JPD-HA200

Electronic Blood Pressure Monitor

User Manual

Version: V1. 0

Date of revision: January 2019


Foreword
Dear user:

Thank you for purchasing our Electronic Blood Pressure


Monitor. Please read this manual carefully and use correctly.
Please keep the user manual properly for checking and
reference at any time .The Electronic Blood Pressure Monitor is
suitable for adults.

Product Profile:

The full name of the product is electronic blood pressure


monitor. Its product structure consists of the main body and a
cuff. It is suitable for measurement of human systolic pressure,
diastolic blood pressure and pulse. The blood pressure values
obtained by this equipment are equivalent to that measured with
the auscultatory method.

The electronic blood pressure monitor adopts intelligent


pressure mode, which can automatically pressure to an
appropriate pressure value according to your blood pressure to
effectively reduce the discomfort caused by incorrect pressure,
shorten the measuring time, and prolong the service life of the
cuff.

The electronic blood pressure monitor is equipped with


the memory function of 99 groups of measurement data for
each of two users, and the data of the two users can be saved
respectively.
Contents
1、Unpacking Inspection...........................................................- 1 -
2、Packing List..........................................................................- 1 -
3、After-sales Service............................................................... - 1 -
4、Safety Precautions................................................................ - 2 -
5、Product Composition............................................................- 3 -
6、Intended for Use................................................................... - 3 -
7、Product Parts.........................................................................- 3 -
8、Installation Before Use.........................................................- 4 -
9、Function Setting Mode......................................................... - 6 -
10、Method for Correct Use......................................................- 9 -
11、General Knowledge of Blood Pressure............................ - 14 -
13、Common Q & A on Blood Pressure................................. - 20 -
14、Abnormal Phenomena and Handling............................... - 22 -
15、Cleaning and Disinfection................................................ - 24 -
17、Precautions for Keeping................................................... - 26 -
18、Product Specification....................................................... - 27 -
19、Statement of Electromagnetic Compatibility................... - 29 -
1、Unpacking Inspection

Before use, please open the package carefully, check


whether all the parts are available according to the following
packing list and whether the parts are damaged during
transportation, and then install and operate in strict accordance
with the manual. If there is any damage or doubt on operation,
please contact our distributors or our company directly. For
contacts mentioned above, you need to provide the following
information, the model of the equipment, the serial number, the
purchase date and your contact information and address.

2、Packing List
No. Name Quantity
Electronic blood
1 1
pressure monitor
2 Cuff 1

3 User manual 1

3、After-sales Service
After-sales service unit: Shenzhen Jumper Medical Equipment Co., Ltd.
Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan,
Shenzhen ,Guangdong,China
Tel: +86-755-26692192, 26696279
Fax: +86-755-26852025
Website:www.jumper-medical.com
Postal Code: 518103

- 1 -
Authorized European Representative:
Wellkang Ltd
Suite B, 29 Harley Street , London W1G 9QR,UK

4、Safety Precautions

The warnings and illustrations shown in the manual are


intended to enable you to use the product safely and correctly,
thus preventing harm to you and others, specific meanings of
which are shown as follows:

Legend, mark and meaning

Warning information

BF-type anti –shock degree for the application part

When the product life expires and the product is to


be discarded, it has to be taken to the designated
places for disposal according to requirements of
the local environmental protection department.

Consult the instructions for use.

This product complies with the MDD93/42/EEC


requirements.

IP21 Degree of protection against the ingress of water.

- 2 -
5、Product Composition

This product is composed of host machine and cuff of an


electronic blood pressure monitor.

6、Intended for Use

The electronic blood pressure monitor is intended to measure


the systolic pressure and diastolic pressure, as well as the pulse
rate of adult person via non-invasive oscillometric technique at
medical facilities or at home.

7、Product Parts
(1) Main body

Air jack

User key

Start / Pause key

USB Memory key


external
power
supply
interface

- 3 -
(2) Display screen

User icon Low battery icon

Date / time display

Unit
Systolic pressure

Blood Pressure
Diastolic pressure Level Indicator

Heartbeat icon Memory icon

Pulse Memory Number

Arrhythmia Average value


Symbol

8、Installation Before Use

(1) Install the battery

Open the battery cover and install


the batteries in the battery
compartment.
The battery compartment is positioned at the bottom of the
electronic blood pressure monitor.

