Humacount 30 80TS
Humacount 30 80TS
Humacount 30 80TS
HumaCount 80TS
| User Manual
System Version
Copyright
Copyright 2020, Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden,
Germany. All rights reserved.
No part of this documentation may be reproduced in any form, nor processed, copied or
distributed by means of electronic systems, without prior permission of HUMAN in
writing. Since all precautionary measures were taken into account in producing these operating
instructions, the manufacturer accepts no responsibility for any errors or omissions. This
includes any liability for damage that could arise from possible incorrect operation based on this
information. Subject to changes without notice as result of technical development.
Table of Contents
1 Safety Instructions 7
1.1 Introduction 7
1.2 User Warranty 7
1.3 Intended Use of the Instrument 7
1.4 General Safety Warnings 8
1.5 Biohazard Warning 9
1.6 Biohazard Warning 9
1.7 Instrument Disinfection 9
1.8 Who should use this manual 10
1.9 Special Symbols used in this Manual and on the Analyzer 10
1.10 General Precautions 10
2 System Description 13
2.1 The Instrument 13
2.1.1 Patient Testing 13
2.1.2 Reagents 14
2.1.3 Technical Operation 17
2.1.4 Calibration 17
2.2 Unpacking and Installation 17
2.2.1 Connect reagents 19
2.2.2 Environmental factors 20
2.2.3 Turning the Instrument ON 24
2.2.4 Turning the Instrument OFF 25
2.2.5 Preparing for shipment 26
2.2.6 Emergency shut-down 27
2.2.7 Standby 28
2.3 Built-in Printer 28
2.4 Accessories 30
2.4.1 Scope of supply of HumaCount 30 TS
30
2.4.2 Scope of supply of HumaCount 80TS 30
2.4.3 Optional Accessories 31
2.5 Parts of the Analyzer HumaCount 30TS/80TS 32
2.5.1 Front panel 32
2.5.2 Main Subunits 33
2.6 Control Material 34
2.7 User Interface (Software) 34
2.7.1 User Interface 34
2.7.2 Menu System 36
2.7.3 Menu Tree 37
2.8 Method of Measurements 40
2.8.1 Photometric Light Absorbance Method 40
2.8.2 Volumetric Impedance Method 40
2.8.3 3-Part Differential Analysis Method 42
2.8.4 Parameters 43
2.8.5 Absolute and Linearity Ranges of Parameters 45
2.9 Quality control 45
2.9.1 QC database 46
2.9.2 Entering New Lot of QC material 47
2.9.3 RUNNING BLANK MEASUREMENT 47
2.9.4 Levy-Jennings QC chart 48
2.9.5 Interpretation of Results 49
4 ADVANCED OPERATION 71
4.1 Diagnostic menu 71
4.1.1 Device Information 71
4.1.2 Device statistics 72
4.1.3 Self-test 72
Contents
5 Troubleshooting 127
5.1 Error messages 127
5.2 Pneumatic/mechanical error messages 128
5.3 Known Limitations 128
5.4 Specifications 128
5.4.1 Specifications HumaCount 30 TS
128
5.4.2 Specifications HumaCount 80TS 128
5.4.3 HumaCount 30TS and HumaCount 80TS 129
5.5 The Measurement Process 132
5.6 Fluidic Schematics 133
5.6.1 Fluidic Schematics of HumaCount 30TS 133
5.6.2 Fluidic Schematics of HumaCount 80TS 133
5.7 Sample Identification 134
6 APPENDIX 135
6.1 Reagent Solution and Consumption 135
6.1.1 Average reagent consumption HumaCount 30 TS
135
6.1.2 Average reagent consumption HumaCount 80TS 136
6.2 APPENDIX “A”: FINDING THE CORRECT [LYSE] VOLUME FOR
VETERINARY APPLICATIONS 136
6.3 Human Software update and settings 139
Contents
Safety Instructions 7
1 Safety Instructions
1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read
the ”GENERAL SAFETY WARNINGS”, describing the appropriate operating pro-
cedures. Please contact your HUMAN authorised local Technical Service in the
event of instrument failure or other difficulties with the instrument.
