Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Sound Absorption Characterisation of Woven Materials. Case Study: Auditorium Restoration

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

th

Proceedings of 20 International Congress on Acoustics, ICA 2010


23-27 August 2010, Sydney, Australia

Sound absorption characterisation of woven materials.


Case study: auditorium restoration
Paola Ricciardi (1), Mariagiovanna Lenti (2)

(1) University of Pavia, Via Ferrata 1, 27100 Pavia, Italy


(2) Tecnasfalti Srl, Via dell’Industria, 12, loc. Francolino 20080 Carpiano (Mi), Italy

PACS: 43.55.Ev, 43.55.Br, 43.55.Fw, 43.55.Dx, 43.55.Ka

ABSTRACT
Nowadays building conservation and refurbishment draw the attention of the world we live in. In particular, in the
public sector, the change of occupancy is commonly used in order to maintain the existing functional layout of spaces
and the original structure of the building. Further improvements need to be also considered in order to save the indoor
environmental quality. A case study is provided below by the analysis of acoustical performances of an auditorium in
Italy, the historical S. Giorgio Palace in Genoa. The palace was built in 1260 and it was the most important public
palace in the town; afterwards it became the headquarters of the Port Authority in 1903. Although the high reflective
materials covering the interior surfaces provide high values of reverberation time, the hall is mainly used as a
conference hall. The acoustical project of restoration, approved by the Ministry of Italian Cultural Heritage, allows
only the application of woven materials for floor and curtains, which can be easily removed in case of a change of
destination to respect the historical and architectural value of the hall. Acoustical measurements, by means of the
impedance tube, have been performed up to now in order to define the best woven materials to improve the overall
acoustic performances of the hall. The normal incidence sound absorption coefficient of different samples of carpet
have been tested. A procedure for the samples location in impedance tube measurements has been outlined. Carpet is
a textile material with a good sound absorption, mainly at high frequencies. In order to improve its acoustic properties
at low frequencies a multilayer system composed of carpet and felt having different characteristics have been
experimentally investigated and the optimal configuration has been defined. .

1. INTRODUCTION 2. THE PRELIMINARY ACOUSTIC PROJECT


OF RESTORATION
The Headquarters of the Port Authority of Genoa, are hosted
in S. Giorgio Palace, located in the centre of the medieval The hall (figure 1, 2) is located on the first floor of S. Giorgio
port area of the city. In San Giorgio Palace, there is the Hall palace and has a rectangular plan with a volume of 6000 m3
“Sala delle Compere”. The palace was built in 1260, and it and a floor area of 500 m2.
was the most important public palace of Genoa and
afterwards it became the headquarters of the Port Authority
in 1903. Even though the hall has an average value of
reverberation time of 6 s. [1], it is used as a conference hall.

Under the supervision of the Ministry of Italian Cultural


Heritage, in accordance with [2, 3] an experimental
measurement campaign on the spot was carried out in order
to determine the hall acoustic behaviour and to define the
acoustical properties of the existing components of the
interior envelope. Furthermore, an analysis of suitable
materials and some numerical simulations [1] was performed Figure 1. Interior view of Sala delle Compere
with the aim of enhancing the interior acoustic performance
in the hall with the application of different sound-absorbing
materials. Due to the high level of reversibility offered by
their application, woven materials are being considered as
covering material for floor surfaces and curtains.

Even though some analysis [4, 5, 6, 7] have been undertaken


of these particular types of materials, a deeper knowledge of
sound propagation through textile materials is of prime
importance for evaluating the noise absorption capacities of
woven fabrics which can serve as absorbent materials and
noise control elements in a wide range of applications.
Figure 2. Ceiling of Sala delle Compere

ICA 2010 1
23-27 August 2010, Sydney, Australia Proceedings of 20th International Congress on Acoustics, ICA 2010

