Scope CC 2576 - 1
Scope CC 2576 - 1
Scope CC 2576 - 1
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
Permanent Facility
ELECTRO-
TECHNICAL-
Using Fluke 6½ DMM
ALTERNATING
1 AC Current @ 50 Hz 8846A by Direct 50 µA to 10 A 0.87 % to 0.6 %
CURRENT (< 1
Method
GHZ)
(Measure)
ELECTRO-
TECHNICAL-
Using Fluke 6½ DMM
ALTERNATING
2 AC Voltage @ 50 Hz 8846A by Direct 1 mV to 1000 V 4.4 % to 0.52 %
CURRENT (< 1
Method
GHZ)
(Measure)
ELECTRO-
Using Zeal 5½
TECHNICAL-
Digital Multi-function
3 ALTERNATING AC Current @ 50 Hz 0.2 mA to 20 A 3 % to 3.87 %
Calibrator By Direct
CURRENT (< 1
Method
GHZ) (Source)
ELECTRO- Using Zeal 5½
TECHNICAL- Digital Multi-function
4 ALTERNATING AC Current @ 50 Hz Calibrator with 20 A to 990 A 3 % to 3.87 %
CURRENT (< 1 Current Coil by
GHZ) (Source) Direct Method
ELECTRO-
Using Zeal 5½
TECHNICAL-
Digital Multi-function
5 ALTERNATING AC Voltage @ 50 Hz 5 mV to 990 V 1.81 % to 0.32 %
Calibrator by Direct
CURRENT (< 1
Method
GHZ) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 2 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
ELECTRO-
TECHNICAL-
Capacitance @ 1 Using Sigma Decade
6 ALTERNATING 100 pf to 1 µf 1.81%
kHz Capacitance Box
CURRENT (< 1
GHZ) (Source)
ELECTRO-
Using Sigma RSI-6
TECHNICAL-
Decade Inductance
7 ALTERNATING Inductance @ 1 kHz 100 µH to 10 H 1.81%
Box by Direct
CURRENT (< 1
Method
GHZ) (Source)
ELECTRO-
TECHNICAL- Using Fluke 6½ DMM
8 DIRECT DC Current 8846A by Direct 10 µA to 10 A 4.03 % to 0.98 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using Zeal HV Probe
9 DIRECT DC High Voltage with Fluke DMM by 1 kV to 10 kV 2.6%
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using Fluke 6½ DMM
10 DIRECT DC Voltage 8846A by Direct 1 mV to 1000 V 0.71%
CURRENT Method
(Measure)
ELECTRO- Using Zeal 5½
TECHNICAL- Digital Multi-function
11 DIRECT DC Current Calibrator with 0.2 mA to 990 A 4.03 % to 1.11 %
CURRENT Current Coil by
(Source) Direct Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 3 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 4 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
ELECTRO-
Using Radix
TECHNICAL- Temperature
Universal Calibrator
18 TEMPERATURE Indicator / Controller 0 °C to 1600 °C 2.7°C
by Direct Method as
SIMULATION / Recorder - R Type
per IS 1248
(Source)
ELECTRO-
Temperature Using Radix
TECHNICAL-
Indicator / Controller Universal Calibrator
19 TEMPERATURE (-) 100 °C to 450 °C 1.3°C
/ Recorder - RTD by Direct Method as
SIMULATION
Type per IS 1248
(Source)
ELECTRO-
Using Radix
TECHNICAL- Temperature
Universal Calibrator
20 TEMPERATURE Indicator / Controller 0 °C to 1700 °C 2.1°C
by Direct Method as
SIMULATION / Recorder - S Type
per IS 1248
(Source)
ELECTRO-
Using Radix
TECHNICAL- Temperature
Universal Calibrator
21 TEMPERATURE Indicator / Controller (-) 100 °C to 400 °C 2.1°C
by Direct Method as
SIMULATION / Recorder - T Type
per IS 1248
(Source)
ELECTRO- Using Countron
TECHNICAL- Digital Timer / Stop 1050 Digital Timer