- 4 -
a) Open the battery cover according to the method shown in the
figure (as shown in the above figure)

b) Place 4 AAA dry batteries in the battery compartment, and


pay attention to the electrode indication of the batteries. Install
the battery as indicated in the picture right under this sentence.

(2) Battery power indication and replacement

If the low battery symbol appears on the screen after the


product is turned on , the measurement cannot be performed
and the batteries must be replaced.

Please use 4 AAA alkaline dry batteries ;

Do not use any expired battery;

If the product is not used for a long time(over 3 months),


please take out the batteries.

(3) USB connection for power supply

The product power can be supplied by plugging into DC 5V


- 5 -
external power supply through a USB interface.

(4) Cuff

The applicable arm circumference rang of the cuff is 22-36cm


or 22-42cm.
Connection: Insert the air plug on the cuff into the air jack on
the left side of the electronic blood pressure monitor. As shown
in the following figure :
-The cuff is the application part.

Air jack

Air plug
Cuff

Air tube

9、Function Setting Mode


(1) User mode
In the shutdown state, press the "User " key for 3 seconds,
the screen displays "User 1 " or "User 2 ", and flickers
continuously. Press the "Memory " key to switch to the other
user. (System default user: User 1 )
User 1 User 2

Press the Switch over


"Memory" key

- 6 -
(Note: In the shutdown state, press "User " key to view the
current time and the current user)

(2) Year & month / Date setting


In the shutdown state, press the "User " key for 3 seconds to
enter the setting mode. Press the "User " key to enter the
year setting mode ( or automatically enter the year setting mode
after the batteries are reloaded and the product is powered on ),
the screen displays 20XX and flickers continuously. press the
"Memory " key to increase the value which can be switched
between 2019 and 2039. press the "User " key for
confirmation and enter the month setting mode. After entering
the month setting mode, the screen will display xx-xx and flash
on the month. Press the "Memory " key to increase the value
which can be switched between 1 and 12, press the "User "
key for confirmation and enter the date setting mode. After
entering the date setting mode, XX-XX is displayed on the
screen and flickers on the date display. Press the "Memory "
key to increase the value which can be switched between 01
and 31, press the "User " key for confirmation and enter the
time setting mode. After entering the time setting mode, the
screen will display XX: XX and it will flicker continuously,
press the "Memory " key to increase the value which can be
switched between 0 and 23, press the "User " key for
confirmation and enter the minute setting mode, press the
"Memory " key to increase the value which can be switched
between 00 and 59, press the "User " key for confirmation
and enter the talking function setting.

- 7 -
(3) Talking function setting (voice version only)
After entering talking function setting mode, " " will
show up at the SYS area, with the current voice status
" " / " " and voice icon (not displayed, when
talking function mode is off) at the DIA area. Press the
"Memory " key to switch the voice status, and press
the "User " key for confirmation and enter the unit
display setting.

Press the
"Memory" key Switchover

Press the Confirm


"User" key

- 8 -
(4) Unit display setting
After entering the unit setting mode, the SYS area will display
" " and unit area will display mmHg, it means the unit
mmHg is selected (system default unit: mmHg), At this time,
press of the "Memory " key, the unit will be switched, it
means the unit kPa is selected. After selecting the unit, press the
"User " key again to confirm the saving.

mmHg mode KPa mode


10、Method for Correct Use
Requirement:Please read the user manuals carefully before
use , user must follow user manuals to use.
(1) Measurement of blood pressure-preparation
before measurement
--Before measurement, avoid eating, smoking or any form of
heavy physical labor, as it will affect the measurement result.
--Before measurement, sit on a chair to have a rest for about
10min in a quiet environment.
--Before measurement, remove the clothing fettering the
measured arm.
--Measure the same arm (generally the left arm).
--Perform fixed-time measurement daily, because the blood
pressure on the same day is also changing.

(2) Measuring blood pressure - common factors


leading to erroneous measurements
- 9 -
--Comparative blood pressure measurement must be carried out
under the same conditions! Generally, the measurement should
be carried out in a quiet environment!
--Exert strengths on the arm may increase your blood pressure,
so make sure you're in a comfortable and relaxed state before
measurement, and do not exercise the muscles of the arm
during the measurement. If necessary, a cushion can be used to
support your arm.
--If the cuff is placed below or above the heart, the
measurement results will be incorrect.