2.1. Experimental campaign of Sala delle Compere while there is a great difference in the frequency domain. For
example Reverberation Time T20 reaches a minimum value
Through experimental measurements, carried out in of 1 s at 16000 Hz and a maximum value of 6 s at 63 Hz. In
accordance with ISO 3382 [2], the acoustic behavior of the the frequency domain more interesting for speaking, between
hall has been determined. In order to understand the spatial 125 to 4000 Hz, the medium reverberation time is above 5
distribution of the acoustical parameters, 9 receiver positions seconds, while in environments for listening comprehension,
have been investigated (figure 3). The experimental results of the optimal reverberation time should be around 1 s [7]. Also
T20, D50 and C50 as a function of frequency for all the 9 C50 and D 50 do not have a significant spatial variance,
positions are reported respectively in figure 4, 5 and 6. especially at medium frequencies. But their values are not
suitable for a conference hall: D50 at medium frequencies is
around 10% while optimal values should be above 50%; .C50
varies from -14 dB up to + 4 dB, only at 16000 Hz, while
optimal values should be above 3 dB.

2.2. Numerical simulations

The acoustic project of the hall restoration, approved by the


Ministry of Italian Cultural Heritage, has allowed the
following modifications:
• Acoustic plaster in the lateral walls (extension of 890
m2);
• Curtains to cover the existing windows (extension of 84
m2);
Figure 3. Receivers position • Carpet on the existing marble floor reproducing the same
design in black and white (extension of 484 m2);
7.00

6.00
In order to investigate the efficacy of the identified project
numerical simulations with RAMSETE pyramid tracing
5.00
software have been carried out. The tridimensional model
used for the simulation is presented in figure 7.
T20 [s]

4.00

3.00

2.00

1.00

0.00
63 125 250 500 1000 2000 4000 8000 16000
Frequenza [Hz]

P.1 P.2 P.3 P.4 P.5 P.6 P.7 P.8 P.9

Figure 4. T20 versus frequency in the 9 positions


80.00

70.00

60.00

50.00
D50 [%]

40.00

30.00
Figure 7. Tridimensional model used for numerical
20.00
simulation
10.00

0.00
The absorption coefficient used for numerical simulations are
63 125 250 500 1000
Frequenza [Hz]
2000 4000 8000 16000
reported in table one. Data were collected by literature [7].
P.1 P.2 P.3 P.4 P.5 P.6 P.7 P.8 P.9 The materials with the star (*) are the ones used to simulate
Figure 5. D50 versus frequency in the 9 positions the acoustic behaviour of the hall after restoration.

10.00 Table 1. Absorption coefficient as function of frequency


used in the numerical simulations. (*material used in the
5.00
acoustic project of restoration)
0.00
63 125 250 500 1000 2000 4000 8000 16000
Absorption coefficient as function of
Materials S [m2] frequency [Hz]
C50 [dB]

-5.00
250 500 1000 2000 4000
-10.00 Plaster
(2 m) 205.31 0.05 0.06 0.08 0.04 0.06
-15.00
Marble 50.78 0.05 0.05 0.05 0.05 0.05
Wood 36.24 0.20 0.15 0.10 0.15 0.10
-20.00
Frequenza [Hz] Gypsum 593.11 0.15 0.10 0.05 0.05 0.05
P.1 P.2 P.3 P.4 P.5 P.6 P.7 P.8 P.9 Glass 80.17 0.06 0.04 0.03 0.02 0.02
Figure 6. C50 versus frequency in the 9 positions Carpet* 484.15 0.10 0.30 0.50 0.65 0.70
Curtains* 83.65 0.15 0.35 0.40 0.50 0.50
Acoustic
There is not a significant variation of the acoustical Plaster * 889.81 0.08 0.15 0.29 0.29 0.33
parameters, T20, C50 and D50 among the 9 receiver position,

2 ICA 2010
23-27 August 2010, Sydney, Australia Proceedings of 20th International Congress on Acoustics, ICA 2010

Results of the simulation model of the hall in the existing


state have been compared with experimental ones and a good
accordance was found. In figure 8 and 9, respectively, spatial
values of T20 of numerical simulations before and after
restoration are reported. T20 before restoration varies from
5.72 to 5.82 s, while after restoration T20 is considerely
reduced up to 1 s. Therefore the optimum values of
reverberation time [7] for conference hall are reached with
the outlined project.