22 TIME & Watch - Analog / by Comparison 10 s to 6 hour 1.02 % to 0.11 %
FREQUENCY Digital Method as per IS
(Measure) 1248
ELECTRO-
TECHNICAL- Using Fluke 6½ DMM
23 TIME & Frequency 8846A by Direct 10 Hz to 1 MHz 2.86%
FREQUENCY Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 5 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
ELECTRO-
Using Zeal Multi-
TECHNICAL-
function Generator &
24 TIME & Frequency 10 Hz to 3 MHz 1.22%
Sigma 3 MHz by
FREQUENCY
Direct Method
(Source)
MECHANICAL- RPM Measurement
Using Digital 500 RPM to 4800
25 ACCELERATION of Centrifuge (Non- 2.48%
Techometer RPM
AND SPEED Contact Type)
Using Sound Level
MECHANICAL- Sound Level Meter
26 Calibrator by Direct 94 dB & 114 dB 0.77dB
ACOUSTICS @ 1 kHz
Method
MECHANICAL-
DIMENSION
Using Gauge Block
(BASIC
27 2 Pin Dial Caliper Set, Slip Gauge 10 mm to 200 mm 8µm
MEASURING
Accessories Set
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Gauge Blocks,
(BASIC Sine Bar, Dial
28 Angle Gauge Up to 45 Degree 10minute
MEASURING Indicator & Surface
INSTRUMENT, Plate
GAUGE ETC.)
MECHANICAL-
Using Electronic
DIMENSION
Angle Plate / Box Level, Surface Plate
(BASIC Up to 300 mm x 300
29 Angle Plate - Jack & Dial Indicator, 4µm
MEASURING mm
Flatness Cylindrical Square,
INSTRUMENT,
Gauge Block
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 6 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 7 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION Using Electronic
(BASIC Cylindrical Setting Probe with
35 3 mm to 200 mm 3.3µm
MEASURING Master Comparatpr Stand,
INSTRUMENT, Gauge Blocks
GAUGE ETC.)
MECHANICAL-
DIMENSION Depth Gauge - Using Gauge Blocks,
(BASIC Vernier / Dial / V-Block & Dial
36 0 to 600 mm 13µm
MEASURING Electronic (L.C.: 0.01 Indicator & Surface
INSTRUMENT, mm & coarser) Plate
GAUGE ETC.)
MECHANICAL-
DIMENSION Depth Micrometer - Using Gauge Block
(BASIC Analog / Digital Set, Surface Plate,
37 0 to 300 mm 5µm
MEASURING (L.C.: 0.001 mm & V-Block & Dial
INSTRUMENT, coarser) Indicator
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Dial
(BASIC Dial Bore Gauge Calibration Tester,
38 Up to 2 mm 2.5µm
MEASURING (L.C.: 0.001 mm) Electronic Probe,
INSTRUMENT, Digital Micrometer
GAUGE ETC.)
MECHANICAL-
Dial Calibration
DIMENSION
Tester / Dial Gauge Using Electronic
(BASIC
39 Calibrator (L.C.: Probe & Gauge 0 to 25 mm 1.5µm
MEASURING
0.0002 mm & Block
INSTRUMENT,
coarser)
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 8 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION
(BASIC Dial Snap Gauge
40 Using Gauge Blocks 1 mm to 200 mm 3µm
MEASURING (L.C.: 0.001 mm)
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Dial Thickness
(BASIC Gauge - Analog / Using Gauge Block
41 0 to 100 mm 6µm
MEASURING Digital (L.C.: 0.001 Set
INSTRUMENT, mm & coarser)
GAUGE ETC.)
MECHANICAL-
Electronic Probe
DIMENSION
with DRO /
(BASIC
42 Comparator with Using Slip Gauge 0 to 25 mm 0.9µm
MEASURING
Stand - Dial &
INSTRUMENT,
Electronic
GAUGE ETC.)
