UM Coag - 4D EN USER
UM Coag - 4D EN USER
UM Coag - 4D EN USER
DIAGON COAG4D
FOUR-CHANNEL SEMI-AUTOMATED
COAGULOMETER
2
1 INTRODUCTION ......................................................................................................................... 5
3 INSTALLATION ......................................................................................................................... 12
4 SETTINGS ................................................................................................................................. 14
7 CALIBRATION........................................................................................................................... 30
3
9 QUALITY CONTROL .................................................................................................................. 38
10 MAINTENANCE ........................................................................................................................ 43
APPENDIX A ...................................................................................................................................... 45
APPENDIX B ...................................................................................................................................... 46
APPENDIX C ...................................................................................................................................... 47
MENU STRUCTURE................................................................................................................................... 47
APPENDIX D ...................................................................................................................................... 48
4
1 Introduction
* The ATIII and D-Dimer tests are available only in Coag 4D with g-Coag4DDi reference
number.
If you have Coag 4D with g-Coag4Dm reference number and you want to measure ATIII and
D-Dimer please contact your dealer.
5
1.3 The structure of the device
Printer cover
opening button
Printer cover
2nd master
USB connector
Touch screen
Incubated reagent
positions Incubation area
6
On/off switch
This switch is used to switch the device on and off.
Touch screen
By touching the appropriate part of the screen the user can
operate the analyzer program. The display is a color graphic
liquid crystal panel.
Incubation area
The incubation area is a monolithic aluminum block regulated to
37 °C. It includes 20 preheated cuvette positions, 4 measure-
ment locations and 2 reagent positions for 10 ml reagent vessels,
which is able to receive 4-ml reagent vessels applying the
adapter.
Reagent positions
There is a rotating magnetic field at the reagent positions,
which, by rotating the magnet rod placed in the reagent vessel,
intermittently mixes the reagents. The mixing can be switched
off.
Measurement locations
Four nephelometric measurement positions are available for the
measurement of coagulation parameters, measurement positions
1 and 4 are also suitable for photometric measurements.
Indicator LED:
There are four indicator LEDs associated with the measurement
positions to monitor their status.
Built-in printer:
This thermal printer prints out the results and other data. The
rubber paper forwarding roller opens up along with the cover to
make paper replacement simple.
Serial port:
9 pin connector for serial communication.
External power supply unit connector
GTM21097-4509 type power supply unit, with input voltage
100-240V AC, and output voltage and current of 9V DC 5A.
7
Warning!
The device may only be operated with its original power supply unit, otherwise the
precision of the measurements cannot be guaranteed! The use of a non-approved
power supply unit may cause radio frequency interference, and may lead to the
malfunctioning of the device!
8
2 Operating principle
9
values can be determined from the measured coagulation time.
The points of the calibration curve can be given by the reagent
manufacturer (at certain manufacturer tests), in the lack of this
they have to be determined by the user from a dilution series.
INR (Calibrated)
A calibration curve may also be used to determine the INR
value, which is a function of the INR and the coagulation time.
The points of the curve are provided by the reagent
manufacturer on the basis of a known INR value calibration
series. The input of the points is the same as the input of the
points of the coagulation time-concentration function points.
10
manufacturer (at certain manufacturer tests), in the lack of this
they have to be determined by the user from a dilution series.
11
3 Installation
Installation may be carried out by trained laboratory personnel
or by the service staff of the distributors authorized by Diagon
Ltd.
Check that the input voltage of the power unit complies with the
voltage of the local mains, and that the mains cable connector
fits into the wall power socket. If there is any discrepancy please
inform Diagon Ltd’s service department or the distributor.
Warning!
The device can only be operated with an undamaged, factory-provided cable, power
unit cover and device cable! Damaged cables may cause electric shocks or faults in
the device!
Damaged insulation can cause electric shock!
12
Warning!
The device must not be placed in direct sunlight!
13
4 Settings
Select the “Setup” function in the main menu, and then select
the test to be set up on the following screen (see figure 4.1).
After selecting the test the test parameter setting screen appears.
- On touching the “Print” button the parameters of the selected
test are printed out.
- The name of the selected test appears in the field with a
black background, the parameters that may be adjusted
appear with an orange background.
- Using the up and down arrows you can select parameters to
be changed.