This manual is for both types of instruments, the HumaCount 30TS and Huma-
Count 80TS. All screens which are equal on both instruments are not shown for
both types of instruments, but only for HumaCount 30TS.
! Note: This manual is for both
types of instruments, Huma-
Count 30TS and HumaCount 80TS.
If a chapter/table or screen is only for one analyzer type, it is marked in the title
and explanations. If it is applicable for both instruments it will be named Huma-
Count 30TS/ HumaCount 80TS.
Figure 1
Biological Hazard Symbol
The sampling needle and other components inside the analyzer may
Table 2
cause injury, or can get damaged if handled incorrectly. Only certified
personnel should open the covers. Running samples with opened cov-
er is not recommended due to the risk of possible injury. Always wear
safety gloves while performing maintenance actions.
The analyzer should be moved holding the analyzer in an upright po-
sition.
Make sure to retain the original packaging material for safe transpor-
tation and storage in the future.
To prepare the analyzer for shipping, storage or extended periods of
inactivity, please drain the reagents (run Preparing for shipment pro-
cedure) and repackage it in its original packaging.
Do not expose the HumaCount 30TS/HumaCount 80TS to direct sun-
light, extreme temperature or humidity over 80%.
2 System Description
The HumaCount 30TS/HumaCount 80TS are fully automated 3-part differential
cell counters designed for in vitro diagnostic use, developed for small clinics and
point-of-care lab offices.
RDW and PDW parameters have two forms of representation: CV and SD. Both
parameters describe the distribution width, but from different aspects. User can
select the units to use for displaying RDW and PDW parameters.
2.1.2 Reagents
Use only reagents supplied by HUMAN with the analyzer, otherwise accuracy
cannot be guaranteed.
2.1.2.1 Contents
HC-DILUENT Isotonic saline solution, used to dilute whole blood sam-
(Cat.No. 17400/11) ples and to rinse the fluidic system between measuring
procedures.
HC-LYSE Creates hemolysate for 3‑part WBC differential and for
(Cat.No. 17400/22) total WBC and HGB.
HC-CLEANER For cleaning process of the fluidics
(Cat.No. 17400/31)
HC-Control Stabilized blood cell control
(Cat.No. 17400/40)
HC-Calibrator Stabilized blood cell calibrator
(Cat.No. 17400/50)
HC-Regular Enzymatic cleaner
(Cat. No. 17400/55)
HC-Acute Hypochlorite based cleaner for disinfection of the instrument
(Cat. No. 17400/56)
Storage:
HC-Diluent (Cat.No. 17400/11), HC-Lyse (Cat.No. 17400/22), HC-Cleaner (Cat.
No. 17400/31) may be stored at 2-42°C.
HC-Control (Cat.No. 17400/40) and HC-Calibrator (Cat.No. 17400/50) may be
stored at 2 – 8°C.
HC-Regular (Cat. No. 17400/55) and HC-Acute HC-Acute (Cat. No. 17400/56)
may be stored at 2-25°C.
2.1.2.2 Stability
The Open Vial Stability for following reagents is 3 months after open-
ing: 17400/11, 17400/22, 17400/31
The Open Vial Stability for following reagents is 18 months after
opening: 17400/55, 17400/56
Hazard statements
The following hazard statements are valid for the respective REF numbers.
[REF] 17400 HC-Lyse CF
H412 Harmful to aquatic life with long lasting effects.
[REF] 17400/56 HC-Acute
H315 Causes skin irritation.
H318 Causes serious eye damage.
H411 Toxic to aquatic life with long lasting effects.
Precautionary statements
The following precautionary statements apply to the following REF numbers.
[REF] 17400/11 HC-Diluent
[REF] 17400/22 HC-Lyse CF
[REF] 17400/31 HC-Cleaner
[REF] 17400/55 HC-Regular
[REF] 17400/56 HC-Acute
The following information about dangerous components are valid for the fol-
lowing REF numbers.