MECHANICAL-
DIMENSION
External Micrometer Using Gauge Block
(BASIC
43 - Analog / Digital Set, Optical Flat & 4 0 to 300 mm 4µm
MEASURING
(L.C.: 0.001 mm) Optical Parallels
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
External Micrometer Using Gauge Block
(BASIC
44 - Analog / Digital Sst, Optical Flat & 4 300 mm to 700 mm 12µm
MEASURING
(L.C.: 0.01 mm) Optical Parallels
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 9 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION
Using Electronic
(BASIC
45 Feeler Gauge Probe with Up to 1 mm 2.9µm
MEASURING
Comparator
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Fillet Welding Gauge Using Profile
46 Up to 200 mm 10µm
MEASURING - Linear Projector
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Optical Flat,
(BASIC Gauge Block Measuring Pin,
47 0 to 50 mm 2µm
MEASURING Accessories - Length Electronic Probe,
INSTRUMENT, Surface Plate
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Using Electronic
48 Hegman Gauge Up to 0.1 mm 3.8µm
MEASURING Probe with DRO
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Height Gauge - Using Caliper
(BASIC Vernier / Dial / Checker, Gauge
49 0 to 600 mm 9.9µm
MEASURING Electronic (L.C.: 0.01 Blocks, Electronic
INSTRUMENT, mm & coarser) Probe, Surface Plate
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 10 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION Using Gauge Block,
(BASIC Height Master (L.C.: Electronic
50 0 to 300 mm 5µm
MEASURING 0.001 mm) Comparator &
INSTRUMENT, Surface Plate
GAUGE ETC.)
MECHANICAL-
DIMENSION Internal Micrometer
Using Gauge Block &
(BASIC Stick Type - Analog /
51 Slip Gauge 50 mm to 325 mm 9.9µm
MEASURING Digital (L.C.: 0.001
Accessories Set
INSTRUMENT, mm & coarser)
GAUGE ETC.)
MECHANICAL-
Using Electronic
DIMENSION
Probe with
(BASIC Length Bar /
52 Comparator Stand, 100 mm to 500 mm 11.2µm
MEASURING Standard Length
Gauge Block, Length
INSTRUMENT,
Bar & Surface Plate
GAUGE ETC.)
MECHANICAL-
DIMENSION Lever Type Dial
Using Dial
(BASIC Gauge - Analog /
53 Calibration Tester & Up to 1 mm 2.3µm
MEASURING Digital (L.C.: 0.01
Electronic Probe
INSTRUMENT, mm)
GAUGE ETC.)
MECHANICAL-
Using Gauge Blocks,
DIMENSION
Electronic Probe
(BASIC Measuring Pin /
54 with Comparator 1 mm to 20 mm 3µm
MEASURING Standard Pin
Stand & Digital
INSTRUMENT,
Micrometer
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 11 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION Measuring Scale /
(BASIC Steel Scale / Taper Using Profile
55 0 to 150 mm 350µm
MEASURING Scale (L.C.: 0.5 mm Projector
INSTRUMENT, & coarser)
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Gauge Block,
(BASIC Micrometer Head
56 Electronic Probe & 0 to 25 mm 1.4µm
MEASURING (L.C.: 0.0002 mm)
Optical Flat
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Electronic
DIMENSION
Probe with
(BASIC Micrometer Setting
57 Comparator Stand, 25 mm to 500 mm 6.8µm
MEASURING Standard
Gauge Block, Length
INSTRUMENT,
Bar & Surface Plate
GAUGE ETC.)