(3) Measuring blood pressure - precautions


--Please use the original cuff that has passed the clinical test!If
you use a component that is not supplied by the manufacturer, it
may cause measurement errors.
--Loose cuff or exposed air bag will lead to wrong
measurement values.
--Repeated measurements can cause blood concentration in the
arm, which will affect the measurement results. So you should
rest for another 3 minutes or raise your arm for a while, so that
the concentrated blood flows away, and then repeat the
measurements.

( 4 ) Measuring blood pressure - how to use the cuff


1. Lay the cuff flat on the table with the hook and loop down,
and pass the end of the cuff through the metal ring to form a
circle. At this time, the hook and loop of the cuff will face the
outside (skip this step if the the ring has already been formed by
the cuff).

2. Put the cuff into the upper arm to be tested, wear the cuff
correctly according to the downward icon " " of the cuff, and
- 10 -
make the cuff air plug point to the lower arm.

3. Apply the cuff to upper arm to ensure that the lower edge of
the cuff is 2-3 cm from the elbow and the air tube is located at
the inner side of the arm.

4. Tighten the free edge of the cuff and glue the hook and loop.

5. The cuff should be comfortably wrapped around the upper


arm, with the elasticity space of two fingers. Remove any
clothing (e.g pullover) that restricts your arm before
measurement.

6. Place the lower arm on the table with the palm facing upward.
Sit with straight back, and make the center of the cuff and the
heart at the same height. Please note that the rubber tube on the
cuff cannot be folded or bent.

Note: If your left arm cannot be used for measurement, please


use your right arm. But in any case, all measurements must be
made on the same arm for the convenience of measurement
comparison.

(1) Winding the (1-1) Wound cuff (2) Wear the cuff (3) The cuff is 2-3cm away from
cuff the elbow

The cuff is on the


same level as the
heart

(4) Stick the (5) Adjust the (6) Measuring (7) Measure with the
hook & loop cuff posture same arm

- 11 -
(5) Starting measurement
Start measurement after wearing the cuff:
1. Press the "On / Off " key, the machine will automatically
return to zero. The air pump starts to inflate the cuff, and the
change of air pressure in the cuff will be displayed on the
screen.
2. When inflated to stable pressure, stop inflating, the pressure
in the cuff will decrease gradually and be displayed on the
screen. When the air pressure is insufficient, the machine will
automatically reinflate to higher pressure.
3. When the pulse is detected, the screen will display the
"heart " symbol and start flashing.
4. At the end of the measurement, the systolic pressure,
diastolic pressure, and pulse measurements will be displayed on
the screen.
5. The screen will continue to display the measurement results,
unless you press the "On / Off " key to turn off the machine.
In no operation state, the machine will be shutdown
automatically after 30 seconds.

Note:After measurement, the measurement results and


information (systolic pressure, diastolic pressure, pulse,
measurement completion time, current user group, whether the
heart rate is irregular) will automatically be saved to the
memory, the group of which is 01. The number of memory
groups is done in the same manner.

※The default measurement results show that the unit is mmHg.

※When the measurement results are wrong, Err is displayed in


the SYS area and the numerical value is displayed in the DIA
area.

※ In the measurement process. If it is in the process of


- 12 -
inflation/automatic deflation, press the "User " key and the
"Memory " key, there will be no response.

※In the measurement process, press the "On / Off " key to
turn off the machine.

※ The automatic shutdown time starts after the value of the


measurement is displayed.

※After the measurement is completed, press the "Memory "


key to switch to the memory mode, and press the "User " key
to switch to another user group.

(6) Using the memory function


The blood pressure monitor can automatically store 99 groups
of memory values for two users, and automatically replace the
99 groups of memory data when the memory is full. When the
blood pressure monitor is used several times, press the
"Memory " key in the shutdown state to display the average
blood pressure of the last three times, and then press the
"Memory " key again to display the first group of blood
pressure measurement values and look up the memory group
number. At the same time, press the "User " key to look
down the memory group number until the last group of blood
pressure measurement values are shown.