[REF] 17400/56 HC-Acute
CAS: 7681-52-9 sodium hypochlorite, solution ≥2.5-<5%
EINECS: 231-668-3
CAS: 1310-73-2 sodium hydroxide ≥0.5-<2%
EINECS: 215-185-5
2.1.4 Calibration
The HumaCount 30TS/HumaCount 80TS arrive to your laboratory factory-calibrat-
ed and ready to use. However, calibration needs updating whenever you find
that the results have slightly changed, or a different or new control material is
used. With each control material you receive for the instrument, you will find
a control sheet listing the parameters the instrument should match. Perform
these calibrations as explained in a later chapter (see chapter 4.9).
Figure 2
Figure 3
Recommended placement of
Reagents and Waste
-- Connect the reagent tubes to the analyzer, use the supplied connecting
tubes and special bottle caps. Do not block the small air vent hole on the
container caps to provide air flow.
Figure 4
Vent hole
-- Be sure that the colour on each tube, cap and connector in the back of the
instrument match;
-- Before putting the drop-down tubes into the reagent containers, wipe the
dust off from them using a lint-free wet paper wipe
-- Use separate Waste lines for the waste tank connections as indicated below:
20
Figure 5
Establishing correct
waste connection
The input socket is a standard power cable connection and the output is a DC
jack.
Figure 6
Solar panels generate DC power, being stored in Batteries (12 volt). Power sup-
plies use 220 volt and convert this AC-voltage to 12 volt DC for the instrument.
Conventional systems with inverter lose energy in 2 ways - converting 12 volt
to 220 volt and then converting 220 volt to 12 volt.
The HumaPower system uses just 12 volt directly applying it to the HumaCount
30TS/HumaCount 80TS.
HumaPower can be charged by solar power, a generator or by ordinary mains.
HumaPower has a connector, being attached directly to the analyzer replacing
Figure 7
Space Requirements
2.2.2.3 Peripherals
Connect external peripherals only when both the instrument and the peripheral
device are off. Possible peripherals are:
2D Barcode reader
-- Optional 2D Barcode reader ([REF] 16430/11) allows you to scan-in all assay
values at the control.
External printer
-- The printer must be recommended by authorized technician
-- The printer must be approved and listed
-- The printer must have a CE mark
A so called printer box ([REF] 16430/13) allows you to connect almost every
printer to the instrument. ! Note: Reagents may cause
corrosion and skin irritation. If
any of the liquids leaked onto the
External keyboard cover of analyzer or the furniture,
-- The external keyboard must be approved wipe it off immediately. In case of
-- The external keyboard must have a USB port or suitable adapter skin contact, rinse the liquid off
with plenty of water.
Link to host computer via USB port
-- Serial link cable must be approved by technician
-- USB B port (linking to host computer) requires a USB A-B cable and USB driv-
er software (contact service for availability)
Therefore always follow the instructions below when switching the instrument
off. To shut down HumaCount 30TS/HumaCount 80TS, press EXIT key in the Main
menu. The following screen appears.
Press Shutdown.
Figure 12
Figure 13
Figure 14
Figure 15
2.2.7 Standby
The analyzer automatically changes into
Figure 16 the standby routine and perfoms a clean
cycle.
-- Open the paper lid (pull the lid upwards by the Figure 18
handle)
-- Gently drop the new roll into the holder of the Figure 19
printer, hold “starting edge” with your hand,
and make sure it comes out on the front of the
printer as indicated in the picture.
2.4 Accessories
In case of damage or missing item, please contact the supplier immediately.
Figure 21
Reagent Tubing Kit
Figure 22
Cleaning Tubing Kit
Figure 23
Human SolarCharge
(110 watt Solar panel)
32
Figure 24
HumaPower (rechargable
12 volt battery system)
Figure 25
Front View
4
2
Figure 26
Rear View
1
1 Power switch
2 USB A connector
3 USB B connector
4 Ethernet connector
2 5 Reagent key (HK) connector
6 Power source connector
7 External Grounding
Connector
8 Reagent connectorsl
3
9 Instrument label
4
(Serial number)
9 5
6
8
The following message will appear if the control material is expired, you have
to apply a new LOT. Press Start to continue anyway or press Cancel to stop the
process.