MECHANICAL-
DIMENSION Parallel Thread Plug
Using Floating
(BASIC Gauge / Wear Check
58 Carriage Diameter 2 mm to 100 mm 4.9µm
MEASURING Plug Gauge -
Measuring Machine
INSTRUMENT, Effective Diameter
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC
59 Plain Plug Gauge Using Gauge Blocks Up to 200 mm 3µm
MEASURING
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 12 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION
(BASIC
60 Plain Ring Gauge Using LMM 3 mm to 100 mm 2.9µm
MEASURING
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Dial
DIMENSION Plunger type Dial
Calibration Tester,
(BASIC Gauge - Analog /
61 Electronic Probe, Up to 25 mm 3.1µm
MEASURING Digital (L.C.: 0.001
Comparator Stand &
INSTRUMENT, mm)
Slip Gauge
GAUGE ETC.)
MECHANICAL-
DIMENSION
Radius Gauge /
(BASIC Using Profile
62 Radius 0.6 mm to 25 mm 8.6µm
MEASURING Projector
Measurement
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Electronic
DIMENSION
Probe with
(BASIC
63 Riser Block - Length Comparator Stand, 0 to 600 mm 6µm
MEASURING
Gauge Block &
INSTRUMENT,
Surface Plate
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC
64 Snap Gauge Using Gauge Blocks 3 mm to 400 mm 3.3µm
MEASURING
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 13 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
Using Surface Plate,
DIMENSION
Spirit Level Tilting Table,
(BASIC 300 mm (base
65 (Sensitivity: 20 Electronic Level, 12µm/m
MEASURING length)
µm/m) Cylindrical Square
INSTRUMENT,
with Dial Indicator
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Surface Plate,
(BASIC Straight Edge - 4 SQRT L/125µm, L
66 Dial Indicator & Up to 5 m
MEASURING Straightness in mm
Electronic Level
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Taper Thread Plug
Using Floating
(BASIC Gauge / Wear Check
67 Carriage Diameter 2 mm to 100 mm 5µm
MEASURING Plug Gauge -
Measuring Machine
INSTRUMENT, Effective Diameter
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Using Digital Vernier
68 Test Sieve 10 mm to 125 mm 25µm
MEASURING Caliper
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Using Profile
69 Test Sieve 45 µm to 10 mm 10.7µm
MEASURING Projector
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 14 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION Using Electronic
(BASIC Probe with
70 Thickness Foil 10 µm to 2 mm 1.7µm
MEASURING Comparator Stand,
INSTRUMENT, Digital Micrometer
GAUGE ETC.)
MECHANICAL-
Using Gauge Block,
DIMENSION
Electronic Probe
(BASIC Thread Measuring 0.17 mm to 6.35
71 with Comparator 2.6µm
MEASURING Wire mm
Stand & Digital
INSTRUMENT,
Micrometer
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Thread Pitch Gauge - Using Profile
72 0.4 mm to 6 mm 6µm
MEASURING Pitch Projector
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Thread Ring Gauge -
73 Using LMM 3 mm to 100 mm 2.8µm
MEASURING Effective Diameter
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Ultrasonic Thickness Using Steel Slip
74 10 µm to 200 mm 8.4µm
MEASURING Gauge Gauge
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 15 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION Using Test Mandrel,
(BASIC Puppy Dial & Surface
75 V Block - Flatness Up to 200 mm 3.2µm
MEASURING Plate by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Test Mandrel,