Figure 10. Spatial values of D50 obtained by the numerical


simulation “ante operam” of the hall

Figure 8. Spatial values of T20 obtained by the numerical


simulation “ante operam” of the hall

Figure 11. Spatial values of D50 obtained by the numerical


simulation “post operam” of the hall

values of D50 are confirming the effectiveness of the defined


solution, since future values of D50 are improving the hall
definition. From values of 0.14 – 0.30 %, in the actual state,
D50 is incresing up to 82% after restoration.

Clarity values of C50 in the existing and future layout have


been simulated and results are reported in figures 12 and 13.
Also clarity index is ammeliorating after restoration: from
minimum values of -8 dB and maximum values of -2 dB in
“ante operam” state, C50 achieves values of + 6.5 dB in the
“post operam” configuration.
Figure 9. Spatial values of T20 obtained by the numerical
simulation “post operam” of the hall In conclusion, simulated values of T20, D50 and C50 are all
demonstrating the acoustic efficiency of the outlined project
Values of D50 before and after the project of restoration are
of restoration.
shown in figures 10 and 11. Also post operam simulated

ICA 2010 3
23-27 August 2010, Sydney, Australia Proceedings of 20th International Congress on Acoustics, ICA 2010

3.1. Carpet

In order to comply with requirements set by the Ministry of


Italian Cultural Heritage were selected carpet manufacturers
able to reproduce the design of the floor in black and white
of the existing marble floor (figure 1). Two manufacturers
were found and 6 different carpet samples were tested,
diverse for composition, thickness and mass per unit area. In
table 2 the characteristics of each carpet sample are
described.

Table 2. Description of the analysed samples of carpet


Mass per
Total unit area
Test
Composition Thickness [g/m2] of Colour
Sample
[mm] Fibre +
Support
Brown
A 100% Wool 12 3934 (Spotted)
Cow
Dark
B 100% Wool 8 2925 Blue
Plain
80% Wool Brown &
C 14 2752
20% Polyamide Green
80% Wool Purple
D 12 3809
Figure 12. Spatial values of C50 obtained by the numerical 20% Polyamide Veined
simulation “ante operam” of the hall 80% Wool Ochre
E 10 3296
20% Polyamide Plain
Electric
100% Blue with
F 12 2548
Polyamide Yellow
Pictures

As an example, in figure 14 the sample B is represented, in


diameters 100 mm and 30 mm.

Figure 14. Sample B

3.2 Felt

Two different types of felt, have been tested. The


characteristics of each felt are reported in table 3.

Table 3. Description of the analysed samples of felt


Test Composition Total Mass per Colour
Sample Thickness unit area
[mm] [g/m2]
of Fibre +
Figure 13. Spatial values of C50 obtained by the numerical Support
simulation “post operam” of the hall
100%
NW3 Polyester 3 800
Raw Marlin
Needle-
3. ANALYSED SAMPLES OF WOVEN punched
NW4 4 950
MATERIALS
With the aim to better define the best woven materials to be As an example, in figure 15 the felt sample of 4 mm
used in the restoration, experimental measurements have thickness (NW4) is represented, in diameters 100 mm and 30
been carried out in order to provide a better understanding of mm.
the acoustical properties of woven materials in their
application as covering for floors. In order to improve
absorption at low frequencies a multilayer system has been
experimentally investigated, coupling two different types of
materials: carpets and felts. In addition, in this particular case
study of “Sala delle Compere”, the existing marble floor had Figure 15. Felt sample NW4 (4 mm thickness)
to be protected and the felts will also fulfill this necessity.

4 ICA 2010
23-27 August 2010, Sydney, Australia Proceedings of 20th International Congress on Acoustics, ICA 2010

4. ABSORPTION COEFFICIENT 5.1.1. Thickness effects on αn


MEASUREMENT METHOD
In figure 17 a comparison between samples, made of the
The normal incidence absorption coefficient, αn, has been same material (80% wool, 20% polyamide) but characterised
experimentally determined according to EN ISO 10534-2: by different thickness (10 - 12 - 14 mm), is shown.
2001 [8], using an impedance tube with two microphones.
The spectral range of interest is achieved by two different
measurement setup of Kundt tube, summarized in table 4.
The analyzed frequencies range is from 400 Hz to 4200 Hz.
The lower working frequency is due to the accuracy of the
signal processing equipment. The upper working frequency
has been chosen to avoid the occurrence of non-plane wave
mode propagation.