Figure 27
The touch screen allows selection of menu items and ‘pressing’ of buttons by
gently tapping the surface over the selected item. The touch-screen device
used in the HumaCount 30TS/HumaCount 80TS analyzer is selected for reliability,
endurance and compatibility with the medical laboratory environment. Simple
operations are supported, but multi-touch ‘gestures’ or multiple tapping are not
supported. The surface glass material is resistant to liquids, however it should
be dry when in use, otherwise it may not work reliably. Use a damp cloth to
clean the glass surface when the analyzer is turned off.
2 3
1 4
Figure 28
Graphical User
Interface Sections
1 Icon of menu
2 Title of the menu or function
5 3 Warning message indicator
4 Current time
5 Selectable menu buttons
6 Context sensitive function
7
buttons
7 Background image
Figure 29
HumaCount 30 /HumaCount
TS
Figure 33
2.7.3.4 Maintenance menu
-- Cleaning
-- Calibration
-- Diagnostics
-- Reagent Status
-- Back to Main menu
Figure 36
Photometric Light
Absorbance Method
Figure 37
Volumetric Impedance Method
A constant direct current flows between the electrodes on both sides of the
aperture. Each cell passing through the aperture causes a change in the
electrical impedance of the conductive blood cell suspension (diluted blood).
This impedance change is detected by the HumaCount 30TS/HumaCount 80TS
electronics and converted to an electrical voltage pulse. The number of pulses
is proportional to the number of particles in the diluted sample.
Figure 38 Figure 39
RBC histogram WBC histogram
42
Data of particles are used to present a size distribution histogram of the WBC
populations. Since size distribution is the only data available for the WBC histo-
gram, there is no enough resolution to accurately differentiate more than three
distinct WBC populations. Based on this fact, all possible normal and abnormal
WBC cell types fall into those three histogram populations.
This being the case, there is also not enough resolution to generate accu-
rately specific morphological flags when abnormal WBC populations are
present in the sample. This is a known limitation of all 3-part differential
WBC cell counter systems that are in the market today. Based on the above
facts, HumaCount 30TS/HumaCount 80TS does not have the ability to generate
specific morphological flags.
2.8.4 Parameters
HumaCount 30TS/ HumaCount 80TS measures and calculates 22 parameters, list-
ed below. For each parameter we list the name, abbreviation and measurement
unit in the first column. Short description for each parameter is in the second
column.
MID
MID, MID% : count and % of
medium size cells
GRA, GRA%: count and % of
granulocytes
Percentages calculated from the absolute
WBC value.
PCT (%)
MPV (fL) = x 10000
PLT (x10³/µL)
! Target values of the control material should be set only once, when you start
using the given lot. Resetting parameters deletes previous QC results of that
level. Any change in the QC material setting deletes previous QC results. It is
strongly recommended to print QC results prior to changes.
46
In QC menu the operator can enter a new lot of controls (by scanning the bar-
code, loading from data file or by entering the assay data from the assay sheet
manually), review data of previous control lots, and print Levy-Jennings graphs.
For instructions on running QC samples, refer to chapter 3.2.7.
2.9.1 QC database
QC database contains the results of the QC measurements in both table and
Levy-Jennings formats.
To access a QC Database, select the Quality Control in Main Menu. All QC files
will be displayed (as shown on the left).
Figure 43 Use and to browse up/down in the
list.
Select the lot number by tapping the cir-
cle in front of the lot number, and tap
to display Levy-Jennings chart.
Press New lot if you wish to enter a new
control lot.
Figure 44
After selecting QC lots, Delete will delete all data related to the selected QC lots,
Print will print, Export will save data onto a USB pen drive.
Press Back to return to Main menu.
Figure 46
Figure 47
Press Measure to start a new QC run of the selected lot, indicated in the header.
Parameters measured:
Parameter names, values, units, and
evaluation if those are lower, inside
or higher than the normal range.
Probe voltage:
This section shows electrode voltages
for WBC and RBC counting, min. and
max. values, for service purpose only.
WBC Lyse volume: 0.9 ml.
It shows amount of lyse used for the
measurement. Lyse setting can be
modi fied by the user before meas-
urement run, but only for extrem
samples. Be aware that target values
of control will not match if changed.
See section 6.2 for details on lyse
volume setting.
