AD229e Operation Manual P6
AD229e Operation Manual P6
AD229e Operation Manual P6
Instruct the patient that he will hear tones with different frequencies
and that he is supposed to push the patient response button
whenever a tone is audible to him.
To view the thresholds stored, hold down "shift" (10) while activating
“disp thr” (5). The thresholds for the various frequencies will now be
displayed on the LCD. To get back to normal mode, hold down “shift”
(10) and “disp thr” (5) again.
In the setup item no. 3 you can choose to have some of the
frequencies deactivated in the Auto Threshold mode.
Please note:
In setup item no. 7 the Hughson Westlake threshold test can
be set up to run with 2 correct responses out of 3 or 3 correct
responses out of 5.
Instruct the patient that he will hear tones with different frequencies
and that he is supposed to push the patient response button
whenever a tone is audible to him.
or
To view the thresholds stored hold down “shift” (10) while activating
“disp thr” (5). Then activate the “Auto Threshold” button (7). The
thresholds for the various frequencies will now be displayed on the
LCD.
In the setup item no. 4 you can choose to have some of the
frequencies deactivated in the Auto Threshold mode.
Please Note:
In the setup item no. 6 the OSHA test can be changed from
“OSHA” to “Békésy” or “HW”. This gives the operator the
option to choose which threshold test that he wants to power
up when Auto Threshold is activated.