(BASIC Puppy Dial & Surface
76 V Block - Parallelism Up to 200 mm 2.8µm
MEASURING Plate by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Test Mandrel,
(BASIC V Block - Puppy Dial & Surface
77 Up to 200 mm 6.1µm
MEASURING Symmetricity Plate by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Microscope - Linear Up to 30 mm x 30
78 Using Glass Scale 4.5µm
(PRECISION & Magnification mm
INSTRUMENTS)
MECHANICAL-
Profile Projector -
DIMENSION Using Angle Scale &
79 Angular (L.C.: 1 Up to 360 Degree 78.6second
(PRECISION Angle Graticule
second)
INSTRUMENTS)
MECHANICAL-
Profile Projector -
DIMENSION Up to X: 150 mm &
80 Linear (L.C.: 0.001 Using Glass Scale 4.7µm
(PRECISION Y: 100 mm
mm)
INSTRUMENTS)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 16 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
DIMENSION Profile Projector -
81 Using Glass Scale Up to 50X 8%
(PRECISION Magnification
INSTRUMENTS)
Using Rubber
MECHANICAL- Rubber Hardness
82 Hardness Calibrator 0 to 100 Shore A 1.9Shore A
DUROMETER Tester
by Direct Method
Using Rubber
MECHANICAL- Rubber Hardness
83 Hardness Calibrator 0 to 100 Shore D 1.9Shore D
DUROMETER Tester
by Direct Method
Analog / Digital
MECHANICAL- Using Digital
Pressure Gauge /
PRESSURE Compound (Pressure
84 Indicator with Up to 10 bar 0.02bar
INDICATING / Vacuum) Gauge as
Transducer -
DEVICES per DKD R 6-1
Hydraulic
Analog / Digital Using Digital
MECHANICAL-
Pressure Gauge / Pressure Gauge &
PRESSURE
85 Indicator with Oil-based Up to 700 bar 1.4bar
INDICATING
Transducer - Comparator Pump
DEVICES
Hydraulic as per DKD R 6-1
Analog / Digital
MECHANICAL- Using Digital
Vacuum Gauge /
PRESSURE Compound (Pressure
86 Indicator with (-) 0.8 bar to 0 bar 0.01bar
INDICATING / Vacuum) Gauge as
Transducer, Switch -
DEVICES per DKD R 6-1
Hydraulic
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 17 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
Using Standard
Weights of Class E2
with Precision
Volumetric Flask,
MECHANICAL- Balance (0 to 1 kg
87 Measuring Cylinder, 100 ml to 1000 ml 6ml
VOLUME with readability 1
Beaker
mg) as per ISO
4787: 2010 by
Gravimetric Method
Using Standard
Weights of Class E2
Volumetric Pipette, with Precision
MECHANICAL- Burette, Flask, Balance (0 to 200 g
88 0.1 ml to 100 ml 5.5µl
VOLUME Beaker, Measuring with readability 0.1
Cylinder mg) as per ISO
4787: 2010 by
Gravimetric Method
MECHANICAL-
Using M1 Class
WEIGHING Weighing Balance
89 Weights as per OIML Up to 100 kg 6.1g
SCALE AND (readability: 0.01 kg)
R-76-1
BALANCE
MECHANICAL-
Weighing Balance Using F1 & M1 Class
WEIGHING
90 (readability: 0.5 g & Weights as per OIML 1 kg to 30 kg 1.2g
SCALE AND
coarser) R-76-1
BALANCE
MECHANICAL-
Weighing Balance Using F1 & M1 Class
WEIGHING
91 (readability: 1 mg & Weights as per OIML 1 mg to 1 kg 20mg
SCALE AND
coarser) R-76-1
BALANCE
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 18 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
Weighing Balance Using E2 Class
WEIGHING
92 (readability: 1 mg & Weights as per OIML 1 mg to 200 g 0.7mg
SCALE AND
coarser) R-76-1
BALANCE
RTD, Thermocouple Using Low
with or without Temperature Dry
THERMAL- Indicator / Block Calibrator,
93 (-) 25 °C to 100 °C 0.46°C
TEMPERATURE Controller, PT-100 RTD (4 Wire)