- On touching the keyboard icon a keyboard appears with
which the parameter values can be adjusted. The buttons of
the keyboard change depending on which parameter is to be
adjusted.
Warning!
Before setting up any changes (first pressing the keyboard icon) user has to enter user
code on the virtual keyboard. Allowance is in effect until next switching off the
instrument. Ask the default user code from the Diagon authorized distributor or
service engineer!
14
- On touching the “X” icon the program exits without saving
any of the changes, but on touching the “ “ icon the
changes are saved when exiting.
15
Algorithm
It is a calculation way of evaluating the coagulation curve,
where:
1. for PT, APTT, Factors and Inhibitors
2. for TT
3. for Fibrinogen
Dimension
Measurement unit, it can be time, percentage, INR Power, INR
Calib. or concentration. Either INR Power or INR Calib., are
further processed as INR.
Parallel
In order to reduce the imprecision results originating from the
manual pipetting the measurement can be performed on two
parallel channels. If the difference of the two results is within a
given percent range the reported result will be the average of the
two. If out of the range the result will be flagged by “D”. The
button (right side on the middle) showing the given percentage
value for the range can be used to set the desired range up to
50%.
16
5 System menu
If you touch the “System” button, the system settings menu is displayed. The possible
settings are as below:
Language English Chinese French
Hungarian Russian Polish
Portuguese Spanish German
Italiano Vietnamese
Measure
17
Identification
There are two possible modes of sample identification:
ID mode: Sample identification on the basis of ID, it can be
typed in manually or with the help of a barcode reader.
SN mode: The device automatically increases the serial number.
It is not advisable to change it because of the safety of sample
identification.
If the automatically generated ID number will be over written
the ID for the next sample will be requested to be entered
manually.
Speed
The incubation positions can be used for incubation and the
cuvettes have to be moved to measurement positions by rows
when the incubation time is up. In normal mode only the four
measurement position are used for measuring process.
See details in chapters 6.4-5.
Error messages
List of flagged measurements sorted by time. When touching the
“i” button program displays explanations of the flags.
18
6 Sample measurement
19
6.2 Measurement window
The user must check that the correct date and time are displayed
on the screen by the analyzer just before starting the
measurement in order to be able to properly retrieve the results
later on at any time.
Sample measurement can be actuated by touching the
“Measure” button in the main menu when the sample selecting
input window comes up (figure 6.1). After selecting the desired
test, in the case of “ID Mode” the display will be changed
immediately to “Measuring” window, while in “SN Mode” only
after entering the starting ID. The areas of this window are the
following.
Sample Identifier
This window shows the sample ID in the current sample
position. If the “Double mode” is on only two sample identifiers
can be stated at the same time, suiting the two channels on the
left and the two channels on the right.
Time counter (sec)
It is the time of incubation or coagulation or measurement.
Status row
It shows either the actual status of measurement process or
required action from user.
20
Result
The results appear here according to the setup of “System“ and
“Setup” menu. The last result is always on the lowest position.
Header of result
Result
Sample Identifier
Status row
Time counter
Header of result
Head of the columns of results
IDs of incubated samples
The ID can be entered. Only in “Quick mode” (see figure 6.5.)
has an additional row. Transferring the sample to measurement
position the ID will be automatically transposed to “Sample
Identifier” field.
Incubation control
It is visible only in “Quick mode” (see figure 6.5). This is used
to start and display incubation time and after transferring the
row of cuvettes to measuring positions then touching this control
field (which shows the current incubation time in second), the
value will be written into time counter.
21
6.3 Sample identification
SN Mode
Measurements are to be started by touching the “Measure”
button and choosing test in test selection screen. If SN mode
was selected previously in “System-Measure-Speed” menu, a
virtual keyboard comes up for entering the first identifier of
measured samples. The further identifiers will be generated
automatically; the program increases the previous identifier by 1
sequentially.
The identifier can contain normal and capital letters, characters
such as _ . / - ( # ) and spaces, which can be accessed by
pressing the “ABC”.
If the automatically generated ID number will be overwritten the
ID for the next sample will be requested to be entered manually.
A similar virtual keyboard will be displayed always when ID
entering field is selected. If barcode reader is used for sample
identification the code should be read here.
22
ID Mode
The sample identifier window is blank. Step into this field and
the virtual keyboard will come up and the ID can be entered. If
you place a sample into measuring position without specifying
the ID the instrument warns with question mark.