(7) Memory clearance function


1: Delete all memory values of the current user
In the shutdown state, long press the "Memory " key for 3
seconds, and then long press the "User " key for 3 seconds
until " " is displayed in the high-voltage area and " " is
displayed in the low-voltage area, thus all of the memory values
- 13 -
of the current user are deleted.

11、General Knowledge of Blood Pressure


(1) General knowledge of blood pressure-definition
of blood pressure
Blood pressure is the pressure that the blood is transmitted to
the artery through the heart beat (systole and diastole).
When systole, the pressure at which the blood is transmitted to
the artery is called systolic pressure, i.e. the high pressure; the
blood circulating the whole body will backflow to the heart,
and when diastole, the pressure is called diastolic pressure, i.e.
the low pressure.

(2) Classification standard of blood pressure

- 14 -
WHO standard blood pressure values are shown below.
Hypotension has not been defined. It is generally considered
that high pressure lower than l00mmHg (13. 3kPa) is defined as
hypotension.

12、Contraindications, Precautions, Warnings and


Prompt Instructions

(1) Warnings:
Do not let a child below 12 years old and the people
who can’t express one’s intention .When it used by the
people of 12~18 years old ,it should accompanied by the
- 15 -
adult .This device has not been tested for use by pregnant
women .
Make sure there is no connection tubing kinking before
start measuring to avoid can not be measured or other injury.
For any patient, do not measure more than 3 times
continuously, it should be at least above 5 minutes of interval
rest between any two measurements,otherwise will cause
extravasated blood.
Do not apply the cuff over a wound as this can cause
further injury.
Do not use cuff in the arm which has invasive treatment
device or arteriovenous shunt,otherwise will cause hurt.
When using the mismatched cuff.,it will make the
monitoring equipment failure.
Please use the cuff follow the use manual,pay attention to
the use of the hose,avoid twining due to excessive length.
Do not use the cuff for a long time,avoid allergies.
If an incorrect reading appears, check the instructions and
check for equipment failure.
Do not make self-diagnosis according to the measurement
results. Please consult your professional doctor with the
measurement result record. The treatment based on the
self-judgement of measurement results will be very dangerous;
Please give the results of the measurement to the doctor
who knows your health and receive the diagnosis;
If you take some medicine, please follow the doctor's
advice;
- 16 -
For those with peripheral circulation disorder caused by
certain diseases, there might be significant difference in the
blood pressure value measured on the wrist and upper arm;
Do not make measurements in 1h after meals or
immediately after strenuous exercise, shower, smoking, alcohol
drinking, coffee drinking or black tea drinking;
Do not use the Electronic Blood Pressure Monitor for the
purpose other than blood pressure measurement; otherwise, it
might cause accident or failure;
Please use the Electronic Blood Pressure Monitor under
the guidance of doctor;
This model of Blood Pressure Monitor is suitable for the
arm circumference range of 22-36cm or 22-42cm, and if
beyond this range, you might not obtain the correct measured
value of blood pressure;
This model of Blood Pressure Monitor is not suitable for
newborns or young children;
The blood pressure is constantly changing. You cannot
rely on only one blood pressure measurement result to judge the
blood pressure condition. The repeated measurement data over
a period of time will be more reliable;
Do not bind too tightly. It will be the most appropriate if
you can relax and can put one finger in. If you do not bind
correctly or put it in place, you might not obtain the correct
measured value of blood pressure;
Do not measure the non-invasive blood pressure (NIBP)
for the patients with sickle cell disease or the patients with skin
injury of limb for blood pressure measurement.

- 17 -
Please do not use mobile phone, computer,electric kettle
and the other can cause interference device around the
device.Please do not use the device around the magnetic field.
Please keep to maintain the device ,during use it.