Temperature Gauge, & S-Type
Recorder Thermocouple
RTD, Thermocouple
Using Dry Block
with or without
Calibrator, PT-100
THERMAL- Indicator /
94 RTD (4 Wire) & S- 100 °C to 400 °C 1.01°C
TEMPERATURE Controller,
Type Thermocouple,
Temperature Gauge,
Dry Block Calibrator
Recorder
RTD, Thermocouple
with or without Using Dry Block
THERMAL- Indicator / Calibrator, PT-100
95 400 °C to 1100 °C 4.74°C
TEMPERATURE Controller, RTD (4 Wire) & S-
Temperature Gauge, Type Thermocouple
Recorder
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 19 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
Site Facility
ELECTRO-
TECHNICAL-
Using Fluke 6½ DMM
ALTERNATING
1 AC Current @ 50 Hz 8846A by Direct 50 µA to 10 A 0.87 % to 0.6 %
CURRENT (< 1
Method
GHZ)
(Measure)
ELECTRO-
TECHNICAL-
Using Zeal HV Probe
ALTERNATING
2 AC High Voltage with Fluke DMM by 1 kV to 10 kV 3.82%
CURRENT (< 1
Direct Method
GHZ)
(Measure)
ELECTRO-
TECHNICAL-
Using Fluke 6½ DMM
ALTERNATING
3 AC Voltage @ 50 Hz 8846A by Direct 1 mV to 1000 V 4.4 % to 0.52 %
CURRENT (< 1
Method
GHZ)
(Measure)
ELECTRO-
Using Zeal 5½
TECHNICAL-
Digital Multi-function
4 ALTERNATING AC Current @ 50 Hz 0.2 mA to 20 A 3 % to 3.87 %
Calibrator By Direct
CURRENT (< 1
Method
GHZ) (Source)
ELECTRO- Using Zeal 5½
TECHNICAL- Digital Multi-function
5 ALTERNATING AC Current @ 50 Hz Calibrator with 20 A to 990 A 3 % to 3.87 %
CURRENT (< 1 Current Coil by
GHZ) (Source) Direct Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 20 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
ELECTRO-
Using Zeal 5½
TECHNICAL-
Digital Multi-function
6 ALTERNATING AC Voltage @ 50 Hz 5 mV to 990 V 1.81 % to 0.32 %
Calibrator by Direct
CURRENT (< 1
Method
GHZ) (Source)
ELECTRO-
TECHNICAL-
Capacitance @ 1 Using Sigma Decade
7 ALTERNATING 100 pf to 1 µf 1.81%
kHz Capacitance Box
CURRENT (< 1
GHZ) (Source)
ELECTRO-
Using Sigma RSI-6
TECHNICAL-
Decade Inductance
8 ALTERNATING Inductance @ 1 kHz 100 µH to 10 H 1.81%
Box by Direct
CURRENT (< 1
Method
GHZ) (Source)
ELECTRO-
TECHNICAL- Using Fluke 6½ DMM
9 DIRECT DC Current 8846A by Direct 10 µA to 10 A 4.03 % to 0.98 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using Zeal HV Probe
10 DIRECT DC High Voltage with Fluke DMM by 1 kV to 10 kV 2.6%
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using Fluke 6½ DMM
11 DIRECT DC Voltage 8846A by Direct 1 mV to 1000 V 0.71%
CURRENT Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 21 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 22 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
ELECTRO-
Using Radix
TECHNICAL- Temperature
Universal Calibrator (-) 200 °C to 1300
18 TEMPERATURE Indicator / Controller 2°C
by Direct Method as °C
SIMULATION / Recorder - K Type
per IS 1248
(Source)
ELECTRO-
Using Radix
TECHNICAL- Temperature
Universal Calibrator
19 TEMPERATURE Indicator / Controller 0 °C to 1600 °C 2.7°C
by Direct Method as
SIMULATION / Recorder - R Type
per IS 1248
(Source)
ELECTRO-
Temperature Using Radix
TECHNICAL-
Indicator / Controller Universal Calibrator
20 TEMPERATURE (-) 100 °C to 450 °C 1.3°C
/ Recorder - RTD by Direct Method as
SIMULATION
Type per IS 1248
(Source)
ELECTRO-
Using Radix
TECHNICAL- Temperature
Universal Calibrator
21 TEMPERATURE Indicator / Controller 0 °C to 1700 °C 2.1°C
by Direct Method as