Double (parallel) mode
Only two sample identifiers can be stated at the same time,
suiting the two channels on the left and the two channels on the
right. In other respects the measurement process is the same as
above. If this mode is off each measurement position has its own
ID.
23
Status indicator
Time counter Status row
LED
1. Ready for testing Green zero Add Tube * Continuous green
2. Add cuvette(s) to
Green zero Add Sample Blinking green
sample area
3. Add sample
according to the Continuous
Start incubation time Incubating
protocol descri-bed orange
in the appendix
4. Incubation time is
Incubation time in red Add Reag. Blinking orange
over (sound signal)
5. Add reagent Start measuring time Measuring Continuous red
6. Test completed Measuring time in
Tube Out * Blinking red
white
7. Remove cuvette(s) Green zero Add Tube * Continuous green
Warning!
When dispensing the starter reagent the pipette tip should not touch the cuvette.
Hold the pipette over the centre of the cuvette and dispense the reagent with an
energetic pressure.
24
6.5 Quick measurement („System→Measure→Speed→<Quick>”)
In order to make the whole measuring process more effectively
with those assays where the measuring time is less than the
incubation time the incubation area with the 5 rows can be used
for incubating the samples.
Note: this operating mode requires experience and special
attention whereas the incubation area has no sensor for
cuvette detection. The operator must keep the same order of
rows on the screen and in the incubation area. Furthermore it
is important that the four samples which are in the same row
have to be transferred to the measuring position keeping the
same order.
The identification for samples which are in measurement
positions should be performed as described in chapter 6.4 and
the measurement cycle should be accomplished according to the
table below.
Note: until the incubation time for measured samples has not
been started the fields of the additional row are blank.
In the example (see figure 6.5) the additional row on the screen
contains the information of the samples are in the first row of
the incubation area. The windows of this row are blank until the
3rd step (see table below) has not been completed for all samples
are in the measurement positions.
25
Status indicator
Time counter Status row
LED
1. Ready for testing Green zero Add Tube Continuous green
2. Add cuvette(s) to
sample and Green zero Add Sample Blinking green
incubation area
3. Add sample
according to the Continuous
Start incubation time Incubate
protocol described orange
in the appendix
4. Incubation time is
Incubation time in red Add Reag. Blinking orange
over (sound signal)
5. Add reagent Start measuring time Measure Continuous red
6. Test completed Measuring time in
Tube Out Blinking red
white
7. Remove cuvette Green zero Add Tube Continuous green
Warning!
When dispensing the starter reagent the pipette tip should not touch the cuvette.
Hold the pipette over the centre of the cuvette and dispense the reagent with an
energetic pressure.
26
be increased by the instrument for the next 4 subsequent
samples whereas the instrument can store IDs of the row.
Note: the automatically generated ID number can be over
written after touching the “IDs of Incubated Sample” field
using the virtual keyboard.
In case “ID” mode was selected previously in the “System”
menu (“System→Measure→Mode”) by touching the “IDs of
Incubated Sample” field (default content is “?”) the sample ID
can be entered using the virtual keyboard.
Note: until ID has not been entered the measurement cannot
be started and the field contains a “?”.
C. Add cuvettes to the incubation and sample positions (see
figure 6.6) and then add the prepared and adequate quantity of
sample into the cuvettes as prescribed in the measurement
protocol (see appendix).
D. The measuring process for those samples which are in
the measuring positions is the same as it is in normal mode (see
table below).
Note: the first column of the table below contains the
different steps of the testing cycle of the sample and the rest
of the columns shows the state of status indicators after
performing the actual step.
E. After performing Step 3 (see table below) – depending
on the ID mode – the “IDs of Incubated Sample” field contains
“?” or the automatically incremented IDs and the “Incubation
control” field changes from blank to “Start” label.
If the instrument is in “ID mode” the IDs can be entered using
the virtual keyboard after touching the”?” field.
By pressing the field containing ”Start” label it will change to a
timer which count the incubation time.
F. Transfer the cuvettes from the incubation area to the
measuring place. The current incubation time and IDs will be
automatically transferred to the “Sample Identifier” and “Time
Counter” fields while the “Incubation Control” field will be
changed to “Start” and the “IDs of Incubated Sample” field
will show “?” or the next incremented IDs.
27
G. The above described row on the screen can be used for
entering IDs and measuring incubation time of the next row in
the incubation area (see example, marked 2, figure 6.6)
according to paragraph B.