Warning flags are displayed for the
measurement. See section 3.3.2
below for details on warning flags.
If a value of the parameters is out of the normal range, the displays of Figure 50
and Figure 51 will appear.
Please, refer to chapter 3.3 for result interpretation, and understanding further
flagging.
Figure 52
52
Cancel
Accept q-flag on parameters of control and sample reporting
Redo QC
In case you believe QC measurement was not as it is supposed to be done.
Validation
You will be able to suppress parameters which are out of the target value range
of the control.
Calibration
Starts the calibration anyway (note, only “primary” parameters can be calibrat-
ed directly).
For further Information about parameter flags, see section 3.3.2 below.
For further information about suppressing parameters see chapter 4.7.12
Figure 53
Care must be taken when using homemade containers pre-dosed with EDTA.
If the container is not filled with enough blood, the ratio of EDTA to blood may
reach a level, which results in osmotic transfer from the RBCs which shrinks
! Note: If you hurt yourself
during analysis, biohazard
substances can cause infection!
them. The ratio of EDTA to blood should not exceed 3 mg/ml. Generally, we Always take special care to sharp
suggest using pre-manufactured sample tubes containing the necessary objects and always use rubber
amount of EDTA. However, make sure not to get short draws. Also, when taking gloves!
blood, please make sure that requirements attached to sample tubes are met.
To initiate analysis:
Invert the closed sample tube at least 8 times to achieve a homogenous
sample. Do not shake the sample, because micro-bubbles can form inside which
may cause erroneous sampling!
54
Figure 54
Tubes used in Vacutainer adapter
Vacutainer with sample blood Sample tube with 5 ml control blood
(cap removed)
Below you can see 3 types of microtainer tubes used in micro adapter. These
are only examples given by us, you can try to use other type of microtainers as
well.
Figure 55
Tubes used in micro adapter
Figure 56
Tubes used in control adapter
The sample rotor will turn the vial into the inside of the instrument and needle
draws sample from the tube. The aspirating needle is retracted, while its outer
surface is automatically rinsed with diluent by needle wash head. This insures
! Note: Remove the cap!! It is very
important because the tip will
not pierce the cap!
the low carry-over between samples. After a few seconds, the rotor turns out.
Now you can remove the sample tube from the adapter.
56
Ensure that:
Figure 57
-- All reagents and the waste are properly
connected
-- All environmental conditions are met
-- Blank measurement has been run and
accepted (see section 3.2.3 below).
-- Quality Control material have been tested
and produced acceptable results, so that
calibration of the analyzer is valid (see chap-
ter 2.9).
-- Specimen is ready to use, mixed well and it is
in the recommended time for measurement
(see chapter 3.1).
Figure 59
Figure 62 Figure 63
Values within the normal range are displayed in white, and outside the normal
range are displayed in red, and flagged:
L – if lower than the normal range, or
H – if higher than the normal range specified by the measurement type.
Q - if control value of primary parameter could not be recalled.
q - if control value of secondary parameter could not be recalled.
! Tip: Flagged results should be D WBC/RBC The RBC-WBC discriminator is over 45 fl on the his-
repeated. threshold togram.
Interference Rerun sample with higher lyse volume setting.
! Tip: Flagged results should be d RBC/PLT threshold The analytical software could not separate RBC’s
repeated. Interference from PLT’s because of a population overlap. Possi-
bly clumped PLTs.
Possible Disease Flag (low) Normal Flag (high) Possible Disease Table 13
Leukopenia WBC- WBC WBC+ Leukocytosis Diagnostic flags and reference
Neutropenia GRA- GRA GRA+ Neutrophilia range of parameters
Lymphopenia LYM- LYM LYM+ Lymphocytosis
MON MON+ Monocytosis
RBC RBC+ Erythrocytosis
Anemia HGB- HGB
Microcytosis MCV- MCV MCV+ Macrocytosis
RDW RDW+ Anisocytosis
Hypochromia MCHC- MCHC
Thrombocytopenia PLT- PLT PLT+ Thrombocytosis
PDW PDW+ Macro PLT
The diagnostic flags are only a hint supporting the interpretation of results, but
need always to be verified by a clinician.