(2) Precautions:

Do not repair, disassemble or modify the electronic blood


pressure monitor without permission;
Do not collide or fall down the host to avoid collision or
strong impact of the machine;
Do not use the electronic blood pressure monitor
immediately after it is taken out from the place below 0°
C ,which should be at least placed for more than 1h in the
environment at the temperature of 5°C-40°C and the
humidity of 80%;
When the device is taken into a ambient where
temperature is 20°C from the highest storage temperature
55°C ,make sure the device has been in the ambient
environment for more than 1 hour before used.
There is no need to check whether the operation of the
sphygmomanometer will have a lasting effect on the patient's
blood circulation,if used according to the instructions.
Except the cloth stained with warm water or soapy water
for wiping the host shell, do not use other articles for wiping;
Do not pull the cuff forcefully or use the cuff as a rag;
Do not be contaminated by the dust or corrosive liquid;

- 18 -
Do not touch the host when measurement;
Remove the battery when the product is left unused for a
long time;
Do not store the electronic blood pressure monitor in a
place with high temperature, moisture or direct sunlight;
Calibrate the electronic blood pressure monitor
periodically.
Do not combine old and new batteries,different brands
batteries for using.

(3) Indicative description

The patient is an intended operator.


Before measurement, calibrate the electronic blood
pressure monitor, check the function of each component, and
maintain the good performance of electronic blood pressure
monitor .
During blood pressure measurement, the measured limb
shall be at the same level at the heart.
Indoor temperature, environment, noise, subject's body
position, speech or exercise might have effect on the blood
pressure measurement. The cuff wearing, improper sensor
position, body movement or presence of electric field might
have effect on the measurement.
We do not recommend that patients with weak pulse,
severe upper gastrointestinal bleeding or severe arteriosclerosis
or elderly with poor peripheral circulation, particularly patients
with peripheral vascular lesion caused by complicated severe
diabetes use the product.
- 19 -
In case of rescue of seriously ill patients, diagnosis of
hypertension and direction of hypertensive medication, use the
electronic blood pressure monitor with caution.
Keep the power supply enough.
After product packaging, it is allowed to use common
vehicle transportation, but rain, moisture, squeezing or
mechanical collision should be avoided.

13、Common Q & A on Blood Pressure

Q1. Why is the blood pressure measured at home lower


than that measured in hospitals?

1. Blood pressure measured at home is 20 mmHg-30 mmHg (2.


7 kPa-4. 0 kPa) lower than that measured in hospitals. It is
because people are nervous when measured in hospitals, while
they tend to be in a stable mood when measured at home. It is
important to understand the normal blood pressure when you
are at home.
2. If the cuff is placed in a position above the heart, the
measured blood pressure is low, so please measure in the
correct posture.

Q2. Why is the blood pressure measured at home higher


than that measured in hospitals?
1. The anti-hypertensive drugs lose the efficacy.
→Please accept the doctor's instruction.
2. The cuff is not twined correctly.
If the cuff position is wrong, no arterial signal will be caught,
and the measured blood pressure value will be higher.
→Please confirm whether the cuff position is correct.

- 20 -
3. The cuff is rolled too loose.
If the cuff is rolled too loose, If the cuff is rolled too loose,
the compression force might fail to be transmitted to the artery,
resulting that the measured blood pressure value is higher than
the actual value.
→Adjust the gap between the arm and the cuff and tighten the
cuff.
4. The posture is incorrect when measurement.
If bending down, sitting cross-legged or bending over the sofa
or low table when measurement, the blood pressure might be
increased due to the abdominal pressure or due to the arm
position lower than the heart.
→ Please make measurement in correct posture.

Q3. Why I feel painful or numb due to tightening of the cuff


when measurement?
When blood pressure measurement, the cuff shall be tighten
until the arterial blood flow is stopped temporarily. So you
might feel painful or numb for a while, which is not harmful for
your body.And you can always stop the device from inflating
the cuff by pressing the on/off button,or by manually removing
the cuff from the user’s arm to assure your safety.

Q4. When will the measurement be better?


Please make measurement in the morning after you get up to
urinate or when your body and mood are in stable state. It is
better to make measurement at the same time every day.

Q5.Why the blood pressure value measured each time is


different?
1.When systole each time, the blood pressure will change to
some extent. For example, a person with the pulse of 70 beats
per minute will have 100,800 blood pressure changes every day.
Because the blood pressure is constantly changing, it is difficult
- 21 -
to obtain the correct blood pressure value by measurement only
once. Please make measurement for 2~3 times. The first
measurement will generally be higher due to nervousness or
inadequate preparation, and then when the second measurement,
the nervous emotion will be slightly alleviated, so generally, the
second measurement will be 5mmHg-10mmHg (0.7kPa-1.3kPa)
lower than the first time. This will be more obvious for those
with higher blood pressure.