Table 4. Range of reliability of impedance tube


Frequency Frequency
Diameter Spacing
ƒmin ƒmax
[mm] [mm]
[Hz] [Hz]
50 400 2000
100 Figure 17. Comparison between αn of samples with same
100 172 2000
30 20 858 4200 composition (80% Wool, 20% Polyamide), but different
thickness (10 - 12 - 14 mm).
During the experimental measurements the following Figure 18 shows the differences between two 100% wool
procedures have been followed, relating to the positioning of carpets having thickness of 8 and 12 mm.
the sample material inside the impedance tube:
• samples thickness has been measured with the aid of a
gauge and the same distance has been kept between the
back plate of the sample holder and the specimen front
surface in order to avoid excessive pressure on the
samples while mounting them,
• the homogeneity of the carpet hair on the entire sample
surface has been checked.

5. ABSORPTION COEFFICIENT RESULTS OF


THE SINGLE LAYER SYSTEM
5.1. Experimental characterization of carpet
Figure 18. Comparison of αn of samples with same
Each single sample of carpet described in table 2 has been composition (100% Wool) but different thickness (8 -
tested and the values of the normal incidence absorption 12 mm).
coefficient are shown in figure 16. All the samples provide
good performance at high frequencies, but they are not very It can be noticed that the effect of the increasing thickness is
efficient below 1000 Hz. more regular when the carpet is entirely made of the same
material (100% wool), rather than two (80% Wool, 20%
Polyamide).

5.1.2. Fibre composition effects on αn

Figure 19 shows that, for high frequencies, 100% wool fibre


(sample A) returns an αn higher than 100% polyamide fibre
(sample F), though having the same thickness.

Figure 16. Normal incidence absorption coefficient for the


various samples of carpet

The most performing type of carpet is the one represented by


sample A, which is the thickest of the 100 % wool samples. Figure 19. Comparison of αn of samples with same thickness
The influence on the variation of thickness and composition (12 mm), but different composition (100% Wool and 100%
have been investigated. Polyamide).

ICA 2010 5
23-27 August 2010, Sydney, Australia Proceedings of 20th International Congress on Acoustics, ICA 2010

The absorption behaviour is coincident up to 3000 Hz, after


this frequency the sample F, made of 100% polyamide,
decreases its acoustic performances while sample A, made of
100% wool, are augmenting.

5.1.3. Consideration on sample F

Test carpet F (Fig. 20) is a particular kind of sample, because


the carpet is coupled with a resilient felt. Comparing its
absorbing behavior with the other samples of carpet (Fig. 16),
it can be affirmed that the absorption coefficient of the
sample F is the lowest, although the felt was already applied
to the support.

Figure 22. Comparative spectra of αn for felt samples in the


frequency range 400 - 4200 Hz.

Figure 23 shows the results obtained from sound absorption


tests performed on each sample of carpet, coupled with a felt
polyester fiber thickness of 3 mm.

Figure 20. Sample F

In addition, for this sample the αn curves measured with high


frequency tube and low frequency tube do not match. The
gap between αlow and αhigh can be noticed in Figure 21.

This is probably due to the fact that this sample is made of


three different layers: the carpet, the support and the felt.
Therefore, between the rigid back of the sample holder and
the rigid carpet support there’s a resilient layer. The
measured αn is probably affected by some resonance
frequency of the free vibrating rigid carpet support.
Figure 23. Absorption coefficient spectrum for all the 6
coupled system carpet + 3 mm thick felt in the frequency
range 400 - 4200 Hz.

Figure 24 shows, in the 3rd octave frequency bands, the


results obtained by sound absorption tests performed on each
sample of carpet, coupled with a felt polyester fiber thickness
of 4 mm.

Figure 21. Sample F: gap between αlow and αhigh curves


around 1000 Hz frequency.

