Determination of Silica in Fluorspar by Silico-Molybdate Visible Spectrophotometry
Determination of Silica in Fluorspar by Silico-Molybdate Visible Spectrophotometry
Determination of Silica in Fluorspar by Silico-Molybdate Visible Spectrophotometry
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
E463 − 14a
temperature to 1050 °C. Maintain at 1050 °C for at least 5 min. crucible to the beaker. Cover the beaker with a polyethylene
Cool to room temperature in a desiccator. sheet and secure it to the beaker with a rubber band.
7.5 Sodium Borate(Na2B4O7)—Anhydrous powder, low- NOTE 3—Hold the crucible while cooling to avoid contamination
silica content. problems. This prevents damage to the beaker.
NOTE 1—If low silica sodium borate is not available, prepare the 10.8 Place the beaker on a steam bath and swirl occasionally
reagent as follows: Transfer 247 g of boric acid to a large platinum dish. until the melt is completely dissolved (Note 4). Cool, remove
Expel water by gradually increasing the temperature to about 1000 °C. and rinse the platinum crucible and cover, and add the
When effervescence ceases, gradually introduce 106 g of sodium carbon-
ate into the molten mass. Maintain at a temperature of about 1000 °C until washings to the beaker. Transfer the solution to a 250-mL
a clear melt is obtained. volumetric flask. Rinse the beaker and add the rinsings to the
flask. Dilute to volume, mix, and examine the solution for any
8. Hazards insoluble material (Note 4). Transfer the solution to a dry
8.1 For precautions to be observed in this method, refer to polyethylene bottle.
Practices E50. NOTE 4—Complete dissolution of the melt requires about 2 h. In cases
of incomplete dissolution, a new sample must be taken.
9. Sample Preparation
10.9 Transfer 50-mL aliquots of the blank and sample
9.1 The analytical sample shall be pulverized, if necessary, solutions to 200-mL polyethylene or TFE-fluorocarbon bea-
to pass a 150-µm sieve (see Test Method E276). Dry at 105 °C kers.
to 110 °C for a minimum of 1 h.
10.10 Transfer (10.0, 20.0, 30.0, 40.0, and 50.0)-mL ali-
10. Procedure quots of the standard solutions to 200-mL polyethylene or
10.1 Transfer 7 g of Na2B4O7 to each of six 25-mL platinum TFE-fluorocarbon beakers. Dilute, if necessary, to a 50-mL
crucibles. Form a cavity in the center of the flux. volume with the remaining blank solution.
10.2 Into Crucibles 1 and 2, weigh 0.100 g to 1 g of the dry NOTE 5—Since commercially available Na2B4O7 frequently contains
sample. Choose sample weights to provide from 5 mg to 10 mg appreciable amounts of silica, each standard and sample solution must
contain the same amounts of this reagent. The dilution of the aliquots to
of SiO2. a 50-mL volume should, therefore, be carried out using a buret.
10.3 Into Crucibles 3 and 4, weigh 10.0 mg of SiO2 reagent. 10.11 Determine the pH of the blank, standard, and sample
10.4 Crucibles 5 and 6 serve as blanks. solutions using a pH meter. If the pH of the solutions lies
10.5 Mix the contents of the crucibles with a platinum or between 0.5 and 0.9 and within 0.1 units of each other, proceed
polyethylene rod. Transfer adhering particles to the crucible. to 10.12. If the pH lies outside these parameters, adjust the pH
with HCl (1 + 1).
10.6 Cover the crucible and heat gently until moisture is
expelled. Increase the temperature until complete fusion re- NOTE 6—Accurate pH adjustments are essential for maximum color
sults. development and color stability.
RSD 5 s 100/X̄ d œ od 2
/sn 2 1d (2)
where:
X̄ = average concentration, %,
d = difference of the determination from the mean, and
n = number of determinations.
2
E463 − 14a
separatory funnel, shake for 30 s, allow the phases to separate, 13. Calculation
and discard the acid layer. Repeat the washing twice more.
13.1 Subtract the average absorbance of the blank solutions
10.14 Transfer the butyl alcohol phase to a dry 100-mL from absorbances of the sample solutions. Convert the net
volumetric flask. Wash the separatory funnel twice with 1-mL absorbance of the sample solution to milligrams of SiO2 by
or 2-mL portions of butyl alcohol, and add the washings to the means of the calibration curve. Calculate the percentage of
volumetric flask. Add 1 mL of ethyl alcohol, dilute to volume SiO2 as follows:
with butyl alcohol, and mix.
Silica, % 5 A/ ~ B 3 10! (1)
11. Photometry
A = silica found in the aliquot used, mg, and
11.1 Concentration Range:
B = sample represented by the aliquot, g.
11.1.1 The recommended concentration range is from
0.2 mg to 2.0 mg of silica per 100 mL of solution, using a 1-cm
cell. 14. Precision and Bias
NOTE 7—Cells having other dimensions may be used, provided suitable 14.1 Precision—Table 1 indicates the precision of the test
adjustments can be made in the amounts of sample and reagent used. method between laboratories.
11.2 Adjust the photometer to the initial setting using water 14.2 Bias—No information on the accuracy of this test
as the reference solution. While maintaining this setting, take method is known. The accuracy of this test method may be
photometric readings of the blank, standard, and sample judged by comparison of accepted values for standard refer-
solutions using a light band centered at approximately 400 nm. ence materials with the mean determined through interlabora-
11.3 Stability of Color: tory testing.
11.3.1 After the addition of the ammonium molybdate, color
is fully developed within 10 min and is stable after extraction 15. Keywords
with butyl alcohol
15.1 fluorspar; photometric; silica; silico-molybdate magne-
12. Preparation of Calibration Curve sium
12.1 Subtract the average absorbance of the blank solution
from the average absorbance of each standard solution and plot
the net absorbances against milligrams of silica per 100 mL of
solution.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/