SIMULATION / Recorder - S Type
per IS 1248
(Source)
ELECTRO-
Using Radix
TECHNICAL- Temperature
Universal Calibrator
22 TEMPERATURE Indicator / Controller (-) 100 °C to 400 °C 2.1°C
by Direct Method as
SIMULATION / Recorder - T Type
per IS 1248
(Source)
ELECTRO- Using Countron
TECHNICAL- Digital Timer / Stop 1050 Digital Timer
23 TIME & Watch - Analog / by Comparison 10 s to 6 hour 1.02 % to 0.11 %
FREQUENCY Digital Method as per IS
(Measure) 1248
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 23 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
ELECTRO-
TECHNICAL- Using Fluke 6½ DMM
24 TIME & Frequency 8846A by Direct 10 Hz to 1 MHz 2.86%
FREQUENCY Method
(Measure)
ELECTRO-
Using Zeal Multi-
TECHNICAL-
function Generator &
25 TIME & Frequency 10 Hz to 3 MHz 1.22%
Sigma 3 MHz by
FREQUENCY
Direct Method
(Source)
Using Sound Level
MECHANICAL- Sound Level Meter
26 Calibrator by Direct 94 dB & 114 dB 0.77dB
ACOUSTICS @ 1 kHz
Method
MECHANICAL-
DIMENSION
Bench Centre - Using Test Mandrel
(BASIC
27 Parallelism & Co- & Lever Dial Gauge 0 to 300 mm 11.1µm
MEASURING
axility as per IS 5980
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Microscope - Linear Up to 30 mm x 30
28 Using Glass Scale 4.5µm
(PRECISION & Magnification mm
INSTRUMENTS)
MECHANICAL-
Profile Projector -
DIMENSION Using Angle Scale &
29 Angular (L.C.: 1 Up to 360 Degree 78.6second
(PRECISION Angle Graticule
second)
INSTRUMENTS)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 24 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
Profile Projector -
DIMENSION Up to X: 150 mm &
30 Linear (L.C.: 0.001 Using Glass Scale 4.7µm
(PRECISION Y: 100 mm
mm)
INSTRUMENTS)
MECHANICAL-
DIMENSION Profile Projector -
31 Using Glass Scale Up to 50X 8%
(PRECISION Magnification
INSTRUMENTS)
MECHANICAL-
3.7 SQRT
DIMENSION Surface Plate - Cast Using Electronic
32 Up to 2 meter {(L+W)/125}µm, L
(PRECISION Iron / Granite Level Meter
& W in mm
INSTRUMENTS)
Analog / Digital
MECHANICAL- Using Digital
Pressure Gauge /
PRESSURE Compound (Pressure
33 Indicator with Up to 10 bar 0.02bar
INDICATING / Vacuum) Gauge as
Transducer -
DEVICES per DKD R 6-1
Hydraulic
Analog / Digital Using Digital
MECHANICAL-
Pressure Gauge / Pressure Gauge &
PRESSURE
34 Indicator with Oil-based Up to 700 bar 1.4bar
INDICATING
Transducer - Comparator Pump
DEVICES
Hydraulic as per DKD R 6-1
Analog / Digital
MECHANICAL- Using Digital
Vacuum Gauge /
PRESSURE Compound (Pressure
35 Indicator with (-) 0.8 bar to 0 bar 0.01bar
INDICATING / Vacuum) Gauge as
Transducer, Switch -
DEVICES per DKD R 6-1
Hydraulic
This is annexure to 'Certificate of Accreditation' and does not require any signature.
PINPOINT PRECISION CALIBRATION LAB PVT. LTD., LG-33, ARORA SHOPPERS PARK,
Laboratory Name :
SHAKTI KHAND-2, INDIRAPURAM, GHAZIABAD, GHAZIABAD, UTTAR PRADESH, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2576 Page No 25 of 25
Validity 23/06/2020 to 19/02/2022 Last Amended on 07/07/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrum
MECHANICAL-
Using M1 Class
WEIGHING Weighing Balance
36 Weights as per OIML Up to 100 kg 6.1g
SCALE AND (readability: 0.01 kg)
R-76-1
BALANCE
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.
This is annexure to 'Certificate of Accreditation' and does not require any signature.