In order to modify the normal ranges refer to chapter 4.7.5.
68
Default sampling depth is: 0 mm, indicated by ‘–––‘ in the selection list, and it
means approximately 3 mm clearance between the bottom of sample tube and
the tip of the needle.
Example to set up sampling depth correctly: if the bottom of the tube is lifted by
12 mm, select +15 mm setting to make sure the needle will not intend to pierce
the bottom of the sample tube.
70
4 ADVANCED OPERATION
Figure 75
72
4.1.3 Self-test
Run self-test to verify operation of essential components of the analyzer.
Self-test should be performed:
Figure 77 -- After installation.
-- After replacing any component.
-- After extended time not in use.
Figure 78
4.2 Database
Patient results are stored in the analyzer’s memory in chronological order, and
can be viewed in the database at any time. Memory capacity usually is 10.000
patient records, including all parameters, histograms, flags, sample ID and date/
! Note: If storage space is full,
the latest (actual) record will
overwrite the oldest record in the
time of measurement. It could be upgraded to 40.000 by a service engineer. Database.
Details button allows the user to view results with histograms, the same results
screen like after measurement.
Total number of records, and the number of selected records are displayed
below the list.
If more than one records are selected, the Statistics button will be activated, and
the SW will be capable of calculating statistics of the parameters (mean, SD, CV,
Levy-Jennings chart).
Print will start printing selected records on the actually selected printer.
Filter button allows searching/selecting records by specific criteria.
Manage button allows the user Delete, Send, or Backup selected records.
Back button will return to Main menu.
A numerical input pop-up screen will appear after pressing e.g. Date to field.
Figure 86 Type the desired date and press Enter. ‘/’
character will appear automatically. Use
Delete to delete the last character.
Figure 87
Figure 88
78
Figure 92
After selecting the communication port, a
progress bar will appear, showing the link
as the selected data are being sent.
After opening the record database from pen drive, the following menu appears.
The header will show the file name.
82
Figure 96
Individual records can be viewed by De-
tails button.
4.4.6 Export
Figure 98
Figure 101
Print jobs screen (indicated on the left) will
show the actual printer status and list of
printer jobs pending.
Figure 103
Printout on external printer
Figure 102
Thermal paper printout
Figure 104
Database Table Printout
Automatic printing: If enabled, the analyzer will print all measurement results
automatically after each measurement.
Figure 109
Press Accept to save settings and exit, or press Cancel to exit without saving
changes.
90
Figure 114
Figure 115
Table 16
Dog Cat Horse Cattle
low high low high low high low high
Pre-defined set-
WBC 109/l 6.0 17.0 5.5 19.5 5.4 14.3 4 12
tings for normal
RBC 1012/l 5.5 8.5 5 10 6.8 12.9 5 10
ranges in veterinary HGB g/l 120 180 80 150 110 190 80 150
mode HCT % 37 55 24 45 32 53 24 46
MCV fl 60 77 39 55 37 59 40 60
MCH pg 19.5 24.5 12.5 17.9 12.3 19.7 11 17
MCHC g/l 310 340 300 360 310 390 300 360
PLT 109/l 200 500 300 800 100 400 100 800
MPV fl 3.9 11.4 12 17 - - - -
LYM 109/l 1.0 4.8 1.5 7.0 1.5 7.7 2.5 7.5
MON 109/l 0.2 1.5 0.0 1.5 0.0 1.5 0.0 0.84
GRA 109/l 3 12 2.5 14.0 2.3 9.5 0.6 6.7
LYM % 12 30 20 55 17 68 45 75
MON % 2.0 4.0 1.0 3.0 0.0 14 2 7
GRA % 62 87 35 80 22 80 15 65
Figure 118
Figure 122
Tap Communication in Settings menu to
enter Communication settings menu.
USB-Serial Baud Rate: To connect PC via USB B emulated serial link, set up baud
rate here identically to the setting of the connected PC serial link baud rate.
Automatic send: This setting can be one or combination of links enabled. Af-
ter a measurement is completed, data record will be sent automatically to the
ports selected here.
Press Accept to save settings and exit or Cancel to exit without saving changes.