--When continuous measurement, please note that:


There might be extravasated blood because the arm is
compressed, resulting that the finger tip blood does not flow
smoothly, If you continue the measurement in case of
extravasated blood, you cannot obtain the correct measured
value. Loosen the arm band, raise your hand over the head, and
grasp and stretch your left and right palms for 15 times
repeatedly. Then the extravasated blood can be dissolved, and
you can continue the blood pressure measurement.

2.Cuff position and twining method. The measured value varies


with the cuff size. Particularly, if the cuff is twined round the
elbow, you cannot obtain the correct measured value.
--Please use the correct cuff twining method for measurement.
The arm circumference range of the enclosed cuff is 22~36cm
(center of the upper arm).If the model is inconsistent, please
purchase separately.

14、Abnormal Phenomena and Handling


* When the measurement is abnormal, the following symbols
will appear. Please use the correct method for measurement.
LCD display information Reason / Meaning

Err is displayed in the The leakage is too fast or the pulse


- 22 -
SYS area, signal is too weak. Please check the
cuff, tie it up and try again
1 is displayed in the
DIA area
Err is displayed in the high
The blood pressure signal cannot
SYS area,
be detected due to too much noises.
Please remove the noise sources
2 is displayed in the
and measure again
DIA area.
Err is displayed in the
SYS area,
The test result is abnormal. Please
measure again
3 is displayed in the
DIA area
Err is displayed in the
SYS area,
The inflation fails. Please check the
cuff, tie it up and try again.
P is displayed in the
DIA area.
HI is displayed in the The inflation pressure is greater
DIA area than 307±8 mmHg ( 41 kPa±1
kPa ), Please measure again.

*Trouble shooting
Abnormal
Item to be checked Measure
phenomenon
Replace the
batteries with new
No display by
Check whether the ones or plug the
pressing the
battery level is low machine into a
power switch
Micro USB for
power supply.

- 23 -
Install in
Are the positive and accordance with
negative poles of the the positive and
batteries reversed? negative poles of
the batteries
Insert the tracheal
Is the tracheal plug
plug into the bore
tight?
No pressure tightly
Is the trachea broken
Buy a new cuff
or leaking?
Keep your arms
Whether your arm is
and body free
Error display, moved when you apply
from random
unable to measure pressure
movements
Whether you talks in Keep quiet and
the course of measure blood
measurement pressure
Is the cuff wrapped too Please tighten the
loose? cuff
The cuff leaks
The airbag of the cuff Replace with a
is broken new cuff

If blood pressure still cannot be measured normally with the


above method, please contact the distributor and do not
disassemble it by yourself!

15、Cleaning and Disinfection


(1) Cleaning:
Recommended cleaning agent:
①Household neutral cleaning agent.
Cleaning steps:
1) Remove the batteries before cleaning.
2) The body of the blood pressure monitor can be cleaned with
- 24 -
clean soft cloth.
3) The body can be wiped using a soft and clean cloth with a
small amount of neutral cleaner or water.
Do not use corrosive detergents. During the cleaning,
be careful not to immerse any part of the blood pressure
monitor in the liquid to prevent liquid from flowing into
the machine.

It is suggested to clean once every week. Complete the


cleaning in 3min each time. The number of repeated
cleaning each time shall not exceed 3 times.

(2) Disinfection:
Recommended disinfecting agent:
① Isopropanol solution with 70% concentration
Disinfection steps:
1)Wipe the machine body with a clean soft cloth stained with a
small amount of the above disinfectant, and dry it immediately.
2)The body of the Electronic Blood Pressure Monitor can be
wiped with a clean cloth stained with a small amount of 75%
medical alcohol for disinfection.

Do not disinfect by such method as high temperature


steam or ultraviolet irradiation, which might damage
the instrument or accelerate the aging!
It is suggested to disinfect the electronic blood
pressure monitor before and after use each time. Each
time of disinfection shall be completed within 1min. The
number of repeated disinfection each time shall not exceed 2
times.
Cleaning and disinfection shall be carried out in the
following environment: temperature: +5°C~+40°C
- 25 -
(
50 -104 ), relative humidity: 15%~80%RH, non-condensing,
atmospheric pressure: 70kPa~106 kPa.