5.2. Experimental characterization of felt

Figure 22 shows the αn coefficient obtained from sound


absorption tests performed on the felt samples described in
Table 3. As foreseeable for porous materials [9], the
increasing thickness of the felt corresponds to an increasing
Figure 24. Absorption coefficient spectrum for all the 6
absorption coefficient effect.
coupled system carpet + 4mm thick felt, in the frequency
6. ABSORPTION COEFFICIENT OF A DOUBLE range 400 - 4200 Hz.
LAYER SYSTEM
The choice of the most appropriate coupled-system to be
Particularly interesting are the results arising by coupling the used for the acoustic correction of the specific case study
original carpet with felts made of polyester fibres. As a first depends heavily by the objective to be achieved. In order to
step, all the samples with the two types of felt are analyzed select the optimal configuration in tables 5 and 6 report the
together, then, as a second step, considerations on every percentage of increase [%] of the absorption coefficient αn,
single sample, by itself and in the double layer system, are obtained by the double layer system compared to the one
carried out. layer system.

6 ICA 2010
23-27 August 2010, Sydney, Australia Proceedings of 20th International Congress on Acoustics, ICA 2010

(α carpet + felt − α carpet )


%∆α n = (1)
α carpet
Where:

αcarpet+felt is the measured absorption coefficient of the


coupled system carpet and felt (double layer system)

αcarpet is the measured absorption coefficient of the carpet by


itself (single layer system).

Table 5. Percentage increase of sound absorption measured


on the coupling carpet / felt (3mm) Figure 26. Absorption coefficient spectrum for sample F at
different configurations in the frequency range 400 - 4200 Hz
Test
A B C D E F
Sample
% ∆αn 49 57 84 40 11 53

Table 6. Percentage increase of sound absorption measured


on the coupling carpet / felt (4mm)
Test
A B C D E F
Sample
% ∆αn 58 68 102 81 96 50

As it can be remarked in tables 5 and 6, sample C guarantees


the highest percentage increase of absorption in both
configurations, with the felt of 3 and 4 mm. Figure 25 shows
the comparison of the absorption coefficient of sample C by
itself and by adding the two different types of felt.
Figure 27. Absorption coefficient spectrum for sample A in
different configurations in the frequency range 400 - 4200 Hz

Figure 25. Absorption coefficient spectrum for sample C at


different configurations in the frequency range 400 - 4200 Hz
Figure 28. Absorption coefficient spectrum for sample B in
Therefore, if the aim is to accomplish good performance at different configurations in the frequency range400 - 4200 Hz
frequencies in the range of speech, then the optimal solution
is the one represented in figure 25, referring to the coupled-
system carpet (sample C) plus 4mm thick felt (NW4). This
configuration ensures a high value and a proper distribution
of the absorption coefficient.

Aiming to analyze the effect of the two layers system in


comparison with the one layer system, all the different
sample in the three configurations are presented from figure
26 to figure 30.

As a tendency, the double layers systems present always an


increase of absorption in lower frequency domain,
ameliorating the acoustic performance of the single layer
carpet. A singular behavior is reached by the sample F
(Figure 26), which is the only one already coupled with a Figure 29 Absorption coefficient spectrum for sample D at
resilient material: the addition of another layer does not different configurations in the frequency range 400 - 4200 Hz
improve its performances.

ICA 2010 7
23-27 August 2010, Sydney, Australia Proceedings of 20th International Congress on Acoustics, ICA 2010

absorption behaviour is coincident up to 3000 Hz, after this


frequency the sample made of 100% polyamide, decreases its
acoustic performances while sample made of 100% wool, are
augmenting. A deeper study on the fibre orientation should
be necessary. The worse behaviour is obtained by the carpet
sample F, even though it is already coupled with a support
and a resilient layer. This is probably due to the fact that the
measured absorption coefficient is probably affected by some
resonance frequency of the free vibrating rigid carpet
support.