100
Figure 125
By tapping automatic send button, the ac-
tive/enabled ports can be configured for au-
tomatic data transmission after every meas-
urement. Select one of the options on the
screen for automatic sending of data records
after measurement:
4.7.12.2 Logging in
The operator has to log in using its pass-
word in order to use the analyzer. Users Figure 127
Figure 128
Log in screen will appear.
Tap the box over Login name to select from
available users, enter the password, then
press Enter.
Figure 129
Figure 132
Log in as Admin and press Automatic
login set. The following menu appears.
The following configuration steps are necessary, before the 2D-barcode reader
will work on the instrument.
Installation:
Connect the 2D Barcode reader via the USB cable to an USB port of the analyzer
(or alternatively a PC).
The configuration of the 2D Barcode reader can be done by scanning of the fol-
lowing barcodes in the order as shown below.
Step 1:
Step 2:
Step 3:
Step 4:
Step 5:
Step 6:
Step 7:
106
Step 8:
Step 9:
Step 10:
Step 11:
Step 12:
Step 13:
The following screen will appear if you used a barcode that is not for the Huma-
count 30TS or HumaCount 80TS instrument. Click OK to continue.
Figure 133
Now the 2D Barcode reader is ready for use on the analyzer. Before up-load the
target values of the controls, make sure you have the correct assay value sheet
and Data-Matrix-Code for the instrument. Data-Matrix-Codes from another in-
strument types cannot be uploaded.
4.8.1 Cleaning
Cleaning: Starts a wash cycle using the
Figure 135 cleaning reagent connected to cleaner in-
put. A cleaning cycle is automatically per-
formed during a start-up and a shutdown
procedure.
Figure 136
Hard Clean Regular
Figure 137
Hard Clean Acute
Figure 135 will show up if you have pressed the Acute Hard-Clean by hypochlo-
rite Cleaner button. You have to place a tube with HC-Hardcleanacute cleaning
solution into the sampler rotor. Press OK to start the cleaning process or press
Cancel to cancel the process.
Needle wash head cleaning: For instructions on how to clean the needle wash
head see chapter 4.8.3.
Figure 138
Figure 139
Needle wash head cleaning: If you press the Needle wash head cleaning but-
ton, the following screen will appear. Follow the instructions and finalize it by
clicking OK.
Figure 140
Any salt build-up on the lower surface may cause malfunction during operation.
Use a soft cloth or wiper dampened with water to clean this area. You can see
the washing head indicated in the following figure:
Figure 141
Parts of measuring block
1 Washing head
1
1. Exit Measure menu. Open the side door after the needle has stopped
moving.
2. Gently rub the lower surface of the washing head with a damp cloth or wiper
to remove the salt build-up.
3. Close the side door.
4.9 Calibration
The analyzer stability can be monitored with HC-CONTROL control blood
([REF] 17400/40). Performing QC determinations regularly verifies contin-
ued optimal performance. The analyzer can be calibrated by HC-calibrator
([REF]17400/50).
Figure 142
Press Factors to check or modify Calibra-
tion factors manually.
If you have a 2D bar code reader connected, simply scan the 2D (Q-R) barcode
on the assay sheet of the control to upload lot number, expiration date, and
target values for the calibration material.
In the absence of a bar-code reader, enter these values manually, one after an-
other by tapping on the value boxes.
Figure 145
After calibrator measurement is com-
plete, the results will be displayed. If
the results look good, and there are no
warning flags displayed, Run calibrator
again until the preset number of runs
(3 or 7) is reached.
Prediluted calibration factors and history menus. See chapters 4.9.2 and 4.9.4
for details on use.
Figure 155
When you push Accept in case of HC-Lyse CF reagent, the instrument will look
for a reagent key (HK). This key must be inserted into “HW Key for Reagent” slot
at the back of the analyzer.
This screen appears to remind you that the reagents must be replaced soon.
Click OK to continue.
Below the stepwise instruction for the replacement of the first and second bot-
tle. It must be done in two different steps, to get the correct open bottle stabil-
ity for each of the two lyse bottles.
If this warning is shows up, replace the first lyse bottle by following the steps
below.