16、Upkeep and Maintenance


- Always keep the surface of
Blood Pressure Monitor clean

Water or neutral
detergents
and tidy, helpful to prolong the
service life of Blood Pressure
Monitor.
- If the host is dirty, please wipe
with a dry soft cloth. If the dirt
cannot be eliminated easily, wipe
with a soft cloth stained with
water or neutral detergent, and
then dry with a dry cloth.
-No maintenance or maintenance
required when using the device .
Do not allow water or other liquids to enter the main
body.
17、Precautions for Keeping
- Do not wipe the host and cuff - Do not clean or moisten the
with volatile oil, diluent or cuff.
gasoline.
Volatile oil
Diluent

Gasoline

- 26 -
- Do not store the product at a - If the electronic blood
place which has direct sunlight, pressure monitor is not in use
high temperature, humidity, for a long time (over 3
dust or corrosive gas. months), take out of the
batteries from the host.

If you fail to observe the above precautions and other proper usage
methods to cause machine failure, the company will not be
responsible for the quality issues.

18、Product Specification
Product name Electronic Blood Pressure Monitor
Model JPD-HA200
Display mode Digital display
Measuring Oscillographic method
method
Measuring Upper arm
part
Blood 30mmHg-255mmHg
Measuring Pressure (5.3kPa-37.3kPa)
range value
Pulse rate 40-199 pulse rates / minute
Static Pressure ±3mmHg(±0. 4kPa)
measurement value
accuracy Pulse rate ±5% of the read value
LCD display Pressure Unit: mmHg / kPa.

- 27 -
Pulse Pulse rate per minute, showing
three digits
Memory 99 groups of measurement data for each of two
capacity users
4 AAA dry batteries or DC 5V USB external
Power supply
power supply
Power off Manual shutdown / 30s delayed automatic
mode shutdown
Body weight About 195g±5g
(excluding
batteries)
Dimension 120mm(length)* 90( width)* 43mm(height)
Screen size 59mm(length)* 47mm( width)
Upper arm cuff (measuring arm circumference
Cuff
22-36cm or 22-42cm )
Accessories Cuff, User manual
High performance dry batteries can be used
Battery life about 300 times under normal temperature
conditions
Service life 5 years
Date of See the label
environment
Temperatur If the product is
5℃-40℃
e condition stored or used
Humidity 15%-90%R outside the
Working condition H specified
environment temperature and
Atmospheri 70kPa-106k humidity range, it
c condition Pa will not work
properly.
Storage and Strong impact, direct collision, exposure to the
transportation sunlight or rain should be avoided during
environment transportation. The packaged blood pressure
- 28 -
monitor should be stored in an environment with
temperature ranging from - 20 ℃ to 55 ℃ ,
relative humidity ranging from 10% to 93%, and
atmospheric pressure ranging from 70 kPa to 106
kPa, with no corrosive gas and positive
ventilation.
19、Statement of Electromagnetic Compatibility
Guidance and manufacturer´s declaration – Electromagnetic emission
The Model JPD-HA200 Electronic Blood Pressure Monitor is intended for use in the
electromagnetic environment specified below.The customer or the user of the Model
JPD-HA200 Electronic Blood Pressure Monitor should assure that it is used in such an
environment.
Emissions Compliance Electromagnetic environment - guidance

The Model JPD-HA200 Electronic Blood Pressure


Monitor uses RF energy only for its internal
RF emissions function. Therefore, its RF emissions are very low
Group 1
CISPR 11 and are not likely to cause any interference in
nearby electronic equipment.
The Model JPD-HA200 Electronic Blood Pressure
Monitor is suitable for use in all
RF emissions establishments,including domestic establishments
Class B and those directly connected to the public
CISPR 11
low-voltage power supply network that supplies
buildings used for domestic purposes.