In order to improve the acoustic performances of carpets at


the low frequency range, a two layers system, coupling the
selected carpet with two different thickness of felt, has been
Figure 30. Absorption coefficient spectrum for sample E in experimentally investigated. Except for sample F, which is a
the frequency range 400 - 4200 Hz. particular sample already a three layers system, all the
coupling system present higher values with the thicker felt (4
mm). In terms of increasing percentage of absorption
coefficient between the two layers and the one layer system,
8. CONCLUSIONS sample C, which is the thickest among all (14 mm) and made
by 80% Wool and by 20% Polyamide coupled with the 4 mm
A case study of an acoustic restoration of an historic hall by felt, reaches the highest values. Therefore this configuration
means of application of woven materials has been presented. can be defined as the optimal one, presenting a minimum of
Experimental campaign results demonstrated the inadequate 0.2 at 400 Hz and a maximum of 0.7 at 4200 Hz.
acoustic response of the hall in its existing state. A
restoration project has been outlined in accordance with the Normal incidence method for measurement of the absorption
Ministry of Italian Cultural Heritage, mainly allowing only coefficient (αn) has proved to be a powerful tool to acquire
temporary solutions with woven materials. Numerical part of the data for a preliminary qualitative and quantitative
simulation with a pyramid tracing software has been carried selection of carpet samples. The optimal configuration should
out in order to define the effectiveness of the defined be tested in a reverberant room aiming to confirm its efficacy
solutions. In the restoration configuration reverberation times also in a wide frequency range and for random incident
is considerably reduced from 6 to 1 s, as medium value in the waves.
hall. Also C50 and D50 are ameliorating. In order to better
define the acoustic characteristic of the woven materials to
put in practice, experimental measurements of normal
incident absorption coefficient have been carried out. The 9. ACKNOWLEDGMENT
scope of the research is also to provide a better understanding
of the acoustical properties of woven materials in their The authors wish to thank Gianmaria Civaschi and Rubina
application as covering for floors. Ramponi for their useful help during the experimental
campaign.
Six samples of carpet were selected, different for
composition, thickness and mass per unit area. All the
analyzed samples were characterized by the manufacturer
possibility to reproduce the existing design of the floor, as 10. REFERENCES
requested by the by the Ministry of Italian Cultural Heritage.
1. P. Ricciardi, R. Ramponi, A. Pieracci, “La sala delle
The normal incident absorption coefficient has been compere di palazzo s. Giorgio di Genova: analisi
determined by means of the impedance tube method with two sperimentali e progetto di bonifica acustica” XXXII
microphones, in the frequencies range from 400 Hz to 4200 National Congress of the Italian Acoustic Association,
Hz.. A procedure for the samples location in impedance tube Sicily 26-28 May (2010)
measurements has been outlined for this particular type of 2. ISO 3382:1997, Acoustics – measurement of the
woven materials. reverberation time of rooms with reference to other
acoustical parameters
By analysing the sample experimental results it can be 3. F. Martellotta, E. Cirillo, A. Carbonari, P. Ricciardi,
noticed that carpets, as all the porous materials, present an “Guidelines for acoustical measurements in churches”
absorption coefficient increasing with frequency, around 0.1 Applied Acoustics 70, 378-388 (2009)
at 400 Hz up to 0.5 or 0.8 at 4200 Hz. The most performing 4. Y. Shoshani, "Studies of textile assemblies used for
carpet is the sample A, made of 100% wall and having the acoustic control” Tech. Text. Int. 2 (3), 32 (1993)
highest thickness of 12 mm. The influence on the variation of 5. Y. K. Shoshani, "Noise absorption by a combination of
thickness and composition have been investigated. Normally, woven and nonwoven fabrics" J. Text. Inst. 82 (4), 500
an increase of thickness corresponds to an augmentation of (1991)
the absorption coefficient. While this tendency is evident in 6. P. Ricciardi, Elementi di acustica e di illuminotecnica,
the carpet sample entirely made of the same material (100 % (Mc Graw, 2009)
wool), this behaviour is not linear in the samples made of two 7. J.S. Bradley, “Uniform Derivation for Speech in Rooms”
material (80% wool and 20%polyammide). The thinner Report BRN 239, ISSN 0701-5232, (1985)
sample (10 mm) of 80% wool and 20% polyammide presents 8. UNI EN ISO 10534-2:2001, Acoustic - Determination of
the highest values of absorption in the medium frequency sound absorption coefficient and impedance in
range (1400 – 2000 Hz) with a decrease in the higher impedance tubes – Part 2: Transfer-function method.
frequency (2400 – 4200 Hz). The fibre composition has also 9. J.F. Allard, N. Atalla, Propagation of sound in Porous
been investigated: by analysing the same thickness sample Media, Modelling Sound Absorbing Materials (Wiley,
but with different composition it can be affirmed that the 2009)

8 ICA 2010

You might also like