Instrument Datasheet

LT-Inclibus
Digital Inclinometer
DES CR IP T I ON

The LT-Inclibus gauge is able to monitor local tilting mechanical joints, while the gauges are connected in
along a line, assuring the alignment, distance and a RS485 chain. The logger can read both the
measuring axis orientation between the gauges. inclination of the two axis in engineering units and
The standard segment is composed by a 2m fibre the internal sensor diagnostics for each measuring
glass rod with two biaxial waterproof gauges, 1m point (temperature and voltage supply).
spaced.
Customers can use any electronic device compatible
The chain of LT-Inclibus can be installed within a with RS485 and Modbus RTU protocol as a logger.
borehole, mounted to a surface of a structure, laid The LT-Inclibus gives a complete and transparent
along a trench or buried in a concrete mass. It is array of data in engineering unit, as a result.
possible to have one or four gauges on the 2m rod
upon request. The rods are connected through

A P P L IC AT I ON S FE AT U RES

✔ Embankments ✔ Light and flexible array


✔ Unstable slopes ✔ Simple and fast to install
✔ Settlements ✔ Number of measuring points customisable
✔ LNG tanks ✔ Each measuring point is individually calibrated
following high level procedures
✔ Deep excavations
✔ Tunnelling
✔ Dam slope stability
✔ Deck bridges deformation
TECH N IC A L S PECI F I C ATI ONS

PRODUCT CODE 0LTIBV20102 0LTIBH20102

(1)
TILTMETER

Application Vertical, biaxial Horizontal, biaxial

Measurement principle TRIAXIAL MEMS inclinometer TRIAXIAL MEMS inclinometer


2 axis used 2 axis used

Measuring range standard ±10° (other ranges available on request)

Sensor resolution 0.0002°

Sensor repeatability <±0.008°

Sensitivity (2) see the Calibration Report

Sensor accuracy
Lin. MPE(3) <±0.10% F.S.
Pol. MPE(3) <±0.05% F.S.

Sensor 24h stability (4) < ±0.1mm/m

Sensor mechanical bandwidth 10 Hz

Sensor offset temperaturedependancy A axis: <±0.01°/°C A axis: <±0.004°/°C


B axis: <±0.004°/°C B axis: <±0.004°/°C

Temperature operating range from -30°C to +70°C

TEMPERATURE SENSOR (5) Embedded on electronic board

Measuring range - 40°C to +125°C

Accuracy ±1°C with temperaturerange -10°C to +85°C

SUPPLY VOLTAGE MONITOR (5) Embedded on electronic board

Measuring range 0 to 36 V

Accuracy ±5% FS

ELECTRICAL SPECIFICATIONS

Signal output RS485 non-optoisolatedcommunication with MODBUS RTU protocol (6)

Maximum reading frequency 2.5 sec for each reading (7)

Power supply from 8 to 28 Vdc

Average consumption 3.2 mA @ 24 Vdc, 4.6 mA @ 12 Vdc

Max cable length to logger 1000 m

(1) Technical characteristics are referred to ±10° measuring range


(2) Sensitivity is a specific parameter different for every gauge. The sensitivity is calculated during gauge calibration test and inserted into the Calibration
Report.
(3) MPE is the MaximumPermitted Error on the measuring range (FSR). In the Calibration Report, the accuracies of the gauge are calculated using both linear
regression (≤ Lin. MPE) and polynomial correction (≤ Pol. MPE)
(4) Stability calculated as difference after a 24 h period under repeatability conditions (ISO18674-3).
(5) These sensors are installed on the internal electronic board for sensor diagnostics.
(6) RS485 not-optoisolated Modbus communication with RTU Protocol. Legacy mode is not supported by this instrument Default output is sen α, other units
available are degree, mm/mand inch/feet (to be requested at order).
(7) Faster reading are available under request, but the performance of the system will be lower due to the increment of noises.
P H YS I C A L FE ATURES

GAUGE FIBERGLASS ROD

Material polycarbonate stainless steel joint tips and fibre-glass rod

IP class IP68 up to 1.0 MPa (2.0 MPa on request) --

Dimensions 145mm x 35mm x 35mm 2m length, rod Ø 7 mm

Weight 200 gr 200 gr

145 mm fibreglass rod Ø 7mm gauge section 35x35mm

500 mm 1000 mm 500 mm

2000 mm

MEASURING AXIS IN HORIZONTAL APPLICATION

A-channel B-channel

MEASURING AXIS IN VERTICAL APPLICATION

A-channel B-channel
A C CE S S O RI ES AN D SPARE PARTS

Available in different lengths (2m, 5m, 10m, 15m), it is composed by a


CABLE WITH CONNECTOR
signal cable with IP68 connector to link the nearest gauge to local logger,
0S400HD00MT
wireless node or junction box.

TERMINATION RESISTANCE Resistance ending device with connector, needed to close every digital IPI
0ETERMRESI0 chain. The value of resistor depends on the layout of the system.

2M FG ELONGATION ROD Fibre glass rod for chain elongation, 2m length, including mechanical joints
0LTIBROD020 and signal cable with connectors assembled at factory.

DIGITAL CABLE
LSZH cable for connecting digital gauge chains to OMNIAlog data logger.
0WE606IPDZH

Kit composed by one 120 Ohm, two 240 Ohm, three 360 Ohm and four 480
RESISTANCES KIT (SPARE) Ohm resistance ending devices. Each one has a M12 5-pin connector for
0ERESIKIT00 linking to digital gauges. Check the compatibility with old digital gauges,
consulting your Sales Representative.

Logger
Cable with Termination
connector resistance

Signal Mechanical Fibreglass Signal LT INCLIBUS Gauge Foundation


cable joint rod connectors segment concrete slab

GEOMOTION.COM.AU | INFO@GEOMOTION.COM.AU

You might also like