Harmonic
emissions N.A.
IEC61000-3-2
Voltage
fluctuations/fli
cker N.A.
emissions
IEC61000-3-3

Guidance and manufacturer´s declaration – Electromagnetic immunity

- 29 -
The Model JPD-HA200 Electronic Blood Pressure Monitor is intended for use in the
electromagnetic environment specified below.The customer or the user of The Model
JPD-HA200 Electronic Blood Pressure Monitor should assure that it is used in such an
environment.
Immunity IEC 60601 Compliance Electromagnetic environment-
test test level level guidance
Electrostatic ±8kV contact ±8 kV contact Floors should be wood, concrete
discharge ±2 kV, ±2 kV, or ceramic tile. If floors are
(ESD) ±4kV, ±4 kV, covered with synthetic material,
IEC the relative humidity should be at
±8 kV, ±8 kV,
61000-4-2 least 30 %.
±15KV ±15 KV
air air
Power Power frequency magnetic
frequency fields should be at levels
(50/60 Hz) characteristic of a typical
30A/m, 30A/m,
magnetic location in a typical
50/60Hz 50/60Hz
field commercial or hospital
IEC environment.
61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level

Guidance and manufacturer´s declaration – Electromagnetic immunity


The Model JPD-HA200 Electronic Blood Pressure Monitor is intended for use in the
electromagnetic environment specified below. The customer or the user ofThe Model
JPD-HA200 Electronic Blood Pressure Monitor should assure that it is used in such an
environment.
Immunity test IEC 60601 Compliance Electromagnetic environment
test level level -guidance
Portable and mobile RF
communications equipment
should be used no closer to any
3 Vrms150 part of the JPD-HA200
kHz to 80 Electronic Blood Pressure
MHz Monitor ,including cables, than
6 Vrms the recommended separation
Radiated RF
150 kHz to N/A distance calculated from the
IEC 61000-4-3
80 MHZ equation applicable to the
outside frequency of the transmitter.
ISM Recommended separation
bandsa distance
 3.5 
d   P
 V1 

- 30 -
 3 .5 
d  p
 E1  80MHz to
800MHz
7
d   P
 E1 800MHz to
2.7GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
10 V/m transmitter manufacturer
Radiated RF and d is the recommended
80 MHz to 10 V/m
IEC 61000-4-3 separation distance in metres(m).
2.7 GHz
Field strengths from fixed RF
transmitters,as determined by an
electromagnetic site survey, a
should be less than the
compliance level in each
frequency range b Interference
may occur in the vicinity
of equipment marked with the
following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by
absorption and reflection from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are
6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283MHz; and
40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8
MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7MHz to 7,3 MHz, 10,1
MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17MHZ, 21,0 MHz to 21,4
MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
b The compliance levels in the ISM frequency bands between 150 kHz and 80
MHz and in the frequency range 80 MHz to 2,7 GHz are intended to decrease
the likelihood that mobile/portable communications equipment could cause
interference if it is inadvertently brought into patient areas. For this reason, an
additional factor of 10/3 has been incorporated into the formulae used in
calculating the recommended separation distance for transmitters in these
frequency ranges.
c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy.To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Model JPD-HA200
- 31 -
Electronic Blood Pressure Monitor is used exceeds the applicable RF compliance level
above, the Model JPD-HA200 Series Electronic Blood Pressure Monitor should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the Model JPD-HA200 Electronic Blood
Pressure Monitor.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distances between portable and mobile RF

communications equipment and Model JPD-HA200 Electronic Blood Pressure

Monitor
The Model JPD-HA200 Series Electronic Blood Pressure Monitor is intended for use in
an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Model JPD-HA200 Series Electronic Blood Pressure Monitor
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
Model JPD-HA200 Series Electronic Blood Pressure Monitor as recommended below,
according to the maximum output power of the communications equipment.
Rated Separation distance according to frequency of transmitter
maximum m
output power 150 kHz to 80 80 MHz to 800 800 MHz to 2.7
of transmitter MHz MHz GHz
W  3.5   3 .5  7
d  p  
d   P d   P
 V1   E1   E1
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

- 32 -
Shenzhen Jumper Medical Equipment Co., Ltd.
Address: D Building, No. 71, Xintian Road,
Fuyong Street, Baoan, Shenzhen ,Guangdong,
China
Tel: +86-755-26692192, 26696279
Fax: +86-755-26852025
Website:www.jumper-medical.com
Postal Code: 518103