179 Replace transport access panel. See flowchart in 8-3 of 969 HQ
Service Manual.
P601 Printer malfunction. See flowchart in 8-3 of this Addendum.
P602 Printer malfunction. See flowchart in 8-3 of this Addendum.
P603 Printer malfunction. See flowchart in 8-3 of this Addendum.
P604 Printer malfunction. See flowchart in 8-3 of this Addendum.
P605 Printer malfunction. See flowchart in 8-3 of this Addendum.
P606 Printer malfunction. See flowchart in 8-3 of this Addendum.
P607 Selftest interrupted. Printing will wait until selftest This message alternates with another
complete. message that indicates the cause of
the interruption. Refer to the flowchart
for the other message.
P608 Printer malfunction. See flowchart in 8-3 of this Addendum.
P609 Printer malfunction. See flowchart in 8-3 of this Addendum.
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-7
Message
Local Panel Message Corrective Action
Number
P620 Printer malfunction. See flowchart in 8-3 of this Addendum.
P910 Printer malfunction. See flowchart in 8-3 of this Addendum.
P912 Error not corrected during print. See flowchart in 8-3 of this Addendum.
P913 Error not corrected during print. See flowchart in 8-3 of this Addendum.
P914 Error corrected during print. Status message; no action required.
P915 Error corrected during print. Status message; no action required.
Message
User Keypad Message Corrective Action
Number
K59 Image signal error. Try acquiring again. See flowchart for L59 in 8-3 of this
Addendum.
K65 Image acquire failed. Try acquiring again. See flowchart for L59 in 8–3 of this
Addendum.
K68 Image acquire failed. Try acquiring again. See flowchart for L68 in 8–3 of this
Addendum.
K72 Memory error. Try acquiring again. Refer to related log message (L64 or
L67) flowchart in 8-3 of this Addendum.
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-8
8-3. Troubleshooting Flowcharts and 8-3-2. Log and Local Panel Message
Supporting Procedures Flowcharts
This subsection contains troubleshooting Flowcharts are provided for the following log and
flowcharts and supporting procedures. The local panel messages. Where trouble analysis
flowcharts are keyed to error messages that procedures are identical for different error codes,
apply only to the 969 HQT. These error a single flowchart is provided for multiple codes.
messages relate wholly (or partially) to IMS
L48 EOF_TIMEOUT (see page 8-9)
functions. For flowcharts relating to error codes
L59 EIB Parameter Write Error (see page 8-10)
that pertain to the MCS (and thus are common
L59 ACQUERR_EIB (see page 8-11)
to both the HQ and HQT), see Section 8 of the
969 HQ Service Manual. L64 ACQ_PARITY_EIB (see page 8-12)
L67 FIFO_OVERFLOW (see page 8-13)
8-3-1. Connecting the MPC to the IMS L68 TIMEOUT_EIB (see page 8-14)
for Troubleshooting L85 MEM_FULL (see page 8-15)
1. Turn off power on the MPC and (if possible) L200 PARITY_ERR (see page 8-16)
the imager. L211 EOL_ERROR (see page 8-9)
L212 HANDSHK_ERROR (see page 8-17)
2. Remove the lower access panel from the L214 LINELENGTH_ERROR (see page 8-17)
rear of the imager. L215 OUTPUT_ERROR (see page 8-9)
3. Connect a straight-through serial cable L601/P601 INVALID_POWER_MONITOR_OFFSET
((9-pin female to 9-pin male) between the (see page 8-18)
COM 1 serial port of the MPC and the lower L602/P602 INVALID_POWER_MONITOR_RANGE
(see page 8-18)
serial port on the rear of the IMS enclosure.
L603/P603 INVALID_OPT_DENSITY_RANGE
(see page 8-18)
L604/P604 INVALID_LASER_DYN_RANGE
(see page 8-18)
L605/P605 ATTEN_TEST_FAIL (see page 8-18)
P606 OPT_CMD_FAIL (see page 8-18)
P608 OPT_UNK_RESP (see page 8-19)
P609 OPT_INV_DATA (see page 8-19)
L620/P620 NO_FILM_MODEL (see page 8-20)
P910 No communications between IMS and MCS
(see page 8-21)
L912/P912 NO_CLK_RECEIVED (see page 8-9)
L913/P913 PARITY_ERROR (see page 8-9)
L914/P914 SINGLE_PARITY_ERROR
(see page 8-22)
L915/P915 SINGLE_NO_CLK_RECEIVED
(see page 8-22)
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-9
L48 EOF_TIMEOUT
L211 EOL_ERROR
L215 OUTPUT_ERROR
L912/P912 NO_CLK_RECEIVED
L913/P913 PARITY_ERROR
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-10
L59 ACQERR_EIB
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-12
L64 ACQ_PARITY_EIB
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-13
L67 FIFO_OVERFLOW
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-14
L68 TIMEOUT_EIB
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-15
L85 MEM_FULL
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-16
L200 PARITY_ERR
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-17
L212 HANDSHK_ERROR
L214 LINELENGTH_ERROR
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-18
L601/P601 INVALID_POWER_MONITOR_OFFSET
L602/P602 INVALID_POWER_MONITOR_RANGE
L603/P603 INVALID_OPT_DENSITY_RANGE
L604/P604 INVALID_LASER_DYN_RANGE
L605/P605 ATTEN_TEST_FAIL
P606 OPT_CMD_FAIL
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-19
P608 OPT_UNK_RESP
P609 OPT_INV_DATA
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-20
L620/P620 NO_FILM_MODEL
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-21
L914/P914 SINGLE_PARITY_ERROR
L915/P915 SINGLE_NO_CLK_RECEIVED
969 HQT Laser Imager
Service Manual Addendum
Troubleshooting EImation 1998 Rev. A 8-23
8-3-5. IMS Power Supply Checks 8-3-8. Fiber Optic Cable Checks
Not required for the 969 HQT.
9-1. General
The Illustrated Parts Breakdown provided in this section covers the IMS asssembly of the 969 HQT.
The following subsections of the IPB in Section 9 of the 969 HQ/HQS Service Manual do not apply to
the 969 HQT:
9-21 HQ (Multi-User) IMS
9-22 HQS (Single-User) IMS
9-23 External System Interfaces/Cabling
9-24 Keypad
9-27 Dual Printer Interface
969 HQT Laser Imager
Service Manual Addendum
Illustrated Parts Breakdown EImation 1998 Rev. A 10-2
Notes:
* A quantity of 1 to 4 of either a 16 Mbyte or 32 MByte MDB may be used. The two types cannot be
mixed.
** Either a VIB or DIB may be installed.