Figure 156
Figure 157
Clicking on Cancel will not allow any measurement of samples without chang-
ing the HC-Diluent. Clicking on Start will automatically guide the user to the
reagent replacement menu. In case a measurement is started and the HC-Lyse
is expired, the following message will occur:
Figure 158
Clicking on Cancel will not allow any measurement of samples without chang-
ing the HC-Lyse. Clicking on Start will automatically guide the user to the rea-
gent replacement menu. In case a measurement is started and the HC-Cleaner
is expired, the following message will occur:
Figure 159
Clicking on Cancel will not allow any measurement of samples without chang-
ing the HC-Cleaner. Clicking on Start will automatically guide the user to the
reagent replacement menu.
Figure 161
For each lyse bottle you have 1050 measurement credits. 1050 tests is a maxi-
mum, which you will only achieve if you run 400 tests per day on one instrument.
122
Follow the steps below to load the measurement credits from the HK module
to the analyzer:
! Do not connect the reagent Hardware Key into USB ports, because the
analyzer cannot read the credits via USB. Use the ‘HW Key for reagent
lock’ connector on the back only.
!
Every measurement cycle will decrease the number of measurement
credits, except for initialization, standby and wake up cycles.
! It is possible that lyse still remains in the reagent container when the
counter gets to 0. This is caused by the extra amount of lyse in the lyse
container to make sure that lyse will be enough for extra maintenance
and/or service procedures.
Example: if there are 100 remaining measurements on the analyzer, and the op-
erator connects and uploads the content of a new HK (2100 credits), then there
will be 2150 measurement credits on the analyzer and all other credits will be
deleted from HK.
Figure 168
5 Troubleshooting
HumaCount 30TS/HumaCount 80TS provides all the necessary software mainte-
nance functions required to keep the analyzer operating in optimum working
order. Despite excellent maintenance, the analyzer can still encounter problems
where operator intervention is required. HumaCount 30TS/HumaCount 80TS
analyzer software informs the operator of situations when further help is re-
quired. These are called error messages.
5.4 Specifications
5.4.3.2 Accuracy
5.4.3.3 Reproducibility
Table 24
Sampling method Open tube system with automatic sample rotor.
Sample types Human (general), male, female, baby, toddler and
child and veterian samples as defined by the soft-
ware.
Clog prevention High-voltage pulse on aperture in each analysis cy-
cle, chemical cleaning and high pressure back-flush
of the aperture using Cleaner reagent.
Cleaning procedure High-voltage burst of the aperture, high-pressure
back-flush, chemical cleaning of the aperture using
cleaner reagent.
Figure 173
Figure 174
134
6 APPENDIX
The manufacturer cannot guarantee that pre-set volumes will always give the
best results. It is therefore strongly recommended that the user oneself shall
find the necessary volumes for the veterinary application. This requires some
experiments, but the result validated a user doing vet practice will be worth it.
The two most important parameters influencing lysing are lysing time and
[LYSE] quantity. You cannot change the lysing time, as it is fine-tuned for the
[LYSE] supplied by HUMAN GmbH.
Above you could read that [LYSE] quantity can be adjusted either at patient lim-
its, or right before measurement. In the following figures, you can see examples
of over- or under-lysed samples.
Increase the [LYSE] quantity (+0.1 ml) if the separation between lysed RBCs and
WBC populations is poorly differentiated, resulting in increased WBC and LYM
counts.
Figure 175
Figure 176
Figure 177
138
Figure 178
An over lysed sample will have all its cells shifted down towards the smaller cells.
It is because of too much lysing reagent is shrinking all cells to the size of their
nuclei.
With [LYSE] quantity decreased, the cells will not shrink that much, and can be
differentiated easily.
https://www.human.de/sw-hc30ts
respectively
https://www.human.de/sw-hc80ts
Or by scanning the QR Code with a mobile device which supports QR Codes. You
also find this barcode as a label on the device.
Humacount 30TS
HumaCount 80TS
If the information is not accessible via the internet, they can be obtained free of
charge from your local distributor.
In case you do not have the latest software or settings installed, please contact
your local distributor.
140