H. If the incubation time of transferred samples are up
continue the measuring process from paragraph 5 in the table
above.
I. The cycle starting with paragraph B can be continued
until samples need to be tested.
ID of incubated samples
Incubation control
1
Note: until the incubation time for measured samples has not
been started the fields of the additional row are blank.
28
5
29
7 Calibration
In the case of the coagulometer calibration means entering
numerical data on the basis of which the program of the device
determines the test result from the measured coagulation time in
the measurement unit prescribed in the settings. For determining
the measured result it is necessary and it is possible to record the
LOT number of the reagents and calibrators used during the
measurement too.
The program of the device makes it possible to enter the
calibration data in the device using a barcode reader rather than
the touch screen, as a result of which the time between changing
the reagents can be reduced as well as the possibility of input
errors. If necessary, the content of the barcode can also be typed
in the program directly. The barcode can contain 16 number or
letter characters in each one of the three lines. Only the data of
the barcode system used by Diagon is interpreted correctly by
the program!
In the Setup menu, the measurement unit defined in the
Dimension line determines the system of the data needed for
calibration.
- In the case of selecting Rate setting the mean coagulation time
of normal plasma, that is the Normal Value, e.g. Mean Normal
Prothrombin Time (MNPT) in the case of PT.
- In the case of INR setting Normal Value and ISI.
- In the case of concentration, percentage or INR CALIB., two
pieces of data of maximum 6 calibration points.
30
valid. These are the name, the LOT number, the year and the
month of expiry date of reagent.
2. Select the Barcode menu point to enter the content of the
barcode with the help of the barcode reader connected to the
USB port, or by typing it in. After pressing the OK button the
barcode data becomes valid and the program fills in the
calibration data.
3. Selecting INR the Normal Value (MNPT) and ISI value
should be entered for determining the INR using the power
method.
31
an already existing data needs to be modified, the red mark can
be moved to the next line by entering and then exiting using the
ESC button. If fewer than six items of data are available, the last
lines must be filled with 0,0.
5. In the case of concentration and INR CALIB. output the same
procedure must be followed.
6. The final results are calculated according to the selection
(done by “Fit” button) of the type of calibration curve which is
calculated by the entered data pairs.
The type of calibration curves are the following: Lin-Lin, Lin-
Inv, Log-Log, Lin-Lin p-p, Lin-Inv p-p Log-Log p-p.
The factory set is Lin-Inv for PT and Log-Log for Fibrinogen.
The calibration curve can be drawn on the screen using the
Graph function.
The calibration curve menu is shown in figure 7.2.
32
7.4 Producing dilution series
The reagent manufacturer may provide the data of the
calibration curves specific to the device. If it is not available, the
dilution series needed for determining the points must be
produced by the user from normal plasma. The coagulation time
of the elements of the line must be measured one by one and
entered in the above program point. The reagent manufacturers
distribute the buffer solutions needed for producing dilutions.
33
8 Retrieval of measured data
In its present construction the device stores the data of 1 000
patients in the format shown in figure 8.1.
34
8.1 Filtering
By using the “Filter” function on the left, the series of displayed
samples can be narrowed down with the help of the menu shown
in figure 8.2. The tests with the same date or the same sample
identifier can be collected from the whole table, and the list can
be restricted to one test. The program only stores the month and
day of the measurement, when making a search it asks for the
Month and the Day separately. If you select the Sample ID
setting, the input screen of the sample identifier is displayed as
shown in figure 6.1.
By touching the Test button repeatedly you can select a built-in
test. The program gives a list of the tests suiting all the
conditions set in three lines of the main filter screen at the same
time.
35
8.2 Detailed display
By selecting the “enlarge function” (magnifier) the detailed
description of the current test (with an orange background) can
be displayed as shown in figure 8.3. On the screen the times
before averaging (e.g. Raw1), the measurement channel number,
the possible error code are displayed, and it can be seen whether
the given test has been printed out or marked. When the window
is displayed, the function buttons at the bottom change, you can
delete tests by pressing the “dustbin” button, or you can print
out a given test by pressing the printer button. If you print from
here, you can start detailed printing as shown on the screen. You
can exit using the “minimize” button.
36
8.3 Action window
“Action” is the icon on the extreme right of the result display
(Figure 8.1). With this icon you can open the action window,
where the measurements can be printed out one by one or in a
group, or sent to an external computer, copied onto a USB
storage device or deleted. These actions can be performed on the
current measurements, marked measurements, all measurements
or on not printed measurements.
37
9 Quality control
Quality of measurements can be ensured only by systematic
quality control procedure. Users have to measure and register
results of the dedicated control material. It is recommended to
use QC material delivered by Diagon or its distributors. If users
apply the instrument software regularly, it stores the results of
QC measurements, the measured data are able to be retrieved,
and processed.
The dedicated way of measuring QC samples is the QC menu,
results measured in QC menu are stored and processed
separately from patient results.
38
Figure 9.2. Control selection screen
39
the OK button. Choosing each test the corresponding target
value definition screen appears (fig.9.4)
When touching the „Name” button repeatedly the abbreviated
names of tests appear one by one. When the desired name is
displayed, the minimum and maximum values of accepted range
of QC result can be entered. Besides the „Dimension” button
names of dimensions appears, which were defined in Settings
menu, in chapter 4. By pressing „Dimension” button the yellow
backgrounded current column can be chosen, and by touching
the „Minimum” and „Maximum” button the upper and lower
limit of acceptance range can be entered using virtual keypad.
40
the same as process of patient sample measurement, but results
are stored in QC result database, which is accessible from
Results function of QC main menu.
41
Diagram
function
42
10 Maintenance
43
10.4 Cleaning the housing
The housing of the analyzer and power supply can be cleaned
after switching off and unplugging the device, using a towel
wetted with water containing a neutral washing agent.
Warning!
Removing the housing of the analyzer and the power supply may result in the failure
of the device or may cause an electric shock! Only skilled personnel may service the
device!
Warning!
In the case of observed overheated block even if the displayed value is within the
acceptable range switch off the analyzer immediately! Failure to do this can cause the
damage of the instrument.
10.6 Scrap
Instrument scrapping must be performed according to the local
legal disposal of waste regulation.
44
Appendix A
Error Messages
D: Difference error: If the difference between parallel measurements is greater
than the permissible (Maximum difference).
C: Curve error: If there is a jump in the optical curve.
L: External light error: If it is too big a change in optical signal and the signal
out of the detection window.
T: Out of range error: If the measurement time is outside the measuring range
(Min time – Max time).
R: Calibration data error: If there is no calibration data in the calibration menu.
I: Incubation is too long: If the measurement does not start in time, because the
incubation time is too long.
O: Incubation was overheated: If the thermostated block is hot.
E: Expired lot: If the used reagent has expired.
R: Barcode type error: If the reagent type has changed (eg. Dia-PT ->Dia-PT
LIQUID), but the control range still belongs to the old reagent type.
Q: Out of QC: If the control result is out of control range.
45
Appendix B
46
Appendix C
Menu structure
Coag4D Manual
Software
Language
Test categories Date ID Test Sec Ratio % INR G/L
0501 111 PT 12,5 1,0 100 1,00 -
0401 101 Fib 6,5 2,5
Date/time
User defined
Screening tests Factors Inhibitors D-dimer ATIII tests
Format
Date
PT II, V, VII, X APC
Time
Sound
Fibrinogen Protein S
Mixer
Printer
TT LA
Measure
Setup
Setup Turbidimetry
Measurement Calibration QC Identification
Clotting Chromogenic - SN
- ID
Speed
Reagent Min. time First time - Normal
ID ... sec Lot Nr Measure - Quick
Expiry date
Max. time Second time
Ratio
Mean Normal
Error
ID ... sec
value Min. time Reagent nr. messages
%
-Calibrator lot Nr Lag time Incubation time
-Curve type
-Values Result
Reagent nr. Dimension
INR
-ISI
-Calibrator lot Nr Incubation time
-Values
G/L Settings
-Calibrator lot Nr Algorithm
-Curve type
-Values -Lot Nr
-Levels Dimension
µgFEU/mL -Dimension
-Calibrator lot Nr -Ranges
-Curve type Parallel
-Values
Barcode
Barcode Date Level % Range Note Mean CV
0501 1 105 85-115 in 115 12,3
0401 1 125 85-115 out
47
Appendix D
“New” waste
Keep dry
This way up
Catalogue number
Serial number
Manufacturer
Non-sterile
Do not reuse
